Abstract
The hollow fiber membrane bundle is the functional component of artificial lungs, transferring oxygen to and carbon dioxide from the blood. It is also the primary location of blood clot formation and propagation in these devices. The geometric design of fiber bundles is defined by a narrow set of parameters that determine gas exchange efficiency and blood flow resistance, principally: fiber packing density, path length, and frontal area. These same parameters also affect thrombosis. This study investigated the effect of these parameters on clot formation using 3D printed flow chambers that mimic the geometry and blood flow patterns of fiber bundles. Hollow fibers were represented by an array of vertical micro‐rods (380 μm diameter) arranged with three packing densities (40%, 50%, and 60%) and two path lengths (2 and 4 cm). Blood was pumped through these devices corresponding to three mean blood flow velocities (16, 20, and 25 cm/min). Results showed that (1) clot formation decreases dramatically with decreasing packing density and increasing blood flow velocity, (2) clot formation at the outlet of the fiber bundle enhances deposition upstream, and consequently (3) greater path length provides greater clot‐free fiber surface area for gas exchange than a shorter path length. These results can help guide the design of less thrombogenic, more efficient artificial lung designs.
Keywords: artificial lung, clot formation, ECMO, fiber bundle, thrombosis
Translational Impact Statements.
Fiber bundle parameters, such as decreased packing density, increased blood flow velocity, and a longer path length, can be used to design a less thrombogenic, more efficient artificial lung to extend functionality.
1. INTRODUCTION
Thrombus formation at the blood–material interface of medical devices has been an ongoing concern for more than 60 years. 6 Of medical devices on the market, artificial lungs have one of the greatest blood contacting surface areas: ranging from 0.8 to 1.8 m2. The fiber bundles are densely packed for maximum gas exchanging efficiency, with blood flow external to the fibers and a mix of oxygen flowing within the fiber lumen. The blood‐contacting surfaces of the fiber bundles experience almost immediate protein adsorption, leading to direct activation of coagulation cascade factors and platelet adhesion and activation. The coagulation cascade and platelet activation, in turn, eventually will lead to the formation of solid clot within the fiber bundle, thereby reducing the gas exchange efficiency and increasing blood flow resistance. Furthermore, embolization of clots may result in thromboembolic occlusion of blood vessels within major organs. To prevent this, patients are prescribed a variety of systemic anticoagulant drugs, which in turn cause increased risk of bleeding and thus must be used with caution. Despite best efforts to mitigate thrombosis, typical clinical artificial lungs clot and fail within 1–3 weeks. 1 , 2 , 3 , 4 , 5 , 6 , 7 , 8
The fluid dynamic engineering of artificial lungs has long been used as a means to reduce clot formation by avoiding areas of blood stagnation or recirculation that are known to lead to accelerated clot formation. This design process has focused primarily on the inflow and outflow housings that collect and direct blood to and from the fiber bundle. There is currently no definitive relationship between the geometric design of the fiber bundle shape and fiber packing density and the speed of clot formation, despite this being the site of the most rapid clot formation. It is hypothesized that these fiber bundle parameters could be optimized to reduce thrombus formation and increase the useful lifetime of these devices.
There is also no current standard artificial lung fiber bundle geometry. Some fiber bundles are rectangular, such as the Maquet (Getinge) Quadrox, while others are cylindrical, such as the LivaNova EOS. Despite this, the shape and dimension of these fiber bundles can be simplified to the following parameters, namely: (1) the path length (L), the distance from the inlet face of the fiber bundle to the outlet in the direction of blood flow, (2) the frontal area (A f), the area of the fiber bundle perpendicular to direction of blood flow; and (3) the packing density, the percentage of the fiber bundle volume that is occupied by the hollow fibers. Commercially available oxygenator fiber bundles utilize a wide range of these parameters, and there is no consensus on what combination makes the least thrombogenic fiber bundle from clinical data or computational simulations. 9 , 10 , 11
In this experimental study, microfluidic flow chambers that mimic the local fluid mechanics within the fiber bundles of commercial oxygenators were constructed. The geometry of the chambers was varied to simulate a range of fiber bundle packing densities, path lengths, and frontal areas. 12 Freshly drawn, lightly heparinized, human whole blood was pumped through these microfluidic chambers, and clot formation was examined using confocal microscopy to map local platelet and fibrin deposition. Micro computed tomography (μCT) was employed to reconstruct the gross, 3D clot geometry.
