Abstract.
Significance
The ability to observe and monitor cell density and morphology has been imperative for assessing the health of a cell culture and for producing high quality, high yield cell cultures for decades. Microcarrier-based cultures, used for large-scale cellular expansion processes, are not compatible with traditional visualization-based methods, such as widefield microscopy, due to their thickness and material composition.
Aim
Here, we assess the optical imaging compatibilities of commercial polystyrene microcarriers versus custom-fabricated gelatin methacryloyl (gelMA) microcarriers for non-destructive and non-invasive visualization of the entire microcarrier surface, direct cell enumeration, and sub-cellular visualization of mesenchymal stem/stromal cells.
Approach
Mie scattering and wavefront error simulations of the polystyrene and gelMA microcarriers were performed to assess the potential for elastic scattering-based imaging of adherent cells. A Zeiss Z.1 light-sheet microscope was adapted to perform light-sheet tomography using label-free elastic scattering contrast from planar side illumination to achieve optical sectioning and permit non-invasive and non-destructive, in toto, three-dimensional, high-resolution visualization of cells cultured on microcarriers.
Results
The polystyrene microcarrier prevents visualization of cells on the distal half of the microcarrier using either fluorescence or elastic scattering contrast, whereas the gelMA microcarrier allows for high fidelity visualization of cell morphology and quantification of cell density using light-sheet fluorescence microscopy and tomography.
Conclusions
The combination of optical-quality gelMA microcarriers and label-free light-sheet tomography will facilitate enhanced control of bioreactor-microcarrier cell culture processes.
Keywords: light-sheet tomography, microscopy, elastic scattering, three-dimensional cell culture, microcarriers
1. Introduction
There has been great interest in developing therapeutics based on mesenchymal stem cells (MSCs) due to their anti-inflammatory and immunomodulatory properties.1–3 However, the use of cell-based therapies, or cytotherapies, in the clinic will require manufacturing batches of trillions of cells and the development of efficient large-scale stem cell manufacturing protocols to meet this demand.4 One promising large-scale expansion strategy is to culture MSCs on spherical microcarriers maintained in suspension,5–7 which can produce a higher cell yield than monolayer cultures while maintaining the viability, identity, and functional potential of MSCs.8–12 Microcarriers can be made from different materials including solids such as polystyrene and glass or hydrogels such as gelatin methacryloyl (gelMA).13,14 Polystyrene microcarriers are readily available but are non-degradable and require a filtration step to harvest cells after expansion; gelMA microcarriers, however, are biodegradable, which improves cell harvest and manufacturing scalability.
Cell culture health is typically assessed through visualization-based methods; unfortunately, standard widefield microscopy techniques used to observe monolayer cultures do not readily translate to three-dimensional (3D) microcarrier-based cultures due to the microcarrier thickness (100+ ) and high refractive index of traditional microcarrier materials (polystyrene, glass).15,16 Furthermore, the assessment of cell culture parameters, such as cell density and morphology, requires a 3D imaging technique to capture the entire microcarrier surface.17 Due to the large diameters and relatively high refractive indices of microcarriers, the high-resolution visualization of cells along the entire microcarrier surface requires a volumetric optical imaging technique capable of optical sectioning to reconstruct the 3D object.13,18–22
Light-sheet fluorescence microscopy (LSFM) is a fast and photo-gentle camera-based imaging method that utilizes fluorescence contrast for image formation. Unfortunately, LSFM requires exogenous labels for fluorescence contrast or higher laser power for autofluorescence imaging, both of which can be toxic for live cells.23 Light-sheet tomography, however, uses elastic scattering for contrast and allows for rapid and direct 3D visualization of individual cells on microcarriers without the need for contrast agents. Light-sheet tomography thus enables robust non-invasive and non-destructive monitoring of cell culture attributes to assess culture quality.24 Because elastic scattering contrast relies on refractive-index heterogeneities in the sample, microcarriers comprised of materials with high refractive indices may scatter light intensely and preclude the visualization of weak-scattering cells on their surface.
