Skip to main content
Cartilage logoLink to Cartilage
. 2024 Mar 1:19476035241234315. Online ahead of print. doi: 10.1177/19476035241234315

Biomechanical Comparison of All-Suture, All-Inside Meniscus Repair Devices in a Human Cadaveric Meniscus Model

Patrick A Massey 1,, Wayne Scalisi 1, Carver Montgomery 1, Drayton Daily 1, James Robinson 1, Giovanni F Solitro 1
PMCID: PMC11569520  PMID: 38426452

Abstract

Objective

Newer all-suture, all-inside meniscus repair devices utilize soft suture anchors. The purpose of this study was to compare the biomechanical performance of 4 meniscus repair devices in human cadaver menisci: the JuggerStitch (all-suture, all-inside), the FiberStitch (all-suture, all-inside), a polyether ether ketone (PEEK) all-inside, and an inside-out device.

Design

Forty human cadaver menisci were tested after creating 20 mm longitudinal tears in the posterior meniscus. Each knee was randomized to 1 of 4 meniscus repair groups: JuggerStitch (all-suture, all-inside), FiberStitch (all-suture, all-inside), FAST-FIX 360 (PEEK-based anchor all-inside), and inside-out (with BroadbandTM tape meniscus needles). For each meniscus, 2 devices were used to prepare vertical mattress repair construct. The specimens were tested by pre-conditioning 20 cycles between 5 N and 30 N and then the tear diastasis was measured, followed by distraction to failure phase after imposing a displacement at a rate of 0.5 mm/s.

Results

Ten menisci were tested in each of the 4 groups. After pre-conditioning, there was no significant difference in the gap formation among groups (P = 0.212). The average failure load for the JuggerStitch, FiberStitch, PEEK all-inside, and inside-out was 384 N, 311 N, 207 N, and 261 N, respectively, with a significant difference between groups (P = 0.034). Post hoc analysis showed the JuggerStitch failure load was higher than the PEEK all-inside and inside-out (P = 0.005, and P = 0.045, respectively). There was no significant difference between the failure load of the JuggerStitch and FiberStitch (P = 0.225).

Conclusion

The JuggerStitch all-suture device, FiberStitch all-suture device, PEEK all-inside, and inside-out devices have similar biomechanical properties for gapping and stiffness. The JuggerStitch all-suture, all-inside device has superior failure load compared with the PEEK all-inside and inside-out repair for longitudinal meniscus tear repair.

Keywords: biomechanics, general, knee meniscus, repair, all-inside

Introduction

Recent studies have highlighted the importance of meniscal preservation in the treatment of meniscal tears.1-6 The menisci play an important role in dynamic load distribution, shock absorption, joint congruity, joint lubrication, and proprioception.7-9 Assuming successful healing, meniscal repair offers preservation of the above-mentioned functions of the menisci. Degenerative changes in the knee have been shown to develop following meniscectomy.10,11 As a result, many advances have been made in meniscus repair techniques and technology.3,12

Many techniques have been developed for meniscus repair, including inside-out, outside-in, and all-inside techniques. The inside-out technique has long been considered the gold standard of meniscus repair. 12 However, the all-inside all-arthroscopic technique avoids the technical demands and morbidity associated with open surgery required of inside-out and outside-in techniques. 13 In addition, multiple studies have found no difference in clinical healing between all-inside versus inside-out repairs.12,14-16 Significant advances have been made for all-inside meniscus repair devices leading to increasing indications for meniscal repair.17-19 All-inside devices with PEEK (polyether ether ketone) anchors have been shown to have similar biomechanical properties and clinical healing to inside-out repairs. 12 While many all-inside devices utilize PEEK-based anchors, a recent study demonstrated increased meniscal cyst formation with all-inside PEEK anchor devices compared with inside-out repairs. 20 Newer all-suture, all-inside anchors have subsequently been developed. Unlike PEEK anchor devices, these newer all-suture devices utilize soft anchors made of suture material. These newer all-suture, all-inside devices are deployed posterior to the capsule and expand when the repair is tensioned.

The purpose of this study was to compare the biomechanical performance of 4 meniscus repair devices in human cadaver menisci: the JuggerStitch (all-suture, all-inside), the FiberStitch (all-suture, all-inside), a PEEK all-inside, and an inside-out device. The hypothesis was that the biomechanical properties would be similar among all-suture, all-inside devices, a current PEEK anchor all-inside meniscal device, and inside-out meniscus repair in response to cyclic loading and load to failure testing.

