Abstract
In patients with atrophic age-related macular degeneration, subretinal photovoltaic implant (PRIMA) provided visual acuity up to 20/440, matching its 100μm pixels size. Next-generation implants with smaller pixels should significantly improve the acuity. This study in rats evaluates removal of a subretinal implant, replacement with a newer device, and the resulting grating acuity in-vivo. Six weeks after the initial implantation with planar and 3-dimensional devices, the retina was re-detached, and the devices were successfully removed. Histology demonstrated a preserved inner nuclear layer. Re-implantation of new devices into the same location demonstrated retinal re-attachment to a new implant. New devices with 22μm pixels increased the grating acuity from the 100μm capability of PRIMA implants to 28μm, reaching the limit of natural resolution in rats. Reimplanted devices exhibited the same stimulation threshold as for the first implantation of the same implants in a control group. This study demonstrates the feasibility of safely upgrading the subretinal photovoltaic implants to improve prosthetic visual acuity.
Keywords: Retinal Prosthesis, vision restoration, neuronal migration, Age-related macular degeneration, photoreceptor death, explantation surgery
Introduction
Loss of photoreceptors in inherited retinal diseases or age-related macular degeneration (AMD) are among the leading causes for irreversible blindness. The inner retinal neurons remain to a large extent [1], [2], [3], albeit with some level of retinal circuits remodeling [4], [5]. We developed a wireless photovoltaic subretinal implant capable of eliciting visual percepts through electrical stimulation of the second order neurons, bipolar cells (BC), originally receiving an input from photoreceptors [6], [7]. A camera on augmented reality glasses captures the surrounding visual scene at eye level, and processed images are projected onto the implant using pulsed near infrared (NIR, 880 nm) light. This design confers several advantages compared to wired implants, including: relative ease of implantation, reduced risk of post-operative complications, the projected NIR does not interfere with the remaining peripheral natural vision, network mediated retinal stimulation (through bipolar cells) preserves many features of the natural signal encoding in the retina, and others [6], [8].
This approach has been tested in clinical trials with AMD patients, where PRIMA implants (Pixium Vision SA, Paris, France) with 100 μm pixels were implanted in the area of geographic atrophy. These patients, previously with no foveal light perception, gained monochromatic formed vision with a prosthetic visual acuity ranging from 20/438 to 20/550 [9], [10], closely matching the limit of this pixel size (20/420). The patients simultaneously perceive central prosthetic and natural peripheral vision [10]. While these landmark results are an important proof of concept for our technology, the number of AMD patients who would benefit from this technology would grow significantly if the implant resolution would be increased. For an acuity of 20/200 or better, pixels of 50 μm or smaller would be needed. However, with planar pixels having return electrode surrounding the active electrode in each pixel, such as in the PRIMA array (Figure 1A,B), the distance between the active and return electrodes determines the penetration depth of the electric field into the retina [11]. Therefore, the stimulation threshold rapidly increases with a decreasing pixel size, and the needed light intensity exceeds the optical and electrochemical safety limits with pixels smaller than 75 μm for human retina [12], [13].
We developed 3 different approaches to overcome this limitation; a planar monopolar (MP) array, and two 3-dimensional geometries: honeycomb (HC) and pillar (PIL) devices with pixels sizes ranging from 55 μm to 22 μm. MP devices contain central active electrodes in each pixel and a common return electrode along the edge of the implant (Figure 1C,D). During full-field illumination, the electric field is oriented vertically near the implant surface, resulting in a stimulation threshold 30 times lower (0.059 mW/mm2) than with bipolar pixels of 40 μm [12]. High contrast and confinement of the electric fields can still be achieved in this configuration by current steering, using active electrodes on non-illuminated pixels as transient returns [14]. This strategy allows scaling the pixel size down to 22 μm in rats [15]. in the 3-D honeycomb configuration, the return electrode is elevated on top of 30 μm high isolating wall around each pixel, with the active electrode at the bottom of the well. The vertical electric field inside such cavities matches the direction of bipolar cells that migrate into the wells [11], [16]. This effect reduces the stimulation threshold and decouples it from the pixel size. The final 3-D configuration involves elevating the active electrodes on top of pillar-like structures to achieve closer proximity to the BC in the INL after implantation. This approach might be the most beneficial for AMD patients, where INL is separated from the subretinal implant by about 40 μm of the debris layer [17].
We have characterized the retinal migration into the voids in subretinal implants in rats [18], [19] and demonstrated that this process did not cause cell loss nor negatively affect the retinal excitability throughout the lifetime of the animals [16], [20]. Our PRIMA implants are designed to last throughout the patients’ lifetime, and already have shown stability in the subretinal space for over 4 years [9].
In this study, we demonstrate a possibility of safely removing the subretinal implants in all 3 configurations, in case they are no longer needed. Furthermore, we demonstrate feasibility of implanting the next-generation device in the same location to provide higher visual acuity. This opens the door to upgrading the current patients having the first-generation PRIMA arrays with 100 μm pixels, to the next-generation implant having 22 μm pixels (Figure 1). We demonstrate that such a procedure preserves the retinal tissue and retains the stimulation thresholds, while increasing the visual acuity to the level of the new implant.
