Skip to main content
Clinical Mass Spectrometry logoLink to Clinical Mass Spectrometry
. 2018 Aug 17;10:25–32. doi: 10.1016/j.clinms.2018.08.004

Urine analysis of 28 designer benzodiazepines by liquid chromatography–high-resolution mass spectrometry

Madeleine Pettersson Bergstrand a,b,, Olof Beck a,c, Anders Helander a,b,c
PMCID: PMC11322785  PMID: 39193360

Highlights

  • An LC-HRMS method covering 28 designer benzodiazepines was developed.

  • The method included screening in FS (gradient run 3.5 min).

  • Positive samples were confirmed using MS/MS (gradient run 4.5 min).

  • The method was applied to patient samples from emergency wards and routine drug testing.

  • Around 30% of the patient samples contained an analyte covered in the LC-HRMS method.

Keywords: Designer benzodiazepines, LC–HRMS, Method validation, New psychoactive substances, Screening and confirmation

Abstract

Hundreds of new psychoactive substances (NPS) covering most drugs-of-abuse classes have been introduced to the recreational drug market in recent years. One class of NPS drugs that has become more common recently is “designer” benzodiazepines. Due to a close structural resemblance with prescription benzodiazepines, some of these substances may elicit a positive response (i.e. cross react) in immunoassay screening. Consequently, it is increasingly important to include NPS benzodiazepines during method confirmation to ensure accurate identification of closely-related compounds as well as detection of the benzodiazepines themselves.

Here, we present our efforts to develop a screening and confirmation method for detection of 28 NPS benzodiazepines in urine using reversed-phase liquid chromatographic separation in combination with high-resolution mass spectrometry (LC–HRMS). MS was performed in positive electrospray mode on a Thermo Fischer Scientific Q Exactive Orbitrap instrument using either full scan (for screening) or parallel reaction monitoring (for confirmation).

We found the lower quantification limit of the method to range from 5 to 50 ng/mL. Analytical precision and accuracy were ≤15% for both screening and confirmation for all except one analyte. The method was used to analyze patient urine samples from routine drug testing and samples from acute intoxication cases presenting in emergency wards. Altogether, 16 of the 28 benzodiazepines (i.e., clobazam, clonazolam, deschloroetizolam, diclazepam, estazolam, etizolam, flubromazepam, flubromazolam, flunitrazolam, 3-hydroxyflubromazepam, 3-hydroxyphenazepam, ketazolam, meclonazepam, metizolam, nifoxipam, and pyrazolam) were detected in the urine samples.

The results from patient sample analysis indicate a high prevalence of NPS benzodiazepine use, emphasizing the importance of including novel drugs of abuse in drug testing menus.

1. Introduction

In recent years, hundreds of new psychoactive substances (NPS) have been introduced to the recreational drug market [1]. NPS are typically variants of classical drugs of abuse that have been structurally modified (i.e., “designed”) to circumvent legislative control and allow open sales online [2]. All common drug classes, including cannabinoids, stimulants, tryptamines, opioids and benzodiazepines, have experienced this phenomenon [2], [3]. Because NPS differ in chemical structure from their classical counterparts, intake may go undetected if standard drug testing by immunoassay screening is used. Additionally, NPS experience a high structural turnover on the market [3], requiring analytical methods be updated continuously to address the latest developments. Liquid chromatographic separation with mass spectrometric detection (LC–MS) provides an analytical methodology that allows the flexibility to reasonably maintain a relevant drug testing menu that will address the rapidly changing landscape of NPS [4].

In the Swedish STRIDA project on NPS [2], one drug class that has become more regularly used is benzodiazepines. Benzodiazepines are prescription drugs for treating anxiety, for example, but are also commonly abused due to their propensity to cause dependence, tolerance and addiction [5]. Of the benzodiazepines introduced as NPS on the EU drug market, some are already registered medicines outside Europe, whereas others are novel, designed variants [6]. The first NPS benzodiazepine of the latter category was pyrazolam, which appeared for sale online in 2012 [6]; it has since been followed by many more [7], [8], [9], [10], [11].

Due to structural similarity, and subsequent antibody cross-reactivity, NPS benzodiazepines are often detectable with immunoassays that target the standard set of prescription benzodiazepines [12], [13]. Accordingly, to avoid “false positive” screening results, confirmation methods should also include these NPS. Recently published methods for NPS benzodiazepines have used e.g. MS detection coupled to ultra-high performance LC, capillary electrophoretic, or gas chromatographic separation [14], [15], [16], [17]. However, an additional analytical problem with the NPS benzodiazepines is that some of these are precursors to, or metabolites of, prescription benzodiazepines [10], [18], which places high demands on the final interpretation of test results.

Here, we have undertaken to develop an LC–high-resolution MS (LC–HRMS) multi-component method for determination of designer benzodiazepines in urine, using direct dilution as the only sample preparation. This offers a sensitive, selective and flexible method and enables rapid inclusion of new analytes. The method is based on a published method for screening (using LC–HRMS) and confirmation (using LC–HRMS/MS) of several different NPS classes [19]. The benzodiazepines were selected for this study based on (i) frequency of mention in drug discussion forums, (ii) presence in the online marketplace, (iii) notification by the European Monitoring Centre for Drugs and Drug Addiction (EMCDDA), and (iv) not being prescribed in Sweden (i.e. NPS, deregistered pharmaceuticals, or pharmaceuticals approved in other countries) [6], [7], [8], [9], [10], [11], [20], [21], [22], [23], [24], [25], [26]. Method performance was evaluated on patient urine samples from routine drug testing and acute intoxication cases suspected to involve NPS benzodiazepines (STRIDA project cases).

