Abstract
Children with type 1 diabetes and their caregivers face numerous challenges navigating the unpredictability of this complex disease. Although the burden of managing diabetes remains significant, new technology has eased some of the load and allowed children with type 1 diabetes to achieve tighter glycaemic management without fear of excess hypoglycaemia. Continuous glucose monitor use alone improves outcomes and is considered standard of care for paediatric type 1 diabetes management. Similarly, automated insulin delivery (AID) systems have proven to be safe and effective for children as young as 2 years of age. AID use improves not only blood glucose levels but also quality of life for children with type 1 diabetes and their caregivers and should be strongly considered for all youth with type 1 diabetes if available and affordable. Here, we review key data on the use of diabetes technology in the paediatric population and discuss management issues unique to children and adolescents.
Keywords: Automated insulin delivery, Continuous glucose monitoring, Diabetes technology, Medical technology, Paediatric care, Review, Type 1 diabetes
Introduction
Diabetes is one of the most common chronic health conditions seen in children, and recent studies show that the incidence of both type 1 and type 2 diabetes is increasing in childhood [1, 2]. The need for therapeutic and technological advances to improve glycaemic outcomes in the paediatric population cannot be understated, as duration of diabetes and younger age at diagnosis are linked to excess mortality and increased risk for CVD later in life [3].
Conventional therapy for children with type 1 diabetes consists of either multiple daily insulin injections (MDI) or continuous insulin infusion via an insulin pump, with the dose of insulin for meals calculated based on the glucose level prior to eating and the carbohydrate content of the food to be consumed. The management of type 1 diabetes is quite difficult in growing children as their insulin requirements change continuously, often increasing as they grow. Insulin sensitivity is impacted by growth hormone and cortisol release [4]. Hormones fluctuate by time of day and vary by age, creating unique patterns of insulin needs throughout childhood. Puberty, a time of significant insulin resistance, is particularly challenging as insulin needs rise rapidly. Consequently, frequent adjustments to insulin doses, including in between regular visits to a paediatric endocrinologist every 3 months, are necessary to achieve adequate glycaemic management in the paediatric type 1 diabetes population.
Over the past two decades, diabetes management has been revolutionised by the use of technology, including continuous glucose monitors (CGMs) and automated insulin delivery (AID) systems. Its use is rapidly increasing in children in many areas of the world and the safety and efficacy of AID is now well documented [5, 6]. CGM use alone improves metabolic outcomes, even when not combined with a pump or AID system, and AID clearly outperforms traditional open-loop insulin pumps in children with type 1 diabetes [7, 8]. In this review, we will provide an update on technology used in the paediatric diabetes population, discussing the impact of technology on both glycaemic and patient-reported outcomes and the application of technology use in the real world. We will also review barriers that remain to increased technology uptake and discuss where research in this field is headed.
Paediatric glycaemic goals
While it is well known that repeated, severe hypoglycaemia is harmful to the developing brain, the negative impact of both acute and chronic hyperglycaemia on brain development and function in childhood has also recently become an area of concern [9]. Adults with diabetes are at increased risk of dementia and cognitive dysfunction, speculated to be related to changes in brain development in individuals with childhood-onset type 1 diabetes [10]. Recently, a longitudinal study confirmed lower brain volume and cognitive scores in children with type 1 diabetes compared with their matched peers without diabetes, the degree of which was associated with hyperglycaemia and glucose variability [11].
New technology has allowed children with type 1 diabetes to safely achieve tighter glycaemic management, and the glycaemic targets for children have shifted accordingly. Nearly two decades ago, HbA1c goals as recommended by the ADA for children with type 1 diabetes were 58–69 mmol/mol (7.5–8.5%) for children <6 years of age, <64 mmol/mol (<8%) for children 6–12 years old and <58 mmol/mol (<7.5%) for adolescents [12]. In 2019, the recommendation was <58 mmol/mol (<7.5%) across all paediatric ages and today the goal of <53 mmol/mol (<7%) is recommended for most children by both the ADA and the International Society for Paediatric and Adolescent Diabetes [13, 14]. Importantly, that goal applies only to children who have access to advanced technologies and specialist diabetes care. Despite the decreasing HbA1c target over time, studies consistently show that, on average, the HbA1c remains well above this goal in children, particularly in early adolescence, even as therapies have improved [15].
Although HbA1c remains a gold standard marker of glucose management, CGM data allows for a more in-depth understanding of an individual’s metabolic excursions, giving important information on blood glucose variability as well as time spent in hypoglycaemia that is lacking with HbA1c measurement alone. CGM consensus guidelines published in 2019 established recommended goals for CGM metrics across all ages including time in range (TIR; glucose levels 3.9–10 mmol/l [70–180 mg/dl]; goal >70%) and time below range (<3.9 mmol/l [<70 mg/dl]; goal <4%) [16]. However, given the many challenges of type 1 diabetes management in childhood, it is not surprising that children are less likely to meet these glycaemic targets as compared with adults [15, 17].
Rising technology use
CGM use in paediatrics has rapidly expanded over the past 15 years as accuracy, availability and insurance coverage increased. Centres across the USA participating in the T1D Exchange Registry reported more than a tenfold increase in CGM use in the 2016–2018 cohort compared with the 2010–2012 cohort [15]. Data from the SWEET Registry, which includes centres in Europe, North America, Asia/Middle East and Australia, also revealed increasing CGM and pump use in children around the same time [8]. A recent population-based study from Australia reported a remarkable increase in CGM uptake from 5% to 79% following the introduction of universal subsidised CGM funding for people with type 1 diabetes under the age of 21 years [18].
AID use is similarly increasing in children with type 1 diabetes, although this technology is not as widespread as CGM and access is highly variable across the world. Multiple AID systems recently became commercially available in the USA and Europe, and uptake in the USA is increasing fairly rapidly based on early reports [19]. Interestingly, technology use, including pump use, is highly variable across clinical centres even within the same country [20].
AID in school-aged children and adolescents
As paediatric AID research has moved from the inpatient setting to supervised hotel studies, to extended use at summer/winter camps and then prolonged home use, it has become very clear that AID systems are safe and effective for children with type 1 diabetes. It is beyond the scope of this review to include literature leading up to the commercialisation of AID systems. Data from paediatric pivotal studies of currently available AID systems (Table 1) show improved glycaemic outcomes with AID [21–28]. Of note, direct comparisons of systems cannot be made due to variable study design and differing CGMs used. Additionally, multiple open-source or ‘do-it-yourself’ systems are also available but are not covered in this review.
