Abstract
Wearables offer a promising solution for enhancing remote monitoring (RM) of heart failure (HF) patients by tracking key physiological parameters. Despite their potential, their clinical integration faces challenges due to the lack of rigorous evaluations. This review aims to summarize the current evidence and assess the readiness of wearables for clinical practice using the Medical Device Readiness Level (MDRL). A systematic search identified 99 studies from 3112 found articles, with only eight being randomized controlled trials. Accelerometery was the most used measurement technique. Consumer-grade wearables, repurposed for HF monitoring, dominated the studies with most of them in the feasibility testing stage (MDRL 6). Only two of the described wearables were specifically designed for HF RM, and received FDA approval. Consequently, the actual impact of wearables on HF management remains uncertain due to limited robust evidence, posing a significant barrier to their integration into HF care.
Subject terms: Heart failure, Hardware and infrastructure, Diagnosis
Introduction
Heart failure (HF) impacts around 63 million individuals worldwide, significantly affecting patients and their caregivers1. Moreover, it places a significant strain on the healthcare system, primarily due to the necessity for frequent outpatient visits and recurrent hospitalizations2. This escalation in demand for services and resources is especially challenging in today’s healthcare environment, which is already facing issues with limited capacity, staff shortages and high workload3. Remote monitoring (RM) has been proposed as a solution to reduce this burden, with a recent meta-analysis highlighting its effectiveness showing that both invasive and non-invasive RM technologies can significantly lower mortality and hospitalization rates among HF patients4. However, the landscape of RM modalities is remarkably diverse, ranging from non-invasive blood pressure measurements to invasive hemodynamic sensors (e.g., CardioMems and Cordella device)4–6. Moreover, the adoption of RM technologies faces obstacles due to the absence of standardized methodologies and external validation7. These challenges contribute to a notable gap in determining the appropriate technology for specific patient categories. As a consequence, the present HF guidelines offer a limited endorsement for incorporating RM in the care of HF patients8.
Simultaneously to this development, there has been a rapid surge in a large array of (commercially available) health technology, including wearable devices ranging from smartwatches, rings and accessories incorporated into clothing9. Leveraging these wearable devices to monitor physiological variables offers a personalized and empowering experience for patients, that might become an important chain in modern HF-management. However, the efficacy of most of these wearables have poorly been studied10. Thus, as physicians increasingly embrace wearables for monitoring, critical questions persist regarding their safety, readiness, and validity11. Consequently, regulatory bodies have taken steps to ensure the safe and effective application of these devices for medical purposes. For instance, the European Union’s Medical Device Regulation (MDR) and U.S. Food and Drug Administration (FDA) have established classifications and guidelines to regulate wearable devices that may have medical applications12,13. Addressing these knowledge gaps and informing the HF community about the integration of these devices in clinical practice is of paramount importance. Therefore, we performed a comprehensive scoping review to provide an overview of all wearable devices currently being tested and used in HF management. This review aims to elucidate their functionalities, applications, and evaluate their developmental progress in the HF population by obtaining their MDR/FDA classification and by using the Medical Device Readiness Level (MDRL) framework14.
Results
A total of 3112 articles were identified and after screening for eligibility, 99 studies were included, involving 13.879 patients (Fig. 1)15–113. A detailed overview of the characteristics of the included studies is provided in Table 1 and Supplementary Table 1. Over time, the number of studies including wearables has steadily increased. Most studies included wrist-worn devices (n = 43, 43.4%), such as accelerometers, followed by vests (n = 23, 23.2%) and hip-worn devices (n = 20, 20.2%). (Table 2) The majority of these studies were designed as prospective studies, accounting for 91.9% of the included studies, and were predominantly conducted in the United States, Western Europe, and Japan. Our search identified only 8 RCTs that involved wearable technology26,28,36,38,41,45,78,107. However, no articles were found that tested the effectiveness of a wearable device in a large randomized setting.
Fig. 1. Description of the Medical Device Readiness Level (MDRL).
In this figure, each stage of the Medical Device Readiness Level (MDRL) is briefly described. A more detailed description of each stage and MDRL itself can be found in the article of Ruiz Seva et al.14.
Table 1.
Study characteristics
| Authors, year | n | Country | Study type | Device | Type of device | Measure techniques | Functions | Medical grade | MDRL |
|---|---|---|---|---|---|---|---|---|---|
| Single measurement technique | |||||||||
| Alosco et al.18 | 96 | USA | OBS | ActiGraph GT1M | Wrist-worn device | ACC | PAa | MDR C I, FDA C II | 6 |
| Alosco et al.16 | 65 | USA | OBS | ActiGraph GT1M | Wrist-worn device | ACC | PAa | MDR C I, FDA C II | 6 |
| Alosco et al.17 | 57 | USA | OBS | Actigraph GT1M | Wrist-worn device | ACC | PAa | MDR C I, FDA C II | 6 |
| Baril et al.21 | 50 | CAN | OBS | Fitbit Flex | Wrist-worn device | ACC | PA | - | 6 |
| Blomqvist et al.24 | 605 | SWE | OBS | ActiGraph GT9X | Wrist-worn device | ACC | PAa | MDR C I, FDA C II | 6 |
| Braun et al.25 | 61 | DEU | OBS | wearable IMU (inertial measurement units) system (Rehagait®, Hasomed, Magdeburg, Germany | Ankles-worn | ACC | PA | - | 6 |
| Butler et al.28 | 69 | USA | RCT | ActiGraph GT9X | Wrist-worn device | ACC | PAa | MDR C I, FDA C II | 6 |
| Da Silva et al.30 | 16 | BRA | OBS | ActiGraph GT3X | Wrist-worn device | ACC | PAa | MDR C I, FDA C II | 6 |
| Dibben et al.35 | 22 | GBR | OBS | GENEActiv, Activinsights Ltd. | Wrist- and waist-worn device | ACC | PA | - | 5 |
| Dibben et al.36 | 247 | GBR | RCT | GENEActiv, Activinsights Ltd. | Wrist-worn device | ACC | PA | - | 7 |
| Dontje et al.37 | 68 | NLD | OBS | SenseWear® Pro3 Armband | Wrist-worn device | ACC | PA | - | 6 |
| Edwards et al.39 | 190 | USA | OBS | ActiGraph AM-7164 | Wrist-worn device | ACC | PAa | MDR C I, FDA C II | 6 |
| Evangalista et al.41 | 38 | USA | RCT | Sportline Pedometer Model 330 | Hip-worn | ACC | PA | - | 6 |
| Floegel et al.43 | 27 | USA | OBS | ActivPAL3c™ and Tractivity® | Wrist-worn device | ACC | PA | - | 6 |
| Floegel et al.42 | 29 | USA | OBS | ActivPAL3c™ and Tractivity® | Wrist-worn device | ACC | PA | - | 6 |
| Fulcher et al.44 | 93 | USA | OBS | ActiGraph GT1M | Wrist-worn device | ACC | PAa | MDR C I, FDA C II | 6 |
| German et al.47 | 58 | USA | OBS | Kenz Lifecorder Ex | Hip-worn | ACC | PA | - | 6 |
| Güder et al.48 | 12 | DEU | OBS | Polar M430 | Wrist-worn device | ACC | PA | - | 6 |
| Holber et al.52 | 531 | USA | OBS | SenseWear® Pro3 | Wrist-worn device | ACC | PA | - | 7 |
| Howie-Esquivel et al.53 | 32 | USA | OBS | Micro Care Timeliness Monitors | Hip- and ankle worn | ACC | PA | - | 5 |
| Izawa et al.59 | 261 | JPN | OBS | Kenz Lifercorder | Hip-worn | ACC | PA | - | 6 |
| Izawa et al.58 | 477 | JPN | OBS | Kenz Lifercorder | Hip-worn | ACC | PA | - | 6 |
| Jehn et al.61 | 55 | DEU | OBS | StayHealthy Inc. RT3, AiperMotion 300 PfH, OMRON Walking Style Pro | Wrist-worn device | ACC | PA | - | 5 |
| Jehn et al.60 | 710 | DEU | OBSb | AiperMotion 300 PfH | Hip-worn | ACC | PA | - | 6 |
| Klompstra et al.66 | 32 | SWE | PT | DirectLife TracmorD, Philips | Wrist-worn device | ACC | PA | - | 6 |
| Klompstra et al.65 | 64 | SWE | OBSb | ActiGraph GT9X | Hip-worn | ACC | PAa | MDR C I, FDA C II | 6 |
| Li et al.69 | 1140 | USA | OBS | Actical, Philips | Patch | ACC | PA, Sleepa | FDA C II | 6 |
| Lin et al.71 | 124 | GBR | OBS | Axivity AX3 (Axivity Ltd) | Vest | ACC | PA | - | 6 |
| McCarthy et al.73 | 20 | USA | OBS | New Lifestyles NL-800 | Hip-worn | ACC | PA | - | 6 |
| Melczer et al.74 | 17 | HUN | OBS | ActiGraph GT3X | Wrist-worn device | ACC | PAa | MDR C I, FDA C II | 6 |
| Melin et al.75 | 60 | SWE | OBS | ActiGraph GT3X | Wrist-worn device | ACC | PAa | MDR C I, FDA C II | 6 |
| Miyahara et al.76 | 70 | JPN | OBS | Omron HJA-350IT | Hip-worn | ACC | PA | - | 6 |
| Nelson et al.79 | 93 | USA | OBS | Actigraph GT3X | Wrist-worn device | ACC | PAa | MDR C I, FDA C II | 6 |