2. Materials and methods

2.1. Chemicals

Adinazolam, bentazepam, clonazolam, cloniprazepam, diclazepam, deschloroetizolam, flubromazolam, flunitrazolam, 3-hydroxyflubromazepam, ketazolam, meclonazepam, metizolam, nifoxipam, nitrazolam, pivoxazepam and pyrazolam were obtained as 1 mg/mL standard solutions from Chiron (Trondheim, Norway). Bromazepam, clobazam, etizolam, estazolam, flurazepam, α-hydroxytriazolam-d4 (internal standard, IS), 3-hydroxyphenazepam, lorazepam-d4 (IS), N-desmethylflunitrazepam, nimetazepam, phenazepam and prazepam were obtained as 1 mg/mL solutions, and α-hydroxyalprazolam-d5, estazolam-d5, flunitrazepam-d7, nordiazepam-d5, oxazepam-d5 and temazepam-d5 (all IS) as 0.1 mg/mL solutions, from Cerilliant (Round Rock, TX, USA). Flubromazepam was obtained as a 1 mg/mL solution and tetrazepam as a 0.1 mg/mL solution from LGC (Teddington, UK). All solutions were obtained in acetonitrile or methanol.

Ammonium formate (LC–MS ultra grade), E. coli β-glucuronidase solution (80 U/mL at 25 °C), ammonia solution (25%), acetic acid (pro analysis grade) and formic acid (MS grade) were from Sigma–Aldrich (St. Louis, MO, USA), acetonitrile (HPLC gradient grade) from VWR International (Radnor, PA, USA), and ammonium acetate (≥98% purity) and methanol (LC–MS grade) from Thermo Fisher Scientific (Waltham, MA, USA). Ultra-pure water (>18 MΩ/cm) was prepared in house using an Integral 3 Milli-Q water system (Merck, Darmstadt, Germany).

2.2. Urine samples

Blank urines were obtained from laboratory staff members who self-reported that they had not taken benzodiazepines, and the samples were confirmed to be negative using the CEDIA benzodiazepine immunoassay. Patient urine samples were surplus aliquots collected from the routine drug testing service at the Department of Clinical Pharmacology, Karolinska University Laboratory (Stockholm, Sweden). These samples were collected consecutively from August 2015 until February 2017. The samples were stored for a maximum of 4 weeks at 4 °C and, thereafter, kept frozen at −20 °C. Patient samples were also obtained from the STRIDA project on NPS [2] (collected in 2012–2016) and these were stored at −20 °C upon arrival in the laboratory.

Ethical approval (No. 00-230 and No. 2013/116-31/2) was obtained from the regional ethical review board.

2.3. Preparation of solutions

2.3.1. Standard and quality control solutions

Four different stock solutions of benzodiazepines (0.1 mg/mL of each substance) were prepared in methanol. Solution A contained diclazepam, estazolam, etizolam, flubromazepam, meclonazepam, phenazepam and pyrazolam; solution B contained bentazepam, bromazepam, clonazolam, deschloroetizolam, flubromazolam, 3-hydroxyphenazepam, and nifoxipam; solution C contained adinazolam, cloniprazepam, fonazepam/N-desmethylflunitrazepam, 3-hydroxyflubromazepam, metizolam, nitrazolam, and pivoxazepam; and solution D contained clobazam, flunitrazolam, flurazepam, ketazolam, nimetazepam, and prazepam.

Urine standards were prepared at 1000 ng/mL in two different sets by diluting either solutions A and B (standard C1) or solutions C, D and 0.1 mg/mL tetrazepam (standard C2) with blank urine. These solutions were then further serially diluted with blank urine to obtain C1 and C2 standards with concentrations of 700, 300, 100, 50, 10, 5 and 1 ng/mL. In addition, single analyte standard solutions were prepared to study selectivity. These single analyte solutions were prepared in blank urine to a final concentration of 500 ng/mL.

Quality control (QC) solutions at 900, 300, 75, 50, 10 and 5 ng/mL were prepared similarly from separate stock solutions. Solutions A, B, C and D were stored at −20 °C and urine standards and QC at −80 °C.

Solutions used for studying stability and matrix effects were prepared in 6 different blank urines as described above. Additional solutions were prepared in 10 mmol/L ammonium acetate (pH 6.0) and used for comparison in the matrix experiments.

2.3.2. Internal standard working solution

The IS working solution containing 100 ng/mL each of α-hydroxyalprazolam-d5, α-hydroxytriazolam-d4, estazolam-d5, flunitrazepam-d7, lorazepam-d4, nordiazepam-d5, oxazepam-d5 and temazepam-d5 was prepared in 10 mmol/L ammonium acetate (pH 6.0) and stored at 4 °C for a maximum of 2 months.

2.4. Sample preparation

Samples were prepared in 96-well polypropylene plates (Waters, Milford, MA, USA) by adding 170 µL IS working solution, 30 µL β-glucuronidase solution, and 50 µL urine sample to each well. The plate was then gently shaking for 30 s, placed at room temperature (20 °C) for 20 min, and, then, directly transferred to the autosampler followed by injection of samples into the LC–HRMS system.

