. 2024 Oct 26;24:681. doi: 10.1186/s12887-024-05110-5

Hemodynamic monitoring during weaning from mechanical ventilation in critically ill pediatric patients: a prospective observational study

Tarek Ahmed Abdelgawad 1, Hanan M Ibrahim 1, Eman Mohamed Elsayed 2, Nehad Salah Abdelhamid 1,, Somia Abdel Hamid Bawady 3, Ahmed R Rezk 1
PMCID: PMC11515244  PMID: 39462317

Abstract

Background

Cardiovascular dysfunction is a significant factor contributing to weaning failure in mechanically ventilated children. Understanding the cardiopulmonary pathophysiological changes that occur during weaning is a prerequisite for the early recognition of weaning failure of cardiovascular origin. This study aimed to assess the effect of weaning trials on central hemodynamics and to identify the indices predictive of cardiac-related weaning failure.

Methods

This prospective observational study was conducted in the Pediatric Intensive Care Unit (PICU) and included mechanically ventilated patients aged between 2 and 30 months who were on minimal ventilatory settings and ready for weaning. Patients who were hemodynamically unstable, diagnosed with neuromuscular diseases, or diagnosed with cardiac diseases were excluded. Hemodynamic parameters were evaluated during weaning from ventilation via echocardiography and noninvasive cardiometry during pressure support (PS) ventilation and at the end of the spontaneous breathing trial (SBT).

Results

The study included 50 patients, comprising 30 males (60%) and 20 females (40%) with ages ranging from 2 to 30 months. Echocardiography revealed a significant increase in the cardiac index (CI), tricuspid annular plane systolic excursion (TAPSE), and the E/A ratio at the end of SBT. Moreover, right ventricular systolic pressure (RVSP) significantly decreased. Noninvasive cardiometry revealed a significant increase in the index of contractility (ICON) and CI at the end of SBT (p-value = 0.023 and < 0.001, respectively). Of the 12 (25%) patients who failed their first extubation trial, they exhibited a significantly lower CI and TAPSE (p values = 0.001 and 0.001, respectively).

Conclusion

This study identified that weaning from mechanical ventilation in children is associated with hemodynamic changes, which can impact weaning success and reveal potential ventricular dysfunction. Bedside echocardiography was found to detect cardiac dysfunctions during weaning, and noninvasive cardiometry was considered a reliable tool that supports echocardiography for detecting changing trends in CI in PICUs. However, accurate values should be confirmed by echocardiography.

Keywords: Pediatric intensive care unit, Echocardiography, Mechanical ventilation, Weaning, Noninvasive cardiometry, Weaning failure, Pediatrics, Hemodynamic changes

Introduction

Mechanical ventilation is a commonly used procedure in critically ill children admitted to pediatric intensive care units (PICUs), yet it is associated with potential major side effects. Prolonged mechanical ventilation increases the risk of infections, morbidity, and mortality [1]. Understanding the cardiopulmonary pathophysiology changes during mechanical ventilation and its withdrawal is essential for early identification of weaning failure of cardiovascular origin. Weaning success depends on the ability of the cardiorespiratory system to tolerate these changes [2]. Focused critical care echocardiography (FCCE) is increasingly being performed and interpreted by trained clinicians as an essential bedside tool [3]. Noninvasive technologies such as thoracic electrical bioimpedance, thoracic bioreactance, vascular unloading technique, pulse wave transit time, and radial artery applanation tonometry are currently available for cardiac output monitoring [4]. This study aimed to assess different hemodynamic changes occurring during the weaning of the pediatric population from mechanical ventilation, and to determine cardiovascular-related predictors of weaning failure.

Methods

Subjects

This prospective observational study included 50 children, aged 2 to 30 months, admitted to the PICU at Ain Shams University Children’s Hospital between May 2021 and September 2022. The study population included mechanically ventilated pediatric patients ready to be weaned according to the following criteria: (i) Clinical criteria: enrolled patients were awake and responsive with an intact cough reflex; no sedation was administered at the time of examination; hemodynamically stable (i.e., capillary refill time < 3 s; age-appropriate blood pressure; heart rate with no or minimum vasopressor support); physiological and metabolic status (i.e., normal ranges of serum electrolytes and glucose, body temperature < 38 °C, hemoglobin level ≥ 8–10 g/dL), adequate oxygenation; pulsed oxygen saturation (SpO2 > 95%); and fair nutritional status. Additional criteria included (ii) arterial blood gases: normal pH, partial pressure of arterial oxygen (PaO2) > 60 mmHg, and partial pressure of arterial carbon dioxide (PaCO2) < 50 mmHg; and (iii) respiratory mechanics: fractionated inspired oxygen (FiO2) < 40%, PaO2/FiO2 > 200, positive end-expiratory pressure (PEEP) ≤ 5 cm H2O, respiratory rate (RR) < 40 breaths/minute, and tidal volume (VT) > 5 ml/kg. Patients were excluded if they were hemodynamically unstable, diagnosed with neuromuscular diseases not predicted to be weaned, or diagnosed with any cardiac disease.