| O’Donnell et al.80 | 596 | GBR | OBS | Axivity AX3 | Wrist-worn device | ACC | PA | - | 6 |
| Okwose et al.81 | 20 | GBR | OBS | Omron HJ-321-E | Hip-worn | ACC | PA | - | 6 |
| Omar et al.82 | 63 | DNK | OBS | Actigraph GT3X | Wrist-worn device | ACC | PAa | MDR C I, FDA C II | 6 |
| Pozehl et al.85 | 182 | USA | OBS | ActiGraph GT3X | Wrist-worn device | ACC | PAa | MDR C I, FDA C II | 6 |
| Prescher et al.86 | 155 | DEU | OBSb | AiperMotion 300 PfH | Wrist-worn device | ACC | PA | - | 6 |
| Radhakrishnan et al.87 | 10 | USA | OBS | Withings Go activity tracker | Wrist-worn device | ACC | PA | - | 6 |
| Rullman et al.89 | 94 | SWE | OBS | Actigraph GT3X | Wrist-worn device | ACC | PAa | MDR C I, FDA C II | 6 |
| Schmidt et al.92 | 22 | PRT | OBS | ActiGraph GT3X | Wrist-worn device | ACC | PAa | MDR C I, FDA C II | 6 |
| Schoemaker et al.97 | 22 | USA | OBS | Stayhealthy Inc. RT3 | Hip-worn | ACC | PA | - | 5 |
| Schwendinger et al.93 | 56 | USA | OBS | GENEActiv, Activinsights Ltd. | Wrist-worn device | ACC | PA | - | 6 |
| Shen et al.95 | 40 | CHN | OBS | ActiGraph GT3X | Wrist-worn device | ACC | PAa | MDR C I, FDA C II | 6 |
| Shiraishi et al.96 | 31 | JPN | OBS | Active Style Pro HJA-750C | Waist-Worn | ACC | PA | - | 6 |
| Snipelisky et al.99 | 110 | USA | OBSb | Kinetic Activity Monitors | Hip-worn | ACC | PA | - | 6 |
| van den Berg-Emons et al.104 | 5 | NLD | OBS | Custom build system | Body- and hip-worn | ACC | PA | - | 5 |
| van den Berg-Emons et al.105 | 36 | NLD | OBS | Custom build system | Body- and hip-worn | ACC | PA | - | 6 |
| Vetrovsky et al.106 | 26 | CZE | OBS | Garmin vívofit | Wrist- and hip-worn device | ACC | PA | - | 6 |
| Vetrovsky et al.107 | 26 | CZE | RCT | Garmin vívofit, Actigraph wGT3X-BT | Wrist-worn device | ACC | PAa | ActiGraph: FDA C II, MDR C I | Garmin: 6, ActiGraph: 6 |
| Waring et al.109 | 50 | USA | OBS | ActiGraph GT3X+ | Wrist-worn device | ACC | PAa | MDR C I, FDA C II | 6 |
| Witham et al.110 | 82 | GBR | OBS | Stayhealthy Inc. RT3 | Hip-worn | ACC | PA | - | 6 |
| Young et al.113 | 100 | USA | OBS | Actigraph wGT3X-BT | Hip-worn | ACC | PAa | MDR C I, FDA C II | 6 |
| Alvarez-Garcia et al.19 | 100 | USA | RCT | ReDS™ Wearable System | Vest | B-IMP | CONG | FDA C II | 6 |
| Amir et al.20 | 50 | ISR | OBS | ReDS™ Wearable System | Vest | B-IMP | CONG | FDA C II | 6 |
| Bensimhon et al. 22 | 108 | USA | PT | ReDS™ Wearable System | Vest | B-IMP | CONG | FDA C II | 6 |
| Curtain et al.29 | 66 | GBR | OBS | Sensinel CPM System by Analog Devices, Inc. | Patches | B-IMP | CONG | - | 6 |
| Lala et al.68 | 220 | USA | OBS | ReDS™ Wearable System | Vest | B-IMP | CONG | FDA C II | 6 |
| Polcz et al.84 | 155 | USA | OBS | ZOE® | vest | B-IMP | CONG | - | 6 |
| Ueno et al.103 | 12 | JPN | OBS | ReDS™ Wearable System | Vest | B-IMP | CONG | FDA C II | 6 |
| Aamodt et al.15 | 10 | NOR | OBS | CardioSet Edema Guard Monitor | Vest | B-IMP | CONG | - | 6 |
| Guo et al.49 | 66 | CHN | OBS | BECG1200A, Thoth Medical Technology Co | Patch | ECG | HR | - | 6 |
| Kikuchi et al.64 | 10 | JPN | OBS | hitoe®, TORAY | Vest | ECG | HR, Rhythm | - | 5 |
| Dagan et al.31 | 29 | ISR | OBS | BB-613WP | Wrist-worn device | PPG | HR, SpO2, Respiratory Rate, PTT | - | 6 |
| Lin et al.72 | 40 | TWN | POC | Prototype of Smart Clothes | Wrist-worn device | SCG | Cardiac Time Intervals | - | 4 |
| Combined measurement techniques | |||||||||
| Yates et al.112 | 29 | USA | OBS | Actiheart, Camntech limited | Wrist-worn device | ACC, ECG | PAa | FDA C II | 7 |
| Smeets et al.98 | 36 | BEL | OBS | Wearable Bioimpedance Monitor (IMEC) | Vest | ACC, ECG, B-IMP | CONG | - | 5 |
| Stehlik et al.102 | 100 | USA | OBS | VitalPatch® RTM (VitalConnect®) | Patch | ACC, ECG, B-IMP, Thermometer | PA, HRa, HRV, RR interval, Skin humidity | FDA C II | Insufficient information |
| Kovisto et al.67 | 20 | FIN | POC | Custom-build multi-sensor device | Multi-sensor device | ACC, ECG, SCG | HR, Rhythm, CAB | - | All: 3 |
| Blockhaus et al.23 | 140 | DEU | OBS | WCD ZOLL LifeVest® | Vest | ACC, ECG, SCG | PA | FDA C III | 6 |
| Burch et al.26 | 197 | USA | RCT | WCD ZOLL LifeVest® | Vest | ACC, ECG, SCG | PA, HRa | FDA C III | PA:6, HR: 9 |
| Burkhoff et al.27 | 1066 | USA | OBS | WCD ZOLL LifeVest® | Vest | ACC, ECG, SCG | HRa, CAB | FDA C III | HR: 9, CAB: 5 |
| Erath et al.40 | 671 | USA | OBS | WCD ZOLL LifeVest® | Vest | ACC, ECG, SCG | HRa, Rhythma, CAB | FDA C III | HR, Rhythm: 9; CAB: 5 |
| Garcia et al.45 | 1013 | FRA | OBS | WCD ZOLL LifeVest® | Vest | ACC, ECG, SCG | HRa | FDA C III | 9 |
| Hillmann et al.51 | 276 | DEU | OBS | WCD ZOLL LifeVest® | Vest | ACC, ECG, SCG | PA, HRa | FDA C III | PA:6, HR: 9 |
| Iliodromitis et al.54 | 77 | DEU | OBS | WCD ZOLL LifeVest® | Vest | ACC, ECG, SCG | PA | FDA C III | 6 |
| Jungbauer et al.62 | 1353 | DEU | OBS | WCD ZOLL LifeVest® | Vest | ACC, ECG, SCG | HRa | FDA C III | 9 |
| Röger et al.88 | 105 | DEU | OBS | WCD ZOLL LifeVest® | Vest | ACC, ECG, SCG | HRa | FDA C III | 9 |
| Mlakar et al.77 | 24 | SVN | OBS | prototype of multi- sensor wearable | Multi-sensor device | ACC, ECG, Thermometer | PA, HR, Rhythm, EE, Temperature, skin-humidity | - | Insufficient information |
| Solar et al.101 | NA | ESP | POC | Wearable platform | Body- and hip-worn | ACC, ECG, Thermometer, Humidity sensor | PA, HR, HRV, Temperature, Skin humidity, Potassium, EE | - | 4 |
| Deka et al.33 | 30 | USA | OBS | Fitbit Charge HR | Wrist-worn device | ACC, PPG | PA, HRa | MDR C IIa, FDA C II | PA: 6 HR: 7 |
| Dorsch et al.38 | 83 | USA | RCT | Fitbit Charge 2 | Wrist-worn device | ACC, PPG | PA, HRa | MDR C IIa, FDA C II | All functions: 6 |
| Gardner et al.46 | 39 | USA | OBS | Nonin WristOx2 model 3150, Actisleep+ | Wrist-worn device | ACC, PPG | PAa, Sleep, SpO2a | ActiGraph: FDA CII, MDR C I; Nonin: FDA C II | All functions: 5 |
| Herkert et al.50 | 19 | NLD | OBS | Fitbit Charge 2 & Mio Slice | Wrist-worn device | ACC, PPG | PA, HRa, EE | Fitbit: MDR C IIa, FDA C II | Fitbit: PA: 6, HR: 6, EE: 5; MioSlice: PA: 6, HR: 6, EE: 5 |
| Nagatomi et al.78 | 30 | JPN | RCT | Fitbit Inspire HR | Wrist-worn device | ACC, PPG | PA, HRa | MDR C IIa, FDA C II | PA: 6 HR: 6 |
| Sohn et al.100 | 20 | USA | OBS | Fitbit Charge 2 | Wrist-worn device | ACC, PPG | PA | MDR C IIa, FDA C II | 6 |
| Vetrovsky et al.108 | 29 | CZE | OBS | Withings Go, Fitbit Charge 2, Garmin vivofit, Garmin vivofit 3, Omron HJ-322U-E, smartLAB walk+ | Wrist-worn device | ACC, PPG | PA | Fitbit: MDR C IIa, FDA C II | All devices: 6 |
| Iqbal et al.57 | NA | USA | POC | Custom-built wearable belt | Hip-worn | ACC, PPG, ECG, B-IMP | PA, HR, Sleep, Rhythm | - | All functions: 4 |
| Iqbal et al.56 | 10 | USA | POC | Custom-built wearable belt | Hip-worn | ACC, PPG, ECG, B-IMP | PA, HR, Sleep, Rhythm | - | All functions: 5 |
| Wong et al.111 | 20 | CHN | OBS | Biofourmis Everion | Armband | ACC, PPG, ECG, Thermometer, Barometer | PA, HR, HRV, Temperature, Blood pulse wave, SpO2, Respiratory Rate | - | Insufficient information |
| Darling et al.32 | 106 | USA | OBS | Philips fluid accumulation vest | Vest | B-IMP, ECG | CONG | - | 6 |
| Pan et al.83 | 37 | USA | OBS | Philips fluid accumulation vest | Vest | B-IMP, ECG | CONG | - | 5 |
| Sanchez-Perez et al.90 | 24 | USA | POC | Multimodal sensing system | Multi-sensor device | B-IMP, microphone | CONG, Rhythm | - | All functions: 5 |
| Li et al.70 | 76 | CHN | OBS | WENXIN® device | Patch | ECG, Microphone | Rhythm, CAB | - | All functions: 5 |
| Inan et al.55 | 45 | USA | OBS | Prototype sensing patch | Patch | ECG, SCG | HF status (compensated vs. decompensated) | - | 4 |
| Shandhi et al.94 | 59 | USA | OBS | Custom-built wearable patch | Patch | ECG, SCG | VO2 | - | 5 |
| Kaneko et al.63 | 49 | JPN | OBS | AUDICOR AM-RT | Patch | PPG, ECG | CONG, HR, CAB | - | All functions: 6 |
| Di Rienzo et al.34 | NA | ITA | POC | SeisMote | Vest | PPG, ECG, SCG | HR, HRV, PTT | - | All functions: 4 |
| Savoldelli et al.91 | 15 | ITA | POC | Hodwy Senior of Comftech© | Vest | PPG, ECG, Thermometer | HR, Rhythm, Temperature | - | All functions: 5 |
NOR Norway, USA United States of America, ISR Israel, SWE Sweden, DEU Germany, AUT Austria, BRA Brazil, ITA Italy, GBR United Kingdom, NLD The Netherlands, FRA France, CHN China, JPN Japan, FIN Finland, TWN Taiwan, SVN Slovenia, DNK Denmark, PRT Portugal, CZE Czech Republic, OBS Observational study, RCT Randomized controlled trial, POC Proof-of-concept, PT Pilot trial, ACC Accelerometry, B-IMP Bio-impedance, ECG Electrocardiography, SCG Seismocardiography, PPG Photoplethysmography, PA Physical activity, HR Heart rate, HRV Heart rate variability, PTT Pulse transit time, EE Energy expenditure, CONG Pulmonary Congestion, (C)HF (Chronic) heart failure, MDR Medical device regulations, FDA Food and drug administration.