Results
Three different subretinal implants: planar, honeycomb (HC) and pillar (PIL) arrays (Fig. 2 A–C) were fabricated and implanted into the subretinal space of RCS rats (6 months old, n = 33) for 6 weeks, to ensure completion of the migration process of the inner nuclear layer (INL) [16]. Reconstructed confocal acquisitions of whole-mounted retina-implant-sclera complex were used to visualize cross-sectional views of 4',6-diamidino-2-phenylindole (DAPI) labelled nuclei and their relative position. With planar devices, the whole INL was resting on top of the implant, with the outer most layer of cells in contact with the pixels (Fig. 2D). The honeycomb walls (Fig. 2E) and the pillars (Fig. 2F) maintained their mechanical integrity, with all structures intact after the implantation surgery, 6 weeks in vivo, and the following tissue preparations including its staining and imaging procedures. Retinal cells migrated and filled the HC cavities and the spaces between pillars, indicating a strong physical interaction with the implants.
Surgical explantation
To explore the feasibility of extracting the subretinal prosthesis, we developed a new surgical procedure. After the primary implantation, conjunctiva is sutured and the transscleral, transretinal cut underneath heals over 6 weeks. To avoid creating a second 1.5 mm cut in the 6 mm diameter eye, the incision for explantation was performed at the location of the healed scar of the primary surgery (see supplemental video). The retina was detached by gentle injection of balanced salt solution (BSS), followed by injection of viscoelastic gel to keep the subretinal space open. While visualizing the implant through the cornea, a 30-gauge blunt canula was inserted into the subretinal space next to the implant. A gentle stream of BSS was maintained until the retina had been detached from the edge of the implant. The canula was progressively moved between the retina and the implant until a complete detachment was observed. A second canula was inserted to coat the implant with viscoelastic gel as a form of lubrication to allow removing the implant smoothly. A 27-gauge Grieshaber surgical grasping forceps (Alcon) were inserted into the subretinal space to grab the edge of the implant and slowly pull it out, making sure that the retina remained undisturbed (Fig. 6A–C). The subretinal space was rinsed with BSS and the conjunctiva sutured. The surgeries for the 3 different implant geometries were comparable. The planar implants required less BSS to detach from the retina and were removed smoothly. HC implants took longer to detach but eventually, they could be dragged out of the subretinal space as smoothly as planar implants. PIL devices detached smoothly from the retina but required a continuous application of viscoelastic gel to separate retina from the top of the pillars, to avoid the retina getting stuck and damaged while dragging the device out. All implants remained undamaged and with preserved 3-D structures (Fig. 3 A–F), and the explantation surgery did not leave any prosthetic pieces and materials in the subretinal space.
Retinal integrity post device explantation
Removal of prosthetic devices that have integrated with biological tissue poses the risk of damaging the tissue and potentially aggravating the disease condition. To ensure that parts of the INL were not damaged and to assess the amount of biological tissue left on the devices after explantation, the implants were fixed in 4% paraformaldehyde (PFA) overnight and processed for scanning electron microscopy (SEM) imaging. Figure 3 demonstrates the debris on surface of the implants of all 3 geometries, which is the residual biological material, compared to pristine pre-implanted devices (Fig. 2A–C). Cell-shaped structures with rounded cell bodies and extending dendrite-like structures were observed in the higher magnification SEM images (red arrows; Fig. 3D–F). Acellular debris (yellow arrows) were also observed on the planar surface, inside the HC wells and between the pillars. Presence of cells on the extracted implants was confirmed by confocal imaging with DAPI nuclei staining (Fig. 3G–I). Due to their structural resemblance to immune cells, IBA1 staining was used to identify the cell type. Most of the cell nuclei co-localized with IBA1 (green; Fig. 3J–L), indicating that they were microglial cells rather than being from neuronal origin. These cells had attached to- and stayed on the implants after retinal detachment. Moreover, there was no apparent difference in the distribution of cells remaining on the implants for all three geometries.
The retinal integrity and recovery after removal of the prosthetic devices was monitored in vivo by funduscopic examination and by optical coherence tomography (OCT) imaging. The position of retinal blood vessels in the fundus relative to the original location of the implants was used to orientate and monitor the correct area. Six weeks after implantation, the subretinal arrays were visible through a clear retina (Fig. 4A, G, M). OCT images confirmed the appearance of the inner retina above the implants. The INL was in close proximity to the surface of the planar implant (Fig. 4D), and with the 3-D implants, only part of the INL was visible above the implants, indicating complete migration of the bipolar cells into the cavities (Fig. 4J, P). After extraction of the implants, the retina had dark areas indicating the presence of blood or subretinal debris (Fig. 4 B, H, N). The cross-sectional views of OCT confirmed the presence of subretinal materials (yellow arrows) preventing full reattachment of the retina to the retinal pigment epithelium (RPE) layer (Fig. 4 E, K, Q). After six weeks of recovery, the retina appeared clear funduscopically (Fig 4. C, I, O), similar to the non-implanted RCS rat retina. The inner retinal layers were well preserved in all three geometries. However, a thick hyper-reflective layer (red arrows) was present in the subretinal space, and occasionally, a dark ‘pocket’ was observed where the implant used to be (double white asterisks).
