Skip to main content

This is a preprint.

It has not yet been peer reviewed by a journal.

The National Library of Medicine is running a pilot to include preprints that result from research funded by NIH in PMC and PubMed.

ArXiv logoLink to ArXiv
[Preprint]. 2024 May 6:arXiv:2405.03147v1. [Version 1]

Data Format Standardization and DICOM Integration for Hyperpolarized 13C MRI

Ernesto Diaz 1, Renuka Sriram 1, Jeremy W Gordon 1, Avantika Sinha 1, Xiaoxi Liu 1, Sule Sahin 1, Jason Crane 1, Marram P Olson 1, Hsin-Yu Chen 1, Jenna Bernard 1, Daniel B Vigneron 1,2, Zhen Jane Wang 1, Duan Xu 1,2, Peder E Z Larson 1,2
PMCID: PMC11100919  PMID: 38764595

Abstract

Hyperpolarized (HP) 13C MRI has shown promise as a valuable modality for in vivo measurements of metabolism and is currently in human trials at 15 research sites worldwide. With this growth it is important to adopt standardized data storage practices as it will allow sites to meaningfully compare data.

In this paper we (1) describe data that we believe should be stored and (2) demonstrate pipelines and methods that utilize the Digital Imaging and Communications in Medicine (DICOM) standard. This includes proposing a set of minimum set of information that is specific to HP 13C MRI studies. We then show where the majority of these can be fit into existing DICOM Attributes, primarily via the “Contrast/Bolus” module.

We also demonstrate pipelines for utilizing DICOM for HP 13C MRI. DICOM is the most common standard for clinical medical image storage and provides the flexibility to accommodate the unique aspects of HP 13C MRI, including the HP agent information but also spectroscopic and metabolite dimensions. The pipelines shown include creating DICOM objects for studies on human and animal imaging systems with various pulse sequences. We also show a python-based method to efficiently modify DICOM objects to incorporate the unique HP 13C MRI information that is not captured by existing pipelines. Moreover, we propose best practices for HP 13C MRI data storage that will support future multi-site trials, research studies and technical developments of this imaging technique.

Keywords: Hyperpolarized 13C MRI, metabolic imaging, DICOM format, Experiment metadata

Introduction

Hyperpolarized (HP) 13C MRI has shown promise as a valuable modality particularly for in vivo measurements of metabolism1, and is currently in human trials at 15 research sites worldwide with over 75 human research publications2. It is based on intravenous injection of a hyperpolarized contrast agent that has been enriched with 13C. The most widely used agent is 13C-pyruvate for measuring metabolism, but there are also a range of other promising agents undergoing clinical translation including 13C-urea (perfusion)3, 13C-alpha-ketoglutarate (metabolism)4, and 13C-fumarate (necrosis)5. To date, human studies have been performed to characterize normal metabolism, and clinical studies have been performed to study in altered metabolism in prostate cancer, brain tumors, kidney tumors, pancreatic cancer, metastatic disease, liver disease, and heart disease2.

As HP 13C MRI continues to grow, it is important to adopt standardized data storage practices, allowing sites to compare data. The Digital Imaging and Communications in Medicine (DICOM) format is an attractive standard format as it is commonly used in digital imaging, medical imaging, and communications. The DICOM file type consists of the image and the metadata of the image packed into a single file. The information in the metadata is organized as a constant and is standardized by a series of data elements. By extracting the data elements, we can access important information regarding the patient demographics and study parameters that are crucial for study interpretation. The DICOM standard is actively supported and updated. It also supports the interchange of information between computer systems such as picture archiving and communication system (PACS).

The goal of this work is to describe an approach for HP 13C MRI data standardization utilizing the DICOM format. We first describe the data requirements, focusing on the unique aspects of HP 13C MRI. We then describe current processing pipelines that support the creation of DICOM objects from multiple vendors and pulse sequences. Finally, we demonstrate how this information can be incorporated into DICOM format.

Hyperpolarized 13C MRI Data

The data requirements for HP 13C MRI are outlined briefly below. This includes performing imaging to spatially resolve signals, encoding of multiple metabolites to measure metabolic conversion, performing dynamic measurements to capture the rapid kinetics, and characteristics of the HP agent, all of which that will affect the results and analysis.

