Abstract
Renal dialysis patients are well known to have a high incidence of hepatitis B carrier state. In studying a group of 63 long-term dialysis patients, 10 were found to be positive for hepatitis B surface antigen by radioimmunoassay (RIA). Surprisingly, however, only three of these RIA positive patients were positive by counter immunoelectrophoresis (CIEP). The discrepancy could not be accounted for by the difference in sensitivity of the two methods. The cause for the negative reactions by CIEP in seven patients was found to be the marked excess surface antigen in these sera which produced false negative results by the postzone phenomenon. After dilution all seven sera were positive by CIEP, requiring a dilution up to 1/20 to produce a positive result. Also, all seven sera were positive by the less sensitive Ouchterlony double diffusion.
Full text
PDF

Images in this article
Selected References
These references are in PubMed. This may not be the complete list of references from this article.
- BLUMBERG B. S., ALTER H. J., VISNICH S. A "NEW" ANTIGEN IN LEUKEMIA SERA. JAMA. 1965 Feb 15;191:541–546. doi: 10.1001/jama.1965.03080070025007. [DOI] [PubMed] [Google Scholar]
- Boulton-Jones J. M., Vick R., Cameron J. S., Black P. J. Immune responses in uremia. Clin Nephrol. 1973 Nov-Dec;1(6):351–360. [PubMed] [Google Scholar]
- Chiaramonte M., Heathcote J., Crees M., Sherlock S. Detection, by three techniques, of hepatitis B surface antigen (HBsAg) and determination of HBsAg and anti-HBs titres in patients with chronic liver disease. Gut. 1977 Jan;18(1):1–6. doi: 10.1136/gut.18.1.1. [DOI] [PMC free article] [PubMed] [Google Scholar]