Key Points
Question
Can self-fitting strategies—self-adjustment and in situ audiometry—provide similar outcomes for adults with self-perceived mild to moderate hearing difficulties using self-fitting over-the-counter (OTC-SF) hearing aids?
Findings
This crossover trial found that both strategies produced equivalent overall Abbreviated Profile of Hearing Aid Benefit (APHAB), overall International Outcome Inventory for Hearing Aids (IOI-HA) satisfaction, speech-in-noise performance and real-ear measurements. However, self-adjustment resulted in meaningfully higher satisfaction and daily use on 2 IOI-HA subscales, underscoring the benefits of user participation.
Meaning
Although OTC-SF hearing aids fitted through self-adjustment and in situ audiometry both yield similar overall APHAB and overall IOI-HA satisfaction, self-adjustment was found to offer additional user satisfaction and encourage more consistent use.
This pseudorandomized clinical trial examines Abbreviated Profile of Hearing Aid Benefit scores for self-adjustment vs in situ audiometric fitting strategies for over-the -counter hearing aids among adults with hearing loss.
Abstract
Importance
Fewer than 20% of US adults with hearing loss use hearing aids due to barriers like high cost. Over-the-counter (OTC) hearing aids offer a potential solution, incorporating self-fitting strategies via smartphone apps. Self-fitting strategies have been validated for Food and Drug Administration (FDA)–approved OTC hearing aids compared with prescription-based approaches. However, no direct comparative analysis exists between in situ audiometry and self-adjustment strategies using self-fitting OTC (OTC-SF) hearing aids.
Objective
To compare self-adjustment and in situ audiometry self-fitting strategies in OTC-SF hearing aids for adults with mild to moderate hearing difficulties.
Design Settings and Participants
A crossover, within-participant pseudorandomized clinical trial was conducted between July and November 2023. Twenty-eight participants were pseudo-randomly assigned to 1 of the 2 self-fitting strategies, and they experienced both interventions for 4 consecutive weeks.
Interventions
The self-adjustment group manually adjusted settings, including overall gain and spectral tilt, using Lexie B2 hearing aids, while the in situ audiometry group used Lexie B2 Plus hearing aids (Lexie Hearing by hearX Group), with an automated fitting based on in situ tests conducted through the app.
Main Outcomes and Measures
The primary outcome was Abbreviated Profile of Hearing Aid Benefit (APHAB). Secondary outcomes were International Outcome Inventory for Hearing Aids (IOI-HA), speech-in-noise tests (DIN and QuickSIN), and real-ear measurements (REMs). Measures were completed at baseline and after the 4-week field trial using each strategy.
Results
Twenty-eight participants (mean [SD] age, 60.2 [12.0] years) were included; 14 men and 14 women. Self-adjustment and in situ audiometry strategies produced no clinically meaningful differences across various outcome measures, including overall APHAB benefit (Cohen d = 0.2; 95% CI, −0.2 to 0.6) and overall IOI-HA satisfaction (Rosenthal r = 0.0; 95% CI, −0.3 to 0.2). Self-adjustment users reported higher satisfaction (Rosenthal r = −0.4; 95% CI, −0.6 to −0.1) and longer daily use (Rosenthal r = −0.3; 95% CI, –0.5 to 0.0) compared with those using in situ audiometry. No clinically meaningful differences were observed in speech-in-noise benefit and real-ear measurements.
Conclusion and Relevance
In this clinical trial of OTC-SF hearing aids, self-adjustment and in situ audiometry strategies resulted in similar outcomes. However, self-adjustment may produce higher satisfaction and longer daily use, highlighting the potential advantages of active user involvement in the fitting process. Further investigation is needed for long-term outcomes.
Trial registration
ClinicalTrials.gov Identifier: NCT05782153
Introduction
Hearing loss is a prevalent health condition affecting millions worldwide, often left untreated.1 Less than 20% of adults with hearing loss in the US use hearing aids due to barriers, such as the high cost.2,3 However, the introduction of over-the-counter (OTC) hearing aids has expanded service delivery models to improve the accessibility and affordability of hearing health care.4 OTC hearing aids enable individuals with self-perceived mild to moderate hearing loss (or hearing difficulties) to proactively manage their hearing health without needing medical examinations, prescriptions, or audiologist fittings, thus reducing some barriers to hearing aid uptake.4
The Food and Drug Administration (FDA) established a regulatory category for OTCs in 2022, including 2 subcategories: OTCs with preset programs (OTC-PS) and self-fitting OTC (OTC-SF) hearing aids.4 OTC-PS hearing aids offer standardized output profiles tailored to common hearing loss patterns. OTC-SF hearing aids are self-fit devices, adjustable via smartphone apps to customize parameters like frequency response.4 One of these strategies is self-adjustment, which allows the user to select signal processing parameters (eg, gain and compression) without obtaining an audiogram.5 Another strategy is in situ audiometry, which involves a hearing test administered through hearing aids with a smartphone app to better match the individual’s hearing profile.6
The concept of self-fitting hearing aids was introduced more than a decade ago,7 and several studies have investigated the effectiveness of self-fitting strategies, such as self-adjustment5,8,9 and in situ audiometry.6,10,11,12 Recent studies have evaluated OTC-SF hearing aids against audiologist-fit counterparts and found comparable outcomes. For instance, Sabin et al5 used a simple self-fitting strategy using a mobile app, where users adjusted signal processing parameters, resulting in improved self-reported sound quality comparable to the audiologist-fit group.5 An observational study demonstrated positive outcomes for OTC-SF compared with prescription hearing aids.13 Moreover, De Sousa et al14 used an in situ audiometry fitting strategy and found 6-week outcomes of OTC-SF hearing aids to be comparable to fitting by an audiologist fit.
