Proposal of an Alternative Near-Minimum Isodose Surface DV-0.01 cc Equally Minimizing Gross Tumor Volume Below the Relevant Dose as the Basis for Dose Prescription and Evaluation of Stereotactic Radiosurgery for Brain Metastases

Kazuhiro Ohtakara 1,2,, Kojiro Suzuki 2
Editors: Alexander Muacevic, John R Adler
PMCID: PMC11069632  PMID: 38707120

Abstract

Introduction

In stereotactic radiosurgery (SRS) for brain metastasis (BM), the prescribed dose is generally reported as a minimum dose to cover a specific percentage (e.g. D98%) of the gross tumor volume (GTV) with or without a margin or an unspecified intended marginal dose to the GTV boundary. In dose prescription to a margin-added planning target volume (PTV), the GTV marginal dose is likely variable and unclear. This study aimed to reveal major flaws of dose prescription to a fixed % coverage of a target volume (TV), such as GTV D98% or PTV D95%, and to propose an alternative.

Materials and methods

Seven quasi-spherical models with volumes ranging from 1.00 to 15.00 cc were assumed as GTVs. The GTVs and the volumes generated by adding isotropic 1- and 2-mm margins to the GTV boundaries (GTV + 1 and 2 mm) were used for SRS planning, dose prescription, and evaluation. Volumetric-modulated arcs with a 5-mm leaf-width multileaf collimator were used to optimize each SRS plan to ensure the steepest dose gradient outside each TV boundary. In dose prescription to the GTV D98%, 0.02-0.3 cc of the GTV is below the prescribed dose, and the volume increases with larger GTVs. The volume below the prescribed dose should be less than the equivalent of a 3-mm-diameter lesion, i.e. 0.01 cc. Therefore, DV-0.01 cc was defined as an alternative near-minimum dose for which the TV below a relevant dose is less than 0.01 cc. Four different dose prescriptions, including the GTV DV-0.01 cc, were compared using specific doses in 1, 3, and 5 fractions, equivalent to 80, 60, and 50 Gy, respectively, as biologically effective doses (BEDs) to the boundaries of GTV, GTV + 1 mm, and GTV + 2 mm, respectively.

Results

Dose prescription to the GTV DV-0.01 cc corresponds to 95.0, 98.0, and 99.0-99.93% coverages for the GTV of 0.20, 0.50, and 1.00-15.00 cc, respectively. The GTV DV-0.01 cc varied substantially and decreased significantly as the GTV increased in dose prescriptions to the GTV D98%, GTV + 1 mm D95%, and GTV + 2 mm D95%. The GTV + 2 mm DV-0.01 cc increased significantly as the GTV increased, except for the dose prescription to the GTV + 2 mm D95% with a decreasing tendency. When comparing BED-based specific dose prescriptions, dose prescription to the GTV DV-0.01 cc was optimal in terms of the following: 1) consistency of the near-minimum dose of GTV; 2) the highest BED at 2 mm outside the GTV, except for 1.00 cc GTV, and the rational increase with increasing GTV; and 3) the highest BED at 2 mm inside the GTV. In dose prescription with the BED of 80 Gy in 1 fraction and 5 fractions to the GTV DV-0.01 cc, the GTV limits were ≤1.40 and ≤8.46 cc, respectively, in order for the irradiated isodose volume not to exceed the proposed thresholds for minimizing the risk of brain radionecrosis.

Conclusions

Dose prescription to a fixed % coverage of a GTV with or without a margin leads to the substantially varied near-minimum dose at the GTV boundary, which significantly decreases with increasing GTV. Alternatively, GTV DV-0.01 cc with a variable coverage (D>95%) for >0.20 cc GTV and fixed D95% for ≤0.20 cc GTV is recommended as the basis for dose prescription and evaluation, along with supplemental evaluation of the marginal dose of the GTV plus a margin (e.g. GTV + 2 mm) to demonstrate the appropriateness of dose attenuation outside the GTV boundary.

Keywords: dose inhomogeneity, marginal dose, near-minimum dose, dose evaluation, dose distribution, target definition, volumetric-modulated arc therapy, dose prescription, stereotactic radiosurgery, brain metastasis

Introduction

Stereotactic radiosurgery (SRS) for brain metastases (BM) is performed using a variety of devices and irradiation techniques. The more precipitous dose falloff outside a target volume (TV) boundary with excellent normal tissue sparing is closely associated with the extremely inhomogeneous TV dose and vice versa [1-3]. Thus, a steep dose gradient outside and inside the TV boundary is an intrinsic characteristic of the dose distribution for SRS and is quite different from those for conventional photon radiotherapy with the homogeneous TV dose. This dose distribution characteristic needs to be tailored to SRS for BM to maximize effectiveness and safety. Volumetric-modulated arcs (VMAs) without unnecessary dose constraints within the TV can provide the optimal dose distribution for SRS in terms of dose conformity, normal tissue sparing, and concentrically laminated steep dose increase inside the TV boundary, compared to other techniques such as dynamic conformal arcs [3-5].

In the target definition and dose prescription, a different margin (e.g. 0-3 mm) is added to a gross tumor volume (GTV) to generate the planning target volume (PTV), and various PTV coverage percentages are used based on a prescribed dose [2,5,6]. The prescribed dose has been reported in various forms, from an unspecified intended marginal dose to a specific % coverage of a TV, such as D98%, where DX% is a minimum dose covering ≥X% of a TV [6-9]. A minimum dose covering a TV reduced by 0.035 cc (DV-0.035 cc) is also recommended as the representative near-minimum dose for <2 cc TV [7]. A minimum dose of 0.001 cc units is uncommon since it is highly sensitive to even the slightest differences in contouring. The differences in TV definition and dose prescription have made it difficult to compare treatment outcomes among facilities and reach a consensus on the optimal dose. In dose prescription to a margin-added PTV, the GTV marginal dose is likely variable and unclear. Therefore, there has been a growing tendency to refer to the GTV marginal dose (e.g. GTV D98%) [3,10]. The reproducibility of near-minimum or near-maximum absorbed doses of GTV is relatively higher in the recent sophisticated image-guided SRS of BM compared to that of lung or liver tumors [11].

The local tumor control probability generally decreases with larger tumors [6,12]. Larger tumors tend to infiltrate into the surrounding brain more frequently and deeply, expand the internal hypoxic region, and increase radioresistance [3,13]. To maintain anti-tumor efficacy regardless of GTV and to attain a high rate of induction of nearly complete local remission after SRS, it is necessary to increase the biologically effective dose (BED) and target coverage for large tumors [14]. The current general policy of lowering the marginal dose for larger tumors and adhering to ≤5 fractions is expected to result in decreased local control of larger tumors [6]. Since 2018, our facility has maintained the GTV marginal dose (D≥98%) at BED ≥80 Gy in principle, using flexible dose fractions ranging from 3 to 15 to improve both the efficacy and safety of the treatment, where the BED calculation is based on the linear-quadratic formula with an alpha/beta ratio of 10 (BED10) [15-17]. The number of cases of partial response and subsequent regrowth has decreased markedly; however, there is still scope for improvement [17-19].

This study was conducted to reconsider the optimal method for dose prescription to ensure the equivalent BED at the GTV boundary, regardless of GTV, and to determine the appropriate representative near-minimum dose at the GTV boundary for dose evaluation and reporting. Specifically, we aimed to reveal fundamental flaws of general dose prescription to a fixed % coverage of a GTV with or without a margin and propose an alternative method through comparisons of various target definitions and prescription methods using VMA.

