Evaluation of the effect of surface roughness and modification of implant abutment axial wall on the retention of zirconium oxide copings before and after thermocycling

Parviz Amini 1, Maryam Hejazi 2,, Masumeh Taghva 2, Sina Safari 3
PMCID: PMC11078501  PMID: 38725819

Abstract

Background

The purpose of this study was to determine the effect of one-wall elimination of the abutment and also the surface treatment of the abutment on the retention of cement-retained, implant-supported zirconium oxide copings.

Material and Methods

In this experimental study, four straight abutments were connected to four implant analogs (DIO, UF, Busan, Korea) with 35 Ncm torque. They are mounted vertically in resin blocks. Abutments were prepared as following groups: A) abutment was used in its intact standard form as a control group. B) 4 mm of the flat wall was removed to produce an abutment with 3 walls. C) The abutment surface was abraded with 50 µm AL2O3 powder. D) 4 mm of a flat wall of the abutment was removed, then the abutment surface was abraded with 50 µm AL2O3 airborne particle.10 zirconium oxide copings were made. Samples were cemented with temp bond NE to abutments. The retention of copings was measured before and after incubation using the universal testing machine. T-test, one-way ANOVA, and Post Hoc Tukey Test were used for statistical analysis of data.

Results

In all groups, retention was decreased after thermocycling (P ≤0.001). 3 wall abutments had less retention than the control group before thermocycling. A significant difference was detected between 3 wall abutments and 4 walls of sandblasted abutments before thermocycling. After thermocycling, no difference in retention was seen between groups.

Conclusions

Thermocycling significantly reduces the retention of implant-supported ceramic copings. Sandblasting abutments with 50 µm AL2O3 air-borne particles did not increase the retention of ziconium oxide copings which were cemented with temp bond NE. One wall elimination of abutment decreased the retention of zirconia copings.

Key words:Retention, crown, abutment, zirconia coping, sandblasting.

Introduction

A revolution in the treatment of edentulous areas occurred with the introduction of implants (1). In recent years, zirconium oxide ceramic has been introduced, and it improved the esthetic in implant treatments (2).

Zirconia stabilized with yttrium became popular as a suitable material for the base of the prosthetic superstructures on implants. Polycrystalline ceramics present high mechanical properties and esthetic and biological benefits such as reducing the accumulation of plaque. Besides, manufacturing ceramic restorations with CAD / CAM has benefits like high marginal accuracy (3).

Now controversy about the selection of screw-retained or cemented restorations exists (1). Cemented crowns are more popular so more sensitive Patients prefer them due to the lack of screw access holes and more similarities to natural teeth (2,4). Other advantages of cemented implant-supported crowns include Fewer side effects, higher fracture resistance (4), more tight occlusion, appropriate loading along the longitudinal axis of the implant (5), better passive fitness, smaller occlusal table due to the absence of screw access hole, lower cost, more easily clinical practice in a shorter time especially in the posterior of the mouth (3), reducing the probability of porcelain fracture (6), more uniform stress distribution (7). The only significant advantage of screw-retained prosthesis is it’s easier retrieval possibility. The ability to bring out the restorations on the implants is needed in the periodic replacement of prosthetic components, loosening or breaking the abutment screw, the abutment fracture, changing the prosthesis because of an implant failure, and further surgical intervention (8).

According to Goodacre et al., retention loss of the crown is the most common mechanical failure in cement-retained implant restorations (9).

The retention of cemented restorations depends on the abutment geometry, surface roughness, and the type of cement (10,11).

Cement selection is one of the important factors controlling the retention of restorations (12-16) choosing more retentive cement like zinc phosphate, polycarboxylate, glass ionomer, and resin-modified glass ionomer can damage the implant fixture, implant abutment, and the screw of abutment or prosthesis if aggressive technique is used to remove the restoration (17). It is recommended to cement all implant-supported restorations with temporary cement to retrieve restoration easily without damage (7). Sometimes temporary cements cannot provide the minimum required retention for implant prostheses so other ways are needed to improve the retention (18).