Materials and Methods

The Institutional Review Board (IRB) approved this study. After IRB review, 24 human cadaver knees were dissected leaving only the tibia, the posterior capsule, and the medial and lateral menisci with roots still attached to the tibia. The menisci were not removed from the tibia during the biomechanical testing performance. Cadaver knees were excluded from the study if they had a previously existing meniscus tear. Four knees were found to have a meniscus tear. This yielded 40 menisci total (20 medial, 20 lateral). The following meniscal dimensions were obtained: length, width, mid body width, and thickness. The average age of the cadavers was 67 years old ± 13 (SD) (range, 23-81). The weight and side of each cadaver knee was recorded. Posterior horn longitudinal full-thickness cuts were simulated 4 mm from the capsule, 20 mm in length.

Each knee was designated to 1 of 4 meniscus repair groups: FiberStitch all-inside (Arthrex, Naples, FL), JuggerStitch all-inside (Zimmer-Biomet, Warsaw, IN), PEEK all-inside (FAST-FIX 360, Smith and Nephew, Andover, MA), and inside-out with Broadband tape (Zimmer-Biomet, Warsaw, IN) meniscus needles. The FiberStitch is constructed of a 2-0 coreless FiberWire (polyester and ultra-high-molecular-weight polyethylene [UHMWPE]) suture with a pre-tied knot, anchored by 2 suture sleeves, that are deployed by a 1.8 mm, 12° up curve hollow needle. The curved JuggerStitch uses 2 soft polyester suture sleeve anchors connected by a 2-0 UHMWPE adjustable locking suture that is deployed with a meniscal repair device that uses a 1.6-mm hollow needle with a beveled tip. The FAST-FIX 360 curved meniscal repair device deploys 2 PEEK anchors (1 mm × 5 mm and 1.5 × 5 mm) that are adjoined by number 2-0 braided UHMWPE suture that has a pre-tied locking and sliding knot. These PEEK anchors are deployed through a hollow 1.5-mm needle which has a beveled tip. The Broadband tape meniscus needles were made of a 30″ tape (1.0 mm) that transitions to 2-0 suture connected to 10″ long 0.6-mm solid needles.

Both the FiberStitch and JuggerStitch are all-inside devices that are anchored with a polyester sleeve which expands when the suture is tensioned. For each meniscus, 2 longitudinal repairs were performed with 10-mm horizontal spacing between the repairs. The 2 separate repairs were performed with separate devices composed of the same implants for each group. For each all-inside repair, the first anchor was inserted in the inner meniscus to a depth of 18 mm and anchor deployed behind the capsule, then the second anchor was placed into the periphery close to the menisco-capsular junction to a depth of 18 mm and second anchor deployed behind the meniscus rim and capsule as well. The inside-out repairs were performed similarly, in a vertical mattress configuration, with the knots tied on the capsular side with a surgeon’s knot and 6 half hitches. All knots were reduced and tied by a single, board-certified orthopedic surgeon who completed a fellowship in sports medicine (P.A.M.). The tension to cinch all of the knots for repair was measured using a dynamometer (Mark-10, Copiague, NY) and recorded. This was done by attaching a dynamometer to the all-inside meniscus devices single suture stand for final tensioning. For the inside-out, this was done by attaching the dynameter to the post, when tying a surgeon’s knot. Following a previously developed protocol, 21 polyester fiber tapes (Mersilene, Ethicon, New Brunswick, NJ) 5 mm wide were placed in the center of the cut with one posterior loop and one anterior loop to load the meniscus after repair.

Mechanical testing was performed with a custom-built attachment made for the Instron 8874 biaxial servohydraulic fatigue testing system (Instron, Norwood, MA), with linear-torsion testing of ±25 kN and ±100 Nm of torque capacity. The attachment converted the axial displacement of the Instron actuator in the symmetrical distraction of the meniscal cut along the medial/lateral axis. In order to obtain a pure distraction of the cut, the tibia was mounted on an X-Y stage allowing free movement in the transverse plane and a custom apparatus pulled symmetrically on the Mersilene tapes (see Fig. 1 ).

Figure 1.

Figure 1.

Testing apparatus. Left tibia medial meniscus with a simulated 2 cm longitudinal meniscus tear is mounted on the x-y stage that allowed free movement while the longitudinal meniscus tear is symmetrically distracted by the custom-made fixture mounted on the Instron 8872 frame. The x-y stage allows for 2 degrees of freedom. The blue arrows represent the symmetric pull of the Mersilene tape by the Instron machine.