Existing studies have validated OTC-SF hearing aids using different self-fitting strategies against prescription-based approaches. However, a direct comparative analysis between in situ audiometry and self-adjustment strategies used on OTC-SF hearing aids remains unexplored. Given the proliferation of available self-fitting OTCs, there is a recognized need for research comparing the efficacy of self-fitting strategies.15 Therefore, this study aimed to bridge this gap by directly comparing 2 self-fitting strategies in OTC-SF hearing aids, self-adjustment and in situ audiometry, among adults with self-perceived mild to moderate hearing difficulties.
Methods
Study Design
The trial protocol is available in Supplement 1, and the statistical analysis plan is in Supplement 2. A crossover, within-participant study using pseudorandomization methods was conducted between July and November 2023. The Consolidated Standards of Reporting Trials (CONSORT) extension to randomized crossover trials was used in reporting this trial (eTables 2 and 3 in Supplement 3). Institutional review board approval was obtained from the University of Pretoria Humanities Research Ethics Committee. The study flow diagram is available in eFigure 1 in Supplement 3. Twenty-eight participants were randomly assigned to 1 of the 2 self-fitting strategies (self-adjustment or in situ audiometry) using Lexie B2 and Lexie B2 Plus Powered by Bose hearing aids. Participants sequentially experienced each intervention for 4 weeks before transitioning to the other, ensuring a comprehensive comparative analysis for both self-fitting strategies.
Participants
Study participation was advertised on social media platforms, and interested individuals were asked to complete a Google Form to determine their eligibility. We enrolled adult participants (>18 years) who self-reported mild to moderate hearing difficulties, had no active ear pathologies, and owned smartphones with data capabilities in line with the use indications for OTC-SF hearing aids. On the Google Form, participants were asked to describe their hearing (without a hearing aid) by selecting one of the following options: (1) my hearing is good, (2) I have little trouble, (3) I have a lot of trouble, or (4) I cannot hear at all. Participants were also asked to rate their hearing difficulties as slight, mild, moderate, or severe. To be included in the study, participants needed to describe their hearing as having little trouble or a lot of trouble and rate their hearing difficulties as mild or moderate. In addition, participants were required to have a high level of English proficiency, determined through an online test,16 to accurately comprehend and follow instructions for self-fitting the hearing aids and complete the standardized outcome measures.
Initially, 103 individuals applied via the Google Form, and the first 30 who met our eligibility criteria were enrolled. This sample size was comparable to FDA-approved studies of similar devices.4 Participants were randomly assigned to 1 of the 2 self-fitting strategies (self-adjustment or in situ audiometry). Randomization was determined using an artificial intelligence generator (ChatGPT, version 3.5; Open AI) and was not concealed because the primary author (M.K.) enrolled participants and assigned interventions accordingly. Due to a technical issue in the manufacturer’s research app, the in situ audiometry self-fitting process defaulted to a preset level rather than individualized thresholds, necessitating the disqualification of the first 9 participants. We addressed potential bias by replacing them with our pool’s next 9 eligible candidates. This technical issue resulted in a departure from true randomization, leading to a pseudorandomized design. In pseudorandomization, participants are assigned to groups in a manner that aims to mimic randomness but can be influenced by systematic processes like technical challenges. The technical issue with the research app was addressed prior to subsequent data collection. The research app is not used commercially, minimizing the likelihood of such issues in actual self-fitting scenarios.
Throughout the trial, 2 participants withdrew. One participant withdrew due to financial difficulties hindering appointment attendance. Another, already a hearing aid user, withdrew due to difficulty wearing different aids for an extended period.
Data Collection
A qualified audiologist conducted baseline hearing assessments for all participants, which included otoscopic examination, tympanometry, pure-tone audiometry, and unaided speech-in-noise tests, such as Digits-In-Noise (DIN) and Quick Speech-in-Noise (QuickSIN). These tests were conducted in a soundproof booth. Participants also completed the Abbreviated Profile of Hearing Aid Benefit (APHAB) in unaided conditions.