This study was approved by the Clinical Research Review Board of Kainan Hospital Aichi Prefectural Welfare Federation of Agricultural Cooperatives (20220727-1).

Materials and methods

This was a planning study for clinical scenarios of a single BM and was intended to address the issues revealed in the previous study [3]. In this study, the GTV was assumed as a contrast-enhanced lesion that is almost equal to or slightly larger than the mass on T2-weighted images (T2-mass) if visible, without notable T1/T2 mismatch [20,21], or a T2-mass itself in cases with paradoxical T1/T2 mismatch (T2-mass > enhancing lesion) [22], which can be deemed pathologically equivalent to a tumor mass excluding microscopic brain invasion [23,24].

We first considered how small a part of the GTV can be effectively controlled, even by a steep dose attenuation margin outside the clinically prescribed isodose surface (IDS). A BM ≥3 mm in diameter is distinguishable from the contrast-enhanced blood vessels and can be a candidate for SRS, particularly in multiple lesions. The volumes of spheres and ellipsoids with a diameter of 2-7 mm and the typical GTV at the largest diameter are shown in Table 1.

Table 1. Spheres, ellipsoids, and typical gross tumor volumes with a maximum diameter of 2–7 mm.

The ratio of the diameters of the ellipsoids is 1:0.9:0.8

GTV: gross tumor volume

Maximum diameter (mm) 2 3 4 5 6 7
Sphere (cc) 0.004 0.014 0.034 0.065 0.113 0.180
Ellipsoid (1:0.9:0.8) (cc) 0.003 0.010 0.024 0.047 0.081 0.129
Typical example of GTV (cc) 0.004 0.01 0.03 0.06 0.10 0.15

The GTVs with volumes of 0.01, 0.02, and 0.035 cc correspond to spheres with diameters of 2.7, 3.4, and 4.1 mm, respectively [7]. The cubic volumes with diameters of 2 and 3 mm are 0.008 and 0.027 cc, respectively. Assuming that one cancer cell is a 20-µm sphere, it may be estimated that 0.01 cc of cancer tissue contains approximately 10 million cancer cells. Considering further the possibilities of tumor growth and/or displacement between the image acquisition and the initiation of irradiation and intra- and/or inter-fractional tumor displacement [15], a part of GTV that falls below the prescribed dose should preferably be less than 0.01 cc. Table 2 shows the parts of GTVs that are less than the prescribed dose in the dose prescription of D98% for GTVs of 1-15 cc.

Table 2. Two percent of GTV from 1 to 15 cc and the corresponding sphere diameter.

*The dose prescription to the GTV D98% results in less than 2% of the GTV being below the prescribed dose.​​​​​ GTV D2% and above irradiate a volume equivalent to 2% of GTV

**The diameter (to the first decimal place) of a sphere corresponding to the 2% volume of GTV

GTV: gross tumor volume; DX%: a minimum dose covering at least X% of a target volume

GTV (cc) 1.00 2.50 5.00 7.50 10.00 12.50 15.00
2% of GTV (cc)* 0.02 0.05 0.10 0.15 0.20 0.25 0.30
Diameter (mm)** 3.4 4.6 5.8 6.6 7.3 7.8 8.3

The volume that accounts for less than 2% of GTV is equivalent to less than a 3.4 mm diameter sphere, even for a 1-cc GTV, which inevitably increases as GTV increases. For a 10-cc GTV, the equivalent of a 7.2-mm-diameter sphere (<0.20 cc) is less than the prescribed dose. Therefore, an alternative near-minimum dose that corresponds to a minimum dose covering a TV reduced by 0.01 cc was defined as DV-0.01 cc in this study. Table 3 shows the GTV coverages (%) that are necessary to meet the GTV DV-0.01 cc requirement, including GTVs of <1 and 100 cc.

Table 3. GTV coverage with the minimum dose to cover GTV minus 0.01 cc.

The coverage value of a GTV by a minimum dose covering the volume subtracted by 0.01 cc from the GTV (GTV DV-0.01 cc). The coverage rate (%) of a ≥5.00 cc GTV (or ≥99.8% coverage) should be specified to the second decimal place, since 0.1% of the 5.00 and 10.00 cc GTVs are 0.005 and 0.01 cc, respectively. Rounding to the third decimal place for GTVs at approximately 0.005 cc affects whether the GTV is considered approximately 0.01 cc

GTV: gross tumor volume; DV-0.01 cc: a minimum dose to cover a target volume minus 0.01 cc

GTV (cc) 0.20 0.50 1.00 2.50 5.00 7.50 10.00 12.50 15.00 100.00
Coverage (%) 95.0 98.0 99.0 99.6 99.80 99.87 99.90 99.92 99.93 99.99

The GTV D95% and D98% are sufficient for GTVs of ≤0.2 and 0.5 cc, respectively, while a ≥99% coverage of a GTV is required for a ≥1 cc GTV with increasing GTV. In this study, GTV DV-0.01 cc was positioned and evaluated as the most relevant metric representing the GTV marginal dose. Similarly, DV-0.01 cc and DV-0.035 cc of the margin-added volumes (GTV + 1 and 2 mm) were also evaluated as reference values [7] since an isotropic margin-added volume can vary depending on the software algorithm being used even though it is the same at 2 mm.

In this study, we compared the following four dose prescription methods: GTV D98%, GTV DV-0.01 cc, GTV + 1 mm D95%, and GTV + 2 mm D95%, as shown in Table 4.

Table 4. Dose prescription methods compared in this study and specific prescribed doses based on biologically effective doses by a number of fractions.

Absolute doses and the corresponding BEDs (in parentheses) up to the first decimal places by a number of fractions. The BED is based on the linear-quadratic formula with an alpha/beta ratio of 10 (BED10)

fr: fraction(s); GTV: gross tumor volume: GTV + X mm: GTV evenly expanded by X mm; DX%: a minimum dose covering at least X% of a target volume; a in Pa: alternative; DV-0.01 cc: a minimum dose to cover a target volume minus 0.01 cc

Dose prescription Target volume Prescribed isodose 1 fr 3 fr 5 fr
GTV_P GTV D98% 23.8 Gy (80.4 Gy) 36.3 Gy (80.2 Gy) 43.0 Gy (80.0 Gy)
GTV_Pa DV-0.01 cc
GTV+1_P GTV + 1 mm D95% 20.0 Gy (60.0 Gy) 30.0 Gy (60.0 Gy) 35.2 Gy (60.0 Gy)
GTV+2_P GTV + 2 mm D95% 18.0 Gy (50.4 Gy) 26.6 Gy (50.2 Gy) 30.9 Gy (50.0 Gy)

Seven quasi-spherical GTV models were generated with slight manual modifications using the sphere drawing tool in 1-mm increments in a dedicated software tool, MIM MaestroTM (MIM Software, Cleveland, Ohio), so that the volume was precisely 1.00-15.00 cc [3]. The volumes equally expanded by 1 and 2 mm of the GTV boundaries were also generated as GTV + 1 mm and GTV + 2 mm using the MIM Maestro software. The volumes of GTV + 1 and 2 mm and the coverage values for the DV-0.01 cc and DV-0.035 cc are shown in Table 5.

Table 5. The volume of GTV evenly expanded by 1 and 2 mm and the % coverage with the minimum dose to cover the volume minus 0.01 or 0.035 cc.

