Abstract
Electromagnetic tracking (EMT) can benefit image-guided interventions in cases where line of sight is unavailable. However, EMT can suffer from electromagnetic distortion in the presence of metal instruments. Metal instruments are widely used in laparoscopic surgery, ENT surgery, arthroscopy and many other clinical applications. In this work, we investigate the feasibility of tracking such metal instruments by placing the inductive sensor within the instrument shaft. We propose a magnetostatic model of the field within the instrument, and verify the results experimentally for frequencies from 6 kHz to 60 kHz. The impact of the instrument’s dimensions, conductivity and transmitting field frequency is quantified for ranges representative of typical metal instruments used in image-guided interventions. We then performed tracking using the open-source Anser EMT system and quantify the error caused by the presence of the rod as a function of the frequency of the eight emitting coils for the system. The work clearly demonstrates why smaller tool diameters (less than 8 mm) are less susceptible to distortion, as well as identifying optimal frequencies (1 kHz to 2 kHz) for transmitter design to minimise for distortion in larger instruments.
Keywords: electromagnetic tracking, surgical navigation, inductive sensor, magnetic modelling
1. Introduction
When line of sight is unavailable, electromagnetic tracking (EMT) is the gold standard for surgical navigation. It has seen use across numerous clinical applications such as bronchoscopy [1,2,3], orthopaedic surgery [4] and cardiovascular surgery [5]. Furthermore, a number of commercial EMT systems are available, including the NDI Aurora and Trakstar (Northern Digital Inc., Waterloo, ON, Canada) and the Polhemus Fastrak (Polhemus Inc., Colchester, VT, USA). In ideal environments, these systems can routinely achieve sub-millimetre tracking accuracy [6]. Difficulties arise in applications which necessitate the tracking of metallic instruments [7,8]. These procedures include needle tracking [9], the tracking of ultrasound probes [10,11] and orthopaedic procedures [12,13]. Such instruments can cause distortion of the magnetic field which is used for sensor localisation, resulting in larger tracking errors. A number of approaches have been proposed to minimise the impact of magnetic field distortion in metallic instrument tracking. One possibility is to offset the electromagnetic sensor from the instrument shaft [14]. However, this can lead to amplified tracking errors in orientation estimates [15] and increased instrument diameters. Another possibility is the fusion between EMT and optical tracking [16]; however, the resulting system is still susceptible to visual occlusion. In this paper, we investigate the feasibility of performing EMT with a sensor placed within the shaft of metallic surgical instruments. To this end, we investigate the shielding effect of a nonmagnetic conductive rod using analytical and experimental approaches. The tracking accuracy in the presence of metallic surgical instruments is also quantified.
2. Mathematical Formalism
Electromagnetic fields are governed by Maxwell’s equations which, in the case of linear, isotropic and nonmagnetic media, are given by [17]
(1) |
(2) |
(3) |
where is the magnetic flux density, is the electric field, is the conduction current density and and are, respectively, the electrical permittivity and magnetic permeability of free space. However, in electromagnetic tracking, we are only interested in slowly oscillating magnetic fields. Therefore, in this work, we limit our analysis to quasistatic magnetic fields. In this regime, Ampere’s law reduces to
(4) |
To investigate the shielding effect of a surgical instrument, we approximate it as an infinite cylindrical rod with inner radius , outer radius and conductivity . Such a cylindrical rod can be seen in Figure 1. We adopt the obvious choice of a cylindrical coordinate system with the rod axis oriented along the z axis. In light of (3), we introduce a magnetic vector potential such that
(5) |
Next, we express Maxwell’s equations in terms of this magnetic vector potential. Equation (1) implies the existence of a scalar potential such that
(6) |
By combining (4) and (6) and utilising Ohm’s law , we have
(7) |
In the case where the externally applied field is oscillating sinusoidally at frequency f, we can adopt a phasor convention such that . The phasor transformation combined with a choice of gauge given by reduces (7) to the vector Helmholtz equation
(8) |
where .
