Abstract
Introduction
Arthrodesis of a (diseased) ankle joint is usually performed to achieve pain relief and stability. One basic principle of arthrodesis techniques includes rigid fixation of the surfaces until union. It seems plausible that stable anchoring and homogeneous pressure distribution should be advantageous, however, it has not been investigated yet. The aim is to achieve uniform compression, as this is expected to produce favorable results for the bony fusion of the intended arthrodesis. Numerous implants with different biomechanical concepts can be used for ankle fusion. In this study, headless compression screws (HCS, DePuy Synthes, Zuchwil, Switzerland) were compared biomechanically to an alternative fixation System, the IOFix device (Extremity Medical, Parsippany, NJ, USA) in regard to the distribution of the compression force (area of contact) and peak compression in a sawbone arthrodesis-model (Sawbones® Pacific Research Laboratories, Vashon, WA, USA).
This study aims to quantify the area of contact between the bone interface that can be obtained using headless compression screws compared to the IOFix. In current literature, it is assumed, that a large contact surface with sufficient pressure between the bones brings good clinical results. However, there are no clinical or biomechanical studies, that describe the optimal compression pressure for an arthrodesis.
Material and methods
Two standardized sawbone blocks were placed above each other in a custom-made jig. IOFix and headless compression screws were inserted pairwise parallel to each other using a template for a uniform drilling pattern. All screws were inserted with a predefined torque of 0.5 Nm. Pressure transducers positioned between the two sawbone blocks were compressed for the measurement of peak compression force, compression distribution, and area of contact.
Results
With the IOFix, the compression force was distributed over significantly larger areas compared to the contact area of the HCS screws, resulting in a more homogenous contact area over the entire arthrodesis surface. Maximum compression force showed no significant difference.
Conclusion
The IOFix system distributes the compression pressure over a much larger area, resulting in more evenly spread compression at the surface. Clinical studies must show whether this leads to a lower pseudarthrosis rate.
Keywords: IOFix, Headless compression screws, Area of contact, Arthrodesis ankle joint, Compression force, Biomechanics
Introduction
Arthrodesis of the ankle or foot joints (i.e. triple arthrodesis, transverse tarsal joint arthrodesis) is commonly performed in the operative treatment of symptomatic, end-stage osteoarthritis and/or acquired flatfoot deformity. These conditions may be posttraumatic/degenerative, malformation- or infect-related. The fusion of a diseased joint is performed to achieve pain relief and stability. One frequent complication following ankle arthrodesis is non-union/pseudarthrosis. In current literature, variable union rates in ankle arthrodesis following open or arthroscopic approaches with different fixation techniques have been reported. Nonunion rates of ankle arthrodesis range from 2 to 47% [1–5]. Non-union of the ankle arthrodesis can be caused by various factors. These include patient-specific problems, such as smoking, arteriosclerosis, or diabetes, and can not be influenced by the surgeon. Nevertheless, factors that can be influenced by the surgeon for good and satisfactory clinical results include the surgical technique to perform an arthrodesis and the selection of an adequate implant to achieve compression. Another factor that could be of importance in the occurrence of pseudarthrosis is the failure to obtain and maintain compression across the fusion during the healing process due to implant failure. [6, 7]
One basic principle of arthrodesis includes, among others, rigid fixation of the surfaces until union, which can be achieved using variable fixation devices with different biomechanical properties [8, 9]. Numerous biomechanical studies investigated the stiffness of different implants used for ankle fusion [10–13]. In these studies, an arthrodesis was performed or simulated in either sawbone models or cadaver specimens with different implant constructs (double plating versus screws; external fixator versus crossed-screw; compressed external fixator versus an uncompressed interlocking nail versus a compressed interlocking nail versus two different three-screw techniques; External ring fixation versus screw fixation) and were biomechanically compared. Using material testing machines, their focus was on investigating the load to failure (N), stiffness of the constructs (N/mm), the ultimate bend, torque, and compression as well as quantifying the bending and torsional stiffness of each arthrodesed joint.
Additionally, some biomechanical studies focused on screw design and their mechanical performance or their pullout strength [1, 14, 15]. In other biomechanical studies, the static performance of different implants and their properties (thread length, cannulation, etc.) were investigated. The overall goal was to determine how screw thread geometry, tapping, and cannulation affect the holding power of screws in cancellous bone and determine whether current designs achieve maximum holding strength.
