Abstract
Purpose
Minimally Invasive Osteosynthesis (MIO) developed to be a key technique in orthopedic trauma surgery, offering a less invasive alternative for managing fractures in various anatomical regions. However, standardized guidelines for its application are lacking. This study aims to establish comprehensive principles for MIO to guide surgeons in treating different types of fractures and its locations.
Methods
A retrospective analysis including 57 fractures in 53 patients was conducted. All patients were treated with MIO. Study range - March 2017 to January 2022 at a Level-I trauma university hospital. The analysis covered various fracture types, focusing on surgical approaches, reduction techniques, plate insertion methods, and its outcomes. The efficacy and safety of MIO was evaluated by analyzing complications, fracture healing time, and necessary revision surgeries.
Results
MIO is a versatile and effective fracture treatment that minimized soft tissue damage and ensured stable osteosynthetic results. Articular fractures typically used a “one way up” plate insertion technique, while non-articular fractures applied an “inside-up-and-down” approach. Low complication rates confirmed the safety and efficiency of MIO.
Conclusion
This research established generalized principles for MIO, offering a systematic approach that can be applied for various fracture types and its locations, by overall enhancing the surgical efficiency as well as patient recovery, without compromising outcomes.
Level of evidence
This study is classified as Level III evidence according to “The Oxford 2011 Levels of Evidence”.
Keywords: Minimally invasive, Osteosynthesis, Fracture, Surgical guide
Introduction
MIO
The application of minimal invasive surgery has steadily increased over the past years [1–3]. In orthopedic trauma surgery, minimal invasive osteosynthesis (MIO) was described for most anatomic regions and its indication is constantly growing ( [4–12]). A MIO approach in fracture care is applicable in every region of the human body. Although MIO was only used sporadically before, the development of locking compression plates (LCP) has enlarged its spectrum of application [13–15]. While the insertion of drill sleeves added another operational step in plate fixation to provide mono axial locking, the sleeves perfectly permit percutaneous locking through stab incisions. Standard approaches usually depend on plate length, but the use of drill sleeves, especially in the distal femur, made this obsolete [16, 17].
In common standard methodologies, the length of the surgical incision permitted all necessary steps for osteosynthesis, ranging from fracture control and reduction, to temporary fixation and plate insertion. In MIO, these steps should be planned independently from each other. Fracture control (visual or digital), reduction using instruments, and plate insertion may all need an individual operative step. In certain scenarios, all steps can be combined in one minimal invasive approach. In the worst case, each step needs an individual single minimal invasive approach.
The advantages and limitations of MIO were discussed extensively in the literature and it can be summarized that it is an effective tool in fracture care [18, 19]. If fractures with severe soft tissue damage, like complex- or open fractures, would undergo a standard operative technique, the risk of soft tissue loss and/or infection would further increase. Therefore, MIO can help limit the damage by placing small incisions into safe zones and by protecting regions that were initially sutured up after an open fracture occurred [20]. The same concept applies for osteoporotic bone fractures [9, 21].
General guidelines for the application of MIO remain scarce in the existing literature. In order to establish these guidelines, we conducted an analysis of MIO-treated fractures across various anatomical regions. The objective of this study was to delineate universally applicable principles for MIO according to diverse fracture types across different anatomical regions. The aim of this research is to provide a comprehensive framework for MIO surgical procedures by having a minimal incision size for a standard, stable osteosynthesis.
Methods
Between 03/2017 and 01/2022, 57 fractures of 53 patients (mean age 59.94 years) treated with MIO were identified at a Level-I-trauma university hospital in Salzburg, Austria. A positive ethical approval was granted by the Ethics Committee of the State of Salzburg (EK No: 1138/2023). Informed consent has been obtained.
Fractures of different anatomic regions were analyzed to determine general rules and principles of MIO (Table 1).
Table 1.
N | |
---|---|
Patients | 53 |
Encountered overall fractures | 57 |
-Articular fractures | 33 |
-Non-articular fractures | 24 |
-Patients with 2 different fractures | 2 |
-Patients with lower extremity fractures in the tibia and fibula* | 2 |
-Fractures in geriatric patients (> 65 years old) | 22 |
-Open fractures | 5 |
-External fixation prior to surgery | 11 |
Left Side | 18 |
Right Side | 39 |
Male | 24 |
Female | 28 |
*Patients with fractures in the lower extremity were mentioned separately, as the fractured bones were evaluated separately
Inclusion criteria:
Fractures treated with MIO.
