Evaluation of an in-use chest CT protocol in lung cancer screening - A single institutional study

Salma Naimi 1, Mercy Afadzi Tetteh 2, Haseem Ashraf 2,3, Safora Johansen 1,4,5,
PMCID: PMC11265249  PMID: 39044837

Abstract

Background

Lung cancer is the most common cause of cancer-related death worldwide and therefore there has been a growing demand for low-dose computed tomography (LDCT) protocols.

Purpose

To investigate and evaluate the dose and image quality of patients undergoing lung cancer screening (LCS) using LDCT in Norway.

Materials and Methods

Retrospective dosimetry data, volumetric CT dose index (CTDIvol) and dose-length product (DLP), from 70 average-size and 70 large-size patients who underwent LDCT scan for LCS were included in the survey. Effective dose and size-specific dose were calculated for each examination and were compared with the American Association of Physicists in Medicine (AAPM) requirement. For a quantitative image quality analysis, noise, signal-to-noise ratio (SNR), and contrast-to-noise ratio (CNR) were determined for different regions in the chest with two iterative reconstruction techniques, iDose and Iterative Model Reconstruction. Differences in dose and image quality between average-size and large-size patients were evaluated by Independent sample t test, and Wilcoxon signed rank test within the same patient group.

Results

The independent sample t test revealed significant differences (p < .05) in dose values between average-size and large-size patients. Mean CTDIvol and DLP for average-size patients were 2.8 mGy and 115 mGy.cm, respectively, with appropriate increment for the large-size patients. Image quality (image noise, SNR, and CNR) did not significantly differ between patient groups when images were reconstructed with a model based iterative reconstruction algorithm.

Conclusion

The screening protocol assessed in this study resulted in CTDIvol values that were compliant with AAPM recommendation. No significant differences in objective image quality were found between patient groups.

Keywords: Low-dose CT, lung cancer screening, CT dose index, American Association of Physicists in Medicine

Introduction

Lung cancer is one of the leading causes of cancer in both men and women worldwide,1,2 and while not always the most diagnosed, lung cancer has the highest mortality rate in the world.3,4 In 2018, 2100 people died of lung cancer in Norway. 5 Early detection of lung cancer is currently the most effective way to reduce the total mortality of lung cancer. Patients with lung cancer diagnosed at stage I have a 5-year survival of more than 80% while the total overall survival is 15%–20%. 6 It is reported that lung cancer screening using low-dose computed tomography (LDCT) results in increased detection and decreased mortality. 7 In large prospective studies as the National Lung Cancer Screening Trial (NLST) in 2011 and the Dutch-Belgian in a lung cancer screening trial in 2020 confirmed that computed tomography (CT) screening can reduce lung cancer mortality by 20%.8,9 A growing number of countries are implementing lung cancer screening programs; however, there is a concern that exposure to the ionizing radiation of LDCT in lung cancer screening might increase the risk of developing solid cancers and leukemia. 1 Current lung cancer screening (LCS) protocols recommend repeated screening at either annual or biannual frequency for eligible individuals, which can result in over 25 scans over a lifetime.1012 Literature has shown that patient dose varies across regions and countries due to several factors, including CT scanner design, diagnostic protocols, and local-based choices of technical parameters.1214 This will also affect if the image quality is sufficient for detection and measurement for small pulmonary nodules in LCS programs.

There exists a need to implement screening programs for lung cancer around the world.15,16 The American College of Radiology (ACR), Society of Thoracic Radiology, and the European Society of Thoracic Imaging have published practice guidelines and technical standards to assist radiologists and medical physicists in developing local CT lung cancer screening protocols.1719 In addition, the American Association of Physicists in Medicine (AAPM) has published a recommended set of lung cancer screening protocols for a range of scanners. 20 These protocols are based on the experience gained from the NLST study and other screening studies by the working group. These protocols and the ACR guidelines result in a radiation dose (CTDIvol) ≤ 3 mGy (≤1.0 mSv) for a standardized patient of 70 kg and a height of 170 cm. However, radiation dose may vary from 0.25 to 5.6 mGy for patient of 50 to 120 kg. 20 The aim of this study is to investigate and evaluate the dose and image quality of patients undergoing LCS using LDCT at a local hospital in Norway. This will provide valuable information for protocol optimization to further reduce patient radiation dose in the implementation of a future LCS.

