Skip to main content
Scientific Reports logoLink to Scientific Reports
. 2024 May 24;14:11943. doi: 10.1038/s41598-024-61987-3

Periodic open and closed resonators as a biosensor using two computational methods

Zaky A Zaky 1,, M Al-Dossari 2, Ahmed S Hendy 3, Wael M Badawy 4,5, Arafa H Aly 1
PMCID: PMC11126621  PMID: 38789449

Abstract

The volatile particles and molecules in our dry exhaled breath can reveal enormous information about the health of any person, such as the person’s respiratory and metabolic functioning. Beyond the carbon dioxide level is an indicator of life, it provides important health-related data like people’s metabolic rate. This study considers periodic open and closed resonators for measuring carbon dioxide concentration in dry exhaled breath. Transfer matrix and green methods are used to simulate the interaction between acoustic waves and the proposed sensor. The band gaps using the green method coincide with the transmittance spectra by the transfer matrix. The suggested sensor recorded a sensitivity of 5.3Hz.m-1.s, a figure of merit of 10,254 m-1.s, a detection limit of 5×10-6m.s-1, and a quality factor of 3×106. Furthermore, the efficiency shows that the proposed design is appropriate as a diagnostic sensor for different diseases such as chronic obstructive pulmonary. Besides, cylindrical-adapted sensors are urgently needed in medicine, industry, and biology because they can simultaneously be used for fluid transport and detection.

Keywords: Acoustic waves, Gas sensor, Parallel resonators, Phononic crystal, Carbon dioxide concentration

Subject terms: Biomedical engineering, Computational science, Computational methods, Acoustics

Introduction

Biosensor system detects and quantifies a specific analyte in samples using physicochemical components1,2. A biomaterial can be a specific recognition element for analytes, such as enzymes, antibodies, cells, and tissues3,4. As a result of the recognition event, the transducer undergoes a physical or chemical change, that yields a detectable optical or electrical signal5,6. Many biosensor applications include medical diagnosis, environmental monitoring, and food safety79. There are several applications for biosensors, including diagnostics, food safety, and environmental defense. Detecting diseases like diabetes, cancer, and infectious diseases has been made possible by biosensors in clinical diagnosis. Food toxins and contaminants, such as pesticides and mycotoxins, have been detected by biosensors in the food industry10. As part of environmental monitoring, biosensors have been used to measure pollutants in air, water, and soil1113.

In recent decades, biosensors have been employed to diagnose diseases14. Using conventional check-ups or more sophisticated methods like biosensors is necessary to track changes in an individual's health status over time. Biosensors are widely used to monitor chronic diseases like diabetes, heart disease, cancer, etc. Infectious diseases can also be detected with them15. Due to their unique optical properties, photonic crystals (PCs) have recently gained significant attention for their use as biosensors16. PCs are periodic nanostructures that control light flow by creating a photonic band gap (PBG), which restricts light propagation at specific frequencies. This property enhances the sensitivity and specificity of biosensors.

An enzyme or antibody can be attached to the surface of a PC biosensor to recognize biological elements16. When the target molecule binds to the recognition element at the surface, the refractive index (RI) at the surface changes. This change in RI affects the transmission spectrum of the crystal, allowing for the quantification of the target molecule present in a sample17. Using PCs as RI sensors is one of the most promising applications of PCs in biosensors. The transmission or reflection of light through PCs can be used to measure RI changes in biological samples, such as changes in the concentration of analytes. Biosensors based on this property have been developed for various biological analytes, including proteins, DNA, and small molecules18. In addition, PC biosensors have been demonstrated in several studies to detect biomarkers for cancer19, toxins, and pathogens, as well as food analysis and environmental monitoring16.

In 1988, Tamura et al.20 developed phononic crystals (PnCs) after being inspired by PCs to control how elastic waves propagate in periodic designs. PnCs are manufactured shapes with periodic variations in material properties such as mass, stiffness, tension, etc.21,22. This regular change in acoustic properties leads to a phononic band gap (PnBG) analogous to the PBG in PCs23,24. Within the PnBG, phonons with specific frequencies aren’t allowed to propagate. Because of the capabilities of PnCs to manipulate acoustic spectra, they are used in various applications, such as acoustic shielding25, waveguiding26, and biosensors27. Therefore, the vast majority of these applications strongly depend on the PnBG. For example, Zaky et al.27 investigated a model using regular tubes with different cross-sections for hazardous gas detection. A sensitivity of 2.5495 Hzm-1s, Q-factor of 4077, and FoM of 9.16 m-1s were recorded.

Dry exhaled breath (DEB) primarily comprises oxygen, nitrogen, carbon dioxide, and inert gases28. Other small volatile organic compounds (VOC) ratios comprise the rest of the DEB. VOCs in DEB are either formed by cellular metabolic processes or absorbed/inhaled from environmental sources29. For example, chronic obstructive pulmonary (COP)30, Lung cancer cells28, congestive heart failure (CHF)30, etc., may produce these VOCs. Due to the low solubility of VOCs in blood, they spread quickly into the exhaled breath31.

Detecting carbon dioxide (CO2) concentration in DEB can give non-invasive information about an individual's respiratory and metabolic functioning31. Bhagya et al.30 proposed a capnographic sensor to sense CO2 concentration in DEB based on an acoustic virial equation to diagnose COP and CHF diseases. Oort et al.32 investigated the vital role of analyzing DEB pneumonia detection. Mazzone et al.28 studied the detection of VOC in DEB as a lung cancer biomarker.

