Abstract
Background/Objectives: This study aimed to characterize the visual performance in patients with bilateral cataracts and a unilateral epiretinal membrane (ERM) undergoing cataract surgery with non-diffractive extended depth of focus (EDoF) intraocular lens (IOL) placement in both eyes and phacovitrectomy in the eye with ERM. Methods: This was a prospective, descriptive, single-arm study. Visual outcomes were measured in monocular and binocular conditions in patients with bilateral cataracts and unilateral ERM stages 2 or 3 implanted with an EDoF IOL. At 6 months, visual acuity (uncorrected and corrected at different distances and contrast levels), contrast sensitivity, and visual disturbances assessed using a Light Distortion Analyzer were determined. Results: We included 22 patients (50% females, mean age of 71.4 ± 5.8 years). Mean monocular best-corrected visual acuities at 100% contrast were 0.07 ± 0.09, 0.23 ± 0.10, and 0.48 ± 0.14 logMAR for eyes with ERM, and 0.02 ± 0.08, 0.19 ± 0.11, and 0.41 ± 0.18 logMAR for fellow eyes for far, intermediate, and near distance vision, respectively. No differences were observed in visual outcomes, contrast sensitivity, or visual disturbances between ERM and non-ERM eyes, except for near distance. No unexpected adverse events were observed. Conclusions: Non-diffractive EDoF IOL can be considered in patients with cataracts and stage 2 or 3 ERM pursuing spectacle independence.
Keywords: cataract, epiretinal membrane, extended depth of focus, intraocular lens, pars plana vitrectomy, phacoemulsification, phacovitrectomy
1. Introduction
Idiopathic epiretinal membrane (ERM) is a common macular disorder in older adults, with a prevalence of 2% to 20% [1]. Epiretinal membranes are thought to be the consequence of fibroglial proliferation on the inner surface of the retina associated with small tears in the inner limiting membrane during a posterior vitreous detachment. Anteroposterior forces result in vertical traction and increased retinal thickness. In contrast, tangential forces pull the superficial retinal layers away from their original position [2], straighten or twist retinal vessels, and reduce the foveal avascular zone [3] and macular edema [4]. As a result, patients experience metamorphopsia, micropsia, monocular diplopia, or visual acuity loss [2]. Its management includes observation or, in the 30% of cases in which there is progression [5], vitrectomy surgery, with a reasonable success rate.
Today, all patients have high expectations of enhanced vision and reduced dependence on glasses following cataract surgery. However, numerous studies have shown that multifocal intraocular lenses (IOLs) can lead to a decline in contrast sensitivity as a consequence of incident light dispersion [6,7,8,9]. Several studies [10,11], including a meta-analysis [9], have found that individual patient satisfaction is related to many factors, including the level of uncorrected visual acuity, contrast sensitivity, or halos/glare.
Recent advances in IOL designs have led to the development of an extended depth of focus (EDoF). These IOLs extend the range of focus, with the goal of offering excellent vision not only at far but also at intermediate distances, thereby also enhancing near vision. Several studies that have compared the optical quality of EDoF IOL vs. bifocal or trifocal multifocal lenses have confirmed the optimal vision quality obtained with EDoF lenses [12,13,14,15]. This illustrates that the indications for EDoF IOLs could perhaps be broadened to encompass selected cases in which contrast sensitivity is a concern, such as those with ERM, mild to moderate glaucoma, or subjects post-LASIK.
The AcrySof® IQ Vivity® lens (Alcon Healthcare, Fort Worth, TX, USA) is a multifocal EDoF lens that does not split the light entering the eye. Consequently, it does not compromise contrast sensitivity or cause as many dysphotopic phenomena as other multifocal IOLs [14,16]. Also, it offers excellent vision at all distances, though glasses are needed for near tasks [16].
This study assessed the visual performance after the implant of the EDoF AcrySof IQ Vivity® lens in both eyes of patients with ERM in one eye and cataracts in both eyes.