*The volumes (GTV + 1 mm, GTV + 2 mm) were generated using the MIM Maestro software. The volumes vary depending on the system algorithm used, even for the same GTV

The coverages of the GTV + 1 and 2 mm by the minimum dose that covers the volume subtracted by 0.01 and 0.035 cc from the GTV + 1 mm and 2 mm

GTV: gross tumor volume; DV-0.01 cc: a minimum dose to cover a target volume minus 0.01 cc; DV-0.035 cc: a minimum dose to cover a target volume minus 0.035 cc

GTV (cc) 1.00 2.50 5.00 7.50 10.00 12.50 15.00
GTV + 1 mm (cc)* 1.52 3.43 6.43 9.35 12.25 15.05 17.91
Coverage (%) of GTV + 1 mm for DV-0.01 cc 99.35 99.71 99.85 99.89 99.92 99.93 99.94
Coverage (%) of GTV + 1 mm for DV-0.035 cc 97.70 98.98 99.46 99.63 99.71 99.77 99.80
GTV + 2 mm (cc)* 2.29 4.71 8.37 11.80 15.15 18.43 21.60
Coverage (%) of GTV + 2 mm for DV-0.01 cc 99.56 99.79 99.88 99.92 99.93 99.95 99.95
Coverage (%) of GTV + 2 mm for DV-0.035 cc 98.47 99.26 99.58 99.70 99.77 99.81 99.84

To evaluate the dose gradient inside the GTV boundary, the volumes equally reduced by 2 mm of the GTV boundaries were generated as GTV - 2 mm using the MIM Maestro software. The volumes of GTV - 2 mm and the GTV coverage values corresponding to the GTV - 2 mm DV-0.01 cc are shown in Table 6.

Table 6. The volume of GTV evenly reduced by 2 mm (GTV – 2 mm) and the GTV coverage with the minimum dose to cover the volume (GTV – 2 mm) minus 0.01 cc.

*The volumes (GTV - 2 mm) were generated using the MIM Maestro software

**These values for GTV itself are completely different from the coverage values (96.7%–99.90%) for the GTV – 2 mm DV-0.01 cc

GTV: gross tumor volume; DV-0.01 cc: a minimum dose to cover a target volume minus 0.01 cc

GTV (cc) 1.00 2.50 5.00 7.50 10.00 12.50 15.00
GTV - 2 mm (cc)* 0.31 1.10 2.66 4.37 6.13 7.97 9.88
GTV coverage (%)** for DV-0.01 cc of GTV - 2 mm 30.20 43.60 53.06 58.15 61.20 63.64 65.78

Three VMA-based SRS (VMARS) plans were generated for the three different TV boundaries (GTV, GTV + 1 mm, GTV + 2 mm) of each GTV. The treatment platform was a 5-mm multileaf collimator Agility® (Elekta AB, Stockholm, Sweden) mounted in a linac Infinity® (Elekta AB, Stockholm, Sweden) with a flattening filter-free mode of a 6 MV X-ray beam, which provides a dose rate of up to 1400 monitor unit per minute [3]. A planning system, Monaco® (Elekta AB, Stockholm, Sweden), was used to optimize VMARS [3]. In Monaco, the difference between the contoured TV and the TV in the dose-volume histogram (DVH) may differ by ≥0.01 cc, depending on the TV. In that case, the coverage value of the DV-0.01 cc was applied instead of reducing 0.01 cc to the TV in the DVH. The isocenters were set at the exact coordinates near the GTV center. The head computed tomography images (1-mm slice thickness) and GTV localization (the right lateral thalamus) used were identical to those employed in the previous study [3]. The arc arrangement consisted of one coplanar arc with an arc length of 360º and two non-coplanar arcs with each arc length of 180º, which are allocated to evenly divide the cranial hemisphere. The collimator angles for each arc are separately set to be 0, 45, and 90º. The VMARS plans were optimized with the Pareto mode, using the simplest and unified combination of cost functions (CFs) by slightly modifying the previously reported method [3]. The minimization of surrounding tissue dose (steepest dose gradient) outside the GTV was prioritized, for which only three CFs were adopted for VMARS optimization, without imposing any dose constraints on the GTV internal dose, as shown in Table 7.

Table 7. CFs used to optimize volumetric-modulated arc-based radiosurgery planning.

*The option of multicriterial optimization is to continue to drive normal tissue sparing while maintaining the target dose

The only three CFs with the optional settings were uniformly applied to GTV, GTV + 1 mm, and GTV + 2 mm for the GTVs of 1.00–15.00 cc

TV: target volume; fr: fractions; RMS: root mean square; GTV: gross tumor volume; GTV + X mm: GTV evenly expanded by X mm

Structures Cost functions Basic settings
TV Target penalty Prescription, e.g. 43.0 Gy/5 fr to GTV; Minimum volume ≥99.00%
Body contour (Patient) Quadratic overdose Maximum dose, e.g. 43.0 Gy; RMS dose excess: 0.02 Gy; Multicriterial* +; TV margin 0.15 cm
Conformality Relative isoconstraint 0.01; Margin around target: 4 cm; Multicriterial* +

The dose calculation, including tissue heterogeneity correction, was based on the X-ray voxel Monte Carlo algorithm with a grid spacing of 1.0 mm and a statistical uncertainty of 1.0% per calculation. The segment shape optimization was included with the highest plan quality (20), the maximum control points of 1024 per arc, the minimum segment width of 0.5 cm, and the medium fluence smoothing.

Following the completion of optimization, each prescribed dose was rescaled according to the dose prescription methods shown in Table 4. The dose distributions based on the four dose prescriptions were compared. Specifically, the tumor volume dependence and the homogeneity of the marginal doses at the GTV boundaries and at 2 mm outside and inside the GTV boundaries were compared. Additionally, the specific dose prescriptions based on BED10 in 5 fractions shown in Table 4, were adopted to compare the anti-tumor efficacy, and the irradiated isodose volume (IIV) at 1, 3, and 5 fractions was used to compare the safety of the treatment. Approximately 18-24 Gy is commonly used in a single fraction (BED10 50.4-81.6 Gy), 27-30 Gy in 3 fractions (BED10 51.3-60.0 Gy), and 30-35 Gy in 5 fractions (BED10 48.0-59.5 Gy). Doses equivalent to the BED10s of 80, 60, and 50 Gy were prescribed to the boundaries of GTV, GTV + 1 mm, and GTV + 2 mm, respectively. The GTV D0.01 cc was used as a metric for the near-maximum (central) dose instead of the maximum dose based on a 0.001-cc unit, D2%, or D0.035 cc [7].

An X Gy volume was defined as an IIV receiving at least X Gy, including the TV [25]. A 24 Gy volume in 5 fractions and a 20 Gy volume in 3 fractions of <20 cc are associated with <10% risk of any brain radionecrosis, while a 12 Gy volume in 1 fraction of ≥5 and ≥10 cc are associated with the risks of symptomatic brain radionecrosis of approximately 10% and 15%, respectively [26,27]. Standard dose conformity and gradient indices were not used to evaluate the quality of treatment plans since these indices consist of ratios of multiple volumes and are strongly influenced by the TV and its coverage value by the prescribed dose [28,29]. Larger TVs are generally associated with better values and vice versa [4,28,29]. Even within the same plan, when the coverage value by the prescribed dose differs, these indices vary significantly [28,29].

For statistical analyses, paired nonparametric tests were used, considering the distributions of the variables. The Spearman’s rank correlation coefficient (SRCC) was used to evaluate the correlations between two numerical variables. Friedman’s test (FT) with Kendall’s coefficient of concordance (KCC) and Scheffe’s post hoc test (SPHT) were used to compare three or more numerical variables. Significance was considered at P < 0.05 (*), P < 0.01 (**), and P < 0.001 (***).

