Abstract
Cardiac pacing to treat bradyarrhythmias has evolved in recent decades. Recognition that a substantial proportion of pacemaker-dependent patients can develop heart failure due to electrical and mechanical dyssynchrony from traditional right ventricular apical pacing has led to development of more physiologic pacing methods that better mimic normal cardiac conduction and provide synchronized ventricular contraction. Conventional biventricular pacing has been shown to benefit patients with heart failure and conduction system disease but can be limited by scarring and fibrosis. His bundle pacing and left bundle branch area pacing are novel techniques that can provide more physiologic ventricular activation as an alternative to conventional or biventricular pacing. Leadless pacing has emerged as another alternative pacing technique to overcome limitations in conventional transvenous pacemaker systems. Our objective is to review the evolution of cardiac pacing and explore these new advances in pacing strategies.
Keywords: pacemaker, conduction system pacing, leadless pacing, His bundle pacing, left bundle pacing, cardiac resynchronization therapy
INTRODUCTION
Cardiac pacing has been the main treatment for patients with bradycardia due to various pathologies, such as sinus node dysfunction or atrioventricular (AV) block. Technological advances have been geared toward determining the optimal ventricular pacing site to best mimic normal human ventricular physiology and conduction. Long-term right ventricular (RV) pacing (RVP) can lead to pacing-induced cardiomyopathy due to left ventricular (LV) mechanical dyssynchrony in 5.9–39% of patients (1). This risk has motivated the emergence of new pacing strategies for more physiologic pacing (Figure 1a-c). These in turn have also been applied to patients with heart failure (HF) with dyssynchronous LV contraction related to conduction system disease. Conventional pacing systems are also limited by the use of transvenous leads, as well as subcutaneously placed pulse generators that can be prone to infection, vascular access challenges, and mechanical stressors that can cause lead malfunction. Leadless pacing systems have been developed (Figure 1d), which are progressing from single to dual chamber pacing capability and can now be integrated with defibrillation and transvenous components.
Figure 1.
Types of cardiac pacing: (a) CRT-P device, (b) His bundle lead, (c) left bundle branch area pacing, (d) leadless pacing. Abbreviations: AV, atrioventricular; CRT-P, cardiac resynchronization therapy pacing; SA, sinoatrial. Figure reprinted with permission, Cleveland Clinic Foundation ©2023. All rights reserved.
CARDIAC PHYSIOLOGIC PACING
RVP was the initial pacing strategy for those requiring permanent pacemakers. However, since its development, studies have shown that RVP produces nonphysiologic LV activation with creation of left bundle branch block (LBBB)-like activation. Retrospective analysis of the Mode Selection Trial (MOST) reported that the risk of HF hospitalization and atrial fibrillation significantly increased with higher pacing burden irrespective of pacing mode (2). The Dual Chamber and VVI Implantable Defibrillator (DAVID) trial highlighted the negative consequences of ventricular dyssynchrony with RVP in patients with reduced LV systolic function. One-year survival free of the composite endpoint was 84% for patients treated with ventricular backup pacing at 40 beats/min compared with 73% for patients with dual-chamber rate-responsive pacing at 70 beats/min, with higher rates of mortality and HF hospitalization in the dual-chamber rate-responsive pacing group (3). The DAVID trial contributed to wide adoption of RVP avoidance algorithms in patients who are not dependent on ventricular pacing. In addition to reduction of unnecessary pacing, clinical practice has sought alternative strategies for more physiologic pacing to improve LV dyssynchrony and clinical outcomes. These have included cardiac resynchronization therapy (CRT) and more recently conduction system pacing (CSP), which includes His bundle pacing (HBP) and left bundle branch area pacing (LBBAP).
Cardiac Resynchronization Therapy
CRT has emerged as one of the most remarkable advances in HF therapy and management. Biventricular pacing (BiVP) with short atrioventricular delay was proposed as a supplementary treatment of advanced HF in the 1990s (4, 5). Conduction disease, including LBBB or intraventricular conduction delay, is present in a significant proportion of patients with HF, leading to dyssynchronous ventricular contraction. CRT is typically applied with BiVP, most commonly with a RV endocardial lead and LV pacing delivered via a lead in a LV branch of the coronary sinus or less commonly a lead placed on the LV epicardium. CRT has been shown to improve functional status and improve survival outcomes in selected patients with cardiomyopathy and widened QRS durations reflective of LBBB or intraventricular conduction delay (6). Over the past few decades, the evaluation of CRT has progressed rapidly from case studies to randomized controlled trials (Table 1).
Table 1.
Major clinical trials of cardiac resynchronization therapy
| Study | Year | Sample size |
Endpoint | Design | Findings |
|---|---|---|---|---|---|
| MUSTIC-SR (7) | 2001 | 48 | 6MWT, QOL, pVO2, Hosp | Single-blinded, randomized controlled, crossover, 6 months | CRT-P improved 6MWT, QOL, pVO2; reduced Hosp |
| MIRACLE (8) | 2002 | 453 | NYHA, QOL, pVO2 | Prospective, randomized, double-blind, parallel controlled | CRT-P improved NYHA, pVO2, 6MWT |
| MUSTIC-AF (71) | 2002 | 131 | 6MWT, QOL, pVO2, Hosp | Single-blinded, controlled, crossover, 6 months | CRT-P (high dropout rate) improved all; reduced Hosp |
| PATH-CHF (64) | 2002 | 42 | 6MWT, pVO2 | Single-blinded, randomized, crossover controlled | CRT-P improved 6MWT, pVO2 |
| MIRACLE-ICD (72) | 2003 | 369 | 6MWT, QOL, Hosp | Prospective, multicenter, randomized, double-blind, parallel controlled | CRT-D improved all compared to baseline |
| MIRACLE-ICD II (73) | 2004 | 186 | VE/CO2, pVO2, NYHA, QOL, 6MWT, LV volumes/LVEF | Double-blinded, ICD versus CRT-D, 6 months | CRT-D improved NYHA, VE/CO2, volumes, LVEF |
| COMPANION (10) | 2004 | 1,520 | All-cause death or Hosp | Prospective, multicenter, randomized controlled | CRT-P/CRT-D reduced all-cause death and Hosp |
| CARE-HF (9) | 2005 | 813 | All-cause death or Hosp for major CV event and death from any cause | Prospective, multicenter, randomized controlled | CRT-P reduced all-cause death or Hosp from major CV event and death from any cause |
| REVERSE (74) | 2008 | 610 | HF clinical composite response, LVESVi, HFH, mortality | Double-blinded, randomized controlled, GDMT, CRT-P ± ICD, 12 months | No change in HF clinical composite response or mortality, reduction in LVESVi and HFH |
| MADIT-CRT (75) | 2009 | 1,817 | HF events or death and LVESVi | Randomized controlled, CRT-P, CRT-D, 2.4 years | CRT-D reduced HF events or death and LVESVi but no change in mortality |
| RAFT (76) | 2010 | 1,798 | Death from any cause or HFH, death from any cause, death from CV cause and HFH | Randomized controlled, ICD alone versus CRT-D, 40 months | CRT-D reduced death from any cause or HFH, death from any cause, and HFH |
Abbreviations: 6MWT, 6-min walk test; CRT, cardiac resynchronization therapy; CRT-D, biventricular pacer with a defibrillator; CRT-P, biventricular pacemaker; CV, cardiovascular; GDMT, guideline-directed medical therapy; HF, heart failure; HFH, heart failure hospitalization; Hosp, hospitalizations; ICD, implantable cardioverter-defibrillator; LV left ventricular; LVEF, left ventricular ejection fraction; LVESVi, left ventricular end-systolic volume index; NYHA, New York Heart Association functional class; pVO2, peak oxygen consumption; QOL, quality of life; VE/CO2, ventilation/carbon dioxide ratio.
