Abstract
A novel Fe2Mo3O8/MoO2@MoS2 nanocomposite is synthesized for extremely sensitive detection of NH3 in the breath of kidney disease patients at room temperature. Compared to MoS2, α‐Fe2O3/MoS2, and MoO2@MoS2, it shows the optimal gas‐sensing performance by optimizing the formation of Fe2Mo3O8 at 900 °C. The annealed Fe2Mo3O8/MoO2@MoS2 nanocomposite (Fe2Mo3O8/MoO2@MoS2‐900 °C) sensor demonstrates a remarkably high selectivity of NH3 with a response of 875% to 30 ppm NH3 and an ultralow detection limit of 3.7 ppb. This sensor demonstrates excellent linearity, repeatability, and long‐term stability. Furthermore, it effectively differentiates between patients at varying stages of kidney disease through quantitative NH3 measurements. The sensing mechanism is elucidated through the analysis of alterations in X‐ray photoelectron spectroscopy (XPS) signals, which is supported by density functional theory (DFT) calculations illustrating the NH3 adsorption and oxidation pathways and their effects on charge transfer, resulting in the conductivity change as the sensing signal. The excellent performance is mainly attributed to the heterojunction among MoS2, MoO2, and Fe2Mo3O8 and the exceptional adsorption and catalytic activity of Fe2Mo3O8/MoO2@MoS2‐900 °C for NH3. This research presents a promising new material optimized for detecting NH3 in exhaled breath and a new strategy for the early diagnosis and management of kidney disease.
Keywords: ammonia, exhaled breath, Fe2Mo3O8/MoO2@MoS2 nanocomposite, kidney disease, room‐temperature
An ultrasensitive NH3 gas sensor is developed based on Fe2Mo3O8/MoO2@MoS2‐900 °C nanocomposite. The sensor effectively distinguishes between patients with early‐ and late‐stage kidney disease by quantitative analysis of NH3 in their exhaled breath. This novel approach offers a promising method for the early diagnosis and management of kidney disease.

1. Introduction
Ammonia (NH3) is a key component in amino acids and proteins. It originates from a variety of sources, such as animal and plant substances, organic decomposition, and industrial wastewater.[ 1 ] Although NH3 can negatively affect the environment and human health, its detection has many practical applications. In particular, it can be used as a breath signature for early screening or indirect diagnosis of some diseases. Poor liver function can delay the metabolism of NH3, while impaired kidney function can delay the excretion of metabolic end products such as urea. Patients with such disease could have a high level of NH3, detectable in their breath.
Conventional methods of diagnosing kidney disease, such as blood measurements, urine tests, renal biopsies, B‐ultrasounds, computed tomography scans, or magnetic resonance imaging, are typically invasive, uncomfortable, time‐consuming, and costly. Moreover, these methods are only suitable for a limited range of patients. Hence, the use of NH3 as a breath signature offers a convenient, low‐cost method for the early screening and diagnosis of kidney disease, even in a domestic environment. Previous studies have reported that NH3 concentrations of >10 ppb in the breath of a patient may indicate liver or kidney failure.[ 2 ] Therefore, it is a great challenge to develop a cost‐effective and portable but still highly sensitive and selective NH3 sensor with a detection limit better than 10 ppb. Such devices can also be adapted for industrial applications such as environmental protection, public safety, and chemical leakage monitoring.
A variety of techniques have been used to detect and monitor NH3, including optical, chromatographic, spectroscopic, and mass spectrometric techniques.[ 3 ] Nevertheless, these methods require complex pretreatments, bulky equipment, and intricate processes at high costs, which have hindered their widespread application.[ 4 , 5 ] Gas sensors, particularly metal‐oxide‐semiconductor sensors, could offer sensitive, portable, user‐friendly, and cost‐effective NH3 detection, making them highly suitable for noninvasive screening applications. Nguyen et al. sputter deposited SnO2/Pt/WO3 ternary films on silicon oxide substrates, which achieved a response of 46.3 for 100 ppm of NH3 at 250 °C.[ 6 ] Kan et al. constructed a dual‐mode foam sensor based on Ti3C2Tx/In2O3 nanocomposite for noninvasive detection of kidney disease by measuring the NH3 concentration. The results showed that the resistance of the sensor changed by at least 0.1 MΩ in a simulated exhaled breath test with a detection limit of 1 ppm at room temperature.[ 7 ] Wang et al. developed a wearable respiratory sensor using CeO2@polyaniline nanocomposite to detect trace levels of NH3 with a relatively low response of 300% for 30 ppm NH3.[ 8 ] Zhao et al. proposed a proton‐conductive gas sensor based on polyvinylpyrrolidone to detect 0.5 ppm NH3 of exhaled breath in simulated environments with an estimated detection limit of 36 ppb.[ 9 ] Thus, the difficulty remains in developing highly sensitive NH3 detectors that are capable of measuring the NH3 concentration as low as 10 ppb in order to provide early diagnosis of kidney diseases before the presence of other typical symptoms. Hence it is urgently needed to develop new material that is sensitive enough to detect NH3 with good selectivity.
MoS2 has been identified as a promising candidate for detecting molecules containing nitrogen due to its strong interaction between Mo and N.[ 10 ] Zhang et al. fabricated MoS2/Co3O4 thin film sensors on interdigitated electrode substrates with 62.4% response to 5 ppm NH3 and fast response/recovery for NH3 detection.[ 11 ] Wang et al. prepared a SnO2/MoS2 nanocomposite, which showed a good response of 2080.36 for 200 ppm of NH3 and a fast response/recovery time of 23/1.6 s for 50 ppm of NH3 at 25 °C.[ 12 ] However, the detection of NH3 in the exhaled breath for patients with kidney disease using gas sensors remains limited due to the ultra‐low NH3 concentration and potential interference with other gases, including moisture. As a result, most of the current gas sensors only offer qualitative detection of ammonia gas in exhaled breath, which provides less accurate and meaningful information. The difficulty in detecting low NH3 concentration for early diagnosis needs to be addressed, which is essential for the early treatment of kidney disease.
In this study, an ultrasensitive NH3 gas sensor was constructed using a Fe2Mo3O8/MoO2@MoS2 nanocomposite. The optimal gas‐sensing performance was achieved by optimizing the formation Fe2Mo3O8 at 900 °C under a low oxygen partial pressure. This optimal sample is named Fe2Mo3O8/MoO2@MoS2‐900 °C, which shows a high response of 875% for 30 ppm of NH3, with an ultralow detection limit of 3.7 ppb, when operating at room temperature. It also shows excellent selectivity for NH3 with long‐term stability observed in a four‐week test. Such a superior detection limit for NH3 is critical for the application in early diagnosis of kidney disease.
