Abstract
Objective:
The mechanical properties of corneal tissues play a crucial role in determining corneal shape and have significant implications in vision care. This study aimed to address the challenge of obtaining accurate in vivo data for the human cornea.
Methods:
We have developed a high-frequency optical coherence elastography (OCE) technique using shear-like antisymmetric (A0)-mode Lamb waves at frequencies above 10 kHz.
Results:
By incorporating an anisotropic, nonlinear constitutive model and utilizing the acoustoelastic theory, we gained quantitative insights into the influence of corneal tension on wave speeds and elastic moduli. Our study revealed significant spatial variations in the shear modulus of the corneal stroma on healthy subjects for the first time. Over an age span from 21 to 34 (N=6), the central corneas exhibited a mean shear modulus of 87 kPa, while the corneal periphery showed a significant decrease to 44 kPa. The central cornea’s shear modulus decreases with age with a slope of −19 +/− 8 kPa per decade, whereas the periphery showed non-significant age dependence. The limbus demonstrated an increased shear modulus exceeding 100 kPa. We obtained wave displacement profiles that are consistent with highly anisotropic corneal tissues.
Conclusion:
Our approach enabled precise measurement of corneal tissue elastic moduli in situ with high precision (< 7%) and high spatial resolution (< 1 mm). Our results revealed significant stiffness variation from the central to peripheral corneas.
Significance:
The high-frequency OCE technique holds promise for biomechanical evaluation in clinical settings, providing valuable information for refractive surgeries, degenerative disorder diagnoses, and intraocular pressure assessments.
Keywords: Elastography, optical coherence elastography, surface acoustic waves, corneal biomechanics
I. Introduction
The mechanical properties of the cornea play a crucial role in determining its response to mechanical stress and have significant implications in vision care. Achieving optimal refractive outcomes in refractive surgeries and accurately measuring intraocular pressure (IOP) in tonometry relies on understanding corneal mechanics while considering corneal stiffness variability. Corneal protrusion in keratoconus (KC) [1, 2], a degenerative disorder, serves as a potential diagnostic marker for localized mechanical degradation [3, 4]. Corneal crosslinking (CXL), a treatment for KC and corneal ectasia, aims to increase stromal elastic modulus and regenerate collagen fibers. Therefore, the ability to measure corneal tissue stiffness is highly valuable in these medical procedures.
While various mechanical tools have been developed for characterizing corneal tissues ex vivo [5–7], in vivo measurements pose significant challenges. Some promising approaches include commercial instruments like the Ocular Response Analyzer and Corvis ST provide overall corneal stiffness indices without spatial resolution [8, 9]. Brillouin microscopy maps longitudinal elastic modulus with high resolution [10, 11] but lacks shear and tensile moduli measurements required for describing corneal deformation.
Several methods have been developed to measure corneal elasticity using elastic waves generated in the cornea. These methods utilize techniques such as air puff [12, 13], ultrasound shear wave [14, 15], and acoustic radiation force [16, 17]. Optical coherence elastography (OCE) is an attractive modality for corneal biomechanical characterization [13, 18–23]. OCE has employed various stimuli mentioned above [24]. In our previous work, we have conducted the first-in-human OCE measurement and obtained shear moduli in the central cornea with wave frequency up to 16 kHz [21].
In this study, we present a more detailed investigation of the human cornea using advanced wave analysis. By applying the acoustoelastic theory to a corneal model accounting for IOP-induced tension, nonlinearities, anisotropy, and spatial variations of tissue stiffness, we have obtained new in vivo data. Our findings reveal a lower shear modulus in the peripheral cornea compared to the central cornea. Additionally, we quantitatively derive the ratio of tensile to shear moduli to reveal the mechanical anisotropy. This work represents a substantial advancement of our conference presentation [25].
II. Theory and Methods
A. Optical coherence elastography system
We utilized a custom-built optical coherence tomography (OCT) system equipped with a swept laser source centered at 1300 nm [21, 26]. The system operated at an A-line rate of 43.2 kHz, delivering an optical power of less than 10 mW to the cornea. Elastic waves were generated using a contact probe, covering a broad frequency range from 2 to 30 kHz. Figure 1a presents a schematic of the experimental setup. The probe consisted of a piezoelectric transducer (PZT) and a tip with a radius of curvature of 0.4 mm. A gentle contact force of approximately 0.01 N was applied as the tip contacted the corneal surface.
Figure 1.

(a) Schematic of optical coherence elastography using a contact probe. The inset shows a photo of the probe consisting of a piezoelectric (PZT) transducer and a tip. (b) Biomechanical model of the cornea and an excited elastic wave. Schematic (i) illustrates the microstructure of the stroma under intraocular pressure (IOP) and in-plane tension σ. Schematic (ii) demonstrates the A0 mode wave along the cornea, involving shear deformation.
At each location of the OCT beam, we acquired 360 A-lines (M-scan). The acquired M-scan data underwent Fourier transformation to extract the wave amplitudes and phases along the depth at the respective beam location. Depending on the wave frequency, data were collected at 96 transverse points. The scan length was 10 mm at 2, 4, and 6 kHz, varied inversely proportionally to the frequency from 10 mm at 6 kHz to 3.3 mm at 20 kHz, and then set to 3 mm at 25 and 30 kHz. Details about the data processing can be found in Ref [21]. In brief, we extracted displacement profiles over time , and then performed a Fourier transform to move the data from the time () domain to the frequency () domain. The phase and amplitude of the mechanical vibration were filtered at the driving frequency. To calculate the wave speeds, we obtained 1-dimensional displacement profiles from the outer surface (epithelium) of the cornea, and applied a Fourier transform to measure the wavenumber ݇of each wave. The phase velocity () was then calculated using the equation . The acquired dataset was used to generate a wave image, representing a snapshot of the wave at a specific phase of oscillation.