Method and materials

Patient data

The first lung cancer screening study in Norway started in August 2022. A total of 125,000 Norwegian inhabitants in the age group of 60–79 years were invited to participate in this study using LDCT protocol. Fourteen thousands out of 125,000 agreed to participate and 1,000 of these were selected as participants for the screening study over a 3-year period from 2022 until 2025. Data were collected from the Picture Archiving and Communication Systems, Carestream Vue PACS version 12.2.20105, Philips Healthcare, Best, Netherlands, between February and March 2023.

Patient selection

Patients between 60 and 79 years, with more than 35 pack years or a PLCOm2012 risk above 2.6%, were eligible for the study. One pack year is one pack of cigarettes (20 cigarettes) a day for one year. PLCOm2012 is a validated risk calculator used for inclusion in lung cancer screening programs. It provides a 6-year risk of developing lung cancer using different which apart from smoking exposure also take into account other factors like family history, BMI, and education level. To ensure a high enough risk group, a threshold of >2.6% was considered. 21 Cases with severe respiratory motion or prosthetic artifacts were excluded from this study. We retrospectively recorded data from 140 patients who had under gone screening between August and September 2022. Patient weight and height were not available; hence, patients were grouped based on AP chest thickness measured in axial slices at the carina. This was determined by measuring the AP thickness of 42 patients who had under gone contrast enhanced chest pulmonary angiogram examination in December 2022. Body mass index (BMI) was calculated and compared with the thorax AP thickness for each of the 42 patients as shown in Table 1. BMI for average-size and large-size was defined based on the BMI classification described in the AAPM protocols. 20 Average-size patients were defined with a body mass index ≥18.5 to 24.9 kg/m2 and large-size ≥25 kg/m2. 22 The exclusion criteria were patients with a body mass index less than 18.5 kg/m2. AP thickness from 20 to 24.9 cm and 25 to 30 cm was therefore defined as average-size and large-size, respectively.

Table 1.

AP thickness of average-size and large-size patients scanned with a contrast enhanced chest pulmonary angiogram protocol used for inclusion range of AP thickness used in this study.

Number of patient Average-size body mass index ≥18.5–24.9 kg/m2 Large-size body mass index ≥25 kg/m2
Weight (kg) Height (cm) Age Body mass index kg/m2 AP (cm) Weight (kg) Height (cm) Age Body mass index kg/m2 AP (cm)
1 69 179 73 21.5 22.6 100 178 75 31.6 26.4
2 64 163 69 24.1 21.4 95 178 70 30 27.6
3 70 174 72 23.1 20 119 184 59 35.1 28
4 70 170 56 24.2 23.5 95 172 85 32.1 26.8
5 64 170 73 22.1 21.1 80 166 70 29 25
6 78 177 63 24.9 22.9 105 167 33 37.6 26.7
7 65 168 78 23 20.4 102 170 68 35.3 26.6
8 65 166 67 23.6 21.6 125 169 32 43.8 29.5
9 78 178 88 24.6 22.6 117 164 48 43.5 25.7
10 77 165 58 21.3 20.4 90 168 65 31.9 25.8
11 76 186 79 22 21 88 180 80 27.2 25.4
12 54 154 89 22.8 20.9 52 168 98 34.7 25.2
13 64 175 77 20.9 22.4 114 172 39 38.5 25.9
14 60 160 57 23.4 20.5 104 170 43 36 25.1
15 60 160 84 23.4 23.5 94 186 60 27.2 25.3
16 70 176 64 22.6 22.9 97 187 68 27.7 25.3
17 69 170 80 23.9 23.8 100 187 78 28.6 25.9
18 56 161 75 21.6 22.7 156 195 59 41 29.2
19 78 183 59 23.3 23.9 130 186 48 37.6 29.6
20 80 181 78 24.4 24.5 92 177 61 29.4 26.1
21 73 175 74 23.8 22.3 82 161 83 31.6 25.2

CT acquisition parameters

CT examinations of the thorax were performed using a Brilliance iCT 256 scanner (Royal Philips, 2012, Amsterdam, The Netherlands) in supine position with arms raised above the shoulders to prevent artifacts. The patients were provided instructions to prevent any voluntary motion and to cautiously follow the breath-hold instructions. The scanner characteristics and protocol parameters are shown in Table 2. Raw data was reconstructed with two iterative reconstruction techniques iDose and Iterative Model Reconstruction (IMR). IDose is a hybrid or statistical iterative reconstruction algorithm whereas IMR is a model based iterative reconstruction algorithm. 23

Table 2.