Moreover, our Fe2Mo3O8/MoO2@MoS2‐900 °C sensor was successfully employed to differentiate patients with early‐ and late‐stage kidney disease by quantitative analysis of NH3 in the exhaled breath. This confirms that our sensor is sensitive enough to identify the early stage of kidney disease with good reliability, providing the opportunity for early medical intervention. The sensing mechanism was elucidated through the analysis of the interaction between Fe2Mo3O8 and NH3, the adsorption and oxidation of NH3, and the charge transfer mechanism, which is manifested through the drastic change of the conductivity upon exposure to NH3. This was supported by Ab Initio density functional theory (DFT) calculations of the adsorption energy of NH3 on sensing materials and the free energy changes during NH3 oxidation. This work provides a new material that offers ultra‐sensitivity to NH3 detection in the exhaled breath of patients with kidney disease. The application of this technology can potentially provide a useful domestic medical device for early diagnosis of kidney functioning.
2. Results and Discussion
2.1. Material Characterization
The Fe2Mo3O8/MoO2@MoS2‐900 °C nanocomposite was prepared through a secondary hydrothermal method and an annealing process and subsequently utilized for the fabrication of an NH3 gas sensor, as illustrated in Figure 1 .
Figure 1.

Schematic diagram of the synthesis of Fe2Mo3O8/MoO2@MoS2‐900 °C nanocomposite and the fabrication of Fe2Mo3O8/MoO2@MoS2‐900 °C gas sensor.
Field‐emission scanning electron microscopy (SEM) and high‐resolution transmission electron microscopy (HR‐TEM) were employed to examine the surface morphology and lattice structure of the as‐prepared Fe2Mo3O8/MoO2@MoS2‐900 °C nanocomposite. In Figure 2a, the SEM image shows the Fe2Mo3O8/MoO2@MoS2‐900 °C nanocomposite, revealing MoO2@MoS2 disks with an average diameter of ≈100 nm. Within these disks, Fe2Mo3O8 particles with regular polyhedral structures and an average diameter of 300 nm could be observed. Figure 2b presents the SEM energy dispersive spectrometer (EDS) mapping of the Fe2Mo3O8/MoO2@MoS2‐900 °C nanocomposite, clearly distinguishing the Fe2Mo3O8 and MoO2@MoS2 compounds based on their distinct morphologies. Elemental maps of Mo, S, Fe, and O (Figure 2c–f) corresponded to the regions identified in Figure 2b. The Mo and S elements were evenly distributed throughout the composite, indicating the predominance of MoS2. Fe and O elements were concentrated at the large particles Fe2Mo3O8. Additionally, a small amount of O was observed outside of the Fe2Mo3O8 particles, which could be attributed to the substitution of S by O in the MoS2 structure, resulting in the formation of the MoO2@MoS2 phase. SEM and EDS mapping images proved the formation of Fe2Mo3O8/MoO2@MoS2 ternary heterojunction, which could enhance the material gas‐sensing performance.
Figure 2.

a) SEM image and b) EDS mapping of Fe2Mo3O8/MoO2@MoS2‐900 °C, c–f) Elemental maps of Mo, S, Fe, and O in Fe2Mo3O8/MoO2@MoS2‐900 °C. g) TEM image of Fe2Mo3O8/MoO2@MoS2‐900 °C. HR‐TEM images of h) MoO2@MoS2 and i) Fe2Mo3O8 in Fe2Mo3O8/MoO2@MoS2‐900 °C.
The TEM image of the as‐prepared Fe2Mo3O8/MoO2@MoS2‐900 °C nanocomposite is shown in Figure 2g. The lattice structure of MoO2@MoS2 was observed in Figure 2h, with lattice stripe spacing of 0.62, 0.26, and 0.28 nm corresponding to (002), and (101) planes of MoS2, and (101) plane of MoO2, respectively. The presence of alternating S and O atoms in MoO2@MoS2 resulted in lattice stress and significant lattice distortion, which in turn provided a large number of active sites for gas‐sensing response. Figure 2i shows the lattice structure of the Fe2Mo3O8, revealing a lattice stripe spacing of 0.18 nm corresponding to the (204) plane. This observation confirms the presence of Fe2Mo3O8 in the Fe2Mo3O8/MoO2@MoS2‐900 °C nanocomposite.
X‐ray diffraction (XRD) was used to analyze the structures of α‐Fe2O3/MoS2 nanocomposite annealed at 600, 750, 900, and 1050 °C (labeled FeMoOS‐600, FeMoOS‐750, Fe2Mo3O8/MoO2@MoS2‐900, and FeMoOS‐1050 °C) respectively. The XRD from the Fe2Mo3O8/MoO2@MoS2‐900 °C composite is shown in Figure 3a. The diffraction peaks at 17.6, 25.1, and 35.9° are assigned to the (002), (102), and (112) crystal planes of the Fe2Mo3O8. The diffraction peaks at 26.0, 36.8, 36.9, 37.0, 37.4, and 49.5° were assigned to the (−111), (200), (111), (−211), (−202), and (−301) crystal planes of monoclinic MoO2. The diffraction peaks at 14.4, 32.7, 39.5, 49.8, and 58.3 were assigned to the (002), (100), (103), (105), and (110) crystal planes of MoS2. Comparing the peak intensities and positions of the samples indicated that the nanocomposite was composed of Fe2Mo3O8, MoO2, and MoS2. Figure 3b illustrates the evolution of the nanocomposites annealed at different temperatures. It was obvious that MoO2 was formed in FeMoOS‐750, FeMoOS‐1050, and Fe2Mo3O8/MoO2@MoS2‐900 °C, while Fe2Mo3O8/MoO2 only appeared in Fe2Mo3O8/MoO2@MoS2‐900 °C, identified by the diffraction peaks at 17.6 and 25.1°. Based on XRD quantitative analysis, the mass percentages of Fe2Mo3O8, MoO2, and MoS2 in Fe2Mo3O8/MoO2@MoS2‐900 °C are 1.0%, 6.3%, and 92.7%, respectively.
Figure 3.

XRD spectra of a,b) Fe2Mo3O8/MoO2@MoS2‐900, FeMoOS‐600, FeMoOS‐750, and FeMoOS‐1050 °C. XPS spectra of c) Mo, d) S, e) O, and f) Fe peaks.