A baseline load of 10 N was applied with the apparatus and baseline gapping was measured. Measurements were obtained including gapping at baseline, gapping after preconditioning, and the failure load. Following a previously published protocol, 1 the specimens were pre-conditioned imposing initially a sinusoidal load in amplitude ranging between 10 N and 30 N for 20 cycles and followed by a distraction to disruption. The amount of 20 cycles was chosen based on a previous biomechanical study on meniscus repair. 22 The cut gapping was measured before and after the cycling phase through the measure of the maximal distance in correspondence of the cuts measured an imaging program called ImageJ (Oracle, Redwood Shores, CA) on calibrated images acquired with a 12MP camera. Gapping distances were obtained averaging the distances across the longitudinal cuts in correspondence of the middle and the margins of the Mersilene tape, similar to a previously described method (see Fig. 2 ). 23

Figure 2.

Figure 2.

Gap measuring illustration. The gaps can be appreciated between the vertical mattress sutures as distraction forces are applied by the anterior and posterior Mersilene tapes. Gapping distances were obtained averaging the distances across the longitudinal tears in correspondence of the middle (distance from B to B') and the margins of the Mersilene tape (distance from A to A' and C to C', respectively).

The repair site distraction was performed imposing a displacement at a rate of 0.5 mm/s while acquiring load-displacement data at 100 Hz. During the distraction phase, peak load was measured, and failure load was defined as the sudden load reduction greater than 10%. Modes of failure were considered to be device pullout, suture breakage, suture pull-through, and suture slippage. They were evaluated and recorded immediately after experimentation. Stiffness of the construct was measured trough the angular coefficient of the linear regression fitting the linear portion of the load-displacement curve.

Statistical Analysis

For comparison of numerical data such as the dimensions of the meniscus, load to failure, and gapping, an analysis of variance (ANOVA) was performed. Statistical analysis was performed with SPSS version 29 (IBM, Armonk, NY). A significance level of 0.05 was set for all comparisons. After ANOVA, post hoc testing was performed with Fisher’s least significant difference test. For comparing categorical data such as the mode of failure, a chi-square was done.

In order to determine the number of specimens for each group, a priori power analysis was performed after a small sample of specimens was tested. After 5 samples were tested in each group, there was an average failure load of µ1 = 406 N, µ2 = 197, µ3 = 183, and µ4 = 288 (for the JuggerStitch, FiberStitch, PEEK all-inside, and inside-out groups, respectively), with a pooled SD of 132. For a 1 – β or power of 0.8, a sample size needed was calculated to be 7 for each group. To provide a buffer, it was determined that 10 menisci in each group were selected for a total of 40 menisci.

Results

After removing 4 knees from the study due to previous meniscus tears, 20 knees with 40 intact menisci were utilized. There were 5 medial meniscus and 5 lateral menisci tested in each group. When comparing the medial menisci length, width, mid body width, and thickness, there was no difference in dimensions across all 4 groups (P = 0.27, P = 0.14, P = 0.98, and P = 0.89, respectively, see Table 1 ). When comparing the lateral menisci length, width, mid body width, and thickness, there was no difference in dimensions across all 4 groups (P = 0.75, P = 0.69, P = 0.5, and P = 0.91, respectively, see Table 2 ). There was also no difference in the age, weight, or side among the 4 groups (P = 0.85, P = 0.9, and P = 0.66, respectively, see Table 3 ).

Table 1.

Medial Meniscus Dimensions of 4 Groups.

Group Length SDa Width SD Mid-Body Width SD Thickness SD
JuggerStitch 43.6 4.8 31.8 2.6 9.6 0.5 5.6 0.9
FiberStitch 45.2 3.2 34.2 2.8 9.4 1.5 5.8 0.4
FastFix360 40.6 2.6 30.8 2.3 9.4 2.4 5.6 1.1
InsideOut 43.6 3.4 30.8 2.2 9.8 1.9 5.4 0.5
P-value 0.27 0.14 0.98 0.89

SD = standard deviation.

Table 2.

Lateral Meniscus Dimensions of 4 Groups.

Group Length SDa Width SDa Mid-Body Width SD Thickness SD
JuggerStitch 36.8 4.6 29.6 1.7 10.8 0.8 5.8 1.6
FiberStitch 36.2 1.5 30.6 5.2 9.4 2.1 6.2 1.3
PEEK all-inside 35 1.2 28.2 0.4 9.4 1.9 5.6 1.1
InsideOut 35 4 29 3.4 10.8 2.7 5.8 1.3
P-value 0.75 0.69 0.5 0.91

PEEK = polyether ether ketone; SD = standard deviation.

Table 3.

Demographic Data of Cadaver Specimens.

