Skip to main content

This is a preprint.

It has not yet been peer reviewed by a journal.

The National Library of Medicine is running a pilot to include preprints that result from research funded by NIH in PMC and PubMed.

ArXiv logoLink to ArXiv
[Preprint]. 2024 Jun 26:arXiv:2406.18426v1. [Version 1]

Fast 3D 31P B1+ mapping with a weighted stack of spiral trajectory at 7 Tesla

Mark Widmaier 1,2,3, Antonia Kaiser 1,2, Salome Baup 1,2, Daniel Wenz 1,2, Katarzyna Pierzchala 1,2, Ying Xiao 1,2,3, Zhiwei Huang 1,2, Yun Jiang 5, Lijing Xin 1,2,4,*
PMCID: PMC11230352  PMID: 38979490

Abstract

Purpose:

Phosphorus Magnetic Resonance Spectroscopy (31P MRS) enables non-invasive assessment of energy metabolism, yet its application is hindered by sensitivity limitations. To overcome this, often high magnetic fields are used, leading to challenges such as spatial B1+ inhomogeneity and therefore the need for accurate flip angle determination in accelerated acquisitions with short repetition times TR). In response to these challenges, we propose a novel short TR and look-up table-based Double-Angle Method for fast 3D 31P B1+ mapping (fDAM).

Methods:

Our method incorporates 3D weighted stack of spiral gradient echo acquisitions and a frequency-selective pulse to enable efficient B1+ mapping based on the phosphocreatine signal at 7T. Protocols were optimised using simulations and validated through phantom experiments. The method was validated in phantom experiments and skeletal muscle applications using a birdcage 1H/31P volume coil.

Results:

The results of fDAM were compared to the classical DAM (cDAM). A good correlation (r=0.94) was obtained between the two B1+ maps. A 3D 31P B1+ mapping in the human calf muscle was achieved in about 10 min using a birdcage volume coil, with a 20% extended coverage relative to that of the cDAM (24 min). fDAM also enabled the first full brain coverage 31P 3D B1+ mapping in approx. 10 min using a 1 Tx/ 32 Rx coil.

Conclusion:

fDAM is an efficient method for 31P 3D B1+ mapping, showing promise for future applications in rapid 31P MRSI.

Introduction:

Phosphorus (31P) Magnetic Resonance Spectroscopic Imaging (MRSI) enables in vivo probing of energy metabolism1,2. Although it was demonstrated that 31P MRSI can be a useful method to study e.g. cancer3, neuropsychiatric4,5 and neurodegenerative diseases68, its clinical application is constrained by long acquisition times resulting from inherently low 31P NMR sensitivity. To partially address this issue, ultra-high magnetic fields (UHF), such as 7T, are used1,912. However, higher B0 results in a less homogenous transmit field B1+13,14, and consequentially in a higher uncertainty of the estimated flip angle (FA) distribution. The FA distribution is critical to assess the concentration changes of different metabolites using 31P MRSI in a reliable manner12,1521. For this purpose, in vivo B1+-mapping, which can be quite time-consuming, is performed. Therefore, it is of highest relevance to reduce the acquisition time needed for this step.

Various B1+ mapping methods for 1H MRI have previously been proposed, including the Double-Angle Method (DAM)22,23, SA2RAGE24, saturated DAM25, actual flip angle method (AFI)26, and phase-dependent methods27,28 such as Bloch Siegert (BS) shift method2931. However, their direct translation to 31P or other X-nuclei B1+ mapping is challenging due to differences in sensitivity, relaxation properties, and deviating resonance peaks at various chemical shifts. Initial adaptations of some of these methods have been successfully applied in 31P B1+ mapping, such as DAM32, AFI15, and BS33.

Nevertheless, these methods exhibit several drawbacks. DAM, from now on referred to as classical DAM (cDAM), is often considered as the standard reference method, but is hampered by prolonged acquisition times due to the necessity of long TR5T1 required for a fully relaxed condition. Especially 31P metabolites exhibit extended T1 relaxation times, leading to prolonged acquisition times, limited spatial resolution, or potential errors when using reduced TRs. Short TR DAM methods have been proposed for 1H, relying on a steady state acquisition. However, the method of Ishimori et al. is limited to a range close to the nominal flip angle, leading to bias in the B1 estimation, and has not been demonstrated in vivo34. The other method from Bouhrara et al. is challenging to apply in X-Nuclei imaging suffering from low SNR with proposed flip angles25. Other methods, as the recently proposed 2D dual-TR35, the AFI15 method and a 3D BS-based method33 are also operating in steady state to reduce the acquisition time. Even though the BS method has proven to be a promising method as it is independent of T1 relaxation times, it needs a dedicated sequence, with an additional Fermi pulse. This pulse introduces additional RF power deposition and prolonged TE, which might lead to sensitivity loss due to weighting in the effective transversal relaxation time. In the dual-TR and BS method, the localisation is based on a chemical shift imaging (CSI) readout, which limits the spatial and temporal resolution. In a different work, a 31P AFI MRSI approach15 was implemented using the phosphocreatine (PCr) signal by applying a frequency-selective pulse. However, the approximation of a much shorter TR,1 and TR,2 compared to T1 comes at costs of SNR with a flip angle (FA) of 60°15,26 or if TR,2 is close to T1 with a bias in the B1+ estimation. In all presented works, validation of the method was only performed by surface coils to overcome the low signal-to-noise ratio (SNR), which limits the application to a small surface area.

To address these problems, we propose a short TR and look-up table based fast DAM (fDAM) for time-efficient 31P B1+ mapping. 3D weighted stack of spiral gradient echo (GRE) acquisitions and a frequency selective pulse were incorporated to enable an efficient 31P B1+-mapping approach based on the PCr signal at 7T using a 31P/1H birdcage volume coil and a 31P/1H head coil array.

