Abstract
Objectives. To derive a collection efficiency formula,, for cylindrical ionization chambers in pulsed radiation beams from a volume recombination model of Boag et al (1996 Phys. Med. Biol. 41 885–97) including free electrons. To validate and a parallel plate chamber formula using an ion transport code and calculate changes in collection efficiencies caused by electric field charge screening at 0.1–100 mGy doses-per-pulse. And to determine collection efficiencies predicted at infinite voltage in the absence of avalanche effects by fitting scaled formulae to efficiencies computed for 100–400 V chamber voltages and 10 and 100 mGy doses-per-pulse. Approach. Calculations were performed for an idealized parallel plate chamber with 2 mm electrode separation , and for an idealized cylindrical chamber with 0.5 and 2.333 mm inner and electrode radii and . Main results. and predict the same collection efficiencies for cylindrical and parallel plate chambers satisfying , an equivalence condition met by the chambers studied. Without charge screening, efficiencies computed using the code equalled and . With screening, efficiencies changed by ⩽0.03%, ⩽1.1% and ⩽21.3% at 1, 10 and 100 mGy doses-per-pulse, and differed between the chambers by ⩽0.9% and ⩽19.6% at ⩽10 and 100 mGy dose-per-pulse. For fits of and , values were ⩽1.2% and ⩽17.6% from unity at 10 and 100 mGy per pulse respectively, closer than for other formulae tested. Significance. Allowing for screening, and described computed collection efficiencies to within 0.03%, 1.1% and 21.3% at doses-per-pulse ⩽1, 10 and 100 mGy. Equivalence of the two chambers broke down at 100 mGy per pulse. Departures of values from unity suggest that collection efficiencies determined experimentally by fitting or to readings made at multiple voltages will be accurate to within 1.2% and 17.6% at 10 and 100 mGy per pulse respectively.
Keywords: ionization chamber, ion recombination, dose-per-pulse, collection efficiency, charge screening
1. Introduction
Doses-per-pulse are typically around 0.3 mGy in photon beams generated for radiotherapy (RT) treatments by linear accelerators (linacs) with flattening filters. In comparison, doses-per-pulse delivered by modern filter-free linacs, intra-operative electron beams and novel FLASH electron systems are respectively factors of 3, 30–300 and 30–15 000 higher. These larger doses-per-pulse lead to increased levels of ion recombination and reduced charge collection in radiation detectors, and this has renewed and intensified interest in recombination effects in dosimetry (Di Martino et al 2005, Laitano et al 2006, Christensen et al 2016, Petersson et al 2017, Gotz et al 2017, McManus et al 2020, Kranzer et al 2021).
Boag (1950) formulated a model of volume recombination between uniform clouds of positive and negative ions generated by a pulse of radiation, as they drift towards the cathode and anode of a parallel plate ionization chamber. From the model Boag obtained a formula f for chamber collection efficiency, defined as the ratio of charge collected to charge generated within the chamber. He also derived a geometric condition for equivalent recombination in cylindrical and parallel plate ionization chambers (Boag 1950, Boag and Currant 1980). Specifically, if the inner and outer electrode radii r in and r out of a cylindrical chamber and the electrode gap d of a parallel plate chamber satisfy the condition
then according to Boag’s 1950 model the collection efficiencies of the chambers should be the same when the same voltage is applied to each. If correct, this relationship allows collection efficiencies calculated for parallel plate ionization chambers to be carried across to equivalent cylindrical chambers, which is practically important since cylindrical chambers are used more often in RT dosimetry.
Subsequently, Boag et al (1996) extended the volume recombination model to account for free electron effects. The extended model describes the initial formation of uniform clouds of positive ions and free electrons, followed by rapid collection at the anode of a fraction p of the electrons and capture of the rest by oxygen molecules, forming a non-uniform cloud of negative ions. The negative ions drift towards the anode and are either collected there or recombine with positive ions first. Three approximate solutions of this model were derived for parallel plate chambers, together with associated collection efficiency formulae , and . These approximate formulae predict collection efficiencies larger than those given by Boag’s 1950 f formula which excluded free electron effects.
Recently, the extended model was solved exactly for parallel plate chambers, across which the initial negative ion density varies exponentially (Fenwick and Kumar 2023). A collection efficiency formula f exp was obtained that gave efficiencies in agreement with values calculated using a numerical code describing ion transport and recombination. Here we solve the same model exactly for a cylindrical geometry, obtaining a formula for the collection efficiency of cylindrical chambers, in which the initial negative ion density varies according to a Gaussian function. We also derive three approximate collection efficiency formulae , and analogous to the three approximate formulae derived by Boag et al (1996) for the parallel plate geometry.
From the and f exp formulae we identify a condition for parallel plate and cylindrical chambers to have the same collection efficiencies, and we compare it to the condition found by Boag et al (1950) from his recombination model which excluded free electron effects.
The analytical development of recombination models provides more general results than can be obtained numerically. For models that include free electron effects, however, it involves two simplifications. First, it ignores screening of the electric field E within ionization chambers due to charge imbalances (ICRU 1982). This simplification is present in the derivations of collection efficiency formulae in this study and those of Boag et al (1996) and Fenwick and Kumar (2023). And second, it treats the rate-constant γ for attachment of free electrons to oxygen molecules as being constant throughout a chamber. For parallel plate chambers this is largely valid, but in cylindrical chambers the electric field strength varies with distance from the chamber axis, and γ is known to depend on E (Hochhäuser et al 1994, Boissonnat 2015).
To determine the effect of these additional factors, we calculate collection efficiencies for chambers using a numerical code that describes ion transport in the presence of E-field charge screening and γ(E) variation. As the derivation of is lengthy the code is initially used with charge screening and γ(E) variation turned off, to check the formula in the conditions for which it was derived. Then we study by how much computed collection efficiencies change when γ(E) variation and charge screening are turned on in a cylindrical chamber, initially adding γ(E) variation and then screening to explore the relative impact of each factor. We also calculate changes in efficiencies computed for parallel plate chambers when screening is turned on. Further, we determine how well the analytical formulae describe the collection efficiencies computed with these effects turned on, and whether the equivalence condition between parallel plate and cylindrical chambers remains the same.
The code is also used to assess the accuracy of a common experimental method for determining chamber collection efficiencies, in which readings are taken at multiple chamber voltages and analyzed using recombination models (Boag 1950, Boag and Currant 1980, Boag et al 1996, Bruggmoser et al 2018a, Kranzer et al 2021). These multi-voltage methods relate the chamber reading obtained for a fixed dose D at voltage V to the collection efficiency at that voltage CE(V) via
Here is the reading in the absence of recombination, which is estimated by fitting readings made at practical voltages using the expression in which is a closed-form collection efficiency formula such as f (Boag 1950). Given a fitted value, collection efficiencies can be calculated directly from readings using equation (2).
The accuracy of this approach is governed by the accuracy of the fitted value. In the final part of this study, we use a technique described in the materials and methods to assess the accuracy of estimation, working from collection efficiencies obtained from the ion transport code. Efficiencies calculated with and without E-field screening and variation are analyzed to characterize the impact of these effects on the experimental estimation of collection efficiencies. Several formulae are used for the term in the multi-voltage fitting process, and we investigate which performs best.
To make the study as accessible as possible, physical models of charge transport and recombination in ionization chambers are summarized in the materials and methods , together with approaches used in the derivation and analysis of collection efficiency formulae. The derivations are set out in the appendix, and efficiencies obtained from the formulae and numerical code are presented and analyzed in the results .
2. Materials and methods
2.1. Nomenclature
Quantities appearing outside the appendix are denoted by the following symbols—
t, time after a radiation pulse
te, time by which no free electrons remain in an ionization chamber
V, potential difference across a chamber
, electric field and field strength
, distance from the cathode towards the anode of a parallel plate chamber
d, electrode spacing in a parallel plate chamber
r, radius of a circle centred on the long axis of a cylindrical chamber
r in, r out, radii of cylindrical chamber inner and outer electrodes
γ, rate-constant for attachment of free electrons to oxygen
ve, drift velocity of free electrons
k1, k2, ke , mobilities of positive and negative ions and electrons
a, drift coefficient for free electrons
p, fraction of free electrons collected at the anode
, dimensionless parameter
n0 , initial uniform density of positively charged ions or free electrons
ne , density distribution of free electrons
, density distributions of positively and negatively charged ions
α, rate-constant for recombination between positive and negative ions
e, electronic charge
, permittivity of air at 20 °C and 101.325 kPa
, dimensionless combination of parameters
, dimensionless combination of parameters
g2 , geometric factor equal to d2 or
, empirical ionization chamber collection efficiency formula
, Boag ionization chamber collection efficiency without free electron effects
, , , , parallel plate chamber efficiency formulae with free electrons
, , , , cylindrical chamber efficiency formulae with free electrons
, chamber reading for dose D at voltage V
, multi-voltage estimate of chamber reading with no recombination
, model of collection efficiency variation with voltage
, estimate of collection efficiency at V from multi-voltage data fit
E1 , exponential integral function
The polarity of cylindrical ionization chambers is described as positive/negative when the central electrode is at a higher/lower potential than the outer one and forms the anode/cathode. A glossary in the appendix lists further key symbols used there.