Subsequently, individual Lexie profiles were established for each participant using their email address and a secure password. Participants were instructed to download the Lexie smartphone app. During the fitting appointment, they received login credentials to connect their hearing aids via the app. The B2 and B2 plus hearing aids are FDA-cleared, rechargeable, receiver in the canal OTC-SF hearing aids powered by Bose. The B2 hearing aid uses a self-adjustment self-fitting strategy (also referred to as a direct adjustment), whereas B2 Plus hearing aid uses an in situ audiometry self-fitting strategy. The participants were given the task of independently fitting their own hearing aids but were given the flexibility to involve a significant other or family member for support if necessary. In case they could not complete the process with the provided options, the researcher (M.K.) was available to offer help.
Participants were pseudorandomly assigned to a self-fitting strategy. Those assigned to the self-adjustment self-fit group received a pair of B2 hearing aids, which they paired with their smartphones via Bluetooth through the app. In this approach, participants were required to manually adjust the hearing aids’ settings using the app. The app interface provided a set of intuitive controls conceptualized as wheels, which allowed users to modify key acoustic parameters in increments (eFigure 2 in Supplement 3), including world volume, ie, overall gain (or amplification level) and spectral tilt (the balance of bass and treble frequencies), similar to Sabin et al.5 The self-adjustment strategy is fully user-driven, obliging the user to decide on the hearing aid settings, guided solely by personal preference and comfort, without any preset starting point.
Participants in the in situ audiometry self-fit group received a pair of the B2 Plus hearing aids, which they connected to their smartphones via the app. This method used an automated fitting protocol based on in situ hearing threshold assessments conducted through the hearing aids, which automatically recommended specific gain settings across different frequencies. This strategy is designed to provide a tailored fit based on the individual’s unique hearing profile. However, after applying the recommended settings based on the in situ audiometry, participants could still adjust the hearing aids’ overall gain (or amplification level) and spectral tilt (the balance of bass and treble frequencies) in the app if they wanted to (eFigure 3 in Supplement 3).
Once the participants were satisfied with their settings, the audiologist performed real-ear measurements (REMs) and aided speech-in-noise testing (ie, DIN and QuickSIN). Participants were subsequently encouraged to wear the hearing aids daily for 28 days, ensuring they did so for as much of the day as was comfortable and practical. After the 28-day period, participants returned for follow-up assessments, including the APHAB (with hearing aids), International Outcomes Inventory for Hearing Aids (IOI-HA), REMs, and aided speech-in-noise tests. Subsequently, each participant received the alternative device and fitting strategy. After the second 28-day period, the same outcome measures were re-evaluated. The primary outcome measure was the APHAB. The benefit was calculated by subtracting the aided score from the unaided score, where a higher score indicates a greater benefit. Secondary measures included the IOI-HA, speech-in-noise tests, and REMs.
Data Analysis
Data were analyzed using SPSS Statistics software (version 29.0; IBM). We used paired sample t tests for normally distributed continuous variables and Wilcoxon signed-rank tests for non–normally distributed or ordinal variables to assess differences between self-fitting strategies within participants. We identified significant differences with clinical relevance by examining effect size and 95% CIs. Cohen d was used for t test findings with interpretation categorized as small (d ≤0.2), small to medium (d, 0.2 to <0.5), medium (d = 0.5), medium to large (d, >0.5 to <0.8), and large (d ≥0.8).17 For the Wilcoxon signed rank test, Rosenthal r was calculated as:
![]() |
and its interpretation included small (r ≤0.1), small to medium (r, >0.1 to <0.30), medium (r = 0.3), medium to large (r, >0.3 to <0.5), and large (r ≥0.5).17 The effect sizes would be considered clinically meaningful if they were medium or larger.
Results
Baseline Assessments
The study included 28 participants (Table 1) with equal sex distribution and a mean (SD) age of 60.2 (12.0) years. Overall, 14 participants (50%) self-reported having a little hearing trouble, while the other half (n = 14; 50%) reported having a lot of trouble. Seven participants (25%) reported mild hearing difficulties, while 21 (75%) reported moderate hearing difficulties. The mean (SD) 4-frequency pure-tone average (PTA) was 32.4 (14.9) dB and 36.8 (16.3) dB for the left and right ears, respectively (eFigure 4 in Supplement 3). Participants had a mean (SD) unaided DIN score of −8.2 (1.8) and a QuickSIN score of 7.0 (4.8).
Table 1. Participant Characteristics (N = 28).
| Characteristic | No. (%) |
|---|---|
| Sex | |
| Female | 14 (50) |
| Male | 14 (50) |
| Age, mean (SD), y | 60.2 (12) |
| Self-perceived hearing difficulty | |
| I have a little trouble | 14 (50) |
| I have a lot of trouble | 14 (50) |
| Self-perceived degree of hearing loss | |
| Mild | 7 (25) |
| Moderate | 21 (75) |
| PTA for 0.5, 1, 2, 4 kHz, mean (SD) | |
| Left | 32.4 (14.9) |
| Right | 36.8 (16.3) |
| Unaided DIN score | −8.2 (1.8) |
| Unaided QuickSIN | 7.0 (4.8) |
Abbreviations: PTA, pure-tone average; DIN, digits-in-noise; QuickSIN, quick speech-in-noise.
