Abstract
Purpose: Assess the comparative accuracy and reliability of postoperative toric intraocular lens (TIOL) alignment measurement methods: Casia2 and Adobe Photoshop with digital slit lamp images (PS method).
Methods: In a study of 41 subjects with 58 eyes postimplantation of TIOL, we independently measured TIOL alignment with Casia2 and PS methods. The intraclass correlation coefficient (ICC1,1) was employed to assess the repeatability of the Casia2 method. While ICC2,1 for absolute agreement and Bland–Altman analysis were utilized to determine the interdevice agreement between the two methods, the regression analysis was conducted to identify any proportional bias.
Results: Casia2 demonstrated excellent intradevice repeatability with an ICC1,1 of 0.998. The absolute agreement between Casia2 and PS was very high with an ICC2,1 of 0.999. The average discrepancy between the two measurement methods was −0.828°, with a 95% confidence interval (CI) ranging from −1.623° to −0.032°. The 95% limits of agreement (LoA) were between −6.761° and 5.105°, indicating a strong concordance in TIOL alignment measurements. Casia2 was capable of accurately measuring the TIOL axis alignment under conditions of pupil diameters (PDs) of 4 mm or greater.
Conclusion: Casia2 and PS demonstrated significant concordance in measuring postoperative TIOL alignment, with Casia2 offering a more straightforward and accessible alternative, particularly beneficial for patients with suboptimal pupil dilation.
Keywords: Adobe Photoshop, alignment, Casia2, digital slit lamp retroillumination, interdevice agreement, pupil diameter, repeatability, toric intraocular lens
1. Introduction
Corneal astigmatism, which affects 34.8% of cataract patients, is a major refractive error [1]. One effective method for reducing preexisting astigmatism is the implantation of toric intraocular lens (TIOL) [2]. However, the toric lens is prone to rotation in the first week after cataract surgery, which can cause a decrease in the correction of astigmatism by 3.3% for every degree of off-axis misalignment [3]. If the misalignment exceeds 10°, surgical repositioning is recommended [4], ideally 2-3 weeks after IOL implantation, prior to the adhesion formation between the TIOL and the capsular bag [5, 6]. Therefore, early detection of IOL rotation is crucial and the method used for measuring postoperative TIOL axial alignment should be accurate and easy to perform.
Currently, the gold standard for evaluating postoperative axial alignment of TIOL is Adobe Photoshop with digital slit lamp images (PS method), which involves taking digital slit lamp retroillumination images combined with Adobe Photoshop software. This method takes into account the tilt of the patient's head and the cyclotorsion of the eye by referring to the episcleral vessel [7, 8]. However, the relative accuracy of the PS method is subject to the observer's experience and bias. In addition, pupils must be sufficiently dilated to visualize one or more toric marks on both sides. With a TIOL optic diameter of 6.0 mm and an overall diameter of 13.0 mm, the minimum pupillary diameter required is 6.5 mm [9].
The Casia2 imaging instrument (Tomey Corporation, Nagoya, Japan) is the second-generation swept-source Fourier-domain anterior segment optical coherence tomography (SS-ASOCT) that has been commercially available since 2015. Compared with the previous generation Casia SS-1000 OCT (Tomey Corporation, Nagoya, Japan), it offers several advantages, including faster scanning speed, higher resolution, and better penetration with a wavelength of 1310 nm, scanning speed of 50,000 A-scans per second, scanning depth of 13 mm, and scanning width of 16 mm. As a result, Casia2 can provide detailed parameters of the classic cornea, anterior chamber, pupil, and lens [10, 11]. In addition, Casia2 has been demonstrated good repeatability and reproducibility under both nonmydriatic and mydriatic conditions [12, 13]. Previous studies had demonstrated the excellent accuracy and simplicity of Casia2 for measuring the tilt and decentration of crystalline lens and IOL automatically [14, 15]. Moreover, the device has built-in software that allows for the measurement of the TIOL axis. However, to our knowledge, there are no reports on the application of Casia2 for measuring the TIOL axis to date.