The proposed model is unique in the diagnostic field based on detecting CO2 in DEB. Furthermore, the 1D-PnC with open and closed periodic parallel resonators shows promising results in the diagnostic field. Also, the suggested detector is able to be utilized for monitoring the level of dangerous pollutants flowing through industrial chimneys on a continual basis. This type of detector is easier to fabricate than nanosensors, which need sophisticated and costly manufacturing processes. Besides, cylindrically adapted sensors are urgently needed in medicine, industry, and biology because they can be used simultaneously for fluid transport and detection.

Materials and mathematical methods

Figure 1 shows a geometrical representation of the 1D-PnC with open and closed periodic parallel resonators of PVC polymer. The proposed design is composed of a duct with a cross-section of S1, the closed parallel resonators on one side have a cross-section of S2, and the opened parallel resonators on the other side have a cross-section of S3. The length of the duct of each cell is d1, the length of the closed branch is d2, and the length of the opened branch is d3. A defect guide with the same cross-section of the main duct and length of dd is at the center of the 1D-PnC sensor with open and closed periodic parallel resonators. The main duct, closed branches, and opened branches will be filled with a stationary sample of exhaled breath containing different concentrations of CO2. These ducts and branches can be made of aluminium and nylon33.

Figure 1.

Figure 1

Geometrical representation of the 1D-PnC with open and closed periodic parallel resonators of PVC polymer.

This model will neglect the effects of temperature gradients and viscosity by considering the sample is stationary “relatively static” at room temperature34. Firstly, all opened resonators will be closed. Then, the DEB sample will fill the model. After that, the resonators will be opened. As a result, the excess gas will leak through the opened branches until the equilibrium state (the outside pressure equals the inside pressure). Currently, the sample is stationary.

Using the transfer matrix method

As the transfer matrix method (TMM) is easy, fast, and gives the exact dispersion relation of the Green method35, the simulations of the response of 1D-PnC with opened and closed periodic parallel resonators of the proposed model to the acoustic waves will be carried out by the TMM14,36:

Mi=AiBiCiDi10yOC1AiBiCiDi, 1
Ai=coskdi2, 2
Bi=jZisinkdi2, 3
Ci=jZisinkdi2, 4
Di=Ai. 5

where c is the acoustic speed, k=ω/c, ρ is the density, and the impedance is calculated as the following:

Zi=1/Yi=ρc/Si. 6

The total acoustic admittance (yOC) of the opened (yO) and closed (yC) resonators to the acoustic wave are calculated as:

yO=-j1Z3cotkd3, 7
yC=j1Z2tankd2, 8
yOC=yO+yC. 9

For the defect cell:

MD=AdBdCdDd, 10
Ad=coskdd2, 11
Bd=jZdsinkdd2, 12
Cd=jZdsinkdd2, 13
Dd=Ad, 14

where Zd=ρc/Sd. The transmittance (T) of the 1D-PnC with open and closed periodic parallel resonators is computed as the following:

t=2Y1A11+A12Y1Y1+A21+A22Y1, 15
T%=100×t2. 16

Using Green method

Using Bloch’s theorem, the band structure of 1D-PnC with open and closed periodic parallel resonators was investigated and matched the TMM results35. In the case of 1D-PnC with open and closed periodic parallel resonators, the function of Green’s surface (GS) of an elementary cell gR-10,0 is:

gR-10,0=gc-10,0+go-10,0, 17

where gc-10,0, and go-10,0 are the function of GS of one cell of closed and opened resonator as the following:

gc-10,0=-jy2tankd2, 18
go-10,0=-jy2Cotankd2. 19

The dispersion for the infinite waveguide is:

cosKd=coskd1-12z1ωsinkd1gR-10,0, 20
cosKd=coskd1+j2Z1YOCsinkd1, 21

where K and k are the Bloch and wave vectors, respectively. All results are performed using MATLAB software on a laptop with the configuration: Intel(R) Core (TM) i5 CPU, 16:00 G RAM, and 2:67 GHz, with 64 bits operating system. The elapsed computation time for every run is about 8 s.

Laly et al.37 studied homogeneous and heterogeneous acoustic absorbing metamaterials using TMM and the finite element method (FEM). In their studies, the results using TMM and FEM were coinciding. Almeida et al.38 designed intake mufflers with different cross-sections. Almeida et al. compared the numerical results using TMM and experimental results. The numerical results recorded high matches.

The effective acoustic speed of the mixture of gases (cmix) of different concentrations (α) can be calculated as the following39:

cmix=αiρiCiαiρi, 22

where ρ is the mass density and c is the acoustic speed of each gas sample. The effective mass density of the mixture of gases (ρmix) can be calculated as the following40:

ρmix=αiρi. 23

Results and discussions

At first, the geometrical dimensions of the proposed sensors' main duct, the closed branches, and the opened branches will be N = 10, d1 = 10 cm, d2 = d3  = 6 cm, dd = 33 cm, Sd = S1  = 1 cm2, S2= S3=0.9 cm2. For normal humans, the DEB is composed of a mixture of N2 (78%), O2 (16%), CO2 (5%), and Ar (1%). Table 1 contains c and ρ of pure gas samples to calculate the effective c and ρ of the DEB samples according to Eqs. (22) and (23).