Abbreviated Profile of Hearing Aid Benefit
No clinically meaningful differences were found between self-adjustment and in situ self-fitting strategies within participants for any APHAB subscales (ease of communication [EC]: Rosenthal r = −0.1; 95% CI, −0.4 to 0.1) (background noise [BN]: Cohen d = −0.1; 95% CI, −0.5 to 0.2) (reverberation [RV]: Cohen d = −0.3; 95% CI, −0.6 to 0.1) (aversion [AV]: Rosenthal r = −0.2; 95% CI, −0.5 to 0.0) or global scores (Cohen d = −0.2, 95% CI, −0.6 to 0.2) after the 4-week field trial. Overall APHAB benefit from baseline did not differ meaningfully between the 2 self-fitting methods (EC: Rosenthal r = −0.1; 95% CI, −0.4 to 0.1) (BN: Cohen d = 0.1; 95% CI, −0.2 to 0.5) (RV: Cohen d = 0.3; 95% CI, −0.1 to 0.6) (AV: Rosenthal r = −0.2; 95% CI, −0.5 to 0.0) (Global: Cohen d = 0.2; 95% CI, −0.2 to 0.6) (Table 2 and Figure 1).
Table 2. APHAB Scores for the Unaided Baseline, After 4-Week Trial and Benefit for the Self-Adjustment and In Situ Audiometry Self-fitting Strategies (N = 28).
| APHAB subscale | Median (range) | Effect size (95% CI)c | Median (range) | Effect size (95% CI)c | ||||
|---|---|---|---|---|---|---|---|---|
| Unaided baselinea | SA self-fitting (post 4-wk trial)a | IA self-fitting (post 4-wk trial)a | SA benefit (unaided – aided)b | IA benefit (unaided – aided)b | ||||
| EC | 29.0 (5.0 to 97.0) | 11.5 (1.0 to 71.0) | 11.0 (1.0 to 96.0) | Rosenthal r = −0.1 (−0.4 to 0.1) | 15.5 (−21.0 to 86.0) | 15.0 (−19.0 to 85.0) | Rosenthal r = −0.1 (−0.4 to 0.1) | |
| BN | 60.0 (11.0 to 95.0) | 26.0 (5.0 to 81.0) | 26.0 (1.0 to 64.0) | Cohen d = −0.1 (−0.5 to 0.2) | 28.5 (−20.0 to 69.0) | 32.0 (−10.0 to 67.0) | Cohen d = 0.1 (−0.2 to 0.5) | |
| RV | 46.0 (10.0 to 83.0) | 23.0 (5.0 to 56.0) | 19.0 (1.0 to 52.0) | Cohen d = −0.3 (−0.6 to 0.1) | 19.0 (−7.0 to 62.0) | 22.5 (−21.0 to 60.0) | Cohen d = 0.3 (−0.1 to 0.6) | |
| AV | 28.0 (1.0 to 91.0) | 40.0 (5.0 to 95.0) | 40.0 (3.0 to 91.0) | Rosenthal r = −0.2 (−0.5 to 0.0)e | −12.0 (−82.0 to 33.0) | −2.0 (−86.0 to 52.0) | Rosenthal r = −0.2 (−0.5 to 0.0)e | |
| Globald | 45.5 (13.0 to 92.0) | 20.0 (7.0 to 60.0) | 19.5 (1.0 to 54.0) | Cohen d = −0.2 (−0.6 to 0.2) | 22.0 (−14.0 to 71.0) | 22.5 (−4.0 to 65.0) | Cohen d = 0.2 (−0.2 to 0.6) | |
Abbreviations: APHAB, Abbreviated Profile of Hearing Aid Benefit; EC, ease of communication; BN, background noise; RV, reverberation; AV, aversion; SA, self-adjustment; IA, in situ audiometry.
Unaided indicates without hearing aid. A lower score indicates less communication difficulty and a higher score indicates greater communication difficulty.
When calculating benefit (unaided-aided), a higher score indicates a higher degree of benefit.
Effect size for normally distributed variables was calculated using Cohen d and Rosenthal r for nonnormal distribution.
The mean score for all subscales, excluding aversiveness.
The upper limit of the 95% CI is reported as 0.0 owing to rounding but is slightly greater than zero.
Figure 1. Outcome Measures Across the Trial For the In Situ Audiometry (IA) Self-Fitting and Self-Adjustment (SA) Self-Fitting.

A, Abbreviated Profile of Hearing Aid Benefit (APHAB) scores ranging from 1% to 99%, with higher scores indicating better outcomes. B, International Outcome Inventory for Hearing Aids (IOI-HA) total scores ranging from 1 to 35, with higher scores indicating better outcomes. C, Digits-in-noise (DIN) benefit scores after fitting ranging from −22.5 to 22.5, with higher scores indicating better outcomes. D, DIN benefit scores after field trial ranging from −22.5 to 22.5, with higher scores indicating better outcomes. E, QuickSIN benefit scores after fitting ranging from −25.5 to 25.5, with higher scores indicating better outcomes. F, QuickSIN benefit scores after field trial ranging from −25.5 to 25.5, with higher scores indicating better outcomes. Violin plots indicate kernel probability density. Boxes are IQR with median, and whiskers are 1.5 times the IQR.
