Skip to main content
Journal of the Society for Cardiovascular Angiography & Interventions logoLink to Journal of the Society for Cardiovascular Angiography & Interventions
editorial
. 2024 Mar 22;3(4):101358. doi: 10.1016/j.jscai.2024.101358

“Steering” Toward Complete Left Atrial Appendage Closure

Safi U Khan 1, Sachin S Goel 1,
PMCID: PMC11308719  PMID: 39130184

Left atrial appendage closure (LAAC) has emerged as an important alternative for stroke prevention in patients with nonvalvular atrial fibrillation who are at increased risk for stroke but not good candidates for long-term anticoagulation.1,2 However, there are safety concerns related to device-related thrombus and peri-device leak (PDL), both of which are associated with increased stroke risk.1 Anatomical variations of the left atrial appendage (LAA) heighten the complexity of achieving desired outcomes. For instance, achieving coaxial alignment between the delivery sheath and the LAA neck is paramount, a task sometimes complicated by challenging LAA morphologies.1 Delivery sheaths used routinely are not deflectable or steerable. The advent of steerable delivery sheaths (SDS), offering adjustable distal end angles from 0° to 120°, provides a potential solution to navigate these anatomical challenges.3,4 Small series and case reports have highlighted the safety and potential for improved outcomes using SDS in LAAC.3, 4, 5 Chang et al4 assessed the feasibility of FlexCath Steerable Sheath (Medtronic) in implanting Watchman (Boston Scientific) and Amulet (Abbott) devices, achieving successful implantation with lower PDL rates through optimal coaxial alignment.

In this issue of JSCAI, Amabile et al6 provide further insights by exploring the utilization of SDS for LAAC with the Amulet device in patients with significantly enlarged left atria on computed tomography (LA volume index >90 mL/m2). This single-center study analyzed the procedural outcomes of 47 patients—22 using the SDS and 25 with a standard delivery sheath. The primary end point was procedural success (defined as the exclusion of LAA without any device-related complications, no leak >5 mm on intraprocedural color Doppler transesophageal echocardiogram, and no procedure-related complications), and secondary end points were the presence of disc/lobe misalignment, patent LAA, and PDL on postprocedural CT scan. Both groups exhibited similar demographics, comorbidities, and procedural characteristics in the analysis. The study demonstrated comparable procedural success and an absolute 46% reduction in residual patent LAA, a 48% decrease in PDL, and a 27% reduction in the risk of off-axis device position with SDS compared with the standard sheath. However, these findings should be approached with caution due to potential selection bias stemming from narrow inclusion criteria, the significant reduction—from 387 to 47—of eligible patients, and the possibility of attrition bias, as only 41 patients received a follow-up CT scan.

With these considerations in view, it is essential to contextualize these results within the broader landscape of the safety challenges of LAAC, especially regarding PDL. In the PROTECT-AF trial (WATCHMAN Left Atrial Appendage System for Embolic PROTECTion in Patients With Atrial Fibrillation), 40.9% of patients had PDL at 45 days, decreasing to 32.1% at 1 year, with leaks >3 mm affecting 13.3% initially and 11.8% at 1 year.7 The Amulet IDE trial reported 37% PDL at 45 days with the Amulet device, compared with 54% with the Watchman device; leaks >3 mm were detected in 10% and 25% of patients, respectively.8 The NCDR LAAO registry noted PDL in 26.6% of patients at 45 days, linking small leaks (<5 mm) to an increased risk of ischemic stroke and transient ischemic attack over 1 year.9 A large meta-analysis including 61,666 patients who underwent LAAC found that transesophageal echocardiography-detected PDL after LAAC was associated with an increased risk of thromboembolic adverse events. This risk had a positive graded association across cut-offs of PDL size (0, 1, 3, and 5 mm).10 These data highlight the significance of minimizing PDL in LAAC procedures, and the advances in new occluder devices and the introduction of SDS promise improvements in sealing the LAA, aiming for better procedural outcomes.1

Furthermore, a substantial proportion of the patients undergoing mitral transcatheter edge-to-edge repair (mTEER) have atrial fibrillation and face a high bleeding risk.3 Combining mTEER with LAAC is practical once access to the left atrium is achieved through a transeptal procedure. Typically, LAAC is performed using an inferoposterior transseptal puncture, whereas MitraClip (Abbott) procedures rely on a high and posterior puncture approach. The SDS technology could enable LAAC device placement through a higher puncture site, similar to that used for MitraClip, enhancing procedural feasibility and efficiency.7

In conclusion, this study by Amabile et al6 adds to the growing experience of SDS in successful device deployment for LAAC. This preliminary evidence paves the way for expanded research and clinical integration. Future studies should pursue broader, multicenter trials with extended follow-up and diverse populations to refine generalizability, focusing on different LAA anatomical variations and comorbidity profiles. Advancements in guidance tools and imaging, including 3D imaging and advanced navigation, are essential for navigating complex LAA anatomies. Collaborative efforts among specialists would be instrumental in integrating these innovations, steering toward safer, more effective LAAC strategies.

