A new dynamic in vitro model for evaluating antimicrobial activity against bacterial biofilms on central venous catheters

Liangyan Fang 1,2, Yunqian Qiao 1,2, Xiuting Li 1,2, Changbin Wang 1,2, Chunqiao Li 1,2, Tongqing Luan 1,2, Wenqing Wang 1,2,
Editor: Renato Kovacs3
PMCID: PMC11370253  PMID: 39058030

ABSTRACT

Central venous catheters (CVCs) are widely used for intravenous medication administration. However, biofilm formation along the catheter surface is the main most important cause of catheter-related bloodstream infections. Nowadays, several antimicrobial-coated catheters are available to prevent biofilm development. In this study, we introduced a new dynamic in vitro model to evaluate the antimicrobial activity against bacterial biofilms on CVCs. Rifampicin–minocycline-coated catheters and control catheters without antimicrobial component were assembled into the model to test the antimicrobial activity on external surface and internal surface. After 1 h irrigation of Staphylococcus epidermidis or Staphylococcus aureus preculture and 23 h irrigation of Trypticase Soy Broth, the viable adherent organism was collected and counted. The enumeration results showed that the number of bacteria attached to antibacterial catheter was significantly less than that of the control catheter, both on external surface (P < 0.05) and internal surface (P < 0.05). The results were further confirmed by the scanning electron microscopy. In conclusion, the dynamic in vitro model can be applied to evaluate the antimicrobial activity against bacterial biofilms grown on the external and internal surfaces of CVCs used in clinical practice.

IMPORTANCE

For the first time, a new dynamic in vitro model was constructed to evaluate the antimicrobial activity against bacterial biofilms on central venous catheters (CVCs) on both external surface and internal surface. This model could be applied to evaluate the antimicrobial activity against bacterial biofilms not only on CVCs but also other types of catheters.

KEYWORDS: central venous catheter, bloodstream infection, in vitro model, antimicrobial, biofilm inhibition

INTRODUCTION

Central venous catheters (CVCs) are widely used for intravenous medication administration, parenteral nutrition, hemodynamics surveillance, and fluid replacement (1). It is widely used in clinical practice, especially in intensive care units (ICUs) (2). However, the utilization of CVCs may result in catheter-related bloodstream infection (CRBSI), leading to significant morbidity and mortality, increased cost for hospitalized patients, and extended hospital stays (3). In China, it is reported that the incidence of CRBSI varied from 1.67 to 16.96 per 1,000 catheter days (4).

Microbial colonization and biofilm formation along catheter surfaces are the main causes of CRBSI. Biofilm is a complex organic material consisting of microorganisms growing in colonies within an extracellular mucopolysaccharide substance, which they produce, allowing the bacteria to be tolerant to antiseptics and antibiotics (5). Microorganisms may colonize both external and luminal surfaces of CVCs, and it is reported that short-term (<10 days) catheters in place are more heavily colonized on the external surfaces while catheters for up to 30 days tend to be more heavily colonized within the lumen (6). Bacteria that colonize CVCs may originate from the skin at the insertion site and migrate along the catheter’s external surface (named external pathway), or originate through hematogenous seeding of the device from another infected nidus in the catheterized patient (named luminal pathway). The most common microorganisms recovered from CVCs include coagulase negative staphylococci, Staphylococcus aureus, Pseudomonas aeruginosa, Enterococcus faecalis, and Candida species (7). Nowadays, different antimicrobial-coated catheters have been developed to prevent biofilm development, which can be divided into release-based and contact-based catheters based on their antimicrobial strategies (7).

At present, several approaches have been reported to assess the antimicrobial effect of catheters using in vitro models, including static models (6, 8) and dynamic models (9, 10). The static models test the antimicrobial activity against microorganism growing on 96-well polystyrene plates, without reflecting the biofilm conditions on medical devices. Several flow systems have been studied to assess in vitro anti-biofilm efficacy by simulating the in vivo situations, but most of them have the limitation that the device has to be shaped in a determined form and size for biofilm formation (11, 12). Besides, a dynamic in vitro model for evaluating biofilm inhibition activity using clinical-used catheters was reported (9), but the system could only mimic extraluminal biofilm formation. Thus, an in vitro model that can mimic luminal biofilm formation is still needed.