Burchard et al. and Roth et al. investigated intramedullary fixation in arthrodesis of the 1st metatarsophalangeal joint. They aimed to analyze the load to failure, cycles to failure, and stiffness compared to traditional fusion methods, such as plates and screws [16, 17]. However, the optimal compression force in an arthrodesis was not investigated.
Due to the design of different screws, the compression force can reach variable peaks during implantation, but more importantly, they may spread the compression more uniformly and thus produce a larger contact area at the arthrodesis site of the ankle joint [18]. Implant design has an influence on mechanical properties e.g. compression force [19].
So far, there are no biomechanical studies that describe the optimal pressure for arthrodesis or osteosynthesis. Insufficient contact surface and lack of compression can be the cause of pseudarthrosis [20, 21]. It is therefore important to investigate implants with different properties regarding their ability to achieve compression and contact surface. It is assumed that a higher compression force at the arthrodesis site and greater distribution of this force (contact area) could have a positive effect on the outcome and will therefore be investigated in this study.
Material and methods
Implants and test set up
For testing, two types of screws, which are commonly used for fixation of arthrodesis of the ankle joint, with different compression principles, were selected for this biomechanical analysis. Included were Headless Compression Screws (HCS, DePuy Synthes, Zuchwil, Switzerland) with two different diameters (4.5 mm and 6.5 mm), and the IOFix system in three sizes (Extremity medical, Parsippany, NJ, USA). Both screw-types are shown in Fig. 1. The IOFix system is a fixed-angle device, comprising two screws. The first screw called the X-post, is inserted parallel to the arthrodesis plane. The second screw is a locking screw, which is inserted through the prefabricated eyelet of the X-post’s head at a 60° angle, shown in Fig. 2.
In total, 5 groups were biomechanically tested and compared to each other: (1) 4.5 mm HCS screws, (2) 6.5 mm HCS screws, (3) IOFix small with the diameters X-Post 6.6 mm and locking screw 4.0 mm, (4) IOFix medium with diameters X-Post 8.0 mm and locking screw 5.0 mm, and (5) IOFix large with diameters X-Post 9.5 mm and locking screw 6.5 mm. The description of the HCS and IOFix screws is shown in Table 1a and b.
Table 1.
Group | Pilot hole | Length (L), thread length (TL), thread pitch increments (TP) |
---|---|---|
(a) Description of the Headless Compression screws | ||
HCS 4.5 mm | 3.2 mm | L 50 mm, TL 20 mm, TP 2 mm increments |
HCS 6.5 mm | 5.0 mm | L 50 mm, TL 16 mm, TP 5 mm increments |
Group | X post: diameter, pilot hole, length |
Locking screw: Diameter (D), length (L), thread length (TL), thread pitch increments (TP) |
---|---|---|
(b) Description of the IOFix screws | ||
IOFix small | 6.6 mm, 3.4 mm, 30 mm | D 4.0 mm, L 50 mm, TL 15 mm, TP 5 mm increments |
IOFix medium | 8.0 mm, 4.5 mm, 30 mm | D 5.0 mm, L 50 mm, TL 15 mm, TP 5 mm increments |
IOFix large | 9.5 mm, 4.5 mm, 30 mm | D 6.5 mm, L 50 mm, TL 16 mm, TP 5 mm increments |
For the experimental setup shown in Fig. 3—which is based on the set up by Mueller et al. [22]—two stacked sawbone blocks (Sawbones® Pacific Research Laboratories, Vashon, WA, USA) were stabilized in a custom-made container to prevent the blocks from tilting. Two pressure sensor transducers (K-Scan 4000, Tekscan, Inc., Boston, MA, USA) were placed between the sawbone blocks and held in place by the clamps attached to each side of the container, to prevent the sensors from slipping.
A template was positioned on the upper sawbone to ensure a standardized uniform height and distance between the screws to be inserted. The K-wires were inserted through the prefabricated holes of the template (with a defined height of 20 mm to the arthrodesis plane and 20 mm distance between the screws) in the template. After removing the template, the screw holes were pre-drilled by over-drilling the guidewires with cannulated drill bits. According to the manufacturer’s instructions, the pilot hole for the 4.5 mm HCS screws was drilled using a 3.2 mm cannulated drilling bit and a 5.0 mm cannulated drilling bit for the 6.5 mm HCS screws.