Imaging

HP 13C experiments can be spatially resolved using MRI techniques, providing important localization to visualize metabolic activity within the body. Therefore, a format to store multi-dimensional data needs to be supported.

Metabolite encoding

HP 13C MRI has the unique ability to encode signals from multiple metabolites, distinguishing the injected agent, or substrate, from resulting metabolic products that are converted in vivo. The most common combination of metabolites is [1-13C]pyruvate (substrate), [1-13C]lactate (product), [1-13C]alanine (product), and 13C-bicarbonate (product). Thus, the data should support a metabolite encoding dimension. Metabolite encoding in the acquisition is typically done with MR spectroscopy methods, chemical shift encoding (e.g. IDEAL) methods, or metabolite-specific imaging6. The data must support metabolite encoding through identification of the individual compounds and/or the resonant frequencies, or with a spectroscopy dimension.

Dynamic Measurements

In a HP 13C MRI experiment, rapid metabolite kinetics following injection, including perfusion and metabolic conversion, that are typically captured with time-resolved imaging. Thus, the data should support dynamic measurements and record timings of the measurements.

HP 13C Agents

To analyze HP 13C data, it is important to know the characteristics of the HP 13C agent, how it was prepared, and how it was used. Table 1 lists such relevant HP 13C agent information. Thus, the data should support capturing of this information.

Table 1:

Hyperpolarized 13C agent information that we typically keep track of during our studies.

HP 13C Agent Information Purpose/Description
HP 13C agent(s) injected Describe the composition of the injected agents. Note that this can include simultaneous injection of multiple agents
Administration route of agent How was the HP solution delivered
Total Injected Volume How much HP solution was injected in total
Concentration of each HP agent Concentration of each HP agent in HP solution
Start Time of injection Describe the timing of the start of injection, important to know relative to the data acquisition timing
Total injection duration Over how long was the HP solution injected
Polarization What was the measured polarization of the HP agent(s). Typically back-calculated to the time of dissolution.
Polarization Measurement Timing timing of the polarization measurement relative to the data acquisition
Agent Relaxation Rates (e.g. T1) Expected or measured relaxation rates used to calculate polarization
Dissolution Timing Timing of the dissolution of the HP agent, most important to know relative to the data acquisition

DICOM Processing Pipelines

Our current processing pipelines support DICOM creation from multiple vendors and pulse sequences, listed in Table 2. The spectroscopy sequences use an open-source standards-based software framework(SIVIC) for DICOM creation which includes support for DICOM Spectroscopy (7,8).

Table 2:

Currently supported sequences with DICOM creation pipelines in use at our institution in alphabetical order.

Scanner Vendor Sequences Raw Data Tools Output
Bruker Slab-selective MRS, chemical shift imaging (CSI) and Echo-planar spectroscopic imaging MRS Raw File or Dad File SIVIC
MATLAB
MR Spectroscopy DICOMs
Bruker Metabolite-Specific Imaging FID Bruker Paravision
MATLAB
Metabolite image, AUC, kinetic rate DICOMs
GE Healthcare Slab-selective MRS, chemical shift imaging (CSI) and Echo-planar spectroscopic imaging Pfile or ScanArchive SIVIC MR Spectroscopy DICOMs
GE Healthcare Metabolite-specific EPI Pfile or ScanArchive GE Orchestra Toolbox
MATLAB
Metabolite images, AUC ratio and kinetic rate DICOMs
RTHawk Research (Vista.ai) Metabolite-specific spiral and bSSFP RTHawk raw data
Real-time reconstructed DICOM
MATLAB to improve data Metabolite image DICOMs

MR Spectroscopy and Spectroscopic Imaging with SIVIC

MR spectroscopy and spectroscopic imaging (MRS/I) for HP 13C MRI includes data sampling at various time delays to resolve a spectrum and separate 13C-labeled compounds. For processing of MRS/I data, we utilize the libraries in SIVIC for data processing and DICOM creation, including support of the DICOM MRS standard9. This software can reads in multiple vendor MRS raw data files (Bruker, GE, Philips, Siemens, Varian), and it for HP 13C MRI studies on GE Healthcare and Bruker MRI scanners.

