Robots on the Stage: A Snapshot of the American Robotic Total Knee Arthroplasty Market

Avinash Inabathula 1,a, Dimitar I Semerdzhiev 2, Anand Srinivasan 1, Farid Amirouche 1, Lalit Puri 1, Hristo Piponov 1
PMCID: PMC11368221  PMID: 39238881

Abstract

  • » Computer-assisted robots aid orthopaedic surgeons in implant positioning and bony resection. Surgeons selecting a robot for their practice are faced with numerous options. This study aims to make the choice less daunting by reviewing the most commonly used Food and Drug Administration-approved robotic total knee arthroplasty platforms in the American arthroplasty market.

  • » Modern total knee arthroplasty (TKA) robots use computer guidance to create a virtual knee model that serves as the surgeon's canvas for resection planning.

  • » Most available robotic TKA (rTKA) systems are closed semiactive systems that restrict implant use to those of the manufacturer.

  • » Each system has distinct imaging requirements, safety features, resection methods, and operating room footprints that will affect a surgeon's technique and practice.

  • » Robots carry different purchase, maintenance, and equipment costs that will influence patient access across different socioeconomic groups.

  • » Some studies show improved early patient-reported outcomes with rTKA, but long-term studies have yet to show clinical superiority over manual TKA.

Introduction

Robot-assisted surgery is used in many medical procedures, and knee arthroplasty is no exception. Robodoc (THINK Surgical) was the first robotic system used in orthopaedic surgery for total hip arthroplasty1-3. Advances in computer science have since led to improvements in navigation and robotic arm control. Popularity and use of robotic total knee arthroplasty (rTKA) systems have surged in recent years, comprising 13% of all TKAs according to the American Joint Replacement Registry 2023 report, a 3-fold increase from 20184,5.

In the modern arthroplasty context and for the purpose of this article, rTKA refers to surgery performed with a computer-guided resection tool, with or without an attached mechanical arm, with haptic or motion sensing feedback. Robotic systems are successors to previous “computer navigation” systems (sometimes referred to as just “navigation”) that gave feedback to the surgeon without guiding implant positioning or resection6. Systems with computer navigation only or those lacking resection control through sensor feedback are outside the scope of this article. Modern arthroplasty robots create a virtual model of the joint that enables surgical planning6-8. They also help precisely execute the surgeon's plan using controlled resection tools8-13.

Classifying Computer-Assisted Robots in Arthroplasty

Computer-assisted systems are broadly classified as passive, active, or semiactive (also known as haptic)7,14. Passive systems do not have an independent moving element. The computer only provides information regarding cut angles and implant position, but resection is manual. In active systems, a computer-controlled robot (usually an arm with multiple joints and a resection tool) autonomously performs resection based on entered parameters7,14. Semiactive systems combine elements of both passive and active: The computer helps guide planning, and a robotic resection tool is controlled by the surgeon. However, the computer also provides sensory feedback to help increase accuracy. Whether these systems translate to better safety and reliability is being studied11,15,16.

Robotic systems are also subdivided into open vs. closed platforms. Open platform systems are compatible with a wide array of implants regardless of manufacturer. Closed platform systems are only compatible with the manufacturing vendor's implants8. This means adopting a closed robotic system also requires switching to that manufacturer's implants. There are currently multiple companies that offer Food & Drug Administration (FDA)-approved closed systems: robotic surgical assistant (ROSA) (Zimmer-Biomet), core of real intelligence (CORI) (Smith & Nephew), Velys (Depuy-Synthes), and Mako (Stryker). OMNIbotics (Corin) is also FDA approved in the United States for TKA17.

Notably, the TMINI and TSolution-One (THINK Surgical) are the only available open platform systems. The TMINI features a motorized optically guided drill for traditional cutting block pins. TSolution-One, the successor to ROBODOC, is a computed tomography (CT)-based active robot with an autonomous milling tool. According to several authors, the TMINI and OMNIbotics are better classified as motorized computer navigation systems rather than robotic tools with haptic feedback17. Thus, these 2 systems are beyond the scope of this study.

Performing TKA With Computer Guidance

The process of performing robotic TKA can seem confusing at first, as it has several additional steps. It can be performed with image-based or imageless systems. To ensure familiarity with their devices, manufacturers require surgeons take rTKA certification courses before use.

Most computers in robotic arthroplasty receive information about the outside world by way of optical arrays. These arrays each contain reflective disks or spheres arranged in unique configurations that can be distinguished by the computer via a camera. A probe and resection tools are also attached to optical arrays. LEDs and reflective optical encoders (electromechanical devices that convert angular or linear displacement into electrical signals) are built into the camera and various moving parts, which convey information to the computer. Tibia and femur arrays are affixed to the patient via bone pins18. The mathematics and engineering underlying optical tracking and modeling are beyond the scope of this article. Simply, the computer obtains the position of the instruments, tibia, ankle, femur, hip center of rotation, and resection tools by recognizing the position and orientation of each array. Joint surface topography is also manually captured with the probe. The computer uses these aggregate data to reconstruct a virtual representation of the patient's tibia and femur19. Imageless systems use only intraoperative data. Image-based systems augment this process with radiograph, magnetic resonance imaging (MRI), or CT scan, although this adds cost, time, and radiation20. In such systems, the computer superimposes in vivo surface points onto a 3D model generated from the preoperative image. This orients and positions the virtual model correctly in space. Imageless robotic systems require registering far more surface points to generate a topographic joint map from scratch, at the benefit of minimizing cost and radiation. Coronal plane imbalance is captured in the model by moving the knee through full range of motion and applying varus/valgus stress. This incorporates ligament laxities into the flexion/extension gaps. Bony resection is then simulated on this virtual model. The proprietary computer of each vendor's system is preloaded with their implants' exact polyethylene, tibial, and femoral component dimensions. The computer can simulate the resultant coronal plane gaps for each resection plane size of the implant. The plan is finally fine-tuned, and resection is performed21-23. Soft-tissue releases can still be performed at the surgeon's discretion. The surgeon can also verify coronal balance by restressing with trial implants in place.

One exception to the optical guidance strategy is the TSolution-One, which uses 2 stabilizing arms, a registration arm with a probe and a resection arm with a milling tool all directly linked to the computer21,24. Since the knee is immobile in this workflow and components are directly connected, there is no need for optical tracking.

Errors in Computer Guidance

Handheld saw blades can change angle or deflect when running through sclerotic bone, introducing variability into the resection plane. Manual cutting blocks can also shift due to retraction or pin motion. Moreover, cutting block positioning relies on bony landmarks, an imprecise method22. Resection depth and angle errors are much smaller in rTKA, though not eliminated23. Several studies show lower cut variability in rTKA compared with manual TKA (mTKA)25-27, thanks, in part, to built-in haptic controls and safety measures. Despite safeguards within each system, undetectable errors such as inaccurate topographic mapping, pin-site loosening during surgery, or pin motion can still create discrepancies between the virtual model and real patient anatomy25,28. Many vendors require presence of product specialists to help troubleshoot intraoperative errors. Regardless, surgeons must be prepared to revert to mTKA in the event of major malfunctions.