2. Materials and Methods
This was a prospective, single-arm descriptive study. Participants were patients with surgical ERM (stages 2 or 3) in one eye and cataracts in both eyes. The eye with ERM and cataracts was first treated in two separate operations, and less than one month later, surgery was performed in the eye with cataract only. The lens implanted in both eyes was the AcrySof IQ Vivity®. Surgeries were carried out at the OMIQ (Oftalmologia Mèdica i Quirúrgica) Research facilities in Barcelona (Spain). The study protocol adhered to the principles outlined in the Declaration of Helsinki and received approval from the Ethics Committee of the Grupo Hospitalario Quirónsalud-Catalunya. All participants signed an informed consent form before undergoing any procedure or test.
2.1. Patients
Participants were required to have both bilateral cataracts and primary or idiopathic unilateral stage 2 or 3 ERM. The latter was defined as follows. Stage 2: presence of ERM associated with progressive retinal distortion, foveal depression loss, and characteristic stretching of the outer nuclear layer, with all retinal layers defined and clearly identified using optical coherence tomography (OCT). Stage 3: the presence of an ERM with continuous ectopic inner foveal layers anomalously crossing the central foveal area, foveal depression absent, and all retinal layers clearly identified using OCT [17]. Further inclusion criteria were good expected postoperative distance-corrected visual acuity (VA ≤ 0.2 logMAR or Snellen equivalent ≥ 20/32), age over 50 years, and follow-up available at 6 months post-surgery. Exclusion criteria were other concomitant posterior segment disorders; myopia < −5.00 D; presence or history of any systemic disorder, condition, or disease that could interfere with procedures or interpretation of results; previous intraocular surgery (vitrectomy in the fellow eye, glaucoma surgery, etc.); and any systemic co-morbidity which, based on the investigator’s expert medical opinion, could confound the results of the study or increase the risk for the subject.
The number of participants considered sufficient to describe the visual results expected from this study was 20. We therefore recruited 22 individuals in case some patients withdrew from the study before completing 6 months of follow-up.
2.2. Intraocular Lens
The AcrySof IQ Vivity® lens is manufactured as a single piece utilizing a hydrophobic acrylic material with a high refractive index (n = 1.55). The lens is a non-diffractive EDoF IOL with wavefront-shaping X-WAVE technology that incorporates two smoothly transitioning elements designed to manipulate and redirect the wavefront. One of these elements is a slightly elevated smooth plateau (~1 µm high) that extends the wavefront, creating a continuous focal range. The other element is a curvature change (across the 2.2 mm region), effectively shifting the wavefront to maximize the utilization of all light energy. This innovative design ensures the efficient use of virtually all transmitted light energy passing through the IOL [18].
The lens anterior surface is intentionally configured with a negative spherical aberration of −0.20 µm for a 6 mm pupil, compensating for the positive spherical aberration typically found in the natural human cornea [19]. Lastly, a blue light-filtering chromophore was added to filter light similarly to the human crystalline lens and reduce the transmission of potentially harmful ultraviolet and blue light wavelengths.
2.3. Clinical Tests
Each participant underwent a thorough preoperative eye examination, including both objective and subjective refraction, uncorrected distance (4 m) visual acuity (UCDVA), and corrected distance VA (CDVA) using optotypes from the Early Treatment Diabetic Retinopathy Study (ETDRS test), as well as examinations of the anterior and posterior segments. Additional preoperative assessments included corneal topography (Pentacam®, Oculus Inc), spectral domain optical coherence tomography (OCT) of the macula and peripapillary region (Spectralis HRA+OCT®, Heidelberg Engineering; Heidelberg, Germany), and ocular biometry (Lenstar LS 900®, Haag-Streit; Köniz, Switzerland).
Intraocular lens powers were determined using the Barrett Universal II formula targeting post-surgery emmetropia in all eyes. Follow-up visits were scheduled for 1 day, 1 week, 1 month, and 3 and 6 months post-surgery. All surgeries were conducted by two experienced surgeons (LS and MG). The outcome variables for this study were measured at the 6-month visit.