Articular and non-articular fractures.
Fractures treated with LCP.
Closed and open fractures.
Comminuted and simple fractures.
Periprosthetic fractures.
Geriatric fractures > 65 years.
Exclusion criteria:
Fractures treated with intramedullary nails or K-wires.
Fractures treated with open reduction and internal fixation (ORIF).
Spinal and pelvic fractures.
Patients younger than 18 years of age.
Pregnancy.
The main parameters for MIO evaluated were the surgical approach, the reduction used, and the insertion of the plates.
Secondary parameters included patient´s positioning, feasibility, bony healing, and post-operative complications.
Surgical techniques applied
Various surgical techniques used for articular and non-articular fractures were analyzed.
The surgical approach itself was defined as an incision, other than a stab incision, to lock plates or for the percutaneous application of instruments (reduction clamps, K-wires or percutaneous lag screws). Stab incisions for instruments were recorded separately.
Surgical approaches used included fracture control, control of the articular surface, plate insertion, or an additional incision for reduction. Therefore, these approaches were evaluated according their location and further divided into four categories:
Surgical approaches:
Above the articular surface and fracture.
Above the fracture.
For plate insertion.
Additional approaches (for reduction).
Reduction and temporary fixation
Closed reduction techniques were applied in all fractures. The reduction technique itself was a reduction using the plate as a template by temporary fixation with K-wires through drill sleeves.
Instruments used for reduction and temporary fixation included: Reduction clamps used percutaneously or through the surgical approach, K-wires used percutaneously or through the approach, mini-plates, and cerclage wires.
The reduction techniques used differ depending on the anatomical region and are described in detail below.
MIO reduction technique for articular fractures: The aim is the anatomical reduction of the articular surface, which is why an incision above the joint is necessary in order to be able to inspect and palpate it. After the incision, reduction is performed using reduction forceps and K-wires. Care must be taken to ensure that these do not interfere with the planned plate osteosynthesis. Once the articular surface has been reduced, the previously selected plate osteosynthesis is inserted via the MIO incision remote from the joint and the plate is aligned with the aid of drill sleeves fixed in the fixed-angle holes. The screw holes are then drilled as standard using the familiar ORIF principles but via stab incisions, only. After fixation of the plate osteosynthesis, the reduction forceps and the K-wires are removed.
MIO reduction technique for non-articular fractures: The aim is to restore the correct length, rotation and axes. An incision over the fracture is not compulsory, but can be helpful for reduction. Primarily, the plate bed is created with a raspatory and the selected plate osteosynthesis is inserted via the MIO incision. The plate is aligned with the aid of drill sleeves fixed in the fixed-angle holes. The screw holes are then drilled as standard using the familiar ORIF principles bute via stab incisions, only.
Plate insertion
Two methods for plate insertion were identified and hence evaluated: Either the plate was inserted “one way up” (or down), or “inside up and down”.
In the “one way up” technique, the plate was inserted through the approach and slid up (or down) along the bony shaft over the fracture.
In the “inside up and down” technique, the plate was inserted through the approach and slid one way up and then the other way down.
Articular fractures occurred mainly in the distal radius, proximal and distal humerus, olecranon, distal femur, and proximal and distal tibia. Non-articular fractures presented in the forearm, clavicle, femur, tibia, distal metaphyseal tibia, and malleolar fibula. The metaphyseal tibia and the malleolar fibula (AO classification code 43 and 44, respectively) were considered as non-articular, due to their intact articular surface. The operated anatomic regions are summarized in Table 2.
Table 2.
N | ||
---|---|---|
Articular Fractures | Proximal Humerus | 8 |
Distal Humerus | 6 | |
Olecranon | 1 | |
Distal Radius | 5 | |
Distal Femur | 4 | |
Proximal Tibia | 8 | |
Calcaneus | 1 | |
Non-articular Fractures | Clavicle | 7 |
Forearm | 7 | |
Femoral Shaft | 3 | |
Tibial Shaft | 2 | |
Distal Tibia | 3 | |
Malleolar (Fibula) | 2 | |
Total | 57 |
Eleven fractures were stabilized with an external fixation before MIO and five fractures were classified as open fractures.