  1. Reads in raw data

  2. Performs MR spectroscopy or MR spectroscopic imaging reconstructions

  3. Writes to DICOM MRS file

  4. Optional – extract peak amplitudes of areas from spectra and create individual metabolite DICOM files

  5. Optional - write raw data or partially processed data to DICOM

  6. Optional - secondary capture of MRS and anatomical visualization to DICOM

GE Metabolite-specific Imaging

Metabolite-specific imaging for HP 13C MRI uses a spectral-spatial RF pulse to selectively excite a single metabolite10 and is followed by fast imaging readout such as echo-planar imaging (EPI) or spirals. We have implemented EPI-based metabolite-specific imaging for GE Healthcare MRI scanners along with a reconstruction pipeline based on the GE Orchestra reconstruction toolbox11. This leverages the EPI processing and DICOM creation functionality that is utilized by GE product pulse sequences.

  1. Reads in raw data (Pfile or ScanArchive)

  2. Perform EPI reconstruction including Nyquist ghost correction

  3. Perform HP-specific coil combination12

  4. Create area-under-curve (AUC) and kinetic rate maps

  5. Write metabolite images and parameter maps to DICOM

RTHawk Metabolite-specific Imaging

We have implemented metabolite-specific imaging with EPI and spirals, as well as metabolite-specific balanced steady-state free-precession (bSSFP) sequences1315 on the RTHawk Research platform. This is a vendor-neutral platform that allows for advanced control of the MRI system including real-time reconstructions and feedback as well as pulse sequence programming.

  1. We utilize real-time reconstruction built into RTHawk during the scanning process, resulting in individual metabolite DICOM images at each timepoint.(Note: This uses the sum of squares coil combination technique.)

  2. The DICOMs Images are exported when the data is transferred from the RTHawk workstation to the local network.

  3. A second, improved image reconstruction is performed using raw complex-valued data and a more advanced coil combination method to enhance the data quality9.

  4. To integrate the improved data into DICOM, a MATLAB script is employed that reads the metadata from the real-time reconstructed DICOMs Images, replaces the original voxel data with the results of the second, optimized reconstruction, and creates new DICOMs Images with appropriate new unique identifiers (UIDs).

Bruker Metabolite-specific Imaging

Similar to the GE metabolite-specific imaging sequence, it uses a spectral-spatial RF pulse to excite a single metabolite10 , and is followed by a fast echo-planar imaging (EPI) readout16.

  1. Bruker scanner has its own built-in EPI reconstruction and an option to write reconstructed metabolite maps to DICOM files.

  2. A MATLAB script is used to read in the DICOM files. From that data, we generate AUC maps and kinetic rate maps, and write back to DICOM files.

  3. Optional: Reads in Bruker raw data using a MATLAB script, use a custom reconstruction script and write to DICOM.

DICOM HP Agent Integration

Many of the HP agent metadata that are desirable for analyses (Table 1) can be integrated into existing DICOM attributes, particularly within the “Contrast/Bolus” module. However, the above DICOM pipelines by default do not include support for these metadata. To address this, we built a tool to modify HP 13C DICOM objects based on user input to include this HP agent metadata.

Currently, the HP 13C images created by MRI scanners or in-house reconstruction pipelines do not utilize the Contrast/Bolus Attributes. We seek to leverage the existing Contrast/Bolus Attributes in DICOM to store information to identify the study and experimental parameters that are valuable for HP 13C data analysis by researchers and clinicians. Table 3 lists the attributes we are adding to our images.

Table 3:

DICOM attributes proposed for addition to HP 13C MRI study data. This captures the majority but not all the desired parameters listed in Table 1.Note that the “Contrast/Bolus Ingredient” attributes listed can include more than one entry to support multiple simultaneously polarized HP 13C agents17 such as pyruvate/urea3,14. The proposed 13C naming conventions are in a style that is consistent with the style used for PET in DICOM.