Tests were carried out monocularly and binocularly. Photopic (~85 cd/m2) VA evaluations employed the ETDRS test at distances of 4 m, 66 cm, and 40 cm for far, intermediate, and near vision, respectively. Contrast sensitivity was gauged with the Pelli Robson test, and VA tests were conducted at various contrast levels (100% for far distance and 25% for intermediate and near distances). The Light Distortion Analyzer (LDA; Bynarytarget Lda, Braga, Portugal) was employed for non-invasive objective measurement of dysphotopsias. This device had a central white LED surrounded by 240 small white LEDs arranged in 24 hemimeridians with a separation angle of 15°, covering a 10° area at a testing distance of 2 m. Peripheral stimuli were presented to the subject from the central zone in all hemimeridians. The speed of stimulus presentation remained constant throughout.
Sitting 2 m from the device in a dark room, subjects were instructed to call out as soon as a small LED appeared differently from the central LED light. This test was repeated three times under both monocular and binocular conditions, and the result was taken as the mean of the three measures. The light distortion index (LDI) was calculated as the ratio between the area of points missed by the subject and the entire area examined, expressed as a percentage (%) of the area covered by the central LED-induced halo. The best-fit circle radius (BFCRad) was defined as the circle that best fit into the area of distortion, delineated by a line connecting all the points in each meridian. This variable is expressed in millimeters.
2.4. Statistical Analysis
Data analysis was performed using SPSS software for Windows version 27.0 (IBM, Armonk, NY, USA). Descriptive statistics were used to analyze the data. We used the Shapiro–Wilk test to assess the normality of each study metric. A descriptive analysis was performed, including mean and standard deviation (SD) for each variable recorded. When parametric analysis was possible, we used Student’s t-test for paired data to compare results between eyes with and without ERM, between consecutive visits, or between pre- and postoperative values when appropriate. When parametric analysis was not possible, we used the Wilcoxon test to compare the variables. Significance was set at p < 0.05 and is represented by the symbol * in the results.
3. Results
A total of 22 patients were enrolled, with an equal distribution between genders (50% female, n = 11). The mean age was 71.4 ± 5.8 years and ranged from 62 to 80 years. Half of the patients had an ERM in the right eye, and the remaining half had an ERM in the left eye. Among the ERM, 27% were classified as stage 2 (n = 12) and 73% as stage 3 (n = 32). Additionally, 55% of the ERMs (n = 24) were characterized as regular, and 86% (n = 38) showed no outer retinal changes. Following surgical ERM correction, 94% (n = 41) showed improvement in the ectopic inner foveal layer. Foveal recovery included 83% (n = 37) of individuals with stage 2 ERM with a restored foveal depression; only 31% (n = 14) of those with stage 3 ERM showed similar recovery at 6 months. Table 1 compares the biometric data between eyes with and without an epiretinal membrane.
Table 1.
Eyes with ERM Mean ± SD (Range) |
Eyes Without ERM Mean ± SD (Range) |
p-Value | |
---|---|---|---|
Preoperative SE (D) | 0.90 ± 1.56 (−3.75 to 3.50) |
0.85 ± 1.36 (−1.75 to 3.50) |
0.995 |
Preoperative Km (D) | 43.67 ± 1.50 (40.84 to 45.78) |
43.95 ± 1.62 (40.94 to 46.84) |
0.208 |
Axial length (mm) | 23.43 ± 0.69 (22.30 to 25.17) |
23.40 ± 0.71 (22.02 to 24.81) |
0.907 |
Anterior chamber depth (mm) | 3.05 ± 0.28 (2.66 to 3.69) |
3.09 ± 0.29 (2.60 to 3.64) |
0.803 |
Implanted IOL (D) | 21.77 ± 1.93 (18.50 to 27.00) |
21.09 ± 1.01 (20.00 to 23.50) |
0.07 |
Preoperative IOP (mmHg) | 14.18 ± 2.92 (10.00 to 20.00) |
14.27 ± 2.60 (10.00 to 21.00) |
0.507 |
Pre-CMT (µm) | 423.32 ± 80.87 (289.00 to 564.00) |
282.45 ± 30.59 (185.00 to 337.00) |
<0.001 * |
Pre-RNFL (µm) | 103.74 ± 12.46 (81.00 to 126.00) |
99.16 ± 9.06 (79.00 to 111.00) |
0.353 |
Pre-FCP (µm) | 399.91 ± 101.41 (211.00 to 567.00) |
236.36 ± 36.44 (146.00 to 337.00) |
<0.001 * |
Pre-volume (mm3) | 9.85 ± 1.25 (8.42 to 12.81) |
8.57 ± 0.46 (7.73 to 9.42) |
<0.001 * |
Table 2 and Table 3 provide the results obtained for monocular and binocular visual acuity, respectively. At 6 months of follow up, significant differences between groups (ERM vs. non-ERM) were observed only in near vision, with those undergoing phacovitrectomy showing worse acuity (p ≤ 0.048). At 6 months post-surgery, only one patient had a spherical equivalent greater than 0.5 D in the non-ERM group, and two patients in the ERM group had a spherical equivalent greater than 0.5 D.