Two fractures in geriatric patients (n = 22) were initially stabilized using external fixation and included a combined tibial head/shaft fracture and a femoral shaft fracture.
Surgical technique
Before the first incision, anatomic landmarks were marked on the skin using a sterile pen. If the landmarks could not be palpated safely, the radiological c-arm was used for planning. The goal was to perform a perfect fracture reduction to keep the incisions as small as possible. The orientation of the approaches was chosen according to standard approaches.
The plate bed was made using a raspatory tool before reduction. Plates were inserted and fixed using drill sleeves. The eccentrically inserted drill sleeves were used as a handle to slide the plate under the soft tissues. Two sleeves permitted the manipulation in all directions. The contact of the tip of the plate to the bone could be felt and the plate positioned accordingly. A drill sleeve was inserted in the last plate hole percutaneously and the plate fixed temporarily with K-wires. The position of the plate was confirmed fluoroscopically and adjustments were made if necessary.
A guiding arm for plate insertion was used for the femoral LCP and the calcaneus plate only.
After insertion and temporary fixation of the plate with K-wires through the drill sleeves, a plate dependent lag screw was used to pull the plate to the bone shaft as necessary. The final locking of the plates was made through the existing approach and percutaneously.
Postoperative protocol
All fractures were treated postoperatively according to AO guidelines. After the soft tissues recovered, early mobilization was encouraged. All patients with articular fractures received pre- and postoperatively CT scans in addition to prior x-ray diagnostics.
Results
The location of the applied surgical approaches according to the encountered fractures was summarized in Table 3. Some surgical approaches were combined.
Table 3.
Total | FAS* | Fracture | Plate | Additional | |
---|---|---|---|---|---|
Articular Fractures | 33 | 33 | 6 | 1 | 5 |
Non-Articular Fractures | 24 | 0 | 21 | 7 | 2 |
Total | 57 | 33 | 27 | 8 | 7 |
*FAS: Approach above the Fracture and the Articular Surface
Within the encountered articular fractures, all surgical approaches were above the articular surface and fracture itself. In non-articular fractures, most approaches were above the fracture. Out of 27 approaches above the fracture, 21 (77.8%) occurred in non-articular fractures.
Analyzing all fractures, including articular and shaft fractures, the surgical approach was placed above the fracture (33 above the articular surface and fracture + 21 above the fracture), except in three shaft fractures (closed reduction). This included 2 extraarticular tibial fractures (1 shaft, 1 metaphysis) and 1 femoral shaft fracture (periprosthetic).
A single surgical approach was used in 38 fractures (66.7%), two surgical approaches in 16 (28.1%) patients, and three approaches in 2 (3.5%) patients.
Plate locking was carried out percutaneously through stab incisions.
All approaches except 2 (3.5%) were incised according to previously established standardized surgical approaches, but were limited in their extend. This was best described for the standard palmar approach of the distal radius, which applies to the same location as the palmar Henry approach [14, 22]. The goal of this minimal invasive approach was to control the fractured palmar cortex, the key landmark for reduction. The 2 exceptions for using limited standard approaches included the distal humerus (n = 6; described by Hofmann et al.), and the calcaneus, where an Ollier approach was used (n = 1) [23].
Various techniques and instruments used for fracture reduction and temporary fixation was summarized in Table 4. Some of these techniques were combined.
Table 4.
N | |
---|---|
Closed Reduction | 57 |
Reduction Clamp | 33 |
K-wire | 30 |
Plate Reduction | 29 |
PC* K-wire | 15 |
PC Reduction Clamp | 9 |
Mini Plates | 2 |
Cerclage | 1 |
*PC = percutaneous
Reduction could be achieved in all fractures except for one, as confirmed by intra- and post-operative x-ray imaging, confirmed by CT scans for articular fractures.
The “one way up” technique of plate insertion was used in all articular fractures, except for the calcaneus (97.0%) (Figs. 1, 2 and 3). The “inside up and down” technique was used mostly for non-articular fractures (13 out of 15: 86.7%), except for the mentioned calcaneus and a medial locking plate for the tibial metaphysis in a combined tibial head and shaft fracture. The two plate insertion techniques according to fracture type were summarized in Table 5.
Table 5.