Table 1.

Acoustic speed and density of pure gases23,41.

Gas sample Density (kg/m3) Acoustic speed (m/s)
CO2 1.8393 267
Ar 1.661 319
O2 1.314 326
Air 1.2047 343
N2 1.165 349

As clear in Fig. 2, the acoustic speed and density of the typical DEB are 338.51 m/s and 1.2275 kg/m3, respectively. By increasing the concentration of CO2 higher than the usual ratio in DEB, the c of the mixture regularly decreases, and the ρ of the mixture regularly increases. Figure 3A shows the band structure of a unit cell of the proposed 1D-PnC with open and closed periodic parallel resonators using an air sample. A DEB sample will be pushed inside the structure. The excess in the DEB sample inside the structure will leak through the opened tubes (stationary state). The DEB sample will go through a dryer to dry water and absorb other organic components from the sample, as clear in the graphical abstract.

Figure 2.

Figure 2

The acoustical velocity of the mixture and its mass density versus the ratio of CO2.

Figure 3.

Figure 3

(A) Band structure of a unit cell using the Green method, (B) Transmittance of perfect open and closed periodic parallel resonators using a sample of air, (C) the transmittance of the defected 1D-PnC with open and closed periodic parallel resonators using a sample of air, and (D) Transmittance of defected open and closed periodic parallel resonators using a sample of exhaled breath with different concentrations of CO2.

The band structure using Bloch’s theorem consists of PnBGs that coincide with the PnBGs in the transmittance spectrum using TMM in Fig. 3B. Due to the step change in the acoustic impedance inside the sensor before and at the resonators, multiple Bragg scattering forms the PnBGs. A defect peak penetrates the third PnBG at 1887.59 Hz by including a horizontal defect cell because of the acoustic wave's localization, as evident in Fig. 3C. Figure 3D clarifies that the defect peak has red-shifted from 1862.89 to 1861.71 Hz, 1860.53 Hz, 1859.36 Hz, 1858.19 Hz, and 1857.02 Hz due to the variation of the CO2 concentration from 0 to 100 ppm in the DEB above the normal ratio. According to Table 1 and the standing wave relation, increasing the CO2 concentration decreases the acoustic speed, and the resonant frequency decreases. As a result, the resonant peaks are red-shifted to lower frequencies.

2d=ncfR. 24

The main goal of the following calculations is to select the relatively best conditions. Therefore, the most important parameters to evaluate the quality of sensors are the transmittance of resonance (T), sensitivity (S), the bandwidth of the resonant peak (FWHM), the figure of merit (FoM), quality factor (Q), and detection limit (LoD), as the following42:

S=ΔfRΔc, 25
FoM=SFWHM, 26
Q=fRFWHM, 27
LoD=fR20SQ, 28

where fR is the peak frequency. The signal-to-noise ratio (SNR) and resolution (RS) will be calculated as follows43:

SNR=ΔfRFWHM,
RS=2FWHM3(SNR)0.25.

The resolution of any sensor device reflects the smallest change in the measurable peak that can be detected accurately.

In Fig. 4A–D, T, S, FWHM, Q, FoM, LoD, SNR and RS are calculated versus dd. Transmittance and FWHM of defect mode at a concentration of CO2 of 0 ppm concentration indicates the intensity and bandwidth of peaks, as clear in Fig. 4A. As dd varies from 20 to 150 cm, the transmittance has a value very close to 100%, and FWHM changes from 0.558 to 0.006 Hz. The narrower FWHM of 0.003 Hz is achieved at dd = 100 cm. The sensitivity is studied based on the change of fR with the change c of CO2 concentration from 0 to 100 ppm, as shown in Fig. 4B. At the thickness dd = 100 cm, the resonant peak has a high transmittance of 99.6% and the lowest FWHM of 0.003 Hz. Besides, the FoM, Q-factor, and SNR have the highest value, but LoD and RS have the lowest value. So, dd = 100 cm is the selected thickness.

Figure 4.

Figure 4

(A) Transmittance at 0 ppm (left axis) and FWHM (right axis), (B) Sensitivity (left axis) and FoM (right axis), (C) Q-factor (left axis) and LoD (right axis), and (D) SNR (left axis) and RS (right axis) versus the length of dd.

In Fig. 5A, at d1 of 10 cm, 15 cm, and 25 cm, the transmittance becomes 100%. But at d1 of 20 cm and 30 cm, the transmittance becomes 71% and 5%, respectively. Moreover, the FWHM of the resonant peak is also given on the right axis of Fig. 5A. At d1 of 20 cm, the resonant peak has the narrowest FWHM of 0.0005 Hz. As explicit in Fig. 5B–D, the FoM, Q-factor, and SNR have the highest performance of 10,254 m-1.s, 3,470,288, and 10,930, respectively, at d1 of 20 cm, but LoD and RS have the lowest value of 5 ×10-3m.s-1, and 3 ×10-5. For sensitivity, the highest values are achieved at d1 of 10 cm and then 20 cm. As a result, d1 of 20 cm will be selected.

Figure 5.

Figure 5

(A) Transmittance at 0 ppm (left axis) and FWHM (right axis), (B) Sensitivity (left axis) and FoM (right axis), (C) Q-factor (left axis) and LoD (right axis), and (D) SNR (left axis) and RS (right axis) versus the length of d1.