Characteristics of the CT scanners and acquisition parameters included in the survey.

Manufactures Philips
Model Brilliance iCT 256
Install year 2012
Detector configuration (mm) a 128 × 0.625
Iterative reconstruction (level) iDose (3) and IMR (1)
Tube voltage 120
AEC settings Min/max mA25/200, DRI b 9
Acquired/recon slice (mm) 0.9/0.45
Pitch 1.171
Rot time (s) 0.33
Kernel iDose: Y-sharp / IMR: Soft tissue
Window width/window level (HU) iDose: 1600/-600 / IMR: 400/40
Software version 4.1.7

aNotation of the detector configuration: the number of active detector rows x the detector element thickness.

bDose right index.

Radiation dose assessment

Volume computed tomography dose index (CTDIvol) and dose-length product (DLP) values were collected for each examination retrospectively. The estimated effective dose was calculated as DLP multiplied by a k-factor of 0.014 mSv*mGy−1 cm−1 for the chest. 24 Size-specific dose estimate (SSDE) was calculated as described in the AAPM Report 204. 25 Thus, SSDE was calculated for each examination based on the recorded CTDIvol values and the corresponding AP chest thickness conversion coefficient. 24

Objective image analysis

Standardized 30-mm-diameter circular regions of interest (ROI) were used to record signal and noise, which represented mean attenuation value and standard deviation (SD) in Hounsfield units (HU). The ROIs were placed in the axial slices at the level of tracheal bifurcation in the subcutaneous fat, peripheral lung parenchyma, air within the trachea, truncus pulmonalis, aorta descendent, and paravertebral muscle, 26 as shown in Figure 1. Moreover, the signal-to-noise ratio (SNR) was calculated in all ROIs as follows 27 :

SNR=HU1SD1

Figure 1.

Figure 1.

Placement of the regions of interest for an objective image quality assessment: the subcutaneous fat, peripheral lung parenchyma, air within the trachea, truncus pulmonalis, aorta descendent, and paravertebral muscle.

The paravertebral muscle measurements were used as the background attenuation to calculate contrast-to-noise ratio (CNR) as follows 27 :

CNR=(HU1HU2)SD2

where HU1 is the attenuation value measured in peripheral lung parenchyma or aorta descendent or truncus pulmonalis and HU2 and SD2 are the attenuation value and noise measured in the paravertebral muscle.

Statistical analysis

Data were recorded in Excel (Microsoft Office 2010) and were analyzed with the Statistical Package for the Social Sciences software version 18.0 (IBM Corp.). The Kolmogorov–Smirnov test was performed to determine whether data were normally distributed. Descriptive statistical analysis of the collected dosimetry parameters, CTDIvol, DLP, SSDE, image noise, SNR, and CNR values was performed for each patient to determine the mean and standard deviation. Independent sample t test was used to compare the differences between average-size and large-size patients. Two-way ANOVA was used to investigate gender differences between the same size groups. To compare the difference in image quality within the same patient group, the Wilcoxon signed ranks test was used. The p value <.05 was considered as statistically significant.

Ethical considerations

This study was approved by Data Protection Office/institutional review board at the involved hospital defined as a quality assurance study. The need for informed consent for the use of existing CT scan images, including raw data, was waived. No sensitive patient data was recorded in the study.

Result

Patient selection

Of 140 patients, 78 (56%) were men and 62 (44%) were women, and the mean age of the participants was found to be 70 years (age range, 62–81) years (Table 3).

Table 3.

Basic characteristics of the participants.

Characteristics Average-size n (%) Large-size n (%)
Number of study participants 70 (50) 70 (50)
Gender
 Male 22 (30) 56 (80)
 Female 48 (70) 14 (20)
Age (years), median ± SD 70 ± 5 69 ± 5
 Range (years) 62 - 81 62 - 81

