Abstract
Purpose:
This goal of this study was to optimize spectrally selective 1H MRS methods for large volume acquisition of low concentration metabolites with downfield resonances at 7T and 3T, with particular attention paid to detection of nicotinamide adenine dinucleotide (NAD+) and tryptophan.
Methods:
Spectrally selective excitation was used to avoid magnetization transfer effects with water, and various sinc pulses were compared to a pure-phase E-BURP pulse. Localization using a single slice selective pulse was compared to voxel-based localization that used three orthogonal refocusing pulses, and low bandwidth refocusing pulses were used to take advantage of the chemical shift displacement of water. A technique for water sideband removal was added, and a method of coil channel combination for large volumes was introduced.
Results:
Proposed methods were compared qualitatively to previously-reported techniques at 7T. Sinc pulses resulted in reduced water signal excitation and improved spectral quality, with a symmetric, low bandwidth-time product pulse performing best. Single slice localization allowed shorter TEs with large volumes, enhancing signal, while low bandwidth slice selective localization greatly reduced the observed water signal. Gradient cycling helped remove water sidebands, and frequency aligning and pruning individual channels narrowed spectral linewidths. High quality brain spectra of NAD+ and tryptophan are shown in four subjects at 3T.
Conclusion:
Improved spectral quality with higher downfield signal, shorter TE, lower nuisance signal, reduced artifacts, and narrower peaks was realized at 7T. These methodological improvements allowed for previously unachievable detection of NAD+ and tryptophan in human brain at 3T in under five minutes.
Keywords: downfield proton spectroscopy, large volume, brain, skeletal muscle, NAD, tryptophan
INTRODUCTION
Metabolites detectable by their downfield protons (>4.7 ppm) such as nicotinamide adenine dinucleotide (NAD+), tryptophan (Trp), and nicotinamide riboside (NR) hold significant potential for clinical applications. NAD+, a crucial coenzyme in redox reactions, plays a key role in energy metabolism, cell survival, and pathways that modify aging (1,2). Tryptophan, an essential amino acid, is a precursor to serotonin and kynurenine pathway metabolites (3–5). Nicotinamide riboside, a NAD+ precursor, has potential in treating age-related degenerative diseases due to its role in boosting NAD+ levels, thus improving mitochondrial function and cellular resilience (6).
There are three main reasons why detection of these metabolites with downfield resonances is challenging. First, their concentrations are lower than the common upfield metabolites that have concentrations in the millimolar range. NAD+ has been estimated to be between 200–400 μM in brain with 31P (7,8) and downfield 1H MRS (9,10) and 290–410 μM/kg wet weight from biopsy of skeletal muscle (11,12). Tryptophan concentrations in postmortem human brain at autopsy have been reported as 20–60 μmol/kg wet weight (13). Second, downfield resonances have often been shown to have shorter T2 relaxation time constants than those found in metabolites with upfield resonances (14–16). Third, a number of downfield peaks experience magnetization transfer effects due to cross relaxation or chemical exchange with water, which reduce their NMR visibility in the spectrum when acquired with conventional spectroscopy sequences utilizing water presaturation (17–19).
Specialized sequences have been proposed to combat these issues. Spectrally selectivity excitation (9,19) and metabolite cycling (20,21) have been used to avoid magnetization transfer-related signal loss. Localization has been done with 2D/3D ISIS (15,18,22) and a 1D LASER-analogue (9) to minimize T2 signal decay. Most of these studies have been at ultra-high fields, with only a single study showing detectability of NAD+ at 3T (21), and each of these studies have used large volume acquisitions (whether voxel or slice localized) to efficiently enhance SNR. When large volumes are used, regional dependence is lost, and signal can only be described as a global measurement (23). In that case, there is no a priori requirement to localize a voxel as opposed to localizing a slice or even performing unlocalized acquisition in order to minimize TE and increase the excited volume.
Our group has previously used a modified PRESS, triple spin echo sequence with spectrally selective excitation to detect NAD+ and Trp in brain as well as NAD+, nicotinamide mononucleotide NMN, and NR in skeletal muscle at 7T (10,24,25). This paper focuses on advancing the optimization of this spin echo sequence and its slice-selective counterpart, alongside refining processing methods for the effective application in downfield spectroscopy of metabolites with low signal resonances. By combining these optimized methodologies, we demonstrate significant enhancements in spectral quality in both the brain and calf at 7T compared to our earlier research. Notably, we achieve the unprecedented detection of NAD+ and Trp in the brain at 3T within a scan duration of under five minutes.
METHODS
This study was acquired under an approved institutional review board protocol at the University of Pennsylvania. All participants gave written, informed consent. Seven healthy volunteers were scanned at 7T and 3T (age range 23 to 39).
Acquisition
Pulse sequence
A comparison of the proposed slice-selective sequence with our previous voxel-localized sequence is shown is Figure 1. As described in more detail below, we compared several spectrally-selective excitation pulses, different localization schemes, outer volume suppression (OVS), and a technique for water sideband removal.
Figure 1:

Pulse sequence diagrams of the previous sequence (A) and the proposed sequence (B). The proposed sequence uses a single slice selective spin echo sequence with a sinc pulse for excitation, an SLR pulse for localization, and incorporates gradient cycling of the crushers surrounding the localization pulse. The previous sequence uses a voxel localization formed by the intersection of three orthogonal slices localized by the same low bandwidth SLR refocusing pulse, while excitation used a uniform response, pure phase E-BURP pulse.
Localization
Localization was performed in a single shot using spatially-selective spin echoes following excitation. In order to localize a single voxel, three orthogonal slices need to be refocused. Crusher gradient pairs were applied surrounding each refocusing pulse. These crushers had reduced duration and amplitude compared to the vendor-provided sequences in order to reduce the minimum TE without generating eddy current effect. Because of the large voxel sizes and an optimized 16-step phase cycle, no additional, unwanted coherences were generated in the time domain signal (26). This sequence is illustrated in Figure 1A.