For the IOFix system, the pilot hole for the X-post is predrilled—according to the manufacturer’s manual—with 2.0 mm/3.4 mm/4.5 mm cannulated drill bits, respectively for IOFix small/medium/large. After insertion of the X-post parallel to the arthrodesis plane, the locking screw is inserted in the eyelet of the X-post’s head and is then passed across the arthrodesis site with a 60° angulation. The lag screw gets engaged in the eyelet of the X-post, resulting in a more uniform compression across the fusion site. [18] Three different sizes of the IOFix device were tested—IOFix small, medium, and large (Table 1).
For the respective group, two identical HCS screws and two identical IOFix devices were inserted pairwise parallel to each other in a straight pattern. A predefined torque was applied to all screws (0.5 Nm) simultaneously to their insertion, which was detected by using a digital screwdriver with an integrated digital sensor (PCE-TM 80, PCE GmbH, Meschede, Germany). Rotational stability was given for all five groups.
Seven measurements per group (n = 35) were performed on pre-assembled saw bone blocks with a density of 15 pounds per cubic foot (solid foam; personalized blocks). The compression force at the arthrodesis site was measured and recorded using two flexiforce pressure transducers, which were connected to a measuring device (Fig. 4). The pressure sensors were calibrated to display compression force and the contact area—which is created when the blocks are pressed together—in Megapascal (MPa) and in square millimeters (mm2).
Each (articulating) surface of the sawbone block has a surface area of 832 mm2 (26 × 32 mm), as seen in Fig. 5. Because in this study an arthrodesis is imitated, lamination of the sawbone blocks to simulate the base plate was omitted.
A calibration was carried out at the beginning of each series of measurements. After final tightening of the screws, the pressure transducer was initiated to start the digital recording of the compression force and measurement of the force distribution (area of contact). This was achieved by the pressure-sensitive films being positioned within the arthrodesis and, therefore, active to determine the average contact area over a time span of 10 s.
Data analysis and statistics
A post hoc power analysis using G*Power (version 3.1.9.7; Franz Paul, Kiel, Germany) was performed to determine the power of the study. Based on the results of the F-test and ANOVA, an effect size of 0.90 was calculated. With this effect size, an α of 0.05, and the sample size of 35, a power of 0.98 was calculated.
For statistical analysis, all further data processing was performed with MATLAB (MathWorks Inc., Natick, MA, USA). Peak pressure was calculated by averaging the values which were measured in a time span of 10 s over a window with the eight surrounding values, to avoid artifacts (according to [23]). The contact area reflected the number of pixels that had a pressure value greater than 0 MPa.
Quantitative parameters are given as mean, median, and standard deviation, and were then calculated for the seven sawbones per screw-pair. Statistical analysis was performed using SPSS (IBM SPSS Statistics 27). In order to compare the HCS 4.5 mm/HCS 6.5 mm groups with the IOFix small/medium/large groups, a test for variance homogeneity was calculated first, followed by an analysis by means of one-factor analysis of variance (ANOVA). The results were presented with boxplots. The significance level of p < 0.05 was accepted as a statistically significant difference.
Results
Compression force
With the predefined torque of 0.5 Nm for every inserted screw, there is no significant difference in the maximum generated compression force (peak compression), when comparing the HCS 4.5 mm, HCS 6.5 mm, IOFix small and IOFix medium.
However, when comparing the peak compression force that was generated by the IOFix large device, compared to either HCS 4.5 mm or HCS 6.5 mm screws, significant differences were detected: While the HCS 4.5 mm screws generated 0.22 MPa and the HCS 6.5 mm 0.41 MPa of compression force, the IOFix large device produced a compression force of 0.84 MPa, being significantly larger than the compression obtained by HCS 4.5/6.5 mm (p < 0.001, p < 0.012). The results are shown in Table 2.
Table 2.
