Abstract
Introduction
Photoplethysmography (PPG) measures are important in monitoring peripheral oxygen saturation (SpO2). Another parameter is a derived quantity referred to as the peripheral perfusion index (PPI). It is calculated as the ratio of the peak-to-peak pulse amplitude of a PPG signal (PAPPG) to the non-pulsating part of the total PPG signal. The PPI has been used as a marker of blood perfusion states in a variety of clinical settings but has not been systematically and directly compared to measures of local blood perfusion. This study’s purpose was to investigate this issue to provide initial data on the relationship between finger skin blood perfusion, measured by laser Doppler blood perfusion flux (LDF) and PAPPG.
Methods
Ten subjects (five male), recruited from medical students with an average age of 26 years, participated. While supine for 30 minutes, skin blood perfusion was recorded using laser Doppler flux (LDF) on the ring finger pulp of the non-dominant hand, and the photoplethysmography pulse (PPG) was recorded from the index finger of the same hand. The recorded data was searched sequentially manually to locate the first 30-pulse sequence in which the PPG amplitude of at least six PPG pulses was less than or equal to 60% of the maximum pulse amplitude in the sequence. The primary PPG parameter of interest was PAPPG. For the LDF signal, the pulse amplitude is designated as PALDF, the total LDF for each pulse is designated as LDFTOT, and the LDF pulsatile component is designated as PF. To investigate the relationship between LDF parameters and PAPPG a linear regression analysis of each 30-pulse sequence was done with PAPPG as the independent variable and each of the three LDF parameters individually (PALDF, LDFTOT, and PF) as dependent variables.
Results
There was a statistically significant direct relationship between PAPPG and all three measures of blood perfusion (p<0.05). Correlation coefficients (R) varied among subjects but within-subject variations versus PAPPG were similar, having mean values that ranged from 0.665 to 0.694. The results also provided evidence in support of a direct relationship between the LDF pulsatility index, defined as the ratio of PF to its mean value., and PAPPG (R=0.779).
Conclusions
When finger PPG pulse amplitudes are measured in individual subjects there is a moderate-to-strong correlation between the PPG pulse amplitude changes and skin blood perfusion changes. This fact impacts the confidence in using the widely available PPG parameter, peripheral perfusion index, as an indicator of changes in tissue perfusion. However, differences in the PPG pulse amplitude among subjects were less reliable indicators of differences in blood perfusion among subjects. The findings also indicate that a related parameter, the LDF pulsatility index, is also highly correlated with the PPG pulse amplitude and may serve as a useful parameter for future clinical investigations.
Keywords: laser doppler, ldf, oxygen saturation, perfusion index, photoplethysmography, ppg, pulsatility index, skin blood perfusion, spo2
Introduction
Photoplethysmographic (PPG) assessments have an important place in monitoring peripheral oxygen saturation (SpO2) in multiple settings and conditions [1,2]. Frequently such assessments are done using a sensor placed on a finger to measure SpO2 and other newly implemented parameters such as blood pressure [3,4] and pulse wave velocity [5,6]. An additional parameter now available in most devices is a derived quantity referred to as the peripheral perfusion index (PPI). It is calculated as the ratio of the peak-to-peak pulse amplitude of a PPG signal (PAPPG) to the non-pulsating part of the total PPG signal [7]. The PPI is not to be confused with the pulsatility index (PLI) which is a parameter expressed as the ratio of pulsatile blood velocity to the corresponding mean or a reference value and used as a parameter to characterize the extent of peripheral arterial disease [8,9] and also used for a range of other conditions related to heart, pulmonary and cerebral features [10-12].
Contrastingly, initial work on the PPI was targeted to investigate its use it as an indicator of tissue perfusion using its value compared to that of the core-to-toe skin temperature difference (DT) in a group of 37 patients [13]. A DT ≥ 7oC was arbitrarily defined as an abnormal overall perfusion. According to this criterion, measurements in the same patients at times when they had normal vs. abnormal perfusion were reported. Based on the generally lower PPI value these researchers found during the abnormal perfusion times, it was suggested that PPI might be a good indicator of tissue perfusion. Because low perfusion states may impact the accuracy and precision of SpO2 measurements [14,15], a more recent study used a PPI value of ≤1% as a criterion to indicate low perfusion in persons with Fitzpatrick classes from I through VI [16]. Based on measurements using two different finger PPG devices, these researchers concluded that both PPI and darker skin were associated with clinically significant high-reading errors. The possibility of using PPI as a detector of, or discriminator for, low perfusion states has also been evaluated for a specific device [17]. In this study a PPI value of 1% was used as an arbitrary cutoff value and SpO2 values were reported accurate for both normal and low perfusion values independent of skin tone.
Although the PPI value may be construed to be at least in part an indicator of tissue blood perfusion an important unanswered question is the extent to which PAPPG, the main variable in the PPI value [7], is related to actual pulsatile tissue blood perfusion. The purpose of the present pilot study was to investigate this issue to provide initial data on the extent of the relationship between finger skin blood perfusion, measured by laser Doppler blood perfusion flux (LDF) and PAPPG.