aThese functions are medical graded.
bThese studies are sub-studies of randomized controlled trials in which a wearable is used as measuring device in both the intervention and the control arm.
Table 2.
Summarized information per measurement technique
| Total | Accelerometer | Bio-impedance | ECG | PPG | SCG | Thermometer | Other techniquesc | |
|---|---|---|---|---|---|---|---|---|
| Single measurement techniques | N = 65 | N = 53 | N = 8 | N = 2 | N = 1 | N = 1 | N = 0 | N = 0 |
| Type of device, n (%)a | ||||||||
| Wrist-worn device | 35 (53.8) | 33 (62.3) | - | - | 1 (100) | 1 (100) | - | - |
| Hip-worn device | 17 (26.2) | 12 (22.6) | - | - | - | - | - | - |
| Vest | 9 (13.8) | 1 (1.8) | 1 (12.5) | 1 (50.0) | - | - | - | - |
| Patch(es) | 3 (4.6) | 1 (1.8) | 7 (87.5) | 1 (50.0) | - | - | - | - |
| Multi-sensor device | - | - | - | - | - | - | - | - |
| Other typesb | 6 (9.2) | 5 (9.4) | - | - | - | - | - | - |
| Study type, n (%) | ||||||||
| Randomized controlled trials | 5 (7.7) | 4 (7.5) | 1 (12.5) | - | - | - | - | - |
| Observational studies | 57 (87.7) | 48 (90.5) | 6 (75.0) | 2 (100) | 1 (100) | - | - | - |
| Proof-of-concept | 1 (1.5) | - | - | - | - | 1 (100) | - | - |
| Pilot trial | 2 (3.1) | 1 (1.8) | 1 (12.5) | - | - | - | - | - |
| Geographical region, n (%) | ||||||||
| Asia | 9 (13.8) | 5 (9.4) | 1 (12.5) | 2 (100) | - | 1 (100) | - | - |
| Europe | 26 (40.0) | 24 (45.3) | 2 (25.0) | - | - | - | - | - |
| North America | 27(41.5) | 23 (43.4) | 4 (50.0) | - | - | - | - | - |
| Latin America | 1 (1.5) | 1 (1.8) | - | - | - | - | - | - |
| Middle Eastern | 2 (3.1) | - | 1 (12.5) | - | 1 (100) | - | - | - |
| Complex measurement techniques | N = 34 | N = 25 | N = 7 | N = 26 | N = 13 | N = 13 | N = 5 | N = 4 |
|---|---|---|---|---|---|---|---|---|
| Type of device, n (%)a | ||||||||
| Wrist-worn device | 8 (23.5) | 8 (32.0) | - | 1 (3.8) | 7 (53.8) | - | - | - |
| Hip-worn device | 3 (8.8) | 3 (12.0) | 2 (28.6) | 3 (11.5) | 2 (15.4) | - | 1 (20.0) | 1 (25.0) |
| Vest | 14 (41.2) | 10 (40.0) | 3 (42.9) | 14 (53.8) | 2 (15.4) | 10 (76.9) | 1 (20.0) | - |
| Patch(es) | 5 (14.8) | 1 (4.0) | 1 (14.3) | 5 (19.2) | 1 (7.7) | 2 (15.4) | 1 (20.0) | 1 (25.0) |
| Multi-sensor device | 3 (8.8) | 2 (8.0) | 1 (14.3) | 2 (7.7) | - | 1 (7.7) | 1 (20.0) | 1 (25.0) |
| Other typesb | 2 (5.9) | 2 (8.0) | - | 1 (3.8) | 1 (7.7) | - | 2 (40.0) | 2 (50.0) |
| Study type, n (%) | ||||||||
| Randomized controlled trials | 3 (8.8) | 3 (12.0) | - | 1 (3.8) | 2 (15.4) | 1 (7.7) | - | - |
| Observational studies | 24 (70.6) | 18 (72.0) | 4 (57.1) | 19 (73.1) | 7 (53.8) | 2 (15.4) | 3 (60.0) | 2 (50.0) |
| Proof-of-concept | 7 (20.6) | 4 (10.0) | 3 (42.9) | 6 (23.1) | 4 (30.8) | 10 (76.9) | 2 (40.0) | 2 (50.0) |
| Pilot trial | - | - | - | - | - | - | - | - |
| Geographical region, n (%) | ||||||||
| Asia | 4 (11.8) | 2 (8.0) | - | 3 (11.5) | 3 (23.1) | - | 1 (20.0) | 2 (50.0) |
| Europe | 14 (41.2) | 12 (48.0) | 1 (14.3) | 12 (46.2) | 4 (30.8) | 8 (61.5) | 3 (60.0) | 1 (25.0) |
| North America | 16 (47.1) | 11 (44.0) | 6 (85.7) | 11 (42.3) | 6 (46.2) | 5 (38.5) | 1 (20.0) | 1 (25.0) |
| Latin America | - | - | - | - | - | - | - | - |
| Middle Eastern | - | - | - | - | - | - | - | - |
PPG Photoplethysmography, ECG Electrocardiography, SCG Seismocardiography.
aType device: In some more than one wearable is used, and therefore this percentage does in some techniques not add up to 100%.
bOther types include: ankle-worn devices (n = 2), a bracelet worn on the upper arm (n = 1), body-worn devices (n = 3), and waist-worn devices (n = 3).
cOther techniques include: barometry (n = 1), humidity sensor (n = 1), microphone (n = 2).
Single measurement techniques
In 65 studies included in this review, a wearable device using a single measurement technique was employed (Table 1). Among these, 53 used accelerometery for measuring PA, with the devices predominantly worn on the wrist (n = 33, 62.3%) or hip (n = 12, 22.6%). Notably, only the ActiGraph devices – the most frequent used accelerometery wearable (n = 53, 81.5%) - have a medical grade for measuring PA (MDR Class I and FDA Class II clearance). Most of these single measurement techniques wearables were considered having a MDRL of 6, indicating that most are in the feasibility phase for the intended purpose (Table 1 and Fig. 2). Most of these studies were observational in nature (90.1%) and only five were RCTs (7.7%). However, none of these studies primarily focused on the efficacy of RM through the use of accelerometers as they were either embedded in a home-based rehabilitation intervention or were part of a substudy in a larger clinical trial. A substantial number of included studies employing wearable accelerometers have demonstrated that HF patients exhibit lower PA levels compared to the healthy population, and experience worse outcomes when their PA is reduced18,37,58,75,76,80,82,86,109. Other studies have shown that PA, as measured by accelerometery, is associated with disease severity, as well as cognitive and executive function16,17,25,44,71,99,104. Additionally, three studies used accelerometers as part of a lifestyle intervention to motivate HF patients to increase their exercise81,87,107.
Fig. 2. PRISMA flowchart.
The flow diagram shows the number of studies identified and in- and excluded at the different stages of this scoping review.
Apart from accelerometery, another technique that can be used in RM for HF is bio-impedance, which is briefly explained in Supplementary Table 2. This technique, which measures pulmonary congestion is used in eight studies (7.9%). The majority of these studies employed the ReDS ™ Wearable system, currently the only wearable with a medical grade (FDA Class II) for measuring pulmonary congestion19,20,22,68,103. Most studies using this technique were observational, except for one RCT and one pilot trial. Within the relatively small RCT (n = 100), a ReDS-guided treatment strategy was assessed in a hospital environment showcasing a reduction in clinical events at 30 days19. Currently, only one study of limited size (n = 10), described the use of this wearable in a remote environment, concluding that the use of such device is feasible in the first 30 days after discharge, and improves self-care behavior15. Consequently, the MDRL was estimated at 6 for all devices that rely solely on bio-impedance.
In two other studies, wearables that solely use ECG signals to measure heart rate and rhythm were utilized. In one study (n = 66), a patch (MDRL: 6) designed for extended wear was utilized as part of a Hospital-Community-Family-Based Telehealth Program49. In the other study (n = 11), the feasibility and safety of a telerehabilitation program were assessed in which a wireless ECG vest (MDRL: 5) was used to remotely monitor the HF patient during a cardiac rehabilitation session64. Lastly, one wearable, a prototype wearable with MDRL 4, used seismocardiography as technique to measure cardiac time intervals, which could contain valuable for HF monitoring, in a proof-of-concept study72.