Methods:

Parameter optimisation

The cDAM involves the acquisition of two images, one with flip angle (FA) α1 and the other with FA α222,23. The achieved FA αe can be estimated by

αe=acosSα22Sα1. (1)

This formula can be derived for the asymptotic value of TR. The signal intensity Sα of one image obtained using a GRE sequence is hereby described as

Sα=Kρ(1exp(σ))sin(α)1cos(α)exp(σ)expTET2*, (2)

where K is a scaling factor, ρ is the spin density, T2* the apparent relaxation time, σ=TR/T1, and TE is the echo time. The normalised signal intensity ratio r between the two GRE images can be described as

r=α1Sα2α2Sα1=α1sinCB1α2α2sinCB1α11cosCB1α1exp(σ)1cosCB1α2exp(σ), (3)

where CB1 is the unitless B1+ scaling factor and can be interpreted as the scaling between the actual and nominal flip angle. CB1 is used as metric in this work to interpretate the results. CB1 can be transformed to B1+ values, using the hard pulse equivalent of the same nominal α_1 in radians by

B1+=CB1α1,rad1060.431TpγμT, (4)

where Tp is the pulse length, γ the gyromagnetic ratio and the factor 0.431 is the hard pulse transformation factor for the Gaussian pulse used in this work. As TR,α1 and α2 are known for a given T1 value, a look-up table can be generated connecting different signal ratios r to the B1+ scaling factor CB1. The factor α1/α2 in Eq. 3 normalises the look-up table to a maximum value of 1. Figure 1b displays different signal ratios for different σ at different actual FAs α1CB1). The estimated FA is determined by searching the closest value of the measured signal ratio r in the look-up table. Throughout this work, we set α2=2α1, as in the cDAM approach.

Figure 1:

Figure 1:

(a) The function between the actual FA and the estimated FA by cDAM is plotted for different σ values. When using a small σ, cDAM develops an increasing bias in the FA estimation. The signal ratio r can be used to generate look up tables connecting it to a specific FA. In (b) look up tables for different σ values are shown for fDAM. Values of the objective function O over a range of α1 and σ is shown in (c). The optimal α1 for a given σ and vice-versa is indicated by the grey line. At the maximum value of the optimisation function, indicated by the white square, the optimal parameters α1,opt and σopt are found.

For an optimal α1, three factors are considered in this work: 1) temporal SNR gain Gα1 of the GRE for α1; 2) Gα2 of the GRE for α2 and 3) the sensitivity of the look-up table. Gα is the temporal SNR gain relative to that of a GRE at fully relaxed condition with σ=5 and α1=90°:

Gα=SNRα(σ,α)SNR905,90=Sα(σ,α)S90(5,90)5σ. (5)

The sensitivity of the look-up table is defined as

s=α1drdα1. (6)

The multiplication of α1 hereby scales the derivative to make s independent of α1. As a detected 1° change, choosing α1=10° is a 10% change in B1+, whereas it is only a 2% change for a α1=50°. To achieve the same sensitivity as α1=50° the derivative when choosing α1=10° must be 5 times larger. The values of the objective function O for the optimization are found by a multiplication of all 3 attributes,

O=Gα1Gα2s (7)

and is shown in Figure 1c. The optimal α1 for a given σ is found by the maximum value of O and is indicated in Figure 1c. The overall optimal α1 was found at α1,opt=59° and σopt=1.6.

31P MRI sequence and reconstruction

Phantom and in vivo calf muscle experiments were conducted on a 7T/68 cm MR scanner (Siemens Healthineers, Erlangen, Germany) with a 28 cm diameter 31P/1H circularly polarized (CP) birdcage coil (Stark Contrast, Erlangen, Germany). Additionally, an in vivo brain experiment was performed on a 7T Terra X system (Siemens Healthineers, Erlangen, Germany) with a double tuned Tx/Rx 31P/1H birdcage coil (26 cm diameter) and a 32 channel 31P-Rx only phased array (Rapid Biomedical, Rimpar, Germany). The 3D spiral 31P GRE sequence diagram (version VB17 and XA60 available for sequence transfer via c2p agreement) is depicted in the Figure 2. An 8 ms frequency selective Gaussian pulse with a bandwidth of 340Hz was used to excite phosphocreatine (PCr). For the VB17A version a 21.08 ms variable-density spiral trajectory36 (Figure 2c) after a TE=200μs was deployed. Due to gradient stimulation limits the spiral trajectory had to be redesigned for the XA60 version. The slew rate limits were reduced from 150 T/(m s)−1 to 100 T/(m s) −1. The duration of the non-uniform spiral for the XA60 version is 18.72 ms and a TE=410μs (Figure 2d). The encoding along kz was achieved by a stack of spiral with a Hamming weighted averaging scheme (Figure 2b). One 3D k-space sampling was achieved with a total of 13 kz encoded spiral readouts. The total acquisition time was, therefore, TA=2nrep13TR, with nrep the number of repetitions of a 3D volume acquisition for SNR enhancement and the factor 2 as two GRE images must be sampled. The matrix size of the reconstructed image space was 16 × 16 × 11 in phantom and calf muscle and 32 × 32 × 11 in human brain for a 230 × 230 × 220 mm3 field of view (FOV). The vector size of one spiral without the rewinding is L=2061 (VB17A) and L=1782 (XA60), respectively. The uniform spiral trajectories of both versions, with characteristic dense sampling in the k-space centre, provided sufficient k-space coverage for a single shot in-plane sampling, with a mean oversampling ratio of 8 for the VB17A and 1.7 for the XA60 version, respectively. All data was processed using a custom script in MATLAB (The MathWorks, Inc., Natick, Massachusetts, USA). A 1D-FFT was applied along slice dimension before regridding. The density compensation function (DCF) was calculated based on the Voronoi diagram3739. Before regridding, the DCF was multiplied by a low-pass filter (LP)