Multisite Stimulation in Cardiomyopathy (MUSTIC) was the first randomized controlled study on CRT. It showed that BiVP significantly improves exercise tolerance and quality of life in patients with chronic HF and intraventricular conduction delay. Overall, 48 patients who had severe HF [New York Heart Association (NYHA) class III] with normal sinus rhythm and QRS duration of more than 150 ms were enrolled. They underwent CRT placement with leads in the right atrium and each ventricle with a 3-month period of inactive pacing and a 3-month period of active pacing. The primary endpoint, distance walked in 6 min, was significantly greater with active pacing; secondary endpoints of quality of life measured by questionnaire, peak oxygen consumption, and hospitalizations for HF were also superior in the active pacing group (7). Subsequently, the Multicenter InSync Randomized Clinical Evaluation (MIRACLE) study was published, with blinded assessment of the effects of CRT, including symptoms, quality of life, and HF status. Similar to the MUSTIC trial, MIRACLE showed significant clinical improvement in patients who underwent CRT pacing, when randomized to blinded BiVP “on” versus “off” for 6 months (8).
The paradigm shift to use of CRT in selected populations of patients with HF was solidified after large randomized clinical trials demonstrated improvements in primary endpoints of all-cause mortality or HF hospitalizations. In the Cardiac Resynchronization—Heart Failure (CARE HF) and Comparison of Medical Therapy, Pacing, and Defibrillation in Heart Failure (COMPANION) trials (9, 10), patients with NYHA class III and IV HF due to LV systolic dysfunction and ventricular dyssynchrony with QRS duration >120 ms on optimal guideline-directed medical therapy were enrolled. CARE HF evaluated the role of CRT pacing (CRT-P), whereas COMPANION investigated CRT with a defibrillator (CRT-D), with both endpoints of death from any cause and unplanned hospitalizations for HF. These trials showed that CRT decreases HF hospitalizations and overall mortality (9, 10). Technical developments that have improved delivery of CRT include multipolar leads, which enable delivery of LV pacing from different electrode positions, and combinations, which can be advantageous in avoiding diaphragmatic stimulation and achieving more basal stimulation.
The European Society of Cardiology and American College of Cardiology/American Heart Association HF guidelines, as modified in 2013, assigned a class I recommendation for CRT in patients with LV ejection fraction (LVEF) ≤35%, LBBB, QRS duration ≥150 ms, and NYHA functional class III or ambulatory class IV chronic HF despite adequate medical treatment (11, 12). The 2023 Heart Rhythm Society (HRS)/Asia Pacific Heart Rhythm Society (APHRS)/Latin American Heart Rhythm Society (LAHRS) Guideline on Cardiac Physiologic Pacing for the Avoidance and Mitigation of Heart Failure (1) extended class II indications for CRT to select patients with narrower QRS durations or LVEF 36–50% (Figures 2 and 3) (1).
Figure 2.
Cardiac physiologic pacing strategies for patients with pacing indications. Adapted from Reference 1 with permission. Abbreviations: CM, cardiomyopathy; CRT, cardiac resynchronization therapy; HBP, His bundle pacing; LBBAP, left bundle branch area pacing; LBBB, left bundle branch block; LVEF, left ventricular ejection fraction; V, ventricular.
Figure 3.
Cardiac physiologic pacing strategies for patients with heart failure indications. Adapted from Reference 1 with permission. Abbreviations: CRT, cardiac resynchronization therapy; HBP, His bundle pacing; LBBAP, left bundle branch area pacing; LBBB, left bundle branch block; LVEF, left ventricular ejection fraction; NYHA, New York Heart Association functional class.
Conduction System Pacing: His Bundle Pacing
About 30% of patients who undergo implantation of a CRT device do not improve their HF symptoms, and the inability to stimulate severely diseased myocardium or myocardial scar without a large stimulus to QRS latency is a major obstacle for CRT delivery (13). CSP that can reduce intraventricular and atrioventricular dyssynchrony by providing a more physiologic pattern of ventricular electrical activation has gained popularity over the past few years and has emerged as a potential alternative to RVP and in some situations CRT.
The His-Purkinje system is essential for the maintenance of the synchronous ventricular contraction via endocardial to epicardial and apical to basal electrical activation. HBP emerged as an alternative to BiVP with the goal of maintaining a physiologic pattern of ventricular activation via the native His-Purkinje system (14). The His bundle lies within the membranous portion of the interventricular septum surrounded by fibrous connective tissue and divides to form right and left bundles as it enters the muscular septum (15). Initial studies were conducted in animal models (16, 17) and first described in patients in early 2000 by Deshmukh et al., who demonstrated that implanting a transvenous pacing lead to capture the His bundle has the potential to improve cardiac function and LV dimensions compared to RVP (18).
With HBP, an active fixation pacemaker lead is affixed into the proximal intraventricular septum (Figure 1b). There can be selective capture of the His bundle, in which only the His bundle is stimulated, or nonselective capture, involving fused capture of the His bundle and adjacent ventricular tissues (19). Selective capture criteria include (a) pacing stimulus to QRS onset interval equal to native His to QRS onset interval, (b) discrete local ventricular electrogram on the pacing lead, (c) paced QRS morphology matching native QRS morphology, and (d) single capture threshold (His capture). Nonselective capture criteria include (a) pacing stimulus to QRS onset interval usually zero, as there is no isoelectric interval between pacing stimulus and QRS, which can lead to an appearance of a pseudodelta wave from local myocardial capture; (b) local ventricular electrogram directly captured by pacing stimulus without a discrete component; (c) paced QRS duration longer than native QRS duration with concordant paced QRS electrical axis and intrinsic QRS axis; and (d) two distinct capture thresholds—RV and His capture (20). Selective capture results in greater reduction in QRS width; however, nonselective capture appears to have a similar LV activation time and pattern (21).
HBP for patients needing a pacemaker.
Retrospective and prospective studies report overall success rates of 65–94% for HBP used in patients with pacemaker indications (Table 2) (22). Pastore et al. sought to study the effects of HBP on left atrial (LA) function in patients with normal cardiac function who needed permanent pacemaker implantation (23). Patients first underwent 3 months of His area pacing followed by 3 months of RVP with various echo parameters such as systole-diastole (S-D) LV electromechanical delay, S-D intra-LV dyssynchrony, myocardial performance index and mitral annular tissue Doppler early diastolic velocity (E′), and LA function evaluated after each 3-month period. His area pacing showed more physiologic LV electromechanical activation/relaxation with higher LA volumes pre–atrial contraction and improved total emptying fraction and lower risk of atrial fibrillation progression compared to RVP (23). Another large retrospective observational study examined clinical outcomes of HBP at one hospital versus RVP at a sister hospital (24). HBP was successful in 92% of patients. The primary composite endpoint of death, HF hospitalizations, or upgrade to BiVP was significantly lower in the HBP group; this difference was primarily observed in patients with ventricular pacing burden >20% (24). Another multicenter study of 844 patients receiving HBP for pacemaker indications demonstrated reasonable long-term success rates but highlighted high pacing thresholds as a limitation of HBP, which necessitated device replacement due to battery depletion in 19.6% of patients over a mean follow-up of 5.9 years and led to interruption of HBP in 7.6% (25).
Table 2.