Study design and data collection

The primary objective of this study was to evaluate various hemodynamic changes that occur during the weaning process from mechanical ventilation in the pediatric population and to identify cardiovascular predictors of weaning failure. The secondary objective was to evaluate the effectiveness of noninvasive cardiometry in assessing hemodynamic changes compared to echocardiography. All enrolled patients were subjected to a full, detailed medical history and thorough clinical examination.

Regarding the weaning process, we utilized PS ventilation with settings of PEEP 5 cm H2O, PS 6–8 cm H2O, and FiO2 30–40% to provide minimal yet effective ventilatory assistance to spontaneously breathing patients. We conducted a 30-minute spontaneous breathing trial (SBT) with zero pressure support (PS), 5 cm H2O positive end-expiratory pressure (PEEP), low trigger settings, tube compensation, and FiO2 of 30–40%. The use of zero PS over T-piece SBT aims to leverage the benefits of maintaining patient-ventilator connection, ensuring adequate humidification, monitoring expiratory volume, and allowing rapid introduction of ventilation if needed [5]. All patients were extubated on nasal continuous positive airway pressure (CPAP) as per our ICU protocol. Ventilator-withdrawal failure was defined as the requirement to end SBT before completing 30 min or the need for re-intubation within 48 h of extubation. Clinical failure criteria included signs of increased respiratory muscle effort, reduced level of consciousness, uncontrollable psychomotor agitation requiring sedation, excessive sweating, and cyanosis. Additionally, failure was indicated by PaO2 levels below 50–60 mmHg, PaCO2 levels above 50 mmHg or an increase of more than 8 mmHg, SaO2 below 88–90% with FiO2 above 0.5, pH below 7.32 or a decrease of more than 0.07, respiratory rate exceeding 40 breaths/min, tachycardia or arrhythmias, high blood pressure for age, or increased blood pressure of ≥ 20% [6]. Extubation failure was defined as the need for re-intubation within 48 h of extubation.

Mechanical ventilator settings, including FiO2, peak inspiratory pressure (PIP), PEEP, mean airway pressure (MAP), RR, VT, and minute ventilation (VT × RR), were recorded for all patients before initiation and at the termination of the 30-minute SBT. All patients were subjected to the following tools before initiating and just after ending a 30-minute SBT. Transthoracic echocardiography was performed by a single observer under the supervision of a pediatric cardiologist using a 3–8 MHz pediatric cardiac transducer using the Samsung HM70A Ultrasound System to measure the left ventricular ejection fraction (LVEF), left ventricular (LV) stroke volume (SV), cardiac output, cardiac index (CI) was calculated, tricuspid annular plane systolic excursion (TAPSE), right ventricular (RV) and LV end-diastolic areas (EDA) to calculate the RVEDA/LVEDA ratio, maximal velocities of mitral E and A waves to calculate the E/A ratio, and right ventricular systolic pressure (RVSP). Noninvasive cardiometry was performed through electrical bioimpedance using the Cardiometer ICON® to measure CI, the index of contractility (ICON), systemic vascular resistance (SVR), corrected flow time (FTc), stroke volume variation (SVV), and thoracic fluid content (TFC).

Data analysis

The data were collected, revised, coded, and entered into the Statistical Package for Social Science (IBM SPSS) version 23. The quantitative data with a parametric distribution are presented as the mean, standard deviation, and range. Qualitative variables are presented as numbers and percentages. The comparison between groups regarding qualitative data was performed using the chi-square test and/or Fisher’s exact test when the expected count in any cell was less than 5. Comparisons between two independent groups with quantitative data and parametric distributions were performed using an independent t test, while comparisons of nonparametric data were performed using the Mann‒Whitney test. Normality was assessed using the Shapiro-Wilk test. Spearman correlation coefficients were calculated to assess the correlation between two quantitative parameters in the same group. A receiver operating characteristic (ROC) curve was used to assess the cutoff points for sensitivity, specificity, positive predictive value (PPV), negative predictive value (NPV), and area under the curve (AUC). Univariate logistic regression analysis was used to assess the odds ratio (OR) and 95% confidence interval (CI) of the most important factors associated with extubation failure. The confidence interval was set to 95%, and the margin of error accepted was set to 5%. PASS11 program for sample size calculation was used, assuming a weaning failure rate of 40%. This percentage was based on the extubation failure rate in our ICU in the year preceding study initiation. The sensitivity of echocardiology parameters for the prediction of failure was set at 80%, and specificity was set at 80%. With a sample size of 50 patients, it is possible to detect sensitivity with a power of 80% and specificity with a power of 94%, while setting the alpha error (type I error) at 0.05. Therefore, a p value > 0.05 was considered nonsignificant, a p value < 0.05 was considered significant, and a p value < 0.01 was considered highly significant.