2. MATERIALS AND METHODS
2.1. Experimental design of the flow chamber
To simulate the local conditions of an artificial lung hollow fiber bundle, miniature flow chambers were geometrically modeled using parametric computer‐aided design software (Solidworks, MA) to represent a small section of the fiber bundle. These flow chambers were 7.3 mm wide by 3 mm tall and had varied combinations of the physical fiber bundle parameters, packing density and path length. All fibers were simulated as 380 μm cylinders (rods) oriented perpendicular to the flow direction. These rods were arrayed evenly spaced to create packing densities of 40%, 50%, or 60% (Figure 1). These rods spanned the entire 3 mm height of the chamber to create blood flow patterns similar to the hollow fibers of an artificial lung. The length of the flow chamber was also varied to examine the effect of path length on clot formation. Lastly, the blood flow velocity was varied to simulate the effect of the fiber bundle frontal area on the local fluid mechanics: inasmuch as larger frontal area reduces average blood flow velocities. Thus, the blood flow velocities were set at 16, 20, and 25 cm/min to simulate, for example, 2 L/min of blood flow through a full sized adult artificial lung with 80, 100, and 120 cm2 frontal areas, or 4 L/min through frontal areas half that size.
Each chamber was 3D‐printed (Ember Autodesk, San Rafael, CA) using clear urethane acrylate resin (PR‐48, Colorado Polymer Solutions, CO) for ease of repeatability and uniformity. Uncured resin was rinsed out using 60 mL of isopropyl alcohol, post‐cured with a UV lamp for 20 min, and then sterilized using UV‐ozone for 15 min. The packing density, path length, and velocity of flow of the chambers were varied by perturbing a baseline design with 50% packing density, 20 cm/min blood velocity, and 2 cm path length (Figure 1). As described previously, the packing density was adjusted to either 40% or 60%; the velocity was adjusted to 16 and 25 cm/min; and the path length was tested at 2 and 4 cm.
2.2. Experimental evaluation of microchannel devices
The estimated, clot‐free, baseline resistance and Reynolds number was computed for each type of device using a glycerol–water mixture. During whole blood experiments, clot were observed to form instantaneously upon blood contacting the device surfaces. Thus, a clot‐free reference resistance was needed to effectively interpret the role of clot in increasing device resistance during whole blood studies. The glycerol–water mixture was titrated to a 3.4 cP viscosity using a viscometer (9721‐B56, Cannon‐Fenske, USA) and was pumped through at a rate of 3.65, 4.38, and 5.46 mL/min using a syringe pump (NE‐300, New Era Pump Systems, Inc., NY) to achieve an average velocity of 16, 20, and 25 cm/min through the simulated fiber bundle. Pressure drop was measured with a manometer (CR410, Ehdis, China), and the resistance was calculated using the standard formula:
in which R is resistance (mmHg*min/mL), ΔP is the pressure drop (mmHg) across the device, and Q was the flow rate (mL/min). Pressure measurements were made in triplicate for each flow rate and averaged. The Reynolds number was calculated using the standard formula:
in which ρ is density, u is velocity, D h is hydraulic diameter, and μ is dynamic viscosity.
Human blood was obtained from 18 consenting donors under an approved Institutional Review Board protocol, of whom nine were male and nine were female. Between 250 and 550 mL blood was drawn from each participant with an 18 gauge catheter, as per standard venipuncture procedures using a gravity drain. The blood was anticoagulated with 0.1 U/mL of heparin. After the full volume of blood had been collected in a sterile blood bag, it was transferred to 60 mL syringes for the experiment within 10 min after the blood draw. A syringe pump (NE‐300, New Era Pump Systems, Inc., NY) was used to infuse blood through each of the devices in a single pass through 1/16″ Tygon tubing (E‐3603, Saint‐Gobain, Malvern, PA), as seen in Figure 1. To minimize the effect of donor‐to‐donor variability, each experimental run was performed with all six configurations of flow chambers (three packing densities and two path lengths) utilizing blood from the same donor.
Differential pressure across the device was measured every 3 min over a total of 15 min using a manometer (CR410, Ehdis, China), and then used to calculate resistance using the equation above. Blood samples were collected at the inlet and outlet for measuring blood cell count (2 mL in K2EDTA) and aPTT/PT (3 mL in 1:9 citrate) using a clinical hematology analyzer (Diagnostica Stago Start 4, Siemens, Germany). 13 If aPTT was outside the range of 20–50 s, the experimental data was not used. After 15 min of flow, the pump was stopped, the device was detached from the circuit and immediately gently rinsed with heparinized saline (2 U/mL) to halt the progression of clot formation. The flow chamber was then filled with 4% (v/v) paraformaldehyde and fixed overnight at 4°C.