International Outcome Inventory for Hearing Aids
There was no clinically meaningful difference between the 2 self-fitting strategies (Rosenthal r = 0.0; 95% CI,−0.3 to 0.2) for IOI-HA total scores. However, the self-adjustment self-fitting method demonstrated meaningfully longer daily use (Rosenthal r = −0.3; 95% CI, −0.5 to 0.0) and higher satisfaction (Rosenthal r = −0.4; 95% CI, −0.6 to −0.1) compared with the in situ audiometry self-fitting strategy. No meaningful differences were found for the other subscales (eFigure 6 in Supplement 3; Table 3; Figure 1).
Table 3. IOI-HA Scores After Self-Adjustment and In Situ Audiometry Self-Fitting Among 28 Participants.
| IOI-HA subscale | Median (range) | Rosenthal r (95% CI)b | |
|---|---|---|---|
| SA self-fitting (after 4-wk field trial)a | IA self-fitting (after 4-wk field trial)a | ||
| Use | 4.0 (3.0 to 5.0) | 4.0 (2.0 to 5.0) | −0.3 (−0.5 to 0.0)c,e |
| Benefit | 4.0 (3.0 to 5.0) | 4.0 (3.0 to 5.0) | −0.1 (−0.3 to 0.1) |
| Residual activity limitation | 4.0 (3.0 to 5.0) | 4.0 (3.0 to 5.0) | −0.2 (−0.4 to 0.1) |
| Satisfaction | 5.0 (3.0 to 5.0) | 4.0 (2.0 to 5.0) | −0.4 (−0.6 to −0.1)e |
| Residual participation restriction | 4.0 (1.0 to 5.0) | 4.0 (2.0 to 5.0) | −0.1 (−0.4 to 0.1) |
| Impact on others | 5.0 (2.0 to 5.0) | 5.0 (3.0 to 5.0) | −0.2 (−0.4 to 0.1) |
| Quality of life | 4.0 (3.0 to 5.0) | 4.0 (2.0 to 5.0) | −0.1 (−0.3 to 0.2) |
| Totald | 29.0 (25.0 to 33.0) | 29.0 (23.0 to 34.0) | 0.0 (−0.3 to 0.2) |
Abbreviations: IOI-HA, International Outcomes Inventory for Hearing Aids; SA, self-adjustment; IA, in situ audiometry.
Benefit is rated using 5 ordinal response categories with a lower score indicating poorer outcomes and a higher score indicating better outcomes.
Effect size calculated using Rosenthal r.
The upper limit of the 95% CI is reported as 0.0 owing to rounding but is slightly greater than zero.
Calculated as the sum of all 7 IOI-HA items.
Clinically meaningful effect size; r ≥0.3.
Speech-In-Noise Tests
Benefit, calculated as the difference in aided from unaided scores, showed no clinically meaningful differences between self-fitting strategies in either the DIN (Cohen d = 0.0; 95% CI, −0.4 to 0.4) or QuickSIN (Cohen d = −0.1; 95% CI, −0.4 to 0.3) immediately after fitting. Benefit scores after 4 weeks of hearing aid use also showed no clinically meaningful differences between self-fitting strategies in DIN (Cohen d = 0.2; 95% CI, −0.1 to 0.6) or QuickSIN scores (Rosenthal r = −0.1; 95% CI, −0.3 to 0.2) (Figure 1; eTable 1 in Supplement 3).
Real-Ear Measurements
For both self-fitting strategies, the average difference in output between 125 and 8000 Hz was within 5 dB of the National Acoustic Laboratories–Non-Linear 2 (NAL-NL2)18 targets for average (65 dB sound pressure level [SPL]) and loud speech (75 dB SPL) immediately after fitting and following 4 weeks of using the hearing aids. The mean (SD) differences for soft speech (55 dB SPL) between 125 and 8000 Hz slightly exceeded 5 dB (after fitting: in situ audiometry 5.9 (4.4) dB, self-adjustment 6.6 (5.8) dB; after trial: in situ audiometry 6.8 (5.7) dB, self-adjustment 6.8 (6.0) dB. No clinically meaningful differences were observed between average speech (65 dB SPL) immediately after fitting (Cohen d = −0.2; 95% CI, −0.6 to 0.2) or after the field trial (Cohen d = −0.1; 95% CI, −0.4 to 0.3) as illustrated in Figure 2. Similar findings were observed for soft speech (55 dB SPL) immediately after fitting (Cohen d = −0.2; 95% CI, −0.5 to 0.1]) and after the field trial (Cohen d = −0.0; 95% CI, −0.3 to 0.3) as well as for loud speech (75 dB SPL) immediately after fitting (Cohen d = −0.1; 95% CI, −0.4 to 0.1) and after the field trial (Cohen d = −0.1; 95% CI, −0.3 to 0.2). See eFigures 6 and 7 in Supplement 3.
