Skip to main content
Respiratory Care logoLink to Respiratory Care
. 2024 Nov;69(11):1432–1443. doi: 10.4187/respcare.11745

Effects of Lung Injury and Abdominal Insufflation on Respiratory Mechanics and Lung Volume During Time-Controlled Adaptive Ventilation

Harry Ramcharran 1, Gregory Wetmore 2, Scott Cooper 3, Jacob Herrmann 4, Andrea Fonseca da Cruz 5, David W Kaczka 6, Joshua Satalin 7, Sarah Blair 8, Penny L Andrews 9, Nader M Habashi 10, Gary F Nieman 11, Michaela Kollisch-Singule 12,
PMCID: PMC11549635  PMID: 38408775

Abstract

BACKGROUD:

Lung volume measurements are important for monitoring functional aeration and recruitment and may help guide adjustments in ventilator settings. The expiratory phase of airway pressure release ventilation (APRV) may provide physiologic information about lung volume based on the expiratory flow-time slope, angle, and time to approach a no-flow state (expiratory time [TE]). We hypothesized that expiratory flow would correlate with estimated lung volume (ELV) as measured using a modified nitrogen washout/washin technique in a large-animal lung injury model.

METHODS:

Eight pigs (35.2 ± 1.0 kg) were mechanically ventilated using an Engström Carescape R860 on the APRV mode. All settings were held constant except the expiratory duration, which was adjusted based on the expiratory flow curve. Abdominal pressure was increased to 15 mm Hg in normal and injured lungs to replicate a combination of pulmonary and extrapulmonary lung injury. ELV was estimated using the Carescape FRC INview tool. The expiratory flow-time slope and TE were measured from the expiratory flow profile.

RESULTS:

Lung elastance increased with induced lung injury from 29.3 ± 7.3 cm H2O/L to 39.9 ± 15.1cm H2O/L, and chest wall elastance increased with increasing intra-abdominal pressures (IAPs) from 15.3 ± 4.1 cm H2O/L to 25.7 ± 10.0 cm H2O/L in the normal lung and 15.8 ± 6.0 cm H2O/L to 33.0 ± 6.2 cm H2O/L in the injured lung (P = .39). ELV decreased from 1.90 ± 0.83 L in the injured lung to 0.67 ± 0.10 L by increasing IAP to 15 mm Hg. This had a significant correlation with a TE decrease from 2.3 ± 0.8 s to 1.0 ± 0.1 s in the injured group with increasing insufflation pressures (ρ = 0.95) and with the expiratory flow-time slope, which increased from 0.29 ± 0.06 L/s2 to 0.63 ± 0.05 L/s2 (ρ = 0.78).

CONCLUSIONS:

Changes in ELV over time, and the TE and flow-time slope, could be used to demonstrate evolving lung injury during APRV. Using the slope to infer changes in functional lung volume represents a unique, reproducible, real-time, bedside technique that does not interrupt ventilation and may be used for clinical interpretation.

Keywords: end-expiratory lung volume, expiratory flow, airway pressure release ventilation, time-controlled adaptive ventilation

Introduction

ARDS is characterized by a loss of functional lung volume due to air space edema and atelectasis and thus may lead to hypoxemia.1 The loss of functional residual capacity (FRC) can be decreased by as much as 45% in severe ARDS.2 Under normal conditions, FRC is determined at end expiration by the balance of forces between the inward elastic recoil of the lung and outward recoil of the chest wall.3 FRC can be measured as the end-expiratory lung volume (EELV) in spontaneously breathing patients. However, in mechanically ventilated patients, EELV will be equal to FRC only when end-expiratory flow PEEP is zero.4 When PEEP is applied to a ventilated patient, EELV will generally be greater than the FRC.

The measurement of EELV during mechanical ventilation at a specified PEEP provides a physiologic assessment of developing lung injury and may be more sensitive than changes in respiratory system compliance.5-6 Thus, it may be useful for monitoring a disease course, to determine if the lung is being recruited,7 and perhaps to guide the setting of optimal PEEP4 and tidal volumes (VT).8 EELV has been measured using computed tomography scan volumetry9-11 and gas dilution techniques12-16 and has been inferred using static pressure/volume curves and inflection points17-19 and electrical impedance tomography.20,21 Despite the value of monitoring EELV, it is not routinely assessed in mechanically ventilated patients since currently available techniques are logistically challenging; require transport; involve radiation; cannot be measured repeatedly; or are costly, inaccurate, or interfere with ventilation.22-24 More recently, a modified nitrogen washout/washin technique of measuring EELV has been validated using a commercially available ventilator.11

One potential ventilator mode for which the EELV might be inferred is airway pressure release ventilation (APRV). The APRV expiratory phase duration using the time-controlled adaptive ventilation (TCAV) method may provide physiologic information about functional lung volume, although this has not been validated clinically or experimentally. The rate at which the lung decompresses during expiration should be related to the severity of lung injury. For example, a steep decay in the expiratory flow may indicate increased lung or chest wall elastance and/or decreased functional lung volume. This rate of decay may be represented by the slope and the acuity of the expiratory flow-time angle (φ), as well as the time it takes for the lung to reach a zero-flow state (TE). This experiment aimed to validate whether these elements of the expiratory flow curve (slope and TE) in APRV can be used as physiologic markers of lung injury severity and lung volume alterations during mechanical ventilation. The experiment was designed using pulmonary and extrapulmonary models of lung injury, with combinations of detergent instillation and abdominal hypertension. The APRV expiratory flow curve parameters (slope and TE) were compared against the GE Carescape R860 FRC INview tool for measuring lung volume.2,11,25,26 The use of slope and/or TE as a surrogate for lung volume represents a unique, reproducible technique that does not halt ventilation and is available on most mechanical ventilators and thus may be exploited for clinical interpretation.

QUICK LOOK.