Complex measurement techniques
A total of 34 studies (34.3%) utilized wearable devices or a combination of wearables employing multiple measurement techniques. Most of the studies included were of observational nature (n = 24, 70.6%), with only three being RCTs (8.8%). Among these studies, the WCD Zoll LifeVest® (n = 9, 26.5%) and smartwatches from various brands (n = 9, 26.5%) were the most commonly used wearables. In the studies where the WCD Zoll LifeVest® was utilized, heart rate was the most often measured function23,26,27,40,45,51,54,62,88. The MDRL for the heart rate function was classified as 9, reflecting its core functionality in detecting life-threatening arrythmias and the ability to deliver therapy if necessary (medical grade: FDA Class III). This device can also be used to measure physical activity (MDRL: 6) and cardiac acoustic biomarkers (MDRL: 5)23,26,27,40,51,54. Multiple studies describe that the LifeVest can be used for RM of HF and thereby potentially improve clinical outcomes23,27,62 and provide valuable information regarding titration and treatment response of GDMT51,54,62.
Smartwatches and other wrist-worn devices form another group of wearables that utilize a combination of measurement techniques. All these devices incorporate accelerometers combined with either PPG, ECG or both. Among these wearables, Fitbit devices were the most frequently used (n = 6, 17.6%) and have a medical grade (MDR Class IIa, FDA Class II) for heart rate measurement. Within this group of devices, the MDRL ranged from 5 to 7, with the lowest MDRL for the functionality energy expenditure in both the Fitbit Charge 2 and Mio Slice devices, and the highest for heart rate in the Fitbit Charge HR. In studies employing these wrist-worn devices, most have demonstrated utility for clinical purposes in HF patients108, though these devices lacked sufficient accuracy in estimating the patients’ energy expenditure50. Additionally, other studies indicate that adherence to and usability of these devices are generally high46,100. There are a limited number of studies focusing on enhancing physical activity (PA) in HF patients33,78. One study focused on self-monitoring and self-management in recently admitted HF patients, which showed improved quality of life in the short-term; however, this effect did not persist to 12 weeks38.
Lastly, there are wearables that are specifically designed for the monitoring of HF patients. Those devices are described and used in 11 studies and ranged from prototype devices to more developed devices with MDRLs ranging from 3 to 6. These wearables utilize various combinations of measurement techniques as described in Table 1. The studies in which these less developed devices are described focus on the ability of measuring variables relevant to HF, such as pulmonary congestion and potassium blood content obtained from ECG signals56,57,90,101. Four studies showcased the ability of two wearables in tracking changes in pulmonary fluid status32,83,90,98. One of these studies demonstrated the usability of their device in a remote setting32. Two other studies demonstrate the correlation between cardiac acoustic biomarkers and absolute pulmonary artery pressures and HF state63,67.
Discussion
In this scoping review, we evaluated the use of non-invasive wearables to monitor HF patients which resulted in the following main findings; (I) Currently, a diverse array of non-invasive wearables are used for monitoring HF patients, with a primary focus on physical activity assessment. Additionally, wearables encompassing ECG, PPG, SCG, and bio-impedance based sensors, facilitating the transmission of parameters such as heart rhythm, heart rate, and pulmonary congestion, have been studied. (II) Most of the wearables intended for RM of HF patients are currently in the feasibility phase (MDRL 6). However, many of these wearables were originally designed for different purposes or different populations and have been adapted for HF monitoring. (III) The majority of these devices are consumer-grade wearables and have not received FDA- or MDR approval for the monitoring of HF patients. (IV) While multiple devices demonstrated safety, reliability, and efficacy in small, predominantly observational, studies, there is a lack of large-scale RCTs to firmly establish the clinical advantages of wearable RM. Overall, our findings highlight the potential of wearables in RM for HF, but more so emphasize the need for further research to establish their clinical benefits before wider adoption and clinical implementation.
In the evolving landscape of remote HF management, the advent of non-invasive wearable technologies offers a promising solution to improve patient monitoring and care, while potentially easing the strain on healthcare systems4,114. In this study, accelerometers were identified as the wearable technology most thoroughly researched, mainly for tracking physical activity. Among them, the Actigraph accelerometer was the most frequently used, reaching a MDRL of 6 and achieving approval from both the FDA and MDR. However, the latter was not specifically for the RM of HF patients. The focus on accelerometers is particularly relevant given its availability and given the current reliance on the subjective New York Heart Association (NYHA) classification to assess functional status in HF patients, a method with a number of intrinsic limitations115. Our review has identified several studies describing a significant association between decreased physical activity, as measured by accelerometers, and key clinical outcomes, including reduced exercise capacity and quality of life, with some studies even linking it to increased mortality58,76,79. While accelerometers show promise in objectively measuring physical activity and potentially complementing traditional clinical assessments, the direct impact of such measurements on patient care, particularly in improving outcomes, remains an area requiring further investigation.
Current RM strategies surpass mere physical activity tracking, aiming primarily at identifying clinical deterioration, especially congestion, to avert hospital admissions or readmissions116. The challenge of detecting early congestion in HF patients, particularly during stages when symptoms are not present, is significant. This challenge is exacerbated by the limitations of physical examinations, laboratory markers, and patient-reported symptoms in predicting HF readmissions117,118. Such limitations underscore the urgent need for innovative, non-invasive methods capable of accurately assessing a patient’s volume status to guide diuretic treatment effectively. Our review has identified various wearable devices designed for the direct non-invasive monitoring of congestion. Notable innovations include the ReDS non-invasive vest and the CardioSet Edema Guard Monitor, both of which employ bio-impedance for direct congestion assessment. The ReDS system, in particular, has attracted the most interest with a considerable amount of evidence supporting its use. Currently, one RCT (NCT03586336) is in progress, exploring the ReDS system’s practicality and effectiveness in a real-world clinical context. Simultaneously, there is an emerging trend towards the development of more pragmatic wearables, such as smartwatches and small patches which are capable of estimating patients’ volume status or hemodynamic parameters associated with HF, including cardiac output and pulmonary wedge pressure119. The development of these indirect measurement techniques often involves the use of machine learning models integrating various variables, such as HR, HRV, pulse pressure timing (PPT), and physical activity (PA) in their estimation of the target parameter. Currently, these innovative approaches are still in the nascent stages of development, typically around MDRL 4–5, indicating a preliminary phase compared to the aforementioned devices40,55. Despite the potential of these ML-driven systems, the complexity and the black-box nature of the underlying algorithms pose challenges in clinical interpretation and acceptance120. To overcome these barriers and ensure the successful integration of these wearables into clinical practice, it is critical to advance clinical validation efforts. This includes conducting large-scale RCTs to ascertain their effectiveness and incorporating explainable AI approaches to demystify the decision-making processes of these technologies121.
Modern consumer wearables can perform measurements with a degree of accuracy comparable to regulated medical instruments. Consequently, the line between wearables designed for consumer use and those intended for medical applications is increasingly blurred. The general public is now more than ever using this data to monitor and improve their health122. A similar trend is evident in scientific research, as highlighted in this review, with the majority of the wearables under investigation being consumer-grade designed for different purposes or different populations and adapted for HF monitoring. While the use of consumer-grade wearables democratizes access to physiological data, it also poses challenges for healthcare providers. It’s essential to recognize that the majority of consumer-grade wearables (or part of their functionalities) have not undergone thorough validation, and even when they have, this validation has primarily been conducted on young and healthy individuals. Hence, clinicians should be cautious, recognizing that these wearables should not be employed for this purpose beyond research settings to safeguard patient well-being123,124. Moreover, the accuracy reported depends on the choice of the gold standard utilized. This underscores the importance of establishing standardized protocols and measures to conduct a robust assessment of the accuracy of these devices, as well as to define their operational limitations10. In this regard, it is striking that a MDRL below 6 hardly occurs in this category of wearables. In contrast, most devices specifically developed for HF patients were still in the prototype phase with the exception of the ReDS™ Wearable System which has a MDRL of 620,22,45,68,103. When specifically developing a technology for HF patients, consideration was given from the start to which parameters could have clinical value, preliminary studies and validation were performed in the intended population. As such, the included prototypes a MDRL of 4 may therefore actually be further along in the process towards clinical application than a number of already commercially available wearables that have now been used on HF patients for the first time. This is also exemplified by the fact that the VitalPatch and ReDS™ Wearable System both have received FDA approval for the monitoring of HF patients, whereas none of the consumer-grade wearables have achieved such recognition20,22,45,68,102,103. If the results of both wearables are replicated in a larger RCT, the devices would obtain a MDRL of 7 or 8. However, the development, validation, and production of wearables tailored to specific purposes often entail higher costs, which can restrict their global adoption and availability. Although we did not include or report the costs associated with each wearable, our findings clearly illustrate this trade-off. Furthermore, they underscore the constraints of the MDRL, emphasizing that iterations should not only emphasize the significance of rigorous validation but also stress that it must be conducted within the target population.