Results

Fifty pediatric patients were included in this study–30 males (60%) and 20 females (40%). The ages of the patients ranged between 3 and 30 months, with a median (IQR) of 21 (8–28) months. Pneumonia was the most common cause of mechanical ventilation and was recorded in 16 patients (32%). Viral sepsis and septic shock were both recorded in 16% of the patient population. Thirty-eight patients (76%) had successful extubation during the first trial; however, 12 patients (24%) did not succeed during the first trial.

The mean (± SD) heart rate significantly increased at the end of SBT (130.10 ± 12.96) compared to that during PS (120.42 ± 16.25) (p-value < 0.001). Regarding echocardiography parameters, there were significant increases in CI, TAPSE, TAPSE Z score, and E/A ratio at the end of SBT compared to those during PS (p values < 0.001, < 0.001, < 0.001, and 0.002, respectively). In addition, there was a significant decrease in RVSP with a p value < 0.001 (Table 1).

Table 1.

Echocardiography parameters during PS and at the end of SBT

PS (n = 50)
(mean ± SD)
SBT (n = 50)
(mean ± SD)
Test
value
P-value
LVEF 60.20 ± 9.17 60.00 ± 7.56 0.227• 0.821
CI (L/m/m2) 4.53 ± 0.51 5.04 ± 0.59 -7.917• 0.000
TAPSE (cm) 1.72 ± 0.20 1.77 ± 0.21 -4.466• 0.000
Z-score of TAPSE* 1.4 (1–1.8) 1.65 (1.22–2.1) -4.676‡ 0.000
LV stroke volume (ml) 20.30 ± 8.41 20.33 ± 8.14 -0.031• 0.976
RVSP (mmHg) 21.48 ± 6.44 16.79 ± 5.17 7.693• 0.000
LVEDA/RVEDA 1.46 ± 0.30 1.53 ± 0.36 -1.586• 0.119
E/A 1.32 ± 0.35 1.45 ± 0.36 -3.201 0.002

• Independent t-test; ‡ Mann Whitney test

* Median (IQR)

Patients who failed extubation had lower TAPSE and TAPSE Z score during PS (p values = 0.011 and 0.039, respectively) and at the end of SBT (p values = 0.001 and 0.004, respectively) with lower CIs at the end of SBT (p value = 0.001) (Tables 2 and 3).

Table 2.

Relationships between extubation outcomes and echocardiography parameters during PS

Echocardiography parameters Success (n = 38)
mean ± SD
Failed (n = 12)
mean ± SD
Test value P-value
LVEF 61.31 ± 9.85 56.67 ± 5.52 1.552• 0.127
CI (L/m/m2) 4.56 ± 0.45 4.43 ± 0.69 0.813 0.420
TAPSE (cm) 1.76 ± 0.17 1.60 ± 0.24 2.629• 0.011
Z-score* 1.44 (1–1.83) 1 (0.5–1.45) -2.062‡ 0.039
LV stroke volume (ml) 20.58 ± 7.75 19.42 ± 10.59 0.414• 0.681
RVSP (mmHg) 21.37 ± 6.42 21.83 ± 6.75 -0.216• 0.830
LVEDA/RVEDA 1.46 ± 0.27 1.45 ± 0.40 0.065• 0.948
E/A 1.32 ± 0.33 1.35 ± 0.44 -0.259 0.797

• Independent t-test; ‡ Mann Whitney test

* Median (IQR)

Table 3.

Relationships between extubation results and echocardiography parameters at the end of SBT

Echocardiography parameter Success (n = 38)
(mean ± SD)
Failed (n = 12)
(mean ± SD)
Test value P-value
LVEF 60.87 ± 7.68 57.25 ± 6.72 1.463• 0.150
CI (L/m/m2) 5.19 ± 0.50 4.56 ± 0.63 3.559 0.001
TAPSE (cm) 1.82 ± 0.16 1.60 ± 0.25 3.508• 0.001
Z-score* 1.69 (1.4–2.1) 0.94 (0.52–1.55) -2.913‡ 0.004
LV stroke volume (ml) 20.41 ± 7.76 20.08 ± 9.63 0.119• 0.906
RVSP (mmHg) 16.49 ± 5.13 17.75 ± 5.40 -0.732• 0.468
LVEDA/RVEDA 1.55 ± 0.32 1.44 ± 0.47 0.919• 0.362
E/A 1.44 ± 0.34 1.49 ± 0.44 -0.480 0.634