Robotic TKA Compared with Manual TKA

Patient Satisfaction

Patient satisfaction, alongside complication rates, is a primary determinant of rTKA value. An estimated 19% of TKA patients are dissatisfied with their outcome26. Surveys suggest this is multifactorial, with persistent postoperative knee pain and continued difficulty with activities of daily living being most quoted26. Researchers hope the added precision and accuracy of robotics will translate to better patient outcomes and satisfaction27,29,49.

Importantly, patients often have higher levels of expectation from rTKAs compared with mTKAs due to proinnovation bias, which affects reporting31. Prospective blinded studies would eliminate these biases but present ethical and logistical concerns. A multicenter prospective study on rTKA reported that 9 of the 10 measures assessed by the Knee Society Scoring (KSS) System improved at 3 months32. In a study from the Function and Outcomes Research for Comparative Effectiveness in total joint replacement registry, patients assessed vvia the New Knee Society Score 2 years after rTKA reported greater satisfaction with their surgery, improved scores of pain, higher levels of recreation, and improved quality of life compared with mTKA33. However, this comparison does not control for significant variability in techniques, perioperative practices, and patient populations. Another prospective study found higher patient satisfaction on a Likert scale and KSS survey at 1 year with rTKA34. When satisfaction scores were reviewed from multiple comparative studies, the authors found variable results at the final follow-up ranging from 3 months to 2 years12. This is a difficult area of study with mostly low-level evidence in publication. The ongoing Robotic Arthroplasty Clinical and Cost-Effectiveness Randomized Controlled Trial-knee in the United Kingdom aims to compare 1-year Forgotten Joint Score (FJS) between rTKA and mTKA and may present high-level evidence35.

Clinical Outcomes and Survivorship

A recent retrospective multicenter study of 861 rTKAs demonstrated reproducible increases in Forgotten Joint Score (FJS) and Knee Injury and Osteoarthritis Outcomes Score for Joint Replacement across institutions36. These improvements are comparable with those of mTKA. When comparing rTKAs with mTKA at 2 years, one study reported greater pain reduction and function on the Western Ontario and McMaster Universities Arthritis Index (WOMAC) in cementless rTKAs than mTKAs37. Randomized controlled studies are scarce in this area, but one such study with 10-year minimum follow-up found no difference in University of California-Los Angeles activity, WOMAC scores, or Knee society scores in a South Korean population18. A meta-analysis of clinical and radiological outcomes in rTKA found a mean clinical follow-up time of 33.8 months for all 20 studies, with >5 years data available in only 5 studies12. Four of these studies were from South Korea using the ROBODOC, which is no longer available in the United States. The authors' combined analysis of studies with over 1-year follow-up showed higher hospital for special surgery and WOMAC scores with rTKA but no differences in range of motion or KSS12. Only one study had 11 years of prospective randomized data, which showed no difference in implant survivorship15.

TKA revision rates, although low, are projected to increase in the future38. Advocates of rTKA cite reduced revision rates as a potential benefit. Kim et al. followed 71 robotic-assisted TKAs for 15 years and found an annual all-cause revision probability of 0.28% for rTKA, not statistically different from 0.49% in mTKA15. They also reported no difference in aseptic loosening at 10 years. Another study compared rates of progressive radiolucency in 80 rTKAs vs. 80 mTKAs and found no difference37. Recently, an analysis of rTKA in 9,220 Medicare beneficiaries (older than 65 years) from the American Joint Replacement Registry was performed39. After controlling for confounders such as surgeon, location, and comorbidities, the authors found no difference in odds of revision at 2 years. Unlike most studies, they saw an increased odds of instability (OR 1.6, 95% CI 1.0-2.4, p = 0.04) and pain (OR 2.1, p < 0.001) at 2 years in the rTKA cohort. This study is limited by small sample size for a registry study and only considers the Medicare age group, but the results contrast to existing studies.

Accuracy of Cuts

In multiple studies, implant alignment seems to be more precise and reproducible with haptic controlled robotic resection compared with conventional cutting blocks11,12. A meta-analysis of 1 robot also showed more consistent cut angles for robotic unicompartmental knee arthroplasty (UKA), although this did not translate to improved short-term patient outcomes8. The same findings were observed in multiple studies comparing cut angles in TKA10,13,40. A prospective randomized study from Song et al. saw fewer mechanical axis outliers (>±3°) in the robotic surgery group (0 vs. 12, p < 0.001)10. Another randomized study found that TKA angles with robotic guidance had far less variability when striving for ±3° of neutral mechanical alignment (p < 0.01)11. While neither study found specific alignment values translated to better patient-reported outcomes, they demonstrate that rTKA allows more consistent cuts. Xu et al. also showed rTKA resulted in far fewer lateral tibial component angle outliers, 3.0% vs. 29.4% in mTKA40. The meta-analysis by Agarwal et al. also found that rTKA had significantly fewer radiographic outliers in their combined analysis of 8 studies, despite 2 showing no difference12.

Soft-Tissue Balancing and Alignment

Another crucial component of performing TKA is ligamentous balancing. Instability due to coronal plane imbalance is one of the leading reasons for revision TKA41. Numerous studies have shown that ligamentous imbalance can cause instability, implant loosening, and joint stiffness38,41. With traditional mechanical resection, soft-tissue releases are often required to compensate for coronal imbalances. rTKA aims to minimize the need for releases by accounting for preresection balance in the planning stage. Surgeons can plan resection angles that compensate for laxity/tightness and reassess balance between releases. Standardized digital tensioners are currently under development to reduce variability in laxity measurement.

Limb alignment and joint line obliquity changes also have a profound influence on knee kinematics and patient satisfaction. Computer-assisted robots allow resection planning with varying alignment philosophies (mechanical, kinematic, functional, etc), with or without restrictions42. The ability to quantify the effects of resections/releases may eliminate some variability when comparing TKAs performed with different alignment philosophies. This may benefit surgeons looking to study outcomes over time for different methodologies.

Cost of Robotic Technology

While rTKA is a useful tool, it comes with steep costs. A study tracking 6 high-volume TKA surgeons over 2 years reviewed the costs of TKA per case. On average, direct cost per case for rTKA was $11,615 while mTKA costs $8,674 (calculated excluding preoperative imaging, labor, acute stay, and supply/implant cost)27. While cost of rTKA was higher, 30-day readmission rate was only 1.2% with rTKAs but 4.9% for conventional TKAs27. Importantly, the authors acknowledge readmissions were likely confounded by greater utilization management exposure in rTKA cases, which predominantly occurred after the removal of TKA from the “inpatient only” Medicare/Medicaid list. In many cases, hospitals or ambulatory surgery centers purchase robotic systems for surgeons and, as stakeholders, will expect a return on investment as the robot depreciates over time.