Therefore, our study aims to introduce the Casia2 method for measuring the alignment of postoperative TIOL and assess the interdevice agreement of the TIOL alignment obtained from Casia2 and PS methods.
2. Materials and Methods
2.1. Subjects
This retrospective study was performed at the Department of Ophthalmology at Guangdong Provincial People's Hospital. All subjects received toric IOL implantation from January 2021 to December 2022, with models including the AcrySof IQ toric SN6AT IOL (Alcon) and the AT TORBI 709M IOL or 909M multifocal IOL (Carl Zeiss Meditec AG). The TORIC IOLs had an optic diameter of 6.0 mm for an overall length of 13 mm. Patients were excluded from the study if pupils could not be sufficiently dilated to visualize toric marks on both sides, or eyes could not focus on the fixation target during the measurement resulting in unsuccessful detection of the alignment of TIOL. This study was approved by the Ethics Committee of Guangdong Provincial People's Hospital and adhered to the tenets of the Declaration of Helsinki. Written informed consent was obtained from each subject before enrollment in this study.
Prior to cataract surgery, biometric measurements such as corneal curvature, corneal astigmatism, axial length, anterior chamber depth, and lens thickness were obtained using IOL Master 700 (Carl Zeiss Meditec, Germany). All postoperative subjects were scanned with both Casia2 and retroillumination slit lamp photography. Following the administration of mydriatic eye drops, the TIOL axis was first assessed using CASIA2 and subsequently using PS once the pupil dilated sufficiently. The measurements were conducted according to standardized specifications to ensure accuracy and reliability, with subjects positioned in a straight and upright position on the chin rest of the Casia2 or slit lamp devices. All measurements were performed by a single experienced examiner to reduce variability. To assess measurement repeatability, each subject underwent three repeated measurements in a dimly lit room.
2.2. Casia2 Method
TIOL were measured using the IOL scan mode, which captures eight distinct AS-OCT images from eight different angles (0–180, 90–270, 23–203, 113–293, 45–225, 135–315, 68–248, and 158–338). The outlines of the TIOL were automatically recognized and then were checked for accuracy by the same examiner, who selected the “Toric” mode and adjusted the IOL outlines as needed. Subsequently, the examiner pressed “Semi-Auto Trace Start” and then “Save” to generate 3D analyses of tilt, decentration, and toric axis relative to the corneal topographic axis (see Figure 1). To ensure consistency, each subject was measured three times by the same examiner using the Casia2 method. And the TIOL axis position was obtained by averaging the three repeated measurements.
2.3. PS Method
The PS method for measuring the TIOL alignment was based on an objective image analysis technique described by Shah [8]. In brief, high-resolution, slit lamp digital retroillumination photographs were captured with a sufficiently dilated pupil, which is 6.5 mm or greater. A grid ring was used to measure the TIOL axis, which could be zoomed in or out by rolling the mouse to adapt to the size of the measured objects. The geometric center of the grid ring was overlapped onto that of TIOL, and the measurement was performed in triplicate to ensure that the orientation (in degrees) and distance of the major episcleral vessel from the geometric center of the TIOL were consistent. TIOL axis positions were obtained by averaging the three repeated measurements, with deviation caused by head tilt and eyeball cyclotorsion minimized by referencing episcleral vessels (Figure 2). These measures were performed by an experienced examiner to ensure consistency and accuracy.