LP1n=cos2π*nLenL, (8)

a combination of a half periodic Hann filter and an exponential filter to reduce the influence of high-frequency components on the image acquired in the phantom and calf muscle. To cope with the lower SNR in the brain, a stronger smoothing filter,

LP2(n)=cos2π*n900 (9)

was used and the remaining L900 points were zero filled. The regridding was performed using a Kaiser-Bessel kernel40,41 to transform the data in to the Cartesian k-space. Lastly, a 2D-FFT transformed the GRE data into the image domain. The data acquired with the 1Tx/32Rx coil was processed for each channel separately and combined using adaptive combine42 in the image domain. The GRE images are then used for CB1 estimations as described above.

Figure 2:

Figure 2:

(a) Schematics of the 31P spiral GRE sequence. (b) The hamming weighted acquisition results with different numbers of averages per kz plain, (c) the spiral readout trajectory for the human calf muscle (VB17A: 21.08 ms, 10 μs sampling rate) and (d) brain (XA60: 18.72 ms, 10 μs sampling rate) experiment.

Phantom validation

The previously described method was validated in a spherical phantom with a diameter of 17 cm, filled with 50 mM Pi solution (Carl Roth GmbH & Co. KG, Karlsruhe, Germany). The longitudinal relaxation time was measured by an inversion recovery method (T1=7.2s).CB1 maps were acquired for σ=2,1.6,1,0.5 with α1=63°,59°,53°,40° and TR=14.4,11.52,7.2,3.6s respectively, to examine the FA optimisation dependent on σ. The acquisition time per GRE was fixed to 13:26 min including dummy scans (4, 5, 8, 16 RF pulse repetitions respectively). The repetitions of one 3D acquisition for each σ were adjusted accordingly with nrep=4,5,8,16. The reference voltage Vref was set to 250 V. B1+ maps acquired by the cDAM (σ=5,α1=65°) were used as reference images (nrep=2,16:30 min acquisition time (TA)). To evaluate the difference between the CB1 maps obtained by fDAM and cDAM, the difference maps were generated by calculating the percentage difference per voxel in slice 6. For σopt=1.6 the estimation performance was further evaluated by the Band-Altman plot and Pearson correlations. To calculate the SNR in a voxel, the phantom was masked, and the signal intensity of the voxel was divided by the noise. The noise is defined as the standard deviation (STD) over all voxels inside the mask of the difference between the first and second half of the acquired averages43.

T1 effect

The accuracy of the look-up table approach depends on the knowledge of σ and thus T1. To investigate the effect of wrong T1 assumptions, simulations were conducted. A look-up table for σopt=1.6 and α1,opt=59° was calculated, with CB1 was ranging from 0.02 to 2.6 (0.02/step). Simulated signals with CB1 ranging from 0.1 to 2.0 (0.02/step) and a deviation of σ ranging from −50% to 50% (0.1%/step) were processed in the estimation process. The estimation error was calculated as the percentage difference of the estimated CB1 from the actual CB1. Further, the simulation of T1 effect was validated in the phantom results for σopt=1.6 by altering the assumed T1 in the look-up table to 50%, 75%, 125%, 150% of the actual T1. The error maps were calculated from the relative difference of these CB1 maps from those obtained with the actual measured T1=7.2s. The T1 effect is also CB1 dependent according to the simulation results. Therefore, two different CB1 values were used in the phantom experiments by conducting experiments at two different Vref=250V and Vref=175V.

In vivo validation in the calf muscle

In vivo data were acquired in the calf muscle of two participants (2 female, 28 and 29 years old), with written informed consent provided. fDAM CB1 maps were acquired, and cDAM CB1 maps were acquired as reference. For the fDAM method the sequence parameters are set with σopt=1.6,(TR=5.7s, assuming T1 of PCr is 3.55 s10,15), α1=59°,α2=118° and nrep=6 (TA 15:45 min; including 5 dummy scans for each GRE). The cDAM CB1 maps were acquired with α1=65°,nrep=3,TR=18s, with a TA of 23:25 min. For both sequences, the transmit voltage was set to 300 V. To investigate a potential TA reduction from the fDAM with nrep=6, subsets of nrep=4 and nrep=2 were taken to generate the CB1 maps “Rep. 4” and “Rep. 2”, respectively. The scan time of these reduced data sets resemble a TA of 10:50 min and 5:55 min respectively. Data was compared with relative difference maps, where fDAM CB1 maps with nrep=6 were taken as references. Due to the low sensitivity of the birdcage coil in the peripheral regions, only voxels with an SNR > 3 were considered (Monte Carlo simulation in the supplementary files: Figure S1). The acquired maps were then interpolated to 32 × 32 × 11 and masked with accordingly down sampled 1H GRE images (128 × 128 × 11, TE=3.37ms,TR=15ms).

In vivo validation in the human brain

In addition to the calf muscle data, CB1 maps in the brain were acquired by fDAM from one participant (male,43 years old, written informed consent provided). The sequence parameters of the fDAM were adapted to the brain with TR=5.4 s, (assuming the T1 of PCr is 3.4 s9,44), α1=59°,α2=118°(nrep=6; TA 15:00 min; including 5 dummy scans for each GRE). Reference cDAM CB1 maps were acquired with α1=65°,nrep=3,TR=18s and a TA of 23:25 min. For both sequences the transmit voltage was set to 400 V. A potential TA reduction from the fDAM with nrep=6 was investigated as described for the calf muscle. The scan time of the subsamples resemble a TA of 10:15 min for nrep=4 and 5:35 min for nrep=2. The acquired maps were masked with manually created brain masks from 1H GRE images (128 × 128 × 11, TE=3.37ms,TR=15ms).

