Evaluation of 3D Stack-of-Spiral Turbo FLASH Acquisitions for PCASL- and VSASL-Derived Brain Perfusion Mapping

Dan Zhu 1,2, Feng Xu 1,2, Dapeng Liu 1,2, Argye Elizabeth Hillis 3, Doris Lin 2, Peter CM van Zijl 1,2, Qin Qin 1,2
PMCID: PMC11054979  NIHMSID: NIHMS1987148  PMID: 37125611

Abstract

Purpose:

The most-used 3D acquisitions for ASL are GRASE and stack-of-spiral (SOS) based FSE, which require multiple shots. Alternatively, turbo FLASH (TFL) allows longer echo trains, and SOS-TFL has the potential to reduce the number of shots to even single-shot, thus improving the temporal resolution. Here we compare the performance of 3D SOS-TFL and 3D GRASE for ASL at 3 T.

Methods:

The 3D SOS-TFL readout was optimized with respect to fat suppression and excitation flip angles for PCASL- and VSASL-derived CBF mapping as well as for VSASL-derived CBV mapping. Results were compared with 3D GRASE readout on healthy volunteers in terms of perfusion quantification and temporal SNR (tSNR) efficiency. CBF and CBV mapping derived from 3D SOS-TFL-based ASL was demonstrated on one stroke patient and their potential for single-shot acquisitions was exemplified.

Results:

SOS-TFL with a 15° flip angle resulted in adequate tSNR efficiency with negligible image blurring. Selective water excitation was necessary to eliminate fat-induced artifacts. For PCASL- and VSASL-based CBF and CBV mapping, compared to the employed 4-shot 3D GRASE with an acceleration factor of 2, the fully sampled 3D SOS-TFL delivered comparable performance (with a similar scan time) using 3 shots, which could be further undersampled to achieve single-shot acquisition with higher tSNR efficiency. SOS-TFL had reduced CSF contamination for VSASL-CBF.

Conclusion:

3D SOS-TFL acquisition was found to be a viable substitute for 3D GRASE for ASL with sufficient tSNR efficiency, minimal relaxation-induced blurring, reduced CSF contamination, and the potential of single-shot, especially for VSASL.

Keywords: cerebral blood flow, cerebral blood volume, velocity-selective arterial spin labeling, pseudo-continuous ASL, stack-of-spiral turbo FLASH

Introduction:

In the first consensus paper on pseudo-continuous arterial spin labeling (PCASL) (1), the recommended acquisition schemes for cerebral blood flow (CBF) mapping are the segmented 3D readout with EPI-based gradient and spin echo (GRASE) or the stack-of-spiral (SOS) based fast spin echo (FSE). 3D readout with whole-brain coverage is also recommended for velocity-selective arterial spin labeling (VSASL) based CBF mapping in its first guideline paper (2), and has been demonstrated using 3D GRASE (35). The 3D GRASE readout has also been utilized for VSASL-based measurement of cerebral blood volume (CBV) (6).

Compared to EPI-based Cartesian trajectories, spiral trajectories offer several advantages: higher acquisition efficiency (7), possibility of more incoherent undersampling for accelerated compressed-sensing (CS) reconstruction (8), and robustness to motion artifacts (9). Spin-echo-based acquisitions have restricted echo train duration due to T2-decay-induced through-slice blurring and SNR efficiency loss (10), thus usually requiring multiple segments (or shots) to cover the full k-space. 3D arterial spin labeling (ASL) scans acquired with a reduced number of segments or even a single shot are desired for higher temporal resolution for functional ASL or mitigating inter-shot motion (11). Implementation of single-shot 3D PCASL was reported using a pseudo-golden-angle rotated SOS-FSE acquisition with a high undersampling factor of 13 with CS reconstruction for brain connectivity analysis (12). The low constant refocusing flip angles of FSE employed to reduce the effect of T2 decay during a relatively long echo train would yield a suboptimal pseudo-steady-state signal (13). The spiral-out readouts typically employed for FSE have used less than half of the echo spacing, limiting its efficiency. Instead, a cylindrical spiral-in/out trajectory was designed to make full use of the echo spacing (14), which may face complications from 3D gridding. Alternatively, image acquisition with turbo FLASH (TFL) allows a longer echo train without severe blurring effects because T1 is typically larger than T2 and T1-induced signal decay is often slower (15). TFL acquisition can also incorporate a spiral-out trajectory starting immediately after the excitation with high acquisition efficiency, which results in a short TE with minimal T2* effect. Segmented 3D SOS-TFL combined with PCASL (16) and VSASL (17) have been demonstrated for fMRI, however, without direct comparison with 3D GRASE acquisition.

In the current study, for both PCASL and VSASL-derived CBF mapping as well as for VSASL-derived CBV mapping at 3 T, we first adjusted the 3D SOS-TFL readout with respect to fat suppression and excitation flip angles. Subsequently, we compared the performance of the optimized protocols with 3D GRASE readout on healthy volunteers, then demonstrated CBF and CBV mapping derived from 3D SOS-TFL based ASL on a stroke patient to exemplify the potential of the single-shot acquisitions.

Methods:

Experiments

All experiments were performed on a 3T scanner (Achieva; Philips Healthcare, Best, The Netherlands) using a dual-transmit body coil (maximum amplitude 13.5 μT) for RF transmission and a 32-channel head coil for signal reception. When using the maximum slew rate of 200 mT/m/ms, the maximum strength of the gradients is 40 mT/m. Brain images were acquired from 12 healthy subjects (42 ± 14 years old, 6 females), 1 stroke patient three months post-onset. The study was approved by the Johns Hopkins School of Medicine Institutional Review Board and written consent was obtained from all subjects.

We evaluated the feasibility of the SOS-TFL ASL imaging for three applications: PCASL-derived CBF mapping (PCASL-CBF) (18), VSASL-derived CBF mapping (VSASL-CBF) (3,19) and VSASL-derived CBV mapping (VSASL-CBV) (6,20,21).