Table 1.
Study detail | Medtronic 670G [21] | Tandem Control-IQ [23, 24] | Omnipod 5 [25] | Medtronic 780G [26] | iLet [27] | CamAPS FX [22, 28] | ||
---|---|---|---|---|---|---|---|---|
Date | 2016 | 2019 | 2020 | 2021 | 2022 | 2022 | 2018 | 2022 |
Study design | Single arm | RCT | RCT | Single arm | Single arm | RCT | RCT | RCT |
Comparator | 2 week run-in | CGM + pump | CGM + pump | 2 week run-in | 2 week run-in | Usual care (~30% on AID) | CGM + pump | CGM + pump |
AID duration, months | 3 | 6 | 4 | 3 | 3 | 3 | 4 | 6 |
No. of participants | 124 | 168 | 101 | 240 | 157 | 326 | 86 | 133 |
Participant age, years | 14–75 | 14–71 | 6–13 | 6–70 | 14–75 | 6 and older | 6 and older | 6–18 |
Baseline HbA1c, mmol/mol (%) | 57 | 57 | 58 / 61 | 55 | 58 | 61 / 63 | 67 / 66 | 66 67 |
Baseline HbA1ca, % | 7.4 | 7.4 | 7.5 / 7.7 | 7.2 | 7.5 | 7.7 / 7.9 | 8.3 / 8.2 | 8.2 / 8.3 |
ΔTIR, % | +5 | +11 | +12.4 | +9 | +5.7 | +11 | +10.8 | +6.7 |
ΔHbA1c, % | −0.5 | −0.33 | −0.42 | −0.38 | −0.5 | −0.5 | −0.36 | −0.32 |
aWhere two HbA1c values are given, the papers did not report a mean baseline HbA1c value for the entire group, but rather listed baseline HbA1c values separately for the closed-loop and control groups, respectively
Although AID use improves glycaemic outcomes and eases the burden of keeping up with increasing insulin needs as children grow and go through puberty, challenges in this age range persist. Current AID options are all hybrid closed-loop systems, meaning that the user is expected to announce meals and in most cases provide the carbohydrate content. One of the newest AID systems on the market allows for an estimation of carbohydrate intake (usual/more/less) but the need for an announcement of a meal remains [27]. Missed or late meal announcements are common in older children and adolescents and might occur more often in adolescents on AID as users learn that the system will compensate by increasing the insulin delivery in response to postprandial hyperglycaemia [29–31].
Children and adolescents who are already doing well prior to AID initiation tend to achieve the highest TIR on AID, while those who have worse glucose management at baseline exhibit the greatest improvement in TIR [32]. This highlights the important points that essentially all children benefit from AID and that the vast majority are candidates for AID use even if they traditionally would not have been thought of as a ‘good’ candidate for an open-loop pump in the past.
AID in toddlers and preschoolers
Children between the ages of 2 and 5 years present unique challenges to diabetes management. Toddlers are often unpredictable when it comes to meal intake and activity. They have low insulin needs and can be exquisitely sensitive to small doses of insulin, limiting the ability of even a half-unit insulin pen or syringe to match their needs. Young children often have hypoglycaemia unawareness or might not be unable to communicate their symptoms of hypoglycaemia. Parents and caregivers of young children are understandably fearful of hypoglycaemia and sometimes contribute to excess hyperglycaemia by over-treating lows, under-dosing insulin or providing extra uncovered snacks when they might not be needed. This age group is entirely reliant on their parents and caregivers for their diabetes management, which often means better glycaemic control as compared with older children and adolescents due to infrequent missed meal boluses. However, this around-the-clock care can be quite burdensome to parents and greatly limits their childcare options [33].
Despite these many challenges, AID systems work well for most young children, even those with low total daily insulin needs, and there are now multiple commercially available AID options for children with type 1 diabetes down to age 2 years in the USA and 1 year in Europe. Table 2 summarises the pivotal study data for each system in this age group [34–37].
Table 2.
Study detail | Omnipod 5 [34] | Medtronic 670G [35] | CamAPS FX [36] | Tandem Control-IQ [37] |
---|---|---|---|---|
Date | 2022 | 2022 | 2022 | 2023 |
Study design | Single arm | Single arm | RCT, crossover | RCT |
Comparator | 2 week run-in | 2 week run-in | CGM + pump | CGM + pump / MDI |
AID duration | 13 weeks | 12 weeks | 16 weeks | 13 weeks |
No. of participants | 80 | 46 | 74 | 102 |
Participant age, years | 2–5.9 | 2–6 | 1–7 | 2–5.9 |
Baseline HbA1c, mmol/mol | 57 | 64 | 56 | 58 |
Baseline HbA1c, % | 7.4 | 8 | 7.3 | 7.5 |
ΔTIR, % | +10.9 | +8.1 | +8.7 | +12.4 |
ΔHbA1c, % | −0.55 | −0.5 | −0.4 | −0.42 |
Patient-reported outcomes
The impact of diabetes technology on the lives of children with type 1 diabetes and their families extends well beyond improved glycaemic outcomes. Early CGM introduction following type 1 diabetes diagnosis in children is associated with reduced hypoglycaemia avoidance behaviours in parents [38]. Parents of children using AID report a reduction in the burden of diabetes, reduced fear of hypoglycaemia and improved quality of sleep in multiple studies [39–42]. In addition to survey data, qualitative data from parental interviews is overwhelmingly positive, highlighting improvements in parental mental health and quality of life, as well as increased parental confidence in outsourcing childcare, with AID use in the child [40, 43, 44].
Children and adolescents with type 1 diabetes generally self-report high satisfaction and acceptability of AID, although evidence of statistically significant improvements in other child-reported psychosocial metrics is harder to find [45, 46]. Other studies have shown clear benefits in child-reported metrics including decreased fear of hypoglycaemia, decreased diabetes distress and increased well-being [47]. Qualitative data from interviews of children and adolescents on AID have also been positive, with children reporting feeling more rested and less worried about their illness after starting AID [43, 44, 47].
Application of technology in real life and practical considerations
Early commercially available AID systems were found to have a high rate of discontinuation and suboptimal time in closed loop when used by children with type 1 diabetes [48]. However, AID usability has improved over time. Numerous recent publications of real-world data show that children with type 1 diabetes are able to achieve a consistently high rate of time in closed loop and achieve glycaemic benefits similar to those observed in RCTs across all AID systems [17, 49–52]. The rate of AID discontinuation reported in these real-world studies is reassuringly low.