The aim of this study was to develop an in vitro model that enables the evaluation of the effect of antimicrobial-coated catheters on biofilms on both external and luminal surfaces. The experimental parameters were designed to simulate the colonization of microorganism and biofilm formation on CVCs.

MATERIALS AND METHODS

Bacterial strains and culture conditions

S. aureus ATCC 6538 and Staphylococcus epidermidis ATCC 12228 were used in this study. Frozen stock culture of strain was first inoculated onto Trypticase Soy Agar (TSA, BD, USA) at 35°C for 24 h. Single bacterial colonies were taken from the culture and transferred into Trypticase Soy Broth (TSB, BD, USA) until the mid-exponential phase.

Catheters

Commercially available antimicrobial CVCs (made of polyurethane, external diameter: 2.1 mm) [Lepu Medical Technology (Beijing) Co., Ltd, China] were tested in the study. Disposable CVCs (made of polyurethane, external diameter: 2.1 mm) (Beijing Target Medical Technologies, Inc, China) with no antimicrobial agent were served as control. According to the manufacturer’s instructions, the antimicrobial CVCs are coated with rifampicin (150 µg/cm) and minocycline hydrochloride (150 µg/cm), both on the external and luminal surfaces. And the manufacturer claims that the loading density per square cm was equal between external and luminal surfaces.

Zone of inhibition test

The inhibition zone assay protocol was based on Clinical and Laboratory Standards Institute guidelines (13). The zone of inhibition assay was performed for both external and luminal surfaces. For the external surface, the antimicrobial and control catheters were cut into 0.5 cm long segments and vertically embedded in TSA plates, which were swab-inoculated with one of the bacteria (appropriate 105 CFU/plate). The plates were incubated overnight at 30–35°C. Diameters of the inhibition zones (excluding the catheter diameters) were measured to reflect the efficacy of bacterial inhibition. To assay antibiotic leaching from the luminal surface, we cut the catheter segments in half longitudinally and then vertically embedded in TSA plates swab inoculated with bacteria.

Biofilm reactor model system design

The in vitro biofilm reactor model developed by our group is designed as a dynamic system. The device contains six independent reactors, inside which six catheter segments are implanted independently, three reactors for antimicrobial CVCs and three for control CVCs. For each reactor, a temperature control compartment is set outside the reactor by connecting with a temperature control water circuit. The water flow is parallel to the axis of all samples (catheter segments). This model is designed for extraluminal and intraluminal antimicrobial effect testing, respectively. For intraluminal antimicrobial effect test, the implanted catheter segments are connected with the medium pipeline to allow liquids flow through (Fig. 1A and B). For evaluating the extraluminal antimicrobial effect, catheter segments with both ends sealed with nails are placed in the reactor aseptically and exposed to the flow medium (Fig. 1C and D). The biofilm reactor model is sterilized by high-pressure steam treatment.

FIG 1.

FIG 1

The schematic diagram of biofilm reactor model system for CVCs. The model system for internal (A) and external (C) surface of CVCs. The reactor detail for internal surface (B) and external surface (D) of CVCs.

Biofilm formation

The bacterial culture (overnight culture) in TSB was simultaneously pumped through the six reactors in the device at the same speed of 40 mL/h for 1 h. Sterilized TSB medium, which simulates the blood flow was pumped to establish biofilm with a continuous flow of nutrients, at the speed of 20 mL/h for an additional 23 h. The control catheters were used to check normal biofilm formation.

Enumeration of viable adherent microorganisms

After irrigation by bacterial suspension for 1 h, followed by TSB medium for 23 h, catheter segments were removed from the reactor model system and cut into 1 cm long segments. The catheter segments were washed gently with 20 mL phosphate-buffered saline (PBS, pH 7.2) for three times to remove free and loosely adhering bacteria and then moved to 15 mL sterilized centrifuge tubes prefilled with 5 mL PBS. The high-speed vortex was performed for 100 s, followed by sonication for 5 min at 40 kHz to safely and completely recover adherent cells from catheter segments. While for the bacteria on the luminal surface, the catheter segments were sliced in half longitudinally in order to expose the luminal surface, and then washed with 20 mL PBS for 3 min (9, 14). The bacterial suspension was subjected to suitable gradient dilution with PBS, 1 mL of which was transferred into separate 90 mm Petri dishes. 15–20 mL TSA, cooled to 42–45°C, was then poured into dishes. The agar plates were incubated at 30–35°C for 48 h to count the number of bacterial colonies. The counting results were standardized to the catheter surface area (CFU/cm2).