2.5. LC–HRMS system

The LC–HRMS system was comprised of a Dionex Ultimate UHPLC System coupled to a Q Exactive MS equipped with a heated electrospray ionization (HESI)-II source (Thermo Fisher Scientific, Waltham, MA, USA). The analytical column was a YMC UltraHT Hydrosphere C18 column (2.0 µm, 100 mm × 2.0 mm) combined with a Hydrosphere C18 guard column, both from YMC Europe GmbH (Dinslaken, Germany). The autosampler temperature was 12 °C, the column temperature 60 °C, and the injection volume 2 µL. Mobile phase A was composed of 10 mmol/L aqueous ammonium formate with 0.005% formic acid (pH 4.8) and mobile phase B of 10 mmol/L ammonium formate in Milli-Q water and methanol (10:90, v:v) with 0.005% formic acid. The flow rate was 0.5 mL/min. A mixture of formic acid, acetonitrile, and Milli-Q water (5:900:95, v:v:v) was used for needle wash, and methanol and Milli-Q water (10:90, v:v) for seal wash.

The MS operated in positive electrospray ionization mode with the following settings: spray voltage 3 kV, capillary temperature 300 °C, sheath gas 60 AU, auxiliary gas 18 AU, heater temperature 450 °C, and S-Lens RF level 70.

The software used for data evaluation was Trace Finder Clinical Research 4.1 (Thermo Fisher Scientific).

2.5.1. Screening method

The screening method used the following gradient: 0–0.50 min, 4% B; 0.51–2.45 min, ramp from 50% to 95% B; 2.45–2.50 min, 95% B; 2.51–3.50 min, 4% B. The exact masses monitored are shown in Table 1. The MS settings were as follows: scanning m/z 100–650, 70,000 resolution, AGC target 1 × 106, maximum IT 100 ms, and microscans 1. Identification criteria for the patient samples included a retention time of ±0.05 min from the standards and a mass of ±10 ppm from the exact mass of the analyte.

Table 1.

Chemical attributes (molecular formula, exact masses) and settings utilized in the screening and confirmation methods (MS/MS transitions, normalized collision energy (NCE) of the monitored MS/MS transitions, monitoring time in each parallel reaction monitoring method (PRM)).

Analyte Molecular formula Exact mass (M) Exact mass (M+H +) Quantifier transition (m/z) Qualifier transition (m/z) NCE (V) Time segment (min)
α-Hydroxyalprazolam-d5 C17H82H5ClN4O 329.1086 330.1165
α-Hydroxytriazolam-d4 C17H82H4Cl2N4O 362.0634 363.0712
Adinazolam C19H18ClN5 351.1245 352.1324 352.1324 > 58.0658 352.1324 > 295.0745 50 2.77–3.17a
Bentazepam C17H16N2OS 296.0978 297.1056 297.1056 > 166.0677 297.1056 > 269.1094 51 2.90–4.40c
Bromazepam C14H10BrN3O 315.0002 316.0080 316.0080 > 182.0839 316.0080 > 209.0948 51 2.00–2.60b
Clobazam C16H13ClN2O2 300.0660 301.0738 301.0738 > 259.0633 301.0738 > 224.0935 52 2.34–2.74a
Clonazolam C17H12ClN5O2 353.0674 354.0752 354.0752 > 326.0563 354.0752 > 319.1064 40 2.00–2.60c
Cloniprazepam C19H16ClN3O3 369.0875 370.0953 370.0953 > 316.0484 370.0953 > 302.0453 46 3.00–4.40a
Deschloroetizolam C17H16N4S 308.1090 309.1168 309.1168 > 255.0951 309.1168 > 276.1367 41 2.50–3.10b
Diclazepam C16H12Cl2N2O 318.0321 319.0399 319.0399 > 227.0496 319.0399 > 154.0418 55 2.70–4.40b
Estazolam C16H11ClN4 294.0667 295.0745 295.0745 > 205.0760 295.0745 > 138.0106 60 2.20–2.80b
Estazolam-d5 C16H62H5ClN4 299.0981 300.1059 300.1059 > 272.0872 60 2.20–2.80b,c, 2.31–2.71a
Etizolam C17H15ClN4S 342.0701 343.0779 343.0779 > 314.0388 343.0779 > 259.0216 46 2.50–3.10b
Flubromazepam C15H10BrFN2O 331.9955 333.0033 333.0033 > 226.0889 333.0033 > 183.9751 55 2.50–3.10c
Flubromazolam C17H12BrFN4 370.0224 371.0302 371.0302 > 343.0096 371.0302 > 292.1105 52 2.30–2.90c
Flunitrazepam-d7 C16H52H7FN3O3 320.1297 321.1375 321.1375 > 307.1344 50 0.70–4.40b, 2.10–2.70c, 2.23–2.63a
Flunitrazolam C17H12FN5O2 337.0970 338.1048 338.1048 > 338.1048 338.1048 > 310.0861 40 1.95–2.35a
Flurazepam C21H23ClFN3O 387.1508 388.1586 388.1586 > 315.0695 388.1586 > 317.0852 30 0.70–2.20a
3-Hydroxyflubromazepam C15H10BrFN2O2 347.9904 348.9982 348.9982 > 302.9928 348.9982 > 330.9877 30 2.31–2.71a
3-Hydroxyphenazepam C15H10BrClN2O2 363.9609 364.9687 364.9687 > 318.9632 364.9687 > 273.0022 51 2.35–2.95b
Ketazolam C20H17ClN2O3 368.0922 369.1001
Ketazolam fragment C16H13ClN2O 284.0711 285.0789 285.0780 > 193.0886 285.0780 > 257.0840 50 2.93–3.33a
Meclonazepam C16H12ClN3O3 329.0562 330.0640 330.0640 > 316.0594 330.0640 > 285.0423 42 2.40–3.00c
Metizolam C16H13ClN4S 328.0544 329.0622 329.0622 > 275.0386 329.0622 > 296.0823 40 2.38–2.78a
N-Desmethylflunitrazepam/Fonazepam C15H10FN3O3 299.0701 300.0779 300.0779 > 198.0714 300.0779 > 225.0815 60 2.07–2.47a
Nifoxipam C15H10FN3O4 315.0650 316.0728 316.0728 > 298.0623 316.0728 > 270.0673 26 0.70–2.30b
Nimetazepam C16H13N3O3 295.0951 296.1030 296.1030 > 221.1073 296.1030 > 268.1081 48 2.33–2.73a
Nitrazolam C17H13N5O2 319.1064 320.1142 320.1142 > 292.0955 320.1142 > 198.0900 48 1.98–2.38a
Nordiazepam-d5 C15H62H5ClN2O 275.0868 276.0947 276.0947 > 213.1309 50 2.71–3.11a
Oxazepam-d5 C15H62H5ClN2O2 291.0817 292.0896 292.0896 > 246.0841 40 2.38–2.78a
Phenazepam C15H10BrClN2O 347.9660 348.9738 348.9738 > 183.9751 348.9738 > 242.0593 47 2.60–3.20c
Pivoxazepam C20H19ClN2O3 370.1079 371.1157 371.1157 > 269.0476 371.1157 > 241.0524 15 3.00–4.40a
Prazepam C19H17ClN2O 324.1024 325.1102 325.1102 > 271.0633 325.1102 > 208.0995 52 3.00–4.40a
Pyrazolam C16H12BrN5 353.0271 354.0349 354.0349 > 167.0722 354.0349 > 206.0835 52 0.70–2.35c
Temazepam-d5 C16H82H5ClN2O2 305.0974 306.1052 306.1052 > 288.0947 40 2.52–2.92a, 2.40–3.00b,c
Tetrazepam C16H17ClN2O 288.1024 289.1102 289.1102 > 169.0887 289.1102 > 117.0576 68 3.00–4.40a
a