For PCASL-CBF scans, a labeling duration (τ) of 1800 ms and a post-labeling delay (PLD) of 2000 ms were used. The labeling plane was placed 90 mm inferior to the anterior commissure-posterior commissure (AC-PC) line. Background suppression (BS) consisted of a slab-selective pre-saturation module right before labeling and 4 inversion pulses: a slab-selective inversion during labeling (777 ms after the beginning of labeling) and three non-selective hyperbolic-secant inversion pulses 277/1244/1780 ms post labeling. A vascular crushing module (20 ms) comprised of ±90° hard pulses and double refocused composite pulses (DRCP) (90°x180°y90°x) was applied before image acquisition, with gradient lobes in foot-head (FH) direction to dephase signal in large vessels (strength = 9.3 mT/m, duration = 3.6 ms, ramp time = 0.5 ms, VENC = 2.0 cm/s, cutoff velocity Vcut = 1.0 cm/s).

The VSASL-CBF scans started with a global WET pre-saturation (22) (four sequential RF pulses: 72°, 92°, 126°, 193°, each followed by a spoiling gradient) with a 2000 ms post-saturation delay (Tsat). Then a 64-ms Fourier-transform-based (FT-) VSI label/control module was applied with a bolus duration (τ) of 1400 ms (as recommended in the VSASL guideline paper (2) and previously described as PLD). Detailed parameters of the FT-VSI modules were: 9 hard excitation pulses (20° each); 8 pairs of 180°composite refocusing pulses in between with MLEV-16 phase cycling, surrounded by bipolar/unipolar (for label/control, respectively) gradient lobes in FH direction (strength = 21.8 mT/m, duration = 0.6 ms, ramp time = 0.3 ms, Vcut = 2.0 cm/s as defined in the guideline paper (2)). Three adiabatic BS pulses were applied 460/732/1115 ms after the VSI labeling. The same vascular crushing module as in the PCASL-CBF scan was applied before acquisition.

The VSASL-CBV scans started with the same global WET pre-saturation and 2000 ms Tsat as the VSASL-CBF scan. Then a 96-ms Fourier-transform-based (FT-) VSS label/control module was applied immediately followed by image acquisition. Detailed parameters of the FT-VSS modules were: 9 hard excitation pulses (10° each); 8 pairs of 180°composite refocusing pulses with MLEV-16 phase cycling, surrounded by bipolar/unipolar (for label/control, respectively) gradient lobes in FH direction (strength = 29 mT/m, duration = 1.2 ms, ramp time = 0.6 ms, Vcut = 0.5 cm/s).

For all 12 healthy subjects and the patient, the 3D SOS-TFL perfusion images for PCASL-CBF, VSASL-CBF and VSASL-CBV were acquired using the default parameters in axial view, with FOV = 220 × 220 × 120 mm3 and resolution = 3.4 × 3.4 × 5.0 mm3 in anterior-posterior (AP) × right-left (RL) × FH direction. The flip angles (FA) of the TFL excitation pulses were 15°, constructed as 1-3-3-1 binomial WE pulses. The turbo factor of the TFL sequence is 24, applied to slice (FH) direction with centric ordering, identical to the number of slices. A spiral-out trajectory with a readout duration of 11 ms was used, resulting in TR/TE = 18/2.6 ms and an echo train duration of 424 ms. With one TFL shot covering all through-plane encodings, the k-space for each label or control source image was fully sampled in-plane with 3 spiral interleaves (3-shot), with each shot taking 4.3/3.9/2.6 s for PCASL-CBF /VSASL-CBF/VSASL-CBV scans, respectively. 8 dynamics (label/control pairs) were acquired for averaging and the total scan times were 3.4/3.1/2.2 min for PCASL-CBF/VSASL-CBF /VSASL-CBV scans, respectively.

For comparison, the TFL acquisition was also performed using Cartesian sampling on 3 out of the 12 healthy subjects (Subjects 1–3) with the same FOV and resolution. The Cartesian sampling used a radial ky-kz centric ordering with a turbo factor of 64 and a smaller FA of 10° considering the large turbo factor, with TR/TE = 7/3.5 ms and the same echo train duration of 426 ms. The Cartesian images were 4-fold undersampled with the vendor-provided CS-SENSE reconstruction. One undersampled 3D control or label source image required 4 TFL shots (4-shot), each taking the same 4.3/3.9/2.6 s for PCASL-CBF/VSASL-CBF/VSASL-CBV scans, respectively. The total scan times of the 4-shot Cartesian sampling with 6 dynamics were identical to the 3-shot fully sampled SOS-TFL scans with 8 dynamics.

On Subjects 1–3, use of WE-pulses for fat suppression was compared with conventional excitation pulses with spectral pre-saturation with inversion recovery (SPIR), which was applied only before the first excitation pulse.

To evaluate the performance of SOS-TFL acquisition with different FAs for binomial WE pulses, TFL excitation FAs of 5°, 15° and 25° were compared on subject 4 and the excitation FAs of 9°, 12°, 15° and 18° were compared on 3 healthy subjects (subjects 5–7).

A comparison of the SOS-TFL acquisition with the GRASE acquisition was conducted on all healthy subjects. The GRASE acquisition had the same FOV as the SOS-TFL scan and a similar resolution of 3.4 × 3.7 × 5.0 mm3 (AP × RL × FH). The parameter setting for GRASE followed the recommendation in the 2015 ASL consensus paper (1). Including all 180° refocusing pulses, the echo train length was 12 with an echo spacing of 13 ms in the slice (FH) direction with centric ordering, resulting in an echo train duration of 162 ms. EPI factor was 15 in the phase-encoding direction (RL) and the readout duration was 9 ms. With a SENSE acceleration factor of 2, one 3D source image required 4 shots and TR = 4.1/3.6/2.3 s for PCASL-CBF/VSASL-CBF /VSASL-CBV, respectively. 6 dynamics were acquired, thus the total scan times of 3.7/3.3/2.1 min were close to the total scan times of SOS-TFL acquisitions.