2.1. Conductive Rod in a Transverse Magnetic Field
In the case of the transverse field , where is a time-dependent amplitude, the magnetic vector potential can be expressed as a single z-directed component , given by
(9) |
where r is the radial coordinate and is the azimuthal coordinate. In light of (9), we need only consider the z component of the Helmholtz equation. Therefore, we have and (8) reduces to the scalar Helmholtz equation. We can expand the Laplacian in cylindrical coordinates to give
(10) |
where the derivatives with respect to z are zero in the case of an infinite cylinder. The complete solution for in the three regions of the domain is given by [18]
(11) |
where and are the Bessel functions of the first and second kind, respectively. The constants n, and are to be determined from boundary conditions. The first boundary condition we impose is the finiteness of the solution at the origin. This implies . Next, we demand that as r tends to infinity, must approach the vector potential of the imposed external field. This gives , and .
Next, we impose the continuity of the normal component of the magnetic flux density and continuity of the tangential component of the magnetic field strength along each interface. This yields four conditions, namely
(12) |
(13) |
(14) |
(15) |
where the primes denote derivatives. The solutions to this system of equations are
(16) |
(17) |
(18) |
(19) |
By using Equation (5), we can obtain the complete solution for the magnetic flux density in each region of the domain. The normalised magnitude of the magnetic flux density can be seen in Figure 2. We have evaluated the solution for two different rod geometries for two different frequencies of the externally applied field. As we are primarily interested in the magnetic flux density within the rod, we now turn our attention to the solution in the innermost region. For , we have
(20) |
2.2. Conductive Rod in a Longitudinal Magnetic Field
If the rod is placed in a longitudinal field given by , we can express the magnetic vector potential as
(21) |
With this in mind, we need only consider the component of the Helmholtz equation. Therefore, we have . We can expand the azimuthal component of the vector Laplacian in cylindrical coordinates to give
(22) |
where the derivative with respect to z and are zero.
The complete solution for in all regions is therefore given by
(23) |
The constants and are to be determined from boundary conditions. We again impose the finiteness of the solution at the origin. This implies . Next, we demand that as r tends to infinity, must approach the vector potential of the imposed external field. This yields .
Next, we impose the continuity of and at the metal air interfaces. This gives four equations:
(24) |
(25) |
(26) |
(27) |
The solution to this system of equations is
(28) |
(29) |
(30) |
(31) |
The magnetic flux density in the innermost region, where , is given by
(32) |
2.3. Magnetic Field Solutions in the Innermost Region
In the case of a nonmagnetic conducting rod, the solution for the magnetic field in the innermost region is the same for both transverse and longitudinal fields. Therefore, we conclude that the effect of the rod on a spatially uniform field is independent of orientation.
The field inside the rod is in the same direction as the external field; however, it has a magnitude given by and a phase difference given by . The normalised field magnitude as a function of frequency and conductivity for a number of rod geometries can be seen in Figure 3. In (a), we have fixed the frequency at kHz and varied the conductivity. In (b), we have fixed the conductivity at MS/m and varied the frequency. This is a typical value for the conductivity of medical-grade stainless steel. We note that the shielding effect is strongly dependent on geometry, frequency and conductivity. The phase difference induced by the rod can be seen in Figure 4. Here, we have again fixed the conductivity at MS/m and varied the frequency.
3. Methods
In order to verify the analytical solutions introduced in Section 2, we conducted a number of experiments to measure the shielding effect of stainless steel rods. The details of each rod can be seen in Table 1. In contrast to the magnetic field used for magnetic tracking, the magnetic field used in these experiments is spatially uniform. These experiments yield the range of magnetic field frequencies and corresponding instrument diameters, for which the inductive shielding effects are acceptable for tracking.
Table 1.