PRM method 1.

b

PRM method 2.

c

PRM method 3.

2.5.2. Confirmation method

The confirmation method used the following gradient: 0–0.50 min, 4% B; 0.51–3.45 min, ramp from 50% to 95% B; 3.45–3.50 min, 95% B; 3.51–4.50 min, 4% B. The MS was operated in parallel reaction monitoring (PRM) mode with 17,500 resolution, AGC target 2 × 105, maximum IT 130 ms, microscans 1, and isolation window 1.5 m/z. Three different PRM methods were used (same general settings, but different PRM inclusion lists). The analytes found in each method, their normalized collision energy (NCE), and the monitored transitions and transition times are shown in Table 1. Identification criteria for the patient samples included a retention time of ±0.05 min from the standards, a mass value of ±10 ppm from the expected mass (Table 1), and a correct peak area ion ratio according to Maralikova and Weinmann [27].

2.6. Method validation

2.6.1. Precision, accuracy and limit of detection

The precision and accuracy were determined over five different days by analyzing the QC samples in triplicate. The lower limit of quantification (LLOQ) was defined as having an accuracy of 80–120% and a coefficient of variation (CV) ≤20%. The limit of detection (LOD) was determined from quintuplicate analysis of samples over five days. The LOD of the screening method was defined as the spiked concentration at which ≥75% of the investigated samples had an acceptable peak shape. The LOD of the confirmation method was defined as the concentration at which ≥75% of the investigated samples had a correct ion ratio.

2.6.2. Matrix effects

Matrix effects were determined at the LLOQ and at 900 ng/mL by comparing the mean area of duplicates of six different blank urines to the mean area of duplicates in 10 mmol/L ammonium acetate (pH 6.0), according to Matuszewski et al [28]:

Matrixeffect=meantotalareaofanalyteinurinemeantotalareaofanalyteinbuffer×100-100

2.6.3. Selectivity

Selectivity was studied using six different blank urines and 500 ng/mL urine solutions of each analyte. Possible interference from >100 other NPS [19] was investigated by analyzing urine solutions spiked with a range of analytes included in a substance abuse monitoring program at the Karolinska University Laboratory. Furthermore, the ability to accurately detect each analyte in the presence of a number of interferents was investigated by spiking urine with 100,000 ng/mL ethyl glucuronide, 30,000 ng/mL each of morphine-3-glucuronide, morphine, codeine-6-glucuronide, codeine, and ethylmorphine-6-glucuronide, 20,000 ng/mL ethyl sulfate, 10,000 ng/mL of α-hydroxyalprazolam, α-hydroxymidazolam, α-hydroxytriazolam, 7-aminoclonazepam, 7-aminoflunitrazepam, 7-aminonitrazepam, desmethyldiazepam, diazepam, lorazepam, oxazepam, and temazepam), 8000 ng/mL of amphetamine, methamphetamine, MDMA, and MDA, 1000 ng/mL buprenorphine, 750 ng/mL 6-acetylmorphine, and 450 ng/mL benzoylecgonine.

2.6.4. Stability

The stability of the analytes in methanol stock solutions A, B, C and D at −20 °C was determined by diluting both stored and freshly prepared methanol solutions with 10 mmol/L ammonium acetate (pH 6.0) to concentrations of 900 ng/mL, as described in Section 2.3.1. The solutions were then analyzed in triplicate, and the relative difference in analyte/IS area ratio of the stored stock solutions was compared to freshly prepared solutions. Analytes were considered to have retained stability during storage if the stored methanol stock solutions differed by ≤20% from the freshly prepared solutions.

The stability of the analytes in a urine matrix was investigated in triplicate at the LLOQ and at 700 ng/mL using six different blank urines, by comparison against a standard curve of the corresponding blank urine. Stability was studied at room temperature (∼20 °C), 4 °C, and −80 °C. Stability was also examined during three freeze/thaw cycles at −80 °C, and for prepared samples stored at 12 °C in the autosampler. Analytes were considered to have retained stability if the mean concentrations of the six spiked blank urines were within ±20% of the spiked value at both the LLOQ and 700 ng/mL.