To evaluate the robustness of the GRASE and SOS-TFL acquisition against high background signal. We compared the PCASL-CBF and VSASL-CBF scans with and without BS pulses for both GRASE and SOS-TFL on Subject 10.

On Subjects 1–3, the global pre-saturation pulse was compared with a slab-selective pre-saturation pulse on the GRASE-based VSASL-CBF scan. The pre-saturation slab was placed parallel to the imaging slab with the same FH coverage.

For each subject, the perfusion scans with the same labeling technique were scanned together in randomized order. The order of the three labeling techniques was also random. For each different acquisition, additional scans: a proton-density weighted image (PD, TR = 10 s, for normalization), a double inversion recovery (DIR, TR/TI1/TI2 = 10/3.58/0.48 s, to visualize gray matter), and an image acquired immediately after global saturation (Sat, to record the T1-induced bias of TFL acquisitions for correction of PD and DIR (15,23) with more explanation in the Discussion section).

Image reconstruction

The images using the fully-sampled 3D SOS-TFL (including deblurring), 3D Cartesian-TFL with a CS-SENSE factor of 4, and the 3D GRASE with a SENSE factor of 2 were reconstructed automatically by the vendor, interpolated to a resolution of 1.7 × 1.7 × 5.0 mm3.

For Subject 12, the raw k-space data of the SOS-TFL-based scans (PCASL-CBF, VSASL-CBF, VSASL-CBV, PD, DIR, and Sat) were saved for off-line reconstruction on MATLAB (MathWorks, Natick, MA). One 3-shot image had 3 spiral arms rotated in-plane by 120° and all 3 arms together were reconstructed with inverse non-uniform fast Fourier transform (NUFFT) as a reference. Single-shot PD, DIR and Sat images were reconstructed with only 1 out of the 3 arms with SENSE and CS-SENSE reconstruction. The 3-shot ASL source images with 8 dynamics were regarded as 24 spiral arms rotated by 120° and 8 out of the 24 arms were reconstructed to a set of single-shot ASL source images with 8 dynamics. The total scan times of a single-shot 8-dynamic scan reported by the scanner were 1.2/1.0/0.8 min for PCASL-CBF /VSASL-CBF/VSASL-CBV scans, respectively. Both SENSE and CS-SENSE used the conjugated gradient algorithm and CS-SENSE used a total variation regularization with a weighing factor of 0.1. Spiral deblurring was implemented using the deconvolution process with kernels calculated from the spiral trajectory and a B0 map (24). B0 maps and sensitivity profiles were obtained directly from default preparation scans automatically conducted by the scanner.

Data Analysis

Perfusion data were analyzed in MATLAB. Perfusion-weighted signal (PWS) was calculated with the difference image (SIdiff) normalized with the signal of the PD image (SIPD) and temporal SNR (tSNR) was calculated as the ratio of the mean value over the SD of the difference image along temporal dynamics. Here we adopted the tSNR efficiency for a fairer comparison between images with different acquisition times, i.e., tSNR normalized with the square root of the acquisition time per label/control pair) (2527).

As for perfusion quantification, the PCASL-CBF maps were calculated with (1):

CBF=6000λSIdiffePLDT1,blood2αT1,bloodSIPD1-e-τT1,blood

Here λ = 0.9 mL/g is the brain/blood partition coefficient, T1,blood = 1650 ms is the effective blood T1, and α=0.85×0.95NBS is the labeling efficiency of PCASL, where NBS is the number of BS inversion pulses.

The VSASL-CBF maps were calculated with (2,19):

CBF=6000λSIdiffeTIT1,bloodSIB2ατSIPD

Here α=0.56×0.95NBS is the labeling efficiency and arterial blood magnetization SIB = 1. TI is the time between labeling pulse and readout (2), as the bolus duration (τ = 1400 ms) plus the duration of the vascular crushing module (20 ms).

VSASL-CBV maps were calculated with (6,20,21):

CBV=100λSIdiffSIPDia,v,cxiαiMT1,iMT2,i,label-MT2,i,control

Compartmental fraction/labeling efficiency/T1 for arterial blood and venous blood were xa/αa/T1,a = 0.21/0.55/1888 ms and xv/αv/T1,v = 0.46/0.31/1707 ms, with negligible contribution from capillary blood (xc/αc = 0.33/0). The function M(T1,i)=1-exp-Tsat/T1,i.MT2,i,label-MT2,i,control=k1+k2k3k4TVS/T2,i, where [k1,k2,k3,k4] = [0.12, 0.86, 0.34, 0.38], TVS = 96 ms is the duration of the VSS pulse train and the arterial and venous blood T2 values were set as T2,a = 163 ms and T2,v = 71 ms. All definition and values of the parameter used above for CBV mapping here were found in these references (6,20,21).

For each SOS-TFL based perfusion maps with different FAs, the total variation (TV) along the three orthogonal dimensions were calculated as a quantitative measurement of sharpness (28).

For each subject and acquisition scheme, a region of interest (ROI) of the whole-brain gray matter (GM) was automatically segmented on the DIR images for all different acquisition schemes with a full width at half maximum (FWHM) thresholding, which is a coarse but fast globle thresholding and avoids the bias of manual segmentation. Regional GM ROIs within frontal, temporal, parietal, and occipital lobes were also selected. Global and regional quantitative perfusion metrics and their corresponding tSNR efficiency were averaged within the GM ROIs.