2.4. Statistical Analysis
Statistical analysis was performed using SPSS (version 26) and GraphPad Prism (version 9.0). Continuous variables were summarized with descriptive statistics (mean, standard deviation, median, and range), and categorical variables were expressed as counts and percentages. The intraclass correlation coefficient (ICC1,1) was employed to assess the repeatability of the Casia2 method [16, 17]. The interdevice agreement between Casia2 and PS methods was assessed with ICC2,1 for absolute agreement [17] and Bland–Altman analysis [18, 19]. ICCs were interpreted as poor (< 0.75), moderate (0.75 to < 0.9), and good (≥ 0.9) [17]. The 95% limits of agreement (LoA) were calculated from the mean difference ± 1.96 × SD according to the Bland–Altman method, with the Bland–Altman plot visually illustrating the agreement between the two methods [20]. Univariate and multivariate regression analyses were employed to identify the related factors for proportional bias, excluding sphere power and cylinder power of TIOL to avoid multicollinearity in the multivariate regression analysis. A p value < 0.05 was considered statistically significant.
3. Results
3.1. Demographic and Biometric Characteristics
Fifty-eight eyes of 41 subjects (20 males and 21 females) who had undergone toric IOL implantation were included in this study. The mean age of the subjects was 67.03 years (ranging from 12 to 88 years). Table 1 summarizes the preoperative biometry of the study population, including average keratometry (SE), corneal astigmatism (△K), axial length (AL), and anterior chamber depth (ACD), all which were measured using the IOL master700 (Carl Zeiss Meditec, Germany).
Table 1.
Characteristics | Value |
---|---|
Subjects, (n) | 41 |
Eyes, n (right/left) | 58 (24/34) |
Sex (male/female) | 20/21 |
Age, (y) | |
Mean ± SD | 67.03 ± 20.6 |
Median (range) | 75 (12, 88) |
SE (D) | |
Mean ± SD | 43.56 ± 1.58 |
Median (range) | 43.8 (40.32, 46.90) |
△K (D) | |
Mean ± SD | 2.18 ± 0.74 |
Median (range) | 2.06 (1.10, 5.03) |
AL (mm) | |
Mean ± SD | 24.91 ± 2.12 |
Median (range) | 24.17 (21.22, 30.85) |
ACD (mm) | |
Mean ± SD | 2.98 ± 0.48 |
Median (range) | 2.88 (1.99, 3.85) |
Note: △K: Corneal astigmatism; preoperative biometry was performed by IOL master 700 (Carl Zeiss Meditec, Germany).
Abbreviations: ACD, anterior chamber depth; AL, axis length; SE, average keratometry.
Bilateral eyes were included in 17 patients. To show that the TIOL axis of both eyes were intraindividually independent, we calculated the ICC2,1 between both eyes in the same patient for the measured method of PS and Caisa2 separately, showing no significant correlation between both eyes of the same patient, no matter by PS or Casia2 method (ICC2,1 casia2 = 0.094, PCasia2=0.338; ICC2,1 PS = 0.087, PPS=0.349).
3.2. TIOL Parameter Measurements and Intradevice Agreement
Table 2 shows the implanted TIOL parameters, including the models, sphere power, and cylinder power of the TIOL. Tilt, decentration, and pupil diameter (PD) were measured using Casia2 device, and the PD was that of the Casia2 method in the measurement of the TIOL axis. Casia2 demonstrated excellent intradevice repeatability with an ICC1,1 of 0.998 (range from 0.997 to 0.999).
Table 2.
Parameters | Value |
---|---|
Models | |
Zeiss 709, n (%) | 40 (69.0%) |
Zeiss 909, n (%) | 9 (15.5%) |
Alcon, n (%) | 9 (15.5%) |
Sphere power (D) | |
Mean ± SD | 16.32 ± 5.31 |
Median (range) | 18.25 (5, 26.5) |
Cylinder power (D) | |
Mean ± SD | 3.03 ± 0.84 |
Median (range) | 3.0 (1.5, 6) |
Tilt degree (°) | |
Mean ± SD | 4.92 ± 1.84 |
Median (range) | 4.90 (1.10, 8.70) |
Decentration (mm) | |
Mean ± SD | 0.22 ± 0.15 |
Median (range) | 0.19 (0.01, 0.70) |
PD (mm) | |
Mean ± SD | 5.69 ± 1.19 |
Median (range) | 5.79 (2.61, 8.80) |
Note: PD: pupil diameter of Casia2 method for the measurement of TIOL axis. PD, tilt and decentration were measured by Casia2 (Tomey, Japan).