2.3. Micro‐computed tomography scans
Micro‐computed tomography (μCT) scans were performed to quantify the three dimensional clot deposition within the flow chambers. Prior to the μCT scan, the fixative inside the flow chambers was flushed using 10 mL of saline. Three flow chambers at a time were placed inside a custom‐built holder and then scanned using a Skyscan 1172 (Bruker‐Skyscan, Contich, Belgium) μCT system that has a 3 μm voxel resolution and the following conditions: 40 Vp, 175 μA, 300 ms exposure, 0.4° rotation step, 8 frame averaging. Serial scans were reconstructed into a 3D volume using NRECON (Bruker‐Skyscan, Contich, Belgium) with a 60% beam hardening correction and a ring artifact correction of 20. These volumes were then converted to ND2 files (FIJI, ImageJ 1.51H, NIH, Bethesda, MD) and imported into an open source software package (Slicer 3D 4.10.2) to be transformed, segmented, and cropped. The volume was cropped 0.4 mm from the top and bottom, 1.3 mm from each side, and 1 mm from the inlet side to avoid the edge effects induced by blood activation from the wall. The cropped volume was then imported as a DICOM image back into FIJI.
Thereafter, all of the images (n = 6) for each experimental condition were averaged to create a 3D map demonstrating the probability of clot formation at any voxel in the device. To isolate the clot volume from the background, three‐dimensional reconstructions were created from scans of four clean chambers for each packing density and path length for a total of four masks, one for each testing parameter. These masks were then subtracted from each individual experimental scan, so the volumes analyzed only consisted of clot (Figure 2).
Micro CT volumetric raw clot data was presented in three ways. Firstly, the resulting 3D volume was sliced transversely from the inlet to the outlet and the clot at each slice was extracted using a custom macro (Figure 2). Clot volume at five locations through the length of the chamber were summed for 2 mm segments and then graphed to quantify the percentage of the flow path occupied by thrombus in the chambers. These results were normalized based on the available volume for clot to form in each type of chamber using the formula:
As a second means of normalization to account for differences due to surface‐initiated blood activation, clot formation was measured at a specific distance from the inlet in each flow chamber. This distance was a function of the blood flow rate and packing density to find an equivalent blood‐contacting surface area. This equivalent distance allowed for the clot volume to be compared for a set of conditions that would provide similar gas exchange due to the same surface area. For example, for devices run at 16, 20, and 25 cm/min, clot volume was compared at 6.8, 5.7, and 4.5 mm from the inlet, respectively, where each device had encountered the same fiber surface area. For the 40%, 50%, and 60% packing density devices, clot was compared at 10.0, 5.7, and 3.2 mm from the inlet, respectively.
Thirdly, for ease of display in a 2D image presentation, the 3D volume reconstructions of clot were averaged along the longitudinal axis. This provided a summative view of the probability distribution of blood clot deposited in the flow chamber for further understanding of the spatial distribution of thrombus.
2.4. Immunofluorescent staining and imaging of fibrin and platelet deposition
Additional experiments were performed to obtain microscale visualization of the clot composition within the microchannel devices via platelet and fibrin fluorescent staining. The experiments were exactly the same as those in the previous section with the substitution of a glass coverslip on one side of the flow chamber for visualization.
In these experiments, packing densities of 40% and 60% and velocities of 16 and 25 cm/min were compared. After the 15‐min experiment, the flow chambers were rinsed with 10 mL of 2 U/mL heparin‐saline and fixed overnight in 4% (v/v) paraformaldehyde. This was followed by six rinses with 1× PBS. Blocking was conducted with 1% (w/v) bovine serum albumin and 5% (v/v) goat serum in PBS for 2 h at 4°C. For platelet and fibrin staining, 1:400 diluted CD61 stain (Tyr773 44‐876G, ThermoFisher, USA) and a mouse monoclonal anti‐human fibrin and fibrinogen (F9902, Sigma‐Aldrich, USA) were incubated inside the flow chambers overnight at 4°C. The following day, after six rinses with 1× PBS, the flow chambers were subsequently incubated with the secondary antibodies, goat anti‐rabbit IgG (H + L) Alexa Fluor 555 (A‐21428, ThermoFisher, USA) and goat anti‐mouse IgG1 CF633 (SAB4600335, Sigma‐Aldrich, USA), in a 1:250 ratio for 1 h at room temperature. These flow chambers were then rinsed with 1× PBS three times and imaged on a confocal microscope (A1 R+ HD25 Nikon) using a 16× NA water immersion objective.