2.2. Charge transport and ion recombination models
In Boag’s 1950 model, a radiation pulse generates uniform clouds of free electrons and positively charged ions, both with initial number density n0 . The electrons rapidly attach to oxygen molecules, forming a uniform cloud of negatively charged ions also of density n0 . Subsequently the positive and negative ions drift towards the cathode and anode, with average speeds given by their mobilities k1 and k2 multiplied by the local electric field strength E. During the drift the positive and negative ions recombine at a rate given by the product of their local densities and the rate-constant α. The resulting collection efficiency works out as
and g2 is given by for parallel plate chambers and for cylindrical chambers. The model collection efficiency falls with initial ion density and electrode separation but rises with the potential difference V across the chamber.
In the extended model of Boag et al (1996) the free electrons drift towards the anode with an average speed ve given by their mobility ke multiplied by the electric field strength E. Electron drift speeds are much faster than those of ions, and consequently in analytical models the electrons are considered to have either been collected at the anode or become attached to oxygen almost instantly, by time t≈ 0 following a radiation pulse, before the positive ions have moved from the locations in which they were generated.
The attachment rate is the product of the local electron density ne and the rate-constant γ, from which it follows that in a parallel plate chamber the fraction p of electrons reaching the anode is
The drift constant, a, in equation (5) is the attachment rate-constant per distance travelled by the electrons, equal to and thus since the E-field strength in the chamber is V/d. The electrons that attach to oxygen form a cloud of negative ions, which in a parallel plate chamber has the initial density distribution
where denotes distance from the cathode (Boag et al 1996). The non-uniformity of this distribution makes it more difficult to solve the ion recombination equations. Consequently, Boag et al (1996) obtained collection efficiency formulae for parallel plate chambers based on three approximations to the distribution of equation (6) which describe uniform density levels—
-
•
n0 (1-p) throughout the ionization chamber;
-
•
n0 throughout a region distant pd→d from the cathode, and zero elsewhere;
-
•
in the region → from the cathode where , and zero elsewhere.
For these initial negative ion distributions, Boag et al (1996) derived the approximate collection efficiency formulae
![]() |
![]() |
![]() |
Whereas depends only on the composite parameter , these formulae additionally depend on p and thus . Across the ranges u = 0.01–0.50 and p= 0.1–0.6, Boag et al (1996) found that differences between collection efficiencies predicted by the three approximate formulae reached 5.1% relative.
Solving the extended model exactly for a parallel plate chamber with the initial exponential negative ion distribution of equation (6) instead gives a collection efficiency formula (Fenwick and Kumar 2023)
where
and E 1 is the standard exponential integral function (Gradshteyn and Rhyzik 2007) for which tabulated values are widely available.
Values of electron and ion mobilities ke, k 1, k 2 used in this study are listed in table 1, together with the electron attachment and ion recombination rate-constants and . They are derived from data compiled by Gotz et al (2017). The table also includes a formula describing the variation of γ with E, a value for the permittivity of air, , and an n 0-per-Gy value which when multiplied by the dose-per-pulse gives the initial density n 0 of free electrons and positive ions initially generated by a radiation pulse.
Table 1.
Values of electron and ion mobilities and rate-constants of electron attachment and ion recombination used in collection efficiency calculations. The number of free electrons liberated per Gy and the permittivity of dry air are also listed.
| Parameter | Value a |
|---|---|
| ke | 8.3 × 10−2 m2V−1s−1 |
| k1 | 1.87 × 10−4 m2V−1s−1 |
| k2 | 2.09 × 10−4 m2V−1s−1 |
| γ | [1.1 + 11.3 × exp(−1.04 × 10−5 × E b )] × 107 s−1 for E ⩾0.327 × 105 V m−1 |
| 7.0 × 107 + 657 × E s−1 for E < 0.327 × 105 V m−1 | |
| α | 1.30 × 10−12 m3s−1 |
| n 0-per-Gy | 2.22 × 1017 m−3Gy−1 |
| 8.86 × 10−12 m−3kg−1s4A2 |
All values derived from Gotz et al (2017) except for n0-per-Gy, the initial free electron density per Gy, which follows directly from (W/e) and the density of dry air at 20 °C and 101.325 kPa, and which is the permittivity of air at the same temperature and pressure.
E is electric field-strength measured in V m−1.
2.3. Collection efficiency derivations for the extended model in a cylindrical geometry
Figure 1 shows a schematic cross-section through a cylindrical ionization chamber. The chamber is idealized as being infinitely long rather than having a finite length like practical chambers which employ guard-rings to retain electric field uniformity towards the chamber ends. The same geometry was studied by Boag (1950). The parallel plate chamber geometry studied in this work is similarly idealized as comprising electrodes with infinite cross-sectional area but narrow separation, again reflecting the approach taken by Boag (1950) and Boag et al (1996).
Figure 1.

Schematic cross-section of a cylindrical ionization chamber, showing inner and outer electrodes (black) with radii rin and rout . Annuli of ions of opposite sign are also shown moving inwards and outwards (blue and red). In the exact derivation of collection efficiencies presented in the appendix, and denote the radial locations at which these ions are formed, while (green) denotes the radius at which ions belonging to the two annuli cross over as they drift to the electrodes. In the figure, lies closer to than to . This represents the situation when the ions drifting inwards are positive, since these ions have ∼20% lower mobilities than do negative ions (table 1).
Collection efficiency derivations for the cylindrical geometry are detailed in the appendix. Several times we invoke a counter-intuitive deduction of Boag (1950), that densities of ions travelling towards the inner or outer electrodes of cylindrical ionization chambers do not change with radial distance from the chamber axis, , except due to ion recombination. Clearly, the circumferential length occupied by the thin annulus of ions shown travelling towards the central electrode in figure 1 falls with decreasing . However, the variation of E with causes the leading edge of the ions to move inward faster than the outer edge. This increases the radial thickness of the annulus throughout which the ions are spread, to a degree that exactly offsets the fall in circumferential length.
The gradual broadening of the annulus of ions might be expected to cause it to begin overlapping neighbouring annuli. This would invalidate Boag’s deduction but does not in fact happen. Consider two neighbouring annuli both moving towards the centre of the chamber, the inner one slightly faster on average. Since the ions at the outer edge of the inner annulus and the inner edge of the outer annulus experience the same E-field and therefore move at the same average speed, the two annuli neither overlap nor pull apart as they drift towards the central electrode, despite both widening.
2.4. Electric field calculations and the numerical code
In the absence of charge imbalance, the electric field in a parallel plate chamber runs along the -axis perpendicular to the planes of the electrodes with magnitude , whereas the field in a cylindrical chamber runs radially with magnitude .
These fields were used in the analytical derivations of collection efficiency formulae. However, charge imbalances do exist between positive ions and free electrons generated directly by a radiation pulse and the negative ions formed by attachment of electrons to oxygen. These imbalances partly screen the electric field, which is further modified as the ions drift towards the electrodes (ICRU 1982). Boag et al (1996) have estimated that errors in calculated collection efficiencies due to omission of screening effects should be small at doses-per-pulse up to 10 mGy, though more substantial at ⩾100 mGy.
From Gauss’s law, the electric fields in the presence of charge imbalance are described by
for parallel plate ionization chambers and
for cylindrical chambers (Gotz et al 2017, Kranzer et al 2021). Given the densities of positive and negative ions and electrons , and , the electronic charge e and the boundary condition that the path integral of across a chamber equals V, equations (12) and (13) can be solved numerically. Thus, the coupled charge transport and electric field evolution process can be computed using a repeating code in which the charge drift and recombination over a short period of time are calculated for a previously computed E-field, then E is updated to reflect the resulting change in charge distribution.
We implemented such a code in MATLAB R2022b (Mathworks, Natick, MA), adapting a previous program (Pardo et al 2005) to include free electrons. At early time-points the rapid drift of free electrons and much slower drift of ions were computed together, whereas in analytical derivations of collection efficiency formulae they are typically considered entirely sequentially. Transport and recombination were modelled thus
where denotes a point is space and the equations reflect the processes described in section 2.2 . The E-fields for parallel plate and cylindrical chambers were defined by equations (12) and (13) when considering screening, and by and when screening was not considered.
Within the code, space was discretized in steps . The time-step was 1/25th of the time for the fastest particle being tracked to traverse or . Thus
where te is the time by which no free electrons remain in the chamber, approximated as when the electron density everywhere has fallen to . Following a time-step, each point’s particle density was updated to reflect transport and recombination. The transport component was accounted for by subtracting from the density at each point the difference in densities between that point and the adjacent point from which particles flow into it, multiplied by the ratio of the particle transport distance at that time-step to or . The code is available for download at http://github.com/juancho-pm/Code-for-Fenwick-et-al-PMB2024.
2.5. Collection efficiency calculations
Collection efficiencies were calculated for a parallel plate chamber with an electrode separation of 2 mm, and for a cylindrical chamber with inner and outer electrode radii of 0.5 mm and 2.333 mm. The electrode separation of the parallel plate chamber matches the classic Markus chamber (PTW 23 343, Freiburg, Germany), the NACP Plane Parallel chamber (IBA, Louvain-la-Neuve, Belgium) and the Exradin A10 (Standard Imaging, Madison, WI). The inner radius of the cylindrical chamber matches the Exradin A1SL and A12S chambers, the IBA CC13 and FC23-C chambers (Iwafuchi et al 2022), and standard Farmer chambers including the Exradin A12, IBA FC65-G and NE2571A (Phoenix Dosimetry Ltd, Sandhurst, UK). The outer radius was chosen so that the cylindrical and parallel chambers were equivalent according to the condition of equation (1). It lies between the 2 mm outer radius of the Exradin A1SL chamber and the 3–3.15 mm outer radii of the rest.