Figure 2. Comparison of the National Acoustic Laboratories–Non-Linear 2 (NAL-NL2) Targets and Measured Real-Ear Output in dB Sound Pressure Levels (SPL) for Left and Right Ears Combined.

A, Comparison of the NAL-NL2 target and real ear output in dB SPL for the in situ audiometry self-fitting strategy immediately after fitting. B, Comparison of NAL-NL2 target and real ear output in dB SPL for the self-adjustment self-fitting strategy immediately after fitting. C, Comparison of NAL-NL2 target and real ear output in dB SPL for the in situ audiometry self-fitting strategy after field trial. D, Comparison of NAL-NL2 target and real ear output in dB SPL for the self-adjustment self-fitting strategy after field trial. The stimulus was a 65 dB SPL International Speech Test signal. IA indicates in situ audiometry; REAR, SA, self-adjustment.
Adjustments After In Situ Self-Fitting Field Trial
During the field trial of the in situ fitting process with recommended settings, participants made various adjustments to their hearing aid settings. Of the 28 participants, 8 (28.6%) maintained their original world volume settings (ie, overall gain across the frequencies), while 10 (35.7%) opted to increase the world volume with a mean (SD) increase of 11.1 (9.7) increments. However, 10 participants (35.7%) decreased the world volume by a mean (SD) of 8.2 (3.0) increments. Moreover, 10 participants (35.7%) did not alter the spectral tilt settings. Among participants who made adjustments to the spectral tilt settings, 6 participants (21.4%) opted for enhancing the treble, with a mean (SD) adjustment of 11.7 (9.8) increments. Conversely, 12 participants (42.9%) preferred a shift toward lower bass, with a mean (SD) change of 9.8 (8.0) increments.
Discussion
This crossover trial directly compared 2 self-fitting strategies, self-adjustment and in situ audiometry. Both methods showed no clinically meaningful differences in APHAB benefit, overall IOI-HA satisfaction, speech-in-noise performance, and real-ear measurements.
Although APHAB benefit and overall IOI-HA satisfaction were similar between self-fitting strategies, self-adjustment participants reported meaningfully higher satisfaction and longer daily use on 2 IOI-HA subsections. This could be due to greater control and customization with self-adjustment, allowing users to tailor hearing aid settings to their preferences without a predefined fitting based on an in situ hearing test, potentially leading to a more comfortable perceived listening experience. However, the lower bound of the CIs suggests that in the broader population, these effects might not be as meaningful or may not even exist. Nonetheless, this aligns with previous research demonstrating user preference for self-selected settings.5,19
Participants’ adjustments during the in situ self-fitting field trial highlight the importance of providing users with the flexibility to customize settings based on their unique needs. It is crucial to consider the diverse needs of different consumer groups because specific self-fitting strategies may be more suitable for certain individuals.7,19 For instance, some may prefer self-adjustment strategies if they are comfortable with technology and desire complete control over the fitting process. However, those who are less familiar with technology or require guidance may benefit more from in situ audiometry fitting strategies, which provide a structured approach and assist users in obtaining initial fitting parameters. In some instances, using a combined approach could be beneficial, as it offers a comprehensive solution that addresses diverse needs and preferences.
Consistent with prior research, this study showed that self-fit users prefer a comfort fit, which may result in higher frequency amplification levels slightly below the NAL-NL2 targets.5,14,19 User preferences may lead to self-reported benefits but may not always result in optimal clinical performance. When evaluating hearing aid fitting strategies, it is important to distinguish between subjective user preferences and objective behavioral measures. Although user satisfaction is valuable, it should be balanced with the need to ensure hearing aid settings are optimal for clinical performance, as determined by objective assessments.
The absence of clinically meaningful differences in speech-in-noise scores between self-fitting strategies for both DIN and QuickSIN tests suggests that individuals can achieve comparable speech perception in noise regardless of the self-fitting method. However, there are minimal benefits when it comes to adaptive speech-in-noise testing (eTable 1 in Supplement 3).5,14 Speech-in-noise test performance in a lab setting may not accurately reflect scenarios where speech is heard in a noisy environment. This could be due to several factors, such as the nature of the noise and speech stimuli used in tests like DIN and QuickSIN, which may not completely replicate the complexity of everyday environments, such as a restaurant with ambient noise. Future studies should consider using fixed speech-in-noise testing instead of only adaptive speech-in-noise testing to ensure standardized and comparable assessments of speech perception benefits.
Limitations
The findings of this clinical trial support the effectiveness of self-fitting strategies for OTC-SF hearing aids, aligning with previous studies.5,13,14 However, this study has some limitations. The need to replace participants due to a technical issue with the research app compromised the randomization process, leading to use of a pseudorandomization method. We did not include a washout period, which could have further strengthened the internal validity of the study. Our sample size also limited our ability to conduct robust statistical analyses investigating correlations between outcomes and demographic variables. In addition, the study’s design limited our ability to assess the long-term outcomes of the different self-fitting strategies.