TAG VR Attribute Name 1H MRI Example Values HP 13C Description HP 13C Example Values
(0018,0010) LO Contrast/Bolus Agent CC
MAGNEVIST
ORAL & OMNIPAQUE
HP 13C agent(s) injected HYPERPOLARIZED [1–13C]PYRUVATE
HYPERPOLARIZED [2–13C]PYRUVATE
HYPERPOLARIZED [1–13C]PYRUVATE + [13C,15N]-UREA
(0018,1040) LO Contrast/Bolus Route Administration route of contrast agent IV
(0018,1041) DS Contrast/Bolus Volume Injected volume (mL)
(0018,1042) TM Contrast/Bolus Start Time Define start of injection relative to data acquisition (Not DICOM compliant)
Start Time in HHMMSS.FFFFFF (DICOM compliant)
(0018,1047) DS Contrast/Bolus Flow Duration Total injection duration [s]
(0018,1048) CS Contrast/Bolus Ingredient IODINE
BARIUM
GADOLINIUM
CARBON
DIOXIDE
13C enriched compounds [1-^13^C]PYRUVATE
[2-^13^C]PYRUVATE
[^13^C,^15AN]UREA
[1-^13^C]ALPHA-
KETOGLUTARATE
(0018,1049) DS Contrast/Bolus Ingredient Concentration Concentration of each 13C compound [mg/mL]
(0400,0550) SQ Modified Attributes Sequence Store prior versions of an attributes that were removed or modified
(0400,0562) DT Attribute Modification DateTime Tracked Date and Time when attributes were modified
(0400,0563) LO Modifying System Describes what modified the attributes SIVIC-HP_agent_DICOM_tool.py

In DICOM, each attribute or data element is identified by a unique tag (TAG), which is comprised of a group number and an element number. The value representation (VR) is a code that indicates the data type used to encode the value(s) of the attribute. Additionally, the value multiplicity (VM) indicates how many values can be present in the attribute. These three components, work together to provide a comprehensive understanding of the attribute and its role within the DICOM file, allowing for accurate and efficient manipulation and interpretation of the medical imaging data.

We reviewed the DICOM standards definition and found the listed attributes in Table 3 can capture much of the unique HP agent metadata. This includes ingredients, injection timing, and dosing characteristics. These can all be added in a DICOM compliant fashion.

One important feature captured in this proposal is the inclusion of multiple ingredients, which can capture dual-agent (“co-polarized”) or multi-agent HP studies3,14,17. In these studies, multiple 13C-enriched compounds are simultaneously hyperpolarized and simultaneously injected. The “Contrast/Bolus Agent Sequence”, which can capture use of multiple agents that affect the data in a DICOM object, was not used since when multiple HP compounds are desired, they are combined into a single agent with multiple ingredients. The use of multiple agents is not performed to our knowledge, as it would require rapid dissolution of multiple separate samples following hyperpolarized to inject both before the signal decays.

We also proposed standardized ingredient names that are consistent with the style used in PET. In these, the enriched isotope(s) are denoted with “^” around the nucleus, e.g. “^13^”. We also use a standard chemical naming convention to identify the site of enrichment within the molecule.

We used the PyDicom python toolbox to access the data in DICOM files (Figure 2). PyDicom is a general-purpose DICOM framework whose purpose is to reads and write DICOM attributes. It can also add, delete, and modify any attributes to DICOM objects.

Figure 2:

Figure 2:

This is an example of the PyDicom functions used. It reads a file, then adds an attribute and saves the file.

The basic workflow of the python script created using the packages PyDicom and OS is as follows: User would run the script in the terminal with a parameter of a directory path for DICOMS that need to be modified. The prompts would pop up one by one and the user can write its own inputs or press enter for default values as shown in Figure 3. Once the user has entered the prompts, it will loop through the folder and add all the attributes to each DICOM file.

We have also aimed to capture the modifications made to the DICOM file when the tool is used. For this, we used the Modified Attributes Sequence to describe what attributes we modified. For Modifying System, it shows the name of this tool we used to change the attributes. We used DateTime within the Modified Attributes Sequence to keep track of the date when we modified the attributes, and this shows other researchers when we modified them. This maybe a non-standard use of these attributes.

This tool has been demonstrated on DICOM objects from a metabolite-specific EPI sequence18 on a GE 3T scanner, reconstructed with a custom implementation based on the GE Orchestra package, DICOM from Bruker 2D CSI scan generated by SIVIC, and a DICOM from a GE echo-planar spectroscopic imaging (EPSI) sequence generated by SIVIC.

