Abstract
Background
The Navitor Investigational Device Exemption (IDE) study is a prospective, multicenter, global study assessing the safety and effectiveness of the Navitor valve in a population with severe, symptomatic aortic stenosis who are at high and extreme surgical risk. The impact of pre-existing conduction abnormalities and implantation technique on new permanent pacemaker implantation (PPI) for the Navitor platform is not fully understood. Therefore, the goal of this analysis was to investigate the associations between patient and procedural factors and the 30-day new PPI rate.
Methods
A total of 260 patients who underwent implantation of a Navitor valve in the Navitor IDE study were reviewed. Patients with preprocedural permanent pacemakers (n = 28) were excluded. Baseline risk factors were assessed for statistical significance. Multivariable logistic regression analyses were performed to identify independent predictors of new PPI.
Results
Mean age of the pacemaker-naïve population was 83.3 ± 5.2 years, 58.6% were female, average Society of Thoracic Surgeons score was 3.8% ± 1.9%, median frailty score was 1 (interquartile range 1, 2), and 17.7% were deemed at extreme surgical risk. Pre-existing first-degree atrioventricular block and right bundle branch block significantly increased the risk of new PPI postimplantation, whereas left bundle branch block did not. Membranous septum length in relation to noncoronary cusp implant depth was a significant predictor of new PPI, with higher rates of new PPI observed when noncoronary cusp implant depth exceeded membranous septum length. Analysis of implant depth alone revealed deeper implants were associated with a higher rate of new PPI, regardless of patient baseline conduction abnormality.
Conclusions
The 30-day rate of new PPI in the Navitor IDE study is associated with patient pre-existing baseline conduction disturbances and implantation depth.
Keywords: Aortic stenosis, Conduction abnormality, Navitor, Pacemaker, Transcatheter aortic valve implantation, Transcatheter aortic valve replacement
Introduction
For elderly patients with symptomatic severe aortic stenosis, transcatheter aortic valve replacement (TAVR) is a safe and effective alternative to surgical aortic valve replacement in all surgical risk groups. However, a common post-TAVR complication is the need for new permanent pacemaker implantation (PPI). PPI post-TAVR has been associated with poor outcomes, such as a higher rate of mortality and increased hospitalizations due to heart failure.1,2 Factors that contribute to new PPI include intrinsic patient characteristics such as pre-existing conduction abnormalities and cardiac anatomical characteristics like length of the membranous septum (MS), valve type (self-expanding vs. balloon expandable), and procedural factors such as implant depth.3,4
The Navitor valve (Abbott Structural Heart, Minneapolis, MN, USA) is a next-generation, self-expanding TAVR device. In the Navitor IDE study (also known as PORTICO NG study, Evaluation of the Portico NG [Next Generation] Transcatheter Aortic Valve in High and Extreme Risk Patients With Symptomatic Severe Aortic Stenosis), which supported Food and Drug Administration approval, the rate of new PPI was 19.0%.5 As operators become more experienced with commercially available TAVR systems and as implant techniques are refined, the expectation is that the 30-day rate of new PPI post-TAVR may be expected to decrease, as has been observed with other self-expanding valve platforms.6,7 This report seeks to provide a detailed analysis of patient, anatomical, and procedural factors contributing to the rate of new PPI with the Navitor valve in the Navitor IDE study.
Methods
Study Background
The Navitor IDE study (ClinicalTrials.gov: NCT04011722) was initiated to assess the safety and effectiveness of the Navitor TAVR System, which includes the use of the FlexNav delivery system (Abbott Structural Heart, Minneapolis, MN, USA) for valve delivery, in a population at high and extreme surgical risk. The design of the study, procedural aspects, and safety and hemodynamic outcomes through 1 year have been described in detail previously.5,8
Statistical Methods
Baseline characteristics and medical history were summarized using descriptive statistics. Fisher’s exact method was used to compare patients with or without pre-existing conduction abnormalities that received a pacemaker within 30 days after TAVR and was also used to compare the association of implant depth with the need for new PPI. Implant depth analysis subgroups were defined relative to the target implant depth of 3 mm (2-4 mm), recommended per the Instructions for Use to minimize PPI.