2.6.5. Carry-over

Carry-over was examined for the screening and the confirmation methods by injection of a blank urine immediately after injecting the highest standards (1000 ng/mL).

2.7. Application of the method

The method was applied to the analysis of 126 routine patient urine samples that tested positively in the CEDIA benzodiazepine immunoassay, but were confirmed negative for the standard set of prescription benzodiazepines and metabolites evaluated by the in-house LC–MS/MS confirmation method. The analytes included in the in-house method were α-hydroxyalprazolam, α-hydroxymidazolam, α-hydroxytriazolam, 7-aminoclonazepam, 7-aminoflunitrazepam, 7-aminonitrazepam, desmethyldiazepam, lorazepam, oxazepam, and temazepam. Additionally, 329 patient urine samples from intoxication cases were obtained from the STRIDA project [2] and analyzed.

3. Results and discussion

3.1. Method validation

3.1.1. Precision, accuracy and limit of detection

The LOD (1–50 ng/mL) and measuring ranges (5–1000 ng/mL) of the screening and confirmation methods are shown in Supplementary material 1. A screening method chromatogram displaying all analytes included in the study is presented in Fig. 1. The measuring range of the confirmation method was similar to that of the screening method, but three analytes (i.e., adinazolam, meclonazepam and nimetazepam) showed a higher LLOQ in the confirmation method. The lower sensitivity of these analytes in the confirmation method was due to the signal intensity of the qualifier ion. The measuring ranges obtained for both the screening and confirmation methods corresponded to the substance concentrations previously observed in patient urine samples [29].

Fig. 1.

Fig. 1

Chromatograms showing the retention time (RT) of all NPS benzodiazepines and internal standards included in the method in a screening run with the highest standard (i.e., 1000 ng/mL). The standards were prepared in two sets (left and right). Ketazolam is shown as both the M+H + ion (RT 2.67) and its in-source fragment at m/z 285.0789 (RT 2.66).

The screening method was less sensitive for ketazolam (i.e., LOD > 50 ng/mL) which was assumed to be due to in-source fragmentation into an ion with m/z 285.0789 (i.e., loss of C4H4O2). This supposition was supported by difficulties encountered while optimizing the NCE for this analyte in the confirmation method. Consequently, the m/z 285.0789 ion (herein named “ketazolam fragment”) was monitored together with ketazolam in the screening method and used as the parent ion for the transitions monitoring ketazolam intake in the confirmation method. However, 285.0789 is also the exact mass of the M+H+ ion of diazepam, which has been demonstrated to be a degradation product of ketazolam when analyzed by LC–MS with thermospray ionization [30], and on storage of ketazolam in methanol [31]. Therefore, a patient sample was not considered positive for ketazolam, unless both ketazolam and the fragment ion were detected in the screening method and the fragment in the confirmation method.

The parent M+H+ ion of flunitrazolam was chosen as the quantifier ion in the confirmation method together with the fragment of m/z 310.0861 (qualifier), since using other fragments of flunitrazolam resulted in an ion ratio dependent on the sample concentration. This was eliminated by using the parent mass as quantifier.

The standard curves used in the accuracy and precision experiments comprised 5–7 concentration levels and all coefficients of determination (R2-values) were >0.98. The total CV of the screening (Supplementary material 2) and confirmation (Supplementary material 3) method was ≤15% at all investigated concentrations for all substances. Furthermore, the accuracy was within ±15% at concentrations above the LLOQ and within ±20% at the LLOQ for all analytes, except pivoxazepam. However, in the screening method, the accuracy for pivoxazepam was within ±20% and the total CV was also <15%.

3.1.2. Selectivity

Background peak areas from the six different blank urines were <20% of the peak areas for all analytes at the LLOQ levels, except for flurazepam and tetrazepam where the responses were ≤44%. The background interference for flurazepam was <20% in the confirmation method, whereas the interference for tetrazepam remained at a high level. However, because the response was always less than the LLOQ area, the interference was considered a minor problem.

Metizolam gave a response at the MS trace used for meclonazepam, but they were separated chromatographically (Fig. 1). Flurazepam gave a response for nifoxipam and these were not well separated in the chromatographic systems. The final identity, therefore, had to be based on the ion ratio. Because of this, metizolam and flurazepam were separated from meclonazepam and nifoxipam in the urine standards.

Unfortunately, diazepam was not separated from ketazolam and its in-source fragment, but as the identification criterion for a ketazolam positive sample included finding both ketazolam and its fragment in full scan mode, samples containing only diazepam would not be incorrectly identified as ketazolam. Lorazepam gave a response of 61% of the LLOQ area of 3-hydroxyphenazepam in the screening method, but as no response was noted in the confirmation method this was not considered an issue. Otherwise, the peak response area for classical drugs of abuse and the >100 NPS examined were similar to the results for blank urines and they were, consequently, considered not to cause selectivity problems.

3.1.3. Matrix effect

The analytes showed matrix effects ≤±20% at both the LLOQ and 900 ng/mL in the screening method, except for N-desmetylflunitrazepam, flunitrazolam, ketazolam, nifoxipam, pivoxazepam and tetrazepam where they ranged between −38% and 63%. However, since the CV of the mean calculated concentration was within 20% for all of these compounds, it appears that the IS was able to compensate for any matrix effects. In the confirmation method, all analytes, except nifoxipam, showed a matrix effect of ≤±20% at both the LLOQ and 900 ng/mL. As in the screening method, the CV of the mean calculated nifoxipam concentration was within 20%, indicating that the IS was also able to compensate for matrix effects in the confirmation method.