A recent meta-analysis advocated for the use of RM for HF patients by showing that non-invasive RM of vital signs is associated with a significant reduction in the risk of first and total HF hospitalizations4. Integrating wearables into HF monitoring systems offers the possibility of significantly improving patient care through the continuous and objective tracking of physiological data. The enhanced connectivity of most wearable devices enables the real-time monitoring of changes in cardiac condition, providing a more immediate and comprehensive view of the patient’s health status. This continuous and up-to-date data acquisition has the potential to enhance the timeliness and predictive accuracy as compared to sporadic measurements125. The rationale behind this improvement lies in the ability to extract valuable insights from trends and effectively filter out daily fluctuations. However, this optimistic outlook on wearable technology integration comes with a caveat. Despite the ability to measure a vast array of physiological parameters, a critical question remains if these measurements tangibly benefit HF management and patient outcomes? The reality of healthcare innovation brings to light the complexity of translating data into actionable insights. While wearables hold the potential to foster proactive HF healthcare and encourage patient engagement and self-management, their effectiveness hinges on our ability to identify which measurements are clinically relevant and how they can be used to guide therapeutic decisions. Several challenges and opportunities remain to unlock the potential of RM with wearables. First, as described above, it is crucial to investigate the safety and effectiveness of wearables through randomized studies that not only include a larger patient population but also expand to encompass diverse geographical settings, including low-income countries, which were notably absent in our study126. Despite the proven safety and reliability of various devices in small observational studies, there is a notable lack of large RCTs confirming the clinical benefits of wearable RM. Current research often focuses on the devices’ ability to reliably transmit data, leaving their actual clinical impact largely unexplored due to the high cost of extensive trials. Execution of these trials are problematic because technological advancements in the field of wearables outpace the results of these trials. Novel trail designs are necessary to resolve this problem127,128. Furthermore, it is crucial to diversify study populations beyond the typical demographic of younger, white, wealthier individuals. Including varied socio-economic backgrounds and focusing on underrepresented groups will help ensure that wearable technologies can be equitably beneficial and tailored to the needs of all populations, particularly those in regions that are currently underserved by advanced medical technologies129,130. Rushing to integrate these devices into standard HF care without thorough testing in other demographic groups could result in flawed monitoring, as evidenced by numerous studies highlighting inaccuracies in PPG-derived SpO2 measurements among individuals with darker skin tones131. Furthermore, since these devices are not readily available to other populations, their integration may potentially exacerbate disparities in HF care132. To address this risk, it is critical to implement strategic measures such as expanding insurance reimbursement, shifting towards value-based payment models, and increasing public and private sector investments in wearable technology125,133. As we embrace wearable devices in our healthcare system, it is imperative to maintain an ongoing focus on equity concerns in order to address the pre-existing digital divide. Second, to facilitate reliable, long-term continuous measurements in a wearable form there is a need for further advancement in sensing and sampling technologies. Patients have well over 500.000 heartbeats and over 100.000 breaths a week. Noise, misinterpretation and false positives are thereby unavoidable. This will lead to extra care visits, which have to be weight against the benefits in resource restrained care systems. One way to accomplish this is by employing multimodal and/or multiplexed sensing, which involves using various transducer types which simultaneously measure different signals in a single wearable134. Third, further integration of wearables will create an overload of data that challenges the feasibility of the needed data infrastructure in hospitals. Not only does this require large-scale data storage, but the assessment of the resulting data will also be labor intensive when done by healthcare providers125. Consequently, the expanding use of wearables necessitates a corresponding increase in healthcare provider training programs. These programs should be designed not only to enhance data management skills but also to enable providers to make informed decisions based on the data collected by these devices. By contrast, the development of monitoring protocols, including alarm thresholds, will play an important role to mitigate this work. Additionally, the potential of cloud or fog computing, data mining, and machine learning in managing and interpreting the vast datasets generated by wearables has been highlighted134. These technologies could play a transformative role in automating data interpretation, thereby reducing the burden on healthcare providers and potentially enhancing the scalability of wearable technologies for RM of HF patients135. This could be further enhanced through the integration of feedback loops that would exclude the intervention of a healthcare provider (closed-loop). For example, a wearable device could allow for real time monitoring of the hemodynamic state of the patient and, in case of congestion, could offer an advice to increase the diuretic dosage or even give a subcutaneous bolus of the diuretic. Fourth, social acceptance of wearables must be ensured by informing users about the advantages and disadvantages and by integrating them into application ecosystems and health-care services. Last, the adoption of protocols for data safety and privacy with the establishment of an ethical regulatory framework for wearable data networks could further promote their use136.
Despite the various strengths inherent in this comprehensive scoping review, it is essential to acknowledge its limitations. Firstly, although the MDRL provides a comprehensive depiction of the developmental stage of a wearable device, it has a limitation when evaluating existing wearables, not specially designed for HF monitoring. Existing wearables do not progress to the entire MDRL framework and when tested in the target population automatically receive at least MDRL 5, as they are beyond the prototype phase. However, clinical value is not guaranteed merely by reaching this stage, as measurements might be unreliable in HF patients. Thus, a universally adapted scale is needed to prioritize the intended medical purpose for all stages of development, enabling a fair comparison of wearables’ readiness. Secondly, the inclusion of only English full-text articles and the omission of non-English articles and conference papers represent another limitation. Conferences and preprint servers play a crucial role in the technology and machine learning community. Consequently, this approach might have missed out on some novel developments and valuable insights from conference publications and non-English journals. However, conference papers may lack the same scrutiny of the peer-review process, potentially leading to biased results. Thirdly, it is important to note that both commercially available and medical-grade wearable activity monitors are constantly evolving, with new products being released each year. As a result, findings from studies conducted earlier in the data collection period may hold less relevance to researchers today. However, evolution of wearable technology does not invalidate the insights and knowledge gained from earlier studies. Instead, it adds to the growing body of research, allowing for comparisons, trend analysis, and identification of changes in technology and its impact on outcomes. Last, the objective of this scoping review was to describe the current evidence of the use of wearables in RM of HF patients. As such, we did not give a detailed description of the capabilities of each wearable, their underlying technique and separate functionalities.
In conclusion, this review evaluated the application of wearables in the RM of HF patients, highlighting a significant reliance on consumer-grade devices repurposed for medical studies. While a minority of wearables have been explicitly designed for HF monitoring and have even received FDA approval, the real value of these technologies in enhancing HF care remains under question. Especially, the paucity of large-scale RCTs underscores a critical gap in our understanding of the real-world benefits of wearables in managing HF. Addressing this question is essential for moving beyond “nice-to-have” gadgets to truly impactful tools in healthcare.
Methods
We performed a scoping review with a systematic literature search of both randomized controlled trials (RCTs) and observational studies according to the Preferred Reporting Items for Systematic Reviews and Meta-Analyses for Scoping Reviews (PRISMA-ScR) guidelines137. The PRISMA-ScR checklist is provided in Supplementary Note 1. A prespecified local protocol was available.
Search strategy and selection criteria
In collaboration with an expert librarian specialized in systematic searches a literature search was carried out on the 05/06/2024 including studies that were published up to that date, by using Embase, Medline Ovid, Web of Science and PubMed. Keywords used in the search included: “heart failure”, “wearable electronic devices”, “telemedicine”, and “remote monitoring”. The full search strategy is presented in Supplementary Note 2. Only peer-reviewed full original articles in the English language were included in our study. Studies were included if they contained any form of RM (or intended for RM) using a wearable sensor in chronic HF patients (NYHA class I-III or ambulatory class IV) aged 18 years or above. For this literature search, a wearable was defined as a connected electronic device that can be worn on the body as an accessory or embedded into clothing without burdening or hindering the wearer. In an effort to distinguish among the different types of wearables used in HF management, we define consumer-grade wearables as devices commercially available for general health monitoring, not specifically intended for medical use. Medical-grade wearables, including those used for research, are designed and validated for specific health monitoring purposes and are regulated by health authorities like the FDA or MDR. Non-invasive measure devices such as standard blood pressure monitors, electrocardiograms or handheld ultrasound were not defined as wearables. Additionally, studies about wearables (e.g., wearable cardioverter defibrillator) used for therapy only were excluded if there was no monitoring function used as well. Articles about the cost-benefit analysis of RM that include a wearable in their costs but do not provide detailed information about the used wearable were also excluded. The same applied for studies that solely described healthcare professionals’ and/or patients’ experiences with RM in general unless they provide specific, detailed information about the wearable device itself.
Three independent reviewers (A.R., A.S., and N.S.) independently performed screening of title and/or abstract to identify studies that potentially met the inclusion criteria. Hereafter, the full-text of each selected study was discussed in detail to decide upon the eligibility based on the inclusion and exclusion criteria. In case of any disagreement regarding eligibility and no consensus was reached the final decision was made by the last author (R.B.). If eligible studies described the same population, only the study with the longest follow-up or most recent publication containing the entire population was included, unless different outcomes of interest were studied in each article. Studies describing a subgroup of the same population were excluded. For title and/or abstract and full-text screening of the article, the online available systematic review tool “Covidence” was used.
Data collection and extraction
The following information was extracted from the main study reports: author, year of publication, country, study name, study design, sample size, type of wearable, predictors the device can measure, sensor types of the device and endpoints of the study. Data was extracted in a predefined Microsoft Excel spreadsheet. Additionally, the medical certification for every device was collected from the online available databases of the Food and Drug Administration (FDA) and the Medical Device Regulation (MDR)138,139. In the United States, wearables used for medical purposes require FDA certification and wearables used in the European Union require MDR certification.
Medical device readiness level
The Medical Device Readiness Level (MDRL), an adaption of the Technology Readiness Level, as proposed by Ruiz Seva et al.14 was used to assess the readiness of the wearables to be implemented in a clinical setting for HF monitoring. This measurement system has nine levels with general descriptions (Fig. 3). The MDRL levels ranged from 1, in which an existing medical challenge is addressed by identification of scientific and design principles, to MDRL 9 in which the wearable device is fully accepted in the market14. For this study, we specifically assessed the wearables’ readiness for HF monitoring. To ensure objectivity in determining the MDRL, each study was independently assessed by three researchers (N.S, A.S, R.B), and discrepancies were resolved through structured discussions to achieve consensus. The assessment was solely based on information extracted from the articles obtained through the systematic search. To capture variations in readiness based on specific monitoring functionalities (e.g., physical activity or heart rate monitoring), the MDRL was evaluated per function. This approach acknowledges that the readiness between the different functions may differ, especially when considering its application in the HF population.
Fig. 3. MDRL per variable.
An overview of the variables utilized in various wearables in research for heart failure remote monitoring. For each variable, the Medical Device Readiness Level (MDRL) is provided. MDRL medical device readiness level, HF heart failure, SpO2 blood oxygen saturation, VO2 Oxygen uptake. MDRL presented in the figure represents the highest MDRL of that variable. Figure created using BioRender.
Supplementary information
Acknowledgements
This work was investigator-initiated and did not receive any external funding. N.S. is supported by a grant from the Dutch Research Council (NWO), grant number: 628.011.214 (STRAP). Some figures are created by using BioRender.
Author contributions
N.S. and A.R.: conceptualization, methodology, investigation, writing – original draft, visualization; A.S.: investigation, writing – review & editing; E.B., E.R., J.B., and R.B.: writing - review & editing; N.B.: writing - review & editing, supervision; R.M.A.B.: conceptualization, supervision. All authors have read and approved the manuscript.
Data availability
The data underlying this article can be shared on reasonable request to the corresponding author.