Device Age (Years) Weight (kg) Side
L R
JuggerStitch 62.5 91.5 3 2
FiberStitch 70.0 82.6 2 3
PEEK all-inside 69.2 82.0 3 2
Inside-out 67.2 79.5 2 3
P value 0.85 0.90 0.66

PEEK = polyether ether ketone.

A post hoc analysis showed a non-significant small correlation between increasing age and increasing load to failure (r = 0.14, P = 0.413) and no correlation between age of cadaver and gapping after cycling (r = 0.04, P = 0.842). There was a difference in the force required to reduce the peripheral side implant (P = 0.001). Post hoc analysis showed the peripheral FiberStitch required a force to reduce it higher than the PEEK all-inside, inside-out, and JuggerStitch (P = 0.001, P < 0.001, and P = 0.009, respectively). There was also a difference in the force required to reduce the root-sided implants (the suture closer to the posterior root of the meniscus) among the 4 groups (P = 0.002). Post hoc testing showed that the root-sided FiberStitch required a higher force to reduce than the PEEK all-inside and inside-out (P = 0.033 and P < 0.001, respectively). The root-sided JuggerStitch had a different tension required compared to the inside-out (P = 0.018, see Table 4 ).

Table 4.

Force to Reduce Peripheral Side and Root Side Sutures of 4 Different Repair Devices.

Device Peripheral Side Sutures (N) SDa Root Side Sutures (N) SDa
JuggerStitch 20.5 4.5 22.5 5.6
FiberStitch 25.3 4.3 25.4 2.8
PEEK all-inside 19.1 4.1 21.5 4.4
Inside-out 18.4 2.3 18.1 1.8
P-value 0.001 0.002

PEEK = polyether ether ketone; SD = standard deviation.

After pre-conditioning the menisci, there was no significant difference in the gap formation among groups (P = 0.212, see Fig. 3 ).

Figure 3.

Figure 3.

Meniscus gapping of 4 different devices after cycling. Error bars represent the standard error for each of the 4 different devices. No statistically significant difference was found between the groups (P = 0.212). PEEK = polyether ether ketone–based all-inside device.

The average failure load for the JuggerStitch, FiberStitch, PEEK all-inside, and inside-out was 384 N, 311 N, 207 N, and 261 N, respectively, with a significant difference between groups (P = 0.034, see Fig. 4 ).

Figure 4.

Figure 4.

Failure load of 4 different repair devices. The error bars represent standard error for each of the 4 device’s failure load averages. A significant difference between the groups was found (P = 0.034). *Post hoc analysis showed the JuggerStitch failure load was higher than the PEEK all-inside and inside-out (P = 0.005 and P = 0.045, respectively). PEEK = polyether ether ketone–based all-inside device.

Post hoc analysis showed the JuggerStitch failure load was higher than the PEEK all-inside and inside-out (P = 0.005, and P = 0.045, respectively). The FiberStitch failure load was no different than the JuggerStitch, PEEK all-inside, or inside-out groups (P = 0.225, 0.089, 0.405, respectively). There was no difference with stiffness among the 4 groups (P = 0.468). The failure loads, stiffness, and gapping data are summarized in Table 5 . Fifty-eight percent of all specimens failed due to suture cutout. No difference in stiffness could be demonstrated among the 4 groups (P = 0.346, see Table 6 ).

Table 5.

Failure Data of 4 Devices.

Device Failure Load (N) SD Stiffness (N/mm) SD Average Gapping at Preload (mm) SD Average Gapping after Cycling (mm) SD
JuggerStitch 384.1 160.8 17.5 3.1 0.88 0.5 1.74 0.8
FiberStitch 310.7 131.2 18.3 5.3 1.59 1.2 3.22 2.6
PEEK all-inside 207 88.8 15.3 4.5 1.59 1 2.98 1.5
Inside-out 260.7 140.4 16.3 3.4 1.6 1 2.89 1.3

SD = standard deviation; PEEK = polyether ether ketone.

Table 6.

Failure Modes of 4 Devices.

Device Anchor Pullout, n Suture Breakage, n Suture Cutout, n Suture Slippage, n
JuggerStitch - 4 5 1
FiberStitch 1 1 8 -
PEEK all-inside - 6 4 -
Inside-out 1 2 6 1

PEEK = polyether ether ketone.

Post hoc power analysis for the failure load showed that with a sample size of 10 for each of the 4 groups and an effect size (η2) of 0.221, there was an estimated power of 0.742.