Current knowledge

Lung injury leads to a loss of effective lung volume as a result of atelectasis and edema. This loss of lung volume can be prognostic and inform ventilator adjustments. Several methods of estimating effective lung volume have been proposed, but there are limited rapid and cost-effective methods that can be performed at the bedside.

What this paper contributes to our knowledge

Increasing intra-abdominal pressure with insufflation created a combined pulmonary and extrapulmonary lung injury in a porcine model. The slope of the expiratory flow in airway pressure release ventilation set according to the time-controlled adaptive ventilation method correlated with increasing lung injury as well as the estimated lung volume established using a nitrogen washout/washin method. The expiratory flow slope, therefore, represents a method for monitoring loss of effective lung volume.

Methods

All experiments were conducted with approval from the State University of New York Upstate Medical University Institutional Animal Care and Use Committee in accordance with Animal Research: Reporting of In Vivo Experiments guidelines. Eight female and male Yorkshire pigs (35.2 ± 1.0 kg) were anesthetized with an intravenous ketamine (90 mg/kg) and xylazine (10 mg/kg) solution. A 7.5-Fr endotracheal tube (Sheridan, Teleflex, Morrisville, North Carolina) was placed through a tracheostomy and connected to an Engström Carescape R860 mechanical ventilator (GE Healthcare, Madison, Wisconsin). A central venous catheter was placed in the external jugular vein for fluid and medication administration. The right carotid artery was cannulated for arterial blood pressure monitoring and blood gas measurements (cobas b 221, Roche, Basel, Switzerland). An esophageal balloon-tipped catheter (CooperSurgical, Trumbull, Connecticut) was placed into the distal esophagus to measure esophageal pressure to calculate lung and chest wall elastance. A cystostomy was performed for bladder pressure measurement and decompression.

APRV was selected as the mode of use for this study because the inspiratory and expiratory times are adjusted independently without being restricted to a specific inspiratory to expiratory ratio. An expiratory release pressure (PLow) of 0 cm H2O, as clinically suggested,27 provides an unobstructed picture of the expiratory flow curve. The expiratory phase can then provide physiologic information about the lung by analyzing the lung collapse time and the acuity of the expiratory flow-time angle, which may be reflective of lung volume and stiffness, just prior to the beginning of exhalation.

Animals were paralyzed with rocuronium (0.01–0.02 mg/kg/min) and placed on the ventilator with the following settings: PHigh 25 cm H2O, PLow 0 cm H2O, THigh 4.0 s, and FIO2 0.30. These settings were held constant throughout the experiment (except during lung injury) to determine the isolated effects of expiratory flow and duration after lung injury and with increased abdominal pressure. The TLow was adjusted according to the expiratory flow curve as previously described.27 Briefly, the peak expiratory flow (PEF) was measured and used to determine when to terminate expiratory flow (TEF) by multiplying PEF x 75% (PEF x 75% = TEF).

Lung volume was measured using the FRC INview module integrated within the Carescape as described by other authors.4,25 Briefly, nitrogen washout and washin rates were calculated by measuring end-tidal O2 and CO2 during a 10% change in FIO2. In this system, the measurements are recorded twice, and the average of the washout and washin rates are computed provided the difference between the 2 values is < 20%. Since the expiratory flow does not decay to zero during TCAV, the estimated EELV based on nitrogen washin/washout rates may be biased. Thus, we will hereafter refer to the measured lung volume as the estimated lung volume (ELV) to account for this. Before each ELV measurement, hemodynamic and ventilator parameters were also recorded.

The lung TE was determined as the length of time required for the lung to deflate to its resting lung volume, as indicated by a flow of 0 L/s (Fig. 1). A prolonged TE may indicate an obstructive lung process, whereas a shortened TE may indicate restrictive lung pathology. The expiratory flow-time angle (φ) generated by the PEF and expiratory flow curve is subject to change depending on the aspect ratio of the flow-time curve; therefore, the slope was used for interpretation. The slope was calculated as follows:

Slope=(PEFTEF)TLow Equation 1:

Fig. 1.

Fig. 1.

Expiratory flow in airway pressure release ventilation. After the release of the inspiratory phase, the greatest absolute flow represents the peak expiratory flow (PEF). The expiratory flow is terminated after a time (TLow) that is adjusted to terminate expiratory flow (TEF) at 75% of PEF. The expiratory flow-time slope and resulting angle φ determine the collapse time of the lung (TE) where the expiratory flow approaches 0 L/s. In this illustration, PEF is −0.8 L/s and TEF is −0.6 L/s with a TLow of 0.5 s. The calculated slope is, therefore, 0.4 L/s2. A steeper slope or more acute φ leads to a shorter TE, which requires adjusting the TLow to preserve estimated lung volume, as compared with a shallower slope or more obtuse φ with a longer TE, allowing for a longer TLow.

Figure 1 is an example hypothetical PEF of −0.8 L/s, with TEF of −0.6 L/s and TLow of 0.5 s. The calculated slope is, therefore, 0.4 L/s2.

Lung strain is another marker of progressive lung injury and has been postulated to be a key factor in the development of ventilator-induced lung injury yet cannot be easily measured clinically.8 We calculated lung strain by normalizing the VT of the TCAV breath to ELV:

Lung strain=VTELV Equation 2:

To simulate intra-abdominal hypertension and induce increased chest wall elastance, a 12-mm trocar was placed into the abdomen via the umbilicus. The abdomen was then insufflated with CO2 to pressures of 6, 8, 10, 12, 14, and 15 mm Hg.

