Abstract
A miniaturized photoacoustic spectroscopy-based gas sensor is proposed for the purpose of detecting sub-ppm-level carbonyl sulfide (OCS) using a tunable mid-infrared interband cascade laser (ICL) and a Helmholtz photoacoustic cell. The tuning characteristics of the tunable ICL with a center wavelength of 4823.3 nm were investigated to achieve the optimal driving parameters. A Helmholtz photoacoustic cell with a volume of ∼2.45 mL was designed and optimized to miniaturize the measurement system. By optimizing the modulation parameters and signal processing, the system was verified to have a good linear response to OCS concentration. With a lock-in amplifier integration time of 10 s, the 1σ noise standard deviation in differential mode was 0.84 mV and a minimum detection limit (MDL) of 409.2 ppbV was achieved at atmospheric pressure and room temperature.
Keywords: Photoacoustic spectroscopy, OCS detection, Helmholtz photoacoustic cell
1. Introduction
Carbonyl sulfide (OCS) is a colorless, polar, room-temperature stable inorganic compound with rotten egg smelling at room temperature and pressure [1]. It is frequently found in medical diagnostics, astronomical research, industrial gas emission monitoring, environmental pollutant monitoring, atmospheric chemistry, etc. [2], [3], [4], [5]. Firstly, OCS can indicate the level of human health. The range of exhaled OCS in a healthy body is from 3 pptV to 30 ppbV, while individuals with conditions such as cystic fibrosis or immune rejection reactions can exhale OCS concentration reaching ppmV-level [2], [6]. Moreover, OCS gas is an environmental air pollutant. For example, when the emission of OCS near factories exceeds 23 ppmV, it can lead to toxicity in mammals. So, quantitative analysis of OCS has significant importance for environmental pollutant monitoring [7]. Finally, the presence of OCS has been detected in other astrophysical bodies currently. For instance, the average concentration of OCS observed on comets is approximately 1260 ppmV [8]. The quantitative analysis of OCS trace gas would lead to significant advancements in astrophysical research [8], [9]. Hence, there is a strongly need in developing stable, sensitive OCS sensor for many applications.
Common OCS detection methods can be divided into two main categories: non-optical and optical method. Non-optical method include gas chromatography [10], [11], fluorescence detection [12], [13], etc. For example, Joaquin et al. built an OCS detection system based on flame ionization and chromatography for the detection of OCS gas [11]; Kodai et al. proposed a method based on fluorescence detection and construct a miniature gas analysis system, which has good sensing performance for OCS in ambient air and industrial exhaust gas. [12]. These researches show that the non-optical methods can achieve high-sensitive detection. However, they have limitations for field application: gas chromatography suffers from bulky size and high cost; Fluorescence detection has the disadvantage of a short lifetime due to the loss of fluorescein, and inability to perform real-time detection [14], [15]. In contrast, optical methods such as tunable diode laser absorption spectroscopy (TDLAS) [16], cavity ringdown spectroscopy (CRDS) [17], [18], [19], and photoacoustic spectroscopy (PAS) [20] have the advantages of real-time response, good gas selectivity, and high sensitivity [21], [22], [23], [24]. Roller et al. built a TDLAS-based OCS detection system using a thermoelectrically cooled mid-infrared quantum cascade laser (QCL) with a 36-m optical multipass cell [16]; Wojtas et al. reported a CRDS-based trace gas sensor in which the optical cavity contained two concave mirrors with approximate reflectivity, achieving OCS gas detection [19]. However, when using these direct absorption spectroscopy techniques, the change in laser intensity due to gas absorption is small relative to the incident laser power, making it difficult to observe a small signal from a large background. Meanwhile, high power lasers (such as QCL, CO2 laser, CO laser and so on) have to be selected to ensure adequate sensitivity, leading to high power consumption and complex cooling system. These factors hinder the application of direct spectroscopy in sensitive, miniaturized OCS detection systems. Compared to TDLAS and CRDS, PAS is a zero background detection technique [25], [26], [27], which makes it even more attractive for trace gas detection in sensitive, miniaturized OCS detection systems. So far, few PAS-based OCS sensing systems have been reported. Mohebbifar et al. measured the absorption of OCS at 9600 nm using a 0.35 W CO2 laser and a cylindrical photoacoustic cell (PAC) with a volume of 318.52 mL [20]. However, the cylindrical PAC and light sources used by Mohebbifar et al. are too large to be used in a variety of scenarios, hindering the popularity of the system. Therefore, the realization of a miniaturized, highly sensitive OCS sensor based on PAS has important application value [28].