Major studies of His bundle pacing
| Reference | Year | Sample size |
Endpoint | Design | Findings |
|---|---|---|---|---|---|
| 18 | 2000 | 18 | LVEDD, LVESD, LVEF, functional status | Prospective, patients with chronic AF, LVEF <40%, QRSd <120 ms | Improvement in LV dimensions, functional status, cardiothoracic ratio, LVEF |
| 77 | 2006 | 18 | NYHA, 6MWT, QOL, hemodynamic parameters | Prospective, crossover, patients with chronic AF, AV junction ablation randomized to 6 months of RVP versus para-Hisian HBP | Improvement in NYHA, 6MWT, QOL, hemodynamic parameters |
| 78 | 2010 | 91 | Not reported | Prospective, all patients with AV block as pacing indication | No long-term clinical outcomes |
| 26 | 2013 | 16 | NYHA, LVEF, LV dimensions | Prospective, HBP in patients with failed LV leads | Improvement in NYHA; improvement in LVEF and LV dimensions |
| 79 | 2014 | 38 | LVEF, NYHA, 6MWT, QOL | Prospective, crossover, randomized patients with AV block, narrow QRS, LVEF >40%, 12 months HBP versus RVP | Improvement in LVEF; no significant improvement in NYHA, 6MWT, QOL |
| 80 | 2015 | 38 | Pacing configurations | Prospective, HBP in patients with indication for CRT | Lower pacing thresholds using bipolar HBP lead and RV lead configuration |
| 81 | 2015 | 94 | HFH, mortality and AF | Prospective, patients with PPM indication, HBP versus RVP | Improvement in HFH; no significant improvement in mortality or AF |
| 24 | 2018 | 332 | All-cause mortality and HFH | Retrospective, patients with PPM indication, HBP versus RVP | Improvement in composite end point (all-cause mortality, HFH, and upgrade to BiV pacing) with HBP |
| 27 | 2018 | 39 | NYHA, LVEF, CRT response | Retrospective, multicenter, HBP in patients with RBBB, CRT indication as primary or rescue strategy | Improvement in NYHA, LVEF, overall response rate |
| 82 | 2019 | 74 | NYHA, LVEF, LV dimensions | Prospective, HBP in patients with LBBB and indication for CRT | Improvement in NYHA and LVEF; decrease in LV dimensions |
| 83 | 2019 | 60 | QRSd, LVEF | Retrospective, HBP for CRT in patients with chronic RVP | Improvement in QRSd and LVEF |
| 29 (His-Sync Trial) | 2019 | 41 | QRSd, LVEF at 6 months, time to cardiovascular Hosp or death at 12 months | Prospective, randomized controlled trial, His-CRT versus CS BiV-CRT; crossover mandated for HBP if failed to achieve 20% QRS narrowing, QRS width ≤130 ms, or high pacing thresholds >5 V/1 ms | His-CRT is not superior to BiV-CRT electrocardiographic (QRSd) or echo (LVEF) parameters |
| 28 (His-Alternative Trial) | 2021 | 50 | QRSd, NYHA, LVEF, LVESV | Prospective, randomized trial His-CRT versus BiV-CRT, with crossover permitted in patients with HF | His-CRT provided similar clinical and physical improvement compared to BiV-CRT at expense of higher pacing thresholds |
Abbreviations: 6MWT, 6-min walk test; AF, atrial fibrillation; AV, atrioventricular; BiV, biventricular; CRT, cardiac resynchronization therapy; CS, coronary sinus; HBP, His bundle pacing; HFH, heart failure hospitalization; Hosp, hospitalizations; LBBB, left bundle branch block; LV left ventricular; LVEDD, left ventricular end-diastolic dimension; LVEF, left ventricular ejection fraction; LVESD, left ventricular end-systolic dimension; LVESV, left ventricular end-systolic volume; NYHA, New York Heart Association functional class; PPM, permanent pacemaker; QOL, quality of life; QRSd, QRS duration; RBBB, right bundle branch block; RVP, right ventricular pacing.
HBP as an alternative to CRT.
Several retrospective and prospective observational studies have used HBP in cases where CRT was indicated but where resynchronization via the coronary sinus was not achievable (26). Success rates range from 56% to 95% with significant improvements reported in QRS duration, LVEF, and NYHA functional class (Table 2). A retrospective multicenter study of 106 patients using HBP for resynchronization therapy reported that HBP was able to narrow the QRS in those with LBBB with improvement in LVEF and NYHA functional class (27). In this study, HBP was attempted as a rescue strategy in patients with failed LV lead or nonresponse to BiVP, or as a primary strategy in patients with AV block, bundle branch block, or high ventricular pacing burden as an alternative to BiVP (27).
In light of high nonresponse rates for traditional CRT, a few small randomized clinical trials have evaluated HBP in patients with CRT indications. His Pacing versus Biventricular Pacing in Symptomatic HF Patients with Left Bundle Branch Block (His-Alternative) was a randomized trial comparing two ways of achieving cardiac resynchronization (28). Fifty patients with LVEF ≤35% with LBBB and NYHA class II–IV were randomized to His-CRT versus biventricular CRT (BiV-CRT). For HF patients with LBBB, His-CRT provided similar clinical and physical improvement, including NYHA class improvement, higher LVEF, and lower LV end systolic volume compared to BiV-CRT at the expense of higher pacing thresholds (28). The His-SYNC (His Bundle Pacing versus Coronary Sinus Pacing for Cardiac Resynchronization Therapy) trial was a multicenter randomized controlled trial where patients eligible for conventional CRT were randomized to His-CRT or BiV-CRT with coronary sinus lead placement (29). Primary outcomes were reduction in QRS duration, absolute improvement in LVEF and echocardiographic response (LVEF improvement >5%) at 6 months, and cardiovascular hospitalization or death at 12 months. With a small sample size of 41 patients, study outcomes showed no significant differences in QRS duration between the two groups, similar improvement in LVEF, and no significant differences in cardiac hospitalization or death. HBP patients had higher pacing thresholds compared with BiV-CRT. This trial showed that HBP was not superior to CRT; however, it suggested HBP could be a reasonable alternative for patients undergoing CRT where coronary sinus anatomy is unfavorable (29). HOPE-HF (His Optimized Pacing Evaluated for Heart Failure trial) (30) was a multicenter prospective randomized double-blinded crossover study in 167 patients with LVEF ≤40%, PR ≥ 200 ms, and either QRS duration ≤140 s or right bundle branch block (RBBB). Patients had His bundle leads implanted and were randomized to 6 months of pacing and 6 months of no pacing in a crossover design. Neither the primary endpoint of peak oxygen uptake during symptom-limited exercise nor LVEF was increased with HBP, but quality of life improved significantly with HBP. Combined HBP with BiVP, known as His-optimized CRT (HOT-CRT), was tested in a small feasibility study of 27 patients with indications for CRT; this study reported significantly reduced paced QRS duration compared to BiVP alone and improvements in LVEF, LV volumes, and NYHA functional class (31).
In patients with conduction system disease, the success of HBP depends on the location of disease within the His-Purkinje system and, in patients with LBBB and HF, whether HBP can correct the bundle branch block. A study by Upadhyay et al. (32) showed that among 72 patients with a LBBB pattern, complete conduction block within the proximal left conduction system was observed in 64% of patients and intact Purkinje activation was seen in 36%. HBP corrected wide QRS in 85% of those with complete conduction block [both left intrahisian and proximal left bundle branch (LBB)], whereas no patients with intact Purkinje activation demonstrated correction of QRS with HBP. Disease localized to left intrahisian block was most amenable to corrective HBP by recruitment of latent Purkinje fibers (32). Intracardiac data might be more useful in predicting response to HBP than surface electrocardiography.
