Abstract
To compare the diagnostic value of ultrahigh-resolution CT-angiography (UHR-CTA) compared with high-pitch spiral CTA (HPS-CTA) using a first-generation, dual-source photon-counting CT (PCD-CT) scanner for preprocedural planning of transcatheter aortic valve replacement (TAVR). Clinically referred patients with severe aortic valve stenosis underwent both, retrospective ECG-gated cardiac UHR-CTA (collimation: 120 × 0.2 mm) and prospective ECG-triggered aortoiliac HPS-CTA (collimation: 144 × 0.4 mm, full spectral capabilities) for TAVR planning from August 2022 to March 2023. Radiation dose was extracted from the CT reports, and the effective dose was calculated. Two radiologists analyzed UHR-CTA and HPS-CTA datasets, assessing the image quality of the aortic annulus, with regard to the lumen visibility and margin delineation using a 4-point visual-grading scale (ranges: 4 = ”excellent” to 1 = ”poor”). Aortic annulus area (AAA) measurements were taken for valve prosthesis sizing, with retrospective UHR-CTA serving as reference standard. A total of 64 patients were included (mean age, 81 years ± 7 SD; 28 women) in this retrospective study. HPS-CTA showed a lower radiation dose, 4.1 mSv vs. 12.6 mSv (p < 0.001). UHR-CTA demonstrated higher image quality to HPS-CTA (median score, 4 [IQR, 3–4] vs. 3 [IQR, 2–3]; p < 0.001). Quantitative assessments of AAA from both CTA datasets were strongly positively correlated (mean 477.4 ± 91.1 mm2 on UHR-CTA and mean 476.5 ± 90.4 mm2 on HPS-CTA, Pearson r2 = 0.857, p < 0.001) with a mean error of 22.3 ± 24.6 mm2 and resulted in identical valve prosthesis sizing in the majority of patients (91%). Patients with lower image quality on HPS-CTA (score value 1 or 2, n = 28) were more likely to receive different sizing recommendations (82%). Both UHR-CTA and HPS-CTA acquisitions using photon-counting CT technology provided reliable aortic annular assessments for TAVR planning. While UHR-CTA offers superior image quality, HPS-CTA is associated with lower radiation exposure. However, severely impaired image quality on HPS-CTA may impact on prosthesis sizing, suggesting that immediate post-scan image evaluations may require complementary UHR-CTA scanning.
Supplementary Information
The online version contains supplementary material available at 10.1007/s10554-024-03050-w.
Keywords: Computed tomography angiography, Aortic valve stenosis, Cardiac computed tomography
Introduction
End-stage Aortic valve stenosis is a severe and progressive disease resulting in an obstruction of left ventricular output, heart failure, and ultimately, death due to cardiovascular disease [1]. Without treatment, the mortality in symptomatic patients exceeds 50% within two years [2]. Transcatheter aortic valve replacement (TAVR) represents an established treatment alternative to surgical aortic valve repair and is strongly recommended for patients at high-surgical risk [3, 4]. There is also growing evidence of its beneficial use in patients at intermediate or low-surgical risk [5, 6]. The preprocedural planning for TAVR relies heavily on Electrocardiogram (ECG)-synchronized CT-angiography (CTA), which provides exact aortic annular sizing and the ideal access route evaluation [7], and can be used for post-procedural risk stratification [8]. Given the dynamic changes that the aortic annulus undergoes during the cardiac cycle, manufacturers base their heart valve sizing recommendations on systolic measurements, as this is when the annulus typically reaches its largest dimension [9]. Systolic visualization of the aortic root can be achieved using Dual-Source CT technology either via low-pitch retrospective spiral cardiac CTA or prospective ECG-triggering [10]. However, current TAVR-planning guidelines still favor the use of retrospective low-pitch spiral CTA [11].
Photon-counting detector CT (PCD-CT) is an emerging technology providing improved geometrical dose efficiency, superior image quality, and spatial resolution whilst maintaining high-temporal resolution [12]. The latter is particularly crucial in cardiac CT imaging [13]. This raises the question of whether the currently guideline-recommended retrospective spiral scan remains necessary for TAVR planning in light of novel dual-source PCD-CT technology. Therefore, our study aimed to compare quantitative measurements of aortic annular sizing prior to TAVR using low-pitch, retrospective UHR-CTA, and high-pitch spectral (HPS)-CTA.
Materials and methods
Ethics statement
This analysis is part of a broader prospective study sanctioned by the Institutional Review Board of the University Medical Center Freiburg (approval ID, 21–2469 and approval date, 09/21/2021). It focuses on exploring the functionalities and properties of novel PCD-CT across various clinical applications, adhering strictly to the principles outlined in the Declaration of Helsinki. All study subjects provided informed written consent prior to their inclusion.
Patient sample
In this single-center, retrospective analysis of a prospective cohort, consecutive patients with a referral for CT prior to TAVR between August 2022 and March 2023 were considered for inclusion. All patients had confirmed severe aortic valve stenosis. Additional inclusion criteria consisted of a complete scanning protocol consisting of both—UHR-CTA and HPS-CTA. Patients presenting contraindications to contrast-enhanced CTA or previous aortic valve replacement were excluded from the analysis.
