. 2024 Aug 15;14(11):1788–1804. doi: 10.1002/2211-5463.13873

An image processing pipeline for electron cryo‐tomography in RELION‐5

Alister Burt 1,2, Bogdan Toader 1, Rangana Warshamanage 3,4, Andriko von Kügelgen 1, Euan Pyle 5,6,7, Jasenko Zivanov 1, Dari Kimanius 1,8, Tanmay A M Bharat 1, Sjors H W Scheres 1,
PMCID: PMC11532982  PMID: 39147729

Abstract

Electron tomography of frozen, hydrated samples allows structure determination of macromolecular complexes that are embedded in complex environments. Provided that the target complexes may be localised in noisy, three‐dimensional tomographic reconstructions, averaging images of multiple instances of these molecules can lead to structures with sufficient resolution for de novo atomic modelling. Although many research groups have contributed image processing tools for these tasks, a lack of standardisation and interoperability represents a barrier for newcomers to the field. Here, we present an image processing pipeline for electron tomography data in RELION‐5, with functionality ranging from the import of unprocessed movies to the automated building of atomic models in the final maps. Our explicit definition of metadata items that describe the steps of our pipeline has been designed for interoperability with other software tools and provides a framework for further standardisation.

Keywords: electron microscopy, electron tomography, image processing, software


We present an image processing pipeline for electron cryo‐tomography data in RELION‐5 with functionality ranging from unprocessed movie import to the automated building of atomic models in final maps. The metadata describing the pipeline steps is aimed at standardisation and interoperability with other software, while the performance improvements and visual tools lead to short processing times and ease of use.

graphic file with name FEB4-14-1788-g004.jpg


Abbreviations

CTF

Contrast Transfer function

ET

Electron tomography

FSC

Fourier Shell Correlation

GUI

Graphical User Interface

MPI

Message Passing Interface

MRC

Medical Research Council

2D

Two‐dimensional

3D

Three‐dimensional

In the electron cryo‐tomography (cryo‐ET) approach, a three‐dimensional (3D) reconstruction called a tomogram is calculated from a series of images that are taken as a sample is rotated around a tilt axis in the electron microscope. Because radiation damage limits the dose that can be applied to the specimen, tomograms typically suffer from large amounts of noise. Moreover, samples are typically thin, slab‐like sections, which dictates that some views of the sample cannot be acquired, leading to artefacts in the reconstructed tomogram. Provided that the structures of interest, or particles, can be localised in the noisy tomograms and that they can be brought into register, many instances of a particle can be combined into a single 3D reconstruction with an increased signal‐to‐noise ratio. This approach, commonly referred to as subtomogram averaging, has been used to calculate structures of macromolecular complexes to resolutions sufficient for de novo atomic modelling for macromolecules inside cells. 1 , 2 , 3

Many tools exist for acquiring and analysing cryo‐ET data. SerialEM, 4 , 5 one of the most popular softwares for tilt series acquisition, saves acquisition‐related metadata in text files with the mdoc extension. Tomo5, a commercial software provided by Thermo Fisher, also writes metadata in the mdoc format. Accurate 3D reconstruction from a tilt series requires determining the parameters of a projection model describing the position, orientation and possibly also deformation of the sample in the electron microscope. The imod software package 6 , 7 provides some of the most popular tools for this task, which is commonly referred to as tilt series alignment. Recently, AreTomo 8 has also gained popularity by providing robust and automated tools for tilt series alignment, particularly for in situ data without fiducial markers. Once the parameters of a projection model are determined, a tomogram may be reconstructed from tilt images. Again, many approaches exist, including weighted back‐projection, 9 simultaneous iterative reconstruction technique, 10 , 11 , 12 algebraic reconstruction technique 13 , 14 , 15 and compressed sensing. 16 The reconstruction method used depends on what the tomogram, or subvolumes thereof, will be used for. For example, tomograms with high contrast at lower spatial frequencies are often used for direct visual interpretation and segmentation, while for subtomogram averaging, high‐resolution features need to be preserved in the reconstruction and lower resolution features may be less pronounced. Direct reconstruction of local regions of interest in tomograms 1 , 17 , 18 , 19 , 20 has obviated the need for calculating large, high‐resolution tomograms from which subvolumes are cropped. To assist in feature identification, deep‐learning approaches have gained traction for denoising 21 , 22 , 23 , 24 and template matching. 25 , 26

Despite the advent of near‐complete cryo‐ET image processing workflows, like those implemented in IMOD 7 , PEET, 27 EMAN2, 28 Dynamo, 29 STOPGAP, 30 TomoBEAR, 31 SCIPION 32 , 33 or NextPyp, 19 a lack of standardisation represents a hurdle to newcomers and makes it difficult to effectively use the best parts of each package.

Here, we present a cryo‐ET image processing pipeline inside the free, open‐source software RELION‐5. Closely mirroring similar procedures in single‐particle analysis, our cryo‐ET pipeline starts from unprocessed movies and acquisition metadata in mdoc files and ends with high‐resolution 3D reconstructions and the automated building of atomic models. 34 The pipeline provides wrappers to CTFFIND4 35 for estimation of contrast transfer function (CTF) parameters; to IMOD 7 and AreTomo 8 for tilt series alignment; and to cryoCARE 22 for tomogram denoising. We present graphical tools for the manual curation of images in tilt series and for the picking of particles in reconstructed tomograms. New to RELION‐5 is the option to write out CTF‐premultiplied 2D stacks of subimages that are cropped from the tilt series at the corresponding positions for each particle. 2D particle stacks can then be used in the same alignment and classification procedures that were introduced in RELION‐4. 18 Refinement of 2D particle stacks provides major computational advantages, both in disc space requirements and in processing speed, compared with methods that reconstruct 3D subvolumes, such as the pseudo‐subtomogram approach introduced in RELION‐4. 18 We also provide an explicit definition of the metadata required for and generated in our pipeline, providing a framework for further standardisation that will also be adopted in the CCP‐EM software 36 , 37 (Tom Burnley, personal communication).

Materials and methods

All new developments in the RELION‐5 tomography pipeline are accessible from the graphical user interface (GUI; Fig. 1), which can be launched from the command line using the command ‘relion ‐‐tomo’. The side bar in the top left of the GUI shows the names of all job types. Below, we describe all jobs that have new developments for cryo‐ET in RELION‐5.

Fig. 1.

Fig. 1

The RELION‐5 tomography GUI. The sidebar on the top left provides access to all job types. The top right panel provides multiple tabs with input parameters for each job type. The bottom half of the GUI provides an overview of finished, running and scheduled jobs; how different jobs relate to each other through their input and output; and the output from the currently selected job.

Importing into RELION's data model

Starting from directories with the unprocessed movie files and their corresponding mdoc files, the ‘Import’ job in the cryo‐ET pipeline of RELION‐5 writes metadata files that describe the acquired tilt series in the STAR format. 38 A primary tilt_series.star file (Fig. 2; Table 1) contains a global table with one line per tilt series. The global table provides general information, including the acceleration voltage, pixel size, amplitude contrast and spherical aberration, plus a reference to a STAR file that contains information about individual images in each tilt series (Fig. 3; Table 2). Tilt series STAR files are stored in the tilt_series/ subdirectory of the Import job directory. They contain a table with one line for each image of the tilt series that stores the name of the unprocessed movie, the nominal tilt angle of the stage, the nominal orientation of the tilt axis, the nominal defocus and the accumulated electron dose (Fig. 3; black).