3.1.4. Stability

The stability of analytes in urine, methanol, and as prepared extracts was studied under various conditions and the results are presented in Supplementary material 4. All analytes, except clonazolam and nifoxipam, were stable at all conditions, as confirmed by the measured concentrations being within ±20% of the expected values (Supplementary material 4). For clonazolam, the mean stability at −80 °C over 4 months (i.e., −21%) was just outside the acceptance limit (i.e., ±20%). The concentration of nifoxipam decreased slightly below the acceptance limit after storage in methanol for 15 months, at 4 °C for 1 week, and at −80 °C for 4 months. It was, therefore, not recommended to store patient urine samples longer than 1 week at 4 °C. However, previous studies investigating the stability of NPS benzodiazepines in urine at −20 °C suggested that they are stable for at least 2 months under these storage conditions [17]. Urine standards and QC samples were stable at −80 °C over a 4-month period and could be thawed at least three times without significantly altering the results. The methanol solutions were stable for 15 months at −20 °C, and prepared extracts were stable for 4 days at 12 °C in the autosampler.

3.1.5. Carry-over

The peak area for blank urines injected immediately after the highest standard was <20% of the LLOQ peak area in both the screening and the confirmation method for all analytes, except pyrazolam and tetrazepam. For pyrazolam, the peak areas were always lower than the LLOQ, whereas for tetrazepam a response similar to that for the LLOQ was obtained. Due to this, it is recommended that tetrazepam positive samples be re-analyzed if showing a concentration near the LLOQ and if the sample was analyzed directly after a high concentration sample.

3.2. Application of the method

Both methods were employed for analysis of 126 patient urine samples collected between 2015 and 2017 that had tested positive in the CEDIA benzodiazepine immunoassay, but were confirmed negative for α-hydroxyalprazolam, α-hydroxymidazolam, α-hydroxytriazolam, 7-aminoclonazepam, 7-aminoflunitrazepam, 7-aminonitrazepam, desmethyldiazepam, lorazepam, oxazepam, and temazepam by the routine LC–MS/MS confirmation method. An NPS benzodiazepine was detected in 40 out of the 126 samples (i.e., 32%). The 9 detected compounds were clobazam, estazolam, flubromazepam, flubromazolam, flunitrazolam, 3-hydroxyflubromazepam, 3-hydroxyphenazepam, metizolam, and nifoxipam, and the number of cases of each are shown in Fig. 2. More than half of the positive cases contained flunitrazolam, which became classified as a narcotic in Sweden in January 2017 [32]. Example chromatograms for a flunitrazolam positive case are shown in Fig. 3A, Fig. 3B.

Fig. 2.

Fig. 2

Number of urine samples containing each analyte among 126 patient samples collected from routine drug testing.

Fig. 3A.

Fig. 3A

Chromatograms from the screening method showing a patient sample containing 14 ng/mL flunitrazolam (left) and a standard spiked with 10 ng/mL flunitrazolam (right).

Fig. 3B.

Fig. 3B

Chromatograms from the confirmation method showing a patient sample containing 14 ng/mL flunitrazolam (left) and a standard spiked with 10 ng/mL flunitrazolam (right). Quantification transitions are shown above and qualification transitions below.

Furthermore, 329 urine samples from the STRIDA project on NPS [2], constituting acute intoxication cases presented to emergency wards in Sweden, were investigated. The following 14 NPS benzodiazepines were detected in 95 (i.e., 29%) of the STRIDA samples: clonazolam, deschloroetizolam, diclazepam, estazolam, etizolam, flubromazepam, flubromazolam, 3-hydroxyflubromazepam, 3-hydroxyphenazepam, ketazolam, meclonazepam, metizolam, nifoxipam, and pyrazolam. Many of these 95 samples were multi-intoxication cases (e.g., co-administrations of flubromazepam and pyrazolam).

The concentration ranges found in the routine and STRIDA samples are presented in Table 2. As expected, the substance concentrations were typically higher in the intoxication cases. Many patient samples showed concentrations for NPS benzodiazepines <50 ng/mL, which is the recommended analytical cut-off for prescription benzodiazepines in Sweden. Hence, based on the present and previous [17] results, a lower cut-off value of 10 ng/mL is suggested for urinary analysis of clonazolam, estazolam, flubromazepam, flunitrazolam, meclonazepam and metizolam, whereas 50 ng/mL may be sufficient for the other analytes.

Table 2.

Concentration ranges of NPS benzodiazepines found in urine samples from routine drug testing and from intoxication cases in emergency wards (STRIDA project samples).

Substance Concentration, ng/mL (Routine samples) Number of positive samples (Routine samples) Concentration, ng/mL (STRIDA samples) Number of positive samples (STRIDA samples)
Clobazam 86 1
Clonazolam 10–132 8
Deschloroetizolam 5 1
Diclazepam 8 1
Estazolam 30 1 8 1
Etizolam 40–1000 5
Flubromazepam 5 1 5–89 11
Flubromazolam 15–61 6 5–1082* 52
Flunitrazolam 5–121 24
3-Hydroxyflubromazepam 191–410 2 90–11,580* 10
3-Hydroxyphenazepam 84–16,549* 5 258–12,090* 5
Ketazolam 1
Meclonazepam 5–126 10
Metizolam 5–14 3 8 1
Nifoxipam 149 1 973–3635* 2
Pyrazolam 6–1995* 11
*

Samples with a substance concentration >1000 ng/mL were diluted 1:10 or 1:100 with Milli-Q water prior to re-analysis.