3.3. Interdevice Agreement and Bland–Altman Analysis
The mean value ± standard deviation of the TIOL axis measured was 94.530 ± 69.838° by the Casia2 method and 95.360° ± 70.105° by the PS method, separately. The ICC2,1 for absolute agreement was 0.999 (ranging from 0.998 to 0.999), indicating good agreement between the two methods. The mean difference between the two methods (namely, Casia2 and PS) was −0.828°, with a 95% confidence interval (CI) ranging from −1.623° to −0.032°. The 95% LoA were between −6.761° and 5.105°. The Bland–Altman plot (Figure 3) visually illustrated a strong concordance between the two methods.
3.4. Agreement Under Different PDs
Table 3 presents the ICC2,1 values for assessing the agreement of TIOL axis measurements between Casia2 and PS methods under different PDs. The results showed that Casia2 was capable of accurately measuring the TIOL axis alignment under conditions of PDs of 4 mm or greater.
Table 3.
PD (mm) | Case (n, %) | Casia2 vs. PS | |
---|---|---|---|
ICC2,1 | 95% CI of ICC | ||
< 4 | 3 (5%) | 0.227 | −0.280, 0.955 |
4∼5 | 14 (24%) | 0.999 | 0.998, 1 |
5∼6 | 17 (29%) | 0.999 | 0.997, 1 |
6∼7 | 17 (29%) | 0.999 | 0.997, 1 |
> 7 | 7 (12%) | 1 | 0.998, 1 |
Note: PD: Pupil diameter of the Casia2 method for the measurement of the TIOL axis.
Abbreviations: CI, confidence interval; ICC, intraclass correlation coefficient.
3.5. Univariate and Multivariate Regression Analyses
Table 4 shows the results of univariate and multivariate regression analyses for the absolute difference (AD) as the dependent variable. The results indicated that sex, age, AL, ACD, SE, △K, models of IOL, PD, tilt, and decentration were not significantly associated with the AD. Together, the Casia2 device showed excellent intradevice repeatability and interdevice agreement with the PS method in measuring the TIOL axis. The Casia2 method was accurate for the measurement of the TIOL axis when the PD was 4 mm or greater. No significant factors were found to be associated with AD.
Table 4.
Variables | Univariate analysis | Multivariate analysis | |||
---|---|---|---|---|---|
Regression coefficient (95% CI) | R 2 | p | Regression coefficient (95% CI) | p | |
Sex | −0.098 (−1.150, 0.955) | 0.001 | 0.853 | −0.143 (−1.398, 1.112) | 0.819 |
Age (y) | 0.018 (−0.008, 0.043) | 0.034 | 0.167 | 0.004 (−0.039, 0.048) | 0.843 |
AL (mm) | −0.153 (−0.400, 0.094) | 0.027 | 0.220 | −0.288 (−0.759, 0.184) | 0.226 |
ACD (mm) | −0.529 (−1.633, 0.574) | 0.016 | 0.341 | 1.200 (−1.024, 3.424) | 0.283 |
SE (D) | 0.023 (−0.312, 0.359) | 0 | 0.889 | −0.029 (−0.461, 0.403) | 0.893 |
△K (D) | −0.406 (−1.120, 0.307) | 0.023 | 0.259 | −0.823 (−1.907, 0.262) | 0.134 |
Models of IOL | −0.252 (−0.952, 0.448) | 0.009 | 0.474 | −0.213 (−1.062, 0.635) | 0.615 |
PD (mm) | −0.239 (−0.681, 0.204) | 0.020 | 0.284 | −0.130 (−0.635, 0.375) | 0.606 |
Tilt (degree) | 0.100 (−0.188, 0.388) | 0.009 | 0.488 | 0.217 (−0.168, 0.602) | 0.262 |
Decentration (mm) | −0.895 (−4.500, 2.711) | 0.004 | 0.621 | −0.444 (−4.855, 3.966) | 0.840 |
Note: AD: the absolute difference of the TIOL axis measured by Casia2 and PS, (|Casia2 − PS|).