Materials and methods
Subjects
Ten volunteer subjects (five male), recruited from 1st and 2nd-year medical students, participated in this research after signing a University Institutional Review Board (IRB) approved consent form (#2019-598-NSU). Entry requirements were (1) the ability and agreement to lie supine without significant movements for up to 45 minutes, (2) no history of cardiovascular or neurological conditions, (3) nonsmoker, and (4) willingness to forgo caffeinated beverages on the day of the experiment. Persons with diabetes mellitus were excluded from participation.
Measurements
Skin blood perfusion was measured using laser Doppler flux (LDF) measurements with a multifiber laser integrating probe (Moor DP7A, Moor Instruments, Wilmington DE, USA) attached to the ring finger pulp of the non-dominant hand as shown in Figure 1. The probe was connected to a laser Doppler measuring system (Laserflo Blood Perfusion Monitor, Moor Instruments, Model BPM2). The recorded blood perfusion or flux is the product of the red blood cell (RBC) concentration and RBC velocity. The measurement uses a low-intensity laser light signal that is transmitted into the skin to a depth of about 1-2 mm [18]. The Doppler-shifted return signal contains information about the speed and number density of moving RBCs, which is processed to yield a parameter, RBC perfusion or RBC flux, that is proportional to blood flow. The system is calibrated using a motility standard provided by the company. LDF is expressed as relative units since it cannot be directly expressed in blood flow units and is widely used to measure skin blood perfusion [19-22]. The PPG pulse was recorded from the index finger of the same hand using a matched infrared emitter and photodiode sensor (TSD200, Biopac Systems, Goleta CA USA) that was gently but securely secured to the finger with Velcro as also illustrated in Figure 1. The system works with an infrared excitation of 860 ± 60 nm with a detected wavelength of 800 nm. Blood pulses in the finger tissue modulate the infrared causing changes in the resistance of the sensor producing a time-varying output voltage reflecting these pulsations. The PPG sensor output was coupled to a PPG amplifier (PPG100C, Biopac Systems, Goleta CA USA) that was set to a gain of 50, a low pass filter of 10 Hz, and a high pass filter of 0.05 Hz. The outputs of the PPG amplifier and the laser Doppler system were coupled to an analog/digital conversion device (DataQ Instruments, Akron OH, USA, model DI-720), and each channel was sampled at 1000 samples per second using Windaq recording and playback software (DataQ Instruments, Akron OH, USA) and the two output channels displayed and recorded on a laptop computer. The channel gains were both set to four for all experiments.
Figure 1. Measurement setup.
Laser Doppler flux (LDF) is measured on the pulp of the ring finger using a multifiber LDF probe. The photoplethysmographic (PPG) signal is obtained from the index finger with the PPG sensor secured to the finger firmly but gently with Velcro.
Measurement procedures
With the subject supine on a padded examination table, the laser Doppler probe and the PPG sensor were placed on the target fingers as illustrated in Figure 1. Afterward, the room lights were dimmed, and the two channels were observed for amplitude suitability at the set gain. The gain remained at a value of four for all subjects. After the subject had been lying supine for 10 minutes, the recording was started, and data was acquired for 30 continuous minutes. At the end of the recording interval, the subject’s blood pressure was taken and recorded while they were still supine.
Data extraction
The recorded data was searched sequentially manually to locate the first 30-pulse sequence in which the PPG amplitude of at least six PPG pulses was less than or equal to 60% of the maximum pulse amplitude in the sequence. Two such sequences are illustrated in Figure 2. In one of the data sets such a sequence was not observed until 1496 seconds (~ 25 minutes) from the start of the recording. In the other example data set a sequence satisfying the search criterion was located after only 4.5 seconds from the start of the recording. The reason for the 30-pulse criterion was to avoid potential temporal confounding factors of longer intervals while still including enough pulses to provide a reasonable regression estimate. The criterion of at least six pulses with pulse amplitudes ≤ the maximum pulse amplitude in that interval was to have a sufficient range of pulse amplitudes for a meaningful regression analysis.
Figure 2. Example measurement sequences.
LDF is the laser Doppler flux measured on the pulp of the ring finger. PPG is the photoplethysmography signal from the index finger. Both are shown as a function of time. In A, the 30-pulse sequence that was analyzed was captured 1496 seconds into the experiment. In B. the 30-pulse sequence that was analyzed was captured 4.5 seconds into the experiment. In both cases there is a spontaneous waxing and waning of the PPG signal with associated changes in the pattern of the LDF signal.
Data parameters
The primary PPG parameter is the PPG pulse amplitude which is the peak-to-peak value illustrated in Figure 3 and as previously described is denoted as PAPPG. For the LDF signal, also illustrated in Figure 3, the pulse amplitude is designated as PALDF. The total LDF associated with each pulse is represented by the total area under the curve and is designated as LDFTOT. The pulsatile component of the LDF is designated as PF.
Figure 3. Parameter definitions.