Table 2.
Eyes with ERM Mean ± SD (Range) |
Eyes Without ERM Mean ± SD (Range) |
p-Value | |
---|---|---|---|
Preoperative monocular CDVA | 0.24 ± 0.18 (0.00 to 0.74) |
0.10 ± 0.10 (0.00 to 0.38) |
0.002 * |
Postoperative monocular UDVA, 6 months | 0.12 ± 0.12 (−0.02 to 0.36) |
0.09 ± 0.14 (−0.12 to 0.42) |
0.288 |
Postoperative monocular CDVA, 6 months | 0.07 ± 0.09 (−0.02 to 0.36) |
0.02 ± 0.08 (−0.12 to 0.20) |
0.075 |
Postoperative monocular UIVA, 6 months | 0.22 ± 0.09 (0.02 to 0.34) |
0.18 ± 0.10 (0.02 to 0.36) |
0.991 |
Postoperative monocular DCIVA, 6 months | 0.23 ± 0.10 (0.08 to 0.40) |
0.19 ± 0.11 (0.02 to 0.40) |
0.565 |
Postoperative monocular UNVA, 6 months | 0.46 ± 0.13 (0.14 to 0.78) |
0.37 ± 0.18 (0.10 to 0.70) |
0.037 * |
Postoperative monocular DCNVA, 6 months | 0.48 ± 0.14 (0.16 to 0.78) |
0.41 ± 0.18 (0.10 to 0.70) |
0.048 * |
Table 3.
Binocular Values Mean ± SD (Range) |
|
---|---|
Preoperative binocular CDVA | 0.24 ± 0.18 (0.00 to 0.74) |
Postoperative binocular UDVA, 6 months | 0.03 ± 0.13 (−0.16 to 0.52) |
Postoperative binocular CDVA, 6 months | −0.01 ± 0.07 (−0.16 to 0.20) |
Postoperative binocular UIVA, 6 months | 0.10 ± 0.08 (−0.08 to 0.24) |
Postoperative binocular DCIVA, 6 months | 0.10 ± 0.08 (−0.02 to 0.24) |
Postoperative binocular UNVA, 6 months | 0.28 ± 0.15 (0.04 to 0.60) |
Postoperative binocular DCNVA, 6 months | 0.30 ± 0.15 (0.04 to 0.60) |
The evolution of visual acuity via a surgical procedure (phacoemulsification or phacovitrectomy) is shown in Figure 1. It confirms a progressive improvement in visual acuity in both groups with time but is more marked in eyes with ERM. At month 1, the visual acuities were better in eyes without ERM irrespective of the distance being evaluated, but they became similar at 3 and 6 months and were only different in the near range.
When comparing contrast sensitivity between eyes with and without ERM, no significant differences were found (Table 4). Similarly, no differences were observed in visual disturbances between the two groups of eyes, as indicated in Table S1.
Table 4.
Eyes with ERM Mean ± SD (Range) |
Eyes Without ERM Mean ± SD (Range) |
p-Value | |
---|---|---|---|
Postoperative monocular Pelli-Robson | 1.51 ± 0.12 (1.35 to 1.65) |
1.55 ± 0.11 (1.35 to 1.65) |
0.834 |
Postoperative monocular VA, low contrast (25%), 66 cm | 0.71 ± 0.13 (0.44 to 1.00) |
0.67 ± 0.18 (0.36 to 1.00) |
0.089 |
Postoperative monocular VA, low contrast (25%), 40 cm | 0.80 ± 0.17 (0.54 to 1.10) |
0.73 ± 0.18 (0.44 to 1.04) |
0.970 |
4. Discussion
The implant of a multifocal intraocular lens in adults with a retinal disorder may be considered a contraindication owing to a possible decrease in contrast sensitivity post-surgery despite a good visual acuity outcome [20]. However, it should be noted that risk factors for retinal and cataract disease considerably overlap, indicating that these two conditions often co-occur in older adults [21].