It is important for the clinical team to set realistic expectations and appropriately counsel families at the time of starting any new technology [53]. CGM readings do not always perfectly match glucometer readings, particularly when glucose is changing rapidly. This can be confusing and frustrating for users if they are not appropriately counselled on what to expect and also creates distrust in the technology. Users of technology need to learn how to troubleshoot data gaps and connectivity issues. Additionally, devices do not always stay on the body for the approved number of days and overlay patches and extra skin preparations are often necessary. Children often report pain with device insertion and skin irritation from the adhesive is common, at times limiting usability [54]. Like in open-loop pumps, infusion-set failure remains common in children using AID and can potentially lead to diabetic ketoacidosis (DKA) if not addressed promptly. Importantly, fewer DKA episodes are observed in children using diabetes technology as compared with those who are not [8].
There are a number of paediatric-specific practical issues with technology use, including the need to train school nurses and other caregivers on how to use a new device and respond to alarms. Remote data sharing has certainly improved safety, although parents are often now more involved than ever in the child’s daily diabetes management at school and communicate with the nurse or teacher multiple times per day about their child’s glucose trends, mealtime dosing and management of exercise during the school day [55].
A final consideration is the timing of technology introduction following diagnosis. CGM introduction now occurs quite early for many children with type 1 diabetes, with documented benefits to early use. However, there tends to be a lag in AID introduction for even the most motivated families, largely due in the USA to insurance constraints. Most of the pivotal AID studies excluded children who were diagnosed <6 months prior to study enrolment to avoid overlap with the ‘honeymoon period’ when glycaemic management almost universally improves as endogenous insulin production transiently resumes. Clinicians have often been reluctant to prescribe AID right at the time of diagnosis, largely related to concerns regarding the ability of a family to learn the complexities of AID in addition to the extensive diabetes teaching necessary at the time of diagnosis. However, there are now two large RCTs that have clearly documented successful early AID introduction. Although these studies do not show that beta cell function is preserved with early AID use, they do suggest that early AID introduction is likely safe and clearly effective at improving glycaemic outcomes following diagnosis [56, 57]. Moreover, evidence suggests that early AID use might diminish the negative glycaemic impact of DKA at the time of diagnosis [58]. As such, it is likely that the gap between diabetes diagnosis and AID introduction will continue to shorten over time.
Barriers to technology use
Multiple barriers exist that limit more widespread use of diabetes technology in children. Adolescents and even younger children may be resistant to having a physical device attached to them [59] and young children have a limited body surface area on which to place devices [54, 60]. Although most families are reassured by data, others feel more anxious about diabetes when monitoring real-time CGM data [54]. Similarly, children and adolescents commonly eschew attracting attention to themselves, due to cosmetic issues or the embarrassment of highlighting their diabetes status in front of classmates [61, 62]. The audible alarms for high or low glucose levels are a frequent source of frustration that cause some to discontinue use [63].
When a child resists wearing a pump but is willing to use a CGM, smart insulin pens connected via Bluetooth to phone apps can provide an alternative option to benefit further from technology. Potential benefits of smart pens include insulin dose calculation with adjustment for insulin on board, the ability to set reminders and the generation of data to aid clinicians in making dose adjustments. In a recent small RCT, initiation of a smart pen cap in adolescents and adults with type 1 diabetes was associated with a 5.2% increase in TIR [64]. An earlier observational study reported a reduction in hypoglycaemia but not hyperglycaemia in children with type 1 diabetes after smart pen introduction [65]. More high-quality evidence is needed to determine the impact of smart pen use in children with type 1 diabetes.
The high cost of diabetes technology is also a barrier to use and global coverage by many national healthcare systems remains limited [66]. There has been an increase in coverage in many countries over time. In Europe, and even more so in the USA, there have been gaps in coverage of technology use in families of lower socioeconomic status (SES) [67]. Importantly, a recent study evaluating predictors of success with AID use did not find a difference when comparing US families receiving Medicaid with those having private insurance, suggesting that all children benefit from AID use regardless of SES [68]. Furthermore, recent data suggest that AID systems are actually cost effective compared with standard care due to improved glycaemic outcomes [69, 70].
Healthcare providers can also act as a barrier to technology access, sometimes serving as ‘gatekeepers’ by only offering options to families they feel can handle the increased complexity of technology-based tools [71, 72]. There appear to be racial/ethnic differences in technology use that exist beyond differences in SES. In one study of CGM use among US families with commercial insurance in the first year following type 1 diabetes diagnosis, 96% of non-Hispanic white children vs 73% of black children were using CGM [73].
Given that AID trials have found that the greatest magnitude of benefit occurs in children and adolescents with the lowest baseline TIR [32], there are growing arguments to expand access among all groups, including minority populations that have historically exhibited low rates of technology use [71, 74, 75]. Further, the research performed in this area has over-represented individuals of higher SES, and there are clear needs to learn the predictors of success among under-represented groups to better identify approaches that are likely to be efficacious among all groups [76].
Future research
Future research in this area will focus not only on continued improvement in glycaemic outcomes but also on improvement in patients’ quality of life. A goal of many future systems is to function without user input, referred to as fully closed-loop (FCL) systems. Due to the experimental nature of these, most have been tested predominantly in adults. FCL systems use additional features for detecting food intake that do not require user action (e.g. a smart watch that detects eating and drinking movements, algorithmic assessments of CGM flux likely associated with a meal, system anticipation of meals or more-aggressive insulin administration coupled with glucagon injection) [77–80]. Other approaches have focused on accounting for exercise, either via anticipation or heart-rate information [81, 82].
Head-to-head trials comparing systems used with and without carbohydrate announcement demonstrate that announcing carbohydrate intake still results in greater TIR; however, studies also show that systems function better than existing systems when meals are not announced [78]. Trials have demonstrated in many cases acceptable TIR when using AID in FCL (58–78%), which could improve glycaemic management in children and adolescents who forget or choose not to bolus for meals [78, 80, 83].
Another area where AID systems could be improved is in reducing the delay between sensing an elevation in glucose and the action of the injected insulins. Thus, faster-acting insulins have been a source of hope for improving postprandial glycaemic management. Some studies have suggested minor glycaemic improvements are obtained with use of an ultra-rapid-acting insulin with AID, although no glycaemic improvements have been found in paediatric studies [84, 85]. Some published work indicates that reworking of the controller algorithm or the insulin settings may be necessary to achieve more complete efficacy [86–88].