Here, we compare the light-sheet tomography imaging compatibilities of microcarriers comprised of two different materials with different index of refraction values. Commercial polystyrene microcarriers () are compared with custom-fabricated gelMA microcarriers () and assessed as microcarrier substrates for light-sheet tomography visualization and enumeration of MSCs. Mie scattering and wavefront map simulations of polystyrene and gelMA microcarriers are performed to describe the angular scattering intensity distributions and microcarrier-induced wavefront errors, respectively. The optical properties of polystyrene and gelMA microcarriers are then directly compared for compatibility with light-sheet imaging for the visualization of cells and subcellular features and the enumeration of adherent cells from fluorescence light-sheet and elastic scattering light-sheet images. This work motivates the use of hydrogel microcarriers and light-sheet tomography for the non-destructive characterization of microcarrier-based cell cultures (Fig. 1).
2. Methods
2.1. gelMA Microcarrier Synthesis
The gelMA microcarriers were produced following a previously published protocol.13 Briefly, gelMA was first synthesized by reacting type A porcine gelatin (Sigma) and methacrylic anhydride at 0.6 mL methacrylic anhydride per gram gelatin in phosphate buffered saline (PBS) for 1 h at 50°C. The reaction was quenched with 40°C PBS then dialyzed through a 12 to 14 kDA nitrocellulose membrane against deionized water at 40°C for 7 days. Then, gelMA and photoinitiator lithium phenyl-2,4,6-trimethylbenzoylphosphinate were dissolved at a final concentration of 7.5% (w/v) and 10 mM, respectively, in PBS at 50°C. A custom polydimethylsiloxane (Sylgard 184) microfluidic device was created and bound to a 2″ × 3″ glass slide after 1 min of oxygen plasma treatment and cured overnight at 85°C. Syringe pumps (Cole Parmer) were used to flow the gelMA solution and fluorinated oil (Novec 7500) through the microfluidic device at 2 and , respectively. The gelMA and oil droplets were then exposed to 365 nm light () for 50 s to polymerize the microcarriers. The gelMA microcarriers were collected from the residual oil solution via centrifugation at 3000 for 3 min on a glycerol bed. Finally, the microcarriers were recovered from the glycerol bed and washed with PBS and stored at 4°C.
2.2. Bone Marrow-Derived MSCs Cultured on Polystyrene Microcarriers
Bone marrow-derived human MSCs (BM-hMSCs)25 transfected with red fluorescence protein (dsRed) were expanded on polystyrene microcarriers (P-221, Pall-Solohill) in a 10 mL rotating wall vessel (RWV) bioreactor (Synthecon) and fixed with neutral buffered formalin. The microcarriers have a manufacture-provided size range of 125 to . First, lentiviral transduction was used to make cells express dsRed (ThermoFisher, Waltham, Massachusetts). Then, BM-hMSCs were initially expanded in low-density monolayer cell culture in complete culture medium (CCM) [-minimum essential medium, 20% (w/v) fetal bovine serum, 2 mM L-glutamine, penicillin, and streptomycin] to obtain the required cell numbers.25 Finally, collagen I-coated polystyrene microspheres and BM-hMSCs were incubated at at 37°C for 2 h in 10 mL CCM with orbital mixing at 30 revolutions per minute (RPM). Cells were fixed with 4% paraformaldehyde (PFA) and stored in PBS at a concentration of 3 mg particles/mL PBS. Prior to imaging, fixed microcarrier-cell samples were incubated with a DRAQ-5 DNA buffer at 37°C for 30 min with agitation, then rinsed with PBS, and embedded in 1% agarose gel for imaging.
2.3. ihMSCs Cultured on gelMA Microcarriers
Passage 4 ihMSCs were first expanded in a low-density monolayer cell culture in CCM to obtain the required cell numbers. As previously described elsewhere,24 the ihMSCs26 were cultured in an RWV bioreactor (RCCS-8DQ bioreactor, Synthecon) at 24 RPM on custom diameter gelMA microcarriers at .13 Specimens were suspended in 1 mM concentration of CellTracker Green (CTG) (Thermo Scientific) for 45 min, then fixed with 4% PFA, and stored in PBS. Prior to imaging, samples were incubated with a DRAQ-5 (Thermo Scientific) DNA staining buffer at 37°C for 30 min with agitation and then rinsed with PBS. CTG permits the visualization of the cell cytoplasm and quantification of cell morphology. DRAQ-5 is used for cell nuclei visualization and direct cell enumeration.