2.5. Statistical analysis
All μCT clot volume and resistance values were analyzed for statistical significance using mixed model analysis with a Bonferroni corrected confidence interval using SPSS (IBM, Chicago, USA). For all tested parameters, clot volumes were compared across all flow chambers, with the blood donor ID as the subject variable. The independent variables included: velocity, packing density, and distance from the inlet. The dependent variable was the volume of clot as a function of distance. The aPTT was used as a covariate. For resistance analysis, the blood donor ID was used as a subject variable, the repeated‐measure variable was time, and the independent variable was the condition tested (e.g., velocity or packing density). The dependent variable was set to be ln(resistance) in the SPSS software and aPTT was set as a co‐variate. A one‐way ANOVA compared the clot volumes at the equivalent distance for the packing density and velocity groups. For path length groups, clot volume for a 1 mm thick segment was compared at 10 and 20 mm from the inlet for both devices. Post‐hoc tests were completed using Tukey's test to find the difference between groups. In all cases, a p‐value of <0.05 was regarded as significant.
3. RESULTS
3.1. Inlet and baseline conditions
The donor blood collected in this study was all within normal clinical ranges, albeit slightly on the lower end of normal for hematocrit, hemoglobin, and platelet counts (Table 1). The variation between groups was generally small.
TABLE 1.
aPTT (s) | PT (s) | HCT (%) | Hgb (g/dL) | PLT (k/μL) | |
---|---|---|---|---|---|
Flow rate | 36.9 ± 3.5 | 8.8 ± 0.3 | 38.8 ± 0.9 | 13.7 ± 0.3 | 237 ± 8 |
Path length | 36.0 ± 3.8 | 9.1 ± 0.3 | 39.5 ± 1.9 | 13.4 ± 0.7 | 275 ± 17 |
Porosity | 41.2 ± 3.8 | 8.4 ± 0.1 | 40.9 ± 1.6 | 13.8 ± 0.4 | 240 ± 21 |
Note: All results are within normal clinical ranges. N = 6 for each group.
Table 2 contains the baseline, glycerol–water resistances, and Reynolds numbers for each test condition. Resistance increased only slightly with packing density and was about 1.68 times greater for the 4 cm path length compared to the 2 cm path length. Blood flow resistances above these values were thus indicative of clot formation. The Reynolds numbers for each flow chamber and condition were calculated using a glycerol–water solution as well, and remained in the clinical range as seen in artificial lungs (Table 2).
TABLE 2.
2 cm | 4 cm | |||
---|---|---|---|---|
Packing density | 40% | 50% | 60% | 50% |
Resistance (mmHg/(mL/min)) | 0.333 ± 0.003 | 0.365 ± 0.008 | 0.379 ± 0.005 | 0.615 ± 0.010 |
Reynolds number | 97.4 | 77.9 | 64.9 | 77.9 |
Root mean square shear stress (Pa) | 0.31 | 0.34 | 0.37 | 0.32 |
3.2. Effects of fiber bundle design on clot formation
3.2.1. General patterns of clot formation
Immunofluorescent staining revealed greater fibrin deposition creating bridges between the fibers, rather than on the surface of individual fibers. On the other hand, platelet deposition was greater directly on the surface of the fibers than the space in between them. These surfaces were likely first covered with proteins, which acted as a nucleation surface for fibrinogen and platelet deposition. Fibrin networks formed in the direction of flow, and consequently there was less fibrin deposition on the upstream side of the hollow fibers, and more fibrin on the downstream side as seen in Figure 2b. Computational models performed by Zhang et al. have shown that stagnation points exist on both upstream and downstream center lines of the fibers, hence, these are likely sites of clot nucleation. 11
3.2.2. Effects of packing density
Both μCT (Figure 3) and fluorescence microscopy results (Figure 4) indicated greater formation of clot in the densely packed flow chambers than the loosely packed chambers. As seen in the probability maps, clot formed preferentially downstream, near the outlets of the channels (Figure 3a). The more tightly packed bundle had clot filling the proximal regions throughout the length of the chamber. The looser bundle had denser clot but within a smaller distal region, near the outlet. There was also noticeable capture of clot at the inlet of each device, possibly from clot embolizing from upstream locations. This observation was more prevalent in the 60% packing density flow chambers. The normalized clot volume was significantly greater at 60% packing density (p = 0.024) in the front half and middle of the flow chamber. (Figure 3b). This is reflected in the quantification of clot at locations in the chamber with equivalent surface areas (Figure 3c). In this case, when accounting for surface activated clot, the 60% packing density flow chamber had 3.78 times more clot than the 40% (p = 0.047). The latter device also had more clot than the 50% and 60% device at the outlet, but these data were not significant (p = 0.65 and p = 0.24, respectively).