Screw | Mean ± Standard deviation | p-value | |
---|---|---|---|
HCS 4,5 mm | HCS 6,5 mm | ||
HCS 4,5mm | 0.22 (± 0.17) | - | n.s |
HCS 6,5mm | 0.41 (± 0.21) | n.s | - |
IOFix small | 0.36 (± 0.09) | n.s | n.s |
IOFix medium | 0.41 (± 0.25) | n.s | n.s |
IOFix large | 0.84 (± 0.09) | < 0.001 | < 0.012 |
Statistical analysis of the mean peak pressure is demonstrated: It shows that, there is no significant difference between the peak pressures produced by the HCS 4.5 mm/6.5 mm screws and the IOFix small and medium. However, there is a significant difference in mean peak pressures when IOFix large is implanted (p < 0.001, p < 0.012), compared to mean peak pressures of HCS 4.5/6.5 mm
Distribution of the force (Area of contact)
In the group of HCS 4.5 mm screws, the mean contact area between the saw bone blocks measured 159.5 mm2, which made up 19% of the surface area (Fig. 6a). In the group of HCS 6.5 mm, the mean contact area increased to 171.4 mm2 (20% of the max. surface area). Testing the IOFix small, there is a disproportionately large increase of the contact area, compared to the contact area produced by 6.5 mm HCS screws—an area of 420.6 mm2 of the pressure sensor pixels were activated, which makes up 50% of the surface (Fig. 6b).
Comparing HCS 4.5/6.5 mm screws to IOFix small/medium or large, significant differences could be detected (p = 0.012, p = 0.022; p = 0.022, p = 0.037; p = < 0.001, < 0.011, respectively for HCS 4.5 mm and HCS 6.5 mm), shown in Table 3.
Table 3.
Screw | Mean ± Standard deviation | p-value | |
---|---|---|---|
HCS 4,5 mm | HCS 6,5 mm | ||
HCS 4,5mm | 159.46 (± 111.65) | – | n.s |
HCS 6,5mm | 171.41 (± 78.16) | n.s | – |
IOFix small | 420.57 (± 173.50) | 0.012 | 0.022 |
IOFix medium | 403.21 (± 194.81) | 0.022 | 0.037 |
IOFix large | 660.58 (± 108.19) | < 0.001 | < 0.011 |
The number of activated pixels (mm2) of the pressure sensor foils are shown: A comparison of HCS screws with IOFix screws shows that there is a statistical significance between HCS 4.5 mm to IOFix small/medium/large and HCS 6.5 mm to IOFix small/medium/large. Therefore, the area of contact is greater when implanting IOFix systems, compared to HCS 4.5/6.5 mm screws
Discussion
This study evaluated the influence of different implants on the compression force and area of contact in a sawbone-based ankle-arthrodesis model. Regarding the contact area, the main finding of this study was that the IOFix system spreads the compression force more evenly over the surface, creating a significantly larger contact area between the sawbones at the arthrodesis site than two parallel inserted HCS screws, which is visualized in Figs. 6a and b. The mechanism of IOFix is the additional implantation of the X-Post prior to inserting the angular stable screw, which distributes compression forces across a greater surface area. One further finding of this study was, that the IOFix large also achieved higher values for the compression pressure than HCS screws did.
From these flexiforce pressure transducer foils, two important parameters can be determined. The first parameter indicates how high the maximum built-up compression on the specific single pixel is (peak compression). The second parameter displays how many pixels surrounding the inserted screw are activated (area of contact). There is no scientific parameter that represents the biomechanic situation at the arthrodesis gap better than the activated pixels, i.e. the area of contact.
Since compression of only a small area of the overall surface or also too much compression on this partial surface are risk factors for the development of pseudarthrosis, an ideal arthrodesis should have moderate compression with an even distribution across the matched bone surfaces, to minimize stress at areas of high peak contact, as well as neutralize shear and bending forces. By avoiding uneven compression across imperfectly matched surfaces, areas of high peak contact stress are minimized and could reduce the risk of bone resorption by osteoclasis, failure of fixation, and non-union [8].
Compared to the IOFix screws, which achieved a larger and more uniform pressure distribution across the arthrodesis, 4.5 mm HCS, and 6.5 mm HCS screws only showed localized compression by activating pixels of the pressure sensors around the screw insertion point. The HCS screws tended to concentrate a high peak pressure nearest where they were inserted. This stress may affect bone resorption. Bone resorption in areas of high peak contact stress within an arthrodesis may lead to progressive loss of bone interdigitation, gapping, and non-union at the interface [24]. It is therefore important to know the biomechanical properties of the implants used.