Results

Simulation results

The effect of σ<5 using the cDAM (Equation 1) is shown in Figure 1a. As σ decreases, FAs are more likely to be overestimated, identifying the need for a correction method. Figure 1b illustrates our look-up table approach, successfully connecting the signal ratio r to an actual FA α=CB1α1 for a given σ. To investigate the optimal parameters, values of the optimisation function were plotted as a function of α1 and σ (Figure 1c). With the respective σ, an optimal α1 was found (the grey line in Figure 1c). The highest value of the optimisation function was found at σopt=1.6 (white squared dot in Figure 1c) with α1,opt=59°.

Phantom Validation

The results of the phantom validation are displayed in Figure 3. Figure 3a demonstrates the CB1 maps of cDAM and fDAM with different σ (2.0, 1.6, 1.0, 0.5 for fDAM) and the corresponding CB1 relative difference (Diff.) [%] maps between cDAM and fDAM. The CB1 maps are in good agreement as indicated by the difference plots which are not exceeding 10% (Figure 3b). Note that the difference is lower in the centre of the slice where the SNR and CB1 are increased. With the optimal α1,opt=59° and σopt=1.6, fDAM was compared over all slices with cDAM using voxel-wise Pearson correlations (Figure 3c) and the mean differences were compared using a Bland-Altman plot (Figure 3d). The estimated CB1 obtained by both methods were in excellent agreement with a Pearson correlation coefficient of 0.95. The Bland-Altman plot showed that fDAM demonstrates a small mean overestimation of 0.01 and 95% limits of agreement (1.96 STD) of ± 0.05.

Figure 3:

Figure 3:

Comparison of CB1 maps acquired by cDAM and fDAM in phantom experiments. (a) CB1 maps of the centre slice from cDAM and fDAM of different σ and (b) the difference maps with cDAM as a reference. (c) Correlation plot and (d) BA plot to compare CB1 maps of fDAM with σopt=1.6 and cDAM over voxels of all slices. In (c) and (d), the colour bar shows the SNR of each voxel.

T1 effect

Figure 4a illustrates the simulation results of the influence of T1 values on the CB1 estimation error [%] using α1,opt=59° and σopt=1.6. For CB1>1, the estimation error was negligible (<10%). For CB1<1,CB1 is overestimated when T1 is underestimated and vice versa. Overall, the CB1 bias was within ±25% with a ±50% deviation in T1 and within ±15% with a ±25% deviation in T1.

Figure 4:

Figure 4:

Evaluation of the introduced bias in CB1 estimation due to T1 deviation using the optimised α1,opt and σopt. (a) The simulation results and (b) the phantom results for two different transmit voltages (175V and 250V) and the estimation error [%] when using deviated T1 values (from −50% to 50%) in the look-up table. (c) Comparison of the phantom results for a voxel in the centre marked in orange (175V: CB1=0.81) and blue (250V: CB1=1.15) with the corresponding simulation results.

To validate the impact of T1 on CB1 estimation, additional phantom experiments using the optimal α1,opt=59° and σopt=1.6 were performed with two sets of reference voltages (175V and 250V). The CB1 values were estimated using look-up tables simulated with biased T1 values of ±25% and ± 50%. The CB1 estimation errors relative to the cDAM with the correct T1 are shown in Figure 4b. A voxel was picked to compare the phantom results with the simulations (Figure 4c). For the reference voltage of 250V, CB1 is estimated to be 1.15. Both results indicate that the bias does not exceed 10% in a ±50% deviation range of T1. For the reference voltage of 175V, CB1 is estimated to be 0.81 and the bias does not exceed 17% in a ±50% range and 10% in a ±25% deviation range of T1. The error maps are in good agreement with the simulation results. Note that with smaller CB1, the estimation errors are more sensitive to T1 bias.

In vivo validation in the calf muscle

1H GRE images, 31P GRE images with α1 and α2 in a human calf muscle (left and right leg), cDAM and fDAM CB1 maps, as well as their difference maps, are shown in Figure 5. Slices 3 to 10 are shown due to missing coil sensitivity coverage in slice 1, 2 and 11. cDAM and fDAM showed comparable results, especially in the centre slices, where the coil sensitivity is high. fDAM demonstrated an overall higher SNR coverage (SNR >3) compared to cDAM. In most regions, the deviation between the two methods did not exceed 15%. The regions with large difference had lower signals, as seen in the 31P GRE images.

Figure 5:

Figure 5:

In vivo calf muscle images of participant 1 (a) and 2 (b), including the 1H GRE images, the 31P GRE1 images (for α1), the 31P GRE2 images (for α2), the CB1 maps of cDAM and fDAM, as well as the difference maps (fDAM-cDAM)100/fDAM. Note that for the CB1 and difference maps only voxels with SNR>3 are displayed.

The influence of SNR on CB1 estimation was evaluated in Figure 6. When decreasing the number of repetitions (Rep.), the overall image coverage was reduced. fDAM with nrep=4 showed at least 20% more coverage as cDAM in just 40% of the TA of cDAM. Regions with low signal sensitivity demonstrated a larger difference. Using just nrep=2 (6 min), a good estimation of CB1 could be achieved in most regions in the centre slices with deviation below 15% from that measured with nrep=6.

Figure 6:

Figure 6:

Comparison of CB1 maps acquired by fDAM (nrep=6; 16 min) with those measured with fewer repetitions (nrep=4; 11 min and nrep=4; 6min) and their respective difference maps (participant 1 (a) and 2 (b)). The 3rd and 5th rows show the percentage difference maps of (Rep. 4 - Rep. 6)100/ Rep. 6 and (Rep. 2 - Rep. 6)100/ Rep. 6, respectively.

