. 2024 Oct 26;14:25576. doi: 10.1038/s41598-024-77346-1

Grating-based phase-contrast computed tomography for breast tissue at an inverse compton source

Daniel Berthe 1,2,3,, Lisa Heck 1,2,3, Sandra Resch 1,2,3, Martin Dierolf 1,2, Johannes Brantl 1,2, Benedikt Günther 1,2, Christian Petrich 1,2,3, Klaus Achterhold 1,2, Franz Pfeiffer 1,2,4,5, Susanne Grandl 6, Karin Hellerhoff 6, Julia Herzen 1,2,3
PMCID: PMC11513984  PMID: 39462058

Abstract

The introduction of mammography screening programs has significantly reduced breast cancer mortality rates. Nevertheless, some lesions remain undetected, especially in dense breast tissue. Studies have shown that phase-contrast imaging can improve breast cancer diagnosis by increasing soft tissue contrast. Furthermore, grating-based phase-contrast imaging enables the simultaneous acquisition of absorption, phase-contrast, and scattering, so-called dark-field images. The latter allows the classification of microcalcifications. In addition, breast computed tomography (BCT) systems can identify and discriminate overlapping but clinically relevant structures. This study investigates the benefit of combining grating-based phase-contrast with BCT. We explore the potential of grating-based phase-contrast breast computed tomography (gbpc-BCT) with a breast phantom and a freshly dissected fibroadenoma. Improved image contrast could be achieved with radiation doses comparable to those used in clinical BCT.

Subject terms: Breast cancer, X-rays

Introduction

Globally, breast cancer is the most prevalent cancer type among women. In 2018, an estimated 2.1 million new cases and 627000 deaths were reported, accounting for 18%of all cancer deaths in women1. Early detection through mammography screening programs can reduce breast cancer mortality by up to 49%24. However, differentiation between benign lesions (i.e., cysts or fibroadenomas) remains a challenge leading to recall for additional imaging like ultrasound examination5. This is especially a problem for women with dense breast tissue who already have a significantly higher risk of developing breast cancer6.

A dedicated absorption-based photon counting BCT, which has been commercially available since recent years, could overcome the problem of overlapping structures by providing three-dimensional visualization of the breast at a radiation dose similar to standard mammography7,8. Initial results proved its potential in detecting suspicious masses9. Another advantage of BCT is its increased comfort over mammography8. Nevertheless, the challenge imposed by low soft-tissue contrast remains in absorption-based BCT. This can result in poor image contrast and often requires additional contrast agents10.

Phase-contrast imaging improves soft-tissue contrast without the administration of contrast agents. Unlike attenuation-based X-ray imaging, phase-contrast imaging utilizes the phase information of X-rays by measuring not only attenuation but also the refraction and the small-angle scattering signal induced by the specimen11. Various methods for obtaining the phase information exist, among others, propagation-based phase-contrast imaging1215, Talbot-Lau interferometry16,17, edge illumination18and coherent diffractive imaging19. However, most of these methods, except for the Talbot-Lau interferometer, require a large synchrotron facility or a complex and expensive liquid metal jet X-ray source and can accordingly not be used in a clinical setting. Therefore, the preferred method for potential clinical use is the Talbot-Lau interferometer-based phase-contrast setup, which is used in this work.

Several studies have demonstrated the advantages of grating-based phase-contrast imaging, particularly its contrast enhancement, in two-dimensional mammography2022. The additional phase information and the dark-field image show a diagnostic value by detecting the morphology of microcalcifications as a sign of early tumor malignancy2325. Here, the performance of a gbpc-BCT is investigated. Its advantages could be twofold compared to mammography: It could discriminate clinically relevant structures overlapping in projections while improving soft-tissue contrast. Raupach et al26. already showed that especially a BCT, which operates at a high spatial resolution, may benefit from the additional phase information in terms of reduced patient dose due to the enhanced soft tissue contrast.

Two modern Photon counting detector (PCD)s, which deliver the required resolution and show a high quantum efficiency in the relevant energy range27, were used to perform a gbpc-BCT at a quasi-monochromatic inverse Compton X-ray source, followed by a comparison of the resulting attenuation and phase-contrast images.

