Skip to main content
Respiratory Care logoLink to Respiratory Care
editorial
. 2024 Oct;69(10):1351–1352. doi: 10.4187/respcare.12438

Reassessing the Role of Heated Humidification During Nebulization: Implications for Clinical Practice

Jie Li 1,
PMCID: PMC11469005  PMID: 39327025

Over 2 decades ago, a series of in vitro and in vivo studies compared aerosol delivery using dry gas versus heated humidified gas during invasive mechanical ventilation.1-5 These studies found that aerosol delivery efficiency was greater with dry gas than with heated humidified gas,1-5 likely due to the increased diameter of aerosol particles caused by absorbing water vapor from the heated humidification. As a result, some clinicians and researchers have recommended turning off humidifiers during nebulization, particularly to optimize the delivery of aerosolized antibiotics for mechanically ventilated patients.6 In clinical practice, 8–37% clinicians have reported turning off humidifiers during aerosol delivery for mechanically ventilated patients.7-9 However, concerns have been raised regarding the potential harms of turning off humidification.

In this issue, Lellouche et al10 critically evaluated the hygrometry of dry gas, heated humidified gas, and the dynamic process of turning off humidifiers. They observed a rapid decrease in humidity to below 30 mg H2O/L after turning off humidifiers,10 while 30 mg H2O/L is the recommended minimum humidity level for mechanically ventilated patients to avoid airway injury.11,12 Their findings underscore the potential risks associated with recommendations to turn off humidifiers in order to enhance aerosol delivery.

Notably, there is a disconnect between these recommendations and the actual findings of more recent studies. It is important to recognize that earlier studies compared dry gas with heated humidified gas1-5 rather than the gas conditions that occur after humidifiers are turned off. Until recently, researchers investigated the effects of interrupting humidification on aerosol delivery efficiency.13,14 Lin et al13 reported that the efficiency of aerosol delivery via pressurized metered-dose inhaler during mechanical ventilation was not improved when humidifiers were turned off compared to aerosol delivery during heated humidification. Similarly, Jacquier et al14 found no improvement in aerosol delivery efficiency after humidifiers were turned off, even after 1 h—a finding corroborated by 3 different laboratories using various small-volume nebulizers, ventilators, humidifiers, and ventilation settings. These findings suggest that the practice of turning off humidifiers to promote aerosol delivery lacks sufficient evidence to support its efficacy.

Additionally, recommendations suggest that humidifiers should be turned back on after nebulization is complete.6 However, in clinical practice, there is a real risk that clinicians might forget to do so. Prolonged inhalation of dry gas can lead to severe complications, such as mucous plugs and airway injury.15 Therefore, after thoroughly reviewing the evidence, a group of international experts reached the consensus that humidifiers should not be turned off for aerosol therapy during mechanical ventilation.16

Some may argue that nebulization alone, even with dry gas, could provide sufficient humidity. To improve aerosol delivery, especially for aerosolized antibiotics, it has been suggested to change ventilator circuits to new sets of dry circuits.6 However, this practice carries significant risks, including alveolar de-recruitment, environmental contamination during circuit disconnection, and an increased risk of pneumonia. The potential benefits of improved aerosol delivery must be weighed against these risks, highlighting the need for further evaluation. Furthermore, Jacquier et al observed that gas humidity quickly decreased from 35–40 mg H2O/L to below 30 mg H2O/L within 5–7 min after turning off humidifiers, regardless of the nebulizer type and placement. With dry gas, nebulization alone could not provide humidity levels exceeding 20 mg H2O/L.14 Given that 30 mg H2O/L is the recommended minimum humidity level for mechanically ventilated patients to avoid airway injury,11,12 it is evident that nebulization alone is insufficient to meet these needs.

Another important clinical implication of the findings by Lellouche et al is the emphasis on the importance of heat in the humidification process. Some may suspect that passover humidifiers without heat could still provide adequate humidity. However, Lellouche et al found that while the water inside a passover humidifier could add some humidity compared to dry gas, the humidity level remained below 10 mg H2O/L.10 This finding highlights the fact that the ability of gas to hold water is temperature dependent. It also serves as a reminder that in clinical scenarios where humidifiers are discontinued, such as during patient transport, the use of a heat-and-moisture exchanger is essential.12

This study not only provides crucial evidence on the hygrometry of humidification but also underscores the importance of critically evaluating existing recommendations and generating new evidence where needed. As respiratory therapists and clinicians, it is imperative that we practice evidence-based medicine, but it is equally important that we critically assess the evidence we rely on.17

Footnotes

Dr Li discloses relationships with Fisher & Paykel Healthcare, Aerogen, MEKICS Co, Vincent, American Association for Respiratory Care, COPD Foundation, Rice Foundation, and Heyer. Dr Li is a section editor for Respiratory Care.