From Fig. 6A, the transmittance has maximum values at d2 of 5.8 cm, 6.5 cm, and 7.0 cm. But transmittance of 71% is recorded at d2 of 6.0 cm. At thicknesses of d2 higher or lower than these selected values, the resonant mode disappears. At d2 of 6.0 cm, the resonant peak has the narrowest FWHM of 0.0005 Hz. As explicit in Fig. 6B–D, the FoM, Q-factor, and SNR have the highest performance of 10,254 m-1.s, 3,470,288, and 10,931, respectively, at d2 of 6.0 cm, but LoD and RS have the lowest value of 5×10-6 m.s-1, and 2×10-3. For sensitivity, the highest values are achieved at d2 of 5.8 cm and then 6.0 cm. As a result, d2 of 6.0 cm will be selected.

Figure 6.

Figure 6

(A) Transmittance at 0 ppm (left axis) and FWHM (right axis), (B) Sensitivity (left axis) and FoM (right axis), (C) Q-factor (left axis) and LoD (right axis), and (D) SNR (left axis) and RS (right axis) versus the length of d2.

Transmittance, FWHM, Q, S, FoM, LoD, SNR, and RS are studied versus the length of d3, as clear in Fig. 7A–D. As the thickness of d3 varies from 5.5 to 6.5 cm, the transmittance decreases from 100 to 42%. By increasing d3 from 6.5 to 7 cm and 7.5 cm, the transmittance increases to 100%. The FWHM decreases from 0.0182 to 0.00052 Hz by changing d3 from 5.5 to 6.0 cm. But the FWHM increases from 0.00052 to 0.02771 Hz by changing d3 from 6.0 to 7.5 cm, as apparent in Fig. 7A. Increasing the d3 from 5.5 to 6.0 cm decreases sensitivity from 5.6 to 5.3 Hz.m-1.s. Then, it slightly changes with increasing d3, as clear in Fig. 7B. On the other axis, FoM changes from 309 m-1.s to 10,254 m-1.s, 2310 m-1.s, 1372 m-1.s, and 190 m-1.s, by increasing d3 from 5.5 cm to 6.0 cm, 6.5 cm, 7.0 cm, and 7.5 cm, respectively. As clear in Fig. 7C, the Q-factor changes from 104,563 to 3,470,288, 782,033, 464,169, and 64,447 by increasing d3 from 5.5 cm to 6.0 cm, 6.5 cm, 7.0 cm, and 7.5 cm, respectively. On the other axis, the LoD decreases from 1.6×10-5 to 4.8 ×10-6 m.s-1 by changing d3 from 5.5 to 6.0 cm. But the LoD increases from 4.8 ×10-6 to 2.6 ×10-4m.s-1 by changing d3 from 6.0 to 7.5 cm, as apparent in Fig. 7A. At the thickness d3 = 6.0 cm, the resonant peak has a slightly high transmittance of 71% and the lowest FWHM of 0.0005 Hz. Besides FoM, sensitivity, Q-factor, and SNR have the highest value, but LoD and RS have the lowest value, as clear in Fig. 7B–D. So, d3 = 6.0 cm is the selected thickness.

Figure 7.

Figure 7

(A) Transmittance at 0 ppm (left axis) and FWHM (right axis), (B) Sensitivity (left axis) and FoM (right axis), (C) Q-factor (left axis) and LoD (right axis), and (D) SNR (left axis) and RS (right axis) versus the length of d3.

According to Fig. 8A, by changing Sd from 0.90 cm2 to 0.95 cm2, 1.00 cm2, 1.05 cm2, and 1.10 cm2, the transmittance of resonant frequencies alters from 62 to 13%, 71%, 78%, and 98%, respectively. At the cross-section of Sd of 1.00 cm2, the resonant peak has the narrowest FWHM of 0.0005 Hz. As clear in Fig. 8B–D, the FoM, Q-factor, and SNR have the highest performance of 10,254 m-1s, 3,470,288, and 10,931, respectively, at Sd of 1.00 cm2, but LoD and RS have the lowest value of 5 ×10-6 m.s-1, and 3.4 ×10-5. For sensitivity, by altering S1 from 0.90 to 0.95 cm2, the sensitivity has not changed. As a result, Sd of 1.00 cm2 will be selected.

Figure 8.

Figure 8

(A) Transmittance at 0 ppm (left axis) and FWHM (right axis), (B) Sensitivity (left axis) and FoM (right axis), (C) Q-factor (left axis) and LoD (right axis), and (D) SNR (left axis) and RS (right axis) versus Sd.

In Fig. 9A, by altering the cross-section of S1 from 0.90 cm2 to 0.95 cm2, 1.00 cm2, 1.05 cm2, and 1.10 cm2, the transmittance of resonant frequencies alters from 33 to 8%, 71%, 27%, and 48%, respectively. At the cross-section of S1 of 1.00 cm2, the resonant peak has the narrowest FWHM of 0.0005 Hz. As clear in Fig. 9B–D, the FoM, Q-factor, and SNR have the highest performance of 10,254 m-1s, 3,470,288, and 10,931, respectively, at S1 of 1.00 cm2, but LoD and RS have the lowest value of 5 ×10-6 m.s-1 and 3 ×10-5. For sensitivity, by altering S1 from 0.90 to 0.95 cm2, S slightly increased from 5.31 Hz.m-1.s to 5.32 Hz.m-1.s. Then, the sensitivity does not change. As a result, S1 of 1.00 cm2 will be selected.

