Skip to main content
PLOS Global Public Health logoLink to PLOS Global Public Health
. 2024 Oct 3;4(10):e0003760. doi: 10.1371/journal.pgph.0003760

Performance and usability of cardiometabolic point of care devices in Nepal: A prospective, quantitative, accuracy study

Marina Giachino 1,*,#, Beatrice Vetter 2,#, Sigiriya Aebischer Perone 1, Jorge César Correia 3, Berra Erkosar 2 , Olivia Heller 1, Vijay Kumar Khanal 4, Bruno Lab 1, Zoltan Pataky 3, Sagar Poudel 5, Mamit Rai 6, Sanjib Kumar Sharma 4
Editor: Julia Robinson7
PMCID: PMC11449279  PMID: 39361597

Abstract

Non-communicable diseases (NCDs), such as cardiovascular disease and diabetes, represent a serious global health concern. There is an urgent need for prompt diagnosis and effective monitoring at point of care, especially in low- and middle-income countries. Here we present the results of a study assessing the quantitative accuracy of two devices that may fit the target product profile for a cardiometabolic point-of-care device. This prospective, quantitative, accuracy study (NCT05257564) was conducted between March to May 2022, investigating the performance of the JanaCare Aina Blood Monitoring System (JCAina) and the Tascom SimplexTAS 101 device (TAS101) compared with local standard laboratory methods in rural Nepal. Using fingerstick capillary blood, cardiometabolic parameters were analysed using both devices. The quantitative accuracy was compared against a local laboratory reference assay. System usability was also assessed. For JCAina, the mean absolute biases (Bland-Altman analysis) for glucose, HbA1c and total cholesterol tests were -3.87 mg/dL (95% CI: -7.52–-0.22), 1.34% (95% CI: 1.21–1.47), and -9.52 mg/dL (95% CI: -11.9–-7.2), respectively, corresponding to mean percentage biases of 2.0%, 18.5%, and -6.4%. These indicate clinically small (<10% biases) differences from laboratory results for glucose and cholesterol, and a moderate (10–20%) positive bias for HbA1c. For TAS101, the mean absolute biases for glucose, HbA1c, total cholesterol and creatinine tests were 18.7 mg/dL (95% CI: 15.8–21.5), -0.2% (95% CI: -0.26–-0.14), 29.8 mg/dL (95% CI: 27.0–32.6), and -0.02 mg/dL (95% CI: -0.05–0.01), respectively, corresponding to mean percentage biases of 12.1%, -2.6%, 15.8%, and -4.5%. These indicate clinically small differences for HbA1c and creatinine, and moderate positive biases for glucose and cholesterol. Both systems exhibited usability challenges. The JCAina and TAS101 point-of-care cardiometabolic devices were shown to have promising accuracy in environmental conditions such as in Nepal, though improvements are still needed for some parameters and for ease of use.

Trial registration: NCT05257564 (ClinicalTrials.gov).

Introduction

Non-communicable diseases (NCDs), such as cardiovascular disease (CVD) and diabetes mellitus (DM), represent a serious global health concern, causing an estimated 41 million deaths every year, equivalent to 74% of all disease-related deaths worldwide [1]. Of these deaths, the majority (32 million; 77%) occur in low- and middle-income countries (LMICs), and half are associated with either CVD (17.9 million; 44%) or DM (2.0 million, including those with diabetic kidney disease; 5%) [1]. Screening, timely-diagnosis and treatment form the basis of an effective response to NCDs [1].

To facilitate diagnosis, management and follow-up of patients in remote and rural areas of LMICs, there is an urgent need to ensure that primary healthcare facilities are capable of prompt diagnosing and effective monitoring NCDs at the point of care.

To address this need, FIND, compiled a landscape of point-of-care devices capable of testing basic cardiometabolic clinical chemistry parameters, potentially suitable for supporting the diagnosis and monitoring of cardiometabolic diseases [2]. These devices measure parameters recommended by the World Health Organization for the detection of cardiometabolic risk, including lipids/lipoproteins, glucose, glycated haemoglobin (HbA1c) and serum creatinine (associated with CVD, DM, and kidney disease) [24]. Measuring a combination of parameters on a single point-of-care device has the potential to facilitate a more comprehensive view of a patient’s condition by reducing the risk to forgo necessary tests due to the complexity of managing multiple point-of-care devices and supplies [5, 6].

In consultation with clinical and laboratory experts, FIND developed a target product profile (TPP) for cardiometabolic point-of-care devices intended for use in primary healthcare centres in LMICs, defining both minimal and optimal requirements (41 in total) [7]. Besides measuring the key cardiometabolic parameters, devices should be suitable for use in primary health care in low resource settings, capable of operating at environmental temperatures of ≥35°C, be suitable for use by minimally trained healthcare workers, and involve the use of self-contained cartridges or strips without the need for additional reagents [7]. Using the TPP requirements to assess devices in the current landscape [2], two devices were identified that met several of these criteria (and particularly TPP characteristic number 1–5, 10, 12, 23, 24 and 26, describing requirements of the intended use, target setting and operators, device design, environmental stability, training time, test menu, test cartridges and need for additional consumable [7]): the JanaCare Aina Blood Monitoring System (JCAina) [8, 9] and the Tascom SimplexTAS 101 device (TAS101) [10, 11]. The JCAina device measures glucose, HbA1c, and total cholesterol (creatinine testing is in development) by way of dry test strips using fingerstick capillary whole blood [2, 8, 9]. While the glucose and total cholesterol tests are single-step processes, the HbA1c is a multi-step process involving buffer incubation before applying to the test strip [2, 8, 9]. The JCAina device can operate at up to 40°C, something which is critical for many LMICs, and most reagents do not require refrigeration [2, 8, 9]. The TAS101 device measures glucose, HbA1c, total cholesterol and creatinine, by way of sample cartridges using fingerstick capillary whole blood [2, 10, 11]. The same procedure is followed for each test, and the device has an optional high temperature mode (up to 40°C) [2, 10, 11].

Available manufacturer data indicates high testing accuracy for both systems [9, 11]. However, as with many point-of-care devices, their accuracy has not been rigorously evaluated in their intended use setting and when operated by the intended users. Therefore, their accuracy in primary healthcare settings and routine use is unknown. These data are critical to enable the transition from laboratory testing to point-of-care testing and to ensure consistent patient management at primary health care facilities. Here we present the results of a study conducted to assess the quantitative accuracy of the JCAina and TAS101 devices in their intended use setting, compared with local standard-of-care laboratory testing.