Other approaches tested in adults have focused on adding other diabetes medications to AID use, including pramlintide (which increased TIR by 10% when co-secreted with insulin) [89], semaglutide (which reduced insulin requirements in early type 1 diabetes) [90] and sodium–glucose cotransporter 2 (SGLT2) inhibitors (which increased TIR by 9.9% when taken orally alongside AID use) [91]. Still, with SGLT2s, the potential for unrecognised ketosis likely limits this use in adolescents until continuous ketone monitoring becomes available [92]. It is still unclear whether the use of automated glucagon delivery with an AID system will greatly improve care, as comparisons of TIR between insulin alone or bihormonal systems have been complicated by trial design and optimised systems have not been tested head-to-head [93]. A recent prospective study of an FCL bihormonal system in adults reported impressive glycaemic outcomes (TIR 80.3%), although there was a fairly high dropout rate largely due to the burden required to use the system (two CGMs and two pumps/infusion sets are required) [94].
New horizons
Compared with previous paradigm-shifting approaches, the process towards development, approval and widespread commercial availability of AID systems has progressed rapidly. Still, more rapid advancement may be allowed if there is greater ability for users to switch between individual components. This shift began in 2023 and now multiple AID systems pair with two or more CGMs. This interoperability between brands will allow users to benefit from ongoing improvements between one CGM type and another and allow individuals to select CGM and AID devices that best suit their needs.
Outside of AID control, there has been interest in using predictive algorithms to provide automated advice to patients or physicians. These decision-support systems have met with limited success thus far [95–97], highlighting the complexities of an approach that highly depends on reliable data over time and the user’s willingness to follow through with recommendations [98]. Still, this type of approach should eventually help in optimising insulin dosing variables.
Finally, with respect to the inequalities of use of diabetes technology among under-represented populations, there is a need for research that could lead to these technologies becoming more accessible and available [76]. This includes demonstrations of ease-of-use and efficacy in all populations to convince all levels of decision makers about the need to increase access to all people with diabetes [75].
Conclusions
It is an exciting time for paediatric endocrinologists, diabetes researchers and children with type 1 diabetes and their families as advances in diabetes technology continue to be made. Reliable CGMs opened the door to successful AID development, which in turn has improved glycaemic outcomes and decreased the burden of diabetes for children with type 1 diabetes and their families. Affordability and access for all to technology should be a top priority moving forward. Although the impact of AID use on long-term diabetes complications remains to be determined, a reduction in complications is certainly to be expected given the positive impact this technology has had on glycaemic management in youth with type 1 diabetes.
Abbreviations
- AID
Automated insulin delivery
- CGM
Continuous glucose monitor
- DKA
Diabetic ketoacidosis
- FCL
Fully closed-loop
- MDI
Multiple daily insulin injections
- SES
Socioeconomic status
- SGLT2
Sodium–glucose cotransporter 2
- TIR
Time in range
Funding
All authors are funded by grant U01DK127551 from NIH NIDDK.
Authors’ relationships and activities
MJS reports research funding, paid to her university, from Tandem Diabetes Care and Insulet. MDD reports research funding paid to his university from Dexcom Inc, Tandem Diabetes Inc and Medtronic. MDB reports research funding paid to his university from Dexcom, Tandem Diabetes Care and Novo Nordisk, and non-financial research support from Tandem Diabetes Care and Dexcom. Additionally, MDB has a number of patents with royalties licensed to Dexcom, Sanofi, Tandem Diabetes Care and Novo Nordisk. MDB finally reports consulting/speakership activities with Dexcom, Tandem Diabetes Care, Roche, Portal Insulin LLC, BoydSense and Vertex.
Contribution statement
All authors were responsible for drafting this review article and reviewing it critically for important intellectual content. All authors approved the version to be published.
Footnotes
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
References
- 1.Zhang K, Kan C, Han F et al (2023) Global, regional, and national epidemiology of diabetes in children from 1990 to 2019. JAMA Pediatr 177(8):837–846. 10.1001/jamapediatrics.2023.2029 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2.Lawrence JM, Divers J, Isom S et al (2021) Trends in prevalence of type 1 and type 2 diabetes in children and adolescents in the US, 2001–2017. JAMA 326(8):717–727. 10.1001/jama.2021.11165 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.Rawshani A, Sattar N, Franzén S et al (2018) Excess mortality and cardiovascular disease in young adults with type 1 diabetes in relation to age at onset: a nationwide, register-based cohort study. Lancet (London, England) 392(10146):477–486. 10.1016/s0140-6736(18)31506-x [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.Acerini CL, Williams RM, Dunger DB (2001) Metabolic impact of puberty on the course of type 1 diabetes. Diabetes Metab 27(4 Pt 2):S19-25 [PubMed] [Google Scholar]
- 5.Michou P, Gkiourtzis N, Christoforidis A, Kotanidou EP, Galli-Tsinopoulou A (2023) The efficacy of automated insulin delivery systems in children and adolescents with type 1 diabetes Mellitus: A systematic review and meta-analysis of randomized controlled trials. Diabetes Res Clin Pract 199:110678. 10.1016/j.diabres.2023.110678 [DOI] [PubMed] [Google Scholar]