Implant depths <2 mm were considered ‘shallow’, depth 4 to 7 mm were ‘deep’, and depth >7 mm was ‘very deep’, similar to analyses performed by Jilaihawi et al. 2019.3 Implant depth was site-reported based on distance from the base of the noncoronary cusp (NCC) to the inflow edge of the deployed Navitor stent frame on postimplant angiography. Subannular MS length was measured by a dedicated computed tomography core laboratory.3
Multivariable logistic regression analyses were performed using a stepwise selection process (entry: p <0.15, stay: p < 0.05) to identify independent predictors of new PPI. Analyses were performed using SAS software version 9.4 (SAS Institute, Cary, North Carolina).
Results
Patient Disposition
A total of 260 patients underwent implantation with a Navitor valve between September 2019 and August 2022. Twenty-eight patients had a pre-existing permanent pacemaker and were excluded from the analysis. Therefore, 232 naïve pacemaker patients were assessed for new PPI in this report.
Baseline Characteristics
Baseline clinical and electrocardiographic variables of the pacemaker-naïve population are summarized in Table 1. The mean age was 83.3 ± 5.2 years, and 58.6% of patients were female. The mean Society of Thoracic Surgeons Predicted Risk of Operative Mortality score was 3.8% ± 1.9% and the median frailty was 1 (interquartile range 1, 2). Sixty-four of the 232 (27.6%) PPI naïve patients at baseline had at least one of the following conduction abnormalities: first-degree atrioventricular (AV) block, second-degree AV block (Type I or II), right bundle branch block (RBBB), left bundle branch block (LBBB), left anterior fascicular block (LAFB); 13 patients (5.6%) had 2 or more abnormalities. Common cardiac arrhythmias and conduction abnormalities included atrial fibrillation (23.3%), RBBB (12.1%), first-degree AV block (9.9%), and LBBB (6.5%). Baseline conduction abnormalities (listed above) were more common in patients who received a new PPI, except for LBBB.
Table 1.
Baseline variable | Total (N = 232) | New PPI (N = 44) | No PPI (N = 188) |
---|---|---|---|
Baseline clinical variables | |||
Age, y | 83.3 ± 5.2 | 83.8 ± 6.1 | 83.1 ± 5.0 |
Sex, female | 58.6% | 59.1% | 58.5% |
STS-PROM Score, % | 3.8 ± 1.9 | 3.9 ± 2.2 | 3.8 ± 1.9 |
Total Frailty Score | |||
Median (Q1, Q3) | 1 (1, 2) | 1 (1, 2) | 1 (1, 2) |
Baseline electrocardiographic variables | |||
Persistent atrial fibrillation | 23.3% | 29.5% | 21.8% |
AV conduction abnormality∗ | 33.7% | 60.5% | 25.0% |
Presence of first-degree AV block | 9.9% | 25.0% | 6.4% |
Presence of second-degree AV block - Type I | 0.9% | 2.3% | 0.5% |
Presence of second-degree AV block - Type II | 0.4% | 2.3% | 0% |
Presence of third-degree AV block | 0% | 0% | 0% |
Presence of LAFB | 3.9% | 9.1% | 2.7% |
Presence of LBBB | 6.5% | 2.3% | 7.4% |
Presence of RBBB | 12.1% | 29.5% | 8.0% |
Abbreviations: AV, atrioventricular; LAFB, left anterior fascicular block; LBBB, left bundle branch block; PPI, permanent pacemaker implantation; RBBB, right bundle branch block; STS-PROM, Society of Thoracic Surgeons Predicted Risk of Operative Mortality.
Conduction abnormalities are not mutually exclusive Values are mean ± SD or n (%) that reflect missing values.
Conduction Abnormality Correlation to New PPI
An analysis was performed to assess whether there was a significant difference between patients with vs. without baseline conduction abnormalities known to be associated with new PPI post-TAVR, as presented in Figure 1. Patients with pre-existing first-degree AV block (47.8 vs. 15.8%, p < 0.001) and RBBB (46.4 vs. 15.2%, p < 0.001) had higher rates of new PPI compared to patients without first-degree AV block or RBBB. There was no significant difference between patients with vs. without pre-existing LBBB (6.7 vs. 19.8%, p = 0.314) and new PPI.