In this paper, a highly sensitive gas sensor for the detection of OCS based on PAS is reported. A wavelength-tunable interband cascade laser (ICL) with the output power of ∼5 mW was used in combination with a differential Helmholtz PAC with the volume of ∼2.45 mL to enhance the detection capability of the system, and thus obtain the optimal detection limit at the sub-ppm level. This study provides support to the field of OCS gas detection and provides ideas for miniaturization of PAS trace gas sensors.
2. Basic principles
PAS technique is based on the Lambert-Beer law and can be used in gas sensors for qualitative and quantitative analysis of trace gases [29], [30], [31], [32]. According to the principle of photoacoustic (PA) signal detection, the acoustic wave generated when gas molecules absorbing the light is related to the gas concentration [33]. So, the concentration of the analyzed gas can be obtained by monitoring the photoacoustic signal [34], [35], [36].
PAC is one core component for a PAS system which strongly impacts the sensitivity and minimum detectable limit. A Helmholtz resonator will be used as the PAC due to its small size and low resonant frequency [37], [38]. For improving the detection performance, the differential structure is introduced as shown in Fig. 1. The cell has a symmetrical structure, with the inlet and outlet gas ports located in the middle of two capillaries tubes, which connect the excitation and detection cavities. When laser light is injected into the excitation cavity and absorbed by the sample gas will causes the gas to expand. Then, the capillaries move like pistons, compressing the gas in one cavity and expanding the gas in the other periodically. Acoustic signals with opposite phases will appear in the two chambers [39], [40]. The PA signal of the gas detection system is [41]
| (1) |
Where indicates the adiabatic index, is the effective optical path length, is the quality factor of the PAC, is the sensitivity of the microphone, is the effective optical power, is the absorption coefficient of a specific gas concentration at a selected wavelength, is the angular modulation frequency and is the volume of the Helmholtz resonator.
Fig. 1.
The structure of the Helmholtz resonator.
For the purpose of miniaturising the OCS detection system based on PAS, a miniaturized Helmholtz PAC was designed as shown in the inset of the Fig. 2(a). The Helmholtz PAC in this work was simulated by the finite element simulation software COMSOL Multiphysics, as shown in Fig. 2(a). According to the result of the frequency scanning, the resonance frequency of the Helmholtz PAC is 2318 Hz. And the sound pressure distribution of the Helmholtz PAC at the first-order longitudinal resonance frequency is also plotted in Fig. 2(a). The excitation and detection cavities have a length of 30 mm and a diameter of 6 mm, and the two capillaries have a length of 30 mm and a diameter of 4 mm. Hence, the volume of this Helmholtz PAC is approximately 2.45 mL. The Helmholtz PAC was fabricated by 3D printing technology, the used material was photosensitive resin, as shown in Fig. 2(b).
Fig. 2.
(a) The results of frequency sweep of the Helmholtz resonator. Inset: Helmholtz resonator sound pressure simulation result; (b) 3D printed Helmholtz resonator.