Following these baseline measurements, the animals were transitioned to CPAP at 20 cm H2O prior to bronchoscopic detergent administration (Sigma-Aldrich, Saint Louis, Missouri). TWEEN leads to surfactant dysfunction, simulating this well-known component of ARDS pathophysiology.28 The airway was visualized with a bronchoscope, which was advanced down the right main bronchus past the right middle lobe bronchus, following which a 3% TWEEN-20 detergent solution (0.375 mL/kg) was instilled, as previously described.29,30 The bronchoscope was withdrawn to the carina and then advanced down the left main bronchus just past the left cranial lobe bronchus, and another (0.375 mL/kg) TWEEN-20 dose was administered. This model induces a heterogeneous lung injury targeting the diaphragmatic lobes. The animals were then ventilated for 10 min on a volume control mode with VT 10 mL/kg, PEEP 0 cm H2O, and FIO2 0.30. An arterial blood gas was obtained prior to and after detergent instillation to document lung injury by PaO2/FIO2. Following lung injury, the abdomen was again insufflated to 6, 8, 10, 12, 14, and 15 mm Hg. A 10-min acclimatization period was implemented following each insufflation pressure change. Since both TWEEN injury and abdominal insufflation may induce systemic hypotension, an intravenous bolus of lactated Ringer’s solution was administered prior to detergent instillation.

Lung, chest wall, and total respiratory system elastance values along with intra-abdominal pressures (IAPs) were recorded at baseline and at each stepwise increase in insufflation pressure before and after lung injury. Each of the expiratory flow parameters (flow-time slope, PEF, TEF, TE, and TLow) were also measured. TE was measured by manually extending the duration of TLow until the expiratory flow curve approached 0 L/s. With each increase in insufflation pressure, the TLow was adjusted to maintain TEF at 75% of PEF to generate a time-controlled PEEP.

Statistical Analysis

Repeated-measures analysis of variance was used to determine differences between groups before and after lung injury, with increasing IAP for continuous parameters with a post hoc Tukey test if significance was found in the condition × IAP effect. Pearson correlation coefficient (ρ) was used to determine the strength of the relationship between ELV and the flow-time slope and TE based on the expiratory flow and similarly used to confirm a correlation between the set intra-abdominal insufflation pressure and measured bladder pressure. A t test was used to compare PaO2/FIO2 and VT before and after induced lung injury. A P value of < .05 was considered significant. Data are presented as mean ± SD. Statistical analysis was performed using Prism 9 (GraphPad by Dotmatics, Dotmatics, Boston, Massachusetts).

Results

Baseline PaO2/FIO2 in animals prior to intra-abdominal insufflation and Tween lung injury was 489 ± 17.9 mm Hg. TWEEN instillation led to mild-moderate hypoxemia, with a significant decrease in the PaO2/FIO2 to 220 ± 11.7 mm Hg (P < .001). Following lung injury, the generated VT were significantly decreased from 8.3 ± 0.8 mL/kg/breath to 6.3 ± 1.5 mL/kg/breath (P = .02). The normal and injured lungs had a similar total decrease in VT (normal 75 ± 50 mL and injured 77 ± 50 mL) with increasing IAPs to 15 mm Hg (P = .45) (Table 1). Despite increasing IAP and lung injury, heart rate did not change appreciably (P = .90) (Table 1).

Table 1.

Pulmonary and Hemodynamic Parameters in Normal and Injured Lung With Varying Insufflation Pressures

graphic file with name DE-RESC240055T001.jpg

As IAP increased from 6 mm Hg to 15 mm Hg under baseline conditions, TLow decreased from 0.54 ± 0.07 s to 0.46 ± 0.04 s such that the TEF/PEF ratio could be maintained at 75% (Table 2). After lung injury, the required TLow was 0.46 ± 0.07 s and decreased to 0.34 ± 0.05 s with increasing IAP to 15 mm Hg. These adjustments were necessary to maintain a time controlled PEEP, which was 11.9 ± 2.0 cm H2O in the normal lungs at baseline and remained between 11.9–14.4 cm H2O with increasing IAP, despite a PLow set at 0 cm H2O (Table 1). In the injured lungs, the time controlled PEEP was lower than in the normal lung (P = .004) but remained between 10.4–13.3 cm H2O with increasing IAP. With a PHigh set at 25 cm H2O, the driving pressure was found to be 12.9 ± 1.8 cm H2O and 11.6 ± 1.8 cm H2O before and after lung injury, respectively, with no added abdominal pressure (Table 1). With increasing IAP, the average driving pressure was higher in the injured lung group (P = .001 vs normal lung) but never > 12.9 cm H2O in the normal lung or > 13.8 cm H2O in the injured lung group.

Table 2.

Pulmonary Parameters Relevant to Estimated Lung Volume in Normal and Injured Lung With Varying Insufflation Pressures

graphic file with name DE-RESC240055T002.jpg

Bladder pressures were measured to validate the set insufflation pressure and were found to be within 1 cm H2O of each other at each generated insufflation pressure, with strong positive correlation in both normal (ρ = 0.89) and induced lung injury (ρ = 0.91) (Table 1). With increasing IAPs, chest wall elastance increased from 15.3 ± 4.1 cm H2O/L to 25.7 ± 10.0 cm H2O/L under baseline conditions. Following lung injury, chest wall elastance increased from 15.8 ± 6.0 cm H2O/L to 33.0 ± 6.2 cm H2O/L (Fig. 2 and Table 1). With increasing IAP to 15 mm Hg in the normal lung, the chest wall elastance increased to account for 53% of the total respiratory system elastance (Fig. 2A). Although there was an increase in chest wall elastance in the injured lung with increasing IAP, the contribution to the total respiratory system elastance reached a maximum of 41%, likely due to the concomitant lung injury favoring an increase in the lung elastance (Fig. 2B). Transpulmonary pressure decreased from 16.1 ± 0.9 cm H2O to 10.6 ± 2.5 cm H2O with increasing IAPs in the normal lung and demonstrated a similar decrease from 17.1 ± 0.6 cm H2O to 13.9 ± 0.8 cm H2O (P = .88) (Table 1). Since the plateau pressure was held constant, the pleural pressure was found to have a concordant increase in the normal lung from 8.6 ± 0.9 cm H2O to 13.6 ± 2.4 cm H2O and from 7.3 ± 0.9 cm H2O to 10.2 ± 0.7 cm H2O (P = .92).

