Abstract
Purpose
This study investigated the potential of a commercially available plastic scintillator, the Exradin W2, as a real‐time dosimeter for ultra‐high‐dose‐rate (UHDR) electron beams. This work aimed to characterize this system's performance under UHDR conditions and addressed limitations inherent to other conventional dosimetry systems.
Methods and materials
We assessed the W2's performance as a UHDR electron dosimeter using a 16 MeV UHDR electron beam from the FLASH research extension (FLEX) system. Additionally, the vendor provided a beta firmware upgrade to better handle the processing of the high signal generated in the UHDR environment. We evaluated the W2 regarding dose‐per‐pulse, pulse repetition rate, charge versus distance, and pulse linearity. Absorbed dose measurements were compared against those from a plane‐parallel ionization chamber, optically stimulated luminescent dosimeters and radiochromic film.
Results
We observed that the 1 × 1 mm W2 scintillator with the MAX SD was more suitable for UHDR dosimetry compared to the 1 × 3 mm W2 scintillator, capable of matching film measurements within 2% accuracy for dose‐per‐pulse up to 3.6 Gy/pulse. The W2 accurately ascertained the inverse square relationship regarding charge versus virtual source distance with R2 of ∼1.00 for all channels. Pulse linearity was accurately measured with the W2, demonstrating a proportional response to the delivered pulse number. There was no discernible impact on the measured charge of the W2 when switching between the available repetition rates of the FLEX system (18–180 pulses/s), solidifying consistent beam output across pulse frequencies.
Conclusions
This study tested a commercial plastic scintillator detector in a UHDR electron beam, paving the way for its potential use as a real‐time, patient‐specific dosimetry tool for future FLASH radiotherapy treatments. Further research is warranted to test and improve the signal processing of the W2 dosimetry system to accurately measure in UHDR environments using exceedingly high dose‐per‐pulse and pulse numbers.
Keywords: dosimetry, electron ultra‐high dose rate (eFLASH), FLASH research extension (FLEX), plastic scintillator
1. INTRODUCTION
FLASH radiotherapy, delivering radiation at lightning‐fast speed (> 40 Gy/s), is a promising modality in treating cancer. 1 The “FLASH effect” potentially improves the sparing of healthy organs and tissues for patients with cancer compared to conventional radiotherapy and decreases overall treatment time. 2 Though the science is still evolving, FLASH has shown potential in lab experiments, 3 , 4 animal studies, 5 , 6 and even early human trials. 7 , 8
A significant roadblock between the promise and widespread use of FLASH radiotherapy is measuring how much radiation dose patients receive. Unlike conventional radiotherapy, existing dosimeters insufficiently address the ultra‐high dose rate (UHDR) environment produced by FLASH radiotherapy. While suitable for real‐time monitoring, ionization chambers lose accuracy at high absorbed dose and dose rates. 9 , 10 , 11 , 12 , 13 , 14 Radiochromic film and optically stimulated luminescent dosimeters (OSLDs) are excellent options for passive UHDR dosimetry but lack real‐time feedback and have maximum absorbed dose limitations (e.g., maximum dose of 15 Gy for OSLDs). 15 , 16 Yet to be established is the ideal, real‐time UHDR dosimeter. 17 , 18 In this study, we explored the potential of a real‐time, commercially available plastic scintillator for UHDR electron dosimetry using the novel FLASH research extension (FLEX) system.