X‐ray photoelectron spectroscopy (XPS) was also used to determine the surface composition and chemical state of the Fe2Mo3O8/MoO2@MoS2‐900 °C nanocomposite, as shown in Figure 3c–f. There were two peaks in the Mo 3d spectrum at 229.85 and 232.97 eV, corresponding to the Mo4+ d5/2 and Mo4+ d3/2 in MoS2 and MoO2, respectively. There were two peaks in the S 2p spectrum at 162.73 and 163.89 eV, assigned to the S2− 2p3/2 and S2− 2p1/2, respectively. There were two peaks in the Fe 2p spectrum at 711.48 and 724.92 eV, corresponding to Fe3+ 2p3/2 and Fe3+ 2p1/2, respectively. Finally, there were two peaks at 530.93 and 532.05 eV, which corresponded to lattice oxygen (OL) and oxygen vacancies (OV), respectively. The oxygen vacancies provide active sites for gas adsorption. When the Fe2Mo3O8/MoO2@MoS2‐900 °C sensor was exposed to air, the oxygen vacancies might facilitate the activation of adsorbed O2, forming Upon exposure to NH3 gas, the will oxidize the adsorbed surface NH3. Thus, the combination of oxygen vacancies and lattice oxygen stability is crucial for optimizing the NH3 gas‐sensing performance.[ 13 ]
2.2. Gas‐Sensing Properties
2.2.1. Effect of Annealing Temperatures on Gas‐Sensing Performance
To investigate the effect of annealing temperatures ranging between 600 to 1050 °C on the NH3 gas‐sensing performance, four nanocomposites, namely FeMoOS‐600, FeMoOS‐750, Fe2Mo3O8/MoO2@MoS2‐900, and FeMoOS‐1050 °C, were synthesized. The gas‐sensing response curves of these nanocomposites exposed to 30 ppm NH3 at room temperature in 5% relative humidity (RH) are depicted in Figure 4a. The obtained gas‐sensing responses were 59%, 204%, 875%, and 409% for FeMoOS‐600, FeMoOS‐750, Fe2Mo3O8/MoO2@MoS2‐900, and FeMoOS‐1050 °C, respectively. These results indicated that the Fe2Mo3O8/MoO2@MoS2‐900 °C nanocomposite exhibited the best gas‐sensing performance. From XRD data, at the annealing temperature over 900 °C, the Fe2Mo3O8 phase disappeared, which is responsible for the decrease in the response. Hence, it can be proposed that the formation of Fe2Mo3O8 is essential for enhancing the gas sensing performance. An optimal annealing temperature of 900 °C was selected for further study.
Figure 4.

a) Gas‐sensing responses of FeMoOS‐600, FeMoOS‐750, Fe2Mo3O8/MoO2@MoS2‐900, and FeMoOS‐1050 °C toward 30 ppm of NH3 at room temperature and 5% RH. b) Gas‐sensing responses of MoS2, α‐Fe2O3/MoS2, MoO2@MoS2 and Fe2Mo3O8/MoO2@MoS2‐900 °C to 30 ppm of NH3 at room temperature and 5% RH. c) Sensitivity of Fe2Mo3O8/MoO2@MoS2‐900 °C for NH3 at concentrations of 1–50 ppm at room temperature and 5% RH. d) The calibration curve between NH3 concentration and response for Fe2Mo3O8/MoO2@MoS2‐900 °C.
2.2.2. Gas‐Sensing Performances of the Different Materials
To compare the gas‐sensing performances of different precursors and intermediate materials, gas sensors based on MoS2, α‐Fe2O3/MoS2, MoO2@MoS2, and Fe2Mo3O8/MoO2@MoS2‐900 °C nanocomposites were constructed and exposed to 30 ppm of NH3 at room temperature in 5% RH. Upon exposure to NH3, the resistances of all the materials increased, which indicated a typical p‐type semiconductor behavior. The responses 4%, 13%, 143%, and 875% were achieved from MoS2, α‐Fe2O3/MoS2, MoO2@MoS2, and Fe2Mo3O8/MoO2@MoS2‐900 °C, respectively (Figure 4b). Moreover, the Fe2Mo3O8/MoO2@MoS2‐900 °C sensor exhibited the fastest response and recovery times of ≈3 min (response) and 60 min (90% recovery) at room temperature in 5% RH. In contrast, MoS2 and α‐Fe2O3/MoS2 sensors showed a limited recovery of 30% in 17 min and were unable to recover fully. The MoO2@MoS2 demonstrated a slightly faster recovery rate of ≈60% within 17 min, although a complete recovery was not achieved. These results indicate that the gas‐sensing response and recovery performance of the original MoS2 could be significantly improved by compounding it with α‐Fe2O3 and annealing in low oxygen partial pressure to form MoO2/MoS2 heterojunction coupled with Fe2Mo3O8, resulting in a far greater response than the substrate materials.
2.2.3. Dynamic Sensing Performance of Fe2Mo3O8/MoO2@MoS2‐900 °C Sensor
The temporal dynamic reversible performance of the Fe2Mo3O8/MoO2@MoS2‐900 °C sensor was investigated by exposing it to 1, 5, 10, 20, 30, 40 and 50 ppm of NH3 with the corresponding responses of 57%, 363%, 565%, 728%, 875%, 987%, and 1086%, respectively (Figure 4c). The result suggested that the Fe2Mo3O8/MoO2@MoS2‐900 °C gas sensor was ultrasensitive and capable of detecting NH3 over a wide range of concentrations at room temperature.
A calibration curve of the responses at different NH3 concentrations is plotted in Figure 4d. A strong linear relationship between the sensor response and NH3 concentration was observed with the correlation coefficient R2 of 0.99, indicating a reliable linear response for NH3 concentrations between 1 and 9 ppm. The detection limit (DL) is calculated using the Equation (1):[ 14 ]
| (1) |
where RMS represents the root mean squared of noise. The slope was calculated from the linear response between 1 to 9 ppm NH3.[ 15 ] For the Fe2Mo3O8/MoO2@MoS2‐900 °C sensor, the DL was determined to be 3.7 ppb, which is below the typical ammonia content (1 ppm) found in the exhaled breath of kidney disease patients.[ 9 ] More importantly, it is also lower than the NH3 concentration threshold of 10 ppb, indicating early liver or kidney failure when other symptoms are not shown. Hence, the ultrasensitive Fe2Mo3O8/MoO2@MoS2‐900 °C sensor can be used not just for domestic monitoring of patients with liver or kidney disease but also for early diagnosis for the prevention of disease.