Figure 2 shows box plots of CTDIvol, DLP, effective dose, and SSDE categorized according to patient’s size. The mean CTDIvol of average-size patient was found to be 2.8 ± 0.48 mGy (range, 1.9–3.9) which is significantly lower (p < .001) than that of the large-size patients (4.5 ± 0.96 mGy; range, 2.3–6.6 mGy). Nonetheless, mean CTDIvol for the entire patient group was found to be 3.62 mGy. No significant difference was observed with average scan length of the two patient groups (p = .204). Consequently, mean DLP of the large-size group (193.6 ± 42.37 mGy∙cm; range, 94.2–293.5) was significantly (p < .001) higher than that of the average-size patients (115.22 ± 21.45 mGy∙cm, range, 79–160.2). The same trend was seen for the effective dose and SSDE data, where effective dose and SSDE of the large-size patients were 3.3 ± 0.72 mSv (range, 1.6–5) and 5.3 ± 1.03 mGy (range, 2.9–7.4) and that of average-size patients were 2 ± 0.4 mSv (range, 1.3–2.7) and 3.8 ± 0.67 mGy (range, 2.7–5.2), respectively (Figure 2). No correlation between gender and dose was observed in this study. No statistically significant difference was found between the CTDIvol (p = .28), DLP (p = .32), SSDE (p = .31), and effective dose (p = .3) when comparing the dose in men versus women within the same size group.

Figure 2.

Figure 2.

Box plot showing (a) CTDIvol = volume CT dose index, (b) DLP = dose-length product, (c) effective dose, and (d) SSDE = size-specific dose estimate of lung cancer screening using low-dose computed tomography. Box plot midlines indicate medians, outer lines indicate 25th and 75th percentiles, respectively, and whiskers indicate the range of data points excluding outliers represented by blank circle. p < .05 indicates statistical significance between average-size and large-size.

The objective image analysis results are presented in Table 4. The image noise in the air within the trachea and peripheral lung parenchyma did not significantly differ between average-size and large-size for iDose, except for subcutaneous fat (p = .009), truncus pulmonalis (p = .044), aorta descendent (p = .015), and paravertebral muscle (p = .003) that had a significant difference. Image noise, SNR, and CNR of images reconstructed with IMR did not significantly differ between average-size and large-size for all tissue types (Figure 3). The same trend was seen for images reconstructed with iDose, except CNR of the peripheral lung parenchyma (p = .001). As expected, significant difference was found for image noise, SNR, and CNR of iDose and IMR images within the same patient group for all tissue types, where IMR improves the image quality of LDCT more than hybrid algorithm iDose (p < .001).

Table 4.

Results of the objective image quality assessment: image noise, SNR, and CNR. Results are presented as median ± SD.

Parameter iDose IMR iDose IMR
Average-size Large-size Average-size Large-size Difference between average-size and large-size p-value Difference between average-size and large-size p-value
Image noise (HU)
 Air within the trachea 27.21 ± 3.81 27.49 ± 4.48 13.16 ± 2.18 13.09 ± 2.69 0.700 0.861
 Peripheral lung parenchyma 24.93 ± 4.50 25.46 ± 3.24 12.33 ± 12.07 13.11 ± 2.72 0.441 0.059
 Subcutaneous fat 30.53 ± 5.34 33.04 ± 5.89 11.38 ± 1.44 11.30 ± 1.62 0.009 0.769
 Truncus pulmonalis 34.25 ± 5.60 36.31 ± 5.89 12.60 ±1.85 12.82 ± 1.15 0.044 0.395
 Aorta descendent 35.76 ± 6.21 38.42 ± 6.53 12.43 ± 1.31 12.64 ± 1.67 0.015 0.412
 Paravertebral muscle 36.83 ± 5.90 39.59 ± 5.05 14.26 ± 4.05 13.35 ±2.34 0.003 0.106
SNR
 Peripheral lung parenchyma 37.42 ± 7.19 36.06 ± 5.46 75.74 ± 14.32 71.89 ± 15.13 0.212 0.124
 Truncus pulmonalis 1.40 ± 0.36 1.33 ± 0.30 3.61 ± 0.72 3.70 ± 0.69 0.667 0.422
 Aorta descendent 1.31 ± 0.25 1.29 ± 0.30 3.69 ± 0.69 3.63 ± 0.53 0.243 0.601
CNR
 Peripheral lung parenchyma 26.23 ± 4.54 24.09 ± 3.63 71.19 ± 18.96 72.81 ± 13.41 0.001 0.561
 Truncus pulmonalis 0.42 ± 0.31 0.37 ± 0.32 1.05 ± 0.82 1.17 ±0.96 0.199 0.455
 Aorta descendent 0.41 ± 0.30 0.37 ± 0.27 1.09 ± 0.81 1.02 ± 0.75 0.206 0.587

p < .05. HU = Hounsfield unit, SD = standard deviation, CNR = contrast-to-noise ratio, SNR = signal-to-noise ratio.