A limitation of HBP is its higher pacing thresholds compared to conventional RVP leads, which may lead to lower pacemaker battery longevity, need for lead revision, and/or need for placement of a back-up ventricular lead in patients who are pacemaker dependent.
Conduction System Pacing: Left Bundle Branch Area Pacing
In patients with prolonged His-ventricular interval or LBBB, HBP may not correct left bundle conduction time because of more distal disease, in which case LBBAP may be more useful. LBBAP (Figure 1c) is a newer technique, with a limited number of studies reporting clinical outcomes and relatively short-term follow-up. Nevertheless, LBBAP leads can be easier to place than HBP leads due to the wider target areas, and pacing thresholds are generally excellent, overcoming some of the limitations of HBP. Achieving effective LBBAP with capture of the LBB or its fascicles or capture of the left septum remains important to demonstrate (33).
The initial technique of LBBAP was developed by Huang et al. (34) in 2017. The lead tip is advanced into the deep septum from the RV septal endocardium to the subendocardial LV septum to directly stimulate the left bundle area. Optimal sites of fixation are generally 1–1.5 cm distal to the His bundle. Paced QRS morphology changes from LBBB toward a RBBB pattern as the lead is advanced into the septum (34, 35). LBBAP capture can be confirmed with various criteria, which include (a) paced QRS morphology that shows an r′ pattern in the V1 lead, (b) peak LV activation time measured from onset of pacing spike to peak R wave in leads V5–6 < 80 ms, (c) LBB potential with sharp high-frequency deflection distance 15–30 ms to onset of surface QRS, (d) retrograde His or anterograde distal LBB potentials, (e) demonstration of differential septal and LBB capture thresholds, potentially with programmed stimulation, and (f) differential thresholds of selective LBBAP capturing only LBB or nonselective LBBAP capturing both LBB and adjacent local septal myocardium (36).
LBBAP in patients with pacing indications.
Retrospective or prospective observational studies in patients with pacemaker or CRT indications report success rates of 85–97% for LBBAP with improvements in QRS duration, LV synchrony, NYHA functional class, LVEF, and LV dimensions (Table 3) (37). Long-term safety and feasibility of LBBAP were demonstrated by a prospective study where LBBAP was attempted in 632 patients with a 97.8% success rate. At 2-year follow-up, patients had low lead complication rates, which were defined as a rise in LBB threshold >3V or loss of capture, lead dislodgments, or right bundle branch injury. Patients with LBBB had significant decreases in QRS duration, and postimplant LVEF improved in patients with QRS >120 ms (38). In the Geisinger-Rush CSP registry, 703 patients undergoing pacemaker implantation for bradycardia were divided into LBBAP (n = 321) and RVP groups (n = 382) with a primary outcome of composite all-cause mortality, HF hospitalization, or upgrade to BiVP. LBBAP compared to RVP was associated with lower composite outcomes of mortality and HF hospitalizations among patients with pacing burden >20% (39). LBBAP was evaluated in patients with HF with persistent atrial fibrillation undergoing AV junctional ablation, showing a decrease in HF hospitalizations, death, and inappropriate implantable cardioverter-defibrillator therapies (40). However, randomized trials remain limited to small feasibility studies.
Table 3.
Major studies of left bundle branch area pacing
| Reference | Year | Sample size |
Endpoint | Design | Findings |
|---|---|---|---|---|---|
| 84 | 2019 | 85 | LV mechanical synchrony (on SPECT MPI) | Retrospective, LBBAP versus HBP in patients with AV block | LV mechanical synchrony similar in HBP and LBBAP |
| 85 | 2019 | 33 | LV synchrony (echo strain analysis), QRSd | Prospective, observational in patients with AV block | LV synchrony improved at 3-month follow-up compared to baseline, improved QRSd |
| 43 | 2019 | 11 | NYHA, NTproBNP, LVESD, LVEF | Retrospective, in patients with CRT indication | Improvement in NYHA, NTproBNP, LV size and function |
| 44 | 2021 | 137 | LVEF, NYHA, pacing threshold | Observational, patients with indication for CRT assigned to LBBAP versus HBP versus BiV pacing | HBP and LBBAP had greater improvement in LVEF and NYHA than BiV pacing. Pacing threshold lower in LBBAP than HBP |
| 86 | 2020 | 63 | NYHA, LVESV, LVEF | Prospective multicenter observational study in patients with indication for CRT | Improvement in NYHA, LVESV, LVEF |
| 41 | 2021 | 325 | QRSd, LVEF, clinical (no HFH and improvement in NYHA) and echocardiographic responses (≥5% improvement in LVEF) | Retrospective, multicenter in patients with indication for CRT | Improvement in QRSd, clinical and echocardiographic response |
| 38 | 2021 | 632 | QRSd, LVEF, LBB capture threshold, lead revision | Prospective, observational study, in patients with attempted LBBAP for bradycardia or HF indication | LBBAP had high success rate with low complications during follow-up up to 24 months |
| 87 (LBBP-RESYNC Trial) | 2022 | 40 | LVEF, LVESV, NTproBNP; NYHA, 6MWT, QRSd | Prospective, randomized, LBBAP versus BiV pacing in patients with NICM, LBBB, and NYHA II–IV for 6 months follow-up | LBBAP had better LVEF improvement, greater reduction in LVESV and NTproBNP versus BiV-CRT; NYHA, 6MWT, QRSd similar in both groups |
Abbreviations: 6MWT, 6-min walk test; AF, atrial fibrillation; AV, atrioventricular; BiV, biventricular; CRT, cardiac resynchronization therapy; HBP, His bundle pacing; HF, heart failure; HFH, heart failure hospitalization; LBB, left bundle branch; LBBB, left bundle branch block; LBBAP, left bundle branch area pacing; LV, left ventricular; LVEDD, left ventricular end-diastolic dimension; LVEF, left ventricular ejection fraction; LVESD, left ventricular end-systolic dimension; LVESV, left ventricular end-systolic volume; MPI, myocardial performance index; NICM, nonischemic cardiomyopathy; NTproBNP, N-terminal pro-brain natriuretic peptide; NYHA, New York Heart Association functional class; QRSd, QRS duration; RBBB: right bundle branch block; SPECT, single photon emission computed tomography.
LBBAP as an alternative to CRT.
LBBAP may be comparable or even preferable to CRT in some patients with its ability to correct LBBB and reverse LV dyssynchrony. There have been a few observational studies involving CRT-eligible patients who had LBBAP (Table 3). LBBAP in patients with HF with LBBB achieved reduction in QRS duration, shortened LV activation time, and improved LVEF and LV dimensions (41, 42). In comparison to BiVP, NYHA class, plasma N-terminal pro-brain natriuretic peptide (NTproBNP) levels, and LV size and function were significantly improved. LBBAP achieved a greater increase in LVEF compared to BiVP, with much lower pacing threshold and higher sensing than with HBP (43, 44).
Recent small studies suggesting potential benefit of LBBAP for patients in need of CRT include Left Bundle Branch Pacing Versus Biventricular Pacing in Cardiac Resynchronization Therapy: A Randomized Controlled Pilot Trial (LBBP-RESYNC Trial). This was a prospective study of nonischemic cardiomyopathy patients with LBBB randomized to LBBAP or BiVP with 6-month follow-up. The primary endpoint was difference in LVEF improvement between the two groups. The LBBAP-CRT group demonstrated greater LVEF improvement (mean difference 5.6%) than the BiVP-CRT group with improvement in LV end-systolic volume, NTproBNP level, NYHA class, and 6-min walk distance (45).