Histological analysis of the recovered retina six weeks after the explantation showed preserved INL cells (Fig. 5B–D) when compared to an age-matched non-implanted RCS retina (Fig. 5A). However, after extraction of 3-D implants (Fig. 5C–D), the INL appears to be less organized due to the induced migration into the implants and their removal. The hyper-reflective membrane on OCT (Fig. 4; red arrows) appeared as mostly acellular material in the subretinal space (Fig. 5; red arrows). The ‘dark pockets’ observed on OCT (Fig. 4; double white asterisk), remained unstained (Fig. 5B; black asterisk), suggesting that they were fluid filled pockets. Masson’s trichrome staining (MTS), which differentially stains collagen, demonstrated that the collagen containing acellular layer (Fig. 5 E–H; blue arrows), was only present in explanted locations and not in the control retina. Müller cell activation was labelled with glial fibrillary acidic protein (GFAP) as a marker of retinal response to injury. The retina having undergone explantation surgery displayed similar GFAP expression as the RCS control (Fig. 5I–L), suggesting the baseline activation of Müller cells is related mostly to the disease pathology rather than surgery. A glial membrane was observed (Fig. 5 yellow arrows) in the subretinal space locations where collagenous subretinal fibrosis was identified. However, it was also present in the RCS control.
Implant replacement and upgrade
The ability to safely remove implants from the subretinal space opens the possibility of replacing the implants with improved new generation devices. PRIMA with 100 μm pixels has been an important clinical proof of concept for photovoltaic subretinal prosthesis. Clinical studies with AMD patients demonstrated that prosthetic visual acuity matched the 100 μm pixel size of these implants [9], [10]. Since animal studies with our next generation devices demonstrated much higher resolution, we assessed the feasibility of replacing the PRIMA implants with flat arrays of 22 μm pixels. Six weeks after the initial subretinal implantation in RCS rats, PRIMA implants were extracted following the extraction procedure of planar devices we described. After the primary implant was out, an MP device was coated with viscoelastic gel and inserted between the retina and RPE through the same 1.5 mm incision. While visualizing the implant through the cornea, it was carefully slid into the exact same location as the primary device, defining the implant placement based on location of the retinal blood vessels nearby (Fig. 6A–F). Fundus images shown in Figure 6G,H confirm the correct placement of the arrays (yellow arrows). OCT images showed well preserved retinal layers both, pre-extraction and post- reimplantation. Interestingly, immediate implant replacement did not cause subretinal fibrosis or development of a subretinal pocket, as observed when the retina was allowed to recover for the same amount of time but without an implant (Fig. 4; red arrows).
While the implant replacement surgeries were successful and retained the inner retinal anatomy, retinal function, excitability and visual acuity remain the key measures of success for retinal prostheses. We measured the visually evoked potential (VEP) in response to full field illumination (Fig. 7A) and defined the stimulation threshold - the lowest light intensity that elicits a significant cortical response (Fig. 7B). VEPs were recorded via transcranial electrodes above the visual cortices, with NIR stimuli applied at 2 Hz with 10 ms pulse duration, while varying the irradiance. The upgrades were from the PRIMA 100 μm pixel arrays of 1.5 mm in width to the new generation arrays of the same width, having 22 μm monopolar (MP) pixels. Since the PRIMA and MP arrays have very different field confinement and stimulation thresholds, we compared the retinal excitability before and after the replacement surgery by comparing the responses to the same MP devices, implanted once or re-implanted. Reimplanted MP devices showed very similar responses at all light intensities and importantly, the stimulation threshold of both groups was 0.06 mW/mm2 (Fig. 7E), suggesting that the retina retains its electrical excitability after the second surgery. Furthermore, we assessed the improvement in spatial resolution by measuring the grating acuity using alternating gratings projected on the reimplanted devices (Fig, 7B). The logarithmic fitting curve crossed the noise level at the 28 μm mark (Fig. 7F), matching the natural limit of visual acuity in rats [19], [20]. This represents a significant improvement in visual acuity compared to the 87 μm pixel pitch of PRIMA devices (d=100μm ηcos 30° for hexagonal array).
Discussion
Clinical trials with PRIMA implants having 100 μm pixels (Pixium Vision SA, Paris, France) demonstrated restoration of central vision in patients blinded by atrophic AMD, with highest resolution achieved to date by visual prostheses: up to 20/438, closely matching its pixel size limit of 20/420 [9], [10]. While very promising for the field and life changing for the patients, visual acuity should increase further to exceed the legal blindness threshold of 20/200, and ideally even beyond. Our next generation implants with pixel sizes down to 20μm, have demonstrated feasibility of achieving this goal in rats as well as in computational models of human retina [11], [15], [16], [20], [21]. Therefore, it is necessary to determine the feasibility and safety of the possible explantation of subretinal prosthesis and its exchange with a newer version. It is especially challenging with novel 3-D configurations of the devices that have shown extensive cell migration of the retina into the implants (Fig. 2) [16], [21]. Planar PRIMA implants were designed to last the lifetime of the patients and have shown good stability for over 7 years so far [9]. Similar long-term subretinal stability is expected with 3-D arrays [20]. However, in the event of an implant malfunction, elective removal, or if a new generation device with higher visual acuity limit becomes available, explantation or replacement of these subretinal implants may be necessary. Our study has shown the feasibility of safely removing devices of all 3 configurations, demonstrated re-implantation of the next-generation arrays, and validated the subsequent improvement in their performance.