3. Experiment details
3.1. Selection of detection wavelength
The OCS molecule indicates some strong absorption band in the infrared region (IR) according to the HITRAN database. As shown in Fig. 3, three major OCS absorption bands are R1, R2 and R3. The absorption coefficient of R2 band is two orders of magnitude stronger than elsewhere, and the corresponding lasers have developed maturely and holds excellent performance. Therefore, a better spectral region selection would be R2, as depicted in more detail in the inset of Fig. 3. According to spectral analysis, the common gas molecules in the air (such as CO, CO2, H2O) show no absorption lines in the spectral range R2, and the absorption peaks of CO can also be easily avoided by adjusting the output wavelength of the laser. Finally, the 4826.5 nm is chosen as the target absorption wavelength for the detection of OCS in this work.
Fig. 3.
Absorption band in the infrared region of OCS. Inset: more detailed R2 absorption band of OCS and some common gas.
3.2. Laser characteristics
Current optical-based OCS detection systems use CO laser, CO2 laser, and QCL as the light source, which are not conducive to portability and miniaturization of the system because of the high power consumption and complex cooling system of the chosen light source. Therefore, considering the power consumption, complexity and sensitivity of the system, a wavelength-tunable interband cascade laser (ICL) from Nanoplus with a TO66 package including a collimating lens is chosen as the light source. The wavelength can be tuned by changing the injection current and controlling the temperature of the thermoelectric controller (TEC) which is already encapsulated inside the laser. In order to determine whether the tuning range of the laser covers the chosen wavelength, the tuning characteristic of the laser is tested. The wavelength of the laser at different temperatures with injection current tuning is shown in Fig. 4(a), the spectral characteristic of the laser is shown in the inset of Fig. 4(a). It is shown that the tunable wavelength range is larger than 5 nm at a certain working temperature, and the spectral bandwidth below 3 MHz. In addition, we need a sufficiently high laser output power to ensure that a high system sensitivity is obtained. When the temperature of the TEC is set to 20 ℃, the relationship between the output power of the laser and the injection current is shown in Fig. 4(b). The threshold current of the laser is around 40 mA, and the maximum optical power is approximately 10 mW at the injection current of 100 mA.
Fig. 4.
(a) Current tuning of ICL at different operating temperatures. Inset: ICL output spectrum at 4823.3 nm; (b) the relationship between ICL output power and current at 20 °C.
3.3. Sensor configuration
The schematic of the OCS sensing system based on PAS is shown in Fig. 5, which mainly includes a light source module, a PAC module, a gas distribution module, and a data acquisition and processing module. For the light source module, the ICL is controlled by a high compliance current source (ILX Lightwave, LDX-3232) and a temperature controller (ILX Lightwave, LDT-5525B) to generate the excitation light signal. The key part of the PAC module is a Helmholtz resonator with four CaF2 windows. Rubber ring seals are used between the windows and the resonance chambers to avoid gas leak. The modulated light passes through the CaF2 window into the Helmholtz resonator and absorbed by the OCS gas molecules. The different concentrations of OCS gas molecules can be obtained by the gas distribution module. Then, the PA signal generated in the detection cavity and the reference signal generated in the excitation cavity are captured by two microphones (BSWA, MPA416) respectively. Finally, the signals are fed into the data acquisition and processing module which mainly includes a lock-in amplifier (Stanford Research Systems, SR830) and a data acquisition card to demodulate and obtain the relationship between the PA signal and the gas concentration.
Fig. 5.
The structure of the OCS sensing system. MFC: Mass Flow Controller. DAQ: Data Acquisition.
4. Experimental results and discussion
4.1. Selection of experimental parameters
The static operating parameter must be determined before the modulation parameters are determined. Several factors need to be considered when selecting proper static operating parameters. Firstly, the excitation light power must be sufficiently large to ensure a strong photoacoustic signal. Secondly, the static operating point must be carefully chosen to optimize performance, considering that the modulated signal can affect both the laser power and output wavelength.