2. MATERIALS AND METHODS
2.1. FLEX system
A commercial linear accelerator (Clinac 23EX, Varian Medical Systems, Palo Alto, CA, USA) was converted by the original vendor to the FLEX system, capable of delivering both conventional and UHDR electron beams. This unique system uses controlled pulse numbers (1–99 pulses) and pulse repetition frequencies (18–180 Hz) to generate a 16 MeV UHDR beam with approximately 180 Gy/s at the isocenter, reaching over 680 Gy/s at the gantry upper accessory mount. 19 , 20 , 21
2.2. W2 and MAX SD system
This study investigated the promising potential of a commercial plastic scintillator (Exradin W2 Scintillator, Standard Imaging, Inc., Middleton, WI, USA) for real‐time UHDR electron dosimetry. 22 The W2 is water‐equivalent and ideal for small‐field dosimetry. 22 Paired with the MAX SD optical detection and signal processing unit, both the 1 × 1 mm and 1 × 3 mm W2 scintillators were evaluated in this study. The MAX SD was originally optimized for small signals, aligning with its intended use for small‐field dosimetry. However, UHDR radiation generates a large quantity of visible photons within the scintillator in a very short time. The MAX SD system uses photo‐detectors to convert the visible light signal into an electric signal for which the measured output of the electrometer is current or charge. Because the system was not originally designed to handle large currents, this leads to over‐ranging in the UHDR environment. To address this, the vendor provided a beta‐version firmware upgrade for the MAX SD, enhancing its ability to process the high signals produced by the UHDR environment and enabling it to handle currents up to 35 nA. Overall, the W2 dosimetry system features advanced processing capabilities, user‐friendly interfaces, online control, and easy data interpretation through Čerenkov‐corrected scintillation. 23 , 24
The MAX SD utilizes a chromatic spectral correction method to mitigate the Čerenkov light impact within the W2 detector system. This approach relies on the blue and green channel signal ratio, measured with both maximum and minimum fiber configurations as displayed in Figure 1. 21 The following equation determines the Čerenkov light ratio (CLR):
where () represents the blue (green) channel signal with the maximum irradiated fiber configuration and () represents the blue (green) channel signal with the minimum irradiated fiber configuration (Figure 1). 22 , 24 Jacqmin et al. 22 provide additional details on how the CLR is applied to correct for the Čerenkov light.
FIGURE 1.

CLR measurement setups using the large‐field method for (a) minimized irradiation to the fiber and (b) maximized irradiation to the fiber. The plastic blocks with two different fiber pathways were 3D‐printed using polylactic acid filament (density of ∼1.2 g/cm3).
2.3. Dosimetric measurements
We deployed the W2 scintillator to analyze the beam properties of the FLEX system. Radiochromic film (GafChromic EBT‐XD, Ashland Inc., Wayne, NJ, USA) was used to validate the accuracy of the W2 system in the UHDR environment, particularly the detection efficiency of the W2 with increasing dose‐per‐pulse. Further tests used the W2 to characterize the 16 MeV FLEX system regarding:
Source‐to‐dosimeter distance (SDD) dependence with virtual source correction
Pulse linearity (i.e., output constancy with delivered number of pulses)
Output constancy versus pulse repetition frequency (PRF)
2.3.1. Detection efficiency
We tested the detection efficiency of the W2 plastic scintillators (1 × 1 mm and 1 × 3 mm) for a range of UHDR environments generated by the 16 MeV FLEX system. The gantry was positioned at 0 degrees (IEC scale) for all experiments with a field size of 10 × 10 cm2 without an applicator. To minimize air gaps, the W2 was positioned beneath 1‐cm of water‐equivalent bolus which is made from solid, homogeneous, and tissue‐equivalent gel (CQ Medical, Orange City, IA, USA). Additional water‐equivalent slabs were placed directly on top of the bolus to attain variable depths of measurement (3–9 cm). Furthermore, 10‐cm of water‐equivalent slabs were used downstream of the W2 for adequate backscatter conditions. The W2 system collected charge at an array of source‐to‐detector distances (61.6–132 cm) using a range of pulse numbers (1, 5, 10, and 43 pulses) at the highest PRF (180 Hz). Corresponding radiochromic film measurements of absolute dose enabled the conversion of the W2‐measured charge to dose. Furthermore, we compared measurements with the 1 × 1 mm W2 scintillator against a plane‐parallel ionization chamber (Advance Markus Electron Chamber, PTW, Freiburg, Germany) 13 , 15 , 25 and OSLDs (NanoDot, Landauer Inc., Glenwood, Il, USA).