Conclusions
To our knowledge, this clinical trial is the first to show that for OTC-SF hearing aids, self-adjustment and in situ audiometry fitting strategies were both associated with positive outcomes in overall APHAB, overall IOI-HA satisfaction, speech-in-noise performance, and real-ear measurements. Self-adjustment may meaningfully improve hearing aid satisfaction and daily use on 2 IOI-HA subscales, underscoring the value of user engagement in self-fitting. Further studies with extended follow-up and larger sample sizes are needed to assess the long-term effectiveness of these approaches individually and in combination, as well as to investigate correlations between outcomes, degree of hearing loss, age, and other relevant variables.
Trial Protocol
Statistical Analysis Plan
eFigure 1. Study Flow Diagram
eFigure 2. The Lexie app interface for the self-adjustment self-fitting process
eFigure 3. The Lexie app interface for the in-situ audiometry self-fitting process
eFigure 4. (A) Distribution of conventional pure tone audiometric frequencies for the left ear, (B) Distribution of conventional pure tone audiometric frequencies for the right ear
eFigure 5. International Outcomes Inventory for Hearing Aids (IOI-HA) items for the in-situ audiometry (IA) self-fitting and self-adjustment (SA) self-fitting measured after approximately four weeks of hearing aid use
eFigure 6. (A) Comparison of NAL-NL2 target and real ear output in dB SPL for the in-situ audiometry self-fitting strategy immediately after fitting, (B) Comparison of NAL-NL2 target and real ear output in dB SPL for the self-adjustment self-fitting strategy immediately after fitting, (C) Comparison of NAL-NL2 target and real ear output in dB SPL for the in-situ audiometry self-fitting strategy after field trial, (D) Comparison of NAL-NL2 target and real ear output in dB SPL for the self-adjustment self-fitting strategy after field trial
eFigure 7. (A) Comparison of NAL-NL2 target and real ear output in dB SPL for the in-situ audiometry self-fitting strategy immediately after fitting, (B) Comparison of NAL-NL2 target and real ear output in dB SPL for the self-adjustment self-fitting strategy immediately after fitting, (C) Comparison of NAL-NL2 target and real ear output in dB SPL for the in-situ audiometry self-fitting strategy after field trial, (D) Comparison of NAL-NL2 target and real ear output in dB SPL for the self-adjustment self-fitting strategy after field trial
eTable 1. DIN and QuickSIN Benefit Scores for the Self-Adjustment and In-situ Audiometry Self-fitting Strategies After Fitting and After Field Trial
eTable 2. CONSORT checklist of information to include when reporting randomised crossover trials
eTable 3. Information to include in abstract of report of randomised crossover trial: extension of CONSORT for abstracts checklist
Data Sharing Statement
References
- 1.World Health Organization . World Report on Hearing. 2021. Accessed January 22, 2024. https://www.who.int/publications/i/item/world-report-on-hearing
- 2.Humes LEUS. U.S. population data on hearing loss, trouble hearing, and hearing-device use in adults: National Health and Nutrition Examination Survey, 2011-12, 2015-16, and 2017-20. Trends Hear. 2023;27:23312165231160978. doi: 10.1177/23312165231160978 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.Jilla AM, Johnson CE, Huntington-Klein N. Hearing aid affordability in the United States. Disabil Rehabil Assist Technol. 2023;18(3):246-252. doi: 10.1080/17483107.2020.1822449 [DOI] [PubMed] [Google Scholar]
- 4.Food and Drug Administration . Medical Devices; Ear, Nose, and Throat Devices. Establishing Over-the-Counter Hearing Aids; 2022. [Google Scholar]
- 5.Sabin AT, Van Tasell DJ, Rabinowitz B, Dhar S. Validation of a self-fitting method for over-the-counter hearing aids. Trends Hear. 2020;24:2331216519900589. doi: 10.1177/2331216519900589 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6.Convery E, Keidser G, Seeto M, Yeend I, Freeston K. Factors affecting reliability and validity of self-directed automatic in situ audiometry: implications for self-fitting hearing AIDS. J Am Acad Audiol. 2015;26(1):5-18. doi: 10.3766/jaaa.26.1.2 [DOI] [PubMed] [Google Scholar]