- 6.Zeng B, Gao L, Yang Q, Jia H, Sun F (2023) Automated insulin delivery systems in children and adolescents with type 1 diabetes: a systematic review and meta-analysis of outpatient randomized controlled trials. Diabetes Care 46(12):2300–2307. 10.2337/dc23-0504 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7.Bombaci B, Passanisi S, Alibrandi A et al (2022) One-year real-world study on comparison among different continuous subcutaneous insulin infusion devices for the management of pediatric patients with type 1 diabetes: the supremacy of hybrid closed-loop systems. Int J Environ Res Public Health 19(16):10293. 10.3390/ijerph191610293 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8.Cardona-Hernandez R, Schwandt A, Alkandari H et al (2021) Glycemic outcome associated with insulin pump and glucose sensor use in children and adolescents with type 1 diabetes. Data from the international pediatric registry SWEET. Diabetes Care 44(5):1176–1184. 10.2337/dc20-1674 [DOI] [PubMed] [Google Scholar]
- 9.Nevo-Shenker M, Shalitin S (2021) The impact of hypo- and hyperglycemia on cognition and brain development in young children with type 1 diabetes. Horm Res Paediatr 94(3–4):115–123. 10.1159/000517352 [DOI] [PubMed] [Google Scholar]
- 10.Biessels GJ, Reijmer YD (2014) Brain changes underlying cognitive dysfunction in diabetes: what can we learn from MRI? Diabetes 63(7):2244–2252. 10.2337/db14-0348 [DOI] [PubMed] [Google Scholar]
- 11.Mauras N, Buckingham B, White NH et al (2021) Impact of type 1 diabetes in the developing brain in children: a longitudinal study. Diabetes Care 44(4):983–992. 10.2337/dc20-2125 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.Redondo MJ, Libman I, Maahs DM et al (2021) The evolution of hemoglobin A(1c) targets for youth with type 1 diabetes: rationale and supporting evidence. Diabetes Care 44(2):301–312. 10.2337/dc20-1978 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13.American Diabetes Association Professional Practice Committee (2023) 14. Children and adolescents: standards of care in diabetes—2024. Diabetes Care 47(Supplement_1):S258–S281. 10.2337/dc24-S014 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14.de Bock M, Codner E, Craig ME et al (2022) ISPAD Clinical Practice Consensus Guidelines 2022: Glycemic targets and glucose monitoring for children, adolescents, and young people with diabetes. Pediatr Diabetes 23(8):1270–1276. 10.1111/pedi.13455 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15.Foster NC, Beck RW, Miller KM et al (2019) State of type 1 diabetes management and outcomes from the T1D exchange in 2016–2018. Diabetes Technol Ther 21(2):66–72. 10.1089/dia.2018.0384 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16.Battelino T, Danne T, Bergenstal RM et al (2019) Clinical targets for continuous glucose monitoring data interpretation: recommendations from the international consensus on time in range. Diabetes Care 42(8):1593–1603. 10.2337/dci19-0028 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17.Breton MD, Kovatchev BP (2021) One year real-world use of the control-IQ advanced hybrid closed-loop technology. Diabetes Technol Ther 23(9):601–608. 10.1089/dia.2021.0097 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18.Johnson SR, Holmes-Walker DJ, Chee M et al (2022) Universal subsidized continuous glucose monitoring funding for young people with type 1 diabetes: uptake and outcomes over 2 years, a population-based study. Diabetes Care 45(2):391–397. 10.2337/dc21-1666 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19.Alonso GT, Triolo TM, Akturk HK et al (2023) Increased technology use associated with lower A1C in a large pediatric clinical population. Diabetes Care 46(6):1218–1222. 10.2337/dc22-2121 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20.Prahalad P, Hardison H, Odugbesan O et al (2024) Benchmarking diabetes technology use among 21 U.S. Pediatric diabetes centers. Clin Diabetes 42(1):27–33. 10.2337/cd23-0052 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21.Bergenstal RM, Garg S, Weinzimer SA et al (2016) Safety of a hybrid closed-loop insulin delivery system in patients with type 1 diabetes. JAMA 316(13):1407–1408. 10.1001/jama.2016.11708 [DOI] [PubMed] [Google Scholar]
- 22.Tauschmann M, Thabit H, Bally L et al (2018) Closed-loop insulin delivery in suboptimally controlled type 1 diabetes: a multicentre, 12-week randomised trial. Lancet (London, England) 392(10155):1321–1329. 10.1016/s0140-6736(18)31947-0 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 23.Brown SA, Kovatchev BP, Raghinaru D et al (2019) Six-month randomized, multicenter trial of closed-loop control in type 1 diabetes. N Engl J Med 381(18):1707–1717. 10.1056/NEJMoa1907863 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 24.Breton MD, Kanapka LG, Beck RW et al (2020) A randomized trial of closed-loop control in children with type 1 diabetes. N Engl J Med 383(9):836–845. 10.1056/NEJMoa2004736 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 25.Brown SA, Forlenza GP, Bode BW et al (2021) Multicenter trial of a tubeless, on-body automated insulin delivery system with customizable glycemic targets in pediatric and adult participants with type 1 diabetes. Diabetes Care 44(7):1630–1640. 10.2337/dc21-0172 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 26.Carlson AL, Sherr JL, Shulman DI et al (2022) Safety and glycemic outcomes during the minimed™ advanced hybrid closed-loop system pivotal trial in adolescents and adults with type 1 diabetes. Diabetes Technol Ther 24(3):178–189. 10.1089/dia.2021.0319 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 27.Russell SJ, Beck RW, Damiano ER et al (2022) Multicenter, randomized trial of a bionic pancreas in type 1 diabetes. N Engl J Med 387(13):1161–1172. 10.1056/NEJMoa2205225 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 28.Ware J, Boughton CK, Allen JM et al (2022) Cambridge hybrid closed-loop algorithm in children and adolescents with type 1 diabetes: a multicentre 6-month randomised controlled trial. Lancet Digit Health 4(4):e245–e255. 10.1016/s2589-7500(22)00020-6 [DOI] [PubMed] [Google Scholar]
- 29.Chernavvsky DR, DeBoer MD, Keith-Hynes P et al (2016) Use of an artificial pancreas among adolescents for a missed snack bolus and an underestimated meal bolus. Pediatr Diabetes 17(1):28–35. 10.1111/pedi.12230 [DOI] [PubMed] [Google Scholar]
- 30.Buckingham BA, Bergenstal RM (2024) Decreasing the burden of carbohydrate counting and meal announcement with automated insulin delivery, meal recognition, and autocorrection doses: a case study. Diabetes Technol Ther 26(S3):97–101. 10.1089/dia.2023.0505 [DOI] [PubMed] [Google Scholar]