CT acquisition protocol
All study subjects were scanned on a first-generation, dual-source PCD-CT scanner (NAEOTOM Alpha, software version syngo CT VA50, Siemens Healthcare—Forchheim, Germany). The scanning protocol involved:
A retrospective ECG-gated, low-pitch helical cardiac CTA employing the ultrahigh-resolution Quantum HD Cardiac scan mode (collimation: 120 × 0.2 mm) with ECG pulsing set at 20–80% of the R-R interval in accordance with guideline recommendations for TAVR-CT [11]. This was followed by
A prospective ECG-triggered, high-pitch aortoiliac CTA employing the spectral Quantumplus mode (collimation 144 × 0.4 mm) to evaluate the feasibility of an arterial access route—technical scan parameters were previously described in detail [14]. A systolic triggering of HPS-CTA at 25% of the R–R interval was selected to depict the aortic annulus in the systolic phase.
Using a dual-syringe power injector (Accutron CT-D, Medtron AG—Saarbruecken, Germany), a combined contrast protocol was administered. This consisted of 70 ml of Iopromid (Ultravist 370, 370 mg iodine/ml, Bayer Healthcare, Leverkusen Germany), followed by a chaser consisting of 40 ml isotonic saline and 30 ml Iopromid, each delivered at aat a flow rate of 5.0 ml/sec. The UHR-CCTA was initiated first using bolus tracking with a delay of 10 s after attenuation exceeded 130 Hounsfield Units (HU) in a region of interest (ROI) placed in the aortic root. The HPS-CTA commenced immediately afterward with a minimal delay of approximately 6 s to ensure consistent image quality without administering contrast twice.
CT reconstructions
UHR-CTA multiphase data were reconstructed in steps of 50 ms of the R–R interval facilitating the identification of the optimal systolic phase for each individual. All images were reconstructed using a vascular convolution kernel (Bv48, Quantum Iterative Reconstruction Level 3), a 5122-matrix size, and a field of view restricted to the heart of 180 × 180 mm. For HPS-CTA monoenergetic 60 keV was applied. A slice thickness of 0.6 mm with an increment of 0.6 mm was employed for UHR-CTA and HPS-CTA.
Analysis of CT data
Two radiologists with 2 years (T.K.) and 4 years (M.T.H.) of experience in cardiovascular CT imaging assessed all images independently on a dedicated workstation (syngo.via, version VB60, Siemens Healthcare—Forchheim Germany, and 3mensio, software version 10.1, Pie Medical Imaging—Maastricht, The Netherlands). They were both blinded to clinical data and each other. All patients` Volume CT Dose index (CTDIvol) and Dose Length Product (DLP) were extracted from the CT reports. To calculate the effective dose, the DLP was multiplied with a conversion factor of κ = 0.014 mSv × (mGycm)−1[15]. ECG-reports during CTA acquisition were transferred as DICXOM data, maximum and minimum heart rate as well as medium heart rate were noted. To calculate heart rate variability during CTA acquisition, the following equation was applied, as previously reported: Calculation previously reported [16]: maximum heart rate – minimum heart rate/((maximum heart Rate + minimum heart rate)/2).
Aortic annulus assessment: subjective image quality
The subjective image quality of the aortic annulus was assessed using a 4-point visual grading scale for each patient on UHR-CTA and HPS-CTA images. Five practice cases were utilized to familiarize the evaluators before the read-out, which were not included in the study population. A score of 4 denoted “excellent” image quality, representing a clear and artifact-free depiction of the aortic annulus with perfect lumen and margin visibility. Image quality was deemed “good” (score of 3) for images with minor artifacts. A score of 2 indicated “fair” image quality due to detectable artifacts and a score of 1 indicated “poor” image quality due to the high presence of artifact interference or blurred depiction of the lumen or margin. Examples are given in Figure S1.
Aortic annulus assessment: objective image quality
On both UHR-CTA and HPS-CTA, at the level of the aortic root, a ROI with 15 mm in diameter was placed, and the mean CT attenuation in Hounsfield Units (HU) was registered to account for signal, as well as the standard deviation (SD) of attenuation was noted to quantify the level of noise. An ROI of similar size was placed in the paraaortic mediastinum, and the mean attenuation was registered to account for background. Signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) were calculated using the following equations:
Quantitative measurements of the aortic annulus area and the aortic annulus perimeter
The aortic annulus was characterized using a virtual semi-oval ring intersecting the basal hinge points of the three aortic valve cusps [17]. This plane was established through multiplanar reformation, adjusted to a double-oblique transverse view to intersect the cusps' most inferior attachments of the native aortic valve. Following the accurate positioning of the plane, the aortic annulus area (AAA) and aortic annulus perimeter (AAP) were determined using planimetry, and effective area- as well as perimeter-derived diameters were calculated. The systolic phase of the maximum aortic annular dimension was used to perform quantitative AAA and AAP measurements on UHR-CTA.
Hypothetical aortic valve prosthesis selection
Based on the cross-sectional quantitative measurements of both CTA datasets, a hypothetical prosthesis valve selection was performed. Following the manufacturer’s sizing recommendations, prosthesis sizing for a balloon-expandable Edwards SAPIEN 3 heart valve (Edwards Lifesciences, Irvine, CA, USA) involved cut-off values based on the AAA. For the self-expandable Evolut R heart valve (Medtronic MN, USA), cut-off values were based on the AAP [18].