For both chambers, collection efficiencies were calculated for chamber potential differences of 200 and 400 V, covering the range of commonly used voltages. Doses-per-pulse of 0.1, 1, 10 and 100 mGy were chosen, covering the range delivered by linacs with and without flattening filters and intra-operative electron beams, and the lower part of the range delivered by electron FLASH systems. For the cylindrical chamber, calculations were made for positive and negative polarities. The formula was initially checked against collection efficiencies obtained from the ion transport code run under the conditions for which the formula was derived. Specifically, the E-field strength was defined as and the rate-constant γ throughout the cylindrical chamber was set to its uniform level in the equivalent parallel plate chamber with the same voltage applied, calculated from the formula in table 1 with E defined as = 500 × V Vm−1.
Next, the effect of γ(E) variation was tested by re-running the code for the cylindrical chamber, with E again defined ignoring charge screening but γ varying with E as described in table 1. Finally, the full code was run for parallel plate and cylindrical chambers to check for the effects of screening. This was done for the 200 and 400 V potential differences for which the preceding sets of collection efficiencies were calculated, and for 100, 300, 700 and 1000 V to further characterize the variation of collection efficiency with chamber voltage. Computed efficiencies were plotted using the Origin software (OriginLab, Northampton, MA) and we assessed which of the collection efficiency formulae best described the efficiency data computed under the different conditions.
2.6. Multi-voltage estimation of collection efficiencies
The accuracy with which is estimated by fitting to readings made at multiple voltages V determines the accuracy of experimental multi-voltage methods for estimating collection efficiencies. It can be assessed by taking a series of collection efficiencies computed for different voltages, fitting them using where is a fitted scaling factor, and finding how close the value of is to a collection efficiency of one in the absence of recombination at . Since the collection efficiency formulae used to describe in this study all take values of one at , this is a check of the relative difference between unity and the fitted factor . This is the same as the relative difference between the true value of and the value obtained by fitting to measured readings , since the readings are essentially collection efficiencies multiplied by the true value.
We therefore fitted to collection efficiencies computed for the parallel plate and positively and negatively polarized cylindrical chambers. The parallel plate data were fitted using the , , , and formulae to describe , and the cylindrical chamber data were fitted using the cylindrical equivalents. Along with , values of the u and ad parameters of the parallel plate formulae were fitted to achieve the best description of the data, rather than being defined as in equations (4) and (5) and determined by the chamber geometry and the transport and recombination coefficients listed in table 1. Parameters of the cylindrical chamber formulae were similarly fitted. We also used an empirical logistic collection efficiency formula developed by Petersson et al (2017) which has the form
where DPP is the dose-per-pulse measured in mGy and ϵ and ν are fitted parameters.
Fitting was carried out using the nls function and nl2sol algorithm in R version 4.0.2 (R Foundation, Vienna, Austria). Complete sets of collection efficiencies calculated for chamber voltages of 100–1000 V were fitted, as were subsets calculated for voltages ⩽400 V to test the performance of the multi-voltage approach over a practical range of chamber voltages.
Initially we fitted idealized collection efficiencies calculated using the numerical code with γ(E) variation and E-field charge screening turned off. For cylindrical chambers, we also fitted collection efficiencies calculated with γ(E) variation turned on. For parallel plate and cylindrical chambers, the process was repeated with charge screening also turned on.
3. Results
3.1. Exact and approximate collection efficiency formulae for cylindrical chambers
Derivations of collection efficiency formulae from the model of Boag et al (1996) are detailed in the appendix. For a cylindrical chamber with positive polarity, the initial density distribution of negative ions formed by free electrons is shown to vary with r according to a Gaussian function
For negative polarity, the negative ion density varies as
and in both cases the fraction of free electrons collected is
Thus p has the same form as for parallel plate chambers (equation (5)) but with replacing ad, or equivalently replacing d2 .
Given the initial negative ion density of equation (19), the collection efficiency formula obtained by exactly solving the extended recombination model for a positively polarized cylindrical ionization chamber is
This has the same form as the formula of equation (10) for parallel plate chambers, but in the u and terms of equation (23) the d2 factor that appears in the equivalent u and ad terms of equation (11) is replaced with . Explicit formulae for , and are provided in equations (A37)–(A39) of the appendix. Collection efficiencies for negatively polarized cylindrical chambers are also described by equation (23).
The difference between and Boag’s (1950) formula for collection efficiency is due to the h factor in equation (23). Values of h calculated for the cylindrical chamber geometry and the particle transport and recombination coefficients of table 1 are plotted in figure 2 for doses-per-pulse of 1, 10 and 100 mGy and voltages of 100–1000 V. The electron attachment rate-constant, γ, was set to the level in the equivalent parallel plate chamber with the same voltage applied. The h values are ⩾1, larger for higher doses-per-pulse, and initially rise with increasing voltage before falling again. The initial rise is due to decreasing with increasing electric field strength. If is set to a constant value, h decreases monotonically with voltage.
Figure 2.

Values of the h factor of equation (23) plotted for doses-per-pulse of 1, 10 and 100 mGy and voltages of 100–1000 V. The h values are calculated for the cylindrical chamber with inner and outer electrode radii of 0.5 and 2.333 mm, using the transport and recombination coefficients of table 1. γ was set to the level in the equivalent parallel plate chamber with the same voltage applied.
Approximate collection efficiency formulae , and derived for positively polarized cylindrical chambers have the same forms as , and in equations (7)–(9), but again with replacing d2 in the u and p terms. The same formulae are obtained for negatively polarized chambers.
3.2. Validation of excluding E-field screening and variation
Tables 2 and 3 list collection efficiencies calculated from the formulae and computed from the numerical code for parallel plate and cylindrical chambers. Efficiencies computed without accounting for E-field screening or γ(E) variation were the same for positively and negatively polarized cylindrical chambers. They agreed with values obtained from with a maximum discrepancy of 0.0001 absolute, validating this formula under the conditions for which it was derived. They also agreed with collection efficiencies computed for parallel plate chambers without accounting for E-field screening or γ(E) variation, as expected since the chambers’ dimensions meet the equivalence condition of equation (1).
Table 2.
Collection efficiencies (CE) calculated for a parallel plate chamber with 2 mm electrode separation, irradiated with doses-per-pulse (DPP) of 0.1–100 mGy. Calculations were made for chamber potential differences of 200 and 400 V, using collection efficiency formulae and a numerical code that included charge screening of the E-field or excluded it (‘idealized’). Relative differences (%) between efficiencies calculated using each method and those from the code with E-field screening are shown in parentheses.
| DPP (mGy) | Numerical with E screening | Numerical idealized | |||||
|---|---|---|---|---|---|---|---|
| Parallel plate CE at 200 V | |||||||
| 0.1 | 0.9994 | 0.9994 | 0.9994 | 0.9993 | 0.9993 | 0.9994 | 0.9993 |
| (0%) | (0%) | (−0.01%) | (−0.01%) | (0%) | (−0.01%) | ||
| 1 | 0.9938 | 0.9939 | 0.9939 | 0.9928 | 0.9934 | 0.9939 | 0.9936 |
| (+0.01%) | (+0.01%) | (−0.10%) | (−0.04%) | (+0.01%) | (−0.02%) | ||
| 10 | 0.9412 | 0.9432 | 0.9432 | 0.9336 | 0.9388 | 0.9436 | 0.9412 |
| (+0.21%) | (+0.21%) | (−0.81%) | (−0.25%) | (+0.25%) | (0%) | ||
| 100 | 0.6237 | 0.6627 | 0.6627 | 0.6173 | 0.6419 | 0.6647 | 0.6534 |
| (+6.25%) | (+6.25%) | (−1.03%) | (+2.92%) | (+6.57%) | (+4.76%) | ||
| Parallel plate CE at 400 V | |||||||
| 0.1 | 0.9998 | 0.9998 | 0.9998 | 0.9996 | 0.9997 | 0.9998 | 0.9998 |
| (0%) | (0%) | (−0.02%) | (−0.01%) | (0%) | (0%) | ||
| 1 | 0.9980 | 0.9980 | 0.9980 | 0.9964 | 0.9975 | 0.9983 | 0.9979 |
| (0%) | (0%) | (−0.16%) | (−0.05%) | (+0.03%) | (−0.01%) | ||
| 10 | 0.9802 | 0.9808 | 0.9808 | 0.9653 | 0.9760 | 0.9834 | 0.9801 |
| (+0.06%) | (+0.06%) | (−1.52%) | (−0.43%) | (+0.33%) | (−0.01%) | ||
| 100 | 0.8168 | 0.8527 | 0.8526 | 0.7514 | 0.8212 | 0.8706 | 0.8475 |
| (+4.40%) | (+4.38%) | (−8.01%) | (+0.54%) | (+6.59%) | (+3.76%) | ||
Table 3.