According to Ajay et al. sandblasting and acid etching can improve the retention of Ni-Cr copings made on titanium abutments (19). Oxygen plasma and sandblasting treatment with 50 µm Al2O3 has been shown to enhance the retention of metal copings (20). Furthermore, airborne particle abrasion with 50 µm Al2O3 increased the bond strength between the base abutments and lithium disilicate restorations (21). According to* Kemarly et al. mechanical surface treatment( sandblasting or CoJet silicoating) is more important than chemical surface treatment for bond strength of Lithium disilicate copings to ti-base abutments (22).

Implant geometry is another important factor that affects the retention of cement-type implant restorations (23-26). Kian et al. found that implant abutments with three walls had more retention than those with two walls (25). However, Derafshi, et al. found that three and four wall abutments had the same retention (26).

The effect of sandblasting and the number of implant abutment walls on the retention of zirconia crown is not obvious. This study aims to determine the effect of surface roughness and one-wall elimination of abutment on the retention of cemented zirconia-based implant-supported restorations.

Material and Methods

Four straight abutments were connected to four implant analogs (DIO, UF, Busan, Korea) with 35 Ncm torque. The abutment diameter and height were 4.5 and 5.5 mm, respectively. Each analog was mounted vertically in resin blocks (Acropars 200, Marlic, Iran) with Surveyor (Mariotti, Bravo, Italy) while the abutment-analog interface was 1 mm above the acrylic surface.

Then abutment preparation was done as follows: (Fig. 1)

Figure 1.

Figure 1

Prepared groups.

Group A: intact abutment was considered as a control group. ( 4 walls)

Group B: The 4 mm height of the abutment flat wall was removed with a tapered carbide bur. ( 3 walls)

Group C: the abutment surface was sandblasted with 50 microns of aluminum oxide abrasive particles at a pressure of 2.5 bar and a duration of 15s from 10 mm distance (Rentfert, Germany). ( 4 walls sandblasted)

Group D: The 4 mm height of the abutment flat wall was removed with tapered Carbide bur, then the abutment was sandblasted with 50 microns of aluminum oxide abrasive particles with a pressure of 2.5 bar and a duration of 15s from a distance of 10mm. An acrylic resin index was used to achieve the uniform shape of the removal wall in groups B and D ( 3 walls sandblasted).

Each abutment was scanned with the 3D scanner (3 shape D810, Denmark) then, 10 ceramic copings with a loop on the occlusal surface were designed with the specialized software Dental Designer. Zirconium oxide copings (Dental Direkt (ZW), Germany) were made with CAD-CAM (IMes-icore (450i), Germany). 25-micron space was considered for the cement up to 2 mm above the margin.

All copings were sintered at 1200°C for 12 hours after the preparation. The accuracy of copings was checked with silicon disclosing medium (Dentaco, Germany) and marginal fitness was evaluated with visual and tactile methods using a sharp explorer. Temporary restoration (Cavit, Golchai, Iran) was used to seal the abutment screw access hole.

Tests were done in two stages before thermocycling and after thermocycling:

In the first stage, copings were cemented with temp bond NE (Kerr, Italy) according to manufacturer instructions. The cement was mixed at room temperature for 30 seconds by an operator. Cement was gently applied to the copings then according to ADA specification N.96, they were kept under 5 kg force for 10 min in the chewing simulator (SD Mechatronik, Germany) (Fig. 2). After the initial setting, excessive cement was removed with a plastic explorer. After 24h of resting the samples, they were connected from the loop area to the clamp of the Instron universal testing machine (Testometric, M350-10CT, Germany). Vertical tensile forces were applied at 0.5 mm/min cross-head speed parallel to the longitudinal axis of the samples. The force in which the bond failure had happened was recorded in Newton for each sample.

Figure 2.

Figure 2

The chewing simulator machine.

The copings were checked with the stereo microscope to determine the presence or absence of Crack as well as the bond failure mode. The Failure Mode of cement is described in Table 1.

Table 1. The mode of failure of cement categories.

graphic file with name jced-16-e399-t001.jpg

To perform the second stage of the test, the remaining cement in abutments and copings were cleaned with a plastic explorer and immersed in ultrasonic cleaners (Ultrasonic, Bandelin, Super RK102H, Germany) for 15 minutes. Abutments and copings were washed with water and stored in distilled water for 5 minutes, then they were dried. Finally, the samples were checked visually. After recementation with tempbond NE, the samples were stored in 100% humidity at 37°C for h24. Then the samples were subjected to 1000 thermal cycles, at the temperature 5 to 55°C (TC-300, Iran) with 30 seconds of dwell time. The specimens were dislodged by applying tensile force parallel to the longitudinal axis with the Instron universal testing machine (Testometric, M350-10CT, Germany). The maximum amount of the dislodgement forces was recorded similar to the first stage.