Case volume is an important factor in offsetting the cost. One study reported that, with a low annual volume of TKA (≤10 cases), the total loss is $71,025 for each case43. This cost is reduced to $7,463 per encounter when performing a moderate volume of TKAs (∼100 cases/yr). A high volume of TKA (≥200 cases/yr) was associated with the lowest cost per encounter at $3,93143. Greater case numbers spread out costs, making recuperation easier. Robotic TKA maintains cost-effectiveness, according to the authors' model, if annual revision rates remain less than 1.6% at 1 year and quality-of-life measures remain favorable (on a scale derived from converted short form-36 scores and other quality-of-life measures)43. Future follow-up >10 years may allow a comprehensive cost-benefit analysis accounting for revision-free survival.

The largest cost in implementing rTKA is the cost of the robot itself. Surgeons and stakeholders new to rTKA may obtain cost estimates from their regional sales representatives. Many practices or hospitals enter into negotiated payment agreements with vendors or arrange payment strategies involving rebates. The terms of these contracts are highly variable and usually confidential. Case volume or rebate-based payment agreements can significantly influence a practice's implant choice. Purchase decisions should be considered in the context of a practice's planned rTKA volume and, for closed systems, comfort using the vendor's implants.

The estimated prices of each system are summarized in Table I7,24,44,45. Of note, new production of TSolution-One in the US market has recently been discontinued due to the upcoming release of a new active system. Aftermarket purchase of robots and replacement parts is possible, though little information is available in this regard.

TABLE I.

Summary Table of Discussed Robots*

Robot (Vendor) Platform Type Estimated Price (USD) Guidance Method Resection Method Imaging Safety Measures FDA 510 (k) Status
CORI (Smith & Nephew) Closed Semiactive 500,000 Optical Handheld burr None • Retracting optically tracked burr TKA + UKA (2020), Revision TKA (2022)
MAKO (Stryker) Closed Semiactive 1,000,000 Optical and image Robotic arm with saw CT • Haptically bound arm TKA + UKA (2015)
ROSA (Zimmer-Biomet) Closed Semiactive 700,000 Optical and image Handheld saw and robotic arm with cut block XR or MRI • Arm with cut blocks bound to single plane
• Cut verification tools
TKA (2019), UKA (2021)
Velys (Smith & Nephew) Closed Semiactive No data Optical Robotic arm with saw None • Haptically bound arm
• Optically tracked saw
TKA (2021), UKA (2024)
TSolution-One (THINK Surgical) Open Active 650,000-800,000 Optical Robotic arm with milling tool CT • Virtual boundaries
• Force and motion sensors
TKA (2024)
TMINI (THINK Surgical) Open Cut-block holes only, semiactive No data Optical Manual None • Virtual boundaries
• Stereotactic guidance
TKA (2023)
*

TKA = total knee arthroplasty and UKA = unicompartmental knee arthroplasty. Listed prices are based on publicly available manufacturer suggested retail price data from vendors or distributed vendor publications. Note that retail and maintenance contract prices vary widely and often negotiated in the purchase agreement between the vendor and hospital or practice. Disposables cost around $1,000 per case, but this also varies by system7,25,40,41,80.

There is marketing value in advertising of rTKA, although this should be balanced with setting reasonable patient expectations. A recent study using Google Trends open-source analytics showed patient interest in rTKA online is outpacing interest in conventional TKA46. A retrospective analysis from the Geisinger Health System demonstrated a modest increase in total TKA and UKA (17% and 190%, respectively) volumes and a slight shift toward more UKA after introducing and marketing rTKA43. This study also found robotic TKA was, on average, $176 more expensive per case than mTKA, though it did not capture postdischarge care utilization.

Given these costs, there are large disparities in rTKA access across geographic and socioeconomic patient populations. Many robots are currently in use at larger hospitals with greater capital. This results in lower accessibility for Americans with more comorbidities, those in predominantly White rural regions, or minorities in urban regions47,48. According to the New York statewide reporting system, rTKA was more often used in patients with private insurance over Medicaid (5.9% vs. 2.2%) and at high-volume hospitals49. These trends underscore a broader lapse in accessibility of medical technology for lower socioeconomic status patients. Minimizing cost and ensuring equitable access to rTKA are important goals for surgeons implementing rTKA.

Operative Time and Efficiency

When implementing robotic TKA, there is a learning curve for the entire surgical team. Another downside to rTKA is added surgical time compared with mTKA29,37,45. Increased case duration is primarily from added steps for computer guidance, resection planning, and robot setup50,51. Song et al. found rTKA increased operative time by a mean of 25 minutes10. Newer robot iterations have streamlined their user interface, tray organization, and instrumentation for efficiency.

Studies on rTKA adoption consistently show a learning curve for surgical time52-54. In one study using ROSA, mean operative time during the learning phase was 114 ± 17 minutes, decreasing to 110 ± 20 minutes during the proficiency phase (p = 0.53). A multicenter study using the Stryker Hospital Reported Outcomes database calculated mean operative times (first incision to wound closure for mTKA vs. rTKA) and found time neutrality was achieved after 15 to 20 cases for most surgeons, with high-volume surgeons requiring only 7 cases53. One-third of surgeons were unable to achieve time neutrality after 20 cases, but later achieved this. Studies did not find a difference in early complications upon adoption of rTKA52-54.

Turnover time is also affected by introduction of a new technology. Surgical personnel new to rTKA have a general dissatisfaction with turnover time since it adds new trays, equipment, and processes that may need troubleshooting. One study reported a turnover time of 72 minutes across 20 robotic cases, which represented a small increase from their baseline55. The largest contributors were cleaning of extra equipment, followed by instrumentation setup and retrieval of the patient from preoperative holding55. One group studied their implementation of a new robot using a task allocation and sequencing model borrowed from Formula One racing. They observed a decrease in turnover time from 99.2 to 53.2 minutes after 3 months, very close to their mTKA time56. Surgeons perceived increased room setup time as the main hindrance to efficiency while support surgical staff perceived instrument availability and processing time as main contributors. Interestingly, the perceived time delays exceeded the actual recorded increases documented by the independent researchers. An efficient workflow and retention of rTKA trained staff minimize turnover time56.