Collection efficiencies (CE) calculated for a cylindrical chamber with 0.5 and 2.333 mm inner and outer electrode radii, irradiated with 0.1–100 mGy doses-per-pulse (DPP). Calculations were made for chamber potential differences of 200 and 400 V, using collection efficiency formulae and a numerical code that included E-field screening and/or γ(E) variation or neither (‘idealized’). Relative differences (%) between efficiencies calculated using each method and those from the code with E-field screening and γ(E) variation are shown in parentheses.
| A. Positive chamber polarity. | ||||||||||||||||||
| DPP (mGy) | Numerical with E screening & γ(E) | Numerical with γ(E) | Numerical idealized | |||||||||||||||
| Cylindrical CE at +200 V | ||||||||||||||||||
| 0.1 | 0.9994 | 0.9994 | 0.9994 | 0.9994 | 0.9993 | 0.9993 | 0.9994 | 0.9993 | ||||||||||
| (0%) | (0%) | (0%) | (−0.01%) | (−0.01%) | (0%) | (−0.01%) | ||||||||||||
| 1 | 0.9937 | 0.9938 | 0.9939 | 0.9939 | 0.9928 | 0.9934 | 0.9939 | 0.9936 | ||||||||||
| (+0.01%) | (+0.02%) | (+0.02%) | (−0.09%) | (−0.03%) | (+0.02%) | (−0.01%) | ||||||||||||
| 10 | 0.9380 | 0.9423 | 0.9432 | 0.9432 | 0.9336 | 0.9388 | 0.9436 | 0.9412 | ||||||||||
| (+0.46%) | (+0.55%) | (+0.55%) | (−0.47%) | (+0.09%) | (+0.60%) | (+0.34%) | ||||||||||||
| 100 | 0.5947 | 0.6585 | 0.6627 | 0.6627 | 0.6173 | 0.6419 | 0.6647 | 0.6534 | ||||||||||
| (+10.72%) | (+11.43%) | (+11.43%) | (+3.80%) | (+7.94%) | (+11.77%) | (+9.87%) | ||||||||||||
| Cylindrical CE at +400 V | ||||||||||||||||||
| 0.1 | 0.9998 | 0.9998 | 0.9998 | 0.9998 | 0.9996 | 0.9997 | 0.9998 | 0.9998 | ||||||||||
| (0%) | (0%) | (0%) | (−0.02%) | (−0.01%) | (0%) | (0%) | ||||||||||||
| 1 | 0.9977 | 0.9978 | 0.9980 | 0.9980 | 0.9964 | 0.9975 | 0.9983 | 0.9979 | ||||||||||
| (+0.01%) | (0.03%) | (0.03%) | (−0.13%) | (−0.02%) | (+0.06%) | (+0.02%) | ||||||||||||
| 10 | 0.9783 | 0.9786 | 0.9808 | 0.9808 | 0.9653 | 0.9760 | 0.9834 | 0.9801 | ||||||||||
| (+0.03%) | (+0.26%) | (+0.26%) | (−1.33%) | (−0.24%) | (+0.52%) | (+0.18%) | ||||||||||||
| 100 | 0.8226 | 0.8376 | 0.8527 | 0.8526 | 0.7514 | 0.8212 | 0.8706 | 0.8475 | ||||||||||
| (+1.82%) | (+3.66%) | (+3.65%) | (−8.66%) | (−0.17%) | (+5.84%) | (+3.03%) | ||||||||||||
| B. Negative chamber polarity. | ||||||||||||||||||
| DPP (mGy) | Numerical with E screening and γ(E) | Numerical with γ(E) | Numerical idealized | |||||||||||||||
| Cylindrical CE at −200 V | ||||||||||||||||||
| 0.1 | 0.9994 | 0.9994 | 0.9994 | 0.9994 | 0.9993 | 0.9993 | 0.9994 | 0.9993 | ||||||||||
| (0%) | (0%) | (0%) | (−0.01%) | (−0.01%) | (0%) | (−0.01%) | ||||||||||||
| 1 | 0.9937 | 0.9938 | 0.9939 | 0.9939 | 0.9928 | 0.9934 | 0.9939 | 0.9936 | ||||||||||
| (+0.01%) | (+0.02%) | (+0.02%) | (−0.09%) | (−0.03%) | (+0.02%) | (−0.01%) | ||||||||||||
| 10 | 0.9324 | 0.9423 | 0.9432 | 0.9432 | 0.9336 | 0.9388 | 0.9436 | 0.9412 | ||||||||||
| (+1.06%) | (+1.16%) | (+1.16%) | (+0.13%) | (+0.69%) | (+1.20%) | (+0.94%) | ||||||||||||
| 100 | 0.5215 | 0.6584 | 0.6627 | 0.6627 | 0.6173 | 0.6419 | 0.6647 | 0.6534 | ||||||||||
| (+26.25%) | (+27.07%) | (+27.07%) | (+18.37%) | (+23.09%) | (+27.46%) | (+25.29%) | ||||||||||||
| Cylindrical CE at −400 V | ||||||||||||||||||
| 0.1 | 0.9998 | 0.9998 | 0.9998 | 0.9998 | 0.9996 | 0.9997 | 0.9998 | 0.9998 | ||||||||||
| (0%) | (0%) | (0%) | (−0.02%) | (−0.01%) | (0%) | (0%) | ||||||||||||
| 1 | 0.9977 | 0.9978 | 0.9980 | 0.9980 | 0.9964 | 0.9975 | 0.9983 | 0.9979 | ||||||||||
| (+0.01%) | (0.03%) | (0.03%) | (−0.13%) | (−0.02%) | (+0.06%) | (+0.02%) | ||||||||||||
| 10 | 0.9747 | 0.9786 | 0.9808 | 0.9808 | 0.9653 | 0.9760 | 0.9834 | 0.9801 | ||||||||||
| (+0.40%) | (+0.63%) | (+0.63%) | (−0.96%) | (+0.13%) | (+0.89%) | (+0.55%) | ||||||||||||
| 100 | 0.6889 | 0.8371 | 0.8527 | 0.8526 | 0.7514 | 0.8212 | 0.8706 | 0.8475 | ||||||||||
| (+21.51%) | (+23.78%) | (+23.76%) | (+9.07%) | (+19.20%) | (+26.38%) | (+23.02%) | ||||||||||||
3.3. Effect of γ(E) variation within the cylindrical chamber
Cylindrical chamber collection efficiencies obtained from the numerical code after accounting for γ(E) variation but not E-field screening were slightly lower than those excluding γ(E) variation and E screening (table 3). Differences were greatest at 100 mGy per pulse and 400 V, where efficiencies of 0.8371 and 0.8523 were computed for a negatively polarized chamber with and without γ(E) variation, a relative difference of 1.8%. At doses-per-pulse ⩽10 mGy, the maximum difference was 0.2%. Computed collection efficiencies accounting for γ(E) variation differed marginally between positive and negative chamber polarities, but the maximum difference was only 0.06% relative at 100 mGy per pulse and 400 V chamber voltage.
3.4. Effect of E-field screening
Tables 2 and 3 also list collection efficiencies computed using the code for parallel plate chambers after accounting for E-field screening, and for cylindrical chambers after accounting for both E screening and γ(E) variation. In figure 3 these collection efficiencies are plotted against idealized efficiencies computed without screening or γ(E) variation.
Figure 3.

Collection efficiencies (CE) calculated for the parallel plate and cylindrical chambers after accounting for E-field screening and γ(E) variation, plotted on log and linear scales against efficiencies calculated without turning on E screening or γ(E) variation. The line of equality is shown. The lowest collection efficiencies, with the greatest differences between chambers, were calculated for 100 mGy per pulse at chamber potential differences of 200 and 400 V, and for 10 mGy per pulse at 200 V, as indicated in the linear plot.
Screening of the E-field reduced collection efficiencies. At 100 mGy per pulse and 200 V, the efficiency computed for the parallel plate chamber after allowing for screening was 5.9% relative lower than the value without screening, and efficiencies calculated for the positively and negatively polarized cylindrical chamber including screening and γ(E) variation were respectively 10.3% and 21.3% lower than efficiencies calculated without these factors. For doses-per-pulse ⩽10 mGy the effect was much smaller, with a maximum reduction in collection efficiency due to screening of 1.1% relative at 200 V.
3.5. The equivalence condition
In Supplementary figure 1, collection efficiencies computed for the cylindrical chamber including E-field screening and γ(E) variation are plotted against efficiencies computed for the parallel plate chamber including screening. At doses-per-pulse ⩽1 mGy, almost equal collection efficiencies were calculated for the two chambers. At 10 mGy per pulse, E-field screening and γ(E) variation still had limited impact on the equivalence of the two chambers. For 200 V chamber voltage, collection efficiencies of 0.9412, 0.9380 and 0.9324 respectively were calculated for the parallel plate chamber and the cylindrical chamber with positive and negative polarity, a maximum relative difference between chambers of 0.9%. For 400 V, the chambers’ collection efficiencies differed by a maximum of 0.6%.
At 100 mGy per pulse, the equivalence between the two chambers broke down after accounting for E-field screening and γ(E) variation. At 200 V the collection efficiencies calculated for the parallel plate chamber and the cylindrical chamber with positive and negative polarity were 0.6237, 0.5947 and 0.5215 respectively, a maximum relative difference between chambers of 19.6%. At 400 V the corresponding efficiencies were 0.8168, 0.8226 and 0.6889, a maximum difference of 18.6%. Since γ(E) variation alone made only 1.8% difference to the collection efficiency of the cylindrical chamber at 400 V and 100 mGy per pulse (table 3), the breakdown in equivalence is largely due to E-field screening.
3.6. Which formula describes the computed collection efficiencies best?
Here, collection efficiencies computed using the code are compared with values obtained from the various formulae with parameters determined directly by chamber geometries and the charge transport and recombination coefficients of table 1. Since these coefficients were also used in the code, this gauges how consistent the analytical formulae are with the underlying physics of the computational model, despite the simplifications made in the analytical derivations.