Blood flow resistance measurements confirmed that the rate of clot formation, and thus blood flow occlusion, occurred more rapidly with increasing packing density (Figure 3d). Additionally, the rate of occlusion was found to accelerate at approximately the 8 min mark in the flow chambers with 50% and 60% packing density chambers, but not in those with 40% packing density. The blood flow resistance in the flow chambers with 60% packing density was found to be significantly greater than in those with 50% (p << 0.001) and 40% packing density (p << 0.001). The resistance of the flow chambers with 50% packing density was also significantly greater than in those with 40% packing density (p < 0.001).
Microscale observations revealed that platelet and fibrin adhesion was lower in the flow chambers with 40% packing density than in those with 60% packing density (Figure 4). Additionally, more fibrin strands formed perpendicular to flow in the flow chambers with 60% packing density, connecting adjacent hollow fibers.
3.2.3. Effects of blood flow velocity
Clot formation increased with lower blood flow velocity in the miniature flow chambers. The probability maps revealed more clot formed throughout the entire length of the slowest flow chamber than in the faster flow chamber as depicted in red and white (Figure 5a). Greater clot density was localized at the outlet of each flow chamber, regardless of velocity. The μCT clot volumes throughout the channels were significantly greater under the lowest flow than the greatest flow by 23.5 ± 1.1 mm3 (p = 0.014) (Figure 5b). At an equivalent distance within each chamber at different flowrates, the clot volume was quantified to account for the effect of surface activation of the coagulation cascade (Figure 5c). At these points, the fastest flow had 2.5 times less clot compared to the medium flow (p = 0.313) and slowest flow (p = 0.213).
Blood flow resistance also followed the μCT results and was significantly greater in the lowest velocity flow chamber compared to medium velocity (p < 0.01) and high velocity (p < 0.001) (Figure 5d). In the medium and high velocity flow chambers, there was a decrease in resistance from 3 to 10–12 min and then an increase in resistance thereafter compared to baseline (Table 2). These results indicate that a greater velocity delayed the initiation of clot formation in a hollow fiber bundle.
Confocal microscopy revealed that fibrin networks that formed under lower flow rates had thicker strands of fibrin that formed between the hollow fibers in each flow chamber. Additionally, the fibrin network density was also greater with more interconnections between the fibers. As seen in Figure 6, there are qualitatively more fibrin strands surrounding the rods in the low blood flow case at the inlet, middle, and outlet of the flow chamber.
3.2.4. Effects of path length
The longitudinal distribution of clot for both 2 and 4 cm flow chambers exhibited similar patterns, with the majority of thrombus accumulating at the outlet, while the center of the chamber remained clear. This observation may indicate that the outlet geometry is a major factor for clot deposition by creating areas of lower velocities that encourage the capture of clot and increased residence time of activated coagulation factors. Overall, a shorter path length of 2 cm resulted in a greater volume of deposited clot than in the 4 cm flow chamber. The clot probability map also shows that the shorter flow chamber had more condensed clot through the length of the flow chamber, whereas most of the flow chamber with a 4 cm path length was clear. The clot volume was greatest at the entrance, the first quarter, middle, third quarter, and the downstream end of the 2 cm flow chambers when compared to the 4 cm flow chambers (p = 0.08) (see Figure 7a).
Although the clot volume in the 2 cm flow chamber was greater, the blood flow resistance was not significantly different from the 4 cm flow chamber (p = 0.396, Figure 7b). The resistance increased at an increasing rate over 15 min for both flow chambers with either 2 or 4 cm path lengths in a similar fashion.
To account for the effect of surface area on the two path lengths, the clot volume was also compared at equal distance of 1 and 2 cm from the inlet for both flow chambers (see Figure 7c). Despite equivalent surface area wetted by blood, there was significantly more clot in the 2 cm flow chambers than in the 4 cm flow chambers at a 2 cm measurement (p = 0.004) but not significantly different at 1 cm and (p = 0.393).
4. DISCUSSION
The results from this study indicate that:
Clot formation in a microfluidic hollow fiber bundle decreases with decreasing packing density and increasing blood flow velocity,
Clot formation at the fiber bundle outlet can significantly increase upstream fiber bundle clot formation, and
A longer path length provides a greater proportion of clot‐free fiber surface area for gas exchange than a shorter path length by avoiding areas of clot formation at the outlet.