Recent studies have also demonstrated LBBAP as a viable alternative option for CRT in patients who failed traditional BiVP due to coronary sinus lead implantation failure or failure to respond to BiVP (46). CSP with HBP or LBBAP compared to BiVP was studied in an observational nonrandomized cohort of 477 patients with class I or II indications for CRT; the investigators reported that CSP was associated with significantly lower mortality and HF hospitalization (47). The Left Bundle Branch-Optimized CRT (LOT-CRT) study, from an international LBBAP collaborative group, assessed LBBAP-optimized CRT combined with a coronary venous LV pacing lead with conventional BiVP [RV and coronary sinus (CS) leads] versus LBBAP alone (without a CS lead). LOT-CRT was associated with a narrower QRS duration than BiVP or LBBAP alone and showed improved LVEF, LV volume, and NYHA class compared to baseline. The authors concluded that LOT-CRT is a feasible option and could be an alternative when suboptimal electrical resynchronization is obtained with BiV-CRT (48). Pivotal randomized clinical trials for CSP compared to CRT in patients meeting indications for CRT are ongoing or planned. These include the Left versus Left Randomized Clinical Trial (NCT05650658), which will compare His or LBB pacing to BiVP in >2,000 patients with HF (LVEF ≤ 50%) and either with a wide QRS (≥130 ms) or with/anticipated >40% pacing. The need for continued assessment of long-term outcomes was highlighted by a European MELOS registry of LBBAP outcomes in 2,533 patients that reported an 8.3% incidence of complications specific to the ventricular transseptal pacing route (49).
Though LBBAP has been the latest approach introduced for cardiac physiologic pacing, LBBAP has been emerging as a dominant approach over HBP due to its superior technical performance with lower pacing thresholds, better sensing, and larger implant target. Recommendations for CSP, including HBP and LBBAP, are reported in the 2023 HRS/APHRS/LAHRS Guideline on Cardiac Physiologic Pacing for the Avoidance and Mitigation of Heart Failure and are summarized in Figures 1 and 2 (1).
LEADLESS PACING
Leadless pacemakers were recently introduced to address lead- and pocket-related complications in conventional transvenous pacemaker systems. Transvenous leads can cause complications, such as venous obstruction, tricuspid regurgitation, and endocarditis, with transvenous lead-related endocarditis mortality rates up to 12–31% (50-52). Pocket-related complications can include skin erosion, pocket infection, and potential sepsis (53, 54). To address these issues with transvenous pacemaker systems, a leadless pacemaker was initially conceptualized in the 1970s (55) and implanted in 2014 (56). The first approved devices are single-component systems that can provide single-chamber RVP, sensing, and delivery of rate response. One of the single-chamber devices has an algorithm to sense atrial contractions, enabling a degree of AV synchronous pacing. Leadless pacemakers are implanted via a percutaneous femoral catheter-based approach to advance the pacemaker into the RV.
One of the first trials in leadless pacing enrolled 33 patients at three centers to evaluate clinical safety and performance of a RV leadless pacing system. The device was successfully implanted in 97% of patients with overall complication-free rate of 94%, adequate electrical performance, and no device-related complications at 12-month follow-up (56, 57). A second clinical study enrolled 526 patients with implantation success rate of 96%, with device-related complications (such as pericardial effusion, vascular complications, and device dislodgment) reported in 6.4% of patients (58). At 12-month follow-up, the complication rate was 71% lower than that of transvenous systems (59). Using another RV leadless device, another study enrolled 725 patients; the device was successfully implanted in 99% of patients, with major complications such as cardiac injury and vascular access issues occurring in 4% of patients (60). Short-term complications of leadless pacemakers and transvenous pacemaker systems were comparable (61). Long-term, 12-month follow-up of leadless pacemakers compared to propensity-matched transvenous system controls showed a 48% lower risk of major complications and 82% lower occurrence of revisions or reinterventions (62, 63). Longer-term outcome data are needed for leadless pacemakers, as at the end of service, either devices are extracted and replaced, if possible, or an additional device(s) is placed. A dual-chamber leadless pacemaker system was recently approved by the FDA.
Leadless LV endocardial pacing has also been studied for CRT in prospective multicenter observational studies (63, 64). One such system consists of a subcutaneous ultrasound transmitter/pulse generator and a rice-sized receiver electrode implanted in the LV endocardium. Ultrasonic energy is delivered to the LV endocardial receiver electrode to achieve BiVP. Studies have included patients who failed conventional CRT due to failed CS lead cannulation or non-responders to traditional CRT. However, significant complications include cardioembolic stroke, pericardial effusion, catheter-induced ventricular fibrillation, electrode embolization, and femoral artery access issues (65-70).
FUTURE DIRECTIONS
Conventional CRT has evolved over the years into a highly effective therapy for HF patients with LBBB. CSP has emerged as an alternative pacing strategy for those patients in whom conventional CRT does not produce satisfactory clinical response, when there is difficulty with CS lead placement, or as an alternative to RVP. Recent studies have shown comparable and in some cases more effective electrical and mechanical resynchronization with CSP with HBP or LBBAP compared to BiVP, although the evidence base for CRT includes randomized controlled trials showing improved survival and reduction in HF hospitalizations in selected patients with HF and widened QRS duration, particularly with LBBB and QRS duration ≥150 ms. Although CSP reestablishes physiologic ventricular activation, it has lagged the evidence base of CRT with a lack of large randomized controlled trial data. However, large randomized clinical trials are launching and should provide guidance as to the suitability of CSP for HF patients. Leadless pacemakers have emerged recently with promising efficacy and safety and could be potentially used for dual-chamber pacing or resynchronization therapy. Leadless cardiac pacemakers may potentially transform cardiac pacing; however, long-term data are needed to verify device safety, performance, and extractability.
Glossary
- AV
atrioventricular
- RV
right ventricular
- RVP
right ventricular pacing
- LV
left ventricular
- HF
heart failure
- LBBB
left bundle branch block
- CRT
cardiac resynchronization therapy
- CSP
conduction system pacing
- HBP
His bundle pacing
- LBBAP
left bundle branch area pacing
- BiVP
biventricular pacing
- LVEF
left ventricular ejection fraction
- BiV-CRT
biventricular cardiac resynchronization therapy
Footnotes
DISCLOSURE STATEMENT
The authors are not aware of any affiliations, memberships, funding, or financial holdings that might be perceived as affecting the objectivity of this review.