Statistics
All statistical analyses were performed using IBM SPSS Statistics for Macintosh (version 28.0, Armonk, NY, United States). Q–Q plot and one-sample Shapiro–Wilk test were used to check for the assumption of normal distribution. Depending on their normality, variables were expressed as mean ± SD, or median and interquartile range. Variables were compared with a two-tailed t-test when normally distributed and with Wilcoxon signed-rank test when non-normally distributed. To assess which parameters would have an influence on the aortic annulus image quality, we fitted linear regression models, which included the subjective image quality as the outcome of interest and body-mass-index, CTDIvol, mean heart rate, heart rate variability, average HU attenuation, and CNR as covariates. To assess which parameters have an influence on mean annular area difference measured in both scanning techniques (UHR-CTA and HPS-CTA), a multiple linear regression model involving the same covariates was performed. Pearson’s r was used for the correlation between AAA and AAP measurements of both CTA datasets. After stratification for image quality, Bland–Altman-analysis was performed to analyze the agreement of quantitative aortic annular measurements within both datasets. Agreement between both readers for quantitative AAA and AAP measurements was determined using intraclass correlation coefficient analysis (model: two-way, type: agreement, unit: of analysis: single) and interpreted according to Koo et al. [19]. A two-tailed P value of < 0.05 was considered to infer statistical significance.
Results
Patient sample
64 patients (mean age, 81.4 years ± 6.9 SD; 28 women) who underwent PCD-CT for TAVR planning were included. Patient characteristics are provided in Table 1. Four patients were excluded from the analysis. Information on the workflow and patient inclusion and exclusion is given in Fig. 1.
Table 1.
Characteristics | Values |
---|---|
Number of patients | 64 (100%) |
Gender | |
Male | 36 (56%) |
Female | 28 (44%) |
Age | 81.4 ± 6.9 |
Height (cm) | 168.4 ± 8.5 |
Weight (kg) | 76.1 ± 13.8 |
Sinus Rhythm | 47 (71%) |
Heart Rate (bpm) during CTA | 73 ± 13 (range 36–136) |
Heart Rate Variabilty during CTA | 015 [0.04–0.38] |
Body-Mass-Index (kg/m2) | 26.9 ± 4.6 |
Cardiovascular Risk Factors and Assessment | |
Coronary Artery Disease | 22 (34%) |
Arterial Hypertension | 53 (83%) |
Diabetes Mellitus | 14 (22%) |
Hyperlipidemia | 41 (64%) |
Smoking | 18 (28%) |
History of Stenting | 15 (23%) |
Chronic Kidney Disease (eGFR ≤ 45 ml/min*1.72m2) | 18 (28%) |
Data are presented as numbers and frequencies in parenthesis, mean ± standard deviation
eGFR estimated Glomerular Filtration Rate, LV-EF left ventricular ejection fraction
*Value are represented in median and interquartile range in square brackets. reported: maximum heart rate–minimum heart rate/((maximum heart Rate + minimum heart rate)/2)
Evaluation of subjective and objective image quality
The aortic annulus on UHR-CTA demonstrated a higher subjective image quality (4 [3, 4] vs. 3 [2, 3], p < 0.001). In UHR-CTA, excellent image quality (score value of 4) was achieved most frequently – 44 of 64 subjects (68.8%), while in HPS-CTA, good image quality (score value of 3) was observed in27 of 64 (42.2%) cases. Poor image quality (score value of 1) was registered in two of 64 (3.1%) subjects scanned with UHR-CTA, while for HPS-CTA, poor image quality was noted in 11 of 64 (17.2%) individuals. In multiple linear regression analysis with subjective image quality as the outcome of interest, intraluminal attenuation was the only significant predictor (p = 0.007), as shown in Supplementary Table S1. Notably, HPS-CTA resulted in a reduced image noise (22.0 ± 7.1 vs. 31.4 ± 8.5, p < 0.001), which was reflected in objective image quality parameters, CNR (16.2 ± 5.9 vs. 14.7 ± 4.4, p < 0.001). Additionally, HPS-CTA was advantageous in terms of radiation exposure, with an Effective Dose of 4.1 mSv vs. 12.6 mSv (p < 0.001 respectively). Detailed metrics of subjective and objective image quality and radiation dose are provided in Table 2. An imaging example is given in Fig. 2.
Table 2.