Competing interests
N.S. none. A.R. none. A.S. none. E.B. none. E.R. [consultant for Corsano, a PPG wristband manufacturer]. R.B. reports an independent research grant for ISS from Abbott to the Institute and has had speaker engagements or advisory boards in the past 5 years with Abbott and Boehringer Ingelheim. J.B. received independent research grant from Abbott for ISS and has had speaker engagement or advisory boards in the past 5 years with Astra Zeneca, Abbott, Boehringer-Ingelheim, Bayer, Daiichi Sankyo, Novartis and Vifor. N.B. reports to be Editor-in-Chief at the European Heart Journal – Digtal Health, Topic coordinator Digital Health at the congress program committee of the European Society of Cardiology (ESC), and Vice-Chair of the Digital Health Committee of the ESC. R.M.A.B. reports an independent research grant for ISS from Abbott to the Institute and has had speaker engagements or visory boards in the past 5 years with Abbott and Boehringer Ingelheim. All remaining authors declare no Competing Financial or Non-Financial Interest.
Footnotes
Publisher’s note Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
These authors contributed equally: Annemiek. E. van Ravensberg, Abdul Shakoor.
Supplementary information
The online version contains supplementary material available at 10.1038/s41746-024-01268-5.
References
- 1.Disease, G. B. D., Injury, I. & Prevalence, C. Global, regional, and national incidence, prevalence, and years lived with disability for 354 diseases and injuries for 195 countries and territories, 1990-2017: a systematic analysis for the Global Burden of Disease Study 2017. Lancet392, 1789–1858 (2018). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2.Savarese, G. et al. Global burden of heart failure: a comprehensive and updated review of epidemiology. Cardiovasc. Res.118, 3272–3287 (2023). [DOI] [PubMed] [Google Scholar]
- 3.Jordan-Rios, A. et al. Reshaping care in the aftermath of the pandemic. Implications for cardiology health systems. Eur. J. Intern. Med.109, 4–11 (2023). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.Scholte, N. T. B. et al. Telemonitoring for heart failure: a meta-analysis. Eur. Heart J.44, 2911–2926 (2023). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5.Sharif, F. et al. Twelve-month follow-up results from the SIRONA 2 clinical trial. ESC Heart Fail11, 1133–1143 (2024). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6.Clephas, P. R. D. et al. Efficacy of pulmonary artery pressure monitoring in patients with chronic heart failure: a meta-analysis of three randomized controlled trials. Eur. Heart J.44, 3658–3668 (2023). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7.Koehler, F. & Hindricks, G. Is telemonitoring for heart failure ready after a journey longer than two decades? Eur. Heart J.44, 2927–2929 (2023). [DOI] [PubMed] [Google Scholar]
- 8.McDonagh, T. A. et al. 2021 ESC Guidelines for the diagnosis and treatment of acute and chronic heart failure. Eur. Heart J.42, 3599–3726 (2021). [DOI] [PubMed] [Google Scholar]
- 9.Wearable Computing Devices Market - Growth, Trends, COVID-19 Impact, and Forecastshttps://www.researchandmarkets.com/reports/4787502/wearable-computing-devices-market-growth (2021–2026).
- 10.Jensen, M. T. et al. ESC working group on e-cardiology position paper: use of commercially available wearable technology for heart rate and activity tracking in primary and secondary cardiovascular prevention-in collaboration with the European Heart Rhythm Association, European Association of Preventive Cardiology, Association of Cardiovascular Nursing and Allied Professionals, Patient Forum, and the Digital Health Committee. Eur. Heart J. Digit Health2, 49–59 (2021). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11.Manninger, M. et al. Role of wearable rhythm recordings in clinical decision making-The wEHRAbles project. Clin. Cardiol.43, 1032–1039 (2020). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.Ravizza, A. et al. Comprehensive review on current and future regulatory requirements on wearable sensors in preclinical and clinical testing. Front Bioeng. Biotechnol.7, 313 (2019). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13.Jiang, N., Muck, J. E. & Yetisen, A. K. The regulation of wearable medical devices. Trends Biotechnol.38, 129–133 (2020). [DOI] [PubMed] [Google Scholar]
- 14.Seva, R. R., Tan, A. L. S., Tejero, L. M. S. & Salvacion, M. L. D. S. Multi-dimensional readiness assessment of medical devices. Theor. Issues Erg. Sci.24, 189–205 (2023). [Google Scholar]
- 15.Aamodt, I. T. et al. Self-care monitoring of heart failure symptoms and lung impedance at home following hospital discharge: longitudinal study. J. Med. Internet Res.22, e15445 (2020). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16.Alosco, M. L. et al. Decreased physical activity predicts cognitive dysfunction and reduced cerebral blood flow in heart failure. J. Neurol. Sci.339, 169–175 (2014). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17.Alosco, M. L. et al. Decreases in daily physical activity predict acute decline in attention and executive function in heart failure. J. Card. Fail.21, 339–346 (2015). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18.Alosco, M. L. et al. Depression is associated with reduced physical activity in persons with heart failure. Health Psychol.31, 754–762 (2012). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19.Alvarez-Garcia, J. et al. Remote dielectric sensing before and after discharge in patients with ADHF: the ReDS-SAFE HF trial. JACC Heart Fail12, 695–706 (2024). [DOI] [PubMed] [Google Scholar]
- 20.Amir, O. et al. Evaluation of remote dielectric sensing (ReDS) technology-guided therapy for decreasing heart failure re-hospitalizations. Int. J. Cardiol.240, 279–284 (2017). [DOI] [PubMed] [Google Scholar]
- 21.Baril, J. F. et al. Use of free-living step count monitoring for heart failure functional classification: validation study. JMIR Cardio3, e12122 (2019). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 22.Bensimhon, D. et al. The use of the Reds noninvasive lung fluid monitoring system to assess readiness for discharge in patients hospitalized with acute heart failure: a pilot study. Heart Lung50, 59–64 (2021). [DOI] [PubMed] [Google Scholar]
- 23.Blockhaus, C. et al. Telemonitoring potential of wearable cardioverter-defibrillators during the follow-up of patients with heart failure. J. Cardiovasc. Dev. Dis.9, 175 (2022). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 24.Blomqvist, A., Bäck, M., Klompstra, L., Strömberg, A. & Jaarsma, T. Utility of single-item questions to assess physical inactivity in patients with chronic heart failure. ESC Heart Fail7, 1467–1476 (2020). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 25.Braun, T. et al. Association between heart failure severity and mobility in geriatric patients: an in-clinic study with wearable sensors. J. Geriatr. Cardiol.19, 660–674 (2022). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 26.Burch, A. E. et al. Wearable cardioverter defibrillator-guided 6-min walk test performed at home is accurate and reliable: RESULTS OF THE TRENDS STUDY. J. Cardiopulm. Rehabil. Prev.40, E14–E17 (2020). [DOI] [PubMed] [Google Scholar]
- 27.Burkhoff, D., Bailey, G. & Gimbel, J. R. Characterization of cardiac acoustic biomarkers in patients with heart failure. Ann. Noninvasive Electrocardiol.25, e12717 (2020). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 28.Butler, J. et al. Accelerometer vs. other activity measures in heart failure with preserved ejection fraction: the VITALITY-HFpEF trial. ESC Heart Fail11, 293–298 (2024). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 29.Curtain, J. P. et al. Measuring congestion with a non-invasive monitoring device in heart failure and haemodialysis: CONGEST-HF. Eur. J. Heart Fail26, 1383–1392 (2024). [DOI] [PubMed] [Google Scholar]
- 30.da Silva, V. Z. et al. Association between physical activity measurements and key parameters of cardiopulmonary exercise testing in patients with heart failure. J. Card. Fail19, 635–640 (2013). [DOI] [PubMed] [Google Scholar]
- 31.Dagan, M. et al. Advanced hemodynamic monitoring allows recognition of early response patterns to diuresis in congestive heart failure patients. J. Clin. Med12, 45 (2022). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 32.Darling, C. E. et al. Bioimpedance-based heart failure deterioration prediction using a prototype fluid accumulation vest-mobile phone dyad: an observational study. JMIR Cardio1, e1 (2017). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 33.Deka, P., Pozehl, B., Norman, J. F. & Khazanchi, D. Feasibility of using the Fitbit(®) Charge HR in validating self-reported exercise diaries in a community setting in patients with heart failure. Eur. J. Cardiovasc Nurs.17, 605–611 (2018). [DOI] [PubMed] [Google Scholar]
- 34.Di Rienzo, M., Rizzo, G., Işılay, Z. M. & Lombardi, P. SeisMote: a multi-sensor wireless platform for cardiovascular monitoring in laboratory, daily life, and telemedicine. Sensors20, 680 (2020). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 35.Dibben, G. O. et al. Physical activity assessment by accelerometry in people with heart failure. BMC Sports Sci. Med Rehabil.12, 47 (2020). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 36.Dibben, G. O. et al. Home-based cardiac rehabilitation and physical activity in people with heart failure: a secondary analysis of the REACH-HF randomised controlled trials. BMJ Open13, e063284 (2023). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 37.Dontje, M. L. et al. Daily physical activity in stable heart failure patients. J. Cardiovasc. Nurs.29, 218–226 (2014). [DOI] [PubMed] [Google Scholar]
- 38.Dorsch, M. P., Farris, K. B., Rowell, B. E., Hummel, S. L. & Koelling, T. M. The effects of the ManageHF4Life mobile app on patients with chronic heart failure: randomized controlled trial. JMIR Mhealth Uhealth9, e26185 (2021). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 39.Edwards, M. K. & Loprinzi, P. D. Sedentary behavior & health-related quality of life among congestive heart failure patients. Int. J. Cardiol.220, 520–523 (2016). [DOI] [PubMed] [Google Scholar]
- 40.Erath, J. W. et al. Influence of decompensated heart failure on cardiac acoustic biomarkers: impact on early readmissions. ESC Heart Fail7, 4198–4205 (2020). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 41.Evangelista, L. S. et al. Validity of pedometers for measuring exercise adherence in heart failure patients. J. Card. Fail11, 366–371 (2005). [DOI] [PubMed] [Google Scholar]
- 42.Floegel, T. A., Allen, K. D. & Buman, M. P. A pilot study examining activity monitor use in older adults with heart failure during and after hospitalization. Geriatr. Nurs.40, 185–189 (2019). [DOI] [PubMed] [Google Scholar]
- 43.Floegel, T. A. et al. Association of posture and ambulation with function 30 days after hospital discharge in older adults with heart failure. J. Card. Fail24, 126–130 (2018). [DOI] [PubMed] [Google Scholar]