2.3.2. Source‐to‐dosimeter distance (SDD) dependence
We explored the charge‐to‐distance relationship for both W2 scintillators positioned at 3 cm depth in the water‐equivalent slab phantom. We varied the SDD from 61.6 to 132 cm and delivered a single UHDR pulse at each distance.
2.3.3. Pulse linearity
We tested the pulse linearity of the system by evaluating the collected charge of the 1 × 3 mm W2 scintillator with an increasing number of delivered pulses. The W2 was positioned at 136‐cm SDD and 6‐cm depth in the water‐equivalent slab phantom. We delivered 1, 2, 4, 6, 8, 10, 20, 30, 40, 50, 60, 70, 80, 90, and 99 pulses at 180 Hz PRF to assess the pulse linearity.
2.3.4. Pulse repetition frequency effect
Similarly, we studied the effect of PRF on the system by positioning the 1 × 3 mm W2 scintillator at 136‐cm SDD and 6‐cm depth in the water‐equivalent slab phantom. We tested PRFs between 18 and 180 pulses/s (18, 36, 54, 72, 90, 108, and 180 Hz) by delivering 10 pulses at each PRF setting.
3. RESULTS
3.1. Detection efficiency
The signal from the larger, 1 × 3 mm W2 scintillator saturated at a lower number of delivered pulses and lower dose‐per‐pulse than the smaller, 1 × 1 mm W2 scintillator (Figure 2a). While the 1 × 3 mm W2 scintillator began to deviate from the linear relationship at 0.4 Gy/pulse (43 pulses) and 0.8 Gy/pulse (single pulse), the 1 × 1 mm W2 scintillator performed better, staying linear up to 0.9 Gy/pulse (5, 10, and 43 pulses) and 3.6 Gy/pulse (single pulse) (i.e., the maximum dose‐per‐pulse of the FLEX system). Furthermore, in comparing dose‐per‐pulse measurements across dosimetry systems (Figure 2b), the plane‐parallel ion chamber accurately matched the reference film up to 0.4 Gy/pulse (within 5%), while the 1 × 1 mm W2 scintillator remarkably demonstrated increased accuracy (within 2%) up to 3.6 Gy/pulse.
FIGURE 2.

(a) Dose‐per‐pulse measured by both W2 scintillators with differing delivered number of pulses compared to radiochromic film; and (b) Percentage deviation of the measured dose‐per‐pulse by W2 1 × 1 mm scintillator, plane‐parallel ionization chamber, plane‐parallel ionization chamber (corrected) and OSLD from the reference dose‐per‐pulse determined by radiochromic film (single pulse delivery).
3.2. Source‐to‐dosimeter distance (SDD) dependence
First, we applied the virtual‐source distance correction to the FLEX machine's single pulse data using the 1 × 1 mm W2 scintillator (Figure 3a), and we observed the classical inverse square relationship between the collected charge and SDD. This strong relationship held for all three measurements (i.e., the blue channel, green channel, and Čerenkov‐corrected net output) with correlation coefficients (R 2 ) of 0.999, 0.998, and 0.998, respectively. Second, we analyzed the results of the 1 × 3 mm W2 scintillator for which the green channel continued to show this inverse square relationship. However, the blue channel and net output displayed this relationship only at larger SDDs (>92.5 cm for blue and >103 cm for net output). This discrepancy was due to the saturation of the 1 × 3 mm W2 scintillator at a higher dose‐per‐pulse (Figure 3b), because of the over‐ranging of the MAX SD system at higher currents.
FIGURE 3.

Source‐to‐dosimeter distance (SDD) dependence for the (a) 1 × 1 mm W2 scintillator and (b) 1 × 3 mm W2 scintillator. Shown is the comparison of the expected charge (continuous solid lines) calculated from the inverse square law accounting for the virtual source correction versus the actual measured charge by W2 plastic scintillators (discrete dots) at various SDDs.