Retrospective UHR cardiac CTA | Prospective High-Pitch aortoiliac CTA | p–value | |
---|---|---|---|
Image Quality Score | 4 [3, 4] | 3 [2, 3] | < 0.001 |
Aortic Root attenuation | 435 ± 84 | 325 ± 85 | < 0.001 |
Image noise | 31.4 ± 8.5 | 22.0 ± 7.1 | < 0.001 |
Signal-to-noise ratio | 14.7 ± 4.4 | 16.2 ± 5.9 | < 0.001 |
Contrast-to-noise ratio | 16.7 ± 4.9 | 19.0 ± 6.5 | < 0.001 |
CTDIvol (mGy) | 65.9 [54.6–80.1] | 4.3 [3.8–5.4] | < 0.001 |
DLP (mGy*cm) | 902 [736–1071] | 292 [262–365] | < 0.001 |
Effective Dose (mSv) | 12.6 [10.3–15.0] | 4.1 [3.7–5.1] | < 0.001 |
Data are presented as mean ± standard deviation or median and interquartile range in square brackets. CTDIvol CT Dose Index volume, DLP Dose-Length-Product
Quantitative measurements of the aortic annular area and the aortic annular perimeter
Mean quantitative measurements of AAA were 477.4 ± 91.1 mm2 on UHR-CTA and 476.5 ± 90.4 mm2 on HPS-CTA, with a mean AAA-difference between both acquisition techniques of 22.3 ± 24.6 mm2. A detailed comparison of patient characteristics with AAA differences below and above the mean difference of 22 mm is provided in Supplementary Table S1. The mean AAP for UHR-CTA was 78.3 ± 7.3 mm, whereas HPS-CTA displayed a mean AAP of 78.8 ± 7.3 mm with a mean difference of 1.9 ± 2.2 mm. In multiple linear regression analysis with the magnitude of AAA difference between UHR-CTA and HPS-CTA as the outcome of interest, the aortic annular subjective image quality score was the only predictor with statistical significance (Table 3). AAA-derived diameter measurements obtained from UHR-CTA and HPS-CTA showed strong, positive correlation (Pearson’s r2 = 0.86, p < 0.001) (Fig. 3). In patients with good or excellent annular image quality (n = 36), the correlation was r2 = 0.94, and in patients with fair or poor image quality (n = 28), the correlation was r2 = 0.71 p < 0.001, respectively.
Table 3.
Linear regression | Difference in aortic annular area | |
---|---|---|
β | p-value | |
BMI (kg/m2) | −0.32 | 0.1 |
CTDIvol (mGy) | 0.25 | 0.2 |
HF mean (bpm) | −0.08 | 0.5 |
HF-Variability | −0.005 | 0.97 |
Average HU attenuation | 0.26 | 0.09 |
CNR | −0.07 | 0.6 |
Subjective Image Quality of HPS-CTA | 0.5 | < 0.001 |
The regression model was significant (r2 = 0.23, p = 0.03)
BMI body-mass-index; CTDIvol, Dose CT Volume Index; HF heart frequency, HU Hounsfield Units; CNR contrast-to-noise ratio
HPS-CTA average capture time of the aortic annulus was 209 ± 37 ms. R–R timepoints used for annular assessment on UHR-CTA were on average at 202 ± 48 ms. The mean difference in R–R interval timing between the two methods was 45 ± 39 ms. Differences in R–R interval timing between HPS-CTA and UHR-CTA did not correlate with differences in AAA- and AAP-measurements ((r = −0.09; p = 0.50, and r = −0.09; p = 0.44, respectively).
After stratification for subjective image quality of aortic annulus depiction on HPS-CTA, Bland-Altmann analysis for patients with good or excellent image quality (n = 36) revealed a mean difference magnitude of 0.40 mm and limits of agreement ranging from −1.2 to 1.1 mm, suggesting negligible bias and small variability. However, for patients with fair or impaired image quality (n = 28), the mean difference magnitude was larger (0.82 mm), with wider limits of agreement: −2.2 to 2.3 mm, indicating increased variability between UHR-CTA and HPS-CTA measurements in lower-quality images. The mean of the absolute diameter differences was close to zero, indicating no systematic smaller or larger values with HPS-CTA compared to UHR-CTA for any image quality group (Fig. 4).
Hypothetical aortic valve prosthesis selection
The selection concordance for Sapien 3 valves (AAA-based sizing) was 91% (58 out of 64 patients), while for Evolute R valves (AAP-based sizing), it was at 89% (57 out of 64 patients) (Tables 4 and 5). Patients with diverging AAA-based and AAP-based heart valve prosthesis selections showed reduced subjective image quality (median 3 [IQR 2–3] vs. 2 [IQR 1–2], p = 0.048; median 3 [IQR 2–3] vs. 2 [IQR 1–3], p = 0.049). In subjects with overall lower image quality of HPS-CTA (Score value 1 and 2 n = 28) the concordance for both Sapien 3 and Evolute R prosthesis valves were identical at 82% (23 out of 28 patients). Conversely, in patients with high image quality of HPS-CTA (Score value 3 and 4, n = 36), 97% (35/36 patients) had an identical selection for Sapien 3 valves, and 94% (34/36 patients) for Evolute R valves (Table 6). A dedicated analysis of quantitative measurements and subjective image quality of patients with diverging AAA and AAP-based heart valve prosthesis recommendations is given in Supplementary Table S3 and Supplementary Table S4.
Table 4.
Area based prosthesis sizing for balloon-expandable SAPIEN 3 | |||||
---|---|---|---|---|---|
UHR–CTA | |||||
Valve size | 20 mm | 23 mm | 26 mm | 29 mm | |
HPS–CTA | 20 mm | 1 | 0 | 0 | 0 |
23 mm | 0 | 18 | 2 | 0 | |
26 mm | 0 | 2 | 28 | 2 | |
29 mm | 0 | 0 | 0 | 11 |
Table 5.