2.2.4. Selectivity of Fe2Mo3O8/MoO2@MoS2‐900 °C Sensor
In practical applications, gas sensors must operate in complex environments where multiple gases are present. Therefore, gas sensors must be highly selective to identify target gases in real environments accurately. The selectivity of the Fe2Mo3O8/MoO2@MoS2‐900 °C sensor was measured by exposing it to 30 ppm of NH3, hydrogen sulfide (H2S), nitrogen dioxide (NO2), methanol (CH3OH), carbon dioxide (CO2), ethanol (C2H5OH), acetone (C3H6O), acetaldehyde (CH3CHO), allicin (which has similar odors to NH3) and n‐hexane, at room temperature and 5% RH. The responses were 875%, 93%, 92%, 85%, 85%, 81% 73%, 58%, 22%, and 7%, respectively, as shown in Figure 5 . The response to NH3 was significantly higher than to the other gases. Such an excellent selectivity toward NH3 is critical to the reliability of the sensing result. The relatively low response to other gases might be ascribed to the high activation energy barrier for them on the surface of Fe2Mo3O8/MoO2@MoS2.[ 13 ]
Figure 5.

Responses of the Fe2Mo3O8/MoO2@MoS2‐900 °C sensor to 30 ppm of NH3 and interference gases at room temperature and 5% RH.
2.2.5. Repeatability and Long‐Term Stability of Fe2Mo3O8/MoO2@MoS2‐900 °C Sensor
The repeatability of the Fe2Mo3O8/MoO2@MoS2‐900 °C sensor was assessed by subjecting it to five response–recovery cycles of exposure to 30 ppm of NH3 at room temperature in 5% RH. As shown in Figure 6a, there is no obvious change in responses after five cycles and remained at ≈875%. However, the recovery gradually slowed down, possibly due to the diffusion of the NH3 between the MoS2 layers. However, as the response is maintained constant, it can be assumed that the tail of the background and the response peak are contributed by different mechanisms. Further modeling and structural control will be carried out in the future in order to optimize the recovery performance. Nevertheless, the Fe2Mo3O8/MoO2@MoS2‐900 °C sensor offers excellent repeatability.
Figure 6.

a) Repeatability and b) long‐term stability of Fe2Mo3O8/MoO2@MoS2‐900 °C sensor. Responses of Fe2Mo3O8/MoO2@MoS2‐900 °C sensor to c) 30 ppm of NH3 at 5% and 60% RH and d) 1–5 ppm of NH3 at 90% RH.
The long‐term stability was evaluated by measuring the response of the Fe2Mo3O8/MoO2@MoS2‐900 °C sensor to 30 ppm of NH3 once a week for four consecutive weeks, after each test, keeping the sensor in a dry room temperature air environment (Figure 6b). The results indicated that the response value decreased 19% from 875% to 711% after four weeks. The results further showed that the Fe2Mo3O8/MoO2@MoS2‐900 °C sensor had remarkable repeatability and long‐term stability for the NH3 detection.
2.2.6. Effect of Humidity on Gas‐Sensing Performance of Fe2Mo3O8/MoO2@MoS2‐900 °C Sensor
Humidity is a dominant component that can interfere with the detection of gases in exhaled human breath and affect the performance of gas sensors. To evaluate this, the performance of the Fe2Mo3O8/MoO2@MoS2‐900 °C sensor when exposed to 30 ppm of NH3 was compared at 5% and 60% RH. The performance was also evaluated at 90% RH by exposing to 1–5 ppm NH3. Before introducing NH3, the Fe2Mo3O8/MoO2@MoS2‐900 °C sensor was stabilized to moist air with the desired RH. During the measurement, RH in the gas‐sensing chamber was controlled by passing dry air through a cylindrical water container while introducing dry NH3. As shown in Figure 6c, by increasing the RH to 60%, the response to 30 ppm of NH3 was decreased to 581% with respect to 875% at 5% RH. At 90% RH, which was close to the RH of exhaled human breath, the responses of 32%, 62%, 90%, 110%, and 134% were achieved by exposing to 1, 2, 3, 4, and 5 ppm of NH3, respectively, at room temperature, as shown in Figure 6d.
The above results indicate that the response of the Fe2Mo3O8/MoO2@MoS2‐900 °C sensor to a specific concentration of NH3 decreased with rising RH, although it still preserved a comparatively high response level. At high humidities, water molecules will compete with O2 and NH3 for their adsorption on the active sites, resulting in hindered NH3 oxidation kinetics and reduced sensitivity. Additionally, the adsorbed water could also disrupt the heterostructure and reduce the electron transport ability of the sensor.[ 16 ]
In the realm of 2D material‐based, solid‐state ammonia gas sensors, various strategies such as defect engineering, doping engineering, noble metal modification, and heterojunction structures are usually employed to enhance sensor performance. For instance, the defect‐engineered 2D SnS2 sensor with S vacancies exhibited a response of 420% to 500 ppm NH3 at room temperature with a detection limit of 20 ppm.[ 25 ] NO2‐doped graphene sensor, representing doping engineering, showed a response of 14% to 30 ppm NH3 at room temperature with a relatively low DL of 200 ppb.[ 26 ] For noble metal modification, the Pt‐MoS2 sensor demonstrated a response of 36% to 70 ppm NH3 at room temperature, with a lower DL of 130 ppb.[ 27 ] With a designed heterojunction, the MoS2/ZnO sensor delivered a response of 70% to 30 ppm NH3, with a poor DL of 920 ppb.[ 28 ] Additionally, the GaN/rGO sensor shows a response of 92% to 200 ppm NH3, with a DL of 28 ppb.[ 29 ] More sensor performance comparisons can be found in Table 1 . In our work, the Fe2Mo3O8/MoO2@MoS2‐900 °C sensor combines the advantages of defect engineering through O doping in MoS2 and a heterojunction structure by forming Fe2Mo3O8 on MoO2@MoS2. To our knowledge, it outperformed most of the NH3 sensors in the literature based on a range of technologies, as we summarized here, with an outstanding response of 875% to 30 ppm NH3 at a superior DL of 3.7 ppb. Notably, the Fe2Mo3O8/MoO2@MoS2‐900 °C sensor exhibits exceptional sensitivity, selectivity, and broad dynamic range, which underscore its potential for clinical diagnosis of liver and kidney disease at an early stage. Compared with the literature, our Fe2Mo3O8/MoO2@MoS2‐900 °C sensor is the only device offering a detection limit below the threshold of 10 ppb NH3 found in the breath of patients who might be suffering from liver or kidney failure.[ 2 ]
Table 1.