In vivo results in the human brain

The 1H GRE images, 31P GRE images with α1 and α2, the CB1 maps of fDAM, cDAM and fDAM with nrep=4 and nrep=2 are shown in Figure 7. Their respective difference maps, taking fDAM with nrep=6 as reference, are in the line below the CB1 maps. Slices 9 to 11 are not displayed as no sensitive tissue is covered in these slices. cDAM and fDAM show comparable CB1 maps demonstrated in the difference maps with < 20% in most regions. nrep=4 is in good agreement not exceeding a difference of 20%, indicating a possible scan time reduction to 11:15 min. With further reduction to Rep. 2 (5:35 min), differences in the CB1 maps are increasing in number and severity, however, still doesn’t exceed 20% in most regions.

Figure 7:

Figure 7:

Comparison of CB1 maps acquired by fDAM (nrep=6; 14:30 min) with those measured with less repetitions (nrep=4; 10:15 min and nrep=2; 5:30 min) and their respective difference maps. The reference 1H GRE images, the 31P GRE1 and GRE2 images with nrep=6 were shown on the top. The 6th row shows the percentage difference maps of (fDAM-cDAM)100/fDAM and the 8th and 10th rows show the percentage difference maps of (Rep. 4-Rep. 6)100/Rep. 6 and (Rep. 2-Rep. 6)100/Rep. 6, respectively.

Discussion:

In this work, we demonstrated the effectiveness of fDAM, a short TR double-angle B1+ mapping approach with a look-up table to compensate for T1 saturation effects. An optimised TR=1.6T1 and α1=59° was found by applying an optimisation function. The fDAM was combined with an efficient weighted stack of spiral acquisition and a frequency selective pulse for fast 3D acquisition. The method performance was evaluated in a phantom and the results showed a good correlation (r=0.94) with those of cDAM. For the first time, using a birdcage coil, a 3D CB1-map in the human calf muscle was achieved in 10:30 min, with a 20% extended coverage relative to that of the cDAM (23:30 min). Additionally, fDAM enabled the first full brain coverage 31P 3D CB1-map at 7T in just 10:15 min using a 1 channel Tx/ 32 channel Rx coil. The CB1 maps showed a typical birdcage coil behavior at high fields with higher B1 values in the center than the peripheral. For GRE1, the mean B1+ was 16 μT, 15 μT and 14 μT for participant 1 (calf muscle), participant 2 (calf muscle), and participant 3 (brain), respectively. The related histograms are shown in the supplementary files (Figure S2). The two coils showed differences in transmit efficiency, as the brain experiment was driven with a 33% higher transmit voltage than the muscle experiment. This could attribute to the degraded transmit efficiency due to the accommodation of the 32-channel receive arrays. The reported values and performance45 stayed in agreement with previous reports of a whole body coil (10.4 μT)46.

CSI is commonly used as a technique for 31P B1+ mapping, offering extra chemical shift information, yet impacting the time efficiency of B1+ mapping. Our proposed method uses a frequency selective pulse combined with the hamming weighted stack of spiral readout, which results in an efficient acquisition. The resulting temporal SNR benefit allows for a higher (relative to DAM32, dual-TR35, BS33) or similar (to AFI15) spatial resolution in a shorter TA relative to existing methods. Note that previous methods have been demonstrated on surface or quadrature coils known for their high local sensitivity, whereas fDAM is demonstrated in the calf muscle on a less sensitive 31P/1H single-channel birdcage coil. The short acquisition time of fDAM enables its integration in the scanning protocol to obtain a subject-specific CB1 map. In this work, we demonstrate the method for the 31P nucleus, while the look-up table approach and optimal parameters can also be applied to other nuclei and GRE types of sequences with fast readout trajectory.

As brain PCr concentration is by a factor 8 lower than in the calf muscle9,16, the sensitivity is much lower, making B1+ map measurement even more challenging. In this work, we took use of two countermeasures: 1) a 1 Tx/ 32 Rx coil to increase the sensitivity and 2) an LP with a lower cut off-frequency (LP2). Note that such LP comes with additional smoothing, which reduces the actual resolution. However, B1+ field changes smoothly across space, rarely giving the necessity for high resolution CB1 measurements.

GRE sequences are known to be sensitive to B0 inhomogeneities47, and are subject to signal loss in areas with strong susceptibility gradients such as the tissue air surface close to the sinuses. The use of an ultra-short TE as in the current implementation will ameliorate this effect. A frequency selective pulse is further increasing the sensitivity due to B0 offset. Areas with large resonance offset may lead to inaccurate CB1 estimation due to partial excitation of PCr or signal bleeding from other metabolites. Future work could target this issue by incorporating B0 maps to correct the CB1 estimations.

The proposed method shares a similar limitation as other methods (DAM32, dual-TR35, AFI15,26), being sensitive to T1 values. In this study, a uniform T1 value was used based on values reported in the literature. Prior knowledge of tissue-specific T1 values could enhance CB1 mapping accuracy; however, such information is often unavailable. We therefore investigated the impact of T1 on CB1 accuracy using simulations and phantom results. In a rather rare and worst case (CB1=0.1), the bias of CB1 estimation was below 15% for a ± 25% difference in T1. In a more likely scenario, phantom and simulation studies indicated that the CB1 estimation bias was below 10% in a ± 25% range of T1(CB1=0.81). When CB1 increases, estimation bias was less sensitive to T1, e.g., below 10% in a ± 50% range of T1 for CB1=1.15. This limitation can be further alleviated by increasing the CB1 (i.e., using a higher transmit voltage), as the bias introduced by T1 is much lower for higher CB1 values (>1.25) as shown in Figure 4a. With the 8 ms Gaussian pulse used in this implementation, in vivo SAR limitations were far from being reached (2–3%), which allows the use of a higher transmit voltage. In studies with strong variation of T1 (> 50%), a T1 insensitive method such as BS-based methods29,30,33 might be more favourable. The current fast readout sequence could be further adapted by including a Femi pulse for this purpose.