B. The microstructure and mechanical stress of the cornea
The corneal stroma is composed of lamellae, consisting of finely arranged collagen fibrils along the plane of the cornea [27, 28]. The anterior cornea exhibits more interwoven and undulating collagen bundles, while the mid and posterior corneas display orthogonally arranged lamellae [29]. This unique microstructure gives the cornea its distinctive elastic properties [30, 31]. The cornea is commonly modeled as an anisotropic composite plate (Fig. 1b (i)), with mechanically reinforcing fibers aligned along the plane [32, 33]. These fibers enhance the tissue’s stiffness along the plane. The collagen fibers contribute to nonlinear elasticity or hyper-elasticity, where stiffness increases with strain. However, their influence on the shear modulus is thought to be relatively minimal and independent of the shearing direction. The corneal tissue exhibits significant mechanical anisotropy due to the pronounced differences between in-plane and out-of-plane tensile moduli, as well as between in-plane tensile and shear moduli.
While the compressional stress from IOP exerted on the posterior surface has minimal impact on tensile and shear moduli, IOP induces in-plane tension in the corneal tissue. This tension significantly increases the tensile modulus by stretching the collagen fibers. According to the Young-Laplace equation, the tensional stress parallel to the plane is given by
| (1) |
Here, and represent the radius of curvature and thickness of the cornea, respectively. With and , we find that . For . Such an in-plane tension significantly alters the in-plane tensile modulus of the cornea. Additionally, corneal tension affects the speed of elastic waves utilized in OCE through the acoustoelastic effect, which we will describe in the next section.
C. The acoustoelastic theory
The cornea is a waveguide for elastic waves due to its surrounding air and aqueous humor [21, 34–36]. With our contact vibrational probe, we primarily excite and analyze the fundamental antisymmetric Lamb wave known as the A0 mode. Figure 1b(ii) illustrates the flexural deformation profile of the wave. In a tension-free isotropic material, the wave speed of the A0 mode would be determined by the shear modulus of the material [37]. However, in actual corneal tissues, which are anisotropic, nonlinear, and under tension, the wave speed depends on both shear and in-plane tensile moduli, as well as the tension .
According to the acoustoelastic theory [38, 39], the propagation of a plane elastic wave can be described using three parameters, , , and , which characterize material stiffness and stress of the cornea. It can be shown that (see Appendix A and B)
| (2) |
The stress explicitly affects the propagation of elastic waves. The parameters are related to the out-of-plane shear modulus and in-plane tensile modulus of the prestressed cornea, via (see Appendix C)
| (3) |
When , and reduce to the intrinsic Young’s modulus, , and shear modulus, , of the cornea in the stress-free condition [21].
We adopt the following strategy to determine the elastic moduli. First, we observe that the wave speed approaches approximately at high frequencies when the elastic wavelength is comparable to or smaller than the corneal thickness (Appendix Fig. A1). Hence, is readily determined from high-frequency OCE data. Second, we estimate the ratio from measured wave displacement profiles throughput the depth. Finally, is determined from the relation , where is calculated based on the IOP. Notably, . We will utilize this equation later.
In general, the magnitudes of , , and can vary within the cornea. Conducting a full three-dimensional analysis of these parameters is beyond the scope of this study. Instead, our analysis assumes that the three parameters remain constant within a small region of interest spanning a few wavelengths. We measure the transverse variation by measuring , , and at different locations from the central cornea to the sclera.
To quantitatively describe the wave motion, we employed the acoustoelastic solution derived from a constitutive model by Gasser et al. [40], so called the Holzapfel-Gasser-Ogden (HGO) model, which was originally developed to describe arterial walls. This model accounts for the anisotropic stiffness enhancement caused by collagen fibers while neglecting fiber dispersion in the central cornea. Subsequently, we derive a secular equation using the incremental dynamic theory [38, 39] to determine the dispersion relation and modal shape of the A0 wave. Within the range of validity of the constitutive model, we estimated the nonlinear variation of and as a function of IOP.
D. Finite element analysis
To simulate the cornea’s mechanical behavior, we developed a finite element analysis (FEA) model using Abaqus/standard software (Abaqus 6.12, Dassault Systèmes). In the simulations of elastic wave propagation, we employed a plane strain model in an annulus geometry. The corneal curvature and thickness were obtained from OCT images and incorporated into the model. The aqueous humor was represented as an acoustic medium, initialized with a pressure matching the IOP. The arbitrary Lagrangian-Eulerian (ALE) adaptive mesh was employed to re-mesh the deformed acoustic medium during the initialization. Elastic waves were induced by applying a local surface pressure that mimics the contact probe.
To determine the contact stiffness, we employed an axisymmetric model for the cornea and a rigid spherical shell (radius 0.5 mm, see Appendix Fig. A3) to represent the PZT probe indenter. To account for the increase in intraocular pressure caused by indentation, the aqueous humor was modeled using the fluid-filled cavity feature available in Abaqus/standard with constant inner volume. The contact stiffness was derived from the slope of a force-displacement curve, allowing us to accurately capture the interaction between the cornea and the indenter during the simulation.
In the FEA models, we utilized the same constitutive material model [40] as employed in the analytical analysis. We ensured mesh convergence by verifying that the simulation results were independent of the mesh density.
E. Wave speed map reconstruction
The wave speed map illustrating the shear modulus was reconstructed using the phase gradient algorithm [41, 42] with a window size about half wavelength of the elastic wave. At 16 kHz, the window size is approximately 0.25 mm (see Table 1). Within the window centered at , wave numbers and were calculated along the propagation direction and . The 2-D wave speed map was calculated using .
TABLE I.
TYPICAL ELASTIC WAVELENGTHS AT DIFFERENT FREQUENCIES IN THE CENTRAL CORNEA
| Frequency (kHz) | 2 | 6 | 10 | 16 | 30 |
| Wavelength (mm) | 3.90 | 1.32 | 0.76 | 0.48 | 0.29 |
F. Human study protocols
This study was conducted following a protocol approved by Mass General Brigham Institutional Review Board (IRB #2017P000867, date of approval: 08/21/17). The clinical trial was registered at www.clinicaltrials.gov (National Clinical Trial Identifier: NCT03230981). Six healthy subjects in the 20s (N = 3) and 30s (N = 3) were recruited. Written informed consent was obtained from each subject after providing a detailed explanation of the study’s nature and possible consequences. Only the left eyes of the subjects were scanned. Prior to OCE measurements, proparacaine ophthalmic drops were applied as a topical anesthetic to the left eye.