Results:

Supporting Information Figure S1 shows the DIR with segmented GM ROIs, PWS, and tSNR efficiency images of the Cartesian TFL (FA = 10°) sampling and the SOS-TFL (FA = 15°) on one axial slice of Subject 1. Compared to Cartesian TFL, SOS-TFL resulted in similar PWS values respectively for each labling technique but much less noisy PWS maps for all labeling techniques. Supporting Information Table S1 summarizes the average and SD of perfusion and tSNR efficiency from whole-brain GM ROIs of the three subjects, showing about 2.6/2.9/2.2 times higher tSNR efficiency of SOS-TFL than Cartesian TFL for PCASL-CBF, VSASL-CBF, and VSASL-CBV methods, respectively.

The benefit of WE over SPIR is demonstrated in Supporting Information Figure S2. SPIR scans show severe fat-related artifacts in all control images and PWS of PCASL-CBF of all three subjects, as well as DIR and PWS of VSASL-CBF images of Subjects 2 and 3. Artifacts were not visible on the PWS of VSASL-CBV images. This could be explained by the fast T1 recovery of the fat signal after SPIR during image acquisition. Artifacts were not visible on PWS images of any subject for VSASL-CBV, possibly due to the higher CBV-weighted PWS than CBF-weighted PWS (Control images of CBV scans were shown with a higher colorbar range) and additional suppression of fat by FT-VSS. On WE scans, no obvious artifacts were observed.

Figure 1 and Supporting Information Figure S3 array the DIR, PWS and tSNR efficiency of SOS-TFL with FAs of 9°, 12°, 15° and 18° (Subject 5, Figure 1) and with FAs of 5°, 15° and 25° (Subject 4, Supporting Information Figure S3). PWS maps with larger FA were less noisy but more blurring compared to those with smaller FA, which was most visible comparing FAs of 5° and 25° of Subject 4 in Supporting Information Figure S3. tSNR efficiency for PCASL-CBF and VSASL-CBV increased with FA but that for VSASL-CBF approached maximum at FA = 15° for the two cases demonstrated. For Subjects 4–7, Figure 2 plots the quantitative PWS, tSNR efficiency, and TV of the 3D perfusion map from whole-brain GM ROIs as a function of FA: the PWSvalues were consistent between different FAs; the tSNR efficiency increased with FA; TV of the perfusion maps was negatively correlated to FA indicating decreased sharpness (more blurring) with increased FA.

Figure 1:

Figure 1:

A sagittal slice of the double inversion recovery (DIR), perfusion-weighted signal (PWS) and temporal SNR (tSNR) efficiency images of 3D stack-of-spiral turbo FLASH (SOS-TFL) of one representative subject with FAs of 9°, 12°, 15° and 18° (subject 5). Gray matter (GM) ROI border is drawn on a DIR image (red line). Different FAs generated similar PWS for PCASL-CBF and VSASL-CBF, -CBV methods, with larger FAs resulting in higher tSNR efficiency but more image blurring.

Figure 2:

Figure 2:

Scatter plot of PWS (top row), tSNR efficiency (center row), and the corresponding total variation (TV) representing image sharpness of the perfusion maps (bottom row) derived from SOS-TFL based PCASL-CBF (left), VSASL-CBF (middle), and VSASL-CBV (right) as a function of flip angle (FA) of Subjects 4 (FAs=5°/15°/25°) and 5 to 7 (FAs=9°/12°/15°/18°), with each subject as individual solid lines, showing constant PWS between FAs while increased tSNR efficiency and decreased TV (more blurring) for larger FA.

Figure 3 demonstrates that SOS-TFL (FA = 15°, WE) based CBF and CBV maps are comparable to GRASE on a representative subject (subject 8). Note that SOS-TFL-based VSASL-CBF maps reduced CSF artifacts compared to GRASE (red circles).

Figure 3:

Figure 3:

Double inversion recovery (DIR), perfusion and temporal SNR (tSNR) efficiency maps of Subject 8, showing close associations between 3D GRASE and stack-of-spiral turbo FLASH (SOS-TFL) (water excitation, flip angle = 15°) for PCASL-CBF, VSASL-CBF, and VSASL-CBV methods, with gray matter (GM) ROI borders drawn on DIR images (red lines). Numbers in each subplot indicate perfusion/tSNR efficiency values averaged within the GM ROI. Small dark holes were observed on superior slices of GRASE-based PCASL- and VSASL-CBF (white arrows) but not visible on SOS-TFL-based images. SOS-TFL-based VSASL-CBF maps have reduced CSF artifacts compared to GRASE (red circle).

The 3D SOS-TFL-based perfusion maps (FA = 15°, WE) were successfully obtained from all 12 subjects. The cross-section images in three orthogonal orientations of the remaining subjects were shown in Supporting Information Figure S4, all showing high image quality and good gray and white matter contrast, except low PCASL-CBF for Subject 11, most likely due to slow blood flow, espetially for areas perfused by posterial cerebral arteries.

The correlations of the whole-brain GM perfusion between SOS-TFL and GRASE for all 12 subjects are plotted in Figures 4ac, showing high correlation coefficients (r ≥ 0.95) for all labeling techniques. In Figures 4df, the box plot of whole-brain GM perfusion for all 12 subjects comparing SOS-TFL to GRASE had no significant difference for PCASL-derived CBF (−1.1%, 36.8 ± 9.8 vs. 37.3±10.7 ml/100g/min, p > 0.05), but was significantly lower for VSASL derived CBF (−7.1%, 37.9 ± 8.3 vs. 40.8±9.3 ml/100g/min, p <= 0.01) and significantly higher for VSASL derived CBV (+15.8%, 5.3 ± 1.2 vs. 4.6 ± 1.1 ml/100g, p <= 0.001). In Figures 4gi, the box plot of whole-brain GM tSNR efficiency for all 12 subjects comparing SOS-TFL to GRASE was significantly lower for PCASL-CBF (−29.4%, 0.63 ± 0.18 vs. 0.89 ± 0.23, p <= 0.001), but had no significant difference for VSASL derived CBF (+1.6%, 0.75 ± 0.19 vs. 0.74 ± 0.26, p > 0.05) and was significantly higher for VSASL derived CBV (+14.1%, 1.29 ± 0.22 vs. 1.13 ± 0.24, p <= 0.05). The correlations and comparison of the regional GM perfusion metrics between the two readouts are plotted in Supporting Information Figure S5, revealing their significant differences mainly in parietal and occipital lobes.