Fig. 2.

Fig. 2

Tomograms metadata description. Metadata to describe a set of tomograms is stored in the STAR format. Labels and data columns shown in black are added during import; labels and columns in violet are added during tomogram reconstruction. Corresponding definitions of metadata labels are given in Table 1.

Table 1.

Metadata items for tomograms. Metadata labels, types and definitions for the description of a set of tomograms. Entries in bold are compulsory for sub‐tomogram averaging; entries in regular are not necessary for sub‐tomogram averaging.

Label Type Definition
_rlnTomoName String Name of a tomogram
_rlnVoltage Float Acceleration voltage (in kV)
_rlnSphericalAberration Float Spherical aberration (in mm)
_rlnAmplitudeContrast Float Amplitude contrast ratio
_rlnMicrographOriginalPixelSize Float Pixel size of the original movie, before binning in motion correction (in Å)
_rlnTomoHand Integer Tomogram handedness, i.e. sign of the slope of defocus over the Z‐coordinate (−1 or 1)
_rlnOpticsGroupName String Name of an optics group
_rlnTomoTiltSeriesPixelSize Float Pixel size of the tilt series, after binning in motion correction (in Å)
_rlnTomoTiltSeriesStarFile String Path of the STAR file with metadata about the individual tilt series images
_rlnEtomoDirectiveFile String Path of the ETomo directive file from tilt series alignment
_rlnTomoTomogramBinning Float Binning of the tomogram reconstruction
_rlnTomoSizeX Integer Width of the tomogram in unbinned voxels (i.e. of size _rlnTomoTiltSeriesPixelSize)
_rlnTomoSizeY Integer Length of the tomogram in unbinned voxels (i.e. of size _rlnTomoTiltSeriesPixelSize)
_rlnTomoSizeZ Integer Height of the tomogram in unbinned voxels (i.e. of size _rlnTomoTiltSeriesPixelSize)
_rlnTomoReconstructedTomogramHalf1 String Path of a tomogram reconstructed from even movie frames only
_rlnTomoReconstructedTomogramHalf2 String Path of a tomogram reconstructed from odd movie frames only

Fig. 3.

Fig. 3

Tilt series images metadata description. Metadata to describe a set of images in a tilt series is stored in the STAR format. Labels and data columns shown in black are added during import; labels and columns in red are added during motion correction; labels and columns in green are added during CTF estimation; labels and columns in blue are added during tilt series alignment; labels and columns in violet are added during tomogram reconstruction. Corresponding definitions of metadata labels are given in Table 2.

Table 2.

Metadata items for tilt series images. Metadata labels, types and definitions for the description of a set of images in a tilt series. Entries in bold are compulsory for sub‐tomogram averaging; entries in regular are not necessary for sub‐tomogram averaging.

Label Type Definition
_rlnMicrographMovieName String Path of the movie stack for an individual tilt series image
_rlnTomoTiltMovieFrameCount Integer Number of frames in the movie stack
_rlnTomoNominalStageTiltAngle Float Nominal value for the stage tilt angle (in °)
_rlnTomoNominalTiltAxisAngle Float Nominal value for the angle of the tilt axis with the Y‐axis (in °)
_rlnMicrographPreExposure Float Pre‐exposure dose (in electrons/Å 2 )
_rlnTomoNominalDefocus Float Nominal value for the defocus (in μm, negative values for underfocus)
_rlnCtfPowerSpectrum String Path of the power spectrum image for CTF estimation
_rlnMicrographNameEven String Path of the summed micrograph image from even movie frames only
_rlnMicrographNameOdd String Path of the summed micrograph image from odd movie frames only
_rlnMicrographName String Path of the summed micrograph image from all movie frames
_rlnMicrographMetadata String Path of a STAR file with metadata from motion correction
_rlnAccumMotionTotal Float Accumulated global motion during the entire movie (in Å)
_rlnAccumMotionEarly Float Accumulated global motion during early frames of the movie (in Å)
_rlnAccumMotionLate Float Accumulated global motion during late frames of the movie (in Å)
_rlnCtfImage String Path of the Thon‐ring image from CTF estimation
_rlnDefocusU Float Defocus in the U‐direction (in Å, positive values for underfocus)
_rlnDefocusV Float Defocus in the V‐direction (in Å, positive values for underfocus)
_rlnCtfAstigmatism Float Absolute value of the difference between defocus in U‐ and V‐direction (in Å)
_rlnDefocusAngle Float Angle between X and defocus U direction (in °)
_rlnCtfFigureOfMerit Float Figure of merit from CTF estimation
_rlnCtfMaxResolution Float Maximum resolution (in Å) of fitted Thon rings from CTF estimation
_rlnCtfIceRingDensity Float Accumulated power of the image in the frequency range (0.25–0.28 Å−1)
_rlnTomoXTilt Float Angle for rotation of the tomogram around the X‐axis (in °)
_rlnTomoYTilt Float Angle for rotation of the tomogram around the Y‐axis (in °)
_rlnTomoZRot Float Angle for rotation of the tomogram around the Z‐axis (in °)
_rlnTomoXShiftAngst Float Shift in X‐direction (in Å) to align the projection of a tomogram with the tilt series image
_rlnTomoYShiftAngst Float Shift in Y‐direction (in Å) to align the projection of a tomogram with the tilt series image
_rlnCtfScalefactor Float Linear scale‐factor to be applied to the CTF values

One option that the user needs to decide when importing tilt series is the defocus handedness (rlnTomoHand in Fig. 2; Table 1), which defines the direction in which defocus changes while moving across the tilt axis in tilted images. In particular, when rlnTomoHand is set to 1, a particle with positive X coordinate (in the 3D space) will move closer to focus and a particle with negative X coordinate will move away from focus as the tilt angle of the specimen is increased, while the opposite effect is achieved when rlnTomoHand is set to −1. Although the defocus handedness could be estimated automatically during CTF estimation, 17 , 39 , 40 for now, it needs to be determined by the user using trial and error. For all Thermo Fisher Krios and Glacios microscopes that we have tested so far, the rlnTomoHand was always −1, which is the default option on the GUI (Yes, to invert defocus handedness).

Motion correction

The tilt_series.star file from the ‘Import’ job can be used as input for a ‘Motion correction’ job. Internally, all images of all tilt series are combined into a single list of images that can then be processed in parallel using a combination of message passing interface (MPI) and threads. Either UCSF MotionCor2 41 or RELION's own implementation of that algorithm 42 may be used for movie frame alignment. An option exists to write out summed micrographs that are calculated only from the even or the odd movie frames, which may be used for tomogram denoising as described below. Although an option exists to bin images at this stage, this option will limit the achievable resolution of all downstream steps. Because there is no option for un‐binning (i.e. reverting to the original pixel size) at a later step, binning is only recommended if the images were recorded in super‐resolution mode and the resolution of the average is not expected to reach the physical Nyquist frequency.

The motion correction job adds several columns to the tables in the STAR files of the individual tilt series, including the name of the motion‐corrected micrograph and statistics about the estimated movements in each of the tilt series images (Fig. 3; red).