Perimeter based prosthesis sizing for self-expandable Evolute R | |||||
---|---|---|---|---|---|
UHR–CTA | |||||
Valve size | 23 mm | 26 mm | 29 mm | 34 mm | |
HPS–CTA | 23 mm | 0 | 0 | 0 | 0 |
26 mm | 0 | 12 | 1 | 0 | |
29 mm | 0 | 3 | 30 | 1 | |
34 mm | 0 | 0 | 2 | 15 |
UHR–CTA Ultrahigh-resolution CT-Angiography; HPS–CTA High-pitch spiral CT-Angiography
Table 6.
Identical prosthesis sizing | ||
---|---|---|
Sapien 3 | Evolute R | |
Total (n = 64) | 58/64 (91%) | 57/64 (89%) |
High IQ, (n = 36) | 35/36 (97%) | 34/36 (94%) |
Low IQ (n = 28) | 23/28 (82%) | 23/28 (82%) |
Table presents a comparison of identical prosthesis valve selections for Sapien 3 and Evolute R, based on Ultra-High Resolution Computed Tomography Angiography (UHR-CTA) and High-Pitch Spectral CT-angiography (HPS-CTA). The data is categorized into two groups: patients with high image quality score on HPS-CTA (score values 3 and 4) and patients with low image quality on HPS-CTA (score values 1 and 2). The selection criteria for Sapien 3 valves are based on the aortic annular area, whereas for Evolute R, the criteria depend on the aortic valve perimeter
IQ Image Quality
Discussion
In this retrospective study of a prospective cohort, we aimed to evaluate the UHR-CTA and HPS-CTA acquisition techniques on a first-generation dual-source PCD-CT scanner in the context of preprocedural planning for TAVR. Our study hereby focused on image quality and the effect on quantitative measurements of the aortic annulus with a hypothetical prosthesis sizing: Here, we observed that both techniques provide comparable and reliable quantitative aortic annular assessments in most cases, albeit with superior image quality using the UHR-CTA scan mode with guideline-recommended retrospective ECG-gating, but lower radiation dose for prospectively ECG-triggered HPS-CTA.
Image quality and radiation dose
We observed that UHR-CTA demonstrated superior image quality scores of the aortic annulus compared to HPS-CTA, which aligns with prior studies noting superior image quality and depiction of small details using ultrahigh-resolution photon-counting CTA for coronary arteries [20, 21], for pulmonary imaging [22], and temporal bone imaging [23]. The HPS-CTA features inherent spectral data using novel Quantum Imaging technology and showed reduced image noise compared to UHR-CTA. These results are confirmatory to previous studies focusing on spectral PCD-CT showing image noise reduction [24–26], and to a study recognizing elevated noise levels in ultrahigh-resolution CTA [14]. However, the difference in image noise observed might results from the different nature of the ECG-synchronized acquisition modes: For the ECG-triggered HPS-CTA all the dose is going into a single cardiac phase, while for the retrospective ECG-gated low-pitch UHR-CTA the total dose is distributed over a wider range of cardiac phases, so the images from a certain reconstructed phase only contain a part of the total radiation dose [27]. HPS-CTA using prospective ECG-triggering had a significant advantage in terms of radiation exposure compared to retrospective ECG-gated low-pitch UHR-CTA, confirming well-established research knowledge [28]. Albeit radiation concerns are largely negligible in this elderly patient population, increasing evidence for the beneficial use of TAVR in low-risk individuals [6, 29], which is likely to expand the indication to younger patients [30], who would benefit from low-radiation dose protocols.
Quantitative measurements of the aortic annular area and the aortic annular perimeter
Regarding the AAA and AAP measurements, we noticed a mean difference of 22.3 ± 24.6 mm2 and 1.9 ± 2.2 mm, respectively, with a strong positive correlation (Pearson r2 = 0.857). However, we observed a lower correlation regarding quantitative aortic annulus assessments for patients with reduced image quality of HPS-CTA. Surprisingly, we did not find a significant association between image quality and mean heart rate or heart rate variability using the high-pitch CTA, which differs from the results reported by Capilli et al. using a second-generation dual-source CT-scanner [31]. One explanation could be improved temporal resolution and wider detector coverage that came with third generation dual-source CT scanners and that also apply to dual-source photon-counting CTA and better delineation of the annulus, enabling diagnostic image quality over a larger range of heart frequencies [16, 32]. However, intraluminal attenuation was associated with reduced image quality on HPS-CTA. It remains to be seen whether monoenergetic reconstructions with lower keV might have a beneficial influence on image quality and AAA and AAP measurements for TAVR-planning, as contrast signal and luminal attenuation can be increased by employing this technique [33].
