Abstract
It has been rarely reported that a stuck optical coherence tomography (OCT) catheter can lead to longitudinal stent deformation (LSD). This complication can result in incomplete stent apposition and dissection after stent implantation. In this study, we present a case where a bailout stent was implanted in the distal segment of the left anterior descending artery (LAD) after longitudinal stent deformation caused by a stuck OCT catheter. This approach was taken to prevent acute stent thrombosis, subacute stent thrombosis, in-stent restenosis (ISR), and death. The patients were followed up for one year, and no adverse events were observed.
Keywords: longitudinal stent deformation, optical coherence tomography, complication
1. Introduction
Coronary longitudinal stent deformation (LSD) is a rare complication occurring in approximately 0.2%–0.68% of cases [1, 2]. LSD refers to the local distortion of an implanted coronary stent along its longitudinal axis during coronary intervention, leading to a shortening of the stent by more than 2 mm [3]. Angiography reveals a ring-shaped segment at the site of LSD, caused by overlapping stent beams, often referred to as the “organ effect” or “accordion effect”. Several independent risk factors have been identified for LSD, including complex coronary arteries (with features such as coronary artery calcification, coronary artery tortuosity, long lesions, and open lesions), prolus element stents, postdilatation balloons, and intravascular imaging techniques like intravenous ultrasound (IVUS) [4]. IVUS or optical coherence tomography (OCT) can accurately detect stent beam collapse, malapposition, dissection, and thrombus formation at the LSD segment. It is important to note that not all cases of LSD are benign, as approximately 14% of patients experience serious adverse events including the need for emergency surgery due to a stuck device, acute stent thrombosis, subacute stent thrombosis, in-stent restenosis (ISR), and even death [5]. Limited clinical studies suggest that re-implantation of a stent or postdilatation treatment following the occurrence of LSD may reduce the incidence of adverse events.
2. Materials and Methods
The keywords “longitudinal stent deformation”, “longitudinal stent compression”, “stent deformation”, “concertina”, “stent shortening”, and “accordion” were used for literature search on PubMed. Our study was conducted based on previous studies and a review of LSD cases (Table 1, Ref. [1, 6, 7, 8, 9, 10, 11, 12, 13, 14, 15, 16, 17, 18, 19, 20, 21]). Furthermore, we thoroughly examined the existing literature on LSD.
Table 1.
Longitudinal stent deformation case reports.
Author | Patient’s Age | Causes | LSD stents | LSD position | Disposal methods | Follow-up |
Hernández-Matamoro et al. [16] | 76 | Balloon of stent | Resolute Onyx | Proximal | Stent | Uk |
Kobayashi et al. [9] | 73 | Uk | Platinum chromium | Proximal | No | 18 months: VST |
Si et al. [6] | 60 | Balloon | Promus Element | Proximal | Post-dilation | 12 months: normal |
Buysschaert et al. [15] | 74 | Balloon | Promus Element | Proximal | No | 24 months: VST |
Aung et al. [14] | 41 | Balloon | Promus Element | Proximal | Post-dilation | Uk |
Foerst et al. [13] | 69 | Balloon | Promus Element | Proximal | Post-dilation | 4 days: normal |
Leong et al. [11] | 69 | Rotawire | Synergy | Distal | Stent | Uk |
Panoulas et al. [10] | 78 | Buddy wire | Resolute Onyx | Proximal | Stent | Uk |
85 | Wire | Resolute Onyx | Proximal | Stent | Uk | |
Demir et al. [18] | 67 | Pressure wire | Xience Pro | Proximal | Stent | Uk |
54 | Pressure wire | Resolute Onyx | Proximal | Stent | Uk | |
Vijayvergiya et al. [17] | 70 | Balloon | Promus Element | Proximal | Post-dilation | 3 ddays: subacute ST |
Cuenza et al. [7] | 57 | Uk | Resolute Onyx | Proximal | Stent | Uk |