CTF estimation

The tilt_series.star file from the ‘Motion correction’ job can be used as input for a ‘CTF estimation’ job, where individual images of all tilt series are again processed in parallel using MPI. CTF estimation is performed using a wrapper to CTFFIND4, 35 with two modifications from the equivalent approach for single‐particle analysis that increase robustness for images at high tilt angles, where the signal to noise ratio is low. First, whereas the same minimum and maximum defocus values are used for all micrographs in single‐particle analysis, in the tomography pipeline the user may provide a search range that will be used around the nominal defocus value for each image. Second, the maximum resolution used for CTF estimation may be varied with the accumulated electron dose for each image in the tilt series.

This job again adds information to the STAR files of the individual tilt series, including the estimated defocus values, a figure‐of‐merit for the CTF fit, and the maximum resolution to which a good fit was obtained for each image of the tilt series (Fig. 3; green). The values for rlnCtfIceRingDensity are calculated as the power of the tilt series images between 0.25 and 0.28 Å−1. 43

Selection of tilt series images

Given a tilt_series.star file, the ‘Exclude tilt‐images’ job launches a Napari‐based 44 , 45 viewer that displays the images in a tilt series and allows the user to de‐select unsuitable images, for example images in which the field of view has shifted or is obstructed by the grid bars, or if there is significant radiation induced movement detected. A screenshot of the viewer is shown in Fig. 4A.

Fig. 4.

Fig. 4

Screenshots of Napari‐based picking tools. (A) The tilt‐series selection program opened by the “Exclude tilt‐images” job allows the user to de‐select tilt images for each tilt series. (B) The picker plug‐in opened by the “Pick tomograms” job allows one to annotate the reconstructed tomograms with spheres (shown here) or 1D‐curves (not shown), which will then be used to randomly sample particles with priors that orient the normal to the sphere or along the curves with the Z‐axis. (C) Individual particles can also be annotated manually, or read from a STAR file for visualisation.

The output selected_tilt_series.star file will point towards individual tilt series STAR files that no longer contain lines with de‐selected images.

Tilt series alignment

The ‘Align tilt series’ job implements wrappers for performing tilt series alignment in IMOD 7 and AreTomo. 8 In IMOD, fiducial markers or local image patches are tracked through a tilt series yielding observed 2D positions of a 3D object, and parameters of a projection model are fit to these observations. In AreTomo, a projection matching routine is used to iteratively improve projection model parameters. Each wrapper takes the information from an input tilt_series.star file and prepares the appropriate commands to align tilt series. This wrapper has not been parallelised.

The RELION‐5 projection model at this stage is defined as 5 parameters per tilt image. rlnTomoXTilt, rlnTomoYTilt and rlnTomoZRot constitute a set of extrinsic Euler angles (in °) which rotate the specimen within a fixed microscope coordinate system. In this coordinate system, the optical axis is aligned with the Z‐axis and the stage tilt axis is aligned with the Y‐axis. The centre of rotation is the centre of the tomogram. The first rotation around the X‐axis (rlnTomoXTilt) accounts for nonperpendicularity of the stage tilt axis to the optical (Z) axis. In most cases, this angle will be close to 0°. The second rotation around the Y‐axis (rlnTomoYTilt) is the stage tilt angle. The third rotation around the Z‐axis (rlnTomoZRot) aligns the Y axis to the tilt axis in the projection image. rlnTomoXShiftAngst and rlnTomoYShiftAngst are shifts (in Å from the centre of rotation) applied after rotating the specimen to align its projection to the experimental image data.

The five tilt alignment parameters describe the transformation required to go from 3D coordinates in the tomogram to 2D coordinates in the tilt series images. More specifically, denoting rlnTomoXTilt, rlnTomoYTilt and rlnTomoZRot as θ x , θ y and θ z , and rlnTomoXShiftAngst and rlnTomoYShiftAngst as Δ x and Δ y , respectively, the transformation matrix R that converts (centred) coordinates in the 3D tomogram to (centred) coordinates in the 2D tilt series images is thereby defined as:

R=Rt*Rz*Ry*Rx (1)

with

Rx=10000cosθxsinθx000sinθx0cosθx001;Ry=cosθy0sinθy00100sinθy000cosθy001;Rz=cosθzsinθz00sinθzcosθz0000001001;Rt=100Δx010Δy00001001

Alignment parameters are extracted from output files of IMOD or AreTomo, and written in an output aligned_tilt_series.star file (Fig. 3; blue).

Python application programming interfaces for tilt series alignment procedures in RELION are available as stand‐alone python packages ‘yet‐another‐imod‐wrapper’ (https://github.com/teamtomo/yet‐another‐imod‐wrapper) and ‘lil‐aretomo’ (https://github.com/teamtomo/lil‐aretomo).

Tomogram reconstruction

The ‘Reconstruct tomograms’ job takes an aligned_tilt_series.star file as input and uses real‐space weighted back‐projection, with premultiplication of the CTF using a single defocus value for each tilt image, to reconstruct the corresponding tomograms. As such, these tomograms will have small errors in defocus, which become worse as one moves away from the centre of the reconstruction. In the presented workflow, the reconstructed tomograms are only used for picking particles and subsequent averaging approaches use cropped 2D stacks as individual particles. Tomograms are therefore typically reconstructed with relatively large pixel sizes (e.g. 10 Å). If even/odd tilt series were calculated in the ‘Motion correction’ job, then even/odd tomograms can also be calculated for denoising in the ‘Denoise tomograms’ job below.

In addition, an option exists to apply an overall tilt angle offset to the tilt series, which may be useful, for example, when tilt series were collected on lamellae that were milled from a thick specimen at a preselected angle away from the direction of the electron beam using a focussed ion beam (FIB). While tilt alignment procedures in software like IMOD and AreTomo can solve for the tilt angle offset, the wrappers in RELION do not use this feature, and the tilt angle offset can only be provided at the tomogram reconstruction stage if known, for example, from the geometry of milled lamellae. Using a tilt angle offset to minimise the Z‐height of tomograms and thereby the required disc space is less important in the RELION pipeline than it is in subtomogram averaging pipelines where particles are extracted as sub‐volumes from large tomograms with small pixels. For the same reason, no X‐axis offset is estimated or corrected for in addition to the tilt image‐specific values of the rlnTomoXTilt parameter that is estimated during tilt series alignment.

The output tomograms.star file, besides containing all the information of the input tilt series file, contains links to the names of the reconstructed tomograms, which are saved in MRC format in a sub‐directory of the job called tomograms/. This job also adds a column with a CTF scale‐factor, calculated as the cosine of the tilt angle, to the STAR files of the individual tilt series images (Fig. 3; violet).

Denoising tomograms

The ‘Denoise tomograms’ job takes a tomograms.star file as input and implements a wrapper to the noise2noise denoising program cryo‐CARE. 46 In the first part of the job, a neural network is trained on a selected subset of representative even/odd tomograms that were constructed in the ‘Reconstruct tomograms’ job. In the second part of the job, the trained network is applied to the complete set of tomograms that were calculated from all movie frames.

The output tomograms.star file from this job points to the denoised tomograms, which are again stored in a sub‐directory of the job called tomograms/.