Complications Specific to rTKA

Pin-Site Fractures and Infections

Bone pin placement is an integral part of rTKA that provides fixation for tracking arrays or other attachments. These can be placed intraincisionally or extraincisionally. Pins introduce a small risk of pin-site fractures and infections18,25,57. A systematic review found fracture rates between 0.06% and 4.8%57,58. Interestingly, none required revision surgery or open reduction and internal fixation. Another study reported the use of bicortical diaphyseal pins caused 3 femur fractures requiring intramedullary nailing (0.19%) in 1,571 TKAs59. The authors had 0 fractures when switching to unicortical pins (stopping pin advancement upon feeling contact with the far cortex). Some surgeons advocate metaphyseal pins since the bone is broader and the same-size holes introduce relatively smaller stress risers57,58,60. Metaphyseal pins can be placed intraincisionally but require a slightly larger incision and careful placement so as to not interfere with cutting or implant positioning. Pin size has also been implicated in fracture risk, and some surgeons advocate 3.2-mm over 4.5-mm diaphyseal pins. Smaller diameter metaphyseal unicortical pins seem to safely minimize fracture risk. However, these changes may provide weaker purchase in osteopenic bone.

Infection can occur at extraincisional pin sites, which presents as persistent drainage or delayed wound healing18,25. This could lead to osteomyelitis if untreated. The rate of pin-site infection was 0.5% in an 839-patient study25. The authors reported that all were successfully treated with topical antibiotics without subsequent recurrence, consistent with other studies. One study investigated whether pin diameter influenced infection risk and found no association60. Use of intraincisional pins eliminates this concern.

Iatrogenic Soft-Tissue Injury

Neurovascular injury, while a risk factor of TKAs, is a very rare occurrence. A study compiled a total of 39,990 TKAs from 1998 to 2013 showing 65 nerve injuries as a direct result of the procedure61. When comparing manual with rTKA, the same sources of injury exist. Certainly, femoral nerve or popliteal artery injury risk from tracking pin placement can be inferred from studies on external fixation62. Haptic controls and less aggressive resection tools theoretically provide added protection in rTKA. However, there are no large studies directly comparing rates of neurovascular injury between rTKA and mTKA to date.

The authors of a prospective comparison study found a lower rate of intraoperative iatrogenic soft-tissue and bony injury in 30 rTKAs compared with 30 mTKAs. They attribute this to reduced need for soft-tissue release and less overall bone resection in the robotic group63. Another prospective randomized controlled trial compared serum cytokine levels pre and postoperatively after patients underwent MAKO or CORI TKA vs. mTKA. The authors found decreased serum C-reactive protein (p = 0.004), interleukin-6 (p < 0.001), and tumor necrosis factor-alpha (p = 0.021) levels at postoperative day 7 with rTKA64. Levels were no longer different by day 28. The authors also hypothesized thiwas due to less soft-tissue release with rTKA, leading to less inflammation and postoperative pain, consistent with other data64,65.

Blood Loss

Minimizing total blood loss is a core principle of surgery. Some authors argue rTKA increases blood loss due to increased case duration. A small study compared 50 robotic TKA vs. 50 mTKA patients and found a 23.7% decrease in blood loss and relative risk of transfusion with rTKA (p < 0.01 and p = 0.02, respectively)66. The same study compared robotic UKA with manual UKA and found no difference in blood loss or transfusion risk66. The authors attribute these differences to more tissue releasing and intramedullary drilling in mTKA. Another retrospective review of primary rTKA vs. conventional TKA found no difference in hemoglobin change or transfusions67. Overall, blood loss is a function of hemostatic technique, case duration, tourniquet use, and tranexamic acid administration rather than robotic use.

Increased tourniquet use may cause greater postoperative pain, greater deep vein thrombosis risk, slower rehab, and nerve injuries of varying severity68-71. Increased surgical time may translate to greater inflation time depending on how one chooses to deploy the tourniquet. One comparison study found no difference in tourniquet-attributed complications between mTKA and rTKA groups despite a small increase in tourniquet time (96.8 vs. 91.6, p < 0.001), which equalized during the final 20 of 148 rTKAs51. There is a general trend toward tourniquet-free TKA to avoid these complications entirely.

Comparison of Each System

This section is summarized in Tables I and II.

TABLE II.

OR Footprints of Each Robot*

Robot OR area (m2) Weight (kg)
CORI <1 m2 (single camera tower with display) Tower: <80 kg
MAKO 1.5 m2 (robotic arm, camera tower, separate display station) Total: 460 kg
ROSA 1.42 m2
 • Robot arm: 150 cm tall, 65 × 120.5 cm (0.78 m2)
 • Camera tower with display: 84.5 × 76.1 cm (0.64 m2)
Robot arm: 320 kg
Tower: 140 kg
Velys 0.85 m2
 • Camera tower with display: 206 cm tall, 81 × 56 cm (0.45 m2)
 • Satellite station (w/arm attached): 156 cm tall, 71 × 56 cm (0.40 m2)
 • Detachable robotic arm: 0.22 m2
Satellite station: 107 kg
Tower: 90.7 kg
TSolution-One • ∼1 m2 (∼6 ft tall) Unknown
TMINI <1 m2 Unknown
*

Information was obtained from vendor distributed publications. All of these systems require a level surface for proper function. Most can be adjusted to minimize occupied space for storage. Manufacturers have specific instructions regarding necessary space/conditions21,72,78,80,81.

Guidance Method

With the exception of TSolution-One, the other systems use optical tracking for computer guidance. The optical tracking systems are very accurate but rely on consistent visualization of arrays by the camera. A key challenge is ensuring the line of sight between the arrays and camera around staff or retractors. TSolution-One holds the knee in fixed flexion (recommended between 110° and 120°) via stabilizing pins and clamps, avoiding the need for optical tracking. A resultant limitation is inability to incorporate flexion and extension laxities into the virtual model24. The surgeon must manually assess pre and postresection soft-tissue balance. In addition, some patients may be unable to achieve the desired degree of flexion to use TSolution-One.

The MAKO and TSolution-One require preoperative fine-cut CT scans with a designated robotic protocol. ROSA has the options of preoperative radiograph, MRI, or imageless guidance. No studies comparing ROSA's accuracy with each method have been published. CORI and Velys, on the other hand, are imageless systems, relying solely on intraoperative mapping, saving cost and radiation. Image-guided or imageless options may be released for most systems in the future.

Resection Method

Resection is accomplished using different methods in each system with various available workflows. No cutting jigs or blocks are used with the MAKO, as cut planes are virtually defined by the computer, restricting the arm's motion to those planes72,73. The Velys robot uses a very similar system with virtual boundaries and haptics74.

By contrast, the ROSA features a haptic bound robotic arm attached to a slotted cutting block with tight tolerance75,76. This special guided block also has holes for guided drilling of pins. The surgeon first performs distal femoral resection with a handheld saw through the guided block. The robot positions the block again to drill holes for a traditional 4-in-1 cut block. Surgeons have the option of using the robotic block to either make the tibial cut directly or drill holes for manual tibial cut block placement. This system also has an optional femoral rotation tool and tensioning instrument. In this workflow, the robot will assess the flexion gap based on the surgeon's tensioner adjustment (similar to conventional gap balancing) and suggest a femoral component rotation that would match the extension gap. The robot positions the guided cut block to drill holes for a 4-in-1 block76,77.

