Digital breast tomosynthesis (DBT) systems utilize an isocentric acquisition geometry which introduces imaging artifacts that are deleterious to image reconstructions. The next-generation tomosynthesis (NGT) prototype was designed to incorporate various x-ray source and detector motions for the purpose of investigating alternative acquisition geometries for DBT. Non-isocentric acquisition geometries, acquisitions that vary the image magnification between projection images, are capable of ameliorating aliasing and other artifacts that are intrinsic to conventional DBT. We used virtual clinical trials (VCTs) to develop custom acquisition geometries for the NGT prototype. A high-resolution (5μm voxel size) star pattern test object was simulated to compare the high-frequency performance of isocentric with non-isocentric image reconstructions. A tilted bar pattern test object was also simulated to compare multiplanar reconstructions (MPR) of isocentric and non-isocentric acquisition geometries. Two source- and detector-motion paths were simulated to obtain super-sampled image reconstructions of the test objects. An aliasing-sensitive metric was used to evaluate spatial resolution performance for two orthogonal frequency orientations. Pairwise comparisons were made for the two frequency orientations between the isocentric and non-isocentric acquisition geometries. Non-isocentric acquisition geometries show an improvement over isocentric acquisition geometries. The greatest improvement was 75.2% for frequencies aligned perpendicular to x-ray source motion, which is the direction of frequencies for which DBT is prone to aliasing. Both frequency orientations exhibit super resolution for non-isocentric geometries. MPR of the tilted bar pattern show z- dependent degeneracies for the isocentric acquisition only, whereas MPR of the non-isocentric acquisition entirely exhibits super resolution.
Keywords: digital breast tomosynthesis, non-isocentric tomosynthesis, physics virtual clinical trial, ray tracing, radial fast Fourier transform, super-resolution, multiplanar reconstruction
1. INTRODUCTION
Digital breast tomosynthesis (DBT) acquires a three-dimensional (3D) image of a patient’s breast by acquiring multiple, two-dimensional x-ray images from serial source positions. For a conventional DBT acquisition, the x-ray projections are acquired by translating the x-ray source along an arc in the mediolateral (ML) direction. The projections are spaced equidistantly in the chest-wall plane of the patient. This acquisition geometry is isocentric, because the arc rotates about a single fulcrum in space. Such DBT systems benefit the detection of masses and low-frequency objects in the breast. That said, the detection of high-frequency objects like microcalcifications, is not benefited by DBT when compared with full- field digital mammography (FFDM)1.
Resolution for conventional DBT is anisotropic, favoring the frequencies that are aligned parallel to x-ray source motion () with super resolution. Using super-sampled image reconstructions, resolution for can be increased twofold, whereas the frequencies aligned perpendicular to source motion () are limited to the alias frequency of the detector. Super resolution improves the resolution of periodic objects and the overall appearance and texture of breast tissue and lesions in image reconstructions.
The next-generation tomosynthesis (NGT) system is a DBT prototype that has been developed to improve the detection of microcalcifications in the breast. This prototype system investigates alternative acquisition geometries by introducing novel x-ray source and detector motions. The introduction of posteroanterior (PA) x-ray source motion has been shown to improve the spatial resolution2–5 and volume estimation of anthropomorphic breast phantoms in image reconstructions6,7. Non-isocentric acquisition geometries, acquisitions that increment the detector z-position between each projection, can further improve spatial sampling. Non-isocentric acquisitions can also lessen the impact of imaging artifacts like masking8,9, spatial resolution anisotropies10–12, and aliasing11,13 in image reconstructions.
The primary method for validating the performance of these prototypes is through clinical trials. Clinical trials of medical imaging systems require a large number of patients, who are imaged repeatedly to compare the performance of different systems. The associated cost, duration, and radiation risk represent a significant impediment to the efficient introduction of novel imaging technologies. Virtual clinical trials (VCTs) represent an approach, based upon computer simulation of human anatomy and imaging modalities, which can help develop, optimize, and validate new and existing medical imaging methods. Fast and detailed simulations of device performance are also instrumental in prototyping clinical trials and augmenting them with simulated data, and may also assist with regulatory approval4,14,15.