Picking particles

A tomograms.star file is used as input for the ‘Pick tomograms’ job, with Napari 44 as image viewer and napari‐threedee 47 to provide tools for the interactive 3D annotation of isolated particles, filaments or spheres on tomogram slices, similar to dtmslice from Dynamo. 48 Particles can be taken directly from particle annotations, or sampled from filaments or spheres. Picking on spheres (Fig. 4B) is useful for particles that are arranged, for example, on spherical virus capsids, vesicles or cells. Filaments are picked as 1D curves. An option to sample particles from nonspherical 2D surfaces was planned, but has not yet been implemented. The napari viewer can also be used to visualise sets of particle coordinates after alignment and/or classification in the latter parts of the pipeline (Fig. 4C). Sampling from different geometries is implemented in a stand‐alone python package ‘morphosamplers’ (https://github.com/morphometrics/morphosamplers).

The coordinates of manually picked particles, spheres or filaments of individual tomograms are saved in STAR files in a subdirectory called annotations/. When using spheres or filaments, particles are sampled from the spheres or filaments according to a user‐specified inter‐particle distance. The output particles.star file contains the 3D coordinates of all picked particles in all tomograms, with coordinates defined in Ångstroms relative to the centre of the tomogram (Fig. 5; Table 3). When particles are sampled from spheres or filaments, the output STAR file will also contain Euler angles that define a rotation that orients the Z‐axis of the particle tangential to the surface of the sphere or perpendicular to the long filament axis. Thereby, in subsequent particle alignment and classification, a 90° prior on the tilt angle will orient particles with their Z‐axis normal to the sphere surface or parallel to the long filament axis. Having non‐zero tilt angles prevents gimbal locks in the Euler angles during refinement in RELION.

Fig. 5.

Fig. 5

Particle metadata description after picking. Metadata to describe a set of particles after picking in the Napari GUI. A single STAR file contains particle coordinates (in Ångstroms relative to the centre of the tomogram) for all tomograms. The labels and data columns in violet arise from the sampling of the particles on manually picked spheres, and express information that aligns their Z‐axis with the normal to the surface sphere. Corresponding definitions of metadata labels are given in Table 3.

Table 3.

Metadata items for particle picks. Metadata labels, types and definitions for the description of a set of picked particles. Entries in bold are always present; entries in regular are only present when particles are sampled from spheres or filaments.

Label Type Definition
_rlnTomoName String Name of a tomogram (same as in Table  1 )
_rlnCenteredCoordinateXAngst Float X‐position in a tomogram, in Ångstroms from the centre
_rlnCenteredCoordinateYAngst Float Y‐position in a tomogram, in Ångstroms from the centre
_rlnCenteredCoordinateZAngst Float Z‐position in a tomogram, in Ångstroms from the centre
_rlnTomoSubtomogramRot Float First Euler angle (in °) that pre‐orients subtomograms in the tomogram coordinate system
_rlnTomoSubtomogramTilt Float Second Euler angle (in °) that pre‐orients subtomograms in the tomogram coordinate system
_rlnTomoSubtomogramPsi Float Third Euler angle (in °) that pre‐orients subtomograms in the tomogram coordinate system
_rlnAngleRot Float First Euler angle for particle alignment (rot, in °) as defined in 49
_rlnAngleTilt Float Second Euler angle for particle alignment (tilt, in °) as defined in 49
_rlnAnglePsi Float Third Euler angle for particle alignment (psi, in °) as defined in 49
_rlnAngleTiltPrior Float Centre of the prior on the second Euler angle for particle alignment (in °)
_rlnAnglePsiPrior Float Centre of the prior on the third Euler angle for particle alignment (in °)

Pseudo‐subtomograms and 2D particle stacks

The particles.star file from the ‘Pick tomograms’ job, together with the tomograms.star file from the ‘Reconstruct tomograms’ job, form the input to the ‘Extract subtomos’ job. Because these two files together define the input, they are bundled in a file called optimisation_set.star, which is also written out by the ‘Pick tomograms’ job. The optimisation_set.star contains links to the filenames of both individual STAR files. One can either use the optimisation_set.star file or the two separate STAR files as input for the ‘Extract subtomos’ job. This job will combine the 3D positions of the particles with the alignment parameters of the tomograms either to construct 3D pseudo‐subtomograms (as introduced in RELION‐4 18 ), or to crop the relevant regions around individual particles in the tilt series images and save these as 2D particle stacks. Both the pseudo‐subtomograms and the 2D stacks will be premultiplied with the CTF in images with a user‐defined box size, and then possibly cropped to a smaller box size after the signal delocalisation has been compensated. These calculations are parallelised using both MPI and threads. Because there are typically fewer images in the tilt series (N tilt ) than the cropped boxed size (B), 2D stacks (with N tilt * B 2 pixels) often occupy less disc space than 3D pseudo‐subtomograms (with B 3 voxels). Options exist to only use tilt series images below a user‐defined maximum dose and to only output particles that are visible on a user‐specified minimum number of tilt series images. Writing particles in float16 saves a factor of 2 in disc space compared to writing in the default float32 MRC format, although not all third‐party programs may be able to read such images.

The particle images are saved in a subdirectory of the job called Subtomograms/. The job also saves a particles.star file with the names of the particle images, and (for 2D stacks only) an array with names of images of the tilt series saved in the stacks. The job also writes out a new optimisation_set.star with the filenames of the output particles.star file and the input tomograms.star file.

Subsequent averaging approaches

The optimisation_set.star from the ‘Extract subtomos’ job can be used as input to any of the existing ‘3D initial reference’, ‘3D classification’, ‘3D auto‐refine’ or ‘3D multi‐body’ jobs in the standard RELION pipeline. These methods already existed in single‐particle and tomography pipelines of RELION 18 , 42 and are not described in detail here. Two new features are an option to resize input 3D reference maps and mask if they have a different pixel and/or box size than the input images (the same also works for standard single‐particle analysis), and an option to impose a prior on the tilt angle, which is useful if particles were sampled from spheres or filaments. Each of these jobs will also output optimisation_set.star files with links to the STAR files with the tomograms and the refined particle coordinates.

Bayesian polishing and CTF refinement

The methods for Bayesian polishing, which performs tilt series realignment and estimation of per‐particle motion throughout the tilt series (but does not follow motion of individual particles through individual movie frames), and CTF refinement, which refines the defocus estimates for each tilt series image, have been described previously 18 and will not be discussed here. Besides an input optimisation_set.star file, these methods also need reference half‐maps, a reference mask and a postprocess.star file from a ‘Post‐processing’ job as input. Because these calculations are sensitive to the size and grey‐scale of the reference, the reference maps need to be calculated by the ‘Reconstruct particle’ job, and a binning factor of 1 must be used. A larger binning factor may be used in the preceding ‘3D classification’ or ‘3D auto‐refine’ jobs.

The ‘CTF refinement’ job will update the defocus values for all tilt series images and store links to updated STAR files for the individual tilt series in an updated tomograms.star file. Likewise, the ‘Bayesian polishing’ job will update the tilt series alignment parameters in a new tomograms.star file. If per‐particle motion estimation is performed, then the ‘Bayesian polishing’ job will also output a motion.star file with the estimated per‐particle motion tracks for all particles. A link to these files will also be added to the output optimisation_set.star file.

If one wishes to perform further refinements or classifications after a ‘CTF refinement’ and/or a ‘Bayesian polishing’ job, new particles should be extracted using the ‘Extract subtomos’ job with the updated CTF parameters and/or tilt series alignment parameters and particle positions. If no further refinement or classification is deemed necessary, then calculating a new reconstruction with the ‘Reconstruct particle’ job will suffice.