- 31.Kovatchev BP, Singh H, Mueller L, Gonder-Frederick LA (2022) Biobehavioral changes following transition to automated insulin delivery: a large real-life database analysis. Diabetes Care 45(11):2636–2643. 10.2337/dc22-1217 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 32.Schoelwer MJ, Kanapka LG, Wadwa RP et al (2021) Predictors of time-in-range (70–180 mg/dL) achieved using a closed-loop control system. Diabetes Technol Ther 23(7):475–481. 10.1089/dia.2020.0646 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 33.Dehn-Hindenberg A, Saßmann H, Berndt V et al (2021) Long-term occupational consequences for families of children with type 1 diabetes: the mothers take the burden. Diabetes Care 44(12):2656–2663. 10.2337/dc21-0740 [DOI] [PubMed] [Google Scholar]
- 34.Sherr JL, Bode BW, Forlenza GP et al (2022) Safety and glycemic outcomes with a tubeless automated insulin delivery system in very young children with type 1 diabetes: a single-arm multicenter clinical trial. Diabetes Care 45(8):1907–1910. 10.2337/dc21-2359 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 35.Forlenza GP, Ekhlaspour L, DiMeglio LA et al (2022) Glycemic outcomes of children 2–6 years of age with type 1 diabetes during the pediatric MiniMed™ 670G system trial. Pediatr Diabetes 23(3):324–329. 10.1111/pedi.13312 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 36.Ware J, Allen JM, Boughton CK et al (2022) Randomized trial of closed-loop control in very young children with type 1 diabetes. N Engl J Med 386(3):209–219. 10.1056/NEJMoa2111673 [DOI] [PubMed] [Google Scholar]
- 37.Wadwa RP, Reed ZW, Buckingham BA et al (2023) Trial of hybrid closed-loop control in young children with type 1 diabetes. N Engl J Med 388(11):991–1001. 10.1056/NEJMoa2210834 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 38.Youngkin EM, Majidi S, Noser AE, Stanek KR, Clements MA, Patton SR (2021) Continuous glucose monitoring decreases hypoglycemia avoidance behaviors, but not worry in parents of youth with new onset type 1 diabetes. J Diabetes Sci Technol 15(5):1093–1097. 10.1177/1932296820929420 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 39.Musolino G, Dovc K, Boughton CK et al (2019) Reduced burden of diabetes and improved quality of life: experiences from unrestricted day-and-night hybrid closed-loop use in very young children with type 1 diabetes. Pediatr Diabetes 20(6):794–799. 10.1111/pedi.12872 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 40.Hood KK, Schneider-Utaka AK, Reed ZW et al (2024) Patient reported outcomes (PROs) and user experiences of young children with type 1 diabetes using t:slim X2 insulin pump with control-IQ technology. Diabetes Res Clin Pract 208:111114. 10.1016/j.diabres.2024.111114 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 41.Cobry EC, Bisio A, Wadwa RP, Breton MD (2022) Improvements in parental sleep, fear of hypoglycemia, and diabetes distress with use of an advanced hybrid closed-loop system. Diabetes Care 45(5):1292–1295. 10.2337/dc21-1778 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 42.de Beaufort C, Schierloh U, Thankamony A et al (2022) Cambridge hybrid closed-loop system in very young children with type 1 diabetes reduces caregivers’ fear of hypoglycemia and improves their well-being. Diabetes Care 45(12):3050–3053. 10.2337/dc22-0693 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 43.Kimbell B, Rankin D, Hart RI et al (2022) Parents’ experiences of using a hybrid closed-loop system (CamAPS FX) to care for a very young child with type 1 diabetes: qualitative study. Diabetes Res Clin Pract 187:109877. 10.1016/j.diabres.2022.109877 [DOI] [PubMed] [Google Scholar]
- 44.Rankin D, Kimbell B, Hovorka R, Lawton J (2022) Adolescents’ and their parents’ experiences of using a closed-loop system to manage type 1 diabetes in everyday life: qualitative study. Chronic Illn 18(4):742–756. 10.1177/1742395320985924 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 45.Weissberg-Benchell J, Vesco AT, Shapiro J et al (2023) Psychosocial impact of the insulin-only iLet bionic pancreas for adults, youth, and caregivers of youth with type 1 diabetes. Diabetes Technol Ther 25(10):705–717. 10.1089/dia.2023.0238 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 46.Cobry EC, Kanapka LG, Cengiz E et al (2021) Health-related quality of life and treatment satisfaction in parents and children with type 1 diabetes using closed-loop control. Diabetes Technol Ther 23(6):401–409. 10.1089/dia.2020.0532 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 47.Gianini A, Suklan J, Skela-Savič B et al (2022) Patient reported outcome measures in children and adolescents with type 1 diabetes using advanced hybrid closed loop insulin delivery. Front Endocrinol (Lausanne) 13:967725. 10.3389/fendo.2022.967725 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 48.Berget C, Messer LH, Vigers T et al (2020) Six months of hybrid closed loop in the real-world: an evaluation of children and young adults using the 670G system. Pediatr Diabetes 21(2):310–318. 10.1111/pedi.12962 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 49.Lombardo F, Passanisi S, Alibrandi A et al (2023) MiniMed 780G six-month use in children and adolescents with type 1 diabetes: clinical targets and predictors of optimal glucose control. Diabetes Technol Ther 25(6):404–413. 10.1089/dia.2022.0491 [DOI] [PubMed] [Google Scholar]
- 50.Messer LH, Berget C, Pyle L et al (2021) Real-world use of a new hybrid closed loop improves glycemic control in youth with type 1 diabetes. Diabetes Technol Ther 23(12):837–843. 10.1089/dia.2021.0165 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 51.Arunachalum S, Velado K, Vigersky RA, Cordero TL (2022) Glycemic outcomes during real-world hybrid closed-loop system use by individuals with type 1 diabetes in the United States. J Diabetes Sci Technol 19322968221088608. 10.1177/19322968221088608 [DOI] [PMC free article] [PubMed]
- 52.Silva JD, Lepore G, Battelino T et al (2022) Real-world performance of the MiniMed™ 780G system: first report of outcomes from 4120 users. Diabetes Technol Ther 24(2):113–119. 10.1089/dia.2021.0203 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 53.Sherr JL, Schoelwer M, Dos Santos TJ et al (2022) ISPAD clinical practice consensus guidelines 2022: diabetes technologies: insulin delivery. Pediatr Diabetes 23(8):1406–1431. 10.1111/pedi.13421 [DOI] [PubMed] [Google Scholar]