Physics VCTs, in contrast to traditional VCTs, are intended to evaluate the implementation of the novel designs using simulated image acquisitions of virtual test objects and objective metrics to evaluate device performance. We have used physics VCTs to investigate novel acquisition geometries for the NGT prototype4. We have shown that physics VCTs can be used to investigate novel acquisition geometries for the NGT system without exhausting resources. In this work, we use physics VCTs and high-resolution test objects to compare the performance of isocentric with non-isocentric acquisition geometries for DBT using configurations of the NGT prototype.
2. MATERIALS & METHOD
Virtual Clinical Trial pipeline
Physics VCTs are implemented using the X-ray Physics Lab’s (XPL) physics VCT pipeline. The physics VCT pipeline has been modified from the original VCT pipeline. Physics VCTs simulate the image acquisition of physical test objects and phantoms. However, rather than using a virtual reader phase of the original VCT pipeline, objective measures of physics are used to evaluate the performance of an imaging device. A tomosynthesis acquisition is simulated, in the same manner as the original pipeline, using ray-tracing methods with a dedicated graphics processing unit (GPU). The mechanical configuration of the NGT prototype is modeled using the physics VCT pipeline, where various acquisition geometries can be simulated.
Next-generation tomosynthesis prototype
Alternative acquisition geometries of the NGT prototype are achieved by creating custom x-ray source and detector motion paths for a series of x-ray projection images. The collection of projection images obtained from each custom acquisition geometry are used for tomographic reconstructions. Details of this prototype have been described in previous work3,4,16,17. The origin of the NGT system is defined at the center of the chest-wall edge of the breast support (Figure 1). The x-y plane of the NGT system serves as an address space for the x-ray focal spot to determine an acquisition geometry. The range of the source is ±150mm in x and +180mm in y. The NGT achieves non-isocentric acquisition geometries by translating the detector in z between x-ray projections with a 44.8mm range, translating 3.2 mm between projection over 15 projections. The configuration of the NGT prototype and the two acquisition geometries that are used for the physics VCT simulations in this study are shown in Figure 1. This study is constrained to the conventional x-ray source motion only and does not include acquisition geometries with PA source motion.
The geometric configuration and constraints of the NGT prototype dictate the image acquisition simulation of the physics VCTs. Physics VCTs simulate the radiographic technique for x-ray projection images that are used to render image reconstructions. Tomographic image reconstructions are created by filtered back-projection using commercial reconstruction software (Piccolo version 4.0.5, Real Time Tomography, Villanova, PA).
High-resolution test objects
A voxelized model of a star pattern test object (Model 07-542-1000, Supertech, Elkhart, IN) was created using Matlab (MathWorks, Natick, Massachusetts, version 2018a). This virtual star pattern (Figure 2) consists of 5μm isotropic voxels with three material indices: lead, acrylic, and air. The diameter of the star pattern is 45mm with a total thickness of 30μm. The range of resolution for this star pattern, from the outer edge to the innermost circle, is 1.3–20.0 line pairs per millimeter (lp/mm). The star pattern contains four quadrants, giving two possible frequency orientations for both x and y ( and , respectively).
In addition to the star pattern, a bar pattern test object was created at a constant spatial resolution of 70 μm with a frequency orientation, . The bar pattern is indexed with lead and air and was created at a 45° tilt relative to the breast support (Figure 3). The tilt of the bar pattern was used to compare the spatial resolution at various depths of the image reconstruction using multiplanar reconstructions (MPR). Using the commercial reconstruction software, MPR can be achieved at arbitrary angles18. This allowed us to visualize the plane of the bar pattern test object at a plane that is oriented 45° relative to the detector plane. Conventional DBT is anisotropic in the z dimension, which degenerate periodic signals like a bar pattern or punctate calcifications in the breast, at various intervals of z. Non-isocentric acquisition geometries can reduce these degeneracies.
