. 2024 Aug 6;25(8):e14444. doi: 10.1002/acm2.14444

Correction: “Commissioning and clinical evaluation of the IDENTIFYTM surface imaging system for frameless stereotactic radiosurgery” https://doi.org/10.1002/acm2.14058

Elizabeth L Covington 1,2, Dennis N Stanley 2, Rodney J Sullivan 2, Kristen O Riley 2, John B Fiveash 1, Richard A Popple 2
PMCID: PMC11302797  PMID: 39106407

An incorrect table was inadvertently uploaded into the original manuscript. The below table contains the correct values for the median and interquartile range values for the IDENTIFY™ surface imaging system before beam‐on at non‐zero couch angles.

TABLE 2.

Median and interquartile range (IQR) of IDENTIFY™ reported offsets before beam‐on at non‐zero couch angles.

Couch angle (°) Vertical (IQR) mm Longitudinal (IQR) (mm) Lateral (IQR) (mm)
270 −0.01 (0.47) −0.03 (0.33) 0.16 (0.32)
225 −0.06 (0.17) 0.21 (0.38) −0.16 (0.18)
135 −0.08 (0.16) 0.19 (0.38) −0.41 (0.19)
90 −0.14 (0.20) 0.05 (0.54) −0.25 (0.24)
0° (end of treatment) −0.05 (0.18) −0.03 (0.35) −0.01 (0.15)