Rod | Material | (mm) | (mm) | (MS/m) |
---|---|---|---|---|
1 | 316 L Stainless Steel | 4 | ||
2 | 316 L Stainless Steel | 5 | 6 |
3.1. Helmholtz Coil Driver Design
To quantify this effect, a spatially uniform, transverse magnetic field of known frequency was used. A convenient method for producing a uniform field is a Helmholtz coil. This consists of two identical co-axial coils which are connected in series. The result is a spatially uniform magnetic field in the cylindrical region oriented along the axis of the coils. This work employed a 300 mm diameter Helmholtz coil (Ferronato BH300HF-3-B, Serviciencia S.L.U, Málaga, Spain). Each coil consists of eight turns with the pair of coils having a self-inductance of 93 µH and a field-to-current ratio of 54.2 µT/A. To increase the current that can be driven in the coils, a series capacitor is used to tune the resonant frequency of the circuit. Assuming a simple lumped model for the Helmholtz coil in series with the capacitor, the self-resonant frequency is given by
(33) |
where and are the self-inductances of each of the coils, respectively, and is the chosen series capacitance. A schematic of the circuit used to drive the Helmholtz coil can be seen in Figure 5. A National Instruments USB-6343 data acquisition unit is used to drive the sinusoidal signal, which is then amplified using an LT1210CT7 power amplifier. A relay-switched capacitor bank is used to select the value of the series capacitor. The capacitance is chosen such that the circuit is driven as close as possible to resonance. The generated field was verified using a calibrated triaxial magnetic field probe (ELT-400, Narda Safety Test Solutions GmbH, Pfullingen, Germany).
3.2. Influence of Transmitter Frequency
Using the Helmholtz coil, a homogeneous field was then driven at a given frequency f. The inductive sensor was aligned parallel to the magnetic field and, without any metallic rods present, the voltage induced across the sensor was measured. Measurements were taken over an interval of twenty seconds at a sampling rate of 250 kS/s. This voltage was validated to be linear in both frequency and applied magnetic field strength for the region of interest. Next, the inductive sensor was fixed axially within the centre of each stainless steel rod. The rod was placed into the Helmholtz coil with an angle between the sensor axis and the driven magnetic field. The corresponding voltage was measured. This was repeated for frequencies from 4 kHz to 60 kHz and angles of , and . The experimental setup can be seen in Figure 6. To determine the shielding effect of the rod, we are interested in the ratio , where is the amplitude of the magnetic flux density within rod i and is the amplitude of the magnetic flux density during the control measurement when no rod is present. To obtain the magnitude of the field, we take the FFT of the the measured voltage signal and extract the amplitude corresponding to the frequency of the driven magnetic field.
A similar procedure was followed to measure the phase difference induced by the presence of the rods. First, using the FFT and with no rod present, we calculated the phase difference between the voltage across the inductive sensor and the current through the Helmholtz coils. In the same way, we obtained the phase difference between the voltage across the inductive sensor and the current through the Helmholtz coil in the presence of rod i. We were then interested in the additional phase difference induced by the presence of the rod. This is given by
(34) |
This was repeated for a range of frequencies from 4 kHz to 60 kHz.
3.3. Impact on Electromagnetic Tracking
To assess the feasibility of tracking with a sensor within metallic surgical instruments, we perform grid measurements with an inductive sensor placed within a stainless steel rod. This work employed the open-source Anser EMT tracking system [19]. The Anser EMT system uses a field generator with eight coils to create a spatially unique and time-varying electromagnetic field. Frequency division multiplexing is used with each emitter coil driven at a different frequency.
The Anser EMT tracking system was used to track an inductive sensor with a diameter of 0.45 mm and a length of 8.2 mm (NDI 610158, Northern Digital Inc., Waterloo, ON, Canada). An optical tracking system (NDI Polaris, Northern Digital Inc., Waterloo, ON, Canada) was used as the ground truth. The experimental setup can be seen in Figure 7.
