Clinical laboratory environmental contamination: use of a fluorescence/bacterial tracer.

D A Kennedy 1, J F Stevens 1, A N Horn 1
PMCID: PMC1141736  PMID: 3145290

Abstract

A "Centrifichem" centrifugal analyser was found using a double fluorescence/bacterial tracer method to be a source of airborne and surface contamination in the laboratory. The airborne contamination was controlled by a modification incorporating a filter. Ubiquitous surface contamination generally accompanied the processing of simulated patients' sera. Double tracer samples could be used in laboratories to assess the quality of hygiene practices in the same way as quality control samples are currently used to assess the quality of analytical performance. The findings indicated a lack of understanding of basic hygiene practices in the laboratory.

Full text

PDF
1229

Selected References

These references are in PubMed. This may not be the complete list of references from this article.

  1. Beaumont L. R. The detection of blood on nonporous environmental surfaces: an approach for assessing factors contributing to the risk of occupational exposure to blood in the autopsy suite. Infect Control. 1987 Oct;8(10):424–426. doi: 10.1017/s0195941700066601. [DOI] [PubMed] [Google Scholar]
  2. Darlow H. M., Simmons D. J., Roe F. J. Hazards from experimental skin painting of carcinogens. A study based on use of a spore model. Arch Environ Health. 1969 Jun;18(6):883–893. doi: 10.1080/00039896.1969.10665510. [DOI] [PubMed] [Google Scholar]
  3. Grist N. R., Emslie J. A. Infections in British clinical laboratories, 1984-5. J Clin Pathol. 1987 Aug;40(8):826–829. doi: 10.1136/jcp.40.8.826. [DOI] [PMC free article] [PubMed] [Google Scholar]
  4. Holton J., Prince M. Blood contamination during venepuncture and laboratory manipulations of specimen tubes. J Hosp Infect. 1986 Sep;8(2):178–183. doi: 10.1016/0195-6701(86)90044-7. [DOI] [PubMed] [Google Scholar]
  5. Kennedy D. A., Stevens J. F. Tracing of laboratory contamination: quality control approach. Lancet. 1988 Feb 27;1(8583):471–472. doi: 10.1016/s0140-6736(88)91262-7. [DOI] [PubMed] [Google Scholar]
  6. Lauer J. L., VanDrunen N. A., Washburn J. W., Balfour H. H., Jr Transmission of hepatitis B virus in clinical laboratory areas. J Infect Dis. 1979 Oct;140(4):513–516. doi: 10.1093/infdis/140.4.513. [DOI] [PubMed] [Google Scholar]
  7. May K. R. Multistage liquid impinger. Bacteriol Rev. 1966 Sep;30(3):559–570. doi: 10.1128/br.30.3.559-570.1966. [DOI] [PMC free article] [PubMed] [Google Scholar]
  8. Piazza M., Guadagnino V., Picciotto L., Borgia G., Nappa S. Contamination by hepatitis B surface antigen in dental surgeries. Br Med J (Clin Res Ed) 1987 Aug 22;295(6596):473–474. doi: 10.1136/bmj.295.6596.473. [DOI] [PMC free article] [PubMed] [Google Scholar]

Articles from Journal of Clinical Pathology are provided here courtesy of BMJ Publishing Group

RESOURCES