LDF is the laser Doppler flux measured on the pulp of the ring finger. PPG is the photoplethysmography signal recorded from the index finger of the same hand. Both are shown as a function of time. PALDF is the peak-to-peak amplitude of the LDF and PAPPG is the peak-to-peak amplitude of the PPG pulse. TLDF is the duration of the LDF pulse. The area with the diagonal lines in the LDF signal is the total LDF per pulse (LDFTOT). The area enclosed by the pulsatile part and labeled PF is the pulsatile flux component.
Analysis
To investigate the relationship between LDF and PAPPG a linear regression analysis of each 30-pulse sequence was done with PAPPG as the independent variable and each of the three LDF parameters individually (PALDF, LDFTOT, and PF) as dependent variables. For each regression parameter, the best and the least good correlation was determined based on the least and greatest correlation coefficient (R), and the average correlation coefficient among all 10 subjects was determined. The overall relationship of perfusion parameters to PAPPG was determined via regression analysis based on the combined data of the 300 measured pulses. The possible relationship between the LDF pulsatility index (PLI), defined as the ratio of PALDF to its mean value (MEANLDF), was determined by regression analysis of each subject’s average PLI with each subject’s average PAPPG. P-values < 0.05 were taken as statistically significant.
Results
Subject demographics
Table 1 lists the key parameters for each of the 10 subjects. Based on body mass index (BMI) values, four of the male subjects would be classified as slightly overweight (BMI > 25.0 Kg/m2) with all others classified as normal weight. The blood pressure of four of the subjects would be classified as elevated based on the current American Heart Association standards due to their systolic pressures being not less than 120 mmHg. However, in the present study subject blood pressures were measured during supine lying, not in a seated position as is the standard. Diastolic pressures of all subjects were within the normal range. For the entire group, average values ± SD for age, BMI, systolic BP, diastolic BP, and heart rate were respectively 26.0 ± 1.3 years, 23.2 ± 2.7 Kg/m2, 117.6 ± 7.7 mmHg, 69.6 ± 6.1 mmHg, and 65.6 ± 10.8 bpm.
Table 1. . Participant demographics.
Based on body mass index (BMI) values, four of the male subjects (5, 6, 8 and 9) would be classified as slightly overweight (BMI > 25.0). Blood pressure (BP) and heart rate were measured during supine lying. According to current American Heart Association standards, four subjects would be classified as having elevated BP (subjects 2, 6, 9, and 10).
| Subject | Sex | Age (years) | Height (m) | Weight (Kg) | BMI (Kg/m2) | Systolic BP (mmHg) | Diastolic BP (mmHg) | Heart Rate (bpm) |
| 1 | M | 28 | 1.73 | 69.2 | 23.6 | 116 | 65 | 69 |
| 2 | F | 26 | 1.65 | 61.6 | 23 | 128 | 77 | 56 |
| 3 | M | 26 | 1.63 | 64.7 | 24.9 | 104 | 60 | 69 |
| 4 | F | 24 | 1.65 | 53.6 | 20.1 | 107 | 68 | 78 |
| 5 | M | 26 | 1.73 | 75.9 | 25.8 | 119 | 63 | 55 |
| 6 | M | 26 | 1.70 | 71.4 | 25.1 | 120 | 69 | 54 |
| 7 | F | 26 | 1.73 | 58.0 | 19.8 | 116 | 72 | 78 |
| 8 | F | 28 | 1.78 | 80.4 | 25.8 | 117 | 76 | 51 |
| 9 | M | 26 | 1.80 | 80.4 | 25.1 | 121 | 68 | 68 |
| 10 | F | 24 | 1.60 | 47.3 | 18.8 | 128 | 78 | 79 |
LDF vs. PPG pulse amplitudes
Figure 4 depicts the relationship between PALDF and PAPPG for the data representing the best correlation coefficient (R) case (R = 0.937) and the least good correlation case (R = 0.468) among the 10 subjects listed in Table 1. The best correlation was for subject 10 (p < 0.00001) and the least good, but still statistically significant, for subject 9 (p=0.01). The overall average value for R across the 10 subjects (mean ± SD) was 0.694 ± 0.156.
Figure 4. Best and worst correlations for LDF pulse amplitude .
The pulse amplitude of the laser Doppler flux (PALDF) is shown versus the pulse amplitude of the PPG (PAPPG) along with the linear regression line (dotted) and corresponding regression equation. Part A is the best correlation among the 10 subjects (R = 0.937) and part B shows the least good correlation among the 10 subjects (R = 0.468). Data is for the sequence of 30 pulses, and R is the correlation coefficient.
Total LDF pulsatile perfusion vs. PPG pulse amplitude
Figure 5 depicts the relationship between total LDF (LDFTOT) and PAPPG for the data representing the same two subjects shown in Figure 4. The correlation coefficient (R = 0.948) for subject 10 was the best among all 10 subjects (p < 0.00001). The correlation for subject 9 was less good (R = 0.765) but still highly statistically significant (p<0.0001). R-values across the 10 subjects for this relationship ranged from 0.317 for subject 5 (p = 0.039) to 0.948 (p< 0.0001) for subject 10. The overall average R across the 10 subjects (mean ± SD) was 0.695 ± 0.226.