Yıldırım et al. [8] | 60 | EBU 3.5 | Promus Element | Proximal | CABG | 5 days: normal |
Yew et al. [12] | 60 | Balloon | Promus Elemen | Proximal | Stent | 3 months: normal |
Shannon et al. [20] | 55 | IVUS | Promus Element | Proximal | Post-dilation | Uk |
62 | Balloon | Promus Element | Proximal | No | Uk | |
68 | Wire | Promus Element | Proximal | Stent | Uk | |
70 | Wire | Promus Element | Proximal | Stent | 6 months: normal | |
76 | Balloon | Promus Element | Proximal | Stent | Uk | |
Hanratty et al. [19] | 50 | AL | Promus Element | Proximal | Stent | Uk |
72 | IVUS | Biomatrix | Proximal | Stent | Uk | |
78 | EBU | Resolute Integrity | Proximal | Post-dilation | Uk | |
Cosgrove et al. [21] | 67 | Reverse RG3 | Synergy | Distal | Stent | 12 months: normal |
Williams et al. [1] | 50 | Balloon | Promus Element | Proximal | Stent | 4 months: normal |
63 | Balloon | Promus Element | Proximal | Post-dilation | Uk | |
55 | Catheter | Promus Element | Proximal | Post-dilation | 4 months: normal | |
63 | Balloon | Promus Element | Proximal | Stent | Uk | |
83 | Catheter | Biomatrix | Proximal | No | 2 months: subacute ST | |
72 | Catheter | TAXUS | Proximal | Stent | 6 months: normal | |
55 | Balloon | Promus Element | Proximal | Stent | Uk | |
63 | Catheter | Promus Element | Proximal | Stent | Uk | |
82 | Balloon | Endeavor | Proximal | Stent | Uk |
uk,unknown; CABG, coronary artery bypass grafting; ST, stent thrombosis; VST, very late stent thrombosis; AL, Amplaze guiding; LSD, longitudinal stent deformation; IVUS, intravenous ultrasound.
3. Case Report
A 42-year-old male patient was admitted to the hospital and diagnosed with coronary artery disease, acute inferior wall myocardial infarction, and type 2 diabetes mellitus. Prior to admission, he underwent coronary angiography at a local hospital and thrombus aspiration was performed in the blocked circumflex artery without stent implantation. On admission, the examination revealed sinus rhythm with the presence of Q waves in leads II, III, and Augmented Vector Foot (aVF). Cardiac ultrasound demonstrated mild tricuspid valve regurgitation and a left ventricular ejection fraction (LVEF) of 63%.
Elective transradial coronary angiography (CAG) showed borderline lesions and suspicious niche in the distal segment of the left main (LM) coronary artery, borderline lesions in the proximal to the middle segment of the left anterior descending artery (LAD), and diffused stenosis in the proximal to the middle segment of the circumflex artery (LCX) (up to 90%), but no significant stenosis was observed in the right coronary artery (Fig. 1A,B). OCT (75 mm pullback, St. Jude Medical, Little Canada, MN, USA) guided stents implantation (6F EBU3.5 guiding catheter, RUNTHROUGH in LAD and SION in LCX) was sequentially performed in the proximal LCX (3.0 29 mm, Promus PREMIER, Boston Scientific, Marlborough, MA, USA) and LM-LAD (2.75 33 mm, 4.0 18 mm, Promus PREMIER, Boston Scientific, USA). The proximal segment of LCX stent was about 2 mm away from the ostial of LCX indicating no connection between stents of LAD and LCX. Angiography showed complete stent apposition with TIMI grade III (Fig. 1C). OCT pullback of the LAD showed stent malapposition and stent dissection at the distal segment of the LAD. However, the presence of the guidewire artifact posed challenges in accurately measuring the depth, length, and angle of the dissection (Fig. 2A). Therefore, a re-pullback of OCT was performed to evaluate whether a bailout stent was required at the distal segment of the LAD stent during which the guidewire supporting the OCT catheter was retracted to the proximal LAD in order to minimize any potential artifact influence (Fig. 1E). Scanning revealed incomplete stent apposition at the distal segment of LAD stent and no bailout stent was needed (Fig. 2B).
Fig. 1.