An explanation for decreased clot formation with lower packing density is that the increased distance between individual hollow fibers impedes the bridging of fibrin strands between adjacent fibers. Consequently, the less dense fibrin net reduces the capture of platelets, and by deduction red blood cells, although not measured in this study. Although fiber bundles that are less dense will have prolonged functionality due to lower clot burden, they would provide less oxygenation per total volume due to the lower surface area. Thus low‐density fiber bundles would need to be designed with larger path lengths and/or frontal areas to provide equivalent surface area for gas exchange. Of these two choices, greater frontal area is the least attractive as it reduces blood flow velocity, which increases clot formation. However, the greater path length will also increase resistance, which may induce greater shear stress and in turn may exacerbate platelet activation. 14
Slower blood velocities through the flow chambers resulted in more clot deposition, as expected due to the lower rates of convection and shear cleaning near the biomaterial surface. 15 , 16 The intrinsic pathway of the coagulation cascade was activated at the blood–material interface. With slower blood flow, the flow chamber would more rapidly accumulate procoagulant factors within the densely‐packed, high surface area of the fiber bundle. 17 Lower velocities may also result in lower shear stresses and will reduce surface washing or enzymatic degradation of a forming clot. In this study, calculated shear stresses were comparable between the studied chambers, and the magnitude was not large enough to provoke shear‐induced activation of platelets (Table 2). 18 As a general rule, low fiber bundle blood flow velocities should be avoided. However, shear stresses above 15 dyne/cm2 should also be avoided.
Of note, the majority of clot deposition was consistently found at the outlet of the flow chambers, close to the exit collector, and to a lesser extent at the inlet for all test conditions. This is similar to what is observed clinically. 19 Artificial lung devices are known to frequently capture thromboemboli at the fiber bundle inlet that originate from procoagulant upstream locations, such as small stagnation regions at the circuit connectors or shear induced platelet aggregates from mechanical pumps. Artificial lungs used for several days to weeks also often have more extensive, occlusive thrombus within their outlet housing due to (1) accumulation of activated clotting factors and activated platelets as blood flows through the fiber bundle and (2) regions of stagnation or recirculation in those regions. In this study, procoagulants almost certainly accumulated due to contact system activation, as demonstrated by progressive fibrin formation from inlet to outlet (Figure 4). Additionally, the blood velocity increases when moving from the wider flow chamber to the 1/16″ barb connection at the outlet and could have led to regions of turbulence and velocity gradients. This has been confirmed with computational simulations of blood flow in this geometry. 20 These results suggest that design of the outlet housing is a neglected factor that may mitigate or exacerbate the extent of clot formation.
The path length of the flow chambers affected the formation of clot to a lesser extent than the other parameters studied. For both the 2 and 4 cm flow chambers, a similar volume of clot formation was deposited at the outlet after 15 min, leading to a comparable increase in blood flow resistance. This suggests that the outlet housing was the dominant cause of clot formation in these devices rather than the path length. In the shorter flow chambers, this clot occluded a greater percentage of the entire flow chamber, which would render the oxygenator less efficient. Therefore, for full‐scale artificial lungs, a longer path length could potentially protect a greater proportion of fibers from occlusive clot forming at the outlet and retain more gas exchange functionality. However, a fiber bundle with a longer path length will have greater resistance to blood flow. This, in turn would require greater pump speed to generate the required pressure in an ECMO circuit, leading to increased shear activation of platelets. Thus, it is advisable to carefully design the fiber bundle to avoid stagnation and recirculation and potentially offset the fiber bundle at a distance from the walls of the outlet housing, to avoid clot growing retrograde into the fiber bundle.
Limitations of this study include the use of a miniature flow chamber that was 3D printed with a urethane acrylate—unlike clinical grade artificial lungs that typically employ polymethylpentene gas exchange fibers. Both materials have hydrophobic properties with comparable surface energies and protein deposition rates. 21 , 22 , 23 , 24 , 25 , 26 While the material selection may change the extent of protein adsorption and the rate of clot progression, it is not likely to affect the general mechanisms of clot formation caused by all hydrophobic polymer surfaces. This artifact was further mitigated in this study by performing paired comparisons of the independent variables: packing density, length, and velocity. The relatively short duration of these experiments could be considered a weakness. However, 15 min is sufficient to capture the activation of the contact coagulation pathway, in which proteins adsorb within milliseconds followed by the cascade of activation from Factor XII to cleaving of fibrinogen and insoluble clot. Additionally, the low level of anticoagulation, corresponding to activated partial thromboplastin time (aPTT) between 20 and 50 s, accelerates the thrombosis that would be observed clinically, in which aPTT ranges between 60 and 80 s. Past studies have underscored that early clot formation in full scale artificial lungs are often indicative of later device failure. 27 Despite this study capturing the first 15‐min of blood flow, we still believe a procoagulant nature on the order of minutes will be predictive of a more procoagulant nature on the order of hours or days. Therefore, the experiments provided sufficient signal with which to study the relative effects of the key independent variables. An additional physical limitation was the 3 mm height of these flow chambers, far smaller than the thickness of a full‐sized fiber bundle. This increased edge effects but was necessary to quantify clot inside the chamber with different imaging modalities and was mitigated by eliminating areas near the edges from analysis.