To measure the accuracy over the volume of interest, a robot was used to move the sensor in a grid at a fixed height above the field generator. Each grid comprised 100 points organised in a grid, occupying an area of 25 cm × 25 cm in the plane. At each grid location, EMT and optical measurements were recorded simultaneously. With an inductive sensor fixed co-axially within each rod, grid measurements were taken at heights of 210 mm and 270 mm above the field generator. Grid measurements were repeated for two sensor orientations; z- and y-directed. A full grid measurement was performed with the emitter coils driven at different frequencies from 1 kHz to 20 kHz.
Once measurements were completed, Horn’s absolute orientation algorithm [20] was used to align the optical and EMT reference frames. The error between these aligned grids was then defined as our tracking error.
4. Results and Discussion
4.1. The Influence of Transmitter Frequency
The experimental measurements of the shielding effect and induced phase difference can be seen in Figure 8. The solid curves denote the theoretical curves computed using (32) with the rod parameters from Table 1. In both cases, there is good agreement between the measured results and the expected analytical curves. As expected, in both cases, the rod with larger diameter has a greater effect on the magnetic field. For the range of frequencies employed in electromagnetic tracking, the magnetostatic approximation of Maxwell’s equations accurately predicts for the effect of metal instruments. Relative to a uniform external field, the magnetic field within a cylindrical instrument has an attenuated magnitude and an additional induced phase difference.
4.2. Impact on Electromagnetic Tracking
Calibration of the electromagnetic sensor was performed by collecting EMT and optical measurements in two orientations at a height of 250 mm above the field generator. These measurements were completed without the presence of any distorters. Using a subset of 50 points from this grid, a calibration was performed to ensure accurate tracking in the region of interest. The remaining points were used to measure the accuracy of the Anser EMT system at each frequency.
The same calibration was used when measuring the tracking performance in the presence of each stainless steel rod for 100 points across each xy plane. The Euclidean error at each grid point i is given by
(35) |
where , and are the deviations between the aligned electromagnetic and optical positions. The root-mean-squared error RMSE is then given by
(36) |
and the mean error ME is given by
(37) |
The standard deviation of the errors is given by
(38) |
The results for each rod and each frequency can be seen in Table 2. The cumulative distribution of the tracking error as a function of emitter coil frequency with the sensor fixed within each rod can be seen in Figure 9. The optical system has a reported volumetric accuracy of 0.15 mm RMSE [21]. The tracking error shows a strong dependency on frequency, with much larger errors at increasing frequency. From these results, it is clear that for the distortionless tracking of metal instruments of 8 mm or greater, the frequency of the transmitter magnetic field must be less than 1 kHz. However, such low-frequency tracking may not provide sufficient update rates for some applications. To enable tracking at higher frequencies, compensation for the effects of the instrument is required. In some applications, tracking errors may be acceptable at 5 kHz or 10 kHz. Alternatively, smaller instruments (<8 mm outer diameter) can be accurately tracked at significantly higher frequencies and update rates. Finally, the impact of dynamic field effects (e.g., instrument motion) has not been considered in this work. However, given that such motion in clinical applications is typically highly controlled and limited, it can be reasonably expected that the quasistatic effects outlined in this paper will dominate the distortion of the magnetic field.
Table 2.