Percutaneous coronary intervention procedure. (A) CAG showed borderline lesions and suspicious niche in the distal segment of the LM (white arrow). (B) CAG showed diffused stenosis in the proximal to the middle segment of the LCX (white arrow). (C) Image of the stents after implantation of one LCX stent and two LAD stents, the distal segment of the LAD stents seemed properly expanded in CAG (white arrow). (D) CAG showed the lumen at the edge of the stent and at the middle segment of the LAD decreased in size (white arrow). (E) Wireless OCT catheter in LAD (white arrow indicates retracted guidewire). (F) OCT catheter stuck in retracting (white arrow indicates the position of stuck). (G) LM-LAD stents before stuck of OCT catheter, red dashed line depicted the length of LM-LAD stents. (H) Shortened LM-LAD stents after retraction of stuck OCT catheter, red dashed line depicted the length of LM-LAD stents. (I) CAG after bailout stent implantation in LAD. (J) The length from the ostium of the LAD to the stent of the LM before LSD is 12.9 mm. (K) The length from the ostium of the LAD to the stent of the LM after LSD is 8.1 mm. CAG, coronary angiography; LM, left main artery; LCX, circumflex artery; LAD, left anterior descending artery; OCT, optical coherence tomography.
Fig. 2.
Longitudinal stent deformation at the proximal segment of the left main coronary artery caused by a stuck optical coherence tomography (OCT) catheter. (A) Before the catheter being stuck, OCT shows the possibility of dissection under the condition of the guidewire artifact. (B) Under the condition of no guidewire support, OCT shows no obvious dissection in the distal part of the stent in the middle segment of the LAD and the yellow arrows show the malapposition of stent beam. (C) After the stuck OCT catheter was removed, OCT shows incomplete stent apposition. (D) After another stent was inserted into the mid-distal segment of the LAD and post-dilatation, OCT shows single-layer stent beam in the 9 o’clock-2 o’clock direction and triple-layer stent mesh in the 3 o’clock-6 o’clock direction, indicating that is the area where stent was placed after passing through the previous stent mesh. LAD, left anterior descending artery; OCT, optical coherence tomography.
In retracting the wireless OCT catheter, it stuck and could not be removed. Several attempts to advance, rotate, and withdraw the OCT catheter failed (Fig. 1F). After repeatedly adjusting the OCT catheter, it popped out of the artery together with the guiding catheter. The OCT catheter creased in the distal second marker segment.
Next, the guiding catheter was reintroduced. The angiography showed that the stent at the proximal segment of the left main coronary artery was shortened, and the “wedding ring sign” was observed (Fig. 1H). In addition, the lumen at the edge of the stent and inthe middle segment of the LAD decreased in size, with suspected dissection or thrombus (Fig. 1D). A bailout 2.5 15 mm stent (Promus PREMIER, Boston Scientific, USA) was implanted to overlap the distal part of the stent in the middle segment of the LAD, based on the finding of dissections, white thrombus and a decreased lumen were observed with a new OCT catheter (Fig. 1I). The OCT showed that the stent was well apposed to the wall after post-dilatation, and no dissection was observed in the distal segment of the LAD (Fig. 2C,D). OCT also revealed LSD at the proximal segment of the left main coronary artery, accompanied by stent edge dissection and stent malapposition. The 4.0 mm 8 mm post-dilatation balloon (Quantum Maverick, Boston Scientific, USA) was reintroduced to post-dilate the stent at the left main coronary artery at 20 atmosphere (atm). The length from the ostium of the LAD to the stent of the LM shortened from 12.9 mm before LSD to 8.1 mm on OCT (Fig. 1G,H,J,K).
Tirofiban was given postoperatively, and no acute stent thrombosis occurred. In one-year follow-up, the patient did not experience any episode of angina attacks or acute myocardial infarction.