Responses of different materials to NH3 gas at room temperature.
| Material | NH3 Concentration [ppm] | DL [ppb] | Ref. | ||||
|---|---|---|---|---|---|---|---|
| 1 | 5 | 10 | 20 | 30 | |||
| Response [%] | |||||||
| CeO2 | 200 | 250 | 350 | – | 500 | 500 | [17] |
| Ti3C2 Mxene | – | – | 0.62 | – | – | 10 000 | [18] |
| Graphene | – | – | – | 13 | – | 27 | [19] |
| Ti3C2Tx/WO3 | 22 | – | – | – | – | 1000 | [20] |
| MoS2/MoO3 | 15 | – | 24 | 30 | 39 | 1000 | [21] |
| MoS2/Co3O4 | 26 | 64 | – | – | – | 100 | [11] |
| NiO/MoS2 | 31 | 45 | 63 | 80 | 90 | 250 | [22] |
| MoS2/CuO | – | 8 | – | – | – | 5000 | [23] |
| MoS2/ZnO | 18 | 25 | 37 | – | – | 12 | [24] |
| 2D SnS2 | – | – | – | 150 | – | 20 000 | [25] |
| NO2‐doped graphene | – | – | – | – | 14 | 200 | [26] |
| Pt‐MoS2 | – | 17 | – | – | 24 | 130 | [27] |
| MoS2/ZnO | – | 9 | 21 | 35 | 70 | 920 | [28] |
| GaN/rGO | 23 | 42 | – | 60 | – | 28 | [29] |
| Fe2Mo3O8/MoO2@MoS2‐900 °C | 57 | 363 | 565 | 728 | 875 | 3.7 | This work |
“‐” The response was not given in the literature.
2.2.7. Detection of NH3 in the Exhaled Breath of Kidney Disease Patients by Fe2Mo3O8/MoO2@MoS2‐900 °C Sensor
To evaluate the NH3 sensor's performance in the diagnosis of kidney disease, the NH3 concentration in the exhaled breath was measured using the Fe2Mo3O8/MoO2@MoS2‐900 °C sensor. In this experiment, internationally recognized definitions for the stages of chronic kidney disease (CKD) were adopted. Early‐stage and late‐stage kidney disease patients are referred to the CKD1‐2 and CKD5 stages, respectively. With patients’ consent, the oral exhaled breath was sampled from three healthy people (H1, H2, H3), three early‐stage kidney disease patients (P1, P2, and P3), and three late‐stage kidney disease patients (P4, P5, and P6) using medical grade Tedlar gas bags before their morning tooth brushing routine. The RH of these collected human exhaled gas samples was measured to be approximately 90% using a hygrometer. The Fe2Mo3O8/MoO2@MoS2‐900 °C sensor was preconditioned in a 90% RH air environment. Once the resistance of the sensor was stabilized, the exhaled breath samples were introduced. To measure the NH3 concentration using a homemade gas‐sensing test station,[ 30 , 31 ] as depicted in Figure 7 .
Figure 7.

Schematic diagram of human exhaled breath collection and gas‐sensing test.
The expiratory responses of three healthy people, H1, H2, and H3, were relatively low at 24%, 16%, and 18%, respectively, as illustrated in Figure 8a. Subsequently, we examined the ability of the Fe2Mo3O8/MoO2@MoS2‐900 °C sensor to distinguish between patients with early‐ and late‐stage kidney disease. Figure 8b shows slightly higher responses from three early‐stage patients (P1, P2, and P3) of 38%, 34%, and 36%. Figure 8c presents the response values from three late‐stage patients (P4, P5, and P6) of 81%, 92%, and 72%. The reliability of the reading is demonstrated by the repeated measurements from patient P1 (early‐stage) and patient P4 (late‐stage) three times. Figure 8d,e indicated that the responses for P1 were 38%, 38%, and 40%, while P4 showed responses of 78%, 87%, and 85%. The relative standard deviations are 2.99% and 5.67% for patients P1 and P4, respectively, indicating good consistency and reliability. Significant differentiation in the exhaled breath responses between two groups of patients at CKD1‐2 and CKD5 stages is distinctively observed. Hence, the sensitivity and the dynamic range of the Fe2Mo3O8/MoO2@MoS2‐900 °C sensor are sufficient for distinguishing the different stages of kidney disease, which has the potential for clinical application.
Figure 8.

Application of the Fe2Mo3O8/MoO2@MoS2‐900 °C sensor into exhaled breath detection: Responses to the exhaled breath of a) three healthy people (H1, H2, and H3), b) three early‐stage patients (P1, P2, and P3) and c) three late‐stage patients (P4, P5, and P6). Repeatability of the response to the exhaled breath of d) P1 and e) P4. f) Linear fit of the responses to NH3 concentrations of 1–5 ppm at room temperature and 90% RH, and the responses to the exhaled breath of three healthy people and six patients.
Furthermore, a calibration curve for the sensor's response to NH3 concentrations ranging from 1–5 ppm at 90% RH was established and presented in Figure 8f. The high correlation coefficient (R2 = 0.99) confirmed a reliable linear relationship, which is important for the accuracy of measurements. This calibration curve was then utilized to determine the NH3 concentration in the exhaled breath of kidney disease patients. For healthy people H1‐H3 and patients P1–P6, the NH3 concentrations were determined to be 0.56, 0.24, 0.32, 1.11, 0.95, 1.03, 2.82, 3.25, and 2.46 ppm, aligning with previously reported values in literature.[ 7 ] Therefore, the Fe2Mo3O8/MoO2@MoS2‐900 °C sensor has been demonstrated as an affordable but reliable diagnosis device to monitor kidney disease by measuring the concentration of NH3 in the exhaled breath.
2.2.8. Sensing Mechanism
During the test, it was observed that MoO2@MoS2, Fe2Mo3O8, and Fe2Mo3O8/MoO2@MoS2‐900 °C nanocomposites all showed p‐type gas‐sensing behavior with holes as the major charge carriers. MoO2@MoS2 is the main conductive component in the nanocomposite, while Fe2Mo3O8 offers active catalytic sites for the adsorption and reaction of the NH3 molecule. The band gaps of MoO2@MoS2 and Fe2Mo3O8 were determined as 2.23 and 1.14 eV (Figure S3a, Supporting Information) from the UV–vis absorption measurements. The work functions, W f, of MoO2@MoS2 and Fe2Mo3O8 were identified to be 5.22 and 6.92 eV, using the UV photoelectron spectroscopy and Equation (2):[ 32 ]
| (2) |
Here, hv is the ultraviolet photon energy (21.22 eV), E Cutoff represents the secondary electron cutoff edge, and E Fermi represents the fermi edge (Figure S3b, Supporting Information). At the heterojunction between the MoO2@MoS2 and Fe2Mo3O8, electrons are expected to be transferred from MoO2@MoS2 to Fe2Mo3O8 due to the smaller work function of MoO2@MoS2, resulting in a hole depletion layer on the surface of p‐type Fe2Mo3O8. When the gas sensor is exposed to air, O2 can easily adsorbed on the surface of Fe2Mo3O8, which is electron‐rich. During this process, oxygen molecules acted as electron acceptors. The O2 adsorption increases the carrier (hole) concentration in the p‐type Fe2Mo3O8, which effectively reduces the thickness of the hole depletion layer. Thus, the resistance of the nanocomposite is decreased. When NH3 is adsorbed, it reacts with to generate NO and H2O, while the Fe2Mo3O8 will gain electrons. Consequently, the majority carrier (hole) concentration in the composite decreases, leading to an increase in the thickness of the depletion layer and the resistance of the composite material.