Results

The correlations between the GTV and the relevant marginal dose metrics relative to each prescribed isodose (100%) in the four different dose prescriptions are shown in Figure 1 and Table 8.

Table 8. Correlations between GTV and the marginal doses relative to each prescribed dose in four different dose prescriptions.

The correlations between the GTVs and the marginal dose metrics (%) relative to the prescribed dose (100%) are shown as the results of SRCC

PD: prescribed dose; GTV: gross tumor volume; DV-0.01 cc: a minimum dose to cover a target volume minus 0.01 cc; GTV + X mm: GTV evenly expanded by X mm; DX%: a minimum dose covering at least X% of a target volume; NS: not significant

Dose prescription Dose metrics Relative dose (range, %) to PD Spearman’s rho P value
GTV D98% GTV DV-0.01 cc 95.4-98.8 -0.937 0.002**
GTV + 1 mm DV-0.01 cc 78.9-84.5 0.786 0.036*
GTV + 2 mm DV-0.01 cc 62.4-71.9 0.893 0.007**
GTV DV-0.01 cc GTV D98% 101.2-104.9 0.937 0.002**
GTV + 1 mm DV-0.01 cc 79.8-88.2 0.786 0.036*
GTV + 2 mm DV-0.01 cc 63.1-75.2 0.929 0.003**
GTV + 1 mm D95% GTV DV-0.01 cc 100.8-109.1 -0.901 0.006**
GTV + 1 mm DV-0.01 cc 85.8-94.1 0.750 0.052 (NS)
GTV + 2 mm DV-0.01 cc 68.8-81.6 0.893 0.007**
GTV + 2 mm D95% GTV DV-0.01 cc 113.3-130.7 -0.883 0.009**
GTV + 1 mm DV-0.01 cc 102.3-113.6 -0.685 0.090 (NS)
GTV + 2 mm DV-0.01 cc 89.0-94.8 -0.523 0.229 (NS)

Figure 1. Correlations between GTV and the relevant marginal doses relative to a prescribed dose in four different dose prescriptions.

Figure 1

The scatter plots show the correlations between the GTVs and the marginal dose metrics in dose prescriptions to the GTV D98% (A); GTV DV-0.01 cc (B); GTV + 1 mm D95% (C); and GTV + 2 mm D95% (D) as shown in Table 4

The GTV DV-0.01 cc (A, C, D), GTV D98% (B), GTV + 1 mm DV-0.01 cc (A-D), and GTV + 2 mm DV-0.01 cc (A-D) are shown as the relative doses (%) to each prescribed dose (100%)

PD: prescribed dose; GTV: gross tumor volume; GTV + X mm: GTV evenly expanded by X mm; DX%: a minimum dose covering at least X% of a target volume; DV-0.01 cc: a minimum dose to cover a target volume minus 0.01 cc

 

The GTV DV-0.01 cc significantly decreased as GTV increased, obviously, except for the GTV_Pa prescription, while the GTV D98% significantly increased as GTV increased in the GTV_Pa prescription. The decrease in the GTV DV-0.01 cc with increasing GTV was remarkable, specifically in the GTV+2_P prescription. The GTV + 1 and 2 mm DV-0.01 cc significantly increased as GTV increased in the GTV_P and GTV_Pa prescriptions. In the GTV+1_P prescription, the GTV + 2 mm DV-0.01 cc significantly increased as the GTV increased, while the GTV + 1 mm DV-0.01 cc only showed an increasing trend. However, the GTV + 1 and 2 mm DV-0.01 cc showed a decreasing tendency as GTV increased in the GTV+2_P prescription.

A difference in the GTV dose inhomogeneities caused by the difference in the optimization targets for VMARS and correlations between the GTV and the GTV dose inhomogeneities are shown in Figure 2A.

Figure 2. Differences in the GTV dose inhomogeneities and the correlations between GTV and the marginal doses in four specific dose prescriptions based on BEDs in 5 fractions.

Figure 2

The scatter plots show the correlations between the GTV and the GTV dose inhomogeneities (A); GTV and the BED10 of GTV DV-0.01 cc (B); GTV and the BED10 of GTV + 2 mm DV-0.01 cc (C); and GTV and the BED10 of GTV + 2 mm DV-0.035 cc

A: Differences in the GTV dose inhomogeneities caused by the difference in the optimization targets (GTV, GTV + 1 mm, and GTV + 2 mm) are shown as the correlations between the GTV and the isodose surface (%) of the GTV DV-0.01 cc relative to the GTV D0.01 cc (100%), for which the results of SRCCs are added. B–D: The four specific dose prescriptions followed the 5-fraction examples shown in Table 4

GTV: gross tumor volume; BED: biologically effective dose; BED10: biologically effective dose based on the linear-quadratic model with an alpha/beta ratio of 10; GTV + X mm: GTV evenly expanded by X mm; DV-0.01 cc: a minimum dose to cover a target volume minus 0.01 cc; DV-0.035 cc: a minimum dose to cover a target volume minus 0.035 cc; DX%: a minimum dose covering at least X% of a target volume; D0.01 cc: a minimum dose covering 0.01 cc of a target volume; SRCC: Spearman’s rank correlation coefficient

The difference in the TVs for VMARS optimization without any internal dose constraint resulted in a significant difference in the GTV dose heterogeneities (FT: P = 0.002 **, KCC = 0.878; SPHT: GTV vs. GTV + 2 mm, P = 0.002 **). VMRAS optimization for the GTV boundary without a margin resulted in the most heterogeneous GTV dose and was noticeable at the 5 cc GTV. The GTV dose inhomogeneities decreased as the GTV increased above 5 cc.

The correlations between the GTV and the GTV DV-0.01 cc and the differences in the four different BED10-based specific dose prescriptions according to Table 4 are shown in Figure 2B and Table 9.

Table 9. Correlations between GTV and the GTV DV-0.01 cc in three specific dose prescriptions based on BED.

The specific dose prescriptions followed the 5-fraction examples shown in Table 4. The correlations between the GTV and the BED10 of GTV DV-0.01 cc are shown as the results of SRCC

GTV: gross tumor volume; GTV + X mm: GTV evenly expanded by X mm; DV-0.01 cc: a minimum dose to cover a target volume minus 0.01 cc; DX%: a minimum dose covering at least X% of a target volume; BED10: biologically effective dose based on the linear-quadratic model with an alpha/beta ratio of 10; SRCC: Spearman’s rank correlation coefficient

Dose prescription GTV DV-0.01 cc (range, BED10) Spearman’s rho P value
GTV D98% 74.9-78.9 Gy -0.890 0.007 **
GTV + 1 mm D95% 60.7-67.9 Gy -0.901 0.006 **
GTV + 2 mm D95% 59.5-73.0 Gy -0.964 <0.001 ***

The BED10s of GTV DV-0.01 cc significantly decreased as the GTV increased, obviously, except for the GTV_Pa prescription. The BED10s of GTV DV-0.01 cc for the GTV+1_P and GTV+2_P prescriptions were significantly lower than those for the GTV_Pa prescription despite assertively allowing the dose increases inside the GTV boundary (FT: P < 0.001 ***, KCC = 0.918; SPHT: GTV_Pa vs. GTV+2_P, P = 0.002 **; GTV_Pa vs. GTV+1_P, P = 0.012 *).

The correlations between the GTV and the GTV + 2 mm DV-0.01 cc, and the GTV and the GTV + 2 mm DV-0.035 cc, and the differences in the four different BED10-based specific dose prescriptions are shown in Figures 2C, 2D, and Table 10.

