Abstract
Purpose:
Dynamic glucose enhanced (DGE) MRI studies employ CEST or spin lock (CESL) to study glucose uptake. Currently, these methods are hampered by low effect size and sensitivity to motion. To overcome this, we propose to utilize exchange-based linewidth () broadening of the direct water saturation (DS) curve of the water saturation spectrum (Z-spectrum) during and after glucose infusion (DS-DGE MRI).
Methods:
To estimate the glucose-infusion-induced changes (), Bloch-McConnell simulations were performed for normoglycemia and hyperglycemia in blood, gray matter (GM), white matter (WM), CSF, and malignant tumor tissue. Whole-brain DS-DGE imaging was implemented at 3 T using dynamic Z-spectral acquisitions (1.2 s per offset frequency, 38 s per spectrum) and assessed on four brain tumor patients using infusion of 35 g of D-glucose. To assess , a deep learning-based Lorentzian fitting approach was used on voxel-based DS spectra acquired before, during, and post-infusion. Area-under-the-curve () images, obtained from the dynamic time curves, were compared qualitatively to perfusion-weighted imaging parametric maps.
Results:
In simulations, was 1.3%, 0.30%, 0.29/0.34%, 7.5%, and 13% in arterial blood, venous blood, GM/WM, malignant tumor tissue, and CSF, respectively. In vivo, was approximately 1% in GM/WM, 5% to 20% for different tumor types, and 40% in CSF. The resulting DS-DGE maps clearly outlined lesion areas.
Conclusions:
DS-DGE MRI is highly promising for assessing D-glucose uptake. Initial results in brain tumor patients show high-quality maps of glucose-induced line broadening and DGE-based lesion enhancement similar and/or complementary to perfusion-weighted imaging.
Keywords: CEST, direct saturation (DS), dynamic glucose enhanced (DGE) MRI, glucoCEST, Z-spectra
1 |. INTRODUCTION
Gadolinium (Gd)-based contrast agents (GBCA) play a major role in MRI for both research and clinical routine. However, their use in certain patient groups is limited due to side effects such as nephrogenic systemic fibrosis.1,2 In vivo deposition3–5 has also led to the United States Food and Drug Administration (FDA) issuing a box package warning on GBCAs, which remain under continued review.6 In addition, many malignant tumors show little to no Gd-enhancement.7,8 Consequently, current clinical practice is judicious use of GBCA, particularly in young and vulnerable populations.9 Therefore, there is a need to develop new contrast agents. The availability of chemical exchange saturation transfer (CEST)10–17 and chemical exchange sensitive spin lock (CESL)18–29 approaches has opened the MRI field to non-metallic contrast agents. When D-glucose (D-Glc) is used as an agent, these approaches have been dubbed glucoCEST and glucoCESL, respectively.
Dynamic glucose enhanced (DGE) MRI applies CEST or CESL of sugars dynamically, providing information on contrast agent uptake in tissues in a manner similar to dynamic contrast enhanced (DCE) and dynamic susceptibility contrast (DSC) MRI.13,14,23–25,27,28,30–35 Unfortunately, DGE MRI signal changes at clinical field strengths (3 T) have been on the order of 1%,17,32 resulting in sensitivity to motion artifacts, especially when based on a single signal intensity from one saturation offset frequency per dynamic.36,37 Here, we propose to reduce these problems by utilizing the transverse relaxation effect originating from the chemical shift difference between the hydroxyl and water proton pools. The exchange between these pools leads to spins experiencing different precession frequencies,38,39 resulting in a collective phase dispersion and a linewidth () broadening of the direct saturation (DS) line shape in water saturation spectra (Z-spectra). We exploit this exchange-based relaxation enhancement to assess changes in D-Glc concentration by acquiring DS spectra using RF saturation of short duration and low . This approach is commonly used in the water saturation shift referencing (WASSR) method40 for measuring shifts, since it minimizes contributions from semi-solid magnetization transfer contrast (MTC), CEST, and relayed nuclear Overhauser effects (rNOEs).41 This allows fitting of the full DS spectrum to a Lorentzian,42 a method recently optimized using deep learning (DL).43
This study, therefore, aimed to develop a WASSR-analogous DGE method to utilize glucose-induced increases in the DS linewidth, dubbed DS-DGE MRI. Its feasibility in different types of tissues at 3 T was investigated through simulations and for D-Glc infusions in brain tumor patients. The in vivo DS-DGE effect size was compared qualitatively with DCE- and DSC-MRI parametric maps to investigate whether similar and/or complementary information was obtained.
2 |. METHODS
2.1 |. Simulations
Bloch-McConnell simulations of Z-spectra at 3 T before and after D-Glc infusion were performed using Pulseq-CEST.44 Five tissues were simulated: blood, gray matter (GM), white matter (WM), malignant tumor with blood–brain barrier (BBB) disruption (TUMOR), and CSF. Importantly, hydroxyl exchange properties and water transverse relaxation times may differ between tissue compartments, leading to different signal contributions. Therefore, following a recently established model,45 three tissue compartments were simulated within WM, GM and TUMOR, namely blood (b), extravascular extracellular space (EES), and cell (c). Within blood, we assumed arteriolar (a) and venular (v) compartments. A previous DGE MRI study reported a venous plasma blood glucose average increase of 9.8mM (n=11) for a D-Glc dose of 25 g.30 Since our experiments used a 40% higher dose (35 g), we assumed a 13.7 mM increase in venous blood D-Glc concentration from normoglycemic baseline to hyperglycemia. The resulting compartmental D-Glc concentrations after transport45 (Table 1) were used to calculate the Z-spectral intensities, , using the hydroxyl proton pools at 0.66, 1.28, 2.08, and 2.88 ppm.48 All compartments, except for TUMOR EES, were assumed to have a pH of 7.2, with hydroxyl proton exchange rates of 2900, 6500, 5200, and 14 300 Hz, respectively,48 at 37°C. For TUMOR EES, a pH of 6.8 was assumed, leading to exchange rates of 1500, 3100, 2500, and 6000 Hz, respectively.48 Compartmental Z-spectra were simulated using 41 frequency offsets: ±[10, 5.0, 4.0, 3.0, 2.5, 2.0, 1.75, 1.5, 1.25, 1.0, 0.80, 0.65, 0.55, 0.48, 0.40, 0.33, 0.26, 0.18, 0.11, 0.036, 0.0] ppm. Saturation was applied using 10 consecutive 50-ms sinc-Gaussian pulses, resulting in a total saturation time () of 0.5 s. The simulations were performed using a of .