We measured the broadband wave speed dispersion at the central cornea from two subjects (Subject 1: 30 yo, Subject 2: 32 yo). We positioned the probe near the cornea’s center (refer to Fig. 2a) and generated elastic waves at various frequencies ranging from 2 to 30 kHz with a frequency increment of 2 or 5 kHz. Next, we measured the transverse variation of corneal stiffness on six subjects in the 20s (N = 3) and 30s (N = 3). We maintained the frequency at 16 kHz and successively moved the probe to five locations along the left eye, ranging from the temporal sclera near the limbus to the central cornea, with a step size of approximately 2 mm. Two lateral scans were performed per subject. The wave velocity at each location was computed. Statistical comparisons were made between the two age groups using the two-sample t-test.
Figure 2.

(a) Photograph showing the OCE measurement being performed on a volunteer. (b) A representative OCT image of the cornea. The circular symbol represents the probe location, and an overlaid OCE wave image measured at a frequency of 16 kHz is shown. (c) Wave images at various wave frequencies from 8 to 30 kHz, all captured at the same probe location. As the frequency increases, both the wavelength and propagation distance decrease. (d) Measured elastic wave speeds and calculated shear moduli obtained from two volunteers (Subject 1: 30 yo, Subject 2: 32 yo). The data points (circles) and shades represent the mean and standard deviation over three scan measurements. Low-frequency data below 6 kHz were prone to error and thus were excluded from the fitting. Dashed curves represent theoretical fits to the data above 6 kHz.
III. Results
A. Shear modulus of the cornea in vivo
Figures 2b and 2c depict a representative OCT image of Subject 1 and the corresponding OCE wave images acquired at different frequencies. The wave speeds as a function of frequency for both subjects are presented in Fig. 2d. The elastic wavelengths at different frequencies are listed in Table I, using data from Subject 1 as an example. At 2 kHz, the wavelength is significantly larger than the thickness of the cornea (~ 0.5 mm). At 16 kHz and above, the elastic wavelength becomes shorter than the corneal thickness while generating sufficient vibrational amplitudes for reliable speed measurement.
We employed curve fitting based on the acoustoelastic theory, incorporating various morphological and mechanical parameters reported in Ref [43, 44], with the shear modulus as the sole fitting parameter. By obtaining the best-fit curves, we derived in-situ shear modulus values of 75.5 ± 5.0 kPa for Subject 1 and 72.3 ± 4.6 kPa for Subject 2 (± represents 95% confidence level of the curve fitting). These values align with our previous mean value of 72 kPa, obtained from 12 subjects, exhibiting an interpersonal variability of 14 kPa, through simpler Kevin-Voight model fitting [21].
Using the constitutive model, we estimated the “intrinsic” shear modulus of the cornea under zero tension and zero strain, which theoretically could be measured from excited corneal tissues. These estimated the mean intrinsic shear modulus values () were 64.2 kPa from Subject 1 and 62.5 kPa from Subject 2. As expected, these values were lower than the shear modulus of the cornea in situ and similar to the value reported previously [45].
B. Spatial variation of shear modulus in the cornea
Figure 3 illustrates the wave speeds measured at different locations within the eye. The wave speeds in the sclera are 15.7 ± 0.45 m/s for Subject 1 and 13.7 ± 0.4 m/s for Subject 2, corresponding to approximately 2-fold values compared to the average speeds observed in the cornea and yielding 4-fold higher shear moduli. Notably, the peripheral region in the cornea, located approximately 1 mm away from the limbus, exhibits the lowest wave speeds (6.4 ± 0.2 m/s for Subject 1 and 6.3 ± 0.15 m/s for Subject 2). This corresponds to in-situ shear moduli () of 41 kPa and 40 kPa, respectively, which is nearly half the value (~74 kPa) observed in the central cornea. Additional wave speed data obtained from all the 6 subjects are provided in Appendix Fig. A2. Figure 3c clearly demonstrating the shorter wavelength in the peripheral cornea. Fig. 3d presents a wave speed map for Subject 1, reconstructed using a phase gradient algorithm (see Method). This speed map reveals striking gradients in corneal mechanics, with a spatial resolution of approximately 0.5 mm (the wavelength at 16 kHz).
Figure 3.

Spatial variations across the cornea. (a) Schematic showing the probe locations in five consecutive measurements. (b) Measured wave velocities at 16 kHz from two volunteers (Subject 1: 30 yo, Subject 2: 32 yo). The highest speed is observed in the sclera, while the lowest speed is found in the peripheral cornea. (c) Wave images of the peripheral and central corneas at 16 kHz. (d) Wave velocity map at 16 kHz overlaid on an OCT image (Subject 1). (e) Transverse profile of the measured shear modulus. (f) Contact stiffness profile calculated from the shear modulus map in (e). The slightly greater thickness in the peripheral cornea compensates for the lower shear modulus, result in relatively uniform bending stiffness rather uniform across the entire corneal region.
Figure 4a compares the wave speeds in the central and peripheral cornea between the two age groups. For the 20s group, the wave speed decreases from 9.55f0.53 m/s in the central corneas to 6.33f0.29 m/s in the peripheral cornea. For the 30s group, the wave speed decreases from 8.37f0.26 m/s to 6.58f0.25 m/s. The central corneas exhibited significant decrease between the 20s and 30s, as previously observed [21]. Interestingly, no significant difference was found in the peripheral corneas between the two age groups. Figure 4b shows the shear moduli ( obtained from all six subjects as a function of the age . A linear regression analysis revealed a negative slope of −19.1 ± 8.0 kPa (95% confidence interval) per decade in the central corneas. For the peripheral corneas, a modest slope of ~ 3.5 kPa per decade was observed, with a 95% confidence range between −1.1 and 8.1 kPa per decade.