Discussion

All-inside meniscus repair devices and techniques have advanced significantly. The PEEK anchor all-inside meniscus repair devices have been shown to have biomechanical and clinical results comparable to the traditional inside-out techniques.3,12,24 This study compared the biomechanical properties of these newer all-suture anchored devices to a traditional PEEK anchor device and inside-out repair. The 2 all-suture all-inside devices had comparable biomechanical properties to each other. The JuggerStitch all-inside device has higher failure load than the PEEK all-inside and the inside-out device for longitudinal meniscus repair (P = 0.005 and P = 0.045, respectively). There was no difference in failure load between the 2 all-suture all-inside devices. No difference was found among the 4 devices with respect to gapping or stiffness.

These devices can be used to repair longitudinal meniscus tears without the need for open incisions to tie knots on the capsule. The next advancement of all-inside meniscus devices has led to the development of all-suture anchored devices. These devices utilize a suture sleeve anchor technology, with a suture material that expands after being deployed on the extra-articular side of the knee capsule. Inside-out meniscus repair for longitudinal meniscus tears has long been considered the gold standard for a reliable repair. 25 All-inside devices such as the PEEK all-inside were subsequently developed and have shown comparable results both clinically and biomechanically.3,12,24 A recent systematic review demonstrates no difference in failure rate and clinical outcomes between inside-out or all-inside repair. 12 All-suture anchors with similar technology have been utilized for repair of soft tissue such as the labrum and rotator cuff with promising clinical and biomechanical results.26,27 These all-suture anchors have shown high failure loads in bone and have the advantage of being placed in a smaller hole in the meniscus, then expanding after deployment. 26 The current study specifically focused on comparing the biomechanical data of 4 devices: the JuggerStitch, FiberStitch, PEEK all-inside, and inside-out repair.

Previous studies have evaluated the failure load, gapping, and stiffness of inside-out and all-inside devices with helpful comparisons.3,28,29 While many of these studies separated the whole meniscus from the capsule, this study preserved the capsule peripheral to the meniscus. Many of these devices are designed to anchor behind the capsule and these devices are typically deployed behind the capsule in the clinical setting. Barber et al. 3 compared the FAST-FIX to inside-out, with a failure load of 68.1 N versus 95.8 N, respectively. The authors showed the OrthoCord (Depuy Synthes Mitek, Raynham, MA) inside-out had a higher failure load than the FAST-FIX 360. Another study by Masoudi et al. 28 compared the FAST-FIX to the NovoStitch (Smith and Nephew, Andover, MA) and inside-out. They showed the FAST-FIX had a failure load of 82.4 N, while the inside-out had a failure load of 118.3 N and NovoStitch had a failure load of 111.4. There was a higher failure load for the inside-out and NovoStitch over the FAST-FIX 360. In addition, a study showed the inside-out and newer Omnispan (Depuy Synthes Mitek, Raynham, MA) had higher failure loads than the FAST-FIX 360. 29 Many of these previous studies were done with a single repair device in a longitudinal meniscus cut with the peripheral capsule removed, while this study utilized 2 parallel repairs for each group with an intact peripheral capsule. It was chosen to model this testing with preservation of the peripheral capsule in this manner to better simulate the actual anatomy of the knee and to preserve the attachments of the native meniscus as many times these repair devices are deployed posterior to the intact capsule. This may explain the failure load average of 207 N for the PEEK all-inside group which is approximately double these previous studies. In addition, the meniscus roots and capsule were kept providing more realistic testing of the meniscus.

Another study by Milchtiem et al. 1 utilized 2 repairs (parallel vertical and crossed vertical sutures) for each group similar to this present study and showed no difference between failure load of 2 all-inside PEEK anchor devices (FAST-FIX and Speed Cinch). Their study showed a failure load of 89.6 N and 71.9 N for the FAST-FIX and Speed Cinch, respectively. While comparing suture configuration is beyond the scope of this study, further research is warranted to compare these groups in the current study’s design. Ramappa et al. 30 also performed 2 longitudinal repairs on each meniscus. They compared the FAST-FIX to inside-out and an all-inside running suture PEEK anchor device called the Sequent. Ramappa et al. 30 demonstrate failure loads of 140 N for the FAST-FIX and 188 for the inside-out group, with a higher failure load for the inside-out group. While some previous studies demonstrated the inside-out had a higher failure load than the FAST-FIX 360, the JuggerStitch actually had a higher failure load than both the PEEK all-inside and inside-out.3,28,30 As many of the load failures occurred due to meniscus cutout, it may be hypothesized that the higher load to failure of the Juggerstich is, in part, due to its relatively minimally traumatic insertion within the meniscus. The soft polyester suture anchors and the tapered tip inserter needle may leave more robust surrounding meniscal tissue for the anchor to lodge in.