Additional artifacts could have been generated due to the artificial surfaces of blood wetted components of the circuit upstream of the flow chamber. However, the same circuit was used in each of the studies to minimize this effect. Nevertheless it would be highly advantageous to confirm the results of this in vitro study with full‐scale artificial lungs having different fiber bundle and outlet housing designs in a large animal model, in which anticoagulation could be titrated to typical of ECMO conditions.
5. CONCLUSIONS
Multi‐scale analysis of clot formation in microchannel devices indicated that fiber bundle clot formation is reduced when there is (1) lower fiber bundle packing density, (2) greater blood flow velocity, and (3) a less procoagulant outlet housing. These factors reduce fibrin mesh formation, platelet deposition, and bulk clot, particularly at the outlet of the fiber bundle.
AUTHOR CONTRIBUTIONS
Angela Lai: Data curation; formal analysis; investigation; project administration; writing – original draft. Natsuha Omori: Investigation; methodology; writing – review and editing. Julia E. Napolitano: Formal analysis; investigation; methodology; writing – review and editing. James F. Antaki: Conceptualization; funding acquisition; project administration; resources; supervision; writing – review and editing. Keith E. Cook: Conceptualization; investigation; project administration; supervision; writing – review and editing.
CONFLICT OF INTEREST STATEMENT
The authors declare no conflict of interest.
PEER REVIEW
The peer review history for this article is available at https://www.webofscience.com/api/gateway/wos/peer‐review/10.1002/btm2.10699.
Supporting information
ACKNOWLEDGEMENTS
This study was funded by National Institutes of Health grant R01HL089456.
Lai A, Omori N, Napolitano JE, Antaki JF, Cook KE. Effect of artificial lung fiber bundle geometric design on micro‐ and macro‐scale clot formation. Bioeng Transl Med. 2024;9(6):e10699. doi: 10.1002/btm2.10699
DATA AVAILABILITY STATEMENT
The data that support the findings of this study are available from the corresponding author upon reasonable request.
REFERENCES
- 1. Hohlfelder B, Szumita PM, Lagambina S, Weinhouse G, Degrado JR. Safety of propofol for oxygenator exchange in extracorporeal membrane oxygenation. ASAIO J. 2017;63(2):179‐184. [DOI] [PubMed] [Google Scholar]
- 2. Philipp A, De Somer F, Foltan M, et al. Life span of different extracorporeal membrane systems for severe respiratory failure in the clinical practice. PLoS One. 2018;13(6):e0198392. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3. Haneya A, Philipp A, Mueller T, et al. Extracorporeal circulatory systems as a bridge to lung transplantation at remote transplant centers. Ann Thorac Surg. 2011;91(1):250‐255. [DOI] [PubMed] [Google Scholar]
- 4. Fischer S, Simon AR, Welte T, et al. Bridge to lung transplantation with the novel pumpless interventional lung assist device NovaLung. J Thorac Cardiovasc Surg. 2006;131(3):719‐723. [DOI] [PubMed] [Google Scholar]
- 5. Strueber M, Hoeper MM, Fischer S, et al. Bridge to thoracic organ transplantation in patients with pulmonary arterial hypertension using a pumpless lung assist device. Am J Transplant. 2009;9(4):853‐857. [DOI] [PubMed] [Google Scholar]
- 6. Camboni D, Philipp A, Arlt M, Pfeiffer M, Hilker M, Schmid C. First experience with a paracorporeal artificial lung in humans. ASAIO J. 2009;55(3):304‐306. [DOI] [PubMed] [Google Scholar]
- 7. Maul TM. ECMO Anticoagulation: It's Still the Biggest Challenge Paper presented at: ASAIO 61st Annual Conference 2015; Chicago, IL.