4. Conclusion

The multi-component LC–HRMS(/MS) method offered comprehensive and sensitive detection of a large number of benzodiazepines sold as NPS. Since the screening method is based on LC–HRMS analysis in full scan mode, it enables easy inclusion of new analytes as they appear on the market for recreational drugs. Analysis of NPS benzodiazepines in urine samples from the routine drug testing or acute drug intoxication cases revealed a high prevalence (∼30%) of NPS benzodiazepine use in Sweden in the last years.

Declaration of interest

None to declare.

Footnotes

Appendix A

Supplementary data associated with this article can be found, in the online version, at https://doi.org/10.1016/j.clinms.2018.08.004.

Appendix A. Supplementary data

The following are the Supplementary data to this article:

Supplementary data 1
mmc1.docx (85.1KB, docx)

References

  • 1.European Monitoring Centre for Drugs and Drug Addication: European Drug Report 2017, http://www.emcdda.europa.eu/system/files/publications/4541/TDAT17001ENN.pdf (accessed 02.27.18).
  • 2.Helander A., Beck O., Hagerkvist R., Hulten P. Identification of novel psychoactive drug use in Sweden based on laboratory analysis-initial experiences from the STRIDA project. Scand. J. Clin. Lab. Invest. 2013;73:400–406. doi: 10.3109/00365513.2013.793817. [DOI] [PubMed] [Google Scholar]
  • 3.Helander A., Backberg M. New Psychoactive Substances (NPS) – the Hydra monster of recreational drugs. Clin. Toxicol. (Phila) 2017;55:1–3. doi: 10.1080/15563650.2016.1217003. [DOI] [PubMed] [Google Scholar]
  • 4.Meyer M.R., Maurer H.H. Review: LC coupled to low- and high-resolution mass spectrometry for new psychoactive substance screening in biological matrices – where do we stand today? Anal. Chim. Acta. 2016;927:13–20. doi: 10.1016/j.aca.2016.04.046. [DOI] [PubMed] [Google Scholar]
  • 5.Baldwin D.S., et al. Benzodiazepines: risks and benefits A reconsideration. J. Psychopharmacol. 2013;27:967–971. doi: 10.1177/0269881113503509. [DOI] [PubMed] [Google Scholar]
  • 6.Moosmann B., et al. Characterization of the designer benzodiazepine pyrazolam and its detectability in human serum and urine. Forensic Toxicol. 2013;31:263–271. [Google Scholar]
  • 7.Moosmann B., et al. Detection and identification of the designer benzodiazepine flubromazepam and preliminary data on its metabolism and pharmacokinetics. J. Mass Spectrom. 2013;48:1150–1159. doi: 10.1002/jms.3279. [DOI] [PubMed] [Google Scholar]
  • 8.Moosmann B., Bisel P., Auwarter V. Characterization of the designer benzodiazepine diclazepam and preliminary data on its metabolism and pharmacokinetics. Drug Test. Anal. 2014;6:757–763. doi: 10.1002/dta.1628. [DOI] [PubMed] [Google Scholar]
  • 9.Huppertz L.M., et al. Characterization of the four designer benzodiazepines clonazolam, deschloroetizolam, flubromazolam, and meclonazepam, and identification of their in vitro metabolites. Forensic Toxicol. 2015;33:388–395. [Google Scholar]
  • 10.Moosmann B., Bisel P., Franz F., Huppertz L.M., Auwarter V. Characterization and in vitro phase I microsomal metabolism of designer benzodiazepines – an update comprising adinazolam, cloniprazepam, fonazepam, 3-hydroxyphenazepam, metizolam and nitrazolam. J. Mass Spectrom. 2016;51:1080–1089. doi: 10.1002/jms.3840. [DOI] [PubMed] [Google Scholar]
  • 11.Manchester K.R., Lomas E.C., Waters L., Dempsey F.C., Maskell P.D. The emergence of new psychoactive substance (NPS) benzodiazepines: a review. Drug Test. Anal. 2018;10:37–53. doi: 10.1002/dta.2211. [DOI] [PubMed] [Google Scholar]
  • 12.Pettersson Bergstrand M., Helander A., Hansson T., Beck O. Detectability of designer benzodiazepines in CEDIA, EMIT II Plus, HEIA, and KIMS II immunochemical screening assays. Drug Test. Anal. 2017;9:640–645. doi: 10.1002/dta.2003. [DOI] [PubMed] [Google Scholar]
  • 13.O'Connor L.C., Torrance H.J., McKeown D.A. ELISA detection of phenazepam, etizolam, pyrazolam, flubromazepam, diclazepam and delorazepam in blood using immunalysis(R) Benzodiazepine kit. J. Anal. Toxicol. 2016;40:159–161. doi: 10.1093/jat/bkv122. [DOI] [PubMed] [Google Scholar]
  • 14.Tomkova J., Svidrnoch M., Maier V., Ondra P. Analysis of selected designer benzodiazepines by ultra high performance liquid chromatography with high-resolution time-of-flight mass spectrometry and the estimation of their partition coefficients by micellar electrokinetic chromatography. J. Sep. Sci. 2017;40:2037–2044. doi: 10.1002/jssc.201700069. [DOI] [PubMed] [Google Scholar]