Figure 5. Best and worst correlations for total LDF.
The total LDF per pulse (LDFTOT) is shown versus the pulse amplitude of the PPG (PAPPG) along with the linear regression line (dotted) and corresponding regression equation for the same two subjects (A and B) as shown in Figure 4. Data is for the sequence of 30 pulses, and R is the correlation coefficient.
LDF pulsatile perfusion component vs. PPG pulse amplitude
Figure 6 depicts the relationship between the pulsatile component of LDF (PF) and PAPPG for the same two subjects as shown in Figures 4, 5. The correlation coefficient (R = 0.890) for subject 10 was the best among all 10 subjects (p < 0.00001). The correlation for subject 9 was less good (R = 0.697) but still statistically significant (p = 0.0003). R-values across the 10 subjects for this relationship ranged from 0.257 (p =0.112) for subject 5 to 0.889 ( p < 0.0001) for subject 10. The overall average value for R across the 10 subjects (mean ± SD) was 0.665 ± 0.183.
Figure 6. Best and worst correlations for pulsatile component of LDF.
The pulsatile flow component (PF) is shown versus the pulse amplitude of the PPG (PAPPG) along with the linear regression line (dotted) and corresponding regression equation for the same two subjects (A and B) as shown in Figure 4. Data is for the sequence of 30 pulses, and R is the correlation coefficient. Subjects A and B are the same as shown in Figure 4.
The overall correlation between LDF and PPG pulse amplitudes
Figure 7 shows the overall composite relationship between PALDF and PAPPG considering all subject data. Each data point represents a single PALDF - PAPPG pair (n =300 pulses). A moderate correlation is observed (R = 0.527) that is statistically significant (p< 0.0001). The 95% confidence limits (lower and upper bounds) on the intercept and slope of the regression equation shown in the figure are for the intercept 0.151 to 0.196 and the slope 0.077 to 0.114.
Figure 7. Correlation among all subjects for LDF pulse amplitude.
The pulse amplitude of the laser Doppler flux (PALDF) is shown versus the pulse amplitude of the PPG (PAPPG) along with the linear regression line (dotted) and corresponding regression equation for all subjects. The p-value for the regression is < 0.0001.
The overall correlation between PF and PPG pulse amplitudes
Figure 8 shows the overall composite relationship between PF and PAPPG considering all subject data. Each data point represents a single PF - PAPPG pair (n =300 pulses). A low correlation is observed (R = 0.321) but is statistically significant (p< 0.0001). The 95% confidence limits (lower and upper bounds) on the intercept and slope of the regression equation shown in the figure are for the intercept 0.084 to 0.112 and the slope, 0.021 to 0.043.
Figure 8. Correlation among all subjects for just the pulsatile LDF component.
The pulsatile flow component (PF) is shown versus the pulse amplitude of the PPG (PAPPG) along with the linear regression line (dotted) and corresponding regression equation for all subjects. The p-value for the regression is < 0.0001.
Overall blood perfusion comparison
Figure 9 shows the average LDF during the 30-pulse sampling interval (LDF30-PULSE) for each of the 10 subjects compared to the overall 30-minute average LDF perfusion (LDF30-MINUTE) for that subject. The high correlation between these two (R = 0.892, p < 0.001) is consistent with the notion that the 30-pulse sample is reflective of the overall perfusion feature. The 95% confidence limits (lower and upper bounds) on the intercept and slope of the regression equation shown in the figure are for the intercept -0.232 to 0.330 and for the slope, 0.515 to 1.242. The 95% confidence limits (lower and upper bounds) on the intercept and slope of the regression equation shown in the figure are for the intercept 0.089 to 0.528 and the slope, 0.059 to 0.453.
Figure 9. Average blood perfusion comparison.
Average LDF during the 30-pulse sampling interval (LDF30-PULSE) in relation to the overall 30-minute average LDF perfusion (LDF30-MINUTE) is shown for each of the 10 subjects. The dotted line is the linear regression with the equation shown in the inset. The p-value of the regression is < 0.001. Error bars are the standard error of the mean (SEM).
LDF pulsatility index
Figure 10 shows the LDF pulsatility index (PLI = PALDF/MEANLDF) for each of the 10 subjects compared to their peak-to-peak PPG pulse amplitude (PAPPG) averaged over each subject’s 30-pulse sampling interval. The data indicates a significant correlation between the pulsatility index and the PPG pulse amplitude (R = 0.779, p = 0.014).
Figure 10. LDF pulsatility index .
The LDF pulsatility index (PLI = PALDF/MEANLDF) is shown for each of the 10 subjects compared to their PAPPG averaged over their 30-pulse sampling interval. The dotted line is the linear regression with the equation shown in the inset. The p-value of the regression = 0.014. Error bars are the standard error of the mean (SEM).