Table 10. Correlations between GTV and the GTV + 2 mm DV-0.01 cc and DV-0.035 cc in four specific dose prescriptions based on BED.

The specific dose prescriptions followed the 5-fraction examples shown in Table 4. The correlations between the GTV and the BED10s of the GTV + 2 mm DV-0.01 cc and DV-0.035 cc are shown as the results of SRCC

GTV: gross tumor volume; GTV + X mm: GTV evenly expanded by X mm; DV-0.01 cc: a minimum dose to cover a target volume minus 0.01 cc; DV-0.035 cc: a minimum dose to cover a target volume minus 0.035 cc; DX%: a minimum dose covering at least X% of a target volume; BED10: biologically effective dose based on the linear-quadratic model with an alpha/beta ratio of 10; SRCC: Spearman’s rank correlation coefficient

Prescription isodose GTV + 2 mm DV-0.01 cc (range, BED10) Spearman’s rho P value GTV + 2 mm DV-0.035 cc (range, BED10) Spearman’s rho P value
GTV D98% 41.4-50.2 Gy 0.893 0.007** 45.6-53.8 Gy 0.893 0.007**
GTV DV-0.01 cc 41.9-53.2 Gy 0.929 0.003** 46.3-57.6 Gy 0.821 0.023*
GTV + 1 mm D95% 36.0-45.8 Gy 0.964 <0.001*** 41.9-49.2 Gy 0.764 0.046*
GTV + 2 mm D95% 44.7-46.5 Gy -0.595 0.159 (NS) 48.7-50.8 Gy -0.685 0.090 (NS)

The GTV + 2 mm DV-0.01 cc and DV-0.035 cc significantly increased as the GTV increased, except for the GTV+2_P prescription, which showed a decreasing trend. The BED10 of GTV + 2 mm DV-0.01 cc in the GTV_Pa prescription was highest for GTV ≥2.50 cc and >50 Gy for GTV ≥7.5 cc. The BED10 of GTV + 2 mm DV-0.035 cc in the GTV_Pa prescription was highest and >50 Gy for GTV ≥2.50 cc.

The correlations between the GTV and the GTV D50% and the GTV and the GTV -2 mm DV-0.01 cc are shown in Figure 3 and Table 11. The correlations between the GTV D50% and the GTV - 2 mm DV-0.01 cc in the four different BED10-based specific dose prescriptions are shown in Table 11.

Table 11. Correlations between GTV, GTV D50%, and GTV – 2 mm DV-0.01 in four specific dose prescriptions based on BED in 5 fractions.

The specific dose prescriptions followed the 5-fraction examples shown in Table 4

GTV: gross tumor volume; DX%: a minimum dose covering at least X% of a target volume; GTV – 2 mm: GTV evenly reduced by 2 mm; DV-0.01 cc: a minimum dose to cover a target volume minus 0.01 cc; BED: biologically effective dose; GTV + X mm: GTV evenly expanded by X mm; BED10: biologically effective dose based on the linear-quadratic model with an alpha/beta ratio of 10; vs: versus

Dose prescription GTV D50% (range, BED10) Spearman’s rho (vs GTV) (P value) GTV – 2 mm DV-0.01 cc (range, BED10) Spearman’s rho (vs GTV) (P value) Spearman’s rho (GTV D50% vs DV-0.01 cc) (P value)
GTV D98% 115.1-123.5 Gy 0.775 (P = 0.041 *) 108.2-134.4 Gy -1.000 -0.775 (P = 0.041 *)
GTV DV-0.01 cc 117.0-132.9 Gy 0.857 (P = 0.014 *) 116.5-136.5 Gy -1.000 -0.857 (P = 0.014 *)
GTV + 1 mm D95% 95.8-108.0 Gy -0.964 (P < 0.001 ***) 88.3-124.3 Gy -1.000 0.964 (P < 0.001 ***)
GTV + 2 mm D95% 82.3-98.6 Gy -0.847 (P = 0.016 *) 59.5-73.0 Gy -0.964 (P < 0.001 ***) 0.883 (P = 0.009 **)

Figure 3. Correlations between GTV and the GTV internal doses in four specific dose prescriptions based on BED in 5 fractions.

Figure 3

The scatter plots show the correlations between the GTV and the BED10 of GTV D50% (A) and the GTV and the BED10 of GTV – 2 mm DV-0.01 cc (B)

The specific dose prescriptions followed the 5-fraction examples shown in Table 4

A: The BED10s of the GTV D50% in dose prescriptions to the GTV D98%, GTV DV-0.01 cc, GTV + 1 mm D95%, and GTV + 2 mm D95%

B: The BED10s of the GTV – 2 mm DV-0.01 cc in dose prescriptions to the GTV D98%, GTV DV-0.01 cc, GTV + 1 mm D95%, and GTV + 2 mm D95%

GTV: gross tumor volume; BED: biologically effective dose; GTV + X mm: GTV evenly expanded by X mm; DV-0.01 cc: a minimum dose to cover a target volume minus 0.01 cc; DX%: a minimum dose covering at least X% of a target volume; BED10: biologically effective dose based on the linear-quadratic model with an alpha/beta ratio of 10; GTV – 2 mm: GTV evenly reduced by 2 mm

A significant difference in the BED10s of GTV D50% was observed among the four dose prescriptions (FT: P < 0.001 ***, KCC = 1.000; SPHT: GTV_Pa vs GTV+2_P, P < 0.001 ***; GTV_P vs GTV+2_P, P = 0.038 *; GTV_Pa vs GTV+1_P, P = 0.038 *). The BED10s of GTV D50% significantly increased as the GTV increased in the GTV_P and GTV_Pa prescriptions; however, they significantly decreased in the GTV+1_P and GTV+2_P prescriptions. A significant difference in the BED10s of GTV - 2 mm DV-0.01 cc was observed among the four dose prescriptions (FT: P < 0.001 ***, KCC = 1.000; SPHT: GTV_Pa vs GTV+2_P, P < 0.001 ***; GTV_P vs GTV+2_P, P = 0.038 *; GTV_Pa vs GTV+1_P, P = 0.038 *). The BED10 of the GTV - 2 mm DV-0.01 cc significantly decreased as the GTV increased and was highest in the GTV_Pa prescription. The BED10s of GTV D50% were inversely correlated significantly with those of the GTV - 2 mm DV-0.01 cc in the GTV_P and GTV_Pa prescriptions. 

Examples of the representative isodose distributions of the four different dose prescriptions based on BED10 in 5 fractions to the 7.50 cc GTV are shown in Figure 4.

Figure 4. Comparison of dose distributions of four specific dose prescriptions based on BED to 7.50 cc GTV.

Figure 4

The images show the target definitions (A); the representative isodose distributions for dose prescriptions to the GTV D98% (B); GTV DV-0.01 cc (C); GTV + 1 mm D95% (D); GTV + 2 mm D95% (E); and the representative absolute doses with the corresponding BED10s (F)

The dose prescriptions followed the 5-fraction examples shown in Table 4

The difference between the GTV D98% (B) and the GTV DV-0.01 cc (C) is small compared to the noticeable difference between the GTV D98% (B), GTV + 1 mm D95% (D), and GTV + 2 mm D95% (E)

BED: biologically effective dose; GTV: gross tumor volume; GTV + X mm: GTV evenly expanded by X mm; GTV - 2 mm: GTV evenly reduced by 2 mm; DX%: a minimum dose covering at least X% of a target volume; DV-0.01 cc: a minimum dose covering a target volume minus 0.01 cc; Abs.: absolute; BED10: biologically effective dose based on the linear-quadratic model with an alpha/beta ratio of 10

Compared to the GTV+1_P and GTV+2_P prescriptions, the difference in the dose distribution between the GTV D98% and DV-0.01 cc (D99.87%) was subtle and could be recognized by the parallel comparison.