Fig. 2.

Fig. 2.

Partitioning of the respiratory system elastance into lung and chest wall elastance. The contribution of lung and chest wall elastance to the total respiratory system elastance with increasing intra-abdominal pressure in A: normal lung and B: after induced lung injury. Error bars represent SD.

The average baseline lung elastance was 29.3 ± 7.3 cm H2O/L and increased to 39.9 ± 15.1 cm H2O/L after bronchoscopic TWEEN instillation (Fig. 2, Table 1). There was an initial decrease in lung and respiratory system elastance in both the normal and injured lungs, with increasing insufflation pressures ≤ 8 mm Hg (Fig. 2, Table 1). Thereafter, with IAPs of ≥ 10 mm Hg, both lung and chest wall elastance values increased.

PEF did not change appreciably with increasing IAP (range 0.70–0.73 L/s) nor with lung injury plus increasing abdominal pressure (range 0.63–0.66 L/s) (Table 2). The flow-time slope increased (the φ flow-time angle became more acute) with increasing abdominal pressures from 6 mm Hg to 15 mm Hg: Slope increased from 0.29 ± 0.07 L/s2 to 0.58 ± 0.10 L/s2 in the healthy lungs and from 0.29 ± 0.06 L/s2 to 0.63 ± 0.05 L/s2 in the injured lungs (Fig. 3, Table 2). This change in slope is supported by the increased chest wall elastance associated with increasing IAP, leading to a higher velocity exhalation of gas from the lung. Given the similar PEF but more acutely angled slope, there was an expected decrease in TE as the lung deflated at a higher rate to its end-expiratory volume. TE decreased from 2.6 s to 1.2 s in the healthy lung, with IAP increasing from 0 mm Hg to 15 mm Hg and decreased from 2.3 s to 1.0 s in the injured lung (Fig. 3, Table 2).

Fig. 3.

Fig. 3.

Changes in the expiratory time and flow. The slope (red) increased with increasing intra-abdominal pressure (IAP) in both A: normal lungs and B: after induced lung injury, leading to a more visually acute expiratory flow angle φ. The time at expiratory pressure (TLow) was adjusted (shortened) with increasing IAP to accommodate the changing slope to maintain a termination expiratory flow/peak expiratory flow ratio of 75%. The expiratory time similarly decreased with increasing IAP. Representations of the expiratory φ flow-time angle and shortened TLow are depicted in the green-dashed boxes. Error bars represent SD.

With lung injury, the ELV decreased and was reflected by a decrease in TE (Fig. 4A) and increase in slope (Fig. 4B); however, the relationship with TE was found to be linear, whereas it was curvilinear with slope. Similarly, in both the normal and injured lungs, as the IAP increased the flow-time slope became steeper (or the φ flow-time angle more acute) with a shorter TE and concomitant decrease in ELV. There was a strong correlation between ELV and TE in the normal lung (ρ = 0.88) and injured lung (ρ = 0.95). There was a moderate correlation between ELV and slope in the normal lung (ρ = 0.78) and the injured lung (ρ = 0.78). Lung strain increased with and without lung injury as IAP increased (P = .004) (Table 2).

Fig. 4.

Fig. 4.

Change in expiratory time (TE) and the expiratory slope relative to estimated lung volume (ELV). Changes in A: TE and B: expiratory slope relative to ELV with increasing intra-abdominal pressures from 0 mm Hg to 15 mm Hg in normal and injured lungs. Error bars represent SD. ELV error bars were removed for figure clarity (values reported in Table 2).

Discussion

This was a physiologic study of the relative impact of increasing chest wall elastance with increasing IAP in normal lungs and mild ARDS on ELV. Additionally, the association between this ELV and the expiratory flow parameters of the APRV waveform was studied (Fig. 1). Whereas lung injury increased lung elastance, increasing IAP in normal and injured lungs increased chest wall elastance with an associated increase in pleural pressure (Fig. 2). The expiratory flow slope and time to achieve a no-flow state were found to correlate with this ELV (Fig. 4), suggesting that they may be useful to monitor the progression of lung injury and signal the need for an adjustment in TLow (Fig. 3).

Lung injury, and specifically ARDS, is associated with marked reductions in aerated lung volume.31 The FRC of patients with moderate-to-severe ARDS may be < 1 L12 as compared with ∼3 L in healthy patients.32 This reduction in functional lung serves as a marker of progressive lung injury and has important prognostic implications that can be used to direct recruitment efforts to prevent further lung injury and reestablish lung volumes.26 It follows that EELV is a primary determinant of lung oxygen capacity since oxygen and CO2 exchange is a function of aeratable lung volume.26 EELV has not been standardly incorporated into the clinical assessment of patients due to the lack of readily available, accurate, and cost-effective measurement techniques.33 Thus, there is an appeal to finding a marker of EELV changes using a standard mechanical ventilator, which served as the premise for this study.

Abdominal insufflation ≤ 15 mm Hg is common in laparoscopic procedures34,35 and is known to lead to increases in respiratory system and chest wall elastance, as well as respiratory system resistance.36,37 Increased abdominal pressure can also affect expiration due to diaphragm elevation and inspiration by pushing the rib cage outward at the zone of apposition.38 The pneumoperitoneum and additional patient factors such as position, height, and weight may have considerable effects on intraoperative respiratory system mechanics.36,39 Thus, our findings are relevant to surgical patients with or without lung injury, as well as critically ill patients with extrapulmonary ARDS in whom intra-abdominal hypertension may lead to altered respiratory mechanics.40,41

Measured bladder pressure correlated well with insufflation pressure, confirming that targeted pressures were achieved. In the direct lung injury group, there was an expected increase in lung elastance compared to uninjured lungs. Both groups demonstrated increasing chest wall elastance with increasing IAP, but this effect was more pronounced in the injured group. Lung and respiratory system elastances initially decreased with increasing insufflation pressure, which was consistent with the initial increase in exhaled VT. This was observed until insufflation pressures of 8 mm Hg, suggesting that this is the point at which competing alveolar de-recruitment and subsequent reduction in VT outweigh this effect. Despite increasing abdominal insufflation pressures, the PEF remained unchanged. This suggests that the increased flow expected with rising lung elastance was balanced by the overall reduction in aeratable lung and resulting VT.