TABLE 1.
Relaxation times,* fractional volumes, D-Glc concentrations, and water contents for tissues (b, GM, WM, TUMOR, and CSF) and their compartments (arterial, venous, EES, cell).
| Tissue | ||||||
|---|---|---|---|---|---|---|
| Parameter | Arterial blood** | Venous blood** | GM | WM | TUMOR | CSF |
| (s) | 1.91 | 1.73 | ||||
| (s) | 3.48 | 3.48 | 3.48 | 3.48 | ||
| (s) | 1.08 | 0.65 | 1.02 | |||
| (s) | 0.152 | 0.052 | ||||
| (s) | 2.78 | 2.78 | 2.78 | 2.78 | ||
| (s) | 0.071 | 0.055 | 0.071 | |||
| (mL comp/mL b) | 0.30 | 0.70 | ||||
| (mL b/mL tis) | 0.038 | 0.018 | 0.050 | 0.00 | ||
| (mL EES/mL tis) | 0.22 | 0.22 | 0.50 | 1.00 | ||
| (mL cell/mL tis) | 0.74 | 0.76 | 0.45 | 0.00 | ||
| ngl (mM) | 6.15 | 5.47 | ||||
| ngl (mM) | 2.24 | 2.42 | 5.45 | 3.69 | ||
| ngl (mM) | 0.167 | 0.381 | 0.811 | |||
| hgl (mM) | 19.8 | 19.1 | ||||
| hgl (mM) | 6.10 | 5.71 | 17.6 | 11.9 | ||
| hgl (mM) | 1.16 | 1.24 | 4.41 | |||
| (mL water/mL b) | 0.856 | 0.856 | 0.856 | |||
| (mL water/mL comp) | 0.938 | 0.938 | 0.938 | |||
| (mL water/mL comp) | 0.809 | 0.678 | 0.674 | |||
| (mL water/mL tis) | 0.839 | 0.738 | 0.815 | |||
Note: Values for , and are from Seidemo et al.45 and references therein.
Abbreviations: , arterial; , blood; , cell; , D-Glc concentration; comp, compartment; D-Glc, D-glucose; , EES; EES, extravascular extracellular space; , volume fraction ; GM, gray matter; hgl, hyperglycemia; ngl, normoglycemia; , water volume density; tis, tissue; TUMOR, malignant tumor with blood–brain barrier disruption; , venous; WM, white matter.
OH was set to have a of 1.20 s and of 0.100 s.
For arterial and venous blood, a well-mixed compartment (fast exchange for water between plasma and erythrocytes) was assumed for the 500 ms saturation period. Relaxation times were calculated based on references,46,47 using: https://www.kennedykrieger.org/physiologic-metabolic-anatomic-biomarkers/resources/software-and-databases/blood-t2-t1-hct-and-oxygenation-calculator and the following settings: 3 T; ; Hematocrit (Hct) 0.4; Oxygenations (Y) 0.98 and 0.60 for arterial and venous blood, respectively.
Z-spectra in GM, WM, and TUMOR were calculated by adding the normalized signals of blood, EES, and cell45:
| (1) |
with being the volume fraction for compartment in mL compartment/mL tissue (Table 1). Capillary blood was assumed to have fast deoxygenation, resulting in the blood compartment (b) consisting of only arteriolar (a) and venular (v) subcompartments49 with volume fractions and in mL/mL blood, respectively. Since MR contrast is determined by water-based compartmental concentrations and tissue volume fractions, corrections are included for , the water content per mL compartment or tissue (Table 1). After adding the Z-spectra from each compartment, 2% Rician noise was applied. The resulting tissue Z-spectra were re-sampled at frequencies used in the experiments (28 frequencies from −5 to 5 ppm), followed by Lorentzian fitting using deep-learning.43 Thereafter, the linewidth difference () between normoglycemia (baseline) and hyperglycemia was calculated for each tissue using Eq. 2 below.
2.2 |. Experiments
2.2.1 |. Patients
Four brain tumor patients were studied (one with an isocitrate dehydrogenase (IDH)-wildtype glioblastoma, two with a grade 2 IDH-mutated astrocytoma, and one with anaplastic lymphoma kinase (ALK)-mutated non–small-cell lung cancer metastases). The project was approved by the local Institutional Review Board, and each participant provided written informed consent. Participants were asked to fast 6 h before the study, but clear liquids were permitted. Before the start of the MRI examination, blood was drawn to verify normal baseline glucose levels (3.9–7.0 mM). Supporting Information (Table S1) lists the study’s exclusion criteria.
2.2.2 |. MRI acquisition protocol
Patients were examined on a 3 T Philips Elition RX system (Philips Healthcare). Pre- and post-contrast enhanced -MPRAGE, fluid-attenuated inversion recovery (FLAIR), DS-DGE, DCE and DSC images were acquired. -MPRAGE: TR/TE/flip angle (FA) = 7.5 ms/3.5 ms/8°, FOV = 212 × 212 mm2, resolution = 1.1 × 1.1 × 2.2 mm3, inversion time = 755 ms, acquisition time = 1 min 46 s, and FLAIR: TR/TE = 11000/120 ms, FOV = 212 × 212 mm2, resolution = 1.1 × 1.1 × 2.2 mm3, inversion time = 2800 ms, acquisition time = 3 min 51 s.
DS-DGE images were acquired using 10 consecutive 50-ms sinc-Gaussian pulses (), followed by a whole-brain simultaneous multi-slice EPI readout (multi-band factor 3). A total of 27 slices with FOV of 208 × 208 mm2 and a resolution of 2.2 × 2.2 × 4.4 mm3 were acquired using TR/TE/FA = 1200 ms/17 ms/52°. Thirty-two frequencies were acquired in 38.2 s at offsets: ±[10 (2×), 5.0, 2.5, 2.0, 1.5, 1.2, 1.0, 0.80, 0.70, 0.60, 0.50, 0.40, 0.30, 0.20, 0.10] ppm. In total 40 dynamics (Z-spectra) were acquired. Approximately 5 min into the scan (after the eighth dynamic), D-Glc was administrated intravenously with a power-injector at a rate of 6.25 g/min using hospital-grade D50 glucose (D50, Hospira; 35 g of D-Glc in 70 mL of water sterile solution prepared by the Johns Hopkins pharmacy), followed by a saline rinse. Total experiment duration was 25.5 min.