Alternatively, as shown in Figure 1B, a single slice can be localized by one spatially selective refocusing pulse, preserving more of the signal when TE is minimized. In this case, a four-step EXORCYCLE (27) was applied over the refocusing pulse, and the crusher gradient pairs are played on all three axes simultaneously. Refocusing pulses in both sequences were 180° symmetric, SLR-optimized pulses (28). At 7T, spectra acquired with relatively low bandwidth (3.1 ppm) refocusing pulses were compared with ones acquired with higher bandwidth (13.3 ppm) which is comparable to that reported with semi-LASER sequences (29).
The relatively narrow bandwidth of the refocusing localization pulses contrasts with conventional spectroscopy sequences where higher bandwidth pulses are typically preferred in order to minimize chemical shift displacement artifact (CSDA) (30). Here, though, we utilized the large chemical shift displacement to shift the localized volume of water away from the tissue of interest, further reducing the residual water signal. Figure 2 shows the effective locations of the resulting volumes of interest (VOIs) centered at 9.1 ppm and that of water at 4.7 ppm for a voxel-based localization as well as the difference in volumes over an excitation range of 2 ppm. In the slice-based localization, the water volume is shifted in the slice direction only, while in the voxel-based localization, the water volume is shifted in three directions.
Figure 2:

Prescribed VOI of the 9.1 ppm resonance (white box) along with the displaced VOI of water at 4.7 ppm for a typical bandwidth refocusing pulse (yellow box) and for the low bandwidth localization pulses proposed (red box) in the (A) axial, (B) coronal, and (C) sagittal views. (D), (E), and (F) are the same except that the displaced VOI is of the resonance at 8.1 ppm, representing the edge of the desired excitation profile.
Excitation
To avoid magnetization transfer effects with water, spectrally selective excitation was employed to leave water signal intact. Here, we compared simple, analytically-derived sinc pulses to a precalculated E-BURP uniform phase excitation pulse (31) that we used in our previous implementations (10,24,25,32). E-BURP does not require any additional refocusing and therefore, does not prolong TE. However, it is defined offline with a single bandwidth-time product and can only be scaled and stretched during runtime. Sinc pulses, on the other hand, can be defined analytically using the following formula for the normalized amplitude
| (1) |
where is the half width of the central lobe, is the filter parameter, and and are the number of zero crossings before and after the central maximum, respectively (33). The bandwidth-time product of the pulse is approximated as the total number of zero crossings (). In this study, all excitation pulses had a 2 ppm bandwidth. This resulted in a duration for E-BURP of 7 ms at 7T and 16.3 ms at 3T. Sinc pulses were denoted as and ranged from having a single main lobe only (Sinc1×1) with duration 3.2 ms at 7T to having three zero crossings on each side symmetrically (Sinc3×3) with duration 24.2 ms at 3T and included Hamming filtering with in Eq (1). Asymmetric sinc pulses (Sinc4×1 and Sinc4×2) were also tested experimentally and via Bloch simulation.
Gradient cycling
Mechanical vibrations of gradient coils can induce time varying magnetic fields resulting in frequency modulation of the signal FID (34–36). With high intensity of the unsuppressed water signal, sidebands can be as large as the metabolite signals of interest. Spectrally selective excitation greatly reduces this artifact compared to broadband spectroscopy without water suppression, but sometimes not enough. Since the largest gradient switching in the sequence occurs with the crusher gradients, we applied the gradient cycling technique described by Clayton el. al. (37) in which crusher gradient polarities were reversed every other acquisition. Reversing the gradient polarity causes the sidebands resulting from that gradient to be 180° out of phase so that summation results in significant cancellation of the artifact.
Outer volume saturation
Outer volume suppression (OVS) based on slice-selective saturation can be applied around the voxel or slice of interest. We used a modified version of the vendor-provided package in which a 3 kHz bandwidth sinc pulse was played selecting a 2 cm slice followed by a spoiler of moment 19.2 mT/m ms (9.6 mT/m amplitude and 2 ms duration). Options were added to scale the spoiling moment up and to choose the axis or axes the spoiler was played on relative to the saturated slice.
Acquisition protocol
Scans were acquired on a 7T whole body scanner (MAGNETOM Terra, Siemens Healthcare, Erlangen, Germany) and a 3T whole body scanner (MAGNETOM Prisma, Siemens Healthcare, Erlangen, Germany). For brain scans at 7T, a 1 Tx/32 Rx phased array RF head coil (Nova Medical, Wilmington, MA, USA) was used, while at 3T, a similar 32-channel receive only phased array RF coil was used. Calf muscle scans at 7T used a 1 Tx/28 Rx phased array knee RF coil (Quality Electrodynamics, Mayfield Village, OH, USA). T1-weighted anatomical images were acquired for spectroscopic volume positioning. In brain, ROIs were positioned to cover the superior portion of the head and obliqued away from the frontal sinuses for improved B0 field homogeneity. Thickness was 4 cm at 7T and 6 cm at 3T resulting in total tissue volumes ranging from 650 to 800 cm3 at 7T and 950 to 1250 cm3 at 3T. For the calf muscle, the ROI was prescribed axially near the most muscular part of the lower leg with a tissue volume of 150 cm3 for voxel localization and 500 cm3 for slice localization. Manual shimming was performed utilizing first and second order shims. For voxel-based localization, reported water linewidths were 25–28 Hz, while slice-based localization typically resulted in linewidths 1–3 Hz larger.