CORI is distinct in its use of a haptic enabled and optically tracked handheld burr connected to the computer via a cord78,79. Resection can take longer with a burr compared with saws, but this allows greater freedom of motion. There are several resection options with the CORI. The distal femur cut is always completed using the burr. The surgeon can then choose to finish the remaining cuts with the burr alone or use guidance to burr holes for a conventional 4-in-1 block. Tibial resection can be entirely performed with the burr, but some elect to burr a flat “shelf” of bone and finish with a tibial cut block aligned with that shelf79.

The TSolution-One uses an autonomous milling tool, similar to the CORI burr, which is attached directly to a robotic arm24. The entire resection must be completed autonomously by the computer-guided resection arm. This step may take longer than the other robots since all resection must be precisely performed. Surgeons do have the option of overriding the autonomous resection and using haptic-bound manual resection.

Unfortunately, no head-to-head comparison studies on final alignment and cut accuracy of each robot have been published.

Safety Measures

Added safety is a quoted benefit of robotic TKA. There is a paucity of studies comparing injury risk of individual robots. All the discussed systems provide visual, haptic, and auditory feedback to the surgeon to ensure resection does not move off the plane. This means locking of mechanisms inside the resection arm (for the ROSA, MAKO, Velys, and TSolution-One) or retracting the resection tool (CORI) when excess motion or force is detected. Virtual boundaries created by the computer also help ensure soft tissues are protected, serving as a “guard rail.” For MAKO and Velys, the “tightness” of these boundaries can be adjusted. All the semiactive systems give real-time visual feedback to the surgeon with color schemes representing optimal and excess resection depths72,74,79,80.

In all the above systems, retractors should still protect key structures. In particular, the patella, patellar tendon, and collateral ligaments are not incorporated into virtual models. Thus, surgeons must take care to prevent errant cuts or destabilizing injuries.

FDA Approval

FDA approval is another important variable to consider, especially for new surgeons. Any cross-manufacturer implant use for closed systems is considered off-label in today's environment. Of note, CORI is currently the only FDA-cleared system for revision TKA, partly because it is imageless. Image-guided revision is challenging due to metal artifact and image distortion on CT or MRI, although artifact reduction protocols can help. In CORI revision TKA, the initial implant's surface is registered as if it were the native joint surface and the desired resection depth is set based on the underlying remaining bone surface. The burr resects bone to the desired depth, creating flat surfaces, and augments are figured into the plan by adjusting implant position in the virtual model until balance is achieved. The surgeon then implants components accordingly. Conversion of UKA or revision with any of the other robots is currently off-label use. Other vendors are devising technologies to expand into revision rTKA as well.

Operating Room Space

Having a robot that is too large for an OR can create difficulties with turnover times and transferring between rooms. When positioning personnel and retractors, smaller spaces compound difficulties with ensuring a clear line of sight between the camera and optical arrays. The footprints of discussed robots are summarized in Table II.24,72,78,80,81. CORI and TMINI were designed with a minimal footprint to help minimize line-of-sight obstructions.

Future of Robotic TKA

Vendors are actively competing to improve their robots by integrating new technologies. Many vendors have introduced augmented reality headsets, which project a user interface onto transparent displays in front of the eyes. By incorporating an optical tracking camera into the headset itself, augmented reality may eliminate line-of-sight difficulties and camera/display towers. Another important emerging technology is artificial intelligence in TKA. Machine learning can, in theory, allow a personalized approach to TKA by making specific recommendations based on patient characteristics. Although more concrete and reliable data are required before computers can be trained to make useful predictions, rTKA allows collection of the necessary data2,82. Rudimentary models are already being developed to predict value metrics after TKA82. Surgeons should pay close attention to this technology moving forward.

Summary

The systems explored here comprise the majority of the US rTKA market in 2024. Each robot offers distinct guidance and resection strategies to achieve a desired plan. While adopting robotic TKA can seem daunting, this article aims to provide a condensed overview for guidance. Surgeons may face unique obstacles or prioritize different variables in this endeavor. Perhaps, the largest change surgeons may face when switching to rTKA is implant restriction with closed systems. Currently, TSolution-One and TMINI are the only open platforms available in America.

The current arthroplasty market has met robotics with cautious optimism. New data are constantly being collected regarding the efficacy and limitations of rTKA. Moreover, cost is a significant driver in today's financially strained healthcare market and the true value proposition of rTKA remains to be fully determined. Subsequent improvements in robotics will likely have a deep impact on the future of arthroplasty.

Footnotes

Disclosure: The Disclosure of Potential Conflicts of Interest forms are provided with the online version of the article (http://links.lww.com/JBJSOA/A659).

Contributor Information

Dimitar I. Semerdzhiev, Email: dsemerdzhiev@achehealth.edu.

Anand Srinivasan, Email: asrinivasanmd@gmail.com.

Farid Amirouche, Email: FAmirouche@northshore.org.

Lalit Puri, Email: LPuri@northshore.org.

Hristo Piponov, Email: piponovmd@gmail.com.