4. Discussion
As patients increasingly expect optimal postoperative refractive outcomes with minimal astigmatism after cataract surgery, the application of TIOL is more and more common. Measurement of the postoperative TIOL alignment is essential to evaluate the rotational stability of TIOL and guide the repositioning to achieve satisfactory postoperative outcomes.
For the first time, our study introduced the Casia2 method applied for the measurement of the postoperative TIOL axis and evaluated the agreement of measured values between Casia2 and PS for the postoperative TIOL alignment.
In our study, Casia2 showed good reproducibility (ICC1,1 0.998). Previous literature studies also reported good reproducibility with Casia2 in measuring anterior segment parameters (ICC 0.86–0.99) [21, 22]. The reproducibility in our study was superior to those reported in the literature studies that were probably attributed to our operation performed by the same trained operator following a standardized procedure and the result only for the postoperative TIOL axis.
Our study demonstrated that Casia2 and PS had good agreement for the measurement of postoperative TIOL alignment with the ICC analysis (ICC2,1 0.999) and Bland–Altman plot (mean difference −0.828°). Although the 95% LoA seem wide (−6.761 to 5.105°), rotations less than 10° generally result in less than 0.50D of astigmatism, which may be no clinically significant [23]. Therefore, the measured discrepancy from the two methods could be within permissible limits.
In terms of operation, one advantage of Casia2 over PS is its simplicity and user-friendliness for both clinicians and patients. PS requires subjects' pupils to be dilated for at least 6.5 mm so that the toric marks can be seen. While our study demonstrated that, as long as sufficient light entered the eyes—typically with a PD of 4 mm or greater—Casia2 could obtain accurate measurements of the TIOL axis. And the result is similar with previous literature reports that Casia2 has good repeatability and reproducibility under both nonmydriatic and mydriatic conditions [13, 14]. This makes Casia2 a more accessible tool, especially for patients with contraindicated or insufficient mydriasis. In addition, the Casia2 device has an internal fixating target to better control the subject's eye position, reducing the error or bias induced by eyeball cyclotorsion or tilt.
In the past few decades, assessments of tilt and decentration of the IOL had been conducted using Purkinje imaging technique and Scheimpflug imaging. However, they require image-processing software to calculate IOL tilt and decentration, and good pupil dilation is a prerequisite for them to capture high quality images. In addition, they used the pupillary axis as a reference for calculating the tilt and decentration [24–26]. Anterior segment ocular coherence tomography (AS-OCT) is later employed for measurement of crystalline lens and IOL tilt and decentration. Compared with the first generation AS-OCT of Casia SS-1000, Casia2, the second generation of AS-OCT improves the scanning speed from 3000 A scan per second to 5000 A scan per second and increases the scanning depth from 6 to 13 mm, and thus, Casia2 can obtain the image from the anterior corneal surface to the posterior surface of the lens in one shot [14, 27, 28]. In our study, we employed the device of Casia2 for measuring IOL tilt and decentration. Owing to its good penetration, it can automatically and quantitatively measure the tilt and decentration of crystalline lens and IOL with high repeatability, regardless of whether the eye is under mydriasis or not. And it measures the tilt and decentration relative to the corneal topographic axis. Compared with the pupillary axis, the corneal topographic axis is considered a better reference for assessing IOL tilt and decentration because it is not affected by the shape of the pupil [14, 28, 29].
However, our study has several limitations. Firstly, the study was performed with a small sample size and at a single institution, and the number of different models of TIOL was unbalanced, which could limit the generalizability of the study and may even bias its results. Therefore, further studies with large-scale, multicenter, and more balanced TIOL models are required to validate our findings in this study. Secondly, our study is limited to the evaluation of agreement between the two methods. Further studies including refractive outcomes are desirable to determine the precision and accuracy of the two different methods.