4. Literature Review
LSD refers to the distortion of the coronary stent along its longitudinal axis, either shortening or elongation. In brief, LSD occurs when the implanted coronary stent undergoes local deformation along its long axis, resulting in a variation exceeding 2 mm [3]. It is a rare complication of coronary stent interventions, with an incidence ranging from 0.2% to 0.68% [1, 2]. LSD can be visualized in CAG as a ring-like image caused by the overlapping stent beams and shortening along the long axis of the stent. This phenomenon has been described in literature as the “organ effect” or “accordion effect”. However, it is important to note that the incidence of LSD may be underestimated due to the limitations of CAG in diagnosing this condition. The ability of X-rays to penetrate different coronary stents can vary, and the clinical experience of physicians also plays a role in accurate diagnosis. In a study of coronary computed tomography (CT) angiography [22], the incidence of LSD was found to be 2.6%. Furthermore, the incidence of LSD may increase if IVUS or OCT is routinely performed after coronary stenting. If LSD is suspected after a detailed analysis of angiography after stenting, further diagnostic procedures such as IVUS or OCT should be considered to provide a more definitive diagnosis and guide subsequent interventional management [6, 7].
Compared to the first-generation drug-eluting stents (DES), the second-generation stents are thinner, resulting in reduced longitudinal strength and a 13-fold higher risk of LSD [22]. Among the second-generation DES, the Promus Element stands out with the least longitudinal compression strength due to its thin platinum-based stent and the offset peak-to-peak stent design, featuring two connecting rod racks [2, 23]. Studies conducted by Prabhu et al. [23] and Ormiston et al. [24] have demonstrated that the Promus Element has the highest incidence of LSD among second-generation drug-coated stents.
Approximately 89% of LSD cases can be attributed to the following mechanisms. Firstly, LSD may occur during the insertion or removal of interventional devices such as balloons, guidewires, rota devices, and IVUS catheters, as these devices can exert pressure on the stent. Secondly, it can occur during the removal of interventional devices when the guide catheter or extension catheter applies relative forces that squeeze the stent. A study by Rhee et al. [4] focused on predicting LSD and its clinical outcomes and identified several independent risk factors for LSD, including lesions in the LM or its opening, treatment of bifurcation lesions with double stents or jailed balloon, interventions performed at a distant part of the stent, IVUS examination, and postdilatation. Other related studies have also highlighted the potential contribution of interventional therapy for bridging vessels, jailed guidewires, pressure guidewires, rota wires, protection devices for the distal vascular segment, and non-slip element (NSE) balloons to the occurrence of LSD. However, there have been no reports in the literature of OCT catheters leading to LSD [3, 25]. Additionally, the use of NSE balloons increases the risk of LSD.
Chung et al. [22] concluded that LSD caused by DES was not associated with the development of major adverse cardiovascular events. However, it is important to note that this finding may be influenced by the limited number of patients included in the study. In contrast, other studies have reported that not all cases of LSD result in benign symptoms, with 14% of patients experiencing serious adverse events. These adverse events can include the need for heart bypass surgery due to the inability to remove the lodged medical device after LSD [8], acute stent thrombosis, subacute stent thrombosis, very late stent thrombosis [9], ISR, and death. Limited clinical studies have suggested that re-stenting or postdilatation management after the occurrence of LSD may reduce the incidence of adverse events associated with LSD. Based on some expert experiences, whether “rota ablation” of the deformed stent allows for better wall apposition of the additionally placed stent and further reduces adverse events in LSD remains unclear. Therefore, more reliable clinical study data is needed to address these aspects. Additionally, there is a lack of strong clinical research evidence regarding whether patients with LSD require longer dual antiplatelet therapy (DAPT) or lifelong DAPT. It is worth noting that very late in-stent thrombosis has been reported in patients with LSD even after one year of DAPT. In the absence of clinical evidence, endovascular imaging techniques such as OCT can be utilized to assess stent neointimal coverage, stent apposition, and dissection. This information can help guide the selection of appropriate antithrombotic drug therapy for individual patients (Table 1).