Previous studies have shown the FAST-FIX 360 repairs to have displacement of 1.4 to 11 mm, depending on the study.3,29,30 Rosso et al. 29 demonstrated no difference in gapping after 100,000 cycles among 2 all-inside groups, the FAST-FIX 360 and Omnispan all-inside device. This is similar to the present study, where a difference in gapping was not found among all 4 groups. This study, however, utilized a low number of cycles and further studies should be done on higher cycles to determine whether there is an increase in gapping. If a larger number of cycles leads to a difference in gapping between devices, this would be an important distinction to determine as it is likely that in vivo failure is related to gapping and not catastrophic failure.

Barber found a mean stiffness of 12.1 N/mm for the FAST-FIX, 16.4 N/mm for the AIR meniscal repair device (Stryker, Kalamazoo, MI), and 15.7 N/mm for OrthoCord all-inside. 3 Another study found a vertical FAST-FIX had a stiffness of 14.4 N/mm. 21 An additional study by Masoudi et al. 28 demonstrated similar stiffness between all-inside devices with the FAST-FIX and Inside-out measuring 13.8 N/mm and 14.0 N/mm, respectively. These previous studies have demonstrated stiffness for the inside-out group and PEEK anchor all-inside devices that is similar to the stiffness values found here. The stiffness found of 15.3 N/mm for the PEEK all-inside and 16.3 N/mm for the inside-out is similar to previous studies. For baseline consideration, Markis found that circumferential directed fibers of the meniscus had a varying tensile modulus between 100 and 300 MPa. 31

The mode of failure in this study for the PEEK all-inside was predominantly due to suture rupture, while the FiberStitch was mainly due to meniscus cutout. The JuggerStitch was almost split evenly among meniscus cutout and suture rupture. Previous studies have shown that suture breakage or rupture is the main mode of failure of the FAST-FIX.28,29 This is similar to the present study, where the suture potentially breaks at lower loads before it is able to cut through the meniscus. The JuggerStitch appears to have a lower rate of cutting through the meniscus than the FiberStitch (50% vs 80%). Multiple reasons may account for this observation including the high tensile strength of 2-0 Fiberwire and the taper tip needle design of the JuggerStitch device which imparts less damage to the meniscal tissue than a cutting needle tip design. The Juggerstich 1.6-mm needle inserter has a circumferential tissue disruption of 2.01 mm2 on insertion compared with the 2.54 mm2 circumference. There was no difference in failure load between the 2 all-suture all-inside occupied by the 1.8-mm needle inserter for the FiberStich. The 0.53 mm2 difference in needle circumference may impart additional tissue damage contributing to the increased rate of cutout in the FiberStitch group.

It should also be noted that the FiberStitch required more tension to reduce the implant. This is likely related to the internal design of the implants with more friction occurring with the FiberStitch during reduction of the tear. This extra tension was required on the suture to cause the suture to slide and reduce the tear. This is likely due to the tensioning mechanism of the FiberStitch device, which utilizes friction to tension the repair device. The increased tension needed for tightening this device is clinically significant for surgeon awareness in the operating room to provide tactile feedback and tension on the repair of the tear.

Limitations

One of the study’s main focuses was on failure load, which represents one of the first limitations. This study used 20 cycles (between 5 and 30 N) for pre-conditioning and assessed gapping while other studies have used up to 100,000 cycles (between 5 and 20 N). 29 The amount of 20 cycles was chosen based on a previous biomechanical study on meniscus repair. 22 It is possible there would have been a difference in gapping between groups had more cycles been used.

Another potential limitation was related to the force required to reduce each implant. The dynamometer that was used could have limited the assessment of tension. The FiberStitch required significantly greater tensioning to secure the repairs than the other devices. This could potentially affect the strength of the repair, resulting in falsely lower gapping measurements. It was clear, however, that with lower forces, the FiberStitch would not reduce the meniscus. The authors assumed that the lateral and medial menisci would behave in a similar manner when biomechanically tested. While this may or may not be true, it represents a limitation of the study.

Finally, this testing was completed on cadaveric knees which leaves out the true biomechanics of an in vitro knee. No compressive forces or forces from condylar rollback were placed on the menisci after repair. This study more focused on the force to failure of the implant with distraction of the menisci after repair. This model of comparison was utilized to better compare the different fixation methods in a controlled setting.

The study was performed at room temperature in a non-aqueous environment, representing another limitation. The cadaveric specimens used were older than the appropriate population and the testing load does not exactly replicate the possible mechanisms of meniscal repair failure.