5. Conclusions
In conclusion, our study demonstrates that Casia2 and PS have significant concordance in measuring the postoperative TIOL alignment. The simplicity and user-friendliness of Casia2 make it a more accessible alternative, particularly beneficial for patients with suboptimal pupil dilation.
Acknowledgments
The author thanks Ningshan Li, the postdoctor of Longgang District People's Hospital of Shenzhen, who provided valuable advice on the data analysis. The author also thanks Yanlei Chen and Yongyi Niu of the Department of Ophthalmology, Guangdong Academy of Medical Sciences and Guangdong Provincial People's Hospital, who assisted with data collection.
Nomenclature
- TIOL
Toric intraocular lens
- PS method
Digital slit lamp retroillumination images combined with Adobe Photoshop methods
- ICC
Intraclass correlation coefficients
- LoA
Limits of agreement
- SS-ASOCT
Swept-source Fourier-domain anterior segment optical coherence tomography
- AD
Absolute difference of the TIOL axis measured by Casia2 and PS
- PD
Pupil diameter of Casia2 method in the measurement of TIOL axis
- SE
Average keratometry
- △K
Corneal astigmatism
- AL
Axis length
- ACD
Anterior chamber depth
Data Availability Statement
The data used to support the findings of this study are available from the corresponding author upon request.
Conflicts of Interest
The authors declare no conflicts of interest.
Funding
The research and publication of the article was funded by the Science and Technology Innovation Special Fund of Longgang District of Shenzhen, China (grant number LGKCYLWS2022023); National Natural Science Foundation Incubation Program of Guangdong Provincial People's Hospital (grant number, KY01201146 to Y.Q., and KY01201147 to H.Z.); National Natural Science Foundation of Guangdong Province, China (grant number, 2023A1515011735 to H.Z.); National Natural Science Foundation of China (grant number, 82171036 to H.Z.).
References
- 1.Ferrer-Blasco T., Montés-Micó R., Peixoto-de-Matos S. C., González-Méijome J. M., Cerviño A. Prevalence of Corneal Astigmatism Before Cataract Surgery. Journal of Cataract & Refractive Surgery . 2009;35(1):70–75. doi: 10.1016/j.jcrs.2008.09.027. [DOI] [PubMed] [Google Scholar]
- 2.Shimizu K., Misawa A., Suzuki Y. Toric Intraocular Lenses: Correcting Astigmatism While Controlling Axis Shift. Journal of Cataract & Refractive Surgery . 1994;20(5):523–526. doi: 10.1016/s0886-3350(13)80232-5. [DOI] [PubMed] [Google Scholar]
- 3.Miyake T., Kamiya K., Amano R., Iida Y., Tsunehiro S., Shimizu K. Long-Term Clinical Outcomes of Toric Intraocular Lens Implantation in Cataract Cases With Preexisting Astigmatism. Journal of Cataract & Refractive Surgery . 2014;40(10):1654–1660. doi: 10.1016/j.jcrs.2014.01.044. [DOI] [PubMed] [Google Scholar]
- 4.Visser N., Bauer N. J., Nuijts R. M. Toric Intraocular Lenses: Historical Overview, Patient Selection, IOL Calculation, Surgical Techniques, Clinical Outcomes, and Complications. Journal of Cataract & Refractive Surgery . 2013;39(4):624–637. doi: 10.1016/j.jcrs.2013.02.020. [DOI] [PubMed] [Google Scholar]
- 5.Ma D., Yu W., Cai L., et al. Optimal Timing of Repositioning Surgery for a Plate-Haptic Toric Intraocular Lens: A Multicenter Retrospective Study. Journal of Refractive Surgery . 2023;39(2):120–126. doi: 10.3928/1081597x-20221221-02. [DOI] [PubMed] [Google Scholar]