Figure 4:

Figure 4:

Correlation of whole-brain gray matter (GM) perfusion (a–c), box plot of whole-brain GM perfusion (d–f), and temporal SNR (tSNR) efficiency (g–i) of 12 subjects using 3D GRASE and stack-of-spiral turbo FLASH (SOS-TFL) (water excitation, flip angle = 15°) readouts, derived from PCASL-CBF, VSASL-CBF, and VSASL-CBV methods. Their correlation coefficients (r) are listed for each plot (a–c). */**/**** indicates significant difference in t-test with p <= 0.05/0.01/0.0001, respectively.

3D SOS-TFL-based perfusion images of a patient three months following a hemorrhagic stroke involving the right dorsal insula and deep GM are shown in Figure 5. Decreased CBF was observed in the right parietal lobe region distal to the site of hemorrhage (purple circles).

Figure 5:

Figure 5:

Double inversion recovery (DIR) and perfusion images of a patient three months following a hemorrhagic stroke involving the right dorsal insula and deep gray matter. Decreased CBF was observed in the right parietal lobe region distal to the site of hemorrhage (purple circles).

Figure 6 and Supporting Information Figure S6 present the off-line images reconstructed from the representative undersampled raw data of one of the three interleaved spiral arms with acceleration factor R = 3 for single-shot, as well as the fully sampled 3-shot images as reference. For PD, DIR and ASL source images, the Pearson correlation coefficient (PCC) between undersampled single-shot images and fully sampled 3-shot images were very high for both SENSE (≥ 97%) and CS-SENSE (≥ 98%). The PCCs were slightly lower for perfusion maps but were still sufficiently high for both SENSE (90% to 94%) and CS-SENSE (91% to 95%). Perfusion values averaged within the GM ROIs were quite close between fully sampled and undersampled images for both SENSE (undersampled/fully sampled = 1.05, 1.05, 0.99) and CS-SENSE (undersampled/fully sampled = 1.04, 1.04, 0.99) and the tSNR efficiency of single-shot images were higher than the 3-shot reference for SENSE (undersampled/fully sampled = 1.00, 1.05, 1.23) and CS-SENSE (undersampled/fully sampled = 1.05, 1.09, 1.28), for PCASL-derived CBF, VSASL derived CBF, VSASL derived CBV, respectively. CS-SENSE reconstruction yielded closer perfusion values and higher tSNR efficiency than SENSE.

Figure 6:

Figure 6:

Off-line image reconstruction of the representative raw datasets with 3 shots (all 3 spiral arms, fully sampled) and with a single shot (one spiral arm, acceleration factor R = 3) using SENSE and CS-SENSE image reconstruction for double inversion recovery (DIR) image, as well as perfusion and tSNR efficiency maps. The gray matter (GM) ROI border is drawn on the 3-shot DIR image (red lines) and averaged perfusion or tSNR efficiency values within the GM ROIs are labeled on the top left corner of each quantitative map. Pearson correlation coefficient (PCC) of the single-shot images with the 3-shot as the reference are labeled at the bottom right corner of each image.

Discussion:

We demonstrated the feasibility of 3D SOS-TFL acquisitions for PCASL- and VSASL-based perfusion quantification. Results showed the tSNR benefit of spiral acquisition over Cartesian and the much more complete removal of fat artifacts when using WE instead of SPIR. Compared to the employed 4-shot 3D GRASE with an acceleration factor of 2, the fully sampled 3D SOS-TFL delivered comparable performance (with a similar scan time) using 3 shots, which can be undersampled with SENSE or compressed sensing to achieve single-shot acquisition with higher tSNR efficiency.

SOS-TFL acquisition offers the flexible choice of excitation FA for desired balance between tSNR efficiency and image blurring. For example, for functional ASL, a larger FA might be preferred to improve tSNR. A potential way to reduce image blurring for both GRASE and SOS-TFL is to incorporate carefully designed variable flip angles (VFA) (16,29,30). Using VFA instead of a constant FA has little effect on the tSNR of SOS-TFL theoretically, as it is primarily affected by the first FA (16). In this work, we chose a constant FA of 15° that is easily achievable on the scanner for a feasibility study. The benefit of VFA for VSASL remains to be explored in future studies.

One advantage of SOS-TFL over GRASE for VSASL-derived CBF mapping is the reduction of CSF contamination (Figure 3), as CSF signal labeled due to the pulsatile motion is still a main challenge of the VSASL technique (2). The most likely reason for reduced CSF contamination in the TFL acquisition compared to the GRASE acquisition is the increased T1-weighting. The whole-brain GM CBF averaged from the 12 subjects was 37.9 ± 8.3 ml/100g/min by SOS-TFL-based VSASL, 7.1% lower than the 40.8±9.3 ml/100g/min by GRASE-based VSASL, but not significantly different from the 36.8 ± 9.8 and 37.3±10.7 ml/100g/min by both SOS-TFL and GRASE based PCASL (Figures 4d,e). GRASE-based VSASL might slightly overestimate CBF in brain parenchyma adjacent to subarachnoid space (2). However, the lower CBF measured by SOS-TFL over GRASE only in parietal and occipital lobes but not in frontal and temporal lobes (Supporting Information Figure S5) is intriguing, which could be related to the longer arterial transit time incurred in parietal and occipital lobes than in frontal and temporal lobes. The exact mechanism accounting this effect warrants further investigation.