2.4. Light-Sheet Fluorescence Microscopy
In situ fluorescence imaging of cells attached to microcarriers was performed on a Zeiss Z.1 Lightsheet microscope (LSM) using a 20× 1.0 NA (water) detection objective lens and 10× 0.2 NA illumination objective lenses (air). BM-hMSCs were illuminated with 561 (DsRed) and 638 nm (DRAQ-5). ihMSCs were illuminated with 488 (CTG) and 638 nm (DRAQ-5). The emission filters used were 505 to 545 nm (CTG), 600 to 600 nm (DsRed), and 660+ nm (DRAQ-5). A variable zoom of 1.16× was used for an effective magnification of 23.2×, and the voxel size was to satisfy Nyquist sampling requirements. Even illumination of the entire microcarrier surface was achieved using 5% to 15% power with dual-sided objective illumination with pivot scanning and online maximum intensity fusion. -stacks were acquired at 50 frames per second (FPS) using pco.edge 5.5M sCMOS cameras.
2.5. Light-Sheet Tomography
Light-sheet tomography based on elastic scattering contrast was performed on the Z.1 LSM with the same illumination and detection optics. The laser blocking filter and emission filters were removed from the optical path between the detection objective lens and the camera to allow scattered light to enter the detection path. Samples were illuminated using the 638 nm laser at 0.1% power, which was sufficient to fill the dynamic range of the camera at 50 FPS. Imaging was performed with dual-sided objective illumination with pivot scanning and online maximum intensity fusion.
2.6. Mie Scattering Simulations
The open-source Mie scattering software MiePlot (v4.6.21) was used to simulate the angular distribution of Mie scattering intensity.27 The software accepts refractive index () inputs for the immersion medium and particle, particle size, illumination wavelength, and illumination polarization. The polystyrene particles were modeled as a sphere with a diameter and equal to 1.59,28 and the gelMA microcarriers were spheres with a diameter and equal to 1.35.24 Cells were simulated to have a diameter of and equal to 1.45.29,30 The combined water and agarose immersion media was modeled with equal to 1.34. The refractive indices of the agarose immersion media and gelMA microcarriers were measured using a handheld refractometer (KRUSS Optronic).
2.7. Wavefront Map Simulations
The ray-tracing software ZEMAX was used to simulate the widefield detection arm of the LSM imaging through either the polystyrene or gelMA microcarrier.
The wavefront map simulation was used to calculate the microcarrier-induced wavefront error, represented by peak-valley and root mean square (RMS) error. The objective and tube lenses were simulated as paraxial lenses of focal length equal to 9 and 180 mm, respectively, to represent the 20× 1.0W Olympus objective and Olympus tube lens. The imaging medium was set with equal to 1.33 to simulate the use of the water-dipping objective lens. The cell, or object plane, is situated on the distal hemisphere of the microcarrier, which is placed at the front focal plane of the objective lens. The gelMA microcarrier was simulated as a sphere of equal to 1.34; the polystyrene microcarrier was simulated as a sphere of equal to 1.59; in addition, the gelMA microcarrier was simulated as and the polystyrene as to compare wavefront errors induced by varying the particle size.
2.8. Data Analysis
The open-source image analysis software ImageJ/FIJI was used to generate 2D intensity projections of the 3D microcarrier fluorescence and scattering volumes. ImageJ was also used to characterize fluorescence signal attenuation with increasing depths by creating an orthogonal projection of the microcarriers and extracting intensity profiles at varying depths of the microcarrier. MATLAB 2020b was used for statistical analysis of the normalized maximum intensity depth profiles of the two microcarriers and the simulated 100% confluent microcarrier. The optical surface area per slice of a sphere was calculated by integrating the sphere surface area using the following equation:
3. Results
3.1. Mie Scattering Simulations
To assess the potential for elastic scattering-based contrast imaging of cells expanded on spherical microcarriers, Mie simulations for the two microcarriers were performed using MiePlot. The angular distributions of Mie scattering intensity for the two microcarriers and a cell were simulated using an unpolarized light source [Fig. 2(a)]. In addition, the gelMA microcarrier was simulated as a diameter to match the diameter of and compare it to the polystyrene microcarrier [Fig. 2(b)].
These simulations show that the polystyrene microcarriers scatter light much more intensely than the hydrogel microcarriers and the cells. This phenomenon is due to the refractive index mismatch of the surrounding water immersion media and the microcarrier. The Mie simulations predict the scattering intensity to be two to four orders of magnitude greater for polystyrene than gelMA microcarriers. The cells, with an overall refractive index between water and polystyrene, scatter light near the same intensity range as the hydrogel microcarriers. This phenomenon permits the visualization of cells along the entire gelMA microcarrier surface using both fluorescence and elastic scattering contrast.