- 54.Hilliard ME, Levy W, Anderson BJ et al (2019) Benefits and barriers of continuous glucose monitoring in young children with type 1 diabetes. Diabetes Technol Therap 21(9):493–498. 10.1089/dia.2019.0142 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 55.March CA, Nanni M, Kazmerski TM, Siminerio LM, Miller E, Libman IM (2020) Modern diabetes devices in the school setting: perspectives from school nurses. Pediatr Diabetes 21(5):832–840. 10.1111/pedi.13015 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 56.Boughton CK, Allen JM, Ware J et al (2022) Closed-loop therapy and preservation of C-peptide secretion in type 1 diabetes. N Engl J Med 387(10):882–893. 10.1056/NEJMoa2203496 [DOI] [PubMed] [Google Scholar]
- 57.McVean J, Forlenza GP, Beck RW et al (2023) Effect of tight glycemic control on pancreatic beta cell function in newly diagnosed pediatric type 1 diabetes: a randomized clinical trial. JAMA 329(12):980–989. 10.1001/jama.2023.2063 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 58.Lakshman R, Najami M, Allen JM et al (2024) Diabetic ketoacidosis at onset of type 1 diabetes and glycemic outcomes with closed-loop insulin delivery. Diabetes Technol Ther 26(3):198–202. 10.1089/dia.2023.0307 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 59.Pauley ME, Berget C, Messer LH, Forlenza GP (2021) Barriers to uptake of insulin technologies and novel solutions. Med Devices (Auckl) 14:339–354. 10.2147/mder.S312858 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 60.Nevo-Shenker M, Phillip M, Nimri R, Shalitin S (2020) Type 1 diabetes mellitus management in young children: implementation of current technologies. Pediatr Res 87(4):624–629. 10.1038/s41390-019-0665-4 [DOI] [PubMed] [Google Scholar]
- 61.Forlenza GP, Messer LH, Berget C, Wadwa RP, Driscoll KA (2018) Biopsychosocial factors associated with satisfaction and sustained use of artificial pancreas technology and its components: a call to the technology field. Curr Diab Rep 18(11):114. 10.1007/s11892-018-1078-1 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 62.Kubiak T, Mann CG, Barnard KC, Heinemann L (2016) Psychosocial aspects of continuous glucose monitoring: connecting to the patients’ experience. J Diabetes Sci Technol 10(4):859–863. 10.1177/1932296816651450 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 63.Messer LH, Johnson R, Driscoll KA, Jones J (2018) Best friend or spy: a qualitative meta-synthesis on the impact of continuous glucose monitoring on life with Type 1 diabetes. Diabet Med 35(4):409–418. 10.1111/dme.13568 [DOI] [PubMed] [Google Scholar]
- 64.Gomez-Peralta F, Abreu C, Fernández-Rubio E et al (2023) Efficacy of a connected insulin pen cap in people with noncontrolled type 1 diabetes: a multicenter randomized clinical trial. Diabetes Care 46(1):206–208. 10.2337/dc22-0525 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 65.Adolfsson P, Björnsson V, Hartvig NV, Kaas A, Møller JB, Ogionwo Lange E (2022) Improved glycemic control observed in children with type 1 diabetes following the introduction of smart insulin pens: a real-world study. Diabetes Ther 13(1):43–56. 10.1007/s13300-021-01177-w [DOI] [PMC free article] [PubMed] [Google Scholar]
- 66.Dos Santos TJ, Dave C, MacLeish S, Wood JR (2021) Diabetes technologies for children and adolescents with type 1 diabetes are highly dependent on coverage and reimbursement: results from a worldwide survey. BMJ Open Diabetes Res Care 9(2):e002537. 10.1136/bmjdrc-2021-002537 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 67.Addala A, Auzanneau M, Miller K et al (2021) A decade of disparities in diabetes technology use and HbA(1c) in pediatric type 1 diabetes: a transatlantic comparison. Diabetes Care 44(1):133–140. 10.2337/dc20-0257 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 68.Forlenza GP, Vigers T, Berget C et al (2022) Predicting success with a first-generation hybrid closed-loop artificial pancreas system among children, adolescents, and young adults with type 1 diabetes: a model development and validation study. Diabetes Technol Ther 24(3):157–166. 10.1089/dia.2021.0326 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 69.Fox DS, Ware J, Boughton CK et al (2024) Cost-effectiveness of closed-loop automated insulin delivery using the Cambridge hybrid algorithm in children and adolescents with type 1 diabetes: results from a multicenter 6-month randomized trial. J Diabetes Sci Technol 19322968241231950. 10.1177/19322968241231950 [DOI] [PMC free article] [PubMed]
- 70.Biskupiak JE, Ramos M, Levy CJ et al (2023) Cost-effectiveness of the tubeless automated insulin delivery system vs standard of care in the management of type 1 diabetes in the United States. J Manag Care Spec Pharm 29(7):807–817. 10.18553/jmcp.2023.22331 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 71.Agarwal S, Crespo-Ramos G, Long JA, Miller VA (2021) “I didn’t really have a choice”: qualitative analysis of racial-ethnic disparities in diabetes technology use among young adults with type 1 diabetes. Diabetes Technol Ther 23(9):616–622. 10.1089/dia.2021.0075 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 72.Addala A, Hanes S, Naranjo D, Maahs DM, Hood KK (2021) Provider implicit bias impacts pediatric type 1 diabetes technology recommendations in the united states: findings from the gatekeeper study. J Diabetes Sci Technol 15(5):1027–1033. 10.1177/19322968211006476 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 73.Lai CW, Lipman TH, Willi SM, Hawkes CP (2021) Early racial/ethnic disparities in continuous glucose monitor use in pediatric type 1 diabetes. Diabetes Technol Ther 23(11):763–767. 10.1089/dia.2021.0120 [DOI] [PubMed] [Google Scholar]
- 74.Brown SA (2022) On the path toward expanding treatment options for diabetes. N Engl J Med 387(10):935–936. 10.1056/NEJMe2209840 [DOI] [PubMed] [Google Scholar]
- 75.Akturk HK, Agarwal S, Hoffecker L, Shah VN (2021) Inequity in racial-ethnic representation in randomized controlled trials of diabetes technologies in type 1 diabetes: critical need for new standards. Diabetes Care 44(6):e121–e123. 10.2337/dc20-3063 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 76.García-Tirado J, Villard O, Hall M, Bisio A, Gonder-Frederick L (2023) Under-representation of diverse populations and glycemic outcomes in major clinical trials of automated insulin delivery. Diabetes Technol Ther 25(10):752–754. 10.1089/dia.2023.0197 [DOI] [PubMed] [Google Scholar]