In order to obtain the optimum static driving current, OCS standard gas with the concentration of 50 ppmV is passed through the PAC. The amplitude of the PA signals is measured at different static driving current and same square wave modulation signal (Peak-to-peak modulation amplitude: 2 V, modulation coefficient: 20 mA/V) with the operating temperature keeping at 23 °C, the result is shown in Fig. 6(a). The measurements were carried out at atmospheric pressure and room temperature. It can be seen that the PA signal is not monotonically growing with the driving current increasing. The reason for this phenomenon is that the absorption strength of the OCS molecule around 4826.5 nm shows fluctuating trend (shown in Fig. 3). According to the result of Fig. 6(a), low driving current leads to weak PA signal, and high driving current may lead to exceed the safe working limit (<100 mA) when adds the modulation signal. Hence, 79 mA is selected as the static driving current of the ICL in this work. In addition, the amplitude of the modulated signal is also an important parameter for PA signal. Fig. 6(b) demonstrates the measured PA signal as a linear fit function of the modulation amplitude with error line. It can be seen that the PA signal amplitude increases with the modulation amplitude increasing with R >0.99 which means that the main factor affecting the PA signal is the output power of the ICL rather than the wavelength at the selected static operating point. It is evident that the error in the signal amplitude becomes larger at a peak-to-peak modulation amplitude of 2.2 V. So, a peak-to-peak modulation amplitude of 2 V is chosen for the measurement, with a modulation coefficient of 20 mA/V. Under the condition of a static driving current of 79 mA and a peak-to-peak modulation amplitude of 2 V, the system is primarily intensity-modulated.
Fig. 6.
(a) PA signal amplitude at different static driving current; (b) Relationship between PA signal amplitude and square wave modulated signal amplitude.
Additionally, when the frequency of the PA signal is matching with the resonant frequency of the PAC, the optimal system performance will be achieved. According to the previous studies, it was found that the manufacturing error of the Helmholtz PAC affects the resonance frequency for the detection channel (A mode), the excitation channel (B mode) and the differential mode (A-B mode). Therefore, the experiment was conducted to measure the frequency response in each of the three modes at atmospheric pressure and room temperature. With the PAC being filled 50 ppmV OCS/N2 gas mixture, the frequency response of the PAC was measured by changing the modulation frequency, as shown in Fig. 7(a). The frequency response of the Helmholtz PAC for the A mode, the B mode and the A-B mode are plotted in the Fig. 7(a). It shows that the resonant frequency of the three modes is consistent at 1931 Hz, with a Q-factor (Q=f/Δf) of ∼6.44 at the differential mode. However, a significant difference exists between the measured resonant frequency and the simulated resonant frequency. The most possible reason is the manufacturing error after analyzing other reasons (such as temperature, pressure, background gas…). According to the experience of the manufacturer, the maximum machining error is less than 0.5 mm and the machining error increases as the dimensions decrease. In addition, according to the theory analysis, it shows that the radius of the capillary has a significant effect on the resonant frequency. Hence, considering the manufacturing error, the resonant frequency of the designed Helmholtz resonator is simulated in the capillary radius range of 1.5 mm∼2.5 mm, as shown in Fig. 7(b). Theoretical analysis indicates that the radius of the capillary significantly affects the resonant frequency.
Fig. 7.
(a) Frequency response curves of the system in channel A, channel B and A-B mode; (b) The resonant frequency of the Helmholtz PAC with capillary radius in the range of 1.5 mm∼2.5 mm.
4.2. Sensor system tests
As analyzed in Fig. 3, CO gas is a important interfering substance for OCS detection around 4826.5 nm. Therefore, the effect of CO on the OCS detection has to be excluded. Fig. 8 shows the PA signal of a 25 ppmV OCS/N2 gas mixture with and without a 9600 ppmV CO/N2 gas mixture at atmospheric pressure and room temperature. Obviously, the high concentration of CO has no effect on the OCS detection.
Fig. 8.
The detection signal of the OCS/N2 gas mixture for both incoming and outgoing CO/N2 mixture at 9600 ppmV.
