The impact of dialysate flow rate on haemodialysis adequacy: a systematic review and meta-analysis

Yasmin Iman 1, Ryan Bamforth 2, Ruth Ewhrudjakpor 3, Paul Komenda 4,5,6, Kelley Gorbe 7, Reid Whitlock 8, Clara Bohm 9,10, Navdeep Tangri 11,12, David Collister 13,14,15,
PMCID: PMC11229034  PMID: 38979109

ABSTRACT

Background

Patients with kidney failure treated with maintenance haemodialysis (HD) require appropriate small molecule clearance. Historically, a component of measuring ‘dialysis adequacy’ has been quantified using urea kinetic modelling that is dependent on the HD prescription. However, the impact of dialysate flow rate on urea clearance remains poorly described in vivo and its influence on other patient-important outcomes of adequacy is uncertain.

Methods

We searched Embase, MEDLINE and the Cochrane Library from inception until April 2022 for randomized controlled trials and observational trials comparing a higher dialysate flow rate (800 ml/min) and lower dialysate flow rate (300 ml/min) with a standard dialysis flow rate (500 ml/min) in adults (age ≥18 years) treated with maintenance HD (>90 consecutive days). We conducted a random effects meta-analysis to estimate the pooled mean difference in dialysis adequacy as measured by Kt/V or urea reduction ratio (URR).

Results

A total of 3118 studies were identified. Of those, nine met eligibility criteria and four were included in the meta-analysis. A higher dialysate flow rate (800 ml/min) increased single-pool Kt/V by 0.08 [95% confidence interval (CI) 0.05–0.10, P < .00001] and URR by 3.38 (95% CI 1.97–4.78, P < .00001) compared with a dialysate flow rate of 500 ml/min. Clinically relevant outcomes including symptoms, cognition, physical function and mortality were lacking and studies were generally at a moderate risk of bias due to issues with randomization sequence generation, allocation concealment and blinding.

Conclusion

A higher dialysate flow increased urea-based markers of dialysis adequacy. Additional high-quality research is needed to determine the clinical, economic and environmental impacts of higher dialysate flow rates.

Keywords: adequacy, dialysate flow rate, haemodialysis, Kt/V, urea reduction ratio


KEY LEARNING POINTS.

What was known:

  • Achieving minimum Kt/V targets has been shown to improve patient morbidity and mortality.

  • There are many other clinically relevant outcomes related to dialysis adequacy, including blood pressure, volume status, ultrafiltration rate, inflammation, middle molecule clearance, physical function, cognition and a variety of patient-reported outcome measures.

  • Understanding how to optimize Kt/V by modifying the HD prescription is important, but not all approaches may be feasible when considering patient values and preferences.

This study adds:

  • An understanding of the impact of higher and lower dialysate flow rates, compared with the standard dialysis flow rate of 500 ml/min, on dialysis adequacy as measured by Kt/V, urea reduction ratio and patient-important outcomes, including symptoms, cognition, physical function and quality of life, is needed.

Potential impact:

  • Increasing dialysis flow rates in adults receiving maintenance HD improves dialysis adequacy as measured by urea kinetic modelling metrics, including URR and single-pool Kt/V.

  • More evaluation is necessary in high-quality prospective studies.

INTRODUCTION

Patients with kidney failure undergoing maintenance haemodialysis (HD) require adequate removal of solutes and fluid in order to restore homeostasis. Historically, one key marker of ‘dialysis adequacy’ has focused on urea kinetic modelling quantified by dialyzer clearance of urea (K) multiplied by the duration of the dialysis session (t) adjusted for the volume (V) of distribution [1]. Achieving minimum Kt/V targets have been shown to improve patient morbidity and mortality, but dialysis adequacy is a complex construct that encompasses many traditional targets, including small molecule clearance, potassium, acid–base status, anaemia, mineral and bone disorder and cardiovascular risk [2–6]. In addition, there are many other clinically relevant outcomes related to dialysis adequacy, including blood pressure (BP), volume status, ultrafiltration rate, inflammation, middle molecule clearance, physical function, cognition and a variety of patient-reported outcome measures (PROMs) related to symptoms that are not captured by Kt/V [7]. However, Kt/V is still an important metric in delivering safe and quality HD care and remains an indicator of quality of care for dialysis programs.

Understanding how to optimize Kt/V by modifying the HD prescription is important [2–7], but not all approaches may be feasible when considering patient values and preferences, reimbursement constraints and scheduling of facility-based HD treatments. Contemporary HD systems offering more portability and improved usability for self-care are becoming increasingly available [8–13]. These systems unlock the potential for HD to be performed in a range of settings and a diverse group of users. Some HD systems, due to either the use of batch-based or bagged dialysate or limitations of water systems, are unable to attain dialysate flow rates at or above the standard rate of 500 ml/min provided by conventional HD systems [8, 9]. Whether this limitation affects the ability of the patient to achieve an optimal Kt/V target is unclear and the relative impact of higher or lower dialysate flow rates on dialysis adequacy in clinical settings is unknown.

We performed a systematic review and meta-analysis to summarize the impact of higher and lower dialysate flow rates,

compared with the standard dialysis flow rate of 500 ml/min, on dialysis adequacy as measured by Kt/V, urea reduction ratio (URR) and patient-important outcomes, including symptoms, cognition, physical function and quality of life.