The potential of single-shot 3D ASL using SOS-TFL with accelerated reconstruction is demonstrated for one representative subject in this work (Figure 6, Supporting Information Figure S6). Further technical development is underway to improve its performance by employing variable density spiral trajectories (31), gold angle rotation (12), and better image reconstruction. As recently being done for VSASL with multi-shot GRASE readout (32), the reproducibility of VSASL with single-shot SOS-TFL acquisition will also need to be evaluated. The higher temporal resolution could allow applications such as functional and resting state ASL (12,16,17), along with more effective motion correction.

Based on our previous point spread function analysis (15), the image signal and the prepared longitudinal magnetization before the TFL acquisition obey a linear (y = kx + b and b ≠ 0) relationship with the intercept of the linear function coming from the readout itself as a T1-induced bias. For magnetization-prepared TFL with low-high profile ordering, the k-space filtering effect induced by the T1 relaxation leads to a reduced contrast-to-noise ratio of the signal prepared before the readout and an additive readout-related bias. This additive bias is identical for control, label, PD, and DIR images when using the identical TFL acquisition module. As a result, the subtraction between control and label images fully eliminates the bias. A quick Sat image captured this bias to be subtracted from PD and DIR images, as utilized in an ultrashort B1 mapping protocol (23).

Conclusions:

Different SOS-TFL acquisitions were evaluated for PCASL and VSASL sequences for brain perfusion mapping. Spiral acquisition obtained much higher tSNR compared to Cartesian-based readout. Water excitation was necessary to eliminate fat-induced artifacts for SOS-TFL. The choice of flip angles needs to balance image blurring and tSNR efficiency. SOS-TFL reduced CSF artifacts for VSASL-derived CBF mapping and increased tSNR efficiency for VSASL-derived CBV mapping. For PCASL- and VSASL-based CBF and CBV mapping, especially for VSASL, SOS-TFL acquisition was found to be a viable substitute for GRASE, with sufficient tSNR efficiency and minimal relaxation induced blurring, reduced contamination from CSF, and the potential of further acceleration to even single-shot.

Supplementary Material

supporting document

Supporting Information Table S1: Averaged results of whole-brain gray matter (GM) perfusion and temporal SNR (tSNR) efficiency of 3 healthy subjects using Cartesian (water excitation, flip angle = 10°) and stack-of-spiral turbo FLASH (water excitation, flip angle = 15°) readouts, derived from PCASL-CBF, VSASL-CBF, and VSASL-CBV methods.

Supporting Information Table S2: PWS (%) of the VSASL-CBF Scan with GRASE acquisition for 3 healthy subjects, comparing slab-selective and global pre-saturation.

Supporting Information Figure S1: Double inversion recovery (DIR), perfusion-weighted signal (PWS) and temporal SNR (tSNR) efficiency images of Cartesian turbo FLASH (flip angle=10°) and stack-of-spiral turbo FLASH (SOS-TFL) (flip angle=15°), both applying water excitation, on Subject 1, with borders of gray matter (GM) ROIs drawn on DIR images (red lines). Numbers in each subplot indicate PWS/tSNR efficiency values averaged within the GM ROIs. SOS-TFL resulted in 2.1/2.5/2.1 times higher tSNR efficiency for PCASL-CBF, VSASL-CBF, and VSASL-CBV methods, respectively.

Supporting Information Figure S2: Comparison of fat suppression effects with SPIR and water-excitation (WE) on double inversion recovery (DIR), proton density(PD) weighted, control and perfusion-weighted signal (PWS) images of stack-of-spiral turbo FLASH (SOS-TFL) scans on all three subjects. SPIR scans show severe fat-related artifacts in all control images and PWS of PCASL-CBF of all subjects, as well as DIR and PWS of VSASL-CBF images of Subjects 2 and 3. Artifacts were not visible on the PWS of VSASL-CBV images. No artifacts were observed on any WE scan.

Supporting Information Figure S3: A sagittal slice of the double inversion recovery (DIR), perfusion-weighted signal (PWS) and temporal SNR (tSNR) efficiency images of 3D stack-of-spiral turbo FLASH (SOS-TFL) of the subject with FAs of 5°, 15° and 25° (subject 4). Gray matter (GM) ROI border is drawn on DIR image (red line). Different FAs generated similar PWS for PCASL-CBF and VSASL-CBF, -CBV methods, with larger FAs resulting in higher tSNR efficiency but more image blurring (most visible with 25°).

Supporting Information Figure S4: Quantitative PCASL-CBF, VSASL-CBF, and VSASL-CBV perfusion map of remaining 6 subjects in three orthogonal orientations using stack-of-spiral turbo FLASH (SOS-TFL) (water excitation, flip angle = 15°).

Supporting Information Figure S5: Correlation of regional gray matter (GM) perfusion (a–c) and box plot of regional GM perfusion (d–f) and temporal SNR (tSNR) efficiency (g–i) of 12 subjects using 3D GRASE and stack-of-spiral turbo FLASH (SOS-TFL) (water excitation, flip angle = 15°) readouts, derived from PCASL-CBF, VSASL-CBF, and VSASL-CBV methods. Four regions of GM were compared: frontal (F), parietal (P), temporal (T) and occipital (O) lobes. The correlation coefficients (r) for each of the four GM regions are listed for each plot (a–c). */**/***/**** indicates significant difference in t-test with p <= 0.05/0.01/0.001/0.0001 after Bonferroni correction, respectively.

Supporting Information Figure S6: Off-line image reconstruction of the representative raw data with 3 shots (all 3 spiral arms, fully sampled) and a single shot (one spiral arm, acceleration factor R = 3) using SENSE and CS-SENSE image reconstruction for proton-density (PD) images and ASL source images of the first control/label dynamic. Pearson correlation coefficients (PCC) of the single-shot images with the 3-shot reference are labeled at the bottom right corner of each image.