Hypothetical aortic valve prosthesis selection
Despite the small differences in AAA and AAP measurements, we found that the vast majority of patients had identical valve prosthesis sizing with both acquisition techniques (over 89% agreement in AAP- and over 91% in AAA-based sizing), supporting in principle their robustness and suitability for TAVR-planning. However, the aortic annulus undergoes conformational and pulsatile changes between systole and diastole, which can lead to different valve prosthesis size selections when measurements are not performed in the systole [34]. This is particularly important, as over- or under-sizing is associated with complications of TAVR, such as device migration, annulus rupture, paravalvular regurgitation, and hemodynamically relevant leaflet thrombosis [35–37]. Nonetheless, it is crucial to note that patients with different valve prosthesis recommendations based on AAA and AAP measurements had significantly reduced subjective image quality. Therefore, when the image quality of HPS-CTA is suboptimal, clinical decision-making might be affected, emphasizing that the depiction of the aortic annulus in good image quality is paramount in TAVR-planning. Given the potential implications of image quality on valve prosthesis sizing, efforts should be made to improve image quality, especially for HPS-CTA.
Potential clinical implications
Our study underscores the importance of tailored imaging in TAVR planning, advocating the combined use of UHR-CTA and HPS-CTA—as each offer unique advantages. We suggest a stepwise approach, primarily employing HPS-CTA for the vast majority of cases, with UHR-CTA reserved for instances where enhanced image detail is necessary. UHR-CTA’s superior image quality is invaluable in complex cases for precise valve prosthesis sizing. On the other hand, HPS-CTA’s lower radiation dose is particularly beneficial for younger patients or those requiring multiple scans, aligning with TAVR's expanding indications to lower-risk groups.
Limitations
Several limitations within our study need acknowledgment: First, the small sample size of 64 patients, predominantly of older age (mean age 81.4 ± 6.9 years), might restrict the broader generalizability of our results. Second, being a single-center study, our findings could be influenced by specific protocols, equipment unique to our center. Third, due to the study's retrospective nature, no randomization in applying UHR-CTA and HPS-CTA was performed. Due to the sequential acquisition of both scans there is a resulting difference in bolus timing and consequently variation in lumen attenuation of the aortic annulus, which might influence the comparability. Lastly, hypothetical prosthesis sizing was based on the measurements of AAA and AAP, real-world decision-making regarding prosthesis sizing involves consideration of other patient-specific factors. In line with updated guidelines, annular measurements from retrospectively ECG-synchronized CTA were used as the reference standard in our comparative analysis, although referring to the ultimately selected prosthesis valve size would have been ideal. Furthermore aortic valve configuration and feasibility of trans-arterial access are relevant for pre-TAVR CT evaluation, but were not part of our analysis.
Conclusions
To conclude, UHR-CTA and HPS-CTA provide reliable aortic annular assessments for TAVR planning. UHR-CTA offers superior image quality, while HPS-CTA has a lower radiation dose. However, severely impaired image quality on HPS-CTA may influence prosthesis sizing decisions, suggesting that immediate post-scan evaluations could help determine the need for a complementary UHR-CTA.
Supplementary Information
Below is the link to the electronic supplementary material.
Abbreviations
- UHR-CTA
Ultrahigh-resolution CT angiography
- HPS-CTA
High-pitch spectral CT angiography
- ECG
Electrocardiogram
- PCD-CT
Photon-counting detector CT
- TAVR
Trans-catheter aortic valve replacement
- AAA
Aortic annulus area
- AAP
Aortic annulus perimeter
- ROI
Region of interest
- HU
Hounsfield Units
- CTDIvol
Volume CT Dose index
- DLP
Dose Length Product
Author contributions
MTH: data curation, draft of manuscript, conceptualization, visualization TK: data curation, conceptualization MS: data interpretation, statistical analysis All other authors reviewed the manuscript
Funding
Open Access funding enabled and organized by Projekt DEAL. Funding was provided by Baden-Württemberg Ministry of Economic Affairs, Labor and Tourism (35-4223.10/20).
Data availability
Data generated or analyzed during the study are available from the corresponding author by request.
Declarations
Competing interest
M.T.H Siemens Healthineers AG (speaker‘s bureau). F.B. Siemens Healthineers (unrestricted research grant, speaker’s bureau), Bayer Healthcare (unrestricted research grant, speaker’s bureau, consultancy), C.L.S Siemens Healthineers (speaker’s bureau), S.F. is an employee of Siemens Healthineers AG, All other authors declare that they have no conflicts of interest relating to the work of this study.
Informed consent
IRB: 09/21/2021 (No. 21-2469). Informed written consent was obtained.