A -weighted gradient-echo sequence was used for DCE MRI: TR/TE/FA = 5.1 ms/2.5 ms/26°, FOV = 212 × 212 mm2, resolution = 2.2 × 2.2 × 4.4 mm3. Gadoteridol (ProHance, Bracco Diagnostics, 0.1 mmol/kg) was given at a rate of 5 mL/s via a power injector followed by a saline rinse. The injection delay was 30 s (15 pre-contrast baseline images). Each 15-slice dynamic scan was 2.0 s and a total of 150 dynamics over 5 min was acquired. After the DCE, post-contrast MPRAGE images were obtained.
Approximately 7 min after the first GBCA injection, a second GBCA dose was injected at a 5 mL/s rate via a power injector, followed by a saline rinse. DSC was performed using single-shot EPI with TR/TE/FA = 1344 ms/29 ms/90°, FOV = 212 × 212 mm2, and resolution = 2.2 × 2.2 × 4.4 mm3. A 15-s injection delay (11 pre-contrast baseline images) was set. A total of 100 dynamics were acquired over 2.2 min for 25 slices.
2.3 |. Post-processing
2.3.1 |. DS-DGE MRI
The dynamic DS spectrum signal intensities from each voxel were normalized using the average of the second of two acquisitions at ±10 ppm and then fitted to a Lorentzian line shape using the DL-based single Lorentzian fitting neural network.43 The , defined as in the DL fitting, were used to generate maps (in Hz) for each dynamic. The dynamic maps were rigid motion corrected using Elastix50 and visually inspected for remaining motion artifacts.
A baseline was calculated by averaging the maps obtained before infusion. To remove outliers, baseline greater or less than the average ± 2SD were discarded. was then calculated by averaging the remaining baseline . Dynamic images were obtained by subtracting from each dynamic image, , followed by normalization with :
| (2) |
Normalized area-under-the-curve () was calculated by subtracting from the average of dynamic obtained from infusion start and through the dynamic scan (), followed by normalization with :
| (3) |
Normalized over the infusion block alone was also calculated. Both fitting and calculations were performed in Python.
2.3.2 |. DCE and DSC MRI
DCE and DSC MRI were processed using OLEA Sphere (Olea Medical Solutions). For both sequences, motion correction was applied before the tracer kinetic modeling. The extended Toft model was used for DCE-MRI51 to retrieve interstitial volume () and the volume transfer constant (), a measure combining permeability and perfusion. For DSC-MRI, standard tracer kinetic modeling including leakage correction was applied to obtain leakage corrected cerebral blood volume (corr. CBV), uncorrected cerebral blood volume (uncorr. CBV) and leakage (K2).52,53
3 |. RESULTS
Figure 1 shows the simulated Z-spectra for the different tissues with and without glucose infusion. Simulated baseline were 87 Hz for blood, 57 Hz for arterial blood, 96 Hz for venous blood, 65 Hz for GM, 60 Hz for WM, 42 Hz for TUMOR, and 16 Hz for CSF. Simulated were 0.56% for blood, 1.3% for arterial blood, 0.30% for venous blood, 0.29% for GM, 0.34% for WM, 7.5% for TUMOR, and 13% for CSF.
FIGURE 1.
Normoglycemic () and hyperglycemic () simulated Z-spectra for tissue compartments (left) and total tissue (right). The lower row shows a zoomed-in view comparing the Z-spectral intensities around half-maximum for normoglycemia and hyperglycemia. Only Z-spectra with a sufficiently large change are visualized for tissue compartments (lower left). Saturation parameters: , 10 consecutive 50-ms sinc-Gauss pulses for .
Figure 2 shows a patient with a recurrent IDH-wildtype glioblastoma with thin peripheral contrast-enhancement around the resection cavity and surrounding expansile infiltrative FLAIR hyperintense tumor. Dynamic images obtained on infusing D-Glc are shown (averaged over two dynamics). Note the increases in vascular, CSF, and tumor tissue. For DCE, shows an increase in the same regions, while shows only a slight increase. Both and are hypointense inside the cavity. The DS-DGE map also displays an increase in the surrounding tumor and a hypointense core. For DSC, shows enhancement comparable to DS-DGE. During the D-Glc infusion, the increased to approximately 15% for the contrast enhanced peri-cavity infiltrative tumor region and 40% for ventricular CSF. Figure 3 shows experimental Z-spectra before and after D-Glc infusion together with DL Lorentzian fits for region-of-interests (ROIs) placed in the DS-DGE , with assistance from and FLAIR images, in the anterior cerebral artery, GM, WM, tumor tissue, and CSF (ventricle). Notice that the and their broadening are of a similar order of magnitude to those simulated in Figure 1. However, the in vivo changes in arterial blood and CSF were generally larger than those in the simulations.
FIGURE 2.
Patient with recurrent IDH-wildtype glioblastoma showing thin Gd-enhancement around the resection cavity. (Left) maps during the scan (averaged over a period of 76 s corresponding to two images). (Right top) Anatomical images (, FLAIR), together with parametric maps from DCE MRI (), DSC MRI (corr. CBV, K2) and DS-DGE MRI ( grayscale and color-coded). (Right bottom) Graph of linewidth change versus time obtained from regions of interest (ROIs) placed in the DS-DGE peri-cavity infiltrative tumor region, located anterior to the cavity, and ventricle (purple diamonds and blue squares, respectively). The ROIs are overlayed on the DS-DGE MRI map located in the graph as purple and blue areas, respectively. To visualize the trend in glucose uptake, curves were temporally smoothed with a 3-point moving average (purple and blue lines). , glucose-infusion-induced change; , area-under-curve; corr. , corrected cerebral blood volume; , dynamic contrast enhanced; , dynamic susceptibility contrast; , direct water saturation-dynamic glucose enhanced; , fluid-attenuated inversion recovery; , gadolinium; , leakage; , volume transfer constant; , linewidth, weighted; , interstitial volume.