• Independent t-test; ‡ Mann Whitney test

* Median (IQR)

The cutoff point for TAPSE during PS for differentiating between patients who failed extubation and those who were successfully extubated was ≤ 1.61 cm, with a sensitivity of 66.67%, specificity of 73.68%, and AUC of 0.695. In comparison, the cutoff point for TAPSE at the end of SBT for differentiating between patients who failed extubation and those who were successfully extubated was ≤ 1.65 cm with a sensitivity of 75.0%, specificity of 81.58% and AUC of 0.766 (Fig. 1).

Fig. 1.

Fig. 1

Receiver operating characteristic (ROC) curve of the TAPSE during PS and after SBT as a predictor of extubation failure

The cutoff point for CI by echo at the end of SBT to distinguish between patients who were not successfully extubated was ≤ 4.8 L/m2/min, with a sensitivity of 75.0%, specificity of 78.95%, and AUC of 0.796 (Fig. 2).

Fig. 2.

Fig. 2

Receiver operating characteristic (ROC) curve of CI by echocardiography after SBT as a predictor of failure

Noninvasive cardiometry revealed significant increases in ICON and CI at the end of SBT compared to those during PS (p values = 0.023 and < 0.001, respectively) (Table 4).

Table 4.

Noninvasive cardiometry parameters during PS and at the end of SBT the mean (± SD) CI values measured by noninvasive cardiometry (5.19 ± 0.72) were similar to those measured by echocardiography (5.04 ± 0.59) at the end of SBT (p-value = 0.105). However, the values measured by noninvasive cardiometry (4.89 ± 0.68) were greater than those measured by echocardiography (4.53 ± 0.51) during PS (p-value < 0.001). There was a positive correlation between CI measured by noninvasive cardiometry and echocardiography during PS and at the end of SBT (Fig. 3), and the difference was evaluated via a bland-Altman diagram (Fig. 4)

PS (n = 50)
(mean ± SD)
SBT (n = 50)
(mean ± SD)
Test value• p-value
ICON 83.38 ± 26.94 90.40 ± 29.87 -2.341 0.023
CI (L/m/m2) 4.89 ± 0.68 5.19 ± 0.72 -3.751 0.000
SVV 19.16 ± 7.75 19.24 ± 8.21 -0.102 0.919
TFC 43.34 ± 16.51 41.40 ± 18.60 0.871 0.388
FTC 268.06 ± 47.52 255.28 ± 47.65 1.958 0.056
SVR 3218.84 ± 1709.54 2992.08 ± 1146.67 1.992 0.052

•: Paired t-test

Fig. 3.

Fig. 3

Correlations between CI by echocardiography and noninvasive cardiometry. a shows the correlation between CI measured via echocardiography and noninvasive cardiometry during PS. b shows the correlation between CI measured via echocardiography and noninvasive cardiometry after SBT

Fig. 4.

Fig. 4

The difference between echocardiography and noninvasive cardiometry for measuring CI.:a Bland‒Altman plot showing the difference between echocardiography and noninvasive cardiometry measurements of CI during PS. b: Bland‒Altman plot showing the difference between echocardiography and noninvasive cardiometry measurements of CI after SBT

Discussion

Cardiovascular dysfunction is a recognizable factor contributing to weaning failure, as observed by previous studies in difficult-to-wean adults on mechanical ventilation [79]. The utilization of echocardiography to detect cardiovascular dysfunction in the intensive care setting has grown significantly over the past two decades [10]. This study aimed to assess the different hemodynamic changes that occur during the weaning of pediatric patients from PS mechanical ventilation to SBT. Hemodynamic instability was recognized during the transition from positive-pressure mechanical ventilation to spontaneous ventilation [11]. At the end of the SBT, significant increases in mean heart rate, CI, TAPSE, TAPSE Z score, and E/A ratio were recorded, while RVSP significantly decreased. Patients who failed extubation had lower TAPSE, TAPSE Z scores, and CI, indicating them as potential predictive parameters of extubation failure. Noninvasive cardiometry showed significant increases in ICON and CI at the end of SBT. Positive correlations between CI measured by noninvasive cardiometry and echocardiography were observed.

Heart rate increased significantly at the end of SBT compared to that during PS in our study. This increase was similar to that reported in studies by Trifi et al. [12] and Caille et al. [9], who reported that the transition to T-trial increased HR significantly. The elevation in heart rate during SBT could be attributed to an increase in adrenergic tone, likely resulting from rising serum catecholamine levels during the discontinuation of positive pressure ventilation, as previously documented [13].