- 44.Fulcher, K. K. et al. Greater physical activity is associated with better cognitive function in heart failure. Health Psychol.33, 1337–1343 (2014). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 45.Garcia, R. et al. Dynamic changes in nocturnal heart rate predict short-term cardiovascular events in patients using the wearable cardioverter-defibrillator: from the WEARIT-France cohort study. Europace25, euad062 (2023). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 46.Gardner, C. L. et al. Electronic physiologic and subjective data acquisition in home-dwelling heart failure patients: an assessment of patient use and perception of usability. Int. J. Med Inform.93, 42–48 (2016). [DOI] [PubMed] [Google Scholar]
- 47.German, C. A. et al. Relationships between objectively measured physical activity, exercise capacity, and quality of life in older patients with obese heart failure and preserved ejection fraction. J. Card. Fail27, 635–641 (2021). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 48.Güder, G. et al. Establishing a cardiac training group for patients with heart failure: the “HIP-in-Würzburg” study. Clin. Res. Cardiol.111, 406–415 (2022). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 49.Guo, X. et al. A hospital-community-family-based telehealth program for patients with chronic heart failure: single-arm, prospective feasibility study. JMIR Mhealth Uhealth7, e13229 (2019). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 50.Herkert, C., Kraal, J. J., van Loon, E. M. A., van Hooff, M. & Kemps, H. M. C. Usefulness of modern activity trackers for monitoring exercise behavior in chronic cardiac patients: validation study. JMIR Mhealth Uhealth7, e15045 (2019). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 51.Hillmann, H. A. K. et al. Feasibility and first results of heart failure monitoring using the wearable cardioverter-defibrillator in newly diagnosed heart failure with reduced ejection fraction. Sensors21, 7798 (2021). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 52.Holber, J. P. et al. The relationship between objectively measured step count, clinical characteristics, and quality of life among depressed patients recently hospitalized with systolic heart failure. Psychosom. Med.84, 231–236 (2022). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 53.Howie-Esquivel, J. & Zaharias, E. Using novel technology to determine mobility among hospitalized heart failure patients: a pilot study. Cardiol. Res.4, 15–25 (2013). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 54.Iliodromitis, K. et al. Assessing physical activity with the wearable cardioverter defibrillator in patients with newly diagnosed heart failure. Front Cardiovasc. Med.10, 1176710 (2023). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 55.Inan, O. T. et al. Novel wearable seismocardiography and machine learning algorithms can assess clinical status of heart failure patients. Circ. Heart Fail11, e004313 (2018). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 56.Iqbal, S. M., Leavitt, M. A., Pedilus, G., Mahgoub, I. & Asghar, W. A wearable telehealth system for the monitoring of parameters related to heart failure. Heliyon10, e26841 (2024). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 57.Iqbal, S. M. A., Mahgoub, I., Du, E., Leavitt, M. A. & Asghar, W. Development of a wearable belt with integrated sensors for measuring multiple physiological parameters related to heart failure. Sci. Rep.12, 20264 (2022). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 58.Izawa, K. P. et al. Usefulness of step counts to predict mortality in Japanese patients with heart failure. Am. J. Cardiol.111, 1767–1771 (2013). [DOI] [PubMed] [Google Scholar]
- 59.Izawa, K. P. et al. Association between mental health and physical activity in patients with chronic heart failure. Disabil. Rehabil.36, 250–254 (2014). [DOI] [PubMed] [Google Scholar]
- 60.Jehn, M. et al. Tele-accelerometry as a novel technique for assessing functional status in patients with heart failure: feasibility, reliability and patient safety. Int. J. Cardiol.168, 4723–4728 (2013). [DOI] [PubMed] [Google Scholar]
- 61.Jehn, M. et al. Accelerometer-based quantification of 6-minute walk test performance in patients with chronic heart failure: applicability in telemedicine. J. Card. Fail15, 334–340 (2009). [DOI] [PubMed] [Google Scholar]
- 62.Jungbauer, C. G., Maier, L. S., Emoto, K., Zirille, F. M. & Mirro, M. J. Achieving guideline-directed heart rate controlearly posthospitalization. Am. J. Cardiol.123, 1096–1100 (2019). [DOI] [PubMed] [Google Scholar]
- 63.Kaneko, T. et al. Relationship between cardiac acoustic biomarkers and pulmonary artery pressure in patients with heart failure. J. Clin. Med.11, 6373 (2022). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 64.Kikuchi, A. et al. Feasibility of home-based cardiac rehabilitation using an integrated telerehabilitation platform in elderly patients with heart failure: a pilot study. J. Cardiol.78, 66–71 (2021). [DOI] [PubMed] [Google Scholar]
- 65.Klompstra, L. et al. Objectively measured physical activity in patients with heart failure: a sub-analysis from the HF-Wii study. Eur. J. Cardiovasc. Nurs.21, 499–508 (2022). [DOI] [PubMed] [Google Scholar]
- 66.Klompstra, L., Jaarsma, T. & Strömberg, A. Exergaming to increase the exercise capacity and daily physical activity in heart failure patients: a pilot study. BMC Geriatr.14, 119 (2014). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 67.Koivisto, T. et al. Mechanocardiography-based measurement system indicating changes in heart failure patients during hospital admission and discharge. Sensors22, 9781 (2022). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 68.Lala, A. et al. Early use of remote dielectric sensing after hospitalization to reduce heart failure readmissions. ESC Heart Fail8, 1047–1054 (2021). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 69.Li, P. et al. Objective assessment of daytime napping and incident heart failure in 1140 community-dwelling older adults: a prospective, observational cohort study. J. Am. Heart Assoc.10, e019037 (2021). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 70.Li, X. C. et al. Evaluation of left ventricular systolic function using synchronized analysis of heart sounds and the electrocardiogram. Heart Rhythm17, 876–880 (2020). [DOI] [PubMed] [Google Scholar]
- 71.Lin, H. et al. Clinical and demographic correlates of accelerometer-measured physical activity in participants enrolled in the OPTIMISE HFpEF study. Eur. J. Cardiovasc. Nurs.21, 67–75 (2022). [DOI] [PubMed] [Google Scholar]
- 72.Lin, W. Y. et al. Realization and technology acceptance test of a wearable cardiac health monitoring and early warning system with multi-channel MCGs and ECG. Sensors18, 3538 (2018). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 73.McCarthy, M. M., Dickson, V. V., Katz, S. D. & Chyun, D. A. An exercise counseling intervention in minority adults with heart failure. Rehabil. Nurs.42, 146–156 (2017). [DOI] [PubMed] [Google Scholar]
- 74.Melczer, C., Melczer, L., Goják, I., Oláh, A. & Ács, P. A comparative analysis between external accelerometer and internal accelerometer’s physical activity data from implanted resynchronization devices in patients with heart failure. Eur. J. Integr. Med8, 18–22 (2016). [Google Scholar]
- 75.Melin, M., Hagerman, I., Gonon, A., Gustafsson, T. & Rullman, E. Variability in physical activity assessed with accelerometer is an independent predictor of mortality in CHF patients. PLoS One11, e0153036 (2016). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 76.Miyahara, S. et al. Postdischarge light-intensity physical activity predicts rehospitalization of older Japanese patients with heart failure. J. Cardiopulm. Rehabil. Prev.38, 182–186 (2018). [DOI] [PubMed] [Google Scholar]
- 77.Mlakar, M. et al. Mining telemonitored physiological data and patient-reported outcomes of congestive heart failure patients. PLoS One13, e0190323 (2018). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 78.Nagatomi, Y. et al. Home-based cardiac rehabilitation using information and communication technology for heart failure patients with frailty. ESC Heart Fail9, 2407–2418 (2022). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 79.Nelson, M. B. et al. Physical activity and relationship to physical function, quality of life, and cognitive function in older patients with acute decompensated heart failure. Am. Heart J.256, 85–94 (2023). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 80.O’Donnell, J. et al. Self-reported and objectively measured physical activity in people with and without chronic heart failure: UK Biobank analysis. Open Heart7, e001099 (2020). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 81.Okwose, N. C. et al. Acceptability, feasibility and preliminary evaluation of a novel, personalised, home-based physical activity intervention for chronic heart failure (Active-at-Home-HF): a pilot study. Sports Med. Open5, 45 (2019). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 82.Omar, M. et al. Hemodynamic determinants of activity measured by accelerometer in patients with stable heart failure. JACC Heart Fail.9, 824–835 (2021). [DOI] [PubMed] [Google Scholar]
- 83.Pan, X. et al. Deep cross-modal feature learning applied to predict acutely decompensated heart failure using in-home collected electrocardiography and transthoracic bioimpedance. Artif. Intell. Med140, 102548 (2023). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 84.Polcz, M. et al. Comparison of clinical symptoms and bioimpedance to pulmonary capillary wedge pressure in heart failure. Am. Heart J.15, 100133 (2022). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 85.Pozehl, B. J. et al. Accelerometer-measured daily activity levels and related factors in patients with heart failure. J. Cardiovasc Nurs.33, 329–335 (2018). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 86.Prescher, S. et al. Prognostic value of serial six-minute walk tests using tele-accelerometry in patients with chronic heart failure: A pre-specified sub-study of the TIM-HF-Trial. Eur. J. Prev. Cardiol.23, 21–26 (2016). [DOI] [PubMed] [Google Scholar]
- 87.Radhakrishnan, K. et al. Usability testing of a sensor-controlled digital game to engage older adults with heart failure in physical activity and weight monitoring. Appl. Clin. Inf.11, 873–881 (2020). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 88.Röger, S. et al. Therapy optimization in patients with heart failure: the role of the wearable cardioverter-defibrillator in a real-world setting. BMC Cardiovasc. Disord.18, 52 (2018). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 89.Rullman, E. et al. Circulatory factors associated with function and prognosis in patients with severe heart failure. Clin. Res. Cardiol.109, 655–672 (2020). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 90.Sanchez-Perez, J. A. et al. A wearable multimodal sensing system for tracking changes in pulmonary fluid status, lung sounds, and respiratory markers. Sensors (Basel)22, 1130 (2022). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 91.Savoldelli, A., Vitali, A., Remuzzi, A. & Giudici, V. Improving the user experience of televisits and telemonitoring for heart failure patients in less than 6 months: a methodological approach. Int. J. Med Inf.161, 104717 (2022). [DOI] [PubMed] [Google Scholar]