fDAM operates within SAR and peripheral nerve stimulation limits on a clinically approved UHF scanner. With the SAR limits far from being exceeded, the method could be considered even for nuclei with a lower gyromagnetic ratio. The strength of this method is the easy-to-implement look-up table approach. In combination with the fact that GRE sequences (or Spin Echo sequences) are widely available as standard sequences, fDAM is easy to use for other researchers or clinicians. To make the application more accessible, our sequence is available for Siemens XA60 and VB17A via c2p. The gradient readout can be easily changed using an external text file, and the sequence can be adapted to other body parts, nuclei, or readout trajectories. The reconstruction pipeline with the look-up table estimation is freely available on GitHub (https://github.com/MaSteWi/fDAM-B1-mapping.git). We expect that the proposed strategy along with recent advances in MR fingerprinting12, novel RF coils48 and denoising techniques49,50, will contribute to significant reductions in total acquisition times making 31P MRSI more suitable for clinical settings.

Conclusion

A fast short-TR double angle B1+ mapping method including a weighted stack of spiral trajectory and a frequency selective pulse was successfully implemented and demonstrated for 31P B1+ mapping in the human calf muscle and the brain. The novel method allows single subject 31P CB1 mapping in both human calf muscle and brain in 10 min, with a spatial resolution of 14 × 14 × 20 mm3 and 7 × 7 × 20 mm3 respectively, showing promise for future applications in rapid X-nuclei imaging. This method can be applied to other fast imaging techniques available with the proposed look-up table approach.

Supplementary Material

Supplement 1

Acknowledgment

We acknowledge access to the facilities and expertise of the CIBM Center for Biomedical Imaging, a Swiss research center of excellence founded and supported by Lausanne University Hospital (CHUV), University of Lausanne (UNIL), Ecole Polytechnique Federale de Lausanne (EPFL), University of Geneva (UNIGE), and Geneva University Hospitals (HUG). Open access funding provided by EPFL.

Funding information

This work was supported by the Swiss National Science Foundation (grant no. 320030_189064 and 213769). Yun Jiang was partly supported by the NIH/NCI grant R37CA263583 and by Siemens Healthineers.

Footnotes

Conflict of interest statement

Yun Jiang receives research grant support from Siemens Healthineers.