The linearity of the designed photoacoustic OCS sensor is evaluated by measuring its response to the different OCS concentrations from 0 ppmV to 50 ppmV at atmospheric pressure and room temperature. The gas distribution system is used to produce different concentrations of the OCS/N2 mixture. An 8-minutes interval between two adjacent concentration steps was set to replace the gas mixture in the Helmholtz PAC. To reduce the effect of gas flow rate noise on the PA signal, the total flow rate was set to 200 sccm. Fig. 9 depicts the PA signal amplitude as a function of OCS concentration. Fig. 9(a), (b), (c) show the 3-minutes average values of PA responses of the designed OCS sensor corresponding to the A mode, B mode and A-B mode, respectively. The linear fitting procedure with a R square value of >0.99 confirms the linearity of this sensor response to the OCS concentration at the three mode. The results indicate that the A-B mode exhibits the highest sensitivity, leading to its selection for subsequent experiments. Fig. 9(d) is the results of repetitive measurements at A-B mode. In addition, the experimental results were used to calculate a 1σ noise standard deviation of 0.84 mV in A-B mode and a signal-to-noise ratio (SNR) of 12.22 for 5 ppmV OCS, which means the minimum detection limit (MDL) of 409.2 ppbV for the OCS in A-B mode.
Fig. 9.
(a) Linear response of channel A signal amplitude at different concentrations; (b) Linear response of channel B signal amplitude at different concentrations; (c) Linear response of the signal amplitude in A-B mode at different concentrations; (d) Signal amplitude at different concentrations in A-B mode.
5. Conclusion
In summary, this paper presents a miniaturized trace gas detection system based on PAS technology, combining a mid-infrared ICL and a Helmholtz PAC to achieve high sensitivity for the detection of OCS gas. The wavelength of 4826.5 nm was selected for OCS detection due to its high line strength and no signal interference from N2, CO2, H2O, and CO. The tuning range of the ICL was researched at different temperatures and currents to determine the optimal modulation parameters of the laser. In order to improve the detection SNR, a differential structure PAC was used. As a result, a detection limit of 409.2 ppbV was obtained at the differential mode. The developed PAS OCS sensor afford broad application prospects in respiratory detection, environmental monitoring, astronomical research and atmospheric chemistry, and provides ideas for miniaturization of trace gas detection systems.
CRediT authorship contribution statement
Zijian Gao: Writing – original draft, Validation, Methodology, Investigation, Data curation, Conceptualization. Lei Li: Writing – review & editing, Supervision, Resources, Funding acquisition, Data curation, Conceptualization. Mingyang Feng: Investigation. Yingying Qiao: Supervision, Resources, Project administration, Methodology, Funding acquisition. Minghui Liu: Validation. Shen Tian: Methodology. Chongxin Shan: Supervision, Resources, Funding acquisition, Conceptualization.
Declaration of Competing Interest
The authors declare that they have no known competing financial interests or personal relationships that could have appeared to influence the work reported in this paper.
Acknowledgments
We are grateful for financial support from the National Nature Science Foundation of China (62005247, 62271451), National Key Scientific Instrument and Equipment Development Project of China (62027816), Zhengzhou Collaborative Innovation Major Project (18XTZX12008), Henan Provincial Science and Technology Research Project (222102210163), Henan Provincial Postdoctoral Research Grant (NO. 202102011).
Biographies

Zijian Gao received his Bachelor of Engineering degree from Henan University, Kaifeng, China, in 2021. He is currently studying in the School of Physics at Zhengzhou University, majoring in electronic information. His current research interests are in photoacoustic sensor.

Lei Li received the B.S. degree in electronic engineering from Zhengzhou University, Zhengzhou, China, in 2004, and the Ph.D. degree in electronic engineering from the Institute of Acoustics, Chinese Academy of Sciences, in 2009. He is currently an Associate Professor with the School of Physics and Engineering, Zhengzhou University. His current research interests include intelligent sensor, trace gas detection and signal processing.