Our study demonstrates that the phase image of a gbpc-BCT significantly enhances visual contrast andcontrast-to-noise ratio (CNR) compared to attenuation-based images. We have shown this by measuring a breast phantom and a freshly dissected fibroadenoma using a laboratory gbpc-BCT setup. The results lay the foundation for further investigation of a gbpc-BCT at a polychromatic X-ray source.

Materials and methods

Detectors

The two used detectors were the SANTIS GaAs 0808 HR (75μmpixel size)28 and SANTIS GaAs 0804 ME (150μmpixel size)29 PCDs from DECTRIS Ltd. (Baden, Switzerland). Both detectors are equipped with a 500μmthick gallium arsenide sensor layer. The SANTIS GaAs 0808 HR has two adjustable energy thresholds28, and the SANTIS GaAs 0804 ME has four29. The used GaAs sensor layer has high quantum efficiency, especially at low photon energies up to 50KeVwhich are relevant for mammography, making it ideal for the presented measurements27,30. In addition the PCDs provide a higher resolution compared to the commonly used flat panel detector, which is also crucial for reducing the patient dose26. For the conducted measurement, the images of the SANTIS GaAs 0808 HR were binned twice, resulting in the same pixel size of 150μm for both detectors.

Image acquisition at the MuCLS

The measurements were conducted at the Munich Compact Light Source (MuCLS) (cf. Fig. 1). The setup consists of an inverse Compton source (Lyncean Technologies Inc., Fremont, USA) that produces a highly brilliant, spatially coherent, and quasi-monoenergetic X-ray beam31and a dedicated imaging beamline, developed and installed by the Technical University of Munich, featuring two measuring hutches32 .

Fig. 1.

Fig. 1

Scheme of the experimental setup. On the left, the Compact Light Source, consisting of an electron storage ring and a laser cavity, is sketched. X-ray photons are generated at the interaction point and illuminate the sample 14.93m downstream. Behind the sample, a Talbot-Lau grating interferometer and the detector are installed. Please note, that the proportions are not to scale.

The sample was mounted on a rotation stage 14.93m downstream of the interaction point in a 30mm thick water container to avoid phase-wrapping. The Talbot-Lau grating interferometer consisting of a phase grating G1 and an analyzer grating G2 was set to an inter-grating distance dG of 34.50cm11. The periods of the phase and analyzer grating were 4.9μm and 5.0μm, respectively. The measurements were conducted at an X-ray energy of 35KeV with a visibility of 35%. A movable detector stage was positioned 16.05m downstream of the interaction point and directly behind the analyzer grating, which resulted in an effective pixel size of 139.6μm. The acquisition parameters for the measurements are listed in Tab. 1. The samples were measured using a step-and-shoot approach. This means that for each projection angle, the analyzer grating was moved in five steps over the distance of one grating period, with an image taken at each position33. The resulting sinusoidal intensity curve for each pixel is called stepping-curve.

Table 1.

Acquisition parameters.

Sample Detector Eff. pixel size Exp. Time Reconstruction
Algorithm
Angles Mean glandular dose
Breast Phantom

SANTIS GaAs

0808 HR

139.6 40 FBP 650 41
163 10
Fibroadenoma

SANTIS GaAs

0804 ME

139.6 60 SIR 400 26
140 9

Processing

An expectation-maximization algorithm retrieves the attenuation and phase information from the sinusoidal stepping curve for all pixels34.

Especially for applications such as BCT, the dose for each projection is very low, which presents a challenge in accurately extracting the phase and attenuation information from the stepping curve. One method to improve image quality is to bin multiple pixels, which reduces noise but lowers image resolution. We applied patchwise phase retrieval, which reduces the noise without reducing the number of pixels but at the expense of image sharpness. It relies on the assumption that the sample does not contain sharp edges or other abrupt spatial changes. The assumption allows using the surrounding pixels as extra information in the expectation-maximization algorithm for each central pixel. This means we can calculate the phase and attenuation value for a three-by-three pixel patch and assign the resulting value to the central pixel in the patch.