ELV can be used to estimate lung strain, which is another important factor to consider in terms of prognostication and to herald the need for ventilator adjustments where high levels of dynamic strain have been associated with mortality and lung edema.42 The idea of using ELV and elastance to guide PEEP has been well established to optimize recruitment and reduce strain,43,44 particularly with increased abdominal pressure, such as during laparoscopy.45 Decreasing PEEP in patients undergoing laparoscopy leads to decreased oxygenation and compliance and increased driving pressure and atelectasis, but there is no method of determining optimal PEEP.46 The calculated lung strain revealed a continued increase in strain concomitant with increasing IAP. Since the plateau pressure was held constant throughout the experiment, increasing insufflation pressures led to an accompanying decrease in VT and lower transpulmonary pressure. This combined with the correlate increase in pleural pressure led to a reduction in VT relative to the smaller functional lung, preventing overdistention of the fewer alveoli available to participate in gas exchange.47

Different methods have been proposed to adjust ventilator settings based on lung and/or respiratory system elastance values41 and intratidal variations in global lung mechanics.48 These mechanical properties have been particularly relevant for determining PEEP responsiveness and so-called optimal PEEP.41 Our study found that both the expiratory flow-time slope and TE correlate with ELV and can detect early changes in respiratory physiology. Thus, changes in the slope might inform the clinician about lung recruitability. For example, a steeper expiratory slope suggests a reduction in ELV and thus an opportunity to recruit. Furthermore, this would prompt an adjustment to TLow to maintain a time controlled PEEP sufficient to prevent de-recruitment. In summary, we propose that the expiratory flow pattern in APRV may be added to the repertoire of physiologic information that guides adjustments to ventilator settings in real time.

APRV is available on most ventilators; therefore, finding an ELV corollary that can be readily assessed without the requirement of O2 changes or a prolonged washout maneuver is attractive. The TCAV methodology advocates for an extended THigh to maximize mean airway pressure and alveolar recruitment without requiring a potentially injurious plateau pressure and a brief TLow to prevent alveolar de-recruitment at expiration. Adjusting the TLow requires a cognizant balance between allowing sufficient time for CO2 removal while limiting alveolar collapse time. PLow is generally set at 0 cm H2O to allow maximal CO2 removal, where a PLow > 0 cm H2O would act as a resistor for optimal ventilation. The flow-time slope of the expiratory flow curve, therefore, provides important physiologic information regarding lung elastance, and the TLow may be adjusted using this curve. Clinically, the TLow must be set < TE to ensure the lung does not collapse to its FRC, leading to atelectasis and subsequent injury from recruitment/de-recruitment. Adjusting the TLow such that the TEF is 75% of the PEF has been shown to be the sweet spot in terms of minimizing alveolar strain and heterogeneity30,49 and maximizing ventilation.50,51

Despite lung injury and increasing IAP with a PLow set at 0 cm H2O, the time controlled PEEP was 12–14 cm H2O in the normal lung and 10–13 cm H2O in the injured lung. This is an important finding since the plateau pressure was held constant; the driving pressure was maintained by adjusting the TLow according to the acuity of the flow-time slope. The TE changed on the order of seconds with increasing insufflation pressure, whereas the TLow was adjusted at the level of the 1/100th of a second (centisecond) with increasing insufflation pressures, illustrating the precision that is required to optimize the time controlled PEEP. Maintaining PEEP in the setting of lung injury and increasing chest wall elastance is vital to prevent atelectasis and subsequent recruitment/de-recruitment injury; however, the optimal method of adjusting PEEP using conventional ventilation is unknown.52-55 A patient-directed methodology that factors in the collapse time of the lung subunits allows for adaptive ventilator adjustments according to lung injury severity and is, therefore, ideal.

Study Limitations

The Carescape INview tool has been previously validated but not with APRV.11,25,26 The ELV calculations are predicated on the assumption that the homogeneity of alveolar gas distribution and gas exchange/cellular metabolism is constant throughout the measurement.56 ELV measurements may also be influenced by hemodynamic instability, variable VT,14 and alteration in muscle tone during anesthesia and paralysis.12 One of the advantages of using the expiratory flow waveform is that patient hemodynamics should not influence ELV calculations during mechanical ventilation in contrast to spontaneous breathing. Thus, our animals were paralyzed during the protocol to limit any artifact of spontaneous breathing on measurements. ELV was not validated in multiple settings, such as higher/lower PHigh. Since the PHigh influences the VT, and the objective of the study was to quantify ELV based on the expiratory flow slope, we did not want to obscure the findings by altering multiple settings, particularly those that impact delivered VT. Additionally, ELV and the expiratory flow slope were studied in a model of mild ARDS. Though it is expected that the results of this study could be extrapolated to moderate and severe ARDS, this increasing injury severity was not specifically studied.

Our study was a physiologic investigation and thus does not reflect clinical outcomes or how ELV may vary with progressive lung injury over time. Animal models with defined lung injury represent a controlled environment that does not fully replicate patients with heterogeneous ARDS, whereby variable respiratory resistance (eg, bronchial tone, PEF), patient physiology (eg, VT, fluid status, body mass index), and disease severity may lead to altered ELV measurements.57 This study used sequentially increasing IAPs to mimic the increasing chest wall elastance that develops in critically ill patients with progressive chest wall edema and intra-abdominal hypertension, which may have influenced measurements as compared with a random sequence of applied IAPs. Though caution must be given to interpreting the data, the animal model used in this study was designed to evaluate a combination of pulmonary and extrapulmonary lung injury to allow for generalization across various lung injury phenotypes.