3.3. Pulse linearity
Using the 1 × 3 mm W2 scintillator, the collected charge for the blue channel, green channel, and net output all exhibited a proportional relationship with the number of delivered pulses, (Figure 4a).
FIGURE 4.

Displayed is the (a) pulse linearity test, i.e., measured charge versus number of delivered pulses, and (b) measured charge versus pulse repetition frequency (PRF) using the 1×3 mm W2 scintillator.
3.4. Pulse repetition frequency effect
The plot in Figure 4(b) exhibits a horizontal trend line (i.e., zero slope) across all channels, indicating no significant change in the collected charge across all PRFs (18–180 Hz). This further demonstrates the stable output of the FLEX system at different pulse rates.
4. DISCUSSION
Our study highlighted a promising commercial plastic scintillator for real‐time UHDR electron dosimetry using the FLEX system. The W2 system responded exceptionally well to the UHDR beam characteristics, including pulse number, pulse repetition frequency, distance, and dose‐per‐pulse, particularly for the small W2 scintillator using low pulse settings. This implies that the 1 × 1 mm W2 scintillator is suitable for real‐time monitoring in UHDR applications using the FLEX system. Further research is warranted to test and improve the signal processing of the W2 system (i.e., the MAX SD) to accurately measure in UHDR environments using exceedingly high dose‐per‐pulse and pulse numbers.
In contrast to the W2, many conventional dosimetry systems fall short of accurate, real‐time monitoring of UHDR beams. Table 1 summarizes some advantages and disadvantages of using traditional dosimetry techniques, such as radiochromic film, TLD, OSLD, and ion chamber, in UHDR electron environments.
TABLE 1.
Summary of characteristics for conventional dosimetry systems in UHDR electron environments.
| Dosimetry system | Characteristics | References |
|---|---|---|
| Radiochromic film |
|
6, 11, 25, 28 |
| OSLD |
|
28, 29 |
| TLD |
|
11, 30, 31 |
| Alanine |
|
6, 26, 31, 32 |
| Ion chamber |
|
6, 9, 11, 25, 26, 28, 30 |
Researchers have also been approaching the challenging UHDR environment with innovative, real‐time dosimetry technologies:
Ultra‐thin ionization chambers have demonstrated accuracy up to 5.4 Gy/pulse with minimal recombination losses (1.4%). 10 , 33
Diamond detectors have also exhibited excellent accuracy up to 11.9 Gy/pulse with a deviation of less than 5%. 17 , 34
Calorimeters and silicon carbide detectors hold promise for future real‐time monitoring. 35 , 36
A dual beam‐current transformer offers a novel approach with a linear relationship to the dose rate. 37
While the accuracy of many conventional scintillating dosimeters, such as phosphor point scintillators and some plastic scintillators, suffer at high dose rates, 38 promising alternatives are emerging for accurate and real‐time UHDR beam monitoring. 18 In our study, the original MAX SD firmware proved incompatible with ultra‐high dose rate dosimetry for both W2 scintillators. The resulting dose‐per‐pulse curve rapidly saturated and suffered even in the lower dose range. However, a beta‐version firmware upgrade provided by the vendor considerably improved the W2/MAX SD performance for UHDR beams and increased the current range upper limit from a maximum of 100 pA (standard setting) to 35 nA (beta‐version upgrade). This upgrade demonstrated optimal performance with the smaller W2 scintillator and a low pulse setting. However, the larger W2 scintillator or higher pulse numbers (up to 43 pulses) can still cause over‐ranging issues, resulting in saturation of the system. This issue could be further addressed through additional modifications to the MAX SD, featuring a more comprehensive current range to prevent overloading. Once resolved, this could solidify the W2 plastic scintillator as a viable UHDR dosimetry option. 39 , 40 Lastly, the W2 scintillator has a minor inherent limitation: its optical fiber suffers from radiation degradation of approximately 2% per kGy. 23
5. CONCLUSIONS
We demonstrated the feasibility of using a commercial plastic scintillator with optimized configuration and updated firmware for real‐time UHDR electron dosimetry using the FLEX system. This development paves the way for real‐time, patient‐specific dosimetry verification during FLASH radiotherapy, alongside other potential applications.