One-way ANOVA, t-test, and Post Hoc Tukey Test were used for statistical analysis of data.

Results

The mean, standard deviation, minimum and maximum removal force, and significant differences between the two stages of tests are shown in Table 2. The maximum removal force was related to 4 walls of sandblasted groups which were (150.61± 16.09 N) before incubation and thermocycling and (47.35 ±31.17 N) after incubation and thermocycling. Minimum removal force was related to 3 walls abutment group (107.47±25.4 N) before thermocycling and (23.92± 11.72 N) after the incubation and thermocycling). In all groups, retention was decreased after incubation and thermocycling (P ≤0.001). 3 wall abutments (groups B and D) had less retention than the control group before incubation and thermocycling (p.value = 0.004 and 0.022 respectively). A significant difference was detected between 3 wall abutments (group B) and 4 walls of sandblasted abutment (group C) before thermocycling. Therefore; before thermocyling 4 walls sandblasted abutment had more retention (P.value = 0.000). After the incubation and thermocycling, no difference was seen between groups.

Table 2. Mean, maximum, minimum, and p-value of groups before and after incubation and thermocycling.

graphic file with name jced-16-e399-t002.jpg

The adhesive Failures of cement were observed in all groups. The prevalence of the remaining cement on abutments before and after thermocycling were shown in Figure 3 and 4, (Table 3).

Figure 3.

Figure 3

The prevalence of residual cement on abutment in groups before incubation and thermocycling.

Figure 4.

Figure 4

The prevalence of residual cement on abutment in groups after incubation and thermocycling.

Table 3. Differences in removal force between groups before and after thermocycling.

graphic file with name jced-16-e399-t003.jpg

Discussion

In this study, the force required to remove the implant-supported zirconia copings was evaluated before and after the incubation and thermocycling. The study found that the removal force was significantly reduced after incubation and thermocycling.

The current study showed that the removal of an abutment wall as often necessary to modify the angle of the implant body, reduced the retention of zirconia copings compared to the control group before thermocycling. While sandblasting the 3-wall abutment can increase coping retention, this amount is not significant statistically. Abutments of groups 3 and 4 sandblasted with 50 microns of aluminum oxide particles compared to similar non-blasting groups showed a higher average of removal force but it was not statically significant.

After thermocycling, the values of the tensile strength dramatically decreased and modifications of the abutment did not have a significant positive effect on retention.

In Kokubo et al.,s study, the effect of sandblasting and thermocycling on the retention of zirconia copings on zirconia abutments using 5 temporary cements was evaluated. The minimum retention was reported for temp bond NE cement in all situations (before and after thermocycling and with or without sandblasting). After thermocycling, the retentive force was significantly decreased in all abutments with or without sandblasting. When temp bond NE was used, sandblasting the abutment did not affect the retention of copings (2). The results are similar to this study. In a study done by Wolfart et al., retention of the restorations cemented with some of the cements was affected by sandblasting but the retention of groups that were cemented with free zinc oxide cements was not affected by sandblasting (27). According to Reddy et al. the higher surface roughness does not influence the retention values of eugenol-free zinc oxide or TempBond cements. However, they found airborne particle abrasion increased the retention for specimens which was cemented with ImProv (28). Therefore increasing the retention of restoration by increasing surface roughness could be cement-dependent and the low cohesive strength of some cements may inhibit the booster effect of surface roughness on retention (29).

De Campos et al. determined the influence of abutment’s surface topography on the retention of single implant-supported metal copings. The research showed that sandblasting with 80-micron aluminum oxide particles and circumferential groove on abutment can increase the retention of metal copings significantly relative to the machined abutments if zinc phosphate cement is used. There was no significant difference between the sandblasted abutment and abutment with groove (30). In the present study, the average removal force in the sandblasted group was more than the control group before and after thermocycling, but statistical differences did not exist probably due to the smaller size of the aluminum oxide particles, different types of copings and different cement.