Figure 4.

(a) Statistical comparison of the wave speed in the central and peripheral corneas of the two age groups. ***, p < 0.001. (b) Regression analysis of age dependence of the shear modulus in the central and peripheral corneas. Shaded regions represent 95% confidence intervals.
The spatial heterogeneity in the central and peripheral corneas may be attributed to two factors. Firstly, the peripheral cornea is about 20% thicker than the central cornea (0.64 mm versus 0.53 mm) [43]. Consequently, at a given IOP, the greater thickness results in a lower tensile stress in the peripheral cornea [as per Eq. (1)], leading to a decrease in wave speed via the acoustoelastic effect. Secondly, the peripheral cornea may possess an intrinsic softness compared to the central cornea. To verify this, we conducted finite element simulations. Assuming the cornea had uniform intrinsic material properties, the wave speed in the thicker periphery was only 4% lower than in the central cornea. However, the significant variation in wave speed observed could only be replicated when the shear modulus in the peripheral cornea was estimated to be 64% of that in the central cornea. Previous ex vivo studies on human corneas have also reported similar gradients. For instance, an ultrasound study demonstrated that the shear modulus of the anterior region of the cornea varied from 1.4 kPa in the center to 0.6 kPa in the periphery [44]. Although their modulus values are considerably lower, likely influenced by swelling in cadaver tissues, the ratio of difference appears comparable. Figure 3e exhibits a shear modulus profile derived from the speed gradient map using the acoustoelastic theory. The shear modulus in the limbus is higher than 100 kPa.
Next, we evaluated the contact stiffness, which characterizes the cornea’s response to local surface indentation. This property is extrinsic and influenced by both intrinsic elastic modulus and morphology. We conducted finite element simulations to calculate the displacement slope as a function of indentation load using a rigid sphere (radius 0.5 mm) (see Appendix Fig. A3). As depicted in Fig. 3f, the contact stiffness remains relatively uniform throughout the cornea, primarily due to the contribution of thickness () to bending stiffness (proportional to ) during indention.
C. Anisotropy of tensile and shear moduli
According to the standard acoustic theory for a uniform plate made of isotropic material, the wave profile of the A0 mode would deviate significantly from a uniform shear-like motion. At low frequencies where the wavelength exceeds the thickness, the A0 mode exhibits flexural (bending) motion, while at high frequencies, the wave becomes increasingly localized at the surface. However, the in vivo wave images we measured displayed nearly bulk shear wave-like motion (see Fig. 2c and Fig. 5a). To understand this phenomenon, we conducted finite element analysis based on the acoustoelastic theory, confirming that the distinctive wave profile arises from the anisotropy or the significance difference between and . In our model, for a given dimensionless wave number , the displacement profile is primarily influenced by the ratio but is less sensitive to the ratio (see Appendix Fig. A4).
Figure 5.

Anisotropic property of the corneal stroma. (a) Wave map measured at 16 kHz in the central cornea across the entire thickness. (b) Normalized displacement profile (magenta) across the depth, along with a best-fit simulation result for a uniform, anisotropic tissue under tension (blue), and a theoretical curve for an isotropic material (dashed curve). (c) Same as (b), but for a nonuniform tissue with depth-dependent stiffness (the top 1/3 region has 4 times higher shear modulus than the bottom 2/3 thickness). (d) Simulated wave profile assuming no viscous damping (corresponding to the blue curve in (b)).
Figure 5b shows a typical depth profile measured at 16 kHz in the central cornea. In the finite element simulation, we initially assumed spatially uniform elastic moduli and varied the ratio to fit the experimental data. Figure 5 showcases the wave depth profiles for two representative cases: (3 for nonplanar elastic waves), corresponding to isotropic materials without tension (represented by dashed curve), and the best-fit case with . Using from the IOP, we obtained . With , we estimate to be approximately 1.8 MPa.
Previous mechanical testing of ex vivo corneal tissues (typically without external tension) revealed that the anterior stroma is significantly stiffer than the mid and posterior stroma [30, 31]. Rheometry measurements have reported a shear modulus of 7.7 ± 6.3 kPa in anterior layers, 3–5 times higher than the shear modulus of 2.0 ± 0.45 kPa in middle layers and 1.3 ± 1.0 kPa in the posterior layers [46]. Such a downward gradient of modulus should result in a more localized wave profile in the lower layers. To counteract the tendency, even greater anisotropy is required to produce shear-like waves. We conducted finite element simulation for a stepwise case, where the top 1/3 region had a shear modulus four times higher than the bottom 2/3 region. Notably, the best fit for the nonuniform cornea was obtained with (Fig. 5c), corresponding to and . These estimated values are comparable to the previously reported tensile moduli values of 0.8–2.2 MPa obtained from ex vivo samples through quasi-static stress-strain testing [7].
IV. Discussion
The analysis of acoustoelastic data in our study provided a comprehensive understanding of the corneal wave characteristics by considering the effects of collagen fiber arrangement and IOP-induced tension. The anisotropic and nonlinear HGO mechanical model we used has advantages over previous, simpler corneal models, such as an isotropic model and a nearly incompressible transverse isotropic (NITI) model [32]. The NITI model is a linear elastic model neglecting the nonlinear elasticity of the corneal tissues under tension in vivo. Considering nonlinear elasticity is critical to correctly describe the effect of IOP on elastic wave speeds via the acoustoelastic theory. Furthermore, the NITI model is a phenomenological model which assumes an isotropic mechanical property in the plane (the corneal plane) and low out-of-plane shear modulus. By contrast, the HGO mode considers tissue microstructure consisting of two families of collagen fibrils, which naturally results in larger in-plane stiffness than the outof-plane due to the engagements of the collagen fibrils.