Materials and methods

Study design

This was a prospective, quantitative, accuracy study investigating the performance of two clinical chemistry cardiometabolic point-of-care devices compared with local standard of care laboratory methods in Nepal. The study was conducted between 9 March 2022 and 26 May 2022 at the Dhulabari primary healthcare centre and Kakarvitta health post in rural Jhapa district (Eastern Nepal), as part of the project “Increasing awareness, detection and management of diabetes, hypertension, chronic kidney and cardiovascular diseases in Eastern Nepal” (KHDC) [12]. The study was embedded in the KHDC-Nepal program, which has the objective to improve the health of the Nepalese population in Eastern Nepal through early detection and management of chronic noncommunicable diseases. The program runs community awareness sessions and door-to-door visits to encourage members of the community to attend a screening visit at selected primary health facilities, health posts or at a pre-designated location. During this visit, people are screened for hypertension, diabetes, chronic kidney and cardiovascular disease, including (but not limited to) testing for glucose, lipids, HbA1c and creatinine, therefore the participants of this study are the intended use population for the point of care devices.

Participants

Participants who attended either of the study sites as part of the KHDC programme (for screening or follow-up appointments) [12] were eligible for inclusion in this study if they were 20 years or older, able and willing to provide informed consent, and had haemoglobin levels ≥8 g/dL. Participants were excluded from the study if they were unable to provide sufficient capillary or venous whole blood sample for all tests, had haemoglobin levels <8 g/dL, or were attending the Dhulabari primary healthcare centre or Kakarvitta health post for reasons other than the KHDC programme.

Interventions and procedures

Upon enrolment, fingerstick capillary blood was collected from all eligible participants and analysed immediately using the JCAina and TAS101 devices. Study staff were trained on the use of the JCAina and TAS101 by the manufacturers via video call with on-site supervision by the study manager and the principal investigator. Both devices were installed closely in line with the manufacturers’ instructions and reagents kept at the indicated storage temperatures (2–8°C for TAS101 and 2–30°C for JCAina total cholesterol, 1–30°C for JCAina glucose, 2–8°C for JCAina HbA1c). Quality control testing for each parameter was performed by the study’s laboratory staff, once per week using the manufacturers’ control solutions and guidelines with the target values to be reached. For reference laboratory testing, samples for plasma, whole blood and serum testing were collected immediately after fingerstick whole blood collection (for investigational device testing). Plasma in K3-EDTA collection tubes was prepared from venous blood by centrifugation within 30 minutes of sample collection to minimize glycolysis [13] and serum was prepared by centrifugation after the blood samples had been left to clot for up to 60 minutes. All samples were then stored at 2–8°C until transferred the same day to the B.P. Koirala Institute of Health Sciences laboratory and tested the same day or the next morning, using one Roche Cobas c311 analyser [14], according to the manufacturer’s instructions and performing all local quality control/calibration standard operating procedures. Before processing the samples at the reference laboratory, they were stored in cooling boxes with temperature monitoring throughout the transport and then stored at 2–8°C.

Outcomes

The primary outcome was the accuracy of quantitative measurements of glucose, HbA1c, total cholesterol and creatinine as measured by the JCAina and TAS101 multiparameter point-of-care devices compared with a local laboratory reference assay. Estimates of agreement (Passing Bablok regression), fixed bias (mean difference), and limits of agreement (Bland-Altman analysis) were assessed for each test. Secondary outcomes included measures of operational characteristics (the proportion of invalid tests per parameter for each device) and system usability measured using the System Usability Scale [15]. This scale employs a set of 10 questions using a Likert scale from 1–5 (1: ‘strongly disagree’; 5: ‘strongly agree’) and has a maximum score of 100 [15]. One point is deducted from every odd-numbered question, and for every even-numbered question the score is subtracted from five, then the total is multiplied by 2.5 to obtain the final value out of 100 [15]. System Usability Scale scores are then rated as follows: <51 is considered ‘awful’; 51–68 ‘poor’; 68 ‘okay’; >68–80.3 ‘good’; and >80.3 ‘Excellent’ [15].

Sample size and statistical analysis

Available data from the manufacturers on the performance of the JCAina and TAS101 cite Pearson correlation coefficients greater than 0.98 for all primary outcome markers compared with a reference laboratory method. Precision estimates (coefficient of variation [%CV]) ranged from 1.9–3.9% and 1.8–4.1% for the JCAina and TAS101 devices, respectively, dependent on the markers and their concentration. Using these assumptions, it was estimated that a sample size of 337 participants would be needed to detect a statistically significant difference of ±1.5% (with 80% power using a two-tailed t-test and a 5% significance level) in primary outcome markers compared with the laboratory reference assay. To improve precision and account for the possibility of lower field performances, a sample size of 400 participants was selected.

All statistical and graphical analyses were performed using R (version 4.2.0) and ggplot2 Library, as well as the SAS statistical analysis software suite [1618]. For the analysis of the primary outcome, point estimates (and 95% confidence intervals [CI]) of agreement between the JCAina/TAS101 tests and laboratory tests were calculated using the Passing Bablok regression method, where a slope of 1 (representing proportional difference) and an intercept of 0 (representing systematic difference) constitutes perfect agreement [19]. Systemic deviations of JCAina/TAS101 tests from laboratory tests (mean bias) were calculated using Bland-Altman plots in both absolute and relative units. All other data are presented using descriptive statistics.

Ethics statement

This study was conducted in accordance with the protocol (available under ClinicalTrials.gov), the ethical principles of the Declaration of Helsinki, and all other applicable laws and regulations. The Institutional Review Board in Nepal (Nepal Health Research Council) reviewed and approved the study documents (including the protocol) before the study was initiated and the Geneva Cantonal Commission of Ethics and Research were provided with the same documents and emitted a positive Advisory Opinion. All participants provided written informed consent before participation in the study (prior to haemoglobin screening).

Results

In total, 366 participants were screened, 359 met the eligibility criteria, and 356 underwent all study procedures (Fig 1). Recruitment was terminated prior to reaching the target of 400 participants due to technical problems with the laboratory equipment, which could not be addressed before the expiry date of some of the point-of-care devices’ reagents.

Fig 1. Patient and testing disposition.

Fig 1

Hb, haemoglobin; HbA1c, glycated haemoglobin.

Baseline demographics and characteristics

Participant baseline demographics and characteristics are presented in Table 1. Overall, participant characteristics were similar between study sites, with a mean age of 51.8 (±14.1) years, a mean body mass index of 25.5 (±4.4) kg/m2, and the majority of participants being female (57.9%), non-smokers (69.7%), non-drinkers of alcohol (76.7%), and just over half of participants had a diagnosis of at least one NCD (50.56%).

Table 1. Patient baseline demographics and characteristics.