Minghui Liu received his Bachelor's degree in measurement and control Technology and instrumentation from Henan Polytechnic University in 2017. He is currently studying for a master's degree in the School of Physics, Zhengzhou University. His research interests are sensing trace gases with photoacoustic spectroscopy.

SHEN TIAN received a B.S. degree from North China Electric Power University, Beijng, China, in 2012. In 2013, he worked as a Product Manager at the Research and Development Center of Hanwei Electronics Group Corporation. Currently, he is pursuing a Ph.D. degree in Mechanics at Zhengzhou University, China. His research interests primarily include acoustic sensors, fiber-optic sensors, and their applications.

Mingyang Feng is now pursuing a Ph.D. degree in mechanics in the School of Physics and Microelectronics of Zhengzhou University, China. His research interests include gas sensors, laser spectroscopy techniques, and photoacoustic spectroscopy.

Yingying Qiao received her Ph.D. degree in Electronic Science and Technology from Beijing University of Posts and Telecommunications (BUPT), Beijing, China, in 2019. She is currently a lecture in the School of Physics and Microelectronics, Zhengzhou University, Zhengzhou, China. Her current research interests include fiber-optic sensors and photoacoustic trace gas sensor.

Chongxin Shan is Professor at the School of Physics and Microelectronics, Zhengzhou University, Zhengzhou, China. He received his Ph.D. degree from Changchun Institute of Optics, Fine Mechanics and Physics, Chinese Academy of Sciences in 2004. His activity is focused on diamond photoelectric materials and devices.
Data availability
Data will be made available on request.
References
- 1.González Fá A.J., et al. Adsorption of carbonyl sulfide on Pt-doped vacancy-defected SWCNT: a DFT study. Appl. Surf. Sci. 2020;525 [Google Scholar]
- 2.Liang Q., et al. Ultrasensitive multispecies spectroscopic breath analysis for real-time health monitoring and diagnostics. Proc. Natl. Acad. Sci. 2021;118(40) doi: 10.1073/pnas.2105063118. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.Swearer D.F., et al. Monitoring chemical reactions with terahertz rotational spectroscopy. ACS Photonics. 2018;5(8):3097–3106. [Google Scholar]
- 4.Lennartz S.T., et al. Marine carbonyl sulfide (OCS) and carbon disulfide (CS<sub>2</sub>): a compilation of measurements in seawater and the marine boundary layer. Earth Syst. Sci. Data. 2020;12(1):591–609. [Google Scholar]
- 5.Le Gal R., et al. Molecules with ALMA at Planet-forming Scales (MAPS). XII. Inferring the C/O and S/H ratios in protoplanetary disks with sulfur molecules. Astrophys. J. Suppl. Ser. 2021;257(1) [Google Scholar]
- 6.Stacewicz T., et al. Detection of disease markers in human breath with laser absorption spectroscopy. Opto Electron. Rev. 2016;24(2) [Google Scholar]
- 7.Svoronos P.D.N., Bruno T.J. Carbonyl sulfide: a review of its chemistry and properties. Ind. Eng. Chem. Res. 2002;41(22):5321–5336. [Google Scholar]
- 8.Saki M., et al. Carbonyl sulfide (OCS): detections in comets C/2002 T7 (LINEAR), C/2015 ER61 (PanSTARRS), and 21P/Giacobini–Zinner and stringent upper limits in 46P/Wirtanen. Astron. J. 2020;160(4):184. [Google Scholar]
- 9.Nguyen T., et al. Successive H-atom addition to solid ocs on compact amorphous solid water. Astrophys. J. 2021;922(2) [Google Scholar]
- 10.Al-Attabi Z., D'Arcy B.R., Deeth H.C. Volatile sulphur compounds in UHT milk. Crit. Rev. Food Sci. Nutr. 2009;49(1):28–47. doi: 10.1080/10408390701764187. [DOI] [PubMed] [Google Scholar]