LITERATURE CITED
- 1.Chung MK, Patton KK, Lau C-P, et al. 2023. 2023 HRS/APHRS/LAHRS guideline on cardiac physiologic pacing for the avoidance and mitigation of heart failure. Heart Rhythm 20(9):E17–91 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2.Sweeney MO, Hellkamp AS, Ellenbogen KA, et al. 2003. Adverse effect of ventricular pacing on heart failure and atrial fibrillation among patients with normal baseline QRS duration in a clinical trial of pacemaker therapy for sinus node dysfunction. Circulation 107(23):2932–37 [DOI] [PubMed] [Google Scholar]
- 3.Wilkoff BL, Cook JR, Epstein AE, et al. 2002. Dual-chamber pacing or ventricular backup pacing in patients with an implantable defibrillator: the Dual Chamber and VVI Implantable Defibrillator (DAVID) Trial. JAMA 288(24):3115–23 [DOI] [PubMed] [Google Scholar]
- 4.Mower M. 1990. Method and apparatus for treating hemodynamic dysfunction by simultaneous pacing of both ventricles. US Patent US4928688A
- 5.Hochleitner M, Hörtnagl H, Ng CK, et al. 1990. Usefulness of physiologic dual-chamber pacing in drug-resistant idiopathic dilated cardiomyopathy. Am. J. Cardiol 66(2):198–202 [DOI] [PubMed] [Google Scholar]
- 6.Leclercq C, Cazeau S, Ritter P, et al. 2000. A pilot experience with permanent biventricular pacing to treat advanced heart failure. Am. Heart J 140(6):862–70 [DOI] [PubMed] [Google Scholar]
- 7.Cazeau S, Leclercq C, Lavergne T, et al. 2001. Effects of multisite biventricular pacing in patients with heart failure and intraventricular conduction delay. N. Engl. J. Med 344(12):873–80 [DOI] [PubMed] [Google Scholar]
- 8.Abraham WT, Fisher WG, Smith AL, et al. 2002. Cardiac resynchronization in chronic heart failure. N. Engl. J. Med 346(24):1845–53 [DOI] [PubMed] [Google Scholar]
- 9.Cleland JG, Daubert J-C, Erdmann E, et al. 2005. The effect of cardiac resynchronization on morbidity and mortality in heart failure. N. Engl. J. Med 352(15):1539–49 [DOI] [PubMed] [Google Scholar]
- 10.Bristow MR, Saxon LA, Boehmer J, et al. 2004. Cardiac-resynchronization therapy with or without an implantable defibrillator in advanced chronic heart failure. N. Engl. J. Med 350(21):2140–50 [DOI] [PubMed] [Google Scholar]
- 11.Brignole M, Auricchio A, Baron-Esquivias G, et al. 2013. 2013 ESC guidelines on cardiac pacing and cardiac resynchronization therapy: the task force on cardiac pacing and resynchronization therapy of the European Society of Cardiology (ESC). Developed in collaboration with the European Heart Rhythm Association (EHRA). Europace 15(8):1070–118 [DOI] [PubMed] [Google Scholar]
- 12.Epstein AE, DiMarco JP, Ellenbogen KA, et al. 2013. 2012 ACCF/AHA/HRS focused update incorporated into the ACCF/AHA/HRS 2008 guidelines for device-based therapy of cardiac rhythm abnormalities: a report of the American College of Cardiology Foundation/American Heart Association Task Force on Practice Guidelines and the Heart Rhythm Society. Circulation 127(3):e283–52 [DOI] [PubMed] [Google Scholar]
- 13.Herweg B, Ilercil A, Madramootoo C, et al. 2006. Latency during left ventricular pacing from the lateral cardiac veins: a cause of ineffectual biventricular pacing. Pacing Clin. Electrophysiol 29(6):574–81 [DOI] [PubMed] [Google Scholar]
- 14.Lustgarten DL, Crespo EM, Arkhipova-Jenkins I, et al. 2015. His-bundle pacing versus biventricular pacing in cardiac resynchronization therapy patients: a crossover design comparison. Heart Rhythm 12(7):1548–57 [DOI] [PubMed] [Google Scholar]
- 15.Vijayaraman P, Dandamudi G. 2016. Anatomical approach to permanent His bundle pacing: optimizing His bundle capture. J. Electrocardiol 49(5):649–57 [DOI] [PubMed] [Google Scholar]
- 16.Scherlag BJ, Kosowsky BD, Damato AN. 1967. A technique for ventricular pacing from the His bundle of the intact heart. J. Appl. Physiol 22(3):584–87 [DOI] [PubMed] [Google Scholar]
- 17.Karpawich PP, Gates J, Stokes KB. 1992. Septal His-Purkinje ventricular pacing in canines: a new endocardial electrode approach. Pacing Clin. Electrophysiol 15(11 Pt.2):2011–15 [DOI] [PubMed] [Google Scholar]
- 18.Deshmukh P, Casavant DA, Romanyshyn M, Anderson K. 2000. Permanent, direct His-bundle pacing: a novel approach to cardiac pacing in patients with normal His-Purkinje activation. Circulation 101(8):869–77 [DOI] [PubMed] [Google Scholar]
- 19.Vijayaraman P, Dandamudi G, Zanon F, et al. 2018. Permanent His bundle pacing: recommendations from a multicenter His bundle pacing collaborative working group for standardization of definitions, implant measurements, and follow-up. Heart Rhythm 15(3):460–68 [DOI] [PubMed] [Google Scholar]
- 20.Vijayaraman P, Chung MK, Dandamudi G, et al. 2018. His bundle pacing. J. Am. Coll. Cardiol 72(8):927–47 [DOI] [PubMed] [Google Scholar]
- 21.Upadhyay GA, Tung R. 2017. Selective versus non-selective his bundle pacing for cardiac resynchronization therapy. J. Electrocardiol 50(2):191–94 [DOI] [PubMed] [Google Scholar]
- 22.Sharma PS, Vijayaraman P, Ellenbogen KA. 2020. Permanent His bundle pacing: shaping the future of physiological ventricular pacing. Nat. Rev. Cardiol 17(1):22–36 [DOI] [PubMed] [Google Scholar]
- 23.Pastore G, Aggio S, Baracca E, et al. 2014. Hisian area and right ventricular apical pacing differently affect left atrial function: an intra-patients evaluation. Europace 16(7):1033–39 [DOI] [PubMed] [Google Scholar]
- 24.Abdelrahman M, Subzposh FA, Beer D, et al. 2018. Clinical outcomes of His bundle pacing compared to right ventricular pacing. J. Am. Coll. Cardiol 71(20):2319–30 [DOI] [PubMed] [Google Scholar]
- 25.Zanon F, Abdelrahman M, Marcantoni L, et al. 2019. Long term performance and safety of His bundle pacing: a multicenter experience. J. Cardiovasc. Electrophysiol 30(9):1594–601 [DOI] [PubMed] [Google Scholar]
- 26.Barba-Pichardo R, Sánchez AM, Fernández-Gómez JM, et al. 2013. Ventricular resynchronization therapy by direct His-bundle pacing using an internal cardioverter defibrillator. Europace 15(1):83–88 [DOI] [PubMed] [Google Scholar]
- 27.Sharma PS, Dandamudi G, Herweg B, et al. 2018. Permanent His-bundle pacing as an alternative to biventricular pacing for cardiac resynchronization therapy: a multicenter experience. Heart Rhythm 15(3):413–20 [DOI] [PubMed] [Google Scholar]
- 28.Vinther M, Risum N, Svendsen JH, et al. 2021. A randomized trial of His pacing versus biventricular pacing in symptomatic HF patients with left bundle branch block (His-alternative). JACC Clin. Electrophysiol 7(11):1422–32 [DOI] [PubMed] [Google Scholar]
- 29.Upadhyay GA, Vijayaraman P, Nayak HM, et al. 2019. His corrective pacing or biventricular pacing for cardiac resynchronization in heart failure. J. Am. Coll. Cardiol 74(1):157–59 [DOI] [PubMed] [Google Scholar]