Figure 9.

Figure 9

(A) Transmittance at 0 ppm (left axis) and FWHM (right axis), (B) Sensitivity (left axis) and FoM (right axis), (C) Q-factor (left axis) and LoD (right axis), and (D) SNR (left axis) and RS (right axis) versus S1.

In Fig. 10A, at S2 of 0.7 cm2, 1.0 cm2, and 1.1 cm2, the transmittance becomes 100%. But at S2 of 0.8 cm2 and 0.9 cm2, the transmittance becomes 48% and 71%, respectively. Moreover, the FWHM of the resonant peak is also given on the right axis of Fig. 10A. At S2 of 0.9 cm2, the resonant peak has the narrowest FWHM of 0.0005 Hz. As clear in Fig. 10B–D, the FoM, Q-factor, and SNR have the highest performance of 10,254 m-1s, 3,470,288, and 10,931, respectively, at S2 of 0.9 cm2, but LoD and RS have the lowest value of 5 ×10-6 m.s-1, and 3 ×10-5. For sensitivity, the highest values are achieved at S2 of 1.1 cm2. As a result, S2 of 0.9 cm2 will be selected.

Figure 10.

Figure 10

(A) Transmittance at 0 ppm (left axis) and FWHM (right axis), (B) Sensitivity (left axis) and FoM (right axis), (C) Q-factor (left axis) and LoD (right axis), and (D) SNR (left axis) and RS (right axis) versus S2.

As the cross-section of S3 varies from 0.7 to 0.8 cm2, the transmittance increases from 7 to 100%. Then, the transmittance decreases to 71%, 59%, and 17% by increasing S3 to 0.9 cm2, 1.0 cm2, and 1.1 cm2. The FWHM reduces from 0.00123 to 0.00052 Hz by changing S3 from 0.7 cm2 to 0.9 cm2. But the FWHM grew from 0.00052 to 0.00222 Hz by changing S3 from 0.9 to 1.1 cm2, as apparent in Fig. 11A. In Fig. 11B, sensitivity did not change with increasing the S3 from 0.7 to 1.1 cm2.

Figure 11.

Figure 11

(A) Transmittance at 0 ppm (left axis) and FWHM (right axis), (B) Sensitivity (left axis) and FoM (right axis), and (C) Q-factor (left axis) and LoD (right axis), and (D) SNR (left axis) and RS (right axis) versus S3.

As for FoM increases from 4300 to 10,254 m-1.s by changing S3 from 0.7 to 0.9 cm2, But it decreases from 10,254 to 2398 m-1.s by changing S3 from 0.9 to 1.1 cm2. In Fig. 11C, the Q-factor increases from 1,456,134 to 3,470,288 by changing S3 from 0.7 to 0.9 cm2, But it decreases from 3,470,288 to 811,123 by changing S3 from 0.9 to 1.1 cm2. As for LoD decreases from 1.16 ×10-5to 4.88 ×10-6 m.s-1 by changing S3 from 0.7 to 0.9 cm2, But it increases from 4.88 ×10-6 to 2.08 ×10-5 m.s-1 by changing S3 from 0.9 to 1.1 cm2. As for SNR, it increases from 4584 to 10,931 by changing S3 from 0.7 to 0.9 cm2, But it decreases from 10,931 to 2557 by changing S3 from 0.9 to 1.1 cm2. On the other axis, the RS records the lowest value of 3 ×10-5 at S3 of 0.9 cm2, as clear in Fig. 11D.

Therefore, at the cross-section S3 = 0.9 cm2, the resonant peak has a slightly high transmittance of 71% and the lowest FWHM of 0.0005 Hz. Besides, the Q-factor and FoM have the highest value, and LoD has the lowest value. So, S3 = 0.9 cm2 is the selected cross-section of S3.

Besides, a linear fitting is performed, as clear in Fig. 12, between the CO2 concentration (α) inside the DEB sample and fR. According to (27), the resonant frequency decreases linearly with increasing the concentration of CO2. As a result, the concentration of CO2 within DEB can easily be calculated by determining the position of the resonant frequency.

fR=-0.0567α+1800. 29

Figure 12.

Figure 12

Resonant frequency of localized peak versus CO2 concentration.

Table 2 compares the results using this model (open and closed periodic parallel resonators) and a model using regular tubes with different cross-sections27. Using open and closed parallel resonators enhanced the efficiency of the gas detector. Even though the sensitivity of open and closed parallel resonators is close to the sensitivity of closed parallel resonators and branched open resonators separately, the FoM of the current study is seventy-three times higher than the FoM of the branched open resonators. Besides, the FoM of the current study is thirty times higher than the FoM of the closed parallel resonators. Also, the Q-factor of the current study is seven 100-fold higher than the Q-factor of the branched open resonators, and 30-fold higher than the Q-factor of the closed parallel resonators. Besides, the combination of open and closed resonators enhanced the linearity, as clear in Fig. 11B in Ref. 27 and Fig. 9B in Ref.45.

Table 2.

Comparison study.

Ref S Hz.s.m–1 Q-factor FoM s.m–1 Structure
2022, Ref.27 2.55 4077 9.16 Regular tubes with different cross-sections
2023, Ref.44 1.58 6790 33.7 Ternary-symmetric periodic tubes
2023, Ref.45 5.8 5000 140 Branched open resonators
2024, Ref.46 4.1 113,962 332 Closed parallel resonators
This study 5.3 3,470,288 10,254 Open and closed parallel resonators

The main issue with the proposed structure is that it may be impractical when considering the size of the overall structure. Thus, we will attempt to decrease the model's size in subsequent work.