Discussion
The present study findings appear to be the first to provide some direct evidence of a statistically significant direct relationship between the amplitude of the finger pulse PPG signal and measures of skin blood perfusions determined by laser Doppler measurements. The skin perfusion measures evaluated included the LDF pulse amplitude (PALDF), total LDF perfusion (LDFTOT), and the LDF pulsatile component (PF). The correlation coefficients for each parameter varied among subjects ranging from a nonsignificant low of 0.257 in one subject to a high of 0.948 in another. However, the overall correlation coefficients among these three LDF measures across the 10 subjects evaluated were similar and had mean values that ranged from 0.665 to 0.694. Thus, the overall correlations could be considered moderate to strong [23]. In addition, the present results also for the first time provide evidence in support of a direct relationship between the LDF pulsatility index and the amplitude of the PPG finger pulse. This relationship is characterized by a strong correlation (R = 0.779) indicating a positive direct relationship. The strong correlations apply when considering the dependence of the perfusion parameters on the PPG pulse amplitude in individual subjects. However, when considering the correlations among subjects the correlation is less good as exemplified by the regressions shown in Figures 7, 8. One interpretation of these combined findings is that changes in the PPG pulse amplitude reflect the same directional changes in the skin blood perfusion parameters but that differences in the PPG pulse amplitude among subjects are less reliable indicators of differences in blood perfusion among subjects.
In addition to the new insight into the physiological relationships between PPG pulse amplitude signals and the several blood perfusion parameters herein identified; the present findings may have direct clinically related impacts. One relates to the clinical use of the peripheral perfusion index obtained with PPG-determined SpO2 in patients. The PPI is often stated as being mainly dependent on stroke volume and vascular tone [24]. The implication would be that lower values of either would be associated with lower PPI values. This concept is supported by several clinical studies, although until now direct systematic comparisons linking PPG amplitude to blood perfusion are lacking. PPI values obtained from a single device and their variation among fingers in 391 healthy subjects indicated a range between 3.54% to 4.66% with the middle right finger yielding the largest mean value [25]. PPI measurements in 20 healthy volunteers who experienced simulated blood volume reduction via lower body negative pressure demonstrated reduced PPI values in response [26]. Subjects with higher baseline PPI values produced the greatest reduction presumably due to peripheral vasoconstriction that was even greater when combined with pain induced by opposite-hand cold water immersion. The PPI has also been used to estimate its association with postoperative acute kidney injury [27]. The strongest association appeared to be with PPI values less than 1%. Changes in PPI relative to changes in blood pressure (BP), stroke volume (SV), and cardiac output (CO) during anesthesia induction suggested that PPI values that decrease vs. those that increase during induction may discriminate between large and smaller decreases in BP, SV, and CO during the induction process [28]. Moreover, the results of PPI monitoring of 291 high-risk patients indicated that a low PPI value was associated with serious adverse events [29]. Contrastingly an increase in PPI during epidural anesthesia was reported to be an earlier indicator of sympathectomy effectiveness than either skin temperature or blood pressure changes [30]. These reported clinical correlations are consistent with the present findings directly linking blood perfusion to the PPG pulse amplitude which is the main factor impacting PPI.
A related parameter studied was the LDF pulsatility index defined as the ratio of the LDF pulse component to the LDF mean value and its relationship to the PPG pulse amplitude. The broader use of the pulsatility index has had several clinical uses. One is the pulmonary artery pulsatility index (PAPi) which is defined as the ratio of the pulmonary artery pulse pressure to right atrial pressure [31], and has been used as a predictor of right ventricular failure [32] and as an indicator of right ventricular function following heart surgery [33]. Another type and use of a pulsatility index is related to cerebral small vessel disease [11,34]. For this application, the pulsatility index is determined by transcranial Doppler ultrasound of the middle cerebral artery, and the pulsatility index is defined as the blood velocity pulse amplitude divided by the mean blood velocity. A large pulse index was reported to be associated with greater small vessel compromise, a greater incidence of cognitive impairment [11], a greater cerebral small vessel disease score [34], an association with small vessel-related incidences of lacunar stroke [35], and a strong correlation to the fractional pressure loss across stenotic cerebral arteries [36]. A pulsatility index of > 1.1 has been used to attempt to predict cerebral microangiopathy in persons free of cerebral symptoms [37]. The interpretation of these data is that an increase in the pulsatility index reflects an increase in small vessel vascular resistance. This is similar to the present LDF pulsatility index interpretation since an increase in arteriolar resistance would be associated with a decrease in mean perfusion and an elevation in the pulsatile component. A variety of other types and uses of the pulsatility index have been investigated including for peripheral arterial disease [38,39], cardiac conditions [40-42], and obstetrics [43,44] to name only some. It is unclear if the LDF pulsatility index will provide as useful an index or benefit as these other uses. However, the fact that it is herein shown to be highly correlated with the PPG pulse amplitude opens the door to further exploratory studies in which this possibility may be investigated.