Several guidelines and recommendations highlighted the importance of using echocardiography in a wide spectrum of clinical settings in the ICU, including those for cardiovascular dysfunction [4, 1416]. In this study, CI significantly increased at the end of SBT; however, there was no increase in stroke volume at the end of SBT. Therefore, we believe that this increase in CI was directly impacted by the increase in HR, which eventually influenced cardiac output. Trifi et al. reported that cardiac output significantly increased during SBT due to a significant increase in stroke volume [12]. However, the CI in our study was lower at the end of SBT in patients with failed weaning. Gerbaud et al. observed an increase in CI at the end-SBT in ventilated adult patients who were successfully weaned (3.3 [3.06–3.77] vs. 3 [2.68–3.3] L/min/m2, P < 0.001). CI remained unchanged in patients with weaning failure [17]. We suggest that the cutoff point for CI by echocardiography at the end of SBT to distinguish between failed and successful extubation cases is ≤ 4.8 L/m2/min, with a sensitivity of 75.00% and a specificity of 78.95%. This could provide a simple, noninvasive guidance for predicting weaning success in ventilated pediatric patients.

Diastolic dysfunction is an essential factor contributing to cardiac-related weaning failure. The E/A ratio increased at the end of SBT, suggesting an alteration in diastolic function in our study population. This alteration in diastolic dysfunction could be an impact of suggested factors, including pulmonary edema, ventricular dysfunction, myocardial ischemia, volume overload, or intrathoracic pressure variations during weaning. Bedet et al. [18] observed no difference in the E/A ratio in patients who were successfully weaned during the second SBT (p = 0.736); however, the E/A ratio significantly increased at the end of the second SBT in patients with failed weaning (p = 0.003). In our study, the E/A ratio in patients who achieved successful or failed weaning was similar at the end of SBT. Therefore, our study revealed that regardless of weaning outcome, the E/A ratio was not considered a predictor of weaning failure.

TAPSE significantly increased at the end of SBT, while lower TAPSE values during PS and SBT were associated with weaning failure. Similarly, Roche-Campo et al. [19] reported that TAPSE was also lower in patients with prolonged weaning (p = 0.03). Our study revealed that the cutoff point for TAPSE during PS for differentiating between failed and successful extubation was ≤ 1.61 cm, with a sensitivity of 66.67% and specificity of 73.68%. The cutoff point for TAPSE at the end of SBT for differentiating between failed and successful extubation was ≤ 1.65 cm, with a sensitivity of 75.0% and specificity of 81.58%. An adult study by Papaioannou et al. [18] showed that patients with prolonged weaning had decreased TAPSE compared to those with short weaning (14.59 ± 1.56 versus 19.13 ± 2.59 mm, respectively). They recorded a cutoff point of 1.64 cm, with a sensitivity of 85.00% and specificity of 75.00%, predicting a prolonged weaning process. Right ventricular systolic dysfunction in prolonged weaning patients could be related to left ventricular diastolic dysfunction, unlike what was reported in adult chronic obstructive pulmonary disease (COPD) patients by Daif et al. [20], in which there was no difference in TAPSE values between failed and succeeded patients. Therefore, our study suggests that TAPSE could provide a simple, noninvasive guidance for predicting weaning failure in the pediatric population.

This study also compared echocardiography to noninvasive cardiometry in the assessment of CI during weaning. The absolute CI values measured by noninvasive cardiometry were greater than those measured by echocardiography. Sanders et al. [21] reported that electrical cardiometry cannot replace thermodilution or transthoracic echocardiography for the measurement of absolute cardiac output. However, our study revealed a significant positive correlation between CI measured by noninvasive cardiometry and echocardiography during PS and after SBT. Boet et al. [22] noted that the cardiac output measured by electrical cardiometry was positively correlated with cardiac output measured by echocardiography in hemodynamically stable preterm infants. Xu et al. [23] also concluded that although the absolute values of cardiac output measured by electrical cardiometry and M-mode echocardiography were not interchangeable, the cardiac output distributions in both was similar. On the basis of our findings, electrical cardiometry cannot replace transthoracic echocardiography, considered the standard method for noninvasive cardiac imaging and monitoring hemodynamics. However, electrical cardiometry has proven to be an acceptable operator-independent trending tool for both cardiac output and CI assessment in pediatric patients on mechanical ventilation.

Strengths and limitations

To our knowledge, this study is the first to assess hemodynamic changes during weaning from PS mechanical ventilation and predictors of weaning failure in the pediatric population. However, this study has some limitations that need to be accounted for. Our study did not assess heart rate variability due to the lack of measurement tools, including invasive arterial line insertion, which is not the usual practice in our PICU. Although the sample size may be considered low compared to that of adult studies, the number of patients is fairly reasonable, considering that this was a single-center observational study in a pediatric population. We believe that a larger sample size from more diverse pediatric populations will add valuable results to our findings. Although thermodilution is a more accurate method to assess cardiac output and CI, we could not use it in our study due to its invasive nature and associated complications.