Total | One Way Up | Up and Down | |
---|---|---|---|
Articular Fractures | 33 | 32 | 2 |
Non-Articular Fractures | 24 | 11 | 13 |
Total | 57 | 43 | 15 |
For non-articular shaft fractures the “inside up and down” technique was not possible because the plate was too long. Therefore, the “one way up” alternative was used (7 of 8 plate approaches; 87.5%). This included a long bone shaft fracture where the plate length almost equaled the length of the bone itself. Hence, the plate was inserted from proximal/distal and pushed all the way to down/up (Fig. 4).
Overall, 79 plates were used for fracture stabilization. This included 35 times one plate, and 22 times two plates.
Plate positioning, its feasibility as well as bony healing with its complications were summarized in Table 6.
Table 6.
Positioning | ||
---|---|---|
Articular Fractures | Proximal Humerus | BC |
Distal Humerus | P | |
Olecranon | P | |
Distal Radius | S/A | |
Distal Femur | S | |
Proximal Tibia | S | |
Calcaneus | L | |
Non-articular Fractures | Clavicle | BC |
Forearm | S/A | |
Femoral Shaft | S/L | |
Tibial Shaft | S | |
Distal Tibia | S | |
Malleolar (Fibula) | S |
BC = Beach Chair, P = Prone, S = Supine, A = Arm Table, L = Lateral
Patient positioning on the operating table included: 32 times supine (10 arm tables), 7 times prone, 15 times within a beach chair, and 3 times lateral. Patients position did not differ in patients treated with ORIF alone.
If appropriate fracture control was given, the applied reduction was performed and assessed.
Four fractures healed with complications. In one case, a proximal humerus fracture treated with a locking compression plate (PHILOS) became infected, leading to the necessity of revision arthroplasty with shoulder prosthesis implantation. Another case involving a distal humeral fracture treated with two LCPs resulted in mispositioning of the ulnar LCP, leading to secondary fracture dislocation, attributed to the restricted approach of MIO. Additionally, another distal humeral fracture treated with two LCPs experienced a peri-implant fracture following a secondary trauma occurring 2 weeks post-surgery. Lastly, an 82-year-old geriatric patient expired subsequent to femoral plating, succumbing to septic urinary tract and pulmonary infections.
To illustrate the feasibility of MIO, several examples have been included in this publication.
Example malleolar fracture, fibula
Positioning: standard supine, bump under the buttock.
Approach: directly over the fracture to control the anteroinferior fracture gap to the posterior superior border of the fibula.
Reduction: Closed (traction and supination of the foot), percutaneous pointed clamp on the distal fibula, pointed clamp on the fracture through the approach.
Fixation: Percutaneous anteroposterior lag screw and a one-third-tubular LCP.
Plate insertion: Inside up and down through the approach, percutaneous locking.
Alternatively, a sole percutaneous reduction and a one way up distal plate insertion was described [24, 25]. The direct control of the fracture was preferred to achieve anatomic reduction.
Example tibial shaft fracture
Positioning: standard supine
Approach: Only for plate insertion distal over the medial malleolus.
Reduction: Closed (traction), percutaneous reduction clamp.
Fixation: Metaphyseal 3,5/4,5 mm LCP.
Plate insertion: One way up from the approach of the medial malleolus, percutaneous locking.
In case where reduction is insufficient or direct fracture control is required, an additional approach above the key landmark of the tibia can be applied.
Example olecranon fracture
Positioning: standard prone
Approach: directly over the fracture to control the fracture gap from the ulnar articular surface to the radial. The control of the articular surface is achieved through the fracture.
Reduction: Closed (extension of the elbow), percutaneous K-wires to manipulate the proximal (olecranon) fragment.
Fixation: Temporary: 2 percutaneous crossed K-wires from proximal to distal. Definitive: 3,5 mm olecranon LCP.
Plate insertion: One way up to the shaft from the approach and twisting the plate ´s bend under the triceps muscle, locking through the approach in a proximal- and percutaneously distal- fashion.
Example tibial head fracture
Positioning: standard supine, bump under the buttock
Approach: directly over the articular surface on the fracture site.
Reduction: Closed (varus stress), K-wires to manipulate the lateral fragment, percutaneous pointed reduction clamp (big or the ball-spiked tibial-head-reduction-clamp), raspatory tool or similar to elevate the articular surface.