AUTHOR CONTRIBUTIONS
Kyuhak Oh, Megan A. Hyun, Kyle J. Gallagher, Ying Yan, and Sumin Zhou contributed to the conception and design of the project, as well as conducting all experiments, corresponding data analysis, and writing, reviewing, and editing the manuscript. Ying Yan and Sumin Zhou (Senior author) are the corresponding authors for this work.
CONFLICT OF INTEREST STATEMENT
University of Nebraska Medical Center is a member of the Varian FlashForward Consortium.
ACKNOWLEDGMENTS
This work was supported in part by the University of Nebraska Collaboration Initiative Program Award (to Ying Yan and Sumin Zhou), and a pilot project award (to Ying Yan and Sumin Zhou) from the Great Plains IDeA‐CTR‐Program (NIH/NIGMS/5U54GM115458). We are grateful for the technical support from Eric Garz and Shannon Holmes at Standard Imaging. Additionally, we would like to thank Marty Schmidt from Varian for his continued technical support in regard to the FLEX system and Brian Maas from McGoogan Health Science Library, UNMC, for his 3D printing services. The authors are also grateful for the support from Faith Regional Carson Cancer Center and Varian Medical Systems in regard to the installation of the FLEX research extension.
Oh K, Hyun MA, Gallagher KJ, Yan Y, Zhou S. Characterization of a commercial plastic scintillator for electron FLASH dosimetry. J Appl Clin Med Phys. 2024;25:e14451. 10.1002/acm2.14451
Contributor Information
Ying Yan, Email: yyan@unmc.edu.
Sumin Zhou, Email: szhou@unmc.edu.
REFERENCES
- 1. Bourhis J, Montay‐Gruel P, Gonçalves Jorge P, et al. Clinical translation of FLASH radiotherapy: Why and how? Radiother Oncol. 2019;139:11–17. [DOI] [PubMed] [Google Scholar]
- 2. Rothwell B, Lowe M, Traneus E, Krieger M, Schuemann J. Treatment planning considerations for the development of FLASH proton therapy. Radiother Oncol. 2022;175:222–230. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3. Adrian G, Ruan JL, Paillas S, Cooper CR, Petersson K. In vitro assays for investigating the FLASH effect. Expert Rev Mol Med. 2022;24:e10. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4. Adrian G, Konradsson E, Beyer S, et al. Cancer cells can exhibit a sparing FLASH effect at low doses under normoxic in vitro‐conditions. Front Oncol. 2021;11:686142. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5. Levy K, Natarajan S, Wang J, et al. Abdominal FLASH irradiation reduces radiation‐induced gastrointestinal toxicity for the treatment of ovarian cancer in mice. Sci Rep. 2020;10(1):21600. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6. Moeckli R, Gonçalves Jorge P, Grilj V, et al. Commissioning of an ultra‐high dose rate pulsed electron beam medical LINAC for FLASH RT preclinical animal experiments and future clinical human protocols. Med Phys. 2021;48(6):3134–3142. [DOI] [PubMed] [Google Scholar]
- 7. Bourhis J, Sozzi WJ, Jorge PG, et al. Treatment of a first patient with FLASH‐radiotherapy. Radiother Oncol. 2019;139:18–22. [DOI] [PubMed] [Google Scholar]
- 8. Mascia AE, Daugherty EC, Zhang Y, et al. Proton FLASH radiotherapy for the treatment of symptomatic bone metastases: the FAST‐01 nonrandomized trial. JAMA Oncol. 2023;9(1):62–69. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9. Kranzer R, Poppinga D, Weidner J, et al. Ion collection efficiency of ionization chambers in ultra‐high dose‐per‐pulse electron beams. Med Phys. 2021;48(2):819–830. [DOI] [PubMed] [Google Scholar]