Conclusion

This study found that weaning from mechanical ventilation is associated with hemodynamic changes, which can impact weaning success. Echocardiography is considered a valuable tool that may raise suspicion of ventricular dysfunction, which can contribute to weaning failure and require further evaluation. Additionally, noninvasive electrical cardiometry was highlighted as a supportive tool for monitoring hemodynamic trends in the PICU, though it should be complemented by echocardiography for accurate absolute measurements. This approach highlights the importance of thorough cardiovascular assessment during the weaning process and addresses a critical knowledge gap in pediatric intensive care. Echocardiography should be used as part of a comprehensive evaluation, not as the sole determinant of extubation readiness. Further research with larger, more diverse populations is needed to validate these findings and determine the optimal role of echocardiography in weaning protocols.

Acknowledgements

We would like to express our sincere gratitude to the medical and nursing staff of the Pediatric Intensive Care Unit at Ain Shams University Children’s Hospital for their support throughout the study. We are also grateful to the families of the patients involved in this study for their cooperation and trust.

Abbreviations

AUC

Area Under the Curve

CI

Cardiac Index

COPD

Chronic Obstructive Pulmonary Disease

CPAP

Continuous Positive Airway Pressure

E/A

Ratio of Early (E) to Late (A) Ventricular Filling Velocities

EDA

End-Diastolic Area

FCCE

Focused Critical Care Echocardiography

FTc

Corrected Flow Time

FiO2

Fraction of Inspired Oxygen

HR

Heart Rate

ICON

Index of Contractility

ICU

Intensive Care Unit

IQR

Interquartile Range

LVEDA

Left Ventricular End-Diastolic Area

LVEF

Left Ventricular Ejection Fraction

MAP

Mean Airway Pressure

NPV

Negative Predictive Value

OR

Odds Ratio

PEEP

Positive End-Expiratory Pressure

PICU

Pediatric Intensive Care Unit

PIP

Peak Inspiratory Pressure

PPV

Positive Predictive Value

PS

Pressure Support

PaCO2

Partial Pressure of Arterial Carbon Dioxide

PaO2

Partial Pressure of Arterial Oxygen

ROC

Receiver Operating Characteristic

RR

Respiratory Rate

RVEDA

Right Ventricular End-Diastolic Area

RVSP

Right Ventricular Systolic Pressure

SBT

Spontaneous Breathing Trial

SD

Standard Deviation

SV

Stroke Volume

SVR

Systemic Vascular Resistance

SVV

Stroke Volume Variation

SaO2

Oxygen Saturation

SpO2

Pulsed Oxygen Saturation

TAPSE

Tricuspid Annular Plane Systolic Excursion

TFC

Thoracic Fluid Content

VT

Tidal Volume

Author contributions

Conceptualization, T.A.A., H.M.I., and A.R.R.; methodology, A.R.R., N.S.A., and E.M.E.; software, N.S.A.; validation, H.M.I; formal analysis, N.S.A.; investigation, E.M.E., S.A.B., and N.S.A.; resources, H.M.I and T.A.A.; data curation, N.S.A. and S.A.B.; writing—original draft preparation, N.S.A.; writing—review and editing, A.R.R. and H.M.I.; visualization, T.A.A.; supervision, H.M.I. and A.R.R.; project administration, A.R.R. All authors have read and agreed to the published version of the manuscript.

Funding

Open access funding provided by The Science, Technology & Innovation Funding Authority (STDF) in cooperation with The Egyptian Knowledge Bank (EKB).

Data availability

The data presented in this study are available upon request from the corresponding author. The data are not publicly available due to the use of other unpublished articles based on this database.

Declarations

Ethics approval and consent to participate

The hospital’s institutional review board and the ethics committee of Ain Shams University approved this study (Approval number: FMASU MD 97, date of approval: May 2020). All procedures followed in our study were in accordance with the ethical standards of the hospital’s institutional review board and the University’s ethics committee, as well as with the principles of the Declaration of Helsinki of 1975, as revised in 2000. Written informed consent was obtained from all patient caregivers to participate in this study.

Consent for publication

Not applicable.

Competing interests

The authors declare no competing interests.