- 11.Joaquin H.F., Juan L. Quantification of poisons for Ziegler Natta catalysts and effects on the production of polypropylene by gas chromatographic with simultaneous detection: pulsed discharge helium ionization, mass spectrometry and flame ionization. J. Chromatogr. A. 2020;1614 doi: 10.1016/j.chroma.2019.460736. [DOI] [PubMed] [Google Scholar]
- 12.Abe K., et al. Measurement device for ambient carbonyl sulfide by means of catalytic reduction followed by wet scrubbing/fluorescence detection. ACS Omega. 2020;5(40):25704–25711. doi: 10.1021/acsomega.0c02985. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13.Cerda M.M., et al. Progress toward colorimetric and fluorescent detection of carbonyl sulfide. Chem. Commun. 2020;56(67):9644–9647. doi: 10.1039/d0cc04528d. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14.Yin X., et al. ppb-Level SO(2) photoacoustic sensors with a suppressed absorption-desorption effect by using a 7.41 mum external-cavity quantum cascade laser. ACS Sens. 2020;5(2):549–556. doi: 10.1021/acssensors.9b02448. [DOI] [PubMed] [Google Scholar]
- 15.Yin X., et al. Ppb-level photoacoustic sensor system for saturation-free CO detection of SF6 decomposition by use of a 10 W fiber-amplified near-infrared diode laser. Sens. Actuators B Chem. 2019;282:567–573. [Google Scholar]
- 16.Roller C., et al. Carbonyl sulfide detection with a thermoelectrically cooled midinfrared quantum cascade laser. Opt. Lett. 2003;28(21):2052–2054. doi: 10.1364/ol.28.002052. [DOI] [PubMed] [Google Scholar]
- 17.Halmer D., et al. Mid-infrared cavity leak-out spectroscopy for ultrasensitive detection of carbonyl sulfide. Opt. Lett. 2005;30(17):2314–2316. doi: 10.1364/ol.30.002314. [DOI] [PubMed] [Google Scholar]
- 18.Bielecki Z., et al. Application of quantum cascade lasers to trace gas detection. Bull. Pol. Acad. Sci. Tech. Sci. 2015;63(2):515–525. [Google Scholar]
- 19.Wojtas J. Application of Cavity enhanced absorption spectroscopy to the detection of nitric oxide, carbonyl Sulphide, and ethane--breath biomarkers of serious diseases. Sens. 2015;15(6):14356–14369. doi: 10.3390/s150614356. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20.Mohebbifar M.R. Ppb level detection of carbonyl sulfide and ethene and study resonant frequencies using laser photoacoustic spectroscopy. Microw. Opt. Technol. Lett. 2021;63(10):2517–2522. [Google Scholar]
- 21.Ma Y., Dang H. Corner cube prism-enhanced photoacoustic spectroscopy based gas sensing. Infrared Phys. Technol. 2020;109 [Google Scholar]
- 22.Lou C., et al. Highly sensitive light-induced thermoelastic spectroscopy oxygen sensor with co-coupling photoelectric and thermoelastic effect of quartz tuning fork. Photoacoustics. 2023;31 doi: 10.1016/j.pacs.2023.100515. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 23.Liu K., et al. A novel photoacoustic spectroscopy gas sensor using a low cost polyvinylidene fluoride film. Sens. Actuators B Chem. 2018;277:571–575. [Google Scholar]
- 24.Shang Z., et al. Ppb-level mid-IR quartz-enhanced photoacoustic sensor for sarin simulant detection using a T-shaped tuning fork. Sens. Actuators B Chem. 2023;390 [Google Scholar]
- 25.Zhang C., et al. Differential integrating sphere-based photoacoustic spectroscopy gas sensing. Opt. Lett. 2023;48(19):5089–5092. doi: 10.1364/OL.500214. [DOI] [PubMed] [Google Scholar]