Results

The RELION 5 pipeline has already proven useful for the processing of two cryo‐ET data sets. 50 , 51 Here, we also illustrate its capability on a benchmark data set consisting of five tomograms of virus‐like particles (VLPs) of the capsid and spacer peptide 1 (CA‐SP1) region of the Gag polyprotein in the immature human immunodeficiency virus 1 (HIV‐1). 52 The same subset of five tomograms has previously been used to benchmark various cryo‐ET tools. NovaCTF introduced 3D CTF correction, leading to 3.9 Å resolution 53 ; optimising frame alignment and CTF parameters in Warp led to 3.8 Å 40 ; a combination of Warp‐RELION‐M and Dynamo yielded 3.4 Å 54 ; and RELION‐4 gave 3.2 Å. 18

Starting from the import of raw movie frames and SerialEM mdoc files, all processing steps were performed using the pipeline described in the Methods section. Movie frames were aligned using RELION's own motion‐correction program; tilt series images without signals (dark images) were manually excluded from the first two tilt series; CTF parameters were estimated using CTFFIND4 35 ; and tilt alignment was performed using the fiducial‐based procedure in IMOD. 7 Next, the five tomograms were reconstructed with a binned voxel size of 10 Å and the napari‐based 3D picker was used to manually annotate 47 spheres coinciding with the outer surfaces of the VLPs observed in the tomograms. Using an interparticle distance of 60 Å, 30597 particles were extracted by randomly sampling the spheres. Priors on the orientations of the particles, which position the normal to the sphere parallel to the Z‐axis for each particle at a tilt angle of 90°, allowed us to obtain an initial 3D reference map by simply running a ‘Reconstruct particle’ job with sixfold rotational symmetry. The initial model was subjected to a first ‘3D auto‐refine’ job at a binning factor of 6, followed by a second ‘3D auto‐refine’ job at a binning factor of 2. After removing 8937 duplicated particles (using the corresponding option in a ‘Subset selection’ job), we then ran a ‘3D classification’ job with nine classes, local alignment and regularisation parameter T = 1. Manual selection of the best two classes resulted in a subset of 9053 particles. These particles were used for a 3D auto‐refinement at a binning factor of 1, which gave a map at 4.0 Å resolution. Five iterations of CTF refinement, Bayesian polishing and 3D auto‐refinement at a binning factor of 1 yielded a final map at 3.3 Å. Automated model building with ModelAngelo 34 led to an atomic model comprising 211 (94%) of the 224 residues that were built in the original study. 52

Fourier shell correlation (FSC) curves of some of the intermediate steps and the final reconstruction are shown in Fig. 6A and the final map is shown in Fig. 6B. In Fig. 6C, we show representative regions from the reconstructed map after the first 3D refinement step at binning factor 1 (first row), after the first iteration of CTF refinement, Bayesian polishing and 3D auto‐refinement (second row), and in the final map (third row), as well as the final map with the atomic model as output by the ModelAngelo job, with no additional model refinements. The computational costs of the individual steps in the pipeline are summarised in Table 4. For all 3D auto‐refinements and the 3D classification leading to the final map, individual particles were written as 2D image stacks in the corresponding ‘Extract subtomos’ jobs. For comparison with RELION‐4, we also extracted particles at a binning factor of 1 as 3D pseudo‐subtomograms. This required 20 times more disc space and resulted in an eightfold increase in wall‐clock compute time for 3D autorefinement, compared to extracting particles as 2D image stacks.

Fig. 6.

Fig. 6

Subtomogram averaging of the CA‐SP1 hexamer of the immature HIV1 capsid. (A) FSC curves at intermediate and final refinement steps in the RELION‐5 pipeline. (B) Final reconstructed map, side view (top), top (bottom left) and bottom views (bottom right). (C) Representative regions in the reconstructed map after the first 3D auto‐refinement at binning factor of 1 (blue, top row), after the first iteration of CTF refinement, Bayesian polishing and 3D auto‐refinement (orange, second row), after the fifth iteration (magenta, third row), and the final map with the atomic model generated by ModelAngelo (green, bottom row).

Table 4.

Computational costs. Each row represents a job described in the Results section, reporting its wall‐clock time required for execution; how many MPI processes, threads and/or GPUs were used; and how much disc space the job required. All computations were performed on an (Ubuntu 22.04.3 LTS) Linux workstation with an AMD Ryzen Threadripper PRO 5975WX CPU with 32 cores and 64 threads, 512 GB RAM and two NVIDIA GeForce RTX 4090 GPUs, each with 24GB memory.

Jobs Time Disk space MPI processes, threads, GPUs Other information
Motion correction 6 min 17 GB 32 MPI, 4 threads
CTF estimation <1 min 212 MB 32 MPI CTFFIND‐4.1
Align tilt‐series 6 min 11 GB IMOD
Reconstruct tomograms 3 min 3 GB 5 MPI, 12 threads Binned pixel size 10 Å
Denoise tomograms/train 1 h 22 min 20 MB 1 GPU
Denoise tomograms/predict 1 min 1.5 GB
Extract subtomos/bin6 17 min 8.6 GB 5 MPI, 12 threads 30 597 particles; Binning factor 6; Box size 96
3D auto‐refine/bin6 3 h 31 min 600 MB 5 MPI, 6 threads, 2 GPUs 26 iterations
Extract subtomos/bin2 3 min 16 GB 5 MPI, 12 threads 30 597 particles; Binning factor 2; Box size 128
Reconstruct particle/bin2 5 min 600 MB
3D auto‐refine/bin2 7 h 35 min 754 MB 5 MPI, 6 threads, 2 GPUs 24 iterations
3D classification 1 h 55 min 2.1 GB 21 660 particles; 25 iterations
Extract subtomos/bin1 3 min 11 GB 5 MPI, 12 threads 9053 particles; Binning factor 1; Box size 192
Reconstruct particle/bin1 4 min 4.6 GB
3D auto‐refine/bin1 1 h 41 min 1.6 GB 5 MPI, 6 threads, 2 GPUs 13 iterations
CTF refinement 6 min <1 MB 5 MPI, 12 threads
Bayesian polishing 1 h 32 min 270 MB
Extract subtomos/bin1 (3D pseudo‐subtomos) 24 min 240 GB 5 MPI, 12 threads 9053 particles; Binning factor 1; Box size 192
3D auto‐refine/bin1 (3D pseudo‐subtomos) 13 h 57 min 1.5 GB 5 MPI, 6 threads, 2 GPUs 12 iterations

The results described above have been made available through Zenodo (DOI: 10.5281/zenodo.11068319) and form the basis for a detailed tutorial of the RELION‐5 tomography pipeline, which can be found at http://relion.readthedocs.io. The final map is also available from the EMDB (EMD‐50672).