Table 1 shows sequence parameters for the different protocols compared here. The minimum TE was used for each protocol unless otherwise noted, while TR was kept fixed at 1 s. Readout consisted of 2048 points acquired with a dwell time of 0.25 ms; 128 averages were acquired for 7T scans in 2 min 12 s (including four preparation scans without readout), while 256 averages were acquired for 3T scans in 4 min 20 s. In each case, a water reference scan was also acquired with excitation centered at 4.7 ppm and averages reduced to the total number of phase and gradient cycling steps with all other parameters remaining the same. Before each readout, the ADC was turned on early with the first four time points discarded to mitigate filter ringing artifacts in the initial part of the FID.
Table 1:
Description of the protocols used in each subfigure.
| Figures | Field Strength (T) | Localization | Excitation | Min TE (ms) | |||
|---|---|---|---|---|---|---|---|
| Mode | RF BWa (Hz) | Pulse Shape | RF BW (Hz) | RF duration (ms) | |||
| 3A | 7 | voxel | 4000 | E-BURP | 600 | 7.0 | 18 |
| 3B, 5A, 8, S4Bb | 7 | voxel | 930 | E-BURP | 600 | 7.0 | 18 |
| 3C | 7 | slice | 4000 | Sinc2x2 | 600 | 6.6 | 13 |
| 3D, 5D, 6A, S3, S4A | 7 | slice | 930 | Sinc2x2 | 600 | 6.6 | 13 |
| 5B | 7 | voxel | 930 | Sinc2x2 | 600 | 6.6 | 24 |
| 5C, 7 | 7 | slice | 930 | E-BURP | 600 | 7.0 | 6 |
| 6B | 7 | slice | 930 | Sinc4x2 | 600 | 10.0 | 13 |
| 6C | 7 | slice | 930 | Sinc4x1 | 600 | 8.4 | 10 |
| 9, S2 | 3 | slice | 560 | Sinc3x3 | 246 | 24.2 | 33 |
Localization pulses were all 3 ms in duration except for at 3T where they were 5 ms.
Figures labeled with an ‘S’ appear in the supporting documentation.
Processing
Coil combination
The water signal reference scan was used for scaling and coil combination (38). In addition to deriving zero-order phase correction and channel amplitude weighting (39), we used the water reference data to obtain channel-dependent frequency shifts resulting from B0 inhomogeneity across large volumes and then applied these shifts to the metabolite spectrum. This was done following a procedure similar to that outlined in Mikkelsen et. al. for robust spectral registration (40,41) in which each channel signal was aligned to a reference by minimizing the expression
| (2) |
where and are the fitted amplitude, zero-order phase, and frequency shift for the channel relative to the reference. The zero-order phase and frequency shift correction terms were then applied to each water and metabolite FID. An additional pruning step was applied to remove individual channels with low residual similarity scores after alignment. This process is illustrated for a representative data set in Supporting Information Figure S1.
RESULTS
Localization
Slice-based localization was shown to have higher SNR than voxel-based localization because of both the larger signal volume and the reduced echo time. However, this comes at the expense of greater amounts of residual water excitation. Figure 3 shows how using narrow bandwidth localization pulses and sinc excitation helped mitigate the issue. For the wider bandwidth localization pulses, the prescribed voxel at 9.1 ppm and the water voxel had 31% overlap, while the overlap between them was 68% when using slice-based localization. For the narrow bandwidth localization pulses, the overlap is 0% for both voxel-based and slice-based localization. The three visible NAD+ peaks cover a spectral range of 0.4 ppm which has an overlap of 67% with voxel localization and 88% with slice localization.
Figure 3:

Representative brain spectra from a single session comparing a circumscribed single voxel localization with high bandwidth (A) and low bandwidth (B) localization pulses and single slice localization with the same high (C) and low (D) bandwidth localization pulses. Note that each spectrum is scaled based on its range and that the spectra with high bandwidth localization show significantly higher water signal.
Excitation
Figure 4 shows the waveform and simulated excitation profiles of the E-BURP pulse as well as various sinc pulses. For the sinc pulses, an isodelay equal to the time from waveform peak to its end was included before readout to remove the linear phase profile across the spectrum. With the sinc pulses, the width of the transition bands is notably dependent on the number of zero crossings, while the E-BURP pulse shows significantly greater rippling in both the stopband and the passband. This results in reduced water excitation when sinc pulses are used, and this effect is heightened when doing a full slice localization compared to a voxel localization as shown in calf muscle in Figure 5, where Figures 5A and 5B compare voxel-localized spectra acquired with E-BURP and Sinc2×2 excitation pulses, respectively. Figures 5C and 5D compare single slice localized spectra with the same pulses.
Figure 4:

Pulse waveform (top) and simulated spectral excitation profiles (bottom) for E-BURP (A,F), Sinc3×3 (B,G), Sinc2×2 (C,H), Sinc1×1 (D,I), and Sinc4×1 (E,J). Sinc pulses are named based on the number of zero crossing on each side of the central maximum (e.g. Sinc4×2 has four zero crossing before and two zero crossings after the central maximum). Insets show the spectral range around water at 7T for each profile and are scaled individually.
Figure 5:

Representative calf spectra with different excitation pulses and localization schemes. An inscribed voxel-based localization with (A) E-BURP and (B) Sinc2×2 excitation. Single slice localization with (C) E-BURP and (D) Sinc2×2 excitation. Spectra were acquired in a single session. Note that the spectra are separately scaled to their respective ranges.
Figure 6 shows the results of testing asymmetric sinc pulses compared to a symmetric one. The pulses in Figures 6A and 6B have the same isodelay and therefore minimum TE. However, the symmetric pulse used in Figure 6A shows a bit better quality spectrum with more resolvable NAD+ and NAA peaks. The highly asymmetric pulse used in Figure 6C shows higher signal but less resolvable peaks and more pronounced phase artifacts at the transition band.