Studies examining the behavior of diffractive multifocal lenses have shown that in persons with ERM, these lenses result in poor refractive outcomes, involving reductions in visual acuity and contrast sensitivity [22,23]. Fortunately, the introduction of new generation lenses, such as the AcrySof IQ Vivity® lens, which does not split light, has improved outcomes in these patients, including contrast sensitivity comparable to that achieved with monofocal lenses [24]. This indicates that for persons with ocular diseases like glaucoma or a retinal disorder, the placement of this type of lens will enable them to have spectacle-free vision at various distances with no increase in dysphotopic phenomena.
In a recent study, Jeon et al. [25] analyzed the performance of the AcrySof IQ Vivity® lens in patients with epiretinal membranes. However, their study was retrospective, and ERMs were limited to stages 1 and 2. In contrast, our prospective study explores more advanced stages (2 and 3) and examines the bilateral implant of these lenses. With this approach, we were able to effectively compare outcomes in eyes with and without ERMs.
Six months after the IOL implant, no significant or clinically relevant differences in visual acuity for distance and intermediate vision (66 cm) were noted between eyes with and without ERMs. However, for near vision, better results were obtained in eyes without ERMs, both with and without distance correction. Although significant, the difference was less than one line of vision. The results of Jeon et al. [25], also obtained six months after IOL implant, indicated better intermediate vision (VA 0.14 vs. 0.22) and near vision (VA 0.23 vs. 0.46). However, these results were only presented without correction, and postoperative refractive defects were not specified. Accordingly, a slight myopic defect could explain this difference. Chung et al. [26] examined an advanced monofocal lens, the Tecnis ICB00 model (Johnson & Johnson Vision Care Inc., Irvine, CA, USA), in patients with ERMs. Patients were divided into two groups: Group 1 included stage 1 and 2 ERM, while Group 2 included stage 3. Outcomes in the two groups in terms of distance vision were similar to those obtained in our study. However, our outcomes were better in that there were differences in intermediate vision (0.26 for Group 1 and 0.32 for Group 2, compared to 0.22) and more pronounced differences in near vision (0.71 for Group 1 and 0.78 for Group 2, compared to 0.46).
When comparing our binocular visual results with those obtained in other studies performed on healthy eyes, we find consistent VA to that reported by Gundersen et al. [27] for far (−0.07 vs. −0.01) and intermediate distance (0.00 vs. 0.10). However, our near distance acuities were less favorable (0.07 vs. 0.28), though comparable, to those obtained by Bala et al. [28] (0.31). In any case, our monocular results also indicated differences between the two groups of eyes in near vision. However, there was an improvement due to binocular summation compared to the results obtained monocularly in the eye without an ERM (0.28 vs. 0.37).
As mentioned earlier, adequate contrast sensitivity is crucial in patients with retinal disease, as this factor may be impacted by the condition. Our study revealed no contrast sensitivity difference between the eye with an ERM and the fellow eye in measurements conducted with the Pelli-Robson test or with a low-contrast test (25%) at various distances. Jeon et al. [25] observed worse contrast sensitivity in eyes with ERMs, speculating that the impaired contrast sensitivity in their ERM group might be attributed to the membrane itself and the disorganization of underlying retinal cells, disrupting the penetration of light. In our study, the agreement observed between eyes with and without ERMs, despite a more advanced stage of membrane development, could be attributed to the good postoperative outcomes observed at the retinal level, whereby 94% of cases showed improvement in the ectopic inner foveal layer.
We also compared visual disturbances between eyes with and without ERMs. Hence, although the eyes without a membrane showed slightly better results, differences between the two groups were not significant or clinically relevant. The literature lacks data on this variable with which to compare our results.