FIGURE 3.
(Left) Normoglycemic and hyperglycemic experimental Z-spectra from the glioblastoma patient in Figure 2 and the corresponding DL Lorentzian fits. (Right) A zoomed-in view demonstrating the linewidth difference between normoglycemic and hyperglycemic experimental Z-spectra. ROI locations are displayed in the image and DS-DGE map. , area-under-curve; , direct water saturation-dynamic glucose enhanced; , gadolinium; , hyperglycemic; , normoglycemic; weighted.
Figure 4 shows a patient with a grade 2 IDH-mutated astrocytoma. Interestingly, the enhanced tumor rim in the DS-DGE image corresponds approximately to the spatial difference between the hypointense area in and the hyperintense area in FLAIR. The enhanced areas during infusion only and over the experimental duration are of comparable size. Corrected and also show an increase in the corresponding area. However, and appear normal.
FIGURE 4.
Patient with grade 2 IDH-mutated astrocytoma. Anatomical images (, FLAIR) together with corr. , and DS-DGE MRI maps ( in both grayscale and color-coded). Color-coded calculated from the infusion block only is also shown. A DS-DGE map overlayed on fused /FLAIR is shown for reference. Graph of linewidth change versus time obtained from region of interest (ROI) placed in the DS-DGE contrast-enhanced area (purple overlayed on DS-DGE AUC map). To visualize the trend in glucose uptake, curves (purple dots) were temporally smoothed with a 3-point moving average (purple line). , glucose-infusion-induced change; , area-under-curve; corr. , corrected cerebral blood volume; , direct water saturation-dynamic glucose enhanced; , fluid-attenuated inversion recovery; , gadolinium; , leakage; , volume transfer constant; , linewidth; weighted; , interstitial volume.
Figure 5 shows four slices from a patient with a grade 2 IDH-mutated astrocytoma. shows an increase in the tumor boundary, while shows only a slight increase. Similar to the patient in Figure 1, the lesion is strongly reduced in intensity and area after leakage correction, as reflected in the enhancement. Note that DS-DGE hyperintense and hypointense tumor regions correspond approximately to the hyperintense rims and hypointense cores in the FLAIR images, respectively. The images show a similar trend, but over a smaller area.
FIGURE 5.
Patient with a grade 2 IDH-mutated astrocytoma. Anatomical images (, FLAIR) together with parametric maps from DCE MRI (), DSC MRI (uncorr. , corr. ) and DS-DGE MRI (color-coded ) from four slices. , area-under-curve; corr. , corrected cerebral blood volume; , dynamic contrast enhanced; , dynamic susceptibility contrast; , direct water saturation-dynamic glucose enhanced; , fluid-attenuated inversion recovery; , gadolinium; , leakage; , volume transfer constant; weighted; uncorr. , uncorrected cerebral blood volume; , interstitial volume.
Figure 6 shows results for a patient with ALK-mutated non–small-cell lung cancer brain metastases. and are increased in the Gd-contrast-enhanced lesion area. The DS-DGE map also displays an increase in part of the lesion area, while the WM shows negligible change. Note that leakage correction strongly reduced the elevated uncorrected values in the lesion area. The enhancement observed on the contralateral side in the DS-DGE map is due to ventricular CSF. The time curve shows a continuous increase up to approximately 20% in the DS-DGE contrast-enhanced lesion, resulting in a relatively smaller lesion area enhancement in the DS-DGE map calculated from the infusion block only.
FIGURE 6.
Patient with brain metastasis from anaplastic lymphoma kinase-mutated non–small-cell lung cancer. Anatomical images (, FLAIR) together with , uncorr. , corr. , and DS-DGE MRI maps ( map from third dynamic, maps in both grayscale and color-coded). A color-coded calculated from the infusion block only is also shown. A graph of linewidth change versus time obtained from ROIs placed in the DS-DGE contrast-enhanced area and contralateral frontal WM is shown to the right (purple dots and orange diamonds, respectively). Regions of interest (ROIs) are overlayed on the DS-DGE MRI map in the graph as purple and orange areas, respectively. To visualize the trend in glucose uptake, curves were temporally smoothed with a three-point moving average (purple and orange lines). , glucose-infusion-induced change; , area-under-curve; corr. , corrected cerebral blood volume; , direct water saturation-dynamic glucose enhanced; , fluid-attenuated inversion recovery; Gd, gadolinium; , volume transfer constant; , linewidth; uncorr. , uncorrected cerebral blood volume; weighted; , interstitial volume.
4 |. DISCUSSION
We developed and implemented DS-DGE MRI to dynamically assess D-Glc uptake in brain tumors. This approach samples Z-spectra dynamically, which has the advantages of (1) multiple signal points per dynamic (leading to higher SNR and reduced motion sensitivity due to the Lorentzian fitting); (2) being independent of changes in the voxel between dynamics, for example, such as those due to motion, as the full DS spectrum is fitted; and (3) a water resonance line shape that has minimal contributions from CEST, rNOE, and MTC effects and can be approximated by a Lorentzian curve.40,42,54–56 When applying RF saturation with low , the FWHM of the DS Z-spectral line can be calculated as:
| (4) |
in which (in units of rad/s) and . The presence of exchangeable protons at an offset from the water resonance will increase 38,39 and broaden this line shape, with relative effects expected to be largest for tissue compartments with a long water . This was borne out in the simulations, where EES and CSF exhibited narrow lines and large changes when the D-Glc concentration was increased (Figure 1). When compartments were added proportionally (Eq. [1]) to generate the Z-spectra for GM, WM, and TUMOR, was highest in TUMOR, which is attributed to (1) the large concentration of D-Glc in EES after BBB breakdown; (2) increased blood volume and EES volume; and (3) the lower pH in tumor EES, which reduces the exchange rate and increases the exchange-based transverse relaxivity. For blood, despite having the highest D-Glc concentration, simulations showed a smaller change, which we attribute to the shorter originating from the high protein contents in both plasma (albumin) and erythrocyte (hemoglobin). Notice that an equal water-based concentration of D-Glc in plasma and erythrocytes57–59 was used, removing any effect of microvascular hematocrit on blood D-Glc concentrations. The small effects in WM and GM are attributed to the low D-Glc concentration in EES and even lower in the cells due to facilitated transport over the BBB and the cell membrane, respectively, and metabolism in the cells. The short of the cell compartment, which occupies a large volume fraction of the tissue, further reduces the DS-DGE effect size.