References

  • 1.Shah J, Vyas A, Vyas D. The history of robotics in surgical specialties. Am J Robotic Surg. 2014;1(1):12-20. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2.Beyaz S. A brief history of artificial intelligence and robotic surgery in orthopedics & traumatology and future expectations. Joint Dis Relat Surg. 2020;31(3):653-5. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Mittelstadt B, Paul H, Kazanzides P, Zuhars J, Williamson B, Pettitt R, Cain P, Kloth D, Rose L, Musits B. Development of a surgical robot for cementless total hip replacement. Robotica. 1993;11(6):553-60. [Google Scholar]
  • 4.American Joint Replacement Registry (AJRR). 2023 Annual Report. Rosemont, IL: American Academy of Orthopaedic Surgeons (AAOS), 2023. [Google Scholar]
  • 5.Antonios JK, Korber S, Sivasundaram L, Mayfield C, Kang HP, Oakes DA, Heckmann ND. Trends in computer navigation and robotic assistance for total knee arthroplasty in the United States: an analysis of patient and hospital factors. Arthroplast Today. 2019;5(1):88-95. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Mavrogenis AF, Savvidou OD, Mimidis G, Papanastasiou J, Koulalis D, Demertzis N, Papagelopoulos PJ. Computer-assisted navigation in orthopedic surgery. Orthopedics. 2013;36(8):631-42. [DOI] [PubMed] [Google Scholar]
  • 7.Jacofsky DJ, Allen M. Robotics in arthroplasty: a comprehensive review. J Arthroplasty. 2016;31(10):2353-63. [DOI] [PubMed] [Google Scholar]
  • 8.Zhang J, Ng N, Scott CEH, Blyth MJG, Haddad FS, Macpherson GJ, Patton JT, Clement ND. Robotic arm-assisted versus manual unicompartmental knee arthroplasty: a systematic review and meta-analysis of the MAKO robotic system. Bone Joint J. 2022;104-B(5):541-8. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9.Bell SW, Anthony I, Jones B, MacLean A, Rowe P, Blyth M. Improved accuracy of component positioning with robotic-assisted unicompartmental knee arthroplasty: data from a prospective, randomized controlled study. J Bone Joint Surg Am. 2016;98(8):627-35. [DOI] [PubMed] [Google Scholar]
  • 10.Song E-K, Seon J-K, Yim J-H, Netravali NA, Bargar WL. Robotic-assisted TKA reduces postoperative alignment outliers and improves gap balance compared to conventional TKA. Clin Orthop Relat Res. 2013;471(1):118-26. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Park SE, Lee CT. Comparison of robotic-assisted and conventional manual implantation of a primary total knee arthroplasty. J Arthroplasty. 2007;22(7):1054-9. [DOI] [PubMed] [Google Scholar]
  • 12.Agarwal N, To K, McDonnell S, Khan W. Clinical and radiological outcomes in robotic-assisted total knee arthroplasty: a systematic review and meta-analysis. J Arthroplasty. 2020;35(11):3393-409.e2. [DOI] [PubMed] [Google Scholar]
  • 13.Thienpont E, Fennema P, Price A. Can technology improve alignment during knee arthroplasty. Knee. 2013;20(suppl 1):S21-28. [DOI] [PubMed] [Google Scholar]
  • 14.Scuderi GR, Tria AJ. Minimally Invasive Surgery in Orthopedics. New York, NY: Springer Science & Business Media; 2009. [Google Scholar]
  • 15.Kim Y-H, Yoon S-H, Park J-W. Does robotic-assisted TKA result in better outcome scores or long-term survivorship than conventional TKA? A randomized, controlled trial. Clin Orthop Relat Res. 2020;478(2):266-75. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16.Lang JE, Mannava S, Floyd AJ, Goddard MS, Smith BP, Mofidi A, Seyler TM, Jinnah RH. Robotic systems in orthopaedic surgery. J Bone Joint Surg Br. 2011;93(10):1296-9. [DOI] [PubMed] [Google Scholar]
  • 17.Allen M, Jacofsky D. Evolution of robotics in arthroplasty. In: Lonner JH, ed. Robotics in Knee and Hip Arthroplasty. New York, NY: Springer International Publishing; 2019:13-25. [Google Scholar]
  • 18.Thomas TL, Goh GS, Nguyen MK, Lonner JH. Pin-related complications in computer navigated and robotic-assisted knee arthroplasty: a systematic review. J Arthroplasty. 2022;37(11):2291-307.e2. [DOI] [PubMed] [Google Scholar]
  • 19.Migliorini F, Cuozzo F, Oliva F, Eschweiler J, Hildebrand F, Maffulli N. Imageless navigation for primary total hip arthroplasty: a meta-analysis study. J Orthop Traumatol. 2022;23(1):21. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20.Smith-Bindman R, Lipson J, Marcus R, Kim K-P, Mahesh M, Gould R, Berrington de González A, Miglioretti DL. Radiation dose associated with common computed tomography examinations and the associated lifetime attributable risk of cancer. Arch Intern Med. 2009;169(22):2078-86. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21.Total knee arthroplasty using the THINK surgical robot. 2021. AAOS Orthopedic Video Theater. Available at: https://www.aaos.org/videos/video-detail-page/23038__Videos. Accessed February 18, 2024. [Google Scholar]
  • 22.Fehring TK. Rotational malalignment of the femoral component in total knee arthroplasty. Clin Orthop Relat Res. 2000;380:72-9. [DOI] [PubMed] [Google Scholar]
  • 23.Plaskos C, Hodgson AJ, Inkpen K, McGraw RW. Bone cutting errors in total knee arthroplasty. J Arthroplasty. 2002;17(6):698-705. [DOI] [PubMed] [Google Scholar]
  • 24.Liow MHL, Chin PL, Pang HN, Tay DK-J, Yeo S-J. THINK surgical TSolution-One (Robodoc) total knee arthroplasty. SICOT J. 2017;3:63. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 25.Kamara E, Berliner ZP, Hepinstall MS, Cooper HJ. Pin site complications associated with computer-assisted navigation in hip and knee arthroplasty. J Arthroplasty. 2017;32(9):2842-6. [DOI] [PubMed] [Google Scholar]
  • 26.Bourne RB, Chesworth BM, Davis AM, Mahomed NN, Charron KDJ. Patient satisfaction after total knee arthroplasty: who is satisfied and who is not? Clin Orthop Relat Res. 2010;468(1):57-63. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 27.Tompkins GS, Sypher KS, Li H-F, Griffin TM, Duwelius PJ. Robotic versus manual total knee arthroplasty in high volume surgeons: a comparison of cost and quality metrics. J Arthroplasty. 2022;37(8S):S782-S789. [DOI] [PubMed] [Google Scholar]
  • 28.Nogalo C, Meena A, Abermann E, Fink C. Complications and downsides of the robotic total knee arthroplasty: a systematic review. Knee Surg Sports Traumatol Arthrosc. 2023;31(3):736-50. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 29.St Mart J-P, Goh EL. The current state of robotics in total knee arthroplasty. EFORT Open Rev. 2021;6(4):270-9. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 30.Hoeffel D, Goldstein L, Intwala D, Kaindl L, Dineen A, Patel L, Mayle R. Systematic review and meta-analysis of economic and healthcare resource utilization outcomes for robotic versus manual total knee arthroplasty. J Robotic Surg. 2023;17(6):2899-910. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 31.Rogers EM, Havens AE. Predicting innovativeness. Sociological Inq. 1962;32(1):34-42. [Google Scholar]