Conclusion

The JuggerStitch all-suture device, FiberStitch all-suture device, PEEK all-inside, and inside-out devices have similar biomechanical properties for gapping and stiffness. The JuggerStitch all-suture, all-inside device has superior failure load compared with the PEEK all-inside and inside-out repair for longitudinal meniscus tear repair.

Footnotes

Acknowledgments and Funding: The authors would like to acknowledge Baraa Shihadeh for preparation of manuscript and Mr. Alan Ogden for assistance in mechanical testing.

The author(s) declared no potential conflicts of interest with respect to the research, authorship, and/or publication of this article.

Funding: The author(s) disclosed receipt of the following financial support for the research, authorship, and/or publication of this article: This study was supported by funding from Zimmer Biomet.

Ethical Approval: A determination of not human research was made by the institutional review board.

ORCID iD: Patrick A. Massey Inline graphic https://orcid.org/0000-0001-8637-0122

References

  • 1. Milchteim C, Branch EA, Maughon T, Hughey J, Anz AW. Biomechanical comparison of parallel and crossed suture repair for longitudinal meniscus tears. Orthop J Sports Med. 2016;4(4):640263. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2. Badlani JT, Borrero C, Golla S, Harner CD, Irrgang JJ. The effects of meniscus injury on the development of knee osteoarthritis: data from the osteoarthritis initiative. Am J Sports Med. 2013;41(6):1238-44. [DOI] [PubMed] [Google Scholar]
  • 3. Barber FA, Howard MS, Ashraf W, Spenciner DB. The biomechanical performance of the latest all-inside meniscal repair devices. Arthroscopy. 2020;36(12):3001-7. [DOI] [PubMed] [Google Scholar]
  • 4. Kanto R, Yamaguchi M, Sasaki K, Matsumoto A, Nakayama H, Yoshiya S. Second-look arthroscopic evaluations of meniscal repairs associated with anterior cruciate ligament reconstruction. Arthroscopy. 2019;35(10):2868-77. [DOI] [PubMed] [Google Scholar]
  • 5. Yoon KH, Park JY, Kwon YB, Lee YJ, Kim EJ, Kim SG. Inside-out repair of the meniscus in concomitant anterior cruciate ligament reconstruction: absorbable versus nonabsorbable sutures. Arthroscopy. 2020;36(4):1074-82. [DOI] [PubMed] [Google Scholar]
  • 6. Kim JG. Editorial commentary: save the meniscal root, why not. Arthroscopy. 2019;35(7):2207-10. [DOI] [PubMed] [Google Scholar]
  • 7. Moulton SG, Bhatia S, Civitarese DM, Frank RM, Dean CS, LaPrade RF. Surgical techniques and outcomes of repairing meniscal radial tears: a systematic review. Arthroscopy. 2016;32(9):1919-25. [DOI] [PubMed] [Google Scholar]
  • 8. Ahmed AM, Burke DL. In-vitro measurement of static pressure distribution in synovial joints—part I: tibial surface of the knee. J Biomech Eng. 1983;105(3):216-25. [DOI] [PubMed] [Google Scholar]
  • 9. Voloshin AS, Wosk J. Shock absorption of meniscectomized and painful knees: a comparative in vivo study. J Biomed Eng. 1983;5(2):157-61. [DOI] [PubMed] [Google Scholar]
  • 10. Allen PR, Denham RA, Swan AV. Late degenerative changes after meniscectomy. J Bone Joint Surg Br. 1984;66(5):666-71. [DOI] [PubMed] [Google Scholar]
  • 11. Hede A, Larsen E, Sandberg H. Partial versus total meniscectomy. A prospective, randomised study with long-term follow-up. J Bone Joint Surg Br. 1992;74(1):118-21. [DOI] [PubMed] [Google Scholar]
  • 12. Fillingham YA, Riboh JC, Erickson BJ, Bach BR, Jr, Yanke AB. Inside-out versus all-inside repair of isolated meniscal tears. Am J Sports Med. 2017;45(1):234-42. [DOI] [PubMed] [Google Scholar]
  • 13. Barber FA, Herbert MA, Schroeder FA, Aziz-Jacobo J, Sutker MJ. Biomechanical testing of new meniscal repair techniques containing ultra high-molecular weight polyethylene suture. Arthroscopy. 2009;25(9):959-67. [DOI] [PubMed] [Google Scholar]
  • 14. Grant JA, Wilde J, Miller BS, Bedi A. Comparison of inside-out and all-inside techniques for the repair of isolated meniscal tears: a systematic review. Am J Sports Med. 2012;40(2):459-68. [DOI] [PubMed] [Google Scholar]