The wave speed values measured at frequencies below 6 kHz were excluded from the model fitting because their long wavelengths tended to be affected by finite corneal geometry. Wave reflection from the corneal-sclera boundary is prominent at low frequencies as shown in Appendix Fig. A5. Additionally, the finite corneal curvature can increase the apparent wave speeds at low frequencies.
By utilizing the derived acoustoelastic solution in conjunction with the incremental dynamic theory, we achieved accurate estimation of wave dispersion and modal shape for the A0 wave, while also considering the influence of IOP on the nonlinear variation of the out-of-plane shear modulus and in-plane tensile modulus. This approach allowed for a detailed quantitative description of corneal wave behavior and its relationship with IOP. The acoustoelastic theory allows us to estimate corneal modulus without tension from the actual shear modulus values measured in vivo. According to the constitutive material model, for example, for a shear modulus of 74 kPa under normal IOP, the intrinsic shear modulus without corneal tension (as observed in excited corneas) is estimated to be 63 kPa, with the tension-induced stiffening of collagen fibrils in the stroma accounting for the 11 kPa difference.
Table II lists the measured results of in vivo human cornea elasticity obtained from previous studies. Potential factors contributing to the differences between the present study and previous research include variations in IOPs, and utilization of different models for elasticity estimation. Furthermore, the low frequency elastic waves used in previous research may not have offered adequate resolution and reliability for estimating corneal elasticity.
TABLE II.
Previous studies measuring in vivo human corneal elasticity
| Method | Stimulus (Frequency) | Results | Model | Reference |
|
| ||||
| US | Shaker(100 Hz) | Young’s modulus, 696 ± 113 kPa | Lamb wave | Sit [50] |
| OCE | Contact probe (2 – 16 kHz) | Shear modulus, 72 ± 14 kPa (age: 25 to 67 years old) | Lamb wave | Ramier [21] |
| OCE | Contact probe (2 – 30 kHz) | Shear modulus, 71–103 kPa for central cornea, 40–48 kPa for peripheral cornea | Lamb wave | Present study |
| OCE | Air-puff (~500 Hz) | Young’s modulus 733 ± 164 kPa | Scholte wave | Jin [22] |
| Corvis ST | Air-puff (~30 Hz) | Young’s modulus ~0.40 MPa for posterior, ~1.52 MPa for anterior cornea | Finite element analysis | Kling [51] |
| Corvis ST | Air-puff (~30 Hz) | Young’s modulus, ~0.32 MPa | Modal analysis | Shih [52] |
| Applanation Tonometry |
Applanation (Static) | Young’s modulus, 0.25 ± 0.1 MPa | Orssengo-Pye model | Pye [53] |
| Indentation | Flat indenter (Static) | Tangent modulus, 0.755 ± 0.159 MPa | Thin shell model | Lam [54] |
US = ultrasound elastography, OCE = optical coherence elastography.
The high-frequency OCE technique utilized in this study provided excellent spatial resolution, surpassing 1 mm. We observed a gradual decrease in tissue stiffness from the central cornea to a minimum shear modulus of 41 kPa in the peripheral cornea (~ 1 mm from the limbus). The underlying structural and physiological reasons for the low stiffness in the corneal periphery remain unclear. The shear modulus of the limbus was higher than the cornea but lower than the sclera. It may be worthwhile to investigate the limbal region with higher spatial resolution at ultrasonic frequencies [55]. Conversely, the increased thickness in the peripheral cornea compensates for the reduced shear modulus, resulting in contact stiffness that remains relatively uniform throughout the cornea. This finding has implications for applanation or air-puff tonometry, where significant corneal deformation occurs. Thus, the peripheral region is softer but not weaker from a mechanical standpoint.
Through the acoustoelastic analysis of the measured wave profiles, we estimated the tensile (Young’s) modulus of the central cornea to be approximately 7 MPa when assuming that the anterior cornea is four times stiffer than the mid and posterior regions. This value is subject to some uncertainty within a few MPa, depending on the exact stiffness profile along the depth. A more direct way to measuring tensile stiffness may involve employing a symmetric (S0) Lamb wave in addition to the A0 mode, which is currently under development in our laboratory.
V. Conclusion
We have developed an advanced OCE technique that allows for measuring in situ corneal elastic moduli with high precision (<7%) and high spatial resolution (< 1 mm). Using this method, we observed remarkable stiffness gradients in human cornea in vivo for the first time. The corneal shear modulus was found to be the minimum at the periphery about 1 mm away from the corneal-limbal junction. The peripheral stiffness exhibited much smaller age dependence between the 20s and 30s than the central cornea stiffness. Our technique has potential for clinical applications to obtain patient-specific in vivo biomechanical data from patients.
Acknowledgments
This study is supported by the U.S. National Institutes of Health (NIH) via grants R01EB027653, R01EY033356, and R01EY034857.
APPENDIX
A. Acoustoelastic model for corneal OCE
Here we derive the mechanical model for corneal OCE. To incorporate mechanical anisotropy and mechanical loading, we build our model on the acoustoelastic theory. Readers can refer to Ref [47] for details of the theory.
Wave equation.