Dhulabari public healthcare facility (n = 175) Kakarvitta health post (n = 181) Total (N = 356)
Sex, n (%)
    Female 100 (57.1) 106 (58.6) 206 (57.9)
    Male 75 (42.9) 75 (41.4) 150 (42.1)
Age, years, mean (±SD) 52.9 (±14.4) 50.8 (±13.8) 51.8 (±14.1)
Fasting status, n (%)
    Yes* 8 (4.6) 23 (12.7) 31 (8.7)
    No 167 (95.4) 158 (87.3) 325 (91.3)
Smoking status, n (%)
    Smoker† 34 (19.4) 50 (27.6) 84 (23.6)
    Former smoker 11 (6.3) 11 (6.1) 22 (6.2)
    Non-smoker 130 (74.3) 118 (65.2) 248 (69.7)
    NR 0 2 (1.1) 2 (0.6)
Alcohol intake, n (%)
    Once a day 6 (3.4) 11 (6.1) 17 (4.8)
    Once a week 13 (7.4) 19 (10.5) 32 (9.0)
    Once a month 12 6.9) 20 (11.0) 32 (9.0)
    None 144 (82.3) 129 (71.3) 273 (76.7)
    NR 0 2 (1.1) 2 (0.6)
BMI, kg/m2, mean (±SD) 25.2 (±4.3) 25.8 (±4.6) 25.5 (±4.4)
KHDC diagnosis, n (%)
    Hypertension 75 (42.9) 61 (33.7) 136 (38.2)
    Diabetes mellitus 66 (37.7) 43 (23.8) 109 (30.6)
    Cardiovascular disease 2 (1.1) 3 (1.7) 5 (1.4)
    Chronic kidney disease 1 (0.6) 4 (2.2) 5 (1.4)
    None 65 (37.1) 86 (47.5) 151 (42.4)
    NR 10 (5.7) 14 (7.7) 24 (6.7)
Concomitant medication, n (%)
    BP lowering medication 66 (37.7) 61 (33.7) 127 (35.7)
    Oral antiglycaemic agents 50 (28.6) 38 (21.0) 88 (24.7)
    Traditional local medication‡ 2 (1.1) 0 2 (0.6)
    Lipid lowering medication 8 (4.6) 17 (9.4) 25 (7.0)
    Other glucose lowering medication 3 (1.7) 7 (3.9) 10 (2.8)
    Other 1 (0.6) 13 (7.2) 14 (3.9)
    None 84 (48.0) 95 (52.5) 179 (50.3)

*No food or drink other than water for 8 hours; †Within the past year; ‡such as Hoodia, Gymnema, or Aloe.

BMI, body mass index; BP, blood pressure; NR, not recorded; SD, standard deviation.

Outcomes

Test accuracy

For the JCAina device, Passing Bablok analysis showed slopes of 0.87 (95% CI: 0.84–0.90), 1.10 (95% CI: 1.02–1.19), and 1.08 (95% CI: 1.02–1.15), for glucose, HbA1c and total cholesterol testing, respectively. Intercepts for glucose, HbA1c and total cholesterol were 20.4 mg/dL (95% CI: 16.1–24.4; upper measuring range limit: 600 mg/dL), 0.61% (95% CI: 0.05–1.09; upper measuring range limit: 15.0%), and -23.34 mg/dL (95% CI: -33.7–-13.9; upper measuring range limit: 400 mg/dL), respectively (where a value of 0 would indicate perfect agreement between tests and laboratory values) (Fig 2). The mean absolute biases (Bland-Altman analysis) for glucose, HbA1c and cholesterol tests were -3.87 mg/dL (95% CI: -7.52–-0.22), 1.34% (95% CI: 1.21–1.47), and -9.52 mg/dL (95% CI: -11.9–-7.2), respectively, corresponding to mean percentage biases of 2.0% (95% CI: 0.4–3.5), 18.5% (95% CI: 17.0–20.1), and -6.4% ((95% CI:-7.9 –-5.0); Fig 3). Overall, these data indicate a clinically slight (<10% bias) difference between JCAina and laboratory results for glucose and cholesterol, within or similar to the recommended thresholds for clinically important differences and acceptable analytical performance according to the 2014 Royal College of Pathologists of Australasia (RCPA) and 1992 Clinical Laboratory Improvement Amendments (CLIA) requirements for analytical quality (± 10% for glucose; ± 6–10% for cholesterol) [20, 21]. For HbA1c, a moderate (10–20%) positive bias was observed, which exceeded the RCPA/CLIA recommended thresholds (± 6–8%) [20, 21].

Fig 2. Passing Bablok analysis of JCAina tests (glucose, HbA1c, total cholesterol).

Fig 2

HbA1c, glycosylated haemoglobin; JCAina, JanaCare Aina Blood Monitoring System.

Fig 3. Bland-Altman analysis of JCAina tests (glucose, HbA1c, total cholesterol).

Fig 3

HbA1c, glycosylated haemoglobin; JCAina, JanaCare Aina Blood Monitoring System.

For the TAS101 device, Passing Bablok analysis showed slopes of 1.05 (95% CI: 1.02–1.08), 0.84 (95% CI: 0.80–0.88), 1.30 (95% CI: 1.24–1.38), and 1.15 (95% CI: 1.03–1.32), for glucose, HbA1c, cholesterol and creatinine testing, respectively. Intercepts for glucose, HbA1c, cholesterol and creatinine were 7.78 mg/dL (95% CI: 3.62–12.02; upper measuring range limit: 700 mg/dL), 0.85% (95% CI: 0.60–1.08; upper measuring range limit: 15.0%), -22.1 mg/dL (95% CI: -33.7–-11.1; upper measuring range limit: 500 mg/dL), and -0.17 mg/dL (95% CI: -0.29–-0.07; upper measuring range limit: 25.0 mg/dL), respectively (Fig 4). The mean absolute biases (Bland-Altman analysis) for glucose, HbA1c, total cholesterol and creatinine tests were 18.7 mg/dL (95% CI: 15.8–21.5), -0.2% (95% CI: -0.26–-0.14), 29.8 mg/dL (95% CI: 27.0–32.6), and -0.02 mg/dL (95% CI: -0.05–0.01), respectively, corresponding to mean percentage biases of 12.1% (95% CI: 10.8–13.5), -2.6% (95% CI: -3.4–-1.9), 15.8% (95% CI: 14.5–17.1), and -4.5% ((95% CI: -8.4–-0.7); Fig 5). Overall, these data indicate clinically small differences between TAS101 and laboratory results for HbA1c and creatinine (<10%; both within RCPA/CLIA recommended thresholds of ± 6–8% and ± 8–10%, respectively), and a moderate positive bias (10–20%, exceeding RCPA/CLIA recommended thresholds) for glucose and total cholesterol results [20, 21].

Fig 4. Passing Bablok analysis of TAS101 tests (glucose, HbA1c, total cholesterol, creatinine).

Fig 4

HbA1c, glycosylated haemoglobin; TAS101, Tascom SimplexTAS 101 device.

Fig 5. Bland-Altman analysis of TAS101 tests (glucose, HbA1c, total cholesterol, creatinine).

Fig 5

HbA1c, glycosylated haemoglobin; TAS101, Tascom SimplexTAS 101 device.

Operational characteristics

For the JCAina device, three technical problems were encountered: one blank screen after insertion of the glucose test, and two HbA1c tests where no results or error codes were displayed. In addition, an adaptor failure resulted in 48 glucose tests not being performed, and three HbA1c tests could not be performed as the control solution test was out of range.