- 10. Gómez F, Gonzalez‐Castaño DM, Fernández NG, et al. Development of an ultra‐thin parallel plate ionization chamber for dosimetry in FLASH radiotherapy. Med Phys. 2022;49(7):4705–4714. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11. Konradsson E, Ceberg C, Lempart M, et al. Correction for ion recombination in a built‐in monitor chamber of a clinical linear accelerator at ultra‐high dose rates. Radiat Res. 2020;194(6):580–586. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12. Pardo‐Montero J, Paz‐Martín J, Brualla‐Gónzalez L, Gómez F. Online recombination correction in dosimetry with arrays of ionization chambers: application to FFF and UHDR radiotherapy. Appl Sci. 2021;11(21):10083–10097. [Google Scholar]
- 13. Liu K, Holmes S, Hooten B, Schüler E, Beddar S. Evaluation of ion chamber response for applications in electron FLASH radiotherapy. Med Phys. 2024;51(1):494–508. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14. Ahmed MF, Dogan SK, McNeur J, Tirpak L, Senadheera L. Ultra‐high dose rate 16 Mev electron beam dosimetry using advanced markus chamber in medical linacs. Med Phys. 2023;50(6):e320. [Google Scholar]
- 15. Romano F, Bailat C, Jorge PG, Lerch MLF, Darafsheh A. Ultra‐high dose rate dosimetry: challenges and opportunities for FLASH radiation therapy. Med Phys. 2022;49(7):4912–4932. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16. Gaza R, McKeever SWS, Akselrod MS. Near‐real‐time radiotherapy dosimetry using optically stimulated luminescence of Al2O3:C: mathematical models and preliminary results. Med Phys. 2005;32(4):1094–1102. [DOI] [PubMed] [Google Scholar]
- 17. Verona Rinati G, Felici G, Galante F, et al. Application of a novel diamond detector for commissioning of FLASH radiotherapy electron beams. Med Phys. 2022;49(8):5513–5522. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18. Poirier Y, Xu J, Mossahebi S, Therriault‐Proulx F, Sawant A. Technical note: characterization and practical applications of a novel plastic scintillator for online dosimetry for an ultrahigh dose rate (FLASH). Med Phys. 2022;49(7):4682–4692. [DOI] [PubMed] [Google Scholar]
- 19. Oh K, Gallagher KJ, Yan Y, Zhou S. Commissioning and initial validation of Eclipse eMC algorithm for the electron FLASH research extension (FLEX) system for pre‐clinical studies. J Appl Clin Med Phys. 2024;25(5):e14289. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20. Oh K, Gallagher KJ, Hyun M, et al. Initial experience with an electron FLASH research extension (FLEX) for the Clinac system. J Appl Clin Med Phys. 2024;25(2):e14159. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21. Cetnar AJ, Jain S, Gupta N, Chakravarti A. Technical note: Commissioning of a linear accelerator producing ultra‐high dose rate electrons. Med Phys. 2024;51(2):1415–1420. [DOI] [PubMed] [Google Scholar]
- 22. Jacqmin DJ, Miller JR, Barraclough BA, Labby ZE. Commissioning an Exradin W2 plastic scintillation detector for clinical use in small radiation fields. J Appl Clin Med Phys. 2022;23(8):e13728. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 23. Standard Imaging Inc. MAX SD User Manual (Middleton, WI, USA, 2022).