Figure 6:

Comparison of symmetric and asymmetric sinc excitation pulse waveforms. (A) Symmetric Sinc2×2, (B) asymmetric Sinc4×2, and (C) strongly asymmetric Sinc4×1. TE was held constant at 13 ms between scans, though the Sinc4×1 pulse has a shorter minimum TE of 10 ms. Arrows highlight the improved peak resolution in (A) and the phase artifact in (C).
Gradient cycling
As sideband artifacts are proportional to the excited water signal, their appearance depends on several factors including shimming and volume location/orientation, and they can be reduced by focusing on reducing water signal excitation. However, in cases where the excited water signal was sufficiently large, appreciable sideband artifacts are observed, and the use of gradient cycling reduced artifacts in the range where NAD+ and Trp resonate as shown in Figure 7. Empirically, we have noted that while sideband artifacts are not always prevalent, gradient cycling has no adverse effect on the spectral quality, and therefore, the use of gradient cycling is recommended.
Figure 7:

Brain spectra acquired in the same session (A) without and (B) with gradient cycling that large residual water contamination. Other acquisition parameters were identical. Arrows highlight regions where water sidebands obscure the metabolite signal of NAD+.
Outer volume saturation
A consequence of the low bandwidth-time product of the localization RF pulses is that the selected volume does not have sharp edges. In principle, this could be improved using OVS, as can signal reduction from undesired, out-of-volume tissues (e.g. lipids from the skull or subcutaneous fat in skeletal muscle). However, in practice, with the current protocols, the outer volume artifacts caused by water and lipid are benign in the downfield region of interest, and the addition of OVS actually increased the amount of nuisance signal excitation despite some effort to optimize the module (results not shown). This is most likely due to the unintentional refocusing of unwanted coherences, but further optimization of the OVS module is beyond the scope of this work.
Coil combination
Individual channel spectra from a representative water reference scan are shown without correction (Figure 8A), after frequency alignment (Figure 8B), and after channel pruning (Figure 8C). Before frequency alignment, zero-order phase was corrected using the phase of the first time point of the fid. However, there were still remaining lineshape differences and center frequency shifts between the individual channels. Most individual channel spectra could be sufficiently corrected using Equation 2, and those that could not were pruned. Applying the frequency and phase corrections from the fit of the water signal resulted in higher signal and narrower peaks in both the coil-combined water and metabolite spectra, shown in Figs 8D and 8E, respectively.
Figure 8:

The effect of optimal coil combination. Individual channel water signals (A) without correction, (B) after frequency alignment, and (C) after frequency alignment and low-similarity channel pruning. Coil combined signal of (D) water and (E) metabolites with optimal combination (red) and with standard coil combination (blue).
3T spectra
Using an optimized, slice selective version of the sequence shown in Fig 1B with a Sinc3×3 excitation pulse and an extra narrow bandwidth localizing refocusing pulse, we were able to acquire spectra at 3T that show clear and consistent NAD+ and Trp peaks from human brain in four different subjects as shown in Figure 9 and Supporting Figure S2.
Figure 9:

Downfield spectra acquired at 3T showing NAD+ and Trp in four subjects using the proposed, slice selective sequence and processing modifications. A Sinc3×3 pulse with 246 Hz bandwidth was used for excitation and centered on 9.7 ppm. Each spectrum was acquired in 4 min 20 s.
DISCUSSION
In this study, we conducted a detailed examination of spectrally-selective spin-echo-based pulse sequences and their subsequent data processing for large volume spectroscopy of downfield metabolites in vivo. By analyzing these sequences in a modular fashion, we gained a clearer understanding of how to construct optimal protocols for specific applications. Focusing on the detection of NAD+, we developed optimized protocols for both 7T and 3T for the human brain, and at 7T for human calf muscle. This was achieved by adjusting the waveforms and bandwidths of the excitation and localization pulses, as well as the center frequency of excitation.
In upfield 1H spectroscopy of human brain, 3D-localized voxel scans are often preferred in order to avoid significant skull lipid artifacts that make quantification of certain metabolites (e.g. lactate) difficult. The same is true in calf muscle where overabundance of subcutaneous fat can obscure intra-myocellular lipid measurement (42). However, in downfield spectroscopy with spectrally selective excitation, minimal lipid signal from the methyl or methylene groups is excited, and signal from the downfield methyne group is relatively small and isolated spectrally. Hence, localization that includes large lipid-containing regions is possible. We have also experimented with unlocalized acquisitions (results not shown) but found that spectral quality is significantly degraded due to difficulty shimming over such a large region. Therefore, we opted to use a single-slice localization where shimming results in only mildly broader linewidths compared to a similarly sized 3D-localized voxel.
In moving from a pure phase selective E-BURP pulse to a traditional sinc pulse for spectrally selective excitation, the greatest drawback is the prolongation of TE to accommodate both the isodelay of the sinc pulse and an additional delay to refocus the linear phase across the spectrum. With spatially selective pulses, the time required to refocus this phase accumulation depends on the available gradient strengths and can be much shorter than the isodelay of the pulse. However, for a spectrally selective pulse, refocusing can only be achieved by waiting the full isodelay time of the pulse, which means that the minimum TE increases by twice the isodelay when switching from E-BURP to a sinc pulse. For narrow band excitation, this can be quite large depending on the bandwidth-time product of the pulse. Since the spectral region we are interested in is well separated from water, we have sacrificed a sharp transition band of the sinc pulse in order to keep the isodelay shorter. However, we did find that a sharper transition band was required for 3T compared to 7T due to the reduced spectral dispersion. For a protocol targeting metabolites closer to water, this sacrifice may not be possible, and the increase of the minimum TE might be more severe. In that case, strongly asymmetric pulses should be investigated more thoroughly.