MATERIALS AND METHODS

Data sources and searches

We developed and followed a protocol that considered population, intervention, comparison and outcome criteria for our search strategy. The population of interest was adults (age ≥18 years) with kidney failure on maintenance HD (dialyzing for ≥90 consecutive days). The setting of HD could have been either facility-based or home HD. Studies of interest contained interventions that were a low dialysate flow rate (300 ml/min) or a high dialysate flow rate (800 ml/min) and a comparator of the standard dialysate flow rate (500 ml/min). Any type of HD prescription was permitted, including weekly schedule, duration, vascular access, blood flow rate and dialyzer. Studies considering dialysate flow rates not equal to 300, 500 or 800 ml/min were included but not meta-analysed. Our primary outcome was adequacy of dialysis measured by URR or Kt/V quantified using either single-pool (spKt/V), double-pool (dpKt/V) equilibrated (eKt/V) or standard (stdKt/V) [14–19]. Our secondary outcomes included all-cause mortality and the following PROMs: health-related quality of life as measured by the EuroQol-5 Dimension (EQ-5D), Kidney Disease Quality of Life (KDQoL-36) questionnaire and its short form version (KDQoL-SF); symptom burden as measured by the Edmonton Symptom Assessment Scale (ESAS) or its renal version (ESAS-r); and other disease-specific instruments including but not limited to mood, pruritus, restless legs, gastrointestinal symptoms (nausea, vomiting, appetite), fatigue, cognition (e.g. Mini Mental Status Exam, Montreal Cognitive Assessment) and physical function (e.g. 6-minute walk test, sit to stand, hand grip strength, others). We applied methodological approaches outlined in the Preferred Reporting Items for Systematic Reviews and Meta-Analyses (PRISMA) guidelines as well as those outlined in PRISMA-S (literature search extension) [20, 21]. This study was registered on PROSPERO (CRD42022321877) [22].

In collaboration with a medical librarian, original research articles were identified from the following databases: Embase, MEDLINE and the Cochrane Library. We included articles in the English language only, there was no restriction on sample size and grey literature was excluded (Fig. 1). Our search strategy included keywords and MeSH terms related to ‘dialysis adequacy’, ‘haemodialysis adequacy’, ‘clearance’, ‘urea reduction ratio’, ‘Kt/V’, ‘dialysis dose’ and ‘dialysate flow’. We included randomized controlled trials (RCTs) and observational studies; all other study designs were excluded.

Figure 1:

Figure 1:

PRISMA flow diagram.

Two reviewers (Y.I. and R.E.) independently reviewed each article by title and abstract to select articles for full-text review. The two reviewers (Y.I. and R.E.) then screened the full text of the selected articles to determine if they were eligible for inclusion in the analysis. References of the included studies were also screened for additional studies that met the inclusion criteria. All disagreements were resolved by consensus after consultation with a third reviewer (D.C.).

Data extraction and quality assessment

A data extraction form was created to capture relevant information from the studies that met the inclusion criteria. Two reviewers (Y.I. and R.E.) independently extracted the data from the studies. The following data were extracted from the included studies, depending on availability: article information (title, first author, year, journal, title, location, design), inclusion and exclusion criteria, characteristics of the intervention and comparator groups (sample size, sex/gender, age, vascular access type, HD setting, study duration, primary and secondary outcome measures. Outcomes from studies with multiple arms related to differing baseline HD prescriptions (i.e. weekly schedule, duration, vascular access, blood flow rate and dialyzer) were extracted separately if present and both unscaled and scaled Kt/V outcomes were reported if applicable [23].

Additionally, two reviewers (Y.I. and R.E.) assessed the quality of the included studies using the Cochrane Risk of Bias Tool [24]. This tool consists of six domains (sequence generation, allocation concealment, blinding, incomplete outcome data, selective outcome reporting and ‘other’ sources of bias) and a categorization of the overall risk of bias. Each separate domain was rated as low, unclear or high risk. The non-randomized trials were assessed using the ROBINS-I (Risk Of Bias In Non-randomized Studies – of Interventions) tool. Discrepancies between the two reviewers were resolved with a third reviewer (D.C.) by consensus.

Data synthesis and analysis

We performed random effects meta-analysis employing the generic inverse variance method to estimate mean differences and related 95% confidence intervals (CIs). Statistical heterogeneity was quantified using the I2 statistic, with statistical significance assessed with the χ2 test. Methods outlined in the Cochrane Handbook for Systematic Reviews of Interventions were used to calculate standard deviations (SDs) of the difference, standard errors of the mean (SEM) difference and the mean difference when required [25]. We assumed a correlation coefficient value of 0.5 in the main analyses, with sensitivity analyses considering values of 0.25 and 0.9. In instances where outcome-specific SDs were not available, we imputed the SEM difference based on values from the other included studies. This was only performed when the minority of studies in the meta-analysis required imputation. As exploratory analyses, we performed a random effects network meta-analysis estimating mean differences and 95% CIs for the outcome spKt/V. The standard dialysate flow rate (500 ml/min) was compared with those both greater and less than 500 ml/min. For main and exploratory analyses, two or more studies with the relevant outcome were required. Statistical significance was determined using α = 0.05.

Statistical analysis pertaining to the standard meta-analysis was performed using Review Manager (RevMan version 5.3, Nordic Cochrane Center, Copenhagen, Denmark). Network meta-analysis was performed using the netmeta package in R version 4.1.0 (R Foundation for Statistical Computing, Vienna, Austria) [26].

RESULTS

Study selection

The PRISMA flow diagram is presented in Fig. 1. The initial search strategy yielded 3118 research articles eligible for screening (3279 total minus 161 duplicates). Of those, 361 were selected for full-text review, with 9 meeting the inclusion criteria [7, 27–34]. Of the nine studies, three were randomized crossover trials and six were observational studies (prospective crossover trials). The RCTs were set in Canada (n = 1), Colombia (n = 1) and the USA (n = 1) and the remaining six observational studies were set in Germany (n = 1), Greece (n = 1), Egypt (n = 2), Morocco (n = 1) and Spain (n = 1). Only two of the nine (22%) included studies were >3 months in duration. Primary outcomes that were reported were Kt/V (unspecified type), spKt/V, Kt, eKt/V and URR. Secondary outcomes that were reported were body weight, BP, nutritional status, anaemia and quality of life. All studies included patients with either an arteriovenous fistula or a central venous catheter. The type of vascular access and blood flow rate were kept constant during the crossover period. Study characteristics are described in Tables 1 and 2.