Physics VCTs were used to simulate image acquisition of the virtual star pattern and tilted bar pattern using isocentric and non-isocentric configurations of the NGT prototype. The conventional x-ray source motion path (Figure 1) was simulated with and without z-direction detector motion (the z-detector position is 0 over all projections), to simulate non-isocentric and isocentric acquisition geometries, respectively. Repositioning and reproducibility of the image acquisition was simulated by perturbing the position of the phantom within 2 mm of the detector origin over 10 tomosynthesis acquisitions for each geometry. Poisson noise was also simulated for each projection image. Super-sampled image reconstructions of the star pattern test object were obtained using the simulated 2D projection images of each tomosynthesis acquisition simulation. The images were reconstructed at a 2.0x super-sampling rate (42.5 μm for the 85 μm detector element size).
The slice of the image reconstruction that contained the plane of the star pattern was used to evaluate the in-plane spatial resolution properties of the 20 image reconstructions. The radial fast Fourier transform (RFFT)19 was used to evaluate the in-plane performance of both acquisition geometries. This metric used the star pattern image as an input to produce the contrast transfer function (CTF) and RFFT graphs (e.g. Figure 4), averaged over the 10 reconstruction slices for each geometry. The CTF is analogous to the modulation transfer function (MTF) but is not normalized to unity at zero spatial frequency. The RFFT graph shows modulation of all frequencies contained in a quadrant of a star pattern image reconstruction alongside the modulation of aliased signals. The CTF was used to determine the limit of spatial resolution (LSR). Conventional DBT is spatially anisotropic11; therefore, each image reconstruction is evaluated for and .
The same acquisition geometries were used to generate an MPR of the tilted bar pattern. Images of the tilted bar pattern were reconstructed at a 2x sampling rate. The reconstruction plane was oriented at the tilt of the bar pattern. Plot profiles of a degenerate region in the image reconstruction of the isocentric acquisition geometry was produced and compared with the plot profile of the same region for the non-isocentric acquisition geometry.
3. RESULTS AND DISCUSSION
Star pattern RFFT
An example of a Radial FFT and CTF calculation is shown for a single reconstruction slice in Figure 4. The RFFT and CTF graphs of the non-isocentric acquisition geometries show a reduction of spectral leakage compared with the isocentric acquisition geometries. The Radial FFT was computed for both frequency orientations, and , and averaged over the 10 reconstruction slices for each acquisition geometry.
The non-isocentric acquisition geometry exhibits isotropic super-resolution (Figure 5), achieving a limiting spatial resolution (LSR) above the alias frequency in both and , whereas the isocentric acquisition geometry achieves super resolution for ; however, resolution is limited to the alias frequency for . The isocentric acquisition geometry shows more significant aliasing in the form of Moiré patterns compared with the non-isocentric acquisition geometry. The modulation contrast is higher for the isocentric acquisition geometries at the lowest spatial frequencies of the star pattern, regardless of frequency orientation.
The LSR of the four conditions was measured using images of the star pattern and the CTF. All isocentric image reconstructions exhibited significant aliasing for , and the LSR was limited to near the alias frequency of the detector (5.88 lp/mm). For reconstruction slices where super resolution was achieved, the LSR was measured at a 5% modulation of the CTF. The improvement of the LSR for the non-isocentric over the isocentric geometry was 75.2% for . Both the image of the star pattern and the RFFT graph for the non-isocentric case exhibit residual aliasing artifact, but this artifact does not overcome the modulation of the input frequency.
Tilted bar pattern MPR
The tilted bar pattern was reconstructed with a roll of 45° about the x-axis for each acquisition geometry. The MPR of the isocentric acquisition geometry shows four distinct degenerate regions (e.g. the region bounded by the red box in Figure 6). Super resolution cannot be achieved at the corresponding z-locations in the image reconstruction. The MPR of the non- isocentric acquisition geometry shows multiple regions of decreased modulation, but no fully degenerate regions of the bar-pattern are seen. Super resolution is observed over the entire MPR.