As an alternative to sinc waveforms, spectrally selective SLR pulses could be used and might be superior for certain applications. Here, we chose instead to use sinc pulses for two main reasons. The first is the ease of implementation; sinc pulses can be defined analytically with a simple formula (Eq 1) that readily accommodates changes in bandwidth-time product, asymmetry, and filtering. This can be especially useful when designing new protocols on the scanner. The second reason is the that the excitation profile flatness of the sinc pulse is more robust to B1+ variation than those of E-BURP or SLR pulses which experience larger passband and stopband ripples. B1+ inhomogeneity is particularly problematic at ultrahigh field, especially with large volumes, but it may not be as significant an issue at lower field strengths.
One consequence of the increased CSDA caused by low bandwidth localization pulses is that the direction of the displacement becomes significant. In the head, gradient polarity is chosen such that the water and lipid volumes are shifted superiorly away from the head in the slice direction. In the calf muscle scans, the slice selective gradient polarity was reversed to shift the water and lipid volumes inferiorly towards the foot where leg volume is reduced. As voxel-based localization uses three orthogonal refocusing pulses, the CSDA is experienced in all three directions simultaneously, creating a larger overall separation and lessening the dependence on direction. Supporting Information Figure S3 shows the differences in the water signals in a slice-based acquisition in human brain for alternate polarities of the slice selection gradient.
Our gradient cycling technique is only applied to the crusher gradients surrounding the slice-selective refocusing pulse(s). While it is true that this technique only removes water sidebands originating from vibrations caused by the crusher gradients and that more robust sideband removal techniques exist (36), water sidebands originating from slice selective gradients are expected to be negligible in the present context for two reasons. First, spectrally selective excitation greatly decreases the residual water signal compared to broad band excitation techniques without water suppression. Second, the low bandwidth-time product RF pulses and thick slices mean that the slice selective gradients are lower in amplitude than those typically used in imaging or even conventional spectroscopy. As shown in Supporting Information Figure S3, for low bandwidth refocusing, reversing the polarity of the slice selective gradient in order to cancel water sidebands is not feasible as entirely different volumes with different water content are localized.
In addition to causing spectral sidebands, pulsed gradients can also generate eddy currents in which a time dependent magnetic field causes spurious peaks and lineshape distortions. Again, crusher and spoiler gradients contribute most to this phenomenon. The sequences presented here have reduced crusher amplitudes surrounding refocusing pulses compared to standard, vendor-provided single voxel sequences. In addition, because of the spectrally selective excitation, there are no spoiler gradients as part of a water presaturation module, which is a significant source of eddy currents in conventional 1H spectroscopy. Therefore, we do not expect eddy currents to be a significant source of artifacts in these sequences, and this is supported by each peak of interest being well fit to a single Lorentzian in the HSVD algorithm (43).
The most common method of eddy current correction (ECC) involves collecting a pure, on resonance water signal and subtracting off its time domain phase from the metabolite signal (44). A consequence is that both the zero order phase and frequency shifts of water are removed, similar to our proposed method of coil channel combination based on spectral registration. However, there are a couple of important differences between the two approaches. First, in ECC, each time point in the FID is corrected individually with the corresponding point in the water reference FID. In contrast, spectral registration-based correction uses hundreds of points to solve for only three parameters making it more robust and accurate when shifts are not caused by time varying eddy currents. Second, the proposed method includes a procedure for pruning channels that cannot be well corrected and do not contribute coherently to the signal, while ECC does not. Pruning has been shown to improve spectral quality in large volume spectroscopy (23). Nevertheless, ECC is a viable method to perform channel-wise spectral alignment for large volume spectroscopy, and a comparison with the proposed method is shown in Supporting Information Figure S4.
Because of the reduced spectral dispersion at 3T, water signal is a more significant nuisance than at 7T. Therefore, methods specifically focused on reducing water signal were most beneficial. This included extra narrow bandwidth localization pulses set to mimic the CSDA present at 7T as well as sinc spectrally selective excitation. The bandwidth of the excitation pulse was reduced, its center frequency was shifted further downfield, and the number of zero crossings was increased to sharpen the transition band. Together, these changes reduced residual water signal sufficiently, where the downfield metabolites were readily identifiable. However, a consequence of these changes is that the minimum TE increased from 13 ms at 7T to 33 ms at 3T with a single slice localization. Voxel-based localization with the same pulses has a minimum TE of 48 ms which is prohibitively long for short T2 downfield metabolites. Despite the reduced signal, NAD+ and Trp peaks were still observable in a scan under 5 minutes, marking the first observation of Trp in the 1H spectrum at 3T and a significant acceleration compared to the 32 minute protocol in Dziadosz and Hoefemann et. al. for detection of NAD+ (21).
A major limitation of this study is that the signal from such large volumes can only be considered a global representation and is not sensitive to regional differences. Recent work using MRSI at 3T has shown promise in obtaining downfield spectra from smaller volumes, though it has currently been restricted to larger signal resonances than those discussed here (45,46). Also, in studies across subjects or tracking a single subject over time, differing volume compositions could obfuscate any metabolic changes. In that case, careful positioning of the volume to match compositions should be done using a tool such as ImScribe (47) that coregisters a high resolution anatomic image to a previously-acquired template and returns the location and orientation of the matching spectroscopy volume.
In addition, while the low bandwidth refocusing pulses deliberately exploit the CSDA of the water and lipid signals, undesired CSDA is also elicited over the excited spectral range of interest. For NAD+, the CSDA between the undetected H5 proton at 8.2 ppm and its coupled partners, the H4 and H6 protons, leads to anomalous J-modulation of those peaks which could affect quantitation. For the protocols described here, the minimum TE is much less than 1/J (19); therefore, the effect on quantitation should be mild and could be taken into account using prior-knowledge based fitting. Nevertheless, this technique of low bandwidth localization is best suited when the spectral range of interest is small compared to its spectral separation from nuisance signals.