Table 1:

Study characteristics.

Dialysis dose Outcome
Author (year) Country Study design Age (years) Intervention Comparator n (control/intervention) HD setting Duration of trial Primary Secondary
Ahrenholz et al. (2015) [27] Germany Observational crossover trial Adult population a. 300 ml/minb. 800 ml/min 500 ml/min 18 (18/18) HD 3 weeks spKt/V NR
Alayoud et al. (2012) [28] Morocco Single-centre observational crossover trial 49 ± 17 700 ml/min 500 ml/min 33 (33/33) HD 3 weeks Kt/Va] NR
Albalate et al. (2015) [29] Spain Single-centre observational crossover trial 78 a. 400 ml/minb. 700 ml/min 500 ml/min 31 (31/31) HD 27 weeks Kt (L) NR
Azar, Ahmad (2009) [7] Egypt Observational crossover trial 50.51 ± 15.12 800 ml/min 500 ml/min 138 (138/138) HD 32.14 ± 28.72 months spKt/V, URR NR
Azar et al. (2007) [30] Egypt Observational crossover trial 48.21 ± 13.38 800 ml/min 500 ml/min 134 (134/134) HD 3 months spKt/V, URR Body weight
Molano-Trivino et al. (2019) [31] Colombia Crossover RCT 62.5 400 ml/min 500 ml/min 46 HD 5 years Kt/Va Body weight
Panagoutsos et al. (2001) [32] Greece Observational crossover trial 52.6 ± 14.5 560 ml/min 500 ml/min 34 (34/34) HD 2 years Kt/Va, URR Body weight, BP, nutritional status, anaemia and QoL
Wang et al. (2008) [33] Canada Crossover RCT 56 800 ml/min 500 ml/min 18 (12/12) HD 12 weeks spKt/V QoL
Ward et al. (2011) [34] United States Multicentre crossover RCT 50 800 ml/min 500 ml/min 33 (33/33) HD NR spKt/V, eKt/V NR

NR: not reported.

a

Type of Kt/V was not specified.

Table 2:

Study trial characteristics.

Author (year) AVF/AVG Dialysis schedule Dialysis duration QB Dialyzer Urea assays Kt/V methodology
Ahrenholz et al. (2015) [27] Catheter or AVF 2–3 times/week 3–4 hours 300 ml/min FDY-150 GW, FX 60, VitaPES 150HF NR QB, dialyzer and URR
Alayoud et al. (2012) [28] Catheter or AVF 3 times/week NR 300 ml/min Fresenius 5008 NR Online clearance monitoring
Albalate et al. (2015) [29] Tunnelled central venous catheter or AVF 3 times/week 3–4 hours 374–382 ml/min AK200, Fresenius 5008 NR Recorded directly from monitor/system
Azar (2009) [7] Catheter or AVF 3 times/week 3–4 hours 300 ml/min Fresenius 4008B NR Daugirdas second-generation formula
Azar et al. (2007) [30] Catheter or AVF 3 times/week 3–4 hours 200–350 ml/min NR NR Daugirdas second-generation formula
Molano-Trivino et al. (2019) [31] Catheter or AVF 3 times/week 4 hours NR NR NR NR
Panagoutsos et al. (2001) [32] Catheter or AVF 3 times/week 4 hours 240–410 ml/min NR NR Daugirdas second-generation formula
Wang et al. (2008) [33] Catheter or AVF 3 times/week 4–4.5 hours 350–400 ml/min F80A, Fresenius Pre- and post-dialysis urea samples taken with arterial needle Post-dialysis blood samples, pre-dialysis and post-dialysis urea samples through arterial needle Kt/V = {−ln(urea post/urea pre) + 3[(weight pre – weight post)/weight post]}/[1 − 0.01786(time dialysis)]
Ward et al. (2011) [34] Catheter or AVF 3 times/week 3–4 hours 400–450 ml/min Single-use Polyflux Revaclear, Revaclear MAX, Phoenix dialysis machines Pre- and post-dialysis plasma urea BUN spKt/V and eKt/V were calculated using pre- and post-dialysis BUN concentrations according to Daugirdas second-generation formula

AVF: arteriovenous fistula; AVG: arteriovenous graft; BUN: blood urea nitrogen; NR: not reported.

Risk of bias assessment

Of the nine studies included in the review, only one was found to be at a low risk of bias (Figs. 2 and 3). Of the three RCTs included, none disclosed allocation concealment or whether participants and personnel were blinded. This led to one of the included studies being categorized as high risk of bias. The remaining seven were categorized as moderate risk due to a lack of or unclear information being presented in one or more categories.

Figure 2:

Figure 2:

Risk of bias assessment for randomized trials using the revised Cochrane Risk of Bias Tool.

Figure 3:

Figure 3:

Risk of bias assessment for non-randomized trials using the ROBINS-I risk of bias tool.

Meta-analysis

Only four studies [7, 27, 30, 33] were selected for meta-analysis, as the other five did not include dialysate flow rate comparisons of 300 versus 500 or 500 versus 800 ml/min [28, 29, 31, 32, 34].

High versus standard flow rate and URR

Two studies with a total of five comparisons were included in the meta-analysis comparing URR for dialysate flow rates of 500 and 800 ml/min [7, 30] (Fig. 4). A dialysate flow rate of 800 ml/min increased URR by 3.38 (95% CI 1.97–4.78, P < .00001). Heterogeneity was present as per the I2 statistic (93%) and the χ2 test (P < .00001).

Figure 4:

Figure 4:

Random effects meta-analysis for URR—500 ml/min versus 800 ml/min.