This rare case of a stuck OCT catheter, which causes LSD of the LM, was primarily due to the sharp protrusion of the non-guidewire-track-supported OCT catheter at both ends of the guidewire exit. The sharp protrusion was stuck in the tortuous middle segment of the anterior descending artery and stent beam with incomplete apposition, resulting in a stuck OCT catheter. Despite repeated attempts to forcefully remove the OCT catheter, the condition worsened, resulting in incomplete stent apposition in the middle segment of the anterior descending artery and a coronary artery dissection, ultimately leading to acute in-stent thrombosis. Additionally, the withdrawal of the OCT catheter caused deep insertion of the guide catheter, which affected the stent in the proximal segment of the left main coronary artery, causing LSD and exposing the previously covered dissection. When attempting to re-deliver the guidewire to the anterior descending artery, it was unable to pass through the stent mesh due to severe incomplete stent apposition in the middle segment. As a result, dilatation and stent implantation were performed again in the middle to distal segments. Fortunately, further examination with OCT revealed that the dissection in the left main coronary artery did not penetrate deep into the smooth muscle layer and had a small angle. Therefore, there was no need for an additional stent in the proximal segment. The original stent in the left main coronary artery was post-dilated again using a post-dilatation balloon, and the patient received postoperative intensive antiplatelet therapy. The patient was followed up for one year, and no complications were observed.
5. Conclusions
Due to the absence of support from the guidewire track, the OCT catheter became angled at the exit, resulting in sharp protrusions on both sides. These protrusions became trapped by the stent beam, resulting in malapposition of the stent, which caused the OCT catheter to become lodged. This further exacerbated the issue of stent malapposition. Furthermore, LSD was observed in the proximal segment of the left main coronary artery, exposing the previously covered stent dissection.
This case was introduced to warn people of these conditions. LSD may occur when OCT is performed without guidewire track support, particularly in tortuous lesions, LM opening lesions, long lesions, and lesions with incomplete stent apposition. For patients at risk of LSD, physicians should maintain the guidewire coaxial, avoid inserting the guidewire deep into the muscle layer and using excessive strength, and place the guide catheter at a proper distance from the proximal segment of the stent. When they push in or remove the OCT catheter, stent balloon, and postdilatation balloons, an imaging device must be used, or the protection guidewire or balloon must be removed. When LSD is suspected, OCT or IVUS must be performed as accurately as possible to carry out re-implantation of the stent or re-postdilatation.
Acknowledgment
Not applicable.
Abbreviations
AL, Amplaze guiding; CABG, coronary artery bypass grafting; DAPT, dual antiplatelet therapy; DES, drug-eluting stent; ISR, in-stent restenosis; IVUS, intravenous ultrasound; LAD, left anterior descending artery; LCX, circumflex artery; LM, left main artery; LSD, longitudinal stent deformation; OCT, optical coherence tomography; ST, stent thrombosis; VST, very late stent thrombosis.
Footnotes
Publisher’s Note: IMR Press stays neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Availability of Data and Materials
Data and materials are available on request.
Author Contributions
MZ and XTP performed literature research and drafted the manuscript. YBW, SY, PB and LG edited the manuscript before submission. All authors provided substantial contribution to the discussion of content. All authors contributed to editorial changes in the manuscript. All authors read and approved the final manuscript. All authors have participated sufficiently in the work and agreed to be accountable for all aspects of the work.
Ethics Approval and Consent to Participate
Patients of the case report consented to anonymous publication of their age, sex, past and present medical history and coronary angiograms. Due to local regulations of the hospital, no ethical approval is required for this retrospective case report. The case report complies with the principles of the 1975 declaration of Helsinki.
Funding
This research received no external funding.
Conflict of Interest
The authors declare no conflict of interest.