Figure 3.

Figure 3.

Bar chart showing (a) SNR = signal-to-noise ratio and (b) CNR = contrast-to-noise ratio in mean for average-size and overweight. HU = Hounsfield units.

Discussion

In this study, the impact of a LDCT protocol on dose and image quality in both average-size and large-size patients was investigated. The results of this study showed a significantly higher radiation dose for large-size patient group compared to average-size. However, no significant difference in objective image quality was observed between the two patient groups.

Even though the risk of radiation-induced cancer is low compared to the benefits of LCS, it is important to keep all doses as low as possible, especially for patients undergoing repetitive examinations. 28 According to the ACR and AAPM recommendation, LDCT must be performed with a CTDIvol of ≤3.0 mGy for a standard-sized patient.18,20 In the present study, the mean CTDIvol for the average-sized patient was found to be 2.8 mGy, which is well within the recommendation. For protocols that use AEC methods, the tube output will be dependent on the patient’s anatomy. Our results also reveal that the Philips AEC (DoseRight Z + 3D) used in this study allowed the tube output to be adjusted according to patient size. Thus, mean CTDIvol used for the large-size patients was significantly higher than that of the average-sized patients. Other researchers have also reported CTDIvol < 3 mGy for average-size patients with appropriate reduction and increment for patients with lower and higher BMIs.29,30 No statistically significant difference was found between the radiation dose (CTDIvol, DLP, SSDE, and effective dose) related to gender within the same size group. The results are in agreement with doses reported by Chu et al 31 confirming that dose variation is dependent on patient size and not gender.

Mean CTDIvol observed in this study for average-sized and large-size patients (2.8 and 4.5 mGy, respectively) was higher than what AAPM proposes for a Philips Brilliance iCT 256 scanner (2.7 and 3.6 mGy, respectively). Additionally, mean DLP and ED observed for average-size patients in this study were 115 mGycm and 2 mSv, respectively, which are higher than what AAPM proposes for a standard-sized patient (≤75 mGycm and ≤1.0 mSv, respectively). 20 These discrepancies might be due to difference in scan parameters and scanning techniques. Also, difference in software version can affect the radiation dose according to the vendor. Besides, research has shown that small differences in scan techniques even with the use of AEC can lead to variation in radiation doses for the same patients. 32 Changes in the table height, scan length, and small differences in the patient’s positioning may also affect the patients’ radiation dose, according to research.32,33 For low-dose CT examination, the desired coverage is from the apex to the lung bases. However, technologist will often add a “safe margin” to the scan length to assure complete coverage, which may increase the DLP. 33 Hence, scan length should be adapted strictly to the clinical indication for each individual patient and must be limited to the area of interest in order to avoid unnecessary radiation dose to the patient.34,35 This suggest that there is potential for dose reduction which implies the protocol will benefit from optimization.

In lung screening, it is critically important to ensure high confidence in the detection and volumetric measurements of small lung nodules. 36 The assessment of small lung nodules does not require high CNR but good spatial resolution because of the inherent high contrast present in the chest. 37 Thus, the evaluation of the lung parenchyma can tolerate high noise. Therefore, it is important that careful attention be paid to the selection of exposure parameters in order to achieve the desired level of image quality. 36 Our results revealed no significant differences in image noise between patient sizes when images where reconstructed with IMR and 0.9 mm slice thickness. The reconstructed slice thickness is a critical determinant for the detection and volumetric measurement accuracy of small nodules as well as the ability to reconstruct images in multiple planes. Consequently, thinner slice thickness increases spatial resolution and decreases the partial volume effect even though it increases image noise.38,39 In the present study, slice thickness of 0.9 mm was used for both lung and soft tissue reconstructions. Study findings have shown greater variability and inaccuracy in volume determination of small nodules compared with large nodules with different slice thicknesses.3840 Winer-Muram and colleagues 41 found that an increase in slice thickness leads to an increase in overestimation of nodule volume. Our data indicates that MBIR (IMR) improves the image quality of LDCT more than hybrid algorithm (iDose) which is in accordance with previous studies. 42

The present study has some limitations that need to be addressed. Thus, patient’s weight and height were not available for this study, making it impossible to calculate BMI of the patients. Hence, patient grouping was done based on AP diameter measured in axial slices at the carina and compared with that of previous studies where weight and height were recorded. The study would have benefited from a qualitative image quality assessment; however, this is a retrospective study so all images had been evaluated by one or two radiologist and image quality was considered clinically adequate. Finally, the study was carried out at a single institution. However, our findings are generally in line with previous studies and therefore may well be valid to other centers. In conclusion, this study showed that the CTDIvol values from a lung cancer screening program were within the AAPM recommended dose for average-size patients. It also showed the use of AEC adjusted the dose according to patient size without affecting image quality. However, the protocol might benefit from optimization of scan parameters in order to reduce the radiation dose further.