4. CONCLUSION
We have shown that non-isocentric acquisition geometries can be simulated with high accuracy using physics VCTs. With 10 simulated tomosynthesis image reconstructions, we were able to measure the RFFT and LSR accurately and achieve results that are commensurate with physical experiments4. The LSR improved for every frequency orientation with the introduction of z-detector motion. The highest overall improvement in LSR for the non-isocentric over isocentric acquisition geometry is 75.2% for . All frequency orientations of the non-isocentric reconstructions exhibit super resolution, whereas the isocentric geometry was prone to aliasing. These results suggest that non-isocentric acquisition geometries can improve the detection of high-frequency objects like microcalcifications for DBT. In the future, we will simulate additional non-isocentric acquisition geometries that include PA source motion to optimize super resolution further.
ACKNOWLEDGEMENTS
The authors would like to thank Johnny Kuo, Susan Ng, and Peter Ringer of Real Time Tomography for technical assistance with Piccolo. Andrew D. A. Maidment is a shareholder of Real Time Tomography and is a member of the scientific advisory board.
Support was provided by the following grants: W81XWH-18-1-0082 from the Department of Defense Breast Cancer Research Program, IRSA 1016451 from the Burroughs Wellcome Fund, 1R01CA196528 from the National Institute of Health, and IIR13264610 from Susan G. Komen. In addition, equipment support was provided by Analogic Inc., Barco NV, and Real Time Tomography. The content is solely the responsibility of the authors and does not necessarily represent the official views of the funding agencies.
REFERENCES
- [1].Rafferty E. a, Park JM, Philpotts LE, Poplack SP and Sumkin JH, “Digital Mammography and Breast Tomosynthesis Compared with Digital Mammography Alone : Results of a Multicenter, multireader trial,” Radiology 266(1), 104–113 (2013). [DOI] [PMC free article] [PubMed] [Google Scholar]
- [2].Eben JE, Vent TL, Choi CJ, Yarrabothula S, Chai L, Nolan M, Kobe E, Acciavatti RJ and Maidment ADA, “Development of a next generation tomosynthesis system,” Prog. Biomed. Opt. Imaging - Proc. SPIE 10573 (2018). [Google Scholar]
- [3].Maidment TD, Vent TL, Ferris WS, Wurtele DE, Acciavatti RJ and Maidment ADA, “Comparing the imaging performance of computed super resolution and magnification tomosynthesis,” Med. Imaging 2017 Phys. Med. Imaging 10132, 1013222 (2017). [Google Scholar]
- [4].Vent TL, Barufaldi B and Maidment ADA, “Simulation and experimental validation of high-resolution test objects for evaluating a next-generation digital breast tomosynthesis prototype,” 21 (2019). [Google Scholar]
- [5].Vent TL, Lepore BL and Maidment ADA, “Evaluating the imaging performance of a next-generation digital breast tomosynthesis prototype,” Med. Imaging 2019 Phys. Med. Imaging(March), 19 (2019). [Google Scholar]
- [6].Acciavatti Raymond; Rodriguez-Ruiz Alejandro; Vent Trevor L.; Bakic Predrag R.; Reiser Ingrid; Sechopoulos Ioannis; Maidment ADA, “Analysis of volume overestimation artifacts in the breast outline segmentation in tomosynthesis,” Med. Imaging 2018 Phys. Med. Imaging, 10573 (2018). [DOI] [PMC free article] [PubMed] [Google Scholar]