According to the acoustoelastic theory, the wave equation for small-amplitude plane elastic wave in a uniformly prestressed solid reads [39]
| (A1) |
where we have used and to denote the coordinates that correspond to and in the main text, respectively. The stream function relates to displacement components and via the relation of and . It satisfies , the constraint equation for material incompressibility. denotes the mass density (1060 kg/m3 for corneal tissues). The coefficients , , and are determined by the constitutive law and the stretch ratio ,
| (A2) |
where the fourth-order tensor, , is the Eulerian elasticity tensor and is defined as
| (A3) |
where is the deformation gradient tensor , and is the strain energy function. In Eq. (A3) the Einstein’s summation convention is used. For the Holzapfel-Gasser-Ogden (HGO) model, the strain energy is
| (A4) |
where , , and are constitutive parameters. denotes the shear modulus in stress-free state. The dimension of is the same as , whereas is a dimensionless parameter which determines the nonlinear hardening effect of the collagen fibrils when being stretched. The parameter characterizes the dispersion of collagen fibrils. indicates the collagen fibrils are randomly distributed, resulting in an isotropic fibril network. if the collagen fibrils are ideally aligned [48], which is applicable for the cornea. . and are two invariants related to two families of collagen fibers. Following the coordinates defined in Fig. 1b, the axes of the collagen fibers of the cornea, denoted by unit vectors and , are aligned with and , i.e., and . Then and can be determined by and [40]:
| (A5) |
Then, the coefficients , , and of the wave equation can be obtained by inserting Eqs. (A3) and (A4) into Eq. (A2)
| (A6) |
In the absence of prestress (i.e., or )
| (A7) |
According to Eq. (3), and , which further reduce to and for isotropic materials . Please note is the so-called plane-strain Young’s modulus involved in plane wave motion. For isotropic materials, the plane-strain Young’s modulus relates to shear modulus by , which equals 4 when the Poisson’s ratio is 0.5.
Dispersion relation.
We now consider the guided wave motion in the cornea. The two sides of the cornea are interfaced with the air and aqueous humor, respectively. The aqueous humor is modeled as a semi-infinite fluid layer and the wave equation is
| (A8) |
where and denote the bulk modulus and density of the fluid, respectively. is a potential function related to the displacement of the fluid (denoted by ) through of and . The pressure of the fluid, denoted by , is determined by
| (A9) |
At the interface between cornea and aqueous humor , the following interfacial conditions apply
| (A10) |
At the surface of cornea , the stress-free boundary conditions require
| (A11) |
More details on the derivations of the boundary conditions can be found in previous studies [38, 49].
We seek the plane wave solutions for and , i.e.,
| (A12) |
where is the wavenumber, and is the phase velocity. Inserting Eq. (A12) into Eqs. (A1) and (A8), we get
| (A13) |
The parameters and are determined by
| (A14) |
and
| (A15) |
Substituting and into Eqs. (A10) and (A11), we get
| (A16) |
where the nonzero components of the 5 × 5 matrix are
| (A17) |
In this derivation, we have utilized the identity
| (A18) |
which can be obtained from Eq. (A14).
The dispersion relation can be obtained by solving equation
| (A19) |
The nontrivial solution of gives the modal shape . The vertical displacement is given by:
| (A20) |
B. Relating IOP to stretch ratio
The in-plane stress within the cornea that balances the intraocular pressure can be determined by the Young-Laplace equation: . Here we show how to relate the stress to the stretch ratio , which is involved in the dispersion relation.
The strain energy function relates the deformation to Cauchy stress by
| (A21) |
where is a Lagrange multiplier for material incompressibility and is the Kronecker delta. For cornea the out-of-plane stress is negligible in comparison to the in-plane stress . So we can get
| (A22) |
The stretch ratio can be obtained by solving the nonlinear equation
| (A23) |
Equation (A23) relates IOP to the deformation . Notably, it is straightforward to check
| (A24) |
For normal human subjects with typical of 15 kPa and of 70 kPa, we obtain .
C. Lamb waves in cornea
Appendix Fig. A1 shows representative dispersion relations of the fundamental Lamb waves (A0 and S0) in cornea obtained from our theoretical model. At zero frequency, , the phase velocity of the S0 mode is . When , the phase velocities of A0 and S0 become the Scholte and Rayleigh surface wave speeds, respectively, which are primarily determined by . The two phase velocities, in the absence of prestress, are related to plane-strain Young’s modulus (; here the subscript from has been dropped for simplicity) and shear modulus by and , indicating and reduce to the Young’s modulus and shear modulus, respectively. Here , , and are components of the stiffness matrix of the intrinsic corneal tissue in the stress-free condition.
Inspired by this observation, we can use and to characterize the in-situ tensile and shear stiffness (denoted by and ) of the cornea under tension. According to Figure A1, the dispersion relation of A0 is sensitive to , and the dispersion relation of S0 in the low frequency regime is sensitive to , which suggests we can interrogate shear and tensile stiffness of cornea by probing A0 and S0 Lamb waves, respectively.
Figure A1.

Dimensionless dispersion relations of the A0 and S0 mode Lamb waves. Solid lines, , . Dashed lines, isotropic materials without prestress . The phase velocity of S0 at is , governed by the tensile stiffness . In the high frequency regime, the phase velocities of A0 and S0 reach plateaus, which are Scholte (solid-fluid interface) wave and Rayleigh wave speed, respectively.
Figure A2.

Wave speeds measured along the cornea from six healthy subjects.
Figure A3.

Finite element analysis of the local contact stiffness for the cornea. (a) a presentative corneal deformation. (b) The load-force curve obtained from different locations.
Figure A4.

Modal shapes for . (A) Variation of the modal shape with the ratio when . (B) Variations of the modal shape with the ratio when .
Figure A5.

(a) Wave displacement profiles extracted from the surface of the cornea at different frequencies. (b) Wavenumber domain Fourier transform of the wave displacement. Black arrow: A0-wave. Red arrow: Waves reflected from the corneal-scleral boundaries.
Contributor Information
Guo-Yang Li, Harvard Medical School and Wellman Center for Photomedicine, Massachusetts General Hospital. Guo-Yang Li is currently with the Department of Mechanics and Engineering Science, College of Engineering, Peking University..
Xu Feng, Harvard Medical School and Wellman Center for Photomedicine, Massachusetts General Hospital. Guo-Yang Li is currently with the Department of Mechanics and Engineering Science, College of Engineering, Peking University..
Seok-Hyun Yun, Harvard Medical School and Wellman Center for Photomedicine, Massachusetts General Hospital. Guo-Yang Li is currently with the Department of Mechanics and Engineering Science, College of Engineering, Peking University.; Harvard-MIT Division of Health Sciences and Technology.
