An evaluation of commercial radioisotope methods for the determination of folate and vitamin B12.

D W Dawson, I W Delamore, D I Fish, T A Flaherty, A H Gowenlock, L P Hunt, K Hyde, J E MacIver, J A Thornton, H M Waters
PMCID: PMC1146046  PMID: 7381023

Abstract

Five commercial kits for the determination of folate and six kits for the determination of vitamin B12 were investigated. Their performance has been compared with microbiological methods for the two vitamins and with a non-commercial radioisotopic method for B12. The results show the importance of the determination of the reference range for an individual laboratory for each method. The precision of the kits varied appreciably, as did their performance using specimens from patients with different haematological disorders. In particular, certain kits failed to detect all patients with pernicious anaemia. The relative accuracy of the kits was assessed. Various factors which should be taken into account in the final selection of a satisfactory kit are discussed.

Full text

PDF
234

Selected References

These references are in PubMed. This may not be the complete list of references from this article.

  1. Britt R. P., Bolton F. G., Cull A. C., Spray G. H. Experience with a simplified method of radio-isotopic assay of serum vitamin B 12. Br J Haematol. 1969 May;16(5):457–464. doi: 10.1111/j.1365-2141.1969.tb00424.x. [DOI] [PubMed] [Google Scholar]
  2. Cooper B. A., Whitehead V. M. Evidence that some patients with pernicious anemia are not recognized by radiodilution assay for cobalamin in serum. N Engl J Med. 1978 Oct 12;299(15):816–818. doi: 10.1056/NEJM197810122991506. [DOI] [PubMed] [Google Scholar]
  3. Davis R. E., Nicol D. J., Kelly A. An automated method for the measurement of folate activity. J Clin Pathol. 1970 Feb;23(1):47–53. doi: 10.1136/jcp.23.1.47. [DOI] [PMC free article] [PubMed] [Google Scholar]
  4. Hill P. W., Dawson D. W. Evaluation of a commercial radioisotopic kit for folate assays. J Clin Pathol. 1977 May;30(5):449–453. doi: 10.1136/jcp.30.5.449. [DOI] [PMC free article] [PubMed] [Google Scholar]
  5. Johnson I., Guilford H., Rose M. Measurement of serum folate: experience with 75Se-selenofolate radioassay. J Clin Pathol. 1977 Jul;30(7):645–648. doi: 10.1136/jcp.30.7.645. [DOI] [PMC free article] [PubMed] [Google Scholar]
  6. Kubasik N. P., Volosin M. T., Sine H. E. Comparison of commercial kits for radioimmunoassay: III. Radioassay of serum folate. Clin Chem. 1975 Dec;21(13):1922–1926. [PubMed] [Google Scholar]
  7. Matthews D. M., Gunasegaram R., Linnell J. C. Results with radioisotopic assay of serum B12 using serum binding agent. J Clin Pathol. 1967 Sep;20(5):683–686. doi: 10.1136/jcp.20.5.683. [DOI] [PMC free article] [PubMed] [Google Scholar]
  8. McGown E. L., Lewis C. M., Dong M. H., Sauberlich H. E. Results with commercial radioassay kits compared with microbiological assay of folate in serum and whole-blood. Clin Chem. 1978 Dec;24(12):2186–2191. [PubMed] [Google Scholar]
  9. Raven J. L., Robson M. B. Experience with a commercial kit for the radioisotopic assay of vitamin B12 in serum: the Phadebas B12 test. J Clin Pathol. 1974 Jan;27(1):59–65. doi: 10.1136/jcp.27.1.59. [DOI] [PMC free article] [PubMed] [Google Scholar]
  10. Raven J. L., Robson M. B., Morgan J. O., Hoffbrand A. V. Comparison of three methods for measuring vitamin B 12 in serum: radioisotopic, euglena gracilis and Lactobacillus leichmannii. Br J Haematol. 1972 Jan;22(1):21–31. doi: 10.1111/j.1365-2141.1972.tb08783.x. [DOI] [PubMed] [Google Scholar]
  11. Wagstaff M., Broughton A. A simple routine radioisotope method for the estimation of serum vitamin B12 using DEAE cellulose and human serum binding agent. Br J Haematol. 1971 Nov;21(5):581–593. doi: 10.1111/j.1365-2141.1971.tb02719.x. [DOI] [PubMed] [Google Scholar]
  12. Waxman S., Schreiber C., Rose M., Johnson I., Sheppard R., Sumbler K., Keen A., Guilford H. Measurement of serum folate by 75Se-selenofolate radioassay. Results of a multi-institutional clinical trial. Am J Clin Pathol. 1978 Sep;70(3):359–363. doi: 10.1093/ajcp/70.3.359. [DOI] [PubMed] [Google Scholar]