Discussion

Gaining structural insights from cryo‐ET data is complex and remains an active area of research, with ongoing developments in CTF estimation, 39 tilt series alignment, 8 tomogram denoising, 21 , 23 , 24 , 46 , 55 automated localisation of structures of interest 30 , 56 , 57 , 58 , 59 , 60 , 61 , 62 , 63 , 64 , 65 and approaches to deal with structural heterogeneity among extracted particles, 66 , 67 , 68 , 69 among many others. As a result, cryo‐ET image processing pipelines, including the one described here, will need to be continuously updated in the coming years. Probably, the most urgent extensions to the pipeline in RELION‐5 would include a more robust handling of reconstruction and defocus handedness from the provided metadata, more robust tools for automated tilt series alignment, and the incorporation of tools for automated tomogram segmentation and particle picking. Given the popularity of the combination of Warp/M 1 , 40 with subtomogram averaging in RELION‐3, 70 updated interoperability between Warp/M and RELION‐5 is also desirable. Moreover, the use of napari as a platform for the development of these picking tools introduces complexity, especially for viewing data remotely. To avoid the maintenance burden of this complexity in RELION, the napari‐based tools are intended to become standalone tools in the near future.

However, different research groups will require different tools to fulfil their specific image‐processing needs, and it will remain difficult to include all of these in a single software pipeline. Recognising this, much of the work presented here was implemented as stand‐alone python packages designed to maximise their reusability outside of the RELION framework. It would be helpful if the different groups that use and write cryo‐ET software were to agree on a standardised description that allows passing metadata between different programs. In single‐particle analysis, the exchange of metadata is often performed through files in the STAR format, 38 with geometric definitions originally defined by Heymann et al 49 and then implemented in RELION. 71 The explicit definition of metadata structures for tilt series, tomograms and extracted particles presented in this paper could fulfil a similar role for the cryo‐ET field. A first step on this journey may be the adoption of RELION‐5's STAR files for cryo‐ET data (Figs 2, 3 & 5 and Tables 1, 2, 3) in the upcoming cryo‐ET pipeline of the Collaborative Computing Project for Electron cryo‐Microscopy (CCP‐EM) software. 37 This new software will provide mechanisms for pipelining a wider range of cryo‐ET programs than those available within RELION‐5 (Tom Burnley, personal communication), thus further improving the accessibility for newcomers to this rapidly developing field.

Conclusions

We present a pipeline for the analysis of cryo‐ET data in RELION‐5 that ranges from the import of unprocessed movies to automated atomic modelling in high‐resolution subtomogram averaging maps. The explicit metadata definitions of tilt series, tomograms and extracted particles in RELION‐5 may also serve wider efforts at standardisation and software inter‐operability.

Conflict of interest

The authors declare no conflicts of interest.

Peer review

The peer review history for this article is available at https://www.webofscience.com/api/gateway/wos/peer‐review/10.1002/2211‐5463.13873.

Author contributions

AB, BT, RW, EP, JZ, DK and SHWS wrote computer code; AB, BT, RW and SHWS performed experiments and analysed results. SHWS supervised the project. All authors contributed to writing the manuscript.

Acknowledgements

We are grateful to Jake Grimmett, Toby Darling and Ivan Clayson for help with high‐performance computing, and to Takanori Nakane for helpful discussions and RELION support. AB was a member of the group of David Barford (DB). This work was supported by the Medical Research Council as part of the United Kingdom Research and Innovation (MC_UP_A025_1013 to SHWS; MC_UP_1201/31 to TAMB and MC_UP_1201/6 to DB). AvK is supported by a Philip Leverhulme Prize to TAMB. RW is supported by The Pittsburgh Foundation & The Commonwealth of Pennsylvania Formula Fund to Zachary Freyberg; EP was supported by grants from the European Research Council (ERC‐StG‐2019 grant 852915) and the BBSRC (grant BB/T002670/1) to Giulia Zanetti. For the purpose of open access, the MRC Laboratory of Molecular Biology has applied a CC BY public copyright licence to any Author Accepted Manuscript version arising.

Edited by Cláudio Soares

Alister Burt and Bogdan Toader contributed equally to this article.

Data accessibility

RELION is distributed under a GPLv2 open‐source software licence and can be downloaded for free from http://www.github.com/3dem/relion. All stand‐alone python packages are distributed under a BSD‐3 open‐source software licence. The data set described in the Results section can be downloaded from EMPIAR under accession number 10164, whereas the results themselves can be downloaded from Zenodo (DOI: 10.5281/zenodo.11068319).