The script is available within SIVIC at https://github.com/SIVICLab/sivic/tree/master/applications/dicom_tools.

HP 13C MRI Data Storage Best Practices

We have also created a proposed set of Data Storage Best Practices. These were based upon a survey of research groups at our institution, performing both human and preclinical HP 13C imaging studies at our institution. The proposed set of Data Storage Best Practices aim to ensure that the stored data completely describes the study as is necessary for subsequent processing. This should also be done so that future researchers can easily retrieve and analyze the data without the need to search multiple sources of information.

Storage of essential study data should be done in a centralized location with high-quality back up, using a single folder per study, including:

  • HP spectroscopy/image data (DICOM recommended)

  • HP raw data

  • 1H image data (DICOM recommended)

  • Copy of any processing scripts

  • Metadata – any other information about the study. This can either be directly noted in the data folder, or in a separate database/spreadsheet. At a minimum this should include
    • Image acquisition data (ideally in DICOM files):
    • Flip angle information, especially for variable flip angle or metabolite-specific flip schedules
    • Injection timing characteristics – length of injection, timing of imaging relative to injection(See Table 1)
    • Dose characteristics – volume injected, concentration of agent, subject weight, polarization.(See Table 1)
  • Other Study data - directly recorded in the study folder, or in a separate database/spreadsheet.

  • Study “Key” needed in all data locations to link to other metadata sources.
    • Option 1: Unique ID for each study
    • Option 2: Unique ID for each subject + Unique ID for each study

Discussion and Unmet Needs

Using the DICOM standard has significant advantages for HP 13C MRI data. Given this technology is being translated into clinical studies, DICOM will allow for seamless integration into existing image viewing and analysis platforms, ultimately supporting use in clinical workflows. The use of standardized DICOM objects would also be valuable in the multi-site clinical trials setting, which are likely to begin within the next several years.

We also believe that the DICOM format will be valuable for the preclinical studies given the flexibility of the format. A pre-clinical DICOM standard has recently been proposed19, and a recent assessment preclinical imaging metadata need has also advocated for integration into DICOM20. While adoption is less consistent in preclinical imaging systems, we have demonstrated a pipeline for our Bruker preclinical MRI scanner.

While the Contrast/Bolus module in DICOM supported most HP agent parameters we wanted to save, there were several parameters without an existing attribute. These include the polarization measurements, T1 relaxation rate, and dissolution timing. These could be stored in private data elements. This could also justify the need for a HP Agent MRI Module in DICOM. This could largely follow the examples from PET, which also must capture similar information: polarization is analogous to specific activity, relaxation rates are analogous to half-life, and dissolution timing is analogous to dose creation timing. The PET Isotope Module (C.8.9.2) could serve as a good model.

Several other current features of the DICOM standard may also be relevant for HP MRI. The Enhanced MR Information Object Definitions (IODs) are attractive because they allow for more MRI specific data sources such as raw k-space and spectroscopy data and can be used for “raw” images (e.g. dynamic metabolite images) as well as derived parameter maps (e.g. kinetic rate maps). The Enhanced Contrast/Bolus Module may also be relevant to store additional HP agent parameters.

Hyperpolarization is also used in HP 129Xe MRI, a modality that unique images pulmonary ventilation, gas exchange, and terminal airway morphology22 and received FDA approval in 2021. There are many shared aspects of both HP MRI methods, including the hyperpolarization of an agent, administration of this agent (inhaled for HP 129Xe), relatively rapid imaging of the agent, and measuring agent kinetics. With this in mind, we believe it would be beneficial to coordinate efforts across the HP modalities when developing DICOM integration methods and specifications of metadata.

HP 13C MRS and MRSI acquisition methods are more challenging to use in a standardized fashion compared to imaging-based methods (e.g. metabolite-specific imaging). This is because the adoption of DICOM MRS by vendors is rather limited, leading to the community development of tools such as SIVIC8. If MRS and MRSI methods continue to show value, it would be important for researchers to work closely with vendor partners and ideally have the DICOM MRS standard integrated into the scanners.