- 26.Zhao X., et al. Cantilever enhanced fiber-optic photoacoustic microprobe for diffusion detection of sulfur dioxide. Sens. Actuators B Chem. 2024;405 [Google Scholar]
- 27.Liu K., et al. Multi-resonator photoacoustic spectroscopy. Sens. Actuators B Chem. 2017;251:632–636. [Google Scholar]
- 28.Rouxel J., et al. Miniaturized differential Helmholtz resonators for photoacoustic trace gas detection. Sens. Actuators B Chem. 2016;236:1104–1110. [Google Scholar]
- 29.El-Busaidy S.A.S., et al. Modelling of open photoacoustic resonators. Photoacoustics. 2020;18 doi: 10.1016/j.pacs.2020.100161. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 30.Zhao X., et al. Integrated near-infrared fiber-optic photoacoustic sensing demodulator for ultra-high sensitivity gas detection. Photoacoustics. 2023;33 doi: 10.1016/j.pacs.2023.100560. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 31.Qiao Y., et al. Sensitivity enhanced NIR photoacoustic CO detection with SF(6) promoting vibrational to translational relaxation process. Photoacoustics. 2022;25 doi: 10.1016/j.pacs.2022.100334. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 32.Ma Y., et al. HCl ppb-level detection based on QEPAS sensor using a low resonance frequency quartz tuning fork. Sens. Actuators B Chem. 2016;233:388–393. [Google Scholar]
- 33.Zheng H., et al. Sub-ppb-level CH(4) detection by exploiting a low-noise differential photoacoustic resonator with a room-temperature interband cascade laser. Opt. Express. 2020;28(13):19446–19456. doi: 10.1364/OE.391322. [DOI] [PubMed] [Google Scholar]
- 34.Fu L., et al. Small-volume highly-sensitive all-optical gas sensor using non-resonant photoacoustic spectroscopy with dual silicon cantilever optical microphones. Photoacoustics. 2022;27 doi: 10.1016/j.pacs.2022.100382. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 35.Zhang C., Qiao S., Ma Y. Highly sensitive photoacoustic acetylene detection based on differential photoacoustic cell with retro-reflection-cavity. Photoacoustics. 2023;30 doi: 10.1016/j.pacs.2023.100467. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 36.Cao Y., et al. Humidity enhanced N(2)O photoacoustic sensor with a 4.53 mum quantum cascade laser and Kalman filter. Photoacoustics. 2021;24 doi: 10.1016/j.pacs.2021.100303. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 37.Liu X., Wu H., Dong L. Methodology and applications of acousto-electric analogy in photoacoustic cell design for trace gas analysis. Photoacoustics. 2023;30 doi: 10.1016/j.pacs.2023.100475. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 38.Ma Q., et al. Near-infrared sensitive differential Helmholtz-based hydrogen sulfide photoacoustic sensors. Opt. Express. 2023;31(9) doi: 10.1364/OE.488835. [DOI] [PubMed] [Google Scholar]
- 39.Li Z., et al. Design of a high-sensitivity differential Helmholtz photoacoustic cell and its application in methane detection. Opt. Express. 2022;30(16):28984–28996. doi: 10.1364/OE.465161. [DOI] [PubMed] [Google Scholar]
- 40.Zhang C., et al. Trace gas sensor based on a multi-pass-retro-reflection-enhanced differential Helmholtz photoacoustic cell and a power amplified diode laser. Opt. Express. 2024;32(1):848–856. doi: 10.1364/OE.512104. [DOI] [PubMed] [Google Scholar]
- 41.Miklós A., Hess P., Bozóki Z. Application of acoustic resonators in photoacoustic trace gas analysis and metrology. Rev. Sci. Instrum. 2001;72(4):1937–1955. [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Data Availability Statement
Data will be made available on request.