- 7.Convery E, Keidser G, Dillon H, Hartley L. A self-fitting hearing aid: need and concept. Trends Amplif. 2011;15(4):157-166. doi: 10.1177/1084713811427707 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8.Boothroyd A, Mackersie C. A “Goldilocks” approach to hearing-aid self-fitting: user interactions. Am J Audiol. 2017;26(3S):430-435. doi: 10.1044/2017_AJA-16-0125 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.Mackersie C, Boothroyd A, Lithgow A. A “Goldilocks” approach to hearing aid self-fitting: ear-canal output and speech intelligibility index. Ear Hear. 2019;40(1):107-115. https://journals.lww.com/ear-hearing/fulltext/2019/01000/a__goldilocks__approach_to_hearing_aid.12.aspx. doi: 10.1097/AUD.0000000000000617 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.O’Brien A, Keidser G, Yeend I, Hartley L, Dillon H. Validity and reliability of in-situ air conduction thresholds measured through hearing aids coupled to closed and open instant-fit tips. Int J Audiol. 2010;49(12):868-876. doi: 10.3109/14992027.2010.503664 [DOI] [PubMed] [Google Scholar]
- 11.Boymans M, Dreschler WA. In situ hearing tests for the purpose of a self-fit hearing aid. Audiol Neurootol. 2017;22(1):15-23. doi: 10.1159/000457829 [DOI] [PubMed] [Google Scholar]
- 12.Convery E, Keidser G, Hickson L, Meyer C. Factors associated with successful setup of a self-fitting hearing aid and the need for personalized support. Ear Hear. 2019;40(4):794-804. doi: 10.1097/AUD.0000000000000663 [DOI] [PubMed] [Google Scholar]
- 13.Swanepoel W, Oosthuizen I, Graham MA, Manchaiah V. Comparing hearing aid outcomes in adults using over-the-counter and hearing care professional service delivery models. Am J Audiol. 2023;32(2):314-322. doi: 10.1044/2022_AJA-22-00130 [DOI] [PubMed] [Google Scholar]
- 14.De Sousa KC, Manchaiah V, Moore DR, Graham MA, Swanepoel W. Effectiveness of an over-the-counter self-fitting hearing aid compared with an audiologist-fitted hearing aid: a randomized clinical trial. JAMA Otolaryngol Head Neck Surg. 2023;149(6):522-530. doi: 10.1001/jamaoto.2023.0376 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15.Manchaiah V, Swanepoel W, Sharma A. Prioritizing research on over-the-counter (OTC) hearing aids for age-related hearing loss. Front Aging. 2023;4:1105879. doi: 10.3389/fragi.2023.1105879 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16.EF SET. English test with free certification. 2024. Accessed January 22, 2024. https://www.efset.org
- 17.Tellez A, Garcia Cadena C, Corral-Verdugo V. Effect size, confidence intervals and statistical power in psychological research. Psychol Russ. 2015;8:27-46. doi: 10.11621/pir.2015.0303 [DOI] [Google Scholar]
- 18.Keidser G, Dillon H, Carter L, O’Brien A. NAL-NL2 empirical adjustments. Trends Amplif. 2012;16(4):211-223. doi: 10.1177/1084713812468511 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19.Convery E, Keidser G, Hartley L. Perception of a self-fitting hearing aid among urban-dwelling hearing-impaired adults in a developed country. Trends Amplif. 2011;15(4):175-183. doi: 10.1177/1084713811424886 [DOI] [PMC free article] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
Trial Protocol
Statistical Analysis Plan
eFigure 1. Study Flow Diagram
eFigure 2. The Lexie app interface for the self-adjustment self-fitting process
eFigure 3. The Lexie app interface for the in-situ audiometry self-fitting process
eFigure 4. (A) Distribution of conventional pure tone audiometric frequencies for the left ear, (B) Distribution of conventional pure tone audiometric frequencies for the right ear
eFigure 5. International Outcomes Inventory for Hearing Aids (IOI-HA) items for the in-situ audiometry (IA) self-fitting and self-adjustment (SA) self-fitting measured after approximately four weeks of hearing aid use
eFigure 6. (A) Comparison of NAL-NL2 target and real ear output in dB SPL for the in-situ audiometry self-fitting strategy immediately after fitting, (B) Comparison of NAL-NL2 target and real ear output in dB SPL for the self-adjustment self-fitting strategy immediately after fitting, (C) Comparison of NAL-NL2 target and real ear output in dB SPL for the in-situ audiometry self-fitting strategy after field trial, (D) Comparison of NAL-NL2 target and real ear output in dB SPL for the self-adjustment self-fitting strategy after field trial
eFigure 7. (A) Comparison of NAL-NL2 target and real ear output in dB SPL for the in-situ audiometry self-fitting strategy immediately after fitting, (B) Comparison of NAL-NL2 target and real ear output in dB SPL for the self-adjustment self-fitting strategy immediately after fitting, (C) Comparison of NAL-NL2 target and real ear output in dB SPL for the in-situ audiometry self-fitting strategy after field trial, (D) Comparison of NAL-NL2 target and real ear output in dB SPL for the self-adjustment self-fitting strategy after field trial
eTable 1. DIN and QuickSIN Benefit Scores for the Self-Adjustment and In-situ Audiometry Self-fitting Strategies After Fitting and After Field Trial
eTable 2. CONSORT checklist of information to include when reporting randomised crossover trials
eTable 3. Information to include in abstract of report of randomised crossover trial: extension of CONSORT for abstracts checklist