3.2. Wavefront Map Simulations
To assess the optical aberrations induced by imaging through the microcarrier, the light-sheet detection arm was simulated in ZEMAX with the distal microcarrier surface at the objective lens front focal plane. The wavefront map simulation was used to quantify the microcarrier-induced peak–valley () and RMS wavefront errors, which are key indicators of optical system performance. These simulations show that the gelMA microcarrier causes less optical path differences (OPD) than the polystyrene microcarrier (Fig. 3). In fact, the gelMA RMS error (0.0003 waves) is , which is the Rayleigh criterion that describes the threshold for diffraction-limited imaging performance [Fig. 3(a)]. Increasing the gelMA microcarrier diameter from 120 to increases the RMS wavefront error to 0.0064 waves [Fig. 3(b)]; however, this is still less than the error (2.8707 waves) introduced by the polystyrene microcarrier [Fig. 3(c)]. When the polystyrene microcarrier is simulated as a diameter, the RMS wavefront error (1.7947 waves) is still greater than that of the 120 and gelMA microcarriers. These simulations illustrate that the custom-fabricated gelMA microcarrier induces fewer optical imaging aberrations compared with commercially available polystyrene microcarriers.
3.3. Microcarrier Screening for Label-Free Elastic Scattering-Based Visualization
The polystyrene microcarrier scatters much more intensely than cells cultured on the microcarrier surface. This causes the cell signal to be obscured by the polystyrene microcarrier scattering [Fig. 4(a)]. Only cells situated on the polystyrene microcarrier equator and proximal hemisphere can be visualized using brightfield transmitted illumination [Fig. 4(b)]. This necessitates the use of fluorescence contrast to visualize and distinguish cells from the polystyrene microcarrier surface [Fig. 4(c)]. The gelMA microcarrier, conversely, scatters at intensities near or lower than cells cultured on the gelMA microcarrier surface, which permits the sub-cellular visualization of cells along the entire 3D surface using light-sheet tomography [Fig. 4(d)]. Standard deviation and sum of slices contrast enhancement projections even permit the simultaneous visualization of the microcarrier, cells, and surrounding agarose. Brightfield illumination can also be used to visualize and identify cells from the gelMA microcarrier surface [Fig. 4(e)]. Populated microcarriers can be distinguished from empty microcarriers using brightfield illumination to visualize textural features on the microcarrier surface. Fluorescence contrast similarly provides an equivalent view of cells cultured on the gelMA microcarrier surface [Fig. 4(f)]. These exemplary volumetric projections illustrate the compatibility of the hydrogel microcarrier for the label-free visualization of cells along the entire microcarrier surface.
3.4. Microcarrier Screening for Direct Fluorescence-Based Cell Enumeration
To evaluate the optical imaging performance of the different microcarriers, fixed MSCs attached to commercially available diameter polystyrene and diameter custom-fabricated hydrogel microcarriers were imaged with an LSFM. Representative data of DsRed and DRAQ-5 labeled BM-hMSCs on polystyrene and CTG and DRAQ-5 labeled ihMSCs on gelMA microcarriers are shown in Figs. 5 and 6.
The polystyrene microcarrier inhibits the visualization of cells on the distal half of the microcarrier equator [Figs. 5(a)–5(f)]. There are 30 cell nuclei identified via DRAQ-5 on the microcarrier equator and proximal surface, only one or two additional cell nuclei can be identified on the distal surface near the equator, and no cell nuclei are resolved in the center of the distal microcarrier hemisphere [Figs. 5(b) and 5(d)]. This is because the fluorescence signal intensity is attenuated with depth, and the ability to visualize, resolve, and enumerate cells beyond the microcarrier equator is severely hindered [Figs. 5(e) and 5(f)]. Cells on the distal half of the microcarrier are illuminated, but the polystyrene microcarrier blurs these slices in the stack and the ability to resolve individual nuclei is eliminated [Fig. 5(g)]. A scattering artifact caused by the microcarrier can be seen on the left hemisphere of the microcarrier in Figs. 5(c) and 5(d).
The hydrogel-based gelMA microcarriers, compared with commonly used polystyrene microcarriers, enable high fidelity visualization and semi-automatic direct enumeration of cells along the entire microcarrier surface [Figs. 6(a)–6(d)]. There are seven cell nuclei resolved from the DRAQ-5 contrast [Fig. 6(f)]. The fluorescence intensity is preserved along the entire depth of the microcarrier, and cell nuclei can be visualized on both the proximal and distal hemisphere of the microcarrier [Figs. 6(e)–6(g)].