Energy band diagrams of the Fe2Mo3O8/MoO2@MoS2‐900 °C composite when exposed to different gases are shown in Figure 9a–d. The sequence of the surface reactions could be described in the following:[ 11 ]
| (3) |
| (4) |
| (5) |
| (6) |
Figure 9.

The diagram of the energy band of Fe2Mo3O8/MoO2@MoS2‐900 °C: a) no contact between the materials, b) contact between the materials in a vacuum, c) contact between the materials in air, and d) contact between the materials in NH3. XPS analysis and theoretical calculation based on DFT: e) XPS spectra of N─O peak, f) the adsorption energy of NH3.
To verify the reaction process, XPS analysis was conducted on the Fe2Mo3O8/MoO2@MoS2‐900 °C sensor after exposure to NH3 gas for 0, 2, and 30 min and then recovered for 2 h (Figure 9e). The analysis revealed no significant changes in the XPS signals from Mo 3d, O 1s, and S 2p. However, a new, albeit weak, peak at 402.6 eV associated with the N 1s peak was detected in the sample exposed to NH3 for 30 min, indicating the formation of N─O bonds and the presence of NO moiety.[ 33 ] The intensity of the N─O peak decreased after a 2‐h recovery period, indicating that the adsorption of the reaction product is reduced during the recovery process at room temperature. Thus, the desorption energy barrier must be at a similar scale as the thermal energy at room temperature. Such property is essential for the reproducibility of the response to NH3 at room temperature, which is essential for the sensor to be operated under ambient conditions. The XPS results confirmed that NO is the reaction product for the NH3 catalytically oxidation.
Furthermore, to investigate the influence of Fe2Mo3O8 on the gas‐sensing response of composite materials, DFT was conducted to determine the adsorption energy of NH3 on various materials and the corresponding change in free energy during NH3 oxidation. In Figure 9f, the maximum adsorption energies of NH3 on Fe2Mo3O8/MoO2@MoS2‐900 °C, MoO2/MoS2, α‐Fe2O3/MoS2 and MoS2 were −0.98, −0.74, −0.67 and −0.15 eV, respectively. The negative values of the adsorption energies indicated an exothermic adsorption process. Consequently, NH3 adsorption was most favorable on Fe2Mo3O8/MoO2@MoS2‐900 °C, suggesting that it facilitated the subsequent NH3 oxidation reaction.
The oxidation reaction mechanism and the associated free energy change (ΔG) of NH3 across four catalysts, including Fe2Mo3O8/MoO2@MoS2‐900 °C, MoO2/MoS2, α‐Fe2O3/MoS2 and MoS2 were illustrated in Figure 10 . The reaction involves six steps with corresponding intermediates. 1) NH3 initially adsorbs on the active site of the catalyst to form *NH3. 2) NH3 is partially dehydrogenated to generate *NH2. 3) *NH2 is further dehydrogenated to form *NH. 4) *NH reacts with O to produce *HNO. 5) *HNO releases an H atom, resulting in the formation of *NO. 6) Finally, *NO desorbs to release gaseous NO and recover the active site. For Fe2Mo3O8/MoO2@MoS2‐900 °C and MoS2, the rate‐limiting steps are determined by the calculation of the energy barriers. On the Fe2Mo3O8/MoO2@MoS2‐900 °C and MoS2, the rate‐limiting steps are the first dehydrogenation with the ΔG values of 1.95 eV and 2.75 eV, respectively. On the MoO2@MoS2 and α‐Fe2O3/MoS2, the rate‐limiting steps involve the deprotonation of HNO, exhibiting ΔG values of 2.66 and 3.93 eV, respectively. It was noteworthy that Fe2Mo3O8/MoO2@MoS2‐900 °C shows the lowest ΔG at the rate‐controlling step, indicating its superior catalytic efficiency in NH3 oxidation. The result suggested that Fe2Mo3O8 forms a heterogeneous interface with MoO2@MoS2 to facilitate electron transfer and activate the adsorbed O2, leading to favorable NH3 adsorption and catalytic oxidation at room temperature. Such a unique interface structure guarantees a high response and very low detection limit for NH3, which is required for the diagnosis of early‐stage kidney disease.
Figure 10.

The free energy changes during NH3 oxidation reaction with Fe2Mo3O8/MoO2@MoS2‐900 °C, MoO2/MoS2, α‐Fe2O3/MoS2 and MoS2.
3. Conclusion
In summary, Fe2Mo3O8/MoO2@MoS2‐900 °C nanocomposite was successfully synthesized and optimized for the room‐temperature NH3 gas sensor with potential application in noninvasive diagnosis of different stages of kidney disease. The sensor exhibited an extremely high response with an ultra‐low detection limit of 3.7 ppb for NH3. Through analyzing exhaled gases, the Fe2Mo3O8/MoO2@MoS2‐900 °C sensor demonstrated a unique ability to differentiate between early‐ and late‐stage kidney disease in patients. Moreover, the gas‐sensing mechanism and associated intermediates were elucidated by XPS detections and DFT calculations. The excellent gas‐sensing performance of the sensor was established due to the formation of heterojunctions between MoS2, MoO2, and Fe2Mo3O8 and the exceptional NH3 adsorption and catalytic oxidation activity of the Fe2Mo3O8. The remarkable capabilities of the Fe2Mo3O8/MoO2@MoS2‐900 °C sensor offer a viable method for analyzing exhaled gases in individuals with kidney disease, suggesting a novel strategy for early diagnosis and management of kidney disease.
4. Experimental Section
Chemical Reagents
Ferric chloride hexahydrate (FeCl3•6H2O, analytical research (AR) grade), sodium molybdate dihydrate (Na2MoO4•2H2O, AR grade), and thiourea (CH4N2S, AR grade) were purchased from Shanghai Macklin Biochemical Co., Ltd, China. Citric acid monohydrate (C6H8O7•H2O, AR grade) and deionized water (H2O) were purchased from Shanghai Sinopharm Chemical Reagent Co., Ltd., China. Ethyl alcohol (C2H5OH, AR grade) was purchased from Zhenxing No.1 Chemical Plant, China.