Multivariable Logistic Regression Analysis to Determine Predictors for New PPI
A multivariable model with baseline and procedural covariates was used to examine independent predictors of new PPI post-TAVR. The list of covariates included in the model is presented in Supplemental Table 1. Seven covariates were selected from the univariable analysis for incorporation into the multivariable analysis: pre-existing conduction disturbance, implant depth on the NCC, MS length ≤NCC implant depth, resheathing, QRS interval >120 ms, aortic annulus eccentricity ≥0.73, and aortic valve area. Only 1 variable, MS length ≤NCC implant depth (p = 0.0023, odds ratio of 3.42), was selected by the model as significant in the multivariable analysis.
MS length and its relation to NCC implant depth and new PPI are presented in Figure 2. MS length was not available for 6 patients. The overall rate of new PPI when NCC implant depth was greater than or equal to MS length was 26.8 vs. 9.1% when NCC implant depth was less than MS length (see Panel 1).
Additionally, patients were grouped based on length of MS: short (Panel 2; <2 mm, n = 69), medium (Panel 3; 2-5 mm, n = 109), and long (Panel 4; >5 mm, n = 48). In all subgroups, the rate of new PPI was higher when NCC implant depth was greater than or equal to MS length. Excluding patients with pre-existing first-degree AV block and RBBB from the analysis, the rate of new PPI was <5% in all subgroups when NCC implant depth was less than MS length (Supplemental Figure 1).
NCC Implant Depth
The relationship of implant depth to new PPI, irrespective of MS length, is presented in Figure 3. All pacemaker-naïve patients had a recorded NCC implant depth. The overall rate of new PPI was 14.1% when the Navitor valve was implanted in the target range (Panel 2), 9.5% when implanted shallow (Panel 1), 19.8% when implanted deep (Panel 3), and 50.0% when implanted very deep (Panel 4). Excluding patients with pre-existing first-degree AV block and/or RBBB (dark blue bars), the rate of new PPI was <10% in both the target (7.5%) and shallow (5.7%) subgroups, but >10% in the deep (12.5%) and very deep (47.6%) subgroups. Excluding all patients with pre-existing conduction abnormalities (light blue bars), the rate of new PPI remained similar in each subgroup vs. when only patients with first-degree AV block and/or RBBB were removed.
Discussion
We report a detailed investigation into new PPI post-TAVR for the Navitor valve in the Navitor IDE study. The results from these analyses suggest baseline conduction abnormalities and implant depth were associated with the rate of new PPI (Figure 4).
30-Day New PPI Rate in Context
The rate of new PPI at 30 days was 19.0% in the Navitor IDE study.5 In other premarket studies of next-generation TAVR devices (i.e., with a paravalvular leak mitigation feature), the rate of new PPI is numerically lower and ranges from 11.8 to 15.0% in a similar high- or extreme-risk population.9, 10, 11 However, in a postmarket setting, the rate of new PPI varies widely. For example, the FORWARD PRO registry reported a 30-day new PPI rate of 20.7%, hypothesized to be attributed to a large portion of patients (30%) with pre-existing conduction disturbances.12 The ACURATE neo2 PMCF study and SOURCE 3 registry reported 30-day new PPI rates lower than their respective premarket rates of 6.5 and 12.0%.13,14 Due to the variation of PPI rates across studies, either due to sample size, chance, or other factors, cross-study comparisons should be considered with caution.
Baseline Conduction Disturbance Is Associated With New PPI
Of the 232 pacemaker-naïve patients, 64 (27.6%) had 1 (or more) of the following baseline conduction abnormalities: first-degree AV block, second-degree AV block (Type I or II), RBBB, LBBB, LAFB. As most (45/64, 70.3%) pacemaker-naïve patients with baseline conduction abnormalities had either first-degree AV block and/or RBBB, it is important to understand the potential association between these abnormalities and risk for new PPI post-TAVR. In our analysis, patients with first-degree AV block (47.8 vs. 15.8%) or RBBB (46.4 vs. 15.2%) were roughly 3x as likely to receive a new PPI post-TAVR compared to patients without first-degree AV block or RBBB at baseline. This is consistent with reports of both self-expanding and balloon-expandable valves, where RBBB was found to be an important predictor of new PPI.15,16
Indeed, a 2014 meta-analysis of 41 studies that reported incidence of new PPI post-TAVR revealed that pre-existing first-degree AV block was a predictor of PPI in 6 studies, while RBBB was a predictor of PPI in 17 studies.17 In the same meta-analysis, LBBB was a predictor of PPI in 16 studies, unlike our findings in the Navitor IDE study. A more recent meta-analysis of 78 studies performed in 2021 confirmed earlier findings, revealing baseline conduction abnormalities, including second-degree AV block Type I, LAFB, and RBBB were associated with higher odds of new PPI post-TAVR.18 These meta-analysis data suggest an increased likelihood of new PPI post-TAVR with common baseline AV conduction abnormalities, especially RBBB, irrespective of other factors.