Fixation:
Temporary: K-wires from lateral to medial, inside-out medially if necessary.
Definitive: 3,5 mm proximal tibial LCP.
Plate insertion: One way down to the shaft from the approach, locking through the approach in a proximal- and percutaneously distal- fashion.
Discussion
Multiple MIO techniques for different anatomic regions have been described in the literature, including extensive reviews discussing its advantages and risks [8, 10, 14, 16, 18, 22–28]. Minimal invasive surgery has become widely accepted in all types of disciplines, ranging from visceral to orthopedic surgery [1–3]. New instruments permit percutaneous surgery for the spine and pelvis [29–33]. Nevertheless, to our knowledge, an attempt to generalize an algorithm with guidelines for MIO approaches has not been established yet.
The following considerations for any surgical approach to carry out an MIO need to be taken into consideration. The applied approach should include: A successful fracture reduction with instruments, adequate plate positioning and locking, as well as minimal damage to anatomical structures and tissues.
The surgical approach itself for articular fractures is placed directly onto the fracture and its articular surface. This permits control over both parts and the reduction of the articular surface. Due to the design of anatomic shaped locking plates, the locking of the articular segment of the plate is possible through that approach. The locking of the shaft component of the plate can be achieved through stab incisions. This surgical method was mainly used for articular fractures in this study. In some articular fractures, an additional incision was made to gain better control of the fractured articular surfaces (Fig. 5). However, it can be summarized, that 2 distinct small approaches which are placed rather deliberately, provide better fracture control than a single long one. Therefore, the soft tissue does not need to be pulled and torn.
Most non-articular fractures required a direct approach onto the fracture and percutaneous locking of the plate proximal and distal to the fracture (Fig. 6). This permitted direct fracture control and facilitated reduction (21 out of 24 non-articular fractures).
If fracture reduction was attained through closed methods or percutaneously, the plate was positioned eccentrically, as depicted in Fig. 4. This approach to fracture reduction was employed in only 3 cases out of the 24 non-articular fractures.
The difference between these 2 methods is best described for clavicular fractures. While the most common surgical technique is a double incision without direct control of the fracture, by applying MIO approaches, a small incision is placed slightly lateral to the fracture for reduction and lateral locking, hence locking the plate medially-percutaneously [15, 34].
In our opinion, fracture control is essential for articular- and non-articular fractures. Contrary to that, most published techniques for MIO in non-articular fractures describe to direct fracture contract. Fluoroscopically controlled reduction is timely and technically inferior to open reduction and a small approach placed on a crucial landmark often suffices to optimize fracture reduction. In shaft fractures with a large fracture zone, the key landmark for reduction should be identified and the incision placed right above it.
Mainly standard approaches were used for the location of MIO surgical approaches. It permitted a fallback option for complex fractures in case reduction itself was not able to be sufficiently carried out. In such cases, an extension of the approach would permit for maximal fracture control.
Hence, in articular fractures the main incision was placed to control the articular surface and the fracture. In non-articular fractures, the incision was placed directly onto the fracture.
Closed reduction was used by applying traction and counter traction during surgery as well as adequate patient positioning on the operating table. Many techniques were combined, depending on the type of fracture to achieve an adequate reduction. For the same anatomical regions, similar techniques were used. It can be summarized that a sufficient MIO technique requires a precise planning based on imaging, the percutaneous application of reduction tools, K-wires and reduction clamps. In extensive approaches, reduction tools were usually applied directly to the bone through the main approach. This MIO reduction technique my require some training and a learning curve, especially to avoid soft tissue damage by crushing reduction clamps due to mispositioned percutaneous incisions [35]. Further, in regions with delicate anatomic structures, a thorough knowledge of their anatomic position is required to avoid accidental damage. In distal humeral fractures where the ulnar nerve is exposed in open techniques, special attention must be taken for its protection [23].
In 29 cases, a plate was used as a reduction aid itself (Fig. 7). This technique seems to be specific for MIO. Compared to classical surgical sequences, in MIO the reduction and fixation can be executed simultaneously by using the plate.