CONCLUSION
This study describes 1H spectroscopic techniques specifically targeting low concentration metabolites with downfield resonances using large volume acquisitions. Utilizing novel acquisition and processing methods, we have achieved improved spectral quality at 7T in brain and calf muscle as well as unmatched spectral quality in brain at 3T, detecting whole brain NAD+ and tryptophan in under 5 minutes.
Supplementary Material
Supporting Information Figure S1: Between-channel correlation matrices (top row) and ranked channel-wise similarity score plots (bottom row), computed before (left) and after (right) channel-wise frequency alignment of a water reference scan. The lowest ranked channels (red x), whose similarity scores fall below the dashed red line, are pruned from the coil combination step.
Supporting Information Figure S2: Same data as Figure 9 in manuscript showing full bandwidth spectra at 3T in human brain from four healthy subjects. Note that the spectra are separately scaled to their respective ranges.
Supporting Information Figure S3: Comparison of spectra acquired in brain from the same slice in the same session with only the direction of the CSDA reversed. The water volume is shifted (A) inferiorly or (B) superiorly. Because of the low B1+ of the 1 Tx/64 Rx head coil at the base of the brain, low water signal is observed even with an inferior CSDA, even though the differences in water signal at each location are vast.
Supporting Information Figure S4: Channel-wise corrections of a water reference scan. Uncorrected (black), eddy current corrected (blue), and frequency aligned and pruned (red) for a dataset showing the relative improvement of the proposed method compared to eddy current correction (A) as well as a dataset showing little difference between the two (B). Both corrections show improved peak shape and linewidth compared to the uncorrected signal. All peaks in each subfigure are normalized to have the same area.
FUNDING INFORMATION
This work was supported by the National Institute of Biomedical Imaging and Bioengineering of the National Institutes of Health under award number P41EB029460, by the National Institute of Aging of the National Institutes of Health under award numbers, R01AG071725 and R01AG063869, and by the National Heart, Lung, and Blood Institute of the National Institute of Health under award numbers R01HL137984, R01HL169378 and F31HL158217.
REFERENCES
- 1.ichiro Imai S. The NAD World: A New Systemic Regulatory Network for Metabolism and Aging—Sirt1, Systemic NAD Biosynthesis, and Their Importance. Cell Biochem Biophys. 2009;53(2):65–74. doi: 10.1007/s12013-008-9041-4 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2.Houtkooper RH, Cantó C, Wanders RJ, Auwerx J. The Secret Life of NAD+: An Old Metabolite Controlling New Metabolic Signaling Pathways. Endocr Rev. 2010;31(2):194–223. doi: 10.1210/er.2009-0026 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.Bender DA. Biochemistry of tryptophan in health and disease. Mol Aspects Med. 1983;6(2):101–197. doi: 10.1016/0098-2997(83)90005-5 [DOI] [PubMed] [Google Scholar]
- 4.Badawy AAB. Kynurenine Pathway of Tryptophan Metabolism: Regulatory and Functional Aspects. Int J Tryptophan Res. 2017;10:117864691769193. doi: 10.1177/1178646917691938 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5.Katsyuba E, Mottis A, Zietak M, et al. De novo NAD+ synthesis enhances mitochondrial function and improves health. Nature. 2018;563(7731):354–359. doi: 10.1038/s41586-018-0645-6 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6.Bieganowski P, Brenner C. Discoveries of Nicotinamide Riboside as a Nutrient and Conserved NRK Genes Establish a Preiss-Handler Independent Route to NAD+ in Fungi and Humans. Cell. 2004;117(4):495–502. doi: 10.1016/S0092-8674(04)00416-7 [DOI] [PubMed] [Google Scholar]
- 7.Zhu XH, Lu M, Lee BY, Ugurbil K, Chen W. In vivo NAD assay reveals the intracellular NAD contents and redox state in healthy human brain and their age dependences. Proc Natl Acad Sci. 2015;112(9):2876–2881. doi: 10.1073/pnas.1417921112 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8.Lu M, Zhu XH, Chen W. In vivo 31 P MRS assessment of intracellular NAD metabolites and NAD + /NADH redox state in human brain at 4 T: In vivo 31 P MRS assessment of brain NAD metabolites and redox state at 4T. NMR Biomed. 2016;29(7):1010–1017. doi: 10.1002/nbm.3559 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.De Graaf RA, De Feyter HM, Brown PB, Nixon TW, Rothman DL, Behar KL. Detection of cerebral NAD + in humans at 7T. Magn Reson Med. 2017;78(3):828–835. doi: 10.1002/mrm.26465 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Bagga P, Hariharan H, Wilson NE, et al. Single-Voxel 1 H MR spectroscopy of cerebral nicotinamide adenine dinucleotide (NAD + ) in humans at 7T using a 32-channel volume coil. Magn Reson Med. 2020;83(3):806–814. doi: 10.1002/mrm.27971 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11.Graham T, Sjøgaard G, Löllgen H, Saltin B. NAD in muscle of man at rest and during exercise. Pflüg Arch Eur J Physiol. 1978;376(1):35–39. doi: 10.1007/BF00585245 [DOI] [PubMed] [Google Scholar]