High versus standard flow rate and spKt/V

Three studies with a total of eight comparisons were included in our meta-analysis comparing spKt/V for dialysate flow rates of 500 ml/min and 800 ml/min [7, 27, 33] (Fig. 5). The mean difference was 0.08 (95% CI 0.05–0.10, P < .00001) in favour of a dialysate flow rate of 800 ml/min. The associated I2 statistic was 52% with a χ2 test P-value <.04, suggesting a moderate degree of heterogeneity.

Figure 5:

Figure 5:

Random effects meta-analysis for spKt/V—500 ml/min versus 800 ml/min.

Sensitivity analysis

Sensitivity analysis results considering correlation coefficients of 0.25 are presented in Supplementary Figs. 1 and 2. Comparing 500 ml/min versus 800 ml/min, assuming a correlation coefficient of 0.25, we found mean differences of 3.33 URR (95% CI 1.93–4.73) and 0.08 spKt/V (95% CI 0.05–0.10), both statistically significant in favour of 800 ml/min. In Supplementary Figs. 3 and 4, sensitivity analyses results assuming a correlation coefficient of 0.9 found mean differences of 3.41 URR (95% CI 1.98–4.85) and 0.08 spKt/V (95% CI 0.05–0.10), both statistically significant in favour of 800 ml/min. These results are both similar in direction and magnitude to those found in the primary analyses.

Included studies

Four studies that were excluded from the meta-analysis [28, 29, 31, 32] reported that an increase in dialysate flow rate was associated with an increase in Kt/V (unspecified type) and/or URR. Alayoud et al. [28] compared 500 ml/min versus 700 ml/min in 33 patients and found that the means and SDs for Kt/V (as measured via online clearance monitoring) modestly increased from 1.50 ± 0.16 to 1.52 ± 0.16. Similarly, Molano-Trivano et al. [31] found that Kt/V increased from 1.57 ± 0.25 to 1.59 ± 0.23 when comparing 400 ml/min versus 500 ml/min for a population of 100 patients. Albalate et al. [29] showed that in 31 patients there was a 4% increase in Kt when comparing 400 ml/min versus 500 ml/min and an additional 3% increase in Kt when comparing 500 ml/min versus 700 ml/min. Panagoutsos et al. [32] compared 500 ml/min versus 560 ml/min in 34 patients and found that URR increased from 52.8 ± 8 to 71 ± 7 and Kt/V increased from 0.93 ± 0.19 to 1.55 ± 0.29. It is important to note that throughout the duration of this 2-year trial, blood flow increased from 240 ml/min to 410 ml/min, which may have had a significant impact on these results.

Secondary outcomes

Only one of the studies with 18 participants included a secondary outcome of interest [33]. It compared results from the KDQoL-SF questionnaire as well as the Health Utility Index Mark 2 (HUI2) questionnaire for health-related quality of life (HRQoL) in patients on HD whose dialysate flow was 500 ml/min versus 800 ml/min. The results showed that there was no difference in KDQoL [mean difference −5.25 (95% CI −15.00–4.50)] or HUI2 [mean difference −0.04 (95% CI −0.16–0.08)] between the two groups. No other study included HRQoL, symptom burden, cognition, physical function or mortality.

Exploratory analysis

Four studies with a total of 15 comparisons were included in the network meta-analysis [7, 27, 33, 34] comparing spKt/V for dialysate flow rates of 300, 600 and 800 ml/min with 500 ml/min (four treatments). The 300, 600 and 800 ml/min dialysate flow rates were associated with statistically significant mean differences of −0.18 spKt/V (95% CI −0.22 to −0.15), 0.11 spKt/V (95% CI 0.02–0.19) and 0.08 spKt/V (95% CI 0.06–0.10), respectively, compared with the standard dialysate flow rate (500 ml/min). Although moderate heterogeneity was present, the associated CI was wide [I2 = 40.6% (95% CI 0.0–71.6), P = .0871). As such, inferences regarding the extent of heterogeneity should be made with caution. A visual representation of the network is presented in Supplementary Fig. 5, with results presented in Supplementary Fig. 6 and Supplementary Table 1.

DISCUSSION

In this systematic review of nine studies and meta-analysis of four studies involving 576 adult patients with kidney failure receiving maintenance HD, we found that increasing dialysate flow rates from 500 ml/min to 800 ml/min improved dialysis adequacy as measured by URR (3.38) and spKt/V (0.08). We also found in an exploratory analysis that increasing dialysate flow rates from 300 ml/min to 500 ml/min improved spKt/V by 0.18. There were limited data on other patient-important outcomes, so the impact of increasing dialysis flow rates on symptoms, cognition, physical function and quality of life is unclear. Results of the exploratory analysis generally supported those found in the primary analyses with respect to the positive relationship between dialysate flow rate and dialysis adequacy as measured by spKt/V.

This study addresses an important question with regards to the effectiveness of increasing dialysis flow rates to improve dialysis adequacy and its impact on patient-important outcomes in the clinical HD setting. Although dialyzer specification sheets provide information about small molecule clearance (e.g. urea, phosphate, creatinine, vitamin B12) across a range of dialysate flow rates in vitro, this systematic review and meta-analysis is the first of its kind to summarize the literature in vivo across a range a settings, populations and HD prescriptions. For example, an in vitro study of dialyzer mass transfer area coefficient (KoA) for urea was examined in 22 different models of commercial hollow fibre dialyzers (≈5 per model, total = 107) showed that KoA did not differ by blood flow, but when dialysate flow increased from 504 (6) to 819 (8) ml/min, urea KoA increased by 14 (7%) (range 3–33%) depending on the dialyzer model [35]. Kt/V targets are important to establish, as inadequate dialysis is associated with morbidity and mortality [5, 6, 36, 37]. A study in the USA found that many patients receiving facility-based HD were not meeting Kt/V targets. Underprescription of dialysis was identified as one of the barriers in this study, as many nephrologists were unaware of underdialysis because they did not explicitly calculate prescribed Kt/V [38]. Common aetiologies leading to underdialysis included vascular access dysfunction and inadequate HD treatment duration [39]. The updated 2015 Kidney Disease Outcomes Quality Initiative clinical practice guideline recommends a spKt/V target of 1.4 per HD session for patients treated thrice weekly, with a minimum delivered spKt/V of 1.2 (1b recommendation) that may be reduced in patients with significant residual native kidney function (not graded) [40]. A stdKt/V target of 2.3 per week with a minimum delivered stdKt/V of 2.1 is suggested in patients on HD schedules other than thrice weekly with incorporation of ultrafiltration or residual kidney function (not graded).

