Quantitative myocardial perfusion mapping by cardiac magnetic resonance imaging (CMR) is an emerging non-invasive diagnostic tool for coronary heart disease. Advantages of quantitative analysis over traditional qualitative assessment include the objective assessment of myocardial blood flow in conditions of globally reduced flow such as coronary microvascular dysfunction [1]. Normal values for stress myocardial blood flow (MBF) and myocardial perfusion reserve (MPR) have been established [2,3] and have been used for diagnostic validation, for example, against invasive functional tests for coronary microvascular dysfunction [4]. Thresholds have also been proposed for determining if a stress test is adequate (stress MBF >1.43 ml/g/min) [5] and whether there is obstructive coronary artery disease (stress MBF <1.94 ml/g/min) [6]. It is unclear however, whether field strength (3 T vs 1.5 T) has a significant impact on flow quantitation, and whether adjustments to the thresholds are required depending on the type of scanner used.
We performed a retrospective analysis of 212 stress perfusion CMR scans comparing global stress and resting MBF, stress endocardial:epicardial ratio and MPR. Scans were performed according to standard protocol at 1.5 T (Aera) in 102 patients and 3.0 T (Prisma, Siemens, Erlangen, Germany) in 110 patients. Basal, mid-ventricular and apical short axis perfusion images were acquired during hyperaemia induced by peripheral venous infusion of adenosine (140–210 μg/kg/min), and at rest. Automated inline perfusion mapping was generated by the dual sequence technique described by Kellman et al. [1] Univariate adjustment was perfomed for likely confounders: gender, rate-pressure product and diagnosis (IBM SPSS v.27). Diagnosis was coded for each scan as either: normal, ischemia with no obstructive coronary disease (INOCA), epicardial coronary artery disease (CAD) with ischemia, CAD without ischemia, and non-ischemic cardiomyopathy.
Resting MBF was higher at 1.5 T versus 3 T after adjustment for baseline rate-pressure product, gender and cardiac diagnosis (0.70 ± 0.28 vs. 0.59 ± 0.35 ml/g/min, p = 0.01). Mean stress MBF was also higher at 1.5 T versus 3 T albeit not statistically significant (stress MBF: 1.81 ± 0.72 ml/g/min vs. 1.64 ± 0.70 ml/g/min, p = 0.09). Endocardial:epicardial ratio (0.87 ± 0.15 vs. 0.88 ± 0.12, p = 0.76) and MPR (2.73 ± 1.06 vs. 2.92 ± 0.92, p = 0.18) were not significantly different at 1.5 T versus 3 T.
Amongst the 77 scans coded as ‘normal’, there were differences in both resting MBF (1.5 T: 0.79 ± 0.35 ml/g/min vs. 3 T: 0.60 ± 0.17 ml/g/min, p < 0.01) and stress MBF (1.5 T: 2.27 ± 0.77 ml/g/min vs. 3 T: 1.90 ± 0.66 ml/g/min, p = 0.02). MPR and endocardial: epicardial ratio values were no different between field strengths (Fig. 1).
Fig. 1.
Box-plots show data obtained from ‘normal’ scans. Comparison of means (independent samples t-test) revealed significant differences in absolute MBF values at rest and stress, however ratio variables were no different between field strengths.
Our results indicate myocardial blood flow may differ when assessed at different MRI field strengths, with stress MBF 19 % higher at 1.5 T even after adjustment for gender and hemodynamic status. While this is most likely related to field strength, it may be due to the pulse sequence used, as there was a staggered transition from FLASH to SSFP sequence readouts on the 3 T scanner during the study period. Further investigation to determine scanner and/or sequence specific reference ranges for myocardial blood flow quantification may be necessary before routine clinical use.
In contrast, the use of quantitative ratios (MPR and endocardial:epicardial ratio) is transferrable across field strengths and may be more immediately useful for current clinical investigation and diagnosis.
CRediT authorship contribution statement
George R. Abraham: Writing – original draft, Visualization, Methodology, Formal analysis, Data curation, Conceptualization. Colin Berry: Writing – review & editing, Supervision, Resources. Qing Fu: Writing – review & editing, Data curation. Stephen P. Hoole: Writing – review & editing, Supervision, Resources. Jonathan R. Weir-McCall: Writing – review & editing, Supervision, Methodology, Investigation, Formal analysis, Data curation, Conceptualization.