- 8. Demarest CT, Shoemaker SJ, Salna MP, et al. The time course of clinical oxygenator failure due to clot formation. Medrxiv. 2020. doi: 10.1101/2020.11.20.20235606 [DOI]
- 9. Tabesh H, Amoabediny G, Poorkhalil A, Khachab A, Kashefi A, Mottaghy K. A theoretical model for evaluation of the design of a hollow‐fiber membrane oxygenator. J Artif Organs. 2012;15:347‐356. [DOI] [PubMed] [Google Scholar]
- 10. Mazaheri AR, Ahmadi G. Uniformity of the fluid flow velocities within hollow fiber membranes of blood oxygenation devices. Artif Organs. 2005;30:10‐15. [DOI] [PubMed] [Google Scholar]
- 11. Zhang J, Chen X, Ding J, et al. Computational study of the blood flow in three types of 3D hollow fiber membrane bundles. J Biomech Eng. 2013;135(12):121009. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12. Maul TM, Nelson JS, Wearden PD. Paracorporeal lung devices: thinking outside the box. Front Pediatr. 2018;2018:6. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13. Gosselin R, Grant RP, Adcock DM. Comparison of the effect of the anti‐Xa direct oral anticoagulants apixaban, edoxaban, and rivaroxaban on coagulation assays. Int J Lab Hematol. 2016;38:505‐513. [DOI] [PubMed] [Google Scholar]
- 14. Chow TW, Hellums JD, Moake JL, Kroll MH. Shear stress‐induced von willebrand factor binding to platelet glycoprotein Ib initiates calcium influx associated with aggregation. Blood. 1992;80(1):113‐120. [PubMed] [Google Scholar]
- 15. Wolberg AS, Aleman MM, Leiderman K, Machlus KR. Procoagulant activity in hemostasis and thrombosis: Virchow's triad revisited. Cardiovasc Anesthesiol. 2012;114(2):275‐285. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16. Wu W‐T, Aubry N, Massoudi M, Kim J, Antaki JF. A numerical study of blood flow using mixture theory. Int J Eng Sci. 2014;76:56‐72. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17. Jaffer IH, Fredenburgh JC, Hirsh J, Weitz JI. Medical device‐induced thrombosis: what causes it and how can we prevent it? J Thromb Haemost. 2015;13(Suppl 1):S72‐S81. [DOI] [PubMed] [Google Scholar]
- 18. Cook KEM, Maxhimer J, Leonard DJ, Mavroudis C, Backer CL, Mockros LF. Platelet and leukocyte activation and design consequences for thoracic artificial lungs. ASIAO J. 2002;48(6):620‐630. [DOI] [PubMed] [Google Scholar]
- 19. Conway RG, Zhang J, Jeudy J, et al. Computed tomography angiography as an adjunct to computational fluid dynamics for prediction of oxygenator thrombus formation. Perfusion. 2020;36:285‐292. [DOI] [PubMed] [Google Scholar]
- 20. Méndez Rojano R, Lai A, Zhussupbekov M, Burgreen GW, Cook K, Antaki JF. A fibrin enhanced thrombosis model for medical devices operating at low shear regimes or large surface areas. PLoS Comput Biol. 2022;18(10):e1010277. doi: 10.1371/journal.pcbi.1010277 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21. Lin YH, Liao KH, Chou NK, Wang SS, Chu SH, Hsieh KH. UV‐curable low‐surface‐energy fluorinated poly(urethane‐acrylate)s for biomedical applications. Eur Polym J. 2008;44:2927‐2937. [Google Scholar]
- 22. Brunstedt MR, Ziats NP, Robertson SP, Hiltner A, Anderson JM. Protein adsorption to poly(ether urethane ureas) modified with acrylate and methacrylate polymer and copolymer additives. J Biomed Mater Res. 1993;27:367‐377. [DOI] [PubMed] [Google Scholar]
- 23. Brunstedt MR, Ziats NP, Schubert M, et al. Protein adsorption onto poly(ether urethane ureas) containing methacrol 2138F: a surface‐active amphiphilic additive. J Biomed Mater Res. 1993;27:255‐267. [DOI] [PubMed] [Google Scholar]
- 24. Chauvel‐Lebert DJ, Auroy P, Bonnaure‐Mallet M. Biocompatibility of elastomers. Polymeric biomaterials. CRC Press; 2001. [DOI] [PubMed] [Google Scholar]
- 25. Xu Y, Dong A, Zhao Y, et al. Synthesis, characterization and biomedical properties of UV‐cured polyurethane acrylates containing a phosphorylcholine structure. J Biomater Sci. 2012;23(16):2089‐2104. [DOI] [PubMed] [Google Scholar]
- 26. Zhang Y, Zhang M, Xu X, et al. Anti‐fouling surfaces for extracorporeal membrane oxygenation by surface grafting of hydrophilic sulfoxide polymers. Biomacromolecules. 2022;23:4218‐4326. [DOI] [PubMed] [Google Scholar]
- 27. Skoog DJ, Pohlman JR, Demos DS, et al. Fourteen day in vivo testing of a compliant thoracic artificial lung. ASAIO J. 2017;63(5):644‐649. [DOI] [PMC free article] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
Data Availability Statement
The data that support the findings of this study are available from the corresponding author upon reasonable request.