In Nejatidanesh et al, ‘s study, the retention of zirconium oxide implant-supported restorations using different cements was evaluated. All samples abraded with 110 microns of aluminum oxide particles and 9 cement were compared. All samples were incubated for 24 hours and thermocycled for 5,000 cycles, 5-55°Cwith 30 seconds of dwell time. The highest retention was reported for resin cement and the lowest one for temporary and glass ionomer cements. They suggested that temporary cement and glass ionomers are not suiTable for the bonding of single zirconium oxide crowns (4). In the present study, temp bond NE is not recommended in the mouth as a temporary cement for single implant-supported zirconium oxide restoration because of the very low retention force after thermocycling. The selection of cement that does not provide adequate retention can be a source of retentive failure of restoration.

Farzin et al. evaluated the effect of the one wall removal of abutment and cement type on implant-supported crown retention. The screw access was filled with cotton and composite resin in the control group and abutments with intact walls. In the experimental group, 4 mm of the height of the flat wall of the abutment was removed, the screw was laid with cotton, and the rest of the way was held open. In this study, the removal force of the 3-wall abutment group was significantly higher than the control group when the temp bond cement was consumed. However, the difference was not observed using Dycal (7). These contrary results to the present study may be related to the present study method in which the screw access was filled with a temporary restoration in all groups and removing a wall of abutment reduced the surface area compared to the control group so the retention was decreased. Another similar study was done by Tan et al. in 2012. They concluded that decreasing the number of abutment axial walls from 4 to 3 and from 3 to 2 opposing walls will increase the retention. They did not fill the screw access canal and this can explain the opposite findings to the existing study. Tan et al. related these contradictory findings to the rougher internal axial wall and the vent-like performance of screw access canal for cement which facilitates restoration seating (25).

While Sandblasting with 50 microns of aluminum oxide particles did not increase the retention of Zirconium oxide copings significantly, it is recommended to evaluate the effect of sandblasting with different sizes without abutment weakening on the retention of zirconium oxide copings in future studies.

It is also recommended studies evaluate the maximum forces that could transmit to the screw abutments, implant body, and the implant-bone contact while the removal force is applied to cemented restorations without any damage to components of the implants or the least retentive strength required to prevent displacement to easy retrieval of implant-supported restoration. In vitro studies such as this study do not replace clinical ones so their results should be interpreted clinically with caution.

Conclusions

1) Incubation and thermocycling significantly reduce the retention of implant-supported ceramic copings.

2) Sandblasting abutments with 50 µm AL2O3 air-borne particles did not increase the retention of zirconium oxide copings

3) The retention of the 3-wall abutment group was decreased as compared to the control group and it is not compensated by sandblasting.

4) There was no significant difference between the groups after thermocycling. Since thermocycling reduces significantly the retention of copings, the clinical use of eugenol-free temporary cement (temp bond NE) for single implant-supported zirconia crowns is doubtful.

Acknowledgments

Ackcnowledgements None declared.

${Availability of data and materials} None declared.

Authors contributions None declared.

Ethics None declared.

Conflicts of interest None declared.

Funding None declared.