RMSE (mm) | ME (mm) | STD (mm) | |||||||
---|---|---|---|---|---|---|---|---|---|
Frequency (kHz) | No Rod | Rod 1 | Rod 2 | No Rod | Rod 1 | Rod 2 | No Rod | Rod 1 | Rod 2 |
1 | 0.89 | 1.09 | 1.09 | 0.71 | 0.89 | 0.92 | 0.53 | 0.63 | 0.59 |
5 | 0.98 | 1.16 | 1.55 | 0.79 | 0.94 | 1.29 | 0.58 | 0.68 | 0.85 |
10 | 1.26 | 1.69 | 4.55 | 1.01 | 1.38 | 3.97 | 0.75 | 0.98 | 2.21 |
15 | 1.38 | 2.75 | 10.01 | 1.13 | 2.30 | 8.83 | 0.80 | 1.52 | 4.72 |
20 | 2.33 | 12.74 | 25.53 | 1.89 | 11.08 | 22.17 | 1.37 | 6.30 | 12.68 |
5. Conclusions
In this work, we investigate the feasibility of tracking metal instruments typical of image-guided interventions using electromagnetic tracking with an inductive sensor placed within the instrument shaft. Analytical models of the instrument as an infinite conductive rod investigated the effect of this distorter on the magnetic field within the shaft. The proposed model neglects end-effects introduced by an instrument of finite extent and the external field was assumed to be spatially uniform. However, in the case of the electromagnetic tracking, neither of these criteria are typically encountered. To enable tracking, there must be a gradient present in the generated magnetic field, and hence, the field is not spatially uniform. The rod is found to have an effect which is strongly dependent on the tracking frequency, material properties and the rod geometry. We experimentally verified these results with a homogeneous field.
The impact of distortion was assessed for larger-diameter rods using an inductive sensor within the shaft and the open-source Anser EMT system. For typical conductivities in steel rods with outer diameters less than 8 mm, it was demonstrated that accurate tracking is possible up to around 5 kHz. For small-enough instrument diameters (<8 mm), the RMSE of less than 2 mm at 10 kHz may be sufficiently low that it can be neglected in comparison to image-to-patient registration errors. These registration errors are typically brought about by respiratory motion where, during a normal breathing cycle, abdominal organ movement has an average amplitude of 13 mm [22]. For larger rods of outer diameter greater than 8 mm, lower transmitter frequencies (1 kHz to 2 kHz) are necessary to ensure distortion-free tracking. This work identifies methods to reduce the tracking error brought about by the shielding effect of a metallic instrument. These are a reduction in the outer diameter of the instruments, the use of a suitable material with a lower electrical conductivity, or a reduction in the magnetic field frequency in exchange for lower update rates.
This work presents the first known evidence validating the use of electromagnetic sensors positioned within the shaft of metal instruments for tracking in clinical settings for applications such as laparoscopic surgery, robotics, arthroscopy and orthopaedics. As such, the results will be of considerable interest to users looking to use magnetic tracking in image-guided interventions where the sensor can now be positioned within the metal shaft itself within the design parameters outlined. This work has immediate applications in procedures such as minimally invasive cholecystectomy, where laparoscopic surgery has become the standard of care [23]. For these procedures, typical port diameters range from 2 mm to 12 mm.
Abbreviations
The following abbreviations are used in this manuscript:
EMT | Electromagnetic tracking |
DAC | Digital-to-analog converter |
ADC | Analog-to-digital converter |
IA | Instrumentation amplifier |
FFT | Fast Fourier transform |
IR | Infrared |
RMSE | Root-mean-square error |
ME | Mean error |
STD | Standard deviation |
Author Contributions
Conceptualisation, E.H. and P.C.-M.; methodology, E.H. and D.C.; software, E.H., D.C. and C.v.d.B.; validation, all authors; formal analysis, E.H.; investigation, E.H. and C.v.d.B.; resources, P.C.-M.; data curation, E.H.; writing—original draft preparation, E.H.; writing—review and editing, E.H. and P.C.-M.; visualisation, E.H.; supervision, P.C.-M.; project administration, P.C.-M.; funding acquisition, P.C.-M. All authors have read and agreed to the published version of the manuscript.
Institutional Review Board Statement
Not applicable.
Informed Consent Statement
Not applicable.
Data Availability Statement
Data are contained within the article.
Conflicts of Interest
The authors declare no conflicts of interest.