References

  • 1. Tegunov D, Xue L, Dienemann C, Cramer P and Mahamid J (2021) Multi‐particle cryo‐EM refinement with M visualizes ribosome‐antibiotic complex at 3.5 Å in cells. Nat Methods 18, 186–193. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2. Xing H, Taniguchi R, Khusainov I, Kreysing JP, Welsch S, Turoňová B and Beck M (2023) Translation dynamics in human cells visualized at high resolution reveal cancer drug action. Science 381, 70–75. [DOI] [PubMed] [Google Scholar]
  • 3. Xue L, Lenz S, Zimmermann‐Kogadeeva M, Tegunov D, Cramer P, Bork P, Rappsilber J and Mahamid J (2022) Visualizing translation dynamics at atomic detail inside a bacterial cell. Nature 610, 205–211. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4. Mastronarde DN (2005) Automated electron microscope tomography using robust prediction of specimen movements. J Struct Biol 152, 36–51. [DOI] [PubMed] [Google Scholar]
  • 5. Schorb M, Haberbosch I, Hagen WJH, Schwab Y and Mastronarde DN (2019) Software tools for automated transmission electron microscopy. Nat Methods 16, 471–477. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6. Kremer JR, Mastronarde DN and McIntosh JR (1996) Computer visualization of three‐dimensional image data using IMOD. J Struct Biol 116, 71–76. [DOI] [PubMed] [Google Scholar]
  • 7. Mastronarde DN and Held SR (2017) Automated tilt series alignment and tomographic reconstruction in IMOD. J Struct Biol 197, 102–113. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8. Zheng S, Wolff G, Greenan G, Chen Z, Faas FGA, Bárcena M, Koster AJ, Cheng Y and Agard DA (2022) AreTomo: an integrated software package for automated marker‐free, motion‐corrected cryo‐electron tomographic alignment and reconstruction. J Struct Biol X 6, 100068. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9. Radermacher M (1988) Three‐dimensional reconstruction of single particles from random and nonrandom tilt series. J Electron Microsc Tech 9, 359–394. [DOI] [PubMed] [Google Scholar]
  • 10. Gilbert P (1972) Iterative methods for the three‐dimensional reconstruction of an object from projections. J Theor Biol 36, 105–117. [DOI] [PubMed] [Google Scholar]
  • 11. Agulleiro JI and Fernandez JJ (2011) Fast tomographic reconstruction on multicore computers. Bioinformatics 27, 582–583. [DOI] [PubMed] [Google Scholar]
  • 12. Fernandez J‐J, Li S, Bharat TAM and Agard DA (2018) Cryo‐tomography tilt‐series alignment with consideration of the beam‐induced sample motion. J Struct Biol 202, 200–209. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13. Gordon R, Bender R and Herman GT (1970) Algebraic reconstruction techniques (ART) for three‐dimensional electron microscopy and x‐ray photography. J Theor Biol 29, 471–481. [DOI] [PubMed] [Google Scholar]
  • 14. Marabini R, Herman GT and Carazo JM (1998) 3D reconstruction in electron microscopy using ART with smooth spherically symmetric volume elements (blobs). Ultramicroscopy 72, 53–65. [DOI] [PubMed] [Google Scholar]
  • 15. Kunz M and Frangakis AS (2014) Super‐sampling SART with ordered subsets. J Struct Biol 188, 107–115. [DOI] [PubMed] [Google Scholar]
  • 16. Böhning J, Bharat TAM and Collins SM (2022) Compressed sensing for electron cryotomography and high‐resolution subtomogram averaging of biological specimens. Structure 30, 408–417.e4. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17. Himes BA and Zhang P (2018) emClarity: software for high‐resolution cryo‐electron tomography and subtomogram averaging. Nat Methods 15, 955–961. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18. Zivanov J, Otón J, Ke Z, von Kügelgen A, Pyle E, Qu K, Morado D, Castaño‐Díez D, Zanetti G, Bharat TAM et al. (2022) A Bayesian approach to single‐particle electron cryo‐tomography in RELION‐4.0. eLife 11, e83724. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19. Liu H‐F, Zhou Y, Huang Q, Piland J, Jin W, Mandel J, Du X, Martin J and Bartesaghi A (2023) nextPYP: a comprehensive and scalable platform for characterizing protein variability in situ using single‐particle cryo‐electron tomography. Nat Methods 20, 1909–1919. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20. Bartesaghi A, Lecumberry F, Sapiro G and Subramaniam S (2012) Protein secondary structure determination by constrained single‐particle cryo‐electron tomography. Structure 20, 2003–2013. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21. Liu Y‐T, Zhang H, Wang H, Tao C‐L, Bi G‐Q and Zhou ZH (2022) Isotropic reconstruction for electron tomography with deep learning. Nat Commun 13, 6482. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22. Buchholz T‐O, Krull A, Shahidi R, Pigino G, Jékely G and Jug F (2019) Content‐aware image restoration for electron microscopy. Methods Cell Biol 152, 277–289. [DOI] [PubMed] [Google Scholar]
  • 23. Bepler T, Kelley K, Noble AJ and Berger B (2020) Topaz‐denoise: general deep denoising models for cryoEM and cryoET. Nat Commun 11, 5208. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 24. Zhang H, Li Y, Liu Y, Li D, Wang L, Song K, Bao K and Zhu P (2023) A method for restoring signals and revealing individual macromolecule states in cryo‐ET, REST. Nat Commun 14, 2937. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 25. de Teresa‐Trueba I, Goetz SK, Mattausch A, Stojanovska F, Zimmerli CE, Toro‐Nahuelpan M, Cheng DWC, Tollervey F, Pape C, Beck M et al. (2023) Convolutional networks for supervised mining of molecular patterns within cellular context. Nat Methods 20, 284–294. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 26. Moebel E, Martinez‐Sanchez A, Lamm L, Righetto RD, Wietrzynski W, Albert S, Larivière D, Fourmentin E, Pfeffer S, Ortiz J et al. (2021) Deep learning improves macromolecule identification in 3D cellular cryo‐electron tomograms. Nat Methods 18, 1386–1394. [DOI] [PubMed] [Google Scholar]
  • 27. Heumann JM, Hoenger A and Mastronarde DN (2011) Clustering and variance maps for cryo‐electron tomography using wedge‐masked differences. J Struct Biol 175, 288–299. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 28. Galaz‐Montoya JG, Flanagan J, Schmid MF and Ludtke SJ (2015) Single particle tomography in EMAN2. J Struct Biol 190, 279–290. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 29. Castaño‐Díez D, Kudryashev M, Arheit M and Stahlberg H (2012) Dynamo: a flexible, user‐friendly development tool for subtomogram averaging of cryo‐EM data in high‐performance computing environments. J Struct Biol 178, 139–151. [DOI] [PubMed] [Google Scholar]
  • 30. Wan W, Khavnekar S and Wagner J (2024) STOPGAP: an open-source package for template matching, subtomogram alignment and classification. Acta Cryst, Sect D: Struct Biol 80. doi: 10.1107/S205979832400295X [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 31. Balyschew N, Yushkevich A, Mikirtumov V, Sanchez RM, Sprink T and Kudryashev M (2023) Streamlined structure determination by cryo‐electron tomography and subtomogram averaging using TomoBEAR. Nat Commun 14, 6543. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 32. de la Rosa‐Trevín JM, Quintana A, Del Cano L, Zaldívar A, Foche I, Gutiérrez J, Gómez‐Blanco J, Burguet‐Castell J, Cuenca‐Alba J, Abrishami V et al. (2016) Scipion: a software framework toward integration, reproducibility and validation in 3D electron microscopy. J Struct Biol 195, 93–99. [DOI] [PubMed] [Google Scholar]
  • 33. Jiménez de la Morena J, Conesa P, Fonseca YC, de Isidro‐Gómez FP, Herreros D, Fernández‐Giménez E, Strelak D, Moebel E, Buchholz TO, Jug F et al. (2022) ScipionTomo: towards cryo‐electron tomography software integration, reproducibility, and validation. J Struct Biol 214, 107872. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 34. Jamali K, Käll L, Zhang R, Brown A, Kimanius D and Scheres SHW (2024) Automated model building and protein identification in cryo‐EM maps. Nature 628, 450–457. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 35. Rohou A and Grigorieff N (2015) CTFFIND4: fast and accurate defocus estimation from electron micrographs. J Struct Biol 192, 216–221. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 36. Wood C, Burnley T, Patwardhan A, Scheres S, Topf M, Roseman A and Winn M (2015) Collaborative computational project for electron cryo‐microscopy. Acta Crystallogr D Biol Crystallogr 71, 123–126. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 37. Burnley T, Palmer CM and Winn M (2017) Recent developments in the CCP‐EM software suite. Acta Crystallogr D Struct Biol 73, 469–477. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 38. Hall SR (1991) The STAR file: a new format for electronic data transfer and archiving. J Chem Inf Comput Sci 31, 326–333. [Google Scholar]