Scanning electron microscopy

Scanning electron microscopy was performed to visualize the biofilm formation on the surface of catheter segments. The segments removed from the device were fixed with 2.5% glutaraldehyde in sodium cacodylate buffer for 2 h at 4°C, and then dehydrated in a series of aqueous ethanol (30%, 50%, 70%, 90%, and 100%) (8). For the luminal surfaces, we first sliced catheter segments in half longitudinally in order to expose the luminal surface, followed by fixation and dehydration. Biofilm was examined in a SEM-FEG Hitachi SU8010 (Hitachi High-Technologies Europe GmbH; Krefeld, Germany).

Statistical analysis

Statistical analysis was performed by SPSS software. All testings were performed in triplicate. The data were recorded as mean ± standard deviation. The number of viable microorganisms adhering to the catheters were recorded as CFU/cm2 and compared by the t-test (P < 0.05 was considered statistically significant).

RESULTS

Determination of inhibition zones

Since the loaded antibiotics can be released and diffuse out, the antibacterial activities of CVCs were initially estimated by the agar diffusion assay. For the external surface, the zones of inhibition of each organism were represented in Fig. 2A and B (S. aureus) (S. epidermidis). The diameter of the inhibition zones was recorded and presented in Table 1. The results showed that, compared to control catheters, the antibacterial catheters could inhibit the growth of S. aureus and S. epidermidison agar plates. For the luminal surface, we cut the catheter segments in half longitudinally, and then vertically embedded in TSA plates premixed with bacteria as presented in Fig. 2C and D (S. aureus) (S. epidermidis). We found that the zone of inhibition was still a complete circle, indicating that the inhibition zones of the luminal surface were consistent with that on the external surface.

FIG 2.

FIG 2

The inhibition zones of CVCs against S. aureus (A and C) and S. epidermidis (B and D). Orange sample: antibacterial CVC; white sample: control CVC.

TABLE 1.

The diameter of zones of inhibition

Organism Sample Diameter of zone of inhibition (meana± standard deviation, mm)
S. aureus Control CVC None
Antibacterial CVC 26 ± 0.53
S. epidermidis Control CVC None
Antibacterial CVC 33 ± 0.49
a

Mean of the zones of three segments.

Extraluminal antimicrobial effect test

In order to validate the antimicrobial effect of CVCs using the in vitro model built in this study, extraluminal activity was first tested. The catheters sealed with nails were placed in the reactor and exposed to the flow medium. After 24 h of perfusion, the adherent bacteria represented the biofilm attached to and colonized on the external surface. The viable adhered cells were recovered by validated vortexing and sonication method, and quantified by agar plating. As shown in Fig. 3A, the number of S. aureus attached to the external surface of antibacterial CVCs (9.5 × 106 ± 2.9 × 106 CFU/cm2) was significantly less than that of the control catheters (2.7 × 107 ± 5.0 × 106 CFU/cm2) (P < 0.05). As for S. epidermidis, a significant difference was also presented between the number of bacteria attached to antibacterial catheter surfaces (1.3 × 107 ± 2.7 × 106 CFU/cm2) and that of the control catheters (4.9 × 107 ± 1.4 × 107 CFU/cm2) (P < 0.05) (Fig. 3C). Furthermore, the SEM was performed to observe the bacterial adhesion of the samples after 24 h of co-incubation with S. aureus (Fig. 3B) and S. epidermidis (Fig. 3D). There was less bacterial adhesion on the surface of anti-microbial CVC samples compared with control CVC for both S. aureus and S. epidermidis. The SEM results were consistent with the zones of inhibition and colony counting results.

FIG 3.