Kt/V targets may be achieved by altering many different aspects of the HD prescription, including dialysis frequency, duration, blood flow rate, dialysate flow rate, dialyzer size (including the use of dual dialyzers in parallel or series) and intradialytic exercise, but not dialysate temperature [41–44]. The relative contribution of each of these individual components and how they should be prioritized sequentially is unclear [7], but the HD prescription should be individualized for every patient depending on his/her needs, values and preferences, in addition to other system factors including facility scheduling availability, permitted duration of HD sessions, dialyzer costs and availability of intradialytic exercise [45]. A functional fistula, graft or central venous catheter is critical to optimize blood flow rates without any significant recirculation. Quality improvement programs consisting of feedback, educational workshops and materials and monitoring of dialysis time, dialyzer type, blood flow, dialysate flow and technical assistance have been shown to improve documented review of the dialysis prescription, dialysis at the prescribed treatment time and URR [46].

However, there are many barriers to increasing dialysis treatment time, including transportation; social issues; intradialytic symptoms such as cramping, nausea/vomiting and headache; personal preference and non-adherence [47, 48]. Using dialyzers with a larger surface area is common, especially in men, obesity and other settings with larger volume density, but this may still not be adequate in select patients depending on their HD prescription. Increasing dialysis flow rates from the common standard of 500 ml/min to 800 ml/min could increase spKt/V by 0.08 (95% CI 0.06–0.11) and URR by 3.38 (95% CI 1.97–4.78), with heterogeneity presumably explained on the basis of centre- and patient-related HD prescription factors [49, 50]. However, this benefit must be balanced by the economic and environmental impacts of increased dialysate flow rates. Assuming an HD session time of 4 hours, increasing the dialysate flow rate from 500 ml/min to 800 ml/min would result in an additional use of 72 l of dialysate per treatment and would result in >10 000 l of additional dialysate per year assuming a thrice weekly HD schedule [19]. Given the limited access to water in some countries that is anticipated to be exacerbated globally as climate change progresses, this may not be sustainable on a renal program level or even individual patient level from an environmental perspective unless reverse osmosis water is recycled or reused and other green nephrology initiatives are adopted [51–53]. Lastly, depending on the volume of dialysate required per HD session, consumable costs may be significantly increased as a result of additional bicarbonate or acid concentrates for online generation of dialysate, adding dialysis waste [54, 55].

It should be noted that there is a lack of literature comparing dialysis adequacy associated with a 600 ml/min dialysate flow rate versus 300, 500 or 800 ml/min. As such, there were limited direct comparisons for 600 ml/min in our exploratory network meta-analysis, contributing to low confidence in the point estimates. Nonetheless, the conclusions drawn from the exploratory analysis generally support those of the main analyses, suggesting a positive relationship between dialysate flow rate and adequacy.

The strengths of this systematic review and meta-analysis include its novelty, comprehensive search strategy developed by a medical librarian, inclusion of only RCTs where only dialysate flow rates were modified and other elements of the HD prescription were fixed, limiting confounding, and its focus on widely used clinical benchmarks of low, standard and high dialysate flow rates. Limitations include a small number of studies with small sample sizes and limited reporting of patient-important outcome measures related to dialysis adequacy such as HRQOL and symptom burden and no reporting of other markers of dialysis adequacy such as clearance of β2 microglobulin and other small or middle molecules. In addition, studies were low quality at a high or moderate risk of bias due to underreporting of randomization sequence generation, allocation concealment and blinding methods. Because there were no studies in a home HD population, the generalizability of our findings may not extend beyond facility-based HD. Furthermore, because of the small number of studies, we were unable to perform a meta-regression across the spectrum of HD prescriptions and therefore we cannot comment on the generalizability of the results to specific types of HD prescriptions. Finally, heterogeneity in results is likely explained by different study settings populations, HD prescriptions and urea kinetic modelling approaches that we were unable to explore in meta-regression due to the limited number of studies [56–60].

In summary, increasing dialysis flow rates in adults receiving maintenance HD improves dialysis adequacy as measured by urea kinetic modelling metrics including URR and spKt/V. The impact of these improvements on patient-important outcomes and the economic and environmental impact of increasing dialysis flow rates to 800 ml/min as compared with other strategies to increase dialysis adequacy needs further evaluation in high-quality prospective studies.

Supplementary Material

sfae163_Supplemental_File

Contributor Information

Yasmin Iman, Seven Oaks Hospital Chronic Disease Innovation Centre, Winnipeg, Manitoba, Canada.

Ryan Bamforth, Seven Oaks Hospital Chronic Disease Innovation Centre, Winnipeg, Manitoba, Canada.

Ruth Ewhrudjakpor, Seven Oaks Hospital Chronic Disease Innovation Centre, Winnipeg, Manitoba, Canada.

Paul Komenda, Seven Oaks Hospital Chronic Disease Innovation Centre, Winnipeg, Manitoba, Canada; Quanta Dialysis Technologies, Alcester, UK; University of Manitoba, Rady Faculty of Health Sciences, Department of Internal Medicine, Winnipeg, Manitoba, Canada.

Kelley Gorbe, Quanta Dialysis Technologies, Alcester, UK.

Reid Whitlock, Seven Oaks Hospital Chronic Disease Innovation Centre, Winnipeg, Manitoba, Canada.