References
- [1].Williams PD, Mamas MA, Morgan KP, El-Omar M, Clarke B, Bainbridge A, et al. Longitudinal stent deformation: a retrospective analysis of frequency and mechanisms. EuroIntervention: Journal of EuroPCR in Collaboration with the Working Group on Interventional Cardiology of the European Society of Cardiology . 2012;8:267–274. doi: 10.4244/EIJV8I2A41. [DOI] [PubMed] [Google Scholar]
- [2].Leibundgut G, Gick M, Toma A, Valina C, Löffelhardt N, Büttner HJ, et al. Longitudinal compression of the platinum-chromium everolimus-eluting stent during coronary implantation: predisposing mechanical properties, incidence, and predictors in a large patient cohort. Catheterization and Cardiovascular Interventions: Official Journal of the Society for Cardiac Angiography & Interventions . 2013;81:E206–E214. doi: 10.1002/ccd.24472. [DOI] [PubMed] [Google Scholar]
- [3].Arnous S, Shakhshir N, Wiper A, Ordoubadi FF, Williams P, Clarke B, et al. Incidence and mechanisms of longitudinal stent deformation associated with Biomatrix, Resolute, Element, and Xience stents: Angiographic and case-by-case review of 1,800 PCIs. Catheterization and Cardiovascular Interventions: Official Journal of the Society for Cardiac Angiography & Interventions . 2015;86:1002–1011. doi: 10.1002/ccd.25790. [DOI] [PubMed] [Google Scholar]
- [4].Rhee TM, Park KW, Lee JM, Lee MS, Jeon KH, Kang HJ, et al. Predictors and Long-Term Clinical Outcome of Longitudinal Stent Deformation: Insights From Pooled Analysis of Korean Multicenter Drug-Eluting Stent Cohort. Circulation. Cardiovascular Interventions . 2017;10:e005518. doi: 10.1161/CIRCINTERVENTIONS.117.005518. [DOI] [PubMed] [Google Scholar]
- [5].Mamas MA, Williams PD. Longitudinal stent deformation: insights on mechanisms, treatments and outcomes from the Food and Drug Administration Manufacturer and User Facility Device Experience database. EuroIntervention: Journal of EuroPCR in Collaboration with the Working Group on Interventional Cardiology of the European Society of Cardiology . 2012;8:196–204. doi: 10.4244/EIJV8I2A33. [DOI] [PubMed] [Google Scholar]
- [6].Si D, Yang H, Liu G, Tong Y, He Y. Treatment of longitudinal stent compression under intravenous ultrasound guidance: A case report. Medicine . 2017;96:e9405. doi: 10.1097/MD.0000000000009405. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [7].Cuenza LR, Yeo KK. Stent accordion phenomenon visualized with optical coherence tomography. Cardiovascular Diagnosis and Therapy . 2017;7:640–642. doi: 10.21037/cdt.2017.09.04. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [8].Yıldırım O, Gul M, Inci S, Aktaş H, Dogan P, Bozkır A, et al. A catastrophic nightmare of the interventional cardiologist: Iatrogenic left main artery dissection and longitudinal stent deformation. Intractable & Rare Diseases Research . 2018;7:209–212. doi: 10.5582/irdr.2018.01081. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [9].Kobayashi N, Hata N, Okazaki H, Shimizu W. Longitudinal stent deformation as a cause of very late stent thrombosis: Optical coherence tomography images. International Journal of Cardiology . 2016;202:601–603. doi: 10.1016/j.ijcard.2015.08.126. [DOI] [PubMed] [Google Scholar]
- [10].Panoulas VF, Demir OM, Ruparelia N, Malik I. Longitudinal deformation of a third generation zotarolimus eluting stent: “The concertina returns!”. World Journal of Cardiology . 2017;9:60–64. doi: 10.4330/wjc.v9.i1.60. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [11].Leong AM, Ong PJL, Ho HH, Watson T. Distal longitudinal deformation of a Synergy stent by jailed Rotawire guidewire. Herz . 2017;42:209–210. doi: 10.1007/s00059-016-4455-z. [DOI] [PubMed] [Google Scholar]
- [12].Yew KL, Cham YL. Longitudinal stent compression: a not so new percutaneous coronary angioplasty (PCI) complication. The Medical Journal of Malaysia . 2012;67:123–124. [PubMed] [Google Scholar]