- 30.Whinnett ZI, Shun-Shin MJ, Tanner M, et al. 2023. Effects of haemodynamically atrio-ventricular optimized His bundle pacing on heart failure symptoms and exercise capacity: the His Optimized Pacing Evaluated for Heart Failure (HOPE-HF) randomized, double-blind, cross-over trial. Eur. J. Heart Fail 25(2):274–83 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 31.Vijayaraman P, Herweg B, Ellenbogen KA, Gajek J. 2019. His-optimized cardiac resynchronization therapy to maximize electrical resynchronization. Circ. Arrhythm. Electrophysiol 12(2):e006934. [DOI] [PubMed] [Google Scholar]
- 32.Upadhyay GA, Cherian T, Shatz DY, et al. 2019. Intracardiac delineation of septal conduction in left bundle-branch block patterns. Circulation 139(16):1876–88 [DOI] [PubMed] [Google Scholar]
- 33.Burri H, Jastrzębski M, Cano O, et al. 2023. EHRA clinical consensus statement on conduction system pacing implantation: endorsed by the Asia Pacific Heart Rhythm Society (APHRS), Canadian Heart Rhythm Society (CHRS), and Latin American Heart Rhythm Society (LAHRS). Europace 25(4):1208–36 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 34.Huang W, Su L, Wu S, et al. 2017. A novel pacing strategy with low and stable output: pacing the left bundle branch immediately beyond the conduction block. Can. J. Cardiol 33(12):1736.e1–3 [DOI] [PubMed] [Google Scholar]
- 35.Jastrzębski M, Kiełbasa G, Moskal P, et al. 2021. Fixation beats: a novel marker for reaching the left bundle branch area during deep septal lead implantation. Heart Rhythm 18(4):562–69 [DOI] [PubMed] [Google Scholar]
- 36.Liu P, Wang Q, Sun H, et al. 2021. Left bundle branch pacing: current knowledge and future prospects. Front. Cardiovasc. Med 8:630399. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 37.Muthumala A, Vijayaraman P. 2021. Clinical outcomes of His-Purkinje conduction system pacing. Pacing Clin. Electrophysiol 44(1):5–14 [DOI] [PubMed] [Google Scholar]
- 38.Su L, Wang S, Wu S, et al. 2021. Long-term safety and feasibility of left bundle branch pacing in a large single-center study. Circ. Arrhythm. Electrophysiol 14(2):e009261. [DOI] [PubMed] [Google Scholar]
- 39.Sharma PS, Patel NR, Ravi V, et al. 2022. Clinical outcomes of left bundle branch area pacing compared to right ventricular pacing: results from the Geisinger-Rush Conduction System Pacing Registry. Heart Rhythm 19(1):3–11 [DOI] [PubMed] [Google Scholar]
- 40.Wang S, Wu S, Xu L, et al. 2019. Feasibility and efficacy of His bundle pacing or left bundle pacing combined with atrioventricular node ablation in patients with persistent atrial fibrillation and implantable cardioverter-defibrillator therapy. J. Am. Heart Assoc 8(24):e014253. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 41.Vijayaraman P, Ponnusamy SS, Cano O, et al. 2021. Left bundle branch area pacing for cardiac resynchronization therapy: results from the International LBBAP Collaborative Study Group. JACC Clin. Electrophysiol 7(2):135–47 [DOI] [PubMed] [Google Scholar]
- 42.Vijayaraman P, Subzposh FA, Naperkowski A, et al. 2019. Prospective evaluation of feasibility and electrophysiologic and echocardiographic characteristics of left bundle branch area pacing. Heart Rhythm 16(12):1774–82 [DOI] [PubMed] [Google Scholar]
- 43.Zhang W, Huang J, Qi Y, et al. 2019. Cardiac resynchronization therapy by left bundle branch area pacing in patients with heart failure and left bundle branch block. Heart Rhythm 16(12):1783–90 [DOI] [PubMed] [Google Scholar]
- 44.Wu S, Su L, Vijayaraman P, et al. 2021. Left bundle branch pacing for cardiac resynchronization therapy: nonrandomized on-treatment comparison with His bundle pacing and biventricular pacing. Can. J. Cardiol 37(2):319–28 [DOI] [PubMed] [Google Scholar]
- 45.Wang Y, Zhu H, Hou X, et al. 2022. Randomized trial of left bundle branch versus biventricular pacing for cardiac resynchronization therapy. J. Am. Coll. Cardiol 80(13):1205–16 [DOI] [PubMed] [Google Scholar]
- 46.Vijayaraman P, Herweg B, Verma A, et al. 2022. Rescue left bundle branch area pacing in coronary venous lead failure or nonresponse to biventricular pacing: results from International LBBAP Collaborative Study Group. Heart Rhythm 19(8):1272–80 [DOI] [PubMed] [Google Scholar]
- 47.Vijayaraman P, Zalavadia D, Haseeb A, et al. 2022. Clinical outcomes of conduction system pacing compared to biventricular pacing in patients requiring cardiac resynchronization therapy. Heart Rhythm 19(8):1263–71 [DOI] [PubMed] [Google Scholar]
- 48.Jastrzębski M, Moskal P, Huybrechts W, et al. 2022. Left bundle branch-optimized cardiac resynchronization therapy (LOT-CRT): results from an international LBBAP collaborative study group. Heart Rhythm 19(1):13–21 [DOI] [PubMed] [Google Scholar]
- 49.Jastrzębski M, Kiełbasa G, Cano O, et al. 2022. Left bundle branch area pacing outcomes: the multicentre European MELOS study. Eur. Heart J 43(40):4161–73 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 50.Brunner MP, Cronin EM, Wazni O, et al. 2014. Outcomes of patients requiring emergent surgical or endovascular intervention for catastrophic complications during transvenous lead extraction. Heart Rhythm 11(3):419–25 [DOI] [PubMed] [Google Scholar]
- 51.Tarakji KG, Wazni OM, Harb S, et al. 2014. Risk factors for 1-year mortality among patients with cardiac implantable electronic device infection undergoing transvenous lead extraction: the impact of the infection type and the presence of vegetation on survival. Europace 16(10):1490–95 [DOI] [PubMed] [Google Scholar]
- 52.Harcombe AA, Newell SA, Ludman PF, et al. 1998. Late complications following permanent pacemaker implantation or elective unit replacement. Heart 80(3):240–44 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 53.Klug D, Balde M, Pavin D, et al. 2007. Risk factors related to infections of implanted pacemakers and cardioverter-defibrillators: results of a large prospective study. Circulation 116(12):1349–55 [DOI] [PubMed] [Google Scholar]
- 54.Essebag V, Verma A, Healey JS, et al. 2016. Clinically significant pocket hematoma increases long-term risk of device infection: BRUISE CONTROL INFECTION Study. J. Am. Coll. Cardiol 67(11):1300–8 [DOI] [PubMed] [Google Scholar]
- 55.Spickler JW, Rasor NS, Kezdi P, et al. 1970. Totally self-contained intracardiac pacemaker. J. Electrocardiol 3(3–4):325–31 [DOI] [PubMed] [Google Scholar]
- 56.Reddy VY, Knops RE, Sperzel J, et al. 2014. Permanent leadless cardiac pacing: results of the LEADLESS trial. Circulation 129(14):1466–71 [DOI] [PubMed] [Google Scholar]
- 57.Knops RE, Tjong FVY, Neuzil P, et al. 2015. Chronic performance of a leadless cardiac pacemaker: 1-year follow-up of the LEADLESS trial. J. Am. Coll. Cardiol 65(15):1497–504 [DOI] [PubMed] [Google Scholar]
- 58.Reddy VY, Exner DV, Cantillon DJ, et al. 2015. Percutaneous implantation of an entirely intracardiac leadless pacemaker. N. Engl. J. Med. 373(12):1125–35 [DOI] [PubMed] [Google Scholar]
- 59.Cantillon DJ, Dukkipati SR, Ip JH, et al. 2018. Comparative study of acute and mid-term complications with leadless and transvenous cardiac pacemakers. Heart Rhythm 15(7):1023–30 [DOI] [PubMed] [Google Scholar]