In current literature, there is a total of nine studies that investigate the clinical outcome and biomechanical properties of the IOFix device on foot and ankle joints [16, 17, 25–31]. Seven of these publications report on the fusion of the 1st metatarsophalangeal joint, one case series reports on the fusion of the talonavicular joint, while there is only one biomechanical study performed on cadaveric ankle joints by Parker et al.[18, 31]
Parker et al. [18] had similar findings to those in this study, though, in their study, they compared the IOFix to one single 6.5 mm, partially-threaded, cannulated cancellous lag-screw with a washer in cadaveric ankle joints. However, it needs to be considered that a cadaver study may not necessarily supply reproducible results. In addition, it would not be recommendable to use only one single screw in an ankle-arthrodesis, as a large area has to be treated and there is no rotational stability given. Furthermore, it also remains unclear whether there are (statistically significant) differences between the IOFix devices, if the devices have different sizes (like small, medium, and large). This study provides better comparable results with the variables being kept the same and standardized because individual differences like bone density, age, gender, etc. which are typical for cadaver studies are not being accounted for.
Furthermore, in this study, three different sizes of the IOFix screws (IOFix small/medium/large) and two different sizes of the HCS screws (HCS 4.5 mm and 6.5 mm) were tested and compared to each other, using artificial saw bone blocks, each with the same density and the same size of possible contact area. A template for drilling to guarantee a standardized test set-up was used. Due to the test setup and the comparison of different sizes/diameters of each screw type, the results can also be transferred to smaller or larger joints.
Using sawbones with the same density could be a potential advantage, as the variable, age-dependent bone density of cadavers did not have to be taken into account for evaluation. Also, this may be relevant for the possible influence the bone quality might have on screw stripping. In optimal osseous settings, the differences of the large IOFix screws may be significant in a sawbone model, however, measurements might not be advantageous in the clinical setting. As mentioned before, bone quality must be taken into account in the surgeon’s preoperative planning, e.g. cystic lesions should be filled when in doubt. Another important difference to a cadaver-based study is that the articulating saw-bone planes were flat and perfectly aligned to each other, limiting the risk of an uneven arthrodesis plane/malalignment. This test setup is an idealized arthrodesis model, which can easily be recreated and the results are reproducible. With this test setup, the biomechanical properties of different screws can be tested and compared to each other.
Nevertheless, this study has several limitations. As with all biomechanical studies, it has the inherent problem that only time-zero compression force can be evaluated, which represents the intraoperative situation. The initial period of clinical healing, though, is the most important and this was measured in this study. Many other factors are also important, especially in the early stages of clinical healing, such as insufficient compression or inadequate immobilization [22].
What is important for reproducible biomechanical results, is at the same time a limitation of this study. It is a sawbone test setup, not resembling anatomical variation. This construct was not subject to physiological loading, including compression and shear.
Furthermore, the size of the IOFix large screws could be relevant- in case of a revision surgery due to lack of bony healing, less bone stock is present. Finally, the drilling and insertion of the screws were standardized in the laboratory, not accounting for any technical and visual problems that might occur during drilling in the operating room. However, the aim of this study was to compare different implants used for ankle arthrodesis in terms of their ability to create a contact surface between the bones and that the test setup can provide reproducible results.
Conclusion
The aim of this biomechanical study was to demonstrate that the tightening of various implants builds up different levels of compression and may activate a larger or smaller area of the pressure sensor, which is positioned in an arthrodesis gap. This study demonstrated that the distribution of forces and the activated area surrounding the arthrodesis screw are dependent on the implant used: two parallel inserted IOFix systems distribute the compressive forces across a significantly greater surface area, leading to an optimized distribution of compression forces at the arthrodesis-site. Also, they show a larger area in force distribution compared to two parallel HCS screws. This could offer clinical advantages. Clinical studies must show whether the use of specific implants can reduce the rate of pseudarthrosis.
Acknowledgements
Not applicable
Author contributions
AW—carried out the experiment, wrote the manuscript with support from ACK and AMK. MK—carried out the experiment, wrote the manuscript with support from AW. LB—carried out the experiment, performed the analytic calculations. ACK—wrote the manuscript with support from AW and AMK. AMK—wrote the manuscript with support from AW and ACK. BMH—involved in planning, supervised the work, interpretation of the data. BW—investigated the implants and supervised the findings of this work. All authors provided critical feedback, helped shape the research, analysis and manuscript, discussed the results and contributed to the final manuscript. All authors approve the submitted version of the manuscript.