For the TAS101 device, four initial test failures (no results obtained; no error code provided) occurred with glucose, nine with HbA1c, four with cholesterol and two with creatinine tests; repeat tests also failed in three, four, two and one cases, respectively. All error codes reported were related to the sample collection device (e.g., insufficient/excess sample volume, wrong sample/device used, missing sample).

System usability

The System Usability Scale was completed by five staff members who were trained in the use of the devices and were actively using them. Full details of their responses are presented in Table 2.

Table 2. Summary of system usability scale results for JCAina and TAS101.
Strongly disagree = 1, Strongly agree = 5
JCAina System Usability Scale User 1 User 2 User 3 User 4 User 5 Average Calc.
1. I think that I would like to use this system frequently 4 4 4 1 4 3.4 x-1
2. I found the system unnecessarily complex 2 3 4 2 4 3.0 5-x
3. I thought the system was easy to use 4 4 4 3 3 3.6 x-1
4. I think that I would need the support of a technical person to be able to use the system 3 1 3 1 3 2.2 5-x
5. I found the various functions in this system were well integrated 4 3 3 1 4 3.0 x-1
6. I thought there was too much inconsistency in this system 2 3 3 1 3 2.4 5-x
7. I would imagine that most people would learn to use this system very quickly 4 3 4 1 4 3.2 x-1
8. I found the system very cumbersome to use 2 2 2 3 3 2.4 5-x
9. I felt very confident using the system 5 4 4 4 4 4.2 x-1
10. I needed to learn a lot of things before I could get going with this system 4 4 4 5 4 4.2 5-x
Total System Usability Scale score 70 62.5 57.5 45.0 55.0 58.0 x2.5
System Usability Scale grade B D D F D D
System Usability Scale adjective rating Good Poor Poor Awful Poor Poor
TAS101 System Usability Scale User 1 User 2 User 3 User 4 User 5 Average Calc.
1. I think that I would like to use this system frequently 4 4 4 5 4 4.2 x-1
2. I found the system unnecessarily complex 1 2 2 1 2 1.6 5-x
3. I thought the system was easy to use 5 3 5 5 4 4.4 x-1
4. I think that I would need the support of a technical person to be able to use the system 2 2 2 1 3 2.0 5-x
5. I found the various functions in this system were well integrated 4 4 4 4 4 4.0 x-1
6. I thought there was too much inconsistency in this system 1 3 2 1 3 2.0 5-x
7. I would imagine that most people would learn to use this system very quickly 5 3 4 3 3 3.6 x-1
8. I found the system very cumbersome to use 2 2 2 3 2 2.2 5-x
9. I felt very confident using the system 5 4 5 5 4 4.6 x-1
10. I needed to learn a lot of things before I could get going with this system 5 4 4 5 4 4.4 5-x
Total System Usability Scale score 80 62.5 75 77.5 62.5 71.5 x2.5
System Usability Scale grade B D B B D B
System Usability Scale adjective rating Good Poor Good Good Poor Good

Calc., calculations; JCAina, JanaCare Aina Blood Monitoring System; TAS101, Tascom SimplexTAS 101 device.

For the JCAina device, the mean System Usability Score was 58 (‘poor’) as a result of variable, but critical ratings, primarily related to ease of use. While users felt confidence in using the system, they did not feel it was cumbersome to use, and did not feel that they needed technical support in using it (average ratings of 4.2/5, 2.4/5 and 2.2/5, respectively). There was however a strong agreement that the system required a lot of training before use (average rating 4.2/5), and three of the users felt that there was too much inconsistency in the system (individual ratings of 3/5).

For the TAS101 device, the mean System Usability Score was 71.5 (‘good’), with three out of the five users reporting ‘good’ ratings overall. The ratings profile was similar to the JCAina device, but with better ratings overall. Users felt confident in using the system and thought it was easy to use (average ratings of 4.6/5 and 4.4/5, respectively). There was again a strong agreement that the system required a lot of training before use (average rating 4.4/5), and the two users who rated the system as ‘poor’ felt that there was too much inconsistency in the system (individual ratings of 3/5).

Discussion

In this prospective, quantitative, accuracy study, two multiparameter cardiometabolic point-of-care diagnostic devices, which closely matched the TPP requirements (compared with other devices in the landscape) [2, 7], were compared with local standard-of-care laboratory tests in a real-world, primary healthcare care setting in Nepal. The main aim of the study was to understand the quantitative differences that clinicians may expect when implementing point-of-care testing in primary healthcare facilities, compared with central laboratory testing available in their setting. Overall, both the JCAina and TAS101 devices showed promising accuracy, with the majority of parameters measured with low bias (<10% difference versus laboratory results), within or similar to the recommended thresholds for clinically important differences and acceptable analytical performance according to RCPA/CLIA guidelines [20, 21]. This indicates they could be suitable for point-of-care testing in a primary healthcare setting. However, some parameters did show positive bias (>10% difference versus laboratory results) that exceeded RCPA/CLIA thresholds, such as HbA1c with JCAina and glucose and total cholesterol with TAS101. Measuring HbA1c, for example, has clinical implication in the management and prevention of diabetes complication. False low measurements can erroneously imply that diabetes is under control, while false high measurements can justify stricter control measures that would increase the risk of hypoglycemia and its aftereffects. As such, it is critical that the level of acceptable bias is established before making decisions about point-of-care device implementation and use. For example, while we indicate that JCAina shows ‘moderate’ bias for HbA1c testing, in relative terms this is approximately two-to-three-fold higher than the estimated bias for other dedicated diabetes point-of-care analysers (Afinion [Abbott] and DCA Vantage [Siemens]) reported in a community-based study, while the TAS101 shows approximately two-fold less bias [22]. Given that the Afinion and DCA Vantage devices have been shown to meet the generally accepted performance criteria for HbA1c in a separate laboratory-based study [23], the information presented here could be useful in making decisions on the use of JCAina or TAS101 in specific practices based on the most common testing procedures required.

The International Federation of Clinical Chemistry recommends that “Device performance claims from the package insert should correlate with those obtained from verification studies that are performed by the [point-of-care testing] POCT personnel” [24]. This is critically important, considering variable field performance of point-of-care tests, [2527] the lack of manufacturer-independent evaluations and the high number of on-market cardiometabolic POC devices [28].

The reasons for discrepancies versus laboratory results found in our study are unclear. For JCAina HbA1c testing it could be due to pre-analytical handling variability, as the test requires several manual steps and sample/buffer mixing [2, 8, 9]. For the TAS101 glucose and total cholesterol discrepancies, the reasons are less apparent, though it is likely due to a combination of technical variability with the instrument, user variability, and environmental factors. In particular, the manufacturer’s instructions recommend installation of the TAS101 device away from direct sunlight or draughts with sufficient space around the machine [10], which was not always possible at one of the sites due to large windows providing sunlight access through much of the laboratory. This could have led to variations in the results dependent on the time of day and set up. We can also not exclude that the observed bias in glucose measurement is not a result of rapidly changing glucose levels in participants during the time of sample collection as the majority of participants had their samples taking in non-fasting status. However, the time between fingerstick and venous whole blood sample collection was kept to a minimum as participants immediately proceeded to venous blood collection after point-of-care fingerstick testing, so we would expect this source of potential variability to be minimal. Degradation of glucose through glycolysis in the reference sample could also be a potential source of positive bias observed in the study, however any effect was likely minimal (around 0.1 mmol/l or 1.8 mg/dL), based on findings of a previous study on glucose degradation kinetics in venous blood samples kept at room temperature for 30 min [29].