- [7].Acciavatti RJ, Barufaldi B, Vent TL, Wileyto EP and Maidment ADA, “Personalization of x-ray tube motion in digital breast tomosynthesis using virtual Defrise phantoms,” Med. Imaging 2019 Phys. Med. Imaging(March), 10 (2019). [DOI] [PMC free article] [PubMed] [Google Scholar]
- [8].Friedewald SM, Rafferty EA, Rose SL, Durand MA, Plecha DM, Greenberg JS, Hayes MK, Copit DS, Carlson KL, Cink TM, Barke LD, Greer LN, Miller DP and Conant EF, “Breast cancer screening using tomosynthesis in combination with digital mammography,” JAMA - J. Am. Med. Assoc 311(24), 2499–2507 (2014). [DOI] [PubMed] [Google Scholar]
- [9].Roth RG, Maidment ADA, Weinstein SP, Roth SO and Conant EF, “Digital Breast Tomosynthesis: Lessons Learned from Early Clinical Implementation,” RadioGraphics 34(4), E89–E102 (2014). [DOI] [PMC free article] [PubMed] [Google Scholar]
- [10].Hu Y-H, Zhao B and Zhao W, “Image artifacts in digital breast tomosynthesis: Investigation of the effects of system geometry and reconstruction parameters using a linear system approach,” Med. Phys 35(12), 5242–5252 (2008). [DOI] [PMC free article] [PubMed] [Google Scholar]
- [11].Acciavatti RJ and Maidment ADA, “Observation of super-resolution in digital breast tomosynthesis,” Med. Phys 39(12), 7518–7539 (2012). [DOI] [PMC free article] [PubMed] [Google Scholar]
- [12].Hatt CR, Tomkowiak MT, Dunkerley DAP, Slagowski JM, Funk T, Raval AN and Speidel MA, “Depth-resolved registration of transesophageal echo to x-ray fluoroscopy using an inverse geometry fluoroscopy system,” Med. Phys 42(12), 7022–7033 (2015). [DOI] [PMC free article] [PubMed] [Google Scholar]
- [13].Vent TL, Acciavatti RJ, Kwon YJ and Maidment ADA, “Quantification of resolution in multiplanar reconstructions for digital breast tomosynthesis,” Med. Imaging 2016 Phys. Med. Imaging 9783(March 2016), 978303 (2016). [Google Scholar]
- [14].Bakic PR, Barufaldi B, Pokrajac D, Lago MA and Maidment AD, “Developing populations of software breast phantoms for virtual clinical trials,” 14th Int. Work. Breast Imaging (IWBI 2018)(July 2018), 73 (2018). [Google Scholar]
- [15].Barufaldi B, Bakic PR, Higginbotham D and Maidment ADA, “OpenVCT: a GPU-accelerated virtual clinical trial pipeline for mammography and digital breast tomosynthesis,” Med. Imaging 2018 Phys. Med. Imaging 1057358(March), 194 (2018). [Google Scholar]
- [16].Eben JE, Vent TL, Choi CJ, Yarrabothula S, Chai L, Nolan M, Kobe E, Acciavatti RJ and Maidment ADA, “Development of a Next Generation Tomosynthesis System,” Med. Imaging 2018 Phys. Med. Imaging, 10573 (2018). [Google Scholar]
- [17].Maidment Andrew D. A.; Acciavatti Raymond J.; Vent Trevor L.; Conant Emily F.; Kwon Young Joon; NG Susan; Kuo Jhonny; Ringer Peter A.; Maidment Tristan; Wurtele David; Licata Joseph; Narayan Tejas; Zhang David; Higginbotham D, “Construction of a Prototype Digital Breast Tomosynthesis System with Superior Spatial Resolution,” RSNA Phys. Basic Sci., RSNA (2016). [Google Scholar]
- [18].Vent TL, Acciavatti RJ, Kwon YJ and Maidment ADA, “Quantification of resolution in multiplanar reconstructions for digital breast tomosynthesis,” Med. Imaging 2016 Phys. Med. Imaging 9783, 978303 (2016). [Google Scholar]
- [19].Vent T, Acciavatti R and Maidment A, “Development and Evaluation of a Spatial Resolution Metric for Tomosynthesis: we-g-601–02,” Med. Phys 44(6), 3261 (2017). [Google Scholar]