Footnotes

Publisher’s note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

References

  • 1.Elisa P, Francesca C, Marco P, Davide V, Laura Z, Fabrizio Z et al. Ventilation Weaning and Extubation Readiness in Children in Pediatric Intensive Care Unit: A Review. Front Pediatr [Internet]. 2022 Apr 1 [cited 2024 Jan 25];10. https://pubmed.ncbi.nlm.nih.gov/35433554/ [DOI] [PMC free article] [PubMed]
  • 2.Zieleskiewicz L, Muller L, Lakhal K, Meresse Z, Arbelot C, Bertrand PM et al. Point-of-care ultrasound in intensive care units: assessment of 1073 procedures in a multicentric, prospective, observational study. Intensive Care Med [Internet]. 2015 Sep 29 [cited 2024 Jan 25];41(9):1638–47. https://pubmed.ncbi.nlm.nih.gov/26160727/ [DOI] [PubMed]
  • 3.Saugel B, Cecconi M, Wagner JY, Reuter DA. Noninvasive continuous cardiac output monitoring in perioperative and intensive care medicine. Br J Anaesth [Internet]. 2015 Apr 1 [cited 2024 Jan 25];114(4):562–75. https://pubmed.ncbi.nlm.nih.gov/25596280/ [DOI] [PubMed]
  • 4.Papolos A, Narula J, Bavishi C, Chaudhry FA, Sengupta PP. U.S. Hospital Use of Echocardiography: Insights From the Nationwide Inpatient Sample. J Am Coll Cardiol [Internet]. 2016 Feb 9 [cited 2024 Jan 25];67(5):502–11. https://pubmed.ncbi.nlm.nih.gov/26846948/ [DOI] [PubMed]
  • 5.Burns KEA, Khan J, Phoophiboon V, Trivedi V, Gomez-Builes JC, Giammarioli B et al. Spontaneous Breathing Trial Techniques for Extubating Adults and Children Who Are Critically Ill: A Systematic Review and Meta-Analysis. JAMA Netw Open [Internet]. 2024 Feb 23 [cited 2024 Jun 24];7(2):E2356794. https://pubmed.ncbi.nlm.nih.gov/38393729/ [DOI] [PMC free article] [PubMed]
  • 6.Magnet FS, Heilf E, Huttmann SE, Callegari J, Schwarz SB, Storre JH et al. The spontaneous breathing trial is of low predictive value regarding spontaneous breathing ability in subjects with prolonged, unsuccessful weaning. Med Klin Intensivmed Notfmed [Internet]. 2020 May 1 [cited 2024 Jul 12];115(4):300. [DOI] [PMC free article] [PubMed]
  • 7.Cabello B, Thille AW, Roche-Campo F, Brochard L, Gómez FJ, Mancebo J. Physiological comparison of three spontaneous breathing trials in difficult-to-wean patients. Intensive Care Med [Internet]. 2010 Jul [cited 2024 Jan 30];36(7):1171–9. https://pubmed.ncbi.nlm.nih.gov/20352189/ [DOI] [PubMed]
  • 8.Epstein SK. Etiology of extubation failure and the predictive value of the rapid shallow breathing index. Am J Respir Crit Care Med [Internet]. 1995 [cited 2024 Jan 30];152(2):545–9. https://pubmed.ncbi.nlm.nih.gov/7633705/ [DOI] [PubMed]
  • 9.Caille V, Amiel JB, Charron C, Belliard G, Vieillard-Baron A, Vignon P. Echocardiography: a help in the weaning process. Crit Care [Internet]. 2010 Jun 22 [cited 2024 Jan 27];14(3). https://pubmed.ncbi.nlm.nih.gov/20569504/ [DOI] [PMC free article] [PubMed]
  • 10.Burk RE, Beesley SJ, Grissom CK, Hirshberg EL, Lanspa MJ, Brown SM. Echocardiography in the Intensive Care Unit. Curr Cardiovasc Imaging Rep [Internet]. 2017 Dec 1 [cited 2024 Jan 30];10(12):1–12. https://link.springer.com/article/10.1007/s12410-017-9438-9
  • 11.Frazier SK, Stone KS, Schertel ER, Moser DK, Pratt JW. A comparison of hemodynamic changes during the transition from mechanical ventilation to T-piece, pressure support, and continuous positive airway pressure in canines. Biol Res Nurs [Internet]. 2000 [cited 2024 Jan 30];1(4):253–64. https://pubmed.ncbi.nlm.nih.gov/11232204/ [DOI] [PubMed]
  • 12.Trifi A, Abdellatif S, Ben Ismail K, Daly F, Touil Y, Nasri R et al. Cronicon EC EMERGENCY MEDICINE AND CRITICAL CARE Cardiac Ultrasound Assessment during ventilator-withdrawal process: A Comparative Cohort Study.