3.5. Depth-Dependent Fluorescence Signal Attenuation
To compare the depth-dependent fluorescence signal intensity attenuation caused by the microcarrier material, the maximum DRAQ-5 intensity per frame per microcarrier profile is averaged, normalized, and plotted for both microcarrier materials; in addition, the theoretical illuminated microcarrier surface area is plotted as a reference for a 100% confluent microcarrier. Only single microcarriers with two or more cells and CTG confluency were included to select for moderate and highly confluent single microcarriers and to avoid including aggregated microcarriers in the analysis. The gelMA microcarriers have a more symmetrical profile than the polystyrene microcarriers, and these data show that the gelMA microcarriers cause less depth-dependent fluorescence signal intensity attenuation than the polystyrene microcarriers (Fig. 7). A two-sample Kolmogorov–Smirnov test was performed to quantitatively assess the similarity of the depth-intensity profiles between the polystyrene and gelMA microcarriers. The gelMA and polystyrene microcarriers showed a statistically significant difference from each other (). The gelMA microcarrier profile did not show statistically significant difference from the 100% confluent microcarrier (), whereas the difference between the polystyrene and 100% confluent microcarrier profiles was found to be statistically significant (). Data were considered to be significant if the values were . Both microcarrier material intensity profiles show a decrease in intensity and deviation from the theoretical confluent microcarrier beginning at around 55% depth; this is an interesting phenomenon that may be due to the spherical nature of the microcarriers or a sample size artifact. Mie scattering of spherical objects is well known to have both broad and sharp intensity oscillations termed wiggles and ripples, respectively, that are due to the spherical shape of the scattering object;31,32 the depth-dependent fluorescence intensity oscillations seen here could similarly be related to the spherical nature of the microcarriers.
4. Discussion
The FDA has encouraged therapeutic manufacturers to deploy process analytical technologies (PATs) as a means to monitor and control a manufacturing process by gathering (near) real-time insight into critical attributes of the therapeutic product that can be incorporated into a control strategy to assure product quality.33 The combination of optically compatible microcarriers with rapid, label-free, high-resolution imaging and robust image analysis-based characterization of adherent cell cultures will greatly enhance the ability to monitor large-scale expansion of cells for cell-based therapies.
In this study, we compared the optical imaging compatibilities of commercial polystyrene microcarriers versus custom-fabricated gelMA microcarriers for non-destructive and non-invasive visualization of the entire microcarrier surface, direct cell enumeration, and sub-cellular visualization of MSCs using Mie scattering and wavefront error simulations and empirical light-sheet imaging data. The Mie scattering simulations explain previous results using reflectance confocal microscopy and light-sheet tomography for scattering-based volumetric imaging of microcarrier cell cultures;13,24 the polystyrene microcarrier scattering intensity obscures the cell scattering signal, and fluorescence modalities are unable to visualize the distal hemisphere of the polystyrene microcarrier. Image formation from the distal microcarrier hemisphere requires photons scattered by cells to reach the camera, but these signal-contributing photons are likely to be deflected and/or absorbed by the polystyrene microcarrier before they reach the detector. Light-sheet imaging is traditionally performed with de-coupled and orthogonal illumination and detection paths, and the detection objective used here has a 97.5 deg angular aperture. It is apparent from the Mie scattering plots that the polystyrene microcarrier scatters more intensely than the gelMA microcarrier in the detection objective lens angular aperture range, as well as in the forward direction, which is relevant for transmitted illumination imaging schemes.
The wavefront error simulations further demonstrate that the gelMA microcarrier induces smaller wavefront aberrations compared with polystyrene microcarriers of the same diameter; this permits diffraction-limited, high-resolution imaging through the entire gelMA microcarrier volume and the visualization of cells along the entire microcarrier surface. The polystyrene microcarrier causes a greater wavefront error due to the refractive index mismatch between itself and the water imaging medium, whereas the gelMA microcarrier has a minimal refractive index mismatch with water. The ZEMAX simulation was performed using sequential mode and does not consider scattering from the microcarriers, which would further reduce the image quality from polystyrene microcarriers.