Funding Statement
This work is funded by the European Union (ERC, DeepField, project number 101002225) and Science Foundation Ireland (Career Development Award 17/CDA/4771). Views and opinions expressed are, however, those of the authors only and do not necessarily reflect those of the European Union or the European Research Council Executive Agency. Neither the European Union nor the granting authority can be held responsible for them.
Footnotes
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content.
References
- 1.Schwarz Y., Greif J., Becker H.D., Ernst A., Mehta A. Real-time electromagnetic navigation bronchoscopy to peripheral lung lesions using overlaid CT images: The first human study. Chest. 2006;129:988–994. doi: 10.1378/chest.129.4.988. [DOI] [PubMed] [Google Scholar]
- 2.Khan K.A., Nardelli P., Jaeger A., O’Shea C., Cantillon-Murphy P., Kennedy M.P. Navigational bronchoscopy for early lung cancer: A road to therapy. Adv. Ther. 2016;33:580–596. doi: 10.1007/s12325-016-0319-4. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.Jaeger H.A., Trauzettel F., Nardelli P., Daverieux F., Hofstad E.F., Leira H.O., Kennedy M.P., Langø T., Cantillon-Murphy P. Peripheral tumour targeting using open-source virtual bronchoscopy with electromagnetic tracking: A multi-user pre-clinical study. Minim. Invasive Ther. Allied Technol. 2019;28:363–372. doi: 10.1080/13645706.2018.1544911. [DOI] [PubMed] [Google Scholar]
- 4.Geisbüsch A., Auer C., Dickhaus H., Putz C., Dreher T. Electromagnetic tracking for femoral derotation osteotomy—An in vivo study. J. Orthop. Res. 2017;35:2652–2657. doi: 10.1002/jor.23579. [DOI] [PubMed] [Google Scholar]
- 5.Gaita F., Guerra P.G., Battaglia A., Anselmino M. The dream of near-zero X-rays ablation comes true. Eur. Heart J. 2016;37:2749–2755. doi: 10.1093/eurheartj/ehw223. [DOI] [PubMed] [Google Scholar]
- 6.Duan X. Chapter 10—Magnetic tracking and positioning in endoscopy. In: Tishin A.M., editor. Magnetic Materials and Technologies for Medical Applications. Woodhead Publishing; Sawston, UK: 2022. pp. 287–338. (Woodhead Publishing Series in Electronic and Optical Materials). [Google Scholar]
- 7.Yaniv Z., Wilson E., Lindisch D., Cleary K. Electromagnetic tracking in the clinical environment. Med Phys. 2009;36:876–892. doi: 10.1118/1.3075829. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8.Franz A.M., Haidegger T., Birkfellner W., Cleary K., Peters T.M., Maier-Hein L. Electromagnetic tracking in medicine—A review of technology, validation, and applications. IEEE Trans. Med. Imaging. 2014;33:1702–1725. doi: 10.1109/TMI.2014.2321777. [DOI] [PubMed] [Google Scholar]
- 9.Nagel M., Hoheisel M., Petzold R., Kalender W.A., Krause U.H. Medical Imaging 2007: Visualization and Image-Guided Procedures. Volume 6509. SPIE; Bellingham, WA, USA: 2007. Needle and catheter navigation using electromagnetic tracking for computer-assisted C-arm CT interventions; pp. 184–192. [Google Scholar]
- 10.Liu X., Kang S., Plishker W., Zaki G., Kane T.D., Shekhar R. Laparoscopic stereoscopic augmented reality: Toward a clinically viable electromagnetic tracking solution. J. Med. Imaging. 2016;3:045001. doi: 10.1117/1.JMI.3.4.045001. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11.Gomes-Fonseca J., Veloso F., Queirós S., Morais P., Pinho A.C., Fonseca J.C., Correia-Pinto J., Lima E., Vilaça J.L. Assessment of electromagnetic tracking systems in a surgical environment using ultrasonography and ureteroscopy instruments for percutaneous renal access. Med. Phys. 2020;47:19–26. doi: 10.1002/mp.13879. [DOI] [PubMed] [Google Scholar]