Ultimately, we believe it is important that both a standardized and essential set of DICOM attributes are recorded for HP 13C MRI studies. This should be standardized so as not to introduce confusion, for example due to different naming conventions or use of different DICOM attributes. This should also include an essential set of attributes to allow for robust and reproducible analyses of HP 13C MRI data. The final definitions are beyond the scope of this paper, but we think would be best addressed by a consensus of HP 13C MRI researchers in future discussions as the community recently formed the HP 13C Consensus Group2. Following consensus, any changes to the DICOM standard should then be proposed by drafting a Correction Proposal (CP) and work with the DICOM MR Working Group (WG 16) to submit it to the base standards WG (WG 6).

Conclusion

We have proposed a set of attributes to store HP 13C MRI metadata, particularly regarding the HP agent, and implemented via PyDicom a program to add these a set of these attributes in a DICOM-compliant fashion based on user input. This was shown to work in line with a variety of DICOM creation pipelines that are used for HP 13C data. In addition, we have proposed a broader set of data storage best practices. Providing further data standardization will advance this technology by allowing sites to seamlessly share and compare data, including in the context of multi-site trials, and by using DICOM HP 13C MRI can be easily integrated into other clinical and research workflows.

Figure 1:

Figure 1:

Example of how HP 13C MRI can be used in a workflow for improved targeting of biopsies in prostate cancer11. Using DICOM has been critical in the success of this workflow, as this allows for integration into PACS as well as the fusion biopsy platform, with radiologists and urologists easily able to review the images and resulting annotations outlining the biopsy targets. Figure reproduced with permission from11.

Figure 4:

Figure 4:

The prompts questions that gives option for default values or user input. As well, showing the attributes that has been modified.

Funding

This work was supported by NIH grants P41EB013598, U24CA253377, R01CA262630, and R01CA249909. Authors PEZL, DBV, and JWG received research support from GE Healthcare.

Competing Interests

Financial interests: Authors ED, RS, AS, XL, JB, ZJW, and XU declare they have no financial interests. Authors PEZ, DBV, and JWG received research funding from GE Healthcare

Funding Statement

This work was supported by NIH grants P41EB013598, U24CA253377, R01CA262630, and R01CA249909. Authors PEZL, DBV, and JWG received research support from GE Healthcare.

Footnotes

Ethics approval

This manuscript presents technical methods but does not include any in vivo study results.