In patients, the effect sizes for the DS-DGE differences, , were generally on the same order of magnitude as those obtained from simulations. This can be observed in Figure 6, where the WM uptake curve shows a small increase during and after D-Glc infusion, and GM and WM intensities in the maps are close to zero. The same applies to the DS-DGE maps in Figures 2, 4, and 5. CSF had a smaller in the simulations than sometimes observed in the experimental data (~15% compared to up to 40%, Figure 2). This difference may be due to the deviation from a Lorentzian line shape because of so-called sidebands appearing as distinct patterns when using short RF pulses at high sampling rates.60 Sidebands are prone to occur in tissues with relatively long such as CSF or other liquid environments as necrotic tumor tissue.60 For our simulations, the offset frequencies were selected to minimize sideband occurrence. In the experiments, if the same sidebands occur before and after infusion start, the change, however, may not be significantly affected. Interestingly, the maximum increases in blood and CSF (Figure 2) in vivo were found to be larger than those from the simulations (Figure 1). One potential explanation for blood could be an osmotic increase in blood water content due to the higher D-Glc concentration and a concomitant increase in , previously suggested as a potential contributor to signal changes in CESL.26,61 Another potential explanation is that the increase in plasma blood glucose concentration, and subsequently the CSF, may be greater than what we assumed in our simulations. Previous DGE studies have reported increases up to 15 mM in plasma blood glucose13,30 for a maximum D-Glc dose of 25 g, whereas our study used 35 g. If this increase proves to be reproducible in blood, it could allow the retrieval of an arterial input function for deconvolving the tumor tissue curve, resulting in a dynamic time curve that could provide information about glucose transport and metabolism, while also removing subject variability due to insulin response. However, this is beyond the goal of this technical feasibility study.
The experimental data showed that several tumor tissues had a larger increase than GM and WM (Figures 2–6), confirming the simulations. In all four tumor cases, BBB leakage in the tumor was visible in images or when comparing uncorrected with corrected in areas coinciding spatially with part of the DS-DGE enhancement. However, the DS-DGE enhancement area was generally larger and more pronounced in intensity, possibly reflecting the increased sensitivity to BBB disruption of the small glucose molecule and detectability at the lower pH in tumor EES. and , to some degree, also showed overlap with DS-DGE enhancements, with all three parameters showing largest enhancement in the metastases. The enhancement in tumors could differ temporally, which may also reflect different amounts of BBB disruption (e.g., Figure 6). A rapid uptake would result in similar enhancement for both time periods (without or with post-infusion), while a slow uptake may show a difference.
Additional patient studies are needed to draw more substantial conclusions, and we intend to explore this in greater detail in a future study. However, these first results are highly encouraging, showing the feasibility for DS-DGE to outline tumor tissue with BBB disruption even before Gd-enhancement can detect it. Another important aspect is that the large effect in malignant tumor, combined with a small effect in normal tissue, constitutes an advantage over PET studies of brain malignancies, where phosphorylated 19F-deoxyglucose signal builds up in both brain tumors and healthy GM.62
There are several practical considerations that need to be mentioned. First, similar to other DGE studies, our scan time was long, which increases the risk of motion, which may introduce tissue mixing, resulting in hypointensities and hyperintensities in the DS-DGE maps.36,37 In addition, motion can shift the voxel to another position, resulting in erroneous curves. As motion correction can reduce these errors,35 we applied this to our dynamic data. However, care must be taken since interpolation errors can remove or reduce true effects. Furthermore, while motion correction was applied to the dynamic images, the individual frequency offset images were not corrected for motion, although these can also be affected by motion. Such a frequency-specific motion correction can be more challenging because of the intensity differences between the offsets in the dynamic Z-spectrum, especially close to the direct water saturation.63 Similarly, partial volume effects due to motion can also induce intensity changes at individual frequencies. On the other hand, the Lorentzian fitting over a large number of points is performed using a pre-determined line shape, which may reduce small motion effects on the measured . The risk of motion can be reduced by shortening the experimental duration. However, this can introduce drawbacks such as a less robust baseline measurement and/or an incomplete glucose uptake measurement. Partial volume effects due to physiological effects, such as ventricular swelling, can also introduce signal changes.33 Assuming CSF mixes with WM at the tissue boundary, this would result in an reduction in WM and an increase in CSF. Second, this study used a D-Glc dose of 35 g (0.5 g/kg, maximum of 35 g), while previous studies have used a maximum dose of 25 g, leading to smaller effects. For example, the first DGE MRI study at 3 T by Xu et al.32 gave an effect size of approximately 1.5% in glioblastoma. In a larger patient cohort, Mo et al.64 found effect sizes of 0.5% to 1.5% in low (LGG) and high grade glioma (HGG) enhancement areas using a dose of 25 g. A smaller maximum dose (20 g of D-Glc in 100 mL) was used by Bender et al.29 in LGG and HGG patients resulting in effect sizes of maximum 0.25% in the contrast enhanced tumor area. For metastases, a recent DGE MRI study by Wu et al.35 found up to a 10% increase after D-Glc infusion, whereas in our study, the metastases showed an increase of more than 20%. However, it is difficult to make a quantitative comparison between these studies since they differ in saturation parameters, D-Glc concentration, and contrast origin, that is, CEST including relaxation versus change based on relaxation. In addition, these differences can be caused by individual variations of the tumor structure and properties, warranting a larger population study. Third, since we are acquiring the Z-spectra during a steady state, an additional signal decrease will occur as a function of time during the Z-spectrum acquisition. Although the frequency offsets are acquired in alternating fashion around the water frequency, this continuous signal decrease will still cause the DS to deviate from a true Lorentzian line shape. This may especially affect the narrower line shapes where fitting deviations may occur at the earlier smaller intensity drops in the Z-spectrum (Figures 1 and 3). Fourth, conflicting results have been reported regarding hyperglycemia’s effect on perfusion, with studies showing minimal regional changes65 or no effect.66,67 For our DCE and DSC measurements, we assumed the latter.