Declaration of competing interest
CB is employed by the University of Glasgow which holds consultancy and research agreements for his work with Abbott Vascular, AstraZeneca, Boehringer Ingelheim, Coroventis, HeartFlow, Menarini, MSD, Novartis, Servier, Siemens Healthcare, TherOx, Inc. and Valo Health and receives research funding from the British Heart Foundation grant (RE/18/6134217, BHF/FS/17/26/32744, PG/19/28/34310), Chief Scientist Office, EPSRC (EP/R511705/1, EP/S030875/1), European Union (754946-2), Medical Research Council (MR/S018905/1) and UKRI (MC/PC/20014). All other authors declare no competing interests relevant to this work.
Acknowledgements
This research was funded by the Jon Moulton Charity Trust [GRA, SPH, JWM].
We thank the patients and staff who have supported these studies. Findings from an earlier version of this study were presented as a poster at the British Society of Cardiovascular MRI conference in Liverpool, United Kingdom in October 2022.
Footnotes
Supplementary data to this article can be found online at https://doi.org/10.1016/j.ahjo.2024.100388.
Appendix A. Supplementary data
Supplementary material
References
- 1.Kellman P., Hansen M.S., Nielles-Vallespin S., Nickander J., Themudo R., Ugander M., Xue H. Myocardial perfusion cardiovascular magnetic resonance: optimized dual sequence and reconstruction for quantification. J. Cardiovasc. Magn. Reson. 2017;19(1):1–14. doi: 10.1186/s12968-017-0355-5. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2.Engblom H., Xue H., Akil S., Carlsson M., Hindorf C., Oddstig J.…Arheden H. Fully quantitative cardiovascular magnetic resonance myocardial perfusion ready for clinical use: a comparison between cardiovascular magnetic resonance imaging and positron emission tomography. J. Cardiovasc. Magn. Reson. 2017;19(1):1–9. doi: 10.1186/s12968-017-0388-9. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.Brown L.A., Gulsin G.S., Onciul S.C., Broadbent D.A., Yeo J.L., Wood A.L.…Plein S. Sex-and age-specific normal values for automated quantitative pixel-wise myocardial perfusion cardiovascular magnetic resonance. Eur. Heart J.-Cardiovasc. Imaging. 2023;24(4):426–434. doi: 10.1093/ehjci/jeac231. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.Kotecha T., Martinez-Naharro A., Boldrini M., Knight D., Hawkins P., Kalra S., Patel D., Coghlan G., Moon J., Plein S., Lockie T., Rakhit R., Patel N., Xue H., Kellman P., Fontana M. Automated pixel-wise quantitative myocardial perfusion mapping by CMR to detect obstructive coronary artery disease and coronary microvascular dysfunction: validation against invasive coronary physiology. J. Am. Coll. Cardiol. Img. 2019;12(10):1958–1969. doi: 10.1016/j.jcmg.2018.12.022. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5.Kotecha T., Monteagudo J.M., Martinez-Naharro A., Chacko L., Brown J., Knight D., Knott K.D., Hawkins P., Moon J.C., Plein S., Xue H., Kellman P., Lockie T., Patel N., Rakhit R., Fontana M. Quantitative cardiovascular magnetic resonance myocardial perfusion mapping to assess hyperaemic response to adenosine stress. Eur. Heart J. Cardiovasc. Imaging. 2021;22(3):273–281. doi: 10.1093/ehjci/jeaa252. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6.Kotecha T., Chacko L., Chehab O., O’Reilly N., Martinez-Naharro A., Lazari J., Knott K.D., Brown J., Knight D., Muthurangu V., Hawkins P., Plein S., Moon J.C., Xue H., Kellman P., Rakhit R., Patel N., Fontana M. Assessment of multivessel coronary artery disease using cardiovascular magnetic resonance pixelwise quantitative perfusion mapping. J. Am. Coll. Cardiol. Img. 2020;13(12):2546–2557. doi: 10.1016/j.jcmg.2020.06.041. [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
Supplementary material