Skip to main content
Journal of Clinical Pathology logoLink to Journal of Clinical Pathology
. 1980 Jun;33(6):571–574. doi: 10.1136/jcp.33.6.571

An evaluation of the Replireader in the identification of Enterobacteriaceae isolated from urine and in the recording of sensitivity tests performed by an agar dilution method.

P M Waterworth
PMCID: PMC1146145  PMID: 6995495

Abstract

In the Replireader system for identifying Enterobacteriaceae, plates of biochemical media are inoculated with a replicator and the results are put into a computer. The machine correctly identified 92x2% of 734 strains of Gram-negative bacilli isolated from urine; it was incorrect in 0x8% and failed to recognise 7%. The Replireader was also used to record the results of sensitivity tests using a plate dilution method in which the drugs were provided by impregnated filter papers (Adapads).

Full text

PDF
571

Selected References

These references are in PubMed. This may not be the complete list of references from this article.

  1. Brown S. D., Washington J. A., 2nd Evaluation of the Repliscan system for identification of Enterobacteriaceae. J Clin Microbiol. 1978 Dec;8(6):695–699. doi: 10.1128/jcm.8.6.695-699.1978. [DOI] [PMC free article] [PubMed] [Google Scholar]

Articles from Journal of Clinical Pathology are provided here courtesy of BMJ Publishing Group

RESOURCES