- 12.White AT, Schenk S. NAD + /NADH and skeletal muscle mitochondrial adaptations to exercise. Am J Physiol-Endocrinol Metab. 2012;303(3):E308–E321. doi: 10.1152/ajpendo.00054.2012 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13.Perry TL, Berry K, Hansen S, Diamond S, Mok C. REGIONAL DISTRIBUTION OF AMINO ACIDS IN HUMAN BRAIN OBTAINED AT AUTOPSY. J Neurochem. 1971;18(3):513–519. doi: 10.1111/j.1471-4159.1971.tb11979.x [DOI] [PubMed] [Google Scholar]
- 14.Fichtner ND, Henning A, Zoelch N, Boesch C, Kreis R. Elucidation of the downfield spectrum of human brain at 7 T using multiple inversion recovery delays and echo times. Magn Reson Med. 2017;78(1):11–19. doi: 10.1002/mrm.26343 [DOI] [PubMed] [Google Scholar]
- 15.Gonçalves SI, Ligneul C, Shemesh N. Short echo time relaxation-enhanced MR spectroscopy reveals broad downfield resonances. Magn Reson Med. 2019;82(4):1266–1277. doi: 10.1002/mrm.27806 [DOI] [PubMed] [Google Scholar]
- 16.Borbath T, Murali-Manohar S, Wright AM, Henning A. In vivo characterization of downfield peaks at 9.4 T: T 2 relaxation times, quantification, pH estimation, and assignments. Magn Reson Med. 2021;85(2):587–600. doi: 10.1002/mrm.28442 [DOI] [PubMed] [Google Scholar]
- 17.MacMillan EL, Chong DGQ, Dreher W, Henning A, Boesch C, Kreis R. Magnetization exchange with water and T 1 relaxation of the downfield resonances in human brain spectra at 3.0 T: Exchange and T 1 Relaxation of Downfield Resonances. Magn Reson Med. 2011;65(5):1239–1246. doi: 10.1002/mrm.22813 [DOI] [PubMed] [Google Scholar]
- 18.Shemesh N, Rosenberg JT, Dumez JN, Muniz JA, Grant SC, Frydman L. Metabolic properties in stroked rats revealed by relaxation-enhanced magnetic resonance spectroscopy at ultrahigh fields. Nat Commun. 2014;5(1):4958. doi: 10.1038/ncomms5958 [DOI] [PubMed] [Google Scholar]
- 19.De Graaf RA, Behar KL. Detection of cerebral NAD + by in vivo 1 H NMR spectroscopy. NMR Biomed. 2014;27(7):802–809. doi: 10.1002/nbm.3121 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20.Fichtner ND, Giapitzakis I, Avdievich N, et al. In vivo characterization of the downfield part of 1 H MR spectra of human brain at 9.4 T : Magnetization exchange with water and relation to conventionally determined metabolite content. Magn Reson Med. 2018;79(6):2863–2873. doi: 10.1002/mrm.26968 [DOI] [PubMed] [Google Scholar]
- 21.Dziadosz M, Hoefemann M, Döring A, Marjańska M, Auerbach EJ, Kreis R. Quantification of NAD + in human brain with 1 H MR spectroscopy at 3 T: Comparison of three localization techniques with different handling of water magnetization. Magn Reson Med. 2022;88(3):1027–1038. doi: 10.1002/mrm.29267 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 22.Dziadosz M, Bogner W, Kreis R. Non-water-excitation MR spectroscopy techniques to explore exchanging protons in human brain at 3 T. Magn Reson Med. 2020;84(5):2352–2363. doi: 10.1002/mrm.28322 [DOI] [PubMed] [Google Scholar]
- 23.Hoefemann M, Adalid V, Kreis R. Optimizing acquisition and fitting conditions for 1 H MR spectroscopy investigations in global brain pathology. NMR Biomed. 2019;32(11):e4161. doi: 10.1002/nbm.4161 [DOI] [PubMed] [Google Scholar]
- 24.Nanga RPR, Elliott MA, Swain A, et al. Identification of l -Tryptophan by down-field 1 H MRS: A precursor for brain NAD + and serotonin syntheses. Magn Reson Med. 2022;88(6):2371–2377. doi: 10.1002/mrm.29414 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 25.Nanga RPR, Elliott MA, Swain A, et al. Identification of new resonances in downfield 1 H MRS of human calf muscle in vivo: Potentially metabolite precursors for skeletal muscle NAD +. Magn Reson Med. 2023;90(3):1166–1171. doi: 10.1002/mrm.29687 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 26.Landheer K, Juchem C. Simultaneous optimization of crusher and phase cycling schemes for magnetic resonance spectroscopy: an extension of dephasing optimization through coherence order pathway selection. Magn Reson Med. 2020;83(2):391–402. doi: 10.1002/mrm.27952 [DOI] [PubMed] [Google Scholar]
- 27.Bodenhausen G, Freeman R, Turner DL. Suppression of artifacts in two-dimensional J spectroscopy. J Magn Reson 1969. 1977;27(3):511–514. doi: 10.1016/0022-2364(77)90016-6 [DOI] [Google Scholar]
- 28.Pauly J, Le Roux P, Nishimura D, Macovski A. Parameter relations for the Shinnar-Le Roux selective excitation pulse design algorithm (NMR imaging). IEEE Trans Med Imaging. 1991;10(1):53–65. doi: 10.1109/42.75611 [DOI] [PubMed] [Google Scholar]
- 29.Scheenen TWJ, Heerschap A, Klomp DWJ. Towards 1H-MRSI of the human brain at 7T with slice-selective adiabatic refocusing pulses. Magma N Y N. 2008;21(1–2):95–101. doi: 10.1007/s10334-007-0094-y [DOI] [PMC free article] [PubMed] [Google Scholar]
- 30.Wilson M, Andronesi O, Barker PB, et al. Methodological consensus on clinical proton MRS of the brain: Review and recommendations. Magn Reson Med. 2019;82(2):527–550. doi: 10.1002/mrm.27742 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 31.Geen H, Freeman R. Band-selective radiofrequency pulses. J Magn Reson 1969. 1991;93(1):93–141. doi: 10.1016/0022-2364(91)90034-Q [DOI] [Google Scholar]