To avoid unnecessary damage to retinal vasculature and peripheral cells, the retina was incised and re-detached at the primary incision scar location. Planar implants remained stable in the subretinal space (Fig. 2D &4A), held in place by the forces keeping the retina attached: RPE fluid transport to choroid and by ocular pressure. No resistance was encountered when injecting BSS between the retina and the implant and the device could be readily removed. HC implants were completely attached to the retina and required a slow, steady stream of BSS, while progressively moving the irrigation canula between the retina and top of HC walls. Once detached, its removal was comparable to planar implants, sliding smoothly out of the subretinal space. PIL implants, on the other hand, required more gentle manipulations. The PILs are fragile structures and care must be taken when inserting the cannula between the retina and the pillar tops. Furthermore, even after detachment of the retina from the implant, the PILs had a tendency to stick to the retina and required constant injection of a viscoelastic bolus. The rat lens occupies the majority of the posterior chamber, leaving limited space (approximately 1 mm; supplementary Fig. 2) for retinal detachment, insertion of tools and manipulation. In larger animals and humans, the significantly larger posterior chambers will allow bigger blebs to be made during retinal detachment to facilitate subretinal manipulations. Additionally, lens removal and vitrectomy in the larger eyes could further help the surgeries. For PIL implant removal, a specialized tool that grabs the edge of the implant and has a protective cover separating the top of the PILs from the retina could further facilitate the removal process and reduce the risk of retinal injury.
The promising results with the next generation implants in terms of safety and resolution [15], [20], [21] suggest that they could reach clinical trials soon. If successful as a commercial product indicated for restoring sight, patients with the first-generation PRIMA devices should also be able to benefit from this advancement. The ability to upgrade from PRIMA 100 μm devices to new implants with 22 μm pixels could drastically improve the prosthetic visual acuity of these patients from the current 20/438 limit [10] up to 20/80 [15], [16]. Tiling such implants to the full size the scotoma should increase the field of view, while the digital zoom on the external camera should allow these patients to get close to acuity of 20/20. Our results demonstrate that the retina retains its structure and its electrical excitability after the upgrade surgery from PRIMA to a planar MP device, and that visual acuity with 20 μm pixels reaches the 28 μm limit of visual resolution in rats. Photovoltaic HC and PIL devices are in the design and fabrication phases, and tests of the upgrade from PRIMA to HC and PIL will be conducted once these devices have been fabricated.
While the retinal morphology was preserved after explantation of the devices, the eyes developed subretinal fibrosis at the location of former implant (supplementary Fig.1). Interestingly, subretinal fibrosis was not observed when a secondary implant was immediately placed at the same location in the subretinal space (Fig. 6G, H). The known contributors to subretinal fibrosis include Müller, microglial, infiltrating immune cells, complement activation, pericytes and RPE cells [22], [23], [24], [25]. If Müller and microglial cells were significant contributors after the implant extraction, reimplantation of a new device would still result in a hyper-reflective seal forming between the implant and the retina. Since it was not observed, it seems likely that the mediators were infiltrating from the RPE/choroid side and separation by the implant prevents their migration towards the retina. Due to the implant opacity, OCT cannot visualize the space behind the device, thus further histological studies will be required to assess whether the fibrotic membrane observed is still forming behind the implant after the upgrade. Subretinal fibrosis could diminish the chances of success in later attempts to restoration of sight by other means. If the mediators of subretinal fibrosis after prosthesis explantation can be identified, subretinal drug dosing at the time of the surgery might help prevent the formation of these fibrotic membranes.
In conclusion, our study demonstrated the feasibility of safely removing subretinal implants of planar and 3-D configurations and replacing them with a new generation device. After reimplantation, retinal excitability remained at the original level, while resolution increased up to the natural limit of visual acuity in rats. The developed surgical techniques can be adapted and should be easier to perform in much larger human eyes. The capability of upgrading the retinal protheses may alleviate the potential concern that implantation of the early versions of the device may preclude patients ability to use the next-generation implants.
Materials and Methods
Planar implants
Planar implants of 2 configurations were used in this study: PRIMA implants with photovoltaic pixels of 100 μm containing local return electrodes around each pixel, and monopolar (MP) devices of 22 μm pixels containing global returns around the edge of the implant (Figure 1). The 1.5 mm PRIMA implants were fabricated by and obtained from Pixium Vision [9], [10]. The 1.5 mm monopolar devices were fabricated at Stanford as previously described [26]. Briefly, implants are optimized for 880 nm wavelength: with 30 μm thickness and metallization on the back surface of the device, approximately 90% of light is absorbed in Si. These photodiode arrays consist of 3508 hexagonal pixels of 22 μm in width. About 20% of the pixels at the implanťs periphery are coated with a common return. The fabrication process involves creating a vertical p-n junction, establishing ohmic contact to the electrodes, applying anti-reflection coatings and metal electrodes, adding a high-capacitance coating (SIROF) to the electrodes, and releasing the device [26]. Subsequently, the sidewalls and back side of the devices are coated with titanium for protection from erosion.
Honeycomb-shaped implants
The Inner Nuclear Layer (INL) thickness in RCS rats ranges from 40 to 50μm and consists of 4–5 layers of nuclei. We have previously demonstrated that 25 μm deep wells of the honeycombs allow bipolar cells to migrate into the cavities while keeping amacrine cells (AC) out [16]. For anatomical investigations, passive honeycomb implants were constructed from crystalline silicon using a two-layer mask technique for generating deep silicon etching patterns (Figure 2B), as previously outlined [11].