Collection efficiencies computed using the code with γ(E) variation and E-field screening turned off were described exactly by and . These efficiencies were overestimated by ⩽2.1% relative by the approximate and formulae, and underestimated by ⩽1.4% by and , and ⩽3.7% by and . The formula, which excludes free electron effects, underestimated the efficiencies by up to 11.9% (tables 2 and 3).
Collection efficiencies computed for the cylindrical chamber accounting for γ(E) variation but not E screening were overestimated by ⩽1.8% by . In fact, described these efficiencies better than any other formula up to 1 mGy per pulse, and better than , and at all doses-per-pulse. At 10 and 100 mGy per pulse, however, described the efficiencies best, exceeding them by ⩽1.2% (table 3).
Collection efficiencies computed for parallel plate and cylindrical chambers accounting for both E-field screening and γ(E) variation were still generally described best by and for doses-per-pulse up to 1 mGy. At 10 mGy per pulse, efficiencies calculated for the different chambers, voltages and polarities were described best by an assortment of formulae including , and . At 100 mGy per pulse, E-field screening was substantial and the collection efficiencies were described best by the formula of Boag (1950). This formula did not describe the computed efficiencies particularly well, though, differing from them by up to 8.0% and 18.4% for the parallel plate and cylindrical chambers respectively (tables 2 and 3).
3.7. Multi-voltage estimation of collection efficiencies
Supplementary figure 2 shows five sets of collection efficiencies computed using the code, plotted against chamber voltages. One set was calculated for the parallel plate chamber including free electron effects and E-field screening (‘Parallel plate’). Another two were calculated for positively and negatively polarized cylindrical chambers accounting for free electron effects, E-field screening and γ(E) variation (‘Cylindrical ± pol’). Another set excluded E screening and γ(E) variation (‘Idealized ’), and a further set additionally excluded free electron effects (‘Boag ’). Results for doses-per-pulse of 1, 10 and 100 mGy are graphed separately.
At 1 mGy per pulse the calculated sets of collection efficiencies were all similar except for the ‘Boag ’ set, which was lower than the rest because it did not include direct collection of free electrons by the chamber. At 100 mGy per pulse the sets of calculated efficiencies differed far more, driven by greater E-field screening at higher doses-per-pulse. For a chamber potential difference of 100 V the screening substantially reduced the E-field, leading to slower ion drift speeds and greater recombination. Consequently, all three sets of collection efficiencies that accounted for screening were lower than the set that corresponds to Boag’s formula, despite this latter set excluding free electron effects. At 10 mGy per pulse the data followed a pattern intermediate between 1 and 100 mGy per pulse.
The same data are presented in figure 4 as Jaffé plots of reciprocals of collection efficiencies versus reciprocals of V. For the Boag set, the plots run roughly linearly through a collection efficiency of unity at (1/V) = 0. This result is well known and lies behind a simple experimental method for estimating the factor in multi-voltage determination of chamber collection efficiencies, by taking readings at two or more voltages and linearly extrapolating them to (1/V) = 0 in a Jaffé plot (Khan 1994, Almond et al 1999).
Figure 4.

(a)–(c) Jaffé plots showing reciprocals of computed collection efficiencies (CE) versus reciprocals of chamber voltages V for doses-per-pulse of 1, 10 and 100 mGy. Sets of efficiencies (‘Parallel plate’, ‘Cylindrical ± pol’) were computed for the parallel plate chamber including free electron effects and E-field screening, and for cylindrical chambers including these effects and γ(E) variation. Another set (‘Idealized f exp’) excluded E screening effects and γ(E) variation, and a further set (‘Boag ’) additionally excluded free electron effects. (d) Reciprocals of ‘Parallel plate’ and ‘Boag ’ collection efficiencies computed for 10 mGy per pulse are plotted alongside curves obtained from reciprocals of the f(V) and fexp (V) formulae with parameter values derived from table 1. Reciprocals of fits of scaled versions of these formulae and the logistic model to the whole 100–1000 V ‘Parallel plate’ dataset are also shown. For brevity, this dataset is referred to in panel (d) as ‘Ppl’.
Collection efficiencies that include free electron effects lie along more curved trajectories in Jaffé plots (figure 4). Irrespective of whether these efficiencies account for E-field screening or γ(E) variation, they approach unity at chamber voltages of ∼1000 V, which generate E-fields sufficiently high that large fractions of free electrons are collected directly. As a result, would be overestimated by linearly extrapolating reciprocals of chamber readings to (1/V) = 0, causing collection efficiencies at practical voltages to be underestimated.
Figure 4(d) and supplementary figure 2(d) show fits of scaled nonlinear collection efficiency formulae () to collection efficiencies that include free electron effects calculated for 10 mGy per pulse. The fit of the scaled version of Boag’s formula to collection efficiencies calculated for the parallel plate chamber after accounting for E screening and free electrons overestimated the unit collection efficiency at (1/V) = 0 by around a factor of 1.02. This is equal to the fitted factor which, as outlined in the materials and methods , describes the relative inaccuracy of the multi-voltage method for determining collection efficiencies. Linear extrapolation of reciprocals of collection efficiencies back to (1/V) = 0 overestimated the collection efficiency there to the same extent, since curves are roughly linear in Jaffé plots.
The fit of the scaled formula, which accounts for free electron effects, predicts a collection efficiency much closer to one at (1/V) = 0. Screening of the E-field had only a limited effect at 10 mGy per pulse, and consequently the plotted collection efficiencies were described quite well by despite this formula not accounting for screening. This was even the case when the u and ad parameters in the formula were calculated directly from the values listed in table 1 and with set to one, giving the green curves in figure 4(d) and supplementary figure 2(d). When the u and ad parameters were fitted along with , giving the black dashed curves, the description of collection efficiencies was improved at 100 V where screening has most effect, but slightly worsened at higher voltages.
Figure 5 shows the values obtained by fitting sets of computed collection efficiencies using scaled , , , and formulae, their cylindrical equivalents and . The sets of efficiencies fitted were computed for the parallel plate chamber and the cylindrical chamber with positive and negative polarities, irradiated with 10 and 100 mGy per pulse, including or excluding E screening effects. For the cylindrical chamber all the computed efficiencies accounted for γ(E) variation. The values obtained are further summarized in table 4.
Figure 5.

Swarm plots of values obtained from fits of scaled collection efficiency formulae to efficiencies calculated using the code. The collection efficiencies fitted were calculated with and without E-field screening effects (‘Scr’ or ‘No Scr’) for 10 mGy and 100 mGy per pulse delivered to the parallel plate chamber and the cylindrical chamber with positive and negative polarities (‘+’ or’-‘). Formulae were fitted to whole sets of efficiencies calculated for voltages V of 100–1000 V and to subsets calculated for V ⩽400 V (‘All V’ or ‘⩽ 400 V’).
Table 4.
Summary of the values plotted in figure 5, obtained by fitting scaled collection efficiency formulae to efficiencies computed for parallel plate and cylindrical chambers and doses-per-pulse of 10 and 100 mGy. The fitted efficiencies were computed with and without allowing for the effect of E-field screening. Formulae were fitted to all collected efficiencies computed for voltages of 100–1000 V, and to subsets computed for 100–400 V.
| Formula | Mean | Median | range | a |
|---|---|---|---|---|
| 10 mGy pulse, all fits to 100–1000 V datasets and 100–400 V subsets | ||||
| or | 0.9983 | 0.9994 | 0.9884, 1.0067 | 0.0036 |
or
|
1.0078 | 1.0039 | 0.9985, 1.0260 | 0.0129 |
| or | 0.9922 | 0.9941 | 0.9819, 0.9981 | 0.0098 |
| or | 1.0037 | 0.9987 | 0.9884, 1.0260 | 0.0108 |
| 1.0204 | 1.0185 | 1.0150, 1.0331 | 0.0211 | |
| 1.0194 | 1.0018 | 1.0000, 1.0354 | 0.0219 | |
| 100 mGy pulse, all fits to 100–1000 V datasets and 100–400 V subsets | ||||
| or | 0.9814 | 0.9990 | 0.8237, 1.0423 | 0.0594 |
| or | 1.1022 | 1.0213 | 0.9963, 1.5304 | 0.1886 |
| or | 0.9396 | 0.9606 | 0.7669, 1.0323 | 0.0922 |
| or | 0.9489 | 0.9675 | 0.7792, 1.0352 | 0.0836 |
| 1.2299 | 1.1845 | 1.1004, 1.5495 | 0.2187 | |
| 1.4578 | 1.1723 | 0.9407, 2.5335 | 0.6572 | |
| 10 mGy pulse, fits to 100–400 V data subsets | ||||
| or | 0.9983 | 0.9983 | 0.9884, 1.0067 | 0.0062 |
| or | 1.0166 | 1.0160 | 1.0080, 1.0260 | 0.0183 |
| or | 0.9871 | 0.9888 | 0.9819, 0.9922 | 0.0135 |
| or | 1.0090 | 1.0080 | 0.9884, 1.0260 | 0.0152 |
| 1.0222 | 1.0188 | 1.0158, 1.0331 | 0.0231 | |
| 1.0205 | 1.0220 | 1.0000, 1.0328 | 0.0229 | |
| 100 mGy pulse, fits to 100–400 V data subsets | ||||
| or | 0.9540 | 0.9757 | 0.8237, 1.0133 | 0.0819 |
| or | 1.1223 | 1.1583 | 1.0020, 1.5304 | 0.1388 |
| or | 0.8888 | 0.9165 | 0.7669, 0.9430 | 0.1313 |
| or | 0.9015 | 0.9287 | 0.7792, 0.9509 | 0.1166 |
| 1.2771 | 1.2554 | 1.1004, 1.5495 | 0.3147 | |
| 1.7599 | 1.8182 | 0.9407, 2.5335 | 0.9162 | |
Root mean square difference between individual values and one.