The correlations between the GTV and the prescribed isodose volume reduced by the GTV (the dose spillage volume outside the GTV) are shown in Figure 5A.

Figure 5. Correlations between GTV and the spillage volume of the prescribed dose outside the GTV, and GTV and the irradiated isodose volumes in four specific dose prescriptions based on BED.

Figure 5

The scatter plots show correlations between the GTV and the spillage volume of the prescribed dose outside the GTV (A); the GTV and the irradiated isodose volume (IIV), including the GTV, of 24 Gy in 5 fractions (B); the GTV and the IIV of 20 Gy in 3 fractions (C); and the GTV and the IIV of 12 Gy in 1 fraction (D). The specific dose prescriptions in 1, 3, and 5 fractions followed the examples shown in Table 4

A: The spillage volumes of the prescribed dose outside the GTV in dose prescription to the GTV D98% for the GTVs of 10.00, 12.50, and 15.00 cc are -0.02, -0.04, and -0.04 cc, respectively, and are displayed as zero for convenience

B,C: 24 Gy volume in 5 fractions and 20 Gy volume in 3 fractions of <20 cc (dashed lines) are associated with <10% risk of brain radionecrosis

D: 12 Gy volumes in 1 fraction of ≥5 cc and ≥10 cc (dashed lines) are associated with the risks of symptomatic brain radionecrosis of approximately 10% and 15%, respectively

GTV: gross tumor volume; BED: biologically effective dose; vol.: volume; fr: fraction(s); GTV + X mm: GTV evenly expanded by X mm; DV-0.01 cc: a minimum dose to cover a target volume minus 0.01 cc; DX%: a minimum dose covering at least X% of a target volume; BED10: biologically effective dose based on the linear-quadratic model with an alpha/beta ratio of 10

The prescribed dose spillage volumes outside the GTV were <0 cc for GTV ≥10.00 cc in the GTV_P prescription, while they increased with increasing GTV in the GTV_Pa prescription.

The correlations between the GTVs and the IIVs are shown in Figure 5B-5D. A significant difference in the 24 Gy volumes in 5 fractions was observed among the four dose prescriptions (FT: P = 0.001 **, KCC = 0.755; SPHT: GTV_Pa vs GTV+1_P, P = 0.002 **). A significant difference in the 20 Gy volumes in 3 fractions was observed among the four dose prescriptions (FT: P = 0.002 **, KCC = 0.723; SPHT: GTV_Pa vs GTV+1_P, P = 0.002 **). A significant difference in the 12 Gy volumes in 1 fraction was observed among the four dose prescriptions (FT: P = 0.003 **, KCC = 0.657; SPHT: GTV_Pa vs GTV+1_P, P = 0.006 **). The linear approximation showed significant correlations between the GTV and the IIVs. In the three GTV_Pa prescriptions using specific doses with the BED10 of 80 Gy, the linear approximation equations with the coefficients of determination and the GTVs exceeding each threshold volume for predicting the risk of brain radionecrosis are shown in Table 12.

Table 12. Correlations between GTV and irradiated isodose volumes in BED-based dose prescriptions to GTV DV-0.01 cc and the GTV thresholds at which the risk of brain radionecrosis increases.

*Diameters of spheres corresponding to GTVs above the IIV thresholds relevant to the risk of brain radionecrosis

The suitability of linear approximations for the correlations between the GTVs and the IIVs is shown, along with the coefficients of determination (R2)

GTV: gross tumor volume; BED: biologically effective dose; DV-0.01 cc: a minimum dose to cover a target volume minus 0.01 cc; IIV: irradiated isodose volume (including a target volume); fr: fraction(s); vol.: volume

GTV DV-0.01 cc IIV Threshold vol. (Risk) Linear approximation formula R2 GTV exceeding the threshold volume (Diameter*)
43.0 Gy (5 fr) 24 Gy vol. <20.0 cc (<10%) y = 2.2288x + 1.1373 0.9996 8.47 cc (25.3 mm)
36.3 Gy (3 fr) 20 Gy vol. <20.0 cc (<10%) y = 2.2659x + 1.1634 0.9996 8.32 cc (25.1 mm)
23.8 Gy (1 fr) 12 Gy vol. 5.0 cc (10%) y = 2.5339x + 1.4323 0.9996 1.41 cc (13.9 mm)
10.0 cc (15%) 3.39 cc (18.6 mm)

The GTV thresholds were similar for both 5 and 3 fractions (8.47 vs 8.32 cc).

Discussion

This study revealed significant flaws in dose prescription to a fixed % coverage of either GTV or margin-added PTV: the significant dose decreases at the GTV boundary with increasing GTV. Although the dose conformity index itself tends to be an excellent value in the GTV D≤98% prescription, the fixed % coverage with the prescribed dose can be one of the causes of poor local control for larger lesions. Furthermore, dose prescription to GTV + 2 mm D95% led to a dose reduction of 1-2 mm outside the GTV boundary as the GTV increased. Moderate (not too steep or gradual) dose attenuation outside the GTV boundary and the appropriate dose are also important to cover the potential microscopic brain invasion and inherent uncertainties in irradiation accuracy [3,5,15,24]. In contrast to a fixed % coverage of GTV with or without a margin, dose prescription to GTV DV-0.01 cc ensures the dose consistency of the GTV boundary, regardless of GTV, and allows for the rational dose increase at 2 mm outside the GTV boundary with increasing GTV.

Considering that most of the tissue surrounding the GTV consists of the brain, optimization with the steepest dose gradient just outside the GTV (T2-mass) boundary is optimal for minimizing the dose to the brain. VMARS optimization targeting the GTV boundary without a margin resulted in the most inhomogeneous GTV dose and the steepest dose increase inside the GTV boundary. The highest dose was observed at 2 mm inside the GTV boundary and decreased with increasing GTV. A steep and concentrically laminated dose increase inside the GTV boundary can lead to early and sufficient tumor shrinkage during and after multi-fraction SRS [15]. Given the low correlations between GTV D50% and GTV - 2 mm DV-0.01 cc, GTV - 2 mm DV-0.01 cc may be associated with a more favorable tumor response than GTV D50%, D2%, or D0.01 cc. The decrease in the GTV - 2 mm DV-0.01 cc with increasing GTV seems reasonable, considering the possibility of increased surrounding brain dose due to the ease of tumor shrinkage during ≥5-fraction SRS for larger tumors.

Although VMARS optimization without any internal dose constraint to a GTV generally leads to the steepest dose gradient outside the GTV boundary [3], the dose 2 mm outside the GTV boundary was almost sufficient (BED10 of DV-0.035 cc >50 Gy), except for a small lesion of 1 cc, and was reasonably higher for larger tumors. A dose of 30-35 Gy is commonly used in 5 fractions [6]. However, the BED10 of GTV DV-0.01 cc fell significantly below the 80 Gy in dose prescription, with BED10 values of 50.0 Gy and 60.0 Gy to GTV + 1 and 2 mm D95%, respectively, despite the steep dose increase inside the prescribed isodose. Therefore, dose assignment to GTV DV-0.01 cc is likely pertinent to ensure a BED10 of ≥80 Gy at the GTV boundary, regardless of GTV. GTV DV-0.01 cc with variable coverage (D>95%) for >0.20 cc GTV and fixed D95% for ≤0.20 cc GTV is recommended as the optimal dose prescription method.