Conclusions

Establishing a surrogate for ELV may allow clinicians to trend functional volume loss, establish earlier diagnoses of lung injury, and monitor evolving respiratory pathophysiology. The most important conclusion of this study is that multiple parameters can be assessed on a breath-to-breath basis during TCAV, including TLow, TE, the expiratory flow-time slope, and visual interpretation of the φ flow-time angle, to demonstrate worsening lung injury, increased lung and chest wall elastance, and predict decreasing ELV. This was a physiologic study, but the results may be extrapolated to patients in whom increasing acuity of the flow-time slope and shortening of the TLow may indicate a need to recruit the lost functional lung volume through adjustments in APRV parameters. Since the expiratory flow-time slope informs TLow adjustments, our findings demonstrate the ability of TCAV to adapt to individual patient lung pathophysiology.

Footnotes

Dr Kollisch-Singule discloses a relationship with Dräger Medical Systems. Dr Habashi is the founder of ICON, of which Ms Andrews is an employee. Dr Habashi holds patents on a method of initiating, managing, and/or weaning airway pressure release ventilation, as well as controlling a ventilator in accordance with the same. Drs Kaczka and Herrmann are co-founders and shareholders of OscillaVent, and are co-inventors on a patent involving multifrequency oscillatory ventilation. Drs Kaczka and Herrmann disclose a relationship with ZOLL Medical. Dr Kaczka discloses a relationship with Lungpacer Medical. The remaining authors have disclosed no conflicts of interest. The authors maintain that industry had no role in the design and conduct of the study; the collection, management, analysis, or interpretation of the data; nor the preparation, review, or approval of the manuscript.

Portions of these data were presented at the 2022 Upstate New York American College of Surgeons Committee on Trauma Resident Paper Competition November 2, 2022, in Buffalo, New York; and at the 2023 American Thoracic Society International Conference, held May 19–24, 2023, in Washington, District of Columbia.

This study was supported by NIH R01 HL142702-04 as well as an educational research grant from Dräger Medical Systems. It was supported in part by the Office of the Assistant Secretary of Defense for Health Affairs, Peer Reviewed Medical Research Program awards W81XWH-20-1-0696 and W81XWH-20-1-0507. Opinions, interpretations, conclusions, and recommendations are those of the authors and are not necessarily endorsed by the Department of Defense.