The coefficient of variation (CV) for the reference tests are another source of variability in this study. However, based on the instructions for use by the manufacturer, the Roche cobas tests for glucose, cholesterol and HbA1c used in this study had a CV (repeatability) of <2% and ≤4% for creatinine, suggesting that the effect was minimal with a CV <5% often considered excellent [30, 31].

Lastly, it is possible that difference in detection methods used by the investigational devices and the laboratory reference devices may have contributed to the observed difference. For glucose and total cholesterol, the investigation tests, as well as the reference tests were based on an enzymatic method. For HbA1c detection, JCAina used the boronate affinity method, TAS101 an enzymatic method and the reference test HbA1c detection was an immunoassay. For creatinine, the TAS101 used an enzymatic method and the reference test was based on the Jaffe method. While this may have contributed to the observed differences between the investigational tests and the reference tests, e.g., due to different risks of interfering substances [32], it highlights the challenges that implementers will face when transitioning from central laboratory to point-of-care testing as testing methods are likely to vary in real-life settings.

Both devices also had challenges in terms of usability, something identified in the TPP as a key aspect for successful implementation in primary healthcare [7]. While users mostly felt that most people would learn to use both systems quickly, they judged the learning required to be substantial prior to first system use. These were not just limited to ease of use, device complexity or technical errors and or issues, but also acceptance by the users. This is something that manufacturers would be advised to take into consideration when optimising devices to ensure their long-term use. In addition, while all reagents were temperature monitored in this study, it should be noted that the JCAina device required some reagents to be stored at 2–8°C [2, 8, 9], as well as for all reagents of TAS101 [2, 10, 11], which adds another layer of complexity (i.e., temperature/refrigeration monitoring) and an additional source of error (i.e., reagent quality). It is important to consider the potential influence of factors such as available infrastructure, complexity versus ease of use, and storage, as these would not be apparent in a controlled clinical trial or laboratory environment. This highlights the value of (i) conducting studies in a real-world setting, and (ii) having a comprehensive TPP available during development to identify key areas (beyond testing accuracy) that are important for making devices fit for purpose in LMICs.

In conclusion, the quantitative accuracy of the JCAina and TAS101 point-of-care cardiometabolic devices in intended use settings was shown to be promising, though investments are still needed to improve key aspects such as consistent performance across the whole portfolio of parameters, reagent storage temperature, as well as ease of use.

Supporting information

S1 Checklist. STARD checklist for accuracy studies.

(PDF)

pgph.0003760.s001.pdf (195.1KB, pdf)
S1 Dataset. Final study dataset.

(XLSX)

pgph.0003760.s002.xlsx (240.7KB, xlsx)

Acknowledgments

The authors would like to acknowledge the contributions of all participating staff at the Dhulabari primary healthcare centre and Kakarvitta health post in rural Jhapa district (Eastern Nepal), as well as all study participants. Editorial assistance in the preparation of the manuscript was provided by Stuart Wakelin, PhD.

Data Availability

The full and final dataset of the clinical study is made available under the Supporting Information files.

Funding Statement

This study was funded by FIND (www.finddx.org), through a grant from the International Committee of the Red Cross and the German Federal Ministry of Education and Research (NCT05257564; grant numbers: KFW-TBBU02; ICR-NCDS01). The study or authors were not funded by any commercial companies, no authors received a salary or other funding from commercial companies, and the funders had no role in the study design, data collection and analysis, decision to publish, or the preparation of the manuscript.