Clara Bohm, Seven Oaks Hospital Chronic Disease Innovation Centre, Winnipeg, Manitoba, Canada; University of Manitoba, Rady Faculty of Health Sciences, Department of Internal Medicine, Winnipeg, Manitoba, Canada.

Navdeep Tangri, Seven Oaks Hospital Chronic Disease Innovation Centre, Winnipeg, Manitoba, Canada; University of Manitoba, Rady Faculty of Health Sciences, Department of Internal Medicine, Winnipeg, Manitoba, Canada.

David Collister, Seven Oaks Hospital Chronic Disease Innovation Centre, Winnipeg, Manitoba, Canada; University of Manitoba, Rady Faculty of Health Sciences, Department of Internal Medicine, Winnipeg, Manitoba, Canada; University of Alberta, Faculty of Medicine & Dentistry, Department of Medicine, Edmonton, Alberta, Canada.

FUNDING

This study was supported with funding from Quanta Dialysis Technologies.

DATA AVAILABILITY STATEMENT

The data underlying this article are available in the article itself.

CONFLICT OF INTEREST STATEMENT

P.K. is a consultant to Quanta Dialysis Technologies. C.B. has an ownership interest in Precision Advanced Digital Manufacturing. N.T. has equity in Quanta Dialysis Technologies and reports grants from Tricida, AstraZeneca, Janssen, Boehringer Ingelheim/Eli Lilly, Bayer, the Canadian Institutes for Health Research and the Kidney Foundation of Canada; personal fees from Tricida, AstraZeneca, Janssen, Boehringer Ingelheim/Eli Lilly, Bayer, Otsuka, Renibus, Roche, ClinPredict and Klinrisk; and other payments from ClinPredict, Klinrisk, Tricida and PulseData. ClinPredict and Klinrisk are engaged in efforts to develop and implement models for CKD progression in health systems. D.C. is funded by a KRESCENT New Investigator Award. Y.I., R.B., R.E., K.G. and R.W. report no conflicts of interest.