  • 32.Khlopas A, Sodhi N, Hozack WJ, Chen AF, Mahoney OM, Kinsey T, Orozco F, Mont MA. Patient-reported functional and satisfaction outcomes after robotic-arm-assisted total knee arthroplasty: early results of a prospective multicenter investigation. J Knee Surg. 2020;33(7):685-90. [DOI] [PubMed] [Google Scholar]
  • 33.Blum CL, Lepkowsky E, Hussein A, Wakelin EA, Plaskos C, Koenig JA. Patient expectations and satisfaction in robotic-assisted total knee arthroplasty: a prospective two-year outcome study. Arch Orthop Trauma Surg. 2021;141(12):2155-64. [DOI] [PubMed] [Google Scholar]
  • 34.Smith AF, Eccles CJ, Bhimani SJ, Denehy KM, Bhimani RB, Smith LS, Malkani AL. Improved patient satisfaction following robotic-assisted total knee arthroplasty. J Knee Surg. 2021;34(7):730-8. [DOI] [PubMed] [Google Scholar]
  • 35.Griffin J, Davis ET, Parsons H, Gemperle Mannion E, Khatri C, Ellard DR, Blyth MJ, Clement ND, Deehan D, Flynn N, Fox J, Grant NJ, Haddad FS, Hutchinson CE, Mason J, Mohindru B, Scott CEH, Smith TO, Skinner JA, Toms AD, Rees S, Underwood M, Metcalfe A. Robotic arthroplasty clinical and cost effectiveness randomised controlled trial (RACER-knee): a study protocol. BMJ Open. 2023;13(6):e068255. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 36.Joo PY, Chen AF, Richards J, Law TY, Taylor K, Marchand K, Clark G, Collopy D, Marchand RC, Roche M, Mont MA, Malkani AL. Clinical results and patient-reported outcomes following robotic-assisted primary total knee arthroplasty: a multicentre study. Bone Joint Open. 2022;3(8):589-95. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 37.Marchand KB, Moody R, Scholl LY, Bhowmik-Stoker M, Taylor KB, Mont MA, Marchand RC. Results of robotic-assisted versus manual total knee arthroplasty at 2-year follow-up. J Knee Surg. 2023;36(2):159-66. [DOI] [PubMed] [Google Scholar]
  • 38.Brown ML, Javidan P, Early S, Bugbee W. Evolving etiologies and rates of revision total knee arthroplasty: a 10-year institutional report. Arthroplasty. 2022;4(1):39. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 39.Kirchner GJ, Stambough JB, Jimenez E, Nikkel LE. Robotic-assisted TKA is not associated with decreased odds of early revision: an analysis of the American Joint Replacement Registry. Clin Orthop Relat Res. 2024;482(2):303-10. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 40.Xu J, Li L, Fu J, Xu C, Ni M, Chai W, Hao L, Zhang G, Chen J. Early clinical and radiographic outcomes of robot-assisted versus conventional manual total knee arthroplasty: a randomized controlled study. Orthop Surg. 2022;14(9):1972-80. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 41.Gunaratne R, Pratt DN, Banda J, Fick DP, Khan RJK, Robertson BW. Patient dissatisfaction following total knee arthroplasty: a systematic review of the literature. J Arthroplasty. 2017;32(12):3854-60. [DOI] [PubMed] [Google Scholar]
  • 42.Hiranaka T, Suda Y, Saitoh A, Tanaka A, Arimoto A, Koide M, Fujishiro T, Okamoto K. Current concept of kinematic alignment total knee arthroplasty and its derivatives. Bone Joint Open. 2022;3(5):390-7. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 43.Rajan PV, Khlopas A, Klika A, Molloy R, Krebs V, Piuzzi NS. The cost-effectiveness of robotic-assisted versus manual total knee arthroplasty: a Markov model-based evaluation. J Am Acad Orthop Surg. 2022;30(4):168-76. [DOI] [PubMed] [Google Scholar]
  • 44.Condon A. A breakdown of 7 robots in spine surgery; 2021. Available at: https://www.beckersspine.com/robotics/52042-a-breakdown-of-7-robots-in-spine-surgery.html. Accessed September 10, 2023.
  • 45.Walgrave S, Oussedik S. Comparative assessment of current robotic-assisted systems in primary total knee arthroplasty. Bone Joint Open. 2023;4(1):13-8. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 46.Brinkman JC, Christopher ZK, Moore ML, Pollock JR, Haglin JM, Bingham JS. Patient interest in robotic total joint arthroplasty is exponential: a 10-year Google trends analysis. Arthroplast Today. 2022;15:13-8. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 47.Peterman NJ, Pagani N, Mann R, Li RL, Gasienica J, Naik A, Sun D. Disparities in access to robotic knee arthroplasty: a geospatial analysis. J Arthroplasty. 2024;39(4):864-70. [DOI] [PubMed] [Google Scholar]
  • 48.Callahan LF, Cleveland RJ, Allen KD, Golightly Y. Racial/ethnic, socioeconomic, and geographic disparities in the epidemiology of knee and hip osteoarthritis. Rheum Dis Clin. 2021;47(1):1-20. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 49.Boylan M, Suchman K, Vigdorchik J, Slover J, Bosco J. Technology-assisted hip and knee arthroplasties: an analysis of utilization trends. J Arthroplasty. 2018;33(4):1019-23. [DOI] [PubMed] [Google Scholar]
  • 50.Deckey DG, Verhey JT, Rosenow CS, Doan MK, McQuivey KS, Joseph AM, Schwartz AJ, Clarke HD, Bingham JS. Robotic-assisted total knee arthroplasty allows for trainee involvement and teaching without lengthening operative time. J Arthroplasty. 2022;37(6S):S201-S206. [DOI] [PubMed] [Google Scholar]
  • 51.Mitchell J, Wang J, Bukowski B, Greiner J, Wolford B, Oyer M, Illgen RL. Relative clinical outcomes comparing manual and robotic-assisted total knee arthroplasty at minimum 1-year follow-up. HSS J. 2021;17(3):267-73. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 52.Bolam SM, Tay ML, Zaidi F, Sidaginamale RP, Hanlon M, Munro JT, Monk AP. Introduction of ROSA robotic-arm system for total knee arthroplasty is associated with a minimal learning curve for operative time. J Exp Orthop. 2022;9(1):86. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 53.Chen Z, Bhowmik-Stoker M, Palmer M, Coppolecchia A, Harder B, Mont MA, Marchand RC. Time-based learning curve for robotic-assisted total knee arthroplasty: a multicenter study. J Knee Surg. 2023;36(8):873-7. [DOI] [PubMed] [Google Scholar]
  • 54.Marchand KB, Ehiorobo J, Mathew KK, Marchand RC, Mont MA. Learning curve of robotic-assisted total knee arthroplasty for a high-volume surgeon. J Knee Surg. 2022;35(4):409-15. [DOI] [PubMed] [Google Scholar]
  • 55.Cohen TN, Anger JT, Shamash K, Cohen KA, Nasseri Y, Francis SE, Fimpler M, Avenido R, Gewertz BL, Shouhed D. Discovering the barriers to efficient robotic operating room turnover time: perceptions vs. reality. J Robotic Surg. 2020;14(5):717-24. [DOI] [PubMed] [Google Scholar]