- 6.Lee B. S., Onishi A. C., Chang D. F. Comparison of Rotational Stability and Repositioning Rates of 2 Presbyopia-Correcting and 2 Monofocal Toric Intraocular Lenses. Journal of Cataract & Refractive Surgery . 2021;47(5):622–626. doi: 10.1097/j.jcrs.0000000000000497. [DOI] [PubMed] [Google Scholar]
- 7.Sanders D. R., Sarver E. J., Cooke D. L. Accuracy and Precision of a New System for Measuring Toric Intraocular Lens Axis Rotation. Journal of Cataract & Refractive Surgery . 2013;39(8):1190–1195. doi: 10.1016/j.jcrs.2013.06.006. [DOI] [PubMed] [Google Scholar]
- 8.Shah G. D., Praveen M. R., Vasavada A. R., et al. Software-Based Assessment of Postoperative Rotation of Toric Intraocular Lens. Journal of Cataract & Refractive Surgery . 2009;35(3):413–418. doi: 10.1016/j.jcrs.2008.10.057. [DOI] [PubMed] [Google Scholar]
- 9.Schartmuller D., Schriefl S., Schwarzenbacher L., Leydolt C., Menapace R. True Rotational Stability of a Single-Piece Hydrophobic Intraocular Lens. British Journal of Ophthalmology . 2019;103(2):186–190. doi: 10.1136/bjophthalmol-2017-311797. [DOI] [PubMed] [Google Scholar]
- 10.Corporation T. Fourier Domain OCT CASIA2. 2017. https://tomey.de/images/product_flyer/CASIA2_br_w.pdf .
- 11.Chansangpetch S., Nguyen A., Mora M., et al. Agreement of Anterior Segment Parameters Obtained from Swept-Source Fourier-Domain and Time-Domain Anterior Segment Optical Coherence Tomography. Investigative Ophthalmology & Visual Science . 2018;59(3):1554–1561. doi: 10.1167/iovs.17-23574. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.Shoji T., Kato N., Ishikawa S., et al. In Vivo Crystalline Lens Measurements With Novel Swept-Source Optical Coherent Tomography: An Investigation on Variability of Measurement. BMJ Open Ophthalmology . 2017;1(1):p. e000058. doi: 10.1136/bmjophth-2016-000058. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13.Lin F., Wang Y., Liu Y., Qu X., Zhou X. The Influence of 0.5% Tropicamide on Anterior Segment Parameters With CASIA2 in Emmetropic, Myopic, and Hyperopic Eyes. Frontiers in Physiology . 2022;13:p. 957097. doi: 10.3389/fphys.2022.957097. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14.Kimura S., Morizane Y., Shiode Y., et al. Assessment of Tilt and Decentration of Crystalline Lens and Intraocular Lens Relative to the Corneal Topographic Axis Using Anterior Segment Optical Coherence Tomography. PLoS One . 2017;12(9):p. e0184066. doi: 10.1371/journal.pone.0184066. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15.Gu X., Chen X., Yang G., et al. Determinants of Intraocular Lens Tilt and Decentration After Cataract Surgery. Annals of Translational Medicine . 2020;8(15):p. 921. doi: 10.21037/atm-20-1008. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16.Shrout P. E., Fleiss J. L. Intraclass Correlations: Uses in Assessing Rater Reliability. Psychological Bulletin . 1979;86(2):420–428. doi: 10.1037//0033-2909.86.2.420. [DOI] [PubMed] [Google Scholar]
- 17.Koo T. K., Li M. Y. A Guideline of Selecting and Reporting Intraclass Correlation Coefficients for Reliability Research. Journal of chiropractic medicine . 2016;15(2):155–163. doi: 10.1016/j.jcm.2016.02.012. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18.McAlinden C., Khadka J., Pesudovs K. Statistical Methods for Conducting Agreement (Comparison of Clinical Tests) and Precision (Repeatability or Reproducibility) Studies in Optometry and Ophthalmology. Ophthalmic and Physiological Optics . 2011;31(4):330–338. doi: 10.1111/j.1475-1313.2011.00851.x. [DOI] [PubMed] [Google Scholar]