References
- [1].Ambrosio R et al. , “Integration of Scheimpflug-Based Corneal Tomography and Biomechanical Assessments for Enhancing Ectasia Detection,” Journal of Refractive Surgery, vol. 33, no. 7, pp. 434–443, Jul, 2017. [DOI] [PubMed] [Google Scholar]
- [2].Lopes BT et al. , “Enhanced tomographic assessment to detect corneal ectasia based on artificial intelligence,” American Journal of Ophthalmology, vol. 195, pp. 223–232, Nov, 2018. [DOI] [PubMed] [Google Scholar]
- [3].Andreassen TT et al. , “Biomechanical properties of keratoconus and normal corneas,” Exp Eye Res, vol. 31, no. 4, pp. 435-41, 10/1980, 1980. [DOI] [PubMed] [Google Scholar]
- [4].Dupps WJ Jr, and Wilson SE, “Biomechanics and wound healing in the cornea,” Experimental eye research, vol. 83, no. 4, pp. 709–720, 2006. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [5].Andreassen TT et al. , “Biomechanical properties of keratoconus and normal corneas,” Experimental eye research, vol. 31, no. 4, pp. 435–441, 1980. [DOI] [PubMed] [Google Scholar]
- [6].Cartwright NEK et al. , “Age-related differences in the elasticity of the human cornea,” Investigative ophthalmology & visual science, vol. 52, no. 7, pp. 4324–4329, 2011. [DOI] [PubMed] [Google Scholar]
- [7].Wollensak G et al. , “Stress-strain measurements of human and porcine corneas after riboflavin–ultraviolet-A-induced cross-linking,” Journal of Cataract & Refractive Surgery, vol. 29, no. 9, pp. 1780–1785, 2003. [DOI] [PubMed] [Google Scholar]
- [8].Zhang D et al. , “Exploring the Biomechanical Properties of the Human Cornea In Vivo Based on Corvis ST,” Frontiers in Bioengineering and Biotechnology, vol. 9, pp. 771763, 2021. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [9].Luce DA, “Determining in vivo biomechanical properties of the cornea with an ocular response analyzer,” Journal of Cataract & Refractive Surgery, vol. 31, no. 1, pp. 156–162, 2005. [DOI] [PubMed] [Google Scholar]
- [10].Scarcelli G et al. , “In vivo measurement of age-related stiffening in the crystalline lens by Brillouin optical microscopy,” Biophysical journal, vol. 101, no. 6, pp. 1539–1545, 2011. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [11].Shao P et al. , “Spatially-resolved Brillouin spectroscopy reveals biomechanical abnormalities in mild to advanced keratoconus in vivo,” Scientific reports, vol. 9, pp. 7467, 2019. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [12].Wang S et al. , “A focused air-pulse system for optical-coherence-tomography-based measurements of tissue elasticity,” Laser Physics Letters, vol. 10, no. 7, pp. 075605, 2013. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [13].Lan G et al. , “In vivo human corneal shear-wave optical coherence elastography,” Optometry and Vision Science, vol. 98, no. 1, pp. 58, 2021. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [14].Weng C-C et al. , “High frequency ultrasound elastography for estimating the viscoelastic properties of the cornea using lamb wave model,” IEEE Transactions on Biomedical Engineering, vol. 68, no. 9, pp. 2637–2644, 2020. [DOI] [PubMed] [Google Scholar]
- [15].Chen P-Y et al. , “High-resolution shear wave imaging of the human cornea using a dual-element transducer,” Sensors, vol. 18, no. 12, pp. 4244, 2018. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [16].Ambroziński Ł et al. , “Acoustic micro-tapping for non-contact 4D imaging of tissue elasticity,” Scientific reports, vol. 6, pp. 38967, 2016. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [17].Zvietcovich F et al. , “Confocal air-coupled ultrasonic optical coherence elastography probe for quantitative biomechanics,” Optics Letters, vol. 45, no. 23, pp. 6567–6570, 2020. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [18].Zvietcovich F, and Larin KV, “Wave-based optical coherence elastography: the 10-year perspective,” Progress in Biomedical Engineering, vol. 4, no. 1, pp. 012007, 2022. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [19].Kirby MA et al. , “Optical coherence elastography in ophthalmology,” Journal of biomedical optics, vol. 22, no. 12, pp. 121720, 2017. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [20].Li Y, Moon S et al. , “Ultrahigh-sensitive optical coherence elastography,” Light: Science & Applications, vol. 9, pp. 58, 2020. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [21].Ramier A et al. , “In vivo measurement of shear modulus of the human cornea using optical coherence elastography,” Scientific reports, vol. 10, pp. 17366, 2020. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [22].Jin Z et al. , “In vivo noninvasive measurement of spatially resolved corneal elasticity in human eyes using Lamb wave optical coherence elastography,” Journal of Biophotonics, vol. 13, no. 8, pp. e202000104, 2020. [DOI] [PubMed] [Google Scholar]
- [23].Mekonnen T et al. , “Longitudinal assessment of the effect of alkali burns on corneal biomechanical properties using optical coherence elastography,” Journal of Biophotonics, vol. 15, no. 8, pp. e202200022, 2022. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [24].Kirby MA et al. , “Spatial resolution in dynamic optical coherence elastography,” Journal of biomedical optics, vol. 24, no. 9, pp. 096006, 2019. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [25].Feng X et al. , “In vivo stiffness mapping of the human cornea using high-frequency optical coherence elastography.” p. JW2B. 1. [Google Scholar]
- [26].Feng X et al. , “In vivo stiffness measurement of epidermis, dermis, and hypodermis using broadband Rayleigh-wave optical coherence elastography,” Acta Biomaterialia, vol. 146, pp. 295–305, 2022. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [27].Ruberti JW et al. , “Corneal biomechanics and biomaterials,” Annual review of biomedical engineering, vol. 13, pp. 269–295, 2011. [DOI] [PubMed] [Google Scholar]
- [28].Wang S, and Chester SA, “Multi-physics modeling and finite element formulation of corneal UV cross-linking,” Biomechanics and Modeling in Mechanobiology, vol. 20, no. 4, pp. 1561–1578, 2021. [DOI] [PubMed] [Google Scholar]