Funding
Open Access funding enabled and organized by Projekt DEAL. This research received no external funding. The sawbone blocks were financed by house internal funds.
Data availability
The data presented in this study are available on request from the corresponding author.
Declarations
Competing interests
The authors declare they have no competing interests.
Footnotes
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
References
- 1.DeCoster TA, Heetderks DB, Downey DJ, Ferries JS, Jones W. Optimizing bone screw pullout force. 1990;4(2): 169–74. [DOI] [PubMed]
- 2.Yasui Y, Hannon CP, Seow D, Kennedy JG. Ankle arthrodesis: a systematic approach and review of the literature. World J Orthop. 2016;7(11):700–8. 10.5312/wjo.v7.i11.700 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.Park JH et al. Arthroscopic versus open ankle arthrodesis: a systematic review. Arthroscopy J Arthroscopic Related Surg 2018;34(3): 988–997. [DOI] [PubMed]
- 4.Manke E, Yeo Eng Meng N, Rammelt S. Ankle Arthrodesis—a review of current techniques and results. Acta Chir Orthop Traumatol Cech 2020;87(4):225–236. Artrodéza hlezna – přehled současných technik a výsledků. [PubMed]
- 5.Ferguson Z, Anugraha A, Janghir N, Pillai A. Ankle arthrodesis: a long term review of the literature. J Orthopaedics 2019;16(5): 430–433. [DOI] [PMC free article] [PubMed]
- 6.Clain MR, Baxter DE. Simultaneous calcaneocuboid and talonavicular fusion. Long-term follow-up study. J Bone Joint Surg Br 1994;76(1): 133–6. [PubMed]
- 7.Mann RA, Beaman DN. Double arthrodesis in the adult. Clin Orthop Relat Res. 1999;365:74–80. 10.1097/00003086-199908000-00010 [DOI] [PubMed] [Google Scholar]
- 8.Cunningham JL, Richardson JB, Soriano RM, Kenwright J. A mechanical assessment of applied compression and healing in knee arthrodesis. Clin Orthop Relat Res. 1989;242:256–64. 10.1097/00003086-198905000-00026 [DOI] [PubMed] [Google Scholar]
- 9.Charnley JC. Positive pressure in arthrodesis of the knee joint. J Bone Joint Surg Br. 1948;30b(3): 478–86. [PubMed]
- 10.Betz MM, Benninger EE, Favre PP, Wieser KK, Vich MM, Espinosa N. Primary stability and stiffness in ankle arthrodes-crossed screws versus anterior plating. Foot Ankle Surg. 2013;19(3):168–72. 10.1016/j.fas.2013.04.006 [DOI] [PubMed] [Google Scholar]
- 11.Ogut T, Glisson RR, Chuckpaiwong B, Le IL, Easley ME. External ring fixation versus screw fixation for ankle arthrodesis: a biomechanical comparison. Foot Ankle Int. 2009;30(4):353–60. 10.3113/FAI.2009.0353 [DOI] [PubMed] [Google Scholar]
- 12.Hoover JR, Santrock RD, James WC. Ankle fusion stability: a biomechanical comparison of external versus internal fixation. Orthopedics 2011;34(4). [DOI] [PubMed]
- 13.Mueckley TM, et al. Biomechanical evaluation of primary stiffness of tibiotalar arthrodesis with an intramedullary compression nail and four other fixation devices. Foot Ankle Int. 2006;27(10):814–20. 10.1177/107110070602701011 [DOI] [PubMed] [Google Scholar]
- 14.Chapman JR, Harrington RM, Lee KM, Anderson PA, Tencer AF, Kowalski D. Factors affecting the pullout strength of cancellous bone screws. J Biomech Eng. 1996;118(3):391–8. 10.1115/1.2796022 [DOI] [PubMed] [Google Scholar]
- 15.Muller M, Abel EW, McLeod G, Rowley DI. Pull-out strengths for a range of screws inserted into the calcaneus: a preliminary study. Clin Biomech. 1992;7(2):125–8. 10.1016/0268-0033(92)90026-Z [DOI] [PubMed] [Google Scholar]
- 16.Burchard R et al. Biomechanics of common fixation devices for first tarsometatarsal joint fusion-a comparative study with synthetic bones. J Orthop Surg Res. 2018; 13(1): 176. [DOI] [PMC free article] [PubMed]