References

  • 1.World Health Organization (WHO). Noncommunicable diseases. Available from https://www.who.int/en/news-room/fact-sheets/detail/noncommunicable-diseases (accessed 01 Feb 2023). 2022. [Google Scholar]
  • 2.FIND. Landscape of point-of-care devices for testing of cardiometabolic diseases. Available from https://www.finddx.org/wp-content/uploads/2023/02/20200228_rep_landscape_poc_cardiometabolic_diseases_FV_EN.pdf (accessed 26 March 2023). 2020. [Google Scholar]
  • 3.World Health Organization (WHO). Package of essential noncommunicable (PEN) disease interventions for primary health care in low resource settings. Available from https://apps.who.int/iris/rest/bitstreams/1301957/retrieve (accessed 02 February 2023). 2020. [Google Scholar]
  • 4.World Health Organization (WHO). The selection and use of essential in vitro diagnostics. Available from https://apps.who.int/iris/rest/bitstreams/1329518/retrieve (accessed 02 February 2023). 2021. [Google Scholar]
  • 5.Albitres-Flores L, Perez-Leon S, Bernabe-Ortiz A, Tenorio-Mucha J, Cardenas MK, Vetter B, et al. Co-creation process of an intervention to implement a multiparameter point-of-care testing device in a primary healthcare setting for non-communicable diseases in Peru. BMC Health Services Research. 2024;24(1):401. doi: 10.1186/s12913-024-10809-3 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Tenorio-Mucha J, Busta-Flores P, Lazo-Porras M, Vetter B, Safary E, Moran AE, et al. Facilitators and barriers of the implementation of point-of-care devices for cardiometabolic diseases: a scoping review. BMC Health Serv Res. 2023;23(1):412. Epub 20230428. doi: 10.1186/s12913-023-09419-2 ; PubMed Central PMCID: PMC10144879. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.Vetter B, Beran D, Boulle P, Chua A, de la Tour R, Hattingh L, et al. Development of a target product profile for a point-of-care cardiometabolic device. BMC Cardiovasc Disord. 2021;21(1):486. doi: 10.1186/s12872-021-02298-7 ; PubMed Central PMCID: PMC8501932. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.JanaCare. Aina. Available from https://www.janacare.com/ (accessed 28 January 2023). 2023. [Google Scholar]
  • 9.JanaCare. Aina Blood Monitoring System: Analytical performance summary. Available from https://www.janacare.com/assets/pdf/LBL-AINA-13_C.pdf (accessed 28 January 2023). 2023. [Google Scholar]
  • 10.Tascom. SimplexTAS 101. Available from https://www.tascom.org/ (accessed 26 March 2023). 2023. [Google Scholar]
  • 11.Tascom. SimplexTAS 101. Catalogue information. Available from http://www.apteq.ch/saas/CustomUpload/374O357O340O370O356O369O350O320O367O371O356O368O320O326O/SimplexTAS_catalogue.pdf (accessed 26 March 2023). 2018. [Google Scholar]
  • 12.Hypertension Kidney, Diabetes & Cardiovascular Diseases. About KHDC Nepal. Available from https://www.khdcnepal.org/ (accessed 27 March 2023). 2021. [Google Scholar]
  • 13.Sacks DB, Arnold M, Bakris GL, Bruns DE, Horvath AR, Kirkman MS, et al. Executive Summary: Guidelines and Recommendations for Laboratory Analysis in the Diagnosis and Management of Diabetes Mellitus. Clinical Chemistry. 2011;57(6):793–8. doi: 10.1373/clinchem.2011.163634 [DOI] [PubMed] [Google Scholar]
  • 14.Roche. Cobas c311 analyser: overview and operating instructions. Available from https://diagnostics.roche.com/gb/en/products/instruments/cobas-c-311-ins-2043.html (accessed 27 March 2023). 2023. [Google Scholar]
  • 15.Brooke J. SUS: A quick and dirty usability scale. In: Jordan PW, Thomas B, Weerdmeester BA, McClelland IL, editors. Usability Evaluation in Industry. London: Taylor & Francis; 1996. p. 189–94. [Google Scholar]
  • 16.The R Foundation. The R project for statistical computing. Available from https://www.r-project.org/ (accessed 28 January 2023). 2022. [Google Scholar]
  • 17.Wickham H. ggplot2: Elegant graphics for data analysis. New York: Springer. 2016. [Google Scholar]
  • 18.SAS Institute Inc. SAS statistical analysis software. Available from https://www.sas.com/en_us/software/stat.html (accessed 27 March 2023). 2023. [Google Scholar]
  • 19.Passing H, Bablok. A new biometrical procedure for testing the equality of measurements from two different analytical methods. Application of linear regression procedures for method comparison studies in clinical chemistry, Part I. J Clin Chem Clin Biochem. 1983;21(11):709–20. doi: 10.1515/cclm.1983.21.11.709 . [DOI] [PubMed] [Google Scholar]
  • 20.Westgard J. Clinical Laboratory Improvement Amendments (CLIA) requirements for analytical quality. Available from https://www.westgard.com/clia.htm (accessed 27 March 2023). 1992. [Google Scholar]
  • 21.Westgard J. Royal College of Pathologists of Australasia (RCPA) analytical quality requirements. Available from https://www.westgard.com/rcpa-biochemistry.htm (accessed 27 March 2023). 2014. [Google Scholar]
  • 22.Saxton AT, Miranda JJ, Ortiz EJ, Pan W. Assessment of Two Diabetes Point-of-care Analyzers Measuring Hemoglobin A1c in the Peruvian Amazon. Ann Glob Health. 2018;84(4):618–24. doi: 10.9204/aogh.2368 ; PubMed Central PMCID: PMC6748252. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23.Lenters-Westra E, Slingerland RJ. Three of 7 hemoglobin A1c point-of-care instruments do not meet generally accepted analytical performance criteria. Clin Chem. 2014;60(8):1062–72. doi: 10.1373/clinchem.2014.224311 . [DOI] [PubMed] [Google Scholar]
  • 24.Khan AI, Pratumvinit B, Jacobs E, Kost GJ, Kary H, Balla J, et al. Point-of-care testing performed by healthcare professionals outside the hospital setting: consensus based recommendations from the IFCC Committee on Point-of-Care Testing (IFCC C-POCT). Clin Chem Lab Med. 2023;61(9):1572–9. doi: 10.1515/cclm-2023-0502 [DOI] [PubMed] [Google Scholar]
  • 25.Hirst JA, McLellan JH, Price CP, English E, Feakins BG, Stevens RJ, et al. Performance of point-of-care HbA1c test devices: implications for use in clinical practice—a systematic review and meta-analysis. Clin Chem Lab Med. 2017;55(2):167–80. doi: 10.1515/cclm-2016-0303 . [DOI] [PubMed] [Google Scholar]
  • 26.Nataatmadja M, Fung AWS, Jacobson B, Ferera J, Bernstein E, Komenda P, et al. Performance of StatSensor Point-of-Care Device for Measuring Creatinine in Patients With Chronic Kidney Disease and Postkidney Transplantation. Can J Kidney Health Dis. 2020;7:2054358120970716. Epub 20201112. doi: 10.1177/2054358120970716 ; PubMed Central PMCID: PMC7672720. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 27.Park PH, Chege P, Hagedorn IC, Kwena A, Bloomfield GS, Pastakia SD. Assessing the accuracy of a point-of-care analyzer for hyperlipidaemia in western Kenya. Trop Med Int Health. 2016;21(3):437–44. Epub 20160107. doi: 10.1111/tmi.12653 . [DOI] [PubMed] [Google Scholar]
  • 28.FIND. Landscape of HbA1c Point-of-care testing devices. Available from https://www.finddx.org/wp-content/uploads/2022/12/20211101_lds_hba1c_poc_FV_EN.pdf (accessed 4 July 2024). 2021. [Google Scholar]
  • 29.Stapleton M, Daly N, O’Kelly R, Turner MJ. Time and temperature affect glycolysis in blood samples regardless of fluoride-based preservatives: a potential underestimation of diabetes. Ann Clin Biochem. 2017;54(6):671–6. Epub 20170113. doi: 10.1177/0004563216682978 . [DOI] [PubMed] [Google Scholar]
  • 30.Chapter Carstensen B. 9 Repeatability, Reproducibility and Coefficient of Variation. Comparing Clinical Measurement Methods: A Practical Guide: John Wiley & Sons; 2010. [Google Scholar]
  • 31.Clinical and laboratory standards institute (CLSI). EP05-A3 Evaluation of Precision of Quantitative Measurement Procedures, 3rd Edition. 2014. [Google Scholar]
  • 32.Schmidt RL, Straseski JA, Raphael KL, Adams AH, Lehman CM. A Risk Assessment of the Jaffe vs Enzymatic Method for Creatinine Measurement in an Outpatient Population. PLoS One. 2015;10(11):e0143205. doi: 10.1371/journal.pone.0143205 ; PubMed Central PMCID: PMC4657986. [DOI] [PMC free article] [PubMed] [Google Scholar]
PLOS Glob Public Health. doi: 10.1371/journal.pgph.0003760.r001

Decision Letter 0

Hugh Cowley

6 Mar 2024

PGPH-D-24-00096

Performance and usability of cardiometabolic point of care devices in Nepal: a prospective, quantitative, accuracy study

PLOS Global Public Health

Dear Dr. Giachino,

Thank you for submitting your manuscript to PLOS Global Public Health. After careful consideration, we feel that it has merit but does not fully meet PLOS Global Public Health’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses the points raised during the review process.

Please note that we have only been able to secure a single reviewer to assess your manuscript. We are issuing a decision on your manuscript at this point to prevent further delays in the evaluation of your manuscript. Please be aware that the editor who handles your revised manuscript might find it necessary to invite additional reviewers to assess this work once the revised manuscript is submitted. However, we will aim to proceed on the basis of this single review if possible.

The reviewer has queried a range of aspects of the study design; please ensure you address each of the reviewer's comments when revising your manuscript.