- 60.Reynolds DW, Ritter P. 2016. A leadless intracardiac transcatheter pacing system. N. Engl. J. Med 374(26):2604–5 [DOI] [PubMed] [Google Scholar]
- 61.Tjong FV, Reddy VY. 2017. Permanent leadless cardiac pacemaker therapy: a comprehensive review. Circulation 135(15):1458–70 [DOI] [PubMed] [Google Scholar]
- 62.Duray GZ, Ritter P, El-Chami M, et al. 2017. Long-term performance of a transcatheter pacing system: 12-month results from the Micra Transcatheter Pacing Study. Heart Rhythm 14(5):702–9 [DOI] [PubMed] [Google Scholar]
- 63.Ritter P, Duray GZ, Steinwender C, et al. 2015. Early performance of a miniaturized leadless cardiac pacemaker: the Micra Transcatheter Pacing Study. Eur. Heart J 36(37):2510–19 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 64.Auricchio A, Stellbrink C, Sack S, et al. 2002. Long-term clinical effect of hemodynamically optimized cardiac resynchronization therapy in patients with heart failure and ventricular conduction delay. J. Am. Coll. Cardiol 39(12):2026–33 [DOI] [PubMed] [Google Scholar]
- 65.Auricchio A, Delnoy P-P, Butter C, et al. 2014. Feasibility, safety, and short-term outcome of leadless ultrasound-based endocardial left ventricular resynchronization in heart failure patients: results of the Wireless Stimulation Endocardially for CRT (WiSE-CRT) study. Europace 16(5):681–88 [DOI] [PubMed] [Google Scholar]
- 66.Reddy VY, Miller MA, Neuzil P, et al. 2017. Cardiac resynchronization therapy with wireless left ventricular endocardial pacing: the SELECT-LV study. J. Am. Coll. Cardiol 69(17):2119–29 [DOI] [PubMed] [Google Scholar]
- 67.Gamble JHP, Herring N, Ginks M, et al. 2018. Endocardial left ventricular pacing for cardiac resynchronization: systematic review and meta-analysis. Europace 20(1):73–81 [DOI] [PubMed] [Google Scholar]
- 68.Gamble JHP, Herring N, Ginks MR, et al. 2018. Endocardial left ventricular pacing across the interventricular septum for cardiac resynchronization therapy: clinical results of a pilot study. Heart Rhythm 15(7):1017–22 [DOI] [PubMed] [Google Scholar]
- 69.Biffi M, Defaye P, Jaïs P, et al. 2018. Benefits of left ventricular endocardial pacing comparing failed implants and prior non-responders to conventional cardiac resynchronization therapy: a subanalysis from the ALSYNC study. Int. J. Cardiol 259:88–93 [DOI] [PubMed] [Google Scholar]
- 70.Graham AJ, Providenica R, Honarbakhsh S, et al. 2018. Systematic review and meta-analysis of left ventricular endocardial pacing in advanced heart failure: clinically efficacious but at what cost? Pacing Clin. Electrophysiol 41(4):353–61 [DOI] [PubMed] [Google Scholar]
- 71.Linde C, Leclercq C, Rex S, et al. 2002. Long-term benefits of biventricular pacing in congestive heart failure: results from the MUltisite STimulation in cardiomyopathy (MUSTIC) study. J. Am. Coll. Cardiol 40(1):111–18 [DOI] [PubMed] [Google Scholar]
- 72.Young JB, Abraham WT, Smith AL, et al. 2003. Combined cardiac resynchronization and implantable cardioversion defibrillation in advanced chronic heart failure: the MIRACLE ICD Trial. JAMA 289(20):2685–94 [DOI] [PubMed] [Google Scholar]
- 73.Abraham WT, Young JB, León AR, et al. 2004. Effects of cardiac resynchronization on disease progression in patients with left ventricular systolic dysfunction, an indication for an implantable cardioverter-defibrillator, and mildly symptomatic chronic heart failure. Circulation 110(18):2864–68 [DOI] [PubMed] [Google Scholar]
- 74.Linde C, Abraham WT, Gold MR, et al. 2008. Randomized trial of cardiac resynchronization in mildly symptomatic heart failure patients and in asymptomatic patients with left ventricular dysfunction and previous heart failure symptoms. J. Am. Coll. Cardiol 52(23):1834–43 [DOI] [PubMed] [Google Scholar]
- 75.Zareba W, Klein H, Cygankiewicz I, et al. 2011. Effectiveness of cardiac resynchronization therapy by QRS morphology in the Multicenter Automatic Defibrillator Implantation Trial–Cardiac Resynchronization Therapy (MADIT-CRT). Circulation 123(10):1061–72 [DOI] [PubMed] [Google Scholar]
- 76.Tang ASL, Wells GA, Talajic M, et al. 2010. Cardiac-resynchronization therapy for mild-to-moderate heart failure. N. Engl. J. Med 363(25):2385–95 [DOI] [PubMed] [Google Scholar]
- 77.Occhetta E, Bortnik M, Magnani A, et al. 2006. Prevention of ventricular desynchronization by permanent para-Hisian pacing after atrioventricular node ablation in chronic atrial fibrillation: a crossover, blinded, randomized study versus apical right ventricular pacing. J. Am. Coll. Cardiol 47(10):1938–45 [DOI] [PubMed] [Google Scholar]
- 78.Barba-Pichardo R, Moriña-Vázquez P, Fernández-Gómez JM, et al. 2010. Permanent His-bundle pacing: seeking physiological ventricular pacing. Europace 12(4):527–33 [DOI] [PubMed] [Google Scholar]
- 79.Kronborg MB, Mortensen PT, Poulsen SH, et al. 2014. His or para-His pacing preserves left ventricular function in atrioventricular block: a double-blind, randomized, crossover study. Europace 16(8):1189–96 [DOI] [PubMed] [Google Scholar]
- 80.Su L, Xu L, Wu S-J, Huang W-J. 2015. Pacing and sensing optimization of permanent His-bundle pacing in cardiac resynchronization therapy/implantable cardioverter defibrillators patients: value of integrated bipolar configuration. Europace 18(9):1399–405 [DOI] [PubMed] [Google Scholar]
- 81.Sharma PS, Dandamudi G, Naperkowski A, et al. 2015. Permanent His-bundle pacing is feasible, safe, and superior to right ventricular pacing in routine clinical practice. Heart Rhythm 12(2):305–12 [DOI] [PubMed] [Google Scholar]
- 82.Huang W, Su L, Wu S, et al. 2019. Long-term outcomes of His bundle pacing in patients with heart failure with left bundle branch block. Heart 105(2):137–43 [DOI] [PubMed] [Google Scholar]
- 83.Vijayaraman P, Herweg B, Dandamudi G, et al. 2019. Outcomes of His-bundle pacing upgrade after long-term right ventricular pacing and/or pacing-induced cardiomyopathy: insights into disease progression. Heart Rhythm 16(10):1554–61 [DOI] [PubMed] [Google Scholar]
- 84.Hou X, Qian Z, Wang Y, et al. 2019. Feasibility and cardiac synchrony of permanent left bundle branch pacing through the interventricular septum. Europace 21(11):1694–702 [DOI] [PubMed] [Google Scholar]
- 85.Li X, Li H, Ma W, et al. 2019. Permanent left bundle branch area pacing for atrioventricular block: feasibility, safety, and acute effect. Heart Rhythm 16(12):1766–73 [DOI] [PubMed] [Google Scholar]
- 86.Huang W, Wu S, Vijayaraman P, et al. 2020. Cardiac resynchronization therapy in patients with nonischemic cardiomyopathy using left bundle branch pacing. JACC Clin. Electrophysiol 6(7):849–58 [DOI] [PubMed] [Google Scholar]
- 87.Wang Y, Zhu H, Hou X, et al. 2022. Randomized trial of left bundle branch versus biventricular pacing for cardiac resynchronization therapy. J. Am. Coll. Cardiol 80(13):1205–16 [DOI] [PubMed] [Google Scholar]