- 92.Schmidt, C. et al. Comparison of questionnaire and accelerometer-based assessments of physical activity in patients with heart failure with preserved ejection fraction: clinical and prognostic implications. Scand. Cardiovasc. J.54, 77–83 (2020). [DOI] [PubMed] [Google Scholar]
- 93.Schwendinger, F., Wagner, J., Infanger, D., Schmidt-Trucksäss, A. & Knaier, R. Methodological aspects for accelerometer-based assessment of physical activity in heart failure and health. BMC Med. Res. Methodol.21, 251 (2021). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 94.Shandhi, M. M. H. et al. Wearable patch-based estimation of oxygen uptake and assessment of clinical status during cardiopulmonary exercise testing in patients with heart failure. J. Card. Fail26, 948–958 (2020). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 95.Shen, H. et al. Impaired chronotropic response to physical activities in heart failure patients. BMC Cardiovasc. Disord.17, 136 (2017). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 96.Shiraishi, Y. et al. Assessment of physical activity using waist-worn accelerometers in hospitalized heart failure patients and its relationship with Kansas City cardiomyopathy questionnaire. J. Clin. Med10, 4103 (2021). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 97.Shoemaker, M. J., Curtis, A. B., Vangsnes, E. & Dickinson, M. G. Clinically meaningful change estimates for the six-minute walk test and daily activity in individuals with chronic heart failure. Cardiopulm Phys. Ther. J.24, 21–29 (2013). [PMC free article] [PubMed] [Google Scholar]
- 98.Smeets, C. J. P. et al. The added value of in-hospital tracking of the efficacy of decongestion therapy and prognostic value of a wearable thoracic impedance sensor in acutely decompensated heart failure with volume overload: prospective cohort study. JMIR Cardio4, e12141 (2020). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 99.Snipelisky, D. et al. Accelerometer-measured daily activity in heart failure with preserved ejection fraction: clinical correlates and association with standard heart failure severity indices. Circ. Heart Fail10, e003878 (2017). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 100.Sohn, A. et al. Integrating remote monitoring into heart failure patients’ care regimen: a pilot study. PLoS One15, e0242210 (2020). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 101.Solar, H. et al. A non invasive, wearable sensor platform for multi-parametric remote monitoring in CHF patients. Health Technol.3, 99–109 (2013). [Google Scholar]
- 102.Stehlik, J. et al. Continuous wearable monitoring analytics predict heart failure hospitalization: the LINK-HF multicenter study. Circ. Heart Fail.13, e006513 (2020). [DOI] [PubMed] [Google Scholar]
- 103.Ueno, Y., Imamura, T., Narang, N. & Kinugawa, K. Chronotype of lung fluid levels in patients with chronic heart failure. J. Clin. Med.11, 2714 (2022). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 104.van den Berg-Emons, H. R., Bussmann, J. H., Balk, A., Keijzer-Oster, D. & Stam, H. Level of activities associated with mobility during everyday life in patients with chronic congestive heart failure as measured with an “activity monitor. Phys. Ther.81, 1502–1511 (2001). [PubMed] [Google Scholar]
- 105.van den Berg-Emons, R. J., Bussmann, J. B., Balk, A. H. & Stam, H. J. Factors associated with the level of movement-related everyday activity and quality of life in people with chronic heart failure. Phys. Ther.85, 1340–1348 (2005). [PubMed] [Google Scholar]
- 106.Vetrovsky, T. et al. The detrimental effect of COVID-19 nationwide quarantine on accelerometer-assessed physical activity of heart failure patients. ESC Heart Fail7, 2093–2097 (2020). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 107.Vetrovsky, T. et al. Lifestyle walking intervention for patients with heart failure with reduced ejection fraction: the WATCHFUL Trial. Circulation149, 177–188 (2024). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 108.Vetrovsky, T. et al. Validity of six consumer-level activity monitors for measuring steps in patients with chronic heart failure. PLoS One14, e0222569 (2019). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 109.Waring, T., Gross, K., Soucier, R. & ZuWallack, R. Measured physical activity and 30-day rehospitalization in heart failure patients. J. Cardiopulm. Rehabil. Prev.37, 124–129 (2017). [DOI] [PubMed] [Google Scholar]
- 110.Witham, M. D., Argo, I. S., Johnston, D. W., Struthers, A. D. & McMurdo, M. E. Predictors of exercise capacity and everyday activity in older heart failure patients. Eur. J. Heart Fail8, 203–207 (2006). [DOI] [PubMed] [Google Scholar]
- 111.Wong, C. K. et al. Daily ambulatory remote monitoring system for drug escalation in chronic heart failure with reduced ejection fraction: pilot phase of DAVID-HF study. Eur. Heart J. Digit Health3, 284–295 (2022). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 112.Yates, B. C., Pozehl, B., Kupzyk, K., Epstein, C. M. & Deka, P. Are heart failure and coronary artery bypass surgery patients meeting physical activity guidelines? Rehabil. Nurs.42, 119–124 (2017). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 113.Young, L., Hertzog, M. & Barnason, S. Feasibility of using accelerometer measurements to assess habitual physical activity in rural heart failure patients. Geriatrics2, 23 (2017). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 114.Vestergaard, A. S., Hansen, L., Sorensen, S. S., Jensen, M. B. & Ehlers, L. H. Is telehealthcare for heart failure patients cost-effective? An economic evaluation alongside the Danish TeleCare North heart failure trial. BMJ Open10, e031670 (2020). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 115.Lim, F. Y. et al. Correlation of the New York Heart Association classification and the cardiopulmonary exercise test: a systematic review. Int J. Cardiol.263, 88–93 (2018). [DOI] [PubMed] [Google Scholar]
- 116.Maric, B., Kaan, A., Ignaszewski, A. & Lear, S. A. A systematic review of telemonitoring technologies in heart failure. Eur. J. Heart Fail11, 506–517 (2009). [DOI] [PubMed] [Google Scholar]
- 117.Almeida Junior, G. L., Xavier, S. S., Garcia, M. I. & Clausell, N. Hemodynamic assessment in heart failure: role of physical examination and noninvasive methods. Arq. Bras. Cardiol.98, e15–e21 (2012). [DOI] [PubMed]
- 118.Anderson, K. M. et al. Perceived symptoms as the primary indicators for 30-day heart failure readmission. PLoS One17, e0267820 (2022). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 119.Scholte, N. T. B. et al. Photoplethysmography and intracardiac pressures: early insights from a pilot study. Eur. Heart J. Digit Health5, 379–383 (2024). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 120.van Smeden, M. et al. Critical appraisal of artificial intelligence-based prediction models for cardiovascular disease. Eur. Heart J.43, 2921–2930 (2022). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 121.Gill, S. K. et al. Artificial intelligence to enhance clinical value across the spectrum of cardiovascular healthcare. Eur. Heart J.44, 713–725 (2023). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 122.Kumar, S. et al. Wearables in cardiovascular disease. J. Cardiovasc Transl. Res16, 557–568 (2023). [DOI] [PubMed] [Google Scholar]
- 123.General Wellness: Policy for Low Risk Devices Guidance for Industry and Food and Drug Administration Staff Department: U.S. Department of Health and Human Services Food and Drug Administration Center for Devices and Radiological Health. (ed U.S. Department of Health and Human Services Food and Drug Administration Center for Devices and Radiological Health) FDA-2014-N-1039 (2019) https://www.fda.gov/media/90652/download.
- 124.Bianchini, E. & Mayer, C. C. Medical device regulation: should we care about it? Artery Res.28, 55–60 (2022). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 125.Bayoumy, K. et al. Smart wearable devices in cardiovascular care: where we are and how to move forward. Nat. Rev. Cardiol.18, 581–599 (2021). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 126.Musa, S. M. et al. Paucity of health data in Africa: an obstacle to digital health implementation and evidence-based practice. Public Health Rev.44, 1605821 (2023). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 127.DeVore, A. D., Fudim, M. & Lund, L. H. Novel trial designs in heart failure: using digital health tools to increase pragmatism. Curr. Heart Fail Rep.21, 5–10 (2024). [DOI] [PubMed] [Google Scholar]
- 128.Inan, O. T. et al. Digitizing clinical trials. NPJ Digit Med.3, 101 (2020). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 129.Lee, E. W. J. & Viswanath, K. Big data in context: addressing the twin perils of data absenteeism and chauvinism in the context of health disparities research. J. Med Internet Res.22, e16377 (2020). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 130.Holmes Fee, C. et al. Strategies and solutions to address Digital Determinants of Health (DDOH) across underinvested communities. PLOS Digit Health2, e0000314 (2023). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 131.Bruining, N. et al. Acquisition and analysis of cardiovascular signals on smartphones: potential, pitfalls and perspectives: by the Task Force of the e-Cardiology Working Group of European Society of Cardiology. Eur. J. Prev. Cardiol.21, 4–13 (2014). [DOI] [PubMed] [Google Scholar]
- 132.Haynes, N. et al. “Can you see my screen?” Addressing racial and ethnic disparities in telehealth. Curr. Cardiovasc Risk Rep.15, 23 (2021). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 133.van Kessel, R. et al. Digital health reimbursement strategies of 8 European countries and Israel: scoping review and policy mapping. JMIR Mhealth Uhealth11, e49003 (2023). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 134.Kim, J., Campbell, A. S., de Avila, B. E. & Wang, J. Wearable biosensors for healthcare monitoring. Nat. Biotechnol.37, 389–406 (2019). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 135.Naseri Jahfari, A., Tax, D., Reinders, M. & van der Bilt, I. Machine learning for cardiovascular outcomes from wearable data: systematic review from a technology readiness level point of view. JMIR Med Inf.10, e29434 (2022). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 136.Cohen, I. G. & Mello, M. M. Big data, big tech, and protecting patient privacy. JAMA322, 1141–1142 (2019). [DOI] [PubMed] [Google Scholar]
- 137.Page, M. J. et al. The PRISMA 2020 statement: an updated guideline for reporting systematic reviews. BMJ372, n71 (2021). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 138.AccessGUDID, https://accessgudid.nlm.nih.gov/devices/search.
- 139.Devices/SPPs - EUDAMEDhttps://ec.europa.eu/tools/eudamed/#/screen/search-device.