Finally, it is undeniable that the experimental transmittance may slightly decrease due to viscous losses, thermal losses, or various relaxation effects47,48. However, the sensor's sensitivity will not change because this effect will be constant, and this loss will not affect the resonant peak shift. Besides, the step width in almost all figures is high because the resonant peaks appear at specific frequencies, not continuously.

Conclusion

This study proposed a novel diagnostic biosensor using a periodic band gap design for detecting CO2 in dry exhaled breath as a biomarker of diseases like diabetes, cancer, and infectious diseases. After optimization of different parameters such as d and S of tubes, the sensitivity, FoM, and Q-factor can reach 5.3 Hz.m-1.s, 10,254 m-1.s, and 3,470,288, respectively. Besides, this model achieved high stability in its sensitivity. The suggested sensor has a high sensitivity and simple design, making it suitable for disease detection.

Acknowledgements

The authors extend their appreciation to the Deanship of Scientific Research at King Khalid University for funding this work through large Groups Project under grant number RGP. 2/46/45.

Author contributions

Z.A. Zaky invented the original idea of the study, implemented the computer code, performed the numerical simulations, analyzed the data, wrote and revised the main manuscript text, leader of the team. M Al-Dossari analyzed the data and discussed the results. A. S. Hendy analyzed the data and discussed the results. W. M. Badawy analyzed the data and discussed the results. A. H. Aly analyzed the data and discussed the results. Finally, all Authors developed the final manuscript.

Funding

The current work was assisted financially to the Dean of Science and Research at King Khalid University via the Large Group Project under grant number RGP. 2/46/45.

Data availability

Requests for materials and data should be addressed to Zaky A. Zaky.

Competing interests

The authors declare no competing interests.