References

  • 1.Abbo B, Razzoog ME, Vivas J, Sierraalta M. Resistance to dislodgement of zirconia copings cemented onto titanium abutments of different heights. The Journal of Prosthetic Dentistry. 2008;99:25–9. doi: 10.1016/S0022-3913(08)60005-0. [DOI] [PubMed] [Google Scholar]
  • 2.Kokubo Y, Kano T, Tsumita M, Sakurai S, Itayama A, Fukushima S. Retention of zirconia copings on zirconia implant abutments cemented with provisional luting agents. Journal of oral rehabilitation. 2010;37:48–53. doi: 10.1111/j.1365-2842.2009.02013.x. [DOI] [PubMed] [Google Scholar]
  • 3.Rödiger M, Rinke S, Ehret-Kleinau F, Pohlmeyer F, Lange K, Bürgers R. Evaluation of removal forces of implant-supported zirconia copings depending on abutment geometry, luting agent, and cleaning method during re-cementation. The journal of advanced prosthodontics. 2014;6:233–40. doi: 10.4047/jap.2014.6.3.233. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.Nejatidanesh F, Savabi O, Shahtoosi M. Retention of implant-supported zirconium oxide ceramic restorations using different luting agents. Clinical oral implants research. 2013;24:20–4. doi: 10.1111/j.1600-0501.2011.02358.x. [DOI] [PubMed] [Google Scholar]
  • 5.Bernal G, Okamura M, Munoz CA. The effects of abutment taper, length, and cement type on resistance to dislodgement of cement-retained, implant-supported restorations. Journal of Prosthodontics. 2003;12:111–5. doi: 10.1016/S1059-941X(03)00006-8. [DOI] [PubMed] [Google Scholar]
  • 6.Ganbarzadeh J, Nakhaei MR, Shiezadeh F, Abrisham SM. The Effect of Abutment Surface Roughness on the Retention of Implant-Supported Crowns Cemented with Provisional Luting Cement. Journal of Dental Materials and Techniques. 2012;1:6–10. [Google Scholar]
  • 7.Farzin M, Torabi K, Ahangari AH, Derafshi R. Effect of Abutment Modification and Cement Type on Retention of Cement-Retained Implant Supported Crowns. Journal of Dentistry (Tehran, Iran) 2014;11:256. [PMC free article] [PubMed] [Google Scholar]
  • 8.Elsharkawy S, Shakal M, Elshahawy W. Effect of various surface treatments of implant abutment and metal cope fitting surface on their bond strength to provisional resin cement. Tanta Dental Journal. 2015;12:235–40. [Google Scholar]
  • 9.Goodacre CJ, Bernal G, Rungcharassaeng K, Kan JYJTJopd. Clinical complications with implants and implant prostheses. J Prosthet Dent. 2003;90:121–32. doi: 10.1016/S0022-3913(03)00212-9. [DOI] [PubMed] [Google Scholar]
  • 10.Shayegh S, Salari A, Ayoubi M, Younesi F. Effect of three types of temporary luting agents and abutment surface sandblasting on retentive strength of implant supported fixed prostheses. Journal of Islamic Dental Association of IRAN (JIDAI) Spring. 2015;27 [Google Scholar]
  • 11.Nejatidanesh F, Savabi O, Jabbari E. Effect of surface treatment on the retention of implant-supported zirconia restorations over short abutments. The Journal of Prosthetic Dentistry. 2014;112:38–44. doi: 10.1016/j.prosdent.2013.07.026. [DOI] [PubMed] [Google Scholar]
  • 12.Covey DA, Kent DK, Germain HAS, Koka S. Effects of abutment size and luting cement type on the uniaxial retention force of implant-supported crowns. The Journal of Prosthetic Dentistry. 2000;83:344–8. doi: 10.1016/s0022-3913(00)70138-7. [DOI] [PubMed] [Google Scholar]
  • 13.Dixon DL, Breeding LC, Lilly KR. Use of luting agents with an implant system: Part II. The Journal of Prosthetic Dentistry. 1992;68:885–90. doi: 10.1016/0022-3913(92)90544-k. [DOI] [PubMed] [Google Scholar]
  • 14.Maeyama H, Sawase T, Jimbo R, Kamada K, Suketa N, Fukui J. Retentive strength of metal copings on prefabricated abutments with five different cements. Clinical implant dentistry and related research. 2005;7:229–34. doi: 10.1111/j.1708-8208.2005.tb00068.x. [DOI] [PubMed] [Google Scholar]
  • 15.Mansour A, Ercoli C, Graser G, Tallents R, Moss M. Comparative evaluation of casting retention using the ITI solid abutment with six cements. Clinical oral implants research. 2002;13:343–8. doi: 10.1034/j.1600-0501.2002.130401.x. [DOI] [PubMed] [Google Scholar]