REFERENCES

  • 1. Daugirdas  JT, Blake  PG, Ing  TS, eds. Handbook of Dialysis. Fifth edition. Philadelphia: Lippincott Williams & Wilkins, 2015. [Google Scholar]
  • 2. Gotch  FA, Sargent  JA. A mechanistic analysis of the National Cooperative Dialysis Study (NCDS). Kidney Int  1985;28:526–34. [DOI] [PubMed] [Google Scholar]
  • 3. Eknoyan  G, Greene  T, Depner  TA  et al.  Effect of dialysis dose and membrane flux in maintenance hemodialysis. N Engl J Med  2002;347:2010–9. 10.1056/NEJMoa021583 [DOI] [PubMed] [Google Scholar]
  • 4. Daugirdas  JT. Kt/V (and especially its modifications) remains a useful measure of hemodialysis dose. Kidney Int  2015;88:466–73. 10.1038/ki.2015.204 [DOI] [PubMed] [Google Scholar]
  • 5. Rivara  MB, Ravel  V, Streja  E  et al.  Weekly standard Kt/Vurea and clinical outcomes in home and in-center hemodialysis. Clin J Am Soc Nephrol  2018;13:445–55. 10.2215/CJN.05680517 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6. Sehgal  AR, Dor  A, Tsai  AC. Morbidity and cost implications of inadequate hemodialysis. Am J Kidney Dis  2001;37:1223–31. 10.1053/ajkd.2001.24526 [DOI] [PubMed] [Google Scholar]
  • 7. Azar  AT. Increasing dialysate flow rate increases dialyzer urea clearance and dialysis efficiency: an in vivo study. Saudi J Kidney Dis Transplant  2009;20:1023–9. [PubMed] [Google Scholar]
  • 8. NxStage Medical . Hemodialysis at Home or in-Center. https://www.nxstage.com/ (1 March 2024, date last accessed). [Google Scholar]
  • 9. Outset Medical . Tablo Hemodialysis System. https://www.outsetmedical.com/tablo/ (1 March 2024, date last accessed)
  • 10. Quanta Dialysis Technologies . Quanta: OG Home. https://quantadt.com/us/ (1 March 2024, date last accessed).
  • 11. Clark  WR, Turk  JE. Approaches to quotidian dialysis: the NxStage System One. Semin Dial  2004;17:167–70. 10.1111/j.0894-0959.2004.17220.x [DOI] [PubMed] [Google Scholar]
  • 12. Plumb  TJ, Alvarez  L, Ross  DL  et al.  Safety and efficacy of the Tablo hemodialysis system for in-center and home hemodialysis. Hemodial Int  2020;24:22–8. 10.1111/hdi.12795 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13. Komenda  PVJ, Harper  G, Wilson  LM  et al.  Hemodialysis with the Quanta SC+: efficacy and safety of a self-care hemodialysis machine. Kidney Med  2020;2:724–31.e1. 10.1016/j.xkme.2020.07.007 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14. Gotch  F. The current place of urea kinetic modelling with respect to different dialysis modalities. Nephrol Dial Transplant  1998;13:10–4. 10.1093/ndt/13.suppl_6.10 [DOI] [PubMed] [Google Scholar]
  • 15. Gotch  FA. Evolution of the single-pool urea kinetic model. Semin Dial  2001;14:252–6. 10.1046/j.1525-139X.2001.00064.x [DOI] [PubMed] [Google Scholar]
  • 16. Daugirdas  JT. Factors that affect postdialysis rebound in serum urea concentration, including the rate of dialysis: results from the HEMO Study. J Am Soc Nephrol  2004;15:194–203. 10.1097/01.ASN.0000103871.20736.0C [DOI] [PubMed] [Google Scholar]
  • 17. Tattersall  JE, DeTakats  D, Chamney  P  et al.  The post-hemodialysis rebound: predicting and quantifying its effect on Kt/V. Kidney Int  1996;50:2094–102. 10.1038/ki.1996.534 [DOI] [PubMed] [Google Scholar]
  • 18. Leypoldt  JK. Urea standard Kt/Vurea for assessing dialysis treatment adequacy. Hemodial Int  2004;8:193–7. 10.1111/j.1492-7535.2004.01095.x [DOI] [PubMed] [Google Scholar]
  • 19. Daugirdas  JT, Depner  TA, Greene  T  et al.  Standard Kt/Vurea: a method of calculation that includes effects of fluid removal and residual kidney clearance. Kidney Int  2010;77:637–44. 10.1038/ki.2009.525 [DOI] [PubMed] [Google Scholar]
  • 20. Page  MJ, McKenzie  JE, Bossuyt  PM  et al.  The PRISMA 2020 statement: an updated guideline for reporting systematic reviews. BMJ  2021;372:n71. 10.1136/bmj.n71 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21. Rethlefsen  ML, Kirtley  S, Waffenschmidt  S  et al.  PRISMA-S: an extension to the PRISMA statement for reporting literature searches in systematic reviews. Syst Rev  2021;10:39. 10.1186/s13643-020-01542-z [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22. National Institute for Health and Care Research . PROSPERO: International prospective register of systematic reviews. https://www.crd.york.ac.uk/prospero (1 March 2024, date last accessed).
  • 23. Daugirdas  JT, Levin  NW, Kotanko  P  et al.  Comparison of proposed alternative methods for rescaling dialysis dose: resting energy expenditure, high metabolic rate organ mass, liver size, and body surface area. Semin Dial  2008;21:377–84. 10.1111/j.1525-139X.2008.00483.x [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 24. Higgins  JPT, Altman  DG, Gotzsche  PC  et al.  The Cochrane Collaboration's tool for assessing risk of bias in randomised trials. BMJ  2011;343:d5928. 10.1136/bmj.d5928 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 25. Higgins  JPT, Thomas  J, Chandler  J  et al. (eds). Cochrane Handbook for Systematic Reviews of Interventions Version 5.1.0.  Chapters 7.7.3.3 and 16.4.6.1. www.training.cochrane.org/handbook [Google Scholar]
  • 26. CRAN-R . The Comprehensive R Archive Network. https://cran.r-project.org/index.html (1 March 2024, date last accessed).
  • 27. Ahrenholz  P, Winkler  RE, Zendeh-Zartochti  D. The role of the dialysate flow rate in haemodialysis. In: Suzuki  H (ed.), Updates in Hemodialysis. InTech; 2015. 10.5772/58878 [DOI] [Google Scholar]
  • 28. Alayoud  A, Benyahia  M, Montassir  D  et al.  A model to predict optimal dialysate flow: a model to predict optimal dialysate flow. Ther Apher Dial  2012;16:152–8. 10.1111/j.1744-9987.2011.01040.x [DOI] [PubMed] [Google Scholar]
  • 29. Albalate  M, Pérez-García  R, de Sequera  P  et al.  Is it useful to increase dialysate flow rate to improve the delivered Kt?  BMC Nephrol  2015;16:20. 10.1186/s12882-015-0013-9 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 30. Azar  AT, Wahba  K, Mohamed  ASA  et al.  Association between dialysis dose improvement and nutritional status among hemodialysis patients. Am J Nephrol  2007;27:113–9. 10.1159/000099836 [DOI] [PubMed] [Google Scholar]
  • 31. Molano Trivino  A, Galván  Á, Wancjer Meid  B  et al.  SAT-342 Long term outcomes of lowering dialysate flow (Qd) in a population of chronic hemodialysis in RTS Colombia. Kidney Int Rep  2019;4(Suppl):S151–2. 10.1016/j.ekir.2019.05.390 [DOI] [Google Scholar]