The Importance of the Membranous Septum
Our predictor analysis revealed NCC implant depth relative to MS length as the only independent predictor of new PPI in the Navitor IDE study. The relative length of the MS (short, medium, or long) did not impact the overall trend observed, as this association was noted across all categories of MS length. This is consistent with other analyses of implant depth and MS length in patients who received a self-expandable TAVR valve,3,4 as well as balloon-expandable TAVR.19
Consistent with recent studies, our study shows that MS length is an important patient factor associated with new PPI. Understanding patient-specific cardiac anatomy, specifically measuring the MS, may aid operators in offering better prognostications of the likelihood for the need of a PPI following TAVR. While measuring MS length may allow physicians to have a discussion with patients with regards to the need for pacemaker, it may be challenging to implant with extreme precision. Operators should continue to be cautious about depth of deployment and the need for pacemaker vs. an embolized transcatheter heart valve.
Implant Technique Best Practices Will Help Achieve Target Implant Depth
This study showed that as the depth of valve implantation increases, irrespective of MS length or baseline conduction disturbances, so too does the rate of new PPI. This phenomenon has also been shown for other self-expanding valve platforms.15 This has been hypothesized to be due to the compression of the cardiac conduction system caused by the valve stent frame, leading to unresolvable post-TAVR conduction disturbances.20 To prevent deep implantation of the Navitor valve, Abbott has refined the implant technique to attain a target implant depth of 3 mm (recommended in the Instructions for Use). Key steps of this implant technique are demonstrated in Figure 4 and described below.
First, predilatation is recommended to prepare the annulus for the Navitor valve and reduces the need for resheathing and postdilatation, thus limiting procedural manipulations of the valve. Following predilatation, initial valve positioning is critical. Positioning the inflow edge of the stent frame so it is aligned with the base of the NCC allows for optimal positioning, which will aid in achieving the recommended target implant depth. Third, prior to releasing the valve, confirming the implant depth in both cusp overlap and an alternative view ensures proper mitigation of paravalvular leak and reduces the risk of depth-related PPI. Imaging in the cusp overlap view provides an accurate depth of implant for the NCC, while a subsequent left anterior oblique view with parallax removed confirms the depth for the left coronary cusp. Lastly, following valve deployment, withdrawing the guidewire to a mid-ventricular position, and confirming all 3 retainer tabs have detached, ensures no unintentional snaring or movement of the valve.
Limitations
There are limitations to consider in this study. Implant depth used in these analyses were site-reported and not confirmed by an independent arbitrator. In addition, other factors (i.e., calcium burden, additional procedural manipulations, and new intraprocedural conduction disturbances) have been shown to be associated with new PPI post-TAVR. The authors investigated major signals pertaining to new PPI in this study, and further investigation into other factors influencing new PPI are beyond the scope of this report.
Conclusions
The rate of new PPI at 30 days in the Navitor IDE study is associated with patient pre-existing baseline conduction disturbances and implant depth. With increased operator experience of the Navitor TAVR system and understanding patient-specific anatomic nuances, the rate of new PPI is expected to diminish in real-world practice.
Impact on Daily Practice
Informed discussions with patients surrounding pre-existing conduction disturbances and anatomical considerations will allow implanters to inform patients about their risk of receiving a pacemaker following an implant with the Navitor valve.
Ethics Statement
The study was conducted in accordance with the Declaration of Helsinki and was approved by the appropriate Institutional Review Board/Ethics Committee of each investigational site and by the applicable regulatory authorities. All patients provided informed consent prior to participation.
Funding
The study was sponsored by Abbott.