Footnotes

The author(s) declared no potential conflicts of interest with respect to the research, authorship, and/or publication of this article.

Funding: The author(s) received no financial support for the research, authorship, and/or publication of this article.

ORCID iDs

Salma Naimi https://orcid.org/0009-0003-7993-1624

Safora Johansen https://orcid.org/0000-0003-2307-4179

References

  • 1.Rampinelli C, De Marco P, Origgi D, et al. Exposure to low dose computed tomography for lung cancer screening and risk of cancer: secondary analysis of trial data and risk-benefit analysis. BMJ 2017; 356: j347. DOI: 10.1136/bmj.j347. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2.Bray F, Ferlay J, Soerjomataram I, et al. Global cancer statistics 2018: GLOBOCAN estimates of incidence and mortality worldwide for 36 cancers in 185 countries. CA A Cancer J Clin 2018; 68: 394–424. DOI: 10.3322/caac.21492. [DOI] [PubMed] [Google Scholar]
  • 3.Murugan VA, Kalra MK, Rehani M, et al. Lung cancer screening: computed tomography radiation and protocols. J Thorac Imag 2015; 30: 283–289. DOI: 10.1097/RTI.0000000000000150. [DOI] [PubMed] [Google Scholar]
  • 4.Raji OY, Duffy SW, Agbaje OF, et al. Predictive accuracy of the Liverpool Lung Project risk model for stratifying patients for computed tomography screening for lung cancer: a case-control and cohort validation study. Ann Intern Med 2012; 157: 242–250. DOI: 10.7326/0003-4819-157-4-201208210-00004. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Kreftregisteret . Cancer in Norway 2019. Oslo: Kreftregisteret, 2022. https://www.kreftregisteret.no/globalassets/cancer-in-norway/2019/cin_report.pdf (accessed 17 mai 2023). [Google Scholar]
  • 6.Chansky K, Detterbeck FC, Nicholson AG, IASLC Staging and Prognostic Factors Committee, Advisory Boards, and Participating Institutions , et al. The IASLC lung cancer staging project: external validation of the revision of the TNM stage groupings in the eighth edition of the TNM classification of lung cancer. J Thorac Oncol 2017; 12: 1109–1121. DOI: 10.1016/j.jtho.2017.04.011. [DOI] [PubMed] [Google Scholar]
  • 7.Taekker M, Kristjansdottir B, Graumann O, et al. Diagnostic accuracy of low-dose and ultra-low-dose CT in detection of chest pathology: a systematic review. Clin Imag 2021; 74: 139–148. DOI: 10.1016/j.clinimag.2020.12.041. [DOI] [PubMed] [Google Scholar]
  • 8.National Lung Screening Trial Research Team. Aberle DR, Adams AM, Berg CD, et al. Reduced lung-cancer mortality with low-dose computed tomographic screening. N Engl J Med 2011; 365: 395–409. DOI: 10.1056/NEJMoa1102873. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9.de Koning HJ, van der Aalst CM, de Jong PA, et al. Reduced lung-cancer mortality with volume CT screening in a randomized trial. N Engl J Med 2020; 382: 503–513. DOI: 10.1056/NEJMoa1911793. [DOI] [PubMed] [Google Scholar]
  • 10.Heuvelmans MA, Oudkerk M. Appropriate screening intervals in low-dose CT lung cancer screening. Transl Lung Cancer Res 2018; 7: 281–287. DOI: 10.21037/tlcr.2018.05.08. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Wood DE, Kazerooni EA, Baum SL, et al. Lung cancer screening, version 3.2018, NCCN clinical practice guidelines in oncology. J Natl Compr Cancer Netw 2018; 16: 412–441. DOI: 10.6004/jnccn.2018.0020. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12.Perisinakis K, Seimenis I, Tzedakis A, et al. Radiation burden and associated cancer risk for a typical population to be screened for lung cancer with low-dose CT: a phantom study. Eur Radiol 2018; 28: 4370–4378. DOI: 10.1007/s00330-018-5373-7. [DOI] [PubMed] [Google Scholar]