FIG 3

Extraluminal antimicrobial activities of CVCs. The comparative number of S. aureus (A) and S. epidermidis (C) attached to external surfaces of catheters. The representative scanning electronic micrograph of S. aureus (B) and S. epidermidis (D) biofilm formation on CVCs.

Intraluminal antimicrobial effect test

For intraluminal antimicrobial effect test, the implanted catheter segments were connected with the medium pipeline. After removed from the reactor, 1 cm segments were cut in half longitudinally. For S. aureus, the count of bacteria attached to the inner surface of antibacterial catheters (2.2 × 107 ± 6.6 × 106 CFU/cm2) was significantly reduced than that of control catheters (5.2 × 107 ± 8.2 × 106 CFU/cm2) (P < 0.05) (Fig. 4A). Similarly, the number of S. epidermidis of antibacterial catheter (6.1 × 106 ± 1.3 × 106 CFU/cm2) was significantly decreased relative to the number of bacteria attached to control catheter surface (1.8 × 107 ± 2.5 × 106 CFU/cm2) (P < 0.05) (Fig. 4C). The SEM was performed to observe the bacterial adhesion of the samples after 24 h of co-incubation with S. aureus (Fig. 4B) and S. epidermidis (Fig. 4D). There was less bacterial adhesion on the surface of anti-microbial CVC samples compared with control CVC for both S. aureus and S. epidermidis. The SEM results were consistent with the colony counting results.

FIG 4.

FIG 4

Intraluminal antimicrobial effect test. The comparative number of S. aureus (A) and S. epidermidis (C) attached to external surfaces of catheters. The representative scanning electronic micrograph of S. aureus (B) and S. epidermidis (D) biofilm formation on CVCs.

DISCUSSION

Nowadays, CRBSI is a significant public health problem and has an associated economic impact. In response to the increased prevalence of catheter-related infections, different strategies have been used to protect catheter surfaces from microbial colonization and biofilm development. These include physical/chemical modifications of catheter surfaces (7) and coating/impregnating catheters with antimicrobial agents, such as antibiotics, antiseptic, silver, or chitosan (1517). However, due to the high risk of creating microbial resistance to antimicrobial treatments, the US Food and Drug Administration (18) discourages the abuse of antimicrobial agents on devices unless a strong rationale is provided to demonstrate the clinical benefit outweighs the associated risk (18). FDA recommends medical device manufacturers provide detailed information and pre-clinical testing to verify the safety and efficacy of devices that include antimicrobial agents (18). Thus, in order to promote the proper development and evaluation of the antimicrobial efficacy of medical devices, appropriate in vitro testing tools are needed to mimic the clinical settings and analyze the performance of devices before performing delicate and expensive in vivo studies. With regard to antimicrobial CVC, despite intensive research in biofilm preventive strategies, there is only a limited choice of in vitro evaluation models.

At present, there is no universally accepted in vitro model to reproduce the biofilm formation on catheter. Curtin et al. developed an in vitro model to mimic biofilm formation on luminal surface based on a modified drip flow reactor (14). And Garcia et al. designed a device (called the Sevilla device) that enables antimicrobial activity against bacterial biofilm to be evaluated on the external surface of intravascular catheter (9). In this study, we developed a well-established in vitro model. With the use of this model, we evaluated the antibacterial effect of commercial minocycline–rifampicin coating CVCs on both external surface and luminal surface with continuous culture supply.

In the last two decades, CVCs coated with conjugated minocycline and rifampicin have been widely used because of their additive properties (7, 19). Rifampin inhibits RNA synthesis in bacteria by binding to and inhibiting bacterial RNA polymerase (20), and minocycline inhibits bacterial protein synthesis by preventing the association of tRNA with the bacterial ribosome (21). Both of them are highly potent against key Gram-positive pathogens and some Gram-negative pathogens, but have less potency against other Gram-negative pathogens and fungal pathogens (22). In the present study, minocycline−rifampicin-impregnated CVCs were applied to detect the antibacterial activity.