Please submit your revised manuscript by Apr 19 2024 11:59PM. If you will need more time than this to complete your revisions, please reply to this message or contact the journal office at globalpubhealth@plos.org. When you're ready to submit your revision, log on to https://www.editorialmanager.com/pgph/ and select the 'Submissions Needing Revision' folder to locate your manuscript file.

Please include the following items when submitting your revised manuscript:

  • A rebuttal letter that responds to each point raised by the editor and reviewer(s). You should upload this letter as a separate file labeled 'Response to Reviewers'.

  • A marked-up copy of your manuscript that highlights changes made to the original version. You should upload this as a separate file labeled 'Revised Manuscript with Track Changes'.

  • An unmarked version of your revised paper without tracked changes. You should upload this as a separate file labeled 'Manuscript'.

Guidelines for resubmitting your figure files are available below the reviewer comments at the end of this letter.

We look forward to receiving your revised manuscript.

Kind regards,

Hugh Cowley

Staff Editor

PLOS Global Public Health

Journal Requirements:

1. Please send a completed 'Competing Interests' statement, including any COIs declared by your co-authors. If you have no competing interests to declare, please state "The authors have declared that no competing interests exist". Otherwise please declare all competing interests beginning with the statement "I have read the journal's policy and the authors of this manuscript have the following competing interests:"

2. We have noticed that you have uploaded Supporting Information files, but you have not included a list of legends. Please add a full list of legends for your Supporting Information files after the references list.

3. In the online submission form, you indicated that "The data will be made available by the corresponding author upon reasonable request". All PLOS journals now require all data underlying the findings described in their manuscript to be freely available to other researchers, either 1. In a public repository, 2. Within the manuscript itself, or 3. Uploaded as supplementary information.

This policy applies to all data except where public deposition would breach compliance with the protocol approved by your research ethics board. If your data cannot be made publicly available for ethical or legal reasons (e.g., public availability would compromise patient privacy), please explain your reasons by return email and your exemption request will be escalated to the editor for approval. Your exemption request will be handled independently and will not hold up the peer review process, but will need to be resolved should your manuscript be accepted for publication. One of the Editorial team will then be in touch if there are any issues.

Additional Editor Comments (if provided):

[Note: HTML markup is below. Please do not edit.]

Reviewers' comments:

Reviewer's Responses to Questions

Comments to the Author

1. Does this manuscript meet PLOS Global Public Health’s publication criteria? Is the manuscript technically sound, and do the data support the conclusions? The manuscript must describe methodologically and ethically rigorous research with conclusions that are appropriately drawn based on the data presented.

Reviewer #1: Yes

**********

2. Has the statistical analysis been performed appropriately and rigorously?

Reviewer #1: Yes

**********

3. Have the authors made all data underlying the findings in their manuscript fully available (please refer to the Data Availability Statement at the start of the manuscript PDF file)?

The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception. The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified.

Reviewer #1: No

**********

4. Is the manuscript presented in an intelligible fashion and written in standard English?

PLOS Global Public Health does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here.

Reviewer #1: Yes

**********

5. Review Comments to the Author

Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters)

Reviewer #1: Interesting manuscript addressing an important global need.

To improve scientific clarity, please address below comments

1. all figures are of a poor quality

2. has the protocol been published?

3. has the authors considered the relevance of learning curve?

4. what are the essential characteristics for the point of care devices as presented in TPP. Author mentioned 41 characterstics in total, are there any key characteristics? Would be good to know how these two devices have been chosen

5. how was the quality control carried out? By whom, using what standard to ensure the device is operating safety and accurately?

6. was there any study in terms of how methods and timing treating and analysing samples result in the discrepancies (biases) as observed in this study?

7. what are the consequences of the bias? The author mentioned clinically significant, insignificant, but would be good to know the details. for instance, what would a high HbA1C reading lead to?

8. are the study population the intended patient groups? The inclusion, exclusion criteria are not clear. More details, such as the protocol, would help, laying out the essential PICO question the patients, the user, the comparator, the outcomes for the study.

**********

6. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files.

Do you want your identity to be public for this peer review? If you choose “no”, your identity will remain anonymous but your review may still be made public.

For information about this choice, including consent withdrawal, please see our Privacy Policy.

Reviewer #1: No

**********

[NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files.]

While revising your submission, please upload your figure files to the Preflight Analysis and Conversion Engine (PACE) digital diagnostic tool, https://pacev2.apexcovantage.com/. PACE helps ensure that figures meet PLOS requirements. To use PACE, you must first register as a user. Registration is free. Then, login and navigate to the UPLOAD tab, where you will find detailed instructions on how to use the tool. If you encounter any issues or have any questions when using PACE, please email PLOS at figures@plos.org. Please note that Supporting Information files do not need this step.

PLOS Glob Public Health. doi: 10.1371/journal.pgph.0003760.r003

Decision Letter 1

Johanna Pruller

10 Jun 2024

PGPH-D-24-00096R1

Performance and usability of cardiometabolic point of care devices in Nepal: a prospective, quantitative, accuracy study

PLOS Global Public Health

Dear Dr. Giachino,

Thank you for submitting your manuscript to PLOS Global Public Health. After careful consideration, we feel that it has merit but does not fully meet PLOS Global Public Health’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses the points raised during the review process.

==============================

The manuscript has been evaluated by one reviewer, and their comments are available below. The reviewer has raised additional questions about the methodology and study design. Could you please revise the manuscript to carefully address the concerns raised?

==============================

Please submit your revised manuscript by Jul 22 2024 11:59PM. If you will need more time than this to complete your revisions, please reply to this message or contact the journal office at globalpubhealth@plos.org. When you're ready to submit your revision, log on to https://www.editorialmanager.com/pgph/ and select the 'Submissions Needing Revision' folder to locate your manuscript file.

Please include the following items when submitting your revised manuscript:

  • A rebuttal letter that responds to each point raised by the editor and reviewer(s). You should upload this letter as a separate file labeled 'Response to Reviewers'.

  • A marked-up copy of your manuscript that highlights changes made to the original version. You should upload this as a separate file labeled 'Revised Manuscript with Track Changes'.

  • An unmarked version of your revised paper without tracked changes. You should upload this as a separate file labeled 'Manuscript'.

Guidelines for resubmitting your figure files are available below the reviewer comments at the end of this letter.

We look forward to receiving your revised manuscript.

Kind regards,

Johanna Pruller, Ph.D.

PLOS Staff Editor

PLOS Global Public Health

Journal Requirements:

Please review your reference list to ensure that it is complete and correct. If you have cited papers that have been retracted, please include the rationale for doing so in the manuscript text, or remove these references and replace them with relevant current references. Any changes to the reference list should be mentioned in the rebuttal letter that accompanies your revised manuscript. If you need to cite a retracted article, indicate the article’s retracted status in the References list and also include a citation and full reference for the retraction notice.

Additional Editor Comments (if provided):

[Note: HTML markup is below. Please do not edit.]