In proximal humerus fractures for instance, this technique proved to be efficient, especially when using multiple drill sleeves. Tightening a plate dependent lag screw through the plate that is temporarily fixed to the bone using K-wires, reduces the greater tuberosity to the shaft and restores valgus malalignment or medial dislocation of the shaft. In tibial head fractures, the compression of the head can be applied safely using the anterolateral plate as a pressure converter, similar to washers or plates on the medial side of the tibial head. The lateral wall of the tibial head will then be realigned. An anatomic plate tightened onto the bone will reduce partially some dimensions of fracture dislocation. Applied from one side in long bones, it will reduce side-to-side and varus/valgus malalignment. Further, when the plate is centered to the bone, the ante/recurvation is corrected.
However, it can be assumed that closed reduction techniques are practically more challenging than open ones. This difference will become even more relevant for complex and comminuted fractures in anatomic regions where an anatomic reduction is necessary.
The identification of key landmarks seems important for estimating the possible reduction. The control of these landmarks permits a successful reduction. The palmar cortex for distal radius fractures or the anterior and posterior fracture borders of the fibula in malleolar fractures are such landmarks. Minimal invasive approaches for these landmarks suffice for fracture reduction and a plate can be slid in. This includes distal radius fractures, where 1,5 cm approaches were described [22]. In such cases, the locking and positioning of the plates might be more difficult than the reduction itself.
The condition of soft tissues is crucial to decide upon fracture treatment with MIO. In some severe open fractures with a foreseen primary delay of wound closure, an alternative and less invasive approach might reduce secondary damage to soft tissues and decrease the risk for infection (Fig. 8).
In osteoporotic fractures (Fig. 9) with closed but damaged soft tissues, MIO might be a safer alternative, even when an anatomic reduction is not possible. In these cases, complex reconstruction methods for bone and soft tissues should be avoided, as their failure risk is high due to the patients age, anticoagulation, and comorbidities, such as diabetes, vascular diseases, and polyneuropathy. In some obese patients, a MIO might not be achievable at all.
It can be summarized that combined reduction techniques, percutaneous and open lead to satisfactory fracture reduction results. The plate itself plays a crucial role in the reduction technique itself [28].
The insertion of the plate is especially interesting because it used to be the determinant factor for the length of the approach. Percutaneous screw insertion and locking of the plate has become practicable through the use of drill sleeves. It permits drilling, measuring using the drill bit, and the insertion of the locking screws through stab incisions. The LISS guiding arm was one of the first implants, apart from nails, to promote this locking through stab incisions. Nevertheless, locking through stab incision is feasible with drill sleeves only without using a guiding arm. Hence, it can be concluded that the length of the approach should only be planned for reduction. In articular fractures, this can usually be implemented and plate approaches for plate insertion are mainly needed in non-articular fractures. Only 8 out of 57 fractures required a plate approach, and 7 were non-articular fractures (n = 24). Out of 33 articular fractures, only 1 needed a plate approach. The goal for plate insertion should be to use the existing approach for fracture reduction.
Hence, plate locking doesn’t need direct incisions other than various stab incisions.
Limitations
Notable technical differences to non-MIOs were the percutaneous application of clamps (n = 9) and K-wires (n = 15). The application of pointed reduction clamps needed additional stab incisions for the clamp tine, and K-wire insertion required planning to avoid damage to structures at risk. Although not evaluated, adequate plate positioning is more cumbersome in MIO than in ORIF and usually needs to be verified fluoroscopically.
Other disadvantages, such as increased surgical duration and radiation exposure were widely discussed in the literature, and some benefits outweigh them. A further limitation of this study is its retrospective setup.
Conclusion
It can be concluded that the approach should be above the articular surface and the fracture itself in articular fractures. The approach should be above the fracture in non-articular fractures. In articular fractures, the plate is inserted one way up (down), whereas in non-articular fractures the plate is inserted inside-up-and-down. Plate locking needs stab incisions only for fracture reduction, where the usual techniques need to be adapted to the limited space. The plate serves as a template for reduction. Detailed knowledge of the anatomical structures within the surgical field is essential for the safe execution of the MIO technique. Certain regions of the body are more amenable to MIO than others due to varying anatomical complexities. For instance, in the case of a distal humerus fracture, the anatomy around the elbow poses significant challenges. The proximity of critical nerves and blood vessels to the operative field necessitates a high level of precision and expertise. Therefore, MIO in this area is recommended only for highly experienced surgeons who are well-versed in the local anatomy.