Briefly, a hexamethyldisilazane (HMDS) primed wafer underwent a spin-coating process with 2 μm of negative photoresist (AZ5214-IR). This layer was exposed to UV light through a patterned photomask, leading to the creation of 25 μm deep cavities using a Bosch etch process in the unprotected regions. After removing the honeycomb-defining resist, a second photoresist layer (7.5% SPR 220–7, 68% MEK, and 24.5% PGMEA) was spray-coated onto the wafer to a thickness of 30 μm. This layer was exposed to define the releasing trenches around the 1 mm wide arrays, also utilizing a Bosch process.
Subsequently, the wafer received a protective spray-coating of 60 μm thick photoresist for backside grinding from 500 to 50 μm, starting from the base of the honeycombs. The remaining excess silicon was etched in XeF2 gas to complete the release of the implants. Each implant comprised four quadrants, featuring hexagonal honeycomb patterns of 40, 30, and 22 μm in width with walls of 25 μm height, having thicknesses of 4, 3, and 2 μm, respectively. The fourth quadrant served as a flat control, as depicted in Fig. 3B. To prevent the dissolution of thermally oxidized silicon (300 nm) in-vivo, arrays were sputter-coated with 200 nm of titanium.
Pillar implants
Pillar electrodes were electroplated, as previously described [21]. Briefly, patterns for the active and return electrode structures were created using a Ti:Au layer (50 nm:200 nm) on blank 4-inch silicon wafers (p-doped). The interconnected active and return electrode structures enabled electroplating the 3D devices across the entire wafer simultaneously. Each pixel's disk electrode was electroplated into a pillar with a 16.5 μm diameter for 40 μm pixels. Applying a constant current density of 1 mA/cm² to the patterned wafers resulted in a plating rate of 3 μm per hour, achieving the desired pillar height in gold. The top surface of the electroplated structures was subsequently sputter-coated with Ti:SIROF (40 nm:436 nm) to provide a high-capacitance material for the electro-neural interface. For anatomical integration studies, the whole implant was sputtered with 200nm Ti before implantation into the subretinal space.
Animals and surgical procedures
All experimental protocols received approval from the Administrative Panel on Laboratory Animal Care (APLAC) at Stanford and were executed following institutional guidelines. The procedures adhered to the Statement for the Use of Animals in Ophthalmic and Vision Research, as outlined by the Association for Research in Vision and Ophthalmology (ARVO). Animal care and implantation procedures were conducted in accordance with previously documented methods [6], [27]. Royal College of Surgeons (RCS) rats with a genetic mutation in MERTK gene resulting in complete photoreceptor degeneration between 4–6 months of age, were housed at the Stanford animal facility.
For anatomical studies, a total of N = 9 animals were implanted subretinally with different types of arrays (3 animals per group), and the devices were explanted 6 weeks later. For physiological studies, additional N = 15 animals were implanted with a PRIMA 100 μm chip for 6 weeks and the array was replaced with a 22 μm planar implant and monitored for up to 6 months.
For the surgical procedures, animals were anesthetized with a combination of ketamine (75 mg/kg) and xylazine (5 mg/kg) administered intraperitoneally. A 1.5 mm incision was made through the sclera and choroid approximately 1 mm posterior to the limbus. Retinal detachment was induced with a saline solution injection, and the implant was placed into the subretinal space and positioned away from the incision site. The conjunctiva was sutured using nylon 10–0, and topical antibiotic (bacitracin/polymyxin B) was applied to the eye postoperatively.
Imaging
In-vivo OCT
Surgical success and retinal reattachment on top of the array was confirmed through Optical Coherence Tomography (OCT) (HRA2-Spectralis; Heidelberg Engineering, Heidelberg, Germany). For ease of imaging in rat eyes, the cornea was covered with viscoelastic gel and a coverslip was used to cancel the corneal curvature and its optical power. OCT monitoring was conducted regularly to observe the recovery of the retina after initial surgery, before extraction of the implants and weekly for 6 weeks post-surgery.
Scanning Electron Microscopy (SEM)
Implants of all 3 configurations were imaged before implantation and after extraction using a Zeiss Sigma SEM with Schottky Field Emission (FE) source and GEMINI electron optical column (Ziess, Oberkochen, Germany). To assess the material left on the devices after surgical explantation, the devices were fixed in 4% PFA overnight and then processed for SEM imaging. The samples are put on a sample holder with carbon tape so that the light samples are not blown away during the SEM chamber pump process (FEI Helios NanoLab 600i DualBeam SEM/FIB and Thermo Fisher Scientific Apreo S LoVac Scanning Electron Microscope). The sample holder with implants is mounted in the SEM chamber and pumped down below 9E-5 mbar. Once a vacuum is established, E-beam with 2kV, 43pA is applied on the sample for imaging. Usually, lower beam power is applied for better imaging on the interface level (so e-beam does not penetrate too deep and image the inner layers).