CVC infections are mainly due to coagulase negative staphylococci, S. aureus, Klebsiella pneumoniae, P. aeruginosa, E. faecalis, and Candida species (7). In this study, S. epidermidis and S. aureus were selected to detect the biofilm inhibition effect of antimicrobial CVCs. The antibacterial catheters presented a powerful inhibition effect on the S. aureus and S. epidermidis plates. And the enumeration of viable adherent organisms on antimicrobial catheters was significantly less than that on control catheters (P < 0.05) for both external and luminal surfaces. These results were in accordance with the research that these two antibiotics are powerful agents against Gram-positive pathogens.

Still, there are some limitations in this study. More work is needed to examine the antimicrobial effect on Gram-negative pathogens such as K. pneumoniae and P. aeruginosa. Besides, further work is needed to evaluate the antimicrobial effect for a longer period of time that correlated with the in vivo situation using this model. CVCs with different mechanisms of action will be tested using this model to validate its wide applicability.

In conclusion, we successfully designed a dynamic in vitro model to mimic biofilm formation on CVCs. Using this model, the antimicrobial effect was examined on both external and luminal surfaces of minocycline–rifampicin-coated catheters.

ACKNOWLEDGMENTS

This study was funded by the Shandong Institute of Medical Device and Pharmaceutical Packaging Inspection (NB202315).

Contributor Information

Wenqing Wang, Email: 22741688@qq.com.

Renato Kovacs, University of Debrecen, Debrecen, Hungary.