Disclosure Statement
I. Sultan is a consultant and receives research support from Abbott, Artivion, Atricure, Boston Scientific, Edwards Lifesciences, Medtronic, and Terumo Aortic. M. J. Reardon reports consultant fees and/or institutional research grants from Abbott, Boston Scientific, Medtronic, and Gore Medical. L. Sondergaard is Chief Medical Officer at Abbott Structural Heart and has previously received consultant fees and/or institutional research grants from Abbott, Boston Scientific, Medtronic, and Sahajanand Medical Technologies. B. Chehab reports speaker, consultancy, and proctor fees for Abbott, Edwards Lifesciences, Medtronic, and CSI. D. Smith reports speaker and proctor fees for Abbott, Edwards Lifesciences, and Biosensors. A. S. Walton is a proctor on medical advisory boards and receives grant support from Abbott, Medtronic, and Edwards Lifesciences. S. G. Worthley has received speaker fees and proctorship from Abbott and HighLife Medical, as well as consultancy fees from Edwards Lifesciences. G. Manoharan is a proctor for Abbott and Medtronic. G. Yong is a proctor for Abbott. H. Jilaihawi reports consultant fees and/or institutional research grants from Abbott, Edwards Lifeciences, Medtronic, and Pi-Cardia. F. Asch reports institutional research grants from Abbott, Edwards, Medtronic, BSC, Corcym, and Biotronik. N. Bates is an employee of Abbott. G. P. Fontana is a consultant, proctor, and speaker for Abbott and Medtronic.
Acknowledgments
The authors would like to thank all the investigators and institutions participating in the Navitor IDE study, and Kai Koo, PhD, and Feiyi Jia, PhD (Abbott), for their contributions to data analysis.
Footnotes
Supplemental data for this article can be accessed on the publisher’s website.
Supplementary Material
References
- 1.Jørgensen T.H., De Backer O., Gerds T.A., Bieliauskas G., Svendsen J.H., Søndergaard L. Mortality and heart failure hospitalization in patients with conduction abnormalities after transcatheter aortic valve replacement. JACC Cardiovasc Interv. 2019;12:52–61. doi: 10.1016/j.jcin.2018.10.053. [DOI] [PubMed] [Google Scholar]
- 2.Fadahunsi O.O., Olowoyeye A., Ukaigwe A., et al. Incidence, predictors, and outcomes of permanent pacemaker implantation following transcatheter aortic valve replacement: analysis from the U.S. Society of Thoracic Surgeons/American College of Cardiology TVT Registry. JACC Cardiovasc Interv. 2016;9:2189–2199. doi: 10.1016/j.jcin.2016.07.026. [DOI] [PubMed] [Google Scholar]
- 3.Jilaihawi H., Zhao Z., Du R., et al. Minimizing permanent pacemaker following repositionable self-expanding transcatheter aortic valve replacement. JACC Cardiovasc Interv. 2019;12:1796–1807. doi: 10.1016/j.jcin.2019.05.056. [DOI] [PubMed] [Google Scholar]
- 4.Hamdan A., Guetta V., Klempfner R., et al. Inverse relationship between membranous septal length and the risk of atrioventricular block in patients undergoing transcatheter aortic valve implantation. JACC Cardiovasc Interv. 2015;8:1218–1228. doi: 10.1016/j.jcin.2015.05.010. [DOI] [PubMed] [Google Scholar]
- 5.Reardon M.J., Chehab B., Smith D., et al. 30-Day clinical outcomes of a self-expanding transcatheter aortic valve: the international PORTICO NG study. JACC Cardiovasc Interv. 2023;16:681–689. doi: 10.1016/j.jcin.2023.02.002. [DOI] [PubMed] [Google Scholar]
- 6.Carroll J.D., Mack M.J., Vemulapalli S., et al. STS-ACC TVT registry of transcatheter aortic valve replacement. J Am Coll Cardiol. 2020;76:2492–2516. doi: 10.1016/j.jacc.2020.09.595. [DOI] [PubMed] [Google Scholar]
- 7.Grubb K.J., Gada H., Mittal S., et al. Clinical impact of standardized TAVR technique and care pathway: insights from the optimize PRO study. JACC Cardiovasc Interv. 2023;16:558–570. doi: 10.1016/j.jcin.2023.01.016. [DOI] [PubMed] [Google Scholar]