That improves the accuracy of the fit for the signal retrieval, resulting in less noise in the final attenuation and phase-contrast images35. Afterwards, the projections were reconstructed using a Filtered back projection (FBP) or Statistical iterative reconstruction (SIR)36.

Radiation dose

The applied Mean glandular dose (MGD) was calculated by the SIERRA Monte Carlo simulation-based model developed by Boone et al37. The model was originally developed for mammography. When it comes to BCT, the dose is calculated for a single projection and then multiplied by the number of frames. It takes tabulated monoenergetic normalized glandular dose coefficients (DgN), which depend on the photon energy E, glandularity g, and compressed breast thickness t, and the incident air Kerma K into account to calculate the MGD for arbitrary spectra:

MGD=E=EminEmaxKEDgNE,t,gκ. 1

Here, κ denotes a conversion factor from air kerma to exposure. The DgN values are tabulated for a glandularity of 0%, 50% and 100%. In the following, a glandularity of 50% was assumed. All samples completely filled the Falcon tubes which they were measured in, allowing the tube diameter to be used directly as the sample thickness for the MGD calculation. Furthermore, the X-ray spectrum was measured using an energy-dispersive detector (Amptek X-123, Amptek Inc., USA).

Samples

The study was conducted under the Declaration of Helsinki and approved by the local ethics committee (Ethik-Kommission der Bayerischen Landesärztekammer (BLAEK), number 19063, date of permission 30/09/2019), and informed consent was obtained from the patient. For the conducted gbpc-BCT measurements, a breast-like phantom made from pork neck and a freshly dissected fibroadenoma provided by the Red Cross Hospital Munich were used. The breast phantom consists of a fatty piece of pork to mimic the adipose breast tissue (cf. Fig. 2(a)). It was fixated in a 4% formaldehyde solution before being immersed into a Falcon tube containing a 70% ethanol solution. An Eppendorf tube filled with a 6mg/mLiodine solution (IMERON 400 MCT, Bracco Imaging Deutschland GmbH) was added. The iodine concentration mimics a realistic contrast agent concentration in a patient’s blood vessels38. A polymethyl methacrylate (PMMA) rod was added for a possible energy calibration.

Fig. 2.

Fig. 2

Sample images. A sagittal slice of a high dose phase-contrast CT of the breast phantom with all the added features is shown in (a). Prior to the surgical removal of the fibroadenoma, a mammography image was taken. The fibroadenoma shows up as a bright structure in the mammography image (b) which correlates to a dark region in the ultra sound image (c).

Clinical mammography and ultrasound examination were performed before the surgical removal of the fibroadenoma. The mammography (Fig. 2b) revealed a 3.6cm lesion marked with a blue circle. Although the lesion is also visible on ultrasound (Fig. 2c), further Core-needle biopsy was required to differentiate between a benign fibroadenoma and a malignant carcinoma. After surgical removal, the fibroadenoma was embedded in a 4% formaldehyde solution for fixation, and a PMMA rod for energy calibration was added.

Results

As a first proof of principle, a gbpc-BCT with a breast phantom was conducted with the measurement parameters listed in Table 1. Figure 3 shows the reconstructed axial slices of the attenuation (a, b) and phase-contrast images (c, d). These images were all extracted from the same scan, but with different numbers of projections to modify the dose. The first column (a, c) shows the results obtained with a higher dose of 41mGy, while the low-dose images displayed in the second column (b, d) are achieved with a dose of 10mGy.

Fig. 3.

Fig. 3

Low dose gbpc-BCT with a breast phantom. Axial attenuation-contrast (a, b) and phase-contrast (c, d) reconstructions measured with a MGD of 41mGy in the first column and 10mGy in the second column, respectively. The arrows mark regions, where the contrast enhancement of gbpc-BCT is particularly evident.

The reconstructions show visually a superior contrast of the phase-contrast images compared to the attenuation-contrast images at both radiation doses. This is particularly evident in the region marked with a blue arrow where the dark adipose tissue is indistinguishable from the muscle tissue in the low-dose attenuation image but is visible in the phase-contrast image. Furthermore, the thin structure indicated by the red arrow in the phase-contrast images can still be seen, even in the image with a reduced dose, while it is not visible in the higher or lower dose attenuation-contrast images. The attenuation and phase images are extracted from the same scan with gratings in the beam, which results in almost twice the radiation dose compared to a standard BCT, due to the additional analyser grating between the sample and the detector.