- 19.Carkeet A. A Review of the Use of Confidence Intervals for Bland-Altman Limits of Agreement in Optometry and Vision Science. Optometry and Vision Science . 2020;97(1):3–8. doi: 10.1097/opx.0000000000001465. [DOI] [PubMed] [Google Scholar]
- 20.Bland J. M., Altman D. G. Statistical Methods for Assessing Agreement Between Two Methods of Clinical Measurement. Lancet (London, England) . 1986;1(8476):307–310. [PubMed] [Google Scholar]
- 21.Xu B. Y., Mai D. D., Penteado R. C., Saunders L., Weinreb R. N. Reproducibility and Agreement of Anterior Segment Parameter Measurements Obtained Using the CASIA2 and Spectralis OCT2 Optical Coherence Tomography Devices. Journal of Glaucoma . 2017;26(11):974–979. doi: 10.1097/ijg.0000000000000788. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 22.Dembski M., Nowińska A., Ulfik-Dembska K., Wylęgała E. Swept Source Optical Coherence Tomography Analysis of the Selected Eye’s Anterior Segment Parameters. Journal of Clinical Medicine . 2021;10(5):p. 1094. doi: 10.3390/jcm10051094. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 23.Felipe A., Artigas J. M., Díez-Ajenjo A., García-Domene C., Alcocer P. Residual Astigmatism Produced by Toric Intraocular Lens Rotation. Journal of Cataract & Refractive Surgery . 2011;37(10):1895–1901. doi: 10.1016/j.jcrs.2011.04.036. [DOI] [PubMed] [Google Scholar]
- 24.Nishi Y., Hirnschall N., Crnej A., et al. Reproducibility of Intraocular Lens Decentration and Tilt Measurement Using a Clinical Purkinje Meter. Journal of Cataract & Refractive Surgery . 2010;36(9):1529–1535. doi: 10.1016/j.jcrs.2010.03.043. [DOI] [PubMed] [Google Scholar]
- 25.Sasaki K., Sakamoto Y., Shibata T., Nakaizumi H., Emori Y. Measurement of Postoperative Intraocular Lens Tilting and Decentration Using Scheimpflug Images. Journal of Cataract & Refractive Surgery . 1989;15(4):454–457. doi: 10.1016/s0886-3350(89)80071-9. [DOI] [PubMed] [Google Scholar]
- 26.de Castro A., Rosales P., Marcos S. Tilt and Decentration of Intraocular Lenses In Vivo From Purkinje and Scheimpflug Imaging. Validation Study. Journal of Cataract & Refractive Surgery . 2007;33(3):418–429. doi: 10.1016/j.jcrs.2006.10.054. [DOI] [PubMed] [Google Scholar]
- 27.Peguda R., Kang P., Maseedupally V., Swarbrick H. Accuracy and Repeatability of an Anterior Segment Swept-Source Optical Coherence Tomographer. Eye and Contact Lens: Science and Clinical Practice . 2018;44(1):S300–s306. doi: 10.1097/icl.0000000000000417. [DOI] [PubMed] [Google Scholar]
- 28.Sato T., Shibata S., Yoshida M., Hayashi K. Short-Term Dynamics After Single- and Three-Piece Acrylic Intraocular Lens Implantation: A Swept-Source Anterior Segment Optical Coherence Tomography Study. Scientific Reports . 2018;8(1):p. 10230. doi: 10.1038/s41598-018-28609-1. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 29.Ding X., Wang Q., Chang P., et al. The Repeatability Assessment of Three-Dimensional Capsule-Intraocular Lens Complex Measurements by Means of High-Speed Swept-Source Optical Coherence Tomography. PLoS One . 2015;10(11):p. e0142556. doi: 10.1371/journal.pone.0142556. [DOI] [PMC free article] [PubMed] [Google Scholar]