The limitations of our study include the size of the analyzed sample. Given that this was a pilot study, the sample size was not determined. Further research is needed to calculate the sample size needed to detect specific differences between the study groups based on these results. Another limitation is that there was no control group of individuals with both eyes healthy, aside from cataracts, allowing for the direct comparison of data, particularly with regard to the binocular data collected.
Our results indicate that eyes affected by stage 2 or 3 fovea-involving ERMs could benefit from combined phacovitrectomy with the implantation of non-diffractive EDoF IOL with X-Wave technology. No differences were observed in visual outcomes, contrast sensitivity, or the presence of visual disturbances, except for near visual acuity when compared to eyes without ERM. Based on our findings, the AcrySof IQ Vivity® lens appears to be a viable option for patients with unilateral ERM seeking spectacle independence, irrespective of foveal involvement. The limitations of our pilot study underscore the need for larger-scale research to validate these findings and optimize clinical applications.
Acknowledgments
This article is a revised and expanded version of a paper entitled “A prospective study of the bilateral implantation of an extended depth of focus intraocular lens in patients undergoing phacoemulsifi-cation in one eye and phacovitrectomy for epiretinal membrane in the fellow eye”, which was presented at the ESCRS Meeting (Barcelona, 6–10 September 2024) [29].The current manuscript includes information not presented in the Meeting, like evolution of visual acuity at different timepoints after surgery.
Supplementary Materials
The following supporting information can be downloaded at: https://www.mdpi.com/article/10.3390/biomedicines12112443/s1, Table S1: Monocular visual disturbance measurements obtained in psychophysical tests at 6 months post-surgery in eyes with and without ERM.
Author Contributions
Conceptualization, M.G and L.S.; methodology, L.S. and M.G.; software, M.B.; validation, S.R., M.V., and E.L.; formal analysis, M.G.; investigation, L.S., M.G., S.R., M.V., and E.L.; resources, M.G.; data curation, S.R., M.V., and E.L.; writing—original draft preparation, M.B.; writing—review and editing, L.S., M.G., S.R., M.V., and E.L.; visualization, L.S., M.G., S.R., M.V., and E.L.; supervision, M.B.; project administration, L.S., S.R., and M.B.; funding acquisition, M.G. and L.S. All authors have read and agreed to the published version of the manuscript.
Institutional Review Board Statement
This study was conducted in accordance with the Declaration of Helsinki and approved by the Ethics Committee of Quirónsalud-Catalunya (Ref: 2021/92-OFT-HUGC).
Informed Consent Statement
Informed consent was obtained from all subjects involved in the study.
Data Availability Statement
Data may be available upon reasonable request and after approval by the sponsor.
Conflicts of Interest
Author L. Sararols received a grant from Alcon Healthcare to support this study. The remaining authors declare that the research was conducted in the absence of any commercial or financial relationship that could be construed as a potential conflict of interest. The funder had no role in the design of the study; in the collection, analyses, or interpretation of data; in the writing of the manuscript; or in the decision to publish the results.
Funding Statement
This research was partially funded by an investigator-initiated study, grant number 70523225, from Alcon Healthcare.
Footnotes
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content.