Footnotes

Publisher's note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

References

  • 1.Fang T-Y, Sun X-W, Wen X-D, Li Y-X, Liu X-X, Song T, et al. High-performance phononic crystal sensing structure for acetone solution concentration sensing. Sci. Rep. 2023;13:7057. doi: 10.1038/s41598-023-34226-4. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2.Gueddida A, Zhang V, Carpentier L, Bonhomme J, Bonello B, Pennec Y, et al. Phononic crystal made of silicon ridges on a membrane for liquid sensing. Sensors. 2023;23:2080. doi: 10.3390/s23042080. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Hoyos-Nogués M, Gil F, Mas-Moruno C. Antimicrobial peptides: Powerful biorecognition elements to detect bacteria in biosensing technologies. Molecules. 2018;23:1683. doi: 10.3390/molecules23071683. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.Bereli N, Akgönüllü S, Asliyüce S, Çimen D, Göktürk I, Türkmen D, et al. Molecular imprinting technology for biomimetic assemblies. Hacettepe J. Biol. Chem. 2020;48:575–601. doi: 10.15671/hjbc.801427. [DOI] [Google Scholar]
  • 5.Zaky ZA, Al-Dossari M, Hendy AS, Aly AH. Studying the impact of interface roughness on a layered photonic crystal as a sensor. Physica Scripta. 2023;98(10):105527. doi: 10.1088/1402-4896/acfa4a. [DOI] [Google Scholar]
  • 6.Abdurehman A, Jafer M. Biosensor technology as diagnostic tool for detection of pathogen. J. Microb. Biochem. Technol. 2020;12:439. doi: 10.35248/1948-5948.20.12.439. [DOI] [Google Scholar]
  • 7.Zaky ZA, Hanafy H, Panda A, Pukhrambam PD, Aly AH. Design and analysis of gas sensor using tailorable Fano resonance by coupling between Tamm and defected mode resonance. Plasmonics. 2022;17:2103–2111. doi: 10.1007/s11468-022-01699-4. [DOI] [Google Scholar]
  • 8.Zaky ZA, Al-Dossari M, Zohny EI, Aly AH. Refractive index sensor using Fibonacci sequence of gyroidal graphene and porous silicon based on Tamm plasmon polariton. Opt. Quantum Electron. 2023;55:1–13. doi: 10.1007/s11082-022-04262-x. [DOI] [Google Scholar]
  • 9.Zaky ZA, Amer HA, Suthar B, Aly AH. Gas sensing applications using magnetized cold plasma multilayers. Opt. Quantum Electron. 2022;54:217. doi: 10.1007/s11082-022-03594-y. [DOI] [Google Scholar]
  • 10.Singh S, Kumar V, Dhanjal DS, Datta S, Prasad R, Singh J. Biological biosensors for monitoring and diagnosis. In: Singh J, Vyas A, Wang S, Prasad R, editors. Microbial Biotechnology: Basic Research and Applications. Springer Singapore; 2020. pp. 317–335. [Google Scholar]
  • 11.Zaky, Z. A., Al-Dossari, M., Hendy, A. S., Zayed, M. & Aly, A. H. Gamma radiation detector using Cantor quasi-periodic photonic crystal based on porous silicon doped with polymer. Int. J. Mod. Phys. B. 2450409 (2024).
  • 12.Zaky ZA, Mohaseb MA, Panda A, Amer HA, Farag AM, Kovac J, Pukhrambam PD, Dhasarathan V, Aly AH. Theoretical analysis of porous silicon one-dimensional photonic crystal doped with magnetized cold plasma for hazardous gases sensing applications. Opt. Quantum Electron. 2023;55(7):584. doi: 10.1007/s11082-023-04907-5. [DOI] [Google Scholar]
  • 13.Gavrilaș S, Ursachi CȘ, Perța-Crișan S, Munteanu F-D. Recent trends in biosensors for environmental quality monitoring. Sensors. 2022;22:1513. doi: 10.3390/s22041513. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14.Zaky ZA, Alamri S, Zhaketov V, Aly AH. Refractive index sensor with magnified resonant signal. Sci. Rep. 2022;12:13777. doi: 10.1038/s41598-022-17676-0. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.Tammam MT, Zaky ZA, Sharma A, Matar ZS, Aly A, Mohaseb MA. Defected photonic crystal array using porous GaN as malaria sensor. IOP Conf. Ser. Mater. Sci. Eng. 2021 doi: 10.1088/1757-899X/1171/1/012005. [DOI] [Google Scholar]
  • 16.Inan H, Poyraz M, Inci F, Lifson MA, Baday M, Cunningham BT, et al. Photonic crystals: Emerging biosensors and their promise for point-of-care applications. Chem. Soc. Rev. 2017;46:366–388. doi: 10.1039/C6CS00206D. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17.Gupta A, Singh T, Singh RK, Tiwari A. Numerical analysis of coronavirus detection using photonic crystal fibre-based SPR sensor. Plasmonics. 2023 doi: 10.1007/s11468-022-01761-1. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Nie S, Emory SR. Probing single molecules and single nanoparticles by surface-enhanced Raman scattering. Science. 1997;275:1102–1106. doi: 10.1126/science.275.5303.1102. [DOI] [PubMed] [Google Scholar]
  • 19.Sinibaldi A. Cancer biomarker detection with photonic crystals-based biosensors: An overview. J. Lightw. Technol. 2021;39:3871–3881. doi: 10.1109/JLT.2021.3056225. [DOI] [Google Scholar]
  • 20.Tamura S, Hurley D, Wolfe J. Acoustic-phonon propagation in superlattices. Phys. Rev. B. 1988;38:1427. doi: 10.1103/PhysRevB.38.1427. [DOI] [PubMed] [Google Scholar]
  • 21.Khattab MS, Ben-Ali Y, Barkani J, Yousfi J, Bria D. Induced guided acoustic waves in waveguides and resonators. Mater. Today Proc. 2022;72:3398–3403. doi: 10.1016/j.matpr.2022.07.444. [DOI] [Google Scholar]
  • 22.Khettabi A, Antraoui I. Study of a finite network of one-dimensional periodic expansion chambers by the transfer matrix method and Sylvester theorem. AIP Conf. Proc. 2019 doi: 10.1063/1.5090620. [DOI] [Google Scholar]
  • 23.Imanian H, Noori M, Abbasiyan A. A highly efficient Fabry-Perot based phononic gas sensor. Ultrasonics. 2022;124:106755. doi: 10.1016/j.ultras.2022.106755. [DOI] [PubMed] [Google Scholar]
  • 24.Khattab MS, El Kadmiri I, Ben-Ali Y, Khaled A, Jeffali F, Bria D. Propagation of the acoustic waves in a one-dimensional parallel guides and symmetric/asymmetric resonators. Mater. Today Proc. 2023;72:3319–3325. doi: 10.1016/j.matpr.2022.07.354. [DOI] [Google Scholar]