Typical values from fits of the scaled Boag (1950) formula were 1.02 and 1.2 at 10 and 100 mGy per pulse respectively. Values differed little between parallel plate and cylindrical chambers, and were not consistently related to whether the fitted data spanned the whole 100–1000 V range or were limited to chamber voltages ⩽400 V. For fits based on , values up to 1.035 and 2.534 were obtained at 10 and 100 mGy per pulse.
values closer to one were obtained when , , , , and their cylindrical analogues were used in the fitting. For the sets of collection efficiencies studied, and performed best, achieving mean values of 0.9983 and 0.9540 at doses-per-pulse of 10 and 100 mGy respectively when the data subsets calculated for voltages ⩽400 V were fitted. Furthermore, root mean square (RMS) differences between unity and the values obtained from fits to the various sets of collection efficiencies were lower for and than for the other formulae.
At 10 mGy per pulse, values for fits of the scaled and formulae to the various sets of collection efficiencies ranged from 0.9884 to 1.0067, indicating that errors in collection efficiencies determined using the multi-voltage method should be ⩽1.2%. At 100 mGy per pulse, however, values for fits of these formulae ranged from 0.8237 to 1.0423, making the approach imprecise.
4. Discussion
We have derived a formula for the collection efficiency of cylindrical ionization chambers by exactly solving a volume recombination model of Boag et al (1996) that includes free electron effects. This formula and another, , similarly derived for parallel chambers have been validated using a numerical code that describes charge transport and recombination. They predict equal collection efficiencies for cylindrical and parallel plate chambers that satisfy the equivalence condition of equation (1), which was originally identified by Boag (1950) for a recombination model that did not include free electrons.
E-field screening was not included in the recombination model of Boag et al (1996), nor in the derivation of or , and we therefore used the numerical code to determine by how much collection efficiencies changed when screeing was accounted for. We also calculated collection efficiency changes when the variation of the electron attachment rate-constant with the radially varying E-field within a cylindrical ionization chamber was factored into the code.
At doses-per-pulse of 0.1 and 1 mGy, E-field screening and variation made very little difference to calculated collection efficiencies (⩽0.03%) which were therefore still described well by and . Thus, at these doses-per-pulse the equivalence condition continued to hold well. At 10 mGy per pulse, E screening and variation had greater but still limited impact. Collection efficiencies were reduced by ⩽1.1% relative, with efficiencies of the equivalent parallel plate and cylindrical chambers differing by a maximum of 0.9%. At 100 mGy per pulse, E screening had much more effect and the chamber equivalence broke down. For 200 V potential difference, collection efficiencies computed for the parallel plate chamber and the cylindrical chamber with positive and negative polarity fell by 5.9%, 10.3% and 21.3% respectively when E screening and variation were included in calculations for a 200 V chamber potential difference.
The notable difference between the collection efficiencies calculated for the two polarities of the cylindrical chamber at 100 mGy per pulse mirrors the substantially non-unit polarity correction factors of 1.20 and 1.13 reported by Darafsheh et al (2021) for IBA CC04 and CC13 cylindrical chambers irradiated at 168 mGy per pulse in a synchro-cyclotron proton beam. The higher efficiency calculated for positive polarity is consistent with results presented by Lapsley (1953) for an argon-filled cylindrical chamber with an outer electrode built from uranium. When Lapsley’s chamber was bombarded with neutrons causing fission fragments to enter the cavity, the chamber voltage needed to achieve a fixed collection efficiency was lower for positive than negative chamber polarity, 120 versus 300 V.
Scaled versions of the collection efficiency formulae were fitted to sets of efficiencies calculated for the parallel plate and cylindrical chambers, including and excluding E field screening and variation. Fitted values of CE∞ predicted by the various fits describe the relative underestimation of collection efficiencies determined using the multi-voltage method with these formulae.
At 10 mGy and 100 mGy per pulse, fits of the scaled version of Boag’s (1950) formula gave CE∞ values of 1.015–1.033 and 1.100–1.550 respectively. Thus, collection efficiencies determined experimentally using the multi-voltage method with these formulae would be underestimated by 1.5%–3.2% and 9.1%–35.5% respectively at these doses-per-pulse.
Values of CE∞ obtained from fits of scaled formulae derived from the model of Boag et al (1996) which includes free electron effects were closer to one. For the sets of collection efficiencies studied, fits of and gave CE∞ values with the lowest RMS differences from unity. At 10 mGy per pulse, the CE∞ values of and fits to efficiencies calculated for voltages of 100–400 V ranged from 0.9884–1.0067. This includes CE∞ values from fits to collection efficiencies computed with E-field screening and variation turned on, effects not included in the derivations of the collection efficiency formulae. According to these results, then, collection efficiencies at 10 mGy per pulse can be determined within an error of 1.2% by analyzing detector readings made at chamber voltages of 100–400 V using the multi-voltage approach with the and formulae.
At 100 mGy per pulse, fits of scaled and formulae also gave CE∞ values with the lowest RMS differences from one. However, CE∞ values for the fits ranged from 0.824 to 1.013, indicating errors of up to 21.4% in collection efficiencies obtained using multi-voltage methods with these formulae. Use of the other formulae investigated would lead to larger errors.
CE∞ values from fits of the scaled empirical formula ranged from 1.000 to 1.035 at 10 mGy per pulse, and from 0.941 to 2.531 at 100 mGy. This formula has provided very good fits of ionization chamber measurements made at doses-per-pulse of 10–104 mGy and chamber voltages of 50–300 V (Petersson et al 2017). However, its flexible nature might lead to considerable uncertainties in extrapolations of fits to very high voltages.
Unlike ionization chamber readings, the computational collection efficiency data we fitted did not include random measurement uncertainties which could affect the CE∞ values obtained. However, uncertainties in ionization chamber readings can be held at low levels, around 0.1% (DeWerd and Smith 2021). And much of the fitted data were calculated after accounting for the effects of E-field screening and variation. These effects introduce larger differences into the data, up to 23% at 100 mGy per pulse, and since they were not factored into the derivations of the fitted formulae they will have a greater impact on CE∞ values.
In practice, ionization chamber collection efficiencies reach values greater than unity at very high electric field strengths, a result of charge amplification caused by free electrons moving at high speeds. In Pinpoint cylindrical chambers with electrode gaps of 0.7–1.15 mm (PTW 31 022 and 31 023) collection efficiencies begin to rise due to charge amplification at voltages of 250–400 V (Bruggmoser et al 2018a, 2018b). Similarly, in an experimental parallel plate detector with a 0.25 mm electrode gap, efficiencies start to rise at 375 V (Kranzer et al 2022). However, the increases in this voltage range were small, ∼0.3%, and did not depend on dose-per-pulse. Charge amplification can be included in numerical calculations of collection efficiency (Kranzer et al 2022, Paz-Martín et al 2022) but this is not of interest for our studies. The aim of fitting a recombination model to chamber readings and extrapolating to is to determine the true reading in the absence of both recombination and amplification. Therefore, the ideal model for this purpose should exclude charge amplification and describe recombination rates that fall to zero at high voltages.
In our ion transport code, we assumed that radiation pulses were instantaneous. In reality, pulses have finite durations during which the free electrons and positive ions generated drift to some extent, and this potentially modifies charge distributions and collection efficiencies. However, pulses generated by conventional medical accelerators have durations ⩽5 μs, and differences between efficiencies calculated for instantaneous and 5 μs long pulses were <3% at a high dose-per-pulse of 500 mGy (Paz-Martín et al 2022). Diffusion of charge carriers was also ignored in the code and in the analytic solutions. But even if formally ignored, the numerical solution of the equations introduces a pseudo-diffusion that can be close to the physical charge carrier diffusion (Paz-Martin et al 2022).
Lastly, it should be noted that in this study the parallel plate and cylindrical ionization chambers were idealized as having electrodes with infinite diameters and lengths respectively. Therefore calculated collection efficiencies do not include the effect of any electric field distortion at the edge of the electrode and guard-ring (Kranzer et al 2021).
5. Conclusions
We derived a collection efficiency formula for cylindrical ionization chambers by exactly solving an extended volume recombination model of Boag et al (1996) that includes free electron effects. According to this formula and another derived for parallel chambers, , the two types of ionization chambers should have the same collection efficiencies if they meet the condition .
The formula and the equivalence condition were validated using a numerical code describing ion transport and recombination. The derivations of and did not include E-field screening or variation. When these effects were factored into the code, calculated collection efficiencies fell by up to 0.03%, 1.1% and 21.3% at doses-per-pulse of 1, 10 and 100 mGy per pulse. The efficiencies calculated at 100 mGy per pulse differed by up to 19.6% relative between parallel plate and cylindrical chambers that met the equivalence condition.
Differences from unity of CE∞ values obtained from fits of scaled collection efficiency formulae to efficiencies calculated for 100–400 V chamber voltages correspond to the relative inaccuracies of collection efficiencies determined experimentally by fitting the scaled efficiency formulae to chamber readings made across this voltage range. For the data analyzed, CE∞ values closest to one were obtained from fits of the scaled and formulae, being ⩽1.2% and ⩽17.6% from unity at 10 and 100 mGy per pulse respectively. Thus, for the chamber geometries studied, collection efficiencies and corresponding correction factors should be determinable to within 1.2% accuracy up to 10 mGy per pulse using the multi-voltage approach, but with poor accuracy at 100 mGy per pulse.