References

  • 1.Huang Z, Gambarota G, Xiao Y, Wenz D, Xin L. Apparent diffusion coefficients of 31P metabolites in the human calf muscle at 7 T. Magn Reson Mater Phys Biol Med. 2023;36(2):309–315. doi: 10.1007/s10334-023-01065-3 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2.Xin L, Ipek Ö, Beaumont M, et al. Nutritional Ketosis Increases NAD+/NADH Ratio in Healthy Human Brain: An in Vivo Study by 31P-MRS. Front Nutr. 2018;5. Accessed July 1, 2022. 10.3389/fnut.2018.00062 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Klomp DWJ, van de Bank BL, Raaijmakers A, et al. 31P MRSI and 1H MRS at 7 T: initial results in human breast cancer. NMR Biomed. 2011;24(10):1337–1342. doi: 10.1002/nbm.1696 [DOI] [PubMed] [Google Scholar]
  • 4.Skupienski R, Steullet P, Do KQ, Xin L. Developmental changes in cerebral NAD and neuroenergetics of an antioxidant compromised mouse model of schizophrenia. Transl Psychiatry. 2023;13(1):1–9. doi: 10.1038/s41398-023-02568-2 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Forlenza OV, Wacker P, Nunes PV, et al. Reduced phospholipid breakdown in Alzheimer’s brains: a 31P spectroscopy study. Psychopharmacology (Berl). 2005;180(2):359–365. doi: 10.1007/s00213-005-2168-8 [DOI] [PubMed] [Google Scholar]
  • 6.Du F, Yuksel C, Chouinard VA, et al. Abnormalities in High-Energy Phosphate Metabolism in First-Episode Bipolar Disorder Measured Using 31P-Magnetic Resonance Spectroscopy. Biol Psychiatry. 2018;84(11):797–802. doi: 10.1016/j.biopsych.2017.03.025 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.Du F, Cooper AJ, Thida T, et al. In Vivo Evidence for Cerebral Bioenergetic Abnormalities in Schizophrenia Measured Using 31P Magnetization Transfer Spectroscopy. JAMA Psychiatry. 2014;71(1):19–27. doi: 10.1001/jamapsychiatry.2013.2287 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Yuksel C, Chen X, Chouinard VA, et al. Abnormal Brain Bioenergetics in First-Episode Psychosis. Schizophr Bull Open. 2021;2(1):sgaa073. doi: 10.1093/schizbullopen/sgaa073 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9.Lei H, Zhu XH, Zhang XL, Ugurbil K, Chen W. In vivo 31P magnetic resonance spectroscopy of human brain at 7 T: An initial experience. Magn Reson Med. 2003;49(2):199–205. doi: 10.1002/mrm.10379 [DOI] [PubMed] [Google Scholar]
  • 10.Bogner W, Chmelik M, Schmid A i., Moser E, Trattnig S, Gruber S. Assessment of 31P relaxation times in the human calf muscle: A comparison between 3 T and 7 T in vivo. Magn Reson Med. 2009;62(3):574–582. doi: 10.1002/mrm.22057 [DOI] [PubMed] [Google Scholar]
  • 11.Valkovič L, Chmelík M, Just Kukurova I, et al. Time-resolved phosphorous magnetization transfer of the human calf muscle at 3T and 7T: A feasibility study. Eur J Radiol. 2013;82(5):745–751. doi: 10.1016/j.ejrad.2011.09.024 [DOI] [PubMed] [Google Scholar]
  • 12.Widmaier M, Lim SI, Wenz D, Xin L. Fast in vivo assay of creatine kinase activity in the human brain by 31P magnetic resonance fingerprinting. NMR Biomed. 2023;36(11):e4998. doi: 10.1002/nbm.4998 [DOI] [PubMed] [Google Scholar]
  • 13.Yang QX, Wang J, Zhang X, et al. Analysis of wave behavior in lossy dielectric samples at high field. Magn Reson Med. 2002;47(5):982–989. doi: 10.1002/mrm.10137 [DOI] [PubMed] [Google Scholar]
  • 14.Tofts PS. Standing Waves in Uniform Water Phantoms. J Magn Reson B. 1994;104(2):143–147. doi: 10.1006/jmrb.1994.1067 [DOI] [Google Scholar]
  • 15.Parasoglou P, Xia D, Chang G, Regatte RR. 3D-mapping of phosphocreatine concentration in the human calf muscle at 7 T: Comparison to 3 T. Magn Reson Med. 2013;70(6):1619–1625. doi: 10.1002/mrm.24616 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16.Kemp GJ, Meyerspeer M, Moser E. Absolute quantification of phosphorus metabolite concentrations in human muscle in vivo by 31P MRS: a quantitative review. NMR Biomed. 2007;20(6):555–565. doi: 10.1002/nbm.1192 [DOI] [PubMed] [Google Scholar]
  • 17.Buchli R, Boesiger P. Comparison of methods for the determination of absolute metabolite concentrations in human muscles by 31P MRS. Magn Reson Med. 1993;30(5):552–558. doi: 10.1002/mrm.1910300505 [DOI] [PubMed] [Google Scholar]
  • 18.Buchli R, Martin E, Boesiger P. Comparison of calibration strategies for the in vivo determination of absolute metabolite concentrations in the human brain by 31P MRS. NMR Biomed. 1994;7(5):225–230. doi: 10.1002/nbm.1940070505 [DOI] [PubMed] [Google Scholar]
  • 19.Dunn JF, Kemp GJ, Radda GK. Depth selective quantification of phosphorus metabolites in human calf muscle. NMR Biomed. 1992;5(3):154–160. doi: 10.1002/nbm.1940050309 [DOI] [PubMed] [Google Scholar]
  • 20.Doyle VL, Payne GS, Collins DJ, Verrill MW, Leach MO. Quantification of phosphorus metabolites in human calf muscle and soft-tissue tumours from localized MR spectra acquired using surface coils. Phys Med Biol. 1997;42(4):691. doi: 10.1088/0031-9155/42/4/006 [DOI] [PubMed] [Google Scholar]
  • 21.Bottomley PA, Atalar E, Weiss RG. Human cardiac high-energy phosphate metabolite concentrations by 1D-resolved NMR spectroscopy. Magn Reson Med. 1996;35(5):664–670. doi: 10.1002/mrm.1910350507 [DOI] [PubMed] [Google Scholar]
  • 22.Insko EK, Bolinger L. Mapping of the Radiofrequency Field. J Magn Reson A. 1993;103(1):82–85. doi: 10.1006/jmra.1993.1133 [DOI] [Google Scholar]
  • 23.Pohmann R, Scheffler K. A theoretical and experimental comparison of different techniques for B1 mapping at very high fields. NMR Biomed. 2013;26(3):265–275. doi: 10.1002/nbm.2844 [DOI] [PubMed] [Google Scholar]
  • 24.Eggenschwiler F, Kober T, Magill AW, Gruetter R, Marques JP. SA2RAGE: A new sequence for fast B1+-mapping. Magn Reson Med. 2012;67(6):1609–1619. doi: 10.1002/mrm.23145 [DOI] [PubMed] [Google Scholar]