See the Original Study on Page 1239

REFERENCES

  • 1.Diot P, Morra L, Smaldone GC. Albuterol delivery in a model of mechanical ventilation. Comparison of metered-dose inhaler and nebulizer efficiency. Am J Respir Crit Care Med 1995;152(4 Pt 1):1391-1394. [DOI] [PubMed] [Google Scholar]
  • 2.Fink JB, Dhand R, Duarte AG, Jenne JW, Tobin MJ. Aerosol delivery from a metered-dose inhaler during mechanical ventilation. An in vitro model. Am J Respir Crit Care Med 1996;154(2 Pt 1):382-387. [DOI] [PubMed] [Google Scholar]
  • 3.Fink JB, Dhand R, Grychowski J, Fahey PJ, Tobin MJ. Reconciling in vitro and in vivo measurements of aerosol delivery from a metered-dose inhaler during mechanical ventilation and defining efficiency-enhancing factors. Am J Respir Crit Care Med 1999;159(1):63-68. [DOI] [PubMed] [Google Scholar]
  • 4.Miller DD, Amin MM, Palmer LB, Shah AR, Smaldone GC. Aerosol delivery and modern mechanical ventilation: in vitro/in vivo evaluation. Am J Respir Crit Care Med 2003;168(10):1205-1209. [DOI] [PubMed] [Google Scholar]
  • 5.Ari A, Areabi H, Fink JB. Evaluation of aerosol generator devices at 3 locations in humidified and non-humidified circuits during adult mechanical ventilation. Respir Care 2010;55(7):837-844. [PubMed] [Google Scholar]
  • 6.Monsel A, Torres A, Zhu Y, Pugin J, Rello J, Rouby JJ; on behalf of the European Investigators Network for Nebulized Antibiotics in Ventilator-associated Pneumonia (ENAVAP). Nebulized antibiotics for ventilator-associated pneumonia: methodological framework for future multi-center randomized controlled trials. Curr Opin Infect Dis 2021;34(2):156-168. [DOI] [PubMed] [Google Scholar]
  • 7.Sun Q, Chang W, Liu X, Xie J, Qiu H, Yang Y, et al. Aerosol therapy during mechanical ventilation in intensive care units: a questionnaire-based survey of 2,203 ICU medical staff in China. J Intensive Med 2022;2(3):189-194. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Ehrmann S, Roche-Campo F, Bodet-Contentin L, Razazi K, Dugernier J, Trenado-Alvarez J, et al. ; AT@ICU Study Group. Aerosol therapy in intensive and intermediate care units: prospective observation of 2808 critically ill patients. Intensive Care Med 2016;42(2):192-201. [DOI] [PubMed] [Google Scholar]
  • 9.Lyu S, Li J, Yang L, Du X, Liu X, Chuan L, et al. The utilization of aerosol therapy in mechanical ventilation patients: a prospective multi-center observational cohort study and a review of the current evidence. Ann Transl Med 2020;8(17):1071. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10.Lellouche F, Bouchard PA. Consequences of pausing heated humidification during invasive ventilation. Respir Care 2024;69(10):1239-1244. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.International Organization for Standardization. Particular requirements for basic safety and essential performance of respiratory humidifying equipment (ISO 80601–2-74:2017[en]). Geneva: International Organization for Standardization; 2017. [Google Scholar]
  • 12.Restrepo RD, Walsh BK; American Association for Respiratory Care. Humidification during invasive and noninvasive mechanical ventilation. Respir Care 2012;57(5):782-788. [DOI] [PubMed] [Google Scholar]
  • 13.Lin HL, Fink JB, Zhou Y, Cheng YS. Influence of moisture accumulation in in-line spacer on delivery of aerosol using metered-dose inhaler during mechanical ventilation. Respir Care 2009;54(10):1336-1341. [PubMed] [Google Scholar]
  • 14.Jacquier S, Lin HL, Li J, Sheridan CA, Karabelas P, Liu JF, et al. Effect of interrupting heated humidification on nebulized drug delivery efficiency, temperature, and absolute humidity during mechanical ventilation: a m ulti-lab in vitro study. J Aerosol Med Pulm Drug Deliv 2024;37(3):115-124. [DOI] [PubMed] [Google Scholar]
  • 15.Williams R, Rankin N, Smith T, Galler D, Seakins P. Relationship between the humidity and temperature of inspired gas and the function of the airway mucosa. Crit Care Med 1996;24(11):1920-1929. [DOI] [PubMed] [Google Scholar]
  • 16.Li J, Liu K, Lyu S, Jing G, Dai B, Dhand R, et al. Aerosol therapy in adult critically ill patients: a consensus statement regarding aerosol administration strategies during various modes of respiratory support. Ann Intensive Care 2023;13(1):63. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17.Hess DR. Evidence-based respiratory care. Respir Care 2021;66(7):1105-1119. [DOI] [PubMed] [Google Scholar]

Articles from Respiratory Care are provided here courtesy of Mary Ann Liebert, Inc.

RESOURCES