REFERENCES

  • 1. Broadhurst D, Moureau N, Ullman AJ. 2016. Central venous access devices site care practices: an international survey of 34 countries. J Vasc Access 17:78–86. doi: 10.5301/jva.5000450 [DOI] [PubMed] [Google Scholar]
  • 2. Kwakman PH, Müller MC, Binnekade JM, van den Akker JP, de Borgie CA, Schultz MJ, Zaat SA. 2012. Medical-grade honey does not reduce skin colonization at central venous catheter-insertion sites of critically ill patients: a randomized controlled trial. Crit Care 16:R214. doi: 10.1186/cc11849 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3. Furuya EY, Dick AW, Herzig CTA, Pogorzelska-Maziarz M, Larson EL, Stone PW. 2016. Central line-associated bloodstream infection reduction and bundle compliance in intensive care units: a national study. Infect Control Hosp Epidemiol 37:805–810. doi: 10.1017/ice.2016.67 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4. Peng S, Lu Y. 2013. Clinical epidemiology of central venous catheter-related bloodstream infections in an intensive care unit in China. J Crit Care 28:277–283. doi: 10.1016/j.jcrc.2012.09.007 [DOI] [PubMed] [Google Scholar]
  • 5. Wolcott R. 2021. Biofilm and catheter-related bloodstream infections. Br J Nurs 30:S4–S9. doi: 10.12968/bjon.2021.30.8.S4 [DOI] [PubMed] [Google Scholar]
  • 6. Sharma S, Mohler J, Mahajan SD, Schwartz SA, Bruggemann L, Aalinkeel R. 2023. Microbial biofilm: a review on formation, infection, antibiotic resistance, control measures, and innovative treatment. Microorganisms 11:1614. doi: 10.3390/microorganisms11061614 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7. Ricardo SIC, Anjos IIL, Monge N, Faustino CMC, Ribeiro IAC. 2020. A glance at antimicrobial strategies to prevent catheter-associated medical infections. ACS Infect Dis 6:3109–3130. doi: 10.1021/acsinfecdis.0c00526 [DOI] [PubMed] [Google Scholar]
  • 8. Edmiston CE, Goheen MP, Seabrook GR, Johnson CP, Lewis BD, Brown KR, Towne JB. 2006. Impact of selective antimicrobial agents on staphylococcal adherence to biomedical devices. Am J Surg 192:344–354. doi: 10.1016/j.amjsurg.2006.04.009 [DOI] [PubMed] [Google Scholar]
  • 9. García I, Conejo M del C, Ojeda A, Rodríguez-Baño J, Pascual A. 2010. A dynamic in vitro model for evaluating antimicrobial activity against bacterial biofilms using a new device and clinical-used catheters. J Microbiol Methods 83:307–311. doi: 10.1016/j.mimet.2010.09.017 [DOI] [PubMed] [Google Scholar]
  • 10. Lehman SM, Donlan RM. 2015. Bacteriophage-mediated control of a two-species biofilm formed by microorganisms causing catheter-associated urinary tract infections in an in vitro urinary catheter model. Antimicrob Agents Chemother 59:1127–1137. doi: 10.1128/AAC.03786-14 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11. Gander S, Hayward K, Finch R. 2002. An investigation of the antimicrobial effects of linezolid on bacterial biofilms utilizing an in vitro pharmacokinetic model. J Antimicrob Chemother 49:301–308. doi: 10.1093/jac/49.2.301 [DOI] [PubMed] [Google Scholar]
  • 12. El-Azizi M, Khardori N. 2016. Efficacy of ultraviolet C light at sublethal dose in combination with antistaphylococcal antibiotics to disinfect catheter biofilms of methicillin-susceptible and methicillin-resistant Staphylococcus aureus and Staphylococcus epidermidis in vitro. Infect Drug Resist 9:181–189. doi: 10.2147/IDR.S109343 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13. Balouiri M, Sadiki M, Ibnsouda SK. 2016. Methods for in vitro evaluating antimicrobial activity: a review. J Pharm Anal 6:71–79. doi: 10.1016/j.jpha.2015.11.005 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14. Curtin JJ, Donlan RM. 2006. Using bacteriophages to reduce formation of catheter-associated biofilms by Staphylococcus epidermidis. Antimicrob Agents Chemother 50:1268–1275. doi: 10.1128/AAC.50.4.1268-1275.2006 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15. Mermel LA. 2015. Effectiveness of minocycline/rifampin vs chlorhexidine/silver sulfadiazine-impregnated central venous catheters. J Am Coll Surg 221:891–892. doi: 10.1016/j.jamcollsurg.2015.07.004 [DOI] [PubMed] [Google Scholar]
  • 16. Santos GOD, Milanesi FC, Greggianin BF, Fernandes MI, Oppermann RV, Weidlich P. 2017. Chlorhexidine with or without alcohol against biofilm formation: efficacy, adverse events and taste preference. Braz Oral Res 31:e32. doi: 10.1590/1807-3107BOR-2017.vol31.0032 [DOI] [PubMed] [Google Scholar]
  • 17. Mendoza G, Regiel-Futyra A, Tamayo A, Monzon M, Irusta S, de Gregorio MA, Kyzioł A, Arruebo M. 2018. Chitosan-based coatings in the prevention of intravascular catheter-associated infections. J Biomater Appl 32:725–737. doi: 10.1177/0885328217739199 [DOI] [PubMed] [Google Scholar]
  • 18. Premarket notification [510(k)] submissions for medical devices that include antimicrobial agents. 2007. Department of Health and Human Services Food and Drug Administration, U.S [Google Scholar]
  • 19. Bonne S, Mazuski JE, Sona C, Schallom M, Boyle W, Buchman TG, Bochicchio GV, Coopersmith CM, Schuerer DJE. 2015. Effectiveness of minocycline and rifampin vs chlorhexidine and silver sulfadiazine-impregnated central venous catheters in preventing central line-associated bloodstream infection in a high-volume academic intensive care unit: a before and after trial. J Am Coll Surg 221:739–747. doi: 10.1016/j.jamcollsurg.2015.05.013 [DOI] [PubMed] [Google Scholar]
  • 20. Campbell EA, Korzheva N, Mustaev A, Murakami K, Nair S, Goldfarb A, Darst SA. 2001. Structural mechanism for rifampicin inhibition of bacterial RNA polymerase. Cell 104:901–912. doi: 10.1016/s0092-8674(01)00286-0 [DOI] [PubMed] [Google Scholar]
  • 21. Chopra I, Roberts M. 2001. Tetracycline antibiotics: mode of action, applications, molecular biology, and epidemiology of bacterial resistance. Microbiol Mol Biol Rev 65:232–260. doi: 10.1128/MMBR.65.2.232-260.2001 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22. Viola GM, Rosenblatt J, Raad II. 2017. Drug eluting antimicrobial vascular catheters: progress and promise. Adv Drug Deliv Rev 112:35–47. doi: 10.1016/j.addr.2016.07.011 [DOI] [PubMed] [Google Scholar]

Articles from Microbiology Spectrum are provided here courtesy of American Society for Microbiology (ASM)