- 24. Thrower S, Prajapati S, Holmes S, Schüler E, Beddar S. Characterization of the plastic scintillator detector system Exradin W2 in a high dose rate flattening‐filter‐free photon beam. Sensors. 2022;22(18):6785. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 25. Petersson K, Jaccard M, Germond JF, Buchillier T, Bochud F. High dose‐per‐pulse electron beam dosimetry —A model to correct for the ion recombination in the Advanced Markus ionization chamber. Med Phys. 2017;44(3):1157–1167. [DOI] [PubMed] [Google Scholar]
- 26. Kokurewicz K, Schüller A, Brunetti E, et al. Dosimetry for new radiation therapy approaches using high energy electron accelerators. Front Phys. 2020;8:568302. [Google Scholar]
- 27. Jaccard M, Durán MT, Petersson K, et al. High dose‐per‐pulse electron beam dosimetry: commissioning of the Oriatron eRT6 prototype linear accelerator for preclinical use. Med Phys. 2017;45(2):863–874. [DOI] [PubMed] [Google Scholar]
- 28. Szpala S, Huang V, Zhao Y, et al. Dosimetry with a clinical linac adapted to FLASH electron beams. J Appl Clin Med Phys. 2021;22(6):50–59. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 29. Christensen JB, Togno M, Nesteruk KP, et al. Al2O3:C optically stimulated luminescence dosimeters (OSLDs) for ultra‐high dose rate proton dosimetry. Phys Med Biol. 2021;66(8):085003. [DOI] [PubMed] [Google Scholar]
- 30. Jaccard M, Petersson K, Buchillier T, et al. High dose‐per‐pulse electron beam dosimetry: usability and dose‐rate independence of EBT3 Gafchromic films: usability. Med Phys. 2017;44(2):725–735. [DOI] [PubMed] [Google Scholar]
- 31. Jorge PG, Jaccard M, Petersson K, et al. Dosimetric and preparation procedures for irradiating biological models with pulsed electron beam at ultra‐high dose‐rate. Radiother Oncol. 2019;139:34–39. [DOI] [PubMed] [Google Scholar]
- 32. Kudoh H, Celina IM, Kaye IRJ, Gillen KT, Clough’ RL. Response of alanine dosimeters at very high dose rate. Appl. Radiat. lsot. 1997;48(4):370–382. [Google Scholar]
- 33. Kranzer R, Schüller A, Gómez Rodríguez F, et al. Charge collection efficiency, underlying recombination mechanisms, and the role of electrode distance of vented ionization chambers under ultra‐high dose‐per‐pulse conditions. Phys. Med. 2022;104:10–17. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 34. Kranzer R, Schüller A, Bourgouin A, et al. Response of diamond detectors in ultra‐high dose‐per‐pulse electron beams for dosimetry at FLASH radiotherapy. Phys Med Biol. 2022;67(7):075002. [DOI] [PubMed] [Google Scholar]
- 35. Bourgouin A, Schüller A, Hackel T, et al. Calorimeter for real‐time dosimetry of pulsed ultra‐high dose rate electron beams. Front Phys. 2020;8:567340. [Google Scholar]
- 36. Romano F, Milluzzo G, Di Martino F, et al. First characterization of novel silicon carbide detectors with ultra‐high dose rate electron beams for FLASH radiotherapy. Appl Sci. 2023;13(5):2986. [Google Scholar]
- 37. Liu K, Palmiero A, Chopra N, et al. Dual beam‐current transformer design for monitoring and reporting of electron ultra‐high dose rate (FLASH) beam parameters. J Appl Clin Med Phys. 2023;24(2):e13891. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 38. Vanreusel V, Gasparini A, Galante F, et al. Point scintillator dosimetry in ultra‐high dose rate electron “FLASH” radiation therapy: a first characterization. Phys Med: Eur J Med Phys. 2022;103:127–137. [DOI] [PubMed] [Google Scholar]
- 39. Clift MA, Sutton RA, Webb DV. Water equivalence of plastic organic scintillators in megavoltage radiotherapy bremsstrahlung beams. Phys Med Biol. 2000;45(7):1885. [DOI] [PubMed] [Google Scholar]
- 40. Beddar AS, Mackie TR, Attix FH. Water‐equivalent plastic scintillation detectors for high‐energy beam dosimetry: I. Physical characteristics and theoretical considerations. Phys Med Biol. 1992;37(10):1883. [DOI] [PubMed] [Google Scholar]