For a more realistic investigation of gbpc-BCT, a freshly dissected fibroadenoma was scanned. Figure 4 shows an axial slice of the reconstructed attenuation (a, b) and phase-contrast images (c, d) of the fibroadenoma. The first column depicts a high-dose measurement at 26mGy, which is used as a ground truth for the low-dose measurement in the second column at 9mGy. The lower dose was achieved by reducing the number of angles used for the reconstruction.

Fig. 4.

Fig. 4

Dose-dependent comparison of the fibroadenoma’s attenuation and phase-contrast images. The images were taken at a dose of 26mGy (a, c) and 9mGy (b, d). The first row shows the attenuation-contrast images and the second row depicts the phase-contrast images extracted from the same measurement. Adipose tissue, indicated by the blue arrow, exhibits a strong contrast in both image modalities, whereas low contrast structures, indicated by the red arrow, can only be identified in the phase-contrast images (c) and (d). The green boxes in (c) mark the sample regions used for the calculation of the CNR.

In the attenuation-contrast images in Fig. 4(a, b), no contrast or structure is visible except for the adipose tissue, indicated by a blue arrow in the top right section. This is due to the homogeneous structure of the fibroadenoma. Since its deviation in attenuation is very small compared to the surrounding tissue, it cannot be resolved in the attenuation images.

The high-dose phase-contrast image reveals a cell-rich region in the center of the fibroadenoma, indicated by a red arrow, cf. Fig. 4(c). This region is visible in the low-dose phase-contrast image, although the noise level is higher, whereas it is not in the attenuation-contrast images. The CNR values in Table 2 calculated for this cell-rich region marked in Fig.4(c) support this observation. The CNR values for the attenuation-contrast images are below one, indicating only noise in the region. In contrast the CNR values for the phase-contrast images are more than ten times higher.

Table 2.

CNR analysis of the fibroadenoma. The CNR of the phase-contrast images reveal a significantly increased CNR compared to the attenuation images. Both image modalities are obtained from the same dataset with gratings in the beam, resulting in approximately double the dose compared to a scan without gratings.

Modality 26mGy 9mGy
Attenuation 0.26 0.13
Phase Contrast 3.31 1.91

Discussion and conclusion

We have demonstrated that gbpc-BCT significantly enhances the visual contrast and CNR compared to an attenuation-based BCT. Therefore we measured a breast phantom and a freshly dissected fibroadenoma at a laboratory gbpc-BCT setup with a quasi-monoenergetic X-ray source. Thereby the achieved MGD of 10mGy is only slightly higher than the one of the commercial BCT devices used in clinics7. However, because the method offers three complementary image modalities, namely attenuation, phase contrast and dark field, and consequently a drastic information gain compared to a conventional BCT, the additional dose might be justified for diagnostic examinations, yet further evaluation is needed to prove this.

The phase contrast and attenuation images shown here were obtained from the same measurements with gratings in the beam. The analyzer grating between the sample and detector absorbs about half the radiation dose, resulting in almost twice the MGD compared to a BCT measurement without gratings. However, the phase-contrast image of the fibroadenoma measurement at 9mGy shows a CNR of 1.91, whereas the attenuation channel at 2.8 times the dose has only a CNR of 0.26. Therefore, despite the attenuation image receiving more than twice the dose, the phase-contrast image still exhibits a significantly higher CNR.

The dark-field image is beneficial for detecting and classifying microcalcifications and discriminating between benign and malignant cysts34,39, but was neglected in this work because the samples did not contain small-angle scattering structures such as calcifications, and the focus was on extracting the phase information in a low-dose gbpc-BCT.