- [29].Boote C et al. , “Quantification of collagen organization in the peripheral human cornea at micron-scale resolution,” Biophysical journal, vol. 101, no. 1, pp. 33–42, 2011. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [30].Mikula ER et al. , “Measurement of an elasticity map in the human cornea,” Investigative ophthalmology & visual science, vol. 57, no. 7, pp. 3282–3286, 2016. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [31].Blackburn BJ et al. , “A review of structural and biomechanical changes in the cornea in aging, disease, and photochemical crosslinking,” Frontiers in bioengineering and biotechnology, vol. 7, pp. 66, 2019. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [32].Pitre JJ et al. , “Nearly-incompressible transverse isotropy (NITI) of cornea elasticity: model and experiments with acoustic micro-tapping OCE,” Scientific reports, vol. 10, pp. 12983, 2020. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [33].Eltony AM et al. , “Measuring mechanical anisotropy of the cornea with Brillouin microscopy,” Nature communications, vol. 13, pp. 1354, 2022. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [34].Larin KV, and Sampson DD, “Optical coherence elastography–OCT at work in tissue biomechanics,” Biomedical optics express, vol. 8, no. 2, pp. 1172–1202, 2017. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [35].Ramier A et al. , “Measuring mechanical wave speed, dispersion, and viscoelastic modulus of the cornea using optical coherence elastography,” Optics express, vol. 27, no. 12, pp. 16635–16649, 2019. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [36].Han Z et al. , “Optical coherence elastography assessment of corneal viscoelasticity with a modified Rayleigh-Lamb wave model,” Journal of the mechanical behavior of biomedical materials, vol. 66, pp. 87–94, 2017. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [37].Li G-Y, and Cao Y, “Assessing the mechanical properties of anisotropic soft tissues using guided wave elastography: Inverse method and numerical experiments,” The Journal of the Acoustical Society of America, vol. 142, no. 3, pp. 1526–1536, 2017. [DOI] [PubMed] [Google Scholar]
- [38].Li G-Y et al. , “Guided waves in pre-stressed hyperelastic plates and tubes: Application to the ultrasound elastography of thin-walled soft materials,” Journal of the Mechanics and Physics of Solids, vol. 102, pp. 67–79, 2017. [Google Scholar]
- [39].Destrade M, and Saccomandi G, Waves in nonlinear pre-stressed materials: Springer Science & Business Media; 2007. [Google Scholar]
- [40].Gasser TC et al. , “Hyperelastic modelling of arterial layers with distributed collagen fibre orientations,” Journal of the royal society interface, vol. 3, no. 6, pp. 15–35, 2006. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [41].Li G-Y, and Cao Y, “Mechanics of ultrasound elastography,” Proceedings of the Royal Society A: Mathematical, Physical and Engineering Sciences, vol. 473, no. 2199, pp. 20160841, 2017. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [42].Zvietcovich F et al. , “Comparative study of shear wave-based elastography techniques in optical coherence tomography,” Journal of biomedical optics, vol. 22, no. 3, pp. 035010–035010, 2017. [DOI] [PubMed] [Google Scholar]
- [43].Doughty MJ, and Zaman ML, “Human corneal thickness and its impact on intraocular pressure measures: a review and meta-analysis approach,” Survey of ophthalmology, vol. 44, no. 5, pp. 367–408, 2000. [DOI] [PubMed] [Google Scholar]
- [44].Mikula E et al. , “Measurement of corneal elasticity with an acoustic radiation force elasticity microscope,” Ultrasound in medicine & biology, vol. 40, no. 7, pp. 1671–1679, 2014. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [45].Pandolfi A, “Cornea modelling,” Eye and Vision, vol. 7, pp. 2, 2020. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [46].Petsche SJ et al. , “Depth-dependent transverse shear properties of the human corneal stroma,” Investigative ophthalmology & visual science, vol. 53, no. 2, pp. 873–880, 2012. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [47].Li G-Y et al. , “Non-destructive mapping of stress and strain in soft thin films through sound waves,” Communications Physics, vol. 5, no. 1, pp. 231, 2022. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [48].Holzapfel GA et al. , “A new constitutive framework for arterial wall mechanics and a comparative study of material models,” Journal of elasticity and the physical science of solids, vol. 61, pp. 1–48, 2000. [Google Scholar]
- [49].Otténio MDM, and Ogden RW, “Acoustic waves at the interface of a pre-stressed incompressible elastic solid and a viscous fluid,” Int. J. Non. Linear. Mech, vol. 42, pp. 310–320, 2007. [Google Scholar]
- [50].Sit AJ et al. , “In Vivo Non-Invasive Measurement of Young’s Modulus of Elasticity in Human Eyes: A Feasibility Study,” Journal of glaucoma, vol. 26, no. 11, pp. 967, 2017. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [51].Kling S et al. , “Corneal viscoelastic properties from finite-element analysis of in vivo air-puff deformation,” PloS one, vol. 9, no. 8, pp. e104904, 2014. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [52].Shih PJ et al. , “A corneal elastic dynamic model derived from Scheimpflug imaging technology,” Ophthalmic and Physiological Optics, vol. 35, no. 6, pp. 663–672, 2015. [DOI] [PubMed] [Google Scholar]
- [53].Pye DC, “A clinical method for estimating the modulus of elasticity of the human cornea in vivo,” Plos one, vol. 15, no. 1, pp. e0224824, 2020. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [54].Lam AK et al. , “Repeatability of a novel corneal indentation device for corneal biomechanical measurement,” Ophthalmic and Physiological Optics, vol. 35, no. 4, pp. 455–461, 2015. [DOI] [PubMed] [Google Scholar]
- [55].Feng X et al. , “ Ultra-wideband optical coherence elastography from acoustic to ultrasonic frequencies.” Nature Communications, vol. 14, 4949, 2023. [DOI] [PMC free article] [PubMed] [Google Scholar]