Confocal fluorescence imaging of whole-mount retina
Whole-mount preparations of the retina integrated with the implants were stained with 4',6-diamidino-2-phenylindole (DAPI). After extraction surgery, the implants were fixed in 4% PFA and immune labelled with 1:400 rabbit raised IBA1 antibody (WAKO, Japan) at room temperature for 24 hours. To label the implants, 1:400 donkey raised rabbit secondary antibody conjugated with Alexa Fluor 488 (Thermo Fisher Scientific, Rockford, IL) and DAPI were incubated for 24 hours at room temperature. Three-dimensional imaging of retinal whole mounts was conducted using a Zeiss LSM 880 Confocal Inverted Microscope with Zeiss ZEN Black software. The identification of implant surfaces was achieved by reflecting a 514 nm laser and using a neutral-density beam splitter that allowed 80% transmission and 20% reflection. Images were acquired through the entire inner nuclear layer (INL) thickness using Z-stack, 10 μm above the INL and below the base of the devices. Stacks were obtained at the center of each implant, with a 40x oil-immersion objective and an acquisition area of 225×225 μm at 500 nm z-steps. The Zeiss z-stack correction module was utilized to compensate for lower light intensity within the wells of the implants.
Confocal fluorescence datasets were processed in ImageJ. To address brightness variations at different Z positions within the wells and above the implant, contrast maximization was initially applied to individual XY planes, ensuring 0.3% channel saturation. XY planes underwent de-speckling through the median filter, and background suppression was performed using the rolling-ball algorithm [28]. Subsequent cascades of gamma adjustments and min-max corrections were applied to further mitigate the background, relative to the noise level. Gaussian blurring was specifically implemented for nucleus staining channels to smooth brightness variations within individual cells. The reconstruction of implants involved projecting the implant reflection toward the bottom of the implant.
Histological preparations
After implant extraction, the eyes were allowed to recover for 6 weeks, then enucleated and fixed in a 1.25% glutaraldehyde solution for 24 hours at room temperature. Subsequently, they underwent post-fixation in osmium tetroxide for 2 hours at room temperature and dehydration through a series of graded alcohol and propylene oxide. After overnight infiltration in epoxy (without DMP-30) at room temperature with Electron Microscopy Sciences' Araldite-EMbed (RT13940, Mollenhauer’s kit), the samples underwent a 36-hour curing process in an oven at 70 °C. Epoxy blocks were carefully trimmed until the extracted implant area became visible. The resulting 700 nm thick sections, cut using a Reichart UltracutE, were stained with 0.5% toluidine blue for light microscopy imaging.
Immunohistochemistry in frozen sections
The eyes were enucleated after euthanasia and fixed in 4% (PFA) overnight. After removal of the cornea and lens, the explanted area was selected under a stereoscope and embedded in optimal cutting temperature compound. Once frozen, the block was sectioned at a thickness of 12 μm. The sections underwent permeabilization with triton-x and were incubated overnight at 4 degrees with 1:500 of goat raised glial fibrillary acidic protein primary antibody (GFAP; SC-6170; Santa Cruz Biotechnologies, Santa Cruz, CA). Subsequently, the sections were incubated for 2 hours with 1:400 donkey raised anti goat, Alexa Fluor 594 conjugated secondary antibody (A-11058; Thermo Fisher Scientific, Rockford, IL), along with DAPI. Confocal microscopy (LSM880; Zeiss, Jena, Germany) and Airyscan (LSM880; Zeiss) were used for imaging.
Trichrome staining
A trichrome stain kit (ab150686; Abcam, Boston, MA) was optimized for use on frozen sections. Briefly, the embedding medium was washed with PBS and slides were incubated in preheated Bouin's Fluid for 60 min and cooled for 10 min, followed by 3 rinses in water. Slides were then incubated in Weigerťs Iron Hematoxylin for 5 min, rinsed in water, incubated in Biebrich Scarlet / Acid Fuchsin solution for a further 15 min, rinsed in water, differentiated in phosphomolybdic / phosphotungstic acid solution for 10–15 min, incubated in Aniline Blue solution for 5–10 min and rinsed 3 times in water. Finally, slides were incubated in acetic acid solution for 3–5 min, dehydrated, cleared and mounted for light microscopy imaging.
Electrophysiology
To assess the visually evoked potentials (VEP) elicited in animals after PRIMA 100 μm implants were upgraded to monopolar 22 μm implants, each animal was implanted with three transcranial electrodes. One electrode was placed in the skull above each hemisphere’s visual cortex (V1; 4 mm lateral from midline, 6 mm caudal to bregma), and one reference electrode above the somatosensory cortex (2 mm right of midline and 2 mm anterior to bregma).
Animals were lightly anesthetized and their pupils were dilated. An artificial tear gel and a cover slip were used to cancel the cornea’s optical power. A customized projection system was used for visualizing and projecting the stimulation patterns onto the subretinal implant. It comprised a near-infrared laser with a wavelength of 880 nm (MF_880 nm_400 um, DILAS, Tucson, AZ), collimating optics, and a digital micromirror display (DMD; DLP Light Commander; LOGIC PD, Carlsbad, CA) for generating optical patterns, coupled with a slit lamp (Zeiss SL-120; Carl Zeiss, Thornwood, NY) and a CCD camera (acA1300–60gmNIR; Basler, Ahrensburg, Germany).
Near-infrared (NIR) light with a pulse duration of 10 ms at 2 Hz, and irradiance ranging from 0.002 to 4.7 mW/mm2 was used to determine the stimulation threshold of the first time implanted eyes with monopolar 22 μm arrays and compared to that with the same implants after re-implantation. The irradiance at the retina was calculated and adjusted based on the ratio between the sizes of projected pattern on the retina and on the cornea. VEPs were recorded using the Espion E3 system (Diagnosys LLC, Lowell, MA) at a sampling rate of 2 kHz and averaged over 500 trials. The stimulation threshold was defined as the VEP amplitude exceeding the noise above the 95% confidence interval [15], [16], [26].