- 77.Corbett JP, Hsu L, Brown SA et al (2022) Smartwatch gesture-based meal reminders improve glycaemic control. Diabetes Obes Metab 24(8):1667–1670. 10.1111/dom.14737 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 78.Garcia-Tirado J, Diaz JL, Esquivel-Zuniga R et al (2021) Advanced closed-loop control system improves postprandial glycemic control compared with a hybrid closed-loop system following unannounced meal. Diabetes Care. 10.2337/dc21-0932 [DOI] [PubMed] [Google Scholar]
- 79.Garcia-Tirado J, Colmegna P, Villard O et al (2023) Assessment of meal anticipation for improving fully automated insulin delivery in adults with type 1 diabetes. Diabetes Care 46(9):1652–1658. 10.2337/dc23-0119 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 80.Blauw H, Onvlee AJ, Klaassen M, van Bon AC, DeVries JH (2021) Fully closed loop glucose control with a bihormonal artificial pancreas in adults with type 1 diabetes: an outpatient, randomized, crossover trial. Diabetes Care 44(3):836–838. 10.2337/dc20-2106 [DOI] [PubMed] [Google Scholar]
- 81.Garcia-Tirado J, Brown SA, Laichuthai N et al (2021) Anticipation of historical exercise patterns by a novel artificial pancreas system reduces hypoglycemia during and after moderate-intensity physical activity in people with type 1 diabetes. Diabetes Technol Ther 23(4):277–285. 10.1089/dia.2020.0516 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 82.DeBoer MD, Cherñavvsky DR, Topchyan K, Kovatchev BP, Francis GL, Breton MD (2017) Heart rate informed artificial pancreas system enhances glycemic control during exercise in adolescents with T1D. Pediatr Diabetes 18(7):540–546. 10.1111/pedi.12454 [DOI] [PubMed] [Google Scholar]
- 83.Forlenza GP, Cameron FM, Ly TT et al (2018) Fully closed-loop multiple model probabilistic predictive controller artificial pancreas performance in adolescents and adults in a supervised hotel setting. Diabetes Technol Ther 20(5):335–343. 10.1089/dia.2017.0424 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 84.Ware J, Allen JM, Boughton CK et al (2023) Hybrid closed-loop with faster insulin aspart compared with standard insulin aspart in very young children with type 1 diabetes: a double-blind, multicenter, randomized, crossover study. Diabetes Technol Ther 25(6):431–436. 10.1089/dia.2023.0042 [DOI] [PubMed] [Google Scholar]
- 85.Dovc K, Bergford S, Fröhlich-Reiterer E et al (2023) A comparison of faster insulin aspart with standard insulin aspart using hybrid automated insulin delivery system in active children and adolescents with type 1 diabetes: a randomized double-blind crossover trial. Diabetes Technol Ther 25(9):612–621. 10.1089/dia.2023.0178 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 86.Colmegna P, Cengiz E, Garcia-Tirado J, Kraemer K, Breton MD (2021) Impact of accelerating insulin on an artificial pancreas system without meal announcement: an in silico examination. J Diabetes Sci Technol 15(4):833–41. 10.1177/1932296820928067 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 87.Colmegna P, Diaz CJL, Garcia-Tirado J, DeBoer MD, Breton MD (2024) Adjusting therapy profiles when switching to ultra-rapid lispro in an advanced hybrid closed-loop system: an in silico study. J Diabetes Sci Technol 18(3):676–85. 10.1177/19322968221140401 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 88.Diaz CJL, Colmegna P, Breton MD (2023) Maximizing glycemic benefits of using faster insulin formulations in type 1 diabetes: in silico analysis under open-and closed-loop conditions. Diabetes Technol Ther 25(4):219–30. 10.1089/dia.2022.0468 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 89.Haidar A, Tsoukas MA, Bernier-Twardy S et al (2020) A novel dual-hormone insulin-and-pramlintide artificial pancreas for type 1 diabetes: a randomized controlled crossover trial. Diabetes Care 43(3):597–606. 10.2337/dc19-1922 [DOI] [PubMed] [Google Scholar]
- 90.Dandona P, Chaudhuri A, Ghanim H (2023) Semaglutide in early type 1 diabetes. N Engl J Med 389(10):958–959. 10.1056/NEJMc2302677 [DOI] [PubMed] [Google Scholar]
- 91.Garcia-Tirado J, Farhy L, Nass R et al (2022) Automated insulin delivery with SGLT2i combination therapy in type 1 diabetes. Diabetes Technol Ther 24(7):461–470. 10.1089/dia.2021.0542 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 92.Alva S, Castorino K, Cho H, Ou J (2021) Feasibility of continuous ketone monitoring in subcutaneous tissue using a ketone sensor. J Diabetes Sci Technol 15(4):768–774. 10.1177/19322968211008185 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 93.Weisman A, Bai JW, Cardinez M, Kramer CK, Perkins BA (2017) Effect of artificial pancreas systems on glycaemic control in patients with type 1 diabetes: a systematic review and meta-analysis of outpatient randomised controlled trials. Lancet Diabetes Endocrinol 5(7):501–512. 10.1016/s2213-8587(17)30167-5 [DOI] [PubMed] [Google Scholar]
- 94.van Bon AC, Blauw H, Jansen TJP et al (2024) Bihormonal fully closed-loop system for the treatment of type 1 diabetes: a real-world multicentre, prospective, single-arm trial in the Netherlands. Lancet Digit Health 6(4):e272–e280. 10.1016/s2589-7500(24)00002-5 [DOI] [PubMed] [Google Scholar]
- 95.DeBoer MD, Valdez R, Corbett JP et al (2023) Effect of an automated advice algorithm (CloudConnect) on adolescent-parent diabetes-specific communication and glycemic management: a randomized trial. Diabetes Ther 14(5):899–913. 10.1007/s13300-023-01401-9 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 96.Goyal S, Nunn CA, Rotondi M et al (2017) A mobile app for the self-management of type 1 diabetes among adolescents: a randomized controlled trial. JMIR Mhealth Uhealth 5(6):e82. 10.2196/mhealth.7336 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 97.Bisio A, Anderson S, Norlander L et al (2022) Impact of a novel diabetes support system on a cohort of individuals with type 1 diabetes treated with multiple daily injections: a multicenter randomized study. Diabetes Care 45(1):186–193. 10.2337/dc21-0838 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 98.O’Connor PJ, Sperl-Hillen JM, Fazio CJ, Averbeck BM, Rank BH, Margolis KL (2016) Outpatient diabetes clinical decision support: current status and future directions. Diabet Med 33(6):734–741. 10.1111/dme.13090 [DOI] [PMC free article] [PubMed] [Google Scholar]