5 |. CONCLUSION
We developed DS-DGE MRI to assess D-Glc uptake in brain tumors. Contrary to glucoCESL- and single-frequency glucoCEST-based DGE MRI, this approach is inherently independent of shifts occurring between the dynamics. Whole-brain dynamic Z-spectral images were acquired in less than 40 s, allowing imaging of D-Glc uptake curves in multiple tissues. Early patient data look highly promising, with DS-DGE highlighting lesion areas with information similar and/or complementary to Gd-based perfusion-weighted imaging. DS-DGE MRI can therefore further bridge the gap between research and clinical implementation of using D-Glc as a biodegradable contrast agent.
Supplementary Material
SUPPORTING INFORMATION
Additional supporting information may be found in the online version of the article at the publisher’s website.
TABLE S1. Exclusion criteria for the study.
ACKNOWLEDGMENTS
We are grateful to Terri Brawner, Kathleen Kahl, Ivana Kusevic, and Maia Lee for their assistance with the experiments. This project was supported by National Institutes of Health (RO1 EB034978 and S10OD021648), the Swedish Research Council (2019-03637), and the Swedish Cancer Society (21 1652 Pj).
Funding information
National Institute of Biomedical Imaging and Bioengineering, Grant/Award Numbers: RO1 EB034978, S10OD021648; Vetenskapsrådet, Grant/Award Number: 2019-03637; Cancerfonden, Grant/Award Number: 211652 Pj
Funding Statement
National Institute of Biomedical Imaging and Bioengineering, Grant/Award Numbers: RO1 EB034978, S10OD021648; Vetenskapsrådet, Grant/Award Number: 2019-03637; Cancerfonden, Grant/Award Number: 21 1652 Pj
Footnotes
CONFLICT OF INTEREST STATEMENT
Under a license agreement between Philips and the Johns Hopkins University, L.K.’s spouse, P.C.M.v.Z., and the University are entitled to fees related to an imaging device used in the study discussed in this publication. P.C.M.v.Z. is also a paid lecturer for Philips. This arrangement has been reviewed and approved by the Johns Hopkins University in accordance with its conflict of interest policies.
DATA AVAILABILITY STATEMENT
The data that support the findings of this study are available from the corresponding author on reasonable request.
REFERENCES
- 1.Mathur M, Jones JR, Weinreb JC. Gadolinium deposition and nephrogenic systemic fibrosis: a Radiologist’s primer. Radiographics. 2020;40:153–162. [DOI] [PubMed] [Google Scholar]
- 2.Woolen SA, Shankar PR, Gagnier JJ, MacEachern MP, Singer L, Davenport MS. Risk of nephrogenic systemic fibrosis in patients with stage 4 or 5 chronic kidney disease receiving a group II gadolinium-based contrast agent: a systematic review and meta-analysis. JAMA Intern Med. 2020;180:223–230. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.Gulani V, Calamante F, Shellock FG, Kanal E, Reeder SB. International Society for Magnetic Resonance in M. Gadolinium deposition in the brain: summary of evidence and recommendations. Lancet Neurol. 2017;16:564–570. [DOI] [PubMed] [Google Scholar]
- 4.McDonald RJ, McDonald JS, Kallmes DF, et al. Gadolinium deposition in human brain tissues after contrast-enhanced MR imaging in adult patients without intracranial abnormalities. Radiology. 2017;285:546–554. [DOI] [PubMed] [Google Scholar]
- 5.Lord ML, Chettle DR, Grafe JL, Noseworthy MD, McNeill FE. Observed deposition of gadolinium in bone using a new noninvasive in vivo biomedical device: results of a small pilot feasibility study. Radiology. 2018;287:96–103. [DOI] [PubMed] [Google Scholar]
- 6.U.S. Food and Drug Administration. FDA-approved labeling for healthcare professionals: Boxed warning section. FDA document no. 105513. 2017. www.fda.gov/media/105513/download (accessed: September 23, 2024).
- 7.Eichberg DG, Di L, Morell AA, et al. Incidence of high grade gliomas presenting as radiographically non-enhancing lesions: experience in 111 surgically treated non-enhancing gliomas with tissue diagnosis. J Neuro-Oncol. 2020;147:671–679. [DOI] [PubMed] [Google Scholar]
- 8.Marta GN, da Cunha Colombo Bonadio RR, Martins RE, Zuppani HB, de Castro GJ. Cystic brain metastases in ALK-rearranged non-small cell lung cancer. Ecancermedicalscience. 2018;12:818. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.Falk Delgado A, Van Westen D, Nilsson M, et al. Diagnostic value of alternative techniques to gadolinium-based contrast agents in MR neuroimaging-a comprehensive overview. Insights Imaging. 2019;10:84. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Ward KM, Aletras AH, Balaban RS. A new class of contrast agents for MRI based on proton chemical exchange dependent saturation transfer (CEST). J Magn Reson. 2000; 143:79–87. [DOI] [PubMed] [Google Scholar]
- 11.Chan KW, McMahon MT, Kato Y, et al. Natural D-glucose as a biodegradable MRI contrast agent for detecting cancer. Magn Reson Med. 2012;68:1764–1773. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.Walker-Samuel S, Ramasawmy R, Torrealdea F, et al. In vivo imaging of glucose uptake and metabolism in tumors. Nat Med. 2013;19:1067–1072. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13.Xu X, Yadav NN, Knutsson L, et al. Dynamic glucose-enhanced (DGE) MRI: translation to human scanning and first results in glioma patients. Tomography. 2015;1:105–114. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14.Xu X, Chan KW, Knutsson L, et al. Dynamic glucose enhanced (DGE) MRI for combined imaging of blood-brain barrier break down and increased blood volume in brain cancer. Magn Reson Med. 2015;74:1556–1563. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15.Rivlin M, Navon G. Glucosamine and N-acetyl glucosamine as new CEST MRI agents for molecular imaging of tumors. Sci Rep. 2016;6:32648. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16.Knutsson L, Xu J, Ahlgren A, van Zijl PCM. CEST, ASL, and magnetization transfer contrast: how similar pulse sequences detect different phenomena. Magn Reson Med. 2018;80:1320–1340. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17.Knutsson L, Xu X, van Zijl PCM, Chan KWY. Imaging of sugar-based contrast agents using their hydroxyl proton exchange properties. NMR Biomed. 2022;36:e4784. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18.Jin T, Autio J, Obata T, Kim SG. Spin-locking versus chemical exchange saturation transfer MRI for investigating chemical exchange process between water and labile metabolite protons. Magn Reson Med. 2011;65:1448–1460. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19.Cobb JG, Xie J, Li K, Gochberg DF, Gore JC. Exchange-mediated contrast agents for spin-lock imaging. Magn Reson Med. 2012;67:1427–1433. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20.Yuan J, Zhou J, Ahuja AT, Wang YX. MR chemical exchange imaging with spin-lock technique (CESL): a theoretical analysis of the Z-spectrum using a two-pool R(1rho) relaxation model beyond the fast-exchange limit. Phys Med Biol. 2012;57:8185–8200. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21.Zaiss M, Bachert P. Exchange-dependent relaxation in the rotating frame for slow and intermediate exchange -- modeling off-resonant spin-lock and chemical exchange saturation transfer. NMR Biomed. 2013;26:507–518. [DOI] [PubMed] [Google Scholar]
- 22.Jin T, Mehrens H, Hendrich KS, Kim SG. Mapping brain glucose uptake with chemical exchange-sensitive spin-lock magnetic resonance imaging. J Cereb Blood Flow Metab. 2014;34:1402–1410. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 23.Schuenke P, Paech D, Koehler C, et al. Fast and quantitative T1rho-weighted dynamic glucose enhanced MRI. Sci Rep. 2017;7:42093. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 24.Schuenke P, Koehler C, Korzowski A, et al. Adiabatically prepared spin-lock approach for T1rho-based dynamic glucose enhanced MRI at ultrahigh fields. Magn Reson Med. 2017;78:215–225. [DOI] [PubMed] [Google Scholar]
- 25.Paech D, Schuenke P, Koehler C, et al. T1rho-weighted dynamic glucose-enhanced MR imaging in the human brain. Radiology. 2017;285:914–922. [DOI] [PubMed] [Google Scholar]
- 26.Jin T, Iordanova B, Hitchens TK, et al. Chemical exchange-sensitive spin-lock (CESL) MRI of glucose and analogs in brain tumors. Magn Reson Med. 2018; 80:488–495. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 27.Herz K, Lindig T, Deshmane A, et al. T1rho-based dynamic glucose-enhanced (DGErho) MRI at 3 T: method development and early clinical experience in the human brain. Magn Reson Med. 2019;82:1832–1847. [DOI] [PubMed] [Google Scholar]
- 28.Boyd PS, Breitling J, Zimmermann F, et al. Dynamic glucose-enhanced (DGE) MRI in the human brain at 7 T with reduced motion-induced artifacts based on quantitative R1rho mapping. Magn Reson Med. 2020;84:182–191. [DOI] [PubMed] [Google Scholar]
- 29.Bender B, Herz K, Deshmane A, et al. GLINT: GlucoCEST in neoplastic tumors at 3 T-clinical results of GlucoCEST in gliomas. MAGMA. 2022;35:77–85. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 30.Knutsson L, Seidemo A, Rydhog Scherman A, et al. Arterial input functions and tissue response curves in dynamic glucose-enhanced (DGE) imaging: comparison between glucoCEST and blood glucose sampling in humans. Tomography. 2018;4:164–171. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 31.Xu X, Xu J, Knutsson L, et al. The effect of the mTOR inhibitor rapamycin on glucoCEST signal in a preclinical model of glioblastoma. Magn Reson Med. 2019;81:3798–3807. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 32.Xu X, Sehgal AA, Yadav NN, et al. D-glucose weighted chemical exchange saturation transfer (glucoCEST)-based dynamic glucose enhanced (DGE) MRI at 3T: early experience in healthy volunteers and brain tumor patients. Magn Reson Med. 2020;84:247–262. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 33.Seidemo A, Lehmann PM, Rydhog A, et al. Towards robust glucose chemical exchange saturation transfer imaging in humans at 3 T: arterial input function measurements and the effects of infusion time. NMR Biomed. 2022;35:e4624. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 34.Seidemo A, Wirestam R, Helms G, et al. Tissue response curve-shape analysis of dynamic glucose-enhanced and dynamic contrast-enhanced magnetic resonance imaging in patients with brain tumor. NMR Biomed. 2023;36:e4863. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 35.Wu Y, Derks S, Wood TC, et al. Improved postprocessing of dynamic glucose-enhanced CEST MRI for imaging brain metastases at 3 T. Eur Radiol Exp. 2023;7:78. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 36.Zaiss M, Herz K, Deshmane A, et al. Possible artifacts in dynamic CEST MRI due to motion and field alterations. J Magn Reson. 2019;298:16–22. [DOI] [PubMed] [Google Scholar]
- 37.Lehmann PM, Seidemo A, Andersen M, et al. A numerical human brain phantom for dynamic glucose-enhanced (DGE) MRI: on the influence of head motion at 3T. Magn Reson Med. 2023;89:1871–1887. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 38.Gore JC, Brown MS, Mizumoto CT, Armitage IM. Influence of glycogen on water proton relaxation times. Magn Reson Med. 1986;3:463–466. [DOI] [PubMed] [Google Scholar]