Acknowledgments
J D F is supported by the Radiation Research Unit at the Cancer Research UK City of London Centre Award [C7893/A28990]. J P-M acknowledges support from Xunta de Galicia, Axencia Galega de Innovación (Grant IN607D 2022/02).
Appendix.
Key symbols used in the derivation of the collection efficiency formula are listed below. Numbers of particles are defined per length along the cylindrical chamber axis. Positive and negative ions are abbreviated above as +ve and -ve.
,, original radial locations of electrons and ions respectively
, , current radial locations of electrons and ions
, original radial location of ions that at
cross ions from with the opposite sign
, time at which an electron originating at re reaches r
, time at which +ve and -ve ions originating at and cross as they drift
t no overlap, time when the lagging edges of +ve and -ve ion clouds cross
, density at time t of electrons originating at
, density at time t of +ve ions or -ve ions originating at r
, density at time t of +ve ions or -ve ions currently located at r′
, density at t of ions generated at r multiplied by circumference
, initial total number of electrons in a cylindrical chamber
, initial total number of -ve ions in the chamber
, total number of ve ions initially located between r and the anode
, total number of +ve ions through which -ve ions generated at r pass before reaching the anode
, difference between and
, number of the -ve ions that recombine before reaching the anode
, number of -ve ions collected that originate between rin and r
, total number of -ve ions collected
, number of +ve ions collected that pass through -ve ions
, total number of +ve ions collected
, combination of parameters
, combination of parameters
A1. Free electron transport in the cylindrical geometry
Immediately after a radiation pulse, free electrons are distributed uniformly throughout a cylindrical chamber with density n 0 linearly related to the dose-per-pulse. In chambers with positive polarity the electrons drift inwards towards the anode (figure 1) and at off axis they are moving with average speed
Electrons generated a radial distance re from the chamber axis therefore reach the circle of radius centred on the chamber axis at a time related to re via
In the absence of other processes, the density of the electrons originating at re would not vary with radius as they drift, because although the electrons converge circumferentially as their distance off-axis decreases, they diverge radially to the same extent due to the varying electric field (Boag 1950). However, over time electrons become attached to oxygen with a rate-constant γ. Consequently, on arrival at the density of free electrons that originate at re is less than the initial density n 0 by a factor
where . Throughout the appendix the form denotes the density at time t of particles generated originally at r, whereas n(r’,t) denotes the density at t of particles currently at r′;. Thus nr
(t) and n(r’,t) generally differ but are equal when t =0 and
Negative ions are formed at a rate given by the local free electron density multiplied by γ, so by time t the negative ion density at r off-axis is
![]() |
From (A2) it follows that
and so
in which r max(t) is the radius of origin of the electrons that cross r at time t. Integration up to the time when electrons originating at the radius r out of the outer electrode arrive at r gives
which is presented in slightly different form in equation (19) of the results . Since no electrons reach r from beyond the outer electrode, equation (A6) describes the density distribution of negative ions at the completion of electron transport and before positive or negative ions have moved substantially from their points of origin. This distribution has a Gaussian dependence on r. At this early time the total number of negative ions in the chamber per length along its axis is
In equation (A7) N 0 is the initial number of free electrons in the chamber, equal to . Again, N 0 is defined per length along the axis, as particle numbers are throughout the appendix. The initial number of negative ions, , is equal to the number of uncollected free electrons from which they were formed, given by where p is the fraction of electrons collected. It therefore follows that
Repeating this analysis for free electrons travelling outwards in a chamber with negative polarity leads to the negative ion density distribution
with p again given by equation (A8).
A2. Exact solution of the extended recombination model for a cylindrical geometry
Now reset t = 0 to the time when all the free electrons have just been collected or become attached to oxygen. Then the initial negative ion distribution (equivalently in a chamber with positive polarity is given by equation (A6) while the initial positive ion distribution is uniform with density .
By time later, inward-drifting negative ions initially formed at r have reached off-axis, where they cross outward drifting positive ions initially generated at (figure 1). Changes in density as particles drift through the chamber are due to recombination alone, and thus the change in negative ion density between and is
The transport of positive and negative ions is like that of electrons but with mobilities k 1 and k 2 respectively replacing ke . Consequently,
where the third equality follows by eliminating from the first two. By , the negative ions from have drifted inward to
and are crossing positive ions that have drifted out from , in which both and are negative. Then from (A11)
and (A10) can be written as
As they drift inward to the anode, the negative ions that originated at r cross positive ions that were generated between r and r in. Integration of (A13) between values of r and r in therefore gives an expression for the fraction of these negative ions that reach the anode
Now define as so that is the number of positive or negative ions originating between r and r +dr that remain unrecombined by time t. Also define as the total number of positive ions through which negative ions originating at r pass as they drift to the anode, so that
In equation (A15) is the reduction in between times 0 and , is the total number of positive ions that initially lie between r and the anode, and is the number of these that have recombined before negative ions originating at r pass through them. Then equation (A14) can be rewritten as
where is .
Next, define as the total number of negative ions that initially lie between the anode and r, equal to
, the integral of between r
in and r. And define as the number of these ions that recombine before reaching the anode, from which it follows that is equal to . Now is the number of negative ions that originate between r and r+ dr and recombine before reaching the anode, and is therefore given by
From (A15), (A17) and the equality between R+,r and R-,r
Denoting as the number of negative ions originating between r in and r that reach the anode, then
From (A18) and the differential of (A19) with respect to r an equation is obtained that links to quantities obtained directly from the initial ion distributions
Equation (A22) can be solved by writing
Integration of (A24) gives
and since (1/m(r)) equals , and and are respectively zero and the total number of negative ions reaching the anode, , then
Given the distribution of equation (A9) and that is uniform with density n 0, it follows that—
Substituting these formulae into (A21),
where
and
Integration of between r in and r out proceeds by changing variable to , for which . Thus,
where E1 is the exponential integral standard function
By changing variable from r to , the integral of between r in and r out is . Substituting these components of the integral of into (A26) gives
The collection efficiency for the Gaussianly-varying negative ion density of equation (A6) can now be obtained by adding to the number of free electrons collected, , and dividing by the total number of negative electrons initially generated, . The resulting formula is simplified by extracting a multiplicative factor of from the logarithm term in and writing it as a separate additive log term, and further simplified by substituting in the more underlying quantities that make up C1, C2 and C3 . Then works out as
For negative chamber polarity, the collection efficiency can be derived via equations (A10)–(A26) above except that now the positive ions are moving inwards. The result is an equation for the number of positive ions collected, , that has the same form as (A26), but with and swapping places in , and replacing . Solving this equation for and dividing by the initial number of positive ions gives the collection efficiency. Interestingly, it is the same as for positive chamber polarity, again given by equations (A36)–(A39).
A3. Approximate solution of the extended model for a cylindrical geometry
Consider a cylindrical chamber with positive polarity and the exact negative ion distribution of equation (A6) replaced by an approximate initial distribution
where p is the fraction of free electrons collected, described by (A8). The exact distribution is compared in figure A1 to the approximate distributions , and used in the derivation of , and . Densities were calculated for a cylindrical chamber with 0.5 and 2.333 mm inner and outer electrode radii, using the transport and recombination coefficients of table 1.
Figure A1.

The initial distribution of negative ion densities within a positively polarized cylindrical ionization chamber, , plotted against radial distance r from the chamber axis. Also plotted are the approximate distributions , and used to derive the approximate collection efficiency formulae , and . The densities were calculated for a chamber with 0.5 and 2.333 mm inner and outer electrode radii, using the electron and ion transport and recombination coefficients of table 1. They are plotted relative to the initial uniform positive ion density,.
Since and the initial distribution of positive ions are both uniform, rates of change of ion densities within the overlap region between negative and positive ions are also uniform, as are the ion densities and at later times. The rates of change are given by
Because these rates are equal the difference between and is constant, at a level pn0 since the initial density of positive ions is n0 . Therefore
which integrate to
By time t the lagging edge of negative ions formed at the outer edge of the chamber, r out, has drifted inward to a radius . At this distance off-axis the positive and negative ions are moving past each other with speeds and , and the number of positive ions crossing the lagging edge between t and t+ dt is therefore
All these positive ions go on to be collected since no further negative ions lie between them and the cathode. Thus, the total number of positive ions collected is
where t no overlap is the time at which the lagging edges of the positive and negative ion clouds formed at r out and r in cross and no overlap remains. Substituting (A43) into (A45)
Changing variable to and writing as T,
From (A11),
Substituting this into (A47) and dividing by the total initial number of positive ions gives the approximate collection efficiency formula
with u defined as in (A38). When the chamber polarity is negative the drift directions of the positive and negative ions are reversed, but the same set of equations and the same f′ formula are obtained.
A4. Approximate solution of the extended model
Now consider the cylindrical chamber with positive polarity and another approximation to the initial negative ion distribution
In the overlap region the densities of positive and negative ions are equal and their rates of change are
which integrate to
Following the method used to find , the same formula (A45) is obtained for in terms of . However, is now defined by (A52) and t no overlap is the time at which the lagging edges of the positive and negative ion clouds formed at and cross, given by
Therefore
where describes the number of positive ions that originate between and and pass through some of the negative ion cloud and are collected at the cathode. Further positive ions formed between and also reach the cathode but without passing through any of the negative ion cloud. Adding these to gives the total number of positive ions collected
Substituting t no overlap from (A53) and dividing by the total number of positive ions initially formed, , then gives
with u defined by (A38).