The imaging data illustrate the limited optical imaging compatibility of commonly employed polystyrene microcarriers for in toto quantitative monitoring of ihMSC growth on microcarriers. The polystyrene microcarrier does not permit the visualization of the entire 3D surface due to the absorption of light and its refractive index mismatch with the imaging medium, which causes optical aberrations and high-intensity scattering; thus, the observation of cells is limited to the proximal hemisphere of the polystyrene microcarrier. The custom gelMA microcarriers are optically transparent with a refractive index close to water, which allows for the label-free visualization of the entire microcarrier core and surface. These experimental findings are supported by both the Mie scattering and wavefront error simulations, which show the value in refractive index matching of the microcarrier for label-free scattering-based 3D imaging of MSCs. In addition, cell nuclei fluorescently labeled with DRAQ-5 situated anywhere on the microcarrier surface can be visualized with a high spatial resolution and SNR as there is minimal refractive index mismatch between the gelMA microcarrier material and the water imaging medium that would cause optical aberrations. Cell enumeration is, therefore, not limited to the superficial half of the microcarrier surface and does not require the membrane lysis or detachment of cells from the microcarrier surface.
Volumetric optical imaging enables high fidelity visualization of the entire microcarrier surface. Coupled with hydrogel microcarriers, cells can be visualized using fluorescence and elastic scattering contrast. The use of light-sheet tomography and gelMA microcarriers permits faster acquisition and visualization of adherent cells using multiple orders of magnitude less power than is required for fluorescence contrast. The angular scattering intensity distribution profiles of the microcarriers were simulated to assess the potential for elastic scattering-based imaging of the cells cultured on the microcarrier surface. The wavefront error simulations further motivate refractive index matching of the microcarrier material and imaging medium for high-quality imaging data of adherent cells on spherical microcarriers. In addition, the optical imaging compatibilities of polystyrene and gelMA microcarriers were quantitatively compared and show a significant difference in the ability to visualize and characterize the entire microcarrier surface.
5. Conclusion
Optically transparent microcarriers composed of gelMA enable non-invasive and non-destructive in toto quantification of cell density and microcarrier surface area confluency. Conversely, the polystyrene microcarrier causes greater scattering of illumination light than the hydrogel microcarrier, and this prevents the visualization of cells on the distal microcarrier hemisphere using either fluorescence or elastic scattering contrast. The gelMA microcarriers, when combined with light-sheet imaging, permit the high-fidelity visualization of cells in 3D and quantification of critical cell culture parameters, such as cell density and morphology. Label-free light-sheet tomography combined with the gelMA microcarriers enables rapid, non-destructive, in toto imaging of adherent cells. This avoids the use of exogenous fluorescent labels, which are destructive to the cells and thus decrease the yield of the cell culture process. Label-free contrast is, therefore, essential for in- or on-line monitoring of cell cultures not only to preserve the sterility of the culture and minimize the waste of valuable resources and product but also for rapid results to enable real-time decision making in the upstream manufacturing process. The combination of optical-quality gelMA microcarriers and label-free light-sheet tomography will aid in the development of PATs for cytotherapy manufacturing and facilitate the enhanced control of the bioreactor-microcarrier cell culture processes.
Acknowledgments
This work was supported by the Texas A&M President’s Excellence Fund (H.C.G., R.K., and K.C.M.; Award No. 76) and the Silicon Valley Community Foundation through the Chan Zuckerberg Initiative (CZI) Imaging Scientist Program (H.C.G.; Award No. 2019-198168). We thank the Texas A&M University Microscopy and Imaging Center Core Facility for providing the Zeiss Z.1 light-sheet microscope (RRID:SCR_022128).
Biographies
Oscar R. Benavides is a biomedical engineer specializing in biophotonics and optical imaging. He earned his PhD from Texas A&M University, focusing on quantitative monitoring of microcarrier-based cell cultures. He has extensive experience in developing optical imaging systems and analysis methods.
Alex J. Walsh is an assistant professor in the Biomedical Engineering Department at Texas A&M University. She received her MS and PhD degrees in biomedical engineering from Vanderbilt University where she received training in label-free microscopy and quantitative image analysis. She completed her post-doctoral training at the Air Force Research Lab studying the biophysical effects of infrared light. Currently, her lab seeks to improve health through the development of label-free optical technologies.
Biographies of the other authors are not available.
Contributor Information
Oscar R. Benavides, Email: obenavides@tamu.edu.