- [13].Foerst J, Foin N, Hettleman B. Longitudinal stent compression demonstrated by angiographic “wedding band” and 3-dimensional optical coherence tomography. JACC. Cardiovascular Interventions . 2012;5:e39–e40. doi: 10.1016/j.jcin.2012.08.015. [DOI] [PubMed] [Google Scholar]
- [14].Aung SM, Demir AS, Boztosun B. Accordion effect in a thin-strut coronary stent. Perfusion . 2012;27:435–437. doi: 10.1177/0267659112448110. [DOI] [PubMed] [Google Scholar]
- [15].Buysschaert I, Ughi GJ, Adriaenssens T, Vermeersch P. Very late stent thrombosis and longitudinal stent deformation. Acta Cardiologica . 2017;72:216–217. doi: 10.1080/00015385.2017.1291203. [DOI] [PubMed] [Google Scholar]
- [16].Hernández-Matamoros H, Gonzálvez-Garcia A, Moreno-Gómez R, Jiménez-Valero S. Longitudinal Stent Deformation: Precise Diagnosis With Optical Coherence Tomography. The Journal of Invasive Cardiology . 2019;31:E395. [PubMed] [Google Scholar]
- [17].Vijayvergiya R, Sharma P, Gupta A, Goyal P, Panda P. Longitudinal stent deformation during coronary bifurcation stenting. Cardiovascular Revascularization Medicine: Including Molecular Interventions . 2016;17:143–145. doi: 10.1016/j.carrev.2015.12.004. [DOI] [PubMed] [Google Scholar]
- [18].Demir OM, Ruparelia N, Khan M, Baker C. Coronary Intervention Complicated by Pressure Wires Caught Within Stent Struts. The Journal of Invasive Cardiology . 2017;29:E102–E103. [PubMed] [Google Scholar]
- [19].Hanratty CG, Walsh SJ. Longitudinal compression: a “new” complication with modern coronary stent platforms–time to think beyond deliverability. EuroIntervention: Journal of EuroPCR in Collaboration with the Working Group on Interventional Cardiology of the European Society of Cardiology . 2011;7:872–877. doi: 10.4244/EIJV7I7A135. [DOI] [PubMed] [Google Scholar]
- [20].Shannon J, Latib A, Takagi K, Chieffo A, Figini F, Sacco F, et al. “Procedural trauma risks longitudinal shortening of the Promus Element™ stent platform”. Catheterization and Cardiovascular Interventions: Official Journal of the Society for Cardiac Angiography & Interventions . 2013;81:810–817. doi: 10.1002/ccd.24600. [DOI] [PubMed] [Google Scholar]
- [21].Cosgrove C, Marciniak M, Spratt JC. Reverse longitudinal stent deformation during percutaneous intervention via the retrograde approach to a chronic total occlusion of the right coronary artery: a case report. European Heart Journal. Case Reports . 2021;5:ytaa571. doi: 10.1093/ehjcr/ytaa571. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [22].Chung MS, Yang DH, Kim YH, Roh JH, Song J, Kang JW, et al. Stent fracture and longitudinal compression detected on coronary CT angiography in the first- and new-generation drug-eluting stents. The International Journal of Cardiovascular Imaging . 2016;32:637–646. doi: 10.1007/s10554-015-0798-4. [DOI] [PubMed] [Google Scholar]
- [23].Prabhu S, Schikorr T, Mahmoud T, Jacobs J, Potgieter A, Simonton C. Engineering assessment of the longitudinal compression behaviour of contemporary coronary stents. EuroIntervention: Journal of EuroPCR in Collaboration with the Working Group on Interventional Cardiology of the European Society of Cardiology . 2012;8:275–281. doi: 10.4244/EIJV8I2A42. [DOI] [PubMed] [Google Scholar]
- [24].Ormiston JA, Webber B, Webster MWI. Stent longitudinal integrity bench insights into a clinical problem. JACC. Cardiovascular Interventions . 2011;4:1310–1317. doi: 10.1016/j.jcin.2011.11.002. [DOI] [PubMed] [Google Scholar]
- [25].Shibutani H, Akita Y, Oishi Y, Sueyoshi H, Mukai Y, Yutaka K, et al. The potential hazard of a non-slip element balloon causing distal longitudinal stent deformation: the first clinical experience and in vitro assessment. Cardiology Journal . 2019;26:645–652. doi: 10.5603/CJ.a2018.0065. [DOI] [PMC free article] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Data Availability Statement
Data and materials are available on request.