Nevertheless, the dose remains a challenge which needs to be tackled in order to get the method closer to a clinical practice. The MGD for BCT scans is often calculated with special DgNCTvalues, accounting for the rounded shape of the uncompressed breast40. However, since these factors have not been simulated for the required volume and diameter, we used the DgN factors in our study. These factors overestimate the volume and, as a result, the dose. Therefore, the calculated dose represents an upper limit of the actual dose administered due to the exaggerated volume of the sample. Furthermore, we had limited time to scan our clinical samples, because of the need for prompt histological analysis. This resulted in samples with a maximum diameter of 2.5cm, which is significantly smaller than the average breast size. While smaller samples are more dose-efficient, achieving comparable imaging quality in a complete breast would require a higher radiation dose. Additionally, the measurements were conducted using a quasi-monochromatic X-ray source to assess the maximum benefit of gbpc-BCT as a preliminary proof of concept before transitioning to a more challenging polychromatic X-ray source. In particular, for measurements with a polychromatic X-ray source, the efficiency of the interferometer decreases compared to the monoenergetic source. However, Rawlik et al. demonstrated that due to its increased soft tissue contrast a gbpc-BCT can even be more dose-efficient than a conventional BCT41.

Furthermore, self-supervised convolutional denoising algorithms like Noise2Inverse42,which are in already well tested with conventional CT, reduce the MGD while maintaining the CNR, to get closer to a clinically relevant dose. On the other hand, more advanced reconstruction techniques have to be applied, such as intensity-based statistical iterative reconstruction, which requires just one single image per angular position to retrieve the phase and attenuation information43.

In summary, gbpc-BCT combines the advantages of phase contrast imaging and BCT, namely, higher soft tissue contrast, no superimposed tissue structures, and higher patient comfort. Overall, it is a promising solution for breast cancer detection and diagnosis and could, in the future, allow to reduce the dose for X-ray based breast examinations.

Acknowledgements

We acknowledge gratefully Dectris Ltd. for supplying us with the detectors and especially Thomas Thüring for the support and fruitful discussions. The study was funded through the EQAP project, which is part of the One-Munich Strategy of the Bavarian ministry of science and arts.

Author contributions

All authors participated in various stages of manuscript preparation. This includes study design, sample preparation, data collection and processing, and writing and revising of the final manuscript. All authors read and approved the final manuscript.

Funding

Open Access funding enabled and organized by Projekt DEAL.

Data Availability

The data that support the findings of this study are available from the corresponding author on reasonable request.

Declarations

Competing interests

All authors declare no financial or non-financial competing interests.