Surgeons must be familiar with the “safe corridors” for screw placement and the temporary fracture reduction techniques using reduction forceps and K-wires. This ensures that the fixation devices are placed without jeopardizing surrounding neurovascular structures. Mastery of these anatomical nuances and surgical techniques is crucial to minimizing complications and achieving optimal outcomes with MIO. The standard positioning of patients on the operating table is adequate and standard approaches for incisions can be extended if necessary.
Author contributions
F. W. Data collection, manuscript writing, and creation of figures, G. H., V. H. Data collection and manuscript editing, A. D. Data collection, manuscript writing, A.T., C.D. Data collection, manuscript writing, and creation of figures.
Funding
Open access funding provided by Paracelsus Medical University.
Data availability
Data is provided within the manuscript. Further information is available on request from the authors.
Declarations
Competing interests
The authors declare no competing interests.
Footnotes
Publisher’s Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Contributor Information
Florian Wichlas, Email: f.wichlas@salk.at.
Christian Deininger, Email: johannes.deininger@stud.pmu.ac.at.
References
- 1.Gomez D, et al. Minimal invasive surgery for multiple adhesive small bowel obstruction: results of a comparative multicenter study. Cir Cir. 2021;89(6):710–7. [DOI] [PubMed] [Google Scholar]
- 2.Ronsini C, et al. Laparo-assisted vaginal radical hysterectomy as a safe option for minimal invasive surgery in early stage cervical cancer: a systematic review and meta-analysis. Gynecol Oncol. 2022;166(1):188–95. [DOI] [PubMed] [Google Scholar]
- 3.Deininger S, et al. Minimal-invasive management of urological complications after kidney transplantation. Int Urol Nephrol. 2021;53(7):1267–77. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.Green A, et al. Identifying safe corridors for anterior pelvic percutaneous instrumentation using computed tomography-based anatomical relationships. Injury. 2022;53(10):3390–3. [DOI] [PubMed] [Google Scholar]
- 5.Beeres FJ, et al. ORIF versus MIPO for humeral shaft fractures: a meta-analysis and systematic review of randomized clinical trials and observational studies. Injury. 2021;52(4):653–63. [DOI] [PubMed] [Google Scholar]
- 6.Coifman O, et al. Lateral malleolus closed reduction and internal fixation with intramedullary fibular rod using minimal invasive approach for the treatment of ankle fractures. Foot Ankle Surg. 2019;25(1):79–83. [DOI] [PubMed] [Google Scholar]
- 7.Melcher C, et al. A new minimal-invasive approach for total hip replacement in sheep. Technol Health Care; 2023. [DOI] [PubMed]
- 8.Cho CH. Is minimal invasive plate osteosynthesis a viable treatment option for 4-part proximal humerus fractures? Clin Shoulder Elb. 2018;21(4):177–8. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.Ziegler P, et al. Ankle fractures of the geriatric patient: a narrative review. EFORT Open Rev. 2023;8(1):1–10. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Wilkinson BG, Marsh JL. Minimally invasive treatment of displaced intra-articular calcaneal fractures. Orthop Clin North Am. 2020;51(3):325–38. [DOI] [PubMed] [Google Scholar]
- 11.Pranata R, et al. Minimal invasive surgery Instrumented Fusion versus Conventional Open Surgical Instrumented Fusion for the treatment of spinal metastases: a systematic review and Meta-analysis. World Neurosurg. 2021;148:e264–74. [DOI] [PubMed] [Google Scholar]
- 12.Loizides A, et al. Ultrasound-guided minimal invasive carpal tunnel release: an optimized algorithm. Cardiovasc Intervent Radiol. 2021;44(6):976–81. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13.Wenger R, et al. Absolute or relative stability in minimal invasive plate osteosynthesis of simple distal meta or diaphyseal tibia fractures? Injury. 2017;48(6):1217–23. [DOI] [PubMed] [Google Scholar]
- 14.Neubauer T, et al. Minimal invasive plating of distal radius fractures. A safe procedure? Ann Anat. 2019;224:172–8. [DOI] [PubMed] [Google Scholar]