  • 32. Panagoutsos  SA, Yannatos  EV, Passadakis  PS  et al.  The clinical impact of increasing the hemodialysis dose. Hemodial Int  2001;5:51–4. 10.1111/hdi.2001.5.1.51 [DOI] [PubMed] [Google Scholar]
  • 33. Wang  W, Tonelli  M, Hemmelgarn  B  et al.  The effect of increasing dialysis dose in overweight hemodialysis patients on quality of life: a 6-week randomized crossover trial. Am J Kidney Dis  2008;51:796–803. 10.1053/j.ajkd.2007.12.031 [DOI] [PubMed] [Google Scholar]
  • 34. Ward  RA, Idoux  JW, Hamdan  H  et al.  Dialysate flow rate and delivered Kt/Vurea for dialyzers with enhanced dialysate flow distribution. Clin J Am Soc Nephrol  2011;6:2235–9. 10.2215/CJN.02630311 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 35. Leypoldt  JK, Cheung  AK, Agodoa  LY  et al.  Hemodialyzer mass transfer-area coefficients for urea increase at high dialysate flow rates. Kidney Int  1997;51:2013–7. 10.1038/ki.1997.274 [DOI] [PubMed] [Google Scholar]
  • 36. Maduell  F, Ramos  R, Varas  J  et al.  Hemodialysis patients receiving a greater Kt dose than recommended have reduced mortality and hospitalization risk. Kidney Int  2016;90:1332–41. 10.1016/j.kint.2016.08.022 [DOI] [PubMed] [Google Scholar]
  • 37. Béguin  L, Krummel  T, Longlune  N  et al.  Dialysis dose and mortality in haemodialysis: is higher better?  Nephrol Dial Transplant  2021;36:2300–7. 10.1093/ndt/gfab202 [DOI] [PubMed] [Google Scholar]
  • 38. Sehgal  AR, Leon  JB, Siminoff  LA  et al.  Improving the quality of hemodialysis treatment: a community-based randomized controlled trial to overcome patient-specific barriers. JAMA  2002;287:1961. 10.1001/jama.287.15.1961 [DOI] [PubMed] [Google Scholar]
  • 39. Coyne  D, Delmez  J, Spence  G  et al.  Impaired delivery of hemodialysis prescriptions: an analysis of causes and an approach to evaluation. J Am Soc Nephrol  1997;8:1315–8. [DOI] [PubMed] [Google Scholar]
  • 40. Daugirdas  JT, Depner  TA, Inrig  J  et al.  KDOQI clinical practice guideline for hemodialysis adequacy: 2015 update. Am J Kidney Dis  2015;66:884–930. 10.1053/j.ajkd.2015.07.015 [DOI] [PubMed] [Google Scholar]
  • 41. Santoro  A. Confounding factors in the assessment of delivered hemodialysis dose. Kidney Int  2000;58(Suppl 76):S19–27. 10.1046/j.1523-1755.2000.07603.x [DOI] [PubMed] [Google Scholar]
  • 42. Eloot  S, Van Biesen  W, Dhondt  A  et al.  Impact of hemodialysis duration on the removal of uremic retention solutes. Kidney Int  2008;73:765–70. 10.1038/sj.ki.5002750 [DOI] [PubMed] [Google Scholar]
  • 43. Mitchell  RL, Webster  AC, Craig  JC. Cochrane Renal Group Report. Am J Kidney Dis  2004;44:768–72. 10.1053/j.ajkd.2004.07.002 [DOI] [Google Scholar]
  • 44. Kirkman  DL, Scott  M, Kidd  J  et al.  The effects of intradialytic exercise on hemodialysis adequacy: a systematic review. Semin Dial  2019;32:368–78. 10.1111/sdi.12785 [DOI] [PubMed] [Google Scholar]
  • 45. Jones  CB, Chan  CT. Boundaries of frequency and treatment time in conventional hemodialysis: balancing convenience, economics, and health outcomes. Semin Dial  2018;31:537–43. 10.1111/sdi.12742 [DOI] [PubMed] [Google Scholar]
  • 46. McClellan  WM. A randomized evaluation of two health care quality improvement program (HCQIP) interventions to improve the adequacy of hemodialysis care of ESRD patients: feedback alone versus intensive intervention. J Am Soc Nephrol  2004;15:754–60. 10.1097/01.ASN.0000115701.51613.D7 [DOI] [PubMed] [Google Scholar]
  • 47. Haraldsson  B. Higher Kt/V is needed for adequate dialysis if the treatment time is reduced. Nephrol Dial Transplant  1995;10:1845–51. 10.1093/ndt/10.10.1845 [DOI] [PubMed] [Google Scholar]
  • 48. Sehgal  A, Snow  R, Singer  M  et al.  Barriers to adequate delivery of hemodialysis. Am J Kidney Dis  1998;31:593–601. 10.1053/ajkd.1998.v31.pm9531174 [DOI] [PubMed] [Google Scholar]
  • 49. Fink  JC, Blahut  SA, Briglia  AE  et al.  Effect of center- versus patient-specific factors on variations in dialysis adequacy. J Am Soc Nephrol  2001;12:164–9. 10.1681/ASN.V121164 [DOI] [PubMed] [Google Scholar]
  • 50. Tangri  N, Tighiouart  H, Meyer  KB  et al.  Both patient and facility contribute to achieving the Centers for Medicare and Medicaid Services’ pay-for-performance target for dialysis adequacy. J Am Soc Nephrol  2011;22:2296–302. 10.1681/ASN.2010111137 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 51. Burnier  M, Fouque  D. Global warming applied to dialysis: facts and figures. Nephrol Dial Transplant  2021;36:2167–9. 10.1093/ndt/gfab260 [DOI] [PubMed] [Google Scholar]
  • 52. Agar  JWM. Reusing and recycling dialysis reverse osmosis system reject water. Kidney Int  2015;88:653–7. 10.1038/ki.2015.213 [DOI] [PubMed] [Google Scholar]
  • 53. Tarrass  F, Benjelloun  M, Benjelloun  O  et al.  Water conservation: an emerging but vital issue in hemodialysis therapy. Blood Purif  2010;30:181–5. 10.1159/000321485 [DOI] [PubMed] [Google Scholar]
  • 54. Lee  H, Manns  B, Taub  K  et al.  Cost analysis of ongoing care of patients with end-stage renal disease: the impact of dialysis modality and dialysis access. Am J Kidney Dis  2002;40:611–22. 10.1053/ajkd.2002.34924 [DOI] [PubMed] [Google Scholar]
  • 55. Agar  JWM. Green dialysis: the environmental challenges ahead. Semin Dial  2015;28:186–92. 10.1111/sdi.12324 [DOI] [PubMed] [Google Scholar]
  • 56. Vanholder  R, Glorieux  G, Eloot  S. Once upon a time in dialysis: the last days of Kt/V?  Kidney Int  2015;88:460–5. 10.1038/ki.2015.155 [DOI] [PubMed] [Google Scholar]
  • 57. Perl  J, Dember  LM, Bargman  JM  et al.  The use of a multidimensional measure of dialysis adequacy—moving beyond small solute kinetics. Clin J Am Soc Nephrol  2017;12:839–47. 10.2215/CJN.08460816 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 58. Steyaert  S, Holvoet  E, Nagler  E  et al.  Reporting of “dialysis adequacy” as an outcome in randomised trials conducted in adults on haemodialysis. PLoS One  2019;14:e0207045. 10.1371/journal.pone.0207045 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 59. Daugirdas  JT, Tattersall  JE. Automated monitoring of hemodialysis adequacy by dialysis machines: potential benefits to patients and cost savings. Kidney Int  2010;78:833–5. 10.1038/ki.2010.218 [DOI] [PubMed] [Google Scholar]
  • 60. Tattersall  J. Do we need another Kt/V?  Nephrol Dial Transplant  2013;28:1963–6. 10.1093/ndt/gft215 [DOI] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

sfae163_Supplemental_File

Data Availability Statement

The data underlying this article are available in the article itself.