Footnotes

Publisher’s note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

References

  • 1.Wild, C., Weiderpass, E. & Stewart, B. World cancer report 2020: Cancer research for cancer prevention.
  • 2.Duffy, S. W. et al. Beneficial effect of consecutive screening mammography examinations on mortality from breast cancer: A prospective study. Radiology 299, 541–547. 10.1148/radiol.2021203935 (2021). [DOI] [PubMed] [Google Scholar]
  • 3.Siu, A. L. Screening for breast cancer: U.s. preventive services task force recommendation statement. Annals of internal medicine 164, 279–296. 10.7326/M15-2886 (2016). [DOI] [PubMed] [Google Scholar]
  • 4.Hendrick, R. E., Baker, J. A. & Helvie, M. A. Breast cancer deaths averted over 3 decades. Cancer 125, 1482–1488. 10.1002/cncr.31954 (2019). [DOI] [PubMed] [Google Scholar]
  • 5.Berg, W. A. et al. Detection of breast cancer with addition of annual screening ultrasound or a single screening MRI to mammography in women with elevated breast cancer risk. JAMA 307, 1394–1404. 10.1001/jama.2012.388 (2012). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Wang, A. T., Vachon, C. M., Brandt, K. R. & Ghosh, K. Breast density and breast cancer risk: a practical review. Mayo Clinic proceedings 89, 548–557. 10.1016/j.mayocp.2013.12.014 (2014). [DOI] [PubMed] [Google Scholar]
  • 7.Berger, N. et al. Dedicated breast computed tomography with a photon-counting detector: Initial results of clinical in vivo imaging. Investigative radiology 54, 409–418. 10.1097/RLI.0000000000000552 (2019). [DOI] [PubMed] [Google Scholar]
  • 8.Sarno, A., Mettivier, G. & Russo, P. Dedicated breast computed tomography: Basic aspects. Medical physics 42, 2786–2804. 10.1118/1.4919441 (2015). [DOI] [PubMed] [Google Scholar]
  • 9.Lindfors, K. K. et al. Dedicated breast CT: initial clinical experience. Radiology 246, 725–733. 10.1148/radiol.2463070410 (2008). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10.Prionas, N. D. et al. Contrast-enhanced dedicated breast CT: initial clinical experience. Radiology 256, 714–723. 10.1148/radiol.10092311 (2010). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Pfeiffer, F., Weitkamp, T., Bunk, O. & David, C. Phase retrieval and differential phase-contrast imaging with low-brilliance x-ray sources. Nature Physics 2, 258–261. 10.1038/nphys265 (2006). [Google Scholar]
  • 12.Pacilè, S. et al. Free propagation phase-contrast breast CT provides higher image quality than cone-beam breast-CT at low radiation doses: a feasibility study on human mastectomies. Scientific reports 9, 13762. 10.1038/s41598-019-50075-6 (2019). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13.Longo, R. et al. Advancements towards the implementation of clinical phase-contrast breast computed tomography at elettra. Journal of synchrotron radiation 26, 1343–1353. 10.1107/S1600577519005502 (2019). [DOI] [PubMed] [Google Scholar]
  • 14.Tavakoli Taba, S. et al. Comparison of propagation-based CT using synchrotron radiation and conventional cone-beam CT for breast imaging. European radiology 30, 2740–2750. 10.1007/s00330-019-06567-0 (2020). [DOI] [PubMed] [Google Scholar]
  • 15.Tavakoli Taba, S. et al. Propagation-based phase-contrast CT of the breast demonstrates higher quality than conventional absorption-based CT even at lower radiation dose. Academic radiology 28, e20–e26. 10.1016/j.acra.2020.01.009 (2021). [DOI] [PubMed] [Google Scholar]
  • 16.Weitkamp, T., Diaz, A., David, C., Pfeiffer, F. & Stampanoni, M. X-ray phase imaging with a grating interferometer. OPTICS EXPRESS (2005). [DOI] [PubMed]
  • 17.Momose, A. et al. Demonstration of x-ray talbot interferometry. Japanese Journal of Applied Physics 42, L866–L868. 10.1143/JJAP.42.L866 (2003). [Google Scholar]
  • 18.Havariyoun, G. et al. A compact system for intraoperative specimen imaging based on edge illumination x-ray phase contrast. Physics in medicine and biology 64, 235005. 10.1088/1361-6560/ab4912 (2019). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19.Miao, J., Sandberg, R. L. & Song, C. Coherent x-ray diffraction imaging. IEEE Journal of Selected Topics in Quantum Electronics 18, 399–410. 10.1109/JSTQE.2011.2157306 (2012). [Google Scholar]
  • 20.Grandl, S. et al. Evaluation of phase-contrast CT of breast tissue at conventional x-ray sources - presentation of selected findings. Zeitschrift fur medizinische Physik 23, 212–221. 10.1016/j.zemedi.2013.02.005 (2013). [DOI] [PubMed] [Google Scholar]