Footnotes
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
References
- 1.Otto CM, Prendergast B. Aortic-valve stenosis—from patients at risk to severe valve obstruction. N Engl J Med. 2014;371:744–756. doi: 10.1056/NEJMra1313875. [DOI] [PubMed] [Google Scholar]
- 2.Kodali SK, Williams MR, Smith CR, et al. Two-year outcomes after transcatheter or surgical aortic-valve replacement. N Engl J Med. 2012;366:1686–1695. doi: 10.1056/NEJMoa1200384. [DOI] [PubMed] [Google Scholar]
- 3.Beyersdorf F, Vahanian A, Milojevic M, et al. 2021 ESC/EACTS Guidelines for the management of valvular heart disease: developed by the task force for the management of valvular heart disease of the European society of cardiology (ESC) and the European Association for cardio-thoracic surgery (EACTS) Eur J Cardiothorac Surg. 2021;60:727–800. doi: 10.1093/ejcts/ezab389. [DOI] [PubMed] [Google Scholar]
- 4.Otto CM, Nishimura RA, Bonow RO, et al. 2020 ACC/AHA guideline for the management of patients with valvular heart disease: a report of the American college of cardiology/American heart association joint committee on clinical practice guidelines. J Am Coll Cardiol. 2021;77:e25–e197. doi: 10.1016/j.jacc.2020.11.018. [DOI] [PubMed] [Google Scholar]
- 5.Makkar RR, Thourani VH, Mack MJ, et al. Five-year outcomes of transcatheter or surgical aortic-valve replacement. N Engl J Med. 2020;382:799–809. doi: 10.1056/NEJMoa1910555. [DOI] [PubMed] [Google Scholar]
- 6.Mack MJ, Leon MB, Thourani VH, et al. Transcatheter aortic-valve replacement with a balloon-expandable valve in low-risk patients. N Engl J Med. 2019;380:1695–1705. doi: 10.1056/NEJMoa1814052. [DOI] [PubMed] [Google Scholar]
- 7.Rudzinski PN, Leipsic JA, Schoepf UJ, et al. CT in transcatheter-delivered treatment of valvular heart disease. Radiology. 2022;304:4–17. doi: 10.1148/radiol.210567. [DOI] [PubMed] [Google Scholar]
- 8.Milo SM, Toia P, Midiri F, et al. Aortic valve and vascular calcium score in pre-TAVI CT: correlation with early post-procedural complications. Radiol Med (Torino) 2023;128:299–306. doi: 10.1007/s11547-023-01603-y. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.Sucha D, Tuncay V, Prakken NH, et al. Does the aortic annulus undergo conformational change throughout the cardiac cycle? A systematic review. Eur Heart J-Cardiovasc Imaging. 2015;16:1307–1317. doi: 10.1093/ehjci/jev210. [DOI] [PubMed] [Google Scholar]
- 10.Dankerl P, Hammon M, Seuss H, et al. Computer-aided evaluation of low-dose and low-contrast agent third-generation dual-source CT angiography prior to transcatheter aortic valve implantation (TAVI) Int J Comput Assist Radiol Surg. 2017;12:795–802. doi: 10.1007/s11548-016-1470-8. [DOI] [PubMed] [Google Scholar]
- 11.Blanke P, Weir-McCall JR, Achenbach S, et al. Computed tomography imaging in the context of transcatheter aortic valve implantation (TAVI)/transcatheter aortic valve replacement (TAVR): an expert consensus document of the society of cardiovascular computed tomography. JACC Cardiovasc Imaging. 2019;12:1–24. doi: 10.1016/j.jcmg.2018.12.003. [DOI] [PubMed] [Google Scholar]
- 12.Flohr T, Schmidt B. Technical basics and clinical benefits of photon-counting CT. Invest Radiol. 2023;58:441–450. doi: 10.1097/RLI.0000000000000980. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13.Mergen V, Sartoretti T, Cundari G, et al. The Importance of temporal resolution for ultra-high-resolution coronary angiography: evidence from photon-counting detector CT. Invest Radiol. 2023 doi: 10.1097/RLI.0000000000000987. [DOI] [PubMed] [Google Scholar]
- 14.Hagar MT, Soschynski M, Saffar R, et al. Accuracy of ultrahigh-resolution photon-counting CT for detecting coronary artery disease in a high-risk population. Radiology. 2023;307:e223305. doi: 10.1148/radiol.223305. [DOI] [PubMed] [Google Scholar]
- 15.Hausleiter J, Meyer T, Hermann F, et al. Estimated radiation dose associated with cardiac CT angiography. JAMA. 2009;301:500–507. doi: 10.1001/jama.2009.54. [DOI] [PubMed] [Google Scholar]
- 16.Soschynski M, Hagen F, Baumann S, et al. High temporal resolution dual-source photon-counting CT for coronary artery disease: initial multicenter clinical experience. J Clin Med. 2022;11:6003. doi: 10.3390/jcm11206003. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17.Blanke P, Siepe M, Reinöhl J, et al. Assessment of aortic annulus dimensions for Edwards SAPIEN transapical heart valve implantation by computed tomography: calculating average diameter using a virtual ring method. Eur J Cardio-Thorac Surg Off J Eur Assoc Cardio-Thorac Surg. 2010;38:750–758. doi: 10.1016/j.ejcts.2010.03.039. [DOI] [PubMed] [Google Scholar]
- 18.Todaro D, Picci A, Barbanti M. Technical characteristics and evidence to date for FDA-and CE Mark-approved valves. Interv Today. 2017;11:53–58. [Google Scholar]