- 32.Swago S, Elliott MA, Nanga RPR, et al. Quantification of cross-relaxation in downfield 1 H MRS at 7 T in human calf muscle. Magn Reson Med. 2023;90(1):11–20. doi: 10.1002/mrm.29615 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 33.Bernstein MA, King KF, Zhou XJ. Handbook of MRI Pulse Sequences. Elsevier; 2004. [Google Scholar]
- 34.Clayton DB, Elliott MA, Leigh JS, Lenkinski RE. 1H spectroscopy without solvent suppression: characterization of signal modulations at short echo times. J Magn Reson San Diego Calif 1997. 2001;153(2):203–209. doi: 10.1006/jmre.2001.2442 [DOI] [PubMed] [Google Scholar]
- 35.Serrai H, Clayton DB, Senhadji L, Zuo C, Lenkinski RE. Localized proton spectroscopy without water suppression: removal of gradient induced frequency modulations by modulus signal selection. J Magn Reson San Diego Calif 1997. 2002;154(1):53–59. doi: 10.1006/jmre.2001.2462 [DOI] [PubMed] [Google Scholar]
- 36.Dong Z, Dreher W, Leibfritz D. Experimental method to eliminate frequency modulation sidebands in localized in vivo 1 H MR spectra acquired without water suppression. Magn Reson Med. 2004;51(3):602–606. doi: 10.1002/mrm.10716 [DOI] [PubMed] [Google Scholar]
- 37.Clayton DB, Elliott ME, Lenkinski RE. In vivo proton MRS of the human brain at 4T without solvent suppression. In: Proceedings of the 7th Scientific Meeting, International Society of Magnetic Resonance in Medicine, Philadelphia. Vol 1602.; 1999. [Google Scholar]
- 38.Near J, Harris AD, Juchem C, et al. Preprocessing, analysis and quantification in single-voxel magnetic resonance spectroscopy: experts’ consensus recommendations. NMR Biomed. 2021;34(5):e4257. doi: 10.1002/nbm.4257 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 39.Brown MA. Time-domain combination of MR spectroscopy data acquired using phased-array coils. Magn Reson Med. 2004;52(5):1207–1213. doi: 10.1002/mrm.20244 [DOI] [PubMed] [Google Scholar]
- 40.Mikkelsen M, Tapper S, Near J, Mostofsky SH, Puts NAJ, Edden RAE. Correcting frequency and phase offsets in MRS data using robust spectral registration. NMR Biomed. 2020;33(10):e4368. doi: 10.1002/nbm.4368 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 41.Near J, Edden R, Evans CJ, Paquin R, Harris A, Jezzard P. Frequency and phase drift correction of magnetic resonance spectroscopy data by spectral registration in the time domain. Magn Reson Med. 2015;73(1):44–50. doi: 10.1002/mrm.25094 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 42.Boesch C, Slotboom J, Hoppeler H, Kreis R. In vivo determination of intra-myocellular lipids in human muscle by means of localized 1 H-MR-spectroscopy. Magn Reson Med. 1997;37(4):484–493. doi: 10.1002/mrm.1910370403 [DOI] [PubMed] [Google Scholar]
- 43.Barkhuijsen H, De Beer R, Van Ormondt D. Improved algorithm for noniterative time-domain model fitting to exponentially damped magnetic resonance signals. J Magn Reson 1969. 1987;73(3):553–557. doi: 10.1016/0022-2364(87)90023-0 [DOI] [Google Scholar]
- 44.Klose U In vivo proton spectroscopy in presence of eddy currents. Magn Reson Med. 1990;14(1):26–30. doi: 10.1002/mrm.1910140104 [DOI] [PubMed] [Google Scholar]
- 45.Považan M, Schär M, Gillen J, Barker PB. Magnetic resonance spectroscopic imaging of downfield proton resonances in the human brain at 3 T. Magn Reson Med. 2022;87(4):1661–1672. doi: 10.1002/mrm.29142 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 46.Özdemir İ, Ganji S, Gillen J, Etyemez S, Považan M, Barker PB. Downfield proton MRSI with whole-brain coverage at 3T. Magn Reson Med. 2023;90(3):814–822. doi: 10.1002/mrm.29706 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 47.Elliott M. ImScribe – Automated brain MRI FOV determination for Siemens MRI. https://www.med.upenn.edu/CAMIPM/imscribe.html. [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
Supporting Information Figure S1: Between-channel correlation matrices (top row) and ranked channel-wise similarity score plots (bottom row), computed before (left) and after (right) channel-wise frequency alignment of a water reference scan. The lowest ranked channels (red x), whose similarity scores fall below the dashed red line, are pruned from the coil combination step.
Supporting Information Figure S2: Same data as Figure 9 in manuscript showing full bandwidth spectra at 3T in human brain from four healthy subjects. Note that the spectra are separately scaled to their respective ranges.
Supporting Information Figure S3: Comparison of spectra acquired in brain from the same slice in the same session with only the direction of the CSDA reversed. The water volume is shifted (A) inferiorly or (B) superiorly. Because of the low B1+ of the 1 Tx/64 Rx head coil at the base of the brain, low water signal is observed even with an inferior CSDA, even though the differences in water signal at each location are vast.
Supporting Information Figure S4: Channel-wise corrections of a water reference scan. Uncorrected (black), eddy current corrected (blue), and frequency aligned and pruned (red) for a dataset showing the relative improvement of the proposed method compared to eddy current correction (A) as well as a dataset showing little difference between the two (B). Both corrections show improved peak shape and linewidth compared to the uncorrected signal. All peaks in each subfigure are normalized to have the same area.