- 15.Beirer M, et al. Does a minimal invasive approach reduce anterior chest wall numbness and postoperative pain in plate fixation of clavicle fractures? BMC Musculoskelet Disord. 2015;16:128. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16.Nester M, Borrelli J Jr. Distal femur fractures management and evolution in the last century. Int Orthop, 2023. [DOI] [PubMed]
- 17.Frigg R, et al. The development of the distal femur less invasive stabilization system (LISS). Injury. 2001;32(Suppl 3):SC24–31. [DOI] [PubMed] [Google Scholar]
- 18.Mohiuddin K, Swanson SJ. Maximizing the benefit of minimally invasive surgery. J Surg Oncol. 2013;108(5):315–9. [DOI] [PubMed] [Google Scholar]
- 19.Kehlet H, Wilmore DW. Evidence-based surgical care and the evolution of fast-track surgery. Ann Surg. 2008;248(2):189–98. [DOI] [PubMed] [Google Scholar]
- 20.Iacobellis C, et al. Minimally invasive plate osteosynthesis in type B fibular fractures versus open surgery. Musculoskelet Surg. 2013;97(3):229–35. [DOI] [PubMed] [Google Scholar]
- 21.Lampropoulou-Adamidou K, et al. Currents of plate osteosynthesis in osteoporotic bone. Eur J Orthop Surg Traumatol. 2014;24(4):427–33. [DOI] [PubMed] [Google Scholar]
- 22.Lebailly F, et al. Distal radius fixation through a mini-invasive approach of 15 mm. PART 1: a series of 144 cases. Eur J Orthop Surg Traumatol. 2014;24(6):877–90. [DOI] [PubMed] [Google Scholar]
- 23.Hofmann V, et al. Minimally invasive plate osteosynthesis (MIPO) for distal humeral fractures: a cadaveric study and first clinical application. BMC Musculoskelet Disord. 2023;24(1):206. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 24.Park YU, Kim SJ, Kim HN. Minimally invasive plate osteosynthesis using the oblong hole of a locking plate for comminuted distal fibular fractures. J Orthop Surg Res. 2021;16(1):281. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 25.Yanez Arauz JM. Minimally invasive treatment of AO B ankle fractures: Surgical technique and long-term outcomes. Foot Ankle Surg. 2021;27(7):742–9. [DOI] [PubMed] [Google Scholar]
- 26.van de Wall BJM, et al. Minimally invasive plate osteosynthesis: an update of practise. Injury. 2021;52(1):37–42. [DOI] [PubMed] [Google Scholar]
- 27.Dresing K. [Minimally invasive osteosynthesis of pilon fractures]. Oper Orthop Traumatol. 2012;24(4–5):368–82. [DOI] [PubMed] [Google Scholar]
- 28.Muller TS, Sommer C. Minimally invasive plate osteosynthesis of the distal tibia]. Oper Orthop Traumatol. 2012;24(4–5):354–67. [DOI] [PubMed] [Google Scholar]
- 29.Marcia S et al. Feasibility, safety, and efficacy of a new percutaneous interspinous device: a retrospective multicenter study. Neuroradiology, 2024. [DOI] [PubMed]
- 30.Gerich T, et al. The role of the bilateral subcutaneous plate in the minimal invasive stabilization of fragility fractures of the pelvis. Trauma Case Rep. 2022;42:100740. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 31.Furtado M, et al. Percutaneous Pedicle Screw for Thoracolumbar fractures: a long-term follow-up. Rev Bras Ortop (Sao Paulo). 2024;59(1):e101–6. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 32.Faber F. [Osteoporotic fractures of the pelvic ring and their minimally invasive treatment options]. Orthopade. 2021;50(10):871–82. [DOI] [PubMed] [Google Scholar]
- 33.Bozorgmanesh M, Latifi R, Hassannejad N. Innovative technique for posterior fixation of vertically unstable pelvic Ring fracture: a Case Report. Arch Bone Jt Surg. 2023;11(10):658–61. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 34.Al-Sadek TA, Niklev D, Al-Sadek A. Midshaft Clavicular Fractures - osteosynthesis with minimally invasive technique. Open Access Maced J Med Sci. 2016;4(4):647–9. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 35.Bedi H, Hickey B. Learning curve for minimally invasive surgery and how to minimize it. Foot Ankle Clin. 2020;25(3):361–71. [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Data Availability Statement
Data is provided within the manuscript. Further information is available on request from the authors.