Grant support:

NIH: R01 HL138182 (QQ), NIH: R01 HL144751 (QQ); P41 EB031771 (PvZ), P50 HD103538 (PvZ), S10 OD021648 (PvZ);

Reference:

  • 1.Alsop DC, Detre JA, Golay X, et al. Recommended implementation of arterial spin-labeled perfusion MRI for clinical applications: A consensus of the ISMRM perfusion study group and the European consortium for ASL in dementia. Magn. Reson. Med 2015;73:102–116. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2.Qin Q, Alsop DC, Bolar DS, et al. Velocity-selective arterial spin labeling perfusion MRI: A review of the state of the art and recommendations for clinical implementation. Magnetic Resonance in Med 2022;88:1528–1547. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Liu D, Xu F, Li W, Zijl PC, Lin DD, Qin Q. Improved velocity-selective-inversion arterial spin labeling for cerebral blood flow mapping with 3D acquisition. Magn Reson Med 2020;84:2512–2522. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.Liu D, Li W, Xu F, Zhu D, Shin T, Qin Q. Ensuring both velocity and spatial responses robust to field inhomogeneities for velocity-selective arterial spin labeling through dynamic phase-cycling. Magn Reson Med 2021;85:2723–2734. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Liu D, Jiang D, Tekes A, et al. Multi-Parametric Evaluation of Cerebral Hemodynamics in Neonatal Piglets Using Non-Contrast-Enhanced Magnetic Resonance Imaging Methods. J Magn Reson Imaging 2021;54:1053–1065. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Li W, Liu D, Zijl PCM, Qin Q. Three-dimensional whole-brain mapping of cerebral blood volume and venous cerebral blood volume using Fourier transform–based velocity-selective pulse trains. Magnetic Resonance in Med 2021;86:1420–1433. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.Ahn CB, Kim JH, Cho ZH. High-Speed Spiral-Scan Echo Planar NMR Imaging-I. IEEE Trans. Med. Imaging 1986;5:2–7. [DOI] [PubMed] [Google Scholar]
  • 8.Lustig M, Donoho D, Pauly JM. Sparse MRI: The application of compressed sensing for rapid MR imaging. Magn. Reson. Med 2007;58:1182–1195. [DOI] [PubMed] [Google Scholar]
  • 9.Liao J-R, Pauly JM, Brosnan TJ, Pelc NJ. Reduction of motion artifacts in cine MRI using variable-density spiral trajectories. Magn. Reson. Med 1997;37:569–575. [DOI] [PubMed] [Google Scholar]
  • 10.Qin Q Point spread functions of the T2 decay in k-space trajectories with long echo train. Magnetic Resonance Imaging 2012;30:1134–1142. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Hernandez-Garcia L, Aramendía-Vidaurreta V, Bolar DS, et al. Recent Technical Developments in ASL: A Review of the State of the Art. Magnetic Resonance in Med 2022;88:2021–2042. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12.Munsch F, Taso M, Zhao L, et al. Rotated spiral RARE for high spatial and temporal resolution volumetric arterial spin labeling acquisition. NeuroImage 2020;223:117371. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13.Alsop DC. The sensitivity of low flip angle RARE imaging. Magn. Reson. Med 1997;37:176–184. [DOI] [PubMed] [Google Scholar]
  • 14.Li Z, Schär M, Wang D, et al. Arterial spin labeled perfusion imaging using three-dimensional turbo spin echo with a distributed spiral-in/out trajectory. Magn. Reson. Med 2016;75:266–273. [DOI] [PubMed] [Google Scholar]
  • 15.Zhu D, Qin Q. A revisit of the k-space filtering effects of magnetization-prepared 3D FLASH and balanced SSFP acquisitions: Analytical characterization of the point spread functions. Magnetic Resonance Imaging 2022;88:76–88. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16.Nielsen J-F, Hernandez-Garcia L Functional perfusion imaging using pseudocontinuous arterial spin labeling with low-flip-angle segmented 3D spiral readouts. Magn Reson Med 2013;69:382–390. [DOI] [PubMed] [Google Scholar]
  • 17.Hernandez-Garcia L, Nielsen J, Noll DC. Improved sensitivity and temporal resolution in perfusion FMRI using velocity selective inversion ASL. Magn Reson Med 2019;81:1004–1015. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Dai W, Garcia D, de Bazelaire C, Alsop DC. Continuous flow-driven inversion for arterial spin labeling using pulsed radio frequency and gradient fields. Magn. Reson. Med 2008;60:1488–1497. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19.Qin Q, van Zijl PCM. Velocity-selective-inversion prepared arterial spin labeling: Velocity-Selective-Inversion Prepared ASL. Magn. Reson. Med 2016;76:1136–1148. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20.Liu D, Xu F, Lin DD, van Zijl PCM, Qin Q. Quantitative measurement of cerebral blood volume using velocity-selective pulse trains. Magn. Reson. Med 2017;77:92–101. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21.Qin Q, Qu Y, Li W, et al. Cerebral blood volume mapping using Fourier-transform-based velocity-selective saturation pulse trains. Magn Reson Med 2019;81:3544–3554. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22.Ogg RJ, Kingsley RB, Taylor JS. WET, a T1- and B1-Insensitive Water-Suppression Method for in Vivo Localized 1H NMR Spectroscopy. Journal of Magnetic Resonance, Series B 1994;104:1–10. [DOI] [PubMed] [Google Scholar]
  • 23.Zhu D, Schär M, Qin Q. Ultrafast B1 mapping with RF-prepared 3D FLASH acquisition: Correcting the bias due to T1-induced k-space filtering effect. Magnetic Resonance in Med 2022:mrm.29247. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 24.Ahunbay E, Pipe JG. Rapid method for deblurring spiral MR images. Magn. Reson. Med 2000;44:491–494. [DOI] [PubMed] [Google Scholar]
  • 25.Boland M, Stirnberg R, Pracht ED, et al. Accelerated 3D-GRASE imaging improves quantitative multiple post labeling delay arterial spin labeling. Magn. Reson. Med 2018;80:2475–2484. [DOI] [PubMed] [Google Scholar]
  • 26.Guo J, Wong EC. Increased SNR efficiency in velocity selective arterial spin labeling using multiple velocity selective saturation modules (mm-VSASL). Magn. Reson. Med 2015;74:694–705. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 27.Meixner CR, Eisen CK, Schmitter S, et al. Hybrid -shimming and gradient adaptions for improved pseudo-continuous arterial spin labeling at 7 Tesla. Magnetic Resonance in Med 2022;87:207–219. [DOI] [PubMed] [Google Scholar]
  • 28.Vu CT, Phan TD, Chandler DM. S3: A Spectral and Spatial Measure of Local Perceived Sharpness in Natural Images. IEEE Trans. on Image Process 2012;21:934–945. [DOI] [PubMed] [Google Scholar]
  • 29.Liang X, Connelly A, Tournier J-D, Calamante F. A variable flip angle-based method for reducing blurring in 3D GRASE ASL. Phys. Med. Biol 2014;59:5559–5573. [DOI] [PubMed] [Google Scholar]
  • 30.Zhao L, Chang C-D, Alsop DC. Controlling T2 blurring in 3D RARE arterial spin labeling acquisition through optimal combination of variable flip angles and k-space filtering. Magn. Reson. Med 2018;80:1391–1401. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 31.Zi R, Zhu D, Qin Q. Quantitative T2 mapping using accelerated 3D stack-of-spiral gradient echo readout. Magnetic Resonance Imaging 2020;73:138–147. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 32.Xu F, Liu D, Zhu D, et al. Test-retest reliability of 3D velocity-selective arterial spin labeling for detecting normal variations of cerebral blood flow. NeuroImage 2023;271:120039. [DOI] [PMC free article] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