Study limitations and recommendations
One study limitation is that the data was obtained from 10 subjects. Despite this, the number was sufficient to uncover significant correlations among the evaluated parameters and to provide a basis and baseline for further studies. Another aspect that may be viewed as a limitation is the demographics of the subjects evaluated. They were young healthy adults. Questions remain as to whether similar patterns and relationships would be present in healthy adults who were considerably more mature. Such a comparative study would seem to be warranted. Moreover, the utility of the basic relationships herein observed to clinical usage remains to be determined. However, the central finding, that the PPG pulse amplitude is significantly correlated with actual blood perfusion, paves the way for targeted future investigations in selective patient populations.
Conclusions
The findings of this pilot study demonstrate that when finger PPG pulse amplitudes are measured in individual subjects there is a moderate-to-strong correlation between the PPG pulse amplitude and skin blood perfusion. This fact impacts the confidence in using the widely available PPG parameter, peripheral perfusion index, as an indicator of changes in tissue perfusion. The findings also indicate that a related parameter, the LDF pulsatility index, is also highly correlated with the PPG pulse amplitude and may serve as a useful parameter for future clinical investigations.
Disclosures
Human subjects: Consent was obtained or waived by all participants in this study. Nova Southeastern University IRB issued approval (#2019-598-NSU).
Animal subjects: All authors have confirmed that this study did not involve animal subjects or tissue.
Conflicts of interest: In compliance with the ICMJE uniform disclosure form, all authors declare the following:
Payment/services info: All authors have declared that no financial support was received from any organization for the submitted work.
Financial relationships: All authors have declared that they have no financial relationships at present or within the previous three years with any organizations that might have an interest in the submitted work.
Other relationships: All authors have declared that there are no other relationships or activities that could appear to have influenced the submitted work.
Author Contributions
Concept and design: Harvey N. Mayrovitz
Acquisition, analysis, or interpretation of data: Harvey N. Mayrovitz
Drafting of the manuscript: Harvey N. Mayrovitz
Critical review of the manuscript for important intellectual content: Harvey N. Mayrovitz
References
- 1.Photoplethysmography and its application in clinical physiological measurement. Allen J. Physiol Meas. 2007;28:0–39. doi: 10.1088/0967-3334/28/3/R01. [DOI] [PubMed] [Google Scholar]
- 2.Current progress of photoplethysmography and SPO(2) for health monitoring. Tamura T. Biomed Eng Lett. 2019;9:21–36. doi: 10.1007/s13534-019-00097-w. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.A deep learning method for continuous noninvasive blood pressure monitoring using photoplethysmography. Liang H, He W, Xu Z. Physiol Meas. 2023;44 doi: 10.1088/1361-6579/acd164. [DOI] [PubMed] [Google Scholar]
- 4.Blood pressure estimation and classification using a reference signal-less photoplethysmography signal: a deep learning framework. Pankaj Pankaj, Kumar A, Komaragiri R, Kumar M. Phys Eng Sci Med. 2023;46:1589–1605. doi: 10.1007/s13246-023-01322-8. [DOI] [PubMed] [Google Scholar]
- 5.Estimation of carotid-femoral pulse wave velocity from finger photoplethysmography signal. Gentilin A, Tarperi C, Cevese A, Mattioli AV, Schena F. Physiol Meas. 2022;43 doi: 10.1088/1361-6579/ac7a8e. [DOI] [PubMed] [Google Scholar]
- 6.Estimation of aortic stiffness by finger photoplethysmography using enhanced pulse wave analysis and machine learning. Hellqvist H, Karlsson M, Hoffman J, Kahan T, Spaak J. Front Cardiovasc Med. 2024;11:1350726. doi: 10.3389/fcvm.2024.1350726. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7.Perfusion index: Physical principles, physiological meanings and clinical implications in anaesthesia and critical care. Coutrot M, Dudoignon E, Joachim J, Gayat E, Vallée F, Dépret F. Anaesth Crit Care Pain Med. 2021;40:100964. doi: 10.1016/j.accpm.2021.100964. [DOI] [PubMed] [Google Scholar]
- 8.Influence of systemic arterial blood pressure and nonhemodynamic factors on the brachial artery pulsatility index in mild to moderate essential hypertension. Duprez DA, De Buyzere ML, De Backer T, et al. Am J Cardiol. 1993;71:350–353. doi: 10.1016/0002-9149(93)90807-o. [DOI] [PubMed] [Google Scholar]