Visual acuity was measured using projection of alternating grating patterns with various bar widths and determining the narrowest grating bars that elicited VEP signal. NIR light at 2.4 mW/mm2, 4 ms-long pulses, at a carrier frequency of 64 Hz and grating switching cycle of 1 Hz (500 ms per image) were used as the projection parameters. Grating bar widths ranged from 13 μm to 117 μm on the retina. The noise baseline was determined by projecting static gratings with a 120 μm bar width, with the other stimulus parameters unchanged.
VEP data analysis was conducted using a custom code developed in MATLAB and Python. The high-frequency noise corresponding to the pulsed stimuli (64Hz carrier frequency) was filtered out using a spectrum reconstruction algorithm. The intersection of a logarithmic fit of the data points above the noise level with the baseline noise determined the grating visual acuity limit [15], [16].
Supplementary Material
Acknowledgement
The authors would like to thank Bing-Yi Wang for her help with VEP training and during the pilot stages of the study. MB performed surgeries; MB, NM performed in-vivo imaging, immunohistochemistry and confocal imaging; DPH, SVS and MB conducted image processing and statistical analysis; MB, NM and DPH conducted the electrophysiological measurements; AKG developed and edited the codes used for VEP signal processing; NJ performed field mapping on all implants used; AS, EB and LG fabricated subretinal implants under the guidance of KM and TK; RD sectioned the retina; DP guided the research and data analysis; all authors participated in writing and/or reviewing the manuscript. Studies were supported by the National Institutes of Health (Grants R01-EY-035227, and P30-EY-026877), the Department of Defense (Grant W81XWH-22–1-0933), AFOSR (Grant FA9550–19-1–0402), Wu Tsai Institute of Neurosciences at Stanford, and unrestricted grant from the Research to Prevent Blindness. Photovoltaic arrays were fabricated at the Stanford Nano Shared Facilities (SNSF) and Stanford Nanofabrication Facility (SNF), which are supported by the National Science Foundation award ECCS1542152. K.M. was supported by a Royal Academy of Engineering Chair in Emerging Technology, UK.
Footnotes
Competing interests:
Ted Kamins: Consultant (Pixium Vision)
Daniel Palanker: Patent, Consultant (Pixium Vision)
Other authors have no other competing interests to declare.
References
- [1].Humayun M. S. et al. , “Morphometric analysis of the extramacular retina from postmortem eyes with retinitis pigmentosa,” Invest. Ophthalmol. Vis. Sci., vol. 40, no. 1, pp. 143–148, Jan. 1999. [PubMed] [Google Scholar]
- [2].Stone J. L., Barlow W. E., Humayun M. S., de Juan E., and Milam A. H., “Morphometric analysis of macular photoreceptors and ganglion cells in retinas with retinitis pigmentosa,” Arch. Ophthalmol. Chic. Ill 1960, vol. 110, no. 11, pp. 1634–1639, Nov. 1992. [DOI] [PubMed] [Google Scholar]
- [3].Kim S. Y., Sadda S., Humayun M. S., de Juan E., Melia B. M., and Green W. R., “Morphometric analysis of the macula in eyes with geographic atrophy due to age-related macular degeneration,” Retina Phila. Pa, vol. 22, no. 4, pp. 464–470, Aug. 2002, doi: 10.1097/00006982-200208000-00011. [DOI] [PubMed] [Google Scholar]
- [4].Jones B. W., Pfeiffer R. L., Ferrell W. D., Watt C. B., Tucker J., and Marc R. E., “Retinal Remodeling and Metabolic Alterations in Human AMD,” Front. Cell. Neurosci., vol. 10, p. 103, 2016, doi: 10.3389/fncel.2016.00103. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [5].Sekirnjak C. et al. , “Changes in physiological properties of rat ganglion cells during retinal degeneration,” J. Neurophysiol., vol. 105, no. 5, pp. 2560–2571, May 2011, doi: 10.1152/jn.01061.2010. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [6].Lorach H. et al. , “Photovoltaic restoration of sight with high visual acuity,” Nat. Med., vol. 21, no. 5, pp. 476–482, May 2015, doi: 10.1038/nm.3851. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [7].Mathieson K. et al. , “Photovoltaic Retinal Prosthesis with High Pixel Density,” Nat. Photonics, vol. 6, no. 6, pp. 391–397, Jun. 2012, doi: 10.1038/nphoton.2012.104. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [8].Ho E. et al. , “Spatiotemporal characteristics of retinal response to network-mediated photovoltaic stimulation,” J. Neurophysiol., vol. 119, no. 2, pp. 389–400, Feb. 2018, doi: 10.1152/jn.00872.2016. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [9].Palanker D., Le Mer Y., Mohand-Said S., Muqit M., and Sahel J. A., “Photovoltaic Restoration of Central Vision in Atrophic Age-Related Macular Degeneration,” Ophthalmology, vol. 127, no. 8, pp. 1097–1104, Aug. 2020, doi: 10.1016/j.ophtha.2020.02.024. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [10].Palanker D., Le Mer Y., Mohand-Said S., and Sahel J. A., “Simultaneous perception of prosthetic and natural vision in AMD patients,” Nat. Commun., vol. 13, no. 1, Art. no. 1, Jan. 2022, doi: 10.1038/s41467-022-28125-x. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [11].Flores T. et al. , “Honeycomb-shaped electro-neural interface enables cellular-scale pixels in subretinal prosthesis,” Sci. Rep., vol. 9, no. 1, pp. 1–12, Jul. 2019, doi: 10.1038/s41598-019-47082-y. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [12].Wang B.-Y., Chen Z. C., Bhuckory M., Goldstein A. K., and Palanker D., “Pixel size limit of the PRIMA implants: from humans to rodents and back,” J. Neural Eng., vol. 19, no. 5, p. 055003, Sep. 2022, doi: 10.1088/1741-2552/ac8e31. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [13].Ho E. et al. , “Characteristics of prosthetic vision in rats with subretinal flat and pillar electrode arrays,” J. Neural Eng., vol. 16, no. 6, p. 066027, Oct. 2019, doi: 10.1088/1741-2552/ab34b3. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [14].Chen Z. C., Wang B.-Y., and Palanker D., “Real-Time Optimization of the Current Steering for Visual Prosthesis,” in 2021 10th International IEEE/EMBS Conference on Neural Engineering (NER), May 2021, pp. 592–596. doi: 10.1109/NER49283.2021.9441400. [DOI] [Google Scholar]
- [15].Wang B.-Y. et al. , “Electronic photoreceptors enable prosthetic visual acuity matching the natural resolution in rats,” Nat. Commun., vol. 13, no. 1, Art. no. 1, Nov. 2022, doi: 10.1038/s41467-022-34353-y. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [16].Bhuckory M. B. et al. , “Cellular migration into a subretinal honeycomb-shaped prosthesis for high-resolution prosthetic vision,” Proc. Natl. Acad. Sci., vol. 120, no. 42, p. e2307380120, Oct. 2023, doi: 10.1073/pnas.2307380120. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [17].Muqit M. M. K., Mer Y. L., Holz F. G., and Sahel J. A., “Long-term observations of macular thickness after subretinal implantation of a photovoltaic prosthesis in patients with atrophic age-related macular degeneration,” J. Neural Eng., vol. 19, no. 5, p. 055011, Oct. 2022, doi: 10.1088/1741-2552/ac9645. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [18].Palanker D. et al. , “Migration of retinal cells through a perforated membrane: implications for a high-resolution prosthesis,” Invest. Ophthalmol. Vis. Sci., vol. 45, no. 9, pp. 3266–3270, Sep. 2004, doi: 10.1167/iovs.03-1327. [DOI] [PubMed] [Google Scholar]
- [19].Butterwick A., Huie P., Jones B. W., Marc R. E., Marmor M., and Palanker D., “Effect of shape and coating of a subretinal prosthesis on its integration with the retina,” Exp. Eye Res., vol. 88, no. 1, pp. 22–29, Jan. 2009, doi: 10.1016/j.exer.2008.09.018. [DOI] [PubMed] [Google Scholar]
- [20].Bhuckory M. et al. , “3D electronic implants in subretinal space: long-term follow-up in rodents,” BioRxiv Prepr. Serv. Biol., p. 2023.07.25.550561, Jul. 2023, doi: 10.1101/2023.07.25.550561. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [21].Butt E. et al. , “Three-dimensional electro-neural interfaces electroplated on subretinal prostheses.” bioRxiv, p. 2023.11.09.566003, Nov. 13, 2023. doi: 10.1101/2023.11.09.566003. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [22].Kimura K. et al. , “Attenuation of EMT in RPE cells and subretinal fibrosis by an RAR-γ agonist,” J. Mol. Med., vol. 93, no. 7, pp. 749–758, Jul. 2015, doi: 10.1007/s00109-015-1289-8. [DOI] [PubMed] [Google Scholar]
- [23].Luo X. et al. , “Choroidal pericytes promote subretinal fibrosis after experimental photocoagulation,” Dis. Model. Mech., vol. 11, no. 4, p. dmm032060, Apr. 2018, doi: 10.1242/dmm.032060. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [24].Shen W. et al. , “A Combination Therapy Targeting Endoglin and VEGF-A Prevents Subretinal Fibro-Neovascularization Caused by Induced Müller Cell Disruption,” Invest. Ophthalmol. Vis. Sci., vol. 59, no. 15, pp. 6075–6088, Dec. 2018, doi: 10.1167/iovs.18-25628. [DOI] [PubMed] [Google Scholar]
- [25].Tenbrock L. et al. , “Subretinal fibrosis in neovascular age-related macular degeneration: current concepts, therapeutic avenues, and future perspectives,” Cell Tissue Res., vol. 387, no. 3, pp. 361–375, Mar. 2022, doi: 10.1007/s00441-021-03514-8. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [26].Huang T. W. et al. , “Vertical-junction photodiodes for smaller pixels in retinal prostheses,” J. Neural Eng., vol. 18, no. 3, p. 036015, Mar. 2021, doi: 10.1088/1741-2552/abe6b8. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [27].Lorach H. et al. , “Interactions of Prosthetic and Natural Vision in Animals With Local Retinal Degeneration,” Invest. Ophthalmol. Vis. Sci., vol. 56, no. 12, pp. 7444–7450, Nov. 2015, doi: 10.1167/iovs.15-17521. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [28].Hanson A. J., “The rolling ball,” in Graphics Gems III, USA: Academic Press Professional, Inc., 1992, pp. 51–60. [Google Scholar]