  • 15.Svidrnoch M., Boranova B., Tomkova J., Ondra P., Maier V. Simultaneous determination of designer benzodiazepines in human serum using non-aqueous capillary electrophoresis – tandem mass spectrometry with successive multiple ionic – polymer layer coated capillary. Talanta. 2018;176:69–76. doi: 10.1016/j.talanta.2017.08.010. [DOI] [PubMed] [Google Scholar]
  • 16.Meng L., Zhu B., Zheng K., Fu S. Ultrasound-assisted low-density solvent dispersive liquid-liquid microextraction for the determination of 4 designer benzodiazepines in urine samples by gas chromatography-triple quadrupole mass spectrometry. J. Chromatogr. B Analyt. Technol. Biomed. Life Sci. 2017;1053:9–15. doi: 10.1016/j.jchromb.2017.04.008. [DOI] [PubMed] [Google Scholar]
  • 17.Pettersson Bergstrand M., Helander A., Beck O. Development and application of a multi-component LC-MS/MS method for determination of designer benzodiazepines in urine. J. Chromatogr. B Analyt. Technol. Biomed. Life Sci. 2016;1035:104–110. doi: 10.1016/j.jchromb.2016.08.047. [DOI] [PubMed] [Google Scholar]
  • 18.Katselou M., Papoutsis I., Nikolaou P., Spiliopoulou C., Athanaselis S. Metabolites replace the parent drug in the drug arena. The cases of fonazepam and nifoxipam. Forensic Toxicol. 2017;35:1–10. doi: 10.1007/s11419-016-0338-5. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19.Stephanson N.N., Signell P., Helander A., Beck O. Use of LC-HRMS in full scan-XIC mode for multi-analyte urine drug testing – a step towards a ‘black-box’ solution? J. Mass Spectrom. 2017;52:497–506. doi: 10.1002/jms.3946. [DOI] [PubMed] [Google Scholar]
  • 20.Maskell P.D., De Paoli G., Nitin Seetohul L., Pounder D.J. Phenazepam: the drug that came in from the cold. J. Forensic Leg. Med. 2012;19:122–125. doi: 10.1016/j.jflm.2011.12.014. [DOI] [PubMed] [Google Scholar]
  • 21.Mandrioli R., Mercolini L., Raggi M.A. Benzodiazepine metabolism: an analytical perspective. Curr. Drug Metab. 2008;9:827–844. doi: 10.2174/138920008786049258. [DOI] [PubMed] [Google Scholar]
  • 22.Pernea M., Sutcliffe A.G. Clobazam and its use in epilepsy. Pediatr. Rep. 2016;8:6516. doi: 10.4081/pr.2016.6516. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23.Schroeck J.L., et al. Review of safety and efficacy of sleep medicines in older adults. Clin. Ther. 2016;38:2340–2372. doi: 10.1016/j.clinthera.2016.09.010. [DOI] [PubMed] [Google Scholar]
  • 24.Dinis-Oliveira R.J. Metabolic profile of oxazepam and related benzodiazepines: clinical and forensic aspects. Drug Metab. Rev. 2017;49:451–463. doi: 10.1080/03602532.2017.1377223. [DOI] [PubMed] [Google Scholar]
  • 25.Wang Y., et al. Discovery of FDA-approved drugs as inhibitors of fatty acid binding protein 4 using molecular docking screening. J. Chem. Inf. Model. 2014;54:3046–3050. doi: 10.1021/ci500503b. [DOI] [PubMed] [Google Scholar]
  • 26.Proy-Vega B., Aguirre C., de Groot P., Solis-Garcia del Pozo J., Jordan J. On the clinical evidence leading to tetrazepam withdrawal. Expert Opin. Drug Saf. 2014;13:705–712. doi: 10.1517/14740338.2014.915023. [DOI] [PubMed] [Google Scholar]
  • 27.Maralikova B., Weinmann W. Confirmatory analysis for drugs of abuse in plasma and urine by high-performance liquid chromatography-tandem mass spectrometry with respect to criteria for compound identification. J. Chromatogr. B Analyt. Technol. Biomed. Life Sci. 2004;811:21–30. doi: 10.1016/j.jchromb.2004.04.039. [DOI] [PubMed] [Google Scholar]
  • 28.Matuszewski B.K., Constanzer M.L., Chavez-Eng C.M. Strategies for the assessment of matrix effect in quantitative bioanalytical methods based on HPLC-MS/MS. Anal. Chem. 2003;75:3019–3030. doi: 10.1021/ac020361s. [DOI] [PubMed] [Google Scholar]
  • 29.Logan B.K., et al. Reports of adverse events associated with use of novel psychoactive substances, 2013–2016: a review. J. Anal. Toxicol. 2017;41:573–610. doi: 10.1093/jat/bkx031. [DOI] [PubMed] [Google Scholar]
  • 30.Verweij A.M.A., Lipman P.J.L., Zweipfenning P.G.M. Quantitative liquid chromatography, thermospray/tandem mass spectrometry (LC/TSP/MS/MS) analysis of some thermolabile benzodiazepines in whole-blood. Forensic Sci. Int. 1992;54:67–74. doi: 10.1016/0379-0738(92)90081-7. [DOI] [PubMed] [Google Scholar]
  • 31.Melo P., Bastos M.L., Teixeira H.M. Benzodiazepine stability in postmortem samples stored at different temperatures. J. Anal. Toxicol. 2012;36:52–60. doi: 10.1093/jat/bkr008. [DOI] [PubMed] [Google Scholar]
  • 32.Svensk författningssamling: Förordning om ändring i förordningen (1992:1554) om kontroll av narkotika; SFS 2016:1303, https://www.notisum.se/rnp/sls/sfs/20161303.pdf (accessed 01.01.18.).

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

Supplementary data 1
mmc1.docx (85.1KB, docx)

Articles from Clinical Mass Spectrometry are provided here courtesy of Elsevier

RESOURCES