  • 13.Smith-Bindman R, Wang Y, Chu P, et al. International variation in radiation dose for computed tomography examinations: prospective cohort study. BMJ 2019; 364: k4931. DOI: 10.1136/bmj.k4931. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14.Saltybaeva N, Martini K, Frauenfelder T, et al. Organ dose and attributable cancer risk in lung cancer screening with low-dose computed tomography. PLoS One 2016; 11: e0155722. DOI: 10.1371/journal.pone.0155722. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.Vonder M, Dorrius MD, Vliegenthart R. Latest CT technologies in lung cancer screening: protocols and radiation dose reduction. Transl Lung Cancer Res 2021; 10: 1154–1164. DOI: 10.21037/tlcr-20-808. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16.National Research Council (US) Board on Radiation Effects Research . Health risks from exposure to low levels of ionizing radiation: BEIR VII, phase I, letter Report (1998). Washington (DC): National Academies Press, 1998. [PubMed] [Google Scholar]
  • 17.Kazerooni EA, Austin JH, Black WC, American College of Radiology and Society of Thoracic Radiology , et al. ACR-STR practice parameter for the performance and reporting of lung cancer screening thoracic computed tomography (CT): 2014 (Resolution 4). J Thorac Imag 2014; 29: 310–316. DOI: 10.1097/RTI.0000000000000097. [DOI] [PubMed] [Google Scholar]
  • 18.American College of R . Adult lung cancer screening technical specifications. Atlanta, GA: American College of R. 2014, https://www.acraccreditation.org/-/media/ACRAccreditation/Documents/LCS/Lung-Cancer-Screening-Technical-Specifications.pdf?la=en</eref (accessed 17 mai 2023). [Google Scholar]
  • 19.European Society of Thoracic Imaging . Chest CT for lung cancer screening. Vienna: European society of Thoracic Imaging, 2020. https://www.myesti.org/content-esti/uploads/ESTI-LCS-technical-standards_2019-06-14.pdf (accessed 17 mai 2023). [Google Scholar]
  • 20.American Association for Physicists in Medicine (AAPM) . AAPM lung cancer screening CT protocols version 5.1. Lung Cabcer Screening CT Protocols. Alexandria, VA: AAPM, 2019. https://www.aapm.org/pubs/CTProtocols/documents/LungCancerScreeningCT.pdf (accessed 17 mi 2023). [Google Scholar]
  • 21.Tammemagi MC, Katki HA, Hocking WG, et al. Selection criteria for lung-cancer screening. N Engl J Med 2013; 368: 728–736. DOI: 10.1056/NEJMoa1211776. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22.Wei BaJ A. BMI classification percentile and cut off points. In: StatPearls. Treasure Island, FL: StatPearls Publishing, 2022. [PubMed] [Google Scholar]
  • 23.Afadzi M, Fossa K, Andersen HK, et al. Image quality measured from ultra-low dose chest computed tomography examination protocols using 6 different iterative reconstructions from 4 vendors, a phantom study. J Comput Assist Tomogr 2020; 44: 95–101. DOI: 10.1097/RCT.0000000000000947. [DOI] [PubMed] [Google Scholar]
  • 24.Trattner S, Halliburton S, Thompson CM, et al. Cardiac-specific conversion factors to estimate radiation effective dose from dose-length product in computed tomography. JACC Cardiovasc Imaging 2018; 11: 64–74. DOI: 10.1016/j.jcmg.2017.06.006. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 25.American Association of Physicists in Medicine . Size-specific dose estimates (SSDE) in pediatric and adult body CT examinations. Alexandria, VA: AAPM, 2018. https://www.aapm.org/pubs/reports/rpt_204.pdf (accessed 25 Mai 2023). [Google Scholar]
  • 26.Do TD, Rheinheimer S, Kauczor HU, et al. Image quality evaluation of dual-layer spectral CT in comparison to single-layer CT in a reduced-dose setting. Eur Radiol 2020; 30: 5709–5719. DOI: 10.1007/s00330-020-06894-7. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 27.Jensen K, Andersen HK, Smedby O, et al. Quantitative measurements versus receiver operating characteristics and visual grading regression in CT images reconstructed with iterative reconstruction: a phantom study. Acad Radiol 2018; 25: 509–518. DOI: 10.1016/j.acra.2017.10.020. [DOI] [PubMed] [Google Scholar]