  • 39. Elferich J, Kong L, Zottig X and Grigorieff N (2024) CTFFIND5 provides improved insight into quality, tilt and thickness of TEM samples. bioRxiv 2024-02. doi: 10.7554/eLife.97227.1 [DOI] [Google Scholar]
  • 40. Tegunov D and Cramer P (2019) Real‐time cryo‐electron microscopy data preprocessing with Warp. Nat Methods 16, 1146–1152. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 41. Zheng SQ, Palovcak E, Armache J‐P, Verba KA, Cheng Y and Agard DA (2017) MotionCor2: anisotropic correction of beam‐induced motion for improved cryo‐electron microscopy. Nat Methods 14, 331–332. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 42. Zivanov J, Nakane T, Forsberg BO, Kimanius D, Hagen WJ, Lindahl E and Scheres SH (2018) New tools for automated high‐resolution cryo‐EM structure determination in RELION‐3. eLife 7, e42166. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 43. González‐Rodríguez N, Areán‐Ulloa E and Fernández‐Leiro R (2024) A web‐based dashboard for RELION metadata visualization. Acta Cryst D 80, 93–100. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 44. Ahlers J, Althviz Moré D, Amsalem O, Anderson A, Bokota G, Boone P, Bragantini J, Buckley G, Burt A, Bussonnier M et al. (2023) Napari: a multi‐dimensional image viewer for python. Version v0.4.18 (Zenodo). doi:  10.5281/zenodo.8115575 [DOI]
  • 45. Gaifas L, Kirchner MA, Timmins J and Gutsche I (2024) blik: an extensible napari plugin for cryo-ET data visualisation, annotation and analysis. bioRxiv 2023-12.doi: 10.1371/journal.pbio.3002447 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 46. Buchholz T‐O, Jordan M, Pigino G and Jug F (2018) Cryo‐CARE: content‐aware image restoration for Cryo‐transmission electron microscopy data. arXiv [Preprint]. doi: 10.48550/arXiv.1810.05420 [DOI]
  • 47. Yamauchi KA and Burt A (2023) Napari‐threedee: a toolkit for human‐in‐the‐loop 3D image analysis in napari. bioRxiv [Preprint]. doi: 10.1101/2023.07.28.550950 [DOI]
  • 48. Castaño‐Díez D, Kudryashev M and Stahlberg H (2017) Dynamo catalogue: geometrical tools and data management for particle picking in subtomogram averaging of cryo‐electron tomograms. J Struct Biol 197, 135–144. [DOI] [PubMed] [Google Scholar]
  • 49. Heymann JB, Chagoyen M and Belnap DM (2005) Common conventions for interchange and archiving of three‐dimensional electron microscopy information in structural biology. J Struct Biol 151, 196–207. [DOI] [PubMed] [Google Scholar]
  • 50. Ochner H, Böhning J, Wang Z, Tarafder AK, Caspy I and Bharat TAM (2024) Structure of the Pseudomonas aeruginosa PAO1 type IV pilus. bioRxiv [Preprint]. doi: 10.1101/2024.04.09.588664 [DOI]
  • 51. Pyle E and Zanetti G (2024) Cryo‐electron tomography reveals how COPII assembles on cargo‐containing membranes. bioRxiv [Preprint]. doi: 10.1101/2024.01.17.576008 [DOI]
  • 52. Schur FKM, Obr M, Hagen WJH, Wan W, Jakobi AJ, Kirkpatrick JM, Sachse C, Kräusslich H‐G and Briggs JAG (2016) An atomic model of HIV‐1 capsid‐SP1 reveals structures regulating assembly and maturation. Science 353, 506–508. [DOI] [PubMed] [Google Scholar]
  • 53. Turoňová B, Schur FKM, Wan W and Briggs JAG (2017) Efficient 3D‐CTF correction for cryo‐electron tomography using NovaCTF improves subtomogram averaging resolution to 3.4 Å. J Struct Biol 199, 187–195. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 54. Burt A, Gaifas L, Dendooven T and Gutsche I (2021) A flexible framework for multi‐particle refinement in cryo‐electron tomography. PLoS Biol 19, e3001319. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 55. Yang Z, Zang D, Li H, Zhang Z, Zhang F and Han R (2024) Self‐supervised noise modeling and sparsity guided electron tomography volumetric image denoising. Ultramicroscopy 255, 113860. [DOI] [PubMed] [Google Scholar]
  • 56. Heebner JE, Purnell C, Hylton RK, Marsh M, Grillo MA and Swulius MT (2022) Deep learning‐based segmentation of Cryo‐electron tomograms. J Vis Exp doi: 10.3791/64435 [DOI] [PubMed] [Google Scholar]
  • 57. Zhou B, Yu H, Zeng X, Yang X, Zhang J and Xu M (2020) One‐shot learning with attention‐guided segmentation in Cryo‐electron tomography. Front Mol Biosci 7, 613347. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 58. Lamm L, Righetto RD, Wietrzynski W, Pöge M, Martinez‐Sanchez A, Peng T and Engel BD (2022) MemBrain: a deep learning‐aided pipeline for detection of membrane proteins in Cryo‐electron tomograms. Comput Methods Prog Biomed 224, 106990. [DOI] [PubMed] [Google Scholar]
  • 59. Siggel M, Jensen RK, Mahamid J and Kosinski J (2024) ColabSeg: an interactive tool for editing, processing, and visualizing membrane segmentations from cryo-ET data. J Struct Biol 216, 108067. doi:  10.1016/j.jsb.2024.108067 [DOI] [PubMed] [Google Scholar]
  • 60. Sazzed S, Scheible P, He J and Wriggers W (2022) Spaghetti tracer: a framework for tracing semiregular filamentous densities in 3D tomograms. Biomol Ther 12, 1022. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 61. Dimchev G, Amiri B, Fäßler F, Falcke M and Schur FK (2021) Computational toolbox for ultrastructural quantitative analysis of filament networks in cryo-ET data. J Struct Biol 213, 107808. doi: 10.1016/j.jsb.2021.107808 [DOI] [PubMed] [Google Scholar]
  • 62. Pyle E, Hutchings J and Zanetti G (2022) Strategies for picking membrane‐associated particles within subtomogram averaging workflows. Faraday Discuss 240, 101–113. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 63. Rice G, Wagner T, Stabrin M, Sitsel O, Prumbaum D and Raunser S (2023) TomoTwin: generalized 3D localization of macromolecules in cryo‐electron tomograms with structural data mining. Nat Methods 20, 871–880. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 64. Genthe E, Miletic S, Tekkali I, Hennell James R, Marlovits TC and Heuser P (2023) PickYOLO: fast deep learning particle detector for annotation of cryo electron tomograms. J Struct Biol 215, 107990. [DOI] [PubMed] [Google Scholar]
  • 65. Chaillet ML, van der Schot G, Gubins I, Roet S, Veltkamp RC and Förster F (2023) Extensive angular sampling enables the sensitive localization of macromolecules in electron tomograms. Int J Mol Sci 24, 13375. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 66. Powell BM and Davis JH (2024) Learning structural heterogeneity from cryo-electron sub-tomograms with tomoDRGN. Nat Methods 1–2. doi: 10.1038/s41592-024-02210-z [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 67. Harastani M, Eltsov M, Leforestier A and Jonic S (2021) HEMNMA‐3D: Cryo electron tomography method based on Normal mode analysis to study continuous conformational variability of macromolecular complexes. Front Mol Biosci 8, 663121. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 68. Barchet C, Fréchin L, Holvec S, Hazemann I, von Loeffelholz O and Klaholz BP (2023) Focused classifications and refinements in high‐resolution single particle cryo‐EM analysis. J Struct Biol 215, 108015. [DOI] [PubMed] [Google Scholar]
  • 69. Sibert BS, Kim JY, Yang JE, Hannon‐Hatfield A, Ke Z, Garfinkel DJ and Wright ER (2023) Workflow for high‐resolution sub‐volume averaging from heterogenous viral and virus‐like assemblies. Microsc Microanal 29, 943–944. [Google Scholar]
  • 70. Bharat TAM and Scheres SHW (2016) Resolving macromolecular structures from electron cryo‐tomography data using subtomogram averaging in RELION. Nat Protoc 11, 2054–2065. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 71. Scheres SHW (2012) RELION: implementation of a Bayesian approach to cryo‐EM structure determination. J Struct Biol 180, 519–530. [DOI] [PMC free article] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Data Availability Statement

RELION is distributed under a GPLv2 open‐source software licence and can be downloaded for free from http://www.github.com/3dem/relion. All stand‐alone python packages are distributed under a BSD‐3 open‐source software licence. The data set described in the Results section can be downloaded from EMPIAR under accession number 10164, whereas the results themselves can be downloaded from Zenodo (DOI: 10.5281/zenodo.11068319).