- 19.Koo TK, Li MY. A guideline of selecting and reporting intraclass correlation coefficients for reliability research. J Chiropr Med. 2016;15:155–163. doi: 10.1016/j.jcm.2016.02.012. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20.Mergen V, Eberhard M, Manka R, et al. First in-human quantitative plaque characterization with ultra-high resolution coronary photon-counting CT angiography. Front Cardiovasc Med. 2022;9:981012. doi: 10.3389/fcvm.2022.981012. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21.Mergen V, Sartoretti T, Baer-Beck M, et al. Ultra-high-resolution coronary CT angiography with photon-counting detector CT: feasibility and image characterization. Invest Radiol. 2022 doi: 10.1097/RLI.0000000000000897. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 22.Gaillandre Y, Duhamel A, Flohr T, et al. Ultra-high resolution CT imaging of interstitial lung disease: impact of photon-counting CT in 112 patients. Eur Radiol. 2023;33:5528–5539. doi: 10.1007/s00330-023-09616-x. [DOI] [PubMed] [Google Scholar]
- 23.Grunz J-P, Heidenreich JF, Lennartz S, et al. Spectral shaping via tin prefiltration in ultra-high-resolution photon-counting and energy-integrating detector CT of the temporal bone. Invest Radiol. 2022;57:819–825. doi: 10.1097/RLI.0000000000000901. [DOI] [PubMed] [Google Scholar]
- 24.Sartoretti T, Landsmann A, Nakhostin D, et al. Quantum iterative reconstruction for abdominal photoncounting detector CT improves image quality. Radiology. 2022;304:E55. doi: 10.1148/radiol.229013. [DOI] [PubMed] [Google Scholar]
- 25.Si-Mohamed SA, Greffier J, Miailhes J, et al. Comparison of image quality between spectral photon-counting CT and dual-layer CT for the evaluation of lung nodules: a phantom study. Eur Radiol. 2022;32:524–532. doi: 10.1007/s00330-021-08103-5. [DOI] [PubMed] [Google Scholar]
- 26.Euler A, Higashigaito K, Mergen V, et al. High-pitch photon-counting detector computed tomography angiography of the aorta: intraindividual comparison to energy-integrating detector computed tomography at equal radiation dose. Invest Radiol. 2022;57:115–121. doi: 10.1097/RLI.0000000000000816. [DOI] [PubMed] [Google Scholar]
- 27.Jakobs TF, Becker CR, Ohnesorge B, et al (2002) Multislice helical CT of the heart with retrospective ECG gating: reduction of radiation exposure by ECG-controlled tube current modulation. Eur Radiol 12:1081–1086. 10.1007/s00330-001-1278-x [DOI] [PubMed]
- 28.Sun Z, Ng K-H. Prospective versus retrospective ECG-gated multislice CT coronary angiography: a systematic review of radiation dose and diagnostic accuracy. Eur J Radiol. 2012;81:e94–100. doi: 10.1016/j.ejrad.2011.01.070. [DOI] [PubMed] [Google Scholar]
- 29.Popma JJ, Deeb GM, Yakubov SJ, et al. Transcatheter aortic-valve replacement with a self-expanding valve in low-risk patients. N Engl J Med. 2019;380:1706–1715. doi: 10.1056/NEJMoa1816885. [DOI] [PubMed] [Google Scholar]
- 30.Tabata N, Sinning J-M, Kaikita K, et al. Current status and future perspective of structural heart disease intervention. J Cardiol. 2019;74:1–12. doi: 10.1016/j.jjcc.2019.02.022. [DOI] [PubMed] [Google Scholar]
- 31.Capilli F, Benndorf M, Soschynski M, et al. Assessment of aortic annulus dimensions for transcatheter aortic valve replacement (TAVR) with high-pitch dual-source CT: comparison of systolic high-pitch vs. multiphasic data acquisition. Eur J Radiol. 2020;133:109366. doi: 10.1016/j.ejrad.2020.109366. [DOI] [PubMed] [Google Scholar]
- 32.Rotkopf LT, Froelich MF, Riffel P, et al. Influence of heart rate and heart rate variability on the feasibility of ultra-fast, high-pitch coronary photon-counting computed tomography angiography. Int J Cardiovasc Imaging. 2023;39:1065–1073. doi: 10.1007/s10554-023-02808-y. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 33.Higashigaito K, Mergen V, Eberhard M, et al. CT angiography of the aorta using photon-counting detector CT with reduced contrast media volume. Radiol Cardiothorac Imaging. 2023;5:e220140. doi: 10.1148/ryct.220140. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 34.Blanke P, Russe M, Leipsic J, et al. Conformational pulsatile changes of the aortic annulus. JACC Cardiovasc Interv. 2012;5:984–994. doi: 10.1016/j.jcin.2012.05.014. [DOI] [PubMed] [Google Scholar]
- 35.Soschynski M, Hein M, Capilli F, et al. Investigation of factors determining hemodynamic relevance of leaflet thrombosis after TAVI. Eur Heart J Cardiovasc Imaging. 2023 doi: 10.1093/ehjci/jead156. [DOI] [PubMed] [Google Scholar]
- 36.Binder RK, Webb JG, Willson AB, et al. The impact of integration of a multidetector computed tomography annulus area sizing algorithm on outcomes of transcatheter aortic valve replacement: a prospective, multicenter, controlled trial. J Am Coll Cardiol. 2013;62:431–438. doi: 10.1016/j.jacc.2013.04.036. [DOI] [PubMed] [Google Scholar]
- 37.Blanke P, Willson AB, Webb JG, et al. Oversizing in transcatheter aortic valve replacement, a commonly used term but a poorly understood one: dependency on definition and geometrical measurements. J Cardiovasc Comput Tomogr. 2014;8:67–76. doi: 10.1016/j.jcct.2013.12.020. [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
Data Availability Statement
Data generated or analyzed during the study are available from the corresponding author by request.