  • 25.Yu P-L, Cicak K, Kampel N, Tsaturyan Y, Purdy T, Simmonds R, et al. A phononic bandgap shield for high-Q membrane microresonators. Appl. Phys. Lett. 2014;104:023510. doi: 10.1063/1.4862031. [DOI] [Google Scholar]
  • 26.Wang Y, Lee J, Zheng X-Q, Xie Y, Feng PX-L. Hexagonal boron nitride phononic crystal waveguides. ACS Photonics. 2019;6:3225–3232. doi: 10.1021/acsphotonics.9b01094. [DOI] [Google Scholar]
  • 27.Zaky ZA, Alamri S, Zohny EI, Aly AH. Simulation study of gas sensor using periodic phononic crystal tubes to detect hazardous greenhouse gases. Sci. Rep. 2022;12:21553. doi: 10.1038/s41598-022-26079-0. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 28.Mazzone PJ. Analysis of volatile organic compounds in the exhaled breath for the diagnosis of lung cancer. J. Thorac. Oncol. 2008;3:774–780. doi: 10.1097/JTO.0b013e31817c7439. [DOI] [PubMed] [Google Scholar]
  • 29.Moor CC, Oppenheimer JC, Nakshbandi G, Aerts JG, Brinkman P, Maitland-van der Zee A-H, et al. Exhaled breath analysis by use of eNose technology: A novel diagnostic tool for interstitial lung disease. Eur. Respir. J. 2021;57:2002042. doi: 10.1183/13993003.02042-2020. [DOI] [PubMed] [Google Scholar]
  • 30.Bhagya D, Suchetha M. A capnographic sensor using acoustic virial equation for diagnostic applications. IEEE Sens. Lett. 2020;4:1–4. doi: 10.1109/LSENS.2020.3008595. [DOI] [Google Scholar]
  • 31.Scheepers MH, Al-Difaie Z, Brandts L, Peeters A, van Grinsven B, Bouvy ND. Diagnostic performance of electronic noses in cancer diagnoses using exhaled breath: A systematic review and meta-analysis. JAMA Netw. Open. 2022;5:e2219372–e2219372. doi: 10.1001/jamanetworkopen.2022.19372. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 32.van Oort PM, Povoa P, Schnabel R, Dark P, Artigas A, Bergmans DC, et al. The potential role of exhaled breath analysis in the diagnostic process of pneumonia—A systematic review. J. Breath Res. 2018;12:024001. doi: 10.1088/1752-7163/aaa499. [DOI] [PubMed] [Google Scholar]
  • 33.Plisson J, Pelat A, Gautier F, Garcia VR, Bourdon T. Experimental evidence of absolute bandgaps in phononic crystal pipes. Appl. Phys. Lett. 2020 doi: 10.1063/5.0007532. [DOI] [Google Scholar]
  • 34.Antraoui I, Khettabi A. Properties of defect modes in a finite periodic structure with branched open resonators. Mater. Today Proc. 2020;27:3132–3138. doi: 10.1016/j.matpr.2020.04.012. [DOI] [Google Scholar]
  • 35.Elmalki M, Khettabi A. Study of various periodic acoustic lattices by two methods: Transfer matrix and Green’s method. Int. Conf. Electr. Inf. Technol. (ICEIT) 2017;2017:1–3. [Google Scholar]
  • 36.Meradi KA, Tayeboun F, Guerinik A, Zaky ZA, Aly AH. Optical biosensor based on enhanced surface plasmon resonance: Theoretical optimization. Opt. Quantum Electron. 2022;54:1–11. doi: 10.1007/s11082-021-03504-8. [DOI] [Google Scholar]
  • 37.Laly Z, Panneton R, Atalla N. Characterization and development of periodic acoustic metamaterials using a transfer matrix approach. Appl. Acoust. 2022;185:108381. doi: 10.1016/j.apacoust.2021.108381. [DOI] [Google Scholar]
  • 38.Almeida V, Lima V, Arruda J. Applying the phononic crystal concept to the intake muffler of a refrigeration compressor. Appl. Acoust. 2022;194:108800. doi: 10.1016/j.apacoust.2022.108800. [DOI] [Google Scholar]
  • 39.Hallewell G, Crawford G, McShurley D, Oxoby G, Reif R. A sonar-based technique for the ratiometric determination of binary gas mixtures. Nucl. Instrum. Methods Phys. Res. Sect. A Accel. Spectrom. Detect. Assoc. Equip. 1988;264:219–234. doi: 10.1016/0168-9002(88)90912-6. [DOI] [Google Scholar]
  • 40.Mehaney A, Ahmed AM. Theoretical design of porous phononic crystal sensor for detecting CO2 pollutions in air. Phys. E Low-Dimens. Syst. Nanostruct. 2020;124:114353. doi: 10.1016/j.physe.2020.114353. [DOI] [Google Scholar]
  • 41.Imanian H, Noori M, Abbasiyan A. Highly efficient gas sensor based on quasi-periodic phononic crystals. Sens. Actuators B Chem. 2021;345:130418. doi: 10.1016/j.snb.2021.130418. [DOI] [Google Scholar]
  • 42.Ling T, Guo LJ. Sensitivity enhancement in optical micro-tube resonator sensors via mode coupling. Appl. Phys. Lett. 2013;103:013702. doi: 10.1063/1.4813007. [DOI] [Google Scholar]
  • 43.Chowdhury U, Mukherjee R, Maity AR, Kumar S, Maji PS. Self-referenced refractive index sensor utilizing Tamm plasmon in a photonic quasicrystal. Opt. Quantum Electron. 2023;55:869. doi: 10.1007/s11082-023-05160-6. [DOI] [Google Scholar]
  • 44.Zaky ZA, Mohaseb M, Aly AH. Detection of hazardous greenhouse gases and chemicals with topological edge state using periodically arranged cross-sections. Phys. Scr. 2023;98:065002. doi: 10.1088/1402-4896/accedc. [DOI] [Google Scholar]
  • 45.Zaky ZA, Mohaseb M, Hendy AS, Aly AH. Design of phononic crystal using open resonators as harmful gases sensor. Sci. Rep. 2023;13:9346. doi: 10.1038/s41598-023-36216-y. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 46.Zaky ZA, Al-Dossari M, Sharma A, Hendy AS, Aly AH. Theoretical optimisation of a novel gas sensor using periodically closed resonators. Sci. Rep. 2024;14:2462. doi: 10.1038/s41598-024-52851-5. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 47.Yin J, Ruzzene M, Wen J, Yu D, Cai L, Yue L. Band transition and topological interface modes in 1D elastic phononic crystals. Sci. Rep. 2018;8:6806. doi: 10.1038/s41598-018-24952-5. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 48.Lima, V. D., Beli, D. and Arruda, J. R. F. Modeling one-dimensional phononic crystal rods using a state space formulation. In INTER-NOISE and NOISE-CON Congress and Conference Proceedings, pp. 7874–7885 (2019).

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Data Availability Statement

Requests for materials and data should be addressed to Zaky A. Zaky.


Articles from Scientific Reports are provided here courtesy of Nature Publishing Group

RESOURCES