  • 21.Stampanoni, M. et al. The first analysis and clinical evaluation of native breast tissue using differential phase-contrast mammography. Investigative radiology 46, 801–806. 10.1097/RLI.0b013e31822a585f (2011). [DOI] [PubMed] [Google Scholar]
  • 22.Auweter, S. D. et al. X-ray phase-contrast imaging of the breast-advances towards clinical implementation. The British Journal of Radiology 87, 20130606. 10.1259/bjr.20130606 (2014). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23.Bagnall, M. J. et al. Predicting invasion in mammographically detected microcalcification. Clinical radiology 56, 828–832. 10.1053/crad.2001.0779 (2001). [DOI] [PubMed] [Google Scholar]
  • 24.Wang, Z. et al. Non-invasive classification of microcalcifications with phase-contrast x-ray mammography. Nature communications 5, 3797. 10.1038/ncomms4797 (2014). [DOI] [PubMed] [Google Scholar]
  • 25.Scherer, K. et al. Correspondence: Quantitative evaluation of x-ray dark-field images for microcalcification analysis in mammography. Nature communications 7, 10863. 10.1038/ncomms10863 (2016). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 26.Raupach, R. & Flohr, T. Performance evaluation of x-ray differential phase contrast computed tomography (PCT) with respect to medical imaging. Medical physics 39, 4761–4774. 10.1118/1.4736529 (2012). [DOI] [PubMed] [Google Scholar]
  • 27.Scholz, J. et al. Biomedical x-ray imaging with a GaAs photon-counting detector: A comparative study. APL Photonics 5, 106108. 10.1063/5.0020262 (2020). [Google Scholar]
  • 28.Technical specifications santis 0808 GaAs.
  • 29.Technical specifications santis 150 prototype.
  • 30.Gkoumas, S. et al. Dose-independent near-ideal DQE of a 75um pixel GaAs photon counting spectral detector for breast imaging. In Bosmans, H., Chen, G.-H. & Gilat Schmidt, T. (eds.) Medical Imaging 2019: Physics of Medical Imaging, 30, 10.1117/12.2512235 (SPIE, 16.02.2019 - 21.02.2019).
  • 31.Eggl, E. et al. The munich compact light source: initial performance measures. Journal of synchrotron radiation 23, 1137–1142. 10.1107/S160057751600967X (2016). [DOI] [PubMed] [Google Scholar]
  • 32.Günther, B. S. Storage Ring-Based Inverse Compton X-ray Sources: Cavity Design, Beamline Development and X-ray Applications. Springer Theses, Recognizing Outstanding Ph.D. Research (Springer International Publishing and Imprint Springer, Cham, 2023), 1st ed. 2023 edn.
  • 33.Weitkamp, T. et al. X-ray phase imaging with a grating interferometer. OPTICS EXPRESS 13, 6296–6304. 10.1364/OPEX.13.006296 (2005). [DOI] [PubMed] [Google Scholar]
  • 34.Wang, D. et al. A dual detector approach for x-ray differential phase contrast imaging. Radiation Physics and Chemistry 95, 86–90. 10.1016/j.radphyschem.2012.12.027 (2014). [Google Scholar]
  • 35.Haeusele, J. et al. Advanced phase-retrieval for stepping-free x-ray dark-field computed tomography. IEEE Transactions on Medical Imaging 10, 10.1109/TMI.2023.3271413 (2023). [DOI] [PubMed]
  • 36.Hahn, D. et al. Statistical iterative reconstruction algorithm for x-ray phase-contrast CT. Sci Rep 5, 10452. 10.1038/srep10452 (2015). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 37.Boone, J. M. Normalized glandular dose (DgN) coefficients for arbitrary x-ray spectra in mammography: computer-fit values of monte carlo derived data. Medical physics 29, 869–875. 10.1118/1.1472499 (2002). [DOI] [PubMed] [Google Scholar]
  • 38.Wadsworth, G. The blood volume of normal women. Blood 9, 1205–1207. 10.1182/blood.V9.12.1205.1205 (1954). [PubMed] [Google Scholar]
  • 39.Scherer, K. et al. Improved diagnostics by assessing the micromorphology of breast calcifications via x-ray dark-field radiography. Scientific reports 6, 36991. 10.1038/srep36991 (2016). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 40.Boone, J. M., Shah, N. & Nelson, T. R. A comprehensive analysis of DgN(CT) coefficients for pendant-geometry cone-beam breast computed tomography. Medical Physics[SPACE]10.1118/1.1636571 (2004). [DOI] [PubMed] [Google Scholar]
  • 41.Rawlik, M. et al. Increased dose efficiency of breast CT with grating interferometry. Optica 10, 938. 10.1364/OPTICA.487795 (2023). [Google Scholar]
  • 42.Hendriksen, A. A., Pelt, D. M. & Batenburg, K. J. Noise2inverse: Self-supervised deep convolutional denoising for tomography. IEEE Transactions on Computational Imaging 6, 1320–1335. 10.1109/TCI.2020.3019647 (2020). [Google Scholar]
  • 43.von Teuffenbach, M. et al. Grating-based phase-contrast and dark-field computed tomography: a single-shot method. Scientific reports 7, 7476. 10.1038/s41598-017-06729-4 (2017). [DOI] [PMC free article] [PubMed] [Google Scholar]