Berkley P. White, Email: berkley.white@tamu.edu.
Holly C. Gibbs, Email: hgibbs@tamu.edu.
Roland Kaunas, Email: rkaunas@tamu.edu.
Carl A. Gregory, Email: cgregory@tamu.edu.
Kristen C. Maitland, Email: kmaitland@tamu.edu.
Alex J. Walsh, Email: walshaj@tamu.edu.
Disclosures
The authors declare no conflicts of interest.
Code and Data Availability
All relevant data, materials, and software code used in this research are available upon request from the corresponding author.
References
- 1.Hmadcha A., et al. , “Therapeutic potential of mesenchymal stem cells for cancer therapy,” Front. Bioeng. Biotechnol. 8, 43 (2020). 10.3389/fbioe.2020.00043 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2.Jovic D., et al. , “A brief overview of global trends in MSC-based cell therapy,” Stem Cell Rev. Rep. 18(5), 1525–1545 (2022). 10.1007/s12015-022-10369-1 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.Zhuang W. Z., et al. , “Mesenchymal stem/stromal cell-based therapy: mechanism, systemic safety and biodistribution for precision clinical applications,” J. Biomed. Sci. 28(1), 28 (2021). 10.1186/s12929-021-00725-7 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.Connon C. J., Bioprocessing for Cell Based Therapies, pp. 1–254, Wiley; (2016). [Google Scholar]
- 5.McKee C., Chaudhry G. R., “Advances and challenges in stem cell culture,” Colloids Surf. B Biointerfaces 159, 62–77 (2017). 10.1016/j.colsurfb.2017.07.051 [DOI] [PubMed] [Google Scholar]
- 6.Anton D., et al. , “Three-dimensional cell culture: a breakthrough in vivo,” Int. J. Mol. Sci. 16(3), 5517–5527 (2015). 10.3390/IJMS16035517 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7.Jensen C., Teng Y., “Is it time to start transitioning from 2D to 3D cell culture?” Front. Mol. Biosci. 7, 33 (2020). 10.3389/fmolb.2020.00033 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8.Egger D., et al. , “Hypoxic three-dimensional scaffold-free aggregate cultivation of mesenchymal stem cells in a stirred tank reactor,” Bioengineering 4(2), 47 (2017). 10.3390/bioengineering4020047 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.Bin Hassan M. N. F., et al. , “Large-scale expansion of human mesenchymal stem cells,” Stem Cells Int. 2020, 9529465 (2020). 10.1155/2020/9529465 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Lambrechts T., et al. , “Evaluation of a monitored multiplate bioreactor for large-scale expansion of human periosteum derived stem cells for bone tissue engineering applications,” Biochem. Eng. J. 108, 58–68 (2016). 10.1016/j.bej.2015.07.015 [DOI] [Google Scholar]
- 11.Zhou L., et al. , “Ex vivo expansion of bone marrow mesenchymal stem cells using microcarrier beads in a stirred bioreactor,” Biotechnol. Bioprocess Eng. 18(1), 173–184 (2013). 10.1007/s12257-012-0512-5 [DOI] [Google Scholar]
- 12.Cunha B., et al. , “Bioprocess integration for human mesenchymal stem cells: from up to downstream processing scale-up to cell proteome characterization,” J. Biotechnol. 248, 87–98 (2017). 10.1016/j.jbiotec.2017.01.014 [DOI] [PubMed] [Google Scholar]
- 13.Rogers R. E., et al. , “A scalable system for generation of mesenchymal stem cells derived from induced pluripotent cells employing bioreactors and degradable microcarriers,” Stem Cells Transl. Med. 10(12), 1650–1665 (2021). 10.1002/sctm.21-0151 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14.Haskell A., et al. , “Scalable manufacture of therapeutic mesenchymal stromal cell products on customizable microcarriers in vertical wheel bioreactors that improve direct visualization, product harvest, and cost,” Cytotherapy 26(4), 372–382 (2024). 10.1016/j.jcyt.2024.01.009 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15.Van Beylen K., Papantoniou I., Aerts J. M., “Microcarrier screening and evaluation for dynamic expansion of human periosteum-derived progenitor cells in a xenogeneic free medium,” Front. Bioeng. Biotechnol. 9, 624890 (2021). 10.3389/fbioe.2021.624890 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16.McNeill E. P., et al. , “Three-dimensional in vitro modeling of malignant bone disease recapitulates experimentally accessible mechanisms of osteoinhibition,” Cell Death Dis. 9(12), 1161 (2018). 