References

  • 1.Wang ZJ, Ohliger MA, Larson PEZ, et al. Hyperpolarized 13C MRI: State of the Art and Future Directions. Radiology. Published online March 5, 2019:182391. doi: 10.1148/radiol.2019182391 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2.Larson PE, Bernard JM, Bankson JA, et al. Current Methods for Hyperpolarized [1–13C]pyruvate MRI Human Studies. Published online September 7, 2023. doi: 10.48550/arXiv.2309.04040 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Qin H, Tang S, Riselli AM, et al. Clinical translation of hyperpolarized 13C pyruvate and urea MRI for simultaneous metabolic and perfusion imaging. Magnetic Resonance in Medicine. n/a(n/a). doi: 10.1002/mrm.28965 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.Chaumeil MM, Larson PEZ, Yoshihara HAI, et al. Non-invasive in vivo assessment of IDH1 mutational status in glioma. Nat Commun. 2013;4:2429. doi: 10.1038/ncomms3429 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Clatworthy MR, Kettunen MI, Hu DE, et al. Magnetic resonance imaging with hyperpolarized [1,4-(13)C2]fumarate allows detection of early renal acute tubular necrosis. Proc Natl Acad Sci U S A. 2012;109(33):13374–13379. doi: 10.1073/pnas.1205539109 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Gordon JW, Chen H, Dwork N, Tang S, Larson PEZ. Fast Imaging for Hyperpolarized MR Metabolic Imaging. Magnetic Resonance Imaging. 2021;53(3):686–702. doi: 10.1002/jmri.27070 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.Crane JC, Olson MP, Nelson SJ. SIVIC: Open-Source, Standards-Based Software for DICOM MR Spectroscopy Workflows. International Journal of Biomedical Imaging. 2013;2013:e169526. doi: 10.1155/2013/169526 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.SIVIC. doi: 10.5281/zenodo.4777197 [DOI] [Google Scholar]
  • 9.Crane JC, Gordon JW, Chen HY, et al. Hyperpolarized 13C MRI data acquisition and analysis in prostate and brain at University of California, San Francisco. NMR in Biomedicine. 2020;n/a(n/a):e4280. doi: 10.1002/nbm.4280 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10.Cunningham CH, Chen AP, Lustig M, et al. Pulse sequence for dynamic volumetric imaging of hyperpolarized metabolic products. J Magn Reson. 2008;193(1):139–146. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Chen HY, Bok RA, Cooperberg MR, et al. Improving multiparametric MR-transrectal ultrasound guided fusion prostate biopsies with hyperpolarized 13 C pyruvate metabolic imaging: A technical development study. Magn Reson Med. 2022;88(6):2609–2620. doi: 10.1002/mrm.29399 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12.Zhu Z, Zhu X, Ohliger MA, et al. Coil combination methods for multi-channel hyperpolarized 13C imaging data from human studies. Journal of Magnetic Resonance. 2019;301:73–79. doi: 10.1016/j.jmr.2019.01.015 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13.Tang S, Bok R, Qin H, et al. A metabolite-specific 3D stack-of-spiral bSSFP sequence for improved lactate imaging in hyperpolarized [1–13C]pyruvate studies on a 3T clinical scanner. Magnetic Resonance in Medicine. n/a(n/a). doi: 10.1002/mrm.28204 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14.Liu X, Tang S, Mu C, et al. Development of specialized magnetic resonance acquisition techniques for human hyperpolarized [13 C,15 N2]urea + [1–13 C]pyruvate simultaneous perfusion and metabolic imaging. Magn Reson Med. 2022;88(3):1039–1054. doi: 10.1002/mrm.29266 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.Liu X, Tang S, Cui D, et al. A metabolite specific 3D stack-of-spirals bSSFP sequence for improved bicarbonate imaging in hyperpolarized [1–13C]Pyruvate MRI. J Magn Reson. 2023;353:107518. doi: 10.1016/j.jmr.2023.107518 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16.Sahin SI, Ji X, Agarwal S, et al. Metabolite-Specific Echo Planar Imaging for Preclinical Studies with Hyperpolarized 13C-Pyruvate MRI. Tomography. 2023;9(2):736–749. doi: 10.3390/tomography9020059 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17.Wilson DM, Keshari KR, Larson PEZ, et al. Multi-compound Polarization by DNP Allows Simultaneous Assessment of Multiple Enzymatic Activities In Vivo. J Magn Reson. 2010;205(1):141–147. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Gordon JW, Vigneron DB, Larson PEZ. Development of a symmetric echo planar imaging framework for clinical translation of rapid dynamic hyperpolarized (13) C imaging. Magn Reson Med. Published online February 2016. doi: 10.1002/mrm.26123 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19.Kalen JD, Clunie DA, Liu Y, Tatum JL, Jacobs PM, Kirby J, Freymann JB, Wagner U,Smith KE, Suloway C, et al. Design and Implementation of the Pre-Clinical DICOM Standard in Multi-Cohort Murine Studies. Tomography. 2021; 7(1):1–9. 10.3390/tomography7010001 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20.Moore SM, Quirk JD, Lassiter AW, Laforest R, Ayers GD, Badea CT, Fedorov AY, Kinahan PE, Holbrook M, Larson PEZ, et al. Co-Clinical Imaging Metadata Information (CIMI) for Cancer Research to Promote Open Science, Standardization, and Reproducibility in Preclinical Imaging. Tomography. 2023; 9(3):995–1009. 10.3390/tomography9030081 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21.Niedbalski PJ, Hall CS, Castro M, et al. Protocols for multi-site trials using hyperpolarized 129Xe MRI for imaging of ventilation, alveolar-airspace size, and gas exchange: A position paper from the 129Xe MRI clinical trials consortium. Magn Reson Med. 2021; 86: 2966–2986. 10.1002/mrm.28985 [DOI] [PubMed] [Google Scholar]

Articles from ArXiv are provided here courtesy of arXiv

RESOURCES