  • 16.Cano-Batalla J, Soliva-Garriga J, Campillo-Funollet M, Muñoz-Viveros CA, Giner-Tarrida L. Influence of abutment height and surface roughness on in vitro retention of three luting agents. International Journal of Oral & Maxillofacial Implants. 2012;27:36–41. [PubMed] [Google Scholar]
  • 17.Sheets JL, Wilcox C, Wilwerding T. Cement Selection for Cement-Retained Crown Technique with Dental Implants. Journal of Prosthodontics. 2008;17:92–6. doi: 10.1111/j.1532-849X.2007.00262.x. [DOI] [PubMed] [Google Scholar]
  • 18.Taylor TD, Belser U, Mericske-Stern R. Prosthodontic considerations. Clinical Oral Implants Research. 2000;11:101–7. doi: 10.1034/j.1600-0501.2000.011s1101.x. [DOI] [PubMed] [Google Scholar]
  • 19.Ajay R, Suma K, Ali SA, Sivakumar JSK, Rakshagan V, Devaki V. Effect of surface modifications on the retention of cement-retained implant crowns under fatigue loads: An In vitro study. J Pharm Bioallied Sci. 2017;9:S154. doi: 10.4103/jpbs.JPBS_146_17. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20.El-Helbawy NGE D, El-Hatery AAE W, Ahmed MHJIJoO, Implants M. Comparison of Oxygen Plasma Treatment and Sandblasting of Titanium Implant-Abutment Surface on Bond Strength and Surface Topography. Int J Oral Maxillofac Implants. 2016;31:555–62. doi: 10.11607/jomi.4355. [DOI] [PubMed] [Google Scholar]
  • 21.Pitta J, Burkhardt F, Mekki M, Fehmer V, Mojon P, Sailer IJTJoPD. Effect of airborne-particle abrasion of a titanium base abutment on the stability of the bonded interface and retention forces of crowns after artificial aging. J Prosthet Dent. 2021;126:214–21. doi: 10.1016/j.prosdent.2020.06.013. [DOI] [PubMed] [Google Scholar]
  • 22.Kemarly K, Arnason S, Parke A, Lien W, Vandewalle KJOD. Effect of various surface treatments on Ti-base coping retention. Oper Dent. 2020;45:426–34. doi: 10.2341/19-155-LR. [DOI] [PubMed] [Google Scholar]
  • 23.Das A, Alqahtani FA, Abdulaziz F, AlFadel M, AlJabri R, Elkhyat TJWJoD. Impact of Different Implant Abutment Geometry on Retention of Implant-supported Cement Retained Prosthesis: A Comparative Study. 2023;14:259–62. [Google Scholar]
  • 24.Negahdari R, Bohlouli S, Yazdani J, Torab A, Dizaj SMJD, Problems M. Effect of different geometric changes in the dental implant abutment body on the amount of residual excess cement and retention in a cemented implant-supported prosthesis. Dent Med Probl. 2021;58:207–13. doi: 10.17219/dmp/127347. [DOI] [PubMed] [Google Scholar]
  • 25.Tan KM, Masri R, Driscoll CF, Limkangwalmongkol P, Romberg EJTJopd. Effect of axial wall modification on the retention of cement-retained, implant-supported crowns. J Prosthet Dent. 2012;107:80–5. doi: 10.1016/S0022-3913(12)60028-6. [DOI] [PubMed] [Google Scholar]
  • 26.Derafshi R, Ahangari AH, Torabi K, Farzin M. Evaluation of the effect of axial wall modification and coping design on the retention of cement-retained implant-supported crowns. J Dent Res Dent Clin Dent Prospects. 2015;9:35–9. doi: 10.15171/joddd.2015.007. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 27.Wolfart M, Wolfart S, Kern M. Retention forces and seating discrepancies of implant-retained castings after cementation. Int J Oral Maxillofac Implants. 2006;21:519–25. [PubMed] [Google Scholar]
  • 28.Reddy SV, Reddy MS, Reddy CR, Pithani P, Kumar S, Kulkarni GJJoc. The influence of implant abutment surface roughness and the type of cement on retention of implant-supported crowns. J Clin Diagn Res. 2015;9:ZC05. doi: 10.7860/JCDR/2015/12060.5621. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 29.Craig RG. Restorative dental materials. Mosby; 1980. [Google Scholar]
  • 30.de Campos TN, Adachi LK, Masahiro K, Yoshida H, Shinkai RS, Neto PT. Effect of surface topography of implant abutments on retention of cemented single-tooth crowns. Int J Periodontics Restorative Dent. 2010;30:409. [PubMed] [Google Scholar]