REFERENCES

  • 1.Puybasset L, Cluzel P, Chao N, Slutsky AS, Coriat P, Rouby JJ. A computed tomography scan assessment of regional lung volume in acute lung injury. The CT Scan ARDS Study Group. Am J Respir Crit Care Med 1998;158(5 Pt 1):1644-1655. [DOI] [PubMed] [Google Scholar]
  • 2.Rylander C, Hogman M, Perchiazzi G, Magnusson A, Hedenstierna G. Functional residual capacity and respiratory mechanics as indicators of aeration and collapse in experimental lung injury. Anesth Analg 2004;98(3):782-789. [DOI] [PubMed] [Google Scholar]
  • 3.Rahn H, Otis AB, Chadwick LE, Fenn WO. The pressure-volume diagram of the thorax and lung. Am J Physiol 1946;146(2):161-178. [DOI] [PubMed] [Google Scholar]
  • 4.Bikker IG, van Bommel J, Reis Miranda D, Bakker J, Gommers D. End-expiratory lung volume during mechanical ventilation: a comparison with reference values and the effect of positive end-expiratory pressure in intensive care unit patients with different lung conditions. Crit Care 2008;12(6):R145. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Bellani G, Guerra L, Musch G, Zanella A, Patroniti N, Mauri T, et al. Lung regional metabolic activity and gas volume changes induced by tidal ventilation in patients with acute lung injury. Am J Respir Crit Care Med 2011;183(9):1193-1199. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Gonzalez-Lopez A, Garcia-Prieto E, Batalla-Solis E, Amado-Rodriguez L, Avello N, Blanch L, et al. Lung strain and biological response in mechanically ventilated patients. Intensive Care Med 2012;38(2):240-247. [DOI] [PubMed] [Google Scholar]
  • 7.Dyhr T, Laursen N, Larsson A. Effects of lung recruitment maneuver and positive end-expiratory pressure on lung volume, respiratory mechanics, and alveolar gas mixing in patients ventilated after cardiac surgery. Acta Anaesthesiol Scand 2002;46(6):717-725. [DOI] [PubMed] [Google Scholar]
  • 8.Chiumello D, Carlesso E, Cadringher P, Caironi P, Valenza F, Polli F, et al. Lung stress and strain during mechanical ventilation for acute respiratory distress syndrome. Am J Respir Crit Care Med 2008;178(4):346-355. [DOI] [PubMed] [Google Scholar]
  • 9.Malbouisson LM, Muller JC, Constantin JM, Lu Q, Puybasset L, Rouby JJ; CT Scan ARDS Study Group. Computed tomography assessment of positive end-expiratory pressure–induced alveolar recruitment in patients with acute respiratory distress syndrome. Am J Respir Crit Care Med 2001;163(6):1444-1450. [DOI] [PubMed] [Google Scholar]
  • 10.David M, Karmrodt J, Bletz C, David S, Herweling A, Kauczor HU, et al. Analysis of atelectasis, ventilated, and hyperinflated lung during mechanical ventilation by dynamic CT. Chest 2005;128(5):3757-3770. [DOI] [PubMed] [Google Scholar]
  • 11.Chiumello D, Cressoni M, Chierichetti M, Tallarini F, Botticelli M, Berto V, et al. Nitrogen washout/washin, helium dilution, and computed tomography in the assessment of end-expiratory lung volume. Crit Care 2008;12(6):R150. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12.Gommers D. Functional residual capacity and absolute lung volume. Curr Opin Crit Care 2014;20(3):347-351. [DOI] [PubMed] [Google Scholar]
  • 13.Robinson PD, Latzin P, Verbanck S, Hall GL, Horsley A, Gappa M, et al. Consensus statement for inert gas washout measurement using multiple- and single-breath tests. Eur Respir J 2013;41(3):507-522. [DOI] [PubMed] [Google Scholar]
  • 14.Weismann D, Reissmann H, Maisch S, Fullekrug B, Schulte J. Monitoring of functional residual capacity by an oxygen washin/washout; technical description and evaluation. J Clin Monit Comput 2006;20(4):251-260. [DOI] [PubMed] [Google Scholar]
  • 15.Ohman T, Sigmundsson TS, Hallback M, Suarez Sipmann F, Wallin M, Oldner A, et al. Clinical and experimental validation of a capnodynamic method for end-expiratory lung volume assessment. Acta Anaesthesiol Scand 2020;64(5):670-676. [DOI] [PubMed] [Google Scholar]
  • 16.Eichler W, Schumacher J, Roth-Isigkeit A, Braun J, Kuppe H, Klotz KF. Automated evaluation of functional residual capacity by oxygen washout. J Clin Monit Comput 2002;17(3-4):195-201. [DOI] [PubMed] [Google Scholar]
  • 17.Falke KJ, Pontoppidan H, Kumar A, Leith DE, Geffin B, Laver MB. Ventilation with end-expiratory pressure in acute lung disease. J Clin Invest 1972;51(9):2315-2323. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Sondergaard S, Karason S, Wiklund J, Lundin S, Stenqvist O. Alveolar pressure monitoring: an evaluation in a lung model and in patients with acute lung injury. Intensive Care Med 2003;29(6):955-962. [DOI] [PubMed] [Google Scholar]
  • 19.Muramatsu K, Yukitake K. A new method to analyze lung compliance when pressure-volume relationship is nonlinear. Am J Respir Crit Care Med 1999;159(6):2028. [DOI] [PubMed] [Google Scholar]
  • 20.Grivans C, Lundin S, Stenqvist O, Lindgren S. Positive end-expiratory pressure–induced changes in end-expiratory lung volume measured by spirometry and electric impedance tomography. Acta Anaesthesiol Scand 2011;55(9):1068-1077. [DOI] [PubMed] [Google Scholar]
  • 21.Bikker IG, Leonhardt S, Bakker J, Gommers D. Lung volume calculated from electrical impedance tomography in ICU patients at different PEEP levels. Intensive Care Med 2009;35(8):1362-1367. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22.Hewlett AM, Hulands GH, Nunn JF, Minty KB. Functional residual capacity during anesthesia. I: methodology. Br J Anaesth 1974;46(7):479-485. [DOI] [PubMed] [Google Scholar]
  • 23.Fretschner R, Deusch H, Weitnauer A, Brunner JX. A simple method to estimate functional residual capacity in mechanically ventilated patients. Intensive Care Med 1993;19(7):372-376. [DOI] [PubMed] [Google Scholar]
  • 24.Ibanez J, Raurich JM, Moris SG. A simple method for measuring the effect of PEEP on functional residual capacity during mechanical ventilation. Crit Care Med 1982;10(5):332-334. [DOI] [PubMed] [Google Scholar]
  • 25.Olegard C, Sondergaard S, Houltz E, Lundin S, Stenqvist O. Estimation of functional residual capacity at the bedside using standard monitoring equipment: a modified nitrogen washout/washin technique requiring a small change of the inspired oxygen fraction. Anesth Analg 2005;101(1):206-212. [DOI] [PubMed] [Google Scholar]
  • 26.Berger-Estilita J, Haenggi M, Ott D, Berger D. Accuracy of the end-expiratory lung volume measured by the modified nitrogen washout/washin technique: a bench study. J Transl Med 2021;19(1):36. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 27.Habashi NM. Other approaches to open lung ventilation: airway pressure release ventilation. Crit Care Med 2005;33(3 Suppl):S228-240. [DOI] [PubMed] [Google Scholar]
  • 28.Slutsky AS, Ranieri VM. Ventilator-induced lung injury. N Engl J Med 2013;369(22):2126-2136. [DOI] [PubMed] [Google Scholar]
  • 29.Judge EP, Hughes JM, Egan JJ, Maguire M, Molloy EL, O’Dea S. Anatomy and bronchoscopy of the porcine lung. A model for translational respiratory medicine. Am J Respir Cell Mol Biol 2014;51(3):334-343. [DOI] [PubMed] [Google Scholar]