Synthesis of Sensing Materials
A typical procedure was used to synthesize MoS2. First, 484 mg of Na2MoO4•2H2O and 684 mg of CH4N2S were mixed and dissolved in 70 mL of deionized water, and the mixture was stirred for 30 min. Then, 462 mg of C6H8O7•H2O was added to the solution and stirred for an additional 10 min. The resulting solution was transferred into a 100 mL autoclave and heated at 200 °C for 21 h. Finally, the mixture was washed three times with deionized water and anhydrous ethanol using centrifugation, and heated at 60 °C for 12 h.
To synthesize the α‐Fe2O3/MoS2 complex, 460 mg of FeCl3•6H2O was dissolved in 60 mL deionized water and stirred for 30 min. Then, 140 mg of MoS2 was added to the solution, which was stirred for another 20 min. The solution was transferred into a 100 mL autoclave and heated at 180 °C for 12 h.
To synthesize the FeMoOS‐600, FeMoOS‐750, FeMoOS‐1050, and Fe2Mo3O8/MoO2@MoS2‐900 °C, α‐Fe2O3/MoS2 composite was annealed at different temperatures,100 mg samples of α‐Fe2O3/MoS2 were evenly spread on small porcelain boats and placed in a tube furnace. The samples were then annealed at 600, 750, 900, and 1050 °C in a low oxygen partial pressure atmosphere.
Material Characterization
The morphologies of the sensing materials were examined using SEM (S‐4800N, Hitachi, Japan) and TEM (JEM‐2100F, JEOL, Japan). The crystallinity was investigated using XRD (D8 ADVANCE, Bruker, Cu kα, 380 eV, Germany). The composition was determined via XPS (ESCAlab250, ThermoFisher, Al X‐ray source, 1486.6 eV, USA). The concentration of Fe in the nanocomposite was determined by glow discharge mass spectrometry (GD‐MS) (Auto Concept GD 90, Mass Spectrometry Instruments Co., Ltd., UK). The band gap and work function of the materials were measured by UV‐visible spectroscopy (U‐4100, Hitachi, Japan) and UV photoelectron spectroscopy (250Xi, ThermoFisher, He I line, 21.22 eV, USA), respectively.
Gas Sensor Fabrication
The Fe2Mo3O8/MoO2@MoS2‐900 °C sensor was fabricated on interdigital electrodes composed of Fe2Mo3O8/MoO2@MoS2‐900 °C sensing materials and interdigital electrodes, as well as for the other sensors based on MoO2@MoS2, α‐Fe2O3/MoS2, MoS2, FeMoOS‐600, FeMoOS‐750, and FeMoOS‐1050 °C materials. The interdigital electrode is 1 × 1 × 0.05 cm3 in size and composed of an Al2O3 ceramic substrate and Au electrode on it. The interdigital electrodes were cleaned under sonication with acetone, deionized water, and ethanol before use. Then, the annealed Fe2Mo3O8/MoO2@MoS2‐900 °C suspension was carefully dripped onto the surface of the electrode. The electrode was then placed back into the oven at 60 °C for 6 h to form a gas sensor based on the Fe2Mo3O8/MoO2@MoS2‐900 °C sensing material.
Gas‐Sensing Response Measurement
To investigate the gas‐sensing capabilities, a dedicated gas‐sensing apparatus was constructed with a gas cylinder, a plastic conduit, a gas‐sensing chamber, and a resistance acquisition device, as shown in Figure 7. The cylinders were filled with dry air and dry NH3 respectively. Within the gas‐sensing chamber, there was a cylindrical cavity with a volume of 300 mL and several venting orifices. The interdigital electrode loaded with the Fe2Mo3O8/MoO2@MoS2‐900 °C nanocomposite was fixed in the cylindrical cavity using a platinum electrode clamp. The bubbling method was applied to adjust the moisture of the injected gas. Typically, NH3 with different relative humidity was obtained by mixing the wet air with dry NH3, and all the gas flow ratios were regulated by mass flow controllers. The final gas flow rate was fixed at 600 mL min−1. The change in the resistance of the sensor was measured by the Keithley 2701 resistance acquisition device. The sensor response was calculated using the equation Response (R) = |Rg – Ra| / Ra × 100%, where Rg represented the sensor resistance in the presence of NH3, and Ra represented the sensor resistance in air. The gas‐sensing experiments were conducted at room temperature and 5% RH, if without any special indication.
In the exhaled breath test experiments of kidney disease patients, informed consent was given freely and voluntarily by all participants after receiving comprehensive information and explanations. The study was reviewed and approved by the Medical Ethics Committee of Shanghai Pudong New Area People's Hospital, with the approval number K63. The exhaled breath was collected in the mouth using a Tedlar bag (Dupont USA), which is constructed from polyvinyl fluoride film with a film thickness of ≈0.05 mm. The gas sample collection procedure involves opening the gas outlet of the gas sampling bag, positioning the volunteer's mouth toward the outlet, and exhaling slowly for a duration exceeding 20 s. Subsequently, the gas outlet is sealed to conclude the collection process.
Theoretical Calculation
DFT was used as implemented in the Vienna Ab initio simulation package (VASP) for all calculations. The exchange‐correlation potential is described by using the generalized gradient approximation of Perdew‐Burke‐Ernzerhof. The projector augmented‐wave method is employed to treat interactions between ion cores and valence electrons. The plane‐wave cutoff energy was fixed to 400 eV. Given structural models were relaxed until the Hellmann‐Feynman forces were <−0.01 eV Å−1 and the change in energy <10−5 eV was attained. The basic structure of the Fe2Mo3O8/MoO2@MoS2‐900 °C composite for DFT calculation consists of three layers of distinct compounds arranged in a heterostructure. The bottom layer comprises MoS2 with the (002) plane serving as the outer crystal plane, the middle layer is MoO2 with the (−111) plane as the outer crystal plane, and the top layer consists of Fe2Mo3O8 with the (102) plane as the outer crystal plane. This composite material is situated within a vacuum layer with a thickness of ≈3.5 nm. The adsorption energy (E ads) was calculated as:
| (7) |
where E(system), E(catalyst), and E(species) are the total energy of the optimized system with adsorbed species, the isolated catalyst, and species, respectively.
The Gibbs free energy change is defined as:
| (8) |
where ΔE is the electronic energy calculated with VASP, ΔE ZPE, and ΔS are the zero‐point energy difference and the entropy change between the products and reactants, respectively, and T is the temperature (298.15 K).
Conflict of Interest
The authors declare no conflict of interest.
Supporting information
Supporting Information
Acknowledgements
X.L. and W.Z. contributed equally to this work. The authors acknowledge the financial support of the National Natural Science Foundation of China (Grant No. 22274164), the Space Application System of China Manned Space Program, the National Key R&D Program of China (Grant No. 2021YFF0700102), and the Shanghai Technical Platform of Testing on Inorganic Materials (Grant No. 19DZ2290700).