Acknowledgments

Declaration of competing interest

Sachin S. Goel is a consultant at Medtronic, JC Medical, WL Gore & Associates, and is on the speakers’ bureau at Abbott Structural Heart. Safi U. Khan reported no financial interests.

Funding sources

This work was not supported by funding agencies in the public, commercial, or not-for-profit sectors.

References

  • 1.Alkhouli M., Ellis C.R., Daniels M., Coylewright M., Nielsen-Kudsk J.E., Holmes D.R. Left atrial appendage occlusion: current advances and remaining challenges. JACC Adv. 2022;1(5) doi: 10.1016/j.jacadv.2022.100136. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2.Joglar J.A., Chung M.K., Armbruster A.L., et al. 2023 ACC/AHA/ACCP/HRS Guideline for the diagnosis and management of atrial fibrillation: a report of the American College of Cardiology/American Heart Association Joint Committee on Clinical Practice Guidelines. Circulation. 2024;149(1):e1–e156. doi: 10.1161/CIR.0000000000001193. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Maiani S., Nardi G., Ristalli F., Di Mario C., Meucci F. Use of a steerable delivery sheath to obtain coaxial alignment in left atrial appendage occlusion after mitral transcatheter edge-to-edge repair: a case report. Eur Heart J Case Rep. 2024;8(1):ytad592. doi: 10.1093/ehjcr/ytad592. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.Chang S.N., Chiu F.C., Chen J.J., et al. First-in-human experience of using a universal steerable sheath in implanting left atrial appendage closure devices. JACC Asia. 2022;2(6):780–783. doi: 10.1016/j.jacasi.2022.08.003. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Cruz González I., Antúnez Muiños P.J., López Tejero S., et al. Left atrial appendage occlusion using the novel Amplatzer steerable delivery sheath combine with FEops HEARTguide. JACC Cardiovasc Interv. 2021;14(21):e301–e304. doi: 10.1016/j.jcin.2021.08.055. [DOI] [PubMed] [Google Scholar]
  • 6.Amabile N., Belfekih A., Balmette V., Mahmoudi K., Mignot N., Roig C. Steerable delivery sheath for left atrial appendage closure in patients with severely enlarged left atria. J Soc Cardiovasc Angiogr Interv. 2024;3(4) doi: 10.1016/j.jscai.2023.101290. [DOI] [Google Scholar]
  • 7.Viles-Gonzalez J.F., Kar S., Douglas P., et al. The clinical impact of incomplete left atrial appendage closure with the Watchman device in patients with atrial fibrillation: a PROTECT AF (Percutaneous Closure of the Left Atrial Appendage Versus Warfarin Therapy for Prevention of Stroke in Patients With Atrial Fibrillation) substudy. J Am Coll Cardiol. 2012;59(10):923–929. doi: 10.1016/j.jacc.2011.11.028. [DOI] [PubMed] [Google Scholar]
  • 8.Lakkireddy D., Thaler D., Ellis C.R., et al. Amplatzer Amulet Left Atrial Appendage Occluder Versus Watchman Device for Stroke Prophylaxis (Amulet IDE): a randomized, controlled trial. Circulation. 2021;144(19):1543–1552. doi: 10.1161/CIRCULATIONAHA.121.057063. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9.Price M.J., Valderrábano M., Zimmerman S., et al. Periprocedural pericardial effusion complicating transcatheter left atrial appendage occlusion: a report from the NCDR LAAO registry. Circ Cardiovasc Interv. 2022;15(5) doi: 10.1161/CIRCINTERVENTIONS.121.011718. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10.Samaras A., Papazoglou A.S., Balomenakis C., et al. Residual leaks following percutaneous left atrial appendage occlusion and outcomes: a meta-analysis. Eur Heart J. 2023;45(3):214–229. doi: 10.1093/eurheartj/ehad828. [DOI] [PubMed] [Google Scholar]

Articles from Journal of the Society for Cardiovascular Angiography & Interventions are provided here courtesy of Elsevier