Reviewers' comments:

Reviewer's Responses to Questions

Comments to the Author

1. If the authors have adequately addressed your comments raised in a previous round of review and you feel that this manuscript is now acceptable for publication, you may indicate that here to bypass the “Comments to the Author” section, enter your conflict of interest statement in the “Confidential to Editor” section, and submit your "Accept" recommendation.

Reviewer #2: (No Response)

**********

2. Does this manuscript meet PLOS Global Public Health’s publication criteria? Is the manuscript technically sound, and do the data support the conclusions? The manuscript must describe methodologically and ethically rigorous research with conclusions that are appropriately drawn based on the data presented.

Reviewer #2: Yes

**********

3. Has the statistical analysis been performed appropriately and rigorously?

Reviewer #2: Yes

**********

4. Have the authors made all data underlying the findings in their manuscript fully available (please refer to the Data Availability Statement at the start of the manuscript PDF file)?

The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception. The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified.

Reviewer #2: Yes

**********

5. Is the manuscript presented in an intelligible fashion and written in standard English?

PLOS Global Public Health does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here.

Reviewer #2: Yes

**********

6. Review Comments to the Author

Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters)

Reviewer #2: With interest I read the paper by Giochino et al. describing results of a cross-sectional accuracy study of prospectively recruited participants examining the performance of two different point-of-care devices assessing glucose, total cholesterol and glycated hemoglobin. While the concept is not novel, this manuscript provides information on two "new" devices in a distinct patient population from Nepal.

Overall, the manuscript is well written, the methodology seems robust, and the statistical analyses are appropriately done and presented. Results are comprehensive and detailed.

I have some minor additional comments to the authors to improve their manuscript:

1. Please detail the time elapsed between venous blood collection and laboratory assessment of the markers of interest. This is especially important for glucose, as the longer the wait is, the longer chances of glucose decay is. Also, were all analyses performed in one core lab? several sites? If so what was the coefficient of variance of the machines used for analyses in each site?

2. To support agreement for continuous measurements between different devices/methods, please add an analysis for interclass correlation.

3. Were repeat measurements for intraclass correlation performed? Reliability is another important aspect not highlighted in the current manuscript.

4. Please discuss and compared the results to other existing devices as this field is saturated in widely available and relatively cheap devices.

5. Discussion should also focus on the the gap which these devices are meant to bridge. The ability to perform point-of-care measurements of glucose and cholesterol is not novel. Are the devices in focus cheaper than alternatives? Are they easier to use?

**********

7. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files.

Do you want your identity to be public for this peer review? If you choose “no”, your identity will remain anonymous but your review may still be made public.

For information about this choice, including consent withdrawal, please see our Privacy Policy.

Reviewer #2: No

**********

[NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files.]

While revising your submission, please upload your figure files to the Preflight Analysis and Conversion Engine (PACE) digital diagnostic tool, https://pacev2.apexcovantage.com/. PACE helps ensure that figures meet PLOS requirements. To use PACE, you must first register as a user. Registration is free. Then, login and navigate to the UPLOAD tab, where you will find detailed instructions on how to use the tool. If you encounter any issues or have any questions when using PACE, please email PLOS at figures@plos.org. Please note that Supporting Information files do not need this step.

PLOS Glob Public Health. doi: 10.1371/journal.pgph.0003760.r005

Decision Letter 2

Julia Robinson

4 Sep 2024

Performance and usability of cardiometabolic point of care devices in Nepal: a prospective, quantitative, accuracy study

PGPH-D-24-00096R2

Dear Ms Giachino,

We are pleased to inform you that your manuscript 'Performance and usability of cardiometabolic point of care devices in Nepal: a prospective, quantitative, accuracy study' has been provisionally accepted for publication in PLOS Global Public Health.

Before your manuscript can be formally accepted you will need to complete some formatting changes, which you will receive in a follow up email. A member of our team will be in touch with a set of requests.

Please note that your manuscript will not be scheduled for publication until you have made the required changes, so a swift response is appreciated.

IMPORTANT: The editorial review process is now complete. PLOS will only permit corrections to spelling, formatting or significant scientific errors from this point onwards. Requests for major changes, or any which affect the scientific understanding of your work, will cause delays to the publication date of your manuscript.

If your institution or institutions have a press office, please notify them about your upcoming paper to help maximize its impact. If they'll be preparing press materials, please inform our press team as soon as possible -- no later than 48 hours after receiving the formal acceptance. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information, please contact globalpubhealth@plos.org.

Thank you again for supporting Open Access publishing; we are looking forward to publishing your work in PLOS Global Public Health.

Best regards,

Julia Robinson

Executive Editor

PLOS Global Public Health

***********************************************************

Reviewer Comments (if any, and for reference):

Reviewer's Responses to Questions

Comments to the Author

1. If the authors have adequately addressed your comments raised in a previous round of review and you feel that this manuscript is now acceptable for publication, you may indicate that here to bypass the “Comments to the Author” section, enter your conflict of interest statement in the “Confidential to Editor” section, and submit your "Accept" recommendation.

Reviewer #1: All comments have been addressed

**********

2. Does this manuscript meet PLOS Global Public Health’s publication criteria? Is the manuscript technically sound, and do the data support the conclusions? The manuscript must describe methodologically and ethically rigorous research with conclusions that are appropriately drawn based on the data presented.

Reviewer #1: Yes

**********

3. Has the statistical analysis been performed appropriately and rigorously?

Reviewer #1: Yes

**********

4. Have the authors made all data underlying the findings in their manuscript fully available (please refer to the Data Availability Statement at the start of the manuscript PDF file)?

The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception. The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified.

Reviewer #1: (No Response)

**********

5. Is the manuscript presented in an intelligible fashion and written in standard English?

PLOS Global Public Health does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here.

Reviewer #1: Yes

**********

6. Review Comments to the Author

Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters)

Reviewer #1: please update figure 1 which cannot be viewed.

**********

7. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files.

Do you want your identity to be public for this peer review? If you choose “no”, your identity will remain anonymous but your review may still be made public.

For information about this choice, including consent withdrawal, please see our Privacy Policy.

Reviewer #1: No

**********

Associated Data

    This section collects any data citations, data availability statements, or supplementary materials included in this article.

    Supplementary Materials

    S1 Checklist. STARD checklist for accuracy studies.

    (PDF)

    pgph.0003760.s001.pdf (195.1KB, pdf)
    S1 Dataset. Final study dataset.

    (XLSX)

    pgph.0003760.s002.xlsx (240.7KB, xlsx)
    Attachment

    Submitted filename: Response to Reviewers.docx

    pgph.0003760.s003.docx (41.3KB, docx)
    Attachment

    Submitted filename: Response to Reviewers.docx

    pgph.0003760.s004.docx (47.5KB, docx)

    Data Availability Statement

    The full and final dataset of the clinical study is made available under the Supporting Information files.


    Articles from PLOS Global Public Health are provided here courtesy of PLOS

    RESOURCES