- 17.Roth KE, Peters J, Schmidtmann I, Maus U, Stephan D, Augat P. Intraosseous fixation compared to plantar plate fixation for first metatarsocuneiform arthrodesis: a cadaveric biomechanical analysis. Foot Ankle Int. 2014;35(11):1209–16. 10.1177/1071100714547082 [DOI] [PubMed] [Google Scholar]
- 18.Parker L, Ray P, Grechenig S, Grechenig W. Does the IOFIX improve compression in ankle fusion? Foot Ankle Surg. 2014;20(4):258–61. 10.1016/j.fas.2014.06.004 [DOI] [PubMed] [Google Scholar]
- 19.Brown GA, McCarthy T, Bourgeault CA, Callahan DJ. Mechanical performance of standard and cannulated 4.0-mm cancellous bone screws. 2000; 18(2): 307–312. [DOI] [PubMed]
- 20.Cichero MJ, Yates BJ, Joyce ASD, Williamson DM, Walsh TP. Different fixation constructs and the risk of non-union following first metatarsophalangeal joint arthrodesis. Foot Ankle Surg. 2021;27(7):789–92. 10.1016/j.fas.2020.10.006 [DOI] [PubMed] [Google Scholar]
- 21.Cho DC, Sung JK. Analysis of risk factors associated with fusion failure after anterior odontoid screw fixation. Spine. 2012; 37(1): 30–4. [DOI] [PubMed]
- 22.Muller JU et al. Biomechanical comparison of three different compression screws for treatment of odontoid fractures evaluation of a new screw design. Clin Biomech 2020;77: 105049. [DOI] [PubMed]
- 23.Li G, DeFrate LE, Zayontz S, Park SE, Gill TJ. The effect of tibiofemoral joint kinematics on patellofemoral contact pressures under simulated muscle loads. J Orthop Res. 2004;22(4):801–6. 10.1016/j.orthres.2003.11.011 [DOI] [PubMed] [Google Scholar]
- 24.Parker L, Singh D. (i) The principles of foot and ankle arthrodesis. Orthopaedics Trauma 2009;23(6): 385–394.
- 25.Benjamin B, Ryan P, Chechelnitskaya Y, Bayam L, Syed T, Drampalos E. Intraosseous device for arthrodesis in foot and ankle surgery: review of the literature and biomechanical properties. World J Orthop. 2021;12(12):1036–44. 10.5312/wjo.v12.i12.1036 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 26.Segal D, Ohana N, Nyska M, Palmanovich E. Does the IOFix implant improve union rates? BMC Musculoskelet Disord 2020;21(1): 654. [DOI] [PMC free article] [PubMed]
- 27.Patel S, Garg P, Fazal MA, Ray PS. First metatarsophalangeal joint arthrodesis using an intraosseous post and lag screw with immediate bearing of weight. J Foot Ankle Surg. 2019;58(6):1091–4. 10.1053/j.jfas.2019.01.006 [DOI] [PubMed] [Google Scholar]
- 28.Singhal R, Kwaees T, Mohamed M, Argyropoulos M, Amarasinghe P, Toh EM. Result of IOFIX (Intra Osseous FIXation) device for first metatarsophalangeal joint arthrodesis: a single surgeon’s series. Foot Ankle Surg. 2018;24(5):466–70. 10.1016/j.fas.2017.05.003 [DOI] [PubMed] [Google Scholar]
- 29.Drampalos E, Vun SH, Bayam L, Fayyaz I. Early results of an intraosseous device for arthrodesis of the hallux metatarsophalangeal joint. Indian J Orthop. 2017;51(3): 299–303. [DOI] [PMC free article] [PubMed]
- 30.Drampalos E, Vun SH, Fayyaz I. Intramedullary and intra-osseous arthrodesis of the hallux metatarsophalangeal joint. J Orthop Surg. 2016;24(3):358–61. 10.1177/1602400317 [DOI] [PubMed] [Google Scholar]
- 31.Shymon SJ, Moss L, Harris TG. Case series using a novel implant and accelerated rehabilitation for patients undergoing an isolated talonavicular arthrodesis. Foot Ankle Spec. 2016;9(3):227–31. 10.1177/1938640015620638 [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Data Availability Statement
The data presented in this study are available on request from the corresponding author.