- 8.Sondergaard L., Walton A.S., Worthley S.G., et al. Thirty-day and one-year outcomes of the Navitor transcatheter heart valve in patients with aortic stenosis: the prospective, multicentre, global PORTICO NG study. EuroIntervention. 2023;19(3):248–255. doi: 10.4244/EIJ-D-22-01108. [DOI] [PubMed] [Google Scholar]
- 9.Forrest J.K., Mangi A.A., Popma J.J., et al. Early outcomes with the Evolut PRO repositionable self-expanding transcatheter aortic valve with pericardial wrap. JACC Cardiovasc Interv. 2018;11:160–168. doi: 10.1016/j.jcin.2017.10.014. [DOI] [PubMed] [Google Scholar]
- 10.Möllmann H., Holzhey D.M., Hilker M., et al. The ACURATE neo2 valve system for transcatheter aortic valve implantation: 30-day and 1-year outcomes. Clin Res Cardiol. 2021;110:1912–1920. doi: 10.1007/s00392-021-01882-3. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11.Kodali S., Thourani V.H., White J., et al. Early clinical and echocardiographic outcomes after SAPIEN 3 transcatheter aortic valve replacement in inoperable, high-risk and intermediate-risk patients with aortic stenosis. Eur Heart J. 2016;37:2252–2262. doi: 10.1093/eurheartj/ehw112. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.Manoharan G., Grube E., Van Mieghem N.M., et al. Thirty-day clinical outcomes of the Evolut PRO self-expanding transcatheter aortic valve: the international FORWARD PRO study. EuroIntervention. 2020;16:850–857. doi: 10.4244/EIJ-D-20-00279. [DOI] [PubMed] [Google Scholar]
- 13.Kim W.K., Tamburino C., Möllmann H., et al. Clinical outcomes of the ACURATE neo2 transcatheter heart valve: a prospective, multicenter, observational, post-market surveillance study. EuroIntervention. 2022;19(1):83–92. doi: 10.4244/EIJ-D-22-00914. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14.Wendler O., Schymik G., Treede H., et al. SOURCE 3 Registry: design and 30-day results of the European Postapproval Registry of the latest generation of the SAPIEN 3 transcatheter heart valve. Circulation. 2017;135:1123–1132. doi: 10.1161/CIRCULATIONAHA.116.025103. [DOI] [PubMed] [Google Scholar]
- 15.Gada H., Vora A.N., Tang G.H.L., et al. Site-level variation and predictors of permanent pacemaker implantation following TAVR in the Evolut low-risk trial. Cardiovasc Revasc Med. 2023;47:48–54. doi: 10.1016/j.carrev.2022.10.004. [DOI] [PubMed] [Google Scholar]
- 16.Nazif T.M., Dizon J.M., Hahn R.T., et al. Predictors and clinical outcomes of permanent pacemaker implantation after transcatheter aortic valve replacement: the PARTNER (Placement of AoRtic TraNscathetER Valves) trial and registry. JACC Cardiovasc Interv. 2015;8:60–69. doi: 10.1016/j.jcin.2014.07.022. [DOI] [PubMed] [Google Scholar]
- 17.Siontis G.C., Jüni P., Pilgrim T., et al. Predictors of permanent pacemaker implantation in patients with severe aortic stenosis undergoing TAVR: a meta-analysis. J Am Coll Cardiol. 2014;64:129–140. doi: 10.1016/j.jacc.2014.04.033. [DOI] [PubMed] [Google Scholar]
- 18.Ullah W., Zahid S., Zaidi S.R., et al. Predictors of permanent pacemaker implantation in patients undergoing transcatheter aortic valve replacement - a systematic review and meta-analysis. J Am Heart Assoc. 2021;10 doi: 10.1161/JAHA.121.020906. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19.Maeno Y., Abramowitz Y., Kawamori H., et al. A highly predictive risk model for pacemaker implantation after TAVR. JACC Cardiovasc Imaging. 2017;10:1139–1147. doi: 10.1016/j.jcmg.2016.11.020. [DOI] [PubMed] [Google Scholar]
- 20.Nai Fovino L., Cipriani A., Fabris T., et al. Anatomical predictors of pacemaker dependency after transcatheter aortic valve replacement. Circ Arrhythmia Electrophysiol. 2021;14 doi: 10.1161/CIRCEP.120.009028. [DOI] [PubMed] [Google Scholar]