- 9.Relationship between Doppler pulsatility index and direct femoral pressure measurements in the diagnosis of aortoiliac occlusive disease. Johnston KW, Kassam M, Cobbold RS. Ultrasound Med Biol. 1983;9:271–281. doi: 10.1016/0301-5629(83)90061-3. [DOI] [PubMed] [Google Scholar]
- 10.Pulmonary artery pulsatility index is superior to right ventricular stroke work index in predicting elevated right atrial pressure. Benjanuwattra J, Phinyo P, Nair N. Acta Cardiol. 2024:1–3. doi: 10.1080/00015385.2024.2396759. [DOI] [PubMed] [Google Scholar]
- 11.Association of pulsatility index with total burden of cerebral small vessel disease and cognitive impairment. Wu H, Xu L, Zheng X, Gu C, Zhou X, Sun Y, Li X. Brain Behav. 2024;14:0. doi: 10.1002/brb3.3526. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.Pulmonary artery pulsatility index and clinical outcomes in patients undergoing transcatheter tricuspid valve interventions. Tanaka T, Sugiura A, Kavsur R, et al. JACC Cardiovasc Interv. 2024;17:952–954. doi: 10.1016/j.jcin.2024.02.020. [DOI] [PubMed] [Google Scholar]
- 13.Use of a peripheral perfusion index derived from the pulse oximetry signal as a noninvasive indicator of perfusion. Lima AP, Beelen P, Bakker J. Crit Care Med. 2002;30:1210–1213. doi: 10.1097/00003246-200206000-00006. [DOI] [PubMed] [Google Scholar]
- 14.A comparison of the performance of 20 pulse oximeters under conditions of poor perfusion. Clayton DG, Webb RK, Ralston AC, Duthie D, Runciman WB. Anaesthesia. 1991;46:3–10. doi: 10.1111/j.1365-2044.1991.tb09303.x. [DOI] [PubMed] [Google Scholar]
- 15.A comparison of fifteen pulse oximeters. Part I: A clinical comparison; Part II: A test of performance under conditions of poor perfusion. Morris RW, Nairn M, Torda TA. Anaesth Intensive Care. 1989;17:62–73. doi: 10.1177/0310057X8901700113. [DOI] [PubMed] [Google Scholar]
- 16.Low perfusion and missed diagnosis of hypoxemia by pulse oximetry in darkly pigmented skin: a prospective study. Gudelunas MK, Lipnick M, Hendrickson C, et al. Anesth Analg. 2024;138:552–561. doi: 10.1213/ANE.0000000000006755. [DOI] [PubMed] [Google Scholar]
- 17.Racial effects on masimo pulse oximetry: impact of low perfusion index. Sharma V, Barker SJ, Sorci R, Park L, Wilson WC. J Clin Monit Comput. 2024;38:347–354. doi: 10.1007/s10877-023-01113-2. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18.Prediction of sampling depth and photon pathlength in laser Doppler flowmetry. Jakobsson A, Nilsson GE. Med Biol Eng Comput. 1993;31:301–307. doi: 10.1007/BF02458050. [DOI] [PubMed] [Google Scholar]
- 19.Effects of local forearm skin heating on skin properties. Mayrovitz HN. Clin Physiol Funct Imaging. 2020;40:369–376. doi: 10.1111/cpf.12653. [DOI] [PubMed] [Google Scholar]
- 20.Neurovascular responses to sequential deep inspirations assessed via laser-Doppler perfusion changes in dorsal finger skin. Mayrovitz HN, Groseclose EE. Clin Physiol Funct Imaging. 2002;22:49–54. doi: 10.1046/j.1475-097x.2002.00404.x. [DOI] [PubMed] [Google Scholar]
- 21.Inspiration-induced vasoconstrictive responses in dominant versus non-dominant hands. Mayrovitz HN, Groseclose EE. Clin Physiol Funct Imaging. 2005;25:69–74. doi: 10.1111/j.1475-097X.2004.00592.x. [DOI] [PubMed] [Google Scholar]
- 22.Gender differences in facial skin blood perfusion during basal and heated conditions determined by laser Doppler flowmetry. Mayrovitz HN, Regan MB. Microvasc Res. 1993;45:211–218. doi: 10.1006/mvre.1993.1019. [DOI] [PubMed] [Google Scholar]
- 23.User's guide to correlation coefficients. Akoglu H. Turk J Emerg Med. 2018;18:91–93. doi: 10.1016/j.tjem.2018.08.001. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 24.Perfusion index for early identification of regional anesthesia effectiveness: a narrative review. Chu T, Xin Y, Zhou S, Xu A. Minerva Anestesiol. 2023;89:671–679. doi: 10.23736/S0375-9393.23.17065-9. [DOI] [PubMed] [Google Scholar]
- 25.A cross-sectional study on the agreement of perfusion indexes measured on different fingers by a portable pulse oximeter in healthy adults. Swain SM, Lata M, Kumar S, Mondal S, Behera JK, Mondal H. Cureus. 2022;14:0. doi: 10.7759/cureus.24853. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 26.Tissue oxygen saturation and finger perfusion index in central hypovolemia: influence of pain. Høiseth LØ, Hisdal J, Hoff IE, Hagen OA, Landsverk SA, Kirkebøen KA. Crit Care Med. 2015;43:747–756. doi: 10.1097/CCM.0000000000000766. [DOI] [PubMed] [Google Scholar]
- 27.Association between peripheral perfusion index and postoperative acute kidney injury in major noncardiac surgery patients receiving continuous vasopressors: a post hoc exploratory analysis of the VEGA-1 trial. Krone S, Bokoch MP, Kothari R, et al. Br J Anaesth. 2024;132:685–694. doi: 10.1016/j.bja.2023.11.054. [DOI] [PubMed] [Google Scholar]
