Abstract
A microbiological survey of the effectiveness of class II (laminar flow) 'safety' cabinets found in the UK in the last six years is recorded. Only two of the nine units tested approached the containment of aerosols achieved by a good class I (exhaust protective) cabinet. The others were potentially hazardous if used with pathogenic material. The National Sanitation Foundation and the British Standards Institute have now laid down adequate specifications based on biological containment, and hopefully the conforming cabinets will be much better; even so, the purchase of a cabinet must be undertaken with care, and the cabinet requires frequent monitoring during use and after servicing.
Full text
PDF








Images in this article
Selected References
These references are in PubMed. This may not be the complete list of references from this article.
- Clark R. P., Mullan B. J. Airflows in and around linear downflow 'safety' cabinets. J Appl Bacteriol. 1978 Aug;45(1):131–135. doi: 10.1111/j.1365-2672.1978.tb04206.x. [DOI] [PubMed] [Google Scholar]
- Coriell L. L., McGarrity G. J. Biohazard hood to prevent infection during microbiological procedures. Appl Microbiol. 1968 Dec;16(12):1895–1900. doi: 10.1128/am.16.12.1895-1900.1968. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Foord N., Lidwell O. M. Airborne infection in a fully air-conditioned hospital. I. Air transfer between rooms. J Hyg (Lond) 1975 Aug;75(1):15–30. doi: 10.1017/s0022172400047033. [DOI] [PMC free article] [PubMed] [Google Scholar]
- McDade J. J., Sabel F. L., Akers R. L., Walker R. J. Microbiological studies on the performance of a laminar airflow biological cabinet. Appl Microbiol. 1968 Jul;16(7):1086–1092. doi: 10.1128/am.16.7.1086-1092.1968. [DOI] [PMC free article] [PubMed] [Google Scholar]
- McGarrity G. J., Coriell L. L. Modified laminar flow biological safety cabinet. Appl Microbiol. 1974 Oct;28(4):647–650. doi: 10.1128/am.28.4.647-650.1974. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Newsom S. W. A test system for the biological safety cabinet. J Clin Pathol. 1974 Jul;27(7):585–589. doi: 10.1136/jcp.27.7.585. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Rake B. W. Influence of crossdrafts on the performance of a biological safety cabinet. Appl Environ Microbiol. 1978 Aug;36(2):278–283. doi: 10.1128/aem.36.2.278-283.1978. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Staat R. H., Beakley J. W. Evaluation of laminar flow microbiological safety cabinets. Appl Microbiol. 1968 Oct;16(10):1478–1482. doi: 10.1128/am.16.10.1478-1482.1968. [DOI] [PMC free article] [PubMed] [Google Scholar]