  • 25.Bouhrara M, Spencer RG. Steady State Double Angle Method for Rapid B1 Mapping. Magn Reson Med. 2019;82(1):189–201. doi: 10.1002/mrm.27708 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 26.Yarnykh VL. Actual flip-angle imaging in the pulsed steady state: A method for rapid three-dimensional mapping of the transmitted radiofrequency field. Magn Reson Med. 2007;57(1):192–200. doi: 10.1002/mrm.21120 [DOI] [PubMed] [Google Scholar]
  • 27.Morrell GR. A phase-sensitive method of flip angle mapping. Magn Reson Med. 2008;60(4):889–894. doi: 10.1002/mrm.21729 [DOI] [PubMed] [Google Scholar]
  • 28.Jy P. B_1 mapping using phase information created by frequency-modulated pulses. Proc 16th Annu Meet ISMRM Tor Can 2008. 2008;361. Accessed November 14, 2023. https://cir.nii.ac.jp/crid/1574231875896909056 [Google Scholar]
  • 29.Bloch F, Siegert A. Magnetic Resonance for Nonrotating Fields. Phys Rev. 1940;57(6):522–527. doi: 10.1103/PhysRev.57.522 [DOI] [Google Scholar]
  • 30.Sacolick LI, Wiesinger F, Hancu I, Vogel MW. B1 mapping by Bloch-Siegert shift. Magn Reson Med. 2010;63(5):1315–1322. doi: 10.1002/mrm.22357 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 31.Khalighi MM, Rutt BK, Kerr AB. Adiabatic RF pulse design for Bloch-Siegert B mapping. Magn Reson Med. 2013;70(3):829–835. doi: 10.1002/mrm.24507 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 32.Greenman RL, Rakow-Penner R. Evaluation of the RF field uniformity of a double-tuned 31P/1H birdcage RF coil for spin-echo MRI/MRS of the diabetic foot. J Magn Reson Imaging. 2005;22(3):427–432. doi: 10.1002/jmri.20372 [DOI] [PubMed] [Google Scholar]
  • 33.Clarke WT, Robson MD, Rodgers CT. Bloch-Siegert -mapping for human cardiac 31P-MRS at 7 Tesla. Magn Reson Med. 2016;76(4):1047–1058. doi: 10.1002/mrm.26005 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 34.Ishimori Y, Shimanuki T, Kobayashi T, Monma M. Fast B1 Mapping Based on Double-Angle Method with T1 Correction Using Standard Pulse Sequence. J Med Phys. 2022;47(1):93–98. doi: 10.4103/jmp.jmp_78_21 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 35.Chmelík M, Považan M, Jírů F, et al. Flip-angle mapping of 31P coils by steady-state MR spectroscopic imaging. J Magn Reson Imaging. 2014;40(2):391–397. doi: 10.1002/jmri.24401 [DOI] [PubMed] [Google Scholar]
  • 36.Lee JH, Hargreaves BA, Hu BS, Nishimura DG. Fast 3D imaging using variable-density spiral trajectories with applications to limb perfusion. Magn Reson Med. 2003;50(6):1276–1285. doi: 10.1002/mrm.10644 [DOI] [PubMed] [Google Scholar]
  • 37.Rasche V, Proksa R, Sinkus R, Bornert P, Eggers H. Resampling of data between arbitrary grids using convolution interpolation. IEEE Trans Med Imaging. 1999;18(5):385–392. doi: 10.1109/42.774166 [DOI] [PubMed] [Google Scholar]
  • 38.Aurenhammer F. Voronoi diagrams—a survey of a fundamental geometric data structure. ACM Comput Surv. 1991;23(3):345–405. doi: 10.1145/116873.116880 [DOI] [Google Scholar]
  • 39.Ong Meng Sang. Arbitrary Square Bounded Voronoi Diagram. MATLAB Central File Exchange. Published January 19, 2024. Accessed January 19, 2024. https://www.mathworks.com/matlabcentral/fileexchange/30353-arbitrary-square-bounded-voronoi-diagram [Google Scholar]
  • 40.Jackson JI, Meyer CH, Nishimura DG, Macovski A. Selection of a convolution function for Fourier inversion using gridding (computerised tomography application). IEEE Trans Med Imaging. 1991;10(3):473–478. doi: 10.1109/42.97598 [DOI] [PubMed] [Google Scholar]
  • 41.Beatty PJ, Nishimura DG, Pauly JM. Rapid gridding reconstruction with a minimal oversampling ratio. IEEE Trans Med Imaging. 2005;24(6):799–808. doi: 10.1109/TMI.2005.848376 [DOI] [PubMed] [Google Scholar]
  • 42.Walsh DO, Gmitro AF, Marcellin MW. Adaptive reconstruction of phased array MR imagery. Magn Reson Med. 2000;43(5):682–690. doi: [DOI] [PubMed] [Google Scholar]
  • 43.Determination of Signal-to-Noise Ratio (SNR) in Diagnostic Magnetic Resonance Imaging. NEMA. Published April 30, 2015. Accessed January 4, 2024. https://www.nema.org/Standards/view/Determination-of-Signal-to-Noise-Ratio-in-Diagnostic-Magnetic-Resonance-Imaging [Google Scholar]
  • 44.Ren J, Sherry AD, Malloy CR. 31P-MRS of healthy human brain: ATP synthesis, metabolite concentrations, pH, and T1 relaxation times. NMR Biomed. 2015;28(11):1455–1462. doi: 10.1002/nbm.3384 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 45.Tomanek B, Volotovskyy V, Gruwel MLH, McKenzie E, King SB. Double-frequency birdcage volume coils for 4.7T and 7T. Concepts Magn Reson Part B Magn Reson Eng. 2005;26B(1):16–22. doi: 10.1002/cmr.b.20038 [DOI] [Google Scholar]
  • 46.Valkovič L, Dragonu I, Almujayyaz S, et al. Using a whole-body 31P birdcage transmit coil and 16-element receive array for human cardiac metabolic imaging at 7T. PLoS ONE. 2017;12(10):e0187153. doi: 10.1371/journal.pone.0187153 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 47.Wang J, Mao W, Qiu M, Smith MB, Constable RT. Factors influencing flip angle mapping in MRI: RF pulse shape, slice-select gradients, off-resonance excitation, and B0 inhomogeneities. Magn Reson Med. 2006;56(2):463–468. doi: 10.1002/mrm.20947 [DOI] [PubMed] [Google Scholar]
  • 48.Wenz D., Dardano T., Widmaier M., Lim S., Huang Z., Xin L.. Degenerate birdcage coil mixed with bent dipole antennas to enhance central SNR in phosphorus MRI/MRS of human brain at 7T. In: Proc. Intl. Soc. Mag. Reson. Med. 31 (2023). ; 2023. [Google Scholar]
  • 49.Lustig M, Donoho DL, Santos JM, Pauly JM. Compressed Sensing MRI. IEEE Signal Process Mag. 2008;25(2):72–82. doi: 10.1109/MSP.2007.914728 [DOI] [Google Scholar]
  • 50.Mosso J, Simicic D, Şimşek K, Kreis R, Cudalbu C, Jelescu IO. MP-PCA denoising for diffusion MRS data: promises and pitfalls. NeuroImage. 2022;263:119634. doi: 10.1016/j.neuroimage.2022.119634 [DOI] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

Supplement 1

Articles from ArXiv are provided here courtesy of arXiv

RESOURCES