10.1038/s41419-018-1203-8 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17.Odeleye A. O. O., et al. , “Development of an optical system for the non-invasive tracking of stem cell growth on microcarriers,” Biotechnol. Bioeng. 114(9), 2032–2042 (2017). 10.1002/bit.26328 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18.Baradez M. O., Marshall D., “The use of multidimensional image-based analysis to accurately monitor cell growth in 3D bioreactor culture,” PLoS One 6(10), e26104 (2011). 10.1371/journal.pone.0026104 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19.Tharmalingam T., et al. , “Enhanced production of human recombinant proteins from CHO cells grown to high densities in macroporous microcarriers,” Mol. Biotechnol. 49(3), 263–276 (2011). 10.1007/s12033-011-9401-y [DOI] [PubMed] [Google Scholar]
- 20.Duchi S., et al. , “A new holistic 3D non-invasive analysis of cellular distribution and motility on fibroin-alginate microcarriers using light sheet fluorescent microscopy,” PLoS One 12(8), e0183336 (2017). 10.1371/journal.pone.0183336 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21.Conchello J. A., Lichtman J. W., “Optical sectioning microscopy,” Nat. Methods 2(12), 920–931 (2005). 10.1038/nmeth815 [DOI] [PubMed] [Google Scholar]
- 22.Jakob P. H., et al. , “A 3-D cell culture system to study epithelia functions using microcarriers,” Cytotechnology 68(5), 1813–1825 (2016). 10.1007/s10616-015-9935-0 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 23.Icha J., et al. , “Phototoxicity in live fluorescence microscopy, and how to avoid it,” Bioessays 39(8) (2017). 10.1002/bies.201700003 [DOI] [PubMed] [Google Scholar]
- 24.Benavides O. R., et al. , “Volumetric imaging of human mesenchymal stem cells (hMSCs) for non-destructive quantification of 3D cell culture growth,” PLoS One 18(3), e0282298 (2023). 10.1371/journal.pone.0282298 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 25.McNeill E. P., et al. , “Characterization of a pluripotent stem cell-derived matrix with powerful osteoregenerative capabilities,” Nat. Commun. 11(1), 3025 (2020). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 26.Gregory C. A., Prockop D. J., “Fundamentals of culture and characterization of mesenchymal stem/progenitor cells (MSCs) from bone marrow stroma,” in Culture of Human Stem Cells, Freshney R. I., Stacey G. N., Auerbach J. M., Eds., pp. 207–232, Wiley; (2007). [Google Scholar]
- 27.Laven P., “MiePlot,” [Online], http://www.philiplaven.com/mieplot.htm (accessed September 4, 2023).
- 28.Jones S. H., King M. D., Ward A. D., “Determining the unique refractive index properties of solid polystyrene aerosol using broadband Mie scattering from optically trapped beads,” Phys. Chem Chem. Phys. 15(47), 20735–20741 (2013). 10.1039/c3cp53498g [DOI] [PubMed] [Google Scholar]
- 29.Liu P. Y., et al. , “Cell refractive index for cell biology and disease diagnosis: past, present and future,” Lab Chip 16(4), 634–644 (2016). 10.1039/C5LC01445J [DOI] [PubMed] [Google Scholar]
- 30.Gul B., et al. , “Cell refractive index: models, insights, applications and future perspectives,” Photodiagn. Photodyn. Ther. 33, 102096 (2021). 10.1016/j.pdpdt.2020.102096 [DOI] [PubMed] [Google Scholar]
- 31.Brandsrud M. A., et al. , “The effect of deformation of absorbing scatterers on Mie-type signatures in infrared microspectroscopy,” Sci. Rep. 11(1), 4675 (2021). 10.1038/s41598-021-84064-5 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 32.Sorensen C. M., Shi D., “Patterns in the ripple structure of Mie scattering,” JOSA A 19(1), 122–125 (2002). 10.1364/JOSAA.19.000122 [DOI] [PubMed] [Google Scholar]
- 33.Riley B. S., Li X., “Quality by design and process analytical technology for sterile products—where are we now?” AAPS PharmSciTech 12(1), 114 (2011). 10.1208/s12249-010-9566-x [DOI] [PMC free article] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Data Availability Statement
All relevant data, materials, and software code used in this research are available upon request from the corresponding author.