  • 13.Routsi C, Stanopoulos I, Kokkoris S, Sideris A, Zakynthinos S. Weaning failure of cardiovascular origin: how to suspect, detect and treat-a review of the literature. Ann Intensive Care [Internet]. 2019 Dec 1 [cited 2024 Jan 27];9(1). https://pubmed.ncbi.nlm.nih.gov/30627804/ [DOI] [PMC free article] [PubMed]
  • 14.Porter TR, Shillcutt SK, Adams MS, Desjardins G, Glas KE, Olson JJ et al. Guidelines for the use of echocardiography as a monitor for therapeutic intervention in adults: a report from the American Society of Echocardiography. J Am Soc Echocardiogr [Internet]. 2015 [cited 2024 Jan 30];28(1):40–56. https://pubmed.ncbi.nlm.nih.gov/25559474/ [DOI] [PubMed]
  • 15.Via G, Hussain A, Wells M, Reardon R, Elbarbary M, Noble VE et al. International evidence-based recommendations for focused cardiac ultrasound. J Am Soc Echocardiogr [Internet]. 2014 [cited 2024 Jan 30];27(7):683https://pubmed.ncbi.nlm.nih.gov/24951446/. [DOI] [PubMed]
  • 16.Levitov A, Frankel HL, Blaivas M, Kirkpatrick AW, Su E, Evans D et al. Guidelines for the Appropriate Use of Bedside General and Cardiac Ultrasonography in the Evaluation of Critically Ill Patients-Part II: Cardiac Ultrasonography. Crit Care Med [Internet]. 2016 Jun 1 [cited 2024 Jan 30];44(6):1206–27. https://pubmed.ncbi.nlm.nih.gov/27182849/ [DOI] [PubMed]
  • 17.Gerbaud E, Erickson M, Grenouillet-Delacre M, Beauvieux MC, Coste P, Durrieu-Jaïs C et al. Echocardiographic evaluation and N-terminal pro-brain natriuretic peptide measurement of patients hospitalized for heart failure during weaning from mechanical ventilation - PubMed. Minerva Anestesiol [Internet]. 2012 Apr [cited 2024 Jan 30];78(4):415–25. https://pubmed.ncbi.nlm.nih.gov/22310189/ [PubMed]
  • 18.Bedet A, Tomberli F, Prat G, Bailly P, Kouatchet A, Mortaza S et al. Myocardial ischemia during ventilator weaning: a prospective multicenter cohort study. Crit Care [Internet]. 2019 Sep 18 [cited 2024 Jan 27];23(1). https://pubmed.ncbi.nlm.nih.gov/31533788/ [DOI] [PMC free article] [PubMed]
  • 19.Roche-Campo F, Bedet A, Vivier E, Brochard L, Mekontso Dessap A. Cardiac function during weaning failure: the role of diastolic dysfunction. Ann Intensive Care [Internet]. 2018 Dec 1 [cited 2024 Jan 27];8(1). [DOI] [PMC free article] [PubMed]
  • 20.Daif MS, Khalil MM, Salem HM, Abd AM, Moteleb E, Din HE et al. Using echocardiography and chest ultrasound for guidance of management of difficult-to-wean COPD patients. J Cardiol Curr Res [Internet]. 2018 Aug 7 [cited 2024 Jan 27];Volume 11(Issue 4). https://medcraveonline.com/JCCR/JCCR-11-00394.php
  • 21.Sanders M, Servaas S, Slagt C. Accuracy and precision of non-invasive cardiac output monitoring by electrical cardiometry: a systematic review and meta-analysis. J Clin Monit Comput [Internet]. 2020 Jun 1 [cited 2024 Jan 27];34(3):433–60. https://pubmed.ncbi.nlm.nih.gov/31175501/ [DOI] [PMC free article] [PubMed]
  • 22.Boet A, Jourdain G, Demontoux S, De Luca D. Stroke volume and cardiac output evaluation by electrical cardiometry: accuracy and reference nomograms in hemodynamically stable preterm neonates. J Perinatol [Internet]. 2016 Sep 1 [cited 2024 Jan 27];36(9):748–52. https://pubmed.ncbi.nlm.nih.gov/27101386/ [DOI] [PubMed]
  • 23.Xu SH, Zhang J, Zhang Y, Zhang P, Cheng GQ. Non-invasive cardiac output measurement by electrical cardiometry and M-mode echocardiography in the neonate: a prospective observational study of 136 neonatal infants. Transl Pediatr [Internet]. 2021 Jul 1 [cited 2024 Jan 27];10(7):1757–64. https://pubmed.ncbi.nlm.nih.gov/34430424/ [DOI] [PMC free article] [PubMed]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Data Availability Statement

The data presented in this study are available upon request from the corresponding author. The data are not publicly available due to the use of other unpublished articles based on this database.