For negative chamber polarity, charge flow directions are reversed and the initial negative ion distribution is approximated as n 0 between and , and zero elsewhere, and now . When (A54) and (A55) are adjusted to account for the resulting change in t no overlap, and to reflect that now positive ions originating in the region to reach the cathode without passing through the negative ion cloud, the formula of (A56) is again obtained.
A5. Approximate solution of the extended model
Consider once more the cylindrical chamber with positive polarity, now with the exact negative ion distribution replaced by the approximate initial distribution
so that the initial number of negative ions is lower than that of positive ions by a factor of , and the initial density of the negative ions and volume in which they are located are both reduced by factors of compared to the positive ion distribution.
Equations (A41)–(A47) obtained in the derivation of continue to apply, except that the p factor appearing in them is replaced by λ, and t no overlap denotes the time at which the lagging edge of the positive ion distribution crosses the negative ion lagging edge which now originates at rather than r out. From (A11) and (A58)
Making these changes in (A47) gives
Adding to the positive ions that are generated between r out and and reach the cathode without passing through any negative ions, the total number of positive ions collected is
Dividing by the initial number of positive ions then gives the collection efficiency
For negative chamber polarity, the initial negative ion distribution is approximated as between r -,edge and r out, and zero elsewhere, and now
Equations (A41)–(A47) continue to apply, but with λ again replacing p and the lagging edge of the positive ion distribution originating at r out rather than r in. No change results in the expressions for t no overlap and in (A60) and (A61). The number of additional positive ions originating outside the initial negative ion cloud is also unchanged, and the collection efficiency is again described by the formula of (A63).
Data availability statement
All data that support the findings of this study are included within the article (and any supplementary information files).
Conflicts of interest
The authors have no conflicts of interest.
References
- Almond P R, Biggs P J, Coursey B M, Hanson W F, Huq M S, Nath R, Rogers D W O. AAPM’s TG-51 protocol for clinical reference dosimetry of high-energy photon and electron beams. Med. Phys. 1999;26:1847–70. doi: 10.1118/1.598691. [DOI] [PubMed] [Google Scholar]
- Boag J W. Ionization measurements at very high intensity. Br. J. Radiol. 1950;23:601–11. doi: 10.1259/0007-1285-23-274-601. [DOI] [PubMed] [Google Scholar]
- Boag J W, Currant J. Current collection and ionic recombination in small cylindrical ionization chambers exposed to pulsed radiation. Br. J. Radiol. 1980;53:471–8. doi: 10.1259/0007-1285-53-629-471. [DOI] [PubMed] [Google Scholar]
- Boag J W, Hochhäuser E, Balk O A. The effect of free-electron collection on the recombination correction to ionization measurements of pulsed radiation. Phys. Med. Biol. 1996;41:885–97. doi: 10.1088/0031-9155/41/5/005. [DOI] [PubMed] [Google Scholar]
- Boissonnat G. PhD Dissertation. 2015. Chambres d’ionisation en protonthérapie et hadronthérapie. [Google Scholar]
- Bruggmoser G, Saum R, Kranzer R. Determination of recombination and polarity correction factors, kS and kP, for small cylindrical ionization chambers PTW 31021 and PTW 31022 in pulsed filtered and unfiltered beams. Z. Med. Phys. 2018a;28:247–53. doi: 10.1016/j.zemedi.2017.09.010. [DOI] [PubMed] [Google Scholar]
- Bruggmoser G, Saum R, Kranzer R. Addendum to the determination of recombination and polarity correction factors, kS and kP, for small cylindrical ionization chambers PTW 31021 and PTW 31022 in pulsed filtered and unfiltered beams: PTW 31023. Z. Med. Phys. 2018b;28:247–53. doi: 10.1016/j.zemedi.2018.11.001. [DOI] [PubMed] [Google Scholar]
- Christensen J B, Tölli H, Bassler N. A general algorithm for calculation of recombination losses in ionization chambers exposed to ion beams. Med. Phys. 2016;43:5484–92. doi: 10.1118/1.4962483. [DOI] [PubMed] [Google Scholar]
- Darafsheh A, Hao Y, Zhao X, Zwart T, Wagner M, Evans T, Reynoso F, Zhao T. Spread-out Bragg peak proton FLASH irradiation using a clinical synchrocyclotron: proof of concept and ion chamber characterization. Med. Phys. 2021;48:4472–84. doi: 10.1002/mp.15021. [DOI] [PubMed] [Google Scholar]
- DeWerd L A, Smith B R. Ionization chamber dosimetry. In: Darafsheh A, editor. Radiation Therapy Dosimetry: A Practical Handbook. CRC Press; 2021. [Google Scholar]
- Di Martino F, Giannelli M, Traino A C, Lazzeri M. Ion recombination correction for very high dose-per-pulse high-energy electron beams. Med. Phys. 2005;32:2204–10. doi: 10.1118/1.1940167. [DOI] [PubMed] [Google Scholar]
- Fenwick J D, Kumar S. Collection efficiencies of ionization chambers in pulsed radiation beams: an exact solution of an ion recombination model including free electron effects. Phys. Med. Biol. 2023;68:015016. doi: 10.1088/1361-6560/aca74e. [DOI] [PubMed] [Google Scholar]
- Gotz M, Karsch L, Pawelke J. A new model for volume recombination in plane-parallel chambers in pulsed fields of high dose-per-pulse. Phys. Med. Biol. 2017;62:8634–54. doi: 10.1088/1361-6560/aa8985. [DOI] [PubMed] [Google Scholar]
- Gradshteyn I S, Rhyzik I M. In: Table of Integrals, Series and Products. 7th edn. Jeffrey A, Zwillinger D, editors. Academic; 2007. pp. 883–5. [Google Scholar]
- Hochhäuser E, Balk O A, Schneider H, Arnold W. Determination of electron swarm parameters from the radiation pulse response of an ionization chamber. J. Phys. D: Appl. Phys. 1994;27:2312–9. doi: 10.1088/0022-3727/27/11/011. [DOI] [Google Scholar]
- ICRU . ICRU Report 34. International Commission on Radiation Units and Measurement; 1982. The dosimetry of pulsed radiation. [Google Scholar]
- Iwafuchi Y, Oguchi H, Okudaira K, Yamamoto K. Experimental determination of the effective point of measurement for cylindrical ionization chambers in megavoltage photon beams. Radiol. Phys. Technol. 2022;15:291–7. doi: 10.1007/s12194-022-00669-z. [DOI] [PubMed] [Google Scholar]
- Khan F M. The Physics of Radiation Therapy. Lippincott, Williams and Wikins; 1994. p. p 114. [Google Scholar]
- Kranzer R, Poppinga D, Weidner J, Schüller A, Hackel T, Looe H K, Poppe B. Ion collection efficiency of ionization chambers in ultra-high dose-per-pulse electron beams. Med. Phys. 2021;48:819–30. doi: 10.1002/mp.14620. [DOI] [PubMed] [Google Scholar]
- Kranzer R, Schüller A, Rodríguez F G, Weidner J, Paz-Martín J, Looe H K, Poppe B. Charge collection efficiency, underlying recombination mechanisms, and the role of electrode distance of vented ionization chambers under ultra-high dose-per-pulse conditions. Phys. Med. 2022;104:10–17. doi: 10.1016/j.ejmp.2022.10.021. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Laitano R F, Guerra A S, Pimpinella M, Caporali C, Petrucci A. Charge collection efficiency in ionization chambers exposed to electron beams with high dose per pulse. Phys. Med. Biol. 2006;51:6419–36. doi: 10.1088/0031-9155/51/24/009. [DOI] [PubMed] [Google Scholar]
- Lapsley A C. Effect of space charge on saturation properties of ionization chambers. Rev. Sci. Instrum. 1953;24:602–5. doi: 10.1063/1.1770798. [DOI] [Google Scholar]
- McManus M, Romano F, Le N D, Farabolini W, Gilardi A, Royle G, Palmans H, Subiel A. The challenge of ionisation chamber dosimetry in ultra-short pulsed high dose-rate very high energy electron beams. Sci. Rep. 2020;10:9089. doi: 10.1038/s41598-020-65819-y. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Pardo J, Franco L, Gómez F, Iglesias A, Pazos A, Pena J, Lobato R, Mosquera J, Pombar M, Sendón J. Development and operation of a pixel segmented liquid-filled linear array for radiotherapy quality assurance. Phys. Med. Biol. 2005;50:1703–16. doi: 10.1088/0031-9155/50/8/006. [DOI] [PubMed] [Google Scholar]
- Paz-Martín J, Schüller A, Bourgouin A, González-Castaño D M, Gómez-Fernández N, Pardo-Montero J, Gómez F. Numerical modeling of air-vented parallel plate ionization chambers for ultra-high dose rate applications. Phys. Med. 2022;103:147–56. doi: 10.1016/j.ejmp.2022.10.006. [DOI] [PubMed] [Google Scholar]
- Petersson K, Jaccard M, Germond J-F, Buchillier T, Bochud F, Bourhis J, Vozenin M-C, Bailat C. High dose-per-pulse electron beam dosimetry—a model to correct for the ion recombination in the Advanced Markus ionization chamber. Med. Phys. 2017;44:1157–67. doi: 10.1002/mp.12111. [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Data Availability Statement
All data that support the findings of this study are included within the article (and any supplementary information files).