- 12.Chen C.C., Yeh P.C., Cheng C.C., Lin C.K., Wong T.H., Cheng Y.C., Hsu W., Chung T.K. Magnetic Targeting Systems for Endo-Bronchoscope Diagnosis and Intramedullary-Nail Surgery: A Review. IEEE Sensors J. 2021;21:12545–12562. doi: 10.1109/JSEN.2021.3059044. [DOI] [Google Scholar]
- 13.Zhu Y., Chang H., Yu Y., Chen W., Liu S., Zhang Y. Meta-analysis suggests that the electromagnetic technique is better than the free-hand method for the distal locking during intramedullary nailing procedures. Int. Orthop. 2017;41:1041–1048. doi: 10.1007/s00264-016-3230-3. [DOI] [PubMed] [Google Scholar]
- 14.Liu X., Rice C.E., Shekhar R. Fast calibration of electromagnetically tracked oblique-viewing rigid endoscopes. Int. J. Comput. Assist. Radiol. Surg. 2017;12:1685–1695. doi: 10.1007/s11548-017-1623-4. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15.Xiao G., Bonmati E., Thompson S., Evans J., Hipwell J., Nikitichev D., Gurusamy K., Ourselin S., Hawkes D.J., Davidson B., et al. Electromagnetic tracking in image-guided laparoscopic surgery: Comparison with optical tracking and feasibility study of a combined laparoscope and laparoscopic ultrasound system. Med. Phys. 2018;45:5094–5104. doi: 10.1002/mp.13210. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16.Ma L., Zhao Z., Zhang B., Jiang W., Fu L., Zhang X., Liao H. Three-dimensional augmented reality surgical navigation with hybrid optical and electromagnetic tracking for distal intramedullary nail interlocking. Int. J. Med Robot. Comput. Assist. Surg. 2018;14:e1909. doi: 10.1002/rcs.1909. [DOI] [PubMed] [Google Scholar]
- 17.Haus H., Melcher J. Electromagnetic Fields and Energy. Prentice Hall; Englewood Cliffs, NJ, USA: 1989. [Google Scholar]
- 18.Zahn M. Electromagnetic Field Theory: A Problem Solving Approach. Wiley; Hoboken, NJ, USA: 1979. pp. 277–281. [Google Scholar]
- 19.Jaeger H.A., Franz A.M., O’Donoghue K., Seitel A., Trauzettel F., Maier-Hein L., Cantillon-Murphy P. Anser EMT: The first open-source electromagnetic tracking platform for image-guided interventions. Int. J. Comput. Assist. Radiol. Surg. 2017;12:1059–1067. doi: 10.1007/s11548-017-1568-7. [DOI] [PubMed] [Google Scholar]
- 20.Horn B.K.P., Hilden H.M., Negahdaripour S. Closed-form solution of absolute orientation using orthonormal matrices. J. Opt. Soc. Am. A. 1988;5:1127. doi: 10.1364/JOSAA.5.001127. [DOI] [Google Scholar]
- 21.Polaris Vega ST—NDI. [(accessed on 9 July 2024)]. Available online: https://www.ndigital.com/polaris-vega-st/
- 22.Korin H.W., Ehman R.L., Riederer S.J., Felmlee J.P., Grimm R.C. Respiratory kinematics of the upper abdominal organs: A quantitative study. Magn. Reson. Med. 1992;23:172–178. doi: 10.1002/mrm.1910230118. [DOI] [PubMed] [Google Scholar]
- 23.Osborne D.A., Alexander G., Boe B., Zervos E.E. Laparoscopic cholecystectomy: Past, present, and future. Surg. Technol. Int. 2006;15:81–85. [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Data Availability Statement
Data are contained within the article.