References
- 1.Pearlstone A.D. The incidence of idiopathic preretinal macular gliosis. Ann. Ophthalmol. 1985;17:378–380. [PubMed] [Google Scholar]
- 2.Giachos I., Chalkiadaki E., Andreanos K., Symeonidis C., Charonis A., Georgalas I., Rotsos T. Epiretinal membrane-induced intraretinal neovascularization. Am. J. Ophthalmol. Case Rep. 2021;23:101180. doi: 10.1016/j.ajoc.2021.101180. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.Schmitz-Valckenberg S., Holz F.G., Bird A.C., Spaide R.F. Fundus autofluorescence imaging: Review and perspectives. Retina. 2008;28:385–409. doi: 10.1097/IAE.0b013e318164a907. [DOI] [PubMed] [Google Scholar]
- 4.Kadonosono K., Itoh N., Nomura E., Ohno S. Capillary blood flow velocity in patients with idiopathic epiretinal membranes. Retina. 1999;19:536–539. doi: 10.1097/00006982-199911000-00010. [DOI] [PubMed] [Google Scholar]
- 5.Fraser-Bell S., Guzowski M., Rochtchina E., Wang J.J., Mitchell P. Five-year cumulative incidence and progression of epiretinal membranes: The Blue Mountains Eye Study. Ophthalmology. 2003;110:34–40. doi: 10.1016/S0161-6420(02)01443-4. [DOI] [PubMed] [Google Scholar]
- 6.Gil M.A., Varón C., Cardona G., Vega F., Buil J.A. Comparison of far and near contrast sensitivity in patients symmetrically implanted with multifocal and monofocal IOLs. Eur. J. Ophthalmol. 2014;24:44–52. doi: 10.5301/ejo.5000335. [DOI] [PubMed] [Google Scholar]
- 7.Shekhar S., Sureka S.P., Sahu S.K. Multifocal intraocular lens explantation: A case series of 50 eyes. Am. J. Ophthalmol. 2015;159:202. doi: 10.1016/j.ajo.2014.09.032. [DOI] [PubMed] [Google Scholar]
- 8.Kamiya K., Hayashi K., Shimizu K., Negishi K., Sato M., Bissen-Miyajima H. Multifocal intraocular lens explantation: A case series of 50 eyes. Am. J. Ophthalmol. 2014;158:215–220.e1. doi: 10.1016/j.ajo.2014.04.010. [DOI] [PubMed] [Google Scholar]
- 9.Rosen E., Alió J.L., Dick H.B., Dell S., Slade S. Efficacy and safety of multifocal intraocular lenses following cataract and refractive lens exchange: Metaanalysis of peer-reviewed publications. J. Cataract. Refract. Surg. 2016;42:310–328. doi: 10.1016/j.jcrs.2016.01.014. [DOI] [PubMed] [Google Scholar]
- 10.Grzybowski A., Kanclerz P., Muzyka-Woźniak M. Methods for evaluating quality of life and vision in patients undergoing lens refractive surgery. Graefes Arch. Clin. Exp. Ophthalmol. 2019;257:1091–1099. doi: 10.1007/s00417-019-04270-w. [DOI] [PubMed] [Google Scholar]
- 11.Wang S.Y., Stem M.S., Oren G., Shtein R., Lichter P.R. Patient-Centered and visual quality outcomes of premium cataract surgery: A Systematic Review. Eur. J. Ophthalmol. 2017;27:387–401. doi: 10.5301/ejo.5000978. [DOI] [PubMed] [Google Scholar]
- 12.Liu J., Dong Y., Wang Y. Efficacy and safety of extended depth of focus intraocular lenses in cataract surgery: A systematic review and meta-analysis. BMC Ophthalmol. 2019;19:198. doi: 10.1186/s12886-019-1204-0. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13.Cochener B., Boutillier G., Lamard M., Auberger-Zagnoli C. A Comparative evaluation of a new generation of diffractive trifocal and extended depth of focus intraocular lenses. J. Refract. Surg. 2018;34:507–514. doi: 10.3928/1081597X-20180530-02. [DOI] [PubMed] [Google Scholar]
- 14.Schallhorn J.M. Multifocal and extended depth of focus intraocular lenses: A comparison of data from the United States Food and Drug Administration Premarket Approval Trials. J. Refract. Surg. 2021;37:98–104. doi: 10.3928/1081597X-20201111-02. [DOI] [PubMed] [Google Scholar]
- 15.Singh B., Sharma S., Dadia S., Bharti N., Bharti S. Comparative Evaluation of visual outcomes after bilateral implantation of a diffractive trifocal intraocular lens and an extended depth of focus intraocular lens. Eye Contact Lens. 2020;46:314–318. doi: 10.1097/ICL.0000000000000637. [DOI] [PubMed] [Google Scholar]
- 16.Arrigo A., Gambaro G., Fasce F., Aragona E., Figini I., Bandello F. Extended depth-of-focus (EDOF) AcrySof® IQ Vivity® intraocular lens implant: A real-life experience. Graefes Arch. Clin. Exp. Ophthalmol. 2021;259:2717–2722. doi: 10.1007/s00417-021-05245-6. [DOI] [PubMed] [Google Scholar]
- 17.Govetto A., Lalane R.A., Sarraf D., 3rd, Figueroa M.S., Hubschman J.P. Insights into epiretinal membranes: Presence of ectopic inner foveal layers and a new optical coherence tomography staging scheme. Am. J. Ophthalmol. 2017;175:99–113. doi: 10.1016/j.ajo.2016.12.006. [DOI] [PubMed] [Google Scholar]
- 18.Hong X., Milanovic Z., Wei X. Ophthalmic Lens Having an Extended Depth of Focus. 9968440 B2. U.S. Patent. 2018 May 15;
- 19.US Food and Drug Administration Summary of Safety and Effectiveness Data (SSED). AcrySofTM IQ VivityTM Extended Vision Intraocular Lens. Published 2020. [(accessed on 16 November 2023)]; Available online: https://www.accessdata.fda.gov/cdrh_docs/pdf/P930014S126B.pdf.