For dose prescriptions with a BED10 of ≥80 Gy to GTV DV-0.01 cc, the target GTV should be less than those for normal SRS in 3-5 fractions to ensure sufficient safety. The GTV application limits of <1.41 and <8.47 cc for 1- and 5-fraction SRS, respectively, can serve as a guideline. Importantly, ≥6 fractions should be considered for GTVs ≥10 cc or >3 cm in diameter. However, the GTV limit for 3 fractions was close to that for 5 fractions, which is probably because 20 and 24 Gy correspond to 55.1% and 55.8% of 36.3 and 43.0 Gy, respectively, which is only a 0.7% difference. The safe application limits of 3-fraction SRS require further investigation.

A standard metric is also necessary to objectively compare the dose distributions among different devices and irradiation methods. It is difficult to describe a typical inhomogeneous dose distribution of SRS in a few metrics [15]. Although D98%, D50%, and D2% of a PTV (>2 cc) may be appropriate as the reporting metrics for a homogeneous TV dose [7], their clinical significance in SRS for BM is highly questionable, given the high correlations with the GTV. The near-minimum dose of GTV (DV-0.01 cc), dose 2 mm outside the GTV, dose 2 mm inside the GTV, and GTV D0.01 cc may be more relevant to the local tumor control in that order. The dose 1 mm outside the GTV boundary can be approximately estimated as the midpoint between the GTV DV-0.01 cc and the dose 2 mm outside the GTV. In SRS performed using various modalities and methods, if the treatment details are reported using these standard metrics, it could be the first step toward building a consensus on the basic design of optimal dose distribution.

This planning study has several inherent limitations. The validity of BED10 80 Gy, optimal dose fractionation, and suitability of BED10 for the clinical SRS of BM remain highly controversial. Further investigation is needed to determine whether the long-term local tumor control can be improved, without increasing adverse radiation effects, by shifting the dose prescription to the GTV boundary from the D≥98% to DV-0.01 cc, including the need for further dose escalation [17,18,30]. The validity of DV-0.01 cc in dose reporting 1-2 mm outside the GTV boundary remains a subject for further investigation, including the suitability of DV-0.035 cc or other metrics. Although this study was limited to SRS for a single lesion, it is important to note that when multiple lesions are irradiated simultaneously, the individual dose gradient outside and inside the GTV boundary can vary significantly due to dose interference. As the number of lesions increases, aligning the DV-0.01 cc for each GTV may not be easy. In any case, the multiple GTV curves in the DVHs should not be aligned but should differ depending on the GTV to align the GTV DV-0.01 cc as much as possible.

Conclusions

Dose prescription to a fixed % coverage of a GTV with or without a margin leads to a substantially varied near-minimum dose at the GTV boundary, which significantly decreases with increasing GTV. Alternatively, the GTV DV-0.01 cc with a variable coverage (D>95%) for >0.20 cc GTV and fixed D95% for ≤0.20 cc GTV is recommended as the basis for dose prescription and/or objective evaluation, along with supplemental evaluation of the marginal dose of the GTV plus a margin (e.g. GTV + 2 mm) to demonstrate the appropriateness of dose attenuation outside the GTV boundary. The dose 2 mm inside the GTV boundary may be more relevant to tumor response than the GTV D50% or near-maximum dose. To ensure safety, it is recommended that the prescription of a BED10 of ≥80 Gy to the GTV DV-0.01 cc be limited to smaller GTVs than those often used in general applications.

Acknowledgments

The authors are grateful to Dr. Yukihiko Oshima and Ms. Noriko Tazawa (Department of Radiology, Aichi Medical University) for their valuable support and contributions.

The authors have declared that no competing interests exist.

Funding Statement

This study was supported by the Japan Society for the Promotion of Science (JSPS) KAKENHI Grant-in-Aid for Scientific Research with the grant number 21K07561.

Author Contributions

Concept and design:  Kazuhiro Ohtakara

Acquisition, analysis, or interpretation of data:  Kazuhiro Ohtakara, Kojiro Suzuki

Drafting of the manuscript:  Kazuhiro Ohtakara

Critical review of the manuscript for important intellectual content:  Kazuhiro Ohtakara, Kojiro Suzuki

Human Ethics

Consent was obtained or waived by all participants in this study. Clinical Research Review Board of Kainan Hospital Aichi Prefectural Welfare Federation of Agricultural Cooperatives issued approval 20220727-1

Animal Ethics

Animal subjects: All authors have confirmed that this study did not involve animal subjects or tissue.