- 28.The peripheral perfusion index discriminates haemodynamic responses to induction of general anaesthesia. Højlund J, Petersen DR, Agerskov M, Foss NB. J Clin Monit Comput. 2023;37:1533–1540. doi: 10.1007/s10877-023-01035-z. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 29.Continuous peripheral perfusion index in patients admitted to hospital wards - an observational study. Rasmussen PS, Aasvang EK, Olsen RM, Haahr-Raunkjaer C, Elvekjaer M, Sørensen HB, Meyhoff CS. Acta Anaesthesiol Scand. 2021;65:257–265. doi: 10.1111/aas.13711. [DOI] [PubMed] [Google Scholar]
- 30.Pulse oximeter perfusion index as an early indicator of sympathectomy after epidural anesthesia. Ginosar Y, Weiniger CF, Meroz Y, et al. Acta Anaesthesiol Scand. 2009;53:1018–1026. doi: 10.1111/j.1399-6576.2009.01968.x. [DOI] [PubMed] [Google Scholar]
- 31.Pulmonary artery pulsatility index: physiological basis and clinical application. Lim HS, Gustafsson F. Eur J Heart Fail. 2020;22:32–38. doi: 10.1002/ejhf.1679. [DOI] [PubMed] [Google Scholar]
- 32.Pulmonary artery pulsatility index as a predictor of right ventricular failure in left ventricular assist device recipients: A systematic review. Essandoh M, Kumar N, Hussain N, et al. J Heart Lung Transplant. 2022;41:1114–1123. doi: 10.1016/j.healun.2022.04.007. [DOI] [PubMed] [Google Scholar]
- 33.The association of the pulmonary artery pulsatility index and right ventricular function after cardiac surgery. Wei J, Kee A, Dukes R, Franke J, Leonardo V, Flynn BC. Crit Care Res Pract. 2024;2024:5408008. doi: 10.1155/2024/5408008. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 34.Relation between severity of cerebral small vessel disease and pulsatility index of internal carotid artery in small vessel occlusion. Kitagawa T, Mitsumura H, Sato T, et al. Clin Neurol Neurosurg. 2024;237:108127. doi: 10.1016/j.clineuro.2024.108127. [DOI] [PubMed] [Google Scholar]
- 35.Distinct association between cerebral arterial pulsatility and subtypes of cerebral small vessel disease. Nam KW, Kwon HM, Lee YS. PLoS One. 2020;15:0. doi: 10.1371/journal.pone.0236049. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 36.New evidence for fractional pressure ratio prediction by pulsatility index from transcranial Doppler in patients with symptomatic cerebrovascular stenosis disease. Xu X, Raynald Raynald, Li X, et al. Quant Imaging Med Surg. 2024;14:264–272. doi: 10.21037/qims-23-701. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 37.Prevalence and predictors of cerebral microangiopathy determined by pulsatility index in an asymptomatic population from the ILERVAS project. Purroy F, Sánchez E, Lecube A, et al. Front Neurol. 2021;12:785640. doi: 10.3389/fneur.2021.785640. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 38.Pulsatility index-adjusted doppler flow measurement of pedal arteries in peripheral artery disease patients. Li MF, Liang HL, Chiang CL, Wu TH. J Chin Med Assoc. 2023;86:98–104. doi: 10.1097/JCMA.0000000000000843. [DOI] [PubMed] [Google Scholar]
- 39.Post-exercise pulsatility index indicates treatment effects in peripheral arterial occlusive disease (PAOD) Udelnow A, Hawemann M, Buschmann I, Meyer F, Halloul Z. Wien Klin Wochenschr. 2022;134:148–155. doi: 10.1007/s00508-021-01818-x. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 40.Prognostic implication of systemic arterial pulsatility index in patients with heart failure. Akao K, Imamura T. Heart Vessels. 2023;38:1092. doi: 10.1007/s00380-022-02214-9. [DOI] [PubMed] [Google Scholar]
- 41.Aortic pulsatility index: a new haemodynamic measure with prognostic value in advanced heart failure. Deis T, Rossing K, Gustafsson F. Card Fail Rev. 2022;8:0. doi: 10.15420/cfr.2022.09. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 42.The aortic pulsatility index: a new sign of severe acute aortic regurgitation. Sanchez F, Goudelin M, Evrard B, Vignon P. J Am Soc Echocardiogr. 2022;35:1006–1007. doi: 10.1016/j.echo.2022.04.008. [DOI] [PubMed] [Google Scholar]
- 43.Uterine artery pulsatility index for the prediction of obstetrical complications in preterm prelabor rupture of membranes. Duncan JR, Dorsett KM, Vilchez G, Schenone MH, Mari G. J Matern Fetal Neonatal Med. 2021;34:4060–4063. doi: 10.1080/14767058.2019.1702961. [DOI] [PubMed] [Google Scholar]
- 44.Reference values for uterine artery mean pulsatility index throughout pregnancy customized by parity, transvaginal measurement and blood pressure. Oviedo-Cruz H, Carrasco-Blancas ER, Cortés-Martínez MA. Gac Med Mex. 2022;158:48–54. doi: 10.24875/GMM.M22000640. [DOI] [PubMed] [Google Scholar]