  • 30.Jain SV, Kollisch-Singule M, Satalin J, Searles Q, Dombert L, Abdel-Razek O, et al. The role of high airway pressure and dynamic strain on ventilator-induced lung injury in a heterogeneous acute lung injury model. Intensive Care Med Exp 2017;5(1):25. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 31.Brochard L, Martin GS, Blanch L, Pelosi P, Belda FJ, Jubran A, et al. Clinical review: respiratory monitoring in the ICU - a consensus of 16. Crit Care 2012;16(2):219. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 32.Hopkins E, Sharma S. Physiology, functional residual capacity. Treasure Island (Florida): StatPearls; 2022. [PubMed] [Google Scholar]
  • 33.Hedenstierna G. The recording of FRC–is it of importance and can it be made simple? Intensive Care Med 1993;19(7):365-366. [DOI] [PubMed] [Google Scholar]
  • 34.Daskalakis M, Scheffel O, Weiner RA. High-flow insufflation for the maintenance of the pneumoperitoneum during bariatric surgery. Obes Facts 2009;2 Suppl 1(Suppl 1):37-40. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 35.Srivastava A, Niranjan A. Secrets of safe laparoscopic surgery: anesthetic and surgical considerations. J Minim Access Surg 2010;6(4):91-94. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 36.Fahy BG, Barnas GM, Flowers JL, Jacobs SC, Plotkin JS, Delaney PA. Effects of split torso positioning and laparoscopic surgery for donor nephrectomy on respiratory mechanics. J Clin Anesth 1998;10(2):103-108. [DOI] [PubMed] [Google Scholar]
  • 37.Loring SH, Behazin N, Novero A, Novack V, Jones SB, O’Donnell CR, et al. Respiratory mechanical effects of surgical pneumoperitoneum in humans. J Appl Physiol (1985) 2014;117(9):1074-1079. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 38.Loring SH, Mead J. Action of the diaphragm on the rib cage inferred from a force-balance analysis. J Appl Physiol Respir Environ Exerc Physiol 1982;53(3):756-760. [DOI] [PubMed] [Google Scholar]
  • 39.Fahy BG, Barnas GM, Nagle SE, Flowers JL, Njoku MJ, Agarwal M. Effects of Trendelenburg and reverse Trendelenburg postures on lung and chest wall mechanics. J Clin Anesth 1996;8(3):236-244. [DOI] [PubMed] [Google Scholar]
  • 40.Malbrain ML, Chiumello D, Pelosi P, Wilmer A, Brienza N, Malcangi V, et al. Prevalence of intra-abdominal hypertension in critically ill patients: a multi-center epidemiological study. Intensive Care Med 2004;30(5):822-829. [DOI] [PubMed] [Google Scholar]
  • 41.Gattinoni L, Pelosi P, Suter PM, Pedoto A, Vercesi P, Lissoni A. Acute respiratory distress syndrome caused by pulmonary and extrapulmonary disease. Different syndromes? Am J Respir Crit Care Med 1998;158(1):3-11. [DOI] [PubMed] [Google Scholar]
  • 42.Protti A, Andreis DT, Monti M, Santini A, Sparacino CC, Langer T, et al. Lung stress and strain during mechanical ventilation: any difference between statics and dynamics? Crit Care Med 2013;41(4):1046-1055. [DOI] [PubMed] [Google Scholar]
  • 43.Dellamonica J, Lerolle N, Sargentini C, Beduneau G, Di Marco F, Mercat A, et al. PEEP-induced changes in lung volume in acute respiratory distress syndrome. Two methods to estimate alveolar recruitment. Intensive Care Med 2011;37(10):1595-1604. [DOI] [PubMed] [Google Scholar]
  • 44.Lambermont B, Ghuysen A, Janssen N, Morimont P, Hartstein G, Gerard P, et al. Comparison of functional residual capacity and static compliance of the respiratory system during a positive end-expiratory pressure (PEEP) ramp procedure in an experimental model of acute respiratory distress syndrome. Crit Care 2008;12(4):R91. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 45.Tusman G, Wallin M, Acosta C, Santanera B, Portela F, Viotti F, et al. Positive end-expiratory pressure individualization guided by continuous end-expiratory lung volume monitoring during laparoscopic surgery. J Clin Monit Comput 2021;36(5):1557-1567. [DOI] [PubMed] [Google Scholar]
  • 46.De Meyer GRA, Morrison SG, Saldien V, Jorens P, Schepens T. Minimizing lung injury during laparoscopy in head-down tilt: a physiological cohort study. Cardiovasc Pathophys Outcomes 2022. [DOI] [PubMed] [Google Scholar]
  • 47.Marini JJ, Gattinoni L. Improving lung compliance by external compression of the chest wall. Crit Care 2021;25(1):264. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 48.Amini R, Herrmann J, Kaczka DW. Intratidal overdistention and de-recruitment in the injured lung: a simulation study. IEEE Trans Biomed Eng 2017;64(3):681-689. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 49.Kollisch-Singule M, Emr B, Smith B, Roy S, Jain S, Satalin J, et al. Mechanical breath profile of airway pressure release ventilation: the effect on alveolar recruitment and microstrain in acute lung injury. JAMA Surg 2014;149(11):1138-1145. [DOI] [PubMed] [Google Scholar]
  • 50.Roy S, Sadowitz B, Andrews P, Gatto LA, Marx W, Ge L, et al. Early stabilizing alveolar ventilation prevents acute respiratory distress syndrome: a novel timing-based ventilatory intervention to avert lung injury. J Trauma Acute Care Surg 2012;73(2):391-400. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 51.Roy S, Habashi N, Sadowitz B, Andrews P, Ge L, Wang G, et al. Early airway pressure release ventilation prevents ARDS-a novel preventive approach to lung injury. Shock 2013;39(1):28-38. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 52.Seah AS, Grant KA, Aliyeva M, Allen GB, Bates JHT. Quantifying the roles of tidal volume and PEEP in the pathogenesis of ventilator-induced lung injury. Ann Biomed Eng 2011;39(5):1505-1516. [DOI] [PubMed] [Google Scholar]
  • 53.Bergez M, Fritsch N, Tran-Van D, Saghi T, Bounkim T, Gentile A, et al. PEEP titration in moderate to severe ARDS: plateau versus transpulmonary pressure. Ann Intensive Care 2019;9(1):81. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 54.Coruh B, Luks AM. Positive end-expiratory pressure. When more may not be better. Ann Am Thorac Soc 2014;11(8):1327-1331. [DOI] [PubMed] [Google Scholar]
  • 55.Gattinoni L, Collino F, Maiolo G, Rapetti F, Romitti F, Tonetti T, et al. Positive end-expiratory pressure: how to set it at the individual level. Ann Transl Med 2017;5(14):288. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 56.Dellamonica J, Lerolle N, Sargentini C, Beduneau G, Di Marco F, Mercat A, et al. Accuracy and precision of end-expiratory lung-volume measurements by automated nitrogen washout/washin technique in patients with acute respiratory distress syndrome. Crit Care 2011;15(6):R294. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 57.van Drunen EJ, Chiew YS, Chase JG, Shaw GM, Lambermont B, Janssen N, et al. Expiratory model-based method to monitor ARDS disease state. Biomed Eng Online 2013;12:57. [DOI] [PMC free article] [PubMed] [Google Scholar]

Articles from Respiratory Care are provided here courtesy of Mary Ann Liebert, Inc.

RESOURCES