Li X., Zeng W., Zhuo S., Qian B., Chen Q., Luo Q., Qian R., Highly Sensitive Room‐Temperature Detection of Ammonia in the Breath of Kidney Disease Patients Using Fe2Mo3O8/MoO2@MoS2 Nanocomposite Gas Sensor. Adv. Sci. 2024, 11, 2405942. 10.1002/advs.202405942
Data Availability Statement
The data that support the findings of this study are available from the corresponding author upon reasonable request.
References
- 1. Wang J., Li Z., Zhang S., Yan S., Cao B., Wang Z., Fu Y., Sens. Actuators B: Chem. 2018, 255, 862. [Google Scholar]
- 2. Freddi S., Gonzalez M. C. R., Carro P., Sangaletti L., De Feyter S., Angew. Chem., Int. Ed. 2022, 61, e202200115. [DOI] [PubMed] [Google Scholar]
- 3. Li D., Xu X., Li Z., Wang T., Wang C., TrAC‐Trend Anal. Chem. 2020, 127, 115890. [Google Scholar]
- 4. Wang L., Teleki A., Pratsinis S. E., Gouma P. I., Chem. Mater. 2008, 20, 4794. [Google Scholar]
- 5. André L., Desbois N., Gros C. P., Brandès S., Dalton T. 2020, 49, 15161. [DOI] [PubMed] [Google Scholar]
- 6. Van Toan N., Hung C. M., Hoa N. D., Van Duy N., Thi Thanh Le D., Thi Thu Hoa N., Viet N. N., Phuoc P. H., Van Hieu N., J. Hazard. Mater. 2021, 412, 125181. [DOI] [PubMed] [Google Scholar]
- 7. Kan Z., Shi F., Yang L., Zhou Q., Zhang Y., Qi Y., Zhang H., Dong B., Ren L., Song H., Xu L., J. Mater. Chem. A 2023, 11, 24299. [Google Scholar]
- 8. Wang C., Cai Y., Zhou W., Chen P., Xu L., Han T., Hu Y., Xu X., Liu B., Yu X., ACS Appl. Mater. Interfaces 2022, 14, 12630. [DOI] [PubMed] [Google Scholar]
- 9. Zhao H., Liu L., Lin X., Dai J., Liu S., Fei T., Zhang T., ACS Sens. 2020, 5, 346. [DOI] [PubMed] [Google Scholar]
- 10. Yang C., Wang Y., Wu Z., Zhang Z., Hu N., Peng C., Nanomaterials 2022, 12, 901. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11. Zhang D., Jiang C., Li P., Sun Y., ACS Appl. Mater. Interfaces 2017, 9, 6462. [DOI] [PubMed] [Google Scholar]
- 12. Wang W., Zhen Y., Zhang J., Li Y., Zhong H., Jia Z., Xiong Y., Xue Q., Yan Y., Alharbi N. S., Hayat T., Sens. Actuators B: Chem. 2020, 321, 128471. [Google Scholar]
- 13. Al‐Hashem M., Akbar S., Morris P., Sens. Actuators B: Chem. 2019, 301, 126845. [Google Scholar]
- 14. Li J., Lu Y., Ye Q., Cinke M., Han J., Meyyappan M., Nano Lett. 2003, 3, 929. [Google Scholar]
- 15. Zhang H., Zhang Z., Li Z., Han H., Song W., Yi J., Nat. Commun. 2023, 14, 3495. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16. Xiong Y., Liu W., Wu K., Liu T., Chen Y., Wang X., Tian J., J. Alloy. Compd. 2022, 927, 166962. [Google Scholar]
- 17. Li P., Wang B., Qin C., Han C., Sun L., Wang Y., Ceram. Int. 2020, 46, 19232. [Google Scholar]
- 18. Wu M., He M., Hu Q., Wu Q., Sun G., Xie L., Zhang Z., Zhu Z., Zhou A., ACS Sens. 2019, 4, 2763. [DOI] [PubMed] [Google Scholar]
- 19. Li Q., Chen W., Liu W., Sun M., Xu M., Peng H., Wu H., Song S., Li T., Tang X., Appl. Surf. Sci. 2022, 586, 152689. [Google Scholar]
- 20. Guo X., Ding Y., Kuang D., Wu Z., Sun X., Du B., Liang C., Wu Y., Qu W., Xiong L., He Y., J. Colloid. Interf. Sci. 2021, 595, 6. [DOI] [PubMed] [Google Scholar]
- 21. Singh S., Deb J., Sarkar U., Sharma S., ACS Sustainable Chem. Eng. 2021, 9, 7328. [Google Scholar]
- 22. Zhang D., Jin Y., Cao Y., J. Mater. Sci. Mater. Electron. 2019, 30, 573. [Google Scholar]
- 23. Sharma S., Kumar A., Singh N., Kaur D., Sens. Actuators B: Chem. 2018, 275, 499. [Google Scholar]
- 24. Zhang D., Jiang C., Sun Y., J. Alloy. Compd. 2017, 698, 476. [Google Scholar]
- 25. Qin Z., Xu K., Yue H., Wang H., Zhang J., Ouyang C., Xie C., Zeng D., Sens. Actuators B: Chem. 2018, 262, 771. [Google Scholar]
- 26. Zanjani S. M. M., Sadeghi M. M., Holt M., Chowdhury S. F., Tao L., Akinwande D., Appl. Phys. Lett. 2016, 108, 033106. [Google Scholar]
- 27. Park J., Mun J., Shin J., Kang S., R. Soc. 2018, 5, 181462. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 28. Gond R., Shukla P., Prakash B., Rawat B., ACS Appl. Electron. Mater. 2024, 6, 2728. [Google Scholar]
- 29. Li D., Han D., Chen Y., Hong Y., Duan Q., Wang H., He X., Zhao L., Wang W., Sang S., Sens. Actuators B: Chem. 2024, 403, 135209. [Google Scholar]
- 30. Li W., Li H., Qian R., Zhuo S., Ju P., Chen Q., Nanomaterials 2022, 12, 1300. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 31. Zhou L., Qian R., Zhuo S., Chen Q., Wen Z., Li G., J. Saudi. Chem. Soc. 2020, 24, 364. [Google Scholar]
- 32. Pandiyan R., Elhmaidi Z. O., Sekkat Z., Abd‐lefdil M., Khakani M., Appl. Surf. Sci. 2017, 396, 1562. [Google Scholar]
- 33. Herranz T., Deng X., Cabot A., Liu Z., Salmeron M., J. Catal. 2011, 283, 119. [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
Supporting Information
Data Availability Statement
The data that support the findings of this study are available from the corresponding author upon reasonable request.