- 39.Yadav NN, Xu J, Bar-Shir A, et al. Natural D-glucose as a biodegradable MRI relaxation agent. Magn Reson Med. 2014;72:823–828. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 40.Kim M, Gillen J, Landman BA, Zhou J, van Zijl PC. Water saturation shift referencing (WASSR) for chemical exchange saturation transfer (CEST) experiments. Magn Reson Med. 2009;61:1441–1450. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 41.Zhou Y, Bie C, van Zijl PCM, Yadav NN. The relayed nuclear Overhauser effect in magnetization transfer and chemical exchange saturation transfer MRI. NMR Biomed. 2023;36:e4778. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 42.Liu G, Qin Q, Chan KW, et al. Non-invasive temperature mapping using temperature-responsive water saturation shift referencing (T-WASSR) MRI. NMR Biomed. 2014;27:320–331. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 43.Mohammed Ali S, Yadav NN, Wirestam R, et al. Deep learning-based Lorentzian fitting of water saturation shift referencing spectra in MRI. Magn Reson Med. 2023;90:1610–1624. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 44.Herz K, Mueller S, Perlman O, et al. Pulseq-CEST: towards multi-site multi-vendor compatibility and reproducibility of CEST experiments using an open-source sequence standard. Magn Reson Med. 2021;86:1845–1858. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 45.Seidemo A, Knutsson L, Yadav NN, Wirestam R, Sundgren PC, van Zijl PCM. A new kinetic model for characterization of glucose delivery, transport and metabolism for MRS and dynamic glucose enhanced (DGE) MRI of the brain. Proceedings of the International Society of Magnetic Resonance in Medicine. Singapore; 2024:729. [Google Scholar]
- 46.Li W, Grgac K, Huang A, Yadav N, Qin Q, van Zijl PC. Quantitative theory for the longitudinal relaxation time of blood water. Magn Reson Med. 2016;76:270–281. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 47.Li W, van Zijl PCM. Quantitative theory for the transverse relaxation time of blood water. NMR Biomed. 2020;33:e4207 Erratum: Quantitative theory for the transverse relaxation time of blood water. NMR Biomed. 2020;33(8):e4329. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 48.Zaiss M, Anemone A, Goerke S, et al. Quantification of hydroxyl exchange of D-glucose at physiological conditions for optimization of glucoCEST MRI at 3, 7 and 9.4 tesla. NMR Biomed. 2019;32:e4113. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 49.Lu H, Golay X, Pekar JJ, Van Zijl PC. Sustained poststimulus elevation in cerebral oxygen utilization after vascular recovery. J Cereb Blood Flow Metab. 2004;24:764–770. [DOI] [PubMed] [Google Scholar]
- 50.Klein S, Staring M, Murphy K, Viergever MA, Pluim JP. Elastix: a toolbox for intensity-based medical image registration. IEEE Trans Med Imaging. 2010;29:196–205. [DOI] [PubMed] [Google Scholar]
- 51.Tofts PS. Modeling tracer kinetics in dynamic Gd-DTPA MR imaging. J Magn Reson Imaging. 1997;7:91–101. [DOI] [PubMed] [Google Scholar]
- 52.Knutsson L, Stahlberg F, Wirestam R. Absolute quantification of perfusion using dynamic susceptibility contrast MRI: pitfalls and possibilities. MAGMA. 2010;23:1–21. [DOI] [PubMed] [Google Scholar]
- 53.Knutsson L, Lindgren E, Ahlgren A, et al. Dynamic susceptibility contrast MRI with a prebolus contrast agent administration design for improved absolute quantification of perfusion. Magn Reson Med. 2014;72:996–1006. [DOI] [PubMed] [Google Scholar]
- 54.Mulkern RV, Williams ML. The general solution to the Bloch equation with constant rf and relaxation terms: application to saturation and slice selection. Med Phys. 1993;20:5–13. [DOI] [PubMed] [Google Scholar]
- 55.Smith SA, Bulte JW, van Zijl PC. Direct saturation MRI: theory and application to imaging brain iron. Magn Reson Med. 2009;62:384–393. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 56.Lim IA, Li X, Jones CK, Farrell JA, Vikram DS, van Zijl PC. Quantitative magnetic susceptibility mapping without phase unwrapping using WASSR. NeuroImage. 2014;86:265–279. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 57.Neely RD, Kiwanuka JB, Hadden DR. Influence of sample type on the interpretation of the oral glucose tolerance test for gestational diabetes mellitus. Diabet Med. 1991;8:129–134. [DOI] [PubMed] [Google Scholar]
- 58.Farrer M, Albers CJ, Neil HA, Adams PC, Laker MF, Alberti KG. Assessing the impact of blood sample type on the estimated prevalence of impaired glucose tolerance and diabetes mellitus in epidemiological surveys. Diabet Med. 1995;12:325–329. [DOI] [PubMed] [Google Scholar]
- 59.Carstensen B, Lindstrom J, Sundvall J, Borch-Johnsen K, Tuomilehto J, Group DPSS. Measurement of blood glucose: comparison between different types of specimens. Ann Clin Biochem. 2008;45:140–148. [DOI] [PubMed] [Google Scholar]
- 60.Schure JR, Weinmuller S, Kamm L, Herz K, Zaiss M. Sidebands in CEST MR-how to recognize and avoid them. Magn Reson Med. 2024;91:2391–2402. [DOI] [PubMed] [Google Scholar]
- 61.Choi WM, Chung J, Jin T, Kim SG. Effect of osmolality on dynamic glucose enhanced (DGE) MRI. Proceedings of the International Society of Magnetic Resonance in Medicine. Honolulu; 2017:194. [Google Scholar]
- 62.Chen W. Clinical applications of PET in brain tumors. J Nucl Med. 2007;48:1468–1481. [DOI] [PubMed] [Google Scholar]
- 63.Simegn GL, Sun PZ, Zhou J, et al. Motion and magnetic field inhomogeneity correction techniques for chemical exchange saturation transfer (CEST) MRI: a contemporary review. NMR Biomed. 2024;38:e5294. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 64.Mo J, Xu X, Ma A, et al. Dynamic glucose enhanced MRI of gliomas: a preliminary clinical application. NMR Biomed. 2025;38:e5265. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 65.Kochunov P, Wey HY, Fox PT, et al. Changes in cerebral blood flow during an alteration in glycemic state in a large non-human primate (Papio hamadryas sp.). Front Neurosci. 2017;11:49. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 66.Xu F, Liu P, Pascual JM, Xiao G, Huang H, Lu H. Acute effect of glucose on cerebral blood flow, blood oxygenation, and oxidative metabolism. Hum Brain Mapp. 2015;36:707–716. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 67.Hasselbalch SG, Knudsen GM, Capaldo B, Postiglione A, Paulson OB. Blood-brain barrier transport and brain metabolism of glucose during acute hyperglycemia in humans. J Clin Endocrinol Metab. 2001;86:1986–1990. [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Data Availability Statement
The data that support the findings of this study are available from the corresponding author on reasonable request.