- 20.Ahmad B.U., Shah G.K., Hardten D.R. Presbyopia-correcting intraocular lenses and corneal refractive procedures: A review for retinal surgeons. Retina. 2014;34:1046–1054. doi: 10.1097/IAE.0000000000000237. [DOI] [PubMed] [Google Scholar]
- 21.Klein R., Klein B.E. The prevalence of age-related eye diseases and visual impairment in aging: Current estimates. Investig. Ophthalmol. Vis. Sci. 2013;54:orsf5–orsf13. doi: 10.1167/iovs.13-12789. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 22.Kim H., Jeon S. Visual outcomes of epiretinal membrane removal after diffractive-type multifocal intraocular lens implantation. BMC Ophthalmol. 2022;22:423. doi: 10.1186/s12886-022-02649-8. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 23.Patel S.B., Snyder M.E., Riemann C.D., Foster R.E., Sisk R.A. Short-term outcomes of combined pars plana vitrectomy for epiretinal membrane and phacoemulsification surgery with multifocal intraocular lens implantation. Clin. Ophthalmol. 2019;13:723–730. doi: 10.2147/OPTH.S195928. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 24.Guarro M., Sararols L., Londoño G.J., Goñi I., Vázquez M.O., Ruiz S.O., López S.O. Visual disturbances produced after the implantation of 3 EDOF intraocular lenses vs 1 monofocal intraocular lens. J. Cataract. Refract. Surg. 2022;48:1354–1359. doi: 10.1097/j.jcrs.0000000000000988. [DOI] [PubMed] [Google Scholar]
- 25.Jeon S., Choi A., Kwon H. Clinical outcomes after implantation of extended depth-of-focus AcrySof® Vivity® intraocular lens in eyes with low-grade epiretinal membrane. Graefes Arch. Clin. Exp. Ophthalmol. 2022;260:3883–3888. doi: 10.1007/s00417-022-05751-1. [DOI] [PubMed] [Google Scholar]
- 26.Chung H.S., Nam S., Jang J.H., Lee K.E., Kim J.Y., Tchah H., Lee H. Short-term clinical outcomes after implantation of monofocal intraocular lens with enhanced intermediate function in eyes with epiretinal membrane. Sci. Rep. 2023;13:18018. doi: 10.1038/s41598-023-44839-4. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 27.Gundersen K.G., Potvin R. Clinical outcomes and quality of vision associated with bilateral implantation of a wavefront shaping presbyopia correcting intraocular lens. Clin. Ophthalmol. 2021;15:4723–4730. doi: 10.2147/OPTH.S342947. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 28.Bala C., Poyales F., Guarro M., Mesa R.R., Mearza A.M., Varma D.K.B., Jasti S., Lemp-Hull J.D. Multicountry clinical outcomes of a new nondiffractive presbyopia correcting IOL. J. Cataract. Refract. Surg. 2022;48:136–143. doi: 10.1097/j.jcrs.0000000000000712. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 29.Sararols L. A prospective study of the bilateral implantation of an extended depth of focus intraocular lens in patients undergoing phacoemulsification in one eye and phacovitrectomy for epiretinal membrane in the fellow eye; Presented at the 2024 ESCRS Meeting; Barcelona, Spain. 6–10 September 2024; Sep 6–10, [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
Data Availability Statement
Data may be available upon reasonable request and after approval by the sponsor.