  • 28.Demb J, Chu P, Yu S, et al. Analysis of computed tomography radiation doses used for lung cancer screening scans. JAMA Intern Med 2019; 179: 1650–1657. DOI: 10.1001/jamainternmed.2019.3893. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 29.Fujii K, McMillan K, Bostani M, et al. Patient size-specific analysis of dose indexes from CT lung cancer screening. AJR Am J Roentgenol 2017; 208: 144–149. DOI: 10.2214/AJR.16.16082. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 30.Kim EY, Kim TJ, Goo JM, et al. Size-specific dose estimation in the Korean lung cancer screening project: does a 32-cm diameter phantom represent a standard-sized patient in Korean population? Korean J Radiol 2018; 19: 1179–1186. DOI: 10.3348/kjr.2018.19.6.1179. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 31.Chu PW, Kofler C, Haas B, et al. Dose length product to effective dose coefficients in adults. Eur Radiol 2024; 34(4): 2416–2425. DOI: 10.1007/s00330-023-10262-6. [DOI] [PubMed] [Google Scholar]
  • 32.Suntharalingam S, Stecker FF, Guberina N, et al. How much is the dose varying between follow-up CT-examinations performed on the same scanner with the same imaging protocol? PLoS One 2016; 11: e0152961. DOI: 10.1371/journal.pone.0152961. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 33.Huo D, Kiehn M, Scherzinger A. Investigation of low-dose CT lung cancer screening scan “Over-Range” issue using machine learning methods. J Digit Imag 2019; 32: 931–938. DOI: 10.1007/s10278-019-00233-z. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 34.Yu L, Liu X, Leng S, et al. Radiation dose reduction in computed tomography: techniques and future perspective. Imag Med 2009; 1: 65–84. DOI: 10.2217/iim.09.5. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 35.Shahir K, Goodman LR, Lam CA, et al. Dose reduction of 69% for computed tomography pulmonary angiography: reduced z-axis computed tomography pulmonary angiography retains accuracy in those younger than 40 years. J Comput Assist Tomogr 2013; 37: 765–769. DOI: 10.1097/RCT.0b013e31829d198a. [DOI] [PubMed] [Google Scholar]
  • 36.Iball GR, Darby M, Gabe R, et al. Establishing scanning protocols for a CT lung cancer screening trial in the UK. Br J Radiol 2021; 94: 20201343. DOI: 10.1259/bjr.20201343. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 37.Fillon M, Si-Mohamed S, Coulon P, et al. Reduction of patient radiation dose with a new organ based dose modulation technique for thoraco-abdominopelvic computed tomography (CT) (Liver dose right index). Diagn Interv Imaging 2018; 99: 483–492. DOI: 10.1016/j.diii.2018.01.013. [DOI] [PubMed] [Google Scholar]
  • 38.Petrou M, Quint LE, Nan B, et al. Pulmonary nodule volumetric measurement variability as a function of CT slice thickness and nodule morphology. AJR Am J Roentgenol 2007; 188: 306–312. DOI: 10.2214/AJR.05.1063. [DOI] [PubMed] [Google Scholar]
  • 39.Zhao B, Schwartz LH, Moskowitz CS, et al. Pulmonary metastases: effect of CT section thickness on measurement--initial experience. Radiology 2005; 234: 934–939. DOI: 10.1148/radiol.2343040020. [DOI] [PubMed] [Google Scholar]
  • 40.Gomi S, Muramatsu Y, Tsukagoshi S, et al. Low-dose CT screening for lung cancer with automatic exposure control: phantom study. Radiol Phys Technol 2008; 1: 244–250. DOI: 10.1007/s12194-008-0036-z. [DOI] [PubMed] [Google Scholar]
  • 41.Winer-Muram HT, Jennings SG, Meyer CA, et al. Effect of varying CT section width on volumetric measurement of lung tumors and application of compensatory equations. Radiology 2003; 229: 184–194. DOI: 10.1148/radiol.2291020859. [DOI] [PubMed] [Google Scholar]
  • 42.Yuki H, Oda S, Utsunomiya D, et al. Clinical impact of model-based type iterative reconstruction with fast reconstruction time on image quality of low-dose screening chest CT. Acta Radiol 2016; 57(3): 295–302. DOI: 10.1177/0284185115575537. [DOI] [PubMed] [Google Scholar]

Articles from Acta Radiologica Open are provided here courtesy of SAGE Publications