supporting document

Supporting Information Table S1: Averaged results of whole-brain gray matter (GM) perfusion and temporal SNR (tSNR) efficiency of 3 healthy subjects using Cartesian (water excitation, flip angle = 10°) and stack-of-spiral turbo FLASH (water excitation, flip angle = 15°) readouts, derived from PCASL-CBF, VSASL-CBF, and VSASL-CBV methods.

Supporting Information Table S2: PWS (%) of the VSASL-CBF Scan with GRASE acquisition for 3 healthy subjects, comparing slab-selective and global pre-saturation.

Supporting Information Figure S1: Double inversion recovery (DIR), perfusion-weighted signal (PWS) and temporal SNR (tSNR) efficiency images of Cartesian turbo FLASH (flip angle=10°) and stack-of-spiral turbo FLASH (SOS-TFL) (flip angle=15°), both applying water excitation, on Subject 1, with borders of gray matter (GM) ROIs drawn on DIR images (red lines). Numbers in each subplot indicate PWS/tSNR efficiency values averaged within the GM ROIs. SOS-TFL resulted in 2.1/2.5/2.1 times higher tSNR efficiency for PCASL-CBF, VSASL-CBF, and VSASL-CBV methods, respectively.

Supporting Information Figure S2: Comparison of fat suppression effects with SPIR and water-excitation (WE) on double inversion recovery (DIR), proton density(PD) weighted, control and perfusion-weighted signal (PWS) images of stack-of-spiral turbo FLASH (SOS-TFL) scans on all three subjects. SPIR scans show severe fat-related artifacts in all control images and PWS of PCASL-CBF of all subjects, as well as DIR and PWS of VSASL-CBF images of Subjects 2 and 3. Artifacts were not visible on the PWS of VSASL-CBV images. No artifacts were observed on any WE scan.

Supporting Information Figure S3: A sagittal slice of the double inversion recovery (DIR), perfusion-weighted signal (PWS) and temporal SNR (tSNR) efficiency images of 3D stack-of-spiral turbo FLASH (SOS-TFL) of the subject with FAs of 5°, 15° and 25° (subject 4). Gray matter (GM) ROI border is drawn on DIR image (red line). Different FAs generated similar PWS for PCASL-CBF and VSASL-CBF, -CBV methods, with larger FAs resulting in higher tSNR efficiency but more image blurring (most visible with 25°).

Supporting Information Figure S4: Quantitative PCASL-CBF, VSASL-CBF, and VSASL-CBV perfusion map of remaining 6 subjects in three orthogonal orientations using stack-of-spiral turbo FLASH (SOS-TFL) (water excitation, flip angle = 15°).

Supporting Information Figure S5: Correlation of regional gray matter (GM) perfusion (a–c) and box plot of regional GM perfusion (d–f) and temporal SNR (tSNR) efficiency (g–i) of 12 subjects using 3D GRASE and stack-of-spiral turbo FLASH (SOS-TFL) (water excitation, flip angle = 15°) readouts, derived from PCASL-CBF, VSASL-CBF, and VSASL-CBV methods. Four regions of GM were compared: frontal (F), parietal (P), temporal (T) and occipital (O) lobes. The correlation coefficients (r) for each of the four GM regions are listed for each plot (a–c). */**/***/**** indicates significant difference in t-test with p <= 0.05/0.01/0.001/0.0001 after Bonferroni correction, respectively.

Supporting Information Figure S6: Off-line image reconstruction of the representative raw data with 3 shots (all 3 spiral arms, fully sampled) and a single shot (one spiral arm, acceleration factor R = 3) using SENSE and CS-SENSE image reconstruction for proton-density (PD) images and ASL source images of the first control/label dynamic. Pearson correlation coefficients (PCC) of the single-shot images with the 3-shot reference are labeled at the bottom right corner of each image.