References

  • 1.Inhomogeneous tumor dose distribution provides better local control than homogeneous distribution in stereotactic radiotherapy for brain metastases. Lucia F, Key S, Dissaux G, et al. Radiother Oncol. 2019;130:132–138. doi: 10.1016/j.radonc.2018.06.039. [DOI] [PubMed] [Google Scholar]
  • 2.Consideration of optimal isodose surface selection for target coverage in micro-multileaf collimator-based stereotactic radiotherapy for large cystic brain metastases: comparison of 90%, 80% and 70% isodose surface-based planning. Ohtakara K, Hayashi S, Tanaka H, Hoshi H. Br J Radiol. 2012;85:0–6. doi: 10.1259/bjr/21015703. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.An extremely inhomogeneous gross tumor dose is suitable for volumetric modulated arc-based radiosurgery with a 5-mm leaf-width multileaf collimator for single brain metastasis. Ohtakara K, Suzuki K. Cureus. 2023;15:0. doi: 10.7759/cureus.35467. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.Characterisation of dose distribution in linear accelerator-based intracranial stereotactic radiosurgery with the dynamic conformal arc technique: consideration of the optimal method for dose prescription and evaluation. Ohtakara K, Hayashi S, Hoshi H. Br J Radiol. 2012;85:69–76. doi: 10.1259/bjr/20905396. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Modified dynamic conformal arcs with forward planning for radiosurgery of small brain metastasis: each double arc and different to-and-fro leaf margins to optimize dose gradient inside and outside the gross tumor boundary. Ohtakara K, Suzuki K. Cureus. 2023;15:0. doi: 10.7759/cureus.34831. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Tumor control probability of radiosurgery and fractionated stereotactic radiosurgery for brain metastases. Redmond KJ, Gui C, Benedict S, et al. Int J Radiat Oncol Biol Phys. 2021;110:53–67. doi: 10.1016/j.ijrobp.2020.10.034. [DOI] [PubMed] [Google Scholar]
  • 7.ICRU Report 91: prescribing, recording, and reporting of stereotactic treatments with small photon beams. Seuntjens J, Lartigau EF, Cora S, et al. https://journals.sagepub.com/toc/crua/14/2 J ICRU. 2014;14:1–160. [Google Scholar]
  • 8.Reduced-dose WBRT combined with SRS for 1-4 brain metastases aiming at minimizing neurocognitive function deterioration without compromising brain tumor control. Nakano T, Aoyama H, Onodera S, et al. Clin Transl Radiat Oncol. 2022;37:116–129. doi: 10.1016/j.ctro.2022.09.005. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9.Daily irradiation versus irradiation at two- to three-day intervals in stereotactic radiotherapy for patients with 1-5 brain metastases: study protocol for a multicenter open-label randomized phase II trial. Tomita N, Ishiyama H, Makita C, et al. BMC Cancer. 2022;22:1259. doi: 10.1186/s12885-022-10371-3. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10.Significant correlation between gross tumor volume (GTV) D98% and local control in multifraction stereotactic radiotherapy (MF-SRT) for unresected brain metastases. Dupic G, Brun L, Molnar I, et al. Radiother Oncol. 2021;154:260–268. doi: 10.1016/j.radonc.2020.11.021. [DOI] [PubMed] [Google Scholar]
  • 11.Clinical comparison of positional accuracy and stability between dedicated versus conventional masks for immobilization in cranial stereotactic radiotherapy using 6-degree-of-freedom image guidance system-integrated platform. Ohtakara K, Hayashi S, Tanaka H, Hoshi H, Kitahara M, Matsuyama K, Okada H. Radiother Oncol. 2012;102:198–205. doi: 10.1016/j.radonc.2011.10.012. [DOI] [PubMed] [Google Scholar]
  • 12.Local tumor progression treated with Gamma Knife radiosurgery: differences between patients with 2-4 versus 5-10 brain metastases based on an update of a multi-institutional prospective observational study (JLGK0901) Serizawa T, Yamamoto M, Higuchi Y, et al. J Neurosurg. 2019;132:1480–1489. doi: 10.3171/2019.1.JNS183085. [DOI] [PubMed] [Google Scholar]
  • 13.Internal dose escalation is associated with increased local control for non-small cell lung cancer (NSCLC) brain metastases treated with stereotactic radiosurgery (SRS) Abraham C, Garsa A, Badiyan SN, et al. Adv Radiat Oncol. 2018;3:146–153. doi: 10.1016/j.adro.2017.11.003. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14.Clinical outcomes of biological effective dose-based fractionated stereotactic radiation therapy for metastatic brain tumors from non-small cell lung cancer. Matsuyama T, Kogo K, Oya N. Int J Radiat Oncol Biol Phys. 2013;85:984–990. doi: 10.1016/j.ijrobp.2012.09.008. [DOI] [PubMed] [Google Scholar]
  • 15.Correlation of brain metastasis shrinking and deviation during 10-fraction stereotactic radiosurgery with late sequela: suggesting dose ramification between tumor eradication and symptomatic radionecrosis. Ohtakara K, Tanahashi K, Kamomae T, Miyata K, Suzuki K. Cureus. 2023;15:0. doi: 10.7759/cureus.33411. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16.Ten-fraction stereotactic radiosurgery with different gross tumor doses and inhomogeneities for brain metastasis of >10 cc: treatment responses suggesting suitable biological effective dose formula for single and 10 fractions. Ohtakara K, Nakabayashi K, Suzuki K. Cureus. 2023;15:0. doi: 10.7759/cureus.34636. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17.Fifteen-fraction radiosurgery followed by reduced-dose whole-brain irradiation with a total biologically effective dose of >90-100 Gy for a locally invasive brain metastasis from lung adenocarcinoma with a high dissemination potential. Ohtakara K, Ohka F, Tanahashi K, Yamada T, Suzuki K. Cureus. 2023;15:0. doi: 10.7759/cureus.49596. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Local control failure after five-fraction stereotactic radiosurgery alone for symptomatic brain metastasis from squamous cell lung carcinoma despite 43 Gy to gross tumor margin with internal steep dose increase and tumor shrinkage during irradiation. Ohtakara K, Tanahashi K, Kamomae T, Ito E, Suzuki K. Cureus. 2023;15:0. doi: 10.7759/cureus.38645. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19.5-fraction re-radiosurgery for progression following 8-fraction radiosurgery of brain metastases from lung adenocarcinoma: importance of gross tumor coverage with biologically effective dose ≥80 Gy and internal dose increase. Ohtakara K, Tanahashi K, Kamomae T, Suzuki K. Cureus. 2023;15:0. doi: 10.7759/cureus.42299. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20.T1/T2 matching to differentiate tumor growth from radiation effects after stereotactic radiosurgery. Kano H, Kondziolka D, Lobato-Polo J, Zorro O, Flickinger JC, Lunsford LD. Neurosurgery. 2010;66:486–491. doi: 10.1227/01.NEU.0000360391.35749.A5. [DOI] [PubMed] [Google Scholar]
  • 21."Comet tail sign": a pitfall of post-gadolinium magnetic resonance imaging findings for metastatic brain tumors. Mitsuya K, Nakasu Y, Narita Y, et al. J Neurooncol. 2016;127:589–595. doi: 10.1007/s11060-016-2069-1. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22.Volumetric-modulated arc-based re-radiosurgery with simultaneous reduced-dose whole-brain irradiation for local failures following prior radiosurgery of brain oligometastases from small cell lung cancer. Ohtakara K, Arakawa S, Nakao M, Muramatsu H, Suzuki K. Cureus. 2023;15:0. doi: 10.7759/cureus.44492. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23.A pathology-based substrate for target definition in radiosurgery of brain metastases. Baumert BG, Rutten I, Dehing-Oberije C, et al. Int J Radiat Oncol Biol Phys. 2006;66:187–194. doi: 10.1016/j.ijrobp.2006.03.050. [DOI] [PubMed] [Google Scholar]
  • 24.Five-fraction stereotactic radiosurgery with non-contrast-enhanced MRI-based target definition and moderate dose spillage margin for limited brain metastases with impaired renal function. Ohtakara K, Suzuki K. Cureus. 2023;15:0. doi: 10.7759/cureus.37384. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 25.Significance of target location relative to the depth from the brain surface and high-dose irradiated volume in the development of brain radionecrosis after micromultileaf collimator-based stereotactic radiosurgery for brain metastases. Ohtakara K, Hayashi S, Nakayama N, Ohe N, Yano H, Iwama T, Hoshi H. J Neurooncol. 2012;108:201–209. doi: 10.1007/s11060-012-0834-3. [DOI] [PubMed] [Google Scholar]
  • 26.Single- and multifraction stereotactic radiosurgery dose/volume tolerances of the brain. Milano MT, Grimm J, Niemierko A, et al. Int J Radiat Oncol Biol Phys. 2021;110:68–86. doi: 10.1016/j.ijrobp.2020.08.013. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 27.Dose-response effect and dose-toxicity in stereotactic radiotherapy for brain metastases: a review. Loo M, Clavier JB, Attal Khalifa J, Moyal E, Khalifa J. Cancers (Basel) 2021;13:6086. doi: 10.3390/cancers13236086. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 28.The relation between various conformity indices and the influence of the target coverage difference in prescription isodose surface on these values in intracranial stereotactic radiosurgery. Ohtakara K, Hayashi S, Hoshi H. Br J Radiol. 2012;85:0–8. doi: 10.1259/bjr/36606138. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 29.Dose gradient analyses in LINAC-based intracranial stereotactic radiosurgery using Paddick's gradient index: consideration of the optimal method for plan evaluation. Ohtakara K, Hayashi S, Hoshi H. J Radiat Res. 2011;52:592–599. doi: 10.1269/jrr.11005. [DOI] [PubMed] [Google Scholar]
  • 30.Gradual and remarkable tumor shrinkage following seven-fraction stereotactic radiosurgery alone with a marginal dose of 48.3 Gy for large lobar possibly intra-sulcal brain metastasis from renal cell carcinoma. Ohtakara K, Aoki S, Tajima M, Ohno T, Suzuki K. Cureus. 2023;15:0. doi: 10.7759/cureus.36346. [DOI] [PMC free article] [PubMed] [Google Scholar]

Articles from Cureus are provided here courtesy of Cureus Inc.

RESOURCES