  • 15. Choi NH, Kim BY, Hwang Bo BH, Victoroff BN. Suture versus FasT-Fix all-inside meniscus repair at time of anterior cruciate ligament reconstruction. Arthroscopy. 2014;30(10):1280-6. [DOI] [PubMed] [Google Scholar]
  • 16. Ahn JH, Kim CH, Lee SH. Repair of the posterior third of the meniscus during meniscus allograft transplantation: conventional inside-out repair versus FasT-Fix all-inside repair. Arthroscopy. 2016;32(2):295-305. [DOI] [PubMed] [Google Scholar]
  • 17. Chahla J, LaPrade RF. Meniscal root tears. Arthrosc: J Arthrosc Relat Surg. 2019;35(5):1304-5. [DOI] [PubMed] [Google Scholar]
  • 18. LaPrade RF. Editorial commentary: we know we need to fix knee meniscal radial root tears: but how best to perform the repairs. Arthroscopy. 2018;34(4):1069-71. [DOI] [PubMed] [Google Scholar]
  • 19. Barber FA, Click SD. Meniscus repair rehabilitation with concurrent anterior cruciate reconstruction. Arthroscopy. 1997;13(4):433-7. [DOI] [PubMed] [Google Scholar]
  • 20. Nishino K, Hashimoto Y, Nishida Y, Terai S, Takahashi S, Yamasaki S, et al. Incidence and risk factors for meniscal cyst after meniscal repair. Arthroscopy. 2019;35(4):1222-9. [DOI] [PubMed] [Google Scholar]
  • 21. Kocabey Y, Chang HC, Brand JC, Jr, Nawab A, Nyland J, Caborn DN. A biomechanical comparison of the FasT-Fix meniscal repair suture system and the RapidLoc device in cadaver meniscus. Arthroscopy. 2006;22(4):406-13. [DOI] [PubMed] [Google Scholar]
  • 22. Anz AW, Branch EA, Saliman JD. Biomechanical comparison of arthroscopic repair constructs for meniscal root tears. Am J Sports Med. 2014;42(11):2699-706. [DOI] [PubMed] [Google Scholar]
  • 23. Schneider CA, Rasband WS, Eliceiri KW. NIH image to imageJ: 25 years of image analysis. Nat Methods. 2012;9(7):671-5. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 24. Marchetti DC, Phelps BM, Dahl KD, Slette EL, Mikula JD, Dornan GJ, et al. A contact pressure analysis comparing an all-inside and inside-out surgical repair technique for bucket-handle medial meniscus tears. Arthroscopy. 2017;33(10):1840-8. [DOI] [PubMed] [Google Scholar]
  • 25. Rosso C, Kovtun K, Dow W, McKenzie B, Nazarian A, DeAngelis JP, et al. Comparison of all-inside meniscal repair devices with matched inside-out suture repair. Am J Sports Med. 2011;39(12):2634-9. [DOI] [PubMed] [Google Scholar]
  • 26. Erickson J, Chiarappa F, Haskel J, Rice J, Hyatt A, Monica J, et al. Biomechanical comparison of a first- and a second-generation all-soft suture glenoid anchor. Orthop J Sports Med. 2017;5(7):232596. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 27. Agrawal V, Pietrzak WS. Triple labrum tears repaired with the JuggerKnotTM soft anchor: technique and results. Int J Shoulder Surg. 2015;9(3):81-9. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 28. Masoudi A, Beamer BS, Harlow ER, Manoukian OS, Walley KC, Hertz B, et al. Biomechanical evaluation of an all-inside suture-based device for repairing longitudinal Meniscal Tears. Arthroscopy. 2015;31(3):428-34. [DOI] [PubMed] [Google Scholar]
  • 29. Rosso C, Müller S, Buckland DM, et al. All-inside meniscal repair devices compared with their matched inside-out vertical mattress suture repair: introducing 10,000 and 100,000 loading cycles. Am J Sports Med. 2014;42(9):2226-33. [DOI] [PubMed] [Google Scholar]
  • 30. Ramappa AJ, Chen A, Hertz B, Wexler M, Grimaldi Bournissaint L, DeAngelis JP. et al. A biomechanical evaluation of all-inside 2-stitch Meniscal repair devices with matched inside-out suture repair. Am J Sports Med. 2014;42(1):194-9. [DOI] [PubMed] [Google Scholar]
  • 31. Makris EA, Hadidi P, Athanasiou KA. The knee meniscus: structure–function, pathophysiology, current repair techniques, and prospects for regeneration. Biomaterials. 2011;32(30):7411-31. [DOI] [PMC free article] [PubMed] [Google Scholar]

Articles from Cartilage are provided here courtesy of SAGE Publications

RESOURCES