  • 56.Souders CP, Catchpole KR, Wood LN, Solnik JM, Avenido RM, Strauss PL, Eilber KS, Anger JT. Reducing operating room turnover time for robotic surgery using a motor racing pit stop model. World J Surg. 2017;41(8):1943-9. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 57.Beldame J, Boisrenoult P, Beaufils P. Pin track induced fractures around computer-assisted TKA. Orthop Traumatol Surg Res. 2010;96(3):249-55. [DOI] [PubMed] [Google Scholar]
  • 58.Smith TJ, Siddiqi A, Forte SA, Judice A, Sculco PK, Vigdorchik JM, Schwarzkopf R, Springer BD. Periprosthetic fractures through tracking pin sites following computer navigated and robotic total and unicompartmental knee arthroplasty: a systematic review. JBJS Rev. 2021;9(1):e20.00091. [DOI] [PubMed] [Google Scholar]
  • 59.Yun AG, Qutami M, Pasko KBD. Do bicortical diaphyseal array pins create the risk of periprosthetic fracture in robotic-assisted knee arthroplasties? Arthroplasty. 2021;3(1):25. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 60.Desai SS, Kunes JA, Held MB, Ren M, deMeireles AJ, Geller JA, Shah RP, Cooper HJ. A comparison of pin site complications between large and small pin diameters in robotic-assisted total knee arthroplasty. J Exp Orthop. 2023;10(1):22. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 61.Shetty T, Nguyen JT, Sasaki M, Wu A, Bogner E, Burge A, Cogsil T, Dalal A, Halvorsen K, Cummings K, Su EP, Lyman S. Risk factors for acute nerve injury after total knee arthroplasty. Muscle Nerve. 2018;57(6):946-50. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 62.Beltran MJ, Collinge CA, Patzkowski JC, Masini BD, Blease RE, Hsu JR, Skeletal Trauma Research Consortium STReC. The safe zone for external fixator pins in the femur. J Orthop Trauma. 2012;26(11):643-7. [DOI] [PubMed] [Google Scholar]
  • 63.Kayani B, Konan S, Pietrzak JRT, Haddad FS. Iatrogenic bone and soft tissue trauma in robotic-arm assisted total knee arthroplasty compared with conventional jig-based total knee arthroplasty: a prospective cohort study and validation of a new classification system. J Arthroplasty. 2018;33(8):2496-501. [DOI] [PubMed] [Google Scholar]
  • 64.Kayani B, Tahmassebi J, Ayuob A, Konan S, Oussedik S, Haddad FS. A prospective randomized controlled trial comparing the systemic inflammatory response in conventional jig-based total knee arthroplasty versus robotic-arm assisted total knee arthroplasty. Bone Joint J. 2021;103-B(1):113-22. [DOI] [PubMed] [Google Scholar]
  • 65.Khlopas A, Sodhi N, Sultan AA, Chughtai M, Molloy RM, Mont MA. Robotic arm–assisted total knee arthroplasty. J Arthroplasty. 2018;33(7):2002-6. [DOI] [PubMed] [Google Scholar]
  • 66.Khan H, Dhillon K, Mahapatra P, Popat R, Zakieh O, Kim WJ, Nathwani D. Blood loss and transfusion risk in robotic-assisted knee arthroplasty: a retrospective analysis. Int J Med Robot. 2021;17(6):e2308. [DOI] [PubMed] [Google Scholar]
  • 67.Stimson LN, Steelman KR, Hamilton DA, Chen C, Darwiche HF, Mehaidli A. Evaluation of blood loss in conventional vs MAKOplasty total knee arthroplasty. Arthroplast Today. 2022;16:224-8. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 68.Tarwala R. Tourniquet versus no tourniquet in total knee arthroplasty: we don't have a winner yet: commentary on article by Rahul Goel, MD, et al: “Tourniquet use does not affect functional outcomes or pain after total knee arthroplasty. A prospective, double-blinded, randomized controlled trial.” J Bone Joint Surg Am. 2019;101(20):e109. [DOI] [PubMed] [Google Scholar]
  • 69.Arthur JR, Spangehl MJ. Tourniquet use in total knee arthroplasty. J Knee Surg. 2019;32(8):719-29. [DOI] [PubMed] [Google Scholar]
  • 70.Tai T-W, Chang C-W, Lai K-A, Lin C-J, Yang C-Y. Effects of tourniquet use on blood loss and soft-tissue damage in total knee arthroplasty: a randomized controlled trial. J Bone Joint Surg Am. 2012;94(24):2209-15. [DOI] [PubMed] [Google Scholar]
  • 71.McCarthy Deering E, Hu SY, Abdulkarim A. Does tourniquet use in TKA increase postoperative pain? A systematic review and meta-analysis. Clin Orthop Relat Res. 2019;477(3):547-58. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 72.MAKO Robotic Arm System User Guide. Ft Lauderdale, FL: Mako Surgical Corp; 2015. [Google Scholar]
  • 73.Roche M. The MAKO robotic-arm knee arthroplasty system. Arch Orthop Trauma Surg. 2021;141(12):2043-7. [DOI] [PubMed] [Google Scholar]
  • 74.Clatworthy M. Patient-specific TKA with the VELYSTM robotic-assisted solution. Surg Technology Int. 2022;40:315-20. [DOI] [PubMed] [Google Scholar]
  • 75.Knapp PW, Nett MP, Scuderi GR. Optimizing total knee arthroplasty with ROSA® robotic technology. Surg Technol Int. 2022;40:289-96. [DOI] [PubMed] [Google Scholar]
  • 76.ROSA Knee User Manual and Surgical Technique v1.4. Montreal, Canada: Zimmer CAS. [Google Scholar]
  • 77.Batailler C, Hannouche D, Benazzo F, Parratte S. Concepts and techniques of a new robotically assisted technique for total knee arthroplasty: the ROSA knee system. Arch Orthop Trauma Surg. 2021;141(12):2049-58. [DOI] [PubMed] [Google Scholar]
  • 78.Real Intelligence. CORI for Knee Arthroplasty User Manual. Version 500185 REVB. Memphis, TN: Simth & Nephew. [Google Scholar]
  • 79.Adamska O, Modzelewski K, Szymczak J, Świderek J, Maciąg B, Czuchaj P, Poniatowska M, Wnuk A. Robotic-assisted total knee arthroplasty utilizing NAVIO, CORI imageless systems and manual TKA accurately restore femoral rotational alignment and yield satisfactory clinical outcomes: a randomized controlled trial. Medicina (Kaunas, Lithuania). 2023;59(2):236. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 80.2300.6-GLBL-en-ROSA-Knee-System-User-Manual-and-Surg-Tech-digital1.pdf. Available at: https://www.zimmerbiomet.com/content/dam/zb-corporate/en/education-resources/surgical-techniques/specialties/robotics/2300.6-GLBL-en-ROSA-Knee-System-User-Manual-and-Surg-Tech-digital1.pdf. Accessed September 10, 2023.
  • 81.VELYS robotic-assisted solution operating dimensions. Raynham, MA: DePuy Synthes; 2021. p. 1-2. Report No.: 103723271. [Google Scholar]
  • 82.Ramkumar PN, Karnuta JM, Navarro SM, Haeberle HS, Scuderi GR, Mont MA, Krebs VE, Patterson BM. Deep learning preoperatively predicts value metrics for primary total knee arthroplasty: development and validation of an artificial neural